IKA C 1 Operating Instructions Manual

IKA® Calorimeter C 1
20000004724
Operating instructions EN
C 1_042016
EN
We declare under our sole responsibility that this product is in compliance with the regulations
2014/68/EC
(article 3, (3)),
2011/65/EU,
2014/30/EU
and 2014/35/EU
and conforms to the standards or standardized documents: EN 61010-1, EN 61326-1, EN 60529 and EN ISO
12100.
Declaration of Conformity
Explication of warning symbols
User notes
The warranty does not cover worn out parts, nor does it apply to faults resulting from improper use, insucient care or maintenance not carried out in accordance with the instructions in this operating manual.
Please read these operating instructions carefully.
IKA® only
consider themselves to be responsible for the safety, reliability and performance of the appliance when
• the appliance has been operated in accordance with the operating instructions,
• only persons authorised by the manufacturer interfere with the appliance,
• only original parts and original accessories are used for repairs.
The calorimeter system may be opened only by an authorised Service Agent or Customer Service Centre.
If service is required, we recommend that you contact our customer service department. Furthermore, we refer you to the applicable safety conditions and accident requirements.
IKA® is not responsible for damages or costs resulting from
accident, improper use of the device or impermissible modifications, repairs or renovations.
Read the operation instructions completely before starting up and follow the safety instructions.
Keep the operation instructions in a place where they can be accessed by everyone. Ensure that only trained staff work with the appliance.
Follow the safety instructions, guidelines, occupational health and safety and accident prevention regulations.
In this chapter you can find out how to work through these Operating Instructions most effectively to be able to work reliably with the calorimeter system.
The instructions in the section “Safety instructions“ must be complied with.
The chapters are designed for you to work through them in order.
The section “Transport, storage“ is relevant for system reliability and for ensuring high measuring accuracy.
The calorimeter system is ready to perform measurements once you have completed the procedures in the section “Setting up and commissioning“: - “Preparing a measurement“ and “Starting a measurement“.
Numbers 1, 2, 3 etc. indicate operating instructions in the following sections. These must always be carried out in the specified sequence.
Warranty and liability
In accordance with
IKA® warranty conditions, the warranty period
is 12 months. For claims under the warranty please contact your local dealer. You may also send the machine direct to our factory, enclosing the delivery invoice and giving reasons for the claim. You will be liable for freight costs.
DANGER
CAUTION
WARNING
NOTICE
Indicates an imminently hazardous situation, which, if not avoided, will result in death, serious injury.
Indicates a potentially hazardous situation, which, if not avoided, can result in death, serious injury.
Indicates a potentially hazardous situation, which, if not avoided, can result in injury.
Indicates practices which, if not avoided, can result in equipment damage.
Reference to exposure to a hot surface!
DANGER
Page
Declaration of Conformity 02 Explication of warning symbols 02 User notes 02 Contents 03 Safety instructions 04 Correct use 05
System properties 05
Transportation, Storage 06
Transport and Storage conditions 06 Unpacking 06
Scope of delivery 06 Description of the system components 06
Hoses 07
Commissioning 07
Place of installation 07 Cooler connection 08 Venting hose connection 08 Drain hose connection 08 Oxygen supply connection 08
Power supply unit connection 09
Connection of peripherals 09 Handling the unit 09
Control unit with display 10
Dialog box 11
System start 12
Switching on the system (in the open state) 12 Selection of language 12 Pressure test 13 Oxygen pressure 13 System check 13 Heating detected 14 Factory settings 14
Calorimetric measurements 14
Determination of gross calorific value 14 Corrections 15 Notes on the sample 15 Complete combustion 16 Calibration 16 Notes on calibration 16
Overview Main menu 17
Measurements 17 Default settings 17 Settings 17 Maintenance 18 Informations 18
Preparing and performing measurements 18
Performing the measurement 18 Placing the sample 19 Preparing a measurement 20 Close the calorimeter 20 Prerequisite checking for a measurement 21 Measurement procedure 21 Cleaning the inner vessel 23 Follow-up measurement 23 Display the storage overview 23 Use of a disposable crucible 23 Using a scale 24 Permissible weight entries 24 Cancel a measurement 25
Page
Measurements 25
New measurements 25 Last result 25 System test 26 Measurement archive 26 Measurement archive: Edit 27 Measurement archive: Print 27 Measurement archive: Delete 27 Measurement archive: Delete all 28 Calibration archive 28 Calibration archive: Edit 28 Calibration archive: C-Value selectin 29 Calibration archive: Print 29 Calibration archive: Delete 29
Default settings 30 Settings 30
Date / Time 30 Units 30 Display / Keypad 31 Inverted menu view 31 List view 31 Language 31 Accessories 32 Chiller test 32
Maintenance 32
Maintenance menu 32 Change seal 33 Stirrer test 33 O2 System blow 33 Servo up 33 Servo down 33
Informations 33 User information: Memory 34
Prior to input a new measurement 34 After input a new measurement 34
Exchanging off wear parts 34
O2 Valve seal change 34 Change ignition wire 35 Change ignition electrode 35 Change the main seal 35 Change ground electrode 36
Leak test with the system test 36 Fault and troubleshoot 36
E01 STIRRER 36 E02 COMMUNICATION 37 E03 DRIFT ALARM 37 E04 MINTEMPDIFF 37 E05 FILLWATER 37 E06 EMPTYWATER 38 E07 POSTWATERFILLING 38 E10 FILLWATER SENSOR 38 E11 SYSTEM OPEN 38 E12 PRESSURE 38 E13 HEATER 39 E16 MAXTEMPDIFF 39
Accessories and Consumables 39 Maintenance 40 Warranty 40 Technical data 41
Contents
Read the operating instructions in full before start­ing up and follow the safety instructions.
Keep the operating instructions in a place where they can be accessed by everyone. Ensure that only trained staff work with the appliance.
The calorimeter C 1 may only be used for the determination of calorific value of solid and liquid materials according to national and international standards (eg DIN 51900, BS 1016 T5, ISO 1928, ASTM 5468, ASTM 5865 and ASTM 4809).
The pressure vessel is integrated into the calorimeter C 1. The maximum extra energy added to the pressure vessel must not exceed 40,000 J (select the test mass accordingly). The permitted operating pressure of 142 bar (14.2 MPa) must not be exceeded. The maximum permitted operating temperature must not exceed 50 °C.
Only fill calorimeter system C 1 with oxygen to a maximum pressure of 40 bar (4 MPa). Check the pressure setting on the pressure reducer for your oxygen supply. Perform a leakage test before each combustion process (see section “Leaktightness testing with system test“).
Some materials tend to explode when combusted (e.g. due to formation of peroxides), which could cause the decomposition vessel to crack. The
IKA® C 1 calorimeter may not be used
for testing explosive samples.
If the burning behaviour of a material is unknown, it must be tested before combustion in the inner vessel (risk of explosion). If you are burning unknown samples, leave the room or keep your distance from the calorimeter.
Benzoic acid may only be combusted in its pressed form! Flammable dust and powder must be first pressed. Oven-dry dust and powder such as splints, hay, straw etc. explode when combusted! Always wet these materials first! Highly flammable liquids with a low vapour pressure (e.g. tetramethyl dihydrogen disiloxane) must not directly touch the cotton thread!
Please pay attention to the combustion of metal-containing substances, that the permissible total energy input is not exceeded! Furthermore, toxic combustion residue in the form of gases, ashes or condensation, for example, is possible in the inner wall of the inner vessel.
Observe the accident prevention requirements applicable to the activity and the work station.
Wear your personal protective equipment.
DANGER
When handling combustion sam­ples, combustion residue and auxiliary materials, please observe the relevant safety regulations.
The following materials, for example, could pose a risk:
- corrosive
- highly flammable
- explosive
- bacteriologically contaminated
- toxic.
Please observe the relevant regulations when handling oxygen.
Warning: oxygen as a compressed gas is oxidising; intensively aids combustion; can react violently with flammable materials.
Do not use oil or grease!
Tubes and screwed joints for oxygen must be kept free of grease.
DANGER
Combustion gases are hazardous to health, therefore the venting hose must be connected to a suita­ble gas cleaning system or extrac­tion system.
DANGER
Caution! In case of failure Error 04 ”Min.Temp.Diff.“ the combustion chamber could be hot after an ig­nition/combustion, even though the message ”No temperature in­crease!“ is displayed.
At the end of the work period, close the main valve for the oxygen supply.
Perform servicing work only when the equipment is depressurised.
When using stainless steel crucibles thoroughly check their condi­tion after each experiment. If the material gets thinner, the crucible may catch fire and dama­ge the pressure vessel. Crucibles must not be used for more than 25 combustions for safety reasons.
The EC Declaration of Conformity confirms that this pressure ves­sel was manufactured by the Pressure Equipment Directive 2014/68/EC. (Article 3, paragraph 3).
The pressure vessel has undergone a pressure test with test pressure of 203 bar (20.3 MPa) and a leakage test with oxygen at 40 bar.
Decomposition vessels are experiment autoclaves and must be tested by a technical expert after each use. (see capture: “Maintenance“).
Individual use is understood here to include a series of experiments performed under roughly the same conditions in terms of pressure and temperature. Experiment autoclaves must be operated in special chambers.
The pressure vessels must undergo repeated tests (internal tests and pressure tests) performed by the technical expert. The frequency of these tests is to be determined by the operator on the basis of experi
-
ence, type of operation and the material used in the pressure vessel.
The declaration of conformity becomes invalid if mechanical modifications are carried out to the experiment autoclaves or if tightness can no longer be guaranteed as a result of major corro­sion (e.g. pitting by halogens).
In particular the Thread on the lower part of the pressure vessel and the nut are subject to considerable stress and must therefore be checked regularly for wear. (see figure “Description of the sys­tem components“).
The condition and function of the seals must be checked and en­sured by way of a system test (see capture “system check“).
Check the seals for damage before each use (see chapter “Maintenance“.
DANGER
If the maintenance, and especially the pressure testing, is not per­formed or is performed incorrectly, there is a risk the decomposition ves­sel may burst or an uncontrolled in­ternal fire may occur at the elec­trodes which could burn away the seals (oxyacetylene torch effect), thus posing a risk to life and limb!
Only technical experts may perform pressure tests and service work on the decomposition vessel.
Safety instructions
We recommend that you send the C 1 calorimeter system to our factory for inspection, and if necessary, repair after 1000 tests or after one year or sooner depending on use.
For the purposes of these operating instructions a technical expert is someone
1. who guarantees to conduct the tests properly on the basis of his training, knowledge and experience gained through practical work,
2. who is suciently reliable,
3. who is not subject to any instructions in terms of the test activity,
4. who has suitable test equipment if necessary,
5. who can provide appropriate proof of the requirements listed in 1.
National directives and laws must be observed for operating pressure vessels!
Anyone operating a pressure vessel must keep it in a proper condition, operate it properly, supervise it, carry out the necessary maintenance and repair work immediately and implement the safety measures required in the circumstances.
A pressure container must not be operated if it exhibits defects that could endanger those working with it or third parties. You can obtain a copy of the pressure vessel regulation from Beuth Press.
The IKA® C 1 calorimeter system is used for calorific value deter­mination of solid and liquid substances.
This is done by placing a known quantity of a substance in a de­composition vessel which is surrounded by a water jacket. The sample is then combusted in an oxygen atmosphere . The calorific value of the sample can then be calculated from the resulting increase in temperature, the sample mass and the known thermal capacity of the overall system.
The C 1 calorimeter system is subject to the Pressure Vessel Direc­tive
2014/68/EC. Comply with the safety instructions accordingly.
For adaptation to individual laboratory tasks, use original
IKA®
consumables and accessories.
System properties
The
IKA® C 1 calorimeter is a combustion calorimeter for deter-
mining gross calorific values of liquid and solid non-explosive sub­stances. The samples were burned at an excess of oxygen under pressure in a closed vessel. The resulting amount of heat, measured in a previously calibrated system, allows the determination of the calorific value of the sample after preliminary various globally ap
-
plied standards and regulations.
The necessary corrections after combustion in the calorimeter can then, for example, about our CalWin
®
C 6040 calorimeter software (accessories), from the acid corrections up to the calo­rific value according to DIN, ISO, ASTM, GB, and GOST standards typed, corrected and calculated. By a transfer of the results to Excel can quickly and easily adapt special or own calculations.
The worldwide unique patented
IKA® calorimeter C 1 operates on
the globally applied isoperibolic measurement procedure at 22 °C and 30 °C. The temperature values are output every 12 seconds, and the calculation of the temperature correction according to standards Regnault Pfaundler.
Measurement times of the system:
- Sample preparation time: < 1 minute
-
from the start of measurement until the start of the next: 12 minutes
- Pre-experimental period: Response of the system 3 minutes
- Main experiment: after ignition of the sample: 4 minutes
To a supply of cooling water, the calorimeter can with a conven­tional thermostat / cooler operate for example such as the
IKA®
KV 600 (accessory).
Do not use distilled or demineralized water (increased risk of corrosion)!
Fill the cooler according to the cooler operating instructions. Cooling water temperature settings on the thermostat / cooler:
- in mode of opretaion isoperibol 22 °C: 18–21 °C
- in mode of opretaion isoperibol 30 °C: 26–29 °C
Alternatively, the device can also be operated via an optionally available heating C 1.20 (accessories) at a fixed water connection. The accessories for connection to the water pipe or the thermo­stat is contained in the corresponding delivery.
Requirements for the operation of the C 1 with the C 1.20 heating on a water pipe:
- tap water: is recommended in drinking tap water quality. Mix in (max. 1 ml for 4-5 l of water) of the supplied water additive. Thus, the durability of the water is improved.
- temperature range: 12 °C – 28 °C (according to the water temperature, the measuring mode 22 °C or 30 °C must be selected).
- water pressure: max. 1 – 1.5 bar (at higher or constant pressure, we recommend using the C 25 water pressure regulating valve available as an accessory)
- water consumption per measurement: about 4 liters
Automatic internal system checks enable the identification and monitoring
• the presence of oxygen pressure
• the controlled stirrer speed
• the the availability of water in the system
• the input water temperature
• of inner vessel
• the correct shutter of the measuring cell with locking
• Ignition counter with memory function for the next due maintenance
Correct use
Transport and Storage conditions
The system must be protected against mechanical impact, vibra­tions, dust deposits and corrosive ambient air during transporta­tion and storage. It is also important to ensure that the relative humidity does not exceed 80%. The appliance must be completely emptied before storing and transportation. In case of repair the device has to be cleaned and free from any materials which may constitute a health hazard.
Unpacking
Please unpack the system components carefully and check for any damage. When you unpack the equipment, check for any damages which may have occurred during transportation. In the case of any damage a fact report must be sent immediately (post, rail or forwarder).
If you require servicing, return the appliance in its original packag­ing. Storage packaging is not sucient. Please also use suitable transport packaging.
IKA® C 1 Calorimeter
• Power supply unit
• Inflow conduit
• Return conduit
• Discharge hose
• Round cable clamp
• Connection tube
• Venting hose
• Mains cable
• 1 Tool
• 5 O-ring FPM 11.0 x 2.0
• 2 O-ring FPM 6.0 x 2.0
• 2 O-ring FPM 15.0 x 2.0
• 5 O-ring FPM 10.0 x 2.5
• 5 O-ring FPM 8.0 x 2.5
• 5 O-ring FPM 4.0 x 1.5
Interfaces on rear of the unit
Controller unit with display, Multifunctional element and keybord
Inner vessel (IB)
Manual, pivotable lift
A
A-A
Base
Ignition wire
Ground electrode
Ignition electrode
Crucible holder
Sectional view inner vessel
Measuring cell as a pressure vessel
Rotary handle
Threaded nut
Threaded base
• 1 O-ring FPM 3.1 x 1.6
• 5 O-ring FPM 3.0 x 1.5
• 5 O-ring FPM 48.0 x 2.0
• 1 Quad-Ring NBR 92.0 x 4.5
• 1 X-Ring NBR 6,07 x 1,78
C 1.104 water bath additive
C 723 Benzoic acid
C 1.1012 Organizer
• 4 base
• O-ring-grease Molykote 55
• Handle
• Operationg instructions C 1
• Safety instructions C 1/C 6000
• Technical information C 1
• Insert sheet C 1 (4 steps to use)
• Warranty card
C 1.1012 Organizer
Transportation, Storage
Scope of delivery
Description of the system components
Hoses
Discharge hose
Inflow conduit
Return conduit
Venting hose
Connection tube
SW8
Cooler
C 1
C 1
Cooler
C 1
C 1
(O)
C 1
(degas)
(water in)
(water out)
(overflow)
filter 10m
Cooler
Exhaust air
Pressure reducer C 29
Adapter SW10
Place of installation
A constant ambient temperature is an important requirement for ensuring the high measuring accuracy of the system. The following conditions must therefore be fulfilled at the place of installation:
No direct solar radiation
No draughts (e.g. beside windows, doors, air conditioning)
Sucient distance to radiators and other heat sources
The minimum distance between the wall and the rear side of the device must be at least 25 cm.
The system must not have laboratory hardware such as shelves, wire ducts, ring lines, etc. installed above it.
The (constant) room temperature should be around 20 °C ... 25 °C.
The system must be set up on a horizontal surface.
For operation of the system the following must be available at the place of installation:
- A power supply corresponding to the nameplates on the system components,
- An oxygen supply (99.95 % pure oxygen, quality 3.5; pressure 30 bar) with pressure gauge.
A shut-off valve for the oxygen supply must be installed. Observe the instructions on handling oxygen given in the chapter “Safety instructions“.
Observe the instructions on handling oxygen given in the chapter “Safety instructions“. Please read these operating instructions carefully.
IKA® consider themselves responsible for the safety,
reliability and performance of the appliance only when
• the device has been operated in accordance with the operating instructions
• the specifications for the place of installation have been met.
max. 1.5 bar
Commissioning
Cooler connection
1. Connect the inflow pipe to the quick coupling to the “water in“ (on the calorimeter) port. Connect the other end of the radiator side of “water out“ and tighten the hose clamp.
Inflow pipe
Return pipe
2. Connect the return pipe to the quick coupling to the “water out“ (on the calorimeter) port. Connect the other end to “water in“ (on the cooler) and tighten the hose clamp.
Venting hose connection
1. Screw the venting hose to to “degas“ screw coupling (SW8; on the calorimeter) and lay the free end in the extractor. Connect the free end to a gas scrabber gadget.
Venting hose
Observe the applicable safety regulations..
The combustion gases are vented from the decompstion vessel after each combustion test on the vent hose. The vent hose must not be kinked or pinched during installation.
3. Do not use distilled or demineralized water (increased risk of corrosion)! Fill the cooler according to the cooler operating instructions. is recommended in drinking tap water quality. Mix in (max. 1 ml for 4-5 l of water) of the supplied water additive. Thus, the durability of the water is improved..
4. Adjust the cooler temperature corresponding to the calori­ meter operating temperature.
Drain hose connection
1. Insert the drain hose fully into the “tap-water“ connection (on the calorimeter). Lay the other end of the drain hose to the cooler filler neck.
Drain hose
2. Secure the hose end with an attachable adhesive protection.
The hose end must always have free access of air and never immerse themselves in the cool water!
Cooler
Oxygen supply connection
1. Insert the connection tube fully into the coupling (“O
2
in“; on the calorimeter) and connect the free end on the pressure reducer
IKA® C 29 (possibly
with the C 29 adapter, included with).
Connection tube allowable operating pressure 40 bar
IKA
®
C 29 Pressure reducer
Combustion gases are dangerous
to health, so the venting hose must be connected to a suitable gas cleaning and extraction.
DANGER
Power supply unit connection
1. Connect the power supply connecting cable (24 V)
NOTE: Note the correct inser­ tion position (flat side of the connector pointing to the right).
Connecting cable
The desktop power supply (scope of delivery) must not be placed flat on the laboratory table. It must be protected against moisture and water penetration and must not be wet.
2. Tape the supplied rubber feet to the power supply. Place the power supply on the rubber feet. Connect the connection cable to the DC power supply to your supply network (100-240V AC 50/60 Hz).
flat side of the connector
Connection of peripherals
Heating: Control output for the connection of the
IKA®
heating.
Scale: RS232 interface for connection of a scale (Mettler, Ohaus, Sortorius, Kern) Interface parameters: Baud rate: 1200 Data its: 7 Parity: odd Stop its: 1 Handshake: none
PC/Printer: RS232 interface for connecting a PC to control the C 1 (CalWin
®
C 6040) or a
printer to output the measurement data. Interface parameters: Baud rate: 9600 Data its: 8 Parity: none Stop its: 1 Handshake: none
USB-B: USB device interface for connection the calori- meters to the PC (CalWin
®
C 6040). The connection simulates a serial interface on the PC. Installation: After the C 1 was connected with the supplied data cable to the PC, the C 1 tells the Windows operating system, which device driver is required.
24 V =
Programming (Prog) interface
USB-B (USB)
PC / Printer (PC)
Scale (balance)
Heating (ext)
This driver can be downloaded from the
IKA® website.
Find your driver at the following website:
http://www.ika.com/ika/lws/download/usb-driver.zip.
Programming-
(only for service)
interface:
24V= 24 V Voltage input for the included power
supply table.
Handling the unit
1. Opening the unit
The closing of the unit in the reverse order
Q
h
h
1
2
Q
10
2. Removing the inner vessel
Inserting the inner vessel takes place in the reverse order
h
3. Removing the inner vessel base
h
4. Switch on the device
Open the device and switch the device on and off with the front power switch.
h
Display
Numbers bloc
Del button
Multi­function element
Esc button
Display: View system information, test data, as well as
menus and dialog boxes for the data input
Numbers bloc:
Entering numbers and decimal points
Del button: Deleting an entered character string to the left of
the cursor (eg the weight of a combustion sample)
Multifunction element Made possible by turning the scroll and select
menu items, as well as by pressing the Change and confirm input values
Esc button: The Esc-(Escape) function is active in the menu structure, the input window and input fields.
• In the menu structure (if any) change to the parent menu
• In the input window is changed to the higher level menu structure
• When the ESC button is pressed in an editable field, the editing is finished and the original value is restored.
Q
Control unit with display
11
Dialog box
The dialog boxes are equipped with the following elements:
Menu-heading
Selected menu item
(blue)
Scrollbar
Menu options resp. user notification
Editable input field
(yellow)
Status field
‚Back‘: Closing a info window without information input
values and return to the parent menu. Also possible using [ESC].
‚Accept‘: Closing a window. Changes made are saved by the system.
‚Start‘: Starting a measurement.
‚Details‘: Detailed information during a running measurement.
• ‚Graph‘: View graph during a running measurement.
Editable input field:
They can be enabled for value entry by pressing the multi-function element. The background color of the active field is indicated in yellow. By using the numeric keys or by turning the multi-function element, the input values can be changed.
Status field: Display the device status, which is important for
the user and operation.
Menu-heading: The name of the currently displayed window
Selected menu item: The blue background (cursor) marks the selected
item. The position can be changed by turning the multifunction element.
Scrollbar: If a menu window more entries can be displayed as the current is displayed a scrollbar. Turn the multi-function element, the other menu items are visible.
User notifications: Some windows have a field in the user notes will
be displayed.
Menu options: They allow a “getting ahead“ from the current window.
In the following example, the most frequently displayed menu options are listed and described:
• “Break off“: Closing a window, the changes made will not be accepted by the system. Also possible using [ESC].
On confirmation changes will be applied
On confirmation changes will not be applied
12
A menu language has been selected, the splash screen will change to the following information window:
INFO: Exceed the number of ignitions devices, the maximum recommended number of ignition, a pressure test must be performed with the device. An additional information window is displayed.
See capture - “Pressure test“
INFO: If the oxygen pressure is too low, a corresponding user information is displayed.
See capture - “Oxygen pressure too low“
By an verification (pressing the multifunction element) another user information on the proper use of the system is shown.
Refer to the chapter “Care and Maintenance“.
INFO: When the system is closed, a purge routine is started. See capture - “System check“
INFO: Has been selected in the settings menu under accessories
heating option, but this is not detected, user information is displayed.
See capture - “Heating detection - System start“
Switching on the system (in the open state)
After switching on the calorimeter, first the initial screen with the
IKA® logo, the calorimeter-type and the software and firmware
version.
INFO: Upon initial system start, the menu language must be selected.
See capture - “Selection of language“
Selection of language
System start
13
Pressure test
Oxygen pressure
Information screen: oxygen pressure is too low.
NOTE:
The calorimeter C 1 required for functionality an oxygen pressure of at least 20 bar.
Reproducible measurements for a working pressure
of 30 bar is required.
System check
System start when the device is closed. The system is brought through the system check in a safe and completely deflated state. For this, the oxygen and water discharge is performed.
Here, the following information screen is displayed:
INFO: During the system check detected no oxygen or the oxygen pressure is too low, the following error message is displayed:
A pressure test must be performed after a decomposition vessel has reached or exceeded the recommended number of ignition cycles. Once the pressure test has been performed a release code can be entered to enable the unit to be used for further measurements. The warning message will then disappear. The enable code can be in the following screen, by pressing the “Next button“ to be entered.
INFO: Using the unit may be further worked.
Contact for performing the pressure test, the
service department from
IKA® Works.
Observe the safety instructions!
service@ika.de
14
Heating detected
If the use of heating selected C 1.20 (accessories) in the Setting menu, but this is not recognized, the following user information is displayed.
NOTE:
Check that the heater is correctly connected and turned on.
Factory settings
Settings:
Menu: animated
Color palette: white
Key tones: on
Date device: 01.01.2012
Time device: 00:00:00 Uhr
Units: J/g
Heating: off
Disposable crucible: no
Printer: off
Service Info: no
Scale: off
Basic settings:
C-value IB1: 0
C-value IB2: 0
Reference calorific value:
26457
Start temperature: 22 °C
HO Combustible crucible
: 19839
QExternally 1: 50
QExternally 2: 0
Controlled time: off
• O rinsing: off
Cooling: on
Prolonged venting: 0
As delivered, the C 1 calorimeter system is configured as follows:
Test condition
Combustion is carried out in a calorimeter under specific conditions. The
IKA® C 1 is filled with a weighed fuel sample, the fuel sample
is ignited and the temperature increase in the calorimeter system measured. The specific calorific value of the sample is calculated as follows:
• Weight of fuel sample
• Heat capacity (C-value) of calorimeter system
• Calculated temperature increase of water in the calorimeter system
For complete combust on the inner vessel is filled with pure oxygen (quality 3.5). The pressure of the oxygen atmosphere in the inner vessel is max. 40 bar.
The exact determinat on of the gross calorific value of a substance requires that the combust on proceed under precisely defined conditons. The applicable standards are based on the following assumptons:
• Depending on the start temperature that is set, the tempera­ ture of the fuel before the combustion is 20 °C upto 30 °C.
• The water contained in the fuel before the combustion and the water that is formed when compounds comprising the fuel containing hydrogen undergo combustion is present in a liquid state.
• No oxidation of the atmospheric nitrogen has taken place.
• The gaseous products after combustion consist of oxygen, nitrogen, carbon dioxide, sulfur dioxide and the oxidation products of the sample.
• Solid substances may be formed (for example ash).
Often, however, the combustion products that form the basis of the standards are not the only products that are formed. In such cases, an analysis of the fuel sample and the products of combustion are necessary to provide data for a correction cal­culation. The standard gross calorific value is then determined from the measured gross calorific value and the analysis data.
The gross calorifc value Ho is formed as the quotient of the amount of heat liberated upon total combustion of a solid or liquid fuel and the weight of the fuel sample. The compounds comprising the fuel that contain water must be present in liquid state after the combustion.
The heat value Hu is equal to the gross calorific value less the energy of condensation of the water that was contained in the fuel and was formed by the combustion. The heat value is the more important quantity for technical purposes, since in all important technical applications, the heat value is the only quantity that can be evaluated in terms of energy.
For information on the fundamental principles of calcula­tion for gross calorific value and heat value, please refer to the applicable standards (for example: DIN 51 900; ASTM D 240;ASTM D 1989, ISO 1928).
Determination of gross calorific value
Calorimetric measurements
15
Fuel
sample
Burning
aid
Igniter
Formation of
sulphuric acid
Formation of
nitric acid
Heat quantity from:
External energy
Corrections
As determined by the system, during a combustion experiment, not only is heat generated from the combustion of the sample, but heat also occurs in the form of extraneous energy.
This may fluctuate considerably in proportion to the heat energy of the fuel sample.
The combustion heat of the cotton threat that ignites the sample and the electrical ignition energy would result in distorted values of the measurement. This effect is taken into consideration in the calculation with a correction value.
Substances with low flammability or substances that burn poorly undergo combustion in combination with combustion aids. The combustion aid is first weighed and then added to the crucible with the sample. From the weight of the combustion aid and its known specific gross calorific value, it is then possible to determine the quantity of heat that is introduced thereby. The result of the experiment must be corrected by this amount of heat. The C 14 disposable crucible is a combustible crucible that can be used instead of a traditional crucible. The disposable crucible undergoes total combustion with no residue. If a disposable crucible is used, there is then no need for a cotton thread. The crucible is in direct contact and is ignited by the fixed ignition
wire of the inner vessel.
Inner vessel in which the combustible crucible is to be used, must be fitted with an additional overlay (see accessories). The sample is weighed directly into the disposable crucible. In most cases, no additional combustion aid is necessary because the disposable crucible itself acts as the combustion aid.
Acid correction
Almost all substances to be analyzed contain sulfur and nitrogen. Under the conditions that pre-vail during calorimetric measure­ments, sulfur and nitrogen undergo combustion and form SO2, SO
3
and NOX. Sulfuric and nitric acid arise in combination with water resulting from combustion and humidity. Heat of solution is also generated. To obtain the standard gross calorific value, the effect of the heat of solution on the gross calorific value is corrected.
To obtain a defined final status and to record all acids quantitati­vely, approx 5ml distilled water or another suitable absorption solution is placed in the inner vessel. With this absorption fluid and the combustion water, the combustion gases form acids.
After the combustion, the decomposition vessel is thoroughly washed with distilled water so as to collect the condensate that has settled on the inner wall of the vessel as well. The solution obtained in this manner can now be examined with a suitable peripheral detection device for aqueous decomposition into the respective acid content.
For more information on this subject, please contact
IKA®‚ or
your nearest authorized dealer.
When calculating the energy value in C 1 external energies are taken into account from kiln furniture, however, there is no correction of acid. The calorific value is not calculated.
To do this, use the calorimeter software
IKA® CalWin®.
The calorimeter system
IKA® C 1 is a precision measuring instru-
ment for the routine determination of calorific values of solid and liquid substances. Exact measurements can however be achieved only if the individual steps of the trial are performed with care. The procedure as described in the “For your safety“ section 1 and in the following sections must therefore be adhered to exactly.
If you are burning unknown samples,
leave the room or keep well away from the calorimeter!
A few points should be noted in respect of the substances to be combusted:
• Normally, solid substances can be burned directly in powder form. Rapidly burning substances (for example benzoic acid) must not undergo combustion in loose form. Benzoic acid must only be burned in the form of pellets! Com­ bustible dust and powder must be compressed into pellets before combustion. Oven-dry dust and powder such as wood chips, hay, straw etc. burn in an explosive manner! They must be moistened first! Readily combustible liquids with a low va­ por pressure must not be come in direct contact with the cotton thread (e.g. tetramethyl dihydrogen disiloxan).
Notes on the sample
Rapidly burning substances tend to squirt. Such substances must be pressed into tablets before combustion. Suitable for this is e.g. the
IKA
®
pelleting press C 21.
• Most liquid substances can be weighed directly into the cru­ cible. Liquid substances exhibiting turbidity or containing water that will settle out must be dried or homogenized before they are weighed in. The water content of these samples must be determined.
• Highly volatile substances are placed in combustion capsules (gelatin capsules or acetobutyrate capsules, see “Accessories“) and undergo combustion together with the capsules.
• For substances with low flammability or low calorific substances use combustion aids (see “Accessories“). Before the capsules or the combustion bag is fi lled with the substance to be deter­ mined, it must be weighed in order to determine the additional extraneous energy introduced into the system from the weight and the gross calorific value. This is taken into consideration with QExtern2. The amount of combustion aid should be as little as possible.
The external energy must be determined externally.
DANGER
16
To correctly determine the gross calorific value, it is of fundamen­tal importance that the sample undergoes complete combustion. After each experiment, the crucible and all solid residues must be examined for any signs that combustion was not complete.
In case of using substances tending to squirt it is not possible to ensure complete combustion. Substances with low flammability (substances with a high content of mineral matterand low-calorific substances) will often
Complete combustion
undergo complete combustion only with the aid of combustion capsules or combustion bags (C 10/ C 12 see “Accessories“). It is also possible to use liquid combustion aids such as paran oil.
The combustion aids as well (for example the cotton thread) must undergo complete combustion. If unburned residues are left over, the experiment must be repeated.
To ensure precise, reproducible results of the measurement, the calorimeter system is calibrated after it is initially placed in service, after service jobs, after parts are replaced and at specific time inter­vals. During calibration, the heat capacity of the calorimeter system is redetermined.
Regular calibration is absolutely essential to obtain accurate measurements!
For this purpose, a specific quantity of a reference substance undergoes combustion in the
IKA® C 1 under the conditions of
the experiment. Since the gross calorific value of the reference substance is known, it is possible after combustion has occurred
Calibration
to calculate the heat capacity based on the increase in temperature of the calorimeter system. The reference substance for calorimetry on an international level is benzoic acid of the National Bureau of Standards (NBS Standard Sample 39 J) with a guaranteed gross calorific value.
For more detailed information on calibration please see the relevant standards. If the
IKA® C 1 is operated with several different decomposition
vessels, you will need to determine the heat capacity of the system for each decomposition vessel. The parts of each decomposition vessel must then not be swapped with parts from other vessels.
Ensure that calibration is carried out under the same conditions as the subsequent tests. If substances are used in the decomposition vessel in combustion tests (e.g. distilled water or solutions), you must use exactly the same amount of this substance for calibration.
Notes on calibration
For determination of calorific values the increase in temperature must be about as great as for the calibration (± 30%). 1g benzoic acid =˜ 6K. The optimum sample quantity must de determined by several trials where necessary.
• Under the conditions that prevail during calorimetric mea­ surements, sulfur and nitrogen un-dergo combustion and form SO2, SO3 and NOX. Sulfuric and nitric acid arise in combinati­ on with the water resulting form combustion. Heat of solution is also generated. This heat of solution is taken into conside­ ration in calculating the gross calorific value. In order to quanti­ tatively record and determine all acids that have been formed, about 5 ml of distilled water or another suitable substance can be added to the decomposition vessel before the experiment.
The calibration of the system must have been performed in accordance with the instructions!
Remove after measuring the complete inner vessel. Turn the inner container upside down, loosen gently the internal container bot­tom. Rinse thoroughly with distilled water, the inner area of the inner container. The rinsing water and the solution formed from the water that was added are combined and the acid content of the combined solution is examined. If the sulfur content of the combustion aid and the nitric acid correction are known, the water does not need to be analyzed.
These correction values can not be entered in C 1. For this, use the
IKA® calorimeter software CalWin®.
To increase the service life of parts to wear and tear (O rings, seals etc.) we recommend to work with water added to the experiment as a matter of general principle.
Halogen rich substances can cause corrosion on the inner con­tainer.
These substances must not be burned in C 1 with the standard inner vessel. In halogen-rich substances of special halogen-resistant inner vessel C 1.12 must be used.
Handle
h
h
h
17
The user has two display modes are available.
Measurements
The following submenus are available:
New measurement
Last result
System test
Measurement archive
Calibration archive
Memory overview
Default settings
The following submenus are available:
C-value IB1
ID IB1
C-value IB2
ID IB2
Ref. cal. value
Ho Comb. cruc.
Start temperature
QExtern 1
QExtern 2
• O2 rinse
Cool Down
Extend. O2 Emptying
Animated view List view
Settings
The following submenus are available:
Date / Time
• Units
Display / Keypad
• Language
• Accessories
Cooler test
Overview Main menu
18
Maintenance
The following submenus are available:
Change seal
Test stirrer
• O2 System blow
Servo up
Servo down INFO: When performing maintenance programs, the system must be open !
Information
The following informations are available:
• Software
• Firmware
Serial number
• Service
Pressure test
• Ignitions
Next test
• Support
The term “measurements“ below refers to both the measure­ments to calibrate the calorimeter system (calibration measure­ments) and the actual measurements for determining the calorific value. The difference lies mainly in the evaluation, whilst prepara­tion and performance make virtually no difference.
Performing the measurement
Select and confirm “Measurements“ in the main menu with the multi-function element.
In the submenu select and confirm the option “New measurement“.
Preparing and performing measurements
19
Input window NEW MEASUREMENT
Select and confirm the menu item “Sample weight“. Enter the sample weight with numeric keypad and confirm.
INFO:
A distinction is made between the use and non disposable
cup. Use of disposable crucible can be selected in Settings " Accessories. INFO: For automatic input of the weight may be a connected scale can be used
INFO: The inputted weight values are checked for permissibility. INFO: The existing store will be reviewed after the entry of a
new measurement / calibration and a user note displayed if it is not possible to save the measurement / calibration.
NOTE:
Ensure that a maximum energy input of from 40,000 joules is not exceeded.
NOTE:
For more, detailed descriptions of the individual menu items are treated in the chapter “Calorimetric Measurements“.
INFO: If a scale is used in activation of the input field, the weight is automatically queried. It can be pressed on the scale alternatively, the transmit button.
Fill in the sample weight.
Calibration off.
When you are ready with your input, press “OK” button.
Don’t close the C 1 before you have pressed “OK“ button!
During the prerequisite checking, the cotton thread can be attached to the ignition wire, hung the crucible in the provided holder and the sample are contacted with the ignition wire.
Placing the sample
When using a disposable crucible of contact between the ignition wire and the combustible crucible must be present. The fixing of the cotton thread is not necessary.
20
To increase the service life of wearing parts (O-rings, seals, etc.) it is advisable to always work with a water template.
Generally, the test sample must be selected so that the tempe­rature elevation during the measurement is below 10 K and the temperature increase of the calibration comes close (max. energy input: 40,000 J). Such exposure may damage the calorimeter.
When exceeding the maximum energy input is recommended to return the calorimeter.
To determine the potential energy initially very small sample weights (approx. 0.2g) must be chosen when working with unknown substances. If you burn unknown samples, leave the room or keep your distance from the calorimeter.
If distilled water or solutions presented in the decomposition vessel during the combustion experiment, the calibration with the same template must have been performed before.
Close the calorimeter
After the sample is loaded and the contact with the ignition wire is present and the inner vessel was placed, the calorimeter is sealed.
The following screen informs the user whether the system is closed properly and securely.
Are combustion vessel and the complete system properly closed, the measurement starts automatically.
Note:
Can the inner container and / or closure ring with handle heavy wire, please follow the instructions for oiling the seals!
2. Weigh the substance with an accuracy of 0.1 mg directly into the crucible. Where appropriate distilled water or a solution must be presented in the inner vessel.
The input of the maximum possible sample weight is limited to a maximum of 5 g.
1. Attach with a loop a cotton thread centered on the ignition wire (at the open calorimeter).
The inner container base remains in the control of sample prepara­tion in the calorimeter. But you can also use it by pulling it out individually, or for consequential - Analysis - determinations, refer to the closed inner container. The following preparations must be made to prepare the system for a measurement:
Preparing a measurement
C 6 Quartz crucible
h
Impression for distilled water or a solution as a template
21
Prerequisite checking for a measurement
When checking the preconditions for a measurement using a distinction is made with and without C 1.20 heating between. The following points should be checked:
• Existing Ignition
• Sucient oxygen present
• Open water connection
• Tempered water or using a heating their functionality
Ignition: Checking whether ignition contact available
Oxygen pressure: Check whether the system has suciently high oxygen pressure has to make a measurement
Water detected: Check whether the system has detected water
Limit ≤ Water ≤ Start temp: Check whether there is the current water temperature within the permissible temperature range
Without C 1.20 heating:
Heater test: Check the heating functionality
With C 1.20 heating:
Measurement procedure
During a measurement, different phases are passed through. These phases are described below
Phase 1: Filling the calorimeter with water
Phase 2: Filling the calorimeter with oxygen (Note: If the option-O
2
flushing under the main menu " Basic Settings is selec­ ted, the inner vessel is two times with Oxygen rinsed)
NOTE:
Alternatively, to display the graph may also have a detailed view of be selected with the following information:
Current temperature: Mean value of the actual temperatures
Ignition temperature: the currently measured temperature
Measurement time: Display: duration of the measurement
Fill time:
Filling the inner vessel with water. The fill time is user specific and must be kept constant on each measurement. Check the filter where long fill times are used.
NOTE:
After switching on the calorimeter C 1 or after a break of 2 hours during the filling of the calorimeter C 1 is still carried out an intermediate emptying (H2O - purge). Once the water sensor detects water, the stirrer is switched on. After a short waiting period begins phase 2.
22
Phase 3: Filling with oxygen
Phase 7: Venting, result is displayed
Phase 6: Post test
Phase 8: Cooling (if option under the main menu " Basic Settings " cooling is enabled) and water emptying
NOTE:
With the “cooling“ option, the energy is absorbed (by the combustion in the calorimeter entered). Deactivation, the measuring time and the accuracy of the calorimeter affecting (depending on the sample throughput).
Phase 5: The system ignites and thereby starts the main test
INFO: An ongoing measurement can be stopped by selecting “Cancel“ at any stage
Phase 4: Stabilizing the temperature and preliminary
23
Phase 9: The system is completely emptied, and can be opened.
The inner vessel can be taken down. Through a visual check is necessary to ensure that the sample in the crucible was combusted completely.
After each combustion are all parts of the inner vessel that come in contact with the combustion products to clean condensate. It is sucient to parts with an absorbent, lint-free cloth to wipe off. Combustion residues in the inner vessel, for example, soot or ash are eliminated in the same way. Can not be cleaned in the manner described (eg due to burn-in or pitting corrosion) of
Cleaning the inner vessel
the inner vessel, it shall be subjected to no cases of mechanical cleaning.
Please contact in such cases, contact your local service center or send the inner vessel for inspection and cleaning to our factory.
Follow-up measurement
By pressing the multi-function element, the result display returns to the measurement sub-menu and the next attempt can be started.
Use of a disposable crucible
The disposable crucible can optionally be used instead of metal or quartz crucible and burned during the measurement complete (see chap. “Accessories“).
INFO: 50 Joule omitted. With the combustible crucible not a cotton thread is used.
Display the storage overview
If after pressing the “New Measurement“ button, the memory overview is displayed, the capacity of the storage spaces for measurements / calibrations are almost depleted. If the maximum number is reached, no new measurements / calibrations more can be saved for this mode. By pressing the OK key to access the screen “New Measurement“. Delete measurements / calibrations are no longer needed.
Result Display
NOTE:
Because the calorimeter C 1 (all internal parts) still not in a first measurement at operating temperature, the result of the first measurement can differ significantly from the subsequent measurements of a test series. We therefore recommend not to evaluate the outcome and possibly to delete. Alternatively, a system test can be performed, see chap. “System Test“.
24
Example: sample weight in the limit and initial weight of the combustible crucible too high
Using a scale
About Settings " accessories can be selected using a scale.
If this option is selected, the displayed on the scale weight is automatically transmitted to the selected weight input field. Depending on type of balance this function may need to be be activated on the scale or the Transfer button must be pressed.
This applies both to the initial sample weight of the sample and the weight of the disposable crucible.
Permissible weight entries
In a non-permissible weight is the OK button inactive (grayed out) and can not be selected..
Sample weight: There are weights of 0.00001 g to 4.99999 g permitted!
From 2.5 g in addition the message appears that the sample
weight in the limit range is (initial mass in the limit).
Combustible crucible: There are weights of 0.00001 g to 0.99999 g permitted! Weights greater than 0.99999 g are not possible.
Confirm the selection with OK button.
25
Cancel a measurement
An ongoing measurement can be canceld at any time.
After ignition, the system waits for 1 minute with the emptying, as otherwise there is a risk that the sample is not completely burned. After confirm the cancellation, the system automatically emptied and bring it to a safe state. The calorimeter can be opened and you can return to the main menu.
New measurements
New measurement: Preparing and starting a measurement
Last result: Displaying the result of the last measurement
System test: The functionality of the calorimeter is checked
Measurement archive: The system can store up to 100 measurements
Calibration archive:
The system can be used for both modes of operation (22 °C, and 30 °C) for each inner container (1 and 2) each storing 15 calibrations.
Memory Summary: Display the memory summary
Last result
This screen shows the results of the measurement carried out previously successfully.
Measurements
26
System test
Functional testing of the system
Use for the system test one or two benzoic acid tablets. Alternatively, however, for example, an appropriate amount of sugar or an alternative fuel to be used.
The system test is used to test the functionality of the overall system. For this purpose all switching operations are carried out, which take place in a normal measurement. If the system test performed without error, the functionality is guaranteed.
The system test can be used to create a quick ˝warm up“ the system. It is faster than a normal measurement, since there is no analysis and improves the accuracy of the first measurement so that they need not be discarded
Measurement archive
In measurements Archive up to 100 measurements can be stored. A measurement taken is automatically added to the archive.
Displays the line number, the name of the measurement as well as the result.
Further details are displayed by clicking on a measurement.
27
Measurement archive: Edit
By clicking on the “EDIT“ button in the list view another screen for selecting the edit mode.
Print:
• Measurements can be selected and printed
Delete:
• Measurements can be selected and deleted
Delete all:
• All measurements are deleted
By selecting one of these menu items to the list view is displayed again. As a heading of the screen active editing is displayed. The “EDIT“ button is replaced with an “OK“ button.
Measurement archive: Delete
By clicking the “OK“ button, the following dialog appears:
By confirming with “OK“, the selected measurements will be deleted.
Measurement archive: Print
Click on the individual measurements they are selected and marked with a green tick. By clicking again the selection is reversed
By confirming the selected measurements with “OK“ followed by the display of the user note. By confirming with “OK“, the selected measurements are printed.
28
Calibration archive
After selecting the appropriate archive is displayed.
INFO: In each calibration archive up to 15 calibrations can be stored. A calibration performed is automatically added to the appropriate archive
Displays the line number, the name of the calibration and the result (C value). Below the list of the current C-value of the system is shown (in red). By confirming the multifunctional element details concerning the measurement are displayed. For the C-value calculation, select ˝Edit“.
INFO: The C-value can also manually at Main Menu > Basic settings entered alternative.
Calibration archive: Edit
By clicking the “EDIT“ button in the list view, another screen appears to select the edit mode.
Selection C values :
• Calibrations can contribute to the redefinition of the C-value be selected
Print:
• Calibrations can be selected and printed
Delete:
• Calibrations can be selected and deleted
By selecting one of these menu items to the list view is displayed again. When heading of the screen active editing is displayed. The “EDIT“ button is replaced with an “OK“ button.
Measurement archive: Delete all
By clicking the “OK“ button, the following dialog appears:
By confirming with “OK“, all measurements will be deleted.
29
Calibration archive: C-Value selection
Click on the individual calibrations they are selected and marked with a green tick. By clicking again the selection is canceled and the new C-value (mean value) is displayed below the list view directly.
By clicking the “OK“ button, the following user message appears:
By clicking the “OK“ button, the new value is accepted.
INFO: For RSD calculation of at least two C-values must be selected. Otherwise, the result is Infinite and it is “Inf“ is displayed as RSD.
Formula for calculating: X is a set of data (x1, .... xn)
and N the number
Relative Standard Deviation (RSD) = (standard deviation / mean value) x 100
Mean value: X = X
Mean
= (Σ
N
n=1
xn )/N
Standard deviation:
σ =
Σ
N
n=1
(xn - X
Mean
)2/(N - 1)
INFO: The new value is also displayed under Main menu " Basic settings.
Calibration archive: Print
Click on the individual measurements they are selected and marked with a green tick. By clicking again the selection is reversed.
By clicking on the OK button, the following user dialog appears:
By confirming with “OK“, the selected calibrations are printed.
Calibration archive: Delete
Click on the individual measurements they are selected and marked with a green tick. By clicking again the selection is reversed.
INFO: Calibrations that are used to the current C-value calculation, can not be deleted. She are grayed out and can not be selected..
By clicking the “OK“ button, the following user dialog appears:
By confirming with “OK“, the selected calibrations will be deleted.
30
C-value IB1; C-value IB2: C value for the corresponding inner vessel
ID IB1; ID IB2: Identification number of the inner vessel
Ref Energy: Calorific reference value of benzoic acid
HO Combustible crucible: Calorific value of the combustible crucible
Start temperature:
Changing the mode of operation between 22 °C and 30 °C.
QExtern 1; QExtern 2: Default values for the external energies 1 and 2
O2 rinse: The inner vessel can before the measurement of oxygen be rinsed (2x with subsequent filling)
Cool down: Cooling the system after measurement
Extend. O2 Emptying [min]: 0 - 99 adjustable
Settings
Date / Time
The setup menu contains the following menu items
Date / Time
Units
Display / Keypad
Language
Accessories
Chiller temperature
Setting the system time and date. This data is used for automatic generation of a new measurement name.
Units
The user can choose between the listed below unity representations:
• J/g
• cal/g
• BTU/lb
• MJ/kg
NOTE:
When using a gas-washing station, the ventilation time must be extended.
Default settings
31
Display / Keypad
Inverted menu view
List view
Colour range black
Language
Selecting the system language
Colour range white
The user can access this menu
select and change the background color of the display. The background color can be selected either white or black
the key tones on or off
the appearance of the menus between animated view and list view move
32
Selection of accessories
Heating: Use of heating during the measurement C 1.2
Combustible crucible: yes / no Weight input of the combustible crucible in the submenu “New measurement“ can be done manually or by data transfer from the balance.
Printer: Using a serial printer. Selection option between:
• off: no expression
• briefly: expression
• standard: result expression with temperature values
Service Information: Service information is outputted via the serial interface
Balance: Selecting a balance type. Selection option between:
out
Sartorius/Kern
Mettler-Toledo
Ohaus
Accessories
Chiller test
Viewing maintenance programs that can be performed by the user at regular intervals.
Inner vessel and the inner vessel bottom lift out and set aside.
Maintenance menu:
The system must be open to performing the maintenance programs.
The maintenance menu can be reached at Menu > Maintenance and contains the following points
Check whether the cooling water tem-
perature within the specified range of
19.5 +/- 1.5 °C at the wor-
king temperature 22 °C
or
27.5 +/- 1.5 °C at the wor- king temperature 30 °C
is located.
Maintenance
33
Change seal:
1 In the Menu Maintenance press “change seal“ button. This requires oxygen pressure be present at C 1. 2 Unscrew the piston (Pos. 38 III) with screwdriver counterclockwise 3 Remove the O-ring (Pos. 38 VI) 4 Put new O-ring on piston (Pos. 38 VI) 5 Screw in piston and tighten securely 6 “Change seal“ button again actuate pistons move to the home position
38
38
III
Stirrer test
The stirrer is switched on and off for a visual check.
O2 System blow:
1 Lift out the inner vessel from the bottom and set aside 2 After about 20 seconds the button “O2 system blow“ button. Keep openings with a paper towel or similar covered by hand 3 After about 3 seconds, this state switches itself off again.
This any remaining condensed water and dirt is in the line blown out.
This screen shows:
the software and firmware version
the serial number of the device
the date of the last service
the date of the last pressure test
the number of performed ignitions
the number of ignitions in the next pressure test must be performed
the number of performed ignitions and support information
IKA® - Contact information
blow
O-ring
(Water inlet valve
seal)
Stirrer
Servo up
The locking pin is extended.
Servo down
The locking pin is retracted.
VI (8,0 x 2,5)
locking pin
Informations
34
After input a new measurement
Prior to input a new measurement
Information about the available space when you select the menu item “New Measurement“
This information is displayed when there is a value in the critical region is located, that is, from 95 out of 100 possible measurements in memory or from 13 of 15 calibrations in store for IB1, IB2 at 22 °C or 30 °C operating temperature.
After entering the values for a new measurement archive database will be checked again and if no free space is available following message is displayed:
Depending on the selection measurement or calibration, the screen sample will vary.
The user can skip this message, then the following measurement or calibration is not saved. However, it can also switch on the button “Archive“ to the appropriate archive and then delete old or unused measurements or calibrations.
O2-Valve seal change
Use only original
IKA® spare parts!
1 Bushing (Pos. 33) with the enclosed tool SW11 screw out 2 Lift piston (Pos. 32/21) with tweezers 3 Remove the O-ring (Pos. 40), insert a new O-ring and ensure the correct position 4 Set piston (Pos. 32/21) back in the opening 5 Bushing (Pos. 33) screw with the enclosed tool SW11 handtight, paying attention to correct position of the O-ring (Pos. 32/8)
33
32/8
32/21
40
(10,0x 2,5)
(3,0 x 1,5)
User information: Memory
Exchanging of wear parts
35
Change ignition wire
1 To release the ignition wire (Pos. 24) on the ignition electrodes (Pos. 23) are both 6- hexagon nuts (Pos. 13) using spanner SW 5.5 2 Turn down the upper nut (Pos. 13I1) 3 To release the ignition wire to the ground electrode (Pos. 22) crucible holders hold hex nut (Pos. 13/3) and solve with spanner SW 5.5 4 Screw (Pos.25) and remove while holding the crucible holder 5 Remove old ignition wire iand attach new ignition wire 6 Ignition wire (Pos. 24) mounted as shown in figure on the screw (Pos.25) 7 Screw in screw (Pos.25) into the crucible holder, screw hex nut (Pos.13/3) on the screw 8 Complete unit (Pos. 25, 24,10,13) srew in the ground electrode, until the screw the ignition wire clamps 9 Align the crucible holder, no contact to the ignition electrode, with hexagon nut on the ground electrode counter 10 Align the lower nut on the ignition electrode (Pos.13/2) so that the ignition wire is horizontal 11 Ignition wire (Pos. 24) on ignition electrodes (Pos. 23) mount 12 Screw the upper hex nut (Pos. 13/1) and tighten the ignition wire with two nuts by counter holding the lower nut
Change ignition electrode
1 Loosen ignition wire at the ignition electrode as described previously 2 Hex nut (Pos. 12) and remove on the bottom of the inner vessel bottom (Pos. 2) loosen with the spanner and screw down SW7 3 Ignition electrodes (Pos. 23) by pulling it upwards 4 Take O-ring (Pos. 3) and replace with a new one 5 Install new ignition electrode in reverse order, making sure that the insulating bushing (Pos. 6) located in the inner vessel bottom 6 Tighten hex nut (Pos.12) by hand
Caution: Risk of short circuit at the nut is tightened too strong
Change the main seal
1 Remove O-ring with a suitable non-sharp tool
The O-ring can be removed with your fingers, pull the O-ring to two fingers apart, thus it is pulled out of the groove. Now it can be removed with your fingers.
Note: Rub the new O-ring before inserting it into the groove with O-ring grease (O-ring grease supplied).
This facilitates the closing of the C 1 considerably!
13/1
13
13
13
25
22
10
24
23
I
II
III
23
12
(4.0 x 1.5)
Inner vessel
Pos. 15 O-ring (48,0 x 2,0)
Uper part
Pos. 43 Quad-ring (92,0 x 4,5)
2x O-ring
h
Marked with rings must be greased as needed with the supplied grease.
h
1x X-ring
Safety note:
Use only original
IKA
®
spare parts!
In particular, the hex nuts
(Hastelloy!)
36
Change ground electrode
1 Loosen ignition wire as described above on the ground electrode 2 Unscrew the ground electrode (Pos. 22) with spanner SW 7 3 Remove and replace O-ring (Pos. 17) to mount the O-ring on the new ground electrodes 4 Screw in the new ground electrode 5 Install the new O-rings (Pos. 14) to the ground electrode
17
22
14
Run a full system test to check for leaks:
• “Measurement“ menu " menu ˝System test“.
• To perform a leak test with the system test, you must not use sample / calibration substance.
• Use to the cotton thread to check whether the ignition electrode is correctly installed. The cotton thread is not burned. Reason could be a short to ground between the ignition electrode and the inner vessel bottom (e.g. by a wrong-mounted ignition wire).
Follow the instructions of the calorimeter!
Since no sample / calibration substance is used, breaks the system test with the error message “MINTEMPDIFF“.
The tightness test is passed if the cotton thread burned and the inner vessel is dry inside. No water should flow from the C 1.
The calorimeter C 1 is subjected during the manufacture to stringent quality controls. Should any malfunctions occur, refer to the following section for a series of fault and error situations, the suitable remedial measures.
All disturbances and errors during measurement are displayed in a specific error window. There, the user receives a Circumstances of malfunctioning or error, and possible causes.
If your troubleshooting attempts is unsuccessful, contact your authorized from
IKA® Technical Service.
E01 STIRRER
This error appears if the setpoint of the stirrer speed is not reached.
A possible cause for this problem is a defective stirring motor.
Check the stirring function, using the maintenance menu.
(6,0x2,0)
(3,1x1,6)
Leak test with the system test
Fault and troubleshoot
37
E02 COMMUNICATION
If no internal communication between the software and firmware are prepared COMMUNICATION ALARM appears..
E04 MINTEMPDIFF
If 30 seconds is no temperature increase of 0.5 °C takes place after ignition, the alarm is displayed.
The most common causes for this problem are:
the cotton thread is not in contact with the sample
the contact with the fuse wire is not present due to contamination or an not properly tightened screw
the bad combustibility the sample, optionally you must use firing aids
Low oxygen pressure. Check the oxygen supply (30-40 bar)
Short circuit at the ignition electrode, crucible holder is present at the ignition electrode
E05 FILLWATER
Message appears when the system during filling with water after 50 seconds does not detect water in the upper water outlet. Causes for this problem are:
the cooler is not switched on or not connected
Water valve seal defective
no oxygen present (for switching the valves, the system requires pressure)
upper water sensor is defective
the filter in the supply line is dirty
Water valve tightly or defective
the water supply is not connect
the water filter is clogged
E03 DRIFT ALARM
Drift Alarm
In preliminary experiment or or post-experiment the temperature drift criterion is not met.
Check the stirring function.
38
E07 POSTWATERFILLING
After 45 seconds in the post-water-filling state, the current temperature is compared with the selected fill-water temperature. If it falls below the Fill-water temperature, the oxygen filling is started. Otherwise, the system additionally waiting 180 seconds, if the fill-water temperature is not reached.
A possible cause for this problem is set too high cooler temperature.
Check the cooler.
E10 FILLWATER SENSOR
Water is detected too early in outlet.
The cause of this problem are water drops on the sensor.
contact this service.
E11 SYSTEM OPEN
System is not closed properly during an ongoing measurement.
Contact closure defective!
Locking can not engage.
Twisted rotary handle.
E12 PRESSURE
Oxygen pressure is too low (min. 20 bar).
The most common cause for this problem is an unopened or empty oxygen bottle.
Check the oxygen supply!
NOTE:
20 bar, the minimal pressure necessary to keep the system is able to work. A pressure < 30 bar may have measurement errors in consequence.
E06 EMPTYWATER
Occurs when the lower water sensor during the emptying of system after 60 seconds still detects water.
Possible causes for this problem are:
the drain hose for sucking the air is in the water. Check the hose laying on the cooler (see page 8).
the pump is defective. Switch the device off and back on and then run a system test. Disconnect the return flow line and let the water drain into a lower lying container..
39
C 6040 CalWin® software KV 600 d Cooling water supply C 1.10 Combustion chamber C 1.12 Combustion chamber, halogen resistent C 1.12 Halogenresistent C 1.50 Dot-Matrix-Printer C 1.30 Venting station C 1.20 Heater C 21 Pelleting press C 29 Pressure gauge C 4 Quartz dish C 6 Quartz crucible, big C 5 Set of VA combustion crucibles C 710.2 Set of VA combustion crucibles C 15 Paran strips C 16 Parafilm C 17 Paran C 9 Gelatine capsules C 10 Acetobutyrate capsules C 12 A Combustion bags C 12 Combustion bags C 723 Benzoic acid, BIG Pack C 723 Benzoic acid, blister package C 710.4 Cotton thread C 1.103 Ignition wire
E13 HEATER
When using the heater C 1.20 this is not detected.
The most common cause for this problem is a switched off or unplugged heating during measurement.
Check the contact of the heating C 1.20
.
E16 MAXTEMPDIFF
C 1.104 Water bath additive C 5041.10 Connecting cable (for CalWin® C 6040) C 1.101 Set of spare parts 1000 C 25 Pressure regulating valve (when connected to
water pipe above 1.5 bar)
Crucible holder
Filter wrench
Oxygen adapter China
Connecting cable PC/CalW./Kal-Mettler AE
Connecting cable PC/Kal-Mettler AT, AM
Connecting cable PC/Kal-Mettler PR, PS, AG
Connecting cable PC/CalWin
®
/Kal-Sartorius
Print ribbon RC-7000
Printer paper for Star Printer SP712
Serial printer cable for SP700
Temperature increase > = 10 °C detected.
Accessories and Consumables
40
Cleaning the water filter:
Clean the water filter in the water supply line every 2-3 days in the beginning and define the regular intervals, depending on the degree of contamination for cleaning.
Cleaning the combustion chamber:
Only use lint-free cloths for cleaning und drying! Cloth or paper lints clog the filter in the water supply line. Thus, the flow rate is reduced and more error messages occur. This also affects the measuring accuracy!
The device is subject only to the natural wear and tear of com­ponents and their statistical failure rate. According to the
IKA®
manufacturer specifications we recommend an annual mainte­nance with security check and pressure test according to Pressure Equipment Directive!
Cleaning:
For cleaning, disconnect the main plug!
Only use cleaning agents recommended by
IKA®.
Contamination
Cleaning agents
Dyes
isopropyl alcohol
Building materials water containing tenside/isopropyl alcohol
Cosmetics
water containing tenside/isopropyl alcohol
Foodstuffs water containing tenside
Fuels water containing tenside
-
When cleaning, please do not allow moisture to get into the unit.
-
Wear the proper protective gloves during cleaning of the devices.
- Before performing a non-recommanded method of cleaning or
decontamination, the user must ascertain with
IKA® that this
method does not destroy the instrument.
Spare parts order:
When ordering spare parts, please give:
- Machine type
- Serial number, see type plate
- Software Version (second displayed value when the device is on)
-
Item and designation of the spare part please see www.ika.com.
Repair: Please send instrument in for repair only after it has been cleaned and is free from any materials which may consti
-
tute a health hazard.
For this you should request the “Decontamination Certificate” from
IKA®, or use the download printout of it from the
IKA® web-
site www.ika.com. Return the instrument in its original packaging. Storage packaging is not sucient. Also, please use suitable shipping package materials.
In accordance with
IKA® warranty conditions, the warranty period is 12 months. For claims under the warranty please contact your local
dealer. You may also send the machine directly to our factory, enclosing the delivery invoice and giving reasons for the claim. You will be liable for freight costs. The warranty does not cover worn out parts, nor does it apply to faults resulting from improper use, insucient care or maintenance not carried out in accordance with the instructions in this operating manual.
Maintenance
Warranty
41
Power supply unit (external) Design voltage /-frequency 100–240 V AC 50/60 Hz Power consumption max. 120 W
Calorimeter: Design voltage 24 V DC 5A Power consumption max. 120 W
Permissible duration of operation 100 % Protection type acc. DIN EN 60529 IP 20 Protection class I Overvoltage category 2 Contamination level II Permissible ambient temperature 5–40 °C Permissible relative humidity 80 % Operation at a terrestrial altitude max. 2000 m Dimensions (W x D x H) 290 x 300 x 280 mm Weight 15 kg Measuring range (max.) 40,000 J Mode of the measuring isoperibol 22 °C isoperibol 30 °C Measuring time approx 7 min Reproducibility (1g benzoic acid NBS39i) 0.15 % RSD Working temperature 20–30 °C Temperature measurement resolution 0.0001 K Cooling medium temperature
19.5 K +/- 1.5 K for measuring mode isoperibol 22 °C
27.5 K +/- 1.5 K for measuring mode isoperibol 30 °C Cooling medium permissible operating pressure 1.5 bar Cooling medium tap water / drinking water quality Cool art flow rate Flow volume 50 ... 60 l/h Oxygen operating pressure 30 ... 40 bar Interfaces RS232, USB-device
Subject to technical changes!
Technical data
IKA®-Werke GmbH & Co.KG
Janke & Kunkel-Str. 10 D-79219 Staufen Tel. +49 7633 831-0 Fax +49 7633 831-98 sales@ika.de
www.ika.com
20005032d
Loading...