The information contained in this document is subject to change without notice.
Hewlett-Packard makes no warranty of any kind with regard to this material, including
but not limited to, the implied warranties of merchantability and fitness for a particular
purpose. Hewlett-Packard shall not be liable for errors contained herein or for incidental or
consequential damages in connection with the furnishing, performance, or use of this
material.
Assistance
For any assistance, contact your nearest Hewlett-Packard Sales and Service Office.
General Safety Considerations
WARNINGBefore this instrument is switched on, make sure it has been properly
grounded through the protective conductor of the ac power cable to
a socket outlet provided with protective earth contact.
Any interruption of the protective (grounding) conductor, inside or
outside the instrument, or disconnection of the protective earth
terminal can result in personal injury.
WARNINGThere are many points in the instrument which can, if contacted,
cause personal injury. Be extremely careful.
Any adjustments or service procedures that require operation of the
instrument with protective covers removed should be performed
only by trained service personnel.
CAUTIONBefore this instrument is switched on, make sure its primary power circuitry
has been adapted to the voltage of the ac power source.
Failure to set the ac power input to the correct voltage could cause damage to
the instrument when the ac power cable is plugged in.
The HP 3335A Synthesizer Level Generator signal source has become obsolete because
parts used in the manufacture of this instrument are no longer available from suppliers.
To meet the needs of our customers, HP has created new performance verification tests
and adjustment tests that use alternative signal sources.
The calibration and service guides are being revised to add additional procedures that do
not use the HP 3335A. The changes include the addition of signal sources required to
replace the 3335A, changes to the test equipment setup illustrations, and changes in the
steps required to execute the procedures.
Because all of our customers will not need to replace the HP 3335A immediately, we are
releasing the changes to the calibration guide and service guides in the form of two manual
supplements. HP is making these procedure revisions available at this time in electronic
form, only. You can download this manual supplement, and add it to your current manual
as needed. You will not be required to replace your manuals just to incorporate these
changes. These changes will be incorporated in new manuals shipped in the future.
To add these changes to your existing manuals, refer to “Implementation Instructions,” on
page 3.
1-2Chapter1
Using this Adjustment Procedure Supplement
Implementation Instructions
Implementation Instructions
Update your service guide by performing the following steps:
Step 1. After downloading this supplement, print a hard copy of the entire document.
Step 2. Remove page 2-7 and 2-8 from your service guide.
Step 3. Insert page 2-7 and 2-8 from the supplement into the service guide.
Step 4. Remove page 3-1 through 3-8 from your service guide.
Step 5. Insert page 3-1 through 2-8 from the supplement into the service guide.
Step 6. Remove the adjustment procedure chapter from the supplement, and insert it
into your service guide as Chapter 3a, “Manual Adjustment Procedures:
HP 3335A Source Not Available,” immediately after Chapter 3.
Chapter 11-3
Using this Adjustment Procedure Supplement
Implementation Instructions
1-4Chapter1
2Test Equipment Required for
Automated Adjustments
2-1
Test Equipment Required for Automated Adjustments
Introduction
Introduction
This chapter contains the table of test equipment that is required for the automated
adjustments. This table is to replace table Table 2-2., “Test Equipment Required for
Automated Adjustments” in Chapter 2, “Adjustment/Diagnostic Software,” of your service
guide.
2-2Chapter2
Test Equipment Required for Automated Adjustments
Introduction
This page intentionally left blank.
Chapter 22-3
Test Equipment Required for Automated Adjustments
Introduction
This page intentionally left blank.
2-4Chapter2
Test Equipment Required for Automated Adjustments
Introduction
This page intentionally left blank.
Chapter 22-5
Loading...
+ 30 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.