General Description ................................... 3
PRELIMINARY EXAMINATION
Remove the instrument from the packing material and examine it carefully to make sure
that no damage has occurred during shipping. If there is any damage, notify your
Dealer or the nearest Hanna Customer Service Center.
Each meter comes supplied complete with a
9V battery and instruction manual:
Note: Save all packing material until you are
sure that the instrument functions correctly. All defective item must be returned
in the original packing with the supplied
accessories.
GENERAL DESCRIPTION
Functional Description &
Specifications of HI 9056 .......................... 5
Functional Description &
Specifications of HI 9057 .......................... 6
HI 9056 performs temperature measurements using interchangeable K-type thermocouple probes.
HI 9057 performs temperature measurements with different types of thermocouples
(K - J - T). This meter is equipped with a key
that switches range from K-type to J-type or
T-type thermocouples.
The non-linearity of the temperature probe
is linearized by the built-in microprocessor,
in order to achieve a high degree of accuracy and resolution. Also the HI9056 accurately compensates for temperature drifts in
the measurement circuit and the reference
junction.
HI 9056 has two built-in K-type probe connectors. By simply pressing a button, the
meter switches between the two readings on
the display. A "∆T" button will display the
difference between the two probe readings.
32
Standard features include:
• Interchangeable probes
• Clear Liquid Crystal Display
• Two year Warranty
• Possibility of switching between the Centigrade and the Fahrenheit scale (for
HI 9057 only)
• HOLD function that freezes the reading
on the display (for HI 9057 only)
• HI-LO memory recall function (for HI 9057
only)
• Low battery detection.
All these functions are easily accessible
through the keypad.
Self-explanatory symbols are used to remind the user of the operating mode or
condition.
Hanna manufactures several K-type thermocouple probes for different applications.
See pages 14-17 for more details.
FUNCTIONAL DESCRIPTION AND
SPECIFICATIONS FOR HI 9056
1. Liquid Crystal Display
2. ON/OFF Key
3. Reading Memory/Hold Key
4. Measurement Resolution Key (0.1 or 1)
5. Input Probe Selection Key (T1 or T2)
6. Temperatures Difference Key
7. Memory Recalling Key.
Specifications
Range-200.0 to 1370°C
Resolution0.1°C (-200.0 to 999.9°C)
1°C (1000 to 1370°C)
Accuracy±0.2% F.S.
(@20°C / 68°F)for one year, excluding probe error
Typical EMC±3°C
Deviation±6°F
ProbeK-thermocouple (optional)
(see pages 14-17)
Battery9V/300 hours of continuous use
Environment0 to 50°C (32 to 122°F); 95% RH
Dimensions180 x 83 x 40 mm (7.1 x 3.3 x 1.6")
Weight350 g (12.4 oz.)
54
FUNCTIONAL DESCRIPTION AND
SPECIFICATIONS OF HI 9057
OPERATIONAL GUIDE
INITIAL PREPARATION
Remove the battery cover on
the rear of your thermometer, unwrap the battery and
connect it to the clip. Insert
the battery into the compartment and replace the cover.
1. Liquid Crystal Display
2. ON/OFF Key
3. Reading Memory/Hold Key
4. Thermocouple Type Selection Key (K or J or T)
5. Minimum Temperature Recalling Key
6. Maximum Temperature Recalling Key
7. Measuring Scale key (°C or °F).
Specifications
RangeK-200.0 to 1370°C / -300.0 to 2500°F
J-200.0 to 900.0°C / -300.0 to 1650°F
T-200.0 to 400.0°C / -300.0 to 750.0°F
Resolution0.1°C (up to 999.9°C) 0.1°F(up to 999.9°F)
1°C (over 1000°C) 1°F (over 1000°F)
AccuracyK=±0.2% F.S.
(@20°C / 68°F) J =±0.2% F.S.
T =±0.5% F.S.
for one year, excluding probe error
Typical EMC±3°C
Deviation±6°F
ProbeK, J and T thermocouple (optional)
(for K-thermocouple see pages 14-17)
Battery9V/300 hours of continuous use
Environment0 to 50°C (32 to 122°F); 95% RH
Dimensions180 x 83 x 40 mm (7.1 x 3.3 x 1.6")
Weight350 g (12.4 oz.)
To switch on, press ON/OFF
on the front of the unit.
ON
OFF
The thermometer performs a
self diagnostic test. The LCD
dispalys all the segments at
once, to ensure none are
BAT
HOLD K J T
ºC
ºF
HI
LO
missing.
After the test, the thermometer will switch to
the measurement mode.
If a temperature probe is plugged in, the
unit will display the temperature.
If the probe is not plugged
in, the display will show
"----".
To switch your thermometer off, press ON/
OFF again.
MEASURING SCALE (°C/°F) (HI 9057
only)
When the meter is turned on, it automatically defaults to the °C scale. Measurements
can be performed on either the Centigrade
or the Fahrenheit scale.
76
To change the scale, press °C/°F once.
°C
°F
°F
Check the highest or lowest measured temperature by pressing the respective HI or
LO key. The appropriate display indicator,
"HI" or "LO", will light up together with the
value retrieved from memory.
MEASUREMENT RESOLUTION
When switched on, HI 9056
°C
and HI 9057 default to a 0.1
digit resolution (see specifications).
These instruments automatically change the
resolution to 1 degree when the measured
temperature is higher than 999.9°C.
In HI 9056, when the tem-
perature is lower than
999.9°C it is possible to
change the resolution by
pressing the 0.1/1 key.
HOLD MODE (HI 9057 only)
The reading hold function is activated by
the HOLD key.
To freeze the measured temperature on the
display, press and hold the HOLD key.
"HOLD" is displayed when this key is pressed.
°C
°C
HI
LO
Release the key to resume normal operation.
CLEAR HI-LO FUNCTION (HI 9057
only)
By pressing HI and LO simultaneously the
present measurement reading is assigned in
memory as the highest and lowest temperature reading, i.e. both "HI" and "LO" will
have equal temperature readings.
HOLD
As soon as the HOLD key is released, the
meter returns to normal operational mode.
HI-LO FUNCTION (HI 9057 only)
Maximum and minimum temperatures are
monitored throughout a continuous measuring process.
The values may be recalled or cleared at
any time during the measurement.
MEMORY FUNCTION (HI 9056 only)
The memory & hold function is activated by
MEM HOLD. The measured temperature is
stored in memory and frozen on the display
when this function key is pressed and held.
MEM
HOLD
98
The memory value will be overridden each
time the memory & hold function is activated. A blinking "MEM" on display indicates the operating mode.
Press and hold MR to recall previously stored
temperature reading. In this mode of operation, the display indicator "MEM" is shown.
TEMPERATURE DIFFERENCE (HI 9056 only)
The difference between the temperature
measured by the two probes can be dis-
played by simply pressing ∆T. A "∆" will
appear on the left-hand of the display to
indicate this function.
MEM
If no data is found, "----"
will be displayed.
THERMOCOUPLE TYPE SELECTION
(HI 9057 only)
By pressing KJT the user is allowed to set
the thermometer according to the connected
thermocouple type, and the related symbol
will be displayed, i.e.:
K for K-Type, J for J-Type, T for T-Type.
KTJ
PROBE INPUT SELECTION (HI 9056 only)
When switched on, the instrument automatically shows
the temperature measured by
K 1
Probe 1 and a "K1" appears
on the LCD display.
To view the input temperature of Probe 2,
press T1/T2 and "K2" will be displayed.
K 2
∆
LOW BATTERY DETECTION
To save battery life, HI 9056 and HI 9057
will switch off after 5 minutes.
When the battery is low the
user is prompted by the word
"BAT" to substitute the battery (see page 12).
DISPLAY CODES GUIDE
DisplayDescription
----The probe is not connected to the instrument
BATWeak battery, needs to be replaced
HI 9056:
HOLDFreezes current reading on display and stores
MEMthe reading into memory
MEMIndicates the value of the temperature stored
K1 or K2 Temperature at probe 1 or 2 respectively
∆∆
∆Difference between temperatures at the two
∆∆
HI 9057:
°CThe instrument is reading in the Centigrade
°FThe instrument is reading in the Fahrenheit
HIDisplays the highest measured temperature
LODisplays the lowest measured temperature
HOLDFreezes current reading on display
K,J or T Thermocouple probe type in use
or the reading is out of range
into memory
probes
scale
scale
1110
CALIBRATION
SHOCKPROOF RUBBER BOOTS
All Hanna Instruments thermometers have
been accurately pre-calibrated at the factory.
It is generally recommended to have all
thermometers recalibrated at least once a
year.
For an accurate annual recalibration, contact your nearest Hanna Service Center.
BATTERY REPLACEMENT
When the battery is low
the user is warned by the
"BAT" indication on the
instrument's display. Replace the battery immediately.
Battery should only be replaced in a safe
area using the battery type specified in this
instruction manual (see page 17).
Remove the battery compartment cover on the rear
of the meter and replace
the 9V battery with a new
one. Make sure the battery
contact is tight before replacing the cover.
The meter may turn on automatically when a new battery is connected. You can
turn it off by pressing
ON/OFF.
These rubber boots are specially made to
prolong the life of your instruments and to
prevent damage due to accidental falls.
They and are available in two different colors:
HI 710009= blue colored
HI 710010= orange colored
Select the proper color according to your
preference.
1312
ACCESSORIES
K-Type Thermocouple Probes
with detachable handle & mini-connector
(to be plugged into HI 766HD probe handle):
HI 766PA Roller surface probe, max 320°C/600°F
100mm
4"
280mm
11"
7.8"
HI 766PB Surface probe, max 650°C/1200°F
200mm
7.9"
HI 766PC Penetration probe, max 900°C/1650°F
155mm
6.1"
3mm
0.12"
HI 766PE1 General use probe, max 900°C/1650°F
155mm
6.1"
3mm
0.12"
HI 766PE2 General use probe, max 900°C/1650°F
200mm
7.8"
HI 766HD Rugged thermocouple probe handle
with 1 m (3.3' ) cable fitted with a miniconnector
with integral handle, 1 m cable & miniconnector:
HI 766A Roller surface probe, max 320°C/600°F
5mm
0.20"
100mm
4"
HI 766PD Air probe, max 300°C/570°F
250mm
9.8"
3mm
0.12"
280mm
11"
420mm
13.3"
16.5"
HI 766B Surface probe, max 650°C/1200°F
400mm
15.7"
260mm
10.2"
1514
HI 766B1 90° surface probe, max 450°C/840°F
300mm
11.8"
30mm
1.2"
HI 766E2 General use probe, max 900°C/1650°F
360mm
13.8"
220mm
8.5"
5mm
0.2"
HI 766B2 Surface probe, max 900°C/1650°F
8mm
0.3"
130mm
5.1"
HI 766B3 Small surface probe with insulated shaft,
max 200°C/390°F
5mm
0.2"
130mm
5.1"
HI 766C Penetration probe, max 900°C/1650°F
260mm
10.3"
120mm
4.8"
3mm
0.12"
HI 766D Air probe, max 300°C/570°F
385mm
15.2"
245mm
9.6"
3mm
0.12"
HI 766F High temperature wire probe without
handle, max 1100°C/2000°F
255mm
10"
1.5mm
0.06"
HI 766F1 Flexible probe without handle, max
480°C/900°F
2mm
1m
3.3’
0.08"
HI 766TV1 Pipe clamp probe, max 200°C/390°F
35mm
Max.
1.4"
HI 766E1 General use probe, max 900°C/1650°F
260mm
10.2"
120mm
4.7"
3mm
0.12"
Other accessories:
HI 710002Soft carrying case
HI 710009Blue rubber boot
HI 710010Orange rubber boot
1716
WARRANTY
All Hanna Instruments meters are guaranteed
for two years against defects in workmanship
and materials when used for their intended
purpose and maintained according to instructions. The electrodes and the probes are
guaranteed for a period of six months. This
warranty is limited to repair or replacement
free of charge.
Damage due to accident, misuse, tampering
or lack of prescribed maintenance are not
covered.
If service is required, contact the dealer from
whom you purchased the instrument. If under
warranty, report the model number, date of
purchase, serial number and the nature of the
failure. If the repair is not covered by the
warranty, you will be notified of the charges
incurred. If the instrument is to be returned to
Hanna Instruments, first obtain a Returned
Goods Authorization number from the Customer Service department and then send it
with shipping costs prepaid. When shipping
any instrument, make sure it is properly packaged for complete protection.
To validate your warranty, fill out and return
the enclosed warranty card within 14 days
from the date of purchase.
All rights are reserved. Reproduction in whole
or in part is prohibited without the written
consent of the copyright owner,
Hanna Instruments Inc., 584 Park East Drive,
Woonsocket, Rhode Island, 02895 , USA.
Hanna Instruments reserves the right to
modify the design, construction and appearance of its products without advance
notice.
CE DECLARATION OF CONFORMITY
Recommendations for Users
Before using these products, make sure that they are entirely suitable for the
environment in which they are used.
Operation of these instruments in residential areas could cause unacceptable
interferences to radio and TV equipment,requiring the operator to take all necessary steps to correct interferences.
Any variation introduced by the user to the supplied equipment may degrade the
instrument's EMC performance.
In particular cases the instruments could turn off. In such cases they can be
turned on by pressing the ON/OFF key.
To avoid electrical shock, do not use these instruments when voltages at the
measurement surface exceed 24VAC or 60 VDC.
To avoid damages or burns, do not perform any measurement in microwave
ovens.
1918
HANNA LITERATUREHANNA LITERATURE
HANNA LITERATURE
HANNA LITERATUREHANNA LITERATURE
Hanna publishes a wide range of catalogs and
handbooks for an equally wide range of applications.
The reference literature currently covers areas such as:
• Water Treatment
• Process
• Swimming Pools
• Agriculture
• Food
• Laboratory
• Thermometry
and many others. New reference material is constantly
being added to the library.
For these and other catalogs, handbooks and leaflets,
contact your dealer or the Hanna Customer Service
Center nearest to you. To find the Hanna Office in
your vicinity, check our home page at
www.hannainst.com.
http://www.hannainst.com
MAN9056R2 04/00