PRELIMINARY EXAMINATION
Remove the instrument from the packing material and examine it to
make sure that no damage has occurred during shipping. If there is
any damage, immediately notify your dealer.
Each meter is supplied compleet with:
• RH probe:HI 70608/2 for HI 9064HI 70605/2 for HI 9065
• Batteries (4 x 1.5V AA)
• Instruction manual
• Rugged carrying case
Note: Save all packing material until you are sure that the instru-
ment functions correctly. All defective items must be returned to
us in the original packing with the supplied accessories.
CE DECLARATION OF CONFORMITY............................................ 14
GENERAL DESCRIPTION
HI 9064 and HI 9065 are heavy-duty hygrometers specially designed for
field applications such as HVAC.
They are housed in a rugged water-resistant case to provide maximum
protection in the field.
The relative humidity probe houses a thin-film polymer sensor, which
quickly performs all RH readings to 98% accuracy.
A “HOLD” key will instantly freeze the reading on the LCD.
HI 9065 can also measure temperature with an internal sensor in the RH
probe from 0 to 60°C as well as from 32 to 140°F.
You can also connect a separate temperature probe for gas and liquid
measurements (HI 765 series).
Note: The RH probe sensor must never come into contact with water
or other liquids.
23
SPECIFICATIONSFUNCTIONAL DESCRIPTION
1
S
H °C
S
H °C
HI 9065
Thermo
Hygrometer
3
ON/OFF
6
5-95
%
RH %
°C
0-60 °C
°F
32-140 °F
HOLD
°C
°FHOLD
RH
7
89
S
H
HI 9064
Hygrometer
RH% 5 . . . . . . 95
2
5
423
ON/OFF
1) Liquid Crystal Display (LCD)
2) HOLD key, to freeze reading on display
3) ON/OFF key, to turn the meter on and off
4) °F key, to display temperature in Fahrenheit (HI 9065 only)
5) °C key, to display temperature in Centigrade (HI 9065 only)
6) RH key, to display RH measurements (HI 9065 only)
7) Battery compartment
8) RH probe socket
9) Temperature probe socket with protective cap (HI 9065 only) for
HI 765 probe series (see “Accessories” section)
10) Shielded cable of RH probe
11) Low RH calibration trimmer
12) High RH calibration trimmer
13) Polypropylene probe body
14) Perforated protective cap, to prevent damages to the sensor
HI 9064HI 9065
Range5.0 to 95.0% RH
--0.0 to 60.0°C / 32 to 140°F
Resolution0.1% RH
--0.1°C / 1°F
Accuracy±2% RH
--±0.4°C / ±1 °F
(for 1 year, excluding probe error)
Typical EMC±3% RH
Deviation--±2°C / ±4°F
RH CalibrationManual, 2 point
through trimmers on RH probe
ProbeHI 70608/2HI 70605/2
Environment0 to 50°C (32 to 122°F);
RH max 100%
Battery Type / Life
4 x 1.5V AA / approx. 500 hours of continuous use
Dimensions196 x 80 x 60 mm
(7.7 x 3.1 x 2.4")
Weight500 g (1.1 lb.)
45
OPERATIONAL GUIDE
INITIAL PREPARATION
Each meter is supplied complete with four 1.5V
AA batteries. Remove the back cover, unwrap
the batteries and install them while paying
attention to their polarity (see “Battery Replacement” section for details).
Connect the RH probe to the DIN socket on the
top of the meter and fasten the screw tightly.
Note: HI 9065 can be used with just one
probe (RH or temperature) at a time.
RH (RELATIVE HUMIDITY) MEASUREMENTS
Turn the meter on by pressing ON/OFF.
Press the RH key to display the RH% measure-
ment. The "H" symbol on the display indicates
RH measurement mode.
ON/OFF
TEMPERATURE MEASUREMENTS (HI 9065 ONLY)
Connect the temperature probe (HI 765 series, see “Accessories”) to
the socket on the top of the meter and the unit will automatically
display the temperature reading.
Note: HI 9065 can be used with just one probe (RH or temperature)
at a time.
Press the “°C” key and the temperature value is displayed in degrees
Centigrade with the “°C” symbol.
°C
The same reading in degrees Fahrenheit is shown if the “°F” key is
pressed.
°F
For any problem in taking measurements, please contact your dealer
or the nearest HANNA Customer Service Department.
°C
RH
H
• For reliable measurements, the end of the humidity detector
should be exposed to a current of air moving at 0.5 m (20") per
second or more.
• In the absence of air movement, the response can be accelerated
by moving the probe.
• The probe sensor must never come into contact with water
or other liquids. If this should happen, or if condensation
causes drops to form on the surface of the humidity sensor, turn
off the instrument and wait until they have evaporated completely. To accelerate the evaporation process, expose the humidity
sensor to a current of air.
67
HOLD FUNCTION
The reading HOLD function is activated by
the HOLD key.
HOLD
The measured value is frozen on the display
when this function key is pressed. "S" is
displayed to indicate that the reading is also
S
stored.
Press the same key again to retirn to normal mode.
DISPLAY CODES
°C temperature readings in degrees Centigrade (HI 9065 only)
H indicate RH (Relative Humidity) readings
S indicate that displayed reading has been stored
CALIBRATION
All HANNA hygrometers have been precalibrated at the factory.
HANNA instruments® uses state-of-the-art thermal humidity chambers for this purpose.
It is generally recommended to have all hygrometers recalibrated at
least once a year. For an accurate annual recalibration, contact the
nearest Hanna Service Center.
You can also check the status of your hygrometer and perform a quick RH recalibration
(with an accuracy of ±5%) by using the
HANNA calibration chamber HI 7101.
The kit is composed of two thermally isolated chambers, both equipped
with a threaded cap and three bottles containing the appropriate
precalibrated saturated salts to produce a known RH value.
PREPARING THE CALIBRATION SOLUTIONS
• Pour approximately 26 cc of distilled water into a glass container.
• Immerse this container into a bath of ice
and water and shake briefly.
Li Cl
75.4%
11.1%
CALIBRATION KIT
Na Cl
O
26 cc
2
H
H
2
O
• Pour approximately 12 cc of distilled water
into the other chamber marked "RH 75.4%".
Na Cl
Li Cl
75.4%
11.1%
12 cc
CALIBRATION KIT
• Add all the content of the HI 7121 bottle
Na Cl
containing NaCl while continuously shaking
the container to avoid the formation of lumps.
Seal this container well when not in use.
The calibration kit needs approximately 4 hours
for stabilization.
Li Cl
11.1%
CALIBRATION KIT
12
9
Na Cl
75.4%
3
4 h
6
CALIBRATION PROCEDURE
• Bring the calibration kit to a temperature of approximately 20°C.
• Remove the cap from the "RH 11.1%" chamber containing the LiCl solution and insert
the probe paying attention not to dip it into
the liquid.
• Remove the adhesive sticker which covers the
calibration trimmer access holes.
Li Cl
75.4%
11.1%
CALIBRATION KIT
Na Cl
• Slowly add the contents of a HI 7111
bottle containing LiCl into the glass container while shaking.
• When the salts have dissolved completely,
add the content of the second HI 7111
bottle.
• Allow the solution to cool, and pour it into
the chamber marked "RH11.1%", making sure that no residue remains on the
walls of the glass container.
• Seal the chamber well when not in use,
as the LiCl solution is extremely hygroscopic and tends to capture the humidity
present in the air causing the solution to
expand in volume and overflow from the
container.
89
Na Cl
Li Cl
75.4%
11.1%
CALIBRATION KIT
Li Cl
ment on.
• Wait for the measurement to stabilize (this
takes about 4 hours).
ON/OFF
12
9
3
4 h
6
• Adjust the low humidity trimmer to read 0.0% (readings between
0.0% and 1.0% RH are also acceptable).
• Press the ON/OFF key to switch the instru-
H
Note: The central trimmer is related to the temperature calibration,
and you will find it on HI 9065 probe only.
• Remove the probe and tightly seal the chamber containing the
LiCl solution.
BATTERY REPLACEMENT
• Remove the cap from the "RH75.4%" chamber containing the NaCl solution and insert
the probe.
• Wait for the measurement to stabilize (approximately 4 hours)
Li Cl
11.1%
CALIBRATION KIT
12
9
Na Cl
75.4%
3
4 h
6
• Adjust the high humidity trimmer to read 64.3%.
H
• Wait for 1 hour and readjust if necessary.
• Leaving the probe in the "RH 75.4%" chamber, adjust the low
humidity trimmer until the display shows 75.4%.
H
HI 9064 and HI 9065 are provided with the advanced BEPS
(Battery Error Prevention System) technology that detects a low power
condition and automatically turns the instruments off to avoid
erroneous readings.
Battery replacement must only take place in a safe area and using
the battery type specified in this instruction manual.
To replace the batteries, simply remove the
two screws on the rear of the instrument
and replace all four 1.5V AA batteries with
new ones, while paying attention to the
correct polarity.
• The relative humidity calibration is now complete.
1011
ACCESSORIES
WARRANTY
HI 70605/2 RH probe with built-in temperature sensor, 2 m (6.6')
cable and perforated cap (for HI 9065)
HI 70605/5 RH probe with built-in temperature sensor, 5 m
(16.5') cable and perforated cap (for HI 9065)
HI 70607/2 RH probe with 2 m (6.6') cable and sintered cap (for
HI 9065)
HI 70607/5 RH probe with 5 m (16.5') cable and sintered cap (for
HI 9065)
HI 70608/2 RH probe with 2 m (6.6') cable (for HI 9064)
HI 70608/5 RH probe with 5 m (16.5') cable (for HI 9064)
HI 7101Calibration chamber for probes with perforated cap,
complete with O-ring and precalibrated saturated salts
HI7111/PSpare saturation LiCl salts for low humidity calibration
(15 g, 6 pcs)
HI7121/PSpare saturation NaCl salts for high humidity calibra-
tion (33 g, 6 pcs)
HI 721317Rugged carrying case
TEMPEARURE PROBES FOR HI 9065
HI 765ATemperature probe for air measurements
HI 765LTemperature probe for liquid measurements
HI 765PWPenetration temperature probe, white handle (also
available with different handle colour)
HI 765WWire temperature probe, whitout handle, for hard-to-
reach places
All Hanna Instruments meters are guaranteed for two years
against defects in workmanship and materials when used for their
intended purpose and maintained according to instructions.
The probes are guaranteed for a period of six months.
This warranty is limited to repair or replacement free of charge.
Damage due to accident, misuse, tampering or lack of prescribed
maintenance are not covered.
If service is required, contact the dealer from whom you purchased the
instrument. If under warranty, report the model number, date of
purchase, serial number and the nature of the failure.
If the repair is not covered by the warranty, you will be notified of the
charges incurred. If the instrument is to be returned to Hanna
Instruments, first obtain a Returned Goods Authorization number
from the Customer Service department and then send it with shipping
costs prepaid.
When shipping any instrument, make sure it is properly packaged for
complete protection.
All rights are reserved. Reproduction in whole or in part is prohibited
without the written consent of the copyright owner.
Hanna Instruments reserves the right to modify the design, construction and
appearance of its products without advance notice.
1213
CE DECLARATION OF CONFORMITY
Recommendations for Users
Before using these products, make sure that they are entirely suitable for the environment in which they are used.
Operation of these instruments in residential area could cause unacceptable interferences to radio and TV equipments, requiring the operator to take all necessary
steps to correct interferences.
Any variation introduced by the user to the supplied equipment may degrade the
instruments' EMC performance. To avoid electrical shock, do not use these instruments when voltages at the measurement surface exceed 24 Vac or 60 Vdc.
To avoid damages or burns, do not perform any measurement in microwave ovens.
In particular cases the HI 9064 and HI 9065 meters could turn off. In these cases
they can be turned on by pressing the ON/OFF key.
SALES AND TECHNICAL SERVICE CONTACTS
Australia:
Tel. (03) 9769.0666 • Fax (03) 9769.0699
China:
Tel. (10) 88570068 • Fax (10) 88570060
Egypt:
Tel. & Fax (02) 2758.683
Germany:
Tel. (07851) 9129-0 • Fax (07851) 9129-99
Greece:
Tel. (210) 823.5192 • Fax (210) 884.0210
Indonesia:
Tel. (21) 4584.2941 • Fax (21) 4584.2942
Japan:
Tel. (03) 3258.9565 • Fax (03) 3258.9567
Korea:
Tel. (02) 2278.5147 • Fax (02) 2264.1729
Malaysia:
Tel. (603) 5638.9940 • Fax (603) 5638.9829
Singapore:
Tel. 6296.7118 • Fax 6291.6906
South Africa:
Tel. (011) 615.6076 • Fax (011) 615.8582
Taiwan:
Tel. 886.2.2739.3014 • Fax 886.2.2739.2983
Thailand:
Tel. 66.2619.0708 • Fax 66.2619.0061
United Kingdom:
Tel. (01525) 850.855 • Fax (01525) 853.668
USA:
Tel. (401) 765.7500 • Fax (401) 765.7575
For e-mail contacts and complete list of Sales and
Technical offices, please see www.hannainst.com
MANRHWPR3 09/05
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.