Hill-Rom Resuscitaire User manual

OPERATING PRECAUTIONS

GENERAL PRECAUTIONS
.
Federal Law restricts this device to sale by or on order of a physician.
.
Infant radiant warmers should be used only by properly trained personnel as directed by an appropriately
qualified physician aware of currently known hazards and benefits.
.
The functional checkout procedure should be performed before each use and after disassembly for cleaning, servicing or maintenance. Refer to qualified service personnel if the unit does not perform as
specified.
.
The Cart/Bassinet End and Side panels cannot be used for pushing or pulling the Cart/Bassinet and or Warmer.
.
Do not leave the infant unattended in the Cart/Bassinet when the side panels orthefront panel arefolded
down.
.
To avoid overheating or underheating, skin temperature must be continuously monitored and controlled
either manually or automatically. Rectal temperature should never be used to control skin temperature.
.
To avoid overheating or underheating when operating in manual mode, observe the infant constantly and
monitor the temperature using the temperature probe supplied with the equipment or another electronic
thermometer.
.
The skin temperature sensing probe must be in direct contact with the skin to provide accurate monitor-
ing of the infant’s skin temperature. Failure to maintain direct skin contact can result in overheating and possible burning. Check infant’s condition at least every fifteen minutes for correct Sensor attachment, reddened skin areas, and proper skin temperature.
.
The skin temperature sensing probe should be placed at least 1.5 inches from any transcutaneous moni-
(TcPOz
tor should not be placed on an area previously used by a
.
To avoid overheating the skin, the location of the skin temperature probe must be such that the skin around the Sensor is in direct line with the radiation from the warmer. Do not place anything between the radiant warmer and the infant that will interfere with the radiation from the warmer.
.
Radiant warming increases insensible water loss. Appropriate measures to maintain proper fluid bal-
.
For effective heating, make sure the Warmer Module is positioned on the center axis and locked with its detent.
.
This Warmer is intended for use with the Cart/Bassinet. If the Warmer is used to warm the mother im- mediately after delivery, all warnings and instructions should be followed with particular regard to the heater-to-mattress distance and skin probe location. Avoid placement of objects near the mother which can absorb heat and become hot to the touch.
.
Compressed gas cylinders, such as oxygen cylinders, can become hazardous projectiles if the gas is
released rapidly due to damage or other causes. Cylinders must be securely fastened.
.
To avoid overheating the Warmer, do not place objects (equipment, blankets, clothing or sterile packs) on top of the Warmer.
TcPCOz)
or
probe to prevent false temperature indications. The skin temperature probe
TcPOz
or
COPYRIGHT@3
TcPC02
1998, HILL-ROM AIR-SHIELDS, INC.
probe.
OPERATING PRECAUTIONS (Continued)
l
Avoid placement of objects between the infant and the Warmer Module that can block heat transfer or absorb heat. These objects may be heated directly by the heater and can cause skin burns.
a
Temperature uniformity (per IEC 601-2-21) across the mattress surface may not be maintained when the Bassinet is tilted in the
l
During calibration, inspect the secondary reflector directly under the warmer heater element for particles.
If particles are present, replace the heater element. The life expectancy of the heater element is 1000
hours of operation.
0
Evidence that the safety certification of the ACCESSORY has been performed in accordance with the appropriate IEC 601-l and/or IEC 601-l-l harmonized national standard.
RESUSCITATION PRECAUTIONS
A one-way valve is installed at the Patient Outlet connection. This valve opens when pressure in the hose delivering gas to the patient falls below -4 cm unlikely event of failure of the gas supply.
5-
and 1 O-degree positions.
Its purpose is to allow patient inspiration in the
H20.
A humidifier, when used, must be placed between the
Patient Outlet
connection and the patient circuit.
DO NOT PLACE A HUMIDIFIER IN THE SUPPLY LINE. Humidifier must have low series flow resistance and not be placed between exhalation valve and patient
airway. Gas going to the patient should not be super-saturated as evidenced by excessive condensation in the
tubing. Condensation droplets on the inner walls of the tubing are normal. Periodic blood gas measurements should be made to ensure proper levels of ventilation. Always confirm the airway pressure relief valve setting before patient use. Confirm that the oxygen/air blender control of the Blended Gas Supply Module is correctly set prior to
use. Confirm that the patient circuit contains all the parts needed prior to use. Ensure that the patient breathing circuit connections are secure and free of obstructions. The Aux Outlet circuit does not provide adjustable pressure limiting.
HzO.
Auxiliary Gas Supply Pressure is internally limited to 160 cm
Always use a patient resuscitation
circuit that includes appropriate pressure relief. Always monitor Airway Pressure and/or provide appropriate pressure relief during infant resuscitation.
Confirm that the oxygen supply is turned off and that the equipment is disconnected from the oxygen supply when performing cleaning and maintenance procedures; a fire and explosion hazard exists when performing cleaning and/or maintenance procedures in an oxygen-enriched environment.
Gas supplies
(02 and Air) should always be clean and dry. Water/Trap Filters should be used in the sup-
ply lines if necessary. All hoses should be securely fastened to fittings. Hand-tighten to avoid damage to fittings. Gas supplies should be maintained at 40 to 75 psi.
Flow restrictions (e.g., flowmeter, valve, etc.) must not be placed in the supply line. If a compressor is used as the air source, steps should be taken to filter and dehumidify the room air
before introducing it into the Gas Supply Module.
_ ii _
OPERATING PRECAUTIONS (Continued)
ELECTRICAL PRECAUTIONS a
An electrical shock hazard exists within the Warmer Module and Controller. Any substitution of compo-
nents within the Controller or Warmer Module may impair the intrinsic safety of the unit. Service should
be performed by qualified personnel.
0
Connect the power cord only to a properly grounded receptacle that is approved for hospital use and of the correct voltage. DO NOT USE EXTENSION CORDS.
l
Use only with power cords supplied with the Warmer.
EXPLOSION PRECAUTIONS
l
Do not use in the presence of flammable anesthetics.
0
Confirm that the oxygen supply is turned off and that the equipment is disconnected from the oxygen supply when performing cleaning and maintenance procedures; a fire and explosion hazard exists when performing cleaning and/or maintenance procedures in an oxygen-enriched environment.
OXYGEN PRECAUTIONS
l
Improper use of supplemental oxygen may be associated with serious side effects including blindness, brain damage, and death. The risks vary with each infant. The method, the concentration, and the dura- tion of oxygen administration should be prescribed by the attending physician.
.
If it is necessary to diately.
administer oxygen in an emergency, the attending physician should be notified imme-
NOTE:
See the current edition of “Guidelines for Perinatal
Care”of
the American Academy of Pediatrics/The
American College of Obstetricians and Gynecologists.
.
The oxygen concentration inspired by an infant does not predictably determine the partial pressure of
PO2
oxygen
(PO,) in the blood. When deemed advisable by the attending physician, blood
should be
measured by accepted clinical techniques.
.
Oxygen flow rates cannot be used as an accurate indication of oxygen concentrations. Oxygen concen- trations should be measured with acalibrated oxygen analyzer at intervals directed by the attending phy- sician.
.
Keep matches and all other sources of ignition out of the room in which the Warmer is located and keep spark-producing equipment or any sources of ignition out of the Bassinet. Wood, textiles, oils, and other combustibles are easily ignited and burn with great intensity in air enriched with oxygen.
.
Although oxygen compatible materials are used in the oxygen delivery system, special care must be tak- en when high pressure oxygen such as found in a medical oxygen cylinder is used. Violent ignition of oil,
grease, greasy substances, small particles of dust, dirt or other particulate contaminants (even small particles of metal), can occur in the presence of high pressure oxygen if their ignition temperature is
reached. An instantaneous increase in temperature can occur due to friction, particle acceleration, or
adiabatic compression, if the oxygen cylinder valve is opened too rapidly. SERIOUS INJURY MAY RE- SULT! Always observe the following precautions:
-
Oil, grease, greasy substances, dust, dirt and other particulate contaminants must be kept away from oxygen regulators, cylinder valves, tubing and all other oxygen equipment.
-
III
-
OPERATING PRECAUTIONS (Continued)
-
Always open oxygen cylinder shut-off valves very slowly and carefully.
-
On high pressure oxygen cylinders use only pressure regulators or reducing valves approved for oxygen service. Do not use oxygen pressure regulators or reducing valves for air or gases other
than oxygen as they may be hazardous. Operate such devices in strict accordance with the
manufacturer’s recommendations.
-
When new or replacement oxygen cylinders are to be installed, they should have their outlet ports
cleared by cracking the cylinder valve momentarily before attachment to the equipment.
LOW FLOW MICROBLENDER WARNINGS
l
If either the air or oxygen gas source pressure is reduced or increased creating a pressure differential of 30 psi, the microblender alarm will sound. This condition significantly alters the
FiO2
and flow output from the microblender.
l
Always operate the low flow microblender with clean/dry medical grade gases.
.
Air inlet water filters are recommended for use with the low flow microblender.
.
The level of oxygen and the partial pressure of oxygen within the patient’s blood should be monitored.
(Pa02)
iv
--
TABLE OF DEFINITIONS AND SYMBOLS
NOTE, IMPORTANT, PRECAUTION, CAUTION, AND WARNING
NOTE:
IMPORTANT PRECAUTION: A Precaution is supplementalinformation
CAUTION:
SYMBOLS
A Note is inserted in text to point outprocedures or conditions which may otherwise be misinterpreted
or overlooked.
Similar to a Note but used when greater emphasis is required.
equipment.
A Caution is inserted in text to call attention to a procedure which, if not followed exactly, can lead to damage or destruction of the equipment.
WARNING: A Warning is inserted in text to call attention to dangerous or hazardous conditions inherent to the operation, cleaning, and maintenance of the equipment which may result In per- sonal injury or death of the operator or patient.
A Note may also be used to clarify apparent/y contradictory or confusing situations.
user
in the safe and effective use of the
Examination Light
Examination Light Switch
Mode Control Key
Attention: consult accompanying documents.
Type B equipment with an F-type isolated (floating) applied part.
Danger! High Voltage!
to
assist the
/
\
-
o-
I
8
0
A
8
Disposable Suction Bottle
Reusable Suction Bottle
Patient
Heater Element
Suction Line Filter
Load Symbol
00
I
w@
Temperature Override Mode Key
Keypad Lock Key
Set Temperature Keys
Power On/Off Switch
Celsius/Fahrenheit Selection Key
Silence/Reset Key
Procedural Silence Indicator
Apgar Timer Keys
-V-

TABLE OF CONTENTS

SECTION 1
GENERALINFORMATION..
1.1
INTRODUCTION..
1.2 DESCRIPTION
1.3 SPECIFICATIONS
1.4 EQUIPMENTPROVIDED
1.5 FACTORY INSTALLED ACCESSORIES
1.6 FIELD INSTALLED ACCESSORIES (Refer to Section 6 for Part Numbers)
2 INSTALLATION .....................................................................
2.1
UNPACKING
2.2 ASSEMBLY (Refer to Figure 2.1)
2.3 ASSEMBLY OF CART/BASSINET
3 FUNCTIONAL DESCRIPTION
3.1 GENERAL ..............................................................................
3.2 FUNCTIONALDESCRIPTION
3.2.1 WARMERMODULE..
3.2.2 CART/BASSINET
3.2.3 CONTROLLER
3.2.4 BLENDER MODULE (OPTIONAL)
3.2.5 RESUSCITATION MODULE (OPTIONAL)
3.2.6 GAS SUPPLY MODULE (OPTIONAL)
3.2.7 ALARMS .........................................................................
APGARTIMER
3.2.8
4 OPERATION
4.1 CONTROLS, INDICATORS AND CONNECTORS
4.2 OPERATIONAL CHECKOUT PROCEDURE
4.3 MECHANICALCHECKOUT
4.4 RESUSCITATION EQUIPMENT PRE-USE CHECKOUT AND SETUP (OPTIONAL)
4.4.1 SUPPLYPRESSURE ..............................................................
4.4.2 BLENDED GAS SUPPLY PRE-USE CHECKOUT AND SETUP (OPTIONAL)
4.4.3 RESUSCITATION MODULE PRE-USE CHECKOUT AND SETUP (OPTIONAL)
4.5 CONTROLLEROPERATION
4.5.1
4.5.2 MANUALMODE
4.5.3
4.5.4 EXAMINATION LIGHT
4.6 CART/BASSINET ......................................................................
DOCKlNGANDUNDOCKlNGTHECART/BASSlNET
4.7
4.8 X-RAYPROCEDURES
CLEANINGANDMAINTENANCE
5
5.1 GENERAL .............................................................................
5.2 CLEANING
DISASSEMBLYFORCLEANING
5.3
5.4 CLEANINGPROCEDURES
5.4.1 CLEANING AGENTS..
5.4.2 PAINTEDSURFACES
54.3 CLEAR PLASTIC AND ACRYLIC SURFACES
......................................................................
PRE-WARMMODE
BABYMODE
............................................................................
......................................................................
.........................................................................
.......................................................................
...........................................................................
.....................................................................
.........................................................
................................................................
...................................................
..........................................................
.........................................................
........................................................
............................................................
..............................................................
..................................................................
....................................................................
....................................................
.............................................
...............................................
...................................................................
..........................................
-
CONTROLLER
.............................................................
............................................................
...............................................................
..................................................................
.............................................................
......................................
.................................................................
...................................................
.........................................................
.............................................................
............................................................
.............................................................
........................................
...............................
....................
............
.............
..........
PAGE
1
1 1 1
3
4
4
6
6
6 6
8
8 8 8 8 8 8
8 12 13 13
14
14
24 25 27 27 27 28
29 29 29
30 31 32
32
32
33
33
33
33
34
34
34
34
-vi-
TABLE OF CONTENTS (continued)
SECTION
5.4.4 METAL SURFACES
5.4.5 WOODEN SURFACES
5.4.6 SKIN TEMPERATURE PROBE (REUSABLE)
5.5 STERILIZATION
5.6 REASSEMBLY AFTER CLEANING
5.7 CALIBRATION .........................................................................
5.8 TROUBLESHOOTING ..................................................................
6 PARTS LIST .......................................................................
6.1 GENERAL
.............................................................................
.......................................................................
...............................................................
............................................................
.........................................
......................................................
PAGE
34 34 35 35 35 35 35
37
37
-vii-
RESUSCITAIREB
B
IRTHING ROOM
SECTION 1

GENERAL INFORMATION

W
ARMER
.l
INTRODUCTION
1
This manual provides instructions for installation, use, operator maintenance and troubleshooting of the equipment. Hill-Rom Air-Shields cannot be re- sponsible for the performance of the equipment if the user does not operate the equipment in accor- dance with the instructions, fails to follow the main- tenance recommendations in Section 5 of this
manual or effects any repairs with unauthorized components. Calibration and repair should be per- formed only by qualified service personnel. Service
manuals are available from Hill-Rom Air-Shields.
This manual should be read, thoroughly under- stood, and be readily accessible to all personnel who will be working with the equipment. The manual
should be stored with the equipment when not in
use. If there is anything you do not understand,
please contact your Hill-Rom Air-Shields’ repre- sentative for further information.
1.2 DESCRIPTION
ResuscitaireB
The signed specifically for birthing room use. The
suscitaire8
detachable Cart/Bassinet, Warmer, and a Control- ler module which provides heat control, monitoring
of skin temperature and Apgar timing. The
citaire8
optional resuscitation package with suction and oxygen delivery. Other options include a reserve gas supply and blender.
1.3 SPECIFICATIONS Specifications for the
Room
cations are subject to change without notice.
Birthing Room Warmer consists of a
Birthing Room Warmer also includes an
Warmer are provided in Table 1
Birthing Room Warmer
Resus-
ResoscitaireB Birthing
.l
. All specifi-
is de-
Re-
-l-
RESUSCITAIREGI
B
IRTHING
TABLE 1.1 SPECIFICATIONS
R
OOM
W
ARMER
POWER REQUIREMENTS ...........................................
OVERLOAD PROTECTION CHASSIS LEAKAGE CURRENT EXAMINATION LIGHT. ALARMS
High Temperature
Check Patient
Apgar Timer Power Fail
Probe.. .......................................
SystemFail
BabyTemp ..............................................
Electrical Module Audio Alarms
Blender Module Pneumatic Audio Alarm
...................................................
..........................................
................................
................................
............................................
.....................................
.......................................
..............................
totalling 15 minutes. Then the heater is turned Off.
.......................
..................................
12OV,
60 Hz, 750W
12A
Dual
>lOO
Foot Candles (0.12
Activates if Skin Temperature Probe is
attached and the skin temperature
sensor reaches 39.0
Activates in Manual Mode after 10 minutes.
Remains on with audible alarm every
30 seconds for 5 minutes;
Activates at the l-,
minute Apgar Time intervals.
Activates when there is a
Activates if Skin Temperature Probe fails
Indicates system failure; refer unit
Activates if Baby Temperature
“C
fluctuates 1
Tone Frequency: 1.2 KHz maximum
Three-stage sound level: 15 seconcls
low, 15 seconds medium, then high.
above or below set
Circuit
Less than 300
“C.
Resets at 38.5
to service immediately.
Breakers
uA
lumens/cm:z)
“(3.
5-and
loss of power.
(open or short).
Vibrating Reed.
lo-
poirit.
DISPLAYS
Skin Temperature Display
Accuracy.. .................................
Resolution
Apgar Timer Display
Range Resolution
Accuracy .............................................................
MANUAL HEAT CONTROL
DATA PORT
DIMENSIONS AND WEIGHT-FREE STANDING WARMER
Height
Width (side to side)
Depth (front to back) Weight (without tanks) Warmer Head Rotation (right and left)
.............................................................
...............................................
.................................................................
..................................
.....................................................
...........................................................
................................................
................................................
.................................................
....................................... 90 Degrees
-2-
f
0.2
“C
“C
“C
for 31
to 37
0 to 59 minutes, 0 to 59 seconcls
Adjustable in 10% increments
from zero to full power (100
Baud Rate.
(88
0.1
Of0.5second
2400 Bits/second fixed
RS232C
74 inches (188 cm)
28 inches (102 cm)
.40
inches (89 cm)
170 Ibs (77 kg)
“F to 98.6
“C
(0.5
lsecond
Compatible
“F)
“F)
%#)
RESUSCITAIREGO
TABLE 1
DIMENSIONS AND WEIGHT
Height to Mattress
Width (to outside of bumper)
Length (to outside bumper)
Weight
MattressTilt
TOTAL SYSTEM DIMENSIONS AND WEIGHT
Height
Width
Length
Weight
ACCESSORY WEIGHT LIMITS
Infusion Pump/IV Pole Monitor Shelf
ENVIRONMENTAL
Operating Temperature Range Storage Temperature Range Relative Humidity Operating Range
RESUSCITATION
Wall Supply Pressure
Cylinder Pressure
Cylinder Diameter
...........................................................
.........................................................
.............................................................
..........................................................
........................................................
................................................
.................................
...................................................
..........................................................
...................................................
.....................................................
............................................
B
IRTHING ROOM
.I
SPECIFICATIONS (Cont.)
-
Cart/Bassinet
.........................................
..........................................
..................................
..................................
..................
W
ARMER
38.5 inches (98 cm) 24 inches (61 cm) 31 inches (79 cm)
(c73
ambfent
ambient
“C
psi
Kg)
75
max
psi
c
160 Ibs
5 and 10 degrees Reverse Trendelenburg.
Infant’s head should be toward front of warmer.
..74inches(188cm)
28inches (71 cm)
48inches(122cm)
..c320Ibs(c152Kg
5 Ibs (2.2 Kg)
10 Ibs (4.5 Kg)
“C
to 30
18
“C
+45
“C
20
4-5
to
inches(
-
5% RH to 95% RH,
non-condensrng
40 to
2900
1 O-l 2 cm)
)
Patient Gas Supply Airway Pressure Limiting Operator-Adjustable Fixed Pressure Relief, Factory Set Suction Circuit Adjustable Suction Intensity Auxiliary Supply Auxiliary Flow Range Auxiliary Pressure Limit
1.4 EQUIPMENT PROVIDED
.
Bassinet
visibility and access to the infant. The Bassi-
net tilts up in the rear to 5 and 10 degrees and provides for X-ray Cassette tray (optional) in- sertion.
.
Warmer Module
houses a heating element and an Examina- tion Light for special procedures.
-
The Bassinet provides maximum
.....................................................
-
The Warmer Module
................................
....................................
..................................... ..
.
Controller
warm, manual heat control, automatic skin temperature servo-control and contains an Apgar Timer, Skin Temperature monitor and
probe.
.
Resuscitation Module (optional) -The Resus-
citation Module contains a suction circuit, a patient oxygen delivery circuit with airway pressure relief and an auxiliary oxygen deliv- ery circuit.
0 to 50 cm
.
60 cm
0 to 150
160 cm
-
The Controller provides pre-
mmHg
kPa) kPa)
(0 to 20
0 to 15 LPM
kPa)
+
Hz0 (4.9
2
Hz0 (5.9
-c
Hz0 (16
10% 10%
kPa)
10%
-3-
RESUSCITAIR~
1.5 FACTORY INSTALLED OPTIONS
.
Resuscitation Module
.
Integrated Precision Blender
.
Gas Supply Module
-
02 Pipeline and Cylinder
-
Oz/Air Pipeline and Cylinder
B
IRTHING
R
OOM
1.6
FIELD INSTALLED ACCESSORIES (Refer to Section 6 for Part Numbers)
.
X-ray Cassette Tray
.
Air Hose Assembly
.
Oxygen Hose Assembly
.
Monitor Shelf
.
I.V.
Pole
.
Drawer Organizer
W
ARMER
-4-
RESUSCITAIREB
B
IRTHING ROOM
PASS-THROUGH
DRAWER
W
ARMER
CASTER LOCK
DRAWER
LOCK
MONITOR SHELF
3c.
MtwiANISM
X-RAY TRAY
iT
TILT
\
WARMER MODULE
CONTROLLER
/
4
02/AIR
WALL
GAS HOSES
SUCTION BOTTLE
-
CAiTER
GAS TANKS
(NOT SUPPLIED)
LOCK
BLENDED GAS
SUPPLY MODULE
RESUSCITATION
GAS SUPPLY MODULE
FIGURE 1
PASS-THROUGH DRAWER
.I
EQUIPMENT PROVIDED WITH FACTORY INSTALLED OPTIONS AND
FIELD INSTALLED ACCESSORIES
-5-
RESUSCITAIREGQ
B
IRTHING
SECTION 2

INSTALLATION

2.1 UNPACKING
The
ResuscitaireB Birthing Room Warmer is shipped in two cartons. One carton contains the fol- lowing assemblies:
.
Upper Post Assembly
.
Lower Post Assembly
.
Warmer Module Assembly
.
Post Handle
.
Any user installed Accessories that were or- dered
The other carton contains the Cart/Bassinet.
2.2
ASSEMBLY OF THE WARMER (RE-
FER TO FIGURE 2.1)
When removing the equipment from the carton, use care not to scratch or otherwise damage unpro- tected surfaces; remove all packing material.
NOTE: If a Monitor Shelf is to be installed on the
unit, refer to the instructions provided with the Monitor Shelf and install the weights
provided with these accessories in the bot-
tom of the Lower Post.
1.
REMOVE THE CONTROLLER (3) from the
Upper Post (2).
CAUTION: Two persons of sufficient strength
are required when performing Steps 3 through 6.
2.
REST THE UPPER POST (2) on top of the
Lower Post opening.
CONNECT THE SUCTION HOSE (4) to the
3. Suction Hose (11). Release the joined Suction Hoses from the Neat Clip inside the Lower Col- umn.
4.
MOUNT THE UPPER POST (2) on the Lower
3/8
-
Post using four 10 provided.
5.
INSTALL TWO 10
(6) IN THE UPPER HOLES OF THE UPPER POST. Do not tighten the screws.
RAISE THE WARMER (7) above the open end
6. of the Upper Post (2) and insert the Power Cable (10) into the open end of the column.
32 x
-
32 X
inch screws (16)
3/8
INCH SCREWS
R
OOM
7.
SLOWLY LOWER THE WARMER (7)
Upper Post. Tighten the upper screws and install the remaining four 10
screws (6). THREAD THE WARMER POWER CABLE
8.
out through the Controller opening. Connect the Power Cable (10) to connector J4 on the Controller (3).
REMOUNTTHE CONTROLLER on the Upper
9.
Post. MOUNT THE BACK COVER (1) on the Upper
10
Column using the six 8 screws (5).
MOUNT THE HANDLE (12) on the Post using
11. the two 10 Screws, No. 10 Lock Washers and Spacers
(14 1
(13 CONNECT THE LINE CORD to the Power
12. Connector on the rear of the Controller (refer to Figure 4.2).
SECURE THE LINE CORD to the Back Cover
13. (1) using the Cable Clamp (8) and 8 screw (9).
CAUTION: Securing the Line Cord to the Back
14.
INSTALL ANY ACCESSORIES that were or- dered using the installation instructions pro-
vided with the accessory.
2.3
ASSEMBLY OF THE CART/BASSINET
1.
Refer to Paragraph 5.6, Step 4, and install the Rear and Side Panels. Raise the Front Panel
from its storage slot and lock it in place.
2.
ENGAGE THE CART/BASSINET TO THE WARMER by pushing the Cart/Bassinet straight onto the Warmer until it latches. Pull on the Cart/Bassinet to ensure that it is latched properly.
To disengage the Cart/Bassinet from the Warmer, push and hold the release switch (Figure 4.14) and pull the Cart/Bassinet away from the Warmer.
W
ARMER
onto the
3/8
-
32 x
32 x
3/4
-
-
32 x 2.6 inch Phillips Head
).
(15 1
Panel is required to prevent removal of the Controller chassis with the ac power applied.
inch
painted
-
32 x
3/8
-6
RESUSCITAIREB
PARTS LIST
Screw, Screw, 10 Screw, 8 Screw, 8 -32 x No.iOLockWasher
Spacer (Qty 2) . . . . . . . . . . . . . . . .
Cable Clamp Post Handle
lo-32~3/8(QtylO)
-
32 x 2.6 (Qty 2) . . . .
3/4
-
32 x
(Qty 6) . . . .
3/8
(Qty 1) . . . . .
. . . . . . . . . . . . 9912416
...99041
.
. : 1: : : : : : : : 1: : : : : .
B
.99
045 30
.99
032 85
.99
031 38
.99 124 61
..1772564
.81
005 73
IRTHING
36
R
OOM
W
ARMER
FIGURE 2.1 INSTALLATION
-7-
Loading...
+ 30 hidden pages