Instruction Manual
HI 9043 - HI 9044
Portable Microprocessor
K-Type Thermocouple
Thermometers
These Instruments are in
www.hannainst.com
Compliance with CE
Directives
Dear Customer,
Thank you for choosing a Hanna product. This manual will
provide you with the necessary information for correct operation. Please read it carefully before using the meter. If you
need additional technical information, do not hesitate to email us at tech@hannainst.com.
These instruments are in compliance with the directives.
PRELIMINARY EXAMINATION
Remove the instrument from the packing material and
examine it carefully to make sure that no damage has
occurred during shipping. If there is any noticeable damage,
notify your Dealer or the nearest Hanna Office.
Note: Save all packing materials until you are sure that the
instrument functions correctly. Any defective item must
be returned in the original packaging together with
the supplied accessories.
FUNCTIONAL DESCRIPTION
1. Liquid Crystal Display
2. Measuring unit selection key
3. HI/LO values reset key
4. HOLD key
5. ON/OFF key
SPECIFICATIONS
Range (*)
Resolution
Accuracy
(@20°C/68°F)
Typical EMC
Deviation
Display
Battery Type
Life
Probe
Environment
Dimensions
Weight
(*) Range may be limited by probe. The range of HI 9044
is limited to 900°C/1650°F by the fixed probe.
-50.0 to 199.9°C / 200 to 1350°C
-58.0 to 399.9°F / 400 to 2462°F
0.1°C (up to 199.9°C) / 1°C (outside)
0.1°F (up to 399.9°F) / 1°F (outside)
±0.2% F.S.
excluding probe error
±3 °C and ±6 °F
with HI 766 K-Thermocouple probe
Dual-line LCD
1 x 9V (IEC 6LR61) battery
approx. 500 hours of continuous use
HI 766 K-Thermocouple series
• optional for HI 9043
• HI 766E2 fixed for HI 9044
-10 to 50ºC (14 to 122ºF);
max. 95% RH non-condensing
180 x 83 x 40 mm (7.1x3.3x1.6")
226 g (8 oz.) meter only
WARRANTY
All Hanna Instruments meters are warranted for two years against
defects in workmanship and materials when used for their intended purpose and maintained according to instructions. Sensors and probes are
warranted for a period of six months.
This warranty is limited to repair or replacement free of charge.
Damages due to accidents, misuse, tampering or lack of prescribed maintenance are not covered.
If service is required, contact the dealer from whom you purchased the
instrument. If under warranty, report the model number, date of purchase,
serial number and the nature of the failure.
First obtain a Returned Goods Authorization number from the Customer
Service department, then return the instrument with the Authorization #
included along with shipment costs prepaid. If the repair is not covered by
the warranty, you will be notified of the charges.
When shipping any instrument, make sure it is properly packaged for
complete protection.
All rights are reserved. Reproduction in whole or in part is prohibited
without the written consent of the copyright owner, Hanna Instruments
Inc., Woonsocket, Rhode Island, 02895 , USA.
Hanna Instruments reserves the right to modify the design, construction and
appearance of its products without advance notice.
GENERAL DESCRIPTION
HI 9043 and HI 9044 are microprocessor-based thermometers
which allow temperature measurements using K-type thermocouple probes. The non-linearity of the temperature probe is
linearized by the microprocessor.
Standard features include dual-level LCD for simultaneously
displaying of the highest and lowest measured temperatures,
display hold, selectable measurement unit (°C or °F) and
resolution which automatically switches from 0.1° to 1° for
temperatures above 200°C (400°F).
The meters are also provided with low battery detection and
BEPS (Battery Error Preventing System), which turns the
unit off when the battery is discharged avoiding erroneous
readings caused by low battery level.
HI 9043 comes complete with a 9V battery and instruction
manual.
HI 9044 comes complete with HI 766E2 fixed probe, a 9V
battery and instruction manual.
FACTORY RECALIBRATION
All Hanna thermometers have been accurately pre-calibrated at the factory. It is generally recommended to have
all thermometers recalibrated at least once a year.
For an accurate annual recalibration, contact your dealer or
the nearest Hanna Service Center.
1) Current temperature value
2) HOLD indicator
3) Low Battery indicator
4) Minimum temperature value
5) Maximum temperature value
6) K-type probe indicator
7) Measuring unit, °C or °F
OPERATIONAL GUIDE
INITIAL PREPARATION
Remove the battery cover on the back of the thermometer.
Unwrap the supplied 9V battery, connect it to the battery
clip and reattach the cover.
Connect a K-type thermocouple probe to the meter.
Note: In HI 9044, the probe is fixed.
To switch ON, press the ON/OFF key on
the front of the meter.
The thermometer will carry out a self
diagnostic test routine, the LCD will
show all segments for a few seconds
(or as long as ON/OFF is held),
followed by the percentage indication of the remaining battery life.
The thermometer then enters normal measurement mode.
If a temperature probe is plugged in, the meter displays
the measured temperature automatically switching to the
resolution suitable to the temperature range; i.e. 0.1° up
to 199.9°C (399.9°F), or 1° above.
If no probe is plugged in, or if
reading is over-range, the display
shows flashing dashes.
If a measurement is slightly over the range of the meter
specifications, the display flashes the closest full-scale value.
To switch the meter OFF, press the ON/OFF key.
MEASURING SCALE (°C/°F)
The instrument is factory set to the °C scale, but measurements
can be performed in either the Celsius or Fahrenheit scale.
Press the °C/°F button to select the
desired scale.
HOLD MODE
The HOLD function is activated by
pressing the HOLD key.
The measured temperature is held on
the display until HOLD is pressed again.
The "H" tag blinks on the display to
indicate the HOLD mode.
Note: Although the display is frozen, internally the meter
continues measuring and updating Hi and Lo values.
HIGH/LOW TEMPERATURES
The maximum and minimum temperatures are continuously monitored
and displayed on the lower portion
of the LCD.
Note: When reading goes over-range, the Hi and Lo values
display dashes until cleared.
CLEAR FUNCTION
Upon pressing the CLEAR key, the
High/Low values may be cleared at
any time during measurement and
the current reading is assigned to
the highest and lowest temperature
values.
BATTERY REPLACEMENT
The instrument is powered by a 9 V battery and is provided
with the Battery Error Prevention System (BEPS), which
turns the unit off when a low battery signal is detected.
When the remaining battery level is
less than 10%, a warning symbol
blinks on the display to indicate a
low battery condition.
It is recommended to replace the battery as soon as the low
battery condition is detected.
Remove the cover on the meter's back by applying pressure
in the indicated direction. Unplug the rundown battery
and replace it with a new one.
Battery replacement must only take place in a non-hazardous area using a 9V alkaline battery.
ACCESSORIES
K-TYPE THERMOCOUPLE PROBES
with integral handle, 1 m (3.3’) cable & mini-connector:
HI 766A Roller surface probe, max 320°C/600°F
HI 766B Surface probe, max 650°C/1200°F
HI 766B1 90º Surface probe, max 450°C/840°F
HI 766B2 Spring-loaded, surface probe, max
900°C/1650°F
HI 766B3 Spring-loaded, small surface probe with
insulated shaft, max 200°C/390°F
HI 766C Penetration probe, max 900°C/1650°F
HI 766C1 Ultra-fast Penetration probe, max 300°C/
570°F
HI 766D Air probe, max 300°C/570°F
HI 766E1 General purpose probe, max 900°C/1650°F
HI 766E2 General purpose probe, max 900°C/1650°F
HI 766F High temperature, flexible wire probe
without handle, max 1100°C/2000°F
HI 766F1 Flexible wire probe without handle, max
480°C/ 900°F
HI 766TR1 Penetration probe, max 250°C/482°F
HI 766TR2 Penetration long probe, max 250°C/482°F
HI 766TV1 Pipe clamp probe, max 200°C/390°F
with detachable handle & mini-connector (to be used
in conjunction with the HI 766HD probe handle):
HI 766PA Roller surface probe, max 320°C/600°F
HI 766PB Surface probe, max 650°C/1200°F
HI 766PC Penetration probe, max 900°C/1650°F
HI 766PD Air probe, max 300°C/570°F
HI 766PE1 General purpose probe, max 900°C/1650°F
HI 766PE2 General purpose probe, max 900°C/1650°F
grill surface probe:
HI 766B4 Grill surface probe with 70 cm (27.6’’)
cable (protected with stainless steel jacket),
max 250°C/482°F
HI 7664B4S Spare stainless steel sensor for HI766B4 probe
OTHER ACCESSORIES
HI 710002 Soft carrying case
HI 710009/10 Blue/Orange rubber boot
HI 710031 Hard carrying case
HI 721310 9V Battery (10 pcs)
HI 766EX Extension cable for K-type probes
HI 766HD Rugged thermocouple probe handle with
1m (3.3') cable fitted with mini-connector
CE DECLARATION OF CONFORMITY
ACCESSORIES
Recommendations for Users
Before using these products, make sure that they are entirely suitable for the environment in which
they are used.
Operation of these instruments in residential areas could cause unacceptable interference to radio
and TV equipment, requiring the operator to take all necessary steps to correct interference.
Any variation introduced by the user to the supplied equipment may degrade the instruments' EMC
performance.
To avoid electrical shock, do not use these instruments when voltage at the measurement surface
exceeds 24VAC or 60VDC.
To avoid damage or burns, do not perform any measurement in microwave ovens.
ISTR9043 01/03