Century CC® Bed
TABLE OF CONTENTS |
|
|
FEATURES................................................................................................... |
|
1 |
DIMENSIONS............................................................................................. |
|
1 |
BED CONTROLS |
|
|
Hi-Lo....................................................................................................... |
|
2 |
Head With Automatic Contour ......................................................... |
|
2 |
Head Only ............................................................................................ |
|
2 |
Knee...................................................................................................... |
|
3 |
Trendelenburg/Reverse Trendelenburg ........................................... |
|
3 |
Vascular ................................................................................................ |
|
3 |
Cardiac Chair...................................................................................... |
|
3 |
Electric Control Lockouts ................................................................... |
|
3 |
FEATURES |
|
|
Siderails ................................................................................................. |
|
4 |
Central Brake and Steer..................................................................... |
|
4 |
Permanent IV Pole .............................................................................. |
|
4 |
Instant CPR .......................................................................................... |
|
5 |
Headboard Removal ......................................................................... |
|
5 |
Convertible Footboard ...................................................................... |
|
5 |
C-arm Interfacing................................................................................ |
|
6 |
Other Features ..................................................................................... |
|
6 |
OPTIONS |
|
|
Integrated Air Support System........................................................... |
|
7 |
SideCom® Communication System.................................................. |
|
7 |
Low Leakage ....................................................................................... |
|
7 |
Scale ................................................................................................. |
|
8–9 |
ACCESSORIES |
|
|
O2 Tank Holder...................................................................................... |
|
9 |
X-Ray Cassette Holder................................................................ |
|
10–11 |
ISS System ........................................................................................... |
|
12 |
Transducer Holder ............................................................................. |
|
12 |
IV Pole ................................................................................................. |
|
12 |
Trapeze Support Bracket.................................................................. |
|
12 |
Cane Bumpers................................................................................... |
|
12 |
Siderail Pads ....................................................................................... |
|
12 |
Chartholder ....................................................................................... |
|
12 |
CLEANING/REMOVAL OF DYNAMICAIRE™ SLEEP SURFACE |
.............13 |
|
SAFETY TIPS............................................................................. |
Back Cover |
|
NOTE An In-Service videotape is available. Contact your |
|
|
Hill-Rom representative or call 1-800-445-3720. |
|
|
FOR SPECIAL ASSISTANCE: |
|
Hill-Rom Account Manager ________________________________________ Phone |
______________________________ |
Clinical Consultant ________________________________________________ Phone |
______________________________ |
Field Service Technician __________________________________________ Phone |
______________________________ |
Customer Service Representative __________________________________ Phone |
______________________________ |
Siderail |
|
SideCom® |
Siderail Patient |
|
Communtication System |
Controls |
|
||
Nurse Controls |
|
|||
|
(Optional) |
(Both Sides) |
Convertible |
|
(Both Sides) |
|
|||
|
|
|
Footboard |
|
|
|
|
|
|
Restraint Strap Holders; |
|
|
|
|
Head, Thigh, Foot End |
|
|
DynamicAire™ |
|
(Both Sides) |
|
|
Sleep Surface |
|
|
|
|
(Optional) |
|
Removable |
|
|
|
|
Headboard |
|
|
|
Patient Scale |
|
|
|
|
|
|
|
|
|
(Optional) |
IV Pole Sockets |
|
|
|
|
(One At |
|
|
|
|
Each Corner) |
|
|
|
Trendelenburg/ |
|
|
|
|
|
|
|
|
|
Reverse Trendelenburg |
|
|
|
|
Controls |
Permanent |
|
|
|
|
IV Pole |
|
|
|
|
|
|
|
|
Electric Controls |
|
|
|
|
Lockout Panel |
Drainage Bag Hooks |
|
|
|
Integrated Air Support |
(Both Sides |
|
|
|
|
|
|
|
System Controls |
|
And Foot End) |
|
|
|
|
|
|
|
(Optional) |
|
|
|
|
|
|
Head Elevation |
|
|
|
|
Indicator |
|
|
|
Motor Power |
(Both Sides) |
|
|
|
Switch |
IV Pole |
|
|
|
|
Storage Area |
|
Pump/Release Pedals |
Central Brake And Steer |
|
|
|
For Manual Control |
||
|
|
(Both Ends) |
||
|
|
(Both Sides) |
|
|
Manual Control |
|
|
||
|
|
|
||
For Bed Positioning |
|
8" Polyurethane Carpet Casters |
||
Including Instant CPR |
|
(6" Polyurethane |
Carpet Casters Optional)
TuckAway
Siderails
Measurements: Inches (centimeters)
91.0 (231) |
|
|
|
|
|
|
|
|
|
93.5 With Roller Bumpers (243) |
|
|
|
|
38.5 (98) |
|
|
|
|
|
|
|
|
|
|||||
|
|
|
|
|
|
|
|
||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
34.0 (86) |
|
|
|
|
|
|
|
|
|
||||
|
|
|
|
|
15.0
(38) 8" Casters
55.5 (140) High 37.5 (97) Low
6" Casters
54.3 (137) High 36.3 (94) Low
8" Casters
40.5 (102) High/22.5 (57) Low
6" Casters
39.3 (99) High/21.3 (54) Low
8" Casters 6.5 (17)
6" Casters 4.6 (12)
57.0 (145)
23.5 (60)
1
Introduction: The Century CC Bed is designed to use electric and/or manual controls.
ELECTRIC CONTROLS |
|
|
MANUAL CONTROLS |
|
|
|
|
|
|
|
|
|
|
|
Hi-Lo Push Button |
|
|
Manual Foot Pedals |
|
|
|
|
|
|
|
|
|
|
|
Knee And Head |
Mode Selector Dial |
Push Button |
|
NURSE CONTROL PANEL
Knee Function Lockout |
|
|
|
Trendelenburg/ |
||
|
|
|
|
|
||
Hi-Lo Function Lockout |
|
|
|
Reverse Trendelenburg |
||
|
|
|
Angle Indicator |
|||
|
|
|
|
|
||
Head Function Lockout |
|
|
|
|
||
Auto Contour Lockout |
|
|
|
|
||
Hi-Lo Bed Control |
Patient Scale Monitor |
|||||
|
|
|||||
|
|
|
|
(See Page 8–9) |
||
|
|
|
||||
HI-LO |
For Electric: |
For Manual: |
||||
|
• Locate the Hi-Lo control on the outside of either |
• Turn mode selector dial to “Hi-Lo” section… |
||||
|
head end siderail or at the foot end nurse control |
• To raise: pump the green pedal located on either |
||||
|
panel… |
|
side of the bed… |
|||
|
• To raise: push the G button then release… |
• To lower: push the orange pedal on either side of |
||||
|
• To lower: push the H button then release. |
|
the bed. |
SAFETY TIP If the bed is occupied and unattended, the bed should stay in the low position and the mode selector dial should not be in the Hi-Lo mode.
HEAD WITH |
For Electric: |
AUTOMATIC |
• When head of bed is raised, the knee will |
CONTOUR |
automatically elevate. (Automatic contour may |
|
be eliminated by pushing up the contour lockout |
|
at the nurse control panel. See Electric Control |
|
Lockouts)… |
|
• Locate the head control button on the inside of |
|
either head end siderails… |
|
• To raise: push the head G button then release… |
|
• To lower: push the head H button then release. |
For Manual:
• No manual automatic contour control.
SAFETY TIP The automatic contour feature prevents the patient from gravitating to the foot end of the bed, thus reducing patient repositioning
HEAD ONLY |
For Electric: |
For Manual: |
|
• Insure the Automatic Contour lockout control |
• Turn the mode selector dial to the“Head”section… |
|
located at the nurse control panel is pushed up… |
• To raise: pump the green pedal… |
|
• Locate the head control button on the inside of |
• To lower: push the orange pedal. |
|
either head end siderails… |
|
|
• To raise: push the head G button then release… |
|
|
• To lower: push the head H button then release. |
|
NOTE A head elevation gauge located on both sides of the bed indicates the degree of head elevation.
2
BED CONTROLS
KNEE |
For Electric: |
For Manual: |
|
• Locate the knee control button on the inside of |
• Turn the mode selector dial to the“Knee”position… |
|
either head end siderail… |
• To raise: pump the green pedal… |
|
• To raise: push the G button then release… |
• To lower: push the orange pedal. |
|
• To lower: push the H button then release. |
|
|
|
|
REVERSE
TRENDELENBURG
Trendelenburg/Reverse
Trendelenburg Release Handle
SAFETY TIP Before lifting or lowering, make sure both Trendelenburg/Reverse Trendelenburg release handles are fully gripped and touching the bumpers.
For Manual:
•If necessary, raise the bed using the HI-LO Control located at the nurse control panel until bed reaches desired height…
•Grip both of the Trendelenburg/Reverse Trendelenburg handles located at the foot end of the bed…
•Lift for Trendelenburg or lower for Reverse Trendelenburg…
•To adjust for desired degrees, refer to the “Trendelenburg Angle” indicator located to the right on the nurse control panel…
•Release handles to lock position. Confirm locking by attempting to move the sleep surface up or down.
To Level The Sleep Surface:
•Grip both handles to release the sleep surface…
•Raise or lower the foot end of the bed until the “Trendelenburg Angle” indicator reads zero…
•Release handles to lock position. Confirm locking by attempting to move sleep surface up or down.
VASCULAR |
For Manual: |
To Level The Sleep Surface: |
|
||
|
• Grasp the foot end section of the sleep surface |
• Lift the foot section to release the foot rack… |
|
||
|
and lift up (two positions are available)… |
• Grasp the rack and pivot downward… |
|
||
|
• Insure the foot rack is properly engaged. |
• Lower the foot section. |
|
||
|
|
|
|
|
|
CARDIAC |
For Manual or Electric: |
To Level The Sleep Surface: |
|
||
CHAIR |
• Place the bed in 12˚ Reverse Trendelenburg… |
• Manually or electrically lower the knee and head |
|
||
|
• Manually or electrically raise the head of the |
of the bed unit until flat… |
|
||
|
bed to 60˚(See Head)… |
• Place the bed in 0˚ Reverse Trendelenburg. |
|
||
|
• Manually or electrically raise the knee of the bed |
|
|
|
|
|
to 35˚(See Knee). |
|
|
|
|
|
|
|
|
|
|
ELECTRIC |
Siderail Lockout Controls: |
Motor Power Switch: |
|
||
CONTROL |
Located in the nurse control panel, the respective |
Located at the foot end base of the bed, this |
|
||
LOCKOUTS |
lockout function will cancel the HI-LO, KNEE, HEAD |
control will turn off ALL electric controls. |
|
||
|
and CONTOUR electric controls. |
• Press on – electric controls are operable… |
|
||
|
|
|
|||
|
• Push the button up – electric control is |
• Press off – electric controls are inoperable. |
|
||
|
inoperable… |
|
|
|
|
|
• Push the button down – electric control is |
|
|
|
|
|
operable. |
|
|
|
|
|
Patient Control Lockout Indicator: |
|
|
|
|
|
Located below the nurse control panel, the |
|
|
|
|
|
indicator illuminates in red when lockout controls |
|
|
|
|
|
are being used. |
|
|
|
|
|
|
|
|
|
|
Motor Power Switch
SAFETY TIP Whenever a patient should be restricted from operating the patient controls, activate the appropriate lockout control. This is especially true for mentally compromised patients and for patients in traction.
3