The information contained in this document is subject to change without notice.
Hewlett-Packard makes no warranty of any kind with regard to this material, including
but not limited to, the implied warranties of merchantability and fitness for a particular
purpose. Hewlett-Packard shall not be liable for errors contained herein or for incidental or
consequential damages in connection with the furnishing, performance, or use of this
material.
Assistance
For any assistance, contact your nearest Hewlett-Packard Sales and Service Office.
General Safety Considerations
WARNINGBefore this instrument is switched on, make sure it has been properly
grounded through the protective conductor of the ac power cable to
a socket outlet provided with protective earth contact.
Any interruption of the protective (grounding) conductor, inside or
outside the instrument, or disconnection of the protective earth
terminal can result in personal injury.
WARNINGThere are many points in the instrument which can, if contacted,
cause personal injury. Be extremely careful.
Any adjustments or service procedures that require operation of the
instrument with protective covers removed should be performed
only by trained service personnel.
CAUTIONBefore this instrument is switched on, make sure its primary power circuitry
has been adapted to the voltage of the ac power source.
Failure to set the ac power input to the correct voltage could cause damage to
the instrument when the ac power cable is plugged in.
This chapter contains an introduction that explains the need for this service guide
supplement and instructions on how to implement the supplement material.
1-1
Using this Adjustment Procedure Supplement
Introduction
Introduction
The HP 3335A Synthesizer Level Generator signal source has become obsolete because
parts used in the manufacture of this instrument are no longer available from suppliers.
To meet the needs of our customers, HP has created new performance verification tests
and adjustment tests that use alternative signal sources.
The calibration and service guides are being revised to add additional procedures that do
not use the HP 3335A. The changes include the addition of signal sources required to
replace the 3335A, changes to the test equipment setup illustrations, and changes in the
steps required to execute the procedures.
Because all of our customers will not need to replace the HP 3335A immediately, we are
releasing the changes to the calibration guide and service guides in the form of two manual
supplements. HP is making these procedure revisions available at this time in electronic
form, only. You can download this manual supplement, and add it to your current manual
as needed. You will not be required to replace your manuals just to incorporate these
changes. These changes will be incorporated in new manuals shipped in the future.
To add these changes to your existing manuals, refer to “Implementation Instructions,” on
page 3.
1-2Chapter1
Using this Adjustment Procedure Supplement
Implementation Instructions
Implementation Instructions
Update your service guide by performing the following steps:
Step 1. After downloading this supplement, print a hard copy of the entire document.
Step 2. Remove page 2-1 through 2-10 from your service guide.
Step 3. Cross out Table 2-5, “Required Test Equipment for TAM,” on page 2-11.
Step 4. Insert page 2-1 through 2-12 from the supplement into the service guide.
Step 5. Remove the adjustment procedure chapter from the supplement, and insert it
into your service guide as Chapter 2a, “Manual Adjustment Procedures:
HP 3335A Source not Available,” immediately after Chapter 2.
Chapter 11-3
Using this Adjustment Procedure Supplement
Implementation Instructions
1-4Chapter1
2Adjustment Procedures
2-1
Adjustment Procedures
Introduction
Introduction
This chapter contains information on automated and manual adjustment procedures.
Perform the automated procedures using the HP 85629B test and adjustment module
(TAM). Adjustments should be performed after a repair or performance test failure, never
perform adjustments as routine maintenance. Refer to Table 2-1 to for which adjustments
to perform.
23. 600 MHz Reference Adjustment (serial prefix 3406A and above)
NOTEBefore performing any adjustments, allow the spectrum analyzer to warm up
for at least 5 minutes.
Safety Considerations
Although this instrument has been designed in accordance with international safety
standards, this manual contains information, cautions, and warnings which must be
followed to ensure safe operation and to prevent damage to the instrument. Service and
adjustments should be performed only by qualified service personnel.
2-2Chapter2
WARNING
Adjustment Procedures
Introduction
• These servicing instructions are for use by qualified personnel
only. To avoid electrical shock, do not perform any servicing
unless you are qualified to do so.
• The opening of covers or removal of parts is likely to expose
dangerous voltages. Disconnect the product from all voltage
sources while it is being opened.
• Adjustments in this section are performed with power supplied to
the instrument and protective covers removed. There are voltages
at many points in the instrument which can, if contacted, cause
personal injury. Be extremely careful. Adjustments should be
performed only by trained service personnel.
• Power is still applied to this instrument with the
off position. Before removing or installing any assembly or
printed circuit board, remove the line-power cord.
• The power cord is connected to internal capacitors inside that
may remain live for 5 seconds after the instrument has been
disconnected from its source of supply.
• Use a nonmetallic adjustment tool whenever possible.
LINE switch in the
Which Adjustments Should Be Performed?
Table 2-1 lists the manual adjustments that should be performed when an assembly is
repaired or changed. It is important to perform the adjustments in the order indicated to
ensure that the instrument meets its specifications.
Chapter 2a, “Manual Adjustment Procedures: HP 3335A Source not Available,” provides
instructions and procedures for making adjustments when the HP 3335A Synthesizer
Level Generator is not available. If the HP 3335A is not available, substitute those
procedures for the procedures of the same number and name found in this chapter.
Test Equipment
The equipment required for the manual adjustment procedures is listed in Table 1-4,
“Recommended Test Equipment.” Any equipment that satisfies the critical specifications
given in the table may be substituted for the preferred test equipment.
Adjustable and Factory-Selected Components
Table 2-2 lists the adjustable components by reference designation and name. For each
component, the table provides a description and lists the adjustment number.
Refer to Table 2-3 for a complete list of factory-selected components used in the instrument
along with their functions. Factory-selected components are identified with an asterisk on
the schematic diagrams.
Chapter 22-3
Loading...
+ 24 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.