Honeywell ToxiPro CO Reference Manual

Reference Manual
ToxiPro® Single Gas Detector
Part #13-264-N3, Version 4
THE TOXIPRO®-NGSG PERSONAL PORTABLE GAS DETECTORS HAVE BEEN DESIGNED FOR THE DETECTION AND
MEASUREMENT OF POTENTIALLY HAZARDOUS ATMOSPHERIC CONDITIONS.
IN ORDER TO ASSURE THAT THE USER IS PROPERLY WARNED OF POTENTIALLY DANGEROUS ATMOSPHERIC
CONDITIONS, IT IS ESSENTIAL THAT THE INSTRUCTIONS IN THIS REFERENCE MANUAL BE READ, FULLY
UNDERSTOOD, AND FOLLOWED.
ToxiPro®-NGSG Reference Manual
Honeywell Analytics Part Number 13-264-N3
Version 4
Copyright 2015
by
Honeywell Analytics, Inc.
Lincolnshire, Illinois 60069
All rights reserved.
No page or part of this operation manual may be reproduced in any form without written permission of the copyright
owner shown above.
Honeywell Analytics reserves the right to correct typographical errors.
Table of Contents
OPERATING TEMPERATURE RANGE ................................................................................................................................................ 5
CERTIFICATIONS ............................................................................................................................................................................. 5
WARNINGS AND CAUTIONS ............................................................................................................................................................. 6
A. Signal Words .......................................................................................................................................................... 6
B. Warnings ................................................................................................................................................................. 6
AVERTISSEMENTS ET MISES EN GARDE ........................................................................................................................................... 7
A. Marquage ................................................................................................................................................................ 7
B. Avertissements ...................................................................................................................................................... 7
1. OVERVIEW .......................................................................................................................................................................... 9
1.1 Methods of sampling ............................................................................................................................................. 9
1.2 Sensors ................................................................................................................................................................... 9
1.2.1 ToxiPro sensor ranges ........................................................................................................................................ 9
1.2.2 ToxiPro O2 sensor ranges ................................................................................................................................... 9
1.2.3 ToxiPro O2 warm-up phase ................................................................................................................................. 9
1.3 Alarm and warning logic ....................................................................................................................................... 9
1.3.1 Gas alarms ......................................................................................................................................................... 9
1.3.2 Calibration and bump test due notices ................................................................................................................ 9
1.3.3 Battery replacement Warning ............................................................................................................................. 9
1.3.4 Low battery alarms ............................................................................................................................................. 9
1.3.5 Missing sensor during startup ............................................................................................................................. 9
1.3.6 Corrupt memory ................................................................................................................................................ 10
1.3.7 Vibrating Alarm Error ........................................................................................................................................ 10
1.3.8 Heartbeat Feature ............................................................................................................................................. 10
1.4 Functions ................................................................ ................................................................ .............................. 10
1.4.1 Security beep .................................................................................................................................................... 10
1.4.2 Latching alarms ................................................................................................................................................ 10
1.4.3 Enhanced Latching Alarm ................................................................................................................................. 10
1.4.4 Silence warning alarms ..................................................................................................................................... 10
1.5 IQ Dock Compatibility .......................................................................................................................................... 10
1.6 Design components ............................................................................................................................................. 10
1.7 Standard accessories .......................................................................................................................................... 11
1.8 Value pack kits ..................................................................................................................................................... 11
2. FIELD OPERATION ............................................................................................................................................................ 11
2.1 Turning the ToxiPro on ........................................................................................................................................ 11
2.1.1 Calibration due lockout ..................................................................................................................................... 12
2.2 Backlight ................................................................................................................................ ............................... 12
2.3 Functions ................................................................ ................................................................ .............................. 12
2.4 Calibration and bump test due notices .............................................................................................................. 13
2.5 Low battery alarms .............................................................................................................................................. 13
2.6 Sensor Over Range .............................................................................................................................................. 13
2.7 Turning the ToxiPro off ....................................................................................................................................... 14
2.8 Always On Mode .................................................................................................................................................. 14
2.8.1 Turning Off in Always On Mode ........................................................................................................................ 14
2.9 Sampling ................................................................................................................................ ............................... 14
2.9.1 Sample draw kit usage ..................................................................................................................................... 14
3. CALIBRATION ................................................................................................................................................................... 15
3.1 Verification of accuracy ....................................................................................................................................... 15
3.1.1 Verifying accuracy and response: ToxiPro O2 .................................................................................................. 15
3.1.2 Verifying accuracy: ToxiPro with toxic gas sensor ............................................................................................ 15
3.2 Effect of contaminants on ToxiPro sensors ...................................................................................................... 15
3.2.1 Effects of contaminants on O2 sensors ............................................................................................................. 15
3.2.2 Effects of contaminants on toxic gas sensors ................................................................................................... 15
3.3 Fresh air/zero calibration .................................................................................................................................... 16
3.4 Functional (bump) testing (toxic sensor versions) ........................................................................................... 16
3.5 Span Calibration (toxic sensor versions) .......................................................................................................... 17
3.6 Failure to calibrate ............................................................................................................................................... 17
3.6.1 Fresh air/zero calibration failure ........................................................................................................................ 17
3.6.1.1 Causes of fresh air/zero calibration failures ........................................................................................... 18
3.6.1.2 Forced fresh air/zero calibration ............................................................................................................. 18
3.6.2 Span calibration failure ..................................................................................................................................... 18
3.6.2.1 Sensor out of range (no CAL) ................................................................................................................ 18
3.6.2.2 Causes for span cal failure ..................................................................................................................... 18
3.7 Fresh air/zero calibration in a contaminated atmosphere ................................................................................ 18
4. MAINTENANCE .................................................................................................................................................................. 19
4.1 Replacing batteries .............................................................................................................................................. 19
4.2 Replacing sensors ............................................................................................................................................... 20
4.3 Proper Cleaning ................................................................................................................................................... 20
4.4 Storage .................................................................................................................................................................. 20
5. PC-INSTRUMENT COMMUNICATIONS ................................................................................................................................. 20
5.1 Event logging ....................................................................................................................................................... 20
5.2 Instrument firmware upgrades ............................................................................................................................ 21
5.3 Initiating communications ................................................................................................................................... 21
6. EXPLODED VIEW AND BASIC PARTS LIST ............................................................................................................................ 21
APPENDICES ................................................................................................................................................................................. 22
Appendix A Sensor Cross-Sensitivity Chart ..................................................................................................................... 22
Appendix B Replacement Sensor List............................................................................................................................... 22
Appendix C Calibration Frequency Recommendation..................................................................................................... 23
HONEYWELL ANALYTICS WARRANTY GAS DETECTION PRODUCTS ................................................................................................ 24
General 24
Instrument & Accessory Warranty Periods ....................................................................................................................... 24
Sensor Warranty Periods .................................................................................................................................................... 24
Operating Temperature Range
The safe operating temperature range of the gas detector is -40 °C to +50 °C. Use of Honeywell Analytics Gas Detectors outside of the instrument’s specified operating temperature range may result in inaccurate and potentially dangerous readings.
Certifications
The ToxiPro®-NGSG carries the following certifications:
UL Class I, Division 1, Groups A,B,C,D, Temp Code T4 UL Class II, Division 1, Groups E,F,G CSA Class I, Division 1, Groups A,B,C,D, Temp Code T4 CSA Class II, Division 1, Groups E,F,G UL International DEMKO A/S 09 ATEX 0823124 II 1 G Ex ia IIC T4 IECEx UL 09.0011 Ex ia IIC T4 Ga KTL, 13-KB4BO-0147X http://www.honeywellanalytics.com/~/media/honeywell-analytics/products/toxipro/documents/kcs- certificate-
juarez-toxipro-13kb4bo0147x.pdf?la=en
The ToxiPro®-NGSG is in conformity with the following standards:
UL 913, Seventh Edition CSA C22.2 No. 0-M91, C22.2 No. 25-1966, C22.2 No. 157-92 EN 60079-0:2009, EN 60079-11:2012, EN 60079-26:2007 IEC 60079-0:2007, IEC 60079-11:2011, IEC 60079-26:2006
Warnings and Cautions
A. Signal Words
The following signal words, as defined by ANSI Z535.4-1998, are used in the ToxiPro-NGSG Operator’s Guide.
indicates an imminently hazardous situation which, if not avoided, will result in death or serious
injury.
indicates a potentially hazardous situation which, if not avoided, could result in death or serious
injury.
indicates a potentially hazardous situation, which if not avoided, may result in moderate or minor
injury.
CAUTION used without the safety alert symbol indicates a potentially hazardous situation which, if not avoided,
may result in property damage.
B. Warnings
ToxiPro personal, portable gas detectors have been designed for the detection of either oxygen deficiencies or specific toxic gas accumulations. An alarm condition indicates the presence of a potentially life­threatening hazard and should be taken very seriously.
In the event of an alarm condition it is important to follow established procedures. The safest course of action is to immediately leave the affected area, and to return only after further testing determines that the area is once again safe for entry. Failure to immediately leave the area may result in serious injury or death.
Use only Panasonic #CR2 in the ToxiPro.
The accuracy of ToxiPro instruments equipped with toxic gas sensors should be checked periodically with known concentration calibration gas. Failure to check accuracy can lead to inaccurate and potentially dangerous readings. The ToxiPro O2 should be periodically calibrated in fresh air.
A sensor that cannot be calibrated or is found to be out of tolerance must be replaced immediately. An instrument equipped with a toxic gas sensor that fails calibration may not be used until testing with known concentration test gas determines that accuracy has been restored, and the instrument is once again fit for use. Instruments equipped with an oxygen sensor that fail calibration may not be used until testing with fresh air determines that accuracy has been restored and the instrument is once again fit for use.
Do not reset the calibration gas concentration setpoints in the ToxiPro unless the concentrations of your calibration gas differ from the concentrations of the calibration gas that is normally supplied by Honeywell Analytics for use in calibrating the ToxiPro.
Use of non-standard calibration gas and/or calibration kit components when calibrating the ToxiPro can lead to dangerously inaccurate readings and may void the standard Honeywell Analytics warranty.
Honeywell Analytics offers calibration kits and long-lasting cylinders of test gas specifically developed for easy calibration. Customers are strongly urged to use only Honeywell Analytics calibration materials when calibrating the ToxiPro.
Substitution of components may impair intrinsic safety.
For safety reasons the ToxiPro must be operated by qualified personnel only. Read, understand and follow the directions set forth in this reference manual before operating the ToxiPro.
The ToxiPro has been tested for intrinsic safety in Explosive Gas/AIR (max. 21.0% O2).
Avertissements et mises en garde
A. Marquage
La norme ANSI Z535.4-1998 définit certaines conventions quant aux marquages de sécurité. Conformément à ces conventions, le marquage utilisé dans le présent document est le suivant :
situation présentant un danger immédiat qui, s'il n'est pas évité, peut entraîner des blessures
graves, voire mortelles.
situation présentant un danger potentiel qui, s'il n'est pas évité, peut entraîner des blessures graves, voire mortelles.
situation présentant un danger potentiel qui, s'il n'est pas évité, peut entraîner des blessures
légères ou sans gravité.
MISE EN GARDE (sans le symbole d'alerte) situation présentant un danger potentiel qui, s'il n'est pas évité, peut
entraîner des dommages sur les biens.
B. Avertissements
Les détecteurs de gaz portables ToxiPro sont conçus pour déceler le manque d'oxygène ou l'accumulation de gaz toxiques spécifiques. Le déclenchement d'une alarme indique la présence d'un danger potentiellement mortel et doit être pris très au sérieux.
En cas de déclenchement d'une alarme, il est impératif de suivre les procédures établies. Le comportement le plus sûr consiste à quitter immédiatement la zone concernée et à revenir uniquement si les résultats des tests complémentaires effectués démontrent que cette zone est de nouveau sûre et accessible. Le non-respect de ces instructions peut entraîner des blessures graves, voire mortelles.
Le détecteur ToxiPro doit être alimenté uniquement par une pile Panasonic CR2.
La précision des détecteurs ToxiPro équipés de capteurs de gaz toxiques doit être vérifiée régulièrement en appliquant une concentration connue de gaz d'étalonnage. Si cette vérification n'est pas réalisée, les mesures de l'appareil risquent d'être imprécises, entraînant ainsi un danger potentiel pour l'utilisateur. Le détecteur ToxiPro O2 doit être étalonné régulièrement à l'air libre.
Tout capteur ne pouvant être étalonné ou hors de sa plage de tolérance doit être remplacé immédiatement. Si l'étalonnage d'un capteur de gaz toxiques n'est pas concluant, le détecteur concerné ne doit en aucun cas être utilisé. Pour que celui-ci soit de nouveau utilisable, il convient préalablement de le tester en appliquant une concentration connue de gaz de test pour vérifier la précision de l'appareil et déterminer selon les résultats de ce test s'il est opérationnel. De même, si l'étalonnage d'un capteur d'oxygène n'est pas concluant, le détecteur concerné ne doit en aucun cas être utilisé. Pour que celui-ci soit de nouveau utilisable, il convient préalablement de le tester à l'air pur pour vérifier la précision de l'appareil et déterminer selon les résultats de ce test s'il est opérationnel.
Les seuils de concentration de gaz d'étalonnage du détecteur ToxiPro ne doivent jamais être réinitialisés sauf si la concentration du gaz d'étalonnage utilisé diffère de celle du gaz d'étalonnage normalement fourni par Honeywell Analytics pour l'étalonnage de l'appareil.
Toute utilisation de gaz d'étalonnage et/ou de matériel d'étalonnage non standard pour étalonner le détecteur ToxiPro risque d'entraîner des mesures imprécises dangereuses et d'annuler la garantie Honeywell Analytics standard. Honeywell Analytics propose des kits d'étalonnage et des bouteilles de gaz de test longue durée spécialement conçus pour faciliter l'étalonnage. Nous vous recommandons vivement d'utiliser uniquement le matériel d'étalonnage Honeywell Analytics pour étalonner le détecteur ToxiPro.
Tout remplacement des composants du détecteur risque de compromettre la sécurité intrinsèque de l'appareil.
Pour des raisons de sécurité, le détecteur ToxiPro doit être utilisé uniquement par du personnel qualifié. Les instructions du présent manuel doivent impérativement être lues, comprises et suivies avant d'utiliser l'appareil.
Le détecteur ToxiPro a fait l'objet de tests de sécurité intrinsèque avec de l'air/du gaz explosif (21,0% O2 maxi).