Dear Customer,
Thank you for choosing a HANNA instruments® product.
Please read this instruction manual carefully before using the instru-
ment.
This manual will provide you with all the necessary information for the
correct use of the instrument, as well as a precise idea of its versatility
in a wide range of applications.
If you need additional technical support, do not hesitate to e-mail us
at tech@hannainst.com
These instruments are in compliance with the CE directives.
PRELIMINARY EXAMINATION
Remove the instrument from the packing material and examine it to
make sure that no damage has occurred during shipping. If there is
any damage, immediately notify your dealer.
Each meter is supplied compleet with:
• RH probe: HI 70608/2 for HI 9064
HI 70605/2 for HI 9065
• Batteries (4 x 1.5V AA)
• Instruction manual
• Rugged carrying case
Note: Save all packing material until you are sure that the instru-
ment functions correctly. All defective items must be returned to
us in the original packing with the supplied accessories.
TABLE OF CONTENTS
PRELIMINARY EXAMINATION ...................................................... 3
GENERAL DESCRIPTION .............................................................. 3
FUNCTIONAL DESCRIPTION ........................................................ 4
SPECIFICATIONS......................................................................... 5
OPERATIONAL GUIDE ................................................................. 6
CALIBRATION ............................................................................. 8
BATTERY REPLACEMENT ............................................................ 11
ACCESSORIES ........................................................................... 12
WARRANTY .............................................................................. 13
CE DECLARATION OF CONFORMITY............................................ 14
GENERAL DESCRIPTION
HI 9064 and HI 9065 are heavy-duty hygrometers specially designed for
field applications such as HVAC.
They are housed in a rugged water-resistant case to provide maximum
protection in the field.
The relative humidity probe houses a thin-film polymer sensor, which
quickly performs all RH readings to 98% accuracy.
A “HOLD” key will instantly freeze the reading on the LCD.
HI 9065 can also measure temperature with an internal sensor in the RH
probe from 0 to 60°C as well as from 32 to 140°F.
You can also connect a separate temperature probe for gas and liquid
measurements (HI 765 series).
Note: The RH probe sensor must never come into contact with water
or other liquids.
23
SPECIFICATIONSFUNCTIONAL DESCRIPTION
1
S
H °C
S
H °C
HI 9065
Thermo
Hygrometer
3
ON/OFF
6
5-95
%
RH %
°C
0-60 °C
°F
32-140 °F
HOLD
°C
°F HOLD
RH
7
89
S
H
HI 9064
Hygrometer
RH% 5 . . . . . . 95
2
5
423
ON/OFF
1) Liquid Crystal Display (LCD)
2) HOLD key, to freeze reading on display
3) ON/OFF key, to turn the meter on and off
4) °F key, to display temperature in Fahrenheit (HI 9065 only)
5) °C key, to display temperature in Centigrade (HI 9065 only)
6) RH key, to display RH measurements (HI 9065 only)
7) Battery compartment
8) RH probe socket
9) Temperature probe socket with protective cap (HI 9065 only) for
HI 765 probe series (see “Accessories” section)
10) Shielded cable of RH probe
11) Low RH calibration trimmer
12) High RH calibration trimmer
13) Polypropylene probe body
14) Perforated protective cap, to prevent damages to the sensor
HI 9064 HI 9065
Range 5.0 to 95.0% RH
-- 0.0 to 60.0°C / 32 to 140°F
Resolution 0.1% RH
-- 0.1°C / 1°F
Accuracy ±2% RH
-- ±0.4°C / ±1 °F
(for 1 year, excluding probe error)
Typical EMC ±3% RH
Deviation -- ±2°C / ±4°F
RH Calibration Manual, 2 point
through trimmers on RH probe
Probe HI 70608/2 HI 70605/2
Environment 0 to 50°C (32 to 122°F);
RH max 100%
Battery Type / Life
4 x 1.5V AA / approx. 500 hours of continuous use
Dimensions 196 x 80 x 60 mm
(7.7 x 3.1 x 2.4")
Weight 500 g (1.1 lb.)
45