The information contained in this document is subject to change without notice.
Hewlett-Packard makes no warranty of any kind with regard to this material, including
but not limited to, the implied warranties of merchantability and fitness for a particular
purpose. Hewlett-Packard shall not be liable for errors contained herein or for incidental or
consequential damages in connection with the furnishing, performance, or use of this
material.
Assistance
For any assistance, contact your nearest Hewlett-Packard Sales and Service Office.
General Safety Considerations
WARNINGBefore this instrument is switched on, make sure it has been properly
grounded through the protective conductor of the ac power cable to
a socket outlet provided with protective earth contact.
Any interruption of the protective (grounding) conductor, inside or
outside the instrument, or disconnection of the protective earth
terminal can result in personal injury.
WARNINGThere are many points in the instrument which can, if contacted,
cause personal injury. Be extremely careful.
Any adjustments or service procedures that require operation of the
instrument with protective covers removed should be performed
only by trained service personnel.
CAUTIONBefore this instrument is switched on, make sure its primary power circuitry
has been adapted to the voltage of the ac power source.
Failure to set the ac power input to the correct voltage could cause damage to
the instrument when the ac power cable is plugged in.
The HP 3335A Synthesizer Level Generator signal source has become obsolete because
parts used in the manufacture of this instrument are no longer available from suppliers.
To meet the needs of our customers, HP has created new performance verification tests
and adjustment tests that use alternative signal sources.
The calibration and service guides are being revised to add additional procedures that do
not use the HP 3335A. The changes include the addition of signal sources required to
replace the 3335A, changes to the test equipment setup illustrations, and changes in the
steps required to execute the procedures.
Because all of our customers will not need to replace the HP 3335A immediately, we are
releasing the changes to the calibration guide and service guides in the form of two manual
supplements. HP is making these procedure revisions available at this time in electronic
form, only. You can download this manual supplement, and add it to your current manual
as needed. You will not be required to replace your manuals just to incorporate these
changes. These changes will be incorporated in new manuals shipped in the future.
To add these changes to your existing manuals, refer to “Implementation Instructions,” on
page 3.
1-2Chapter1
Page 7
Using this Adjustment Procedure Supplement
Implementation Instructions
Implementation Instructions
Update your service guide by performing the following steps:
Step 1. After downloading this supplement, print a hard copy of the entire document.
Step 2. Remove page 2-7 and 2-8 from your service guide.
Step 3. Insert page 2-7 and 2-8 from the supplement into the service guide.
Step 4. Remove page 3-1 through 3-8 from your service guide.
Step 5. Insert page 3-1 through 2-8 from the supplement into the service guide.
Step 6. Remove the adjustment procedure chapter from the supplement, and insert it
into your service guide as Chapter 3a, “Manual Adjustment Procedures:
HP 3335A Source Not Available,” immediately after Chapter 3.
Chapter 11-3
Page 8
Using this Adjustment Procedure Supplement
Implementation Instructions
1-4Chapter1
Page 9
2Test Equipment Required for
Automated Adjustments
2-1
Page 10
Test Equipment Required for Automated Adjustments
Introduction
Introduction
This chapter contains the table of test equipment that is required for the automated
adjustments. This table is to replace table Table 2-2., “Test Equipment Required for
Automated Adjustments” in Chapter 2, “Adjustment/Diagnostic Software,” of your service
guide.
2-2Chapter2
Page 11
Test Equipment Required for Automated Adjustments
Introduction
This page intentionally left blank.
Chapter 22-3
Page 12
Test Equipment Required for Automated Adjustments
Introduction
This page intentionally left blank.
2-4Chapter2
Page 13
Test Equipment Required for Automated Adjustments
Introduction
This page intentionally left blank.
Chapter 22-5
Page 14
Test Equipment Required for Automated Adjustments
Introduction
This page intentionally left blank.
2-6Chapter2
Page 15
Test Equipment Required for Automated Adjustments
Introduction
This page intentionally left blank.
Chapter 22-7
Page 16
Test Equipment Required for Automated Adjustments
Introduction
Table 2-2. Test Equipment Required for Automated Adjustments
DescriptionHP Model Number
Controller*HP 9000 model 236 (HP 9836) or
HP 9000 model 310 or
HP 9000 model 320 or
HP 9000 model 350
Synthesizer/level generatorHP E4421B or
HP 8663A
Synthesized sweeper
Spectrum analyzer
‡
†
HP 83650A
HP 8566B (preferred) or
HP 8563E (2nd choice) or
HP 8563A or
HP 8562A
Power meterHP 438A
4.2 GHz power sensorHP 8482A
50 GHz power sensor
HP 8487A
Frequency counterHP 5342A
Digital multimeterHP 3478A
Power splitterHP 11667C
Testand Adjustment Module (TAM)HP 85629B
Cable, 50 GHz coax8120-6164
§
Cable, BNC-to-SMB (2 required)85680-60093
Cable, BNC-to-BNC (3 required)HP 10503A
Cable, 36 inch SMA8120-1578
Cable, HP-IB (8 required)HP 10833A/B
* 4 megabytes of free memory is required for the adjustment program.
†
The software menu shows the HP 8340 as the synthesized sweeper, but it drives the required
HP 83650A (programming language set to “analyzer” mode).
‡
The software menu shows the HP 8566 as the spectrum analyzer, but it will drive the other
three spectrum analyzers listed. If an HP 8566 is used, it must be an HP 8566B, not HP 8566A.
§
Must have calibration factors down to 10 MHz; a standard HP 8487A is only calibrated down to
50 MHz.
2-8Chapter2
Page 17
Test Equipment Required for Automated Adjustments
Introduction
Equipment Connections
Computer (Controller) Setup
For HP 9000 model 236 computers, setup instructions are provided in Chapter 1,
“Computer Installation,” of the BASIC Operating Manual. For HP 9000 model 310, 320, or
350 computers, setup information is provided in the Configuration Reference Manual for
the Series 300 computers.
HP-IB Cables
All HP-IB controlled test equipment should be connected to the internal HP-IB of the
controller (select code 7).
Chapter 22-9
Page 18
Test Equipment Required for Automated Adjustments
Introduction
2-10Chapter2
Page 19
3Manual Adjustment Procedures
3-1
Page 20
Manual Adjustment Procedures
Introduction
Introduction
This chapter contains information on manual adjustment procedures. Adjustments should
be performed after a repair or performance test failure, never perform adjustments as
routine maintenance. . For information on which adjustments to perform, refer to
Table 3-1.
Information on automated adjustments can be found in Chapter 2 of this service guide.
Following is a list of the automated adjustments:
NOTEBefore performing any adjustments, allow the spectrum analyzer to warm up
for at least 5 minutes.
Safety Considerations
Although this instrument has been designed in accordance with international safety
standards, this manual contains information, cautions, and warnings which must be
followed to ensure safe operation and to prevent damage to the instrument. Service and
adjustments should be performed only by qualified service personnel.
WARNING
• These servicing instructions are for use by qualified personnel
only. To avoid electrical shock, do not perform any servicing
unless you are qualified to do so.
• The opening of covers or removal of parts is likely to expose
dangerous voltages. Disconnect the product from all voltage
sources while it is being opened.
• Adjustments in this section are performed with power supplied to
the instrument and protective covers removed. There are voltages
at many points in the instrument which can, if contacted, cause
personal injury. Be extremely careful. Adjustments should be
performed only by trained service personnel.
• Power is still applied to this instrument with the
off position. Before removing or installing any assembly or
printed circuit board, remove the line-power cord.
• The power cord is connected to internal capacitors inside that
may remain live for 5 seconds after the instrument has been
disconnected from its source of supply.
• Use a nonmetallic adjustment tool whenever possible.
LINE switch in the
Which Adjustments Should Be Performed?
Table 3-1 lists the manual adjustments that should be performed when an assembly is
repaired or changed. It is important to perform the adjustments in the order indicated to
ensure that the instrument meets its specifications.
Chapter 3a, “Manual Adjustment Procedures: HP 3335A Source Not Available,” provides
instructions and procedures for making adjustments when the HP 3335A Synthesizer
Level Generator is not available. If the HP 3335A is not available, substitute those
procedures for the procedures of the same number and name found in this chapter.
Chapter 33-3
Page 22
Manual Adjustment Procedures
Introduction
Test Equipment
The equipment required for the manual adjustment procedures is listed in Table 1-5,
“Recommended Test Equipment.” Any equipment that satisfies the critical specifications
given in the table may be substituted for the preferred test equipment.
Adjustable and Factory-Selected Components
Table 3-2 lists the adjustable components by reference designation and name. For each
component, the table provides a description and lists the adjustment number.
Refer to Table 3-3. for a complete list of factory-selected components used in the
instrument along with their functions. Factory-selected components are identified with an
asterisk on the schematic diagrams.
Adjustment Tools
For adjustments requiring a nonmetallic tuning tool, use fiber tuning tool,
HP part number 8710-0033.
Two different tuning tools may be necessary for IF bandpass adjustments, depending on
the type of tuning slug used in the slug-tuned inductors. If the tuning slug requires a
slotted tuning tool, use HP part number 8710-1010. If the tuning slug requires a forked
tuning tool, use HP part number 8710-0772.
Never try to force an adjustment control. This is especially critical when tuning variable
capacitors or slug-tuned inductors. Required service accessories, with part numbers, are
listed under “Service Kit” in Chapter 1, “General Information.”
Instrument Service Position
Refer to Chapter 4 for information on removing the spectrum analyzer cover assembly and
accessing all internal assemblies.
3-4Chapter3
Page 23
Table 3-1 Related Adjustments
Manual Adjustment Procedures
Introduction
Assembly
Changed or
Repaired
Perform the Following Related Adjustments
in the Order Listed
Adjustment Number
Manual Automated*
A1A1 keyboardNo related adjustment
A1A2 RPGNo related adjustment
A2 controller16 MHz PLL adjustment15
Display adjustment2
If EEROM from old A2 controller could not be used
in new A2 or if EEROM must be replaced, also
perform the following adjustments:
LOMA adjustments4
External mixer amplitude adjustment or 3rd
amp/2nd IF align
†
136
Front end cal8
A3 interface
Display adjustment (fast zero span)
†
25
Front end cal8
A4 log amp/cal osc
Display adjustment (fast zero span)
†
25
Demodulator adjustment11
A5 IF
IF amplitude adjustment
†
DC log amplifier adjustment
IF bandpass adjustment
IF amplitude adjustment
†
†
†
410
511
39
410
A6 power supplyHigh voltage power supply adjustment1
Display adjustment2
A6A1 HV moduleHigh voltage power supply adjustment1
Display adjustment2
A7 LOMALOMA adjustments4
Front end cal (or perform the frequency response
8
performancetest in theHP 8560 E-Series Spectrum
Analyzer Calibration Guide. The adjustment must
be performed if the performance test fails.)
A8 low band
Front end cal8
mixer
A9 input
attenuator
Front end cal (or perform the frequency response
performancetest in theHP 8560 E-Series Spectrum
8
Analyzer Calibration Guide. The adjustment must
be performed if the performance test fails.)
Calibrator amplitude adjustment or CAL OUT
adjustment
†
87
External mixer bias adjustment12
Sampling oscillator adjustment612
Signal ID oscillator adjustment14
External mixer amplitude adjustment or 3rd
amp/2nd IF align
†
136
Front end cal8
A15U100 sampler
Sampling oscillator adjustment
†
612
A17 CRT driverDisplay adjustment2
A18V1 CRTDisplay adjustment2
A19 HP-IBNo related adjustment
A21 OCXO10 MHz reference adjustment (OCXO)9
*If any automated adjustment is required, you must first perform automated adjustment 1,
Initial Information.
†
Either the manual or the automated adjustment may be performed.
‡
Perform either the manual “YTO adjustment” (7) or the automated “LO frequency” and “YTO
FM coil” adjustments (2 and 3).
3-6Chapter3
Page 25
Table 3-2 Adjustable Components
Manual Adjustment Procedures
Introduction
Reference
Designator
A2R15216 MHz PLL ADJ22Adjusts the free-running frequency of the
A2R206DGTL X GAIN2Adjusts the horizontal gain in the X line
A2R209SWEEP OFFSET2Adjusts the beginning of the trace to the
A2R215DGTL Y GAIN2Adjusts the vertical gain in the Y line
A2R218VIDEO OFFSET2Adjusts the vertical position in fast-analog
A2R262STOP BLANK2Adjusts the blanking at the end of a vector on
A2R263START BLANK2Adjusts the blanking at the start of a vector on
A2R268VIDEO GAIN2Adjusts the vertical gain in fast-analog zero
A2R271SWEEP GAIN2Adjusts the end of the trace to the rightmost
A4C707FM DEMOD16Adjusts the FM demodulation for a peak
A4R445LIMITER PHASE5Adjusts Limiter Phase for peak response.
A4R531LOG AMP TOS5Minimizes error to Top of Screen.
A4R544LIN FIDELITY BOW5Minimizes Linearity Fidelity error.
A4R826CAL OSC AMPTD4Sets calibration oscillator output power
A5L300LC CTR 13Adjusts center frequency of first stage of LC
A5L301LC CTR 23Adjusts center frequency of second stage of LC
A5L700LC CTR 33Adjusts center frequency of third stage of LC
A5L702LC CTR 43Adjusts center frequency of fourth stage of LC
A5R34315 DB ATT4Adjusts the attenuation of the reference 15 dB
Adjustment NameAdjustment
Number
Description
16 MHz CPU clock
generator.
leftmost vertical graticule line in fast-analog
zero-span mode.
generator.
zero span to match the digital zero-span input.
the display.
the display.
span to match with the digital zero-span
input.
vertical-graticule line in fast-analog zero-span
mode.
response.
(nominally −35 dBm). This power is injected
into the IF during the AUTO IF ADJUST
routines.
bandwidth filter to 10.7 MHz.
bandwidth filter to 10.7 MHz.
bandwidth filter to 10.7 MHz.
bandwidth filter to 10.7 MHz.
attenuator for 15 db between minimum and
maximum attenuation.
Chapter 33-7
Page 26
Manual Adjustment Procedures
Introduction
Table 3-2 Adjustable Components (Continued)
Reference
Designator
A5T200XTAL CTR 13Adjusts center frequency of first stage of
A5T202XTAL CTR 23Adjusts center frequency of second stage of
A5T500XTAL CTR 33Adjusts center frequency of third stage of
A5T502XTAL CTR 43Adjusts center frequency of fourth stage of
A6R410HV ADJ1Adjusts the voltage between A6TP405 and
A14R426.01 GHz7Adjusts the main coil tune driver current at a
A14R76FM SPAN7Adjusts the FM span accuracy by affecting the
A14R933.2 GHz7Adjusts the main coil fixed driver current at a
A15C100SMPL MATCH6Transforms the sampler input impedance to
A15C210VCO RANGE6Adjusts the VCO tank capacitance so that 21V
A15C629SIG ID19Fine adjusts the 298 MHz SIG ID oscillator
A15U30210 MHz ADJ15Adjusts frequency of the temperature
A15R561CAL AMPTD13Adjusts amplitude of the 300 MHz calibrator
A15R926EXT BIAS ZERO17Adjusts zero bias point of external mixer bias.
A17R4Z GAIN2Adjusts maximum intensity.
A17R11CUTOFF2Adjusts intensity to turn off blanked lines.
A17R21Z FOCUS2Adjusts focus for lines of different brightness.
A17R26X FOCUS2Adjusts focus at the left and right corners of
A17R34COARSE FOCUS2Adjusts focus at the center of the display.
A17R55X GAIN2Adjusts the horizontal-deflection amplifier
Adjustment NameAdjustment
Number
Description
crystal bandwidth filter to 10.7 MHz.
crystal bandwidth filter to 10.7 MHz.
crystal bandwidth filter to 10.7 MHz.
crystal bandwidth filter to 10.7 MHz.
A6TP401 to the voltage marked on the A6A1
high voltage module.
YTO frequency of 6.01 GHz (near the upper
YTO frequency limit).
sensitivity of the FM coil driver.
YTO frequency of 3.2 GHz (near the lower
YTO frequency limit).
50 ohms over the 285 to 297.2 MHz range.
on the VCO tune line equals 298 MHz VCO
frequency.
frequency to optimize its performance.
compensated crystal oscillator (TCXO) to
10 MHz.
signal to −10.0 dBm.
the display.
gain.
3-8Chapter3
Page 27
Table 3-2 Adjustable Components (Continued)
Manual Adjustment Procedures
Introduction
Reference
Designator
A17R57X POSN2Adjusts the CRT horizontal position.
A17R75Y GAIN2Adjusts the vertical-deflection amplifier gain.
A17R77Y POSN2Adjusts the CRT vertical position.
A17R90TRACE ALIGN2Adjusts the display axis rotation.
A17R92DDD2Adjusts focus of the center of the display.
A17R93ASTIG2Adjusts for the spot roundness on the CRT
Adjustment NameAdjustment
Number
Description
display.
Table 3-3. Factory Selected Components
Reference
Designator
A5C2043Selected to optimize center frequency of LC tank that loads the
A5C2163Selected to optimize center frequency of LC tank that loads the
A5C3263Selected to optimize LC pole center frequency.
Adjustment
Number
Basis of Selection
crystal.
crystal.
A5C3273Selected to optimize LC pole center frequency.
A5C5053Selected to optimize center frequency of LC tank that loads the
crystal.
A5C5163Selected to optimize center frequency of LC tank that loads the
crystal.
A5C7173Selected to optimize LC pole center frequency.
A5C7183Selected to optimize LC pole center frequency.
Chapter 33-9
Page 28
Manual Adjustment Procedures
Introduction
3-10Chapter3
Page 29
3a Manual Adjustment Procedures:
HP 3335A Source Not Available
3a-1
Page 30
Manual Adjustment Procedures: HP 3335A Source Not Available
What You'll Find in This Chapter
What You'll Find in This Chapter
This chapter provides alternative procedures for the adjustment of the spectrum analyzer
that do not require the use of the HP 3335A Synthesizer Level Generator. The HP 3335A
has been discontinued. Bcause of the unavailability of the HP 3335A, new adjustments
procedures are required that use different signal sources. If the HP 3335A is not available,
substitute these procedures for those of the same number found in Chapter 3, “Manual
Adjustment Procedures.”
3a-2Chapter3
Page 31
Manual Adjustment Procedures: HP 3335A Source Not Available
Required Test Equipment
Required Test Equipment
The following table lists the test equipment required to execute the adjustments in this
chapter. These adjustments originally required the use of the HP 3335A Synthesizer Level
Generator.
Table 3a-1Recommended Test Equipment
InstrumentCritical Specifications for Equipment
Substitution
Sources
Synthesized Signal Generator
Cables
Cable, 50 Ω coaxial
(four required)
CableTest Cable85680-60043
Adapters
Adapter
(four required)
Frequency range: 250 kHz to 3 GHz
Frequency resolution: 1 Hz
Attenuator resolution: 0.02 dB
Level accuracy: ±0.5 dB
External 10 MHz Ref. Input
Connectors: BNC (m)
Length: ≥ 122 cm (48 in.)
Type N (m)-to-BNC (f)1250-1476
Recommended
Model
HP E4421B or
HP E4422B,
HP E4432B,
HP E4433B
HP 10503A
AdapterType N (f)-to-2.4 mm (f)HP 11903B
Adapter2.4 mm (f) to BNC (f)1250-2187
Adapter
(Option 026 only)
Adapter
(Option 026 only)
Chapter 33a-3
APC-3.5 (f) to APC-3.5 (f)5061-5311
APC-3.5 (f) to BNC-3.5 (f)1250-1200
Page 32
Manual Adjustment Procedures: HP 3335A Source Not Available
4a. IF Amplitude Adjustments
4a. IF Amplitude Adjustments
The IF amplitude adjustments consist of the cal oscillator amplitude adjustment and the
reference 15 dB attenuator adjustment.
Assembly Adjusted
A4 log amp/cal oscillator A5 IF assembly
Related Performance Tests
IF Gain Uncertainty Scale Fidelity
Description
This adjustment sets the output amplitude of the A4 log amp/cal oscillator and the
absolute amplitude of the reference 15 dB attenuator.
The output of the A4 log amp/cal oscillator is adjusted so that a −55 dBm signal applied to
the 10.7 MHz IF input on the A5 IF assembly (A5J3) causes a displayed signal of −60 dBm.
The effect of this adjustment is visible only after the
complete.
amplitude from the A4 log amp/cal oscillator.
This procedure also sets the attenuator of the reference 15 dB attenuator so that a source
amplitude change of 50 dB combined with a spectrum analyzer reference level change of
50 dB displays an amplitude difference of 50 dB.
Figure 3a-1 IF Amplitude Adjustment Setup
ADJ CURR IF STATE causes the IF gain adjustment to use the “new” output
ADJ CURR IF STATE sequence is
3a-4Chapter3
Page 33
Manual Adjustment Procedures: HP 3335A Source Not Available
NOTEThe 15 dB reference attenuator adjustment is preset at the factory and need
not be done if the entire A5 IF assembly is replaced.
Chapter 33a-5
Page 34
Manual Adjustment Procedures: HP 3335A Source Not Available
4a. IF Amplitude Adjustments
Procedure
1. Press LINE to turn the spectrum analyzer off. Remove the spectrum analyzer cover and
place the spectrum analyzer in the service position as illustrated in Figure 3a-1.
2. Disconnect W29, violet coax cable, from A5J3. Connect the test cable between A5J3 and
the RF output of the HP E4421B. Press
6. Note the marker value. Ideally it should read −60 dBm ±0.1 dB.
7. If the marker reads less than −60.1 dBm, rotate A4R826 CAL OSC AMPTD one-third
turn clockwise for every 0.1 dB less than −60 dBm. See Figure 3a-2 for the location of
A4R826. A change in the displayed amplitude will not be seen until
ADJ CURR IF STATE
is pressed.
8. If the marker reads greater than −59.9 dBm, rotate A4R826 CAL OSC AMPTD
one-third turn counter clockwise for every 0.1 dB greater than −60 dBm. See
Figure 3a-2 for the location of A4R826. A change in the displayed amplitude will not be
seen until
ADJ CURR IF STATE is pressed.
NOTEIf A4R826 has inadequate range, refer to “Inadequate CAL OSC AMPTD
Range” in Chapter 9.
9. On the spectrum analyzer, press
ADJ CURR IF STATE. After allowing the analyzer time
to complete the adjustments, the displayed amplitude and marker reading should
change.
10.Repeat step 7 and step 9 until the marker reads −60 dBm ±0.1 dB.
11.Disconnect the test cable from A5J3 and reconnect W29 to A5J3.
3a-6Chapter3
Page 35
Manual Adjustment Procedures: HP 3335A Source Not Available
4a. IF Amplitude Adjustments
A5 Reference Attenuator Adjustment
1. Set the spectrum analyzer reference level to −60 dBm. If markers are displayed, press
MKR and MARKERS OFF.
2. Set the HP E4421B
AMPLITUDE to −60 dBm.
3. Connect a BNC cable between the RF output of the HP E4421B and the spectrum
analyzer INPUT 50Ω.
4. On the spectrum analyzer, press
CAL and REF LVL ADJ. Use the front panel knob or step
keys to place the peak of the displayed signal 3 dB to 5 dB below the reference level.
5. On the spectrum analyzer, press
PEAK SEARCH and MARKER DELTA. Set the spectrum
analyzer reference level to −10 dBm.
6. Change the HP 4421B
7. On the spectrum analyzer, press
Amplitude to −10 dBm.
CAL.
8. Note the ∆MKR amplitude. Ideally, it should read 50.00 dB ±0.1 dB.
9. If the ∆MKR amplitude is less than 49.9 dB, rotate A5R343 15 dB ATTEN one-half turn
counterclockwise for each 0.1 dB less than 50.00 dB. Do not adjust A5R343 more than
five turns before continuing with the next step.
10.If the ∆MKR amplitude is greater than 50.1 dB, rotate A5R343 15 dB ATTEN one-half
turn clockwise for each 0.1 dB greater than 50.00 dB. Do not adjust A5R343 more than
five turns before continuing with the next step.
11.On the spectrum analyzer, press
ADJ CURR IF STATE. Note the ∆MKR amplitude
reading.
12.Repeat step 1 through step 11 until the ∆MKR amplitude reading is 50.00 dB ±0.1 dB.
Chapter 33a-7
Page 36
Manual Adjustment Procedures: HP 3335A Source Not Available
4a. IF Amplitude Adjustments
A5 Adjustment Verification
1. On the spectrum analyzer, disconnect W29 from A5J3. Connect the test cable between
A5J3 and the RF output of the HP E4421B.
2. Set the spectrum analyzer reference level to −10 dBm.
3. Set the HP E4421B
4. On the spectrum analyzer, press
Amplitude to −5 dBm.
MKR and MARKER NORMAL.
5. The MARKER amplitude should read −10 dBm ±0.13 dB. If the reading is outside of
this range, repeat step 4 of “Procedure,” on page 6 through “A5 Reference Attenuator
Adjustment” step 12.
6. On the spectrum analyzer, reconnect W29 to A5J3. Press
7. Connect a BNC cable between the HP 8563E CAL OUTPUT and INPUT 50Ω.
8. On the spectrum analyzer, press
MKR CAL and REF LVL ADJ.
9. Use the knob or step keys to adjust the REF LEVEL CAL setting until the MKR reads
−10.00 dBm ±0.1 dB.
10.On the spectrum analyzer, press
STORE REF LVL.
3a-8Chapter3
Page 37
Manual Adjustment Procedures: HP 3335A Source Not Available
5a. DC Log Amplifier Adjustments
5a. DC Log Amplifier Adjustments
There are three DC log adjustments; limiter phase, linear fidelity, and log fidelity.
Assembly Adjusted
A4 log amp/cal oscillator
Related Performance Tests
IF Gain Uncertainty Scale Fidelity
Description
These three adjustment need only be done under the following conditions:
Limiter phase Only if a repair is made to blocks F, G, H, I, or J.
Linear fidelity Only if a repair is made to blocks C, D, F, G, H, I, J, K, O, IF gain accuracy,
RBW switching, or log fidelity.
Log fidelityOnly if a repair is made to blocks D, F, H, K, IF gain accuracy, RBW
switching, or log fidelity.
If multiple adjustments are required they should be done in the following order:
1. Limiter Phase
2. Linear Fidelity
3. Log Fidelity
Figure 3a-3 DC Log Adjustment Setup
Chapter 33a-9
Page 38
Manual Adjustment Procedures: HP 3335A Source Not Available
NOTEAdjustments should be made with all of the shields on and only after allowing
at least a 20 minute warmup.
3a-10Chapter3
Page 39
Manual Adjustment Procedures: HP 3335A Source Not Available
5a. DC Log Amplifier Adjustments
A4 Limiter Phase Adjustment
1. Press LINE to turn the spectrum analyzer off. Remove the spectrum analyzer cover and
place the spectrum analyzer in the service position as illustrated in Figure 3a-3. See
Figure 3a-4 for adjustment location.
2. Connect the HP E4421B RF output to the spectrum analyzer 50Ω input. Press
turn the spectrum analyzer on.
5. On the spectrum analyzer, press
complete adjustments then press
CAL and ADJ CURR IF STATE, wait for the analyzer to
MKR.
LINE to
6. Adjust A4R445 for maximum on-screen amplitude. Refer to Figure 3a-4 for the location
of A4R445.
Chapter 33a-11
Page 40
Manual Adjustment Procedures: HP 3335A Source Not Available
5a. DC Log Amplifier Adjustments
A4 Linear Fidelity Adjustment
1. Press LINE to turn the spectrum analyzer off. Remove the spectrum analyzer cover and
place the spectrum analyzer in the service position as illustrated in Figure 3a-3. See
Figure 3a-4 for adjustment location.
2. Connect the HP E4421B RF output to the spectrum analyzer 50Ω input. Press
turn the spectrum analyzer on.
8. If the delta marker amplitude reads −40 dB ±2 dB, no adjustment is necessary.
LINE to
9. If the signal is lower on the screen than expected (delta marker amplitude reads less
than −42dB) then adjust A4R544 (see Figure 3a-4) for an even lower level and press
CAL and ADJ CURR IF STATE. Allow sufficient time for the analyzer to complete the
adjustment.
10.If the signal is higher on the screen than expected (delta marker amplitude reads
greater than −38 dB) then adjust A4R544 for an even higher level signal and press
CAL
and ADJ CURR IF STATE. Allow sufficient time for the analyzer to complete the
adjustment.
11.Repeat step 7 through step 10.
3a-12Chapter3
Page 41
Manual Adjustment Procedures: HP 3335A Source Not Available
5a. DC Log Amplifier Adjustments
A4 LOG Fidelity Adjustment
1. Press LINE to turn the spectrum analyzer off. Remove the spectrum analyzer cover and
place the spectrum analyzer in the service position as illustrated in Figure 3a-3. See
Figure 3a-4 for adjustment location.
2. Connect the HP E4421B RF output to the spectrum analyzer 50Ω input. Press
turn the spectrum analyzer on.
9. If the error is less than ±0.2 dB, no adjustment is necessary.
10.Set the HP E4421B power to −10 dBm.
11.Adjust A4R531 (see Figure 3a-4) to read two times the error. For example, if the
calculated error is +0.75 dB, adjust A4R531 for a delta marker amplitude reading of
+1.5 dB. Press
12.Repeat step 7 through step 11.
CAL and ADJ CURR IF STATE.
Chapter 33a-13
Page 42
Manual Adjustment Procedures: HP 3335A Source Not Available
5a. DC Log Amplifier Adjustments
3a-14Chapter3
Page 43
Index
A
A4 limiter phase adjustment
3a-11
A4 linear fidelity adjustment
3a-12
A4 log fidelity adjustment
adjustment
adjustable components table
3-5
factory selected components
related adjustments table
safety considerations
tools
, 3-4
adjustment procedures
cal oscillator amplitude
DC log amplifiers
IF amplitude
adjustments
recommended test equipment
3a-3
, 3a-4
, 3a-13
, 3-3
, 3-2
, 3a-4
, 3a-9
,
,
,
, 3-5
, 3-4
,
C
components
adjustable components table
3-5
D
DC log amplifer adjustments
3a-9
I
IF amplitude adjustments
implementation instructions
introduction
L
limiter phase adjustment
linear fidelity adjustment
log fidelity adjustment
T
test equipment list
test equipment, automated
adjustments
tools
adjustment
U
using the supplement
, 1-2
, 2-18
, 3-4
, 3a-4
, 3a-11
, 3a-12
, 3a-13
, 3a-3
, 1-1
,
,
, 1-3
IndexIndex-1
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.