Every effort has been made to ensure the accuracy in the
contents of our documents, however, Zellweger Analytics
Limited can assume no responsibility for any errors or
omissions in our documents or their consequences.
Zellweger Analytics Limited would greatly appreciate being
informed of any errors or omissions that may be found in the
contents of any of our documents and to this end we include
the following form for you to photocopy, complete and return to
us so that we may take the appropriate corrective action.
MAN0597ISSUE 09 - 12/05
2302M5030
HELP US TO HELP YOU
To:Marketing Services,
Zellweger Analytics Limited,
Hatch Pond House,
4 Stinsford Road,
Nuffield Estate,
POOLE. Dorset.
BH17 0RZ.
United Kingdom.
I suggest the following corrections/changes be made to:
Chapter ........... Section ...........
Marked up copies attached (as appropriate):Yes / No
Please inform me of the outcome of this change: Yes / No
For Marketing Services, Zellweger Analytics Limited:
Actioned By:Date:
Response:Date:
3
MAN0597ISSUE 09 - 12/05
2302M5030
GENERAL STATEMENT OF LIMITED WARRANTY
GENERAL STATEMENT OF LIMITED WARRANTY
COPYRIGHT
This publication contains information partly derived from
proprietary data of Zellweger Analytics. The main objective
of this information is to assist in the operation and
maintenance of the instrument described herein. The
publication of this information does not convey any right to
reproduce or use the information for any purpose other
than in the operation or maintenance of the equipment
described herein.
Zellweger Analytics shall not be liable for any incidental or
consequential damages in connection with any deletions,
errors or omissions in this Manual.
All products are designed and manufactured to the latest
internationally recognised standards by Zellweger Analytics
under a Quality Management System that is certified to
ISO9001. As such Zellweger Analytics warrants its products
against defective parts and workmanship as detailed below.
1.Zellweger Analytics Limited will repair or (at its option)
replace any Contract Goods which are or may become
defective under proper use within the following timescales
from Delivery where such defects solely arise from faulty
design materials or workmanship (Other than a design
made furnished or specified by the Buyer) provided such
defective goods are returned by the Buyer within the
requisite time limits set out below at its expense to
Zellweger Analytics Limited's premises, which returned
goods must detail the Service Event Number (SE#) clearly
on the package and the Buyer shall obtain an SE# by
telephoning Zellweger Analytics Limited Service
Department.
4
MAN0597ISSUE 09 - 12/05
2302M5030
GENERAL STATEMENT OF LIMITED WARRANTY
2.All products 12 months from delivery (date of original
despatch to the buyer) with the following exceptions:
Impact multigas detector
24 months from delivery
Impact disposable OFCH cartridge
12 months from date of installation into Impact
instrument provided installation takes place
before the stated ‘INSTALL BY’ date on the
cartridge packaging
3.Any such defective goods returned in accordance with the
provisions of this clause must be accompanied by a
detailed report stating the nature of the defect with the
SE#. If no such report is included then Zellweger Analytics
Limited reserve the right to charge a £50 (Sterling)/$75
(US dollars) investigative fee before any repair or
replacement is carried out.
4.The warranties set out in this clause are non pro rata, i.e.
the initial warranty period is not extended by virtue of any
works carried out there under.
5.Where it is impractical to return any defective goods to
Zellweger Analytics Limited’s premises then Zellweger
Analytics Limited should be notified by writing within the
relevant warranty period referred to above and upon being
so notified Zellweger Analytics Limited will despatch a
service engineer to site on a day rate basis (details
available upon request) and if such goods are found to be
defective hereunder then they will be repaired or replaced
free of charge but day rate charges for the service
engineer will apply.
6.Subject to the provisions of this clause Zellweger Analytics
Limited shall not be liable for any loss or damage
whatsoever or howsoever occasioned which may be a
direct or indirect result of the use or operation of the
Contract Goods by the buyer or any Party.
5
MAN0597ISSUE 09 - 12/05
2302M5030
GENERAL STATEMENT OF LIMITED WARRANTY
7.This warranty covers instrument and parts sold to the
Buyer only by authorised distributors, dealers and
representatives as appointed by Zellweger Analytics
Limited.
8.This warranty does not cover consumable items or items
likely to wear in normal operation including but not limited
to dry-cell batteries, filters and fuses.
9.The liability of Zellweger Analytics Limited under this
clause shall be in lieu of any warranty or conditions implied
by law as to the quality or fitness for any particular purpose
of the Contract Goods and (save as provided in this
clause) Zellweger Analytics Limited shall not be under any
liability whether in contract tort or otherwise in respect of
any defects in the Contract Goods or for any injury (other
than personal injury caused by Zellweger Analytics
Limited’s negligence as defined by Section 1 of the Unfair
Contract Terms Act 1977) damage or loss resulting from
such defects or from work done in connection therewith.
10. This warranty supersedes all existing warranty statements
and Zellweger Analytics Limited make no other warranty
expressed or implied except as stated above.
11. If a claim is made against Zellweger Analytics Limited in
respect of Liability under the Consumer Protection Act
1987 in circumstances where a like claim could have been
made against the Buyer in contract the buyer should
indemnify Zellweger Analytics Limited in full against all
damages costs and expenses that may become payable
as the result of such claim.
6
MAN0597ISSUE 09 - 12/05
2302M5030
TOTAL ENVIRONMENTAL SOLUTIONS
TOTAL ENVIRONMENTAL SOLUTIONS
Ensure that you read and understand these Operating
Instructions BEFORE installing or operating any part of the
equipment.
Please pay particular attention to the Safety Warnings.
The dry cell battery holder part no. 2302B0371 or
rechargable battery pack part no. 2302B0842 must not be
removed, replaced or recharged in the hazardous area.
Only the following alkaline dry cell batteries must be used
in the dry cell holder part no. 2302B0371.
Duracell MN1500 or Energizer Intelligent E91.
Rechargeable cells must not be used in the dry cell battery
holder part no. 2302B0371.
WARNINGS
Do not mix rechargeable battery packs and dry cell battery
packs.
The instrument must be serviced only by qualified
personnel trained by Zellweger Analytics or by a Zellweger
Analytics Appointed agent. Servicing must be carried out
only in a non-hazardous area.
The Impact must not be used in an oxygen enriched
atmosphere.
Refer to Section 4. OPERATION for details of restrictions
of use of the IMPACT/IMPACT PRO
The Flammable Sensor requires an oxygen content of
greater than 10%v/v to operate reliably. In circumstances
where the oxygen content of the sample is less than
10%v/v, the reading displayed on the Flammable Channel
should be regarded as suspect. In this situation, a Warning
54 (Low O
- Flam Inaccurate) will be generated on the
2
instrument.
7
MAN0597ISSUE 09 - 12/05
2302M5030
TOTAL ENVIRONMENTAL SOLUTIONS
WARNINGS
The Flammable Sensors sensitivity can be adversely
affected by exposure to certain substances (silicon and
sulphur compounds are examples). Every effort should be
made to avoid exposure to these substances. Following an
H
S alarm or repeated gassing with H2S a check should
2
be performed on the Flammable sensor to verify its
accuracy and a calibration performed if necessary.
If -0.0 %V/V for the CO2 sensor channel is permanently
displayed, a sensor zero calibration needs to be performed
in clean air. The alarm level A1 for the CO
exceed 0.5 %V/V.
range must not
2
Do not calibrate the CO
channel if CO2 is indicated in clean
2
air after the apparatus has been exposed to CO
concentrations above the upper limit of the measuring
range. In this case leave instrument in clean air overnight
before calibration.
The calibration interval for the CO2 sensor shall not exceed
1 month.
Dispose of the spent cartridge and its packaging in
accordance with local regulations. Do not dispose of in fire.
IMPORTANT NOTICE
Zellweger Analytics Limited can take no responsibility for
installation and/or use of its equipment if this is not done in
accordance with the appropriate issue and/or amendment
of the relevant manual.
The user of this manual should ensure that it is appropriate
in all detail to the exact equipment to be installed and/or
operated. If in doubt, the user should contact Zellweger
Analytics Limited for advice.
2
If further details are required which do not appear in this
manual, contact Zellweger Analytics Limited or their agent.
8
MAN0597ISSUE 09 - 12/05
2302M5030
Note: The failure to observe and abide by the above Warnings
and Cautions may render void the intrinsic safety
approval of the IMPACT/IMPACT PRO, and may remove
any right of claim against Zellweger Analytics relating to
product liability or consequential damage to any third
party.
9
MAN0597ISSUE 09 - 12/05
2302M5030
CONTENTS
SectionPage
Contents
1.INTRODUCTION16
1.1Intended use17
1.2Product overview18
2.GETTING STARTED20
2.1How to turn Impact/Impact Pro on and off20
2.2Inserting the cartridge21
2.3Charge for first use22
2.4Dry-cell Battery insertion25
2.5Sampling26
3.INSTRUMENT START-UP28
3.1Instrument information28
3.2Selecting location / operator29
3.3Fresh Air Auto zeroing sensors29
3.4Sensor Warm-up Phase30
4.OPERATION31
4.1Monitoring Condition31
4.1.1Display Screens31
4.1.2Other Displayed Symbols33
4.1.3Confidence Signal34
4.1.4Go/No Go Option34
4.2Atmospheric alarm conditions35
4.3Alarm condition36
4.3.1Latching Alarms (default)36
4.3.2Non-latching Alarms36
4.3.3Vibrating alarm (where fitted)36
4.3.4Resetting an Alarm37
4.4Fault and Warning Condition37
4.4.1Warning37
4.4.2Fault condition38
4.5Safelink38
4.5.1What is Safelink?38
4.5.2Using Safelink39
4.6PUMP (IMPACT PRO ONLY)42
4.6.1Removal of the Pump Adaptor47
10
MAN0597ISSUE 09 - 12/05
2302M5030
CONTENTS
SectionPage
4.7Menus47
4.7.1Flammable Gas Selection48
4.7.2Operator49
4.7.3Calibration50
4.7.4Instrument Details50
4.7.5Safelink51
4.7.6Language51
4.8Datalogging51
4.8.1Installing the PC Software52
4.8.2Event Datalogging52
4.8.3Gas Datalogging (optional)53
4.8.4Output Formats53
4.9Calibration54
4.9.1Contaminants54
4.9.2Flow Calibration – Instrument56
4.9.3Flow Calibration – PC61
4.9.4Enforcer Calibration62
5.FAULT FINDING AND PROCEDURES64
6.ACCESSORIES65
6.1Basestation65
6.2Basestation Plinth66
6.3Basestation PSU Link Cable66
6.4Offline Trickle Charger67
6.5Power supply for Basestation and Trickle
Charger67
6.6Enforcer68
6.7Enforcer gas cylinder68
6.8Flow adaptor69
6.9Metal Belt Clip69
6.10 Instrument Webbing Clip69
6.11 Body Harness Kit70
6.12 Pump Adaptor Kit70
6.13 10m Sample Tube Kit70
6.14 Earpiece71
6.15 Basestation PC Link Cable71
11
MAN0597ISSUE 09 - 12/05
2302M5030
CONTENTS
SectionPage
6.16 Safelink Cable Assembly71
6.17 Safelink Cable Restraint Clip72
6.18 Hand Aspirator Kit72
6.19 Sensing Probe 1m73
6.20 Ball float73
7.ROUTINE MAINTENANCE74
7.1Cleaning74
7.2Filters74
7.3Battery charging / replacement75
7.3.1Rechargeable Battery Pack76
7.3.2Dry cell77
8.ROUTINE SERVICING78
8.1Serviceable Cartridges78
9.SPARE PARTS96
10. GLOSSARY97
11. APPENDIX A101
11.1 Warning Codes101
11.2 Fault Codes104
12. APPENDIX B106
12.1 Warranty106
12.2 Certification Approvals106
12.2.1 DMT Test Report107
12.2.2 Safelink Connection Drawing110
12.3 Technical Specifications111
12.3.1 Instrument Specification111
12.3.2 Charger Specifications112
12.3.3 Specific Speed of Response Data
(Rising Gas Levels)113
12.3.4 Specific Recovery Time Data
(Falling Gas Levels)114
12
MAN0597ISSUE 09 - 12/05
2302M5030
INSTRUMENT LABELS
INSTRUMENT LABELS
EQUIPMENT LABEL
An explanation of the information on the equipment label is
shown below.
Manufacturers Trademark & Address
Poole
BH17 0RZ, UK
Impact Unit Part No. 2302B1000XXX
User Cautions
CAUTION:
Comms Connection only to other certified
Impact detectors in hazardous area
Read and understand instruction manual before
operating or servicing
Replace/Recharge batteries only in Non Hazardous area
Refer to Manual for battery types
CENELEC (ATEX) CERTIFICATION LABEL
An explanation of the information on the Cenelec (ATEX)
certification label is shown below.
Product Name
Explosion Protection
Mark and Equipment
Group Category
Certified Ambient
Temperature Range
EU Explosive
Atmosphere
Symbol and
Certification Code
as EN50014: 1992
DMT Approval
CE Mark Conforms to all
EC Directives
Certification No.
Identification
No. of ATEX
Notified Body
Australian
Surface
Approval
Australian
Mines
Approval
Brazilian
Approval
13
MAN0597ISSUE 09 - 12/05
2302M5030
INSTRUMENT LABELS
This instrument has been assessed by DMT for performance
of Oxygen, Methane, Propane, Carbon Monoxide, Hydrogen
Sulfide and Carbon Dioxide channels.
The label marking indicates this:
DMT 02 ATEX G 001
PFG Nr. 41300502
The instrument has been tested in accordance to the following
European Standards.
•EN50054 & EN50057: 1998 for Combustible Gases
(Methane and Propane).
•EN61779-1: 2000 for Combustible Gases (Methane and
Propane) and EN61779-4: 2000.
•EN50104: 1998 for the measurement of Oxygen.
•EN45544-1 & EN45544-2: 1999 for the measurement of
Carbon Monoxide, Hydrogen Sulfide and Carbon Dioxide.
•EN50271: 2000 for the assessment of Digital Components
and Software.
WARNING
Assessment has only been made in the range of 0 to
100%LEL. Use of other ranges of flammable gas
measurement on this instrument will invalidate this
approval.
14
MAN0597ISSUE 09 - 12/05
UL / CSA Approval
User Warning
Product Name
2302M5030
INSTRUMENT LABELS
UL / CSA CERTIFICATION LABEL
An explanation of the information on the UL / CSA certification
label is shown below.
WARNING
Only the combustible gas detection portion of this
instrument has been assessed by CSA for performance.
Furthermore assessment has only been made in the 0 to
100 %LEL scale. Use of other ranges of flammable gas
measurement on this instrument will invalidate approval.
15
MAN0597ISSUE 09 - 12/05
2302M5030
1. INTRODUCTION
1. INTRODUCTION
The Impact / Impact Pro is a compact, portable gas monitor
designed to be carried or worn without hindering the user. Its
purpose is to monitor the atmosphere continuously for
hazardous levels of up to four gases. These concentrations if
gases are measures using Zellweger Analytics sensors.
Audible and visual alarms alert the user to danger when
hazardous conditions are detected
The instrument is usually supplied with four gas sensors, for
detecting oxygen (enrichment and deficiency), flammable
gases (up to the Lower Explosive Limit) and two toxic gases
(for personal safety) all housed in an easily replaceable
cartridge.
Various sensor technologies are used to achieve this. In the
vast majority of cases, electrochemical technology is used to
detect oxygen and toxic gases while catalytic combustion
technology is used to detect flammable gases.
Two types of cartridge are available. One is disposable where
the cartridge has a fixed life and once this has expired the
cartridge is disposed of. The other is a Serviceable Cartridge
where the sensors can be individually replaced when required.
The serviceable cartridge type can only be used in the
Impact Pro.
Note: Throughout this manual it is assumed that the Impact /
Impact Pro is equipped with a Disposable Four Sensor
Gas Cartridge. References to sensors not fitted in the
users instrument should be ignored.
This manual covers all models - some features are not
available on all models. Not all models are available in
every country.
The instrument is supplied with dry cells and holders as
standard. Rechargeable batteries and charger can be
purchased separately as a kit.
16
MAN0597ISSUE 09 - 12/05
2302M5030
1. INTRODUCTION
1.1INTENDED USE
The Impact/Impact Pro has been designed to alert the user to
potentially hazardous atmospheres whilst carrying out his/her
normal duties. Therefore, the instrument must be kept switched
on and worn as close to the breathing area as possible, and
several accessories are provided to allow the instrument to be
worn in a number of different ways:
a.On the chest
b.On a belt
c.Attached to a body harness
The instrument is provided with various methods to enable the user
to comply safely and easily with confined space regulations.
CAUTIONS
Hand aspirated remote sampling only provides continuous
gas readings as long as the bulb is being operated.
Zellweger Analytics recommend that the instrument be
calibrated at least every 6 months or in accordance with
customer site procedures, whichever is sooner. Correct
operation of the instrument should be confirmed with test
gas of known concentration before each use.
channel if
Do not calibrate the
air after the apparatus has been exposed to
concentrations above the upper limit of the measuring
range. In this case leave the instrument in clean air
overnight before calibration.
The calibration interval for the
1 month.
The use of Enforcer calibration accessory is strongly
recommended as it enables this to be performed quickly
and easily.
CO
2
CO
2
is indicated in clean
CO
2
CO
sensor shall not exceed
2
WARNING
A sensor which cannot be calibrated or which is found to be
out of tolerance should be replaced immediately. For the
disposable cartridge a replacement cartridge must be fitted.
17
MAN0597ISSUE 09 - 12/05
1. Buttons
2. Pump Aperture
3. Cartridge and
Filter Cover
4. Audible Aperture
5. Grille Cover Screws
6. Certification Label
7. Battery Covers
8. Data Connector
9. Tool
9
8
2
3
4
5
1
7
6
▲ (yellow) Moves up through
menu screens, and used to
increase values
✔ (green) used as ‘OK’ in
menus. Accepts alarms,
and resets peak values
(red) On/Off button.
Also used as ‘change’ in
menus.
▼ (yellow) Moves down
through menu screens, and
used to decrease values
2302M5030
1. INTRODUCTION
1.2PRODUCT OVERVIEW
On the top of the unit are four buttons (1). Their functions are
summarised below:
Pressing any key will automatically activate the display
backlight for 10 seconds.
18
MAN0597ISSUE 09 - 12/05
2302M5030
1. INTRODUCTION
There are currently two types of instrument available - Impact
and Impact Pro. The main differences between the two
instruments is that the Impact Pro supports a range of
Serviceable Cartridges (see Section 8.1 ServiceableCartridges of this manual). The other detail differences are the
addition of an internal sample pump, vibrating alarm and
Safelink feature.
The instrument can be further personalised by use of the
Impact Configuration Utility (ICU) PC software, which can be
purchased separately as part of the Data Logging Kit. This
allows the user to change various settings and features of the
instrument including, but not limited to Alarm Levels, Autozero
function, latching alarms, vibrating alarm (where fitted), data
logging settings and Safelink messages.
19
MAN0597ISSUE 09 - 12/05
2302M5030
2. GETTING STARTED
2. GETTING STARTED
2.1HOW TO TURN IMPACT/IMPACT PRO ON
AND OFF
The Impact has been designed for ease of use, and especially
for one-handed operation - only a single button is needed to
turn it on and off.
•To turn the instrument on, press the button until the
instrument activates its audible and visual alarms. It will
follow the start-up sequence described in Section 3.
INSTRUMENT START-UP.
If the instrument displays an error that no cartridge is fitted
then follow the procedure in Section 2.2 Inserting the cartridge.
•To turn the instrument off, press and hold the button for
three seconds, until it switches off. Note that on some
models a password must be entered to switch the unit off.
Failure to enter the correct password will cause the
instrument to continue as though the button had not
been pressed.
20
MAN0597ISSUE 09 - 12/05
2302M5030
2. GETTING STARTED
21
2.2INSERTING THE CARTRIDGE
(1)If the instrument is switched on then switch it off by
pressing and holding the button. Note: if a cartridge
is already fitted, check that the instrument clock is
correct. If it is not, change the instrument clock as
described in section 4.1.1.
(2)Undo the two grille cover screws (5).
(3)If a cartridge is already fitted then remove it by undoing
the central screw.
(4)Insert new cartridge into aperture as shown. Ensure that
the point ‘★’ is located correctly in the pump or moulding
(depending on model).
(5)Gently tighten up the central screw to secure in place.
(6)Check the condition of the filter on the grille cover (3),
and if necessary, replace it.
(7)If the unit is fitted with a pump replace the pump seal.
★
MAN0597ISSUE 09 - 12/05
2302M5030
2. GETTING STARTED
22
(8)Replace the front cover, and retighten the two screws
(5).
(9)Wait at least 20 minutes. Then switch the Impact on by
pressing the button and check no faults are reported
by the instrument. If fault 4 occurs refit the cartridge.
(10)Once the new cartridge is fitted the instrument will
compare it to the cartridge fitted previously. The
instrument will alert the user with warning if:
a)the alarm levels are different
b)the number of sensors is different
c)the mix of the sensors is different.
(11)If the instrument reports a gas alarm, switch instrument
off, wait 20 minutes, and switch unit on again.
2.3CHARGE FOR FIRST USE
Either a rechargeable battery pack or a dry-cell battery source
can power the Impact. For the rechargeable battery it must be
charged before first use, to replace any battery capacity used
during transit and storage.
(1)Ensure the basestation is connected to a suitable power
source.
(2)Place the instrument in the basestation.
The basestation employs a locking mechanism to
ensure that the Impact is retained under most operating
conditions. To ensure this operates correctly the
following procedure should be followed:
MAN0597ISSUE 09 - 12/05
2302M5030
2. GETTING STARTED
23
a)Ensure that the Impact is orientated such that the
instrument lip will be inserted under the locking lip.
b)Insert the front of the Impact at an angle such that
the instrument lip slides under the locking lip
.
Instrument lip
Locking lip
Front
Front
RearFront
Closeup detail
MAN0597ISSUE 09 - 12/05
(Disengaged)Rearlocking catch(Engaged)
2302M5030
2. GETTING STARTED
c)Push down on the rear of the Impact such that the
rear locking catch engages.
d)To remove the Impact press down on the rear
locking catch.
All 4 alarm lights will flash indicating the Impact has
started charging.
(3)Whilst charging the instrument, the instrument will flash
2 red LEDs approximately every 2 seconds. When
charging is complete it will light the green LEDs
constantly. A pair of fully discharged battery packs will
require 7 hours to recharge fully.
WARNING
Do not charge the battery pack in a hazardous area.
24
MAN0597ISSUE 09 - 12/05
2302M5030
2. GETTING STARTED
2.4DRY-CELL BATTERY INSERTION
+
+
(1)Undo the two battery compartments (7) using the tool
(9) provided, on the bottom of each instrument.
(2)Release each dry-cell battery holder and remove the
cells if fitted.
(3)Insert new cells, ensuring correct orientation by
checking that the negative terminal is at the same end
as indicated on the moulding. Ensure they are of the
correct type, to comply with the intrinsic safety
requirements.
(4)Replace dry-cell holders in the battery compartments,
and retighten cover screws.
(5)Instrument is now ready for use.
25
MAN0597ISSUE 09 - 12/05
+
+
+
+
+
2302M5030
2. GETTING STARTED
2.5SAMPLING
In normal operation the Impact is worn on the belt, with its
harness, or held by hand. Once turned on the Impact monitors
the atmosphere continuously, which reaches the sensors by
diffusing through the vents of the grille cover or by being pulled
through by the internal pump (if fitted). For non-pump
instruments normal air movements are sufficient to carry the
sample to the sensors, and the sensors react immediately to
changes in concentrations of the gases being measured in the
atmosphere immediately surrounding the detector. Depending
on your application and the options fitted to the instrument the
environment can be sampled remotely in a variety of ways, as
summarised below:
No pumpFlow hoodHand
With pumpPump adaptorSampling 1m
aspirator
Sampling Ball float
tube
tubesensing
probe
26
MAN0597ISSUE 09 - 12/05
2302M5030
2. GETTING STARTED
27
CAUTIONS
Hand aspirated remote sampling only provides continuous
gas readings while the bulb is being squeezed. Each time
a reading is required, it is necessary to squeeze the bulb
at a rate of one per second until the readings remain
stable.
When using the built-in pump ensure that the sampling
tube is not inserted into a fluid.
MAN0597ISSUE 09 - 12/05
English
Language
French
Italiano
Deutsch
Español
French
Language
Deutsch
Español
Italiano
Nederlands
Flammable Gas
Methane
Oxy Flam
CO H2S
Calibration Due
in 120 days
2302M5030
3. INSTRUMENT START-UP
3. INSTRUMENT START-UP
3.1INSTRUMENT INFORMATION
After turning the instrument on it will display the information in
the following autosequence (depending on model):
The first display identifies the model.
Whilst this is shown the alarms are
tested, If a vibrating alarm is fitted this
is also activated. Whilst this screen is
displayed, pressing the button will
cause the instrument to identify the
version of software installed and the
instrument serial number.
A Language Selection Screen is then
shown (first use only). Use the
▲ and
▼ buttons to highlight the required
language and press the
✔ key to
make the selection. Note that some of
the languages may be 'off screen' use the
▲ and ▼ buttons to access
these.
The language can be changed at a
later time, if required. See section
4.7.6 for details.
The display then shows the gas that
the flammable sensor is set to
monitor, the sensors fitted and when
the calibration is due (if the calibration
is due in less than 30 days).
28
MAN0597ISSUE 09 - 12/05
Location
Default Location
Operator
Default Operator
✓ - OK
- Change
Zero sensors?
✓ - OK
- No
2302M5030
3. INSTRUMENT START-UP
The instrument logs data on the
operator’s exposure, if any, to
measured toxic gases. It achieves this
by requesting the operator to confirm
his/her identity and the location where
the instrument is being used. The
screen displays the last location and
operator. If these are acceptable then
press the
✔ button. Otherwise select
a new location and/or operator.
3.2SELECTING LOCATION / OPERATOR
To change the location and/or operator press the button
and the display will show the current location. Press the
▼ buttons to cycle through the list. Once the correct location
has been found then press
✔, and follow the same procedure
to select the operator.
▲ and
3.3FRESH AIR AUTO ZEROING SENSORS
If enabled the instrument will prompt if
you wish to zero the sensors in fresh
air, to adjust for any natural drift that
may have occurred. If the ✔ button is
pressed the instrument will then ask
the user to ensure that it is being
zeroed in fresh, uncontaminated air.
29
MAN0597ISSUE 09 - 12/05
Are you in
fresh air?
✓ - Yes,- No
2302M5030
3. INSTRUMENT START-UP
If the ✔ button is pressed the
Note: This is a ‘soft’ zero. Any adjustments only remain in
force whilst the instrument is switched on. To
permanently adjust the zero, the zero via the calibration
menu must be used (‘hard zero’).
instrument will zero the sensors
automatically, and display whether the
procedure was successful. The
oxygen reading will be adjusted to
20.9%v/v – the other sensor readings
will be adjusted to 0 ppm and 0%lel
appropriately. If the button is
pressed instead the instrument will
use its current zero values instead
and proceed to the monitoring screen.
3.4SENSOR WARM-UP PHASE
Following display of the Autozero screen it will be noted that
the message 'Cell Self Test Please Wait' is displayed. This is
displayed while the sensors are in the warm-up phase of the
start-up procedure. If the option of Autozero is declined the
main gas measuring screen will be displayed with the
message 'Warm Up' being displayed in place of the gas
concentration readings. The instrument should not be used
until the 'Warm Up' message has disappeared. The amount of
time that the sensors are in warm up will depend on the sensor
type. Table 12.3.1 in section 12 of this manual gives details of
the warm up times for each of the sensors.
30
MAN0597ISSUE 09 - 12/05
2302M5030
4. OPERATION
31
4. OPERATION
Note: Wherever the manual says ‘select’, the procedure is to
use the
▲ and ▼ buttons to cycle through the list, and
then press
✔
to select the required option.
4.1MONITORING CONDITION
4.1.1Display Screens
With no alarm conditions the display will typically show:
The gas sensors and their respective units are displayed,
complete with a battery gauge monitor at the bottom. For an
instrument fitted with less than four sensors each unused
sensor position will show ‘---‘.
Several data screens are available and these can be viewed
by pressing the
▲ and ▼ buttons to cycle through. The symbol
in the centre of the screen will identify which screen is active.
Peak Screen ()
This symbol is displayed when the
instrument is displaying the peak
readings for the sensors, i.e. the
highest readings seen since the
instrument was switched on or since
they were reset. This is useful for preentry checks for confined space entry.
These readings can be reset by
pressing the
✔ button whilst this
display is shown.
This screen will alternate with a screen showing the minimum
value of oxygen. In this case the is replaced by .
020.9
00
02CO%V/V
ppm
Flam
H2S
%LEL
ppm
020.9
00
02CO%V/V
ppm
Flam
H2S
%LEL
ppm
▲
MAN0597ISSUE 09 - 12/05
2302M5030
4. OPERATION
32
STEL Screen ( )
This symbol is displayed when the
instrument is displaying the STEL
readings for the toxic sensors. The
STEL is a time weighted average,
measured over a 15 minute reference
period. It is used to monitor exposure
to toxic gases in line with current
regulations and/or legislation. Until 15
minutes has elapsed these are
projected values.
LTEL Screen ( )
This symbol is displayed when the
instrument is displaying the LTEL
(TWA) readings for the toxic sensors.
The LTEL is a time weighted average,
measured over an 8 hour reference
period. It is used to monitor exposure
to toxic gases in line with current
regulations and/or legislation. Until the
Impact has been monitoring
continuously for 8 hours these are
projected values.
Status Screen
This screen shows the current time,
date and battery status.
It is possible to change the instrument
clock setting as follows: Press the
✔
key while this screen is displayed and
the clock setting function will be activated, with the hour field
highlighted. Use the
▲ and ▼ keys to change the value as
required and press the
✔ key to move to the minute field.
Again use the
▲ and ▼ keys to change the value and continue
until all of the fields have been correctly set. Finally press the
✔ key to store the changes. Note that the key can be
pressed at any time to abort the procedure.
00
COppm
H2S
ppm
00
COppm
H2S
ppm
18:33:00
22/Jan/2001
MAN0597ISSUE 09 - 12/05
2302M5030
4. OPERATION
33
4.1.2Other Displayed Symbols
If the pump is operating correctly
the pump symbol will rotate.
If the pump flow becomes
blocked the instrument will warn
the user and stop the pump to
prevent potential damage. When
the user accepts the warning
message the instrument will
attempt to restart the pump.
There is no need to remove and
refit the Pump Adaptor. At this
point, the user should investigate
the cause of the blockage alarm.
If the blockage is cleared then the
pump will successfully restart. If
the blockage is still present then
the instrument will again warn the
user that the pump flow is still
blocked. This will repeat until the
blockage is removed.
This shows an approximation of the remaining battery
capacity within the instrument. If there is less than
20 minutes remaining battery life then the instrument
will display a ‘Low Battery’ warning.
This will be shown instead of the numerical reading
for any sensor or channel that is faulty, if a zero or
span calibration has failed, in instances of low oxygen
and in some cases of high gas concentrations. Switch
the instrument off and then back on. If this does not
clear the fault then recalibrate the sensor or change
the cartridge.
020.9
00
02CO%V/V
ppm
Flam
H2S
%LEL
ppm
MAN0597ISSUE 09 - 12/05
2302M5030
4. OPERATION
This indicates that the sensor output of the indicated
-0
If -0.0 Vol.% for the CO2 sensor channel is permanently
displayed, a sensor zero (during boot procedure) or a zero
calibration needs to be performed in clean air. The alarm
level A1 for the CO
channel has drifted negative. In severe cases of
sensor negative drift, a Warning 51 (Excessive
Negative Drift) will be shown. Please see Appendix A
– Warning Codes for further details on this warning
message.
WARNING
range must not exceed 0.5 Vol.%.
2
4.1.3Confidence Signal
To ensure correct operation the instrument monitors itself and
will confirm correct operation by giving an audible and green
visual confidence signal once every 30 seconds. The
confidence signal is given when the instrument is able to
detect gas (for example the confidence signal will not be given
during pump or sensor calibration phases, or if the sensors are
in warm-up).
There is a configurable option to silence the audible
confidence signal but the visual confidence signal will still
operate. If the instrument is in a low battery condition the
confidence signal will occur twice every 30 seconds.
Note: The audible confidence signal is the primary indication
that the instrument is functioning correctly. It is therefore
strongly recommended that this feature not be disabled.
4.1.4Go/No Go Option
A configurable option is available which replaces the numeric
values with a
there is or has been an alarm or fault, as shown below. All
alarms operate as normal, but other functions and menus are
disabled.
✔ symbol when everything is OK and ✘ when
34
MAN0597ISSUE 09 - 12/05
✓
✗
2302M5030
4. OPERATION
This display now also shows the battery capacity as follows.
The Go / No Go Option is configured using the Impact
Configuration Utility (ICU) PC software from the Configuration
Screen.
4.2ATMOSPHERIC ALARM CONDITIONS
CAUTION
The Impact portable gas detector has been designed for
the detection of oxygen deficiencies and enrichments,
flammable gas and toxic gas levels. An alarm condition
indicating the presence of one or more of these potentially
life-threatening hazards should be taken seriously.
In the event of an alarm is activated when the measured
gas concentration exceeds the pre-set alarm point.
A rapid increase in reading followed by a declining or
erratic reading may indicate a hazardous combustible gas
concentration that exceeds the measuring range of the
Impact (i.e. greater than 100%LEL or 5.0%v/v Methane).
In the event of the instrument being exposed to a very high
level of flammable gas the following behaviour will be
noted.
•Warning 54 (Low O
indicated. This warning will be latched.
•Warning 52 (sensor over-range) and an alarm
condition will be indicated.
•The flammable reading will be latched to 100%LEL or
5.0%v/v Methane with an alternating cross on the
channel.
For other sensors reading greater than the measurement
range, ‘
displayed.
▲▲▲’ alternating with the full-scale reading will be
) and an alarm condition will be
2
35
MAN0597ISSUE 09 - 12/05
02CO%V/V
ppm
Flam
H2S
%LEL
ppm
12
3
2302M5030
4. OPERATION
4.3ALARM CONDITION
There are two modes of alarm: latching and non-latching.
However, the display will give the same alarm indication:
An alarm symbol will appear in the
relevant section of the display. The
alarm symbol contains a number,
0
indicating the increasing severity of
the alarm, with increasing frequency
of the audible and visual alarms. If
there is a STEL/LTEL alarm the
relevant icon will appear and flash.
Any alarm will cause the backlight to be switched on
automatically.
4.3.1Latching Alarms (default)
In the latched condition, once an alarm occurs both audible
and visual alarms continue to operate even after the
atmospheric hazard has cleared. Pressing any of the
instrument buttons will clear an alarm. Any subsequent alarm
will reactivate the audible and visual alarms.
4.3.2Non-latching Alarms
In this mode, should a gas alarm occur the instrument would
enter an alarm condition. When the readings return to normal
levels the audible and visual alarms will stop.
4.3.3Vibrating alarm (where fitted)
If this option is fitted to the instrument any alarm condition that
activates the audible and visual alarms will also activate the
built-in vibrating alarm.
36
MAN0597ISSUE 09 - 12/05
WARNING
Code - 0
See manual
2302M5030
4. OPERATION
4.3.4Resetting an Alarm
If an alarm condition occurs it is possible to cancel the alarm
by pressing any of the buttons, once the gas measurements
have returned to a safe level. Otherwise, the instrument will
remain in the alarm condition, but the audible alarm will be
muted. Any subsequent alarms that occur 1 second after the
previous alarm has been reset will reactivate the audible
alarm.
4.4FAULT AND WARNING CONDITION
In addition to the gas alarms, the Impact includes a number of
auxiliary alarms to safeguard proper use of the instrument. At
switch on the Impact performs an electronic self-test that
assures the user of proper performance. When the Impact
detects that an electronic fault or failure condition has
occurred, the audible and visual alarms are activated and an
explanatory message will be displayed.
CAUTION
As the Impact is designed to protect from potentially lifethreatening atmospheric conditions, any alarm conditions
must be taken seriously.
4.4.1Warning
The instrument displays a warning
message for situations where a fault
or error has occurred but may be
resolved by the user.
The 'Code - 0' shown here is an
example only. A full list of codes is
given in Appendix A.
37
MAN0597ISSUE 09 - 12/05
FAULT
Code - 0
Please contact
Zellweger Analytics
Z A Service Number
2302M5030
4. OPERATION
4.4.2Fault condition
If a fault condition is detected during
start-up or subsequently, the
instrument will display a fault
message warning the user and giving
a contact number. This will remain
until the instrument is turned off by
pressing the button for at least
3 seconds.
The 'Code - 0' shown here is an
example only. A full list of codes is
given in Appendix A.
4.5SAFELINK
4.5.1What is Safelink?
Safelink is a confined space entry communication system
between instruments, fitted with the Safelink feature. It allows
one instrument (the 'attendant') to display the gas readings
measured by the other connected instrument (the 'entrant'), up
to a maximum cable length of 100m. Safelink also provides an
automatic timed response system requiring the Entrant's
instrument to return a signal, activated by the entrant, within a
user-specified time interval. Failure of the user to respond will
cause an alarm to be raised on the Attendant's instrument.
Additionally, any continued pressing on any button on the
entrants instrument at any time will cause an 'Emergency'
message to appear on the attendant instrument - similar to the
action of a panic button. In Safelink mode the entrant
instrument can neither be switched off, nor can the pump
38
MAN0597ISSUE 09 - 12/05
Safelink Mode
Entrant
Attendant
2302M5030
4. OPERATION
feature (where fitted) be used on either instrument. It will be
found that while the pump is running the Safelink option on the
instrument menu system has been removed. Furthermore, if
an attempt to start the pump is made (on either the attendant
or entrant instrument) during Safelink operation an alarm
condition is generated and Safelink Mode has to be exited.
Safelink mode cannot be entered when the batteries are low
(i.e. there is less than 1 bar on the battery indication meter).
In some instances a Fault or Warning will not be displayed on
the Entrant instrument. In this case it is important to note that
gas alarms on the Entrant instrument are still indicated on the
Entrant and the Attendant instruments.
4.5.2Using Safelink
Connect the Safelink cable between two instruments. Turn on
each instrument and from the user menu on each instrument
select the Safelink mode.
On one instrument select Attendant.
When selecting the Attendant on one
instrument the instrument will display
‘locating’ while it attempts to connect
with the other instrument. Once
established each instrument will
briefly display ‘Configuring’. The instrument will then show the
readings – ensure that the Safelink symbols appear on the
display. To check the integrity of the communications it is
recommended that a message be sent from the Attendant to
the Entrant. Whenever a message must be responded, the
instrument will flash its green LEDs and sound its audible
alarm at a rate of once per second.
Until the Safelink cable is disconnected both instruments will
remain in Safelink mode.
39
MAN0597ISSUE 09 - 12/05
02
Tox1
%V/V
ppm
Flam
Tox2
%LEL
ppm
Emergency
00:12:34
Select Message
Evacuate Area
Message 1
Message 2
Are you OK?
2302M5030
4. OPERATION
4.5.2.1 Attendant Instrument
communications are occurring. When it is flashing the Safelink
has become disconnected and the readings will all show ‘---‘,
until the link is restored or the user exits from Safelink mode.
All display modes from the Entrant’s instrument (peak, STEL,
etc) are available to the Attendant.
020.9
00
The instrument will display the
readings being monitored on the
Entrant instrument. The symbol is
used to indicate the Attendant’s
instrument. When the symbol is static
it indicates that Safelink
The status screen shows the duration
that Safelink has been in operation,
and the status of the connection as
either Normal, Emergency, or Link
failure.
By pressing the
✔ button the
Attendant has access to a menu, to
select a message to send to the
Entrant. These are configurable using
the PC software.
The Select Message menu gives access to two additional
actions:
a.Emergency, which immediately issues a ‘Get Out’
warning to the Entrant.
b.Exit, to enable termination of the Safelink without
removing the cable.
40
MAN0597ISSUE 09 - 12/05
Select Message
Message 2
Message 3
Message 4
Message 1
Exit Safelink
EXIT
Continue
2302M5030
4. OPERATION
4.5.2.2 Entrant instrument
The symbol is used to indicate the Entrant’s instrument.
When the symbol is static it indicates that Safelink
communications are occurring. When it is flashing the Safelink
has become disconnected.
Only the instantaneous values are shown, although all alarm
conditions will operate as normal.
By pressing the
has access to a menu, to select a
message to send to the attendant.
These are configurable using the PC
✔ button the entrant
software, and could be used to
indicate work progress, e.g. ‘Valve
Now Closed’.
4.5.2.3 Timed Response
At a preselected interval the attendant’s instrument will prompt
for the Entrant to be checked. The Entrant must press any
button within a preselected time, otherwise the Safelink
system will assume an emergency situation has occurred and
raise alarms on both instruments. The default interval is
5 minutes. The default time to respond to a message is
30 seconds. Both of these can be changed using the PC
software.
4.5.2.4 Terminating Safelink
To terminate Safelink mode the cable
should be disconnected between the
instruments. Both instruments will
display a menu enabling Safelink
mode to be switched off.
41
MAN0597ISSUE 09 - 12/05
2302M5030
4. OPERATION
Safelink mode must be terminated before attempting to use
the instrument in a basestation (for charging or calibration) or
the Enforcer.
4.6PUMP (IMPACT PRO ONLY)
The pump allows gas to be drawn through tubing across the
sensors. Fitting the Pump Adaptor will automatically switch
the pump on.
If the pump is operating correctly the pump symbol will rotate.
Please refer to Section 4.1.2 Other Displayed Symbols of this
manual on the operation of the instrument under blocked flow
conditions.
The instrument contains a feature that tests and if necessary
dynamically sets the Pump Stall threshold via a calibration
routine. Upon fitting the Pump Adaptor, the user will see the
following sequence of screens. The instrument gives
instructions at each stage informing the user as to what
operations need to be performed.
Please follow the instructions on screen during the Pump Test
and Calibration procedure. Fit the required length of sample
tube of the inlet of the Pump Adaptor together with the
Hydrophobic Filter. Fit the Pump Adaptor to the front of the
instrument.
Pump Test
Please ensure
pump unblocked
Press ✔ to continue
Ensure that there is nothing obstructing the flow into the
sample pump and press the
✔ button.
42
MAN0597ISSUE 09 - 12/05
2302M5030
4. OPERATION
The following screen will be displayed briefly.
Pump Test
Please wait...
Followed by…
Pump Test
Please block pump
Press ✔ to continue
Use a suitable method to block the end of the sample tube and
press the
✔ button.
The instrument will now test the pump. Ensure that the
blockage is kept in place for the duration of this test.
If the pump test is successful the following screen will be
displayed.
Pump Test
OK
Press ✔ to continue
Remove the blockage from the sample tube and press the
✔
button.
The internal sample pump feature is now ready for use.
To stop the sample pump, remove the Pump Adaptor as
described in Section 4.6.1 Removal of the Pump Adaptor.
If the Pump Test fails then the pump calibration procedure will
start.
Pump Calibration
Please ensure
pump unblocked
Press ✔ to continue
43
MAN0597ISSUE 09 - 12/05
2302M5030
4. OPERATION
Ensure that there is nothing obstructing the flow into the
sample pump and press the
The instrument will then perform the first part of the pump
calibration.
✔ button.
Pump Calibration
Please wait...
Pump Calibration
Please block pump
Press ✔ to continue
Use a suitable method to block the end of the sample tube and
press the
✔ button. The pump calibration will now commence.
Pump Calibration
Please wait...
Ensure that the pump is kept in a blocked state until the
following screen is displayed.
Pump Calibration
OK
Press ✔ to continue
Remove the blockage from the sample tube and press the
✔
button.
The internal sample pump feature is now ready for use.
44
MAN0597ISSUE 09 - 12/05
2302M5030
4. OPERATION
If the Pump Calibration fails (please see the diagnostic table at
the end of this section) then use of the sample pump is
inhibited. The Pump Adaptor must be removed from the
instrument and the fault investigated. To correctly remove the
Pump Adaptor, follow the instructions in section 4.6.1.
Removal of the Pump Adaptor will turn the pump off.
Below 0°C the efficiency of the pump will be adversely
affected, requiring a longer sampling time.
The following is a list of possible problems that could cause the
instrument to fail the various tests detailed in this procedure.
Symptom Possible CauseRemedy
The instrument reports
'Pump Test Fail'
The sample pump is
in an uncalibrated
state or the operating
conditions of the
pump have changed
significantly since the
last calibration.
The pump was not
blocked correctly
when required.
A blocked condition
was not detected
within 30 seconds of
the start of the test.
Perform the Pump
Calibration to
recalibrate the
pump.
Remove the Pump
Adaptor and refit to
restart the test.
Ensure that the
pump is blocked
correctly when
instructed to.
Remove the Pump
Adaptor and refit to
restart the test.
Ensure that the
pump is blocked as
soon as the 'Please
block pump screen'
is displayed.
45
MAN0597ISSUE 09 - 12/05
2302M5030
4. OPERATION
Symptom Possible CauseRemedy
The instrument reports
'Pump Calibration Fail'
The pump was not
blocked correctly
when required.
There is a leakage in
the system.
A blocked condition
was not detected
within 30 seconds of
the start of the test.
Remove the Pump
Adaptor and refit to
restart the test.
Ensure that the
pump is blocked
correctly when
instructed to.
Check that the
following items are
correctly fitted and/
or are not damaged.
• Pump Seal
(behind front
cover)
• Sample Tube
• Pump Adaptor
Remove the Pump
Adaptor and refit to
restart the test.
Ensure that the
pump is blocked as
soon as the 'Please
block pump screen'
is displayed.
PUMP FAULT and
WARNING 16 'Pump
Fault' screens are
displayed following the
Pump Calibration.
The Pump Assembly
is faulty.
The Pump
Calibration has
failed.
46
Request a new
Pump Assembly or
contact Zellweger
Analytics for
assistance.
Remove the Pump
Adaptor and
investigate the fault
(see The instrument
reports 'Pump
Calibration Fail'
above).
MAN0597ISSUE 09 - 12/05
User Menu
Flammable
Calibration
Operator
Instrument
Safelink
Language
2302M5030
4. OPERATION
4.6.1Removal of the Pump Adaptor
To ease the removal of the Pump Adaptor, follow the steps
below.
(1)Support the end of the Pump Adaptor close to the inlet
port using the thumb of one hand.
(2)Press on the top of the clip using the thumb of the other
hand until a click is heard.
(3)Lift the Pump Adaptor clear of the instrument.
4.7MENUS
While the gas monitoring screen is displayed, pressing the ✔
button will provide access to menus, depending on the model.
Note: Calibration is not permitted until the sensors have
warmed-up and been self tested or if the pump is
running. If you access the user menu under any of these
circumstances the Calibration option will not be
available.
47
MAN0597ISSUE 09 - 12/05
2302M5030
4. OPERATION
4.7.1Flammable Gas Selection
It is possible to set the display to read for specific flammable
gases. Select the required flammable gas. The instrument will
automatically adjust its internal correction factors.
Note: The above data are applicable only to instruments that
are configured to display the Flammable gas in %LEL.
The above table and the cross-sensitivity feature offered in the
Impact and Impact Pro are supplied for indicative purposes
only. The following points should be noted when using the
table or the software feature.
(1)There is variability in sensor cross-sensitivity between
methane and other flammable compounds. Therefore, if
48
MAN0597ISSUE 09 - 12/05
2302M5030
4. OPERATION
the instrument is calibrated to Methane (including
Enforcer calibrations), the reading when other
Flammable gases are selected will be subject to
variation.
(2)For more accurate detection of non-Methane gases, the
Flammable Channel of the instrument should be
calibrated to Propane, Pentane or Butane (selectable
from the Settings option in the Calibration Menu). In this
instance the reading obtained when Methane is
selected may be subject to inaccuracy.
(3)Maximum accuracy will be obtained by calibrating with
the target gas, and is therefore the preferred method.
Target Gas Recommended Calibration Method
Enforcer
UI Calibration (using Methane as the selected
Methane
Propane
Butane
Pentane
Other
flammable
gases
calibration gas)
PC Calibration (using Methane as the selected
calibration gas)
UI Calibration (using Propane as the selected
calibration gas)
PC Calibration (using Propane as the selected
calibration gas)
UI Calibration (using Butane as the selected
calibration gas)
PC Calibration (using Butane as the selected
calibration gas)
UI Calibration (using Pentane as the selected
calibration gas)
UI Calibration (using Propane, Butane or Pentane as
the selected calibration gas)
PC Calibration (using Propane or Butane as the
selected calibration gas)
4.7.2Operator
This allows a new operator and/or location to be selected
without restarting the instrument.
49
MAN0597ISSUE 09 - 12/05
Software Revision
Version 2.6
Serial Number
0000000000
Calibration Due
in 34 days
02%V/VFlam %LEL
50
80
120
23.0
19.0
17.0
Flammable Gas
Methane
A1
A2
A3
50
100
150
100
50
25
50
100
50
25
A1
A2
A3
STEL
LTEL
COppm
H2S
ppm
Pump
Fitted
Data Logging
Event
Safelink
Fitted
Last Calibration
1 Jan 2001
Autozero
Enabled
Gas Alarms
Latched
Battery
Rechargeable
Vibrating Alarm
Enabled
Language
English
2302M5030
4. OPERATION
4.7.3Calibration
See Section 4.9 Calibration for full details.
4.7.4Instrument Details
Several screens are available detailing the instrument set-up,
such as alarm levels. These can be viewed by pressing the
and ▼ buttons to cycle through them. Some examples are
shown below and these are indicative only. Information shown
will vary according to model, country, product application and /
or specific requirements.
▲
This information is also available at
switch-on of the instrument (see
section 3.1 for details).
These are the alarm level settings for
oxygen and flammable sensors.
There are no associated STEL or
LTEL alarms for these sensors.
↑indicates a rising alarm and ↓
indicates a falling alarm.
This shows the alarm levels for the
toxic sensors.
The following displays indicate whether particular options are
fitted and how the instrument is currently configured.
50
MAN0597ISSUE 09 - 12/05
Cartridge SN
0001138
Date of Manufacture
25/Jun/2001
Boot ROM Version
Impact Boot 1.4+
Cartridge Type
Standard
Install By
26/Dec/2001
Date of Activation
9/Aug/2001
2302M5030
4. OPERATION
The following additional information is also displayed:
4.7.5Safelink
For Impact Pro instruments the additional menu for Safelink is
available. It is used to initiate communication via Safelink to
another Impact Pro instrument.
The Safelink option (where applicable) will not be available in
the following situations.
•When the internal sample pump is running.
•When the batteries are low (i.e. there is less than 1 bar on
the battery indication meter).
4.7.6Language
The Impact is supplied pre-loaded with English, French,
German, Italian, Spanish and Dutch languages, which can be
selected using the
▲, ▼ and ✔ keys of the instrument. In
addition one other alternative language can be uploaded to the
Impact (currently available are Portuguese, Danish, Swedish,
Norwegian and Finnish) using the Impact Configuration Utility
(ICU) PC software, which is supplied separately.
4.8DATALOGGING
There are two types of datalogging. In both cases the
information is accessed by downloading it to a computer using
the PC software, which is supplied as part of the Datalogging
Kit. The PC software enables this data to be downloaded,
51
MAN0597ISSUE 09 - 12/05
2302M5030
4. OPERATION
stored, printed and analysed. The data can be exported in a
format suitable for use with major spreadsheet software
packages. The datalog memory can be automatically cleared
after a successful download – note that this will not change any
of the alarm levels, instrument settings, or calibration gas
settings or values of the instrument. An internal battery will
retain the data for 5 years even if the instrument battery is
disconnected or the instrument is switched off. Refer to
Section 4.8.1 on how to install the PC software.
4.8.1Installing the PC Software
(1)Insert the CD into your CD-ROM drive. If the CD
autoruns press the Esc key to stop.
(2)From the Start menu, choose Run.
(3)In the Run dialog box, type d:\English\ setup.exe, where
d is the letter assigned to your CD-ROM drive.
(4)Click OK, then follow the instructions on your screen.
4.8.2Event Datalogging
All instruments are supplied complete with event datalogging.
This records the time and date whenever an event occurs.
When the datalogging memory is full then the earliest data is
overwritten. In event mode the datalogging memory can store
at least 500 events. An event can be:
•Instrument switch on
•Instrument switch off
•Peak gas reading whilst switched on
•Any gas alarm (A3, A2, A1, STEL, LTEL)
•Low battery
•Fault
52
MAN0597ISSUE 09 - 12/05
2302M5030
4. OPERATION
4.8.3Gas Datalogging (optional)
Set-up by the PC software for instruments fitted with this full
datalogging option it enables the instrument to monitor areas
for gas, sampling either at a regular interval (for instance,
every 15 seconds) or when the readings change by a userselected difference.
This extended datalogging can also store the event logging
information if required.
4.8.4Output Formats
The calibration history, event logging and gas logging are
output in a CSV (Comma Separated Variable format) with
the file extension '.txt'. To view the information in
Microsoft Excel®, follow these steps:
(1)Open the file using Microsoft® Excel. Note that the 'Files
of Type' box must be set to 'All files' to show the files,
which have a '.txt' extension.
(2)Excel will recognise the format as 'delimited text' and
start up a 3 step 'Text Import Wizard'.
(3)For step 1 select 'Delimited'. Then press 'Next'.
(4)For step 2, in the Delimiters box, ensure Tab and
Commas are enabled. Then press 'Next'.
(5)For step 3, ensure Column Data Format is 'General'.
Then press 'Finish'.
(6)The data will now be shown in consecutive columns,
and can be saved, printed and analysed as required.
To create a file, which can be opened directly from Windows
Explorer simply by double-clicking on it, change the file
extension to '.csv' before downloading and saving the file.
53
MAN0597ISSUE 09 - 12/05
2302M5030
4. OPERATION
4.9CALIBRATION
The accuracy of the Impact instrument must be functionally
checked with known concentration calibration gas before each
day’s use. If any sensor fails recalibrate the Impact before using,
or replace the cartridge. It is recommended that the instrument
be calibrated at least every 6 months. The calibration interval for
the CO
The Impact instrument provides three methods of calibration,
offering flexibility for the user. Flow calibration is the traditional
method and can be performed, following instructions displayed
on the instrument itself (Section 4.9.2 Flow Calibration –Instrument), or when the instrument is connected to a PC via a
basestation, on the PC display (See Section 4.9.3 FlowCalibration – PC). An alternative method for instruments fitted
with a combination of oxygen, flammable, carbon monoxide and
hydrogen sulfide sensors only is to use the Enforcer accessory.
For all other gases the flow calibration method must be used.
sensor shall not exceed 1 month.
2
CAUTIONS
Use of non-standard calibration gas and/or calibration
components when calibrating the Impact can lead to
dangerously inaccurate readings.
The calibration should be carried out in a well-ventilated
area to avoid contaminants.
Do not calibrate the CO
clean air after the apparatus has been exposed to CO
concentrations above the upper limit of the measuring
range. In this case leave the instrument in clean air
overnight before calibration.
4.9.1Contaminants
Oxygen sensors may be affected by prolonged exposure to
carbon dioxide, and it is recommended the Impact not be
subjected to use in atmospheres containing more than
channel if CO2 is indicated in
2
2
54
MAN0597ISSUE 09 - 12/05
2302M5030
4. OPERATION
25%v/v carbon dioxide (CO2). Oxygen sensors may read high
in the presence of carbon dioxide.
Flammable sensors may be affected by exposure to Silicone
substances; organic phosphorus containing compounds and
halogenated organic compounds. Although the flammable
sensors used in Impact and Impact Pro have a high resistance
to Hydrogen Sulfide (H
should be expected. Therefore it is strongly recommended
repeated gassing with H
should be followed by checking and if necessary re-calibration
of the Flammable sensor. The sensors on average will lose
approximately 20% of their sensitivity when subjected to forty
applications of H2S for a period of 1 minute each.
S) poisoning some loss in sensitivity
2
S or gas alarms on the H2S channel
2
The toxic sensors are designed to be gas specific, minimising
the effects of common cross-interfering gases. The table
below summarises the effect of various gases on the Carbon
Monoxide and Hydrogen Sulfide sensors.
Applied GasH
Acetone (1000ppm)00
Acetylene (40ppm)080
Ammonia (50ppm)00
Carbon Monoxide (50ppm)050
Carbon Dioxide (5000ppm)00
Chlorine (0.5ppm)00
Ethanol (2000ppm)03
Ethylene (100ppm)085
Hydrogen (100ppm)020
Hydrogen Sulfide (10ppm)100
S Response
2
(ppm)
CO Response
(ppm)
Iso-Propanol (200ppm)00
Nitric Oxide (25ppm)04
Nitrogen Dioxide (3ppm)00.5
Sulfur Dioxide (2ppm)00
55
MAN0597ISSUE 09 - 12/05
Password:
XX . .
Calibration
Span
Settings
Zero
2302M5030
4. OPERATION
4.9.2Flow Calibration – Instrument
The following equipment is required:
•cylinder of calibration gas (either certified individual gas per
sensor or a certified multi-gas mixture)
•flow adaptor: ensure that the tubing is fitted to the port
marked 'IN' (except for Cl
should be fitted to the port marked 'OUT').
•flowmeter/regulator set to 300ml/min
•tubing
cartridges where the tubing
2
To put the instrument into calibration mode the menus are
accessed. If configured it may be necessary to enter a
password before access to calibration is possible.
The keys must be pressed in the
correct sequence, otherwise the
password will be rejected. As each
key is entered the ‘.’ will be replaced
by ‘X’.
The calibration procedure requires a
zero, followed by a span. The span
requires specific concentration of
calibration gas that matches those
held in the settings. Select the
required option.
56
MAN0597ISSUE 09 - 12/05
Zero
O2
Flam
CO
H2S
All
Presswhen
in fresh air?
✓
2302M5030
4. OPERATION
When performing a flammable span gas concentration the
instrument will assume the use of the gas selected. However,
after calibration it will revert to the flammable gas that the
instrument is set to measure. No intervention by the user is
required.
Note: For Methane or Propane calibrations the following
If other values are required, adjustment of the
calibration point should be carried out as detailed in
Section 4.9.2.2 Span of this manual.
4.9.2.1 Zero
If calibrating the CO
section 4.9.
channel, refer to the CAUTIONS in
2
The display will show the sensors
fitted in the cartridge, and offer the
option to ‘hard’ zero an individual
channel or all four simultaneously.
O
is automatically highlighted.
2
Assuming flammable is selected by
pressing the
▼ and ✔ keys, the
display will show:
The zero must be performed in fresh,
uncontaminated air. Alternatively, a
supply of compressed air with an
oxygen content of 20.9%v/v can be
used.
57
MAN0597ISSUE 09 - 12/05
Zero
Please wait . . .
Zero
OK
Pressto continue✓
Zero
Fail
Pressto continue✓
or
Span
O2
Flam
CO
H2S
All
Apply gas at
0.3L per min
✓ - OK,- Abort
2302M5030
4. OPERATION
After pressing the ✔ button the instrument will display the
following whilst zeroing the sensor.
Once the zero is complete the instrument will display whether
the zero was successful.
If the zero was unsuccessful repeat the zero ensuring that the
instrument is in fresh air. If a second failure occurs the
cartridge should be changed. After a successful zero a span
should be performed.
4.9.2.2 Span
The display will show the sensors
fitted in the cartridge, and offer the
option to span an individual channel or
three simultaneously using a gas mix.
Having selected the gas the display
will change to show:
58
MAN0597ISSUE 09 - 12/05
Span
Please wait . . .
Span
OK
Pressto continue✓
Span
Fail
Pressto continue✓
or
2302M5030
4. OPERATION
Ensure that the Span Calibration Setting matches that of the
gas concentration being applied. If not, press the button to
change the setting.
The following table gives the current set allowable range of
Span Gas Concentrations.
Gas TypeAllowable Range
Methane20%LEL to 100%LEL
Carbon Monoxide80ppm to 500ppm
Hydrogen Sulfide15ppm to 50ppm
Carbon Dioxide0.5 %V/V to 2.1 %V/V
Chlorine2ppm to 10ppm
Ammonia10ppm to 100ppm
Sulfur Dioxide5ppm to 20ppm
Ensure that the correct concentration of the gas is available. If
not then press the button to change the settings.
If calibrating the CO
before pressing the
channel, turn on the gas 2 minutes
2
✔ button.
After pressing the
✔ button the
instrument will display this message
whilst spanning the sensor.
Once the span is complete the instrument will display whether
the span was successful.
If the span was unsuccessful repeat the span ensuring that the
calibration gas used is of the correct concentration, there is
sufficient gas in the cylinder and that the flow rate is correct.
59
MAN0597ISSUE 09 - 12/05
Settings
Flam:
CO:
H2S:
50
150
50
%LEL
ppm
ppm
Select Gas
Propane
Butane
Methane
Pentane
2302M5030
4. OPERATION
4.9.2.3 Calibration Gas Settings
WARNING
Verify that the concentration printed on the label of the
calibration gas cylinder that will be used is the same as the
concentration shown on the calibration gas settings
screen. Using the wrong concentration may cause
incorrect adjustment during calibration procedures, and
lead to dangerously inaccurate readings during normal
operation.
Select the channel to change. The
highlighted channel settings can then
be altered using the
▲ (increase) and
▼ (decrease) buttons, holding the
buttons down to change the setting
more rapidly. Press
✔ to store the
new settings value.
Assuming flammable is chosen the instrument will request
which calibration gas is being used, irrespective of what the
instrument is set to measure.
Zellweger Analytics recommends the following calibration gas
concentrations:
Cartridges manufactured before 1/2/2002 have the following
calibration gas limits:
MinimumMaximum
Flammable (%lel):25%100%
CO:100ppm500ppm
H
S:15ppm50ppm
2
:0.5% v/v3.0% v/v
CO
2
Cartridges manufactured after 1/2/2002 have the following
calibration gas limits:
MinimumMaximum
Flammable (%lel):25%100%
Flammable (%vol): 1.2%5.0%
CO:80ppm500ppm
H
S:15ppm50ppm
2
:0.5% v/v3.0% v/v
CO
2
Performing a manual calibration using settings lower than
these specified minima will result in a calibration failure.
4.9.3Flow Calibration – PC
For this function the user will require the Impact Configuration
Utility (ICU) PC software, serial cable, Base Station and Power
supply. These items can be purchased as part of the
DataLogging kit (available separately).
Place the Impact into the basestation and ensure that there is
power connected. Switch the Impact on, and then follow the
instructions on the computer.
The instrument will display the readings whilst the calibration
process is occurring and the symbol will be displayed in the
centre of the display.
The software is designed to be easy to use and has its own
manual and online help.
61
MAN0597ISSUE 09 - 12/05
Press to continue✓
Enforcer Calibration
2302M5030
4. OPERATION
Information, such as gas cylinder serial numbers, can also be
entered for traceability. Also, calibration reports can be
printed.
4.9.4Enforcer Calibration
This is designed for oxygen,
flammable, carbon monoxide and
hydrogen sulfide, using a special-totype gas cylinder. It offers quick,
simple and safe operation by any
user, due to the use of a patented low
pressure, low flow rate system.
Note that the Enforcer cylinder contains dangerous gases.
The instrument is switched on and the autozero procedure
performed. If the instrument was 'hard zeroed' (using the zero
on the calibration menu) prior to a successful Enforcer
calibration the unit will automatically update the next
calibration required by 180 days. If the unit was only 'soft
zeroed' (autozero during switch on) prior to a successful
Enforcer calibration the unit will not adjust the next calibration
date. If the Enforcer calibration fails then no adjustments are
made whether a hard or soft zero was performed.
The instrument, when placed correctly into the Enforcer, will
recognise it and will prompt the user to press the
✔ button
The instrument will then control the calibration process. The
user should hear the distinctive ‘click’ of the Enforcer gas
62
MAN0597ISSUE 09 - 12/05
OK
Pressto continue✓
Enforcer Calibration
Fail
Pressto continue✓
Enforcer Calibration
2302M5030
4. OPERATION
delivery solenoid valve. After completion, the instrument will
display whether calibration was successful.
or
Remove the instrument from the Enforcer. If the Enforcer
calibration failed repeat the process. If it fails again return the
instrument for a calibration under controlled conditions or
replace the cartridge.
The instrument automatically reverts to a flammable gas
setting of methane whilst in use with the Enforcer. On
completion it will revert to the flammable gas that the
instrument is set to measure. No intervention by the user is
required.
63
MAN0597ISSUE 09 - 12/05
2302M5030
5. FAULT FINDING AND PROCEDURES
5. FAULT FINDING AND PROCEDURES
There are two levels of fault within the instrument.
The first level is a warning, from which the user may be able to
recover, e.g. a low battery. The user must press the
to confirm the warning has been seen.
The second level is usually non-recoverable by the user, and
in some cases the only action available is to switch the
instrument off.
In both cases a warning/fault code number is given with a brief
description of the fault, and a contact name/telephone number
to enable the user to further assistance if required. Details of
the warning and fault codes are given in Appendix A.
✔ button
64
MAN0597ISSUE 09 - 12/05
2302M5030
6. ACCESSORIES
6. ACCESSORIES
WARNING
DO NOT ATTEMPT TO CHARGE THE BATTERY PACK
IN THE HAZARDOUS AREA.
The following accessories are available for use with Impact/
Impact Pro.
6.1BASESTATION
(Part Number: P2302B0800)
This unit provides a smart charging
facility for instruments powered by
rechargeable NiMH batteries. The
instrument is simply inserted into the
base station whereby it is fast
charged. The instrument controls the
charging process and will indicate
when it is fully charged, and then
change to a trickle charge mode to
keep the battery topped up until the
instrument is required.
The basestation also provides a facility to connect the
instrument to a PC, for downloading of logged data, viewing or
changing instrument configuration information, or as a means
of instrument calibration.
Ensure that the instrument is not inserted backwards into the
basestation. Firstly locate the lip at the front of the instrument
into the corresponding recess, and then latch the rear catch.
To release pull back the rear catch (see Section 2.3 Charge forfirst use).
If the instrument is switched on whilst placed in the basestation
the instrument will not reach full capacity (typically only 85% if
switched on for the entire fast charge period).
65
MAN0597ISSUE 09 - 12/05
2302M5030
6. ACCESSORIES
The basestation accepts an input of 12Vdc to 32Vdc.
Basestations may be daisy-chain linked. See Section 6.3
Basestation PSU Link Cable.
6.2BASESTATION PLINTH
(Part Number: P2302B0804)
The basestation can also be mounted
desk-mounted or vertically, singly or
stacked using the optional plinth. To
mount the basestation vertically
remove the base plate and rotate. To
mount in a vehicle remove the base
plate to reveal two mounting holes, for
screwing the basestation to the
vehicle bodywork or suitable framing.
6.3BASESTATION PSU LINK CABLE
(Part Number: 2302D0821)
Allows basestations to be linked
together, to a maximum of 5,
minimising cabling and power
sockets, and forming a multibank
charger.
Due to the potential variety of applications and installations
Zellweger Analytics do not supply such power supplies. The
requirements are 12Vdc to 32Vdc, rated at 500mA per
basestation.
WARNING
The standard Power Supply unit must not be used when
connecting multiple Base Stations. Overheating of the
Power Supply and ultimately failure will result.
66
MAN0597ISSUE 09 - 12/05
2302M5030
6. ACCESSORIES
6.4OFFLINE TRICKLE CHARGER
Permits off-instrument charging of 2 or
4 instrument battery packs in 14 hours.
It enables 24 hour availability of
rechargeable instruments with
purchase of additional battery packs.
Battery packs must be charged in
pairs. The trickle charger is supplied
with the required mains PSU.
The red LED indicates the batteries are receiving charge.
Please note that the battery packs must be charged in pairs,
and may be left inserted indefinitely until required.
Power SupplyPart Number
230Vac 50Hz Euro plug format2302B0730
230Vac 50Hz UK plug format2302B0731
120Vac 60Hz USA plug format2302B0732
240Vac 50Hz Australian plug format2302B0733
6.5POWER SUPPLY FOR BASESTATION
AND TRICKLE CHARGER
A range of suitable power adaptors is available to power the
base station and trickle charger:
Power SupplyPart Number
1230Vac 50Hz Euro plug format2302D0816
2230Vac 50Hz UK plug format2302D0818
3120Vac 60Hz USA plug format2302D0819
4240Vac 50Hz Australian plug format2302D0820
512V/24VDC vehicle charger lead2302D0815
67
MAN0597ISSUE 09 - 12/05
12345
2302M5030
6. ACCESSORIES
CAUTION
These individual Power Supplies will only supply power to
one Basestation or one Trickle Charger.
6.6ENFORCER
(Part Number: 2302B0831)
This test and calibration accessory is
only for use with instruments fitted
with any combination of oxygen,
flammable, carbon monoxide and
hydrogen sulfide sensors. Its
operation is described in Section 4.9.4
Enforcer Calibration. This is supplied
complete with an Enforcer disposable,
multigas cylinder.
For EMC/RFI Enforcer is tested to EN50270 Light Industrial
Type 2.
6.7ENFORCER GAS CYLINDER
(Part Number: 2302D0833)
The Enforcer requires a gas cylinder with pre-set
concentration values, and a specific valve to maintain the
required low pressure and flow rate, which enables the
Enforcer to operate quickly and simply.
68
MAN0597ISSUE 09 - 12/05
2302M5030
6. ACCESSORIES
6.8FLOW ADAPTOR
(Part Number: P2302B0810)
Allows aspirator to be fitted to the
instrument. It is also used when
calibrating using the instrument or the
PC. It is coloured grey to differentiate
it from the pump adaptor. See Section
6.12 Pump Adaptor Kit.
6.9METAL BELT CLIP
(Part Number: P2302D0826)
Supplied as standard. Enables the
instrument to be worn on a belt.
6.10INSTRUMENT WEBBING CLIP
(Part Number: P2302B0382)
Allows the instrument to be clipped to
harness webbing.
69
MAN0597ISSUE 09 - 12/05
2302M5030
6. ACCESSORIES
6.11BODY HARNESS KIT
(Part Number: P2302B0822)
Simply clips onto the rear of the
instrument. Allows the instrument to
be worn on the chest within the
breathing zone. Comes with a waist
strap and neck strap.
6.12PUMP ADAPTOR KIT
(Part Number: 2302B0814)
When plugged into the pump aperture
(2)
of an instrument fitted with the
pump, the pump will be automatically
activated by the instrument.
Subsequent removal of the adaptor
will switch the pump off. This
maximises the battery life of the
instrument, as well as prolonging the
life of the pump in the instrument. It is
coloured black to differentiate it from
the flow adaptor. See Section 6.8
Flow adaptor.
6.1310M SAMPLE TUBE KIT
(Part Number: 2302B0828)
Extends the sampling reach of the
instrument, and can be used hand
aspirator or Pump Adaptor Kit.
Supplied with a coupler it can be
extended beyond 10m (30') by
purchasing additional kits. The
maximum recommended length is
20m and the sample time should be
adjusted accordingly.
70
MAN0597ISSUE 09 - 12/05
2302M5030
6. ACCESSORIES
6.14EARPIECE
(Part Number: 2302B0841)
For noisy environments the earpiece
plugs into the audible aperture (4), to
allow any audible alarms or signals to
be heard. The vibrating alarm if fitted
will still operate. The instrument
audible alarm produces a high sound
level. Caution should be exercised
when using the earpiece.
6.15BASESTATION PC LINK CABLE
(Part Number: P2302D0807)
This is supplied to connect the
basestation to a standard 9-pin D-type
RS232 serial port on a PC or laptop,
running the PC software.
6.16SAFELINK CABLE ASSEMBLY
This is available in 4 different lengths
and allows Safelink enabled
instruments to be connected together
for communications, as described in
Section 4.5 Safelink.
10m (33') + 2 x Cable Restraints2302B0735
30m (100') + 2 x Cable Restraints2302B0736
50m (150') + 2 x Cable Restraints2302B0737
100m (300') + 2 x Cable Restraints2302B0746
2302B0746 is supplied with a cable reel and an adaptor.
71
MAN0597ISSUE 09 - 12/05
2302M5030
6. ACCESSORIES
6.17SAFELINK CABLE RESTRAINT CLIP
(Part Number: P2302B0713)
Clips the Safelink cable to a belt to
provide a strain relief.
6.18HAND ASPIRATOR KIT
(Part Number: 2302B0813)
Clips onto the flow hood to provide
manual aspiration.
The bulb should be squeezed at a rate
of one per second, until a stable
reading is obtained. As a guideline,
the following sample tube lengths will
require:
Length m
(approx ft)
1 (3)15
5 (15)20
10 (30)25
15 (50)30
30 (100)40
Sampling Time
(seconds)
72
MAN0597ISSUE 09 - 12/05
2302M5030
6. ACCESSORIES
6.19SENSING PROBE 1M
(Part Number: 2302B0847)
Used with the bulb operated hand
aspirator or the internal pump if fitted,
it allows sampling from above normal
height areas or into areas of low
accessibility. Ensure that the sections
of the sampling probe are tightened to
prevent dilution of the sample.
6.20BALL FLOAT
(Part Number: 2302B0846)
Clips to the end of the sampling tube
being used, and provides the dual
function of preventing liquid entering
the instrument and also breaking the
surface tension of any liquid to release
trapped gas.
73
MAN0597ISSUE 09 - 12/05
2302M5030
7. ROUTINE MAINTENANCE
7. ROUTINE MAINTENANCE
The Impact/Impact Pro is designed to operate almost
maintenance free under most conditions, except for the need
for regular calibration. However, it is recommended that on a
regular basis the instrument is cleaned and the filters changed.
7.1CLEANING
The instrument may need cleaning if exposed to dirty or hostile
conditions. This can be easily achieved by wiping over with a
damp cloth. Do not use bleaching products or products
containing silicon compounds as these can cause damage to
the sensors.
7.2FILTERS
The Goretex filter fitted under the grille cover is manufactured
from a water repelling material to protect the sensors from the
ingress of water and other contaminants. The life of this filter
is dependent on the amount of dust and viscous liquids in the
atmosphere. If it becomes dirty (discoloured) it will act as a
barrier and prevent the diffusion of the atmosphere onto the
sensors, and must be replaced. This is easily achieved by
removing the grille cover screws (Section 1.2 Productoverview (5)),
using the moulded detail to locate it.
discarding the old filter and fitting a new filter,
Note that for instruments fitted with a chlorine sensor a
stainless steel filter (part number P2302D0823) must be used
instead of the Goretex filter.
74
MAN0597ISSUE 09 - 12/05
2302M5030
7. ROUTINE MAINTENANCE
7.3BATTERY CHARGING / REPLACEMENT
WARNING
Battery Charging and replacement must not be carried out
in a hazardous or potentially dangerous area.
A mixture of dry cell and rechargeable battery packs must
not be used - the instrument safety system prevents
operation in such a situation, and the instrument displays
a fault message and activates the audible and visual
alarms.
If the instrument is fitted with a CO
to the instructions in Section 8 – Charging Procedure for
Impact Pro fitted with CO
Variant Cartridges.
2
Cartridge, please refer
2
INFORMATION
To maintain peak battery performance Zellweger Analytics
strongly recommend that the batteries be periodically
(every 4 to 6 months) fully discharged before recharging.
To do this, remove the Impact from the Base Station (if
fitted) and switch the Impact on. Leave the Impact to run
until it switches off. This may take up to 12 hours
depending on the state of charge of the batteries. A
Warning 14 (Battery Low) will be generated towards the
end of this procedure, which can be acknowledged or
ignored. Once the Impact has switched off, fit it into the
Base Station and allow the instrument to fully recharge
before next use.
In severe cases (e.g. when the instrument has not been
used for some time) it may be necessary to repeat this 2 or
three times to restore peak battery performance.
75
MAN0597ISSUE 09 - 12/05
2302M5030
7. ROUTINE MAINTENANCE
INFORMATION
Should the batteries in the Impact instrument become
excessively discharged (e.g. when unused for some time)
there may be occasions when the alarm LEDs and
sounder will operate when attempting to switch the
instrument on before recharging. In this instance please
place the instrument into the powered Base Station and
switch the instrument off when Fault 23 is displayed. The
Impact will then proceed to charge as normal.
7.3.1Rechargeable Battery Pack
CAUTION
The rechargeable battery packs are sealed units. Do not
attempt to remove the cells from the packs as this will
invalidate the safety certification.
Battery packs can be recycled by returning to the nearest
Zellweger Analytics Appointed Distributor.
Remove the battery packs if the instrument will not be used for
some time.
The basestation accessory is used to
charge rechargeable battery packs,
provided it is powered from an
appropriate power supply source. The
instrument is simply placed in the
basestation, which charges the packs
under the control of the instrument.
When first placed in the basestation all four red LEDs will
briefly light, to indicate charging will occur.
Whilst charging, two of the red LEDs will flash at a slow rate.
Once fully charged the red LEDs will be switched off and the
green LEDs will be constantly lit.
76
MAN0597ISSUE 09 - 12/05
2302M5030
7. ROUTINE MAINTENANCE
If the instrument is switched on when it is placed in the
basestation the charging progress will be displayed on the
instrument.
If a dry-cell instrument is inadvertently placed in the
basestation no damage will occur due to a built-in safety
system.
In the event that the rechargeable battery packs needs to be
replaced or the dry-cell adaptors are to be used to power the
instrument, the battery cover screws (see Section 1.2 Product
overview)
Product overview) and the sealed units removed.
Replacement rechargeable packs are inserted into the
compartments and the covers replaced.
are released using the built-in tool (see Section 1.2
The full fast charge time for a pair of fully discharged battery
packs is 7 hours indicated by two of the red alarm LEDs
flashing approximately every 2 seconds. After this period the
instrument will trickle charge the battery packs indefinitely
shown by the green LEDs being constantly illuminated.
7.3.2Dry cell
If dry cells are to be used the specific
adaptors must be used to comply with
certification. Ensure the orientation of
the individual cells is correct, and that
they are all of the following types:
Duracell MN1500
Energizer Intelligent E91.
WARNING
Use of any other dry-cell battery will render the intrinsic
safety approval of the instrument void.
77
MAN0597ISSUE 09 - 12/05
2302M5030
8. ROUTINE SERVICING
8. ROUTINE SERVICING
Apart from the maintenance described in Section 7. ROUTINE
MAINTENANCE the servicing requirements are limited to:
•Calibration
•Changing of the cartridge when required or indicated by the
instrument
•Replacing sensors in serviceable cartridges (i.e. non-
OFCH combinations where supported by the instrument)
8.1SERVICEABLE CARTRIDGES
Instruments that support serviceable cartridges can have the
individual sensors replaced, instead of replacing the whole
cartridge.
1
4
2
The procedure is as follows:
(1)Switch the instrument off and remove the front cover.
(2)Unscrew the central cartridge screw and remove the
cartridge.
(3)On the underneath undo the two screws (1).
(4)Lever out the PCB from the locating tabs (2), on the two
sides adjacent to the screws. Retain the sensor retainer
(3) if fitted.
(5)Pull out the PCB.
78
3
MAN0597ISSUE 09 - 12/05
2302M5030
8. ROUTINE SERVICING
(6)For all sensors except the oxygen, replacing the sensor
requires the user to unplug the existing sensor and
locate the new sensor in the same position.
Please note that CO2 sensors are supplied on a PCB,
which must be removed before fitting the sensor to the
cartridge.
(7)For the oxygen sensor the two screws (4) must be
undone. The tabs of the replacement sensor must be
securely screwed to headers. Ensure each sensor tab is
secured to the correct header.
(8)Replace the PCB in the housing, until it clicks into place.
(9)Replace the two screws (1), ensuring the retainer is
fitted (3) and insert the cartridge into the instrument.
(10)Check the front cover membrane and if necessary
replace it.
(11)Replace the front cover and wait at least 20 minutes.
Then switch the instrument on.
(12)The instrument must be calibrated before use.
79
MAN0597ISSUE 09 - 12/05
2302M5030
8. ROUTINE SERVICING
Special Notes for Cl2 Cartridges
It is important that you read all of the notes below before
installation or use of your new cartridge.
Installation of your Cl
Cartridge
2
Note: Cartridge Installation or Replacement must never be
carried out in a hazardous area.
(1)If a cartridge was previously fitted to the Impact into
which the Cl
cartridge is to be installed, remove the
2
Front Cover Grille using the Allen Key provided in the
bottom of the Impact. Unscrew the fixing in the centre of
the cartridge and lift the cartridge from the Impact.
(2)Place the Cl2 cartridge into the Impact and screw the
cartridge fully home using the Allen Key. Guide the
cartridge if necessary using your fingers to ensure that
it is seated correctly.
(3)Replace the Pump Seal with the one provided with your
cartridge.
(4)Remove the existing Filter Assembly from the inside of
the Front Cover Grille by peeling it away. Take the
Stainless Steel Filter assembly (supplied in the
Cartridge Kit) and remove the protective backing paper.
Locate the Filter assembly onto the locating pegs on the
inside of the Front Cover Grille and lightly press around
the edge of the gasket to secure.
(5)Replace the Front Cover Grille and refit the Allen Key
into the bottom of the Impact.
(6)To complete installation of the cartridge allow at least
20 minutes before operating the Impact.
(7)After this period of time switch the Impact on and accept
the New Configuration when prompted to do so.
80
MAN0597ISSUE 09 - 12/05
2302M5030
8. ROUTINE SERVICING
Using your Cl2 Cartridge
Please ensure that the following points are noted when the
cartridge is in service.
(1)Always use PTFE tubing (maximum length of 500mm)
and a flow rate of 500ml/min when calibrating the Cl
channel.
(2)Apply all gases to the port marked OUTLET (not INLET)
on the Flow Housing.
(3)With regard to the actual sensors fitted, please perform
span calibrations on the sensors of the Impact cartridge
in the following order: Cl
, NH3, H2S, CO, Flammable.
2
2
(4)When performing a Span Calibration on the Cl
it is good practice to flow Cl
calibration span gas
2
channel
2
through connected tubing for a minimum of 15 minutes
without the flow adapter fitted to the Impact before
attempting a span calibration. Failure to do so may
result in a false calibration of the Cl
channel.
2
(5)Do not apply Span Calibration gas to the Impact for
greater than 5 minutes. Again failure to do so may result
in a false calibration of the Cl
channel.
2
(6)This cartridge is not suitable for pumped mode
operation.
(7)Due to the open nature of the Stainless Steel filter, there
is a greater risk of water damage occurring to the
instrument. Every attempt should be made to keep the
instrument dry.
81
MAN0597ISSUE 09 - 12/05
2302M5030
8. ROUTINE SERVICING
(8)Refer to the CAUTIONS given in section 1.1 of this
manual for details on recommended daily checks on
your gas detector.
Applied GasEffect on CO
Sensor
CO (100ppm)100ppm0ppm0%v/v0ppm35ppm1ppm
S (40ppm)0ppm40ppm0%v/v-25ppm150ppm1ppm
H
2
CO2 (0.5%v/v)0ppm0ppm0.5%v/v0ppm0ppmTBA
Cl2 (5ppm)0ppm0ppm0%v/v5ppm0ppm-2ppm1
NH3 (50ppm)0ppm0ppm0%v/v-2ppm150ppmTBA
Hydrogen
(1000ppm)
SO2 (10ppm)0ppm0ppm0%v/v2ppm0ppm10ppm
1
Reading displayed as -0ppm.
200ppm0ppm0%v/vTBA35ppmTBA
Effect on
H2S Sensor
Effect on
CO2 Sensor
Effect on Cl
Sensor
Effect on NH3
2
Sensor
Effect on SO2
Sensor
82
MAN0597ISSUE 09 - 12/05
2302M5030
8. ROUTINE SERVICING
Special Notes for NH3 Cartridges
It is important that you read all of the notes below before
installation or use of your new cartridge.
Installation of your NH
Cartridge
3
To install the cartridge, follow the instructions below.
Note: Cartridge Installation or Replacement must never be
carried out in a hazardous area.
(1)If a cartridge was previously fitted to the Impact into
which the NH
cartridge is to be installed, remove the
3
Front Cover Grill using the Allen Key provided in the
bottom of the Impact. Unscrew the screw in the centre
of the cartridge and lift the cartridge from the Impact.
(2)Place the NH3 cartridge into the Impact and screw the
cartridge fully home using the Allen Key. Guide the
cartridge if necessary using your fingers to ensure that
it is seated correctly.
(3)Replace the Pump Seal with the one provided with your
cartridge and replace the Front Cover. Refit the Allen
Key into the bottom of the Impact.
(4)Allow 20 minutes for the sensors to stabilise before first
use.
(5)To complete installation of the cartridge, switch the
Impact on and accept the New Configuration when
prompted to do so.
Using your NH
Cartridge
3
Please ensure that the following points are noted when the
cartridge is in service.
(1)The Ammonia channel has an extended warm-up time.
Although the other channels of the cartridge will be
83
MAN0597ISSUE 09 - 12/05
2302M5030
8. ROUTINE SERVICING
reading correctly, the detector should not be used to
measure Ammonia until the 'warm up' message has
disappeared.
(2)The operating temperature range of the cartridge
defaults to that of the lowest sensor fitted. For NH
is -20°C to +40°C.
(3)Always use PTFE tubing (maximum length of 500mm)
when calibrating the NH
channel.
3
this
3
(4)When performing a Span Calibration on the NH
channel it is good practice to flow NH3 calibration span
gas through connected tubing for a minimum of
15 minutes without the flow adapter fitted to theImpact before attempting a span calibration. Failure to
do so may result in a false calibration of the NH
channel.
(5)The accuracy of the instrument must be functionally
checked with known concentration calibration gas
before each day's use. If any sensor fails, recalibrate the
Impact before using or alternatively replace the
cartridge. Please refer to the Impact User Manual on
general instructions on how to calibrate.
(6)Apply calibration span gas at a rate of 300ml/min for a
period of 5 minutes before starting the span calibration
on the sensor.
(7)With regard to the actual sensors fitted, please perform
span calibrations on the sensors of the Impact cartridge
in the following order: Cl
, NH3, H2S, CO, Flammable.
2
3
3
(8)Purge the system with air for a period of 15 minutes
between Cl
and NH3 calibrations to avoid interaction
2
between these two gases.
(9)The sensors fitted to Impact may give responses to
gases other then the target gas. The table below gives
typical sensor responses to various gases and should
84
MAN0597ISSUE 09 - 12/05
2302M5030
8. ROUTINE SERVICING
be used as a guide only. Note: Always use the target
gas to perform sensor calibrations.
(10)Your attention is drawn to the cross-sensitivity of the
NH3 sensor in the presence of H2S. This is normal
operation and is due to the chemistry of the NH
sensor.
3
Applied GasEffect on CO
Sensor
CO (100ppm)100ppm0ppm0%v/v0ppm35ppm1ppm
S (40ppm)0ppm40ppm0%v/v-25ppm150ppm1ppm
H
2
CO2 (0.5%v/v)0ppm0ppm0.5%v/v0ppm0ppmTBA
Cl2 (5ppm)0ppm0ppm0%v/v5ppm0ppm-2ppm1
NH3 (50ppm)0ppm0ppm0%v/v-2ppm150ppmTBA
Hydrogen
(1000ppm)
SO2 (10ppm)0ppm0ppm0%v/v2ppm0ppm10ppm
1
Reading displayed as -0ppm.
200ppm0ppm0%v/vTBA35ppmTBA
Effect on
H2S Sensor
Effect on
CO2 Sensor
Effect on Cl
Sensor
2
NH3Sensor
Effect on
Effect on
SO2Sensor
85
MAN0597ISSUE 09 - 12/05
2302M5030
8. ROUTINE SERVICING
Special Notes for SO2 Cartridges
It is important that you read all of the notes below before
installation or use of your new cartridge.
Installation of your SO
Cartridge
2
Note: Cartridge Installation or Replacement must never be
carried out in a hazardous area.
(1)If a cartridge was previously fitted to the Impact Pro into
which the SO
cartridge is to be installed, remove the
2
Front Cover Grill using the Allen Key provided in the
bottom of the Impact Pro. Unscrew the screw in the
centre of the cartridge and lift the cartridge from the
Impact Pro.
(2)Place the SO
cartridge into the Impact Pro and screw
2
the cartridge fully home using the Allen Key. Guide the
cartridge if necessary using your fingers to ensure that
it is seated correctly.
(3)Replace the Pump Seal with the one provided with your
cartridge and replace the Front Cover. Refit the Allen
Key into the bottom of the Impact Pro.
(4)Allow 20 minutes for the sensors to stabilise before first
use.
(5)To complete installation of the cartridge, switch the
Impact Pro on and accept the New Configuration when
prompted to do so.
Using your SO
Cartridge
2
Please ensure that the following points are noted when the
cartridge is in service.
(1)The operating temperature range of the cartridge
defaults to that of the lowest sensor fitted.
86
MAN0597ISSUE 09 - 12/05
2302M5030
8. ROUTINE SERVICING
(2)Always use PTFE tubing (maximum length of 500mm)
when calibrating the SO
channel.
2
(3)When performing a Span Calibration on the SO
channel it is good practice to flow SO2 calibration span
gas through connected tubing for a minimum of
15 minutes without the flow adapter fitted to theImpact before attempting a span calibration. Failure to
do so may result in a false calibration of the SO
channel.
(4)The accuracy of the instrument must be functionally
checked with known concentration calibration gas
before each day's use. If any sensor fails, recalibrate the
Impact before using or alternatively replace the
cartridge. Please refer to the Impact User Manual on
general instructions on how to calibrate.
(5)Apply calibration span gas at a rate of 300ml/min for a
period of 1 minute before starting the span calibration on
the sensor.
(6)With regard to the actual sensors fitted, please perform
span calibrations on the sensors of the Impact cartridge
in the following order: Cl
, NH3, H2S, SO2, CO,
2
Flammable.
2
2
(7)If the cartridge contains either NH
or H2S sensors
3
purge the system with air for a period of 15 minutes
between NH
or H2S and SO2 calibrations to avoid
3
interaction between these gases.
(8)The sensors fitted to Impact may give responses to
gases other then the target gas. The table below gives
typical sensor responses to various gases and should
87
MAN0597ISSUE 09 - 12/05
2302M5030
8. ROUTINE SERVICING
be used as a guide only. Note: Always use the target
gas to perform sensor calibrations.
Applied GasEffect on CO
Sensor
CO (100ppm)100ppm0ppm0%v/v0ppm35ppm1ppm
S (40ppm)0ppm40ppm0%v/v-25ppm150ppm1ppm
H
2
CO2 (0.5%v/v)0ppm0ppm0.5%v/v0ppm0ppmTBA
Cl2 (5ppm)0ppm0ppm0%v/v5ppm0ppm-2ppm1
NH3 (50ppm)0ppm0ppm0%v/v-2ppm150ppmTBA
Hydrogen
(1000ppm)
SO2 (10ppm)0ppm0ppm0%v/v2ppm0ppm10ppm
1
Reading displayed as -0ppm.
200ppm0ppm0%v/vTBA35ppmTBA
Effect on
H2S Sensor
Effect on
CO2 Sensor
Effect on Cl
Sensor
Effect on NH
2
Sensor
Effect on SO2
3
Special Notes for CO2 Cartridges
It is important that you read all of the notes below before
installation or use of your new cartridge.
Installation of your CO
The cartridge is supplied on a Bias PCB to ensure that the
sensor is ready to use when installed.
Cartridge
2
Sensor
To install the cartridge, follow the instructions below.
Note: Cartridge Installation or Replacement must never be
carried out in a hazardous area.
(1)If a cartridge was previously fitted to the Impact into
which the CO
cartridge is to be installed, remove the
2
Front Cover Grill using the Allen Key provided in the
bottom of the Impact. Unscrew the screw in the centre
of the cartridge and lift the cartridge from the Impact.
(2)Again using the Allen Key provided in the base of the
Impact instrument, unscrew the screw in the centre of
the new CO
cartridge. Do not attempt to remove the
2
screw from the bottom of the Bias PCB. Remove the
Bias PCB from the Cartridge.
88
MAN0597ISSUE 09 - 12/05
1
9
3
6
ON
20 min
2302M5030
8. ROUTINE SERVICING
(3)Place the CO2 cartridge into the Impact and screw the
cartridge fully home using the Allen Key. Guide the
cartridge if necessary using your fingers to ensure that
it is seated correctly.
(4)Replace the Pump Seal with the one provided with your
cartridge and replace the Front Cover. Refit the Allen
Key into the bottom of the Impact.
(5)To complete installation of the cartridge, switch the
Impact on and accept the New Configuration when
prompted to do so. Note: This must be done
immediately after installation in order to maintain correct
bias on the CO
sensor.
2
(6)Allow 20 minutes for the sensors to stabilise before use.
Note that the Impact may be in alarm condition during
this time.
Using your CO
Cartridge
2
Please ensure that the following points are noted when the
cartridge is in service.
(1)Always keep the instrument batteries charged or keep
fresh alkaline cells fitted when the cartridge is installed.
Never allow the batteries to become drained.
(2)The operating temperature range of this cartridge is 0°C
to +40°C. Do not use the cartridge outside this range.
The cartridge may be stored safely at temperatures
down to -10 °C.
89
MAN0597ISSUE 09 - 12/05
2302M5030
8. ROUTINE SERVICING
(3)The CO2 sensor is suitable for use in Barometric
Pressures of 1013mBar ± 10%. Over this pressure
range, the reading is directly proportional to Barometric
Pressure.
(4)The calibration interval for the CO
sensor shall not
2
exceed 1 month. Owing to the characteristics of the
sensor a significant increase of reading can be
CO
2
expected at temperatures below +10°C, and a
significant decrease at temperatures above +30 °C. To
ensure safe operation, always calibrate the instrument
between +10 °C and +30 °C.
(5)The accuracy of the instrument must be functionally
checked with known concentration calibration gas
before each day's use. If any sensor fails, recalibrate the
Impact before using or replace the cartridge. Please
refer to the Impact User Manual on how to calibrate.
(6)The sensors fitted to Impact may give responses to
gases other then the target gas. The table below gives
typical sensor responses to various gases and should
be used as a guide only. Note: Always use the target
gas to perform sensor calibrations.
Applied GasEffect on CO
Sensor
CO (100ppm)100ppm0ppm0%v/v0ppm35ppm1ppm
H2S (40ppm)0ppm40ppm0%v/v-25ppm150ppm1ppm
CO2 (0.5%v/v)0ppm0ppm0.5%v/v0ppm0ppmTBA
Cl2 (5ppm)0ppm0ppm0%v/v5ppm0ppm-2ppm1
NH3 (50ppm)0ppm0ppm0%v/v-2ppm150ppmTBA
Hydrogen
(1000ppm)
SO2 (10ppm)0ppm0ppm0%v/v2ppm0ppm10ppm
1
Reading displayed as -0ppm.
200ppm0ppm0%v/vTBA35ppmTBA
Effect on
S Sensor
H
2
Effect on
CO2 Sensor
Effect on Cl2
Sensor
Effect on NH3
Sensor
Effect on SO2
90
Sensor
MAN0597ISSUE 09 - 12/05
2302M5030
8. ROUTINE SERVICING
Charging Procedure for Impact Pro fitted with CO2 Variant
Cartridges
If using Rechargeable Ni-MH Battery Packs, charging should
always be carried out with the batteries removed from the
instrument. This can be achieved by using the Zellweger
Analytics Off-Line Charger. Part numbers of the chargers are
given in this section.
(1)Remove the exhausted batteries using the special tool
found in the base of the instrument.
(2)Immediately place the instrument, with the batteries
removed, onto the Base Station, which is connected to
a Power Supply Unit (PSU). This will ensure that the
CO
sensor is correctly biased while the Battery Packs
2
are not present. Alternatively, fit fully charged batteries.
(3)Place the exhausted batteries into the Off-Line Charger,
ensuring that they are charged in pairs (one each side
of the red 'Charge' LED). The Off-Line Charger can
accept two pairs (4 total) of Rechargeable Battery
Packs. The red 'Charge' LED will illuminate while the
batteries as present.
(4)After charging is complete (12 hours minimum), remove
the Battery Packs from the Off-Line Charger. Remove
the instrument from the Base Station and insert the
Battery Packs. Tighten the retaining screw using the
special tool.
(5)The instrument is now ready for use.
Note: The instrument does not have to be switched on during
this procedure, however if it is switched on a Warning
105 (see Section 11 Appendix A) will be shown. This
should be accepted. In addition Fault 23 may be
reported when the instrument is removed from the Base
Station. If this occurs, the instrument should be
switched off and back on again to reset the instrument.
91
MAN0597ISSUE 09 - 12/05
2302M5030
8. ROUTINE SERVICING
Off-Line Charger and Spare Battery Part Numbers
The following are the part numbers for the Off-Line Charger.
Each variant is supplied complete with the relevant PSU for the
destination country shown.
•Off-Line Charger Kit (Europe)[2302B1415]
•Off-Line Charger Kit (UK)[2302B1414]
•Off-Line Charger Kit (USA)[2302B1416]
•Off-Line Charger Kit (Australia)[2302B1417]
•Pair of Ni-MH Batteries
(UK, Europe, Australia) [2302B0842]
•Pair of Ni-MH Batteries (USA)[2302B2015]
Using Alkaline Dry Cells
If using Alkaline Cells (Duracell MN1500 or Energizer E91)
and they require replacing, ensure that the batteries are not left
out of the instrument for more than 15 minutes. This is to
ensure correct biasing of the CO
sensor.
2
WARNING
The following points should be noted.
(1)The Bias PCB supplied with your CO
Cartridge contains a non-rechargeable lithium
battery. Do not attempt to recharge the battery
under any circumstances.
(2)Do not reuse the Bias PCB.
2
(3)Do not destroy the Bias PCB in fire.
(4)Dispose of the Bias PCB and cartridge
packaging in accordance with Local
Regulations.
92
MAN0597ISSUE 09 - 12/05
2302M5030
8. ROUTINE SERVICING
Special Notes for NO2 Cartridges
It is important that you read all of the notes below before
installation or use of your new cartridge.
Installation of your NO
Cartridge
2
Note: Cartridge Installation or Replacement must never be
carried out in a hazardous area.
(1)If a cartridge was previously fitted to the Impact Pro into
which the NO
cartridge is to be installed, remove the
2
Front Cover Grill using the Allen Key provided in the
bottom of the Impact Pro. Unscrew the screw in the
centre of the cartridge and lift the cartridge from the
Impact Pro.
(2)Place the NO
cartridge into the Impact Pro and screw
2
the cartridge fully home using the Allen Key. Guide the
cartridge if necessary using your fingers to ensure that
it is seated correctly.
(3)Replace the Pump Seal with the one provided with your
cartridge and replace the Front Cover. Refit the Allen
Key into the bottom of the Impact Pro.
(4)Allow 20 minutes for the sensors to stabilise before first
use.
(5)To complete installation of the cartridge, switch the
Impact Pro on and accept the New Configuration when
prompted to do so.
Using your NO
Cartridge
2
Please ensure that the following points are noted when the
cartridge is in service.
93
MAN0597ISSUE 09 - 12/05
2302M5030
8. ROUTINE SERVICING
(1)The operating temperature range of the cartridge
defaults to that of the lowest sensor fitted. The
temperature range of the NO
(2)Always use PTFE tubing (maximum length of 500mm)
when calibrating the NO
2
(3)When sampling remotely (i.e. using the internal sample
pump or hand aspirator together with tubing) users
should be aware of increased sample times. For NO
the sample time will increase by approximately
90 seconds for every 10m of tubing used.
Sensor is -20°C to +50°C.
2
channel.
2
(4)When performing a Span Calibration on the NO
channel it is good practice to flow NO2 calibration span
gas through connected tubing for a minimum of
15 minutes without the flow adapter fitted to theImpact before attempting a span calibration. Failure to
do so may result in a false calibration of the NO
channel.
(5)The accuracy of the instrument must be functionally
checked with known concentration calibration gas
before each day's use. If any sensor fails, recalibrate the
Impact before using or alternatively replace the
cartridge. Please refer to the Impact User Manual on
general instructions on how to calibrate.
(6)Apply calibration span gas at a rate of 300ml/min for a
period of 1 minute before starting the span calibration on
the sensor.
(7)With regard to the actual sensors fitted, please perform
span calibrations on the sensors of the Impact cartridge
in the following order: Cl
, NH3, H2S, NO2, SO2, CO,
2
Flammable.
2
2
(8)If the cartridge contains NH
the system with air for a period of 15 minutes between
NH
, H2S or Cl2 and NO2 calibrations to avoid
3
interaction between these gases.
94
, H2S or Cl2 sensors purge
3
MAN0597ISSUE 09 - 12/05
2302M5030
8. ROUTINE SERVICING
(9)The sensors fitted to Impact may give responses to
gases other then the target gas. The table below gives
typical sensor responses to various gases and should
be used as a guide only. Note: Always use the target
gas to perform sensor calibrations.
Applied GasEffect on
CO Sensor
CO (100ppm)100ppm0ppm0%v/v0ppm35ppm1ppm0ppm
H2S (40ppm)0ppm40ppm0%v/v-25ppm150ppm1ppm-3.2ppm1
CO2 (0.5%v/v)0ppm0ppm0.5%v/v0ppm0ppmTBATBA
Cl2 (5ppm)0ppm0ppm0%v/v5ppm0ppm-2ppm15ppm
NH3 (50ppm)0ppm0ppm0%v/v-2ppm150ppmTBATBA
Hydrogen
(1000ppm)
SO2 (10ppm)0ppm0ppm0%v/v2ppm0ppm10ppm0ppm
NO2 (10ppm)0ppm0ppm0%v/v0ppm-5ppm1-10ppm110ppm
1
Reading displayed as -0ppm.
200ppm0ppm0%v/vTBA35ppmTBATBA
Effect on
H2S
Sensor
Effect on
CO2
Sensor
Effect on
Cl2 Sensor
Effect on
NH3 Sensor
Effect on
SO2 Sensor
Effect on
NO2 Sensor
95
MAN0597ISSUE 09 - 12/05
2302M5030
9. SPARE PARTS
9. SPARE PARTS
The following spare parts are available from Zellweger
Analytics authorised Service Centres, in addition to the
accessories listed in Section 6. ACCESSORIES.
2302B2015
(23)USA Front Cover Grill Assy 2302B2017
Note: In order to meet Certification Requirements, fit only
Zellweger Analytics component approved flammable
sensor.
96
MAN0597ISSUE 09 - 12/05
2302M5030
10. GLOSSARY
10. GLOSSARY
BASEEFABritish Approvals Service for Electrical
Equipment in Flammable Atmospheres
– UK Safety Certification
Catalytic SensorFor detection of combustible gases.
These are made of an electrically
heated platinum wire coil, covered first
with a ceramic base such as alumina
and then with a final outer coating of
palladium or rhidium catalyst dispersed
in a substrate of thorium.
CEIndicates compliance to all relevant
European directives
CellAn individual sensor
CENELECComite Europeen de Normalisation
Electrotechnique – European Safety
Certification
COSHHControl of Substances Hazardous to
Health
CSACanadian Standards Association
dBADecibels, relative to the A weighting
scale (as affected by the human ear).
Electrochemical SensorA gas sensitive electrode, formed by a
permeable membrane and special
electrolyte.
EMCElectromagnetic compatibility.
ESDElectrostatic discharge.
Gas AnalyserNormally refers to equipment used to
measure extremely small
concentrations of gas (low or sub ppm)
or one specific gas in the presence of
several others.
97
MAN0597ISSUE 09 - 12/05
2302M5030
10. GLOSSARY
Gas Detector or Monitor Refers to equipment designed to alert the
user to potentially hazardous concentrations
of gas in the monitored atmosphere.
Hard ZeroWhen a hard zero is performed using
either the built-in calibration menu or
the PC software then any adjustments
will remain in place when the
instrument is turned off. Performing a
hard zero followed by an Enforcer
calibration, instrument menu span or
PC calibration will adjust the calibration
and change the calibration due date.
Hazardous AreasAreas where there is the possibility of
the presence of an explosive mixture of
flammable gas or vapour and air are
known as ‘Hazardous’ and other areas
as ‘safe’ or ‘non-hazardous’. Any
electrical equipment used in hazardous
areas must be tested and approved to
ensure that, in use even under fault
conditions, it can not cause an
explosion.
In Europe, hazardous areas are defined
as follows:
Zone 0: An area where an explosive
mixture is likely to be present at all times,
under normal operating conditions.
Zone 1: An area where an explosive
mixture is likely to occur in normal
operation.
Zone 2: An area where an explosive
mixture is not likely to occur in normal
operation, and if it does it is only for
short periods.
98
MAN0597ISSUE 09 - 12/05
2302M5030
10. GLOSSARY
In the US, hazardous areas are
classified in 2 divisions:
Division 1: Equates to Zone 0 and Zone 1
Division 2: Equates to Zone 2
Intrinsically safe(approval by an appropriate authority to
use the instrument in hazardous area).
IPIngress Protection – a measure of
protection against the ingress of dust
and water
ISIntrinsically Safe (approval by an
appropriate authority to use the
instrument in hazardous area).
LEDLight emitting diode.
LELLower Explosive Limit – is the lowest
concentration of ‘fuel’ in air which will
burn and for most flammable gases and
vapours it is less than 5% by volume.
LEL%Percentage of the Lower Explosive
Limit (for example, 10% LEL of
methane is approx 0.5% by volume).
LTELLong Term Exposure Limit. The 8 hour
LTEL is the time-weighted average
concentration for a normal 8 hour day to
which most workers may be repeatedly
exposed, day after day, without
adverse effect.
OzOunce (weight).
PeakMaximum, or minimum, measurement
since switch on.
PellistorRegistered trade name for a
commercial device – A very small
sensing element used in catalytic
99
MAN0597ISSUE 09 - 12/05
2302M5030
10. GLOSSARY
sensors and sometimes also called a
‘bead’ or a ‘siegistor’.
Poison resistantCapability of a catalytic sensor to
reduce the effect of inhibiting
substances or contaminants, such as
silicones.
PPBParts per billion concentrations in the
atmosphere.
PPMParts per million concentrations in the
atmosphere.
RFIRadio frequency interference.
Safe AreaWork area in which there is no danger
of contamination with explosive gases.
Soft ZeroWhen a soft zero is performed (i.e. the
autozero on startup) any adjustments
will only remain in place whilst the
instrument remains switched on.
Performing a soft zero (autozero)
followed by an Enforcer calibration or
instrument calibration menu span (i.e.
not a menu zero first) will adjust the
calibration whilst the instrument
remains switched on but will not change
the calibration due date.
STELShort Term Exposure Limit, usually
monitored over 15 minute periods
TWATime-Weighted Average
UELUpper Exposure Limit
ULUnderwriters Laboratories (USA)
%VOLConcentration of gas, measured in
percentage by volume.
%V/VAnother way of representing %VOL
100
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.