HP M1165, M1166, M1167, M1175, M1176 User manual

...
HP
Viridia
M1165/66/67/75/76/77A
Component System HP
M1205A
and
Monitorinc
Viridia
26/24
Ser
ΠΠ
M1046-9001K
User’s
Volume
System
Reference
1
Information
Ca
Part Printed
First
HEWLETT® PACKARD
Number
02/99
Edition
M1046-9001K
Manual
Notice
This
document Reserved. prohibited,
Hewlett-Packard 3000 Andover,
(978)
Reproduction,
except
Minuteman
MA
687-1501
01810-1085
contains
as
allowed
Road
proprietary
adaptation,
under
information
or
translation
the
copyright
which
without
laws.
is
protected
prior
by
copyrigt
written
permis
Publication
number
M1046-9001K
Warranty
The
information
Hewlett-Packard not
limited
Hewlett-Packard consequential
material.
Copyright © Hewlett-Packard
to,
the
damages
contained
makes
no
implied
shall
not
in
warranties
be connection
in
this
document
warranty
liable
Company,
is
of
any
kind
or
merchantability
for
errors
with
the
1999
subject
with
to
regard
contained
furnishing,
change
and
herein
without
to
this
fitness
or
performance,
notice
material,
for a partic
for
incidental
or
use
in
«
Printing
New
editions Update pages
rearranged
The
documentation
date
changes incorporated when
extensive
packages
to
be
merged
due
History
of
this
document
may
be
by a revision
to
changes
printing when a new at
reprint
technical
will
issued
between
on a previous
date
edition
do
not
cause
changes
incorporate
editions
date
at
the
page
and
part
is
printed.
the
date
are
incorporated.
number
PFistEditon................................
Important
United
The
EN
Device
States
federal law
M1165/66/75/76A
60601-1-2
and
Directive
carries
(MDD).
restricts
Systems
C€o363
these
comply
Marking
devices
with
all
bottom
are
(Minor
to
change.)
February
to
UL544,
to
Council
material
and
contain
of
the
page.
not
considered
indicate
corrections
The
1999
sale
by
CSA
22.2-125,
Directive
updated
replacement
Note
its
current
and
document
or
on
the
TEC
93/42/EEC,
since
that
revised.
edition.
updates
part
order
601-1,
the
and
page
wi
of
ЕК
Eu
p
*
ni
a
The
M1167/77A
EN
60601-1,
European
The
M1205A EN60601-1, European
The
M1205A
_ ; through
and
Medical
option
and
Medical
option
5,
and
Systems
EN
60601-1-2
Device
020
and
EN60601-1-2
Device
021
complies
VDE
0871
comply
and
Directive
022
and
Directive
level
A.
with
UL2601-1,
carries
(MDD).
complies
carries
(MDD).
with
UL2601,
CSA
《CE&o3s66
with
UL2601,
C€o123
IEC
22.2
Marking
IEC
Marking
601-1,
No.
601.1-M90,
to
601-1,
to
Council
CSA
Council
CSA
C22.2
TEC
Directi
C22.2
Directiv
no.
601-1,
6
n
Electromagnetic
The
electromagnetic
performed
Manufacturer’s
EMC
During standard For
two Immunity, subjected
performance.
|
observed. observed. returned
Immunity ECG
Pleth
IEC the
power
no
anomalies
power
numerics disruption
Testing
the
and
of
some
to a field
However For
to
Levels
801-4
specifies
cord
cord,
and
in
according
Declaration
test
program
HP
proprietary
the
tests,
reduced
At
many
these
normal.
(V/m)
and
were
anomalies
RS-232
RS-232
Compatibility
compatibility
to
international
for
details.
the
M1205A
tests.
IEC
801-3
Radiated
performance
of 3 V/m
at a number
test
These
over a frequency
of
the
test
points
the
reduced
that
the
product
all
I/O
cables
observed.
disruption.
communication
were
However
sometimes
(EMC)
standards
was
During
Immunity
was observed.
frequencies
of
test
points,
radiated
levels
From
26
„175
worst
no
problems
be
subjected
greater
at
observed
At
voltages
was
(M1205A
validation
for
subjected
most
of
range over
increased
field
was
are
shown
to
200
MHz
case
to
than 3 m.
pulse
levels
in
greater
observed.
of
EMC
to
many
the
testing
and
IEC
IEC
of
26
the
specified
reduced
in
high
During
greater
the
than
Viridia
the
M1205A
with
medical
EMC
no
anomalies
801-4
Fast
801-3
specifies
to
1000
MHz
range
width
of
to
the
the
table
below.
From
.220 .391
speed
pulses
and
after
than
form
of
display
300 V applied
26/24
included
tests,
the
ECG
level
200
worst worst
most
680 V appliec
tes
devices.
both
int
wer
Transient
that
with
no
no
anom:
and
at
wh
to
10
case case
up
to
100
of
the
transier
to
the
|
th
d
|
R
In
compliance
satefy
hazard,
with
at
the
IEC
801-4,
reduced
the
M1205A
levels
performs
specified
above.
its
intended
function
withc
System
The
%
patient
|
amplifiers monitors problem
Characteristics
phenomena
monitors
already
in
used
actual
discussed
in
to
in
use
today.
display
use by
use.
above
are
This
performance
the
physiological
customers,
not
unique
signals.
interference
to is
the
M1205A
due
to
the
Among
from
electromagnetic
but
are
very sensitive
the
many
similarly
charact
hig
|
sourc
Avoiding
When
electromagnetic
can
be
taken
Eliminate reduce
Attenuate may power
Reduce to
maximum required. often
Add devices Customer
Electromagnetic
interference
to
mitigate
the
their
the
be
reduced
cord,
the
sensitivity
By
be
eliminated.
external attenuators.
such
Engineer
the
problem.
source.
strength.
coupling.
plugging
sensitivity.
reducing
as
by
moving
an
isolation
can
Possible
the
of
the
If
For
be
Interference
(EMD
is
sources
the
coupling
and/or
M1205A
the
gain
If
of
rearranging
into a different
system.
the
EMI transformer
help
ECG
of
the
becomes
in
In
system
determining
encountered
of
EMI
can
path
is
through
the
all
of
the
amplifier
the
receiving
an
unusually
or a transient
there
be
turned
leads.
circuit
EMC
testing
gain
the
the
need
are a number
off
or
the
patient
If
the
coupling
may
help.
the
was
four
EMI,
the
difficult
suppressor
for
external
problem
mover
leads,
M1205/
times
interf
may
o
t
i:
w
be
d
Intended
Use
Intended
Use
Description
The
HP
M1165/66/67/75/76/77A
Viridia
The battery
HP
26/24
M1205A
power.
Series
Viridia
Monitors
Purpose
The
HP
M1165/66/67/75/76/77A Viridia generates M1165/66/67/75/76/77A 24
Patient
The Viridia
26/24
Series
Series
alarms
Monitors
and
Population
HP
M1165/66/67/75/76/77A
26/24
Series
measures
are
is
Viridia
are
Models
recordings.
Viridia
not
intended
24CT
Viridia
and
Component
therapeutic
Viridia to
Component
network
and
Component
displays
It
exchanges
devices.
Component
be
used
Monitoring
connectable
26
CT
may
Monitoring
multiple
Monitoring
on
physiological
information
Monitoring
adult,
pediatric,
bedside
powered
System
System
patient
by
System
with
and
System
and
and
monitori
either
AC
and
parameters
compatible
the
HP
and
neonatal
the
line
the
an
dev
M120:
the
patie
E
E
E
Environment
The
HP
M1165/66/67/75/76/77A Viridia is
It serial
The Viridia
“United
vi
26/24
Series
not
intended
communicates
1/0
port.
HP
M1165/66/67/75/76/77A
26/24
Series
States
is
for
home
with
Monitors
Federal
intended
use.
devices
are
jaw
restricts
Viridia to
such
Viridia
Component
be
used
as a central
Component
prescription
this
in a clinical
station
devices
device
to
sale
Monitoring
environment
through
Monitoring
and
will
by
or
System
network
System
carry
on
the
and
by
licensed
and
the
order
the
H
interfac
the
E
followi
of a pl
|
Indications
Condition
The
HP
M1165/66/67/75/76/77A Viridia to recordings
Part
The Viridia from
26/24
measure
of
Body
HP
M1165/66/67/75/76/77A
26/24
accessory
Frequency
The
HP
M1165/66/67/75/76/77A Viridia
Physiological
The Viridia treatment,
-
66/67/75/76/77A are
26/24
HP
M1165/66/67/75/76/77A
26/24
to
well
suited
Series
and
display
of
adult,
Series
electrode,
of
Series
Series
assess
Viridia
for
for
Use
Monitors
pediatric,
or
Type
does
Use
Monitors
Purpose
Monitors
adequacy
Component
patient
Viridia
are
multiple
or
of
Tissue
Viridia
not
contact
transducer,
Viridia
are
Viridia
are
of
monitoring.
Component
generally
physiological
neonatal
indicated
patients.
with
Component
the
body
and
sensor
Component
indicated
indicated treatment,
Monitoring
for
Component
when
or
Monitoring
when
parameters
Which
or
use
to
System
the
Monitoring
tissue
of
devices.
Monitoring
when
Monitoring
the
purpose
rule
out
and
System
the
clinician
and
waves,
Device
System
the
patient.
System
prescribed
System
is
to
causes
the
of
HP
and
the
decides
to
gener:
Interac
and
the
Signals
and
the
by a clinici
and
the
gain
informe
symptoms.
M1205A
Vir
]
|
a
|
|
Patient
Adult,
pediatric,
Population
and
neonatal
non-ambulatory
patients.
Indications
for
Use
Prescription
The
HP
M1165/66/67/75/76/77A
Viridia
26/24
Series
Versus
Monitors
Over-the-Counter
Viridia
are
Component
prescription
devices.
Monitoring
System
and
the
H
viii
Warnings,
Warnings, information and section
Warning
A
“warning”
procedures
*
«
e
+
cautions,
the
HP
M1205A
refer
For
continued followed. procedures.
Explosion
Alarms - Do Adjustment patient
combines
This
equipment
professionals.
Cautions,
and
about
the
Viridia
to
the
equipment
calls
attention
are
not
safe
Instructions
Hazard-
not
of
alarm
jeopardy.
close
notes
The
HP
26/24
followed.
use
in
Do
not
rely
exclusively
volume
Remember
personal
is
only
intended
and
Notes
are
used
throughout
M1165/66/67/75/76/77A
Series.
in
general.
to
of
this
this
use
surveillance
The
the
user
equipment,
manual
this
equipment
on
the to a low that
the
for
warnings
of
in
no
audible
level
most
with
use
in
imminent
it
way
or reliable
healthcare
this
Viridia
and
is
necessary
supersede
in
the
alarm
off
during
correct
User’s
Manual
Component
cautions
hazard
that
presence
system
patient method operation
facilities
to
give
Monit
included
to
people
the
listed
established
of
flammab
for
patient
monitorů
of
patient mo:
of
monitorů
by
trained
y
ix
it
in
me
1
The ambulances.
+
This
®
To
reduce
qualified
This interference possible.
product
product
the
personnel.
equipment
is
is
risk
on
not
not
may
the
intended
intended
of
for
for
electrical
interfere
ultrasound
outside
home
shock,
with
ultrasound
display.
hospital
use.
do
NOT
Try
to
use
remove
imaging
keep
the
such
as a helicopter:
any
cover.
equipment
instruments
Refer
by
cau
as
:
fa
Indications
+
Exposure dangerous. supplies, clean
with
department
*
Although recommended this
se
Connecting
powered
interface permanent sure HP
Viridia
*
Do
not
of
the
disrupt
Caution
A
“caution”
user.
for
Use
of
electrical
Electrical
parameter
and
dry.
Thoroughly
liquids.
If
additional
or
Hewlett-Packard
this
equipment
to
equipment.
the
HP
on
is
not
lock-up
damage
the
SDN
interface
monitoring
connect a second
Viridia
calls
Model
Viridia
attention
module
contacts
contacts
module
decontamination
is
avoid
the
Viridia
monitoring
supported.
may
occur.
will
result.
cable
network
rack
24CT
or
communication.
to a condition
or
and
plug-in
dry
any
Response
shielded use
of
electrically
Error
Power
To
prevent
is
properly (SDN).
by a cable
26CT.
connections
to
connections
connections
electrical
connections
is
required
Center.
against
Electromagnetic
radiating
network
codes
cycling
and
the
(SDN)
HP
unintentional
secured
when
using a module
Using a second
or
possible
situation
saline
such
as
and
rack
Viridia
product
at
both
rack
or
other
cable
connections
that
please
contact
Interference
devices
cable
when
monitoring
will
recover
disruption
ends
connected
that
could
liquids
connectors,
mu
become
cor
yow
in
close
pr
the
pro:
neti
the
in
mor
when
conne
rack
docke
by a cz
cause.
a:
p
®
Ventilation
equipment
not
locate
*
Maintenance - Failure
employing
may
+
Do
not
the
internal solutions The
monitor
Requirements - Failure
failure
equipment
the
cause
undue
spray
components
should
and,
use
of
equipment
cleaning
be
should
in
turn,
in
an
on
the
this
equipment
solutions
and applied be
turned
jeopardize
enclosed
part
of
failure
directly
cause
to a cloth
off
during
to
meet
ventilation
the
functions
area
which
the
responsible
to
implement a satisfactory
and
possible
onto
the
equipment
and
malfunction
the
cloth
requirements
could
individual, hospital,
health
monitor.
used
cleaning.
of
automated
restrict
hazards.
Moisture
or
to
wipe
may
mor
heat
dissip:
maintene
drop]
failure.
the
mon
«
c
Cle:
Replacement parts
degradation components
and
accessories
Parts - It
of
performance.
are
listed
is
be
at
highly
used
the
recommended
with
this
Accessories
back
of
the
that
equipment.
and
parts
appropriate
only
Hewlett-Packard
Failure
for
to
individual
section
in
do
this
so
may
module
manu
1
Note—A
may
precede
At
this
time, diagrams, which
will
the
equipment
Support
note
gives
or
follow
Hewlett-Packard
component
assist
the
which
Offices
is
provided
special
the
part
user's
are
classified
instructions applicable
will
lists,
text.
make
descriptions,
appropriate
by
Hewlett-Packard
at
the
end
of
to
highlight
available
calibration
qualified
this
manual.
an
operating
on
request,
instructions,
technical
to
procedure
and
in
English
or
personnel
be
repairable. A list
to
repair
or
on
other
{
{
«
Indications
for
Use
Using
To comprehensive
The parameter
The
This place works,
* * Introducing
+
«
*
This
enable
you
User's
Reference
module
Core
section
for
new
and
the
HP
Getting
Configuring
Other
Document
of
shows
Patients
Manual
to
find
index
the
users
you
the
M1205A
Started
the
information
at
the
back.
Manual
guides.
guide
to
start
how
HP Viridia
System
is
separated
contains
because
to
get
M1165/66/67/75/76/77A
26/24
easily,
there
all
the
general
it
gives
started.
Series
is a contents
into
two
information
an
introduction
Here
is a list
Viridia
parts;
of
Component
list
the
the
core
about
to
the
major
at
the
document
system
the
front
of
an
system.
and
sections:
Monitoring
th
|
tl
*
Alarm
Recording
*
Trends
+
Installation
Care
xii
Functions
Functions
and
Data
and
and
Cleaning
Management
Patient
Safety
Parameter
These
sections monitoring, white
tab
which
Module
each
and
problem
has
Sections
contain
solving
the
title
information
if
you
of
the
section.
for
one
encounter
parameter
difficulty.
module.
Each
This
section
cover
is
sey
Note— available modules information
Note—The therefore
Although
functionality the
The
Notice
77A
Viridia
HP
M1165/66/67/75/76/77A
Gas
Module
This
Manual
B.0
monitor
a.
b.
User’s
for
you
for
screenshots
differ
to
there
Component
the
the
The
the
system.
have
the
from
the
may
may
and
is
only
can
be
Release
suffix
suffix
Reference
ordered,
parameters
displayed
what
User
be
products
not
be
the
HP
applicable
identified
B.0
of
the
of
Manual
This
means,
the
screens
and
actually
in
Monitoring
the
same.
Viridia
M1205A
the
Viridia
for
by:
label on
EPROMpack
© Monitor
EPROMpack
contains
of
will
functions
in
this
appears
your
System
This
Component
Release
the
monitor,
information
course,
manual
area
User’s
26/24
part
that
depending
not
always
is
valid
were
on
your
that
and
the
Reference
Series
B.0
versions
or
number.
Revision”: ‘Show
part
number
apply
for
ail
generated
screen
look
similar
HP
M1205A
Monitoring
only.
of
To
on a Release
for
all
to
the
during
to
Manual
System,
the
monitors
view
the
parameter
on
the
your
system.
systems.
in
demo mode
patient
the
HP
Viridia
is
intended
this
number,
SW
Rev.
B.0
model
moni
M1165
26/24
the
M102
listed
Monito
ar
H
Se
tc
al
pre
Responsibility
of
the
Manufacturer
Responsibility
Hewlett-Packard performance
assembly out by
the
electrical
the
instrument
To
ensure
conjunction
HP
M1026A wherever that
these
available.
parts
Manufacturer’s
For
South
America,
Hewlett-Packard 3000 Andover
MA
01810-1099
only
of
the
operations,
persons
optimum
with
the
Anesthetic
may
Minuteman
equipment
installation
If
of
the
Manufacturer
considers
itself
if:
extensions,
authorized
is
used
usage,
HP
M1165/66/67/75/76/77A
Gas
non-HP
cause
by
of
in
accordance
we
recommend
Module
parts
to
the
HP,
the
and are
HP
Address
North
America
Company
Road
and
responsible
re-adjustments, and
relevant
room
with
that
the
HP
Viridia
the
HP
M1205A
used,
Hewlett
equipment.
Canada:
for
any
effects
modifications
complies
instructions
parts
and
Component
Viridia
Packard
on
safety,
or
with
national
for
use.
accessories
Monitoring
26/24
Series
is
not
liable
reli:
repairs
star
are
us
Mo
for
z
For
all
other
Hewlett-Packard
Herrenberger
71034
Germany
xiv
countries:
Str.
Béblingen
GmbH
130
Condition
Part
FreguencyofUse.....................
Physiological
Paient Prescription
Responsibility
Manufacturers Declaration Declaration
The
HP
Viridia
26/24
The External Hardkey
HP
Viridia
Battery Parameter Operating
Main
Selection Task Getting
The
Viridia M1046A
of
Body
Population
of of
Series
Handheld
Alarm
Eunctions.
Model
Power
Modules
Levels
Screen
Window
WindoW
into
CMS
Computer
or
Type
Purpose
Versus
of
Conformity
Conformity
CMS
Over-the-Counter
the
Manufacturer.
Address.
and
Monitors
Keypad
Device.
26CT/24CT
Supply
the
Operating
Computer
Module
of
Tissue
(according
Power
Levels
Modules
with
to
Supply
Which
ISO/IEC
the
Guide
rr
Device
22
Interacts ea
eee
ekkn
and
EN
45014)
κ
κκ
νε
enken
erkreeer
kk
knne
....
εκ κ κ
serene,
en
εν
M1046B ECG
The
Viridia Operating Performance
M1095AFlatscreenDisplay
M1094A/B
Using
an
Safety
Computer
Output
and
26/24
Series
Rules
to
Remember
Specifications
and
ÎTE
Display
Modulk.........................................
Defibrillator
Parameter
M1092A
..................................,......,......
Marker
......................,...................
of the
HP
.......................................
CRT
Input
Module
Rack
Displays
.............................
Display. . 2.
............................
...........................
22 2
이이다
Getting
Setting
Started
Setting Setting Atlaching
Changing
Selecting a Screen
Freezing What
Changes
MakingChangestotheMainDisplay...............................
Assigning
Selecting a Screen
up
theMonitor(Viridia26/24only)................................
up
the
Parameter
the
Patient. Adjusting Starting Screen Reserving a Channel Power
up
Procedure
you
Procedure
Screen
Monitoring
Messages
Failure
your
Monitor
Display
....
Waves
to
(Viridia
Can
Configure,
the
Configuration
WavestoScreenChannels.....................................
iie
Modules
Contrast
..................,..............................
Screens
..........................................
...........................................,,.
.......................................,....,
CMS
only)
ーー
ーー
トレ ーー
.......,...................................
Procedure
Selecting
Screen
Labels
for
Realtime
Display Screens
Procedure
レス
トー
laa
sk
トレー
kk
skr
レー
n
eens
rerrker
レーー
ira
Contents-2
Selecting
.
Changing
Selecting
Selecting
Displaying OxyCRG
CSA
Wave
Trace
Configuring a Second
Other
Adjusting
Adjusting
Selecting Configuring Other The The Changing Changing
the
Number
Procedure
the
Wave
Procedure
Realtime Procedure Numerics Additional
Procedure
NotesonoxyCRG.....................
Display
Notes
Replace
Procedure
Mode
Procedure
Functions
Procedure
Procedure
Patients Status Monitor
On/Off
an
Application
Split
Display
(Viridia
on
CSA
the
Volume
the
Date
Waves
Module,
LogFuncüon..................
Revision Default the
Patient
of
Waves
..................
Overlap
.
Wave
Speeds
...
Information
Window
.....................
Screen
Trends
CMS
only)
...............
Independent
You
Can
Configure
Control
.....................................................
and
Time
......................
for
Central
Controls
Settings
Recorders . .................................
Bedside
..............................................
Function. . . .
and
Category
1...
0...
........................................
5... Display
and
Central
Patient
Category
レレ
cece
(Viridia
Recordings.
5...
CMS
1...
aaa
esse
nesker
レー
トー
レレ
ーー
ドー
een
eee
eee
only)
.....................
rs
バー
バイ
eee
eee
eens
..............
ea
krkkee
ーーーーーー
tees
aaa
ECG................
eee
Changing
Procedure
Changing
Procedure
The
Test Procedure
Analog Parameter Parameter
the
Configuration
Operating
............
Signals
Function.
................
Output
(CMS Settings Settings
Transfer Transfer
Set
Modes
.......................,.........,..........
.
only)..
..........
Messages
rr
e
rara
Other
OvervieW
HP
Using Extended
Alarm
Alarm
Alarm
Patients
Viridia
Patient
The
Other
Automatic
Configuring
HP
Viridia
Overview
To
View
an
Alert
Notification
Functions
Display
Alarm
Functions Suspending Silencing Alarm
Priorities Individual When
an
Alarm
Setup.
Getting
Changing
Setting The Nurse
into
the
deu
Care
System
Patients
AlarmOtherPatients..................................
Selection
the
Other
Patient
(Viridia
Extended
....................,.......
on
the
Alarms
and
Parameter
the
the
Volume
Call
...............
Resetting
..............
Occurs
Alarms
Alarm
Control
Relay
..............
Window
Patients
Care
Other
Alarms......
Controls
System
CMS
only)
Patients
Control
Alarms.
..........
Selection
Panel.
.
Limits
...............................
with
an
Arrhythmia
...............................,..
Bed
...........,....................
Window
eee
Computer
eee
n
Recording
General
Recorders
Contents-4
Functions
Recorder
Information
Changing Continuing a Timed Inserting a Calibration
Recording Recording Loading
Cemntral
Loading
To
Cleaning
Loading
Cleaning
the
Layouts...........
Status
Messages.
Paper...........
Recorders.
Paper
Replace
Paper
the
Paper
the
Recorder.
Recording
into
the
in
the
Printhead
into
the
Roller
on
............................................
......................................
Signal
.............................,........
Plug-In
Plug-In
in
the
Plug-In
Four
Channel
the
Four
Recorder
Recorder.
Recorder (M1117A)
Channel
ccc
..............................
.............................
.........................
Recorder
(M1117A)
ence
(Viridia
Recorder.
eee
CMS
.............
eee
only)
Admit/Discharge/End
Admitting a Patient
Changing
DischargingaPatient/EndingaCase..................................
Trends
and
Introduction
ViewingPatientData.......................
Trending Viewing Viewing Selecting
ViewingGraphTrends................
Performing
Performing Changing-or
ReviewingCalculations.........................................
Printing
Printing Printing What
Drug
Calculator..........................
Neonatal
Introduction
Patient
Calculations
to
Trends & Calculations
Priority Blood Vital
Parameters
and
Reviewing
Calculations
Reports.
Task
ScheduledReports.................................
to
Do
Event
Review
to
Neonatal
Case
................
Information
Measure-ments
Signs.......................
for
Eniering
Window
If
Your
an
Report
Byent
.......................................
ーー ーー
Graph
Calculations......
...........................................
Input
Value
Reports
Does
Review
PPP
PPP
....................................
öle
ーー ーー ーー トト
.
Trends...
..............................,..
еее
......................................
Not
некие
Print.
νο
.
0...
«ενω
εν
ωνων
ーー
ene
εν εν ν ων
トーーーーーーー
ων
ων ων
ees
a
ως
Contents-6
Viewing
Viewing
Adjusting
Data
DataTransferModule...................................
Types
Combining
Vital
Troubleshooting.........
Performance
Neonatal Manual Graphical Operating
oxyCRG
Operating
Neonatal
Event
Criteria
Event
Details Controls
Controls
Events.
............................................
Storage.
Episodes
...........................................
...............................................
νο
νεο
νεο
εννοω
...........................................
............,.................................
Event
Review
Settings
Transfer
What
is
Transferred
of
Transfer.
ToModule...............................
ToMonitor..............................
TransferringBlood
Data
Time
Conversion.
Signs,
TimeStamp...................
Reports
essere
Blood
....
«νο
Review
.............................................
νε
ενω
ρω ω ρω
ρω
ρε
Analysis
Data...................................
レレ
ーー
トー ーー
and
Graphs
.
Data
Specifications
Transfer
Module.
.........................
..........................................
.
-
νε
ων
レス トー
ων
ωρών
0...
トト
トー
0
ων
ων
εν κ ρω
トー
トト
ών
arara
ストー
トト
εν
ων
ων
8...
ων κ ων νώ ων
iie
eee
トー
トト
ーー
トー
トー
ως
ως
νο
Monitor
Viridia
Maintenance
Installation
and Patient
Safety
CMS.............................,..
Power
Source Grounding Combining Environ-ment Conden-sation
xplanation
Require-ments
the
System
Equipment
...................
iie
of
Symbofs
Checks
...............................................
.
..........................................
.
1...
used
νε
ων
νεο
ilan
εν
εωωο
eee
εννοω
never
ων ών ων
ννο ον
Performing Pressure Performing Performing Performing Performing Performing Performing Performing Performing Performing Tests
Module
for
Vuelink
the
Invasive
Self-Test
the
NBP
the
SpO2/Pleth
the
Cardiac
the
tepO2/AcpCO2
the
CO2
the
TemperatureModuleself-Tesı........................
the
Blood
the
Recorder
the
Data
..........................................
. .
Module
Module
Management
Module
Self-Test
Module
Output
AnalysisModuleSelf-Test......................
Module
and
Self-Test
Module
Module
Self-Test.
Anesthetic
Self-Test
..............................
Self-Test
Database
Self-Test......................
...........................
Self-Test...................
Gas
Module
..................
Contents-10
The
HP
Viridia
CN
26/24
This
chapter
24
Series
A Parameter
HP
Operatinglevels.......................
The The Operating
provides
Monitors.
Modules.....................
Viridia
Model
Viridia Viridia
Rules
Series
an
It
includes
26CT/24CT
CMS
Computer
26/24
Series
to
Remember............
overview
the
Parameter
of
the
following
Power
Modules
Module
Mc
HP
Viri
sectic
Supply
......
The
HP
Viridia
CMS
and
26/24
Ser
=
ο
5
= >
a =
©
=
E
E
2
=
Introduction
©
q
q © N
3
=
Introduction
Ci
The
HP
24
Series patient the
systems needed, unit.
M1165/66/67/75/76/77
Monitors,
monitors
can
or
you
hereafter
with
have
can
interchange
networking
modules
Viridia
referred
and
added
the
modules
CMS
to
data
or
removed
and
the
as
the
“monitor”,
management
between
HP
M120
ca
at a later
syste
2
Viridia
CMS
Note—Some the
HP
Viridia
respective
“Viridia
The
The system module
Note—This an ITE
CMS
following
HP
Viridia
consists
and
1.
M1167/77A Satellite
XGA
compatible
(information
2.
M1165/75/A Integral
3.
M1166/76A Module
features
CMS
sections
only”
system
CMS
of
parameter
Color
Module
system
Technology
Monochrome
Module
color
Rack
explained
and
the
are
marked
or
“Viridia
types
is
available
three
individual
modules:
Flatscreen
Rack
is
also
display
Rack
CRT
in
this
manual
HP
Viridia
are
available with
controller
Equipment)
Display
26/24
throughout
26/24
available:
as a choice
parts; a display
Display
to
CRT
Display and
and
Series
the
Series
of
and
an
External
drive
displays
Computer
are
not
avail:
Monitor manual only’.
three
Computer
commerciali
witl
system
module,
Alarm
(XGA
Computer
Module
T
1-2
The
HP
Viridia
CMS
and
26/24
Series
Monitors
M1167/77A
System
Display Module
Display
Computer
Parameter
Module
Module
Modules
OO
RU
eae
M1095A
M1046B
Satellite
Computer
Module
a
10.4”
Computer
Rack
parameter
Flatscreen
Modules
Display
Module
The
HP
Viridia
CMS
and
26/24
Serie
£a
ο
|
on
EN
q 5
Ts
oc
cu
FP
Introduction
.
M1167/77A
XGA
Display
System
with
External
Alarm
External
Device
Alarm
Device
Computer Module
Display
Computer
Parameter
Module
Module
Modules
o
©
ITE
Display
M1046B
Satellite
+
Computer
Rack
(XGA
Type)
Module
1-4
The
HP
Viridia
CMS
and
26/24
Series
Monitors
M1165/75A
Display Module
Parameter Modules
and
M1166/76A
System
Medel
Types
Display
Computer
Parameter
All
system
types:
Full
Title
The
HP
The
HP
for
Anesthesia
The
HP
for
Neonatal
Note—In
CMS,
the
Module
Module
Modules
types
Viridia
Viridia
Viridia
this
manual,
HP
Viridia
are
Component
Component
Care
Component
Care
also
the
ACMS
M1094A/B/92A
M1046A
Integral
available
-
Monitoring
Monitoring
Monitoring
system
and
will
the
14"
Computer
and/or
Satellite
as a choice
System
System
System
be
referred
HP
Viridia
CRT
Module
of
three
to
NCMS.
Display
Rack
dif
Abb
НРУ
HP
HPA
as
the
\
The
HP
Viridia
CMS
and
26/24
Serie
=
o
E Ki
z >
a =
a
=
E
©
©
a
N
я S N
>
= o
n
ON
Introduction
Display
Modules
Below
are
described
indicator
Screen
brightness
Power
on/off
(also
remote
for
power
labeled
in
more
Power
light
control
switch
on/oft
supply)
diagrams detail
mama
è
in
the
of
the
following
display
sections
modules.
The
co
1-6
The
HP
Viridia
CMS
and
Control Panel
26/24
indicator
Power switch
M1092A / M1094B
Series
Monitors
light
On/Off
Screen' Contrast
Screen Brightness
contro!
CRT
Connector keypad
control
Display
for rer
cable
Module
Loading...
+ 160 hidden pages