Honeywell M5-C01 Instruction Sheet

GasAlertMicro 5/PID/IR Battery Charger [English] Instruction Sheet
D5688/2 Page 1
Trimmed and folded dimensions: 4.125 (w) x 5.875 (H)
BW TECHNOLOGIES
BY HONEYWELL
GasAlertMicro 5/PID/IR Battery Charger
Introduction
The GasAlertMicro 5/PID/IR Battery Charger (“the charger”) charges the GasAlertMicro 5, GasAlertMicro 5 PID, and the GasAlertMicro 5 IR rechargeable battery packs.
Standard GasAlertMicro 5/PID/IR Battery Charger kit includes:
GasAlertMicro 5/PID/IR Battery Charger unit
Universal power adapter
Power cord
Wall mounting bracket
Instruction sheet
Table 1. Order Number
Model No. Description Qty
M5-C01* Cradle charger 1
*
Add suffix (-UK) for United Kingdom mains plug, (-EU) for European mains
plug, (-AU) for Australian mains plug.
D5688/2 iERP: 124011 © BW Technologies 2006, all rights reserved. Printed in Canada. All product names are trademarks of their respective companies.
a Safety Information - Read First
For indoor use only. Warning: Substitution of components may impair Intrinsic Safety of the
detector being charged.
Do not charge the battery pack with any other unapproved charger. Do not
use the charger, change the batteries, or charge the batteries in a hazardous location. Failure to observe these precautions will impair the Intrinsic Safety of the detector and can lead to fire and/or explosion.
Do not expose the charger to electrical shock or severe continuous
mechanical shock.
Ensure the battery pack is locked in place prior to operating the detector. If the charger is damaged or parts are missing, contact
BW Technologies by Honeywell immediately.
To eliminate the risk of electrical shock, unplug the charger whenever
cleaning or other maintenance is required.
Avoid touching the charging cradle contact pins. Use only BW approved batteries. Do not use alkaline or other non-
rechargeable batteries with this charger.
The detector must be deactivated before charging. Do not calibrate the detector during or immediately after charging is
complete.
Table 2. International Symbols
Symbol Meaning
n
Approved to both U.S. and Canadian Standards by the CSA International
a
Caution – risk of danger
dc current
Parts of the GasAlertMicro 5/PID/IR Battery Charger
1 Retaining hooks 2 Charging cradle 3 Battery connectors (contact pins) 4 Charger status LED 5 Mounting access hole 6 Retaining base
GasAlertMicro 5/PID/IR Battery Charger [English] Instruction Sheet
D5688/2 Page 2
Trimmed and folded dimensions: 4.125 (w) x 5.875 (H)
Mounting the Charger
The charger can be easily mounted onto a secure surface. To mount the charger, complete the following:
1. Select a secure location for the charger to be mounted.
2. Measure a distance of 2.36 in. (60 mm) for the retaining screws. Refer to Figure 1.
Figure 1. Mounting Bracket
1
Mounting bracket
2
Philips retaining screws (2)
3. Using the screws provided, mount the bracket.
4. Attach the charger to the bracket by sliding the charger mounting ports down onto the mounting bracket. Refer to Figure 2.
Figure 2. Attaching the Charger Base
1
Charger base
2
Mounting bracket
3
Mounting ports
4
Retaining screw holes
5. Press the charger firmly against the bracket until the release tabs click.
Charging
a Warning
To charge successfully, the temperature must be between 50°F to 95°F (10°C to 35°C). If the detector emits a low battery alarm, charge the battery immediately.
When charging a new battery for the first time, fully charge the battery pack before use (approximately 3.5 hours).
Note
If a fully charged battery is inserted into the charger, the charger status LED lights red for 6 to 10 minutes and then changes to green (does not damage the battery).
The detector can remain in the cradle after charging is complete.
Charging the battery pack in temperatures above 86°F (30°C) greatly reduces the life of the battery pack.
The battery pack may be hot immediately after charging. Approximately three charges are required to achieve a full charge
capacity for a new battery pack.
Charging the Battery Pack with the Detector
To charge the detector, complete the following:
1. Insert the power cord into the charger port. Refer to Figure 3.
Figure 3. GasAlertMicro 5/PID/IR Battery Charger (Back)
1
Back of the charger
2
Power jack port
3
Charger power jack
2. Plug the power cord into a Vac outlet. The charger performs a self-test. The charger LED lights red then
green during the test. When the self-test is complete, the LED deactivates.
3. Deactivate the detector.