CE Declaration of Conformity .................... 63
ISO 9000 Certified
Company since 1992
PRELIMINARY EXAMINATION
Remove the instrument from the packing material and examine it to make sure that no
damage has occurred during shipping. If there
is any damage, notify your Dealer.
Each HI 9032 bench meter is supplied complete with:
•HI 7669/2W Temperature Probe
•HI 7686 Conductivity Probe
•External Power Supply (HI710005 or
HI710006)
•HI 76405 Electrode Holder
•Dust Cover
Note: Save all packing material until you are
sure that the instrument functions correctly. All defective items must be returned in the original package together
with the supplied accessories.
GENERAL DESCRIPTION
Hanna Instruments HI9032 is a microprocessor-based, bench-top, conductivity meter
designed for the laboratory stringent demands,
with the ability to perform five types of measurements:
•Conductivity (4 different ranges: 0.0 to
399.9 µS/cm, 399 to 3999 µS/cm, 3.99 to
39.99 mS/cm, 39.9 to 99.9 mS/cm)
•Fertilizer Concentration (F1 or F2 in
the range 0.00 to 39.9 g/l)
•NaCl of sea water saturation (range 0.0
to 130.0%)
Commands are easily accessible through the
user-friendly keyboard:
Through the COND key the meter is set in
the conductivity range, TDS key selects Total
Dissolved Solids measurements, NaCl key
sets the range to NaCl measurement, and the
FERT key selects 1st or 2nd fertilizer measurements.
Temperature compensation can be either automatic or manual. It can be selected through
the key ATC / MTC respectively.
A temperature coefficient factor (for conductivity measurements) can be also selected
through the %TC key (from 0.0 to 3.0%).
The TDS factor (for TDS measurements) can
be selected through the FTR key (from 0.01
to 0.99).
The unit includes memory storages for all
calibration data even after the unit is shut off.
HI 9032 is equipped with two liquid crystal
displays:
•A large one in two sections (primary and
secondary LCD) indicates conductivity,
TDS, NaCl, Fertilizer and temperature coefficient %;
•A smaller one indicates temperature.
The meter also incorporates a standard R232C
output to allow serial communication with a
host computer; a series of programs can be
selected for the setting of the R232C baud
rate.
THE FRONT PANEL
CALIBRATION
ins
HI9032 microcomputer conductivity meter
LO BAT
SAMPLE
1
HOLD
CONDUCTIVITY
FERTILIZER NaCl
CON
CAL
TEMP.
2
rum n
TDS
F1
BUF
F2
HI
LO
Select
Cond. Fert. NaCl T.D.S.
1
Depress
CAL
s
e
2
Depress
30
Depress
ENTER
4
(refer to instruction manual for details)
When on L.C.D. is blinking, place electrode in reference buffer
5
BUF
When displayed on L.C.D. & ~ on L.C.D. is blinking, equipment confirm that
6
BUF
bufferis within ±15% of theoritical set value, but reading is still not stable yet
m
µ
S
When
7
ppm
%
gm
°C
%TC
BUF
accepting the calibration
Equipment return to OPERATION mode for measurement
8
Display
Main
Sub.
Display
Temp. Display
Cond.
7
T.D.S.
4
FTR
1
ATC
MTC
0
or
to enter calibration parameter
....
9
repeat
& when necessary
3
4
& " CON" displayed on L.C.D., depress
: The Result of measurement
: Conditions of the measurement
: Temperature reading
(Keyboard entry is shown by the blinking of Decimal point)
NaCl
Fert.
9
8
65
°C
3°F2
PGM
%TC
Clear
.
key to confirm for
CON
ON
ENTER
1. Large display for Conductivity, Fertil-
izer, NaCl and TDS
This large LCD displays the conductivity,
fertilizer, NaCl and TDS reading. The
corresponding symbol will be displayed
along with the measured value.
A "----" signal will appear on this display
to indicate an out of
range condition.
2. Temperature display
This LCD displays the temperature mea-
sured by the HI7669/2W probe or the
value that was manually set through the
keyboard. Temperature can be selected
between the Celsius and the Fahrenheit
scale by pressing the °C or the °F key
respectively.
If the probe is not connected when the meter
is turned on, the "25°C" (or "77°F") symbol
with a blinking decimal point will default.
3
OFF
4
CAL
CON
54
If the temperature
probe is disconnected during operation, the decimal point
will start blinking.
It will blink also if
the temperature
range is exceeded.
This LCD also displays the program # during
programming.
• Cond key, to select conductivity mea-
surement mode
• Fert. key, to select fertilizer measure-
ment mode and to select Fertilizer #1
or #2 ("F1" or "F2")
• NaCl key, to select NaCl measure-
ment mode
• T.D.S. key, to select total dissolved
solids measurement mode
• FTR key, to select the TDS factor for
TDS measurements (see page 45)
•°C key, to display temperature in the
Celsius scale
•°F key, to display temperature in the
Fahrenheit scale
• ON key, to switch the instrument on
• OFF key, to switch the instrument off
• CAL key, to enter or exit the calibra-
tion mode
• CON key, to confirm calibration data
• ENTER Key, to confirm/enter the nu-
meric data entry
• 0 to 9 keys, numeric and decimal keys
for numeric data input
• Clear Key, to clear the numeric data
entry
LIQUID CRYSTAL DISPLAYS
Large display for Conductivity, Fertilizer, NaCl
and TDS
Primary Display
LO BAT
SAMPLE
HOLD
CONDUCTIVITY
FERTILIZERNaCl
CON
CAL
Secondary Display
BUF
TDS
F1
F2
m
µ
ppm
gm
°C
%TC
S
%
• ATM/MTC key, to select automatic or
manual temperature compensation
• %T.C. key, to select the temperature
coefficient for conductivity measurements
or to select a different conductivity buffer
during calibration (see page 21)
• PGM key, used in programming
HI
LO
TEMP. display for temperature in °C or °F
and program #
76
THE REAR PANEL
1. DIN female socket for the conductivity probe
(HI7686)
2. Temperature probe socket for use with the
HI7669/2W
3. Reset button
4. Potentiometer for conductivity calibration
(0.0µS)
5. RS 232C socket for serial communication.
6. DC Power Socket (for HI710005 or
HI710006).
Probes connection
Connect the supplied conductivity probe
HI7686 securely to the DIN socket provided (#1) by aligning the pins with the socket,
pushing the plug in and tightening the threaded
ring.
For temperature measurements and auto-
matic temperature compensation connect the
temperature probe HI7669/2W to the °C
socket (#2).
RESET button and restart the entire operation.
Power connection
Plug the 12VDC adapter (HI710005 or
HI710006) into the DC socket (#6).
Plug the adapter into the mains.
Note: The instrument uses the following con-
figuration.
+
It is recommendable to use Hanna HI710005
or HI710006 voltage adapters (supplied with
the meters) that use the proper polarity configuration.
HI9032 can also be used with other adapters. In this case, remember to check the
correct polarity of your adapter before connecting it to the meter.
Note:
•Make sure the main line is protected
by a fuse.
•If you need to unplug the instrument,
press the OFF key before disconnecting the meter from the mains.
-
RESET button
The RESET button (#3) is used when the
instrument displays erroneous messages due
to strong electrical interference or when the
instrument's power supply was disconnected
before the meter was switched off. Press the
98
MEASURE
COND.
FERTILIZER
NaCl
TDS
TEMP.
RESOLUTION
COND.
FERTILIZER
NaCl
TDS
TEMP.
ACCURACY
(@20°C/68°F)
TEMP.
SPECIFICATIONS
0.0-399.9 / 399-3999 µS/cm
3.99-39.99 / 39.9-99.9 mS/cm
0.00 to 39.99 g/l
0.0 to 130.0%
0.0-199.9 / 0-1999 ppm;
0.00 to 19.99 g/l
0.0 to 50.0°C / 32.0 to 122.0°F
0.1 µS/cm; 1 µS/cm;
0.01 mS/cm; 0.1 mS/cm
0.01 g/l
0.1 pmm; 1 ppm; 0.01 g/l
0.1°C / 0.1°F
±1% full scale each range
±0.5°C / ±1°F
excluding probe error
0.1%
OPERATIONAL GUIDE
Make sure that the instrument has been calibrated before taking measurements (see calibration procedures on page 21 for conductivity, page 31 for NaCl and 35 for fertilizer).
Connect the conductivity probe to the
meter securely by aligning the pins
with the socket on the back of the
meter, pushing the plug in and tightening the threaded ring.
Connect the temperature probe to the
back of the meter for automatic temperature compensation or for temperature measurements.
If possible, to minimize any EMC interferences, use plastic beakers for
the solutions.
•Immerse the conductivity probe into the
solution submerging the
holes of the sleeve in the
sample.
Agitate the probe lightly in
the solution to remove any
air bubbles which may
have been trapped inside
the sleeve.
Immerse the temperature
probe as close as possible to the conductivity
probe.
1110
%TC
.
ON
•Switch the instrument on
by pressing the ON key.
The meter automatically defaults to the
conductivity mea-
CONDUCTIVITY
m
surement mode with
2.0% as temperature
coefficient (on the
%TC
large LCD) and the
temperature in the
Celsius scale (on the TEMP. LCD).
•Select a different measurement mode by
pressing the following keys:
Fert.
•the Fert. key for fertilizer
#1
•the Fert. key twice for fer-
tilizer #2
8
Fert.
8
Conductivity measurement mode
•By pressing the Cond. key, the display
will show "CONDUCTIVITY" to confirm the
measurement mode.
m
Cond.
S
7
CONDUCTIVITY
S
%TC
•The meter will automatically default the
temperature coefficient % value to 2.0%
(shown on the secondary display) every
time it is turned on.
•Change the temperature coefficient to the
value of the solution, by pressing the %TC
key. The % value will start blinking on the
secondary LCD.
m
S
CONDUCTIVITY
%TC
•the NaCl key for NaCl of
sea water saturation.
•the T.D.S. key for total
dissolved solids
•the Cond. key to enter the
conductivity measurement
mode again.
NaCl
9
T.D.S.
4
Cond.
7
Enter the desired value by pressing the
numeric keys and then press the ENTER
key to confirm the value.
The stored value will be displayed and it
will stop blinking.
E.g. If you want to enter 2.5% as temperature coefficient value, proceed as follows:
Press the %TC key
Press 2
Press .
Press 5
1312
Press the ENTER key.
%TC
.
°F
2
CONDUCTIVITY
%TC
•If further measurements are desired, rinse
the probe with a small amounts of the next
sample and then proceed.
.
5
ENTER
Fertilizer measurement mode
•By pressing the Fert. key the display will
show "FERTILIZER" to confirm the mea-
m
S
surement mode and the secondary display will show "1", i.e. Fertilizer #1.
%TC
PGM
To clear the numeric entry,
Clear
press the CLEAR key.
See page 44 for the procedure to calculate the correct coefficient value.
If you need to take conductivity measurements without ATC, press the ATC/MTC key.
•HI9032 is an auto-ranging conductivity
meter and the reading automatically
switches from one resolution to the next
(from pure water @ 0.1µS/cm up to
99.9 mS/cm).
•When using HI7686 together with HI7669/
2W all readings are automatically compen-
sated for any temperature difference. Wait
for a few minutes for the temperature sensor to attain thermal equilibrium with the
test solution before taking the measurement. If the temperature of the conductivity
probe and the solution is drastically apart,
a longer time should be allowed before taking readings.
•Once the measurement reading stabilizes
the measurement is complete.
Fert.
FERTILIZER
8
gm
•Select the second fertilizer ("2") by pressing the Fert. key again.
Fert.
8
FERTILIZER
gm
•When using HI7686 together with HI 7669/
2W all readings are automatically com-
pensated for any temperature difference.
Wait for a few minutes for the temperature
sensor to attain thermal equilibrium with
the test solution before taking the measurement. If the temperature of the conductivity probe and the solution is drastically apart, a longer time should be allowed before taking readings.
•Once the measurement reading stabilizes
the measurement is complete.
•If further measurements are desired, rinse
the probe with a small amount of the next
solution and test the sample.
1514
TDS
ppm
NaCl measurement mode
•By pressing the NaCl key the display will
show "NaCl" to confirm the measurement
mode.
NaCl
9
NaCl
%
•The meter will automatically default the
TDS factor value to 0.5 (shown on the
secondary display) every time it is turned
on.
•Change the TDS factor to the value of the
solution, by pressing the FTR key. The
TDS factor value will start blinking on the
secondary LCD.
•When using HI7686 together with HI7669/
2W all readings are automatically compen-
sated for any temperature difference. Wait
for a few minutes for the temperature sensor to attain thermal equilibrium with the
test solution before taking the measurement. If the temperature of the conductivity
probe and the solution is drastically apart,
a longer time should be allowed before taking readings.
•Once the measurement reading stabilizes
the measurement is complete.
•If further measurements are desired, rinse
the probe with a small amount of the next
solution and test the sample.
Total Dissolved Solids measurement mode
•By pressing the TDS key the display will
show "TDS" to confirm the measurement
mode.
FTR
TDS
1
ppm
•Enter the desired value by pressing the
numeric keys and then press the ENTER
key to confirm the value.
The stored value will be displayed and it
will stop blinking.
E.g. If you want to enter 0.9 as TDS
factor, proceed as follows:
Press the FTR key
Press .
Press 9
Press the ENTER key.
FTR
1
%TC
.
NaCl
9
ENTER
T.D.S.
4
ppm
TDS
1716
Cond.
7
•To clear the numeric entry,
PGM
press the CLEAR key.
Clear
See page 45 for further information about
the TDS factor.
•Press the ATC/MTC key and the decimal
point will blink on the TEMP. display (both
when in the Celsius and in the Fahrenheit
scale).
ATC
MTC
0
•HI9032 is an auto-ranging TDS meter,
and the reading automatically switches from
one resolution to the next (from 0.00mg/l
up to 19.99 g/l).
•When using HI7686 together with HI7669/
2W all readings are automatically compen-
sated for any temperature difference. Wait
for a few minutes for the temperature sensor to attain thermal equilibrium with the
test solution before taking the measurement. If the temperature of the conductivity
probe and the solution is drastically apart,
a longer time should be allowed before taking readings.
•Once the measurement reading stabilizes
the measurement is complete.
•If further measurements are desired, rinse
the probe with a small amount of the next
solution and test the sample.
With Manual Temperature Compensation
•Immerse the HI7686 probe in the solution
submerging the holes of the sleeve into
the liquid. Tap the probe lightly on the
bottom of the recipient to remove any air
bubbles which may have being trapped
inside the sleeve.
•Take the temperature of
the solution with a
Checktemp or a glass
thermometer.
°F
2
•Press the °C or
the °F key and
°C
3
the temperature
value will start
blinking.
•Enter the temperature of the solution by
pressing the numeric keys and then press
ENTER key to confirm the value. The stored
value will be displayed and the decimal
point will stop blinking.
E.g. If you want to enter 37.2°C as temperature value, proceed as follows:
Press 3
Press 7
Press .
Press 2
Press the ENTER key and the display will
show "37.2°C".
°C
3
%TC
.
To clear the numeric entry,
press the CLEAR key.
°F
2
ENTER
PGM
Clear
•The measurement reading displayed on
the large LCD will be manually compensated for temperature variations.
1918
Loading...
+ 23 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.