HP 3400a schematic

*TB 9-6625-011-35
SUPERSEDED COPY DATED 30 AUGUST 1982
DEPARTMENT OF THE ARMY TECHNICAL BULLETIN
CALIBRATION PROCEDURE FOR
TRUE RMS VOLTMETER ME-318/U
(HEWLETT-PACKARD, MODEL 3400A)
Headquarters, Department of the Army, Washington, DC
29 May 1992
Approved for public release; distribution is unlimited.
♦♦REPORTING OF ERRORS♦♦
You can help improve this publication by calling attention to errors and by recommending improvements and stating your reasons for the recommendations. Your letter or DA Form 2028, Recommended Changes to Publications, should be mailed directly to Commander, U. S. Army Aviation and Missile Command, ATTN: AMSAM-TMD-EP, Redstone Arsenal, AL 35898-5000. FAX to DSN 788-2313 (commercial 256-842-2313). A reply will be furnished directly to you.
SECTION I INDENTIFICATION AND DESCRIPTION
Test instrument identification.............................. 1 2
Forms, records, and reports.................................. 2 2
Calibration description......................................... 3 2
II EQUIPMENT REQUIREMENTS
Equipment required.............................................. 4 2
Accessories required............................................. 5 3
III CALIBRATION PROCESS
Preliminary instructions ...................................... 6 4
Equipment setup................................................... 7 5
Voltage accuracy................................................... 8 5
Final procedure..................................................... 9 10
*This bulletin supersedes TB 9-6625-011-35, dated 30 August 1982, including all changes.
Paragraph Page
TB 9-6625-011-35

SECTION I

IDENTIFICATION AND DESCRIPTION
1. Test Instrument Identification. This bulletin provides instructions for the
calibration of True RMS Voltmeter ME-318/U (Hewlett-Packard, Model 3400A). The manufacturer's manuals were used as the prime data sources in compiling these instructions. The equipment being calibrated will be referred to as the TI (test instrument) throughout this bulletin.
a. Model Variations. Variations among models are described in text. b. Time and Technique. The time required for this calibration is approximately 1.5
hours, using the dc and low frequency technique.
2. Forms, Records, and Reports. Forms, records, and reports required for calibration personnel at all levels are prescribed by TB 750-25.
3. Calibration Description. TI parameters and performance specifications which pertain to this calibration are listed in table I.
Table 1. Calibration Description
Test instrument parameters Performance specifications
Ac voltage Range: 1.0 mV to 300 V rms in 12 ranges
Frequency: 10 Hz to 10 MHz Accuracy: ±(%) FS
Frequency
10 to 50 Hz................ 5.0
50 Hz to 1.0 MHz............. 1.0
1.0 to 2.0 MHz............. 2.0
2.0 to 3.0 MHz............. 3.0
3.0 to 10 MHz............. 5.0

SECTION II

EQUIPMENT REQUIREMENTS
4. Equipment Required. Table 2 identifies the specific equipment to be used in this
calibration procedure. This equipment is issued with Secondary Transfer Calibration Standards Set AN/GSM-286. Alternate items may be used by the calibrating activity. The items selected must be verified to perform satisfactorily prior to use and must bear evidence of current calibration. The equipment must meet or exceed the minimum use specifications listed in table 2. The accuracies listed in table 2 provide a four-to-one ratio between the standard and TI.
5. Accessories Required. The accessories required for this calibration are common usage accessories issued as indicated in paragraph 4 above and are not listed in this calibration procedure.
2
Table 2. Minimum Specifications of Equipment Required
Common name Minimum use specifications
CALIBRATOR
DIGITAL MULTIMETER
Ac voltage:
Range: 10 mV to 300 V Frequency: 400 Hz and 1 kHz Accuracy: ±.25%
Wideband voltage:
Voltage: 900 µV to 0.9 V Frequency: 20 Hz to 10 MHz (1 kHz reference) Amplitude flatness: ±(%)
Frequency: 20 Hz 1.25
400 Hz to 1 MHz 0.25
2 MHz 0.50 3 MHz 0.75
10 MHz 1.25 Range: -1.0 V dc Accuracy: ±.25%
TB 9-6625-011-35
Manufacturer and model
(part number)
John Fluke, Model 5700A/CT w/option 003 (p/o MIS-35947) w/Power Amplifier, John Fluke, Model 5215A/CT (5215A/CT), w/Ac Divider, John Fluke, Model 7405A­4207 (7405A-4207)
John Fluke, Model 8506A/CT (p/o MIS-
35947)

SECTION III

CALIBRATION PROCESS
6. Preliminary Instructions
a. The instructions outlined in paragraphs 6 and 7 are preparatory to the calibration
process. Personnel should become familiar with the entire bulletin before beginning the calibration.
b. Items of equipment used in this procedure are referenced within the text by common
name as listed in table 2.
c. Unless otherwise specified, verify the results of each test and, whenever the test requirement is not met, take corrective action before continuing with the calibration. Additional maintenance information is contained in the manufacturers' manuals for this TI.
d. Unless otherwise specified, all controls and control settings refer to the TI.
3
TB 9-6625-011-35
7. Equipment Setup
WARNING
HIGH VOLTAGE is used or exposed during the performance of this calibration. DEATH ON CONTACT may result if personnel fail to observe safety precautions. REDUCE OUTPUT(S) to minimum after each step within the performance check where applicable.
a. Remove protective cover from TI only when necessary to make adjustments. Replace cover after completing the adjustments; then repeat the check.
b. Zero meter using front panel adjustment and set RANGE switch to 300 VOLTS.
c Connect to an appropriate ac voltage source and press LINE switch to ON. Allow at
least 30 minutes for warmup.
8. Voltage Accuracy
a. Performance Check
(1) Connect calibrator OUTPUT to TI INPUT terminal and connect digital
multimeter to TI rear panel DC OUT terminal.
(2) Set TI RANGE switch to .01 VOLTS and set calibrator for a 10 mV, 400 Hz
output. If digital multimeter does not indicate -1.0 (±.01) V, perform b (1) and (2) below.
(3) Connect calibrator OUTPUT to TI INPUT terminal.
(4) Set calibrator for a 10 mV, 400 Hz output, then adjust calibrator for a TI indication of 1 on the 0-to-1 scale. Calibrator Error display will be < ±1.0%; if not, perform b(3) through (7) below.
(5) Set TI RANGE switch to 1 VOLTS and set calibrator for a 1.0 V, 400 Hz output, then adjust calibrator for a TI indication of 1 on the 0-to-1 scale. Calibrator Error display will be < ±1.0%; if not, perform b(8) below.
(6) Reset calibrator.
(7) Connect ac divider INPUT (p/o calibrator) to calibrator OUTPUT terminals and ac divider OUTPUT to TI INPUT terminal.
(8) Set TI RANGE switch and calibrator initial output as indicated in table 3. Adjust calibrator for the TI meter indication specified. Calibrator Error display will indicate within the specified limits.
4
TB 9-6625-011-35
(9) Connect TI INPUT to calibrator WIDEBAND output and press calibrator W BND pushbutton.
(10) Set TI RANGE switch to .001 VOLTS.
(11) Set calibrator for an initial 900 µV, 1 kHz, wideband output. Adjust calibrator for a TI indication equal to value recorded in table 3 (.0009 V) to establish a 1 kHz reference. Press calibrator NEW REF pushbutton.
(12) Set calibrator frequency to 20 Hz, then readjust amplitude for TI reference established in (11) above. Calibrator Error display indication will be < ±5.6%.
(13) Repeat technique of (12) above for remaining frequencies listed for the .001 VOLTS RANGE switch settings in table 4. Calibrator Error display will be within specified limits; if not, perform b(9) through (11) below.
(14) Repeat technique of (10) through (13) above for TI RANGE switch settings and calibrator initial voltage listed in table 4. Calibrator Error display indication will be within the limits specified; if not, perform b(12) through (14) below for 1 VOLTS RANGE switch setting and (15) through (17) for .3 VOLTS RANGE switch setting.
Table 3. Range Accuracy
Test instrument Calibrator
RANGE
switch settings
(VOLTS)
.001 1 --- 1.0 V 1.0 kHz 1.0 .003 --- 3 3.0 V 1.0 kHz 1.0
.03 --- 3 30 mV 1.0 kHz 1.0 .1 1 --- 100 mV 1.0 kHz 1.0
.3 --- 3 0.3 V 1.0 kHz 1.0 1 1 --- 1.0 V 1.0 kHz 1.0 1 .8 --- 0.8 V 1.0 kHz 1.25 1 .6 --- 0.6 V 1.0 kHz 1.7 1 .4 --- 0.4 V 1.0 kHz 2.5 1 .2 --- 200 mV 1.0 kHz 5.0 3 --- 3 3.0 V 1.0 kHz 1.0
10 1 --- 10 V 1.0 kHz 1.0 30 --- 3 30 V 1.0 kHz 1.0
100 1 --- 100 V 1.0 kHz 1.0
1 2
b(15) below.
3
below.
300 --- 3 300 V 1.0 kHz 1.0
After performing this check, set calibrator for a .9 V, 1 kHz output and record TI indication (.0009 V) for use in b(9) below. After performing this check, set calibrator for a 0.3 V. 1 kHz output and record TI indication (on the 0-to-1 scale) for used in
After performing this check, set calibrator for a .9 V, 1 kHz output and record TI indication for use in b(l2)
Meter indication scale Initial output
0-to-1 0-to-3 Voltage Frequency
Reset calibrator and remove ac divider from setup
ERROR
indications
< ± (%)
display
1
2 3
5
TB 9-6625-011-35
Table 4. Frequency Response
Test instrument Calibrator
RANGE
switch settings Initial
voltage Frequency .001 --- 400 Hz 1.1 .001 --- 10 kHz 1.1 .001 --- 50 kHz 1.1 .001 --- 100 kHz 1.1 .001 --- 500 kHz 1.1 .001 --- 1.0 MHz 1.1 .001 --- 2.0 MHz 2.2 .001 --- 3.0 MHz 3.3 .001 --- 10 MHz 5.6
1 0.9 V 1.0 kHz N/A 1 --- 20 Hz 5.6 1 --- 10 kHz 1.1 1 --- 50 kHz 1.1 1 --- 100 kHz 1.1 1 --- 500 kHz 1.1 1 --- 1.0 MHz 1.1 1 --- 2.0 MHz 2.2 1 --- 3.0 MHz 3.3 1 --- 10 MHz 5.6
.3 0.3 V 1.0 kHz N/A .3 --- 20 Hz 5.0 .3 --- 400 Hz 1.0 .3 --- 10 kHz 1.0 .3 ... 50 kHz 1.0 .3 --- 100 kHz 1.0 .3 --- 500 kHz 1.0 .3 --- 1.0 MHz 1.0 .3 --- 2.0 MHz 2.0 .3 --- 3.0 MHz 3.0 .3 --- 10 MHz 5.0
Output
Error display limits
±(%)
b. Adjustments
(1) Adjust R4 (R627 for TIs with amplifier board A6, P/N 03400-66512) (fig.1) until
digital multimeter indicates -1.00 V.
(2) If necessary, adjust R6 (fig. 1) for full scale indication on TI. (3) Set calibrator for a 10 mV, 400 Hz output. Adjust R6 (fig.1) until TI indicates
full scale (R).
(4) Set RANGE switch to .1 VOLTS.
6
(5) Adjust R7 (R626 for TIs with amplifier board A6, P/N 03400-66512) (fig. 1) until
TI indicates 0.01 V (1 /10 scale) (R).
(6) Set RANGE switch to .01 VOLTS.
(7) Repeat (3) through (6) above until no further adjustment is necessary.
(8) Set calibrator for a 1.0 V, 400 Hz output. Adjust R104 (fig. 1) until TI indicates
1 (R).
(9) Set calibrator for an initial 900 µV 1 kHz wideband output, then adjust for the
TI indication record for table 3 (.0009 V).
(10) Press calibrator NEW REF pushbutton, then set frequency to 10 MHz. Adjust
C405 (fig. 1) for the 1 kHz reference established b(9) above (R).
(11) Vary calibrator frequency between 3 and 10 MHz and if required, readjust C405
(fig. 1) for best in-tolerance condition.
TB 9-6625-011-35
(12) Set calibrator for an initial 0.9 V, 1 kHz wideband output, then adjust for the TI
indication recorded for table 3 (0.9 V).
(13) Press calibrator NEW REF pushbutton, then set frequency to 100 kHz. Adjust
C102 (fig. 1) for the 1 kHz reference established in b(12) above (R).
(14) Vary calibrator frequency between 100 kHz and 10 MHz and, if required,
readjust C102 (fig. 1) for best intolerance condition.
(15) Set calibrator for an initial 0.3 V, 1 kHz wideband output, then adjust for the TI
indication recorded for table 3 (0.3 V).
(16) Press calibrator NEW REF pushbutton, then set frequency to 3.0 MHz.).
Adjust C303 (fig. 1) for the 1 kHz reference established in b(l5) above (R).
(17) Vary calibrator frequency between 3 and 10 MHz and, if required, readjust C303
(fig. 1) for best in-tolerance condition.
7
TB 9-6625-011-35
9. Final Procedure
a. Deenergize and disconnect all equipment. b. Annotate and affix DA Label/Form in accordance with TB 750-25.
8
TB 9-6625-011-35
By Order of the Secretary of the Army:
GORDON R. SULLIVAN
General, United States Army
Chief of Staff
01527
Distribution: To be distributed in accordance with DA Form 12-34-E, Block No. 2091, requirements for calibration procedure TB 9-6625-011-35.
US GOVERNMENT PRINTING OFFICE: 1992 - 631-006/60238
PIN NO: 015188-000
Errata
3400A Operating & Service Manual
03400-90013
August 1984
Title & Document Type:
Manual Part Number:
Revision Date:
HP References in this Manual
This manual may contain references to HP or Hewlett-Packard. Please note that Hewlett­Packard's former test and measurement, semiconductor products and chemical analysis businesses are now part of Agilent Technologies. We have made no changes to this manual copy. The HP XXXX referred to in this document is now the Agilent XXXX. For example, model number HP8648A is now model number Agilent 8648A.
About this Manual
We’ve added this manual to the Agilent website in an effort to help you support your product. This manual provides the best information we could find. It may be incomplete or contain dated information, and the scan quality may not be idea l. If we find a better copy in the future, we will add it to the Agilent website.
Support for Your Product
Agilent no longer sells or supports this product. You will find any other available product information on the Agilent Test & Measurement website:
www.tm.agilent.com
Search for the model number of this product, and the resulting product page will guide you to any available information. Our service centers may be able to perform calibration if no repair parts are needed, but no other support from Agilent is available.
Loading...