Authorized European Union Representative:
HILL-ROM SAS
B.P. 14 - Z.I. DU TALHOUET
56330 PLUVIGNER
FRANCE
TEL: +33 (0)2 97 50 92 12
No part of this text shall be reproduced or transmitted in any form or by any means, electronic or
mechanical, including photocopying, recording, or by any information or retrieval system without
written permission from Hill-Rom Services, Inc. (Hill-Rom).
The information in this manual is confidential and may not be disclosed to third parties without
the prior written consent of Hill-Rom.
This manual (USR025) was originally released and supplied in English. For a list of available
translations, contact Hill-Rom Technical Support.
Product may be covered by one or more patents. Please consult listing at
www.hill-rom.com/patents for any patent(s).
Sixth Edition, March 2013
First Printing 2000
Printed in the USA
Affinity® is a registered trademark of Hill-Rom Services, Inc.
Beaucoup® is a registered trademark of Ecolab, Inc.
CSA® is a registered trademark of Canadian Standards Association, Inc.
Comfortline® is a registered trademark of Hill-Rom Services, Inc.
Dartex® is a registered trademark of Dartex Coatings Ltd.
EasyGlide® is a registered trademark of Hill-Rom Services, Inc.
Freedom Hill™ is a trademark of Hill-Rom Services, Inc.
Galahad® is a registered trademark of Puritan/Churchill Chemical Co.
Hi-Tor® is a registered trademark of Ecolab, Inc.
Hill-Rom® is a registered trademark of Hill-Rom Services, Inc.
Liberty Hill™ is a trademark of Hill-Rom Services, Inc.
Matar® is a registered trademark of Ecolab, Inc.
Nano Ag+® is a registered trademark of Hill-Rom Services, Inc.
OneStep® is a registered trademark of Hill-Rom Services, Inc.
Quanto® is a registered trademark of Ecolab, Inc.
SideCom® is a registered trademark of Hill-Rom Services, Inc.
Slo-Blo® is a registered trademark of Littelfuse, Inc.
Affinity® Three Birthing Bed and Affinity® Four Birthing Bed User Manual (USR025 REV 6)i
SmartSilver® is a registered trademark of NanoHorizons, Inc.
Stow and Go® is a registered trademark of Hill-Rom Services, Inc.
Surfacide™ is a trademark of Walton-March, Inc.
Tempur-Pedic® is a registered trademark of Dan-Foam A/S Corp.
Vesphene® II is a registered trademark of Steris, Inc.
Virex® is a registered trademark of Diversey, Inc.
The information contained in this manual is subject to change without notice.
Hill-Rom makes no commitment to update or keep current, the information contained in this
manual.
Hill-Rom reserves the right to make changes without notice in design, specifications, and models.
The only warranty Hill-Rom makes is the express written warranty extended on the sale or rental
of its products.
To order additional copies of this manual (USR025), refer to the back cover for contact information. For countries not listed on the back cover, contact your distributor.
NOTE:
The back cover is a comprehensive list of Technical Support contact information for Hill-Rom. The
product discussed in this manual may not be available in all of the countries listed.
NOTE:
Replace this manual (USR025) if it is damaged and/or can not be read.
Reference Documents
Affinity® Three Birthing Bed and Affinity® Four Birthing Bed Service Manual (MAN272)
Affinity® Four Birthing Bed—Unpacking Instructions (IS484)
iiAffinity® Three Birthing Bed and Affinity® Four Birthing Bed User Manual (USR025 REV 6)
Affinity® Three Birthing Bed and Affinity® Four Birthing Bed User Manual (USR025 REV 6)v
viAffinity® Three Birthing Bed and Affinity® Four Birthing Bed User Manual (USR025 REV 6)
Intended Use
The Affinity® Three Birthing Bed and Affinity® Four Birthing Bed are intended to be used as birthing beds
for women of child bearing age in an LDR (Labor, Delivery, Recovery) or LDRP (Labor, Delivery, Recovery,
Postpartum) setting within the acute care Labor and Delivery market. They are not intended for use as a
general hospital beds.
The intended users of this product are healthcare employees who have been trained to use the product,
and who have the physical strength and cognitive skills to operate and control the product. There are some
controls and features on the bed intended to be used by the patients and family members upon
appropriate orientation by the caregiver. Follow facility safety protocols if an intended user does not have
the physical strength or cognitive skills to operate and control the product safely.
The maximum patient weight is 470 lb (213 kg).
The bed safe working load is 500 lb (227 kg).
Safe working load: includes mattresses, headboard, patient weight, any accessories or equipment on the
bed, caregiver weight, and any visitors sharing the bed.
There are no known Contraindications for this product.
Introduction
This manual provides the information required for normal operation of the Affinity® Three Birthing Bed and
Affinity® Four Birthing Bed from Hill-Rom. Before operating the Affinity® Three Birthing Bed and Affinity®
Four Birthing Bed, be sure that you have read and understood in detail the contents of this manual. It is
important that you read and strictly adhere to the aspects of safety contained in this manual. Any reference
to a side of the bed is from the patients’ view lying in the bed on their back.
In this manual, there are references to different bed models. To identify which model of bed you have, look
at the serial number label. The label is located on the main frame at the head end. For example,
P3700BXXXX identifies a B model bed.
Intended Use
Affinity® Three Birthing Bed and Affinity® Four Birthing Bed User Manual (USR025 REV 6)1
Symbols
Symbols
Document Symbols
This symbol is used throughout the manual to identify a WARNING or CAUTION:
–WARNING identifies situations or actions that may have an effect on patient or user safety. To
ignore a warning could cause patient or user injury.
–CAUTION identifies special procedures or precautions that persons must obey to help prevent
equipment damage.
Product Symbols
These symbols are on the product and may or may not be on your model of the Affinity® Three Birthing Bed
and Affinity® Four Birthing Bed:
SymbolDescription
Type B applied part according to IEC 60601-1 (UL 60601-1)
According to IEC 60529, rating for protection against fluid ingress
and identified as equipment that is protected against vertically falling
water when tilted up to 15°. A through D model beds.
According to IEC 60529, rating for protection against fluid ingress
and identified as equipment that is protected against vertically falling
water when tilted up to 15°. E model and newer beds.
WARNING: Consult accompanying documents. Yellow and black.
E model and newer beds.
CAUTION: Consult accompanying documents. White and black.
E model and newer beds.
CAUTION: Consult accompanying documents. A through D model
beds.
2Affinity® Three Birthing Bed and Affinity® Four Birthing Bed User Manual (USR025 REV 6)
SymbolDescription
Conforms to the European Medical Device Directive 93/42/EEC.
The CE mark was first applied in 2000.
Warning: Do Not Use with Oxygen Tents—Must consult accompanying documentation—indicates the use of oxygen administering equipment of the nasal, mask, or ventilator type only or oxygen tents that
can be contained inside the siderails.
Can be green or blue.
Certified by Underwriters Laboratory in accordance with UL 2601-1
(UL 60601-1), CAN/CSA C22.2 NO.601.1, EN 60601-1, IEC 601-238 and IEC 601-1-2.
Room Light On/Off control. A through D model beds.
Symbols
Room Light On/Off control. E model and newer beds.
Back section mattress (lumbar) control. A through D model beds.
Back section mattress (lumbar) control. E model and newer beds.
Bed Up/Bed Down control
Affinity® Three Birthing Bed and Affinity® Four Birthing Bed User Manual (USR025 REV 6)3
Symbols
SymbolDescription
Foot and head sections Up/Down controls
Seat section mattress control (optional). A through D model beds.
Seat section mattress control (optional). E model and newer beds.
TV On/Off/Audio control (optional). A through D model beds.
TV On/Off/Audio control (optional). E model and newer beds.
CPR function—Identifies the release lever that can be used to manually drop the inclined head section, in order that cardiopulmonary
resuscitation can be performed without delay. A through D model
beds.
CPR function—Identifies the release lever that can be used to manually drop the inclined head section, in order that cardiopulmonary
resuscitation can be performed without delay. E model and newer
beds.
Lockout control label—The switch in the up position indicates the
lockout control is on. The switch in the down position indicates the
lockout control is off (refer to the section titled “Lockout Controls” on
page 20 for lockout status information). A through D model beds.
4Affinity® Three Birthing Bed and Affinity® Four Birthing Bed User Manual (USR025 REV 6)
SymbolDescription
Lockout control label—The switch in the up position indicates the
lockout control is on. The switch in the down position indicates the
lockout control is off (refer to the section titled “Lockout Controls” on
page 20 for lockout status information). E model and newer beds.
Battery charge status (refer to the section titled “Battery Backup” on
page 11 for battery status information).
Nurse call (refer to the section titled “Nurse Call (Optional)” on
page 14 for nurse call switch information).
Lockout control status—when the lockout control status light is on,
the lockout function is activated.
Symbols
Identifying mains fuse.
Caution: Identifying power inlet fuse or a battery power inlet fuse.
Consult accompanying documentation. E model and newer beds.
Alternating current
Equipotentiality. Beds with grounding post only. A through E model
beds.
Affinity® Three Birthing Bed and Affinity® Four Birthing Bed User Manual (USR025 REV 6)5
Symbols
SymbolDescription
Warning: Protective earth ground—Must consult accompanying documentation. Beds with grounding post only. E model and newer
beds.
Safe working load. A through D model beds.
= kg
Warning: Bed safe working load. E model and newer beds.
Warning: Maximum patient weight. E model and newer beds.
Warning: Safe working load—foot section only. Must consult accompanying documents. E model and newer beds.
Electric shock hazard. A through D model beds.
Warning: Electrical shock hazard. Must consult accompanying documents. E model and newer beds.
Warning: Use hospital grade receptacle only. Must consult accompanying documents. Beds without grounding post only. E model and
newer beds.
6Affinity® Three Birthing Bed and Affinity® Four Birthing Bed User Manual (USR025 REV 6)
SymbolDescription
Trend-Like function. A through D model beds.
Trend-Like function. E model and newer beds.
Do not throw away—indicates the need to recycle the item in accordance with local regulations.
Pb
Keep feet clear of this area. do not step on the Stow and Go® Foot
Section. Potential for injury. A through D model beds.
Symbols
Do not step—do not step on the Stow and Go® Foot Section.
Blue and black is for A through D model beds.
Red and black is for E model and newer beds.
Warning—Do not step (located on the base frame).
Blue and black is for A through D model beds.
Red and black is for E model and newer beds.
Lock and unlock position of the calf support. A through D model
beds.
Lock and unlock position of the calf support. E model and newer
beds.
Calf support release lever. A through D model beds.
Affinity® Three Birthing Bed and Affinity® Four Birthing Bed User Manual (USR025 REV 6)7
Symbols
SymbolDescription
Calf support release lever. E model and newer beds.
Transport label. A through D model beds.
Warning: Transport label. Must consult accompanying documents.
E model and newer beds.
Shows that you should not hang power cords on the Line Managers.
Do Not Push—do not use the headboard to push the bed. On wood
headboards without integrated push handles.
Warning: Engage foot section latch—Stow and Go® Beds. Must consult accompanying documents. E model and newer beds.
Warning: Lift-off foot section—engagement advisory. Must consult
accompanying documents. E model and newer beds.
8Affinity® Three Birthing Bed and Affinity® Four Birthing Bed User Manual (USR025 REV 6)
SymbolDescription
Caution: Connect to approved communication system only. Consult
accompanying documents. E model and newer beds.
Warning: Must consult accompanying documents. On the pendant
and calf support. E model and newer beds.
Symbols
Affinity® Three Birthing Bed and Affinity® Four Birthing Bed User Manual (USR025 REV 6)9
Features
Features
Caregiver-Side Control
10Affinity® Three Birthing Bed and Affinity® Four Birthing Bed User Manual (USR025 REV 6)
Patient-Side Control
ItemDescriptionItemDescription
AV-cut or straight-cut mattressI8° trend-like position
BLift-off, Stow and Go™ Foot Section or
Slide off foot section
CEasyGlide® Calf SupportsKRemovable headboard
DOneStep® Foot SupportsLLockout indicator
EFluid basinMBattery indicator
FCentral brake and steer pedalNSiderail lighting control (optional)
GStorable labor gripsOAir support system (optional)
HCPR featurePSiderail communications package
a. Slide off foot section no longer available.
Automatic Tilt
Features
JOne-handle siderail release
a
(optional)
•As the head section is raised, the seat section
gradually tilts up from 0° to 15°.
•As the head section is lowered, the seat section
gradually returns to a flat position.
Battery Backup
The Affinity® Three Birthing Bed and Affinity® Four Birthing Bed are designed with battery
backup as a standard feature. The battery allows the hilow, foot, and head motors to be
activated from the siderail controls without power being supplied to the bed. In addition, the
battery backup powers the nurse call function, but it does not power any other bed functions,
such as the optional air support system.
Location: The battery backup indicator is located on the siderail to indicate the battery
condition.
Battery Backup LED Indicators (while connected to AC power):
• ON = Battery is charging.
Battery Backup LED Indicators (while on battery power):
• ON = Battery status is operational.
• FLASHING = Battery needs charging.
• OFF = Battery is discharged below the level required to run the motors.
Battery Backup
Indicators
NOTE:
If the battery has been completely discharged, it may take up to 36 hours to recharge to operational status.
Affinity® Three Birthing Bed and Affinity® Four Birthing Bed User Manual (USR025 REV 6)11
Features
s
CAUTION:
Remove the battery if the bed will not be in service for extended periods of time. Failure to do so could
result in damage to the life of the battery, or damage to the bed. Contact the appropriate maintenance
personnel, and refer to the Affinity® Three Birthing Bed and Affinity® Four Birthing Bed Service Manual
(MAN272).
To ensure the battery is always charged, plug the bed into the appropriate power source whenever possible.
Bed Controls
The bed is electrically operated using siderail or pendant controls. The hilow, head, and foot functions can be
operated by battery backup, if necessary.
Affinity® Three Birthing Bed
Nurse Control Panel
Patient Control Panel
Affinity® Four Birthing Bed (B through D model beds)
Nurse Control Panel
Patient Control Panel
Affinity® Four Birthing Bed (E model beds)
Nurse Control Panel
Back Section Lumbar Mattress Control (Optional)
The Back Section Mattress (Lumbar) control enables you to inflate or deflate the back section
lumbar mattress in order to attain a degree of firmness or softness.
Patient Control Panel
Location: Caregiver and patient siderails (right hand and left hand)
To Activate:
• To inflate the back section, press and hold the Back Firm button (plus sign) until the
desired firmness is attained.
• To deflate the back section, press and hold the Back Deflate button (minus sign) until
the desired softness is attained.
12Affinity® Three Birthing Bed and Affinity® Four Birthing Bed User Manual (USR025 REV 6)
Back Section
Mattress Control