HP 11713A OPERATING AND SERVICE MANUAL

Agilent Technologies
11713A Attenuator/Switch Driver
Operating and Service Manual
Part Number: 11713-90023
Printed in USA
Print Date: July 2001
Supersedes: August 1999
.

Notice

The information contained in this document is subject to change without notice.
Agilent T echnologies makes no warranty of any kind with r egard to this material, including, but not limited to, the implied war ranties of merchantability and fitness for a particular pur pose. Agilent Technologies shall not be liable for errors contained herein or for incidental or consequential damages in connect ion with the furnishing, performance, or use of this material.
Agilent T echnologies assumes no responsibility for the use or relia bility of its software on equipment that is not furnished by Agilent Technologies.
This document contains proprietary information which is protected by copyright. All rights are reserved. No part of this document may be photocopied, reproduced, or translated to another language without prior written consent of Agilent Technologies.
Restricted Rights Legend
Use, duplication, or disc losure by the U.S. Government is subject to restrictions as set forth in subparagraph (c)(1)(ii) of the Rights in Tec hni cal Data and Computer Softw ar e clause at DFARS 252.227-7013 for DOD agencies, and subparagraphs (c)(1) and (c)( 2) of the Commercial Computer Software Restricted Rights clause at FAR
52.227-19 for other agencies. Agilent Technologies
Santa Rosa Systems Division 1400 Fountaingrove Parkway Santa Rosa, CA 95403-1799, U.S.A.
Copyright 1987, 1993, 1999–2001 Agilent Technologies, Inc

Warranty

Certification Agilent Technologies certifies that this product met its published
specifications at the time of shipment from the factory. Agilent Technologies further certifies th at its calibration measurements are traceable to the United States National Institute of Standards and T ec hnology (NIST, formerly NBS), to the extent a llowed by the Institute ’s calibration facility, and to the calibration facilities of other International Standards Organization members.
Warranty This Agilent Technologies system product is warranted against defects
in materials and workmanship for a period corresponding to the individual warranty periods of its compo nent products . Instruments are warranted for a pe riod of one y ear. During the w arra nty pe riod , Agile nt Technologies will, at its option, either repair or replace products that prove to be defective.
Warranty service for products installed by Agilent Technol ogies and certain other products designated by Agilent will be performed at Buyer’s facility at no charge wit hin Agilent T e chhnologies service travel areas. Outside Agilent Techhnologies service travel areas, w arranty service will be performed at Buyer’s facility only upon Agilent Tec hhnologies prior agreement and Buyer shall pay Agilent Techhnologies round trip travel expenses. In all other areas, products must be returned to a service facility designated by Agilent Techhnologies.
For products returned to Agilent Techhnologies for warranty service, Buyer shall prepay shipping charges to Agilent Techhnologies and Agilent T echhnologies shall pay shipping charges to return the product to Buyer. However , Buyer shall pay all shipping charges, duties, and taxes for products returned to Agilent Techhnologies from another country.
Agilent Techhnologies warrants that its software and firmware designated by Agilent Techhnologies for use with an instrument will execute its programming instructions when properly installed on that instrument. Agilent Techhnologies does not warrant that the operation of the instrument, or software, or firmware will be uninterrupted or error free.
LIMITATION OF WARRANTY. The foregoing warranty s hall not apply to defects resulting from improp er or inad equate maintenanc e by Buyer, Buyer-supplied software or interfacing, unauthorized modification or misuse, operation outside of the environmental specifications for the product, or imp roper site preparation or maintenance.
NO OTHER WARRANTY IS EXPRESSED OR IMPLIED. AGILENT TECHNOLOGIES SPECIFICALLY DISCLAIM S THE IMPLIED WARRANTIES OR MERCHANTABILITY AND FITNESS FOR A
11713A Operating and Service Manual iii
PARTICULAR PURPOSE. EXCLUSIVE REMEDIES. THE REMEDIES PROVIDED HEREIN
ARE BUYER’S SOLE AND EXCLUSIVE REMEDIES. AGILENT TECHNOLOGIES SHALL NOT BE LIABLE FOR ANY DIRECT, INDIRECT, SPECIAL, INCIDENTAL, OR CONSEQUENTIAL DAMAGES, WHETHER BASED ON CONTRACT, TORT, OR ANY OTHER LEGAL THEORY.
Assistance Product maintenance agreements and other customer assistance
agreements are available for Agilent Technologies products. For assistance, call your local Agilent Technologies Sales and Service
Office (refer to "Service and Support" on page v).
iv 11713A Operating and Service Manual

Contacting Agilent

Any adjustment, maintenance, or repair of this product must be performed by qualified personnel. Contact your customer engineer through your local Agilent Techhnologies Service Center.
If you do not have access to the Internet, one of these Agilent Technologies centers can direct you to your nearest representative:
Online assistance: www.agilent.com/find/assist United States
(tel) 1 800 452 4844
New Zealand
(tel) 0 800 738 378 (fax) (+64) 4 495 8950
Latin America
(tel) (305) 269 7500 (fax) (305) 269 7599
Japan
(tel) (+81) 426 56 7832 (fax) (+81) 426 56 7840
Canada
(tel) 1 877 894 4414 (fax) (905) 282-6495
Australia
(tel) 1 800 629 485 (fax) (+61) 3 9210 59 47
Europe
(tel) (+31) 20 547 2323 (fax) (+31) 20 547 2390
Asia Call Center Numbers
Country Phone Number Fax Number
Singapore 1-800-375-8100 (65) 836-0252 Malaysia 1-800-828-848 1-800-801664 Philippines (632) 8426802
1-800-16510170 (PLDT Subscriber Only)
Thailand (088) 226-008 (outside Bangkok)
(662) 661-3999 (within Bangkok) Hong Kong 800-930-871 (852) 2506 9233 Taiwan 0800-047-866 (886) 2 25456723
(632) 8426809 1-800-16510288 (PLDT Subscriber Only)
(66) 1-661-3714
People’s Republic of China
India 1-600-11-2929 000-800-650-1101
800-810-0189 (preferred)
10800-650-0021
10800-650-0121
11713A Operating and Service Manual v

Safety and Regulat o ry Inf ormation

Review this product and related documentation to familiarize yourself with safety markings and instructions before you operate the instrument. This product has been designed and tested in accordance with international standards.
WARNING The WARNING notice denotes a hazard. It calls attention to a
procedure, practice, or the like , that, if n ot co rrectly pe rf ormed or adhered to, could result in personal injury. Do not proceed beyond a WARNING notice until the indicated conditions are fully understood and met.
CAUTION The CAUTION notice denotes a hazard. It calls attention to an
operating procedure, practice, or the like, which, if not correctly performed or adhered to , could result in damage to the produc t or loss of important data. Do not proceed beyond a CAUTION notice until the indicated conditions are fully understood and met.
Instrument Markings
When you see this symbol on your instrument, you should refer to the instruments instruction manual for important information.
This symbol indicates hazardous voltages.
This symbol indicates that the instrument requires alternating current (ac) input.
The CE mark is a registered tradema rk of the Europea n Co mmunity. If it is accompanied by a year, it indicates the year the design was proven.
The C-Tick mark is a regi stered t rademark of the Australian Spectrum Management Community.
The CSA mark is a registered trademark of the Canadian Standards Association.
1SM1-A This text indicates that the instrument is an Industrial Scientific and
Medical Group 1 Class A product (CISPER 11, Clause 4). This symbol indicates that the power line switch is ON.
This symbol indicates that the power line switch is in STANDBY position.
vi 11713A Operating and Service Manual
This symbol indicates that the power line switch is OFF
Safety Earth Ground
Before Applying Power
This is a Safety Class I product (provided with a protective earthing terminal). An uninterruptible safety earth ground must be provided from the main power source to the product input wiring terminals , p ower cord, or supplied power cord set. Whenever it is likely that the protection has been impaired, the product must be made inoperative and secured against any unintended operation.
Verify that the product is configured to match the available main power source as described in the input power configuration instructions in this manual. If this product is to be powered by autotransformer, make sure the common terminal is connected to the neutral (grounded) side of the ac power supply.
Compliance With German Noise Requirements
This is to declare that this instrument is in conformance with the German Regulation on Noise Declaration for Machines (Laermangabe nach der Maschinenlaermrerordnung-3.GSGV Deutschland).
Acoustic Noise Emmision/Gerauschemission
LpA <70 dB LpA <70 dB Operator position am Arbeitsplatz Normal position normaler Betrieb per ISO 7779 nach DIN 45635 t.19
11713A Operating and Service Manual vii
WARNING
WARNING
General Safety Considerations
• This product has been designed and tested in accordance with IEC Publication 1010, Safety Requirements for Electronic Measuring Apparatus, and has been supplied In a safe condition. The Instruction documentation contains information and warnings which must be followed by the user to ensure safe operation and to maintain the produc t In a safe condition.
• The ON//OFF switch or the detachable power cord is the instrument disconnecting device. It disconnects the main circuits from the main supply before other parts of the instrument. Alternately, an externally installed switch of circuit breaker, which is readily identifiable and is easily reached by the operator, may be used as a disconnecting device.
• This product is designed for use in Installation Category and P ollution Degree 2 per IEC 1010 and 664 respectively.
• Install the instrument according to the enclosure protection provided. This instrument protects against finger access to hazardous parts within the enclosure. The Instrument does not protect against the ingress of water.
• If this product is not used as specified, the protection provided by the equipment could be impaired. This product must be used in a normal condition (in which all means for protection are intact) only.
• When installing the product in a cabinet, the convection into and out of the product must not be restricted. The ambient temperature (outside the cabinet) must be less than the maximum operating temperature of the product by 4°C for every 100 watts dissipated in the cabinet. If the total power dissipated in the cabinet is greater than 800 wafts forced convection must be used.
viii 11713A Operating and Service Manual
11713A Operating and Service Manual ix

Typeface Conventions

Italics • Used to emphasize important information:
Use this software only with the xxxxxX system.
• Used for the title of a publication:
Refer to the xxxxX, System-Level User’s Guide
• Used to indicate a variable:
Type LOAD BIN filename.
Instrument
Display
• Used to show on-screen prompts and messages that you will see on
the display of an instrument:
The xxxxxX will display the message CAL1 SAVED.
[Keycap] • Used for labeled keys on the front panel of an instrument or on a
computer keyboard:
Press [Return].
{Softkey}
• Used for simulated keys that appear on an instrument display:
Press {Prior Menu}.
User Entry • Used to indicate text that you will enter using the computer
keyboard; text shown in this typeface must be typed exactly as printed:
Type LOAD PARMFILE
• Used for examples of programming code:
#endif / / ifndef NO CLASS
• Used for a subdirectory name or file path: Edit the file usr/1oca1/bin/samp1e.txt
Path Name
Computer
Display
• Used to show messages, prompts, and window labels that appear on a computer monitor:
Edit Parameters window will appear on the screen.
The
• Used for menus, lists , dialog boxes, and button boxes on a computer monitor from which you make selections using the mouse or keyboard:
Double-click
x 11713A Operating and Service Manual
EXIT to quit the program.
Contents
1.Agilent Technologies 11713A Attenuator/Switch Driver
Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-2
Compatible Attenuators and Switches . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-3
2.Installing the Agilent Technologies 11713A Attenuator/Switch Driver
Initial Inspection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-2
Environmental Limitations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-2
Preparing for use. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-2
Verifying Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-3
Figure 2-1 Line Voltage Selection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-4
GPIB Addressing. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-5
Figure 2-2 GPIB Address Selection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-5
Table 2-1 ASCII Address Codes to Decimal Equivalents . . . . . . . . . . . . . . . . . . . . . . . . . . .2-5
Mating Connectors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-6
Figure 2-3 General Purpose Interface Bus Connection . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-6
Connecting to Attenuators and Switches . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-7
Table 2-2 Attenuator Cable Connector Pin Numbers and Wire Color Codes . . . . . . . . . . .2-8
Driving Additional Coaxial Switches. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-9
Figure 2-4 Connections for Coaxial Switches . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-9
Figure 2-5 Connections for Agilent 8761B SPDT Switches . . . . . . . . . . . . . . . . . . . . . . . .2-10
Connecting to Four-Section . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-11
Figure 2-6 Typical Connections for Programmable Four-Section Attenuator. . . . . . . . . .2-11
Connecting to Relays. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-12
Figure 2-7 Relay Driving Circuit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-12
Verifying Performance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-12
3.Specifications
Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-2
Table 3-1 Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-2
Supplemental Characteristics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-2
Table 3-2 Supplemental Characteristics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-2
4.Verification
Recommended Test Equipment. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-2
Table 4-1 Recommended Test Equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-2
Operator’s Check for Local Operation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-3
Figure 4-1 Local Operation Test Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-3
GPIB Interface Check for Remote Operation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-5
Figure 4-2 GPIB Interface Check Test Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-5
Remote Message . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-6
Receiving the Message . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-6
5.Local Operation
Local Control . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-2
Figure 5-l Front Panel Features. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-3
Figure 5-2 Rear Panel Features. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-4
1
Contents
6.Remote Operation
GPIB Compatibility. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-2
Remote Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-2
Local Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-2
Addressing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-3
Data Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-3
Receiving the Data Message . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-3
Sending the Data Message . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-4
Receiving the Trigger Message . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-4
Receiving the Clear Message. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-4
Receiving the Local Lockout Message. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-5
Receiving the Clear Lockout/Set Local Message . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-5
Receiving the Pass Control Message. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-5
Sending the Require Service Message . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-5
Sending the Status Byte Message. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-5
Sending the Status Bit Message . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-5
Table 6-1 GPIB Message Reference Table. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-6
Table 6-2 GPIB Control Statement Results . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-7
Table 6-3 Attenuation Levels and Corresponding Data Strings . . . . . . . . . . . . . . . . . . . . .6-8
7.Replacing Major Assemblies
Replaceable Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-2
Table 7-1 Accessories Supplied . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-2
Table 7-2 Line Fuse Part Numbers. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-2
Table 7-3 Major Assemblies. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-3
Table 7-4 Major Cables. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-3
Table 7-5 Cabinet Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-4
Module Cover Removal/Replacement. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-5
Front Panel Removal/Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-6
Rear Panel Removal/Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-7
A1 Interconnect and Power Supply Assembly Removal/Replacement. . . . . . . . . . . . . . . . . .7-8
A2 Driver Assembly Removal/Replacement. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-9
A3 Latch Assembly Removal/Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-10
A4 Remote Local Logic Assembly Removal/Replacement. . . . . . . . . . . . . . . . . . . . . . . . . . .7-11
A5 GPIB Handshake Data Input Assembly Removal/Replacement. . . . . . . . . . . . . . . . . . .7-12
A6 Interface Assembly Removal/Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-13
A7 Switch Board Assembly Removal/Replacement. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-14
A8 Power Supply/Line Filter Assembly Removal/Replacement . . . . . . . . . . . . . . . . . . . . . .7-15
8.Servicing the Attenuator/Switch Driver
Preparing a Static-Safe Workstation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8-2
Figure 8-1 Static-Safe Work Station. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8-2
Reducing ESD Damage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8-2
Table 8-1 Static-Safe ESD Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8-3
Maintenance and Adjustments. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8-4
Fuse Removal/Replacement. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8-4
Adjustments . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8-4
Troubleshooting. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8-5
2
Contents
Isolating a Malfunction. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8-5
Parts Identification. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8-7
Figure 8-2 11713A Attenuator/Switch Driver Top Cover Removed. . . . . . . . . . . . . . . . . . .8-8
Figure 8-3 11713A Attenuator Switch Driver Bottom Cover Removed. . . . . . . . . . . . . . . .8-9
11713A Simplified Block Diagram . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8-11
Troubleshooting the A1 Interconnect Assembly. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8-13
Troubleshooting the A2 Driver Assembly. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8-14
Troubleshooting the A3 Latch Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8-16
Troubleshooting the A4 Local Logic Assembly. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8-17
Troubleshooting the A5 Handshake and Data Input Assembly . . . . . . . . . . . . . . . . . . . . . .8-18
Troubleshooting the A6 Interface Assembly. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8-20
Troubleshooting the A7 Switch Board Assembly. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8-21
The A8 power supply/line module assembly. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8-23
3
1 Agilent Technologies 11713A
Attenuator/Switch Driver
Overview In this chapter you will find:
• Function, features , and capabilitie s of the 117 13A atte nuator /switch driver
• Compatible Agilent Technologies switches and attenuators
Agilent Technologies 11713A Attenuator/Switch Driver

Overview

Overview
The 11713A attenuator/switch driver is an GPIB compatible instrument designed to provide control of up to two four-section programmable step attenuators and two microwave coaxial switches.
Features • Instrument control is accomplished manually from front panel
pushbuttons or automatically over the GPIB interface bus.
• Programing via the GPIB can be accomplished in simple one line statements.
• The attenuator/switch driver features front panel status monitoring and a self-contained power supply with current limiting.
• The instrument includes solenoid arc suppression diodes and has common terminal supplies of +24 Vdc.
• Connections are made at the rear panel to the ten pairs of current sinking transistors
Five contact pairs per set are present at each rear panel
multi-pin plug. This plug mates with the attenuator drive cable.
The remaining two contact pairs are connected to the banana
jacks mounted on the rear panel. The banana jacks are available for driving the coaxial switches or for optional applications.
Each pair of contacts is individually programmable via GPIB or
from the front panel pushbutton.
The contact pair status is indicated by the front panel LEDs in
the appropriate pushbutton
The contact pairs 9 and 0 can be used to switch between one of
two loads or to reverse the current through a single load.
Options The Option 907 front handle kit contains front panel handles that can
be added to the driver cabinet. The Option 908 rack adapter kit contains the flange and hardware for
mounting the driver in a standard rack of 48.3 cm (19 inches).
1-2 11713A Operating and Service Manual
Agilent Technologies 11713A Attenuator/Switch Driver
Overview

Compatible Attenuators and Switches

The 11713A attenuator/switch driver is designed to drive the following Agilent Technologies attenuators and switches. If you are using attenuators and switches made by another company, check their switching characteristics against those specified in Chapter 3, "Specifications".
Table 1-1 Compatible Agilent Technologies Switches
Agilent Model Number
8761B SPDT 8762,B,C,F (3331A,B,C) Termina ted SPDT 8763A,B,C (33312A,B, C) 1-port Terminated SP3T 8764A,B,C (33313A,B, C) Unterminated SP4T 8765A,B,C,D,F Opt 024 and 100(33314A,B,D) Unterminated SPDT 8766K (33366K) Unterminated SP3T 8767K (33367K) Unterminated SP4T 8768K (33368K) Unterminated SP5T 8769K (33369K) Unterminated SP6T 87104A,B, C Option 100 Terminated SP4T 87106A,B,C Option 100 Termina ted SP6T 87204A,B,C Option 100 Termina ted SP4T 87206A,B,C Option 100 Termina ted SP6T
87222C,D,E Option 100 87406B Option 100 4-port Matrix 87606B Option 1006 6-port Matrix
a
Description
2P2T Transfer
a. See caution on page 2-7.
Table 1-2 Compatible Agilent Technologies Attenuators
Agilent Model Number Description
8494G,H (33320G,H) 11 dB, 1 dB steps 8495G,H,K (33321 G,H,K) 70 dB, 10 dB steps 8496G,H (33322G,H) 110 dB, 10 dB steps 8497K ( 33323K) 90 dB, 10 dB steps 84904K,L (33324K,L) 11 dB, 1 dB steps 84906K,L ( 33326K,L 90 dB, 10 dB steps 84907K,L (33327K,L) 70 dB, 10 dB steps
11713A Operating and Service Manual 1-3
Agilent Technologies 11713A Attenuator/Switch Driver
Overview
Connecting Accessories
Table 1-3 lists accessories that are available for the 11713A attenuator/switch driver. Supplied accessor ies inc lu de two dual Viking connector cables, 12-pin conductor, 60 inches long, and a line power cable. Refer to "Replaceable Parts".
Table 1-3 Available Accessori es
Accessories Available But Not Supplied
Interconnect kit (Type-N connectors) Interconnect kit (7-mm connectors) Interconnect kit (SMA connectors) Attenuator/switch driver rear cord wrap kit
Viking connector to four ribbon cables to connect up to four switches
Viking connector to 10-pin DIP connector, 60 inches long Agilent 11764B 9-Pin conductor cable, to Viking connector (for five
switches), 60 inches long. 12-Pin conductor cable, bare wire to Viking connector
(for five switches, 60 inches long.
a
a
a
b
Agilent 11716A Agilent 11716B Agilent 11716C Agilent 5062-3995 Agilent 11761A
8120-2178
5061-0969
a.Order this kit to connect two programmable step attenuators in series. b.Order this kit to set the driver on the rear panel for bench operation.
Recommended Test Equipment
Recommended test equipment for verifying performance and f or troubleshooting the 11713A attenuator/switch driver is given in Table 4-1.
1-4 11713A Operating and Service Manual
2 Installing the Agilent
Technologies 11713A Attenuator/Switch Driver
Overview In this chapter you will learn about:
• How to unpack and check your instrument
• How to contact Agilent Technologies for service
• Operating requirements for the att enuator/switch driver
• How to set the GPIB address of your instrument
• How to connect to switches, attenuators, and relays
Installing the Agilent Technologies 11713A Attenuator/Switch Driver

Initial Inspection

1. Unpack and inspect the shipping container and its contents thoroughly to ensure that nothing was damaged during shipment. If the shipping container or cushioning material is damaged, the contents should be checked both mechanically and electr ically.
2. If the contents are damaged or defective, contact your nearest Agilent T echnologies Service and Support Office. Refer to the Se rvice and Support information in the front matter of this manual. Agilent Technologies will arrange for repair or replacement of the damaged or defective equipment. Keep the shipping materials for the carrier’s inspection.
3. If you are returning your instrument for service, repackaging the attenuator/switch driver requires or iginal ship ping containers and materials or their equivalents. Agilent Technologies can provide packaging materials identical to the original materi als. Refer to "Service and Support" o n page v for the Agil ent Technologies nearest you.

Environmental Limitations

4. Ensure that you meet the environmental limitations for storage and shipment:
Temperature -55 to 75 °C Humidity < 90% relative Altitude < 15,300 meters (50,000 feet)

Preparing for use

Bench Operation
5. For bench operation, supplied equipment includes:
plastic feet shaped to assure self-alignment of the instruments
when stacked
fold-away tilt stands to raise the front of the instrument for easier
viewing of the control panel
If the driver is to be used in a position that requires setting it on the rear panel, a kit is available (refer to Table 1-3) that includes feet with mounting screws.
Rack Mounting
6. For rack mounting the instrument, use the rack -adapter kit half-module. (Refer to Table 1-3).
To install the kit:
Remove the feet and tilt stands. Remove the side trim strips. Attach the rack mount flange and flange adapter.
2-2 11713A Operating and Service Manual
Installing the Agilent Technologies 11713A Attenuator/Switch Driver

Verifying Requirements

7. Verify that you meet the following requirements.
Power source
100 to 120 volts (+5%, –10%) from 48 to 440 Hz 220 to 240 volts (+5%, –10%) from 48 to 66 Hz, single phase.
Power consum ption
80 VA max
CAUTION Before the instrument is switched on, it must be set t o the voltage of t he
power source or damage to the instrument may result. Refer to Figure 2-1.
WARNING If this instrument is to be energized via an autotransformer for
voltage reduction, make sure the common terminal is connected to the earthed pole power source.
Power Cable
In accordance with international safety standards, a three-wire power cable is provided with this instrument. When it is connected to an appropriate ac power receptacle, the cable grounds the instrument cabinet. The type of power cable plug shipped with each instrument depends on the country of origin.
11713A Operating and Service Manual 2-3
Installing the Agilent Technologies 11713A Attenuator/Switch Driver
Figure 2-1 Line Voltage Selection
WARNING To avoid the possibility of hazardous electrical shock, do not
operate this Instrument at line voltages greater than 126.5 Vac with line frequencies greater than 66 Hz (leakage currents at these line settings may exceed 3.5 mA).
Line Voltage and Fuse Selection a. Open the cover door of the line power module. Push the fuse-pull
lever to the left and remove the fuse. (Refer to Figure 2-1.)
b. Pull out the voltage selector PC board. Position the board so that the
voltage nearest the actual fine voltage level is on the top-left side of the board.
c. Push the board firmly into the module slot. d. Push the fuse-pull back into normal right-hand position. e. Check that the fuse is the correct rating and ty pe for the input ac
line voltage. Fuse ratings for line voltages are indicated below the power module.
f. Insert the fuse into the holder.
2-4 11713A Operating and Service Manual
Installing the Agilent Technologies 11713A Attenuator/Switch Driver

GPIB Addressing

8. Use the five-bit address switches located on the rear panel of the instrument to change the GPIB address.
The 11713A has a factory preset address of decimal 28. The
switch labeled with a one is the least significant bit.
There are 32 possible addresses. Addresses 0 and 31 are
typically reserved for GPIB functions and should not be used. IEE-488.1 limits the number of addressable elements (instruments) to 16.
Figure 2-2 GPIB Address Selection
GPIB Address
Table 2-1 ASCII Address Codes to Decimal Equivalents
ASCII Address Codes - LISTEN
SP 00 0 16
!011 17
“022 18 #033 19
$044 20 %05 5 21 &06 6 22
‘077 23 (088 24 )099 25 *10: 26
+11 ; 27
,12< 28
-13= 29
.14> 30
/15? 31
Decimal Equivalents
ASCII Address Codes - LISTEN
Decimal Equivalents
11713A Operating and Service Manual 2-5
Installing the Agilent Technologies 11713A Attenuator/Switch Driver

Mating Connectors

Mating connectors used with the driver are the 24-pin GPIB male,
the 12-pin Viking Industries connector, and the banana type male.
Figure 2-3 provides interconnection data for the G eneral Purpose Interface Bus.
Figure 2-3 General Purpose Interface Bus Conn ection
Logic levels The interface bus logic levels are TTL compatible, that is, the true (1)
state is 0.0 to +0.4 Vdc and the false (0) state is +2.5 to +5.0 Vdc .
Programming
Refer to Chapter 6, "Remote Operation". and Output data format
Mating connector Agilent 1251-0293 Mating cables
available
CAUTION The maximum accumulative length of connecting cable for an interface
1
Agilent 10833A, 1 meter (3.3 ft.), Agilent 10833B, 2 meters (6.6 ft.)
Agilent 10833C, 4 meters (13.2 ft.), Agilent 10833D, 0.5 meter (1.6 ft.)
bus system is 20 meters (65.5 ft.)
A general purpose interface bus system may contain no more than 2
meter (6.6 ft.) of connecting cable per instrument.
1. The Agilent 10834A adapter is required with these cables to provide additional clearance for the cable connections.
2-6 11713A Operating and Service Manual
Installing the Agilent Technologies 11713A Attenuator/Switch Driver

Connecting to Attenuators and Switches

Use the two plug-in drive cables supplied with the driver to connect
to Attenuators and the multi-pin plugs on the rear panel.
These attenuator cables have 12-pi n connectors that may not be usable on some equipment. In these instances, the cable may be modified by removing a connector. A 9-wire cable with only one connector is also available. (Refer to Table 1-3. )
Table 2-2 describes the cable wire colors/connector pin numbers and corresponding status of the front panel pushbutton indicators.
To drive a simple non-latching relay, only one wire in a pair needs to be used for the ground. Refer to Figure 2-7 for a typical relay-driving schematic.
CAUTION Do not drive the 87222C/D/E using the S9 or S0 outputs from either t he
banana plug outputs, or from pins 3 or 4 within the Atten X and Atten Y Viking sockets, both located on the rear panel of the 11713A.
11713A Operating and Service Manual 2-7
Installing the Agilent Technologies 11713A Attenuator/Switch Driver
Table 2-2 Attenuator Cable Connector Pin N um bers and Wire Color Codes
Front Panel Pushbuttons
ATTENUATOR X, Y, and SWITCHES
Pushbutton
Number
SWITCHE
S 9 ON ATTEN X 3(S9-A) Gray 9 OFF ATTEN X 4(S9-B0 White/Red 0 ON ATT EN Y 3 (S0-A) Grey 0 OFF ATTEN Y 4 (S0-B) White/Red
ATTENUATORS
XY 1 5 OFF 5 Violet
26 OFF 7 Black
3 7 OFF 9 Orange
48 OFF 11 Brown
Pushbutton LEDs Pin Numbers Attenuator Cable Wire
a
ON 6 Yellow
ON 8 Green
ON 10 Blue
ON 12 White
Rear Panel connectors ATTEN X or Y, and
Attenuator Cable
Color Code
1 Red (24 Vdc) 2 White/Brown (Gnd)
Connect to S9 and S0 outputs
a. The ON/OFF status of the pushbutton lamps indicates which cable wire or pin on the
rear panel connector is grounded.For example, if ATTENUATOR X pushbutton 3 is illu­minated, pin 10 of the ATTEN X connector (blue wire of the cable) is grounded and pin 9
floats at a high impedance.
For the SWITCHES, if pin 3 is grounded (lamp on), pin 4 is connected to +24 V.
2-8 11713A Operating and Service Manual
Installing the Agilent Technologies 11713A Attenuator/Switch Driver

Driving Additional Coaxial Switches

Make switch connections to S0 outputs, S9 outputs, or to rear panel
ATTEN X output or ATTEN Y output. Figure 2-4 shows the rear panel connections to S0 outputs and the
GPIB command necessary to change the position of Agilent 8762 or 8765 series coaxial switches. (For GPIB command usage, refer to "Data Message Input Format")
Connections to Agilent 8762 or 8765 series coaxial switches can also be made to the ATTEN X output or ATTEN Y output (refer to Figure 2-6 for connections/pin numbers). Control can be extended to four, eight, or 10 switches if S0 and S9 outputs are als o used.
Figure 2-4 Connections for Coaxial Switches
GPIB command "B0" connects port C and port 1 and turns the front
panel LED for pushbutton 0 OFF.
GPIB command "A0" connects port C and port 2 and turns the front
panel LED for pushbutton 0 ON.
11713A Operating and Service Manual 2-9
Installing the Agilent Technologies 11713A Attenuator/Switch Driver
Figure 2-5 shows the rear panel connections and the GPIB command necessary to change the position of Agilent 8761B SPDT RF switches. (For GPIB command usage, refer to "Data Message Input Format" on page 6-3.)
Figure 2-5 Connections for Agilent 8761B SPDT Switches
GPIB command "B9" connects port C and port 2 and turns the front
panel LED for pushbutton 9 OFF.
GPIB command “A9” connects port C and port 1 and turns the front
panel LED for pushbutton 9 ON.
2-10 11713A Operating and Service Manual
Installing the Agilent Technologies 11713A Attenuator/Switch Driver
Connecting to Four-Section
To use one four-section attenuator assembly, connect an attenuator
cable to either the ATTEN X output (A6J1) or ATTEN Y output (A6J2). Connect both outputs to have more than four attenuator segments.
Figure 2-6 shows a typical connection for a programmable four-section attenuator and illustrates the connection/pin number for each connector: ATTEN X or ATTEN Y Using these same connections to Agilent 8762 or 8765 series coaxial switches, control can be extended to four, eight, or ten switches if S0 and S9 outputs are also used.
Figure 2-6 Typical Connections for Programmable Four-Section Attenuator
11713A Operating and Service Manual 2-11
Installing the Agilent Technologies 11713A Attenuator/Switch Driver

Connecting to Relays

To drive ten devices, connect attenuator cables at ATTEN X and Y
and switch cables to S9 and S0. A total of 10 relays may be on at one time if the total current is less
than 650 mA. However, since there are dual transistor and relay drivers, where one driver is on while the other is off, a total of 20 relays may be controlled.
Figure 2-7 shows the connections for a simplified relay driving circuit. This circuit is adaptable for simple non-latching relays.
CAUTION If the total continuous load current of 650 mA is exceeded, damage may
result.
Figure 2-7 Relay Driving Circuit
NOTE It is also recommended that two 28.7 V zener diodes be connected
back-to-back across the relay coils to reduce voltage transients.
Verifying Performance
9. Use the procedure for checking the electrical per formance of the attenuator/switch driver given in Chapter 4. If there is mechanical damage or a defect, or if the instrument does not pass the electrical performance test, notify the nearest Agilent Technologies office.
WARNING To avoid hazardous electrical shock, do not perform electrical
tests when there are signs of shipping damage to any portions of the outer enclosure (covers, panels, connectors).
2-12 11713A Operating and Service Manual

3 Specifications

Overview In this chapter you will find:
• Specifications which describe warranted performance standards
• Supplemental characteristics which are non-warranted parameters

Specifications

Specifications
Specifications describe war ranted performance over the temperature range 0 to +55 °C after one hour of continuous operation, unless otherwise noted.
Table 3-1 Specifications
Drive Power Supply Voltage + 24 ± 2.0 Vdc Current 1.3 A maximum peak for 1 second
0.65 A maximum continuous current Contact pairs 1 through 8, 9, and 0, maximum
current of 0.65 A continuous through one or all contacts

Supplemental Characteristics

Supplemental characteristics are intended to provide useful information. They are typical but non-warrant ed performance
parameters .
Table 3-2 Supplemental Cha racteris t ics
Power 100 or 120 Vac, +5%, –10% at 48 to 440 Hz
200 or 240 Vac, +5%, –10% at 48 to 66 Hz 80 VA maximum
Response Time 10 µs maximum for contact pairs 1 through 8
20 µs maximum for contact pairs 9 and 0
Driver Life > 2,000,000 switchings at maximum current for
contact pairs 9 and 0
Maximum Load Inductance 500 mH Maximum Load Capacitance < 0,01 µF for contact pairs 9 and 0 Net Weight 4.1 kg (9 lbs) Dimensions Height: 102 mm (4 inches including feet)
Rack height: 89 mm (3.5 inches, half-width module)
Width: 213 mm (8.4 inches) Depth: 295 mm (11.6 inches)
3-2 11713A Operating and Service Manual

4 Verification

Overview In this chapter you will find:
• Recommended test equipment
• How to verify local operation
• How to verify GPIB interface remote operation
Verification

Recommended Test Equipment

Recommended Test Equipment
Recommended Test Table 4-1 lists the test equipment required for performance verification and Equipment troubleshooting. Equipment other than the recommended models can be used provided the minimum specifications are satisfied.
Table 4-1 Recommended Test Equipment
Instrument Type Critical Specifications
Digital voltmeter 0 to + 30 Vdc T Attenuators (2 required) Programmable, 4 sections P, T Switches (2 required) + 24 Vdc drive source P, T Controller GPIB GPIB compatible P, T
a. P = Performance; T = Troubleshooting
Use
a
4-2 11713A Operating and Service Manual
Verification
Operator’s Check for Local Operation
Operator’s Check for Local Operation
This check tests initial driver turn-on and local operation.
Check that the following initial conditions are met before you proceed. Verify that driver power transformer primary is matched to the line
voltage by the Line Voltage Selection Card.
Check the driver power fuse for correct rating. Current fuse ratings
for various line voltages are given on the line power module on rear panel.
Connect the power cable to the power receptacle.
Procedure
1. Connect the equipment that is to be used in your application as shown below.
Figure 4-1 Local Operation Test Setup
11713A Operating and Service Manual 4-3
Verification
Operator’s Check for Local Operation
2. Press the LINE ON/OFF switch to ON. Switch should remain in the ON position and the green switch indicator LED should illuminate. All numbered pushbutton LEDs should be on and LOCAL LED should be on.
3. Depress the 10 numbered pushbuttons on the driver front panel. Each LED should alternate between off and on as each key is pressed.
In addition, if any equipment is connected (attenuators, relays, or switches), an audible click should be heard from the unit actuated. Pressing any numbered pushbutton should not cause any other pushbutton to change state.
4-4 11713A Operating and Service Manual
Verification

GPIB Interface Check for Remote Operation

GPIB Interface Check for Remote Operation
These procedures check the driver’s ability to process the GPIB messages described in the GPIB Message Reference Table 4-1. The checks can be performe d to ge t h e r or se pa r a te ly.
The validity of these checks are based on the following assumptions:
• The driver operates correctly with front panel control (that is, in local mode). This can be verified with the Local Operation check.
• The controller properly executes GPIB operation.
• The GPIB interface properly executes the controller's instructio ns.
If the driver appears to fail any of the GPIB checks, the validity of the above assumptions should be confirmed before servicing the driver.
If all of these checks are successful, the driver's GPIB interface is operating correctly. These procedures do not check all of the driver's program codes that can be executed. However, if the driver works correctly from the front panel, and the GPIB interface works correctly, there is a high probability that the driver will respond to all the program codes.

Initial Setup

Connect the driver to the controller with the GPIB cable as shown in the test setup below. The cable must remain connected for all tests.
Figure 4-2 GPIB Interface Check Test Setup
11713A Operating and Service Manual 4-5
Verification
GPIB Interface Check for Remote Operation
The select code of the controll er’s I/O i s ass umed to be 7. The ad dress of the driver is assumed to be 28 (factory preset). This select code-address combination (728) is not necessary for these checks to be valid. If necessary, modify the program lines presented here with the appropriate combination for yo ur instrument.

Remote Message

This check assumes that the driver is in the local state. If the driver is in the remote state (front panel REMOTE light ON), switch the driver off and then on again.
Description Command
Send remote message to the driver
REMOTE 728
Operator’s Response
Check that the driver’s REMOTE light turns ON. If it does not, the driver failed to process the remote message.

Receiving the Message

This check determines if the driver receives data messages. The driver must be in the remote state. Check that all front panel LEDs are ON except local when the driver first comes on and is set to the remote state.
Description Command
Address the driver to listen and send data messages
Operator’s Check
After the first data message is sent, check that the front-panel LEDs are OFF. After the second data message is sent, check that the LEDs are ON. If not, the driver failed to process the data message.
OUTPUT 728; "B12 34567890"
OUTPUT 728; "A12 34567890"
4-6 11713A Operating and Service Manual
Verification
GPIB Interface Check for Remote Operation
Local Lockout and Clear Lockout/ Set Local Messages
This check requires the bus to be in the remote state. Check that the driver’s REMOTE light is ON.
Description Command
Local Lockout: Send Local Lockout message. LOCAL LOCKOUT 7
Operator’s Response
Press the LOCAL key on the driver front panel. The REMOTE LED should remain ON. If not , the driver f ailed to pro cess the Local Loc kout message.
Description Command
Clear Lockout/Set Local: LOCAL 7 Send Clear Lockout/Set Local message to clear Local Lockout messages and set all devices to
Local
Operator’s Response
Check that the LOCAL light is ON and REMOTE light is OFF. Press numbered pushbutton switches and check that switches change status (LEDs on and off). If not, the driv e r fai led to process th e Cle a r Lockout/Set Local message.
11713A Operating and Service Manual 4-7

5 Local Operation

Overview In this chapter you will find:
• Function and description of front panel co ntrols, connectors, and indicators
• Function and description of rear panel controls, connectors, and indicators
Local Operation

Local Control

Local Control
Local control of the Agilent 11713A switch driver is accomplished manually from front panel pushbuttons.
Figure 5-1 and Figure 5-2 detail front and rear panel controls, connectors, and indicators.
5-2 11713A Operating and Service Manual
Figure 5-l Front Panel Features
Local Operation
Local Control
[1] ATTENUATOR X [4] LOCAL In the local mode, pushbutton switches 1,
2, 3, and 4 activate solid state switches (current sinks) to change the attenuation setting of an attenuator connected to the ATTEN X co nnector on the rear panel.
[2] ATTENUATOR Y In the local mode, pushbutton switches 5, 6, 7, and 8 activate solid state switches (current sinks) to change the attenuation setting of an attenuator connected to the ATTEN Y co nnector on the rear panel.
[3] S9 A,B In the local mode, Pushbutton switches 9 and 0 change the position of a coax ial switch connected to rear panel ba na na jacks S9 A, B and S0 A, B respective y.
Pushbutton switch that resets to the local mode when not in local lockout. LED turns on in LOCAL mode.
[5] REMOTE LED When on, the remote LED indicates that the driver is in remote control.
LINE ON LED When on, this
[6]
LED indicates that power is applied to the attenuator/switch drives.
[7] LINE OFF/ON
The line off/on button controls primary power
11713A Operating and Service Manual 5-3
Local Operation
Local Control
Figure 5-2 Rear Panel Features
[1] ATTEN X [8] Fuse
Connector to accept cable plug going to programmable attenuator X
[2] ATTEN Y
Connector to accept cable plug going to programmable attenuator Y
[3] S9 A,B
Banana jack connectors for hook up to th e coaxial switches
[4] 24 Vdc
Banana jack connector provides the common connection for the +24 Vdc used to drive the coaxial switches.
1.25 A at 110/120 Vac; 750 mA at 220/240 Vac
[9] Line Voltage Selection Card
Matches transformer primary to line voltage [10] Receptacle
Matches transformer primary to line voltage via power cable
[11] Line Power Module Assembly [12]
[5] S0 A,B
Banana jack conn ecto rs fo r hook up t o the coax ial switches
[6] Window
Safety interlock; fuse cannot be removed while power cable is connected to power receptacle
This symbol is used to point out a necessary reference for the user.
[13] GPIB Address
Switches to set the address of the driver (decimal 28 is factory preset)
[7] Fuse Pull Handle [14] GPIB Connector
Mechanical interlock; fuse must be removed before extraction of Line Voltage Selection Card
The interface connector from a source device to a listening device for the remote mode of operation
5-4 11713A Operating and Service Manual
6 Remote Operation
Overview In this chapter you will learn about:
• Remote programming capability of the attenuator/switch driver
• Function and use of GPIB commands in remote mode
• Programming information to set up desired level of attenuation

Remote Operation

11713A Operating and Service Manual 6-1
Remote Operation
The driver can be operated remotely through the General Purpose Interface Bus (GPIB).

GPIB Compatibility

Remote-only functions and all front panel functions (except the LINE switch) are programmable through the GPIB. The operator’s interface check in Chapter 4, "Verification" provides a test of the GPIB to verify that the driver can function with each of the applicable bus messages. Table 6-1 summarizes the driver’s GPIB capabilities in terms of the twelve bus messages.

Remote Mode

The driver accepts commands from the bus in the remote mode. In remote, its front panel co ntrols are d isabled ( except for t he LINE switc h and the LOCAL pushbutton). The LOCAL pushbutton can be disabled by a local lockout command. The driver can be addressed to listen and when addressed to listen, the driver will respond to the following messages: Data, Remote, Local Lockout, and Clear Lockout/Set Local.
Local-to-Remote Changes
The driver switches to remote operation upon receipt of the Remote message. The Remote message has two parts:
• Remote enable bus control line (REN) set true
• Device listen address (MLA) received once (while REN is true) The driver's outputs and all control settings remain unchanged with
the Local-to-Remote transition. The REMOTE LED is ON when the driver is in the remote state.
Local Mode
In local, the driver's front-panel controls are fully operational and the instrument will respond to the Remote message. Whether addressed or not, it will also respond to the Local Lockout and Clear Lockout/Set Local messages.
Remote-to-Local Changes
The driver returns to local control upon re ceipt of the Clear Loc kout/Set Local message (REN to local from remote when the front panel LOCAL key is pressed. With the remote-to-local transition, the driver's outputs remain unchanged.
). If the driver is not in Local Lockout, it switches
6-2 11713A Operating and Service Manual
Remote Operation

Local Lockout

The local lockout condition disables the front panel LOCAL key so that pressing the LOCAL key will not return the driver to local mode. When local lockout is in effect, the driver is returned to local under program control.
(sending Clear Lockout/Set Local) or by turning the d river LINE switch OFF and ON. However, returning to local by switching to OFF and then ON, will defeat the purpose of local lockout and the sys tem controller will lose control.
Addressing
The driver’s listen address is set by the sl ide switches on t he rear panel. Refer to "GPIB Addressing".
Data Messages
The state of the driver is controlled by data message on the GPIB. Data
messages include the driver’s GPIB progr am codes. The GPIB program codes contain information for programming the front panel functions (except for the LINE switch). The driver receives data messages when addressed to listen.
Receiving the Data Message
The driver responds to data messages when in remot e and addres sed to listen.
Data Message Input Format
The programming data string for the driver consists of a string of ASCII coded characters composed of one or both of the following fields:
[Adm] [Bdn]
where: A or a = general ON command; B or b = general OFF command
dm = any digits 0 through 9 (excluding dn) dn = any digits 0 through 9 (excluding dm)
The driver responds to an alphanumeric data string such as "A12B34" when the driver is in the remote state and addressed to listen. (The alpha characters in the data string can be either upper or lower case).
11713A Operating and Service Manual 6-3
Remote Operation
The data string is further defined as follows:
Ax Bx Ay By A/B 9 A/B 0
where: Ax Bx = data string for ATTEN X
(x can be none, one or more of the digits 1, 2, 3, or 4)
Ay By = data string for ATTEN Y
(x can be none, one or more of the digits 5, 6, 7, or 8)
Note that if Ax (Ay) uses a digit, then that digit ma y not be used in Bx (By).
A/B 9 = GPIB command for switch S9 A/B 0 = GPIB command for switch S0
The data string contains the operation selected by the user (desired contact closure, value of attenuation, etc.). The coded command from the controller is converted by the driver into control voltages. The output control voltages drive the device and illuminate front panel lights that indicate the operation selected.
Ta ble 6-2 shows the relationship between ea ch programming code and the
effect it has on the output at each connector/pin.
Table 6-3 shows the programming strings required to set up various three
and four-section attenuators to the desired levels of attenuation.

Program Execution Time

The response time for the driver is defined as the time between data valid (DAV) changing from a high to a low and the output signal at the rear panel connector changing state . Th e switching speed of the Agilent Tec hnologies programmable attenuators is 20 ms including 10 ms for the driver (ATTEN outputs). The switching speed for the RF switches is 20 to 50 ms plus 20 ms for the driver (S9 and S0 outputs).
Sending the Data Message
The driver does not have the capability of sending a data message.
Receiving the Trigger Message
The driver does not respond to the Trigger message.
Receiving the Clear Message
The remote message causes the driver to switch to the remot e mode. The remote message has two parts: the remote enable (REN) and address-to-listen. The driver’s outputs do not change with the local-to-remote transition. The REMOTE LED turns on when the dr iver is in remote mode.
6-4 11713A Operating and Service Manual
Remote Operation
Receiving the Local Message
The driver does not respond to the local message. When the driver is in Local Message remote mode, it can be returned to local mode with a Clear Lockout/Set Local message.
The front panel LOCAL key can also return the driver to local mode. However, pressing the LOCAL key might interrupt a data message to the driver. This would leave the driver in a state unknown to the controller. This situation would be undesirable and is avoided by setting the Local Lockout condition which disabl es the LOCAL key.

Receiving the Local Lockout Message

The LOCAL pushbutton switch is disabled with Local Lockout in effect. The driver is returned to local through program control by the controller.
NOTE The driver may also be returned to local by turning the driver LINE
switch off and then on, but th is defeats the p urpose of the Loc al Lockout and the system controller loses control.
Receiving the Clear Lockout/Set Local Message
The system controller sends a Clear Locko ut/Set Local mes sage to clear the Local Lockout condition and return all devices to local. No instrument settings are changed when going from local loc kout (re mote control) to local control.
Receiving the Pass Control Message
The driver does not respond to the Pass Control message because it cannot act as a controller
Sending the Require Service Message
The driver cannot send a Require Service message.
Sending the Status Byte Message
The driver does not respond to a serial poll and as a r esult does not send the Status Bit message.
Sending the Status Bit Message
Table 6-1 summarizes the driver’s GPIB capabilities in terms of the twelve bus messages.
11713A Operating and Service Manual 6-5
GPIB Message
Remote Operation
Table 6-1 summarizes the driver’s GPIB capabilities in terms of the twelve bus messages.
Table 6-1 GPIB Message Reference Table
Applicable Response Related
Commands and
Controls
a
Interface Functions
Data yes All front panel functions, except LINE
ON/OFF switch, are programmable. The front panel REMOTE indicator turns on when addressed.
Data No The driver does not respond to the
trigger message
Trigger No The driver does not respond to the clear
message.
Clear No The driver does not respond to the clear
message
Remote Yes The driver remote mode is enabled when
the REN bus line is true. However, R remains in local (i.e, the front panel is active) until it is addressed to listen the first time. The output signal is unchanged. The front panel REMOTE indicator turns on when in remote mode.
Local No The driver does not respond to the Local
message. Refer to Clear Lockout/set Local.
Local Lockout Yes LOCAL key is disabled. Only the
controller can return to the driver to local.
Clear Lockout/ Set Local
Pass Control/Take Control
Require Service
Status Byte No The driver does not respond to a parallel
Abort No The driver stops listening IFC T0,L2
Yes Driver goes to local and local lockout is
cleared when REN goes false.
No The driver has no controller capability SRQ SR0
No The driver does not respond to a serial
poll.
poll.
GET DT0
DCL, SDC DC0
REN
GTL
LLO
REN
SPE, SPD T0
TO, L2, AH1, SH0
RL1
RL1
RL1
RL1
RL1
PP0
b
b
b
b
b
a.Commands, control lines and Interface functions are defined in IEEE Std 488 and ANSI Sid MC1.1.
Knowledge of these might not be necessary if your controller’s manual describes programming in terms of the twelve GPIB messages shown in the left column.
b.The driver does not have complete RU capability since it can not process the Go-To-Local (GTL) message.
Complete GPIB capability as defined in IEEE Std 488 and ANSI Std MC1.1 is:
SH0, AH1, T0, TE0, L2, LE0, DC0, DT0, RL1, C0, SR 0, PP0.
6-6 11713A Operating and Service Manual
Remote Operation
Table 6-2 gives the relationship between each programming code and the effect it has on the output at each connector/pin.
Table 6-2 GPIB Control Statement Results
Front Panel Digits
1ATTEN X5 H L
27HL
39HL
411HL
5ATTEN Y5 H L
67HL
79HL
811HL
9S9A +24 VB
Connector Connector
Pin Numbers
6LH
8LH
10 L H
12 L H
6LH
8LH
10 L H
12 L H
Command Result AB
B+24 V
0S0A +24 V
B+24 V
H= Open (High) L =Closed (Low)
=Connected to ground
11713A Operating and Service Manual 6-7
Remote Operation
Table 6-3 shows the programming strings required to set up various three and four-section attenuators to the desired level of attenuation.
Table 6-3 Att enuation Levels and Corresponding Data Strings
0 to 11 dB or 0 to 110 dB Attenuation 4-Section Attenuator
ATT ENUATOR X (Pushbuttons)
Attenuator Section (dB)
Attenuation (in dB)
X Y ATTEN X ATTEN Y X Y ATTEN X 0 0 B1234 B5678 6 60 A24B13 1 10 A1B234 A5B678 7 70 A124B3 2 20 A2134 A6B578 8 80 A34B12 3 30 A12B34 A56B78 9 90 A134B2 4 40 A4B123 A8567 10 100 A234B1 5 50 A14B23 A58B67 11 110 A1234
ATT ENUATOR X (Pushbuttons)
Attenuator Section (dB)
Attenuation (in dB) Attenuation (in dB) GPIB Program Codes
X Y ATTEN X ATTEN Y X or Y ATTEN X 0 0 B123 B567 40 A3B12 10 10 A1B23 A5B67 50 A 13B 2 A67B5 20 20 A2134 A56B7 60 A23B1 A67B5 30 30 A12B3 A56B7 70 A 123 A567
ATT ENUATOR X (Pushbuttons)
Attenuator Section (dB)
Attenuation (in dB) Attenuation (in dB) GPIB Program Codes
X or Y ATTEN X ATTEN Y X or Y ATTEN X ATTEN Y
0 B1234 B5678 40 A24B13 A68B57 10 A1B234 A5B76 50 A124B3 A568B7 20 A4B123 A8B568 60 A234B1 A678B5 30 A14B23 A58B67 70 A1234 A5786
1 2 3 4 ATTENUATOR Y
(Pushbuttons)
1 2 4 4 Attenuator Section
(dB) Attenuation (in dB)
0 to 70 dB Attenuation 3-Section Attenuator 1 2 3 4 ATTENUATOR Y
(Pushbuttons)
10 20 40 Not
Used
0 to 70 dB Attenuation 4-Section Attenuator 1 2 3 4 ATTENUATOR X
10 20 20 20 Attenuator Section
Attenuator Section (dB)
(Pushbuttons)
(dB)
56 7 8
10 20 40 40
ATTEN Y A68B57 A568B7 A78B56 A578B6 A678B5 A5678
5
678
10
20 40 Not
ATTEN Y A7B56
56 7
10 20 20
Used
8
20
6-8 11713A Operating and Service Manual

7 Replacing Major Assemblies

Overview In this chapter you will find:
• Replaceable parts and connecting accessories
• Procedures for the removal and replacement of major assemblies in the 11713A attenuator/switch driver
Module cover Front panel Rear panel A1 interconnect and power supply assembly A2 driver assembly A3 latch assembly A4 remote local logic assembly A5 GPIB handshake and data input assembly A6 interface assembly A7 switch board assembly A8 power module/line filter assembly
Replacing Major Assemblies

Replaceable Parts

Replaceable Parts
Table 7-1 lists the accessories that are supplied with the 11713A attenuator/switch driver. Other available accessories are listed in Table 1-3. For ordering information, refer to "Service and Support" on page v in the front matter of this manual.
Table 7-1 Accessories Supplied
Item AgilentPart Number
Dual Viking connectors, 12-pin cond uctor, 60 inches long (2)
Line power cable 11713A Operating and Service Manual1 1713-90023
8120-2703
Part number will vary
a. The power cable is selected at the factory according to the voltage available in the coun-
try of destination.
Table 7-2 Line Fuse Part Numbers
Line Voltage Line Fuse Agilent Part Number
100 V/120 V Fast blow 1.25 A 2110-0094 220 V/240 V Fast blow 750 mA 2110-0063
a
7-2 11713 Operating and Service Manual
Replacing Major Assemblies
Replaceable Parts
Major assemblies and cables of the 11713A attenuator/switch driver are listed below. For ordering information, refer to "Service and Support" on page v in the front matter of this manual
Table 7-3 Major Assemblies
Assembly Reference Designator Agilent Part Number
A1 interconnect and power supply assembly 11713-60012 A2 driver assembly 11713-60016 A3 latch assembly 11713-60003 A4 remote local logic assembly 11713-60004 A5 GPIB handshake data input assembly 11713-60014 A6 interface assembly 11713-60010 A7 switch board assembly 11713-60007 A8 power module/line filter assembly 0960-0443
Table 7-4 Major Cables
Cable Reference Designator Agilent Part Number
W1 12-conductor 8120-2703 W2 12-conductor 8120-2703 W3 34-conductor 8120-2811 W4 34-conductor 8120-2811 W5 18 awg 3-conductor 8120-1378 W6 line ON/OFF (includes S1) 11713-60013
11713 Operating and Service Manual 7-3
Replacing Major Assemblies
Replaceable Parts
Table 7-5 Cabinet Parts
Reference Designation
MP1 11713-00024 1 Front panel MP2 5021-8413 1 Front panel MP3 11713-00005 1 Sub panel MP4 5021-8413 1 Rear frame MP5 11713-00012 1 Rear panel MP6 5021-5829 2 Side strut MP7 5021-9253 1 Top cover, perforated MP8 5062-3871 1 Bottom cover,
MP9 5040-8801 2 Foot (standard) MP10 5040-8803 2 Foot, non- sk id MP12 5001-0538 2 Side trim MP13 1460-1345 1 Tilt stand MP14 7120-4294 1 Label MP15 7120-1254 1 Nameplate
Agilent Part Number
Qty Description
perforated
7-4 11713 Operating and Service Manual
Replacing Major Assemblies

Module Cover Removal/Replacement

Module Cover Removal/Replacement
CAUTION This module contains components that can be damaged or destroyed by
electrostatic discharge. It should be serviced at a static-safe workstation. Refer to "Preparing a Static-Safe Work Station" in Chapter 8.
NOTE Top an d bottom covers are removed and replaced similarly.
To Remove Module Covers
1. Unscrew the one single screw in the middle of the cover flange that holds the cover to the rear panel.
2. Slide the cover toward the rear of the module.
To Replace Module Covers
1. Slide the cover toward the front of the modul e.
2. Tap on the cover to ensure that all sides fit snugly.
3. Tighten the holding screw at the rear flange.
11713 Operating and Service Manual 7-5
Replacing Major Assemblies

Front Panel Removal/Replacement

Front Panel Removal/Repla cemen t
CAUTION This module contains components that can be damaged or destroyed by
electrostatic discharge. It should be serviced at a static-safe workstation. Refer to "Preparing a Static-Safe Work Station" in Chapter 8.
To Remove the Front Panel
1. Remove the top and bottom covers. Refer to "Mod ule Cov er Removal/Replacement".
2. Remove the four corner screws that hold the front panel to the frame.
3. Disconnect the ribbon cable from the A1 interface/power supply assembly to the front panel.
4. Unscrew the five screws that hold the front panel board onto the front panel frame.
To Replace the Front Panel
1. Install the front panel board onto the front panel frame using five screws.
2. Connect the ribbon cable from the A1 interface/power supply assembly to the front panel
3. Use the four corner screws to secure the front panel board to the frame.
4. Replace the covers.
7-6 11713 Operating and Service Manual
Replacing Major Assemblies

Rear Panel Removal/Replacement

Rear Panel Removal/Replacement
CAUTION This module contains components that can be damaged or destroyed by
electrostatic discharge. It should be serviced at a static-safe workstation. Refer to "Preparing a Static-Safe Work Station" in Chapter 8.
To Remove the Rear Panel
1. Remove the top and bottom covers. Refer to "Mod ule Cov er Removal/Replacement".
2. Remove the two screws at the GPIB connector.
3. Remove the four comer screws that secure the rear panel to the module.
4. Disconnect the ribbon cable from the A1 interconnect/power supply assembly to the rear panel.
5. Separate the rear panel from the module.
To Replace the Rear Panel
1. Position the rear panel onto the rear-panel frame.
2. Connect the ribbon cable from the A1 interconnect/power supply assembly to the rear panel.
3. Secure the rear panel to the module with the four comer screws.
4. Replace the two screws at the GPIB connector.
5. Replace the top and bottom covers.
11713 Operating and Service Manual 7-7
Replacing Major Assemblies

A1 Interconnect and Power Supply Assembly Removal/Replacement

A1 Interconnect and Power Supply Assembly Removal/Replacement
CAUTION This module contains components that can be damaged or destroyed by
electrostatic discharge. It should be serviced at a static-safe workstation. Refer to "Preparing a Static-Safe Work Station" in Chapter 8.
To Remove the A1 Interconnect/Power Supply Assembly
1. Remove the top and bottom covers. Refer to "Mod ule Cov er Removal/Replacement".
2. Remove the front and the rear panel from the module.
3. Using their individual rings, pull the four boards , A2 driver, A3 latch, A4 remote local logic, and A5 GPIB assemblies from their sockets one at a time.
4. Disconnect both ribbon cables from the front and rear panels to the A1 interconnect/power supply assembly.
5. Disconnect the power transformer connector.
6. Remove the power transformer by removing the four mounting screws, nuts, and washers.
7. Unscrew the ten small screws and three large screws to separate th e A1 interconnect/power supply assembly from its frame
To Replace the A1 Interconnect/Power Supply Assembly
1. Replace the 3 large screws and the ten small screws that secure the assembly to the frame.
2. Replace the four long mounting screws with the corresponding nuts and washers to install the power transformer onto the board.
3. Connect the power transformer connector onto its matching connector.
4. Connect both ribbon cables from the A1 interconnect/power supply assembly to the front and rear panels.
5. Assemble the front and rear panel to th e frame u sing four scr ews for each panel.
6. Install all four boards, A2 driver, A3 latch, A4 remote local logic, and A5 GPIB assemblies into their sockets.
7. Replace the module covers.
7-8 11713 Operating and Service Manual
Replacing Major Assemblies

A2 Driver Assembly Removal/Replacement

A2 Driver Assembly Removal/Replacement
CAUTION This module contains components that can be damaged or destroyed by
electrostatic discharge. It should be serviced at a static-safe workstation. Refer to "Preparing a Static-Safe Work Station" in Chapter 8
To Remove the A2 Driver Assembly
1. Remove the top cover.
2. With a strong upward motion, lif t the red ring at the top edge of the A2 driver and remove the assembly.
To Replace the A2 Driver Assembly
1. Gently insert the board into the socket marked A2 with a vertical motion. Use balanced pressure from your thumbs to push the board into the socket.
2. Replace the top cover.
11713 Operating and Service Manual 7-9
Replacing Major Assemblies

A3 Latch Assembly Removal/Replacement

A3 Latch Assembly Removal/Replacement
CAUTION This module contains components that can be damaged or destroyed by
electrostatic discharge. It should be serviced at a static-safe workstation. Refer to "Preparing a Static-Safe Work Station" in Chapter 8.
To Remove the A3 Latch Assembly
1. Remove the top cover.
2. With a strong upward motion, lift the orang e ring at the top edge of the A3 latch a ssembly.
To Replace the A3 Latch Assembly
1. Gently insert the board into the socket marked A3 latch assembly with a vertical motion. Apply balanced pr essure with your thumbs to push the board into the socket.
2. Replace the top cover.
7-10 11713 Operating and Service Manual
Replacing Major Assemblies

A4 Remote Local Logic Assembly Removal/Replacement

A4 Remote Local Logic Assembly Removal/Replacement
CAUTION This module contains components that can be damaged or destroyed by
electrostatic discharge. It should be serviced at a static-safe workstation. Refer to "Preparing a Static-Safe Work Station" in Chapter 8.
To Remove the A4 Remote Local Logic Assembly
1. Remove the top cover.
2. With a strong upward motion, lift the yellow ring at the top edge of the A4 remote local logic assembly
To Replace the A4 Remote Local Logic Assembly
1. Gently insert the board with a vertical motion into the socket marked A4 remote local logic assembly. Apply balanced pressure with your thumbs to push the board into the socket
2. Replace the top cover.
11713 Operating and Service Manual 7-11
Replacing Major Assemblies

A5 GPIB Handshake Data Input Assembly Removal/Replacement

A5 GPIB Handshake Data Input Assembly Removal/Replacement
CAUTION This module contains components that can be damaged or destroyed by
electrostatic discharge. It should be serviced at a static-safe workstation. Refer to "Preparing a Static-Safe Work Station" in Chapter 8.
To Remove the A5 GPIB Assembly
1. Remove the top cover
2. With a strong upward motion, lift the green ring at the top edge of the A5 GPIB handshake data input assembly.
To Replace the A5 GPIB Assembly
1. Gently insert the board with a vertical motion into the socket marked A5 GPIB handshake data input assembly. Apply balanced pressure with your thumbs to push the board into the socket.
2. Replace the top cover
7-12 11713 Operating and Service Manual
Replacing Major Assemblies

A6 Interface Assembly Removal/Replacement

A6 Interface Assembly Removal/Replacement
CAUTION This module contains components that can be damaged or destroyed by
electrostatic discharge. It should be serviced at a static-safe workstation. Refer to "Preparing a Static-Safe Work Station" in Chapter 8.
To Remove the A6 Interface Assembly
1. Remove both covers. Uns crew one screw on eac h cover and then slide the cover toward the rear.
2. Disconnect the ribbon cable from the A1 interconnect assembly to the rear panel.
3. Remove the rear panel from the main frame by removing the four corner screws.
4. Remove the three screws that secure the A6 interface assembly.
To Replace the A6 Interface Assembly
1. Use the three screws to install the A6 interface assembly t o the rear panel.
2. Assemble the rear panel to the main frame using the four corner screws.
3. Connect the ribbon cable from the rear panel to the A1 interconnect assembly.
4. Replace the module covers.
11713 Operating and Service Manual 7-13
Replacing Major Assemblies

A7 Switch Board Assembly Removal/Replacement

A7 Switch Board Assembly Removal/Replacement
CAUTION This module contains components that can be damaged or destroyed by
electrostatic discharge. It should be serviced at a static-safe workstation. Refer to "Preparing a Static-Safe Work Station" in Chapter 8.
To Remove the A7 Switch Board Assembly
1. Remove both covers. Uns crew one screw on each cover and then slide the cover toward the rear.
2. Disconnect the ribbon cable from the A1 interconnect assembly to the rear panel.
3. Remove the front panel from the main frame by removing the four corner screws.
4. Remove the five screws that secure the A7 switch assembly to the front panel and remove the board
To Replace the A7 Switch Board Assembly
1. Replace the board using five screws to install it onto the front panel.
2. Assemble the front panel to the main frame using the four comer screws.
3. Connect the ribbon cable from the rear panel to the A1 interconnect assembly.
4. Replace the module cover.
7-14 11713 Operating and Service Manual
Replacing Major Assemblies

A8 Power Supply/Line Filter Assembly Removal/Replacement

A8 Power Supply/Line Filter Assembly Removal/Replacement
CAUTION This module contains components that can be damaged or destroyed by
electrostatic discharge. It should be serviced at a static-safe workstation. Refer to "Preparing a Static-Safe Work Station" in Chapter 8.
To Remove the A8 Power Supply/Line Filter Assembly
1. Locate the line filter unit behind the power connector where power cable plugs in.
2. Before proceeding, make sure you identif y and record the color of the transformer wire and the corresponding pin number.
3. Desolder the transformer wires and unplug the two wires from the power switch cable.
4. Desolder the ground wire on the metallic cover of the unit.
5. Remove the rear panel from the main frame by removing the four comer screws.
6. Bend the two sides of the assembly to release the lock while gently removing the line filter assembly.
To Replace the A8 Power Supply/Line Filter A ssembly
1. Push the new assembly into the opening.
2. Solder all wires. T ake care to connect the cor rect color of transformer wire to the corresponding pin number.
3. Plug in the two wires from the power switch cable.
4. Solder the ground wire to the previous location on the metallic cover of the unit.
5. Reassemble the rear panel to the module.
6. Replace the module covers.
11713 Operating and Service Manual 7-15
8 Servicing the Attenuator/Switch
Driver
Overview In this chapter you will find:
• How to prepare a static-safe work station
• How to replace the rear-panel fuse
• Troubleshooting hints for diagnosing common problems
• Assembly level troubleshooting
Servicing the Attenuator/Switch Driver
Preparing a Static-Safe Workstation
Preparing a Static-Safe Workstation
Electrostatic discharg e (ESD) can damage or destroy electronic components. All work performed on assemblies consisting of electronic components should be done at a static-safe workstation.
An example of a static-safe work station is shown below using two types of ESD protection:
conductive table mat and wris t strap combination, and conductive floor mat and heel strap combination
These methods may be used together or separately. A list of static-safe accessories and their part numbers is given on the following page.
Figure 8-1 Static-Safe Work Station

Reducing ESD Damage

To help reduce the amount of ESD damage that occurs during installation, testing, or servicing instruments use the following guidelines:
• Be sure that all instruments are prope rly earth- ground ed to preven t buildup of static charge.
• Personnel should be grounded with a resistor-isolated wrist strap before touching the center pin of any connector and before removing any assembly from the instrument.
8-2 11713A Operating and Service Manual
Servicing the Attenuator/Switch Driver
Preparing a Static-Safe Workstation
• Before connecting any coaxial cable to an instrument connector for the first time each day, momentarily ground the center and outer conductor of the cable.
• Handle all PC board assemblies and electronic components only at static-safe work stations.
• Store or transport PC board assemblies and electronic components only in static-shielding containers .
• PC board assembly edge-connector contacts may be clea ned by using a lintfree cloth with a solution of 80% electronics-grade isopro pyl alcohol and 20% deionized water. This should be performed at a static-safe work station.
Table 8-1 Static-Safe ESD Accessories
Agilent Part Number Description
9300-0797 Set includes:
3M static control mat 0.6 m x 1.2 m (2 ft x 4 ft) and 4.6 m (115 ft) ground wire (The wrist-strap and wrist-strap cord
are not included. They must be ordered separately) 9300-0865 Ground wire, 4.6 m (15 ft) 9300-0980 Wrist-strap cord 1.5 m (5 ft) 9300-1367 Wrist-strap, color black, stainless steel, without cord, has
four adjustable links and a 7-mm post-type connection. 9300-1308 ESD heel-strap (reusable 6 to 12 months)
Order the above by calling any Agilent Technologies Sales and Service Office.
11713A Operating and Service Manual 8-3
Servicing the Attenuator/Switch Driver

Maintenance and Adjustments

Maintenance and Adjustments

Fuse Removal/Replacement

User maintenance is limited to replacement of the rear panel fuse. The main ac line fuse is located on the rear panel in the line power module.
Use the following procedure to remove the fuse. Refer to Figure 3-2.
1. Remove the line power cable from its jack.
2. Open the cover door of the line power module.
3. Push the fuse-pull lever to the left and remove the fuse.
4. Replace the fuse with one of the correct rating and type for your selected input ac line voltage. Fuse ratings for line voltage s are indicated below the power module.
WARNING Be sure to select the correct fuse r a ting for the selected line voltage. Do
not use repaired fuses or short circuited fuse-holders. To do so could cause a shock or fire hazard.

Adjustments

The 11713A attenuator/switch driver has no internal electrical or mechanical adjustments.
8-4 11713A Operating and Service Manual
Servicing the Attenuator/Switch Driver

Troubleshooting

Troubleshooting

Isolating a Malfunction

Use the following procedure to help you isolate a malfunction. Recommended test equipment for troubleshooting is listed in Table
4-1.
Figure 8-2 and Figure 8-3 identify the assemblies and test points
specified in the steps below.
Procedure
1. Remove the switch driver top cover. Refer to "Module Cover Removal/Replacement".
2. Turn on the driver and verify that the proper power up status is reached. The driver should come up in local with all pushbutton LEDs on, which corresponds to the m aximum attenuation settings.
3. Measure the voltages at A1TP1 and A1TP2
TP1 = + 5.0 ± 0.5 Vdc TP2 = + 24.0 ± 1.0 Vdc
4. Verify that the + 24.0 ± 1.0 Vdc is present at the rear panel red banana jack.
5. If an attenuator or switch buzzes when actuated by the front panel pushbutton, the output transistor may be shorted, or one of the protection diodes may be shorted.
6. If one of the pushbutton LEDs does not light, determine if the pushbutton is operational by checking to see if the circuitry drives an attenuator or switch properly. If the switching click of the relay in the attenuator or switch can be heard, only the LED may be defective.
7. Check the driver in full local operation before testing in remote operation.
If a pushbutton does not actuate an attenuator section or change the position of a coaxial switch, do the following:
a. Check the output pulses of the A7 switch board assembly either
at the A7 outputs or on the multiplexer inputs of the A4 remote local logic assembly.
b. If the pulses are present when the corresponding front panel
button is pushed, check the output of the A4 multiplexers. Verify that the multiplexers are in local mode of operation.
11713A Operating and Service Manual 8-5
Servicing the Attenuator/Switch Driver
Troubleshooting
c. Verify th at the pul se reac hes the c loc k input of A3 latc h assembly
driver assembly storage register flip-flops. Verify that the Q and NQ outputs change state as the corresponding pushbutton is exercised.
d. Measure input and output levels of inverting drivers on the A2
driver assembly.
e. If all logic levels up to the Darlington transistors operate, check
for shorted or open Darlington transi st ors. For proper operation the voltage across the series base resistor to the Darlington transistors should be greater than or equal to 0.5 Vdc when the transistor is on.
8. To check the driver in remote op eration with no external loads connected, use an appropriate controller program to execute the following commands.
Command Desired Outcome
REMOTE 728 The REMOTE light turns ON. OUTPUT 728;"Al2345 678" All attenuator lights go ON except the local
light.
OUTPUT 728;"Bl2345 678" All attenuator lights go OFF except the remote
light.
NOTE The default address 728 is set at the factory. If you are using a different
address number, replace 728 in the commands above with the appropriate address of your instrument. Refer to "GPIB Addressin g".
8-6 11713A Operating and Service Manual
Servicing the Attenuator/Switch Driver
Troubleshooting

Parts Identification

Figure 8-2 and Figure 8-3 identify the location of the following assemblies and parts referred to in troubleshooting procedures.
Assembly Reference Designator Agilent Part Number
A1 interconnect and power supply assembly 11713-60012 A1C1 capacitor 1300 µF 50 V 0180-2181 A1C1 capacitor 7200 µF 15 V 0180-2221 A1U1 voltage regulator 1826-0181 A1U2 voltage regulator 1826-0559 A2 driver assembly 11713-60016 A3 latch assembly 11713-60003 A4 remote local logic assembly 11713-60004 A5 GPIB handshake data input assembly 11713-60014 A6 interface assembly 11713-60010 A7 switch board assembly 11713-60007 A8 power module/line filter assembly 0960-0443 A8C1 capacitor.1µ F 250 Vac(RMS) 0160-4065 F1 fuse 1.25 A/250 V 2110-0094 MP39 heatsink 11713-00022 T1 transformer 9100-4080 W4 34-conductor 8120-2811 W6line ON/OFF (includes S1) 11713-60013
11713A Operating and Service Manual 8-7
Servicing the Attenuator/Switch Driver
Troubleshooting
Figure 8-2 11713A Attenuator/Switch Driver Top Cover Removed
8-8 11713A Operating and Service Manual
Servicing the Attenuator/Switch Driver
Troubleshooting
Figure 8-3 11713A Attenuator Switch Driver Bottom Cover Removed
11713A Operating and Service Manual 8-9
P/O A6 Interface
A7 Front Panel
Local
Attenuator
X and Y
Switches
S1-S0
+5 V
+5 V
A5 GPIB Interface
+5 V
1-0
One-
Shots
» 17ms
LOCAL
REMOTE
DS1-DS0
LEDS
SW1 through SW0
Remote/Local Light
A4 Remote Local Logic
Remote/Local
2to1
Multiplexer
CLK 1 through CLK 0
A3 Latch
Power
Up
Driver Storage
Registers
CLK
J&K Inputs
2
A2 Driver P/O A6 Interface
Attenuator X-Drivers
ATTEN X
+24 V +24 V Common
1-8
Attenuator Y-Drivers
FF
1 through 0
Switch 9 and 0 Drivers
+5 V
9,0
ATTEN Y
S9
A
B
A
B
S0
A
B
GPIB
Address
GPIB
Connector
REN
DIO1-DIO7
RTN
IRC
DRV NRFD NDAC
+5 V
Address Switches
1 through 5
DIO1-DIO7
Schmitt
Trigger
Inverters
DAV
NRFD
NDAC
Address Comparator
Address
Data Lines NDO1-NDIO5
State
Qualifier
GPIB
Handshake
NDIO1-NDIO5
LLO
Detector
ADDR
NDIO1-NDIO5
NLLO Out
Control Lines
Handshake Out
LLO
FF
Listen
Enable
Remote/
Local Logic
FF
Listen
FF
Aor B
Clock
4 Line
to
10 Line
Demultiplexer
Aor B
FF
Remote/Local
A and B
Remote/Local
Qualifier
R/L
AorB
au401a
S1

11713A Simplified Block Diagram

Servicing the Attenuator/Switch Driver

Troubleshooting the A1 Interconnect Assembly

Troubleshooting the A1 Interconnect Assembly
The A1 interconnect assembly contains the power supply and connectors for other major assemblies.
To Troubleshoot the A1 Interconnect Assembly
1. Pull out the A2 driver, A3 latch, A4 remote local logic, and A5 GPIB board assemblies.
2. Turn the power on, and measure the voltage at TP1 on the A1
interconnect assembly. 5 Vdc ± 5% should be present.
3. Measure the voltage at TP2. 24 Vdc.± 5% should be present.
4. If either voltage is not acceptable, disconnect both ribbon cables to the front and rear panels and remeasure the voltages at TP1 and TP2.
5. If the voltages at the test points are still unacceptable , replace t he Al interconnect assembly.
Assembly Overview
The power supply is enabled when the LINE ON/OFF switch is set to the ON position. The power supply provides +5 Vdc for the logic circuits, and +24 Vd c at 0.65 0 ampe res c ontinuous to d rive att enuato rs , relays, or switches connected to the driver.
A "crowbar" circuit is incorporated that causes the primary line fuse to blow if the ac input voltage exceeds the ratings on the power mains fine module.
In addition, the Al interconnect assembly provides GPIB lines level shifting to meet the IEEE specifications for a listener device.
11713A Operating and Service Manual 8-13
Servicing the Attenuator/Switch Driver

Troubleshooting the A2 Driver Assembly

Troubleshooting the A2 Driver Assembly
The A2 driver assembly is controlled by inputs from the A3 latch assembly which is clocked by either the A7 fron t panel switc hes or from the GPIB through the A4 remote local logic assembly.
To Troubleshoot the A2 Driver Assembly
1. To determine the p roblem board assembly, test them in the following sequence: A1 interconnect, A7 switch, A5 GPIB, A4 remote local logic, A3 latch, and A2 driver.
2. If all the other board assemblies are in working condition, do the following:
a. Turn the power off. b. Op en th e to p cover. c. Pull out the A2 driver assembly and check the board and its
socket for damage. d. Reinstall the board carefully. e. Switch the power on. f. Manually press the attenuator buttons several times. The light
should come on/off each time that a button is activated.
3. If the manual verification is satisfactory, check the computer-controlled remote operation by sending these commands.
Command Desired Outcome
OUTPUT 728;”A12345 678” All attenuator lights go on except the local light OUTPUT 728;”BI2345 678” All attenuator lights go off except the remote light.
4. If the response s satisfactory, the A2 drive assembly should be operational.
5. If the response is unsatisfactory, check the other assemblies first before you decide to replace the A2 driver assembly.
8-14 11713A Operating and Service Manual
Servicing the Attenuator/Switch Driver
Troubleshooting the A2 Driver Assembly
Assembly Overview
The A2 driver assembly contains eight pairs of driver transistors and two driver relays.
One transistor of each pair, for example, Q9 and Q18, is in saturation while the other transistor is in cutoff. The saturated transistor sinks current from the + 24 Vdc supply th rough the load connect ed to the rear panel of the driver and returns it to ground. The open collector outputs are protected against spikes from inductive loads with clamp diodes to both ground and the + 24 Vdc supply. The driver transistors each have a metal tab that is attached to a common heat sink.
The relays K1 and K2 provide reversible polarity for ground and + 24 Vdc necessary to drive switches. The relay contacts are also protected against spikes from inductive loads.
11713A Operating and Service Manual 8-15
Servicing the Attenuator/Switch Driver

Troubleshooting the A3 Latch Assembly

Troubleshooting the A3 Latch Assembly
The A3 latch assembly contains 10 J-K flip-flops and a power-up circuit.
To Troubleshoot the A3 Latch Assembly
1. If any of the following conditions cannot be met, replace the A3 latch assembly.
a. Turn the power off and then turn it on again. All attenuator lights
and the local light should be on.
b. Press each attenuator switch. Only the corresponding light
should turn off. No other light should flicker or change status.
c. Send a remote command to turn the attenuators off and then on.
All attenuators should come on.
d. Send a remote command to individually control each light
OFF/ON. Only the corresponding light should turn off. No other fight should flicker or change status.
2. If these tests are inconclusive, test the Al interconnect, A7 switch, AS GPIB, and A4 remote local logic board assemblies in this sequence before replacing the A3 latch assembly.
Assembly Overview
When the driver is in the local mode, the J and K inputs to all ten flip-flops are in the logic high state. The flip-flops then operate as "toggle" flip-flops and respond to the front pan el pushbuttons. The pushbuttons function as clock pulses for the flip-flops.
In the remote state, the A4 assembly A or B flip-flop controls the J, K inputs. In the A state, the J inputs are in the logic low state and the K inputs are logic high.
The decoder on the A4 assembly generates the clock pulses for each flip-flop. In the A state, flip -flops 1 through 10 are set so the Q = 1 and NQ = 0. In the B state J inputs = 1 and K inputs = 0. Clock impulses from the decoder set flip-flops 1 through 10 on the negative edge as commanded by the bus.
The power-up circuit sets the state of all flip-flops so that the instrument powers up in the local state and with all switches set to the A (LEDs on) condition. Any atte nuators connected to the driver are set to maximum attenuation on power up.
8-16 11713A Operating and Service Manual
Servicing the Attenuator/Switch Driver

Troubleshooting the A4 Local Logic Assembly

Troubleshooting the A4 Local Logic Assembly
The A4 local logic assembly contains a remote/local flip-flop, an A or B flip-flop, a four-line to ten-line decoder, a local lockout flip-flop, listen flip-flop, 10 two-li ne to one-line multipl exers , and the required circui try to generate a clock signal to c ontrol the op eration of t he four- t o ten-line decoder .
To Troubleshoot the A4 Local Logic Assembly
1. Make sure that the A3 latch, A5 GPIB, and A7 switch board assemblies are functioning properly before troubleshooting the A4 local logic assembly.
2. If any of the following conditions cannot be met, r eplace the A4 local logic assembly.
a. Turn the driver off and then on. The local light should be on. The
local light should be on.
b. To check the driver in remote operation, use an appropriate
controller to execute the following commands.
Command Desired Outcome
REMOTE 728 LOCKOUT 7 Press the local button. Nothing happens.
LOCAL 7 The local button regains control.
OUTPUT 728;”Al2345 678” All attenuator lights should be on. OUTPUT 728;”B12345 678” All attenuat or lights should go off.
a. Factory preset GPIB address
a
The remote light is on and the local light off.
The local operation is locked out.
Assembly Overview
The remote/local flip-flop is progr ammed through the GPIB ci rcuitry on the A5 handshake and data input assembly. If the local lockout flip-flop is set to a high, the remote local flip-flop cannot be set to local via the A7 front panel LOCAL pushbutton. In all other cases the A7 local pushbutton will set the instrument to local.
In local operation, the 10 two-line to one-line multiplexers are set to propagate and invert the outputs of the monostable multivibrator s from the A7 front panel assembly. The outputs become the clocks for the A3 latch assembly flip-flops. Bus commands determine the state of the A or B flip-flop which controls the output state of the instrument.
11713A Operating and Service Manual 8-17
Servicing the Attenuator/Switch Driver

Troubleshooting the A5 Handshake and Data Input Assembly

Troubleshooting the A5 Handshake and Data Input Assembly
The A5 handshake and data input assembly provides the interface between the bus and the driv er.
To Troubleshoot the A5 Handshake and Data Input Assembly
1. Make sure that the A4 remote local logic, A3 latch, and A2 driver assemblies are functioning properly and that the A6 interface connections are correct before troubleshooting the A5 handshake and data input assembly.
2. If any of the following conditions cannot be met, r eplace the A4 local logic assembly.
a. Turn the driver off and then on. The local light must be on. The
local light must be on.
b. To check the driver in remote operation, use an appropriate
controller to execute the following commands.
Command Desired Outcome
REMOTE 728 LOCKOUT 7 Press the local button. Nothing happens. The
a
The remote light is on and the local light off.
local operation is locked out.
LOCAL 7 All attenuator lights go on.
OUTPUT 728;”A12345678” OUTPUT 728;”Bl2345 678” All attenuator lights go off.
Factory preset GPIB address
a.
All attenuator lights go on.
Assembly Overview
The three-wire handshake lines to the handshake logic circuits synchronize the operation of the driver remote operation. The lines are ready for data (NRFD), data valid (DV), and data accepted (NDAC).
The handshake logic processes the DAV signal and produces the handshake out (FIS OUT) signal passed to the bus logic and the local lockout logic. The DAV signal and the handshake enable (HS ENABLE) signal combine to output the NRFD and NDAC signals on the bus. The basic purpose of this logic is to signal the other driver circuits that the DIO lines contain a character for possible processing to interface the circuits to the GPIB three-wire handshake system.
8-18 11713A Operating and Service Manual
Servicing the Attenuator/Switch Driver
Troubleshooting the A5 Handshake and Data Input Assembly
The bus logic performs the following functions:
• accepts inputs from the data input/output (DIO) lines and the attention (ATN) signal from the interface bus. These inputs, in conjunction with the HS OUT signal, enable the remote/local logic and the local lockout logic.
• processes the attenuator/switch state codes and select codes present on the DIO lines. It then combines them with handshake out (HS OUT) and remote signals to output the NCLK signal from U11A on the A4 remote /local assembly.
• places the driver into remote operation or local operat ion on command from the interface bus in conjunction with the remote/local logic.
11713A Operating and Service Manual 8-19
Servicing the Attenuator/Switch Driver

Troubleshooting the A6 Interface Assembly

Troubleshooting the A6 Interface Assembly
The A6 interface assembly contains the GPIB address switch, the connectors for the X and Y attenuators, and the wiring for switch outputs 9 and 0.
To Troubleshoot the A6 Interface Assembly
1. To test the circuitry of the A6 interface assembly, check the driver both in manual and remote operation.
a. Manually press the attenuator buttons several times. The light
should come on/off each time that a button is activated.
b. If the manual verification is satisfactory, check the
computer-controlled remote operation by sending these commands.
Command Desired Outcome
OUTPUT 728;”Al2345 678” All attenuator lights go on except the local light. OUTPUT 728;”Bl2345 678” All attenuator lights go off except the remote light.
2. If the response is satisfactory, the A6 interface assembly should be operational.
3. If the response is unsatisfactory, refer to the troubleshooting procedures for the A2 driver assembly.
8-20 11713A Operating and Service Manual
Servicing the Attenuator/Switch Driver

Troubleshooting the A7 Switch Board Assembly

Troubleshooting the A7 Switch Board Assembly
The A7 switch board assembly (front panel) consists of 10 non-retriggerable monostable multivibrators.
To Troubleshoot the A7 Switch Board Assembly
1. Check both manual and remote operation. a. To check the driver in manual operat ion, turn the driver off and
then on. All attenuator lights and the local light should be on.
b. To check the driver in remote operation, use an appropriate
controller program to execute the following commands:
Command Desired Outcome
OUTPUT 728;”Al2345 678” All attenuator lights go on except the local light. OUTPUT 728;”Bl2345 678” All attenuator lights go off except the remote light.
2. If a function cannot be performed both manually and remotely, there may be an open line between the attenuator switc h on the A7 switc h assembly through the A1 interconnect/power supply and to the A3 latch assembly. To help pin point the problem, use the bloc k diagram on page 8-11 to furt her trace the circuitry.
Assembly Overview
The resistor co n n e ct ed to the +5 volts ke eps the input to th e multivibrators at a logic high. The voltage across the input capacitor is 0 volts when the front panel pushbutton switches are open.
When a button is pressed, the input voltage on the one shot instantaneously drops to 0 volts and starts the out put pulse. The output pulse width is controlled by a resistor and capacitor, for example R7 and C5. Nominal pulse width is 17 ms. The pulses are sent to the local/remote multiplexers on the A4 assembly and become the clock inputs for the flip-flops on the A3 latch assembly.
LEDs behind the front panel switches indicate the present out put sta te of the driver as seen by the flip-flops on the A3 latch assembly. The driver is in the A state when LEDs are on and in the B state when they are off.
11713A Operating and Service Manual 8-21
Servicing the Attenuator/Switch Driver
Troubleshooting the A7 Switch Board Assembly
LOCAL and REMOTE LEDs signify the operating mode of the driver. A reset circuit clears the LOCAL-REMOTE flip-flop on the A4 remote local assembly to the local state unless the instrument is in local lockout. The ON/OFF LED is operated from the +5 Vdc supply.
In remote operation, a bus command sets the listen flip-flop to accept further bus commands. The listen f lip-flop resets when a local command is given.
8-22 11713A Operating and Service Manual
Servicing the Attenuator/Switch Driver

The A8 power supply/line module assembly

The A8 power supply/line module assembly
The A8 power supply/line assembly houses the power receptacle, fuse, and voltage selector PC board.
To Troubleshoot the A8 Power Supply/Line Modul e Assembly
1. If the attenuator/switch driver does not power on when the line button is pressed ON, check that the fuse is good.
The main ac line fuse is located on the rear panel in the line power module. Refer to "Fuse Removal/Replacement".
2. If the fuse is not faulty, replace the A8 power supply/line module assembly.
11713A Operating and Service Manual 8-23

Index

A
address codes 2-5 address switches 2-5 adjustments 8-4 assemblies 7-3 attenuation levels 6-8 attenuator/switch driver
local operation 5-1 malfunction 8-5 remote operation 6-1 trouble shooting 8-5 wire color codes 2-8
B
bench operation 2-2
C
cables
internal 7-3
restrictions 2-6 characteristics, supplemental 3-2 clear lockout message 4-7 coaxial switches 2-9 compatible switches/ attenuators 1-3 connecting to attenuator drivers 2-7 connectors
mating 2-6
pin numbers 2-8 continuous current 2-12
D
data message input format 6-3 data messages 6-3 description, general 1-2
E
electrostatic discharge information 8-2 environmental limitations 2-2 ESD 8-2 ESD accessories 8-2
F
factory preset address 2-5 features
front panel 5-3 front panel
control 5-3
features 5-3
pushbuttons 5-3
removal 7-6
fuse
removal 8-4 replacement part numbers 7-2 selection 2-4
G
GPIB
address 2-5, 6-3 bus messages 6-6 compatibility 6-2
I
initial conditions 4-3 initial setup, remote operations 4-5 inspection 2-2 installation 2-1 isolating a malfunction 8-5
L
LEDs 5-3 line voltage selection 2-4 load current 2-12 local control 5-2 local lockout 6-3 local lockout message 4-7 local mode 6-2 local-to-remote 6-2 logic levels 2-6
M
major assemblies 7-1 malfunction 8-5 mating connectors 2-6 module cover 7-5
O
operating requirements 2-3
operator’s check
local operation 4-3 remote operation 4-5
options 1-2
P
power cable 2-3 power requirements 2-3 program execution time 6-4
R
rack mounting 2-2 rear panel
removal 7-7
receiving messages
11713A Operating and Service Manual Index 1
clear 6-4 clear lockout 6-5 data 4-6 local 6-5 lockout 6-5 pass control 6-5 set local 6-5
trigger 6-4 remote mode 6-2 remote operation check 4-5 remote-to-local 6-2 replacing major assemblies 7-1 returning your instrument 2-2
S
sending messages
data 6-4
require service 6-5
status bit 6-5
status byte 6-5 servicing your instrument 2-2 set local message 4-7 supplemental characteristics 3-2
T
test equipment 8-5 Troubleshooting 8-5
V
verification
local operation 4-3
remote operation 4-5
Index 2 11713A Operating and Service Manual
Loading...