HP 8565e schematic

Page 1
HP 8564E and 8565E Spectrum Analyzers
Service Guide Supplement
Using Performance Tests When an
HP 3335A Source Is Not Available
Manufacturing Part Number: 08564-srvsup
Printed in USA
October 1999
© Copyright 1999 Hewlett-Packard Company
Page 2
Notice
The information contained in this document is subject to change without notice. Hewlett-Packard makes no warranty of any kind with regard to this material, including
but not limited to, the implied warranties of merchantability and fitness for a particular purpose. Hewlett-Packard shall not be liable for errors contained herein or for incidental or consequential damages in connection with the furnishing, performance, or use of this material.
Assistance
For any assistance, contact your nearest Hewlett-Packard Sales and Service Office.
General Safety Considerations
WARNING Before this instrument is switched on, make sure it has been properly
grounded through the protective conductor of the ac power cable to a socket outlet provided with protective earth contact.
Any interruption of the protective (grounding) conductor, inside or outside the instrument, or disconnection of the protective earth terminal can result in personal injury.
WARNING There are many points in the instrument which can, if contacted,
cause personal injury. Be extremely careful. Any adjustments or service procedures that require operation of the
instrument with protective covers removed should be performed only by trained service personnel.
CAUTION Before this instrument is switched on, make sure its primary power circuitry
has been adapted to the voltage of the ac power source. Failure to set the ac power input to the correct voltage could cause damage to
the instrument when the ac power cable is plugged in.
Page 3

Contents

1. Using this Adjustment Procedure Supplement
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-2
Implementation Instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-3
2. Test Equipment Required for Automated Adjustments
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-12
3. Manual Adjustment Procedures
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-2
3a. Manual Adjustment Procedures: HP 3335A Source Not Available
What You'll Find in This Chapter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3a-2
Required Test Equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3a-3
4a. IF Amplitude Adjustments . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3a-4
5a. DC Log Amplifier Adjustments . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3a-9
Contents-1
Page 4
Contents
Contents-2
Page 5
1 Using this Adjustment Procedure
Supplement
1-1
Page 6
Using this Adjustment Procedure Supplement

Introduction

Introduction
The HP 3335A Synthesizer Level Generator signal source has become obsolete because parts used in the manufacture of this instrument are no longer available from suppliers. To meet the needs of our customers, HP has created new performance verification tests and adjustment tests that use alternative signal sources.
The calibration and service guides are being revised to add additional procedures that do not use the HP 3335A. The changes include the addition of signal sources required to replace the 3335A, changes to the test equipment setup illustrations, and changes in the steps required to execute the procedures.
Because all of our customers will not need to replace the HP 3335A immediately, we are releasing the changes to the calibration guide and service guides in the form of two manual supplements. HP is making these procedure revisions available at this time in electronic form, only. You can download this manual supplement, and add it to your current manual as needed. You will not be required to replace your manuals just to incorporate these changes. These changes will be incorporated in new manuals shipped in the future.
To add these changes to your existing manuals, refer to “Implementation Instructions,” on
page 3.
1-2 Chapter1
Page 7
Using this Adjustment Procedure Supplement

Implementation Instructions

Implementation Instructions
Update your service guide by performing the following steps:
Step 1. After downloading this supplement, print a hard copy of the entire document. Step 2. Remove page 2-7 and 2-8 from your service guide. Step 3. Insert page 2-7 and 2-8 from the supplement into the service guide. Step 4. Remove page 3-1 through 3-8 from your service guide. Step 5. Insert page 3-1 through 2-8 from the supplement into the service guide. Step 6. Remove the adjustment procedure chapter from the supplement, and insert it
into your service guide as Chapter 3a, “Manual Adjustment Procedures:
HP 3335A Source Not Available,” immediately after Chapter 3.
Chapter 1 1-3
Page 8
Using this Adjustment Procedure Supplement
Implementation Instructions
1-4 Chapter1
Page 9
2 Test Equipment Required for
Automated Adjustments
2-1
Page 10
Test Equipment Required for Automated Adjustments

Introduction

Introduction
This chapter contains the table of test equipment that is required for the automated adjustments. This table is to replace table Table 2-2., “Test Equipment Required for
Automated Adjustments” in Chapter 2, “Adjustment/Diagnostic Software,” of your service
guide.
2-2 Chapter2
Page 11
Test Equipment Required for Automated Adjustments
Introduction
This page intentionally left blank.
Chapter 2 2-3
Page 12
Test Equipment Required for Automated Adjustments
Introduction
This page intentionally left blank.
2-4 Chapter2
Page 13
Test Equipment Required for Automated Adjustments
Introduction
This page intentionally left blank.
Chapter 2 2-5
Page 14
Test Equipment Required for Automated Adjustments
Introduction
This page intentionally left blank.
2-6 Chapter2
Page 15
Test Equipment Required for Automated Adjustments
Introduction
This page intentionally left blank.
Chapter 2 2-7
Page 16
Test Equipment Required for Automated Adjustments
Introduction
Table 2-2. Test Equipment Required for Automated Adjustments
Description HP Model Number
Controller* HP 9000 model 236 (HP 9836) or
HP 9000 model 310 or HP 9000 model 320 or HP 9000 model 350
Synthesizer/level generator HP E4421B or
HP 8663A
Synthesized sweeper
Spectrum analyzer
HP 83650A
HP 8566B (preferred) or HP 8563E (2nd choice) or HP 8563A or HP 8562A
Power meter HP 438A
4.2 GHz power sensor HP 8482A
50 GHz power sensor
HP 8487A
Frequency counter HP 5342A
Digital multimeter HP 3478A
Power splitter HP 11667C
Testand Adjustment Module (TAM) HP 85629B
Cable, 50 GHz coax 8120-6164
§
Cable, BNC-to-SMB (2 required) 85680-60093
Cable, BNC-to-BNC (3 required) HP 10503A
Cable, 36 inch SMA 8120-1578
Cable, HP-IB (8 required) HP 10833A/B
* 4 megabytes of free memory is required for the adjustment program.
The software menu shows the HP 8340 as the synthesized sweeper, but it drives the required
HP 83650A (programming language set to “analyzer” mode).
The software menu shows the HP 8566 as the spectrum analyzer, but it will drive the other
three spectrum analyzers listed. If an HP 8566 is used, it must be an HP 8566B, not HP 8566A.
§
Must have calibration factors down to 10 MHz; a standard HP 8487A is only calibrated down to
50 MHz.
2-8 Chapter2
Page 17
Test Equipment Required for Automated Adjustments
Introduction

Equipment Connections

Computer (Controller) Setup
For HP 9000 model 236 computers, setup instructions are provided in Chapter 1, “Computer Installation,” of the BASIC Operating Manual. For HP 9000 model 310, 320, or 350 computers, setup information is provided in the Configuration Reference Manual for the Series 300 computers.
HP-IB Cables
All HP-IB controlled test equipment should be connected to the internal HP-IB of the controller (select code 7).
Chapter 2 2-9
Page 18
Test Equipment Required for Automated Adjustments
Introduction
2-10 Chapter2
Page 19

3 Manual Adjustment Procedures

3-1
Page 20
Manual Adjustment Procedures

Introduction

Introduction
This chapter contains information on manual adjustment procedures. Adjustments should be performed after a repair or performance test failure, never perform adjustments as routine maintenance. . For information on which adjustments to perform, refer to
Table 3-1.
Information on automated adjustments can be found in Chapter 2 of this service guide. Following is a list of the automated adjustments:
Automated Adjustments
1. Initial Information
2. LO Frequency
3. YTO FM Coil
4. LOMA Adjustments
5. Fast Zero Span
6. 3rd Amp/2nd IF Align
7. Cal Out Adjustment
8. Front End Cal
9. IF Bandpass Poles
10. IF Amplitude
11. DC Log Adjustments
12. Sampling Oscillator
Manual Adjustments
1. High Voltage Power Supply Adjustment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-9
2. Display Adjustment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-11
3. IF Bandpass Adjustment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-17
4. IF Amplitude Adjustments . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-22
5. DC Log Amplifier Adjustments . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-26
6. Sampling Oscillator Adjustment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-30
7. YTO Adjustment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-33
8. Calibrator Amplitude Adjustment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-36
9. 10 MHz Reference Adjustment — OCXO . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-37
10. 10 MHz Reference Adjustment — TCXO (Option 103) . . . . . . . . . . . . . . . . . . . . . . . 3-40
11. Demodulator Adjustment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-42
12. External Mixer Bias Adjustment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-45
13. External Mixer Amplitude Adjustment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-46
14. Signal ID Oscillator Adjustment (serial prefix 3517A and below) . . . . . . . . . . . . . . 3-49
15. 16 MHz PLL Adjustment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-52
16. 600 MHz Reference Adjustment (serial prefix 3406A and above) . . . . . . . . . . . . . . 3-55
3-2 Chapter3
Page 21
Manual Adjustment Procedures
Introduction
NOTE Before performing any adjustments, allow the spectrum analyzer to warm up
for at least 5 minutes.

Safety Considerations

Although this instrument has been designed in accordance with international safety standards, this manual contains information, cautions, and warnings which must be followed to ensure safe operation and to prevent damage to the instrument. Service and adjustments should be performed only by qualified service personnel.
WARNING
• These servicing instructions are for use by qualified personnel only. To avoid electrical shock, do not perform any servicing unless you are qualified to do so.
• The opening of covers or removal of parts is likely to expose dangerous voltages. Disconnect the product from all voltage sources while it is being opened.
• Adjustments in this section are performed with power supplied to the instrument and protective covers removed. There are voltages at many points in the instrument which can, if contacted, cause personal injury. Be extremely careful. Adjustments should be performed only by trained service personnel.
• Power is still applied to this instrument with the off position. Before removing or installing any assembly or printed circuit board, remove the line-power cord.
• The power cord is connected to internal capacitors inside that may remain live for 5 seconds after the instrument has been disconnected from its source of supply.
• Use a nonmetallic adjustment tool whenever possible.
LINE switch in the

Which Adjustments Should Be Performed?

Table 3-1 lists the manual adjustments that should be performed when an assembly is
repaired or changed. It is important to perform the adjustments in the order indicated to ensure that the instrument meets its specifications.
Chapter 3a, “Manual Adjustment Procedures: HP 3335A Source Not Available,” provides
instructions and procedures for making adjustments when the HP 3335A Synthesizer Level Generator is not available. If the HP 3335A is not available, substitute those procedures for the procedures of the same number and name found in this chapter.
Chapter 3 3-3
Page 22
Manual Adjustment Procedures
Introduction

Test Equipment

The equipment required for the manual adjustment procedures is listed in Table 1-5, “Recommended Test Equipment.” Any equipment that satisfies the critical specifications given in the table may be substituted for the preferred test equipment.

Adjustable and Factory-Selected Components

Table 3-2 lists the adjustable components by reference designation and name. For each
component, the table provides a description and lists the adjustment number. Refer to Table 3-3. for a complete list of factory-selected components used in the
instrument along with their functions. Factory-selected components are identified with an asterisk on the schematic diagrams.

Adjustment Tools

For adjustments requiring a nonmetallic tuning tool, use fiber tuning tool, HP part number 8710-0033.
Two different tuning tools may be necessary for IF bandpass adjustments, depending on the type of tuning slug used in the slug-tuned inductors. If the tuning slug requires a slotted tuning tool, use HP part number 8710-1010. If the tuning slug requires a forked tuning tool, use HP part number 8710-0772.
Never try to force an adjustment control. This is especially critical when tuning variable capacitors or slug-tuned inductors. Required service accessories, with part numbers, are listed under “Service Kit” in Chapter 1, “General Information.”

Instrument Service Position

Refer to Chapter 4 for information on removing the spectrum analyzer cover assembly and accessing all internal assemblies.
3-4 Chapter3
Page 23
Table 3-1 Related Adjustments
Manual Adjustment Procedures
Introduction
Assembly
Changed or
Repaired
Perform the Following Related Adjustments
in the Order Listed
Adjustment Number
Manual Automated*
A1A1 keyboard No related adjustment A1A2 RPG No related adjustment A2 controller 16 MHz PLL adjustment 15
Display adjustment 2 If EEROM from old A2 controller could not be used
in new A2 or if EEROM must be replaced, also perform the following adjustments:
LOMA adjustments 4 External mixer amplitude adjustment or 3rd
amp/2nd IF align
13 6
Front end cal 8
A3 interface
Display adjustment (fast zero span)
25
Front end cal 8
A4 log amp/cal osc
Display adjustment (fast zero span)
25
Demodulator adjustment 11
A5 IF
IF amplitude adjustment
DC log amplifier adjustment IF bandpass adjustment IF amplitude adjustment
410 511 39 410
A6 power supply High voltage power supply adjustment 1
Display adjustment 2
A6A1 HV module High voltage power supply adjustment 1
Display adjustment 2
A7 LOMA LOMA adjustments 4
Front end cal (or perform the frequency response
performancetest in theHP 8560 E-Series Spectrum Analyzer Calibration Guide. The adjustment must be performed if the performance test fails.)
A8 low band
Front end cal 8
mixer A9 input
attenuator
Front end cal (or perform the frequency response
performancetest in theHP 8560 E-Series Spectrum
Analyzer Calibration Guide. The adjustment must be performed if the performance test fails.)
Chapter 3 3-5
Page 24
Manual Adjustment Procedures
Introduction
Table 3-1 Related Adjustments (Continued)
Assembly
Changed or
Repaired
A10/A12
Perform the Following Related Adjustments
Adjustment Number
in the Order Listed
Manual Automated*
Front end cal 8
RYTHM/SBTX
A11 YTO
YTO adjustment LO frequency YTO FM coil
A13 2nd converter Front end cal 8 A14 frequency
Display adjustment (fast zero span)
25
control
YTO adjustment LO frequency YTO FM cal
Front end cal 8
A15 RF 10 MHz reference adjustment (TCXO, Option 103)
or CAL OUT adjustment
15 7
Calibrator amplitude adjustment or CAL OUT adjustment
87
External mixer bias adjustment 12 Sampling oscillator adjustment 6 12 Signal ID oscillator adjustment 14 External mixer amplitude adjustment or 3rd
amp/2nd IF align
13 6
Front end cal 8
A15U100 sampler
Sampling oscillator adjustment
612
A17 CRT driver Display adjustment 2 A18V1 CRT Display adjustment 2 A19 HP-IB No related adjustment A21 OCXO 10 MHz reference adjustment (OCXO) 9 *If any automated adjustment is required, you must first perform automated adjustment 1,
Initial Information.
Either the manual or the automated adjustment may be performed.
Perform either the manual “YTO adjustment” (7) or the automated “LO frequency” and “YTO
FM coil” adjustments (2 and 3).
3-6 Chapter3
Page 25
Table 3-2 Adjustable Components
Manual Adjustment Procedures
Introduction
Reference
Designator
A2R152 16 MHz PLL ADJ 22 Adjusts the free-running frequency of the
A2R206 DGTL X GAIN 2 Adjusts the horizontal gain in the X line
A2R209 SWEEP OFFSET 2 Adjusts the beginning of the trace to the
A2R215 DGTL Y GAIN 2 Adjusts the vertical gain in the Y line
A2R218 VIDEO OFFSET 2 Adjusts the vertical position in fast-analog
A2R262 STOP BLANK 2 Adjusts the blanking at the end of a vector on
A2R263 START BLANK 2 Adjusts the blanking at the start of a vector on
A2R268 VIDEO GAIN 2 Adjusts the vertical gain in fast-analog zero
A2R271 SWEEP GAIN 2 Adjusts the end of the trace to the rightmost
A4C707 FM DEMOD 16 Adjusts the FM demodulation for a peak
A4R445 LIMITER PHASE 5 Adjusts Limiter Phase for peak response. A4R531 LOG AMP TOS 5 Minimizes error to Top of Screen. A4R544 LIN FIDELITY BOW 5 Minimizes Linearity Fidelity error. A4R826 CAL OSC AMPTD 4 Sets calibration oscillator output power
A5L300 LC CTR 1 3 Adjusts center frequency of first stage of LC
A5L301 LC CTR 2 3 Adjusts center frequency of second stage of LC
A5L700 LC CTR 3 3 Adjusts center frequency of third stage of LC
A5L702 LC CTR 4 3 Adjusts center frequency of fourth stage of LC
A5R343 15 DB ATT 4 Adjusts the attenuation of the reference 15 dB
Adjustment Name Adjustment
Number
Description
16 MHz CPU clock
generator.
leftmost vertical graticule line in fast-analog zero-span mode.
generator.
zero span to match the digital zero-span input.
the display.
the display.
span to match with the digital zero-span input.
vertical-graticule line in fast-analog zero-span mode.
response.
(nominally 35 dBm). This power is injected into the IF during the AUTO IF ADJUST routines.
bandwidth filter to 10.7 MHz.
bandwidth filter to 10.7 MHz.
bandwidth filter to 10.7 MHz.
bandwidth filter to 10.7 MHz.
attenuator for 15 db between minimum and maximum attenuation.
Chapter 3 3-7
Page 26
Manual Adjustment Procedures
Introduction
Table 3-2 Adjustable Components (Continued)
Reference
Designator
A5T200 XTAL CTR 1 3 Adjusts center frequency of first stage of
A5T202 XTAL CTR 2 3 Adjusts center frequency of second stage of
A5T500 XTAL CTR 3 3 Adjusts center frequency of third stage of
A5T502 XTAL CTR 4 3 Adjusts center frequency of fourth stage of
A6R410 HV ADJ 1 Adjusts the voltage between A6TP405 and
A14R42 6.01 GHz 7 Adjusts the main coil tune driver current at a
A14R76 FM SPAN 7 Adjusts the FM span accuracy by affecting the
A14R93 3.2 GHz 7 Adjusts the main coil fixed driver current at a
A15C100 SMPL MATCH 6 Transforms the sampler input impedance to
A15C210 VCO RANGE 6 Adjusts the VCO tank capacitance so that 21V
A15C629 SIG ID 19 Fine adjusts the 298 MHz SIG ID oscillator
A15U302 10 MHz ADJ 15 Adjusts frequency of the temperature
A15R561 CAL AMPTD 13 Adjusts amplitude of the 300 MHz calibrator
A15R926 EXT BIAS ZERO 17 Adjusts zero bias point of external mixer bias. A17R4 Z GAIN 2 Adjusts maximum intensity. A17R11 CUTOFF 2 Adjusts intensity to turn off blanked lines. A17R21 Z FOCUS 2 Adjusts focus for lines of different brightness. A17R26 X FOCUS 2 Adjusts focus at the left and right corners of
A17R34 COARSE FOCUS 2 Adjusts focus at the center of the display. A17R55 X GAIN 2 Adjusts the horizontal-deflection amplifier
Adjustment Name Adjustment
Number
Description
crystal bandwidth filter to 10.7 MHz.
crystal bandwidth filter to 10.7 MHz.
crystal bandwidth filter to 10.7 MHz.
crystal bandwidth filter to 10.7 MHz.
A6TP401 to the voltage marked on the A6A1 high voltage module.
YTO frequency of 6.01 GHz (near the upper YTO frequency limit).
sensitivity of the FM coil driver.
YTO frequency of 3.2 GHz (near the lower YTO frequency limit).
50 ohms over the 285 to 297.2 MHz range.
on the VCO tune line equals 298 MHz VCO frequency.
frequency to optimize its performance.
compensated crystal oscillator (TCXO) to 10 MHz.
signal to 10.0 dBm.
the display.
gain.
3-8 Chapter3
Page 27
Table 3-2 Adjustable Components (Continued)
Manual Adjustment Procedures
Introduction
Reference
Designator
A17R57 X POSN 2 Adjusts the CRT horizontal position. A17R75 Y GAIN 2 Adjusts the vertical-deflection amplifier gain. A17R77 Y POSN 2 Adjusts the CRT vertical position. A17R90 TRACE ALIGN 2 Adjusts the display axis rotation. A17R92 DDD 2 Adjusts focus of the center of the display. A17R93 ASTIG 2 Adjusts for the spot roundness on the CRT
Adjustment Name Adjustment
Number
Description
display.
Table 3-3. Factory Selected Components
Reference
Designator
A5C204 3 Selected to optimize center frequency of LC tank that loads the
A5C216 3 Selected to optimize center frequency of LC tank that loads the
A5C326 3 Selected to optimize LC pole center frequency.
Adjustment
Number
Basis of Selection
crystal.
crystal.
A5C327 3 Selected to optimize LC pole center frequency. A5C505 3 Selected to optimize center frequency of LC tank that loads the
crystal.
A5C516 3 Selected to optimize center frequency of LC tank that loads the
crystal. A5C717 3 Selected to optimize LC pole center frequency. A5C718 3 Selected to optimize LC pole center frequency.
Chapter 3 3-9
Page 28
Manual Adjustment Procedures
Introduction
3-10 Chapter3
Page 29
3a Manual Adjustment Procedures:
HP 3335A Source Not Available
3a-1
Page 30
Manual Adjustment Procedures: HP 3335A Source Not Available

What You'll Find in This Chapter

What You'll Find in This Chapter
This chapter provides alternative procedures for the adjustment of the spectrum analyzer that do not require the use of the HP 3335A Synthesizer Level Generator. The HP 3335A has been discontinued. Bcause of the unavailability of the HP 3335A, new adjustments procedures are required that use different signal sources. If the HP 3335A is not available, substitute these procedures for those of the same number found in Chapter 3, “Manual
Adjustment Procedures.”
3a-2 Chapter3
Page 31
Manual Adjustment Procedures: HP 3335A Source Not Available

Required Test Equipment

Required Test Equipment
The following table lists the test equipment required to execute the adjustments in this chapter. These adjustments originally required the use of the HP 3335A Synthesizer Level Generator.
Table 3a-1 Recommended Test Equipment
Instrument Critical Specifications for Equipment
Substitution
Sources
Synthesized Signal Generator
Cables
Cable, 50 coaxial
(four required)
Cable Test Cable 85680-60043
Adapters
Adapter
(four required)
Frequency range: 250 kHz to 3 GHz Frequency resolution: 1 Hz Attenuator resolution: 0.02 dB Level accuracy: ±0.5 dB External 10 MHz Ref. Input
Connectors: BNC (m) Length: 122 cm (48 in.)
Type N (m)-to-BNC (f) 1250-1476
Recommended
Model
HP E4421B or HP E4422B, HP E4432B, HP E4433B
HP 10503A
Adapter Type N (f)-to-2.4 mm (f) HP 11903B Adapter 2.4 mm (f) to BNC (f) 1250-2187 Adapter
(Option 026 only)
Adapter
(Option 026 only)
Chapter 3 3a-3
APC-3.5 (f) to APC-3.5 (f) 5061-5311
APC-3.5 (f) to BNC-3.5 (f) 1250-1200
Page 32
Manual Adjustment Procedures: HP 3335A Source Not Available

4a. IF Amplitude Adjustments

4a. IF Amplitude Adjustments
The IF amplitude adjustments consist of the cal oscillator amplitude adjustment and the reference 15 dB attenuator adjustment.

Assembly Adjusted

A4 log amp/cal oscillator A5 IF assembly

Related Performance Tests

IF Gain Uncertainty Scale Fidelity

Description

This adjustment sets the output amplitude of the A4 log amp/cal oscillator and the absolute amplitude of the reference 15 dB attenuator.
The output of the A4 log amp/cal oscillator is adjusted so that a 55 dBm signal applied to the 10.7 MHz IF input on the A5 IF assembly (A5J3) causes a displayed signal of 60 dBm. The effect of this adjustment is visible only after the complete. amplitude from the A4 log amp/cal oscillator.
This procedure also sets the attenuator of the reference 15 dB attenuator so that a source amplitude change of 50 dB combined with a spectrum analyzer reference level change of 50 dB displays an amplitude difference of 50 dB.
Figure 3a-1 IF Amplitude Adjustment Setup
ADJ CURR IF STATE causes the IF gain adjustment to use the “new” output
ADJ CURR IF STATE sequence is
3a-4 Chapter3
Page 33
Manual Adjustment Procedures: HP 3335A Source Not Available
4a. IF Amplitude Adjustments

Equipment

Signal Generator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . HP E4421B
Adapters
Type N (m) to BNC (f) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1250-1476
Type N (f) to 2.4 mm (f) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .HP 11903B
Cables
BNC, 122 cm (48 in) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .HP 10503A
Test cable . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 85680-60093
Figure 3a-2 IF Amplitude Adjustment Locations
NOTE The 15 dB reference attenuator adjustment is preset at the factory and need
not be done if the entire A5 IF assembly is replaced.
Chapter 3 3a-5
Page 34
Manual Adjustment Procedures: HP 3335A Source Not Available
4a. IF Amplitude Adjustments

Procedure

1. Press LINE to turn the spectrum analyzer off. Remove the spectrum analyzer cover and place the spectrum analyzer in the service position as illustrated in Figure 3a-1.
2. Disconnect W29, violet coax cable, from A5J3. Connect the test cable between A5J3 and the RF output of the HP E4421B. Press
3. Set the spectrum analyzer controls as follows:
Center Frequency . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10.7 MHz
Span . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .200 kHz
Reference Level . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .−60 dBm
Attenuator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .0 dB
dB/division . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1 dB/DIV
Resolution bandwidth . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .300 kHz
Video bandwidth . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .100 Hz
LINE to turn the spectrum analyzer on.
4. On the spectrum analyzer, press
MKR, CAL, and IF ADJ ON OFF so OFF is underlined.
5. Set the HP E4421B controls as follows:
Frequency . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10.7 MHz
Amplitude . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .−55 dBm
Mod On/Off. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Off
6. Note the marker value. Ideally it should read 60 dBm ±0.1 dB.
7. If the marker reads less than 60.1 dBm, rotate A4R826 CAL OSC AMPTD one-third turn clockwise for every 0.1 dB less than 60 dBm. See Figure 3a-2 for the location of A4R826. A change in the displayed amplitude will not be seen until
ADJ CURR IF STATE
is pressed.
8. If the marker reads greater than 59.9 dBm, rotate A4R826 CAL OSC AMPTD one-third turn counter clockwise for every 0.1 dB greater than 60 dBm. See
Figure 3a-2 for the location of A4R826. A change in the displayed amplitude will not be
seen until
ADJ CURR IF STATE is pressed.
NOTE If A4R826 has inadequate range, refer to “Inadequate CAL OSC AMPTD
Range” in Chapter 9.
9. On the spectrum analyzer, press
ADJ CURR IF STATE. After allowing the analyzer time
to complete the adjustments, the displayed amplitude and marker reading should change.
10.Repeat step 7 and step 9 until the marker reads 60 dBm ±0.1 dB.
11.Disconnect the test cable from A5J3 and reconnect W29 to A5J3.
3a-6 Chapter3
Page 35
Manual Adjustment Procedures: HP 3335A Source Not Available
4a. IF Amplitude Adjustments

A5 Reference Attenuator Adjustment

1. Set the spectrum analyzer reference level to 60 dBm. If markers are displayed, press
MKR and MARKERS OFF.
2. Set the HP E4421B
AMPLITUDE to 60 dBm.
3. Connect a BNC cable between the RF output of the HP E4421B and the spectrum analyzer INPUT 50Ω.
4. On the spectrum analyzer, press
CAL and REF LVL ADJ. Use the front panel knob or step
keys to place the peak of the displayed signal 3 dB to 5 dB below the reference level.
5. On the spectrum analyzer, press
PEAK SEARCH and MARKER DELTA. Set the spectrum
analyzer reference level to 10 dBm.
6. Change the HP 4421B
7. On the spectrum analyzer, press
Amplitude to 10 dBm.
CAL.
8. Note the MKR amplitude. Ideally, it should read 50.00 dB ±0.1 dB.
9. If the MKR amplitude is less than 49.9 dB, rotate A5R343 15 dB ATTEN one-half turn counterclockwise for each 0.1 dB less than 50.00 dB. Do not adjust A5R343 more than five turns before continuing with the next step.
10.If the MKR amplitude is greater than 50.1 dB, rotate A5R343 15 dB ATTEN one-half turn clockwise for each 0.1 dB greater than 50.00 dB. Do not adjust A5R343 more than five turns before continuing with the next step.
11.On the spectrum analyzer, press
ADJ CURR IF STATE. Note the MKR amplitude
reading.
12.Repeat step 1 through step 11 until the MKR amplitude reading is 50.00 dB ±0.1 dB.
Chapter 3 3a-7
Page 36
Manual Adjustment Procedures: HP 3335A Source Not Available
4a. IF Amplitude Adjustments
A5 Adjustment Verification
1. On the spectrum analyzer, disconnect W29 from A5J3. Connect the test cable between A5J3 and the RF output of the HP E4421B.
2. Set the spectrum analyzer reference level to 10 dBm.
3. Set the HP E4421B
4. On the spectrum analyzer, press
Amplitude to 5 dBm.
MKR and MARKER NORMAL.
5. The MARKER amplitude should read 10 dBm ±0.13 dB. If the reading is outside of this range, repeat step 4 of “Procedure,” on page 6 through “A5 Reference Attenuator
Adjustment” step 12.
6. On the spectrum analyzer, reconnect W29 to A5J3. Press
PRESET and set the controls
as follows:
Center frequency . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 300 MHz
Span . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .0 Hz
Reference level . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .−10 dBm
Resolution bandwidth . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .300 kHz
7. Connect a BNC cable between the HP 8563E CAL OUTPUT and INPUT 50Ω.
8. On the spectrum analyzer, press
MKR CAL and REF LVL ADJ.
9. Use the knob or step keys to adjust the REF LEVEL CAL setting until the MKR reads
10.00 dBm ±0.1 dB.
10.On the spectrum analyzer, press
STORE REF LVL.
3a-8 Chapter3
Page 37
Manual Adjustment Procedures: HP 3335A Source Not Available
5a. DC Log Amplifier Adjustments
5a. DC Log Amplifier Adjustments
There are three DC log adjustments; limiter phase, linear fidelity, and log fidelity.

Assembly Adjusted

A4 log amp/cal oscillator

Related Performance Tests

IF Gain Uncertainty Scale Fidelity

Description

These three adjustment need only be done under the following conditions:
Limiter phase Only if a repair is made to blocks F, G, H, I, or J. Linear fidelity Only if a repair is made to blocks C, D, F, G, H, I, J, K, O, IF gain accuracy,
RBW switching, or log fidelity.
Log fidelity Only if a repair is made to blocks D, F, H, K, IF gain accuracy, RBW
switching, or log fidelity.
If multiple adjustments are required they should be done in the following order:
1. Limiter Phase
2. Linear Fidelity
3. Log Fidelity
Figure 3a-3 DC Log Adjustment Setup
Chapter 3 3a-9
Page 38
Manual Adjustment Procedures: HP 3335A Source Not Available
5a. DC Log Amplifier Adjustments

Equipment

Signal Generator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .HP E4421B
Adapters
Type N (m) to BNC (f) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1250-1476
Type N (f) to 2.4 mm (f) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . HP 11903B
Cables
BNC, 122 cm (48 in) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . HP 10503A
Test cable . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 85680-60093
Figure 3a-4 DC Log Adjustment Locations
NOTE Adjustments should be made with all of the shields on and only after allowing
at least a 20 minute warmup.
3a-10 Chapter3
Page 39
Manual Adjustment Procedures: HP 3335A Source Not Available
5a. DC Log Amplifier Adjustments

A4 Limiter Phase Adjustment

1. Press LINE to turn the spectrum analyzer off. Remove the spectrum analyzer cover and place the spectrum analyzer in the service position as illustrated in Figure 3a-3. See
Figure 3a-4 for adjustment location.
2. Connect the HP E4421B RF output to the spectrum analyzer 50 input. Press turn the spectrum analyzer on.
3. Set the spectrum analyzer controls as follows:
Center frequency . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15 MHz
Span . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0 Hz
Reference level . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . −10 dBm
dB/division . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 dB/DIV
Resolution bandwidth. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 300 kHz
IF ADJ. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . OFF
4. Set up an HP E4421B as follows:
Frequency . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15 MHz
Amplitude . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . −18 dBm
Mod On/Off . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Off
5. On the spectrum analyzer, press complete adjustments then press
CAL and ADJ CURR IF STATE, wait for the analyzer to
MKR.
LINE to
6. Adjust A4R445 for maximum on-screen amplitude. Refer to Figure 3a-4 for the location of A4R445.
Chapter 3 3a-11
Page 40
Manual Adjustment Procedures: HP 3335A Source Not Available
5a. DC Log Amplifier Adjustments

A4 Linear Fidelity Adjustment

1. Press LINE to turn the spectrum analyzer off. Remove the spectrum analyzer cover and place the spectrum analyzer in the service position as illustrated in Figure 3a-3. See
Figure 3a-4 for adjustment location.
2. Connect the HP E4421B RF output to the spectrum analyzer 50 input. Press turn the spectrum analyzer on.
3. On the spectrum analyzer, press
AMPTD UNITS, dBm, CAL, and IF ADJ ON OFF (OFF).
PRESET AMPLITUDE, LINEAR, MORE 1 of 3,
4. Set the spectrum analyzer controls as follows:
Center frequency . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15 MHz
Span . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 MHz
Resolution bandwidth . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .300 kHz
Reference level. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .−10 dBm
5. Set up an HP E4421B as follows:
Frequency. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15 MHz
Amplitude. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .−10 dBm
Mod On/Off. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Off
6. On the spectrum analyzer, press
PEAK SEARCH and MARKER DELTA.
7. Reduce the HP E4421B input power to 58 dBm.
8. If the delta marker amplitude reads 40 dB ±2 dB, no adjustment is necessary.
LINE to
9. If the signal is lower on the screen than expected (delta marker amplitude reads less than 42dB) then adjust A4R544 (see Figure 3a-4) for an even lower level and press
CAL and ADJ CURR IF STATE. Allow sufficient time for the analyzer to complete the
adjustment.
10.If the signal is higher on the screen than expected (delta marker amplitude reads greater than 38 dB) then adjust A4R544 for an even higher level signal and press
CAL
and ADJ CURR IF STATE. Allow sufficient time for the analyzer to complete the adjustment.
11.Repeat step 7 through step 10.
3a-12 Chapter3
Page 41
Manual Adjustment Procedures: HP 3335A Source Not Available
5a. DC Log Amplifier Adjustments

A4 LOG Fidelity Adjustment

1. Press LINE to turn the spectrum analyzer off. Remove the spectrum analyzer cover and place the spectrum analyzer in the service position as illustrated in Figure 3a-3. See
Figure 3a-4 for adjustment location.
2. Connect the HP E4421B RF output to the spectrum analyzer 50 input. Press turn the spectrum analyzer on.
3. On the spectrum analyzer, press
ADJ CURR IF STATE.
PRESET, CAL, IF ADJ ON OFF (OFF), and
4. Set the spectrum analyzer controls as follows:
Center frequency . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15 MHz
Span. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0 MHz
Resolution bandwidth. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 300 kHz
Reference level . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . −10 dBm
5. Set up an HP E4421B as follows:
Frequency . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15 MHz
Amplitude . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . −10 dBm
Mod On/Off . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Off
6. On the spectrum analyzer, press
MKR and MARKER DELTA.
7. Decrease the HP E4421B power to 26 dBm.
LINE to
8. Calculate the error:
Error delta marker reading 16 dB=
9. If the error is less than ±0.2 dB, no adjustment is necessary.
10.Set the HP E4421B power to 10 dBm.
11.Adjust A4R531 (see Figure 3a-4) to read two times the error. For example, if the calculated error is +0.75 dB, adjust A4R531 for a delta marker amplitude reading of +1.5 dB. Press
12.Repeat step 7 through step 11.
CAL and ADJ CURR IF STATE.
Chapter 3 3a-13
Page 42
Manual Adjustment Procedures: HP 3335A Source Not Available
5a. DC Log Amplifier Adjustments
3a-14 Chapter3
Page 43

Index

A
A4 limiter phase adjustment
3a-11
A4 linear fidelity adjustment
3a-12
A4 log fidelity adjustment adjustment
adjustable components table
3-5
factory selected components related adjustments table safety considerations tools
, 3-4
adjustment procedures
cal oscillator amplitude DC log amplifiers IF amplitude
adjustments
recommended test equipment
3a-3
, 3a-4
, 3a-13
, 3-3
, 3-2
, 3a-4
, 3a-9
, ,
,
, 3-5
, 3-4
,
C
components
adjustable components table
3-5
D
DC log amplifer adjustments
3a-9
I
IF amplitude adjustments implementation instructions introduction
L
limiter phase adjustment linear fidelity adjustment log fidelity adjustment
T
test equipment list test equipment, automated
adjustments
tools
adjustment
U
using the supplement
, 1-2
, 2-18
, 3-4
, 3a-4
, 3a-11
, 3a-12
, 3a-13
, 3a-3
, 1-1
,
,
, 1-3
Index Index-1
Loading...