The information contained in this document is subject to change without notice.
Agilent Technologies makes no warranty of any kind with regard to this material,
including but not limited to, the implied warranties of merchantability and fitness for a
particular purpose. Agilent Technologies shall not be liable for errors contained herein or
for incidental or consequential damages in connection with the furnishing, performance, or
use of this material.
Certification
Agilent Technologies certifies that this product met its published specifications at the time
of shipment from the factory. Agilent Technologies further certifies that its calibration
measurements are traceable to the United States National Institute of Standards and
Technology, to the extent allowed by the Institute's calibration facility, and to the
calibration facilities of other International Standards Organization members.
Warranty
This Agilent Technologies instrument product is warranted against defects in material
and workmanship for a period of one year from date of shipment. During the warranty
period, Agilent Technologies will, at its option, either repair or replace products which
prove to be defective.
For warranty service or repair, this product must be returned to a service facility
designated by Agilent Technologies. Buyer shall prepay shipping charges to
Agilent Technologies and Agilent Technologies shall pay shipping charges to return the
product to Buyer. However, Buyer shall pay all shipping charges, duties, and taxes for
products returned to Agilent Technologies from another country.
Agilent Technologies warrants that its software and firmware designated by
Agilent Technologies for use with an instrument will execute its programming instructions
when properly installed on that instrument. Agilent Technologies does not warrant that
the operation of the instrument, or software, or firmware will be uninterrupted or
error-free.
2
Page 3
LIMITATION OF WARRANTY
The foregoing warranty shall not apply to defects resulting from improper or
inadequate maintenance by Buyer, Buyer-supplied software or interfacing,
unauthorized modification or misuse, operation outside of the environmental
specifications for the product, or improper site preparation or maintenance.
NO OTHER WARRANTYISEXPRESSEDORIMPLIED. AGILENT TECHNOLOGIES
SPECIFICALLY DISCLAIMS THE IMPLIED WARRANTIES OF
MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE.
E
XCLUSIVE REMEDIES
THE REMEDIES PROVIDED HEREIN ARE BUYER'S SOLE AND EXCLUSIVE
REMEDIES. AGILENT TECHNOLOGIES SHALL NOT BE LIABLE FOR ANY
DIRECT, INDIRECT, SPECIAL, INCIDENTAL, OR CONSEQUENTIAL DAMAGES,
WHETHER BASED ON CONTRACT, TORT, OR ANY OTHER LEGAL THEORY.
Assistance
Product maintenance agreements and other customer assistance agreements are available
for Agilent Technologies products.
For any assistance, contact your nearest Agilent Technologies Sales and Service Office.
Cleaning
The instrument front and rear panels should be cleaned using a soft cloth with water or a
mild soap and water mixture.
This chapter identifies the performance test procedures which test the
electrical performance of the analyzer. Allow the analyzer to warm up
in accordance with the temperature stability specifications before
performing the tests in this chapter.
None of the test procedures involve removing the cover of the analyzer.
11
Page 12
Calibrating
Calibrating
Calibrating
Calibration
Calibration verifies that the analyzer performance is within all
specifications. It is time consuming and requires extensive test
equipment. Calibration consists of all the performance tests. For a
complete listing of the performance tests, see the performance
verification tests table for your specific analyzer.
Operation Verification
Operation verification only tests the most critical specifications. These
tests are recommended for incoming inspection, troubleshooting, or
after repair. Operation verification requires less time and equipment
than the calibration. See the performance verification tests table for
your analyzer.
Calibration Cycle
The performance tests in Chapter 2 , "Performance Verification
Tests"should be used to check the analyzer against its specifications
once every year. Specifications are listed in this calibration guide.
The 300 MHz frequency of the CAL OUT signal must be checked at the
same time and adjusted if necessary. Refer to the “10 MHz Frequency
Reference Adjustment” procedure in the assembly-level repair service
guide.
When A 3335A Source Is Not Available
The 3335A Synthesizer Level Generator signal source has become
obsolete because parts used in the manufacture of this instrument are
no longer available from suppliers. To meet the need of our customers,
this calibration guide has been revised to add new performance
verification tests that do not use the 3335A Synthesizer Level
Generator. This revision includes the addition of signal sources
required to replace the 3335A, changes to the test equipment setup
illustrations, and changes in the steps required to execute the
procedures.
Since all of our customers will not need to replace their 3335A
Synthesizer Level Generators immediately, the original performance
tests which use the 3335A signal generator have been retained. The
revisions have been incorporated in this calibration guide as Chapter
2a, "PerformanceVerificationTests:If 3335A Source Not Available" and
Chapter 3a, “Performance Test Records: If 3335A Source Not
Available.”
12Chapter1
Page 13
Calibrating
Calibrating
Performance Verification Test Tables
The tables on the following pages list the performance tests in
Chapter 2 and Chapter 2a. Select the analyzer option being calibrated
and perform the tests marked in the option column.
A dot indicates that the test is required for calibration. A diamond
indicates that the test is required for operation verification. Note that
some of the tests are used for both calibration and operation
verification.
Table 1-18591C Performance Verification Tests
Performance Test NameCalibration for
Instrument Option:
a
701
704
011
130
Std
1.10 MHz Reference Output Accuracy•
2.10 MHz Precision Frequency Reference Output Accuracy•••••
4.Frequency Readout and Marker Count Accuracy◊◊◊◊◊◊
6.Noise Sidebands◊◊◊◊◊◊
7.System Related Sidebands••••••
8.
8a.
Frequency Span Readout Accuracy
Frequency Span Readout Accuracy
59.Absolute Amplitude, Vernier, and Power Sweep Accuracy•
62.Tracking Generator Level Flatness•
64.Harmonic Spurious Outputs•
66.Non-Harmonic Spurious Outputs•
68.Tracking Generator Feedthrough•
73.Gate Delay Accuracy and Gate Length Accuracy••••
74.
74a.
Gate Card Insertion Loss
Gate Card Insertion Loss
b
c
••••
••••
75.TV Receiver, Video Tester•
a. Use this column for all other options not listed in this table.
b. If a 3335A source is not available, use the alternative performance test with the
same number found in Chapter 2a, “Performance Verification Tests: If 3335A
Source Not Available.”
c. If a 3335A source is not available, substitute this performance test for the one
with the same number found in Chapter 2 , “Performance Verification Tests.”
d. “Part 2: Third Order Intermodulation Distortion, 50 MHz” is not required for
operation verification.
14Chapter1
Page 15
Calibrating
Calibrating
Table 1-28591E Performance Verification Tests
Performance Test NameCalibration for Instrument Option:
a
001
004
010
011
101
103
105
130
Std
1.10 MHz Reference Output Accuracy•••••••••
2.10 MHz Precision Frequency Reference Output
•
Accuracy
4.Frequency Readout and Marker Count
◊◊◊◊◊◊◊◊◊◊
Accuracy
6.Noise Sidebands◊◊◊◊◊◊◊◊◊◊
7.System Related Sidebands••••••••••
8.
8a.
Frequency Span Readout Accuracy
Frequency Span Readout Accuracy
Performance Test NameCalibration for Instrument Option:
a
001
004
010
011
101
103
105
Std
130
64.Harmonic Spurious Outputs••
66.Non-Harmonic Spurious Outputs••
68.Tracking Generator Feedthrough••
72.
72a.
CISPR Pulse Response
CISPR Pulse Response
73.Gate Delay Accuracy and Gate Length
b
c
•
•
•
Accuracy
74.
74a.
Gate Card Insertion Loss
Gate Card Insertion Loss
b
c
•
•
75.TV Receiver, Video Tester◊
a. Use this column for all other options not listed in this table.
b. If a 3335A source is not available, use the alternative performance test with the
same number found in Chapter 2a, “Performance Verification Tests: If 3335A
Source Not Available.”
c. If a 3335A source is not available, substitute this performance test for the one
with the same number found in Chapter 2 , “Performance Verification Tests.”
d. “Part 2: Third Order Intermodulation Distortion, 50 MHz” is not required for
operation verification.
107
16Chapter1
Page 17
Calibrating
Calibrating
Table 1-38593E Performance Verification Tests
Performance Verification Test NameCalibration for Instrument Option:
a
004
010
026
027
101
103
105
130
Std
1.10 MHz Reference Output Accuracy•••••••••
2.10 MHz Precision Frequency Reference Output
•
Accuracy
3.Comb Generator Frequency Accuracy••••••••••
5.Frequency Readout and Marker Count
◊◊◊◊◊◊◊◊◊◊
Accuracy
6.Noise Sidebands◊◊◊◊◊◊◊◊◊◊
7.System Related Sidebands••••••••••
9.
9a.
Frequency Span Readout Accuracy
Frequency Span Readout Accuracy
Performance Verification Test NameCalibration for Instrument Option:
a
004
010
026
027
101
103
105
130
71.Tracking Generator LO Feedthrough
Std
•
Amplitude
72.
72a.
CISPR Pulse Response
CISPR Pulse Response
73.Gate Delay Accuracy and Gate Length
b
c
•
•
•
Accuracy
74.
74a.
Gate Card Insertion Loss
Gate Card Insertion Loss
b
c
•
•
75.TV Receiver, Video Tester◊
a. Use this column for all other options not listed in this table.
b. If a 3335A source is not available, use the alternative performance test with the
same number found in Chapter 2a, “Performance Verification Tests: If 3335A
Source Not Available.”
c. If a 3335A source is not available, substitute this performance test for the one
with the same number found in Chapter 2 , “Performance Verification Tests.”
d. “Third Order Intermodulation Distortion” is not required for operation
verification.
107
18Chapter1
Page 19
Calibrating
Calibrating
Table 1-48594E Performance Verification Tests
Performance Verification Test NameCalibration for Instrument
Option:
a
004
010
101
103
105
130
Std
1.10 MHz Reference Output Accuracy•••••••
2.10 MHz Precision Frequency Reference Output
•
Accuracy
4.Frequency Readout and Marker Count Accuracy◊◊◊◊◊◊◊◊
6.Noise Sidebands◊◊◊◊◊◊◊◊
7.System Related Sidebands••••••••
9.
9a.
Frequency Span Readout Accuracy
Frequency Span Readout Accuracy
Performance Verification Test NameCalibration for Instrument
Option:
a
004
010
101
103
105
Std
130
71.Tracking Generator LO Feedthrough Amplitude•
72.
72.
CISPR Pulse Response
CISPR Pulse Response
b
c
•
•
73.Gate Delay Accuracy and Gate Length Accuracy•
74.
74a.
Gate Card Insertion Loss
Gate Card Insertion Loss
b
c
•
•
75.TV Receiver, Video Tester◊
a. Use this column for all other options not listed in this table.
b. If a 3335A source is not available, use the alternative performance test with the
same number found in Chapter 2a, “Performance Verification Tests: If 3335A
Source Not Available.”
c. If a 3335A source is not available, substitute this performance test for the one
with the same number found in Chapter 2 , “Performance Verification Tests.”
d. “Third Order Intermodulation Distortion” is not required for operation
verification.
107
20Chapter1
Page 21
Calibrating
Calibrating
Table 1-58594Q Performance Verification Tests
Performance Verification Test NameCalibrationfor
Instrument
Option:
190195704
1.10 MHz Reference Output Accuracy•
2.10 MHz Precision Frequency Reference Output Accuracy••
4.Frequency Readout and Marker Count Accuracy◊◊◊
6.Noise Sidebands◊◊◊
7.System Related Sidebands••
9.
9a.
Frequency Span Readout Accuracy
Frequency Span Readout Accuracy
Performance Verification Test NameCalibration for Instrument
Option:
a
004
010
101
103
105
Std
130
71.Tracking Generator LO Feedthrough Amplitude•
72.
72.
CISPR Pulse Response
CISPR Pulse Response
b
c
•
•
73.Gate Delay Accuracy and Gate Length Accuracy•
74.
74.
Gate Card Insertion Loss
Gate Card Insertion Loss
b
c
•
•
75.TV Receiver, Video Tester◊
a. Use this column for all other options not listed in this table.
b. If a 3335A source is not available, use the alternative performance test with the
same number found in Chapter 2a, “Performance Verification Tests: If 3335A
Source Not Available.”
c. If a 3335A source is not available, substitute this performance test for the one
with the same number found in Chapter 2 , “Performance Verification Tests.”
d. “Third Order Intermodulation Distortion” is not required for operation
verification.
107
Chapter 123
Page 24
Calibrating
Calibrating
Table 1-78596E Performance Verification Tests
Performance Verification Test NameCalibration for Instrument
Option:
a
004
010
101
103
105
130
Std
1.10 MHz Reference Output Accuracy•••••••
2.10 MHz Precision Frequency Reference Output
•
Accuracy
3.Comb Generator Frequency Accuracy◊◊◊◊◊◊◊◊
5.Frequency Readout and Marker Count Accuracy◊◊◊◊◊◊◊◊
6.Noise Sidebands◊◊◊◊◊◊◊◊
7.System Related Sidebands••••••••
9.
9a.
Frequency Span Readout Accuracy
Frequency Span Readout Accuracy
Performance Verification Test NameCalibration for Instrument
Option:
a
004
010
101
103
105
Std
130
71.Tracking Generator LO Feedthrough Amplitude•
72.
72a.
CISPR Pulse Response
CISPR Pulse Response
b
c
•
•
73.Gate Delay Accuracy and Gate Length Accuracy•
74.
74a.
Gate Card Insertion Loss
Gate Card Insertion Loss
b
c
•
•
75.TV Receiver, Video Tester◊
a. Use this column for all other options not listed in this table.
b. If a 3335A source is not available, use the alternative performance test with the
same number found in Chapter 2a, “Performance Verification Tests: If 3335A
Source Not Available.”
c. If a 3335A source is not available, substitute this performance test for the one
with the same number found in Chapter 2 , “Performance Verification Tests.”
d. “Third Order Intermodulation Distortion” is not required for operation
verification.
107
Chapter 125
Page 26
Calibrating
Safety
Safety
Familiarize yourself with the safety symbols marked on the analyzer,
and read the general safety instructions and the symbol definitions
given in Chapter 12 , “Safety and Regulatory Information,” before you
begin verifying performance of the spectrum analyzer.
Before You Start
There are four things you should do before starting a performance
verification test:
• Switch the analyzer on and let it warm up in accordance with the
temperature stability specification.
• Read “Making a Measurement” in your analyzer user's guide.
• After the analyzer has warmed up as specified, perform the
self-calibration procedure documented in “Improving Accuracy With
Self-Calibration Routines” in the 8590 E-Series and L-Series
Spectrum Analyzer User's Guide, 8591C Cable TV Analyzer,
Spectrum Analyzer Reference User's Guide,or8594Q QAM Analyzer
Spectrum Analyzer Reference User's Guide. The performance of the
analyzer is only specified after the analyzer calibration routines
have been run and if the analyzer is autocoupled.
• Read the rest of this section before you start any of the tests, and
make a copy of the Performance Verification Test Record described
below in “Recording the test results.”
Test Equipment You Will Need
Table 1-8 through Table 1-11 list the recommended test equipment for
the performance tests. The tables also list recommended equipment for
the analyzer adjustment procedures which are located in the
8590 Series Analyzers Assembly-Level Repair Service Guide. Any
equipment that meets the critical specifications given in the table can
be substituted for the recommended model.
If a 3335A Synthesizer/Level Generator is not available, see Table 1-8
through Table 1-10 for alternative recommended test equipment,
accessories, and adapters.
26Chapter1
Page 27
Calibrating
Recording the Test Results
Recording the Test Results
Performance verification test records, for each spectrum analyzer, are
provided in Chapter 3 , "Performance Test Records" and Chapter 3a,
"Performance Test Records: If 3335A Source Not Available" following
the tests.
Each test result is identified as a TR Entry in the performance test and
on the performance verification test record. We recommend that you
make a copy of the performance verification test record, record the test
results on the copy, and keep the copy for your calibration test record.
This record could prove valuable in tracking gradual changes in test
results over long periods of time.
Frequency and Amplitude Self-Calibration
Perform the frequency and amplitude self-calibration routines at least
once per day, or if the analyzer fails a verification test. To perform
self-calibration, press
must be up to operating temperature in order for this test to be valid.
Press
continuously fails one or more specifications, complete any remaining
tests and record all test results on a copy of the test record. Then refer
to Chapter 1 for instructions on how to solve the problem.
CAL STORE when the test is complete. If the analyzer
CAL then CAL FREQ & AMPTD. The instrument
Periodically Verifying Operation
The analyzer requires periodic verification of operation. Under most
conditions of use, you should test the analyzer at least once a year with
either operation verification or the complete set of performance
verification tests.
Chapter 127
Page 28
Calibrating
Periodically Verifying Operation
Table 1-8Recommended Test Equipment
EquipmentCritical Specifications for Equipment
Substitution
Digital VoltmeterInput Resistance: ≥10 MΩ
Recommended
Use
Model
3456AP,A,T
Accuracy: ±10 mV on 100 V range
DVM Test LeadsFor use with 3456A34118BA,T
Frequency Counter
Frequency: 10 MHz
5334A/BP,A,T
b
Resolution: ±0.002 Hz
External Timebase
Frequency StandardFrequency: 10 MHz
5061BP,A
Timebase Accuracy (Aging): <1 ×10−9/day
Measuring ReceiverCompatible with Power Sensors
8902AP,A,T
dB Relative Mode
Resolution: 0.01 dB
Reference Accuracy: ±1.2%
Microwave Frequency
Counter
OscilloscopeBandwidth: dc to 100 MHz
Frequency Range: 9 MHz to 7 GHz
Timebase Accuracy (Aging): <5 × 10
−10
5343AP,A,T
/day
54501AT
Vertical Scale Factor of 0.5 V to 5 V/Div
Power MeterPower Range: Calibrated in dBm and dB
436AP,A,T
relative to reference power −70 dBm to
+44 dBm, sensor dependent
a
Power SensorFrequency Range: 100 kHz to 1800 MHz
Maximum SWR:
1.60 (100 kHz to 300 kHz)
1.20 (300 kHz to 1 MHz)
1.1 (1 MHz to 2.0 GHz)
1.30 (2.0 to 2.9 GHz)
Power Sensor
c
Frequency Range: 1 MHz to 2 GHz
Maximum SWR:
1.18 (600 kHz to 2.0 GHz) 75 Ω
Power Sensor,
Low Power
Frequency Range: 300 MHz
Amplitude Range: −20 dBm to −70 dBm
Maximum SWR: 1.1 (300 MHz)
Power Sensor
d
Frequency Range: 50 MHz to 26.5 GHz
Maximum SWR:
1.10 (300 MHz)
1.15 (50 MHz to 100 MHz)
1.10 (100 MHz to 2.0 GHz)
1.15 (2.0 GHz to 12.4 GHz)
1.20 (12.4 GHz to 18.0 GHz)
1.25 (18.0 GHz to 26.5 GHz)
8482AP,A,T
8483AP,A,T
8484AP,A,T
8485AP,A,T
28Chapter1
Page 29
Periodically Verifying Operation
Table 1-8Recommended Test Equipment (Continued)
Calibrating
EquipmentCritical Specifications for Equipment
Substitution
Pulse Generator
e
Period Range: 1 ms to 980 ms ±2%,
single pulse mode
Level −2 V to +2 V
Transition Time: 6 ns ±10%, ±1 ns
Pulse Width: 150 ns to 3 µs ±1% ±1 ns
Pulse GeneratorFrequency: 100 Hz
Duty Cycle: 50%
Output: TTL
Quasi-Peak
Down-Loadable Program (DLP)11946-10001P,A,T
Detector Driver
Signal GeneratorFrequency Range: 1 MHz to 1000 MHz
Amplitude Range: −35 to +16 dBm SSB
Noise: <−120 dBc/Hz at 20 kHz offset
Spectrum Analyzer,
Microwave
Frequency Range: 100 kHz to 7 GHz
Relative Amplitude Accuracy:
100 kHz to 1.8 GHz: <±1.8 dB
Frequency Accuracy: <±10 kHz @ 7 GHz
f
Synthesized Sweeper
Frequency Range: 10 MHz to 22 GHz
Frequency Accuracy (CW): ± 0.02%
Leveling Modes: Internal and External
Modulation Modes: AM Power
Level Range: −35 to +16 dBm
Recommended
Use
Model
8161AP,T
8116AP,T
8640B Option
P,A,T
002 or 8642A
8566A/BP,A,T
8340A/B
P,A,T
or 83630A
a
Synthesizer/Function
Generator
Synthesizer/Level
Generator
Universal Counter
g
h
Frequency Range: 0.1 Hz to 500 Hz
Frequency Accuracy: ±0.02%
Waveform: Triangle
Frequency Range: 1 kHz to 80 MHz
Amplitude Range: +12 to −85 dBm
Flatness: ±0.15 dB
Attenuator Accuracy: ±0.09 dB
Time Interval Range: 25 ms to 100 ms
Single Operation Range:
+2.5 Vdc to −2.5 Vdc
3325BP,T
3335AP,A,T
5316BP,T
Chapter 129
Page 30
Calibrating
Periodically Verifying Operation
Table 1-8Recommended Test Equipment (Continued)
EquipmentCritical Specifications for Equipment
Substitution
Base Band Signal
Source
Video ModulatorDifferential Gain: <2%
When a 3335A source is not available:
Synthesized Signal
Generator
i
Capable of providing the following VIT
signals:
FCC composite
NTC7 composite
or
CCIR 17 and CCIR 330
Differential Phase: <0.5°
Frequency Range: 100 kHz to 2560 MHz8663AP
Recommended
Model
Magni Signal
Creator
8780A,Scientific
Atlanta 6350 or
6351 with
Option FAOC
a. P = Performance Test, A = Adjustment, T = Troubleshooting
b. Precision Frequency Reference only
c. 8591E and 8591C only
d. Not for 8591E or 8591C
e. For Option 103 or 8591C
f. For 8591E, 8591C, 8593E Option 026 or Option 027, 8594E,
8594Q, 8595E, and 8596E
g. If a 3335A source is not available, substitute an 8663A signal generator.
h.For Option 105 and 8591C
i. For Option 107
Use
P,T
P,T
a
30Chapter1
Page 31
Table 1-9Recommended Accessories
Calibrating
Periodically Verifying Operation
EquipmentCritical Specifications for
Accessory Substitution
Active Probe
b
5 Hz to 500 MHz41800AT
Recommended
Model
Active Probe300 kHz to 3 GHz85024AT
Attenuator, 3 dB
c
Type N (m to f) Attenuation: 3 dB
Frequency: dc to 12.4 GHz
8491A
Option 003
Attenuator, 10 dBType N (m to f) Frequency: 300 MHz8491A
Option 010
Attenuator, 20 dB
d
Type N (m to f)
Attenuation: 20 dB
8491A
Option 020
Frequency: dc to 12.4 GHz
Attenuator, 1 dB StepAttenuation Range: 0 to 12 dB
355CP,A
Frequency Range: 50 MHz
Connectors: BNC female
Attenuator, 10 dB StepAttenuation Range: 0 to 30 dB
355DP,A
Frequency Range: 50 MHz
Connectors: BNC female
8721AP,T
Directivity: >40 dB Maximum
VSWR: 1.1:1
Transmission Arm Loss: 6 dB
(nominal)
Coupling Arm Loss: 6 dB (nominal)
Logic PulserTTL voltage and current drive levels546AT
Logic ClipTTL voltage and current drive levels548AT
Directional CouplerFrequency Range: 1.7 GHz to 8 GHz
0955-0125P,T
Coupling: 16 dB (nominal) Max.
Coupling Deviation: ±1dB
Directivity: 14 dB minimum
Flatness: 0.75 dB maximum
VSWR: <1.45
Insertion Loss: <1.3 dB
Low Pass Filter,Cutoff Frequency: 4.4 GHz
11689AP,A
Rejection at 5.5 GHz: >40 dB
Chapter 131
Page 32
Calibrating
Periodically Verifying Operation
Table 1-9Recommended Accessories (Continued)
EquipmentCritical Specifications for
Accessory Substitution
Low Pass Filter, 50 MHzCutoff Frequency: 50 MHz
Rejection at 80 MHz: >50 dB
Low Pass Filter, 300 MHzCutoff Frequency: 300 MHz
Bandpass Insertion Loss: <0.9 dB
at 300 MHz
Stopband Insertion Loss: >40 dB
at 435 MHz
Modulator
Teletech SC35B
Power Splitter
Power Splitter
f
g
Frequency 50 MHz
ON/OFF RATIO >70 dB
Switching Speed 2 ns
Insertion Loss: 5 dB
Frequency Range: 50 kHz to 1.8 GHz
Insertion Loss: 6 dB (nominal)
Output Tracking: <0.25 dB
Equivalent Output SWR: <1.22:1
Frequency Range: 50 kHz to 22 GHz
Insertion Loss: 6 dB (nominal)
Output Tracking: <0.25 dB
Equivalent Output SWR: <1.22:1
Recommended
Model
0955-0306P,T
0955-0455P,A,T
0955-0533P,T
11667AP,A
11667BP,A
Use
a
Termination, 50 ΩImpedance: 50 Ω (nominal)
(2 required for Option 010)
Termination
Termination, 75 Ω
When a 3335A source is not available:
Attenuator/Switch DriverCompatible with 8494G and 8496G
Attenuator Interconnect
Kit
h
i
Impedance: 75 Ω (nominal)
(2 required for option 011)
programmable step attenuators
Mechanically and electrically connects
8494A/G and 8496A/G
a. P = Performance Test, A = Adjustment, T = Troubleshooting
b. 8591E only
c. Option 103 and 8591C only
d. 8593E, 8594E, 8594Q, 8595E, and 8596E
e. Option 107 only
f. 8591C, 8591E, and 8593E
g. 8593E, 8594E, 8594Q, 8595E, and 8596E
h.8595E and 8596E only
i. 8591E and 8591C only
908AP,T
909D
909E Option 201P,T
11713AP, A
11716 SeriesP, A
32Chapter1
Page 33
Table 1-10Recommended Adapters
Calibrating
Periodically Verifying Operation
EquipmentCritical Specifications for Accessory
Substitution
Recommended
Model
Use
a
AdapterAPC 3.5 (f) to APC 3.5 (f)5061-5311P,A,T
Adapter
Adapter
b
c
BNC (f) to dual banana plug1251-1277P,A,T
SMA (f) to SMA (f)1250-1158P,A,T
AdapterBNC (m) to BNC (m)1250-0216P,A,T
AdapterSMA (m) to SMA (m)1250-1159P, A, T
AdapterBNC (m) to BNC (m), 75 Ω1250-1288P,A,T
AdapterBNC (f) to SMB (m)1250-1237A,T
AdapterBNC tee (m) (f) (f)1250-0781T
Adapter
d
MNC (m) to SMA (f)1250-1700P,A,T
AdapterType N (f) to APC 3.5 (f)1250-1745P,A,T
AdapterType N (f) to APC 3.5 (m)1250-1750P,A,T
Adapter
e
Type N (f) to SMA (f)1250-1772P,A,T
AdapterType N (m) to APC 3.5 (m)1250-1743P,A,T
AdapterType N (m) to APC 3.5 (f)1250-1744P,A,T
Adapter
f
Type N (f) to BNC (f)1250-1474P,A,T
AdapterType N (f) to BNC (m)1250-1477P,A,T
AdapterType N (f) to BNC (m), 75 Ω1250-1534P,A,T
AdapterType N (m) to BNC (f) (4 required)1250-1476P,A,T
AdapterType N (m) to BNC (m) (2 required)1250-1473P,A,T
AdapterType N (f) to N (f)1250-1472P,A,T
AdapterType N (m) to N (m)1250-1475P,A,T
AdapterType N (f) to N (f), 75 Ω1250-1529P,A,T
Adapter
g
Type N (f), 75 Ω, to Type N (m), 50 Ω1250-0597P,A,T
AdapterSMB (f) to SMB (f)1250-0692A,T
AdapterSMC (m) to SMC (m)1250-0827A,T
AdapterSMB (m) to SMB (m)1250-0813A,T
Chapter 133
Page 34
Calibrating
Periodically Verifying Operation
Table 1-10Recommended Adapters (Continued)
EquipmentCritical Specifications for Accessory
Substitution
Adapter,
Minimum Loss
Adapter
When a 3335A source is not available:
AdapterBNC (f) to SMA (m)1250-1200P, A, T
AdapterBNC tee (f, m, f)1250-0781P, A, T
h
50 to 75 Ω, matching Frequency Range: dc to
2 GHz Insertion Loss: 5.7 dB
Type N tee (m) (f) (f)1250-0559P,T
Recommended
Model
11852BP,A,T
Use
a
a. P = Performance Test, A = Adjustment, T = Troubleshooting
b. 8591C and 8591E only
c. 8594E, 8594Q, 8595E, and 8596E only
d. 8593E only
e. 8593E, 8594E, 8594Q, 8595E, and 8596E only
f. 8591C, 8591E, 8594E, 8594Q 8595E, and 8596E only
g. 8591E Option 001 and Option 011 only
h.8593E, 8594E, 8595E, and 8596E with Option 010 only
34Chapter1
Page 35
Table 1-11Recommended Cables
Calibrating
Periodically Verifying Operation
EquipmentCritical Specifications for Cable
Substitution
Cable
b
Cal Comb SMA (m) to (m)08592-60061P,A,T
Recommended
Model
Use
CableSMA (m) to (m), 61 cm (18 in)8120-1578P,A,T
Cable Assembly
Length: approximately 15 cm (6 in)
8120-1292A
c
Connectors: BNC (f) to Alligator Clips
Cable AssemblyLength: ≥91 cm (36 in)
11102AA
Connectors: Banana Plug to Alligator Clips
Cable
d
Frequency Range: 10 MHz to 26.5 GHz
8120-4921P,A
Maximum SWR: <1.4 at 26.5 GHz
Length: ≥91 cm (36 in)
Connectors: APC 3.5 (m) both ends
Maximum Insertion Loss 2 dB
(2 required)
CableFrequency Range: 50 MHz to 7 GHz
5061-5458P,A,T
Length: ≥91 cm (36 in)
Connectors: SMA (m) both ends
CableType N, 183 cm (72 in)11500AP,A,T
CableType N, 62 cm (24 in)11500B/CP,A,T
a
CableType N, 152 cm (60 in)11500DP,A,T
CableFrequency Range: dc to 310 MHz
10502AP,A,T
Length: 20 cm (9 in)
Connectors: BNC (m) both ends
Cable
e
BNC, 75 Ω, 30 cm (12 in)5062-6452P,A,T
CableBNC, 75 Ω, 120 cm (48 in)15525-80010P,A,T
Cable, TestLength: ≥91 cm (36 in)
85680-60093A,T
Connectors: SMB (f) to BNC (m)
(2 required)
a. P = Performance Test, A = Adjustment, T = Troubleshooting
b. For 8593E only
c. Not for 8591E
d. For 8593E Option 026 or Option 027, 8594E, 8594Q, 8595E,
8596E only
e. For 8591E Option 001 and Option 011 only
Chapter 135
Page 36
Calibrating
Periodically Verifying Operation
36Chapter1
Page 37
2Performance Verification Tests
These tests verify the electrical performance of the spectrum analyzer.
Allow the spectrum analyzer to warm up in accordance with the
temperature stability specifications before performing the tests.
If a 3335A source is not available, use the alternative performance test
with the same number found in Chapter 2a.
37
Page 38
Performance Verification Tests
Calibrated Attenuator Settings, 8494G and 8496G
Calibrated Attenuator Settings, 8494G and
8496G
Refer to Table 2-1 for each test in Chapter 2 which requires the use of a
calibrated attenuator.
If your instrument is equipped with a Precision Frequency Reference,
perform “10 MHz Precision Frequency Reference Output Accuracy,”
instead.
The settability is measured by changing the setting of the
digital-to-analog converter (DAC) which controls the frequency of the
timebase. The frequency difference per DAC step is calculated and
compared to the specification.
The related adjustment for this performance verification test is the
“10 MHz Frequency Reference Adjustment.”
Equipment Required
Microwave frequency counter
Frequency standard
Cable, BNC, 122 cm (48 in) (2 required)
The test results will be invalid if REF UNLK is displayed at any time
during this test. REF UNLK will be displayed if the internal reference
oscillator is unlocked from the 10 MHz reference. A REF UNLK might
occur if there is a hardware failure or if the jumper between 10 MHz
REF OUTPUT and EXT REF IN on the rear panel is removed.
Output Accuracy, 8590 E-Series Option 004,
8591C, and 8594Q
If the spectrum analyzer is not equipped with a Precision Frequency
Reference, perform “10 MHz Reference Output Accuracy,” instead.
This test measures the warmup characteristics of the 10 MHz reference
oscillator. The ability of the 10 MHz oscillator to meet its warmup
characteristics gives a high level of confidence that it will also meet its
yearly aging specification.
A frequency counter is connected to the 10 MHz REF OUTPUT. After
the spectrum analyzer has been allowed to cool for at least 60 minutes,
the spectrum analyzer is powered on. A frequency measurement is
made five minutes after power is applied and the frequency is recorded.
Another frequency measurement is made 25 minutes later (30 minutes
after power is applied) and the frequency is recorded. A final frequency
measurement is made 60 minutes after power is applied. The difference
between each of the first two frequency measurements and the last
frequency measurement is calculated and recorded.
The related adjustment for this procedure is “10 MHz Precision
Frequency Reference Accuracy Adjustment.”
Equipment Required
Frequency counter
Frequency standard
Cable, BNC, 122 cm (48 in) (2 required)
Procedure
The spectrum analyzer must have been allowed to sit with the power
for at least 60 minutes before performing this procedure. This
adequately simulates a cold start. A cold start is defined as the
spectrum analyzer being powered on after being for at least 60 minutes.
1. Allow the spectrum analyzer to sit with the power off for at least 60
minutes before proceeding. Connect the equipment as shown in
Figure 2-2.
2. Set the spectrum analyzer LINE switch on. Record the Power On
Time below.
Power On Time _____________
42Chapter2
Page 43
Performance Verification Tests
2. 10 MHz Precision Frequency Reference Output Accuracy, 8590 E-Series Option 004, 8591C,
and 8594Q
Figure 2-210 MHz Precision Frequency Reference Accuracy Test Setup
3. Set the frequency counter controls as follows:
FUNCTION/DATA ....................................... FREQ A
INPUT ....................................................................... A
X10 ATTN ............................................................ OFF
AC ......................................................................... OFF
50 Ω Z ................................................................... OFF
AUTO TRIG ............................................................ON
100 kHz FILTER A ...............................................OFF
4. On the frequency counter select a 10 second gate time by pressing
GATE TIME 10 GATE TIME. Offset the displayed frequency by
−10.0 MHz by pressing
CHS/EEX, 6, SELECT/ENTER, SELECT ENTER. The frequency counter
MATH, SELECT/ENTER, CHS/EEX, 10,
should now display the difference between the INPUT A signal and
10.0 MHz with 0.001 Hz resolution.
5. Proceed with the next step 5 minutes after the Power On Time noted
in step 2.
6. Wait at least two periods for the frequency counter to settle. Then
record the frequency counter reading in Table 2-3 as Counter
Reading 1 with 0.001 Hz resolution.
7. Proceed with the next step 30 minutes after the Power On Time
noted in step 2.
8. Record the frequency counter reading in the Table 2-3 as Counter
Reading 2 with 0.001 Hz resolution.
9. Proceed with the next step 60 minutes after the Power On Time
noted in step 2.
Chapter 243
Page 44
Performance Verification Tests
2. 10 MHz Precision Frequency Reference Output Accuracy, 8590 E-Series Option 004, 8591C,
and 8594Q
10.Wait at least two periods for the frequency counter to settle. Record
the frequency counter reading in Table 2-3 as Counter Reading 3
with 0.001 Hz resolution.
14.Record the results as TR Entry 2 of the performance verification test
record.
Performance verification test “10 MHz Precision Frequency Reference
Output Accuracy” is now complete.
44Chapter2
Page 45
Performance Verification Tests
3. Comb Generator Frequency Accuracy, 8593E and 8596E
3. Comb Generator Frequency Accuracy, 8593E
and 8596E
A 100 MHz signal from a synthesized source and the output from a
comb generator are applied to the input of the spectrum analyzer. The
source frequency is adjusted until the two signals appear at the same
frequency. The frequency setting of the source is then equal to the comb
generator frequency and this frequency is compared to the specification.
The related adjustment procedure for this performance verification test
is “Comb Generator Frequency Adjustment.”
Equipment Required
Synthesized sweeper
Power splitter
Cable, APC mm (m) 91 cm (36 in)
Cable, SMA 61 cm (18 in) (m) to (m)
Adapter, Type N (m) to APC 3.5 (m)
Adapter, 3.5 mm (f) to 3.5 mm (f)
Chapter 245
Page 46
Performance Verification Tests
3. Comb Generator Frequency Accuracy, 8593E and 8596E
Procedure
1. Connect the equipment as shown in Figure 2-3.
Option 026 only: Omit the Type N to APC adapter.
Figure 2-3Comb Generator Frequency Accuracy Test Setup
2. Press INSTRUMENT PRESET on the synthesized sweeper, then set
the controls as follows:
PRESET on the spectrum analyzer,then wait for preset routine
to finish. Set the spectrum analyzer by pressing the following keys:
FREQUENCY, 100, MHz
AUX CTRL, COMB GEN ON OFF (ON)
SPAN, 10, MHz
AMPLITUDE, REF LVL, 10, dB
BW, RES BW AUTO MAN, 10, kHz
4. On the spectrum analyzer, press the following keys:
PEAK SEARCH
MKR FCTN, MK TRACK ON OFF (ON)
SPAN, 100, kHz
5. Press AMPLITUDE and adjust the reference-level setting until the
signal peak is 10 dB below the reference level.
6. Set the synthesized sweeper RF on. Adjust the synthesized sweeper
power level until the two signals are the same amplitude.
46Chapter2
Page 47
Performance Verification Tests
3. Comb Generator Frequency Accuracy, 8593E and 8596E
7. Set SCALE LOG LIN (LOG) to 2 dB on the spectrum analyzer.
8. If necessary, readjust the synthesized sweeper power level until the
two signals are the same amplitude.
9. Set the synthesized sweeper CW to 100 MHz. A very unstable signal
will probably appear. The peak amplitude should be at least 3 dB
greater in amplitude than either of the individual signals.
10.Adjust the synthesized sweeper CW setting until a single signal
appears to rise and fall in amplitude at the slowest rate (1 Hz
frequency resolution will be necessary). The signal peak should be
displayed approximately 6 dB above the amplitude of the individual
signals.
11.Record the synthesized sweeper CW frequency setting as TR Entry 1
in the appropriate performance verification test record in Chapter 3.
The frequency should be between 99.993 MHz and 100.007 MHz.
Performance verification test “Comb Generator Frequency Accuracy” is
now complete.
Chapter 247
Page 48
Performance Verification Tests
4. Frequency Readout and Marker Count Accuracy, 8591C, 8591E, 8594E, and 8594Q
4. Frequency Readout and Marker Count
Accuracy, 8591C, 8591E, 8594E, and 8594Q
The frequency readout accuracy of the spectrum analyzer is tested with
an input signal of known frequency. By using the same frequency
standard for the spectrum analyzer and the synthesized sweeper, the
frequency reference error is eliminated.
The related adjustment for this performance test is the “Sampler Match
Adjustment.”
Equipment Required
Synthesized sweeper
Adapter, Type N (f) to APC 3.5 (m)
Adapter, APC 3.5 (f) to APC 3.5 (f)
Cable, Type N, 183 cm (72 in)
Cable, BNC, 122 cm (48 in)
Additional Equipment for 75 Ω Input
Adapter, minimum loss
Adapter, Type N (f) to BNC (m), 75 Ω
Procedure
This performance test consists of two parts:
Part 1: Frequency Readout Accuracy
Part 2: Marker Count Accuracy
Perform “Part 1: Frequency Readout Accuracy” before “Part 2: Marker
Count Accuracy.”
Part 1: Frequency Readout Accuracy
1. Connect the equipment as shown in Figure 2-4 (8591E, 8591C) or as
shown in Figure 2-5 (8594E, 8594Q). Remember to connect the
10 MHz REF OUT of the synthesized sweeper to the EXT REF IN of
the spectrum analyzer.
48Chapter2
Page 49
Performance Verification Tests
4. Frequency Readout and Marker Count Accuracy, 8591C, 8591E, 8594E, and 8594Q
Figure 2-48591E and 8591C Frequency Readout Accuracy Test Setup
Figure 2-58594E and 8594Q Frequency Readout Accuracy Test Setup
CAUTIONUse only 75 Ω cables, connectors, or adapters on instruments with 75 Ω
inputs, or damage to the input connector will occur.
2. Perform the following steps to set up the equipment:
a. Press INSTRUMENT PRESET on the synthesized sweeper, then
4. Frequency Readout and Marker Count Accuracy, 8591C, 8591E, 8594E, and 8594Q
Additional Frequency Readout Accuracy Steps for
Option 130
7. Set the spectrum analyzer by pressing the following keys:
BW, RES BW AUTO MAN, 300, Hz
SPAN, 20, kHz
8. Press PEAK SEARCH on the spectrum analyzer.
9. Record the MKR frequency reading as TR Entry 4 of the
performance verification test record. The reading should be within
the limits of 1.49999924 GHz and 1.50000076 GHz.
“Part 1: Frequency Readout Accuracy” is now complete for the Option
130. Continue with “Part 2: Marker Count Accuracy.”
Part 2: Marker Count Accuracy
Perform “Part 1: Frequency Readout Accuracy” before performing this
procedure.
1. Press
PRESET on the spectrum analyzer, then wait for the preset
routine to finish. Set the spectrum analyzer to measure the marker
count accuracy by pressing the following keys:
FREQUENCY, 1.5, GHz
SPAN, 20, MHz
BW, RES BW AUTO MAN, 300, kHz
MKR FCTN, MK COUNT ON OFF (ON)
More 1 of 2
CNT RES AUTO MAN, 100, Hz
2. Press PEAK SEARCH, then wait for a count be taken (it may take
several seconds).
3. Record the CNTR frequency reading as TR Entry 5 of the
performance verification test record. The reading should be within
the limits of 1.4999989 GHz and 1.5000011 GHz.
4. Change the spectrum analyzer settings by pressing the following
keys:
SPAN, 1, MHz
MKR FCTN, MK COUNT ON OFF (ON)
More 1 of 2
CNT RES AUTO MAN, 10, Hz
Chapter 251
Page 52
Performance Verification Tests
4. Frequency Readout and Marker Count Accuracy, 8591C, 8591E, 8594E, and 8594Q
5. Press PEAK SEARCH, then wait for a count be taken (it may take
several seconds).
6. Record the CNTR frequency reading as TR Entry 6 in the
appropriate performance verification test record in Chapter 3. The
reading should be within the limits of 1.49999989 GHz and
1.50000011 GHz.
If you are testing a spectrum analyzer equipped with Option 130
continue with step 7.
Performance verification test “Frequency Readout and Marker Count
Accuracy” is now complete for all other spectrum analyzers.
Additional Marker Count Accuracy Steps for
Option 130
7. Set the spectrum analyzer by pressing the following keys:
BW, RES BW AUTO MAN, 300, Hz
SPAN, 20, kHz
8. Press PEAK SEARCH on the spectrum analyzer.
9. Record the MKR frequency reading as TR Entry 7 in the appropriate
performance verification test record in Chapter 3. The reading
should be within the limits of 1.49999989 GHz and 1.50000011 GHz.
10.Set the spectrum analyzer by pressing the following keys:
BW, RES BW AUTO MAN, 30, Hz
SPAN, 2, kHz
11.Press PEAK SEARCH, MKR FCTN, Mk Track On Off (ON), then wait
until the count is completed (it may take several seconds).
12.Record the MKR reading as TR Entry 8 in the appropriate
performance verification test record in Chapter 3. The reading
should be within the limits of 1.49999989 and 1.50000011.
Performance verification test “Frequency Readout and Marker Count
Accuracy” is now complete for spectrum analyzers equipped with
Option 130.
52Chapter2
Page 53
Performance Verification Tests
5. Frequency Readout and Marker Count Accuracy, 8593E, 8595E, and 8596E
5. Frequency Readout and Marker Count
Accuracy, 8593E, 8595E, and 8596E
The frequency readout accuracy of the spectrum analyzer is tested with
an input signal of known frequency. By using the same frequency
standard for the spectrum analyzer and the synthesized sweeper, the
frequency reference error is eliminated.
The related adjustments for this performance verification test are
“Sampler Match Adjustment” and “Frequency Reference Adjustment.”
Equipment Required
Synthesized sweeper
Adapter, Type N (f) to APC 3.5 (m)
Adapter, APC 3.5 (f) to APC 3.5 (f)
Cable, APC 3.5, 91 cm (36 in)
Cable, BNC, 122 cm (48 in)
Additional Equipment for Option 026
Adapter, 3.5 mm (f) to 3.5 mm (f)
Procedure
This performance verification test consists of two parts:
Part 1: Frequency Readout Accuracy
Part 2: Marker Count Accuracy
Perform “Part 1: Frequency Readout Accuracy” before “Part 2: Marker
Count Accuracy.”
Chapter 253
Page 54
Performance Verification Tests
5. Frequency Readout and Marker Count Accuracy, 8593E, 8595E, and 8596E
Part 1: Frequency Readout Accuracy
1. Connect the equipment as shown in Figure 2-6. Remember to
connect the 10 MHz REF OUT of the synthesized sweeper to the
EXT REF IN of the spectrum analyzer.
Option 026 only: Use the 3.5 mm adapter to connect the cable to the
spectrum analyzer input.
Figure 2-6Frequency Readout Accuracy Test Setup
2. Perform the following steps to set up the equipment:
a. Press INSTRUMENT PRESET on the synthesized sweeper, then
POWER LEVEL ........................................... −10 dBm
b. Press
PRESET on the spectrum analyzer, then wait for the preset
routine to finish. Set the spectrum analyzer by pressing the
following keys:
FREQUENCY, 1.5, GHz
SPAN, 20, MHz
3. Press PEAK SEARCH on the spectrum analyzer to measure the
frequency readout accuracy.
4. Record the MKR frequency reading in the performance verification
test record as indicated in Table 2-5. The reading should be within
the limits shown.
5. Change to the next spectrum analyzer span setting listed in
Table 2-5.
6. Repeat steps 3 through 5 for each spectrum analyzer frequency and
span setting listed in Table 2-5.
54Chapter2
Page 55
Performance Verification Tests
5. Frequency Readout and Marker Count Accuracy, 8593E, 8595E, and 8596E
If you are testing a spectrum analyzer equipped with Option 130
continue with step 7.
Part 1 of performance verification test “Frequency Readout and Marker
Count Accuracy” is now complete for all other spectrum analyzers.
Additional Frequency Readout Accuracy Steps for
Option 130
7. Set the synthesized sweeper CW to 1.5 GHz.
8. Set the spectrum analyzer by pressing the following keys:
FREQUENCY, 1.5, GHz
BW, 300, Hz
SPAN, 20, kHz
9. Press PEAK SEARCH on the spectrum analyzer.
10.Record the MKR frequency reading as TR Entry 16 of the
performance verification test record. The reading should be within
the limits of 1.49999924 GHz and 1.50000076 GHz.
Part 1 of performance verification test “Frequency Readout and Marker
Count Accuracy” is now complete for spectrum analyzers equipped with
Option 130. Proceed with “Part 2: Marker Count Accuracy.”
Part 2: Marker Count Accuracy
Perform “Part 1: Frequency Readout Accuracy” before performing this
procedure.
1. Press
routine to finish.
2. Set the spectrum analyzer to measure the marker count accuracy by
pressing the following keys:
PRESET on the spectrum analyzer, then wait for the preset
FREQUENCY, 1.5, GHz
SPAN, 20, MHz
BW, RES BW AUTO MAN, 300, kHz
MKR FCTN, MK COUNT ON OFF (ON)
More 1 of 2
CNT RES AUTO MAN, 100, Hz
3. Press PEAK SEARCH, then wait for a count be taken (it may take
several seconds).
Chapter 255
Page 56
Performance Verification Tests
5. Frequency Readout and Marker Count Accuracy, 8593E, 8595E, and 8596E
4. Record the CNTR frequency reading as TR Entry 17 of the
performance verification test record. The reading should be within
the limits shown in Table 2-6.
5. Change the spectrum analyzer settings by pressing the following
keys:
SPAN, 1, MHz
MKR FCTN, MK COUNT ON OFF (ON)
More 1 of 2
CNT RES AUTO MAN, 10, Hz
6. Press PEAK SEARCH, then wait for a count be taken (it may take
several seconds).
7. Record the CNTR frequency reading as TR Entry 18 of the
performance verification test record. The reading should be within
the limits shown in Table 2-6.
8. Repeat step 2 through step 7 for each spectrum analyzer setting
listed in Table 2-6.
If you are testing a spectrum analyzer equipped with Option 130
continue with step 9.
Part 2 of performance verification test “Frequency Readout and Marker
Count Accuracy” is now complete for all other spectrum analyzers.
Additional Marker Count Accuracy Steps for
Option 130
9. Set the synthesized sweeper CW to 1.5 GHz.
10.Set the spectrum analyzer by pressing the following keys:
FREQUENCY, 1.5, GHz
BW, RES BW AUTO MAN, 300, Hz
SPAN, 20, kHz
11.Press PEAK SEARCH on the spectrum analyzer.
12.Record the MKR frequency reading as TR Entry 27 of the
performance verification test record. The reading should be within
the limits of 1.49999989 GHz and 1.50000011 GHz.
13.Set the spectrum analyzer by pressing the following keys:
BW, RES BW AUTO MAN, 30, Hz
SPAN, 2, kHz
56Chapter2
Page 57
5. Frequency Readout and Marker Count Accuracy, 8593E, 8595E, and 8596E
14.Press PEAK SEARCH, MKR FCTN, Mk Track On Off (ON), then wait
until the count is completed (it may take several seconds).
15.Record the MKR reading as TR Entry 28 in the appropriate p
performance verification test record in Chapter 3. The reading
should be within the limits of 1.49999989 and 1.50000011.
Part 2 of performance verification test “Frequency Readout and Marker
Count Accuracy” is now complete for spectrum analyzers equipped with
Option 130.
6. Noise Sidebands, 8590 E-Series, 8591C, and 8594Q
6. Noise Sidebands, 8590 E-Series, 8591C,
and 8594Q
A 500 MHz CW signal is applied to the input of the spectrum analyzer.
The marker functions are used to measure the amplitude of the carrier
and the noise level 10 kHz, 20 kHz, and 30 kHz above and below the
carrier. The difference between these two measurements is compared to
specification after the result is normalized to 1 Hz.
There are no related adjustment procedures for this performance test.
Equipment Required
Signal generator
Cable, Type N, 183 cm (72 in)
Additional Equipment for Option 026
Adapter, APC 3.5 (f) to Type N (f)
Additional Equipment for 75 Ω Input
Adapter, minimum loss
Adapter, Type N (f) to BNC (m), 75 Ω
Procedure
This performance test consists of three parts:
Part 1: Noise Sideband Suppression at 10 kHz
Part 2: Noise Sideband Suppression at 20 kHz
Part 3: Noise Sideband Suppression at 30 kHz
Perform part 1 before performing part 2 or part 3 of this procedure.
A worksheet is provided at the end of this procedure for calculating the
noise sideband suppression.
Chapter 259
Page 60
Performance Verification Tests
6. Noise Sidebands, 8590 E-Series, 8591C, and 8594Q
Part 1: Noise Sideband Suppression at 10 kHz
1. Set the signal generator controls as follows:
FREQUENCY ............................................... 500 MHz
CAUTIONUse only 75 Ω cables, connectors, or adapters on instruments with 75 Ω
inputs, or damage to the input connector will occur.
3. Press
PRESET on the spectrum analyzer, then wait for the preset
routine to finish. Set the spectrum analyzer by pressing the
following keys:
FREQUENCY, 500, MHz
SPAN, 10, MHz
60Chapter2
Page 61
Performance Verification Tests
6. Noise Sidebands, 8590 E-Series, 8591C, and 8594Q
4. Press the following spectrum analyzer keys to measure the carrier
amplitude.
PEAK SEARCH
MKR FCTN, MK TRACK ON OFF (ON)
SPAN, 200, kHz
BW, 1, kHz
VID BW AUTO MAN, 30, Hz
MKR FCTN, MK TRACK ON OFF (OFF)
SGL SWP
5. Wait for the completion of a sweep, then press PEAK SEARCH.
6. Record the MKR amplitude reading in Table 2-7 as the Carrier
Amplitude.
7. Press the following spectrum analyzer keys to measure the noise
sideband level at +10 kHz:
MARKER ∆, 10, kHz
MKR, MARKER NORMAL
8. Record the MKR amplitude reading in Table 2-7 as the Noise
Sideband Level at +10 kHz.
9. Press the following spectrum analyzer keys to measure the noise
sideband level at −10 kHz:
PEAK SEARCH
MARKER ∆, −10, kHz
MKR, MARKER NORMAL
10.Record the MKR amplitude reading in Table 2-7 as the Noise
Sideband Level at −10 kHz.
11.Record the more positive value, either Noise Sideband Level at
+10 kHz or Noise Sideband Level at −10 kHz fromTable 2-7 as the
Maximum Noise Sideband Level.
12.Calculate the Noise Sideband Suppression (NSS) by subtracting the
Carrier Amplitude (Carrier AMP) from the Maximum Noise
Sideband Level (NSL) at 10 kHz as follows:
NSS = Maximum NSL − Carrier AMP
13.Record the Noise Sideband Suppression at 10 kHz in the
performance verification test record as TR Entry 1. The suppression
should be ≤−60 dBc.
Chapter 261
Page 62
Performance Verification Tests
6. Noise Sidebands, 8590 E-Series, 8591C, and 8594Q
Part 1 of performance verification test “Noise Sidebands” is now
complete. Proceed with Part 2 of performance verification test “Noise
Sidebands.”
Part 2: Noise Sideband Suppression at 20 kHz
1. Press the following spectrum analyzer keys to measure the noise
sideband level at +20 kHz:
MKR, MARKER ∆, 20, kHz
MARKER NORMAL
2. Record the MKR amplitude reading in Table 2-7 as the Noise
Sideband Level at +20 kHz.
3. Press the following spectrum analyzer keys to measure the noise
sideband level at −20 kHz:
PEAK SEARCH
MARKER ∆, −20, kHz
MKR, MARKER NORMAL
4. Record the MKR amplitude reading in Table 2-7 as the Noise
Sideband Level at −20 kHz.
5. Record the more positive value, either Noise Sideband Level at
+20 kHz or Noise Sideband Level at −20 kHz from Table 2-7 as the
Maximum Noise Sideband Level.
6. Subtract the Carrier Amplitude (Carrier AMP) from the Maximum
Noise Sideband Level (NSL) at 20 kHz using the equation below.
NSS = Maximum NSL − Carrier AMP
7. Record the Noise Sideband Suppression at 20 kHz in the
performance verification test record as TR Entry 2. The suppression
should be ≤−70 dBc.
Part 2 of performance verification test “Noise Sidebands” is now
complete. Proceed with Part 3 of performance verification test “Noise
Sidebands.”
62Chapter2
Page 63
Performance Verification Tests
6. Noise Sidebands, 8590 E-Series, 8591C, and 8594Q
Part 3: Noise Sideband Suppression at 30 kHz
1. Press the following spectrum analyzer keys to measure the noise
sideband level at +30 kHz:
MKR, MARKER ∆, 30, kHz
MARKER NORMAL
2. Record the MKR amplitude reading in Table 2-7 as the Noise
Sideband Level at +30 kHz.
3. Press the following spectrum analyzer keys to measure the noise
sideband level at −30 kHz:
PEAK SEARCH
MARKER ∆, −30, kHz
MKR, MARKER NORMAL
4. Record the MKR amplitude reading in Table 2-7 as the Noise
Sideband Level at −30 kHz.
5. Record the more positive value, either Noise Sideband Level at
+30 kHz or Noise Sideband Level at −30 kHz from Table 2-7 as the
Maximum Noise Sideband Level.
6. Subtract the Carrier Amplitude (Carrier AMP) from the Maximum
Noise Sideband Level (NSL) at 30 kHz using the equation below.
NSL = Maximum NSL − Carrier AMP
7. Record the Noise Sideband Suppression at 30 kHz in the appropriate
performance verification test record as TR Entry 3. The suppression
should be ≤−75 dBc.
NOTEThe resolution bandwidth is normalized to 1 Hz as follows:
1 Hz noise-power = (noise-power in dBc) − (10 x log[RBW])
For example, −60 dBc in a 1 kHz resolution bandwidth is normalized to
−90 dBc/Hz.
Performance verification test “Noise Sidebands” is now complete.
Chapter 263
Page 64
Performance Verification Tests
6. Noise Sidebands, 8590 E-Series, 8591C, and 8594Q
Table 2-7Noise Sideband Worksheet
DescriptionMeasurement
Carrier Amplitude___________dBm or dBmV
Noise Sideband Level at +10 kHz___________dBm or dBmV
Noise Sideband Level at −10 kHz___________dBm or dBmV
Maximum Noise Sideband Level at ±10 kHz___________dBm or dBmV
Noise Sideband Level at +20 kHz___________dBm or dBmV
Noise Sideband Level at −20 kHz___________dBm or dBmV
Maximum Noise Sideband Level at ±20 kHz___________dBm or dBmV
Noise Sideband Level at +30 kHz___________dBm or dBmV
Noise Sideband Level at −30 kHz___________dBm or dBmV
Maximum Noise Sideband Level at ±30 kHz___________dBm or dBmV
64Chapter2
Page 65
Performance Verification Tests
7. System Related Sidebands, 8590 E-Series, 8591C, and 8594Q
7. System Related Sidebands,
8590 E-Series, 8591C, and 8594Q
A 500 MHz CW signal is applied to the input of the spectrum analyzer.
The marker functions are used to measure the amplitude of the carrier
and the amplitude of any system related sidebands 30 kHz above and
below the carrier. System related sidebands are any internally
generated line related, power supply related or local oscillator related
sidebands.
There are no related adjustment procedures for this performance test.
Equipment Required
Signal generator
Cable, Type N, 183 cm (72 in)
Additional Equipment for Option 026
Adapter, APC 3.5 (f) to Type N (f)
Additional Equipment for 75 Ω Input
Adapter, minimum loss
Adapter, Type N (f) to BNC (m), 75 Ω
Procedure
1. Set the signal generator controls as follows:
FREQUENCY ............................................... 500 MHz
Option 026 only: Use the APC adapter to connect the cable to the
spectrum analyzer input.
Chapter 265
Page 66
Performance Verification Tests
7. System Related Sidebands, 8590 E-Series, 8591C, and 8594Q
Figure 2-8System Related Sidebands Test Setup
CAUTIONUse only 75 Ω cables, connectors, or adapters on instruments with 75 Ω
inputs, or damage to the input connector will occur.
3. Press
PRESET on the spectrum analyzer, then wait for the preset
routine to finish. Set the spectrum analyzer by pressing the
following keys:
FREQUENCY, 500, MHz
SPAN, 10, MHz
4. Set the spectrum analyzer to measure the system related sideband
above the signal as follows:
PEAK SEARCH
MKR FCTN, MK TRACK ON OFF (ON)
SPAN, 200, kHz
BW, 1, kHz
VID BW AUTO MAN, 30, Hz
5. Allow the spectrum analyzer to stabilize for approximately 1 minute,
then press the following keys:
MKR FCTN, MK TRACK ON OFF (OFF)
FREQUENCY, CF STEP AUTO MAN, 130, kHz
6. Press SGL SWP and wait for the completion of the sweep. Then press
PEAK SEARCH, MARKER ∆.
7. On the spectrum analyzer, press
66Chapter2
FREQUENCY, ⇑ (step-up key).
Page 67
Performance Verification Tests
7. System Related Sidebands, 8590 E-Series, 8591C, and 8594Q
8. Measure the system related sideband above the signal by pressing
SGL SWP on the spectrum analyzer.Wait for the completion of a new
sweep, then press
PEAK SEARCH.
9. Record the Marker-∆ Amplitude as TR Entry 1 of the performance
verification test record.
The system related sideband above the signal should be <−65 dB.
10.Set the spectrum analyzer to measure the system related sideband
below the signal by pressing the following spectrum analyzer keys:
⇓ (step-down key)
⇓ (step-down key)
11.Measure the system related sideband below the signal by pressing
SGL SWP. Wait for the completion of a new sweep, then press
PEAK SEARCH.
The system related sideband below the signal should be < −65 dB.
12.Record the Marker-∆ Amplitude as TR Entry 2 in the appropriate
performance verification test record in Chapter 3.
Performance verification test “System Related Sidebands” is now
complete.
Chapter 267
Page 68
Performance Verification Tests
8. Frequency Span Readout Accuracy, 8591E and 8591C
8. Frequency Span Readout Accuracy,
8591E and 8591C
For testing each frequency span, two synthesized sources are used to
provide two precisely-spaced signals. The spectrum analyzer marker
functions are used to measure this frequency difference and the marker
reading is compared to the specification.
There are no related adjustment procedures for this performance test.
Equipment Required
Synthesized sweeper
Synthesizer/level generator
Signal generator
Power splitter
Adapter, Type N (m) to Type N (m)
Adapter, Type N (f) to APC 3.5 (f)
Cable, Type N, 183 cm (72 in)
Cable, Type N, 152 cm (60 in)
Additional Equipment for 75 Ω Input
Adapter, minimum loss
Adapter, Type N (f) to BNC (m), 75 Ω
Procedure
This performance test consists of two parts:
Part 1: 1800 MHz Frequency Span Readout Accuracy
Part 2: 10.1 MHz to 10 kHz Frequency Span Readout Accuracy
Perform “Part 1: 1800 MHz Frequency Span Readout Accuracy” before
“Part 2: 10.1 MHz to 10 kHz Frequency Span Readout Accuracy.”
68Chapter2
Page 69
Performance Verification Tests
8. Frequency Span Readout Accuracy, 8591E and 8591C
Part 1: 1800 MHz Frequency Span Readout Accuracy
1. Connect the equipment as shown in Figure 2-9. Note that the power
splitter is used as a combiner.
Figure 2-91800 MHz Frequency Span Readout Accuracy Test Setup
CAUTIONUse only 75 Ω cables, connectors, or adapters on instruments with 75 Ω
inputs, or damage to the input connector will occur.
2. Press
PRESET on the spectrum analyzer, then wait for the preset
routine to finish.
3. Press INSTRUMENT PRESET on the synthesized sweeper and set
the controls as follows:
5. Adjust the spectrum analyzer center frequency, if necessary, to place
the lower frequency on the second vertical graticule line (one
division from the left-most graticule line).
6. On the spectrum analyzer,press
SGL SWP. Waitfor the completion of
a new sweep, then press the following keys:
PEAK SEARCH, MARKER ∆, NEXT PEAK
Chapter 269
Page 70
Performance Verification Tests
8. Frequency Span Readout Accuracy, 8591E and 8591C
The two markers should be on the signals near the second and tenth
vertical graticule lines (the first graticule line is the left-most).
7. Press
MARKER ∆, then continue pressing NEXT PK RIGHT until the
marker ∆ is on the right-most signal (1700 MHz).
8. Record the MKR ∆ frequency reading as TR Entry 1 of the
performance verification test record.
The MKR reading should be within the 1446 MHz and 1554 MHz.
Part 2: 10.1 MHz to 10 kHz Frequency Span Readout
Accuracy
Perform “Part 1: 1800 MHz Frequency Span Readout Accuracy” before
performing this procedure. An additional step to measure the frequency
span accuracy at 1 kHz is included for spectrum analyzers equipped
with Option 130.
1. Connect the equipment as shown in Figure 2-10. Note that the
power splitter is used as a combiner.
Figure 2-1010.1 MHz to 10 kHz Frequency Span Readout Accuracy Test
Setup
CAUTIONUse only 75 Ω cables, connectors, or adapters on instruments with 75 Ω
inputs, or damage to the input connector will occur.
70Chapter2
Page 71
Performance Verification Tests
8. Frequency Span Readout Accuracy, 8591E and 8591C
2. Press PRESET on the spectrum analyzer, then wait for the preset
routine to finish. Set the spectrum analyzer by pressing the
following keys:
FREQUENCY, 70, MHz
SPAN 10.1 MHz
3. Press INSTRUMENT PRESET on the synthesized sweeper, then set
the controls as follows:
a. This is not a spectrum analyzer specification; however, the 300 Hz span is tested
to ±5% to keep the narrow bandwidth accuracy and residual FM measurement
uncertainty at a minimum. If the 300 Hz span accuracy is >5% the additional
measurement uncertainty may need to be included for the bandwidth accuracy
and residual FM measurement uncertainties.
Chapter 273
Page 74
Performance Verification Tests
9. Frequency Span Readout Accuracy, 8593E, 8594E, 8595E, 8596E, and 8594Q
9. Frequency Span Readout Accuracy,
8593E, 8594E, 8595E, 8596E, and 8594Q
For testing each frequency span, two synthesized sources are used to
provide two precisely-spaced signals. The spectrum analyzer marker
functions are used to measure this frequency difference and the marker
reading is compared to the specification.
There are no related adjustment procedures for this performance test.
Equipment Required
Synthesized sweeper
Synthesizer/level generator
Signal generator
Power splitter
Adapter, Type N (m) to Type N (m)
Adapter, Type N (f) to APC 3.5 (f)
Cable, Type N, 183 cm (72 in)
Cable, Type N, 152 cm (60 in) or Adapter, APC 3.5 (f) to Type N (f)
Additional Equipment for Option 026
Adapter, APC 3.5 (f) to Type N (f)
Procedure
This performance verification test consists of two parts:
Part 1: 1800 MHz Frequency Span Readout Accuracy
Part 2: 10.1 MHz to 10 kHz Frequency Span Readout Accuracy
Perform “Part 1: 1800 MHz Frequency Span Readout Accuracy” before
“Part 2: 10.1 MHz to 10 kHz Frequency Span Readout Accuracy.”
74Chapter2
Page 75
Performance Verification Tests
9. Frequency Span Readout Accuracy, 8593E, 8594E, 8595E, 8596E, and 8594Q
Part 1: 1800 MHz Frequency Span Readout Accuracy
1. Connect the equipment as shown in Figure 2-11, Figure 2-12
for 8594E and 8594Q. Note that the power splitter is used as a
combiner.
Figure 2-111800 MHz Frequency Span Readout Accuracy Test Setup
Figure 2-12For 8594E and 8594Q Only - Frequency Span Readout Test
Setup
2. Press
PRESET on the spectrum analyzer, then wait for the preset
routine to finish. Set the spectrum analyzer by pressing the
following keys:
FREQUENCY, 900, MHz
SPAN, 1800, MHz
Chapter 275
Page 76
Performance Verification Tests
9. Frequency Span Readout Accuracy, 8593E, 8594E, 8595E, 8596E, and 8594Q
3. Press INSTRUMENT PRESET on the synthesized sweeper and set
the controls as follows:
5. Adjust the spectrum analyzer center frequency, if necessary, to place
the lower frequency on the second vertical graticule line (one
division from the left-most graticule line).
6. On the spectrum analyzer,press
a new sweep, then press
PEAK SEARCH, MARKER ∆, NEXT PEAK.
SGL SWP. Waitfor the completion of
The two markers should be on the signals near the second and tenth
vertical graticule lines (the first graticule line is the left-most).
7. Press
MARKER ∆, then continue pressing NEXT PK RIGHT. The
marker ∆ should be on the right-most signal.
8. Record the MKR ∆ frequency reading as TR Entry 1 of the
performance verification test record.
The MKR reading should be within the 1446 MHz and 1554 MHz.
Part 2: 10.1 MHz to 10 kHz Frequency Span Readout
Accuracy
Perform “Part 1: 1800 MHz Frequency Span Readout Accuracy” before
performing this procedure. An additional step to measure the frequency
span accuracy at 1 kHz is included for spectrum analyzers equipped
with Option 130.
1. Connect the equipment as shown in Figure 2-13. Note that the
power splitter is used as a combiner.
Figure 2-1310.1 MHz to 10 kHz Frequency Span Readout Accuracy Test
Setup
76Chapter2
Page 77
Performance Verification Tests
9. Frequency Span Readout Accuracy, 8593E, 8594E, 8595E, 8596E, and 8594Q
2. Press PRESET on the spectrum analyzer, then wait for the preset
routine to finish. Set the spectrum analyzer by pressing the
following keys:
FREQUENCY, 70, MHz
SPAN, 10.1, MHz
3. Press INSTRUMENT PRESET on the synthesized sweeper, then set
the controls as follows:
5. Adjust the spectrum analyzer center frequency to center the two
signals on the display.
6. On the spectrum analyzer,press
SGL SWP. Waitfor the completion of
a new sweep, then press the following keys:
PEAK SEARCH, MARKER ∆, NEXT PEAK
The two markers should be on the signals near the second and tenth
vertical graticule lines (the first graticule line is the left-most).
7. Record the MKR-∆ frequency reading in the performance verification
test record as TR Entry 2. The MKR-∆ frequency reading should be
within the limits shown.
8. Press
MKR, MARKER 1 ON OFF (OFF) on the spectrum analyzer.
9. Change to the next equipment settings listed in Table 2-9.
10.On the spectrum analyzer,press
SGL SWP. Waitfor the completion of
a new sweep, then press the following keys:
PEAK SEARCH, MARKER ∆, NEXT PEAK
11.Record the MKR-∆ frequency reading in the performance verification
test record.
12.Repeat steps 8 through 11 for the remaining spectrum analyzer span
settings listed in Table 2-9.
If you are testing a spectrum analyzer equipped with Option 130
continue with step 13.
Performance verification test “Frequency Span Readout Accuracy” is
now complete for all other spectrum analyzers.
Chapter 277
Page 78
Performance Verification Tests
9. Frequency Span Readout Accuracy, 8593E, 8594E, 8595E, 8596E, and 8594Q
Additional Steps for Option 130
13.Set the spectrum analyzer to measure the frequency span accuracy
at 1 kHz by pressing the following keys:
MKR, More 1 of 2, MARKER ALL OFF
BW, 30, Hz
If necessary, adjust the center frequency to display the two signals.
14.Change to the next spectrum analyzer span setting listed in Table
2-9. Be sure to set the synthesized sweeper CW and synthesizer/level
generator frequencies as shown in the table.
15.On the spectrum analyzer,press
SGL SWP. Waitfor the completion of
a new sweep, then press the following keys:
PEAK SEARCH, MARKER ∆, NEXT PEAK
16.Record the MKR-∆ frequency reading in the performance verification
test record.
17.Repeat steps 14 and 15 for the 300 Hz spectrum analyzer span
setting.
18.Verify that the 300 Hz span setting is within 225 Hz to 255 Hz.
Performance verification test “Frequency Span Readout Accuracy” is
now complete for spectrum analyzers equipped with Option 130.
a. This is not a spectrum analyzer specification; however, the 300 Hz span is tested
to ±5% to keep the narrow bandwidth accuracy and residual FM measurement
uncertainty at a minimum. If the 300 Hz span accuracy is >5% the additional
measurement uncertainty may need to be included for the bandwidth accuracy
and residual FM measurement uncertainties.
78Chapter2
Page 79
Performance Verification Tests
10. Residual FM, 8591E and 8591C
10. Residual FM, 8591E and 8591C
This test measures the inherent short-term instability of the spectrum
analyzer LO system. With the analyzer in zero span, a stable signal is
applied to the input and slope-detected on the linear portion of the IF
bandwidth filter skirt. Any instability in the LO transfers to the IF
signal in the mixing process. The test determines the slope of the IF
filter in Hz/dB and then measures the signal amplitude variation
caused by the residual FM. Multiplying these two values yields the
residual FM in Hz. The narrow bandwidth options use a 300 Hz span.
This span is not specified, however, it is tested in “Frequency Span
Accuracy.”
There are no related adjustment procedures for this performance test.
Equipment Required
Signal generator
Cable, Type N, 183 cm (72 in)
Additional Equipment for 75 Ω Input
Adapter, minimum loss
Adapter, Type N (f) to BNC (m), 75 Ω
Procedure
This performance test consists of two parts:
Part 1: Residual FM
Part 2: Residual FM Measurement for Option 130
Perform part 2 in addition to part 1 only if your spectrum analyzer is
equipped with Option 130. All other spectrum analyzers only perform
part 1.
Chapter 279
Page 80
Performance Verification Tests
10. Residual FM, 8591E and 8591C
Part 1: Residual FM
Determining the IF Filter Slope
1. Connect the equipment as shown in Figure 2-14.
Figure 2-14Residual FM Test Setup
CAUTIONUse only 75 Ω cables, connectors, or adapters on instruments with 75 Ω
inputs, or damage to the input connector will occur.
2. Set the signal generator controls as follows:
FREQUENCY ............................................... 500 MHz
PRESET on the spectrum analyzer, then wait for the preset
routine to finish. Set the spectrum analyzer by pressing the
following keys:
FREQUENCY, 500, MHz
SPAN, 1, MHz
75 Ω input only: Press AMPLITUDE, More 1 of 2, Amptd Units,
then
dBm.
AMPLITUDE, −9, dBm
SCALE LOG LIN (LOG), 1, dB
BW, 1, kHz
80Chapter2
Page 81
Performance Verification Tests
10. Residual FM, 8591E and 8591C
4. On the spectrum analyzer, press the following keys:
PEAK SEARCH
MKR FCTN, MK TRACK ON OFF (ON)
SPAN, 10, kHz
Wait for the AUTO ZOOM message to disappear. Press the following
spectrum analyzer keys:
MKR →, MARKER →REF LVL
MKR, MARKER 1 ON OFF (OFF)
5. On the spectrum analyzer, press the following keys:
SGL SWP
PEAK SEARCH, MARKER ∆
If you have difficulty achieving the ±0.1 dB setting, then make the
following spectrum analyzer settings:
SPAN, 5, kHz
BW, VID BW AUTO MAN, 30, Hz
6. Rotate the spectrum analyzer knob counterclockwise until the
MKR-∆ amplitude reads −1dB±0.1 dB. Press
MARKER ∆. Rotate the
knob counterclockwise until the MKR-∆ amplitude reads −4dB
±0.1 dB.
7. Divide the MKR-∆ frequency in hertz by the MKR-∆ amplitude in dB
to obtain the slope of the resolution bandwidth filter. For example, if
the MKR-∆ frequency is 1.08 kHz and the MKR-∆ amplitude is
3.92 dB, the slope would be equal to 275.5 Hz/dB. Record the result
below:
Slope _____________Hz/ dB
Chapter 281
Page 82
Performance Verification Tests
10. Residual FM, 8591E and 8591C
Measuring the Residual FM
8. On the spectrum analyzer, press MKR, More 1 of 2, MARKER ALL
OFF, PEAK SEARCH, then MARKER ∆. Rotate the knob
counterclockwise until the MKR-∆ amplitude reads −3dB±0.1 dB.
9. On the spectrum analyzer, press the following keys:
MKR, MARKER NORMAL
MKR →, MARKER →CF
SGL SWP
BW, VID BW AUTO MAN, 1, kHz
SPAN,0, Hz
SWEEP, 100, ms
SGL SWP
NOTEThe displayed trace should be about three divisions below the reference
level. If it is not, press
and use the knob to place the displayed trace about three divisions
below the reference level. Press SGL SWP.
TRIG, SWEEP CONT SGL (CONT), FREQUENCY,
10.On the spectrum analyzer, press
MARKER →PK-PK. Read the MKR-∆ amplitude, take its absolute
MKR →, MORE 1of 2,
value, and record the result as the Deviation.
Deviation _____________dB
11.Calculate the Residual FM by multiplying the Slope recorded in
step 7 by the Deviation recorded in step 10.
Record this value as TR Entry 1 in the appropriate performance
verification test record in Chapter 3. The residual FM should be less
than 250 Hz.
If you are testing a spectrum analyzer equipped with Option 130
continue with “Part 2: Residual FM Measurement for Option 130."
Performance verification test “Residual FM” is now complete for all
other spectrum analyzers.
82Chapter2
Page 83
Performance Verification Tests
10. Residual FM, 8591E and 8591C
Part 2: Residual FM Measurement for Option 130
The following procedure is an additional test for testing the residual
FM of spectrum analyzers equipped with Option 130. Perform “Part 1:
Residual FM” before performing this procedure.
Determining the IF Filter Slope
1. Press PRESET on the spectrum analyzer, then wait for the preset
routine to finish. Set the spectrum analyzer by pressing the
following keys:
FREQUENCY, 500, MHz
SPAN, 1, MHz
75 Ω input only: Press AMPLITUDE, More 1 of 3, Amptd Units,
then
dBm.
AMPLITUDE, −9, dBm
SCALE LOG LIN (LOG), 1, dB
2. On the spectrum analyzer, press the following keys:
PEAK SEARCH
MKR FCTN, MK TRACK ON OFF (ON)
SPAN, 300, Hz
3. Wait for the AUTO ZOOM message to disappear. Then press the
following spectrum analyzer keys:
MKR →, MARKER →REF LVL
MKR, MARKER 1 ON OFF (OFF)
BW, 30, Hz
SGL SWP
4. Wait for the completion of a new sweep. Then, on the spectrum
analyzer, press
PEAK SEARCH, MARKER ∆.
5. Rotate the spectrum analyzer knob counterclockwise until the
MKR-∆ amplitude reads −1dB±0.2 dB. Press MARKER ∆. Rotate the
knob counterclockwise until the MKR-∆ amplitude reads −4dB
±0.3 dB.
6. Divide the MKR-∆ frequency in hertz by the MKR-∆ amplitude in dB
to obtain the slope of the resolution bandwidth filter. For example, if
the MKR-∆ frequency is 1.08 kHz and the MKR-∆ amplitude is
3.92 dB, the slope would be equal to 275.5 Hz/dB. Record the result
below:
Slope _____________Hz/ dB
Chapter 283
Page 84
Performance Verification Tests
10. Residual FM, 8591E and 8591C
Measuring the Residual FM
7. On the spectrum analyzer, press the following keys:
TRIG, SWEEP CONT SGL (CONT)
MKR, MARKER 1 ON OFF (OFF)
SPAN, ZERO SPAN
SWEEP, SWP TIME AUTO MAN, 300, ms
8. On the spectrum analyzer, press FREQUENCY.
9. Rotate the spectrum analyzer knob until the displayed trace is
approximately 3 divisions below the reference level, then press
SGL SWEEP.
10.On the spectrum analyzer, press
MARKER →PK-PK. Read the MKR-∆ amplitude, take its absolute
MKR →, More 1 of 2,
value, and record the result as the Deviation.
Deviation _____________dB
11.Calculate the Residual FM by multiplying the Slope recorded in step
6 by the Deviation recorded in step 10. Record this value as TR
Entry 2 in the appropriate performance verification test record in
Chapter 3. The residual FM should be less than 30 Hz.
Performance verification test “Residual FM” is now complete.
84Chapter2
Page 85
Performance Verification Tests
11. Residual FM, 8593E, 8594E, 8595E, 8596E, and 8594Q
11. Residual FM, 8593E, 8594E, 8595E, 8596E,
and 8594Q
This test measures the inherent short-term instability of the spectrum
analyzer LO system. With the analyzer in zero span, a stable signal is
applied to the input and slope-detected on the linear portion of the IF
bandwidth filter skirt. Any instability in the LO transfers to the IF
signal in the mixing process. The test determines the slope of the IF
filter in Hz/dB and then measures the signal amplitude variation
caused by the residual FM. Multiplying these two values yields the
residual FM in Hz. The narrow bandwidth options use a 300 Hz span.
This span is not specified, however, it is tested in “Frequency Span
Accuracy.”
There are no related adjustment procedures for this performance test.
Equipment Required
Signal generator
Cable, Type N, 183 cm (72 in)
Additional Equipment for Option 026
Adapter, APC 3.5 (f) to Type N (f)
Procedure
This performance verification test consists of two parts:
Part 1: Residual FM
Part 2: Residual FM Measurement for Option 130
Perform part 2 in addition to part 1 only if your spectrum analyzer is
equipped with Option 130. All other spectrum analyzers only perform
part 1.
Chapter 285
Page 86
Performance Verification Tests
11. Residual FM, 8593E, 8594E, 8595E, 8596E, and 8594Q
Part 1: Residual FM
Determining the IF Filter Slope
1. Connect the equipment as shown in Figure 2-15.
Figure 2-15Residual FM Test Setup
2. Set the signal generator controls as follows:
FREQUENCY ............................................... 500 MHz
PRESET on the spectrum analyzer, then wait for the preset
routine to finish. Set the spectrum analyzer by pressing the
following keys:
FREQUENCY, 500, MHz
SPAN, 1, MHz
AMPLITUDE, −9, dBm
SCALE LOG LIN (LOG), 1, dB
BW, 1, kHz
4. On the spectrum analyzer, press the following keys:
PEAK SEARCH
MKR FCTN, MK TRACK ON OFF (ON)
SPAN, 10, kHz
5. Wait for the AUTO ZOOM message to disappear. Press the following
spectrum analyzer keys:
MKR →, MARKER →REF LVL
MKR, MARKER 1 ON OFF (OFF)
86Chapter2
Page 87
Performance Verification Tests
11. Residual FM, 8593E, 8594E, 8595E, 8596E, and 8594Q
6. On the spectrum analyzer, press the following keys:
SGL SWP
PEAK SEARCH, MARKER ∆
If you have difficulty achieving the ±0.1 dB setting, press the
following spectrum analyzer keys:
SPAN, 5, kHz
BW, VID BW AUTO MAN, 30, Hz
7. Rotate the spectrum analyzer knob counterclockwise until the
MKR-∆ amplitude reads −1dB±0.1 dB. Press
MARKER ∆. Rotate the
knob counterclockwise until the MKR-∆ amplitude reads −4dB
±0.1 dB.
8. Divide the MKR-∆ frequency in hertz by the MKR-∆ amplitude in dB
to obtain the slope of the resolution bandwidth filter. For example, if
the MKR-∆ frequency is 1.08 kHz and the MKR-∆ amplitude is
3.92 dB, the slope would be equal to 275.5 Hz/dB. Record the result
below:
Slope _____________Hz/ dB
Measuring the Residual FM
9. On the spectrum analyzer, press
PEAK SEARCH, then MARKER ∆. Rotate the knob counterclockwise
MKR, More 1 of 2, MARKER ALL OFF,
until the MKR-∆ amplitude reads −3dB±0.1 dB.
10.On the spectrum analyzer, press the following keys:
MKR, MARKER NORMAL
MKR →, MARKER →CF
SGL SWP
BW, VID BW AUTO MAN, 1, kHz
SPAN, 0, Hz
SWEEP, 100, ms
SGL SWP
NOTEThe displayed trace should be about three divisions below the reference
level. If it is not, press
TRIG, SWEEP CONT SGL (CONT), FREQUENCY,
and use the knob to place the displayed trace about three divisions
below the reference level. Press SGL SWP.
Chapter 287
Page 88
Performance Verification Tests
11. Residual FM, 8593E, 8594E, 8595E, 8596E, and 8594Q
11.On the spectrum analyzer, press MKR →, More 1of 2,
MARKER →PK-PK. Read the MKR-∆ amplitude, take its absolute
value, and record the result as the Deviation.
Deviation _____________dB
12.Calculate the Residual FM by multiplying the Slope recorded in
step 8 by the Deviation recorded in step 11. Record this value as TR
Entry 1 in the appropriate performance verification test record in
Chapter 3. The residual FM should be less than 250 Hz.
If you are testing a spectrum analyzer equipped with Option 130
continue with “Part 2: Residual FM Measurement for Option 130."
Performance verification test “Residual FM” is now complete for all
other spectrum analyzers.
Part 2: Residual FM Measurement for Option 130
The following procedure is an additional test for testing the residual
FM of spectrum analyzers equipped with Option 130. Perform “Part 1:
Residual FM” before performing this procedure.
Determining the IF Filter Slope
1. Press
PRESET on the spectrum analyzer, then wait for the preset
routine to finish. Set the spectrum analyzer by pressing the
following keys:
FREQUENCY, 500, MHz
SPAN, 1, MHz
AMPLITUDE, −9, dBm
SCALE LOG LIN (LOG), 1, dB
2. On the spectrum analyzer, press the following keys:
PEAK SEARCH
MKR FCTN, MK TRACK ON OFF (ON)
SPAN, 300, Hz
3. Wait for the AUTO ZOOM message to disappear. Then press the
following spectrum analyzer keys:
MKR →, MARKER →REF LVL
MKR, MARKER 1 ON OFF (OFF)
BW, 30, Hz
SGL SWP
4. Wait for the completion of a new sweep.
88Chapter2
Page 89
Performance Verification Tests
11. Residual FM, 8593E, 8594E, 8595E, 8596E, and 8594Q
5. On the spectrum analyzer, press PEAK SEARCH, MARKER ∆.
6. Rotate the spectrum analyzer knob counterclockwise until the
MKR-∆ amplitude reads −1dB±0.2 dB. Press MARKER ∆. Rotate the
knob counterclockwise until the MKR-∆ amplitude reads −4dB
±0.3 dB.
7. Divide the MKR-∆ frequency in hertz by the MKR-∆ amplitude in dB
to obtain the slope of the resolution bandwidth filter. For example, if
the MKR-∆ frequency is 1.08 kHz and the MKR-∆ amplitude is
3.92 dB, the slope would be equal to 275.5 Hz/dB. Record the result
below:
Slope _____________Hz/ dB
Measuring the Residual FM
8. On the spectrum analyzer, press the following keys:
TRIG, SWEEP CONT SGL (CONT)
MKR, MARKER 1 ON OFF (OFF)
SPAN, ZERO SPAN
SWEEP, SWP TIME AUTO MAN, 300, ms
FREQUENCY
9. Rotate the spectrum analyzer knob until the displayed trace is
approximately 3 divisions below the reference level. Then press
SGL SWEEP.
10.On the spectrum analyzer, press
MARKER →PK-PK. Record the absolute value of the MKR-∆
MKR →, More 1 of 2,
amplitude as the Deviation below:
Deviation ______________dB
11.Calculate the Residual FM by multiplying the Slope recorded in step
7 by the Deviation recorded in step 10. Record the Residual FM as
TR Entry 2 in the appropriate performance verification test record in
Chapter 3. The residual FM should be less than 30 Hz.
The performance verification test “Residual FM” is now complete.
Chapter 289
Page 90
Performance Verification Tests
12. Sweep Time Accuracy, 8590 E-Series, 8591C, and 8594Q
12. Sweep Time Accuracy,
8590 E-Series, 8591C, and 8594Q
This test uses a synthesizer function generator to amplitude modulate
a 500 MHz CW signal from another signal generator. The spectrum
analyzer demodulates this signal in zero span to display the response in
the time domain. The marker delta frequency function on the spectrum
analyzer is used to read out the sweep time accuracy.
If you are testing a spectrum analyzer equipped with Option 101,
perform “Fast Time Domain Sweeps” in addition to this procedure.
There are no related adjustment procedures for this performance test.
Equipment Required
Synthesizer/function generator
Signal generator
Cable, Type N, 152 cm (60 in)
Cable, BNC, 120 cm (48 in)
Additional Equipment for 75 Ω Input
Adapter, minimum loss
Adapter, Type N (f) to BNC (m), 75 Ω
90Chapter2
Page 91
12. Sweep Time Accuracy, 8590 E-Series, 8591C, and 8594Q
Procedure
If you are testing a spectrum analyzer equipped with Option 101,
perform “Fast Time Domain Sweeps,” in addition to this test.
1. Set the signal generator to output a 500 MHz, −10 dBm, CW signal.
Set the AM and FM controls to off.
75 Ω input only: Set the output to −4 dBm.
2. Set the synthesizer/function generator to output a 500 Hz, +5 dBm
triangle waveform signal.
3. Connect the equipment as shown in Figure 2-16.
Figure 2-16Sweep Time Accuracy Test Setup
Performance Verification Tests
CAUTIONUse only 75 Ω cables, connectors, or adapters on instruments with 75 Ω
inputs, or damage to the input connector will occur.
4. Press
PRESET on the spectrum analyzer, then wait for the preset
routine to finish. Set the spectrum analyzer by pressing the
following keys:
FREQUENCY, 500, MHz
SPAN, 10, MHz
PEAK SEARCH
MKR, FCTN MK TRACK ON OFF (ON)
SPAN, 50, kHz
Chapter 291
Page 92
Performance Verification Tests
12. Sweep Time Accuracy, 8590 E-Series, 8591C, and 8594Q
5. Wait for the AUTO ZOOM routine to finish. Then press the following
spectrum analyzer keys:
SPAN
ZERO SPAN
BW, 3, MHz
SWEEP, 20, ms
AMPLITUDE, SCALE LOG LIN (LIN)
6. Adjust signal amplitude for a midscreen display.
7. Set the signal generator AM switch to the AC position.
8. On the spectrum analyzer, press
trigger so that the spectrum analyzer is sweeping.
9. Press SGL SWP. After the completion of the sweep, press PEAK
SEARCH. If necessary, press NEXT PK LEFT until the marker is on the
left-most signal. This is the “marked signal.”
10.Press MARKER DELTA and press NEXT PK RIGHT 8 times so the
marker delta is on the eighth signal peak from the “marked signal.”
Record the marker ∆ reading in Table 2-10.
11.Repeat steps 9 through 10 for the remaining sweep time settings
listed in Table 2-10.
12.Record the marker ∆ reading in the appropriate performance
verification test record in Chapter 3.
Performance verification test “Sweep Time Accuracy” is now complete.
Table 2-10Sweep Time Accuracy
Spectrum Analyzer
Sweep Time Setting
20 ms500.0 Hz15.4 ms(1)_________16.6 ms
Synthesizer/Functio
n Generator
Frequency
TRIG then VIDEO. Adjust the video
Minimum
Reading
TR Entry
MKR ∆
Maximum
Reading
100 ms100.0 Hz77.0 ms(2)_________83.0 ms
1 s10.0 Hz770.0 ms(3)_________830.0 ms
10 s1.0 Hz7.7 s(4)_________8.3 s
92Chapter2
Page 93
Performance Verification Tests
13. Scale Fidelity,8590 E-Series, 8591C, and 8594Q
13. Scale Fidelity, 8590 E-Series, 8591C,
and 8594Q
A 50 MHz CW signal is applied to the INPUT 50 Ω of the analyzer
through two step attenuators. The attenuators increase the effective
amplitude range of the source. The amplitude of the source is decreased
in 10 dB steps and the analyzer marker functions are used to measure
the amplitude difference between steps. The source's internal
attenuator is used as the reference standard. The test is performed in
both log and linear amplitude scales.
The related adjustment for this performance test is “Log and Linear
Amplitude Adjustment.”
Equipment Required
Synthesizer/level generator
Attenuator, 1 dB step
Attenuator, 10 dB step
Cable, BNC, 122 cm (48 in)
Cable, BNC, 20 cm (9 in)
Adapter, Type N (m) to BNC (f)
Adapter, BNC (m) to BNC (m)
Additional Equipment for Option 026
Adapter, APC 3.5 (f) to Type N (f)
Additional Equipment for 75 Ω Input
Adapter, minimum loss
Adapter, Type N (f) to BNC (m), 75 Ω
Chapter 293
Page 94
Performance Verification Tests
13. Scale Fidelity,8590 E-Series, 8591C, and 8594Q
Procedure
Log Scale
1. Set the synthesizer/level generator controls as follows:
FREQUENCY ................................................. 50 MHz
2. Connect the equipment as shown in Figure 2-17. Set the 10 dB step
attenuator to 10 dB attenuation and the 1 dB step attenuator to
0 dB attenuation.
75 Ω input only: Set the attenuation of the 10 dB step attenuator to
0 dB. Connect the minimum loss pad to the INPUT 75 Ω using
adapters.
Figure 2-17For 8591E Only - Scale Fidelity Test Setup
Figure 2-18Scale Fidelity Test Setup
94Chapter2
Page 95
Performance Verification Tests
13. Scale Fidelity,8590 E-Series, 8591C, and 8594Q
CAUTIONUse only 75 Ω cables, connectors, or adapters on instruments with 75 Ω
inputs, or damage to the input connector will occur.
3. Press
PRESET on the spectrum analyzer, then wait for the preset
routine to finish. Set the spectrum analyzer by pressing the
following keys:
FREQUENCY, 50, MHz
SPAN, 10, MHz
75 Ω input only: Press AMPLITUDE, More 1 of 2, Amptd Units,
then
dBm.
PEAK SEARCH
MKR FCTN, MK TRACK ON OFF (ON)
SPAN, 50, kHz
Wait for the auto zoom routine to finish, then set the resolution
bandwidth and the video bandwidth by pressing the following keys:
BW
RES BW AUTO MAN, 3, kHz
VID BW AUTO MAN, 30, Hz
4. If necessary, adjust the 1 dB step attenuator attenuation until the
MKR amplitude reads between 0 dBm and −1 dBm.
5. On the synthesizer/level generator, press AMPLITUDE and use the
increment keys to adjust the amplitude until the spectrum analyzer
MKR amplitude reads 0 dBm ±0.05 dB.
It may be necessary to decrease the resolution of the amplitude
increment of the synthesizer/level generator to 0.01 dB to obtain a
MKR reading of 0 dBm ±0.05 dB.
6. On the spectrum analyzer, press
PEAK SEARCH, then MARKER ∆.
7. Set the synthesizer/level generator AMPTD INCR to 4 dB.
8. On the synthesizer/level generator, press AMPLITUDE, then
increment down to step the synthesizer/level generator to the next
lowest nominal amplitude listed in Table 2-11.
9. Record the Actual MKR ∆ amplitude reading in the performance
verification test record as indicated in Table 2-11. The MKR
amplitude should be within the limits shown.
10.Repeat steps 8 through 9 for the remaining synthesizer/level
generator Nominal Amplitudes listed in Table 2-11.
Chapter 295
Page 96
Performance Verification Tests
13. Scale Fidelity,8590 E-Series, 8591C, and 8594Q
11.For each Actual MKR ∆ reading recorded in Table 2-11, subtract the
previous Actual MKR ∆ reading. Add 4 dB to the number and record
the result as the incremental error in the performance verification
test record as indicated in Table 2-11. The incremental error should
not exceed 0.4 dB/4 dB.
If you are testing a spectrum analyzer equipped with Option 130,
continue with step 12.
The “Log Scale” portion the performance verification test “Scale
Fidelity” is now complete for all other spectrum analyzers. Proceed
to step 14, “Linear Scale.”
Additional Steps for Option 130
12.Press the following spectrum analyzer keys:
BW, RES BW AUTO MAN, 300, Hz
SPAN, 10, kHz
13.Repeat steps 4 through 11 for the narrow bandwidths. Record the
results as indicated in Table 2-12.
The scale fidelity in log mode is complete for spectrum analyzers
equipped with Option 130. Continue with step 14.
Linear Scale
14.Set the synthesizer/level generator controls as follows:
AMPTD INCR ..................................................0.05 dB
15.Set the 1 dB step attenuator to 0 dB attenuation.
16.Press
PRESET on the spectrum analyzer, then wait for the preset
routine to finish. Set the spectrum analyzer by pressing the
following keys:
AMPLITUDE, SCALE LOG LIN (LIN)
75 Ω input only:
FREQUENCY, 50, MHz
SPAN, 10, MHz
More 1 of 2, INPUT Z 50 Ω 75 Ω (50 Ω)
PEAK SEARCH
MKR FCTN, MK TRACK ON OFF (ON)
SPAN, 50, kHz
96Chapter2
Page 97
Performance Verification Tests
13. Scale Fidelity,8590 E-Series, 8591C, and 8594Q
Wait for the auto zoom routine to finish, then set the resolution
bandwidth and the video bandwidth by pressing the following keys:
BW
RES BW AUTO MAN, 3, kHz
VID BW AUTO MAN, 30, Hz
17.If necessary, adjust the 1 dB step attenuator attenuation until the
MKR reads approximately 223.6 mV. It may be necessary to
decrease the resolution of the amplitude increment of the
synthesizer/level generator to 0.01 dB to obtain a MKR reading of
223.6 mV ± 0.4 mV.
18.On the synthesizer/level generator, press AMPLITUDE, then use the
increment keys to adjust the amplitude until the spectrum analyzer
MKR amplitude reads 223.6 mV ±0.4 mV.
19.On the spectrum analyzer, press
TRACK ON OFF (OFF).
PEAK SEARCH, MKR FCTN, MK
20.Set the synthesizer/level generator amplitude increment to 3 dB.
21.On the synthesizer/level generator, press AMPLITUDE, then
increment down to step the synthesizer/level generator to the next
lowest Nominal Amplitude listed in Table 2-13.
22.Record the MKR amplitude reading in the performance verification
test record as indicated in Table 2-13. The MKR amplitude should be
within the limits shown.
23.Repeat steps 21 and 22 for the remaining synthesizer/level
generator Nominal Amplitudes listed in Table 2-13.
If you are testing a spectrum analyzer equipped with Option 130,
continue with step 24.
The “Linear Scale” portion of the performance verification test “Scale
Fidelity” is now complete for all other spectrum analyzers. Proceed
to step 26, “Log to Linear Switching.”
Additional Steps for Option 130
24.Press the following spectrum analyzer keys:
BW, RES BW AUTO MAN, 300, Hz
SPAN, 10, kHz
25.Repeat steps 17 through 22 for the narrow bandwidths. Record the
results as indicated in Table 2-14.
The “Linear Scale” portion of the performance verification test “Scale
Fidelity” is now complete for spectrum analyzers equipped with
Option 130. Proceed to step 26, “Log to Linear Switching.”
Chapter 297
Page 98
Performance Verification Tests
13. Scale Fidelity,8590 E-Series, 8591C, and 8594Q
Log to Linear Switching
26.Set the 10 dB step attenuator to 10 dB attenuation and the 1 dB
step attenuator to 0 dB attenuation.
27.Set the synthesizer controls as follows:
FREQUENCY ................................................. 50 MHz
routine to finish. Set the spectrum analyzer by pressing the
following keys:
FREQUENCY, 50, MHz
SPAN, 10, MHz
BW, 300, kHz
29.On the spectrum analyzer, press the following keys:
PEAK SEARCH
MKR →, MARKER →REF LVL
PEAK SEARCH
30.Record the peak marker reading in Log mode below.
Log Mode Amplitude Reading _____________dBm
31.Press
then press
AMPLITUDE, SCALE LOG LIN (LIN) to change the scale to linear,
More 1 of 2, Amptd Units, and dBm to set the amplitude
units to dBm.
32.Press PEAK SEARCH, then record the peak marker amplitude
reading in linear mode.
Linear Mode Amplitude Reading __________dBm
33.Subtract the Linear Mode Amplitude Reading from the Log Mode
Amplitude Reading, then record this value as the Log/Linear Error.
Log/Linear Error______________dB
34.If the Log/Linear Error is less than 0 dB, record this value as TR
Entry 73 in the performance verification test record. The absolute
value of the reading should be less than 0.25 dB. If the Log/Linear
Error is greater than 0 dB, continue with the next step.
35.On the spectrum analyzer, press the following keys:
MKR →, MARKER →REF LVL
PEAK SEARCH
98Chapter2
Page 99
Performance Verification Tests
13. Scale Fidelity,8590 E-Series, 8591C, and 8594Q
36.Record the peak marker amplitude reading in linear mode.
Linear Mode Amplitude Reading __________dBm
37.On the spectrum analyzer, press the following keys:
AMPLITUDE, SCALE LOG LIN (LOG)
PEAK SEARCH
38.Record the peak marker reading in Log mode below.
Log Mode Amplitude Reading _____________dBm
39.Subtract the Log Mode Amplitude Reading from the Linear Mode
Amplitude Reading, then record this value as the Linear/Log Error.
Linear/Log Error ____________dB
40.Record the Linear/Log Error as TR Entry 73 in the performance
verification test record. The absolute value of the reading should be
less than 0.25 dB.
If you are testing a spectrum analyzer equipped with Option 130,
continue with step 42.
The performance verification test “Scale Fidelity” is now complete
for all other spectrum analyzers.
Additional Steps for Option 130
41.Press the following spectrum analyzer keys:
AMPLITUDE, SCALE LOG LIN (LOG)
BW, RES BW AUTO MAN, 300, Hz
SPAN, 10, kHz
42.Repeat steps 29 through 39 for the narrow bandwidths. Record the
results in the performance verification test record as TR Entry 74.
Performance verification test “Scale Fidelity” is now complete for
spectrum analyzers equipped with Option 130.
Chapter 299
Page 100
Performance Verification Tests
13. Scale Fidelity,8590 E-Series, 8591C, and 8594Q
Table 2-11Cumulative and Incremental Error, Log Mode