Hanna Instruments HI 9063 User Manual

Instruction Manual
HI 9063
Portable Waterproof
K-Type Thermocouple
Thermometer
www.hannainst.com
Dear Customer, Thank you for choosing a Hanna product. This manual will provide you with the necessary information for correct operation. Please read it carefully before using the meter. If you need additional technical information, do not hesitate to e-mail us at tech@hannainst.com.
This instrument is in compliance with the directives.
PRELIMINARY EXAMINATION
Remove the instrument from the packing material and exam­ine it carefully to make sure that no damage has occurred during shipping. If there is any noticeable damage, notify your Dealer or the nearest Hanna Office.
Note: Save all packing materials until you are sure that the
instrument functions correctly. Any defective item must be returned in the original packaging together with the supplied accessories.
FUNCTIONAL DESCRIPTION SPECIFICATIONS
Range (*)
Resolution
Accuracy
Typical EMC Deviation
Battery Type Life
Probe Environment Dimensions Weight
(*) Range may be limited by probe.
0.1°C (up to 199.9°C) / 1°C (outside)
0.1°F (up to 399.9°F) / 1°F (outside)
approx. 2000 hours of continuous use
K-type thermocouple (see “Accessories”)
-10 to 50ºC (14 to 122ºF); RH 100%
-50.0 to 1350°C
-58.0 to 2462°F
±0.2% F.S.
for 1 year, excluding probe error
±3°C / ±6°F
(with HI 766 probes)
4 x 1.5V AA (IEC LR6)
196 x 80 x 60 mm (7.7x3.1x2.4")
500 g (1.1 lb.)
WARRANTY
All Hanna Instruments meters are warranted for two years against de­fects in workmanship and materials when used for their intended purpose and maintained according to instructions. Sensors and probes are war- ranted for a period of six months. This warranty is limited to repair or replacement free of charge. Damages due to accidents, misuse, tampering or lack of prescribed mainte­nance are not covered. If service is required, contact the dealer from whom you purchased the instru­ment. If under warranty, report the model number, date of purchase, serial number and the nature of the failure. First obtain a Returned Goods Authorization number from the Customer Ser­vice department, then return the instrument with the Authorization # included along with shipment costs prepaid. If the repair is not covered by the war­ranty, you will be notified of the charges. When shipping any instrument, make sure it is properly packaged for com­plete protection.
Hanna Instruments reserves the right to modify the design, construction and
appearance of its products without advance notice.
GENERAL DESCRIPTION
HI 9063 is a waterproof, microprocessor-based, K-type ther­mocouple thermometer, which provides very accurate measurements in a wide range of temperatures.
The meter is also provided with low battery detection and BEPS (Battery Error Preventing System), which turns the unit off when the batteries are discharged avoiding erroneous readings caused by low battery level.
HI 9063 features include autoranging capability, dual-level LCD for simultaneously displaying of maximum and minimum measured temperatures, °C/°F selection button and hold function.
The meter is supplied complete with four 1.5V AA batteries and instruction manual.
1. Temperature probe connector
2. Liquid Crystal Display
3. ON/OFF key
4. Measuring unit selection key
5. HOLD key
6. HI/LO values reset key
1. Current temperature value
2. HOLD indicator
3. Low battery indicator
4. Minimum temperature value
5. Maximum temperature value
6. K-type probe indicator
7. Measuring unit, °C or °F
FACTORY RECALIBRATION
All Hanna Instruments thermometers have been accurately pre-calibrated at the factory.
It is generally recommended to have all thermometers recalibrated at least once a year.
For an accurate annual recalibration, contact your nearest Hanna Service Center.
OPERATIONAL GUIDE
INITIAL PREPARATION
Each meter is supplied complete with batteries.
Remove the back cover, unwrap the batteries and install them while pay­ing attention to their polarity.
Connect a K-type thermocouple probe to the meter.
To switch on, press the ON/OFF key on the front of the meter.
If the meter does not come on, make sure that the batteries are correctly installed.
The thermometer will carry out a self diagnostic test routine, the LCD will show all segments for a few seconds (or as long as ON/OFF is held), followed by the per­centage indication of the remaining battery life.
The thermometer then enters normal measurement mode. If a temperature probe is plugged in, the meter displays the
measured temperature. If no probe is plugged in, or if reading is over-range, the display shows flash­ing dashes. If a measurement is slightly over the range of the meter specifications, the display flashes the closest full-scale value. To switch the thermometer OFF, press the ON/OFF key.
MEASURING SCALE (°C/°F)
The instrument is factory set to the °C scale, but measurements can be performed in either the Celsius or Fahrenheit scale. Press the °C/°F button to select the desired scale.
HOLD MODE
The HOLD function is activated by press­ing the HOLD key. The measured temperature is held on the display until HOLD is pressed again. The "H" tag blinks on the display to indicate the HOLD mode. Note: Although the display is frozen, internally the meter
continues measuring and updating Hi and Lo values.
HIGH/LOW TEMPERATURES
The maximum and minimum tempera­tures are continuously monitored and displayed in the lower portion of the LCD.
Note: When reading goes over-range, the Hi and Lo values
display dashes until cleared.
CLEAR FUNCTION
Upon pressing the CLEAR key, the current reading is assigned to the highest and lowest temperature val­ues. The High/Low values may be cleared at any time during measure­ment.
BATTERY REPLACEMENT
The instrument is powered by four 1.5 V alkaline batteries and is provided with the Battery Error Prevention System (BEPS), which turns the unit off when a low battery signal is detected.
When the remaining battery level is less than 10%, a warning symbol blinks on the display to indicate a low battery condition.
It is recommended to replace the batteries as soon as the low battery condition is detected.
Battery replacement must only take place in a nonhazardous area using four 1.5V alkaline batteries.
In order to replace rundown batteries, simply remove the two screws on the rear cover of the instrument and replace the four batteries with new ones paying attention to the correct polarity.
Reattach the cover and tighten the two screws.
ACCESSORIES
K-TYPE THERMOCOUPLE PROBES
with integral handle, 1 m (3.3’) cable & mini-connector: HI 766A Roller surface probe, max 320°C/600°F HI 766B Surface probe, max 650°C/1200°F HI 766B1 90º Surface probe, max 450°C/840°F HI 766B2 Spring-loaded, surface probe, max
900°C/1650°F
HI 766B3 Spring-loaded, small surface probe with
insulated shaft, max 200°C/390°F
HI 766C Penetration probe, max 900°C/1650°F HI 766C1 Ultra-fast Penetration probe, max 300°C/
570°F
HI 766D Air probe, max 300°C/570°F HI 766E1 General purpose probe, max 900°C/1650°F HI 766E2 General purpose probe, max 900°C/1650°F HI 766F High temperature, flexible wire probe
without handle, max 1100°C/2000°F
HI 766F1 Flexible wire probe without handle, max
480°C/ 900°F
HI 766TR1 Penetration probe, max 250°C/482°F HI 766TR2 Penetration long probe, max 250°C/482°F HI 766TV1 Pipe clamp probe, max 200°C/390°F
with detachable handle & mini-connector (to be used
in conjunction with the HI 766HD probe handle):
HI 766PA Roller surface probe, max 320°C/600°F HI 766PB Surface probe, max 650°C/1200°F HI 766PC Penetration probe, max 900°C/1650°F HI 766PD Air probe, max 300°C/570°F HI 766PE1 General purpose probe, max 900°C/1650°F HI 766PE2 General purpose probe, max 900°C/1650°F
grill surface probe: HI 766B4 Grill surface probe with 70 cm (27.6’’)
cable (protected with stainless steel jacket), max 250°C/482°F
HI 7664B4S Spare stainless steel sensor for HI766B4 probe
OTHER ACCESSORIES
HI 710021 Spare protective case HI 721317 Rugged carrying case HI 766EX Extension cable for K-type probes HI 766HD Rugged thermocouple probe handle with
1m (3.3') cable fitted with mini-connector
CE DECLARATION OF CONFORMITY
Recommendations for Users
Before using this product, make sure that it is entirely suitable for the environment in which it is used. Operation of this instrument in residential areas could cause unacceptable interference to radio and TV equipment, requiring the operator to take all necessary steps to correct interference. Any variation introduced by the user to the supplied equipment may degrade the instruments' EMC performance. To avoid electrical shock, do not use these instruments when voltage at the measurement surface exceeds 24 Vac or 60 Vdc. To avoid damage or burns, do not perform any measurement in microwave ovens.
IST9063R1 06/05