The operator of this instrument is advised that if the equipment is used in a manner not specified in this manual, the protection provided by the equipment may
be impaired.
CAUTION
Only qualified, service-trained personnel who are aware of the hazards involved should
remove the cover from the instrument or connect external wiring to the instrument.
WARNING
PERSONAL INJURY
DO NOT USE these products as safety or emergency stop devices, or in any other
application where failure of the product could result in personal injury.
Failure to comply with these instructions could result in death or serious injury.
This manual will explain the setup, features and operation of 3rd generation SC
Series instruments, specifically the models SC1000, SC2000, SC2001 and
SC3004.
Further information about customer specific programming and setup will be
explained on the Customer Information Sheet that is provided with every instrument.
This manual uses the following conventions to present information:
[TEXT IN BRACKETS]The label of a front panel button.
DISPLAY
->Indicates that what follows is an item from a
Chapter 1, “Introduction”, offers general information about the SC Series and
this instruction manual.
Chapter 2, “Getting Started Quickly”, provides an overview of how to get
started quickly if your instrument and transducers were ordered at the same time,
or if Signature Calibration is used.
Text that appears on the display, such as error
messages or menu items.
sub-menu, such as
SYSTEM MENU
->
DIAGNOSTICS
.
Chapter 3, “Operating Modes”, discusses the significant features of the SC
Series and operation procedures when the instrument is in the INITIALIZE, RUN,
ERROR or SETUP modes.
Chapter 4, “Chassis Models”, explains the differences between the SC1000,
SC2000, SC2001 and SC3004 chassis. Information relating to the hardware
chassis such as panel and rack mounting is given.
Chapter 5, “System Connector”, contains information about wiring to the 25-pin
System connector to access the Limit Outputs, Function Inputs and serial communications.
Chapter 6, “System Menu”, discusses the System Menu which allows you to
examine and change settings that affect the operation of the chassis.
Chapter 7, “Serial Communications”, briefly describes RS-232 and RS-485
communications. It also shows how to use the
settings and test the RS-232 or RS-485 communications.
Chapter 8, “Display Menu”, describes the Display Menu which allows you to
change what is displayed on the lower line of the SC2000 and SC2001 instruments.
Chapter 9, “Limits”, discusses how the limits operate and describes how to alter
their settings via the Limit Menus.
Chapter 10, “Strain-Gage Input Channel”, explains how to wire, configure,
SERIAL COM
Menu to examine the
SC Series Instruction Manual page 9
operate and calibrate Strain-Gage Input Channels with your transducers.
Chapter 11, “AC/AC-LVDT Input Channel”, explains how to wire, configure, operate
and calibrate AC/AC-LVDT Input Channels with your transducers.
Chapter 12, “High-Level Input Channel”, explains how to wire, configure, operate
and calibrate High-Level Input Channels with your amplified transducers, in-line amplifiers, or DC-DC LVDTs.
Chapter 13, “Relay Output Channel”, describes how a Relay Output channel can
enable additional limits in the instrument.
Chapter 14, “DAC Output Channel”, explains the configuration and operation of
additional digital-to-analog voltage or current outputs.
Chapter 15, “Split Display Virtual Channel”, shows how you can display two values
from any of the channels in the instrument at the same time.
Chapter 16, “Mathematics Virtual Channel”, describes the flexibility of customerspecific SensoCode programming.
Chapter 17, “Error Messages”, lists error messages that the instrument may display,
describes their causes and, where possible, suggests solutions.
Chapter 18, “Setup Menu Reference”, is a list of all SETUP menus and a cross-reference to related information in this instruction manual.
page 10 008-0608-00
1.2 Related Documents
Customer Information Sheet
Introduction1
Every instrument is shipped with a Customer Information Sheet which documents
important information specific to each instrument, such as:
•part number,
•date of manufacture,
•list of all installed channels and their setup information,
•customer specific SensoCode programming of Mathematics Virtual channels
and operation notes.
Communications Guide
Supplemental Instructions
The “SC Series Communications Guide”, document 008-0610-00, describes in
detail how to communicate with an SC Series instrument using RS-232 and RS-
485. Wiring diagrams, sample programs, and descriptions of each command are
included.
A printed copy of this document is available for order, or you may download it from
http://www.honeywell.com/sensotec.
If an instrument is configured with Mathematics Virtual channels, one or more sets
of Supplemental Instructions may be included. These instructions contain important information about which indicator lights, Function Input pins and/or Limit Output pins of the System connector are used by the Mathematics Virtual channel.
SC Series Instruction Manual page 11
1.3 What is the SC Series?
The SC Series of Signal Conditioners/Indicators are versatile, multi-channel devices
designed to operate with many different types of sensors. Several different chassis
types, Input channels, and Output channels are available to allow the configuration of
an SC instrument to meet a variety of measurement and control needs. The operation
of an SC instrument is based on digital technology to provide improved accuracy,
superior ease of setup, and a wealth of features.
1.3.1 Features
1.3.2 Chassis Models
1.3.3 Channel Types
Input Channels
The main features of the models SC1000, SC2000, SC2001 and SC3004 are:
•Four alarm limits (optional sixteen), with versatile setup (not available on Model
SC1000)
•Automatic setup, calibration, and scaling of strain-gage sensors through the use
of Signature Calibration
•Field selectable, digital, low-pass filtering (“damping”) on each Input channel
•Up to ±50,000 part resolution
•Field selectable five-, six - or seven-digit (9,999,999 maximum) display
•RS-232 communications standard (RS-485 optional)
•Local or remote setup using the RS-232 or RS-485 port
•Push-button on/off tare feature
Several models (i.e. chassis types) are available:
•SC1000: 1 to 4 physical channels, 3/8 DIN case, no limits or peak detector
•SC2000: 1 to 4 physical channels, 3/8 DIN case
•SC2001: 1 to 4 physical channels, portable case
•SC3004: 1 to 14 physical channels, 19” rack mount case, 1 to 3 quad-line displays
Channels can be one of three types: Input, Output, or Virtual
Input channels are hardware circuit boards with a unique channel number. Currently,
they are available for the following types of sensors:
•Strain-gage sensors, such as unamplified pressure transducers and load cells
•Sensors with voltage outputs, such as transducers with the Option 2a, 2b, 2c, 2d,
2g, 2j, 2k, 2p, 2t or 2y internal amplifier
•Sensors with current outputs, such as transducers with the Option 2j, 2k or 2n
internal amplifier
Output channels are hardware circuit boards with a unique channel number. They
include:
•Relay Output channels, which can add additional limits to the standard four (N/A
on Model SC1000).
•DAC Output channels, which provide additional voltage or current outputs.
Virtual channels are software based devices that occupy a channel number, but not a
physical slot, in an instrument.
•Split Display Virtual channels allow the displaying of any two channel’s track, peak
or valley values at the same time.
•Mathematics Virtual channels run small programs written in an interpretive
language called SensoCode. This provides great flexibility which allows the
SC Series to do many jobs which otherwise requires a personal computer or
PLC.
Introduction1
SC Series Instruction Manual page 13
1.4 What is
Signature Calibration
?
1.4.1 Overview
1.4.2 Benefits
1.4.3 Information
Stored
A small integrated circuit is located either inside the transducer, in an in-line package
between the instrument and the transducer, or in the connector of a cable. All data
necessary to set up the transducer with the instrument are stored (even linearity data),
and setup is automatic when a new transducer is connected to the instrument.
The Strain-Gage Input channel of the SC Series is designed to operate with SignatureCalibration. It will automatically set itself up with transducers which contain the memory device, but can also be set up using a front-panel interactive procedure. The Sig-
nature Calibration module can also be programmed from the instrument’s front panel.
Signature Calibration is only available with unamplified strain-gage transducers.
The benefits are:
•The transducer’s Calibration Record is always located where it is needed most...
with the transducer.
•The instrument is always set up correctly with the transducer.
•Interchanging of transducers and instruments is a quick process.
•A User Calibration Data area that can be altered by customers to fit their requirements.
•A Factory Calibration Sheet Data area, unalterable by the customer, can be copied back into the User Calibration Data.
The following information is stored inside transducers equipped with Signature Cali-bration:
•Full-scale mV/V: The full-scale millivolt-per-volt (mV/V) rating of the transducer
when its full load is applied; also called “calibration factor”.
•Shunt-Cal mV/V: The millivolt-per-volt output of the transducer when the shunt
calibration resistor is placed across its (-)SIGNAL and (-)EXCITATION leads.
•Shunt Resistance: The resistance value, in Ohms, that was used to obtain the
shunt-cal mV/V value above.
•Full-Scale Value: The full scale value of the transducer, in engineering units.
•Engineering Units: The engineering units that the transducer is calibrated in (i.e.
pounds, grams, pascals, inches of water, etc.).
•Serial Number: The serial number of the transducer.
•Excitation Voltage: The magnitude and type of signal used to excite the transducer.
•Linearization Points (optional): These can be used by an instrument using Shunt
Calibration or Millivolt-per-Volt Calibration to correct any non-linearity in the transducer and thus improve the accuracy of the system. An additional “multiple-point
calibration” can be purchased with the transducer that allows linearity correction
information to be placed into its Signature Calibration module.
page 14 008-0608-00
2.1 Locate Required Parts and Information
The following items are required to set up an SC Series instrument with your
transducer:
•SC Series instrument
•Transducers that are to be connected to the instrument
•For each transducer to be connected to the instrument, a connecting cable.
This cable will have a 12-pin, green, plastic connector on one end and the
transducer’s mating connector on the other end. Usually, this cable is ordered
along with the instrument and transducer.
If the connecting cable was not ordered with the instrument, you may need to
make this cable.
•The transducer’s Calibration Record or Certificate of Calibration.
•The Customer Information Sheet that shipped with your instrument. This
sheet describes which cards are installed in each channel.
•Power cord for the instrument.
Chapter 2
Getting Started Quickly
Getting Started Quickly2
2.2 Connect the Transducer to the Correct Channel of the Instrument
For each transducer, attach its connecting cable to the transducer, and then to the
correct 12-pin channel-connector on the instrument.
The Customer Information Sheet indicates which serial numbered transducer is to
be connected to each channel of the instrument.
If the instrument and transducer(s) were not purchased with a connecting cable,
you may need to make this cable. For the transducer’s pin connections, see that
transducer’s calibration record. For the pin connections for that channel of the
instrument, see the appropriate chapter in this manual. For example, if you wish
to wire to a Strain-Gage Input Channel see the chapter “Strain-Gage Input Channel” on page 59.
SC Series Instruction Manual page 15
2.3 Turn on the Instrument
Connect the power cord between the instrument power source and the instrument,
and turn the On/Off switch on the back of the instrument to the On position.
The instrument enters its INITIALIZE mode that lasts a few seconds per channel. As
each channel in the instrument is initialized, the transducer's serial number may be
seen on the display if the transducer has a Signature Calibration Module in it.
NOTICE
If the channel’s display flashes “
instrument has detected a transducer other than the one which was last calibrated with that channel. The instrument is prompting you to apply the
requested load to the transducer so that a “mV/V Calibration” or a “Shunt Calibration” can take place.
After making certain that the correct transducer is connected this channel,
press the [ENTER] button after you have applied the requested load to perform
the calibration.
APPLY 00000.
” (or some other load value), the
NOTICE
If the instrument displays error code 57 or 68, it has detected a transducer
other than the one which was last calibrated with that channel. Furthermore,
the channel cannot perform an automatic calibration because its calibration
type has been set to “Known Load Calibration”.
Either re-connect the correct transducer to the channel, or see “
TYPE
Menu Item” on page 67 to select another calibration type.
CALIBRATION
When the instrument enters its normal operating mode (RUN mode), you will see the
following format on the front panel display:
1• 00000. PSIG
"
where:
•Channel number: “1” is the channel number.
•Value type: The next character indicates the nature of the following value.
A blank character indicates the tracking value. The instrument is “tracking” the signal, continuously updating the display in response to the signal from the transducer.
A “
•" character indicates the peak value (highest value seen since the [CLEAR]
button was last pressed).
A “
–" character indicates the valley value (lowest value seen since the [CLEAR]
button was last pressed).
•Data value: “00000.” displays the value from the transducer in engineering units.
•Units label: “PSIG” indicates the engineering units being used (up to 4 charac-
ters).
•Tare indicator: A “
"“ symbol indicates tare is off; a “Ë“ symbol indicates tare is on.
page 16 008-0608-00
2.4 Use the SETUP Menus to Enter Transducer Information
You can skip this step if:
•You are using a transducer equipped with Signature Calibration, or
•The transducer and instrument were purchased together and set up by Sen-
sotec.
Otherwise, you must enter information about your transducer into the SETUP
menu of the channel to which it is connected. See the appropriate chapter of this
manual for that card type.
2.5 Calibrate the Transducers to Their Channels
You can skip this step if:
•You are using a transducer equipped with Signature Calibration, or
•The transducer and instrument were purchased together and set up by Sen-
sotec.
Otherwise, you must choose a calibration method (e.g. Shunt Calibration, mV/V
Calibration, or Known Load Calibration) appropriate for your application and use
the SETUP menu for that channel to calibrate the channel to the transducer. See
the appropriate chapter of this manual for that card type.
Getting Started Quickly2
2.6 The SC Series Instrument is Ready for Use
See “RUN Mode” on page 20 for information on how to operate the instrument
while it is in the RUN mode.
Apply some test stimulus on the transducer to observe changes in the display.
SC Series Instruction Manual page 17
page 18 008-0608-00
3.1 Operating Modes
3.2 INITIALIZE Mode
Chapter 3
Operating Modes
The SC Series instruments have four modes of operation:
•INITIALIZE, to test the instrument upon power up
•RUN, normal operation
•ERROR, which indicates that an abnormal situation has occurred that
stopped the operation of the instrument
•SETUP, a menu which allows setup and calibration of the chassis and its
channels
Each of these will be described in this chapter.
When the instrument is powered up or otherwise reset, it enters the INITIALIZE
mode. As the instrument enters this mode, all segments of the display and all
front panel indicator lights (if available) illuminate momentarily. Next, each channel
in the instrument is checked for proper operation.
If a problem is detected, the instrument may enter the ERROR mode.
Depending on the type of channel, other actions may occur. For example, a
Strain-Gage channel will calibrate its analog-to-digital converter and attempt to
read the transducer’s Signature Calibration information. If the transducer is
equipped with Signature Calibration, the transducer’s serial number is displayed
momentarily.
Operating Modes3
SC Series Instruction Manual page 19
3.3 RUN Mode
3.3.1 Display
After the INITIALIZE mode finishes, the instrument enters the RUN mode, its normal
mode of operation.
Model SC3004 and upper
line of SC1000, SC2000,
SC2001
Lower Line of SC1000,
SC2000, SC2001
3.3.2 [VALUE] button
The display will show a channel number on the far left, followed by the channel’s operation messages.
For example, a Strain Gage amplifier channel will use the format below:
1• 00000. PSIG
"
where:
•Channel number: “1” is the channel number.
•Value type: The next character indicates the nature of the following value.
A blank character indicates the display is being continuously updated by “tracking”
the input signal to the channel.
A “
•" character indicates the peak value (highest value seen since the [CLEAR]
button was last pressed).
A “
–" character indicates the valley value (lowest value seen since the [CLEAR]
button was last pressed).
•Data value: “00000.” displays the value from the transducer in engineering units.
•Units label: “PSIG” indicates the engineering units being used (up to 4 charac-
ters).
•Tare indicator: A “
The contents of the display’s lower line is selected with the “
"“ symbol indicates tare is off; a “Ë“ symbol indicates tare is on.
DISPLAY MENU
->
LOWER MODE
menu item as either displaying a channel or indicating the status of Limits. See “Indicator Lights” on page 21.
”
3.3.3 [CLEAR] Button
NOTICE
The [VALUE] button is not used by SC1000 instruments in the RUN mode.
After the channel number, the next characters indicate which data value for the displayed channel is shown. There are three (possibly four) data values available from
each channel:
•(blank), tracking data value
•“
•", peak data value, (highest value since the peak/valley detector was cleared)
•“
–", valley data value, (lowest value since the peak/valley detector was cleared)
•“
Û
", percentage of the Analog-to-Digital converter’s full-scale digitizing capability.
This data value is only available when the channel’s “
DIAGNOSTICS->DISPLAY ADC
”
menu item is set to “ON”.
Pressing and releasing the [VALUE] button cycles though the available sources for
data values for the displayed channel.
Pressing and releasing the [CLEAR] button will reset the peak and valley values of the
channel being monitored by the display to the track value. Additionally, any limits in
the instrument that are “latched” will be reset.
NOTICE
The [CLEAR] button is not used by SC1000 instruments in RUN mode.
page 20 008-0608-00
To clear the peak and valley values of all channels simultaneously, use the System connector’s Function Input #2 pin. See “System Connector” on page 35 for
details.
Operating Modes3
To clear the peak and valley values of a single channel regardless of which channel is being monitored by the display, use the channel’s AUX1 or AUX2 control
pins on its connector. See the particular chapter regarding that channel for
details.
3.3.4 [CHANNEL] button
Shunt Calibration Check
3.3.5 [TARE] button
3.3.6 Indicator Lights
If the configuration of the instrument contains more than one channel, the left
most characters of the display indicates which channel the display is monitoring.
Pressing and releasing the [CHANNEL] button will cause the next channel to be
displayed.
On instruments with a dual-line display, the [CHANNEL] button cannot be used to
change which channel the lower line is monitoring. That is selected with the “
->
PLAY MENU
LOWER CHANNEL
” menu item.
DIS-
If the [CHANNEL] button is held down for more than 3 seconds, the present channel’s shunt calibration value (if the channel has shunt calibration available) will be
displayed.
To reset the channel’s display to zero, press the [TARE] button while in RUN
mode. To restore the tare value, press the [TARE] button again. The “Tare” indicator will illuminate when tare is on.
To tare all channels simultaneously, use the System connector’s Function 1 Input
pin. See “System Connector” on page 35.
To tare a single channel regardless of which channel is being monitored by the
display, use the channel’s AUX1 or AUX2 control pins on its connector. See the
particular chapter regarding that channel for details.
Models SC1000, SC2000, SC2001
Model SC3004
A “"“ symbol on the right side of a channel’s display indicates tare is off for that
channel; a “
If the lower line of the display is configured with the “
menu item as “
3 and 4. A “
Ë“ symbol indicates tare is on.
DISPLAY MENU
LIMIT 01-04
"“ symbol indicates a limit is deactivated; a “Ë“ symbol indicates a limit
”, then the lower line will monitor the status of Limits 1, 2,
->
LOWER MODE
is activated. For example:
L1"L2" L3Ë L4
Ë
indicates that Limit 1 and Limit 2 are deactivated and Limit 3 and Limit 4 are activated. Since the SC1000 is not equipped with limits, the lower line of the display
will be blank. See “LOWER MODE Menu Item” on page 51.
These indicators may be overridden by the operation of a Mathematics channel in
special applications.
Front panel lights “L-1”, “L-2”, “L-3”, and “L-4”, monitor the status of Limits 1, 2, 3,
and 4. Model SC3004 instruments also include front panel lights labeled “L-5”, “L6”, “L-7” and “L-8” which monitor the status of optional Limits 5, 6, 7, and 8.
The function of the front panel lights may be overridden by the operation of a
Mathematics channel in special applications.
”
SC Series Instruction Manual page 21
3.4 ERROR mode
The instrument enters the ERROR mode when a critical error occurs that prevents the
instrument from operating. The display alternates between displaying a two-digit code
in the form “ERROR xx ON CH.yy” and a short description of the error. The first two
digits “xx” hold the error code. The last two digits, “yy” is channel number that caused
the error. For example, “
channel 1.
While the instrument is in the ERROR mode, no other operations are taking place
except for limited serial communications capabilities. See “Error Message List” on
page 152 for a list of error codes and their probable causes.
ERROR 60 ON CH.01
” indicates that error number 60 occurred on
page 22 008-0608-00
3.5 SETUP Menu mode
The SETUP Menu mode is used to display or change the settings that control the
operation of the instrument.
Operating Modes3
3.5.1 Available Menus
3.5.2 Entering and
Exiting the SETUP Menu
mode
3.5.3 Moving Through
SETUP Menus and Menu
Items
Each major function of the instrument has its own SETUP Menu. See “Setup Menu
Reference” on page 157.
To enter the SETUP Menu mode, press and hold the [UP] and [DOWN] buttons at the
same time until you see “
see the first SETUP Menu item.
The following table shows how to move though menus and menu items, and how to
display or change the setting of a particular item.
ButtonAction
[UP] or [DOWN]Scroll through available menus, sub-menus or menu items.
[ENTER]Select a menu, sub-menu or menu item
[EXIT]If changing a setting: abandons changes to a menu item
To change the setting of a menu item:
Press [ENTER] to display the current setting, which is preceded by the ‘
Press [UP] or [DOWN] until you see the setting you want. To abandon changes, press
[EXIT]. To accept the new setting and move to the next item, press [ENTER].
SETUP
” on the display. When you release the buttons, you will
Table 3-1: Navigating the Menus
otherwise: leaves the SETUP mode.
*
’ character.
3.5.4 Exiting the SETUP
Menu mode
To leave SETUP mode and return to the RUN mode, press [EXIT] at any time.
SC Series Instruction Manual page 23
page 24 008-0608-00
4.1 Introduction
Chapter 4
Chassis Models
The SC Series of Signal Conditioners/Indicators are available in several different
chassis models. In general, each chassis model operates in an identical fashion
and can be ordered with any type of Input channels, Output channels, or Virtual
channels.
Input channels and Output channels are printed circuit boards that occupy a physical slot inside the instrument’s chassis. 12-pin connectors are located on the rear
panel to connect to each Input or Output Channel. Each channel in the instrument
is identified by a number (channel 01, channel 02, etc.).
A Virtual channel exists in software only; it does not occupy a physical slot inside
the instrument’s chassis. There is no rear-panel connector for a Virtual channel.
However, a Virtual channel does require a channel number.
Chassis Models4
SC Series Instruction Manual page 25
4.2 Specifications
Model
SC1000
GENERAL
# Physical channels1 to 41 to 41 to 41 to 14
# Virtual channels8888
Case MaterialAluminumAluminumAluminumAluminum
PHYSICAL
Form factor3/8 DIN3/8 DINsuitcase2U rack
Mountingbench, panel or rackportablerack
DISPLAY
# characters/line16161620
# lines/display2224
# displays1111, 2 or 3
Display typeVacuum
Fluorescent
ENVIRONMENTAL
Storage temp.-30º C to 90º C-30º C to 90º C-30º C to 90º C-30º C to 90º C
Operating temp.5º C to 40º C5º C to 40º C5º C to 40º C5º C to 40º C
Relative Humidity80% max. for temperatures up to 31º C; decreasing linearly to 50% max. at
Otherindoor use at altitudes up to 2000 m; Pollution Degree 2; Overvoltage Cate-
Total Excitation Drive120 mA max.120 mA max.120 mA max.400 mA max.
100 to 230 VAC,
47 to 63Hz
100 to 230 VAC,
47 to 63Hz
100 to 230 VAC,
47 to 630Hz
page 26 008-0608-00
100 to 230 VAC,
47 to 63Hz
4.3 Models SC1000 and SC2000
Chassis Models4
4.3.1 Differences
4.3.2 External Arrangement
Model SC1000 instruments do not include peak/valley capture or limits features.
All other SC instruments include peak/valley capture and 4 limit (“alarm”) outputs.
The external arrangement of the AC powered SC1000 and SC2000 is given
below.
0123456789ABCDEF
0123456789ABCDEF
4.3.3 Rear Panel
4.3.4 Panel Mounting
Figure 4-1: External Arrangement of AC powered SC1000 and SC2000
The pinout for the 25-pin System connector is provided later in this chapter. The
pinouts for the individual channels are located in the chapter for that channel.
The panel cutout size conforms to the 3/8 DIN standard. Panel-Mounting Jacks
(order code AA928) are available that slide into two slots at the sides of the instrument. Use the following procedure to mount an SC1000 or SC2000 into a panel.
SC Series Instruction Manual page 27
Figure 4-2: Panel Cutout Drawing for SC1000 & SC2000 (not to scale)
Step 1: In the panel or rack, cut a hole as shown above. The panel may be up to 1/4”
in thickness.
Step 2: Use an 0.062" Allen wrench to remove two set screws that hold the Panel-
Mounting Jacks to the case. To do this, insert the wrench into the side slots at
the rear. Remove the screws completely.
Step 3: Remove the Panel-Mounting Jacks by sliding them toward the rear. If the jacks
don't slide easily, tap them gently.
Step 4: Put the instrument through the hole in the panel.
Step 5: Reinsert the Panel-Mounting Jacks into the slide slots. Slide them as far
toward the panel as possible.
Step 6: Reinsert the setscrews and tighten them. This will force the Panel-Mounting
Jacks toward the rear side of the panel, drawing the instrument tightly into
place.
4.3.5 Rack Mounting
4.3.6 Bench Mounting
4.3.7 Case Removal
page 28 008-0608-00
A Rack Mounting Kit (order code AA934) is available for mounting a single SC1000 or
SC2000 into a 19”, 2U rack. It includes the Panel-Mounting Jacks described above as
well as the 19” rack panel.
If bench mounting the instrument, you may want the optional carrying handle/bench
stand (order code AA926).
WARNING
Disconnect the power cord and all cables from the instrument before attempting to remove the case.
Failure to comply with these instructions could result in death or serious
injury.
CAUTION
Use a #0 Phillips screwdriver on the black screws to avoid damaging them.
Failure to comply with these instructions may result in product damage.
Chassis Models4
Removal of rear panel
Remove of case from entire circuit
board assembly
4.3.8 Internal Arrangement
The rear panel must be removed in order to install or remove channels.
Step 1: Remove the four black Phillips-head machine screws that secure the back
panel of the instrument to the case. These screws are located on the rear of
the case, one at each of the four corners.
Step 2: Remove the cable shield connection screw to allow the installation or
remove of channels per “Input or Output Channel Installation Procedure” on
page 42.
Step 1: Remove the two Phillips-head screws that secure the front panel.
Step 2: Remove the front panel and disconnect its two connecting cables.
Step 3: Remove the four black Phillips-head machine screws that secure the back
panel of the instrument to the case. These screws are located on the rear of
the case, one at each of the four corners. Do not remove the cable shield
connection screw in the center.
Step 4: The circuit boards will slide out of the rear of the case as a unit.
The figure below shows the names and locations of the printed circuit boards.
Front Panel/Display Board
Assembly
Microprocessor
Board
Power Supply
Board
SensoBus Backplane
Board
Up to 4 Input or Output
Channel boards
(1 shown)
Rear Panel
Figure 4-3: Internal Arrangement of SC1000 and SC2000
Below is a description of each printed circuit board.
•The SensoBus Backplane Board serves as the connection between all
SC Series Instruction Manual page 29
boards in the instrument.
•The Front Panel/Display Board Assembly contains the display and all front panel
controls.
•The Power Supply Board contains the +15V, -15V and +5V power supplies.
•The Microprocessor Board contains the microprocessor, ROM software storage
chip, and the System connector.
•The Hardware Input/Output Channel Boards plug into the remaining four slots
of the SensoBus Backplane Board.
4.3.9 Cleaning
4.3.10 Vehicle Power
Option
4.3.11 Fuse
Replacement
Turn off the instrument and unplug all connectors. Use a soft cloth or tissue and a
mild cleaner. Do not use liquid or aerosol cleaners. Do not allow any cleaner inside
the instrument.
Model SC1000 and SC2000 instruments are available with a vehicle power option for
operation with batteries and linear DC power supplies. See “Specifications” on
page 26 voltage and power requirements.
NOTICE
Due to the momentary startup inrush current of the instrument’s power supply,
the use of switching power supplies with the SC are not recommended.
The power-line fuses of AC-powered instruments are located within the instrument's
power entry module on the rear panel. Use two 2A, 250V fast-blow fuses (p/n 0293026-00).
page 30 008-0608-00
Loading...
+ 144 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.