Dear Customer, Thank you for choosing a Hanna Product. Please read this instruction
manual carefully before using the instrument. It will provide you with the necessary information for the correct use of the instrument, as well as a precise idea of its versatility.
These instruments are in compliance with the EN 50081-1 and EN 50082-1.
TABLE OF CONTENTSTABLE OF CONTENTS
TABLE OF CONTENTS
TABLE OF CONTENTSTABLE OF CONTENTS
HI
LO
M
PRELIMINARY EXAMINATION....................................................... 3
GENERAL DESCRIPTION .............................................................. 3
FUNCTIONAL DESCRIPTION LCD .................................................. 4
FUNCTIONAL DESCRIPTION HI 9040........................................... 5
FUNCTIONAL DESCRIPTION HI 9041........................................... 6
FUNCTIONAL DESCRIPTION HI 9050 & HI 9051.......................... 7
ON
SPECIFICATIONS ........................................................................ 8
OPERATIONAL GUIDE ............................................................... 10
FUNCTIONAL OPERATION GUIDE............................................... 13
DISPLAY CODES ....................................................................... 14
CALIBRATION........................................................................... 14
BATTERY REPLACEMENT............................................................ 15
TEMPERATURE PROBES & ACCESSORIES.................................... 15
WARRANTY.............................................................................. 17
OTHER PRODUCTS FROM HANNA .............................................. 18
CE DECLARATION OF CONFORMITY............................................ 19
directives
http://www.hannainst.com
These Instruments are in
Compliance with the CE Directives
ISO 9000 Certified Company since 1992
2
Untitled-5
°C
°C
°F
HI
LO
M
°C
°F
HI
LO
M
3
PRELIMINARYPRELIMINARY
PRELIMINARY
PRELIMINARYPRELIMINARY
Remove the instrument from the packing material and examine it carefully to make sure that no damage has occurred during shipment. If noticeable damage is found, contact your Dealer immediately.
All models, except HI 9041, are supplied complete with an HI 765BL temperature probe. A 9V battery is also included.
Note: Save all packing material until you are sure that the meter
functions correctly. Any damaged or defective item must be returned in its original packing with the supplied accessories.
GENERALGENERAL
GENERAL
GENERALGENERAL
HI 9040, HI 9041, HI 9050 and HI 9051 are microprocessor-based thermometers which allow temperature measurement using a cali­brated thermistor probe. The temperature probe is linearized by the built-in microprocessor. HI 9041 has been specifically designed by Hanna Instruments for use in the food industry. Standard features include:
• display and clearing of maximum/minimum temperature mea­surements
• memory entry and hold
• low battery detection
Additional functions (HI 9050 and HI 9051) include:
• sound alarm with preset operating range
• display of constantly updated temperature differential between stored and measured values.
These functions are easily accessible through the splash-proof mem­brane keypad. Self-explanatory symbols are used to remind the user of the operating mode or condition.
EXAMINATIONEXAMINATION
EXAMINATION
EXAMINATIONEXAMINATION
DESCRIPTIONDESCRIPTION
DESCRIPTION
DESCRIPTIONDESCRIPTION
4
FUNCTIONALFUNCTIONAL
FUNCTIONAL
FUNCTIONALFUNCTIONAL
1
1
1) Measuring Scale
2) Maximum and Minimum Temperature Indicators
3) Hold and Memory Indicator
4) Alarm Indicators.
DESCRIPTIONDESCRIPTION
DESCRIPTION
DESCRIPTIONDESCRIPTION
HI 9040
HI 9041
HI 9050 and HI 9051
LCDLCD
LCD
LCDLCD
2
2
3
4
3
3
4
Untitled-5
5
FUNCTIONAL DESCRIPTION HI 9040FUNCTIONAL DESCRIPTION HI 9040
FUNCTIONAL DESCRIPTION HI 9040
FUNCTIONAL DESCRIPTION HI 9040FUNCTIONAL DESCRIPTION HI 9040
6
FUNCTIONAL DESCRIPTION HI 9041FUNCTIONAL DESCRIPTION HI 9041
FUNCTIONAL DESCRIPTION HI 9041
FUNCTIONAL DESCRIPTION HI 9041FUNCTIONAL DESCRIPTION HI 9041
1
1
s
ins
rum n
HI 9040 microcomputer thermometer
3
4
°C °F
HI
ON
OFF
-50 .. .. +150 C
-58 .. .. +302 F
HOLDCLEAR
LO
2
ON/OFF
RANGE
7
65
-50°C to 150°C
ACCURACY
±0.4°C
Foodcare
MICROCOMPUTER THERMOMETER
HI 9041
2
1) Liquid Crystal Display
2) ON/OFF Key
3) Range Selection Key
4) Max./Min. Reset Key
5) Maximum Temperature Key
6) Minimum Temperature Key
7) HOLD Measurement Key.
1) Liquid Crystal Display
2) ON/OFF Key.
5
6
Untitled-5
7
FUNCTIONAL DESCRIPTIONFUNCTIONAL DESCRIPTION
FUNCTIONAL DESCRIPTION
FUNCTIONAL DESCRIPTIONFUNCTIONAL DESCRIPTION
HI 9050 & HI 9051HI 9050 & HI 9051
HI 9050 & HI 9051
HI 9050 & HI 9051HI 9050 & HI 9051
1) Liquid Crystal Display
2) Max/Min Reset Key
3) Max. Temperature/Upper Alarm Value Key
4) Alarm Limits Selection Key
5) Memory Entry and Hold Measurement Key
6) Memory Recall/Temperature Differential Key
7) Activate/Deactivate Alarm Key
8) Min. Temperature/Lower Alarm Value Key
9) ON/OFF Key.
8
SPECIFICATIONSSPECIFICATIONS
SPECIFICATIONS
SPECIFICATIONSSPECIFICATIONS
Range -50.0 to 150.0°C Resolution 0.1°C Accuracy ±0.4°C / ±0.8°F
(@20°C/68°F) for one year, excluding probe error Typical EMC ±0.5°C / ±1°F
Deviation with HI 765BL probe Probe HI 765BL with 1 m (3.3') cable (included) Battery 9V / approx. 300 hours of continuous use Environment 0 to 50°C (32 to 122°F);
Dimensions 180 x 83 x 40 mm (7.1 x 3.3 x 1.6") Weight 350 g (12.4 oz)
Range -50.0 to 150.0°C Resolution 0.1°C Accuracy ±0.4°C (@20°C/68°F) for one year, excluding probe error Typical EMC ±0.5°C Deviation with HI 765BL probe Battery 9 V / 300 approx. hours of continuous use Environment 0 to 50°C (32 to 122°F);
Dimensions 180 x 83 x 40 mm (7.1 x 3.3 x 1.6") Weight 350 g (12.4 oz)
-58.0 to 302°F
0.2°F(-58.0 to 199.8°F)/1°F(200 to 302°F)
max 95% RH non-condensing
max 95% RH non-condensing
HI 9040
HI 9041
7
8
Untitled-5
9
Range -50.0 to 150.0°C Resolution 0.1°C Accuracy ±0.4°C (@20°C/68°F) for one year, excluding probe error Typical EMC ±0.5°C Deviation with HI 765BL probe Probe HI 765BL with 1 m (3.3') cable (included) Battery 9V / 300 approx. hours of continuous use Environment 0 to 50°C (32 to 122°F);
Dimensions 180 x 83 x 40 mm (7.1 x 3.3 x 1.6") Weight 350 g (12.4 oz)
Range -58.0 to 199.8 & -58 to 302°F Resolution 0.2 & ±1°F Accuracy ±0.8°F (@20°C/68°F) for one year, excluding probe error Typical EMC ±1°F Deviation with HI 765BL probe Probe HI 765BL with 1 m (3.3') cable (included) Battery 9V / 300 approx. hours of continuous use Environment 0 to 50°C (32 to 122°F);
Dimensions 180 x 83 x 40 mm (7.1 x 3.3 x 1.6") Weight 350 g (12.4 oz)
max 95% RH non-condensing
max 95% RH non-condensing
HI 9050
HI 9051
10
OPERATIONAL GUIDEOPERATIONAL GUIDE
OPERATIONAL GUIDE
OPERATIONAL GUIDEOPERATIONAL GUIDE
MEASURING SCALE - °C/°F (HI 9040 only)
The meter automatically defaults to the °C scale when turned on. Measurements can be performed in either the Centigrade or Fahren­heit scale. To change the scale, press the "°C/°F" button once.
HOLD MODE (except for HI 9041)
The hold function is activated by either the HOLD (HI 9040) or HOLD/MEM (HI 9050 and HI 9051) key. The measured tempera­ture is frozen on the display when this key is pressed. A blinking "M" on display indicates the operational mode. By pressing the key one more time, the meter returns to measure­ment mode.
HI/LO FUNCTION (except for HI 9041)
The maximum and minimum temperatures are continuously monitored. The contents may be recalled and/or cleared at any time during measurement. Enquires of the highest or lowest temperature measured may be made by pressing and hold­ing down the respective HI or LO key. The appropriate display indicator, "HI" or "LO", will light up together with the value retrieved from memory. Release the key to resume normal operation.
CLEAR FUNCTION (except for HI 9041)
Upon pressing the CLEAR key, the memory is cleared and the present measurement read­ing is assigned to the highest and lowest temperature memories, i.e. the temperature reading is equal for both memories. The "HI" and "LO" display indicators will blink twice to alert the user.
MEM
HOLD
HI LO
HI LO
CLEAR
°C °F
HOLD
M
HI
HI LO
9
10
Untitled-5
11
ALARM (HI 9050 and HI 9051 only)
During the start-up, the alarm is disabled. It may be enabled or disabled again by press­ing the ENABLE key. Both the high and low setpoints are by default set to 25°C (or 77°F). The alarm is triggered when the measured value is greater than the high setpoint or lower than the low setpoint. In such an instance, the corresponding graphics indicator blinks ac­companied by an audible sound. The beeper may be turned off by pressing any key except the ON/OFF key (The ON/OFF key will turn the unit off and clear the memory). It will automatically be reacti­vated once the alarm condition is cleared. Reactivation is also possible by turning the alarm off and on again. Note that the setpoint values will not be affected by the on-off action. To set the alarm limits, press the SET key. The flashing "HI
)
"LO
)" alarm indicators request the user to select the limits.
SET
Press the HI key to enter set mode for high temperature point or the LO key for low tem­perature limit. If the HI key is chosen, the low alarm indicator is switched off leaving the "HI" indicator blinking with its graphics indicator held static. Enter the setpoint value with the ! and " keys. Holding the keys down enables rapid change. Exit from the high alarm limit setting by pressing the SET key for a second time. The user may now proceed to set the low alarm limit or return to the measurement mode by pressing SET again. The setting of low tem­perature limit works in a similar manner. In case of improper setpoint entries, a continuous false alarm beeper will sound.
MEMORY MODE (HI 9050 and HI 9051 only)
There are two modes of operation. The first is the Memory & Hold function which is
activated by the MEM HOLD key. The measured temperature is memorized and frozen on the
11
ENABLE
HI •)
)
LO •)
)
HI LO
HI LO
SET
MEM
HOLD
)
)" and
12
display when this key is pressed. The previously memorized tem­perature will be overridden in this case. A blinking "M" on the display indicates the operating mode. Press the MEM/HOLD Key again to revert back to regular operating mode. The next memory function key, MR/∆M, serves two purposes:
- The first entry enables recall of the latest memorized temperature. If no data is found, "---" will be displayed and the subsequent function will be aborted on the next key entry. In this mode, the "M" indicator also appears on the bot­tom right corner of the LCD.
- The "M" symbol indicates the constantly updated difference between the memory content and the current temperature be­ing measured. Calculation is performed by subtracting the measured value from the memory content, i.e.:
M = (M-T)
where M = memory content
T = measured temperature
A negative result therefore represents a lower memorized value.
HI
Press the MR/M key once again to exit this mode. While in the memory mode, other functions such as HI, LO and
Alarm remain active.
LOW BATTERY DETECTION
The instrument stops operating when a low battery condition is detected. The user is warned by "Eb" displayed on the LCD.
12
M
MR
M
M
Untitled-5
13
FUNCTIONAL OPERATION GUIDEFUNCTIONAL OPERATION GUIDE
FUNCTIONAL OPERATION GUIDE
FUNCTIONAL OPERATION GUIDEFUNCTIONAL OPERATION GUIDE
Function Key Indicator Remarks
Lowest Temperature LO LO Hold down LO to read Highest Temperature HI HI Hold down HI to read Clear HI/LO memories CLEAR "HI/LO" Blinks Twice Hold and Memory HOLD (HI 9040) "M" Store reading in memory and
(HI 9050 and HI 9051 HOLD/MEM freeze it on display only) (HI 9050 and Press again to release display
For HI 9050 and HI 9051 only:
On/Off alarm ENABLE HI •
Set high alarm SET HI •
Set Low alarm SE T HI •
Scroll up/down (for alarm setting only) by 0.1 steps. Hold to accelerate
Memory recall MR/M M Recall stored memory Temperatures MR/∆M M Subtract present temperature
Differential from memory content
Measurement Mode °C/°F °C/°F Toggle between °C and °F (HI9040)
HI 9051) with reading retained in
) )
LO •
) )
LO •
HI HI •
)
! or " HI •)) 3.Press either key to adjust
SET HI •
SET 5.Exit alarm setting. Activate
LO LO •
) )
LO •
)) 1.Wait for selection
LO •
)) )
! or " LO •)) 3.Press either key to choose
LO •
) )
SET HI •
SET 5.Exit alarm setting. Activate
! or " Increase or decrease display
MR/M Return to normal code
memory
) Indicators disappear if key is ) pressed again
) 1. Waiting for selection )
) 2.High alarm set mode
setpoint
) 4.End high alarm set mode ) and wait for next selection
alarm manually
) 2.Low alarm set mode
setpoint
) 4.End low alarm set mode ) and wait for next selection
alarm manually
change
14
DISPLAY CODESDISPLAY CODES
DISPLAY CODES
DISPLAY CODESDISPLAY CODES
Display Description
Er Probe disconnected or temperature out of range Eb Low battery error All segments light up Out of range or no probe connected to the instrument
for more than 3 seconds
HI 9050 and HI 9051 only:
)
HI"•
)" Measured value exceeds HI alarm setpoint
)
LO •
)
HI •)) Measured value falls below LO alarm setpoint LO"•
))"
- - - M Empty memory
CALIBRATIONCALIBRATION
CALIBRATION
CALIBRATIONCALIBRATION
All Hanna thermometers have been accurately pre-calibrated at the factory.
It is, however, recommended to have all thermometers recalibrated at least once a year. For an accurate annual recalibration, contact your nearest Hanna Service Center.
Hanna Test Plugs provide a quick and easy way to test the meter's accuracy by simply connecting these Test Plugs to the probe connector of the meter. If the reading differs by more than ±0.4°C (±0.8°F) from the Test Plug value, the unit is due for recalibration. Choose the Test Plug closest to your measurement range to best suit your needs:
HI 765-18C Calibration key, –18.0°C ±0.4°C HI 765000C Calibration key, 0.0°C ±0.4°C HI 765070C Calibration key, 70.0°C ±0.4°C HI 765-004F Calibration key, –0.4°F ±0.8°F HI 765032F Calibration key, 32.0°F ±0.8°F HI 765158F Calibration key, 158.0°F ±0.8°F
13
14
Untitled-5
15
BATTERY REPLACEMENTBATTERY REPLACEMENT
BATTERY REPLACEMENT
BATTERY REPLACEMENTBATTERY REPLACEMENT
The instrument stops operating when the low battery signal is detected. The user is warned of this by the code "Eb". Battery replacement should only take place in a non-hazardous area using an alka­line 9V battery.
Remove the battery cover at the rear of the meter and replace the 9V battery with a new one. Press in the battery contacts to make sure they are properly inserted be­fore replacing the cover.
TEMPERATURETEMPERATURE
TEMPERATURE
TEMPERATURETEMPERATURE
Hanna Instruments offers a wide range of probes to meet all your requirements from liquids to air and penetration. These probes use highly sensitive thermistor sensors which provide greater accuracy, faster response and a temperature range that is superior to conven­tional thermistor probes. All Hanna temperature probes are supplied pre-calibrated from the factory and are ready to be used with your meter. The probes are easily to your meter with a standard connector. Completely interchangeable, these probes make it possible for you to switch from one to another without wasting time and money going through time-consuming and tedious calibration procedures. The probes are available in different colors to avoid cross contamina­tion during testing:
HI 765A Air probe, with 1 m (3.3') cable and white handle HI 765A/10 Air probe, with 10 m (33') cable and white handle HI 765BL General purpose liquid probe, with 1 m (3.3') cable
HI 765BL/10 General purpose liquid probe, with 10 m (33') cable HI 765L General purpose liquid probe, with 1 m (3.3') cable HI 765L/10 General purpose liquid probe, with 10 m (33') cable HI 765PBL Penetration probe with 1 m (3.3') cable and blue
PROBES & ACCESSORIESPROBES & ACCESSORIES
PROBES & ACCESSORIES
PROBES & ACCESSORIESPROBES & ACCESSORIES
and black handle and black handle and white handle and white handle handle
BATTERY
16
HI 765PBL/10 Penetration probe with 10 m (33') cable and blue HI 765PG Penetration probe with 1 m (3.3') cable and green HI 765PG/10 Penetration probe with 10 m (33') cable and green HI 765PR Penetration probe with 1 m (3.3') cable and red HI 765PR/10 Penetration probe with 10 m (33') cable and red HI 765PW Penetration probe with 1 m 3.3') cable and white HI 765PW/10 Penetration probe with 10 m (33') cable and white HI 765W Wire probe, without handle (hard-to-reach places) HI 765W/10 Wire probe, without handle (hard-to-reach places)
handle handle handle handle handle handle handle with 1 m cable with 10 m cable
HANNA INSTRUMENTS TEMPERATURE PROBES
15
16
Untitled-5
17
OTHER ACCESSORIES
HI 710002 Soft carrying case (formerly BORGM) HI 710009 Blue rubber boot HI 710010 Orange rubber boot HI 710031 Hard carrying case (formerly PKGCASE) HI 721310 9V Battery (10 pcs) MAN40R2 Instruction Manual
WARRANTYWARRANTY
WARRANTY
WARRANTYWARRANTY
All Hanna Instruments meters are warranted for two years against defects in workmanship and materials when used for their intended purpose and maintained according to instructions. The probes are warranted for 6 months.
This warranty is limited to repair or replacement free of charge. Damage due to accidents, misuse, tampering or lack of prescribed maintenance are not covered. If service is required, contact the dealer from whom you purchased the instrument. If under warranty, report the model number, date of purchase, serial number and the nature of the failure. If repair is not covered by the warranty, you will be notified of the charges incurred. If the instrument is to be returned to Hanna Instruments, first obtain a Returned Goods Authorization Number from the Customer Service department and then send it with shipment costs prepaid. When shipping any instrument, make sure it is properly packaged for complete protection. To validate your warranty, fill out and return the enclosed warranty card within 14 days from the date of purchase.
All rights are reserved. Reproduction in whole or in part is prohibited without the written consent of the copyright owner, Hanna Instruments Inc., 584 Park East Drive, Woonsocket, Rhode Island, 02895 , USA.
18
OTHER PRODUCTS FROM HANNAOTHER PRODUCTS FROM HANNA
OTHER PRODUCTS FROM HANNA
OTHER PRODUCTS FROM HANNAOTHER PRODUCTS FROM HANNA
• CALIBRATION AND MAINTENANCE SOLUTIONS
• CHEMICAL TEST KITS
• CHLORINE METERS
• CONDUCTIVITY/TDS METERS
• DISSOLVED OXYGEN METERS
• HYGROMETERS
• ION SPECIFIC METERS (Colorimeters)
• MAGNETIC STIRRERS
• NaCl METERS
• pH/ORP/Na ELECTRODES
• pH/ORP/Na METERS
• PUMPS
• REAGENTS
• SOFTWARE
• PROBES (DO, µS/cm, RH, T, TDS)
• TITRATORS
• TRANSMITTERS
• TURBIDITY METERS
• Wide Range of Accessories
Most Hanna meters are available in the following formats:
• BENCH-TOP METERS
• POCKET-SIZED METERS
• PORTABLE METERS
• PRINTING/LOGGING METERS
• PROCESS METERS (Panel and Wall-mounted)
• WATERPROOF METERS
• METERS FOR FOOD INDUSTRY
For additional information, contact your dealer or the nearest Hanna Customer Service Center. You can also e-mail us at: tech@hannainst.com.
Hanna Instruments reserves the right to modify the design, con­struction and appearance of its products without advance notice.
17
18
Untitled-5
19
CE DECLARATION OF CONFORMITYCE DECLARATION OF CONFORMITY
CE DECLARATION OF CONFORMITY
CE DECLARATION OF CONFORMITYCE DECLARATION OF CONFORMITY
DECLARATION OF CONFORMITY
We
Hanna Instruments Srl V.le delle industrie 12 35010 Ronchi di Villafranca (PD) ITALY
herewith certify that the thermometers
HI 9040 HI 9041 HI 9050 HI 9051
have been tested and found to be in compliance with the following regulations:
IEC 801-2 Electrostatic Discharge IEC 801-3 RF Radiated EN 55022 Radiated, Class B
Date of Issue: 30-11-1995
D.Volpato - Engineering Manager
On behalf of
Hanna Instruments S.r.l.
20
HANNA LITERATUREHANNA LITERATURE
HANNA LITERATURE
HANNA LITERATUREHANNA LITERATURE
LAB RECORDING WATER ANALYSIS HANDBOOK
ENVIROCARE GENERAL CATALOG
These and many others catalogs, handbooks and leaflets are avail­able from Hanna. To receive your free copy, contact your dealer or the nearest Hanna Customer Service Center.
PRINTED IN PORTUGAL
MAN40R2
05/97
Recommendations for Users
Before using these products, make sure that they are entirely suitable for the environment in which they are used. Operation of these instruments in residential areas could cause unacceptable interference to radio and TV equipment,
requiring the operator to take all necessary steps to correct interferences. Any variation introduced by the user to the supplied equipment may degrade the instruments' EMC performance. To avoid electrical shock, do not use these instruments when voltages at the measurement surface exceed 24VAC or 60VDC. To avoid damage or burns, do not perform any measurement in microwave ovens. In particular cases the meters could turn off. In these cases they can be turned on by pressing the ON/OFF key.
19
http://www.hannainst.com
Loading...