For e-mail contacts and a complete list of Sales and
Technical offices, please see www.hannainst.com.
28
MAN9026
01/05
www.hannainst.com
1
Dear Customer,
Thank you for choosing a Hanna Instruments product.
Please read this instruction manual carefully before using the instrument.
This manual will provide you with the necessary information for correct
use of the instrument, as well as a precise idea of its versatility.
If you need additional technical information, do not hesitate to e-mail
us at tech@hannainst.com or turn to the back cover for our
worldwide contact list.
This instrument is in compliance with directives.
WARRANTY
HI 9026 is guaranteed for two years against defects in workmanship
and materials when used for its intended purpose and maintained
according to instructions. Electrodes and probes are guaranteed for six
months. This warranty is limited to repair or replacement free of charge.
Damage due to accidents, misuse, tampering or lack of prescribed
maintenance is not covered.
If service is required, contact the dealer from whom you purchased the
instruments. If under warranty, report the model number, date of
purchase, serial number and the nature of the problem. If the repair
is not covered by the warranty, you will be notified of the charges
incurred. If the instruments are to be returned to Hanna Instruments, first obtain a Returned Goods Authorization number from the
Technical Service department and then send it with shipping costs
prepaid. When shipping any instrument, make sure it is properly
packed for complete protection.
RECOMMENDATIONS FOR USERS
Before using this product, make sure that it is entirely suitable for the
environment in which it is used.
Operation of this instrument in residential areas could cause
unacceptable interferences to radio and TV equipment, requiring
the operator to follow all necessary steps to correct interferences.
The glass bulb at the end of the pH electrode is sensitive to
electrostatic discharges. Avoid touching this glass bulb at all times.
During operation, ESD wrist straps should be worn to avoid possible
damage to the electrode by electrostatic discharges.
Any variation introduced by the user to the supplied equipment may
degrade the instruments EMC performance.
To avoid electrical shock, do not use this instrument when voltages at
the measurement surface exceed 24 VAC or 60 VDC.
To avoid damage or burns, do not perform any measurement in
microwave ovens.
Hanna Instruments reserves the right to modify the design,
construction and appearance of its products without advance notice.
27
OTHER ACCESSORIES
HI 721317Rugged carrying case
HI 740157Plastic electrode refilling pipet (20 pcs)
HI 76405Electrode holder
HI 7662Stainless steel Temperature probe with 1 m (3.3')
screened cable
HI 8427pH and ORP electrode simulator with 1 m (3.3')
coaxial cable ending in female BNC connectors
HI 931001pH and ORP electrode simulator with LCD and 1 m
(3.3') coaxial cable ending in female BNC connectors
PRELIMINARY EXAMINATION
Remove the instrument from the packing material and examine it to
make sure that no damage has occurred during shipping.
If there is any damage, notify your dealer or the nearest Hanna
Customer Service Center.
The meter is supplied with:
HI 1230B combination double-junction, gel pH electrode
HI 7662 stainless steel temperature probe with 1 m (3.3') cable
pH 4.01 & pH 7.01 buffer solutions (20 mL sachet)
100 mL plastic beaker
1.5V AA alkaline batteries (4 pcs)
Instruction manual
Rugged carrying case
Note: Save all packing material until you are sure that the instrument
functions correctly. All defective items must be returned in the
original packing with the supplied accessories.
GENERAL DESCRIPTION
HI 9026 is a state-of-the-art waterproof, heavy-duty pH meter designed to
provide laboratory results and accuracy under harsh industrial conditions.
This meter is provided with a series of new diagnostic features which
add an entirely new dimension to the measuremeny of pH, by
allowing the user to dramatically improve the reliability of the
measurement:
7 memorized buffers (pH 1.68, 4.01, 6.86, 7.01, 9.18, 10.01
and 12.45) for calibration
Custom calibration by entering up to two custom buffers
Messages on the LCD to make the calibration easy and accurate
Diagnostic features to alert the user when the electrode needs
cleaning
Monitoring of the electrode aging
User-selectable calibration time-out to remind when a new
calibration is necessary
HI 9026 can also measure with ORP electrodes, thanks to its capability
to measure mV with a resolution up to 0.1 mV.
Moreover, it offers an extended temperature range from 20 °C (4 °F)
to 120 °C (248 °F), using HI 7662 interchangeable probes.
Plastic-body (Kynar®), double junction, refillable with built-in amplifier,
combination pH electrode. Use: very high humidity.
24
SPECIFICATIONS
5
OPERATIONAL GUIDE
120 mm
4.7"
5mm
0.2"
5mm DIA
0.2"
"S"VERSION
HI 1330
120 mm
4.7"
12 mm
0.5"
HI 1230
75 mm
2.95"
6 mm
0.25"
HI 2031
120 mm
4.7"
12 mm
0.5"
"S"VERSION
HI 1332
210 mm
8.25"
8 mm
0.3"
7.5mm DIA
0.29"
"S"VERSION
HI 1331
INITIAL PREPARATION
The meter is supplied complete with batteries. Remove the back cover,
unwrap the batteries and install them while paying attention to their
polarity (see page 15).
To prepare the instrument for use, connect the pH electrode and the
temperature probe to the BNC and temperature sockets on the top of
the instrument. The temperature probe is used in conjunction with
the pH electrode to utilize the meter's ATC capability, but it can also
be used independently to take temperature measurements. If the
probe is disconnected, temperature can be set manually with the UP
and DOWN arrow keys (see page 7 for details).
Turn the instrument ON by pressing ON/OFF.
At start-up the display will show all the used segments for a few seconds
(or while the button is held), followed by the percentage indication of the
remaining battery charge, then enters measurement mode.
After measurement switch the meter off, clean the electrode and store it
with a few drops of HI 70300 storage solution in the protection cap.
The auto-off feature turns the meter off after 20 minutes with no
button pressed to save battery life. To disable this feature, see Setup
Menu section on page 9.
pH MEASUREMENTS
To take a pH measurement remove the electrode protective cap and simply submerge
the tip (4cm/1½") of the electrode and the
temperature probe into the sample to be tested.
If necessary, press the RANGE key until the
display changes to the pH mode.
Allow for the electrode to adjust and reading
to stabilize (hourglass symbol turns off).
The LCD will show the pH measurement
together with the temperature of the sample.
In order to take more accurate pH measurements, make sure that the
instrument is calibrated (see page 9 for details).
It is recommended that the electrode is always kept wet and rinsed
thoroughly with the sample to be measured before use.
The pH reading is directly affected by temperature. In order for the
meter to measure the pH accurately, temperature must be taken into
consideration. If the sample temperature is different from the temperature at which the pH electrode was kept, allow a few minutes to reach
thermal equilibrium.
To use the meter's Automatic Temperature Compensation feature,
submerge the temperature probe into the sample as close to the
electrode as possible and wait for a few minutes.
If manual temperature compensation (MTC) is desired, the
temperature probe must be disconnected from the instrument.
The display will show the default temperature
of 25 °C or the last temperature set with the
"°C" (or "°F") indicator blinking.
The MTC tag and up & down arrows symbols light up on the LCD to
indicate that the meter is in MTC mode and the arrow keys can be
used to enter the desired temperature value.
Note: When in MTC the user can press and hold the arrow keys, and
the meter will start incrementing/decrementing the temperature
value. The meter keeps measuring and the display is updated
every second.
Glass-body, single junction, refillable, combination pH electrode.
Use: general purpose.
22
ORP MEASUREMENTS
To perform ORP measurements, connect an optional ORP electrode
(see "Accessories" section) to the meter and turn it ON.
If necessary, enter the "mV" mode by pressing RANGE until the
display changes to mV.
Submerge the ORP electrode tip (4cm/1½") into the sample to be
tested and wait a few minutes for the reading to stabilize.
Measurements within the ±699.9 mV range are displayed with 0.1
mV resolution, while outside this range the resolution automatically
switches to 1 mV.
The ATC (or MTC) tag is turned off because mV readings are not
temperature compensated.
7
For accurate ORP measurements, the surface of the electrode must be
clean and smooth. Pretreatment solutions are available to condition
the electrode and improve its response time (see "Accessories" section).
Notes:
When the reading is out of range, the display will flash the closest
full-scale value.
If using pH electrode while in mV mode, the meter will measure
the mV generated by the pH electrode.
MEM & MR FUNCTIONS
From normal measurement mode, pressing the MEM key will freeze
the readings on the display and store into internal memory the
current values (pH and temperature, or mV and temperature)
together with condition and buffer segments.
The MEM tag blinks and the display remains frozen until the MEM
key is pressed again.
Note: While in MEM mode, the user can switch between pH and mV
by pressing the RANGE key.
A stored value can be recalled by pressing MR: the display will show
the value, together with the RCL & MEM tags, as long as the MR key
is pressed.
Note: Pressing MR, only the range that was active at the time MEM
was pressed is shown.
BACKLIGHT FEATURE
The meter is provided with a Backlight feature,
which can be easily toggled on and off through
the keyboard by pressing the LIGHT key.
Note: The backlight automatically shuts off after approximately 1 minute
with no buttons pressed.
ACCESSORIES
pH CALIBRATION SOLUTIONS
HI70004PpH 4.01 Buffer Solution, 20 mL sachet, 25 pcs
HI70006PpH 6.86 Buffer Solution, 20 mL sachet, 25 pcs
HI70007PpH 7.01 Buffer Solution, 20 mL sachet, 25 pcs
HI70009PpH 9.18 Buffer Solution, 20 mL sachet, 25 pcs
HI70010PpH 10.01 Buffer Solution, 20 mL sachet, 25 pcs
HI 7001LpH 1.68 Buffer Solution, 500 mL bottle
HI 7001MpH 1.68 Buffer Solution, 230 mL bottle
HI7004LpH 4.01 Buffer Solution, 500 mL bottle
HI7004MpH 4.01 Buffer Solution, 230 mL bottle
HI7006LpH 6.86 Buffer Solution, 500 mL bottle
HI7006MpH 6.86 Buffer Solution, 230 mL bottle
HI7007MpH 7.01 Buffer Solution, 500 mL bottle
HI7007MpH 7.01 Buffer Solution, 230 mL bottle
HI7009LpH 9.18 Buffer Solution, 500 mL bottle
HI7009MpH 9.18 Buffer Solution, 230 mL bottle
HI7010LpH 10.01 Buffer Solution, 500 mL bottle
HI7010MpH 10.01 Buffer Solution, 230 mL bottle
ELECTRODE STORAGE SOLUTION
HI70300LStorage Solution, 460 mL bottle
HI70300MStorage Solution, 230 mL bottle
ELECTRODE CLEANING SOLUTIONS
HI70000PElectrode Rinse Solution, 20 mL sachet, 25 pcs
HI7061LGeneral Cleaning Solution, 460 mL bottle
HI7061MGeneral Cleaning Solution, 230 mL bottle
HI7073LProtein Cleaning Solution, 460 mL bottle
HI7073MProtein Cleaning Solution, 230 mL bottle
HI7074LInorganic Cleaning Solution, 460 mL bottle
HI7074MInorganic Cleaning Solution, 230 mL bottle
HI7077LOil & Fat Cleaning Solution, 460 mL bottle
HI7077MOil & Fat Cleaning Solution, 230 mL bottle
REFILLING ELECTROLYTE SOLUTIONS (50 mL, 4 pcs)
HI70713.5M KCl+AgCl Electrolyte for single junction electrodes
HI70721M KNO3 Electrolyte
HI70823.5M KCl Electrolyte for double junction electrodes
HI80931M KCl+AgCl Electrolyte
ORP PRETREATMENT SOLUTIONS
HI7091LReducing Pretreatment Solution, 460 mL bottle
HI7091MReducing Pretreatment Solution, 230 mL bottle
HI7092LOxidizing Pretreatment Solution, 460 mL bottle
HI7092MOxidizing Pretreatment Solution, 230 mL bottle
8
21
or
TROUBLESHOOTING GUIDE
SMOTPMYSMELBORPNOITULOS
evissecxe/esnoperwolS
.tfird
.edortceleHpytriDIHnipitedortceleehtkaoS
03rofnoitulos1607
wollofnehtdnasetunim
.erudecorPgninaelCeht
pusetautculfgnidaeR
.)esion(nwoddna
.noitcnujytrid/deggolC
leveletylortcelewoL
sedortceleelballifer(
.)ylno
.edortceleehtnaelC
etylortcelehserfhtiwllifeR
.)ylnosedortceleelballifer(
swohsyalpsiD
elacsllufgniknilb
.eulav
fotuognidaeR
.egnar
sielpmastahtkcehC
;egnarelbarusaemnihtiw
dnaleveletylortcelekcehC
.sutatsedortcelelareneg
fotuoelacsVm
.egnar
yrdroenarbmemyrD
.noitcnuj
00307IHniedortcelekaoS
tsaeltarofnoitulosegarots
.setunim03
swohsyalpsiD
.F°roC°gniknilb
redrofotuO
.eborperutarepmet
erutarepmetecalpeR
.eborp
swohsyalpsiD
.gniknilbNAELC
neewtebecnereffiD
suoiverpdnawen
neebsahnoitarbilac
.detceted
dnaedortcelenaelC
melborpehtfI.etarbilacer
reffubehtkcehc,sniamer
.snoitulos
swohsyalpsiD
yrettabgniknilb
.lobmys
.levelyrettabwoL.seirettabegnahC
krowtonseodreteM
erutarepmethtiw
.eborp
erutarepmetnekorB
.eborp
erutarepmetecalpeR
.eborp
etarbilacotsliafreteM
ytluafsevigro
.sgnidaer
.edortceleHpnekorB.edortceleecalpeR
siLACGNORW
Hpgniruddeyalpsid
.erudecorpnoitarbilac
rognorW
.reffubdetanimatnoc
noitulosreffubtahtkcehC
.hserfdnatcerrocsi
.ffostuhsreteM;seirettabdaeD
sierutaefffo-otuA
,esacsihtni:delbane
retfaffostuhsretem
.esu-nonfonim02
;seirettabegnahC
.FFO/NOsserP
2rE,1rE,0rE
.putratstaegassem
.rorreMORPEEynarorelaedruoytcatnoC
.retneCecivreSannaH
tratstaegassemrlC
.pu
HptluafeddedaoL
.seulavnoitarbilac
.noitarbilacHpmrofreP
pH CALIBRATION
It is recommended to calibrate the instrument frequently, especially if
high accuracy is required.
The pH range should be recalibrated:
a) Whenever the pH electrode is replaced.
b) At least once a week.
c) After testing aggressive chemicals.
d) When extreme accuracy is required.
e) When the calibration time-out is expired (if feature is enabled).
PROCEDURE
The meter offers a choice of 7 memorized buffers (1.68, 4.01, 6.86,
7.01, 9.18, 10.01 and 12.45 pH) and also allows the user to enter
two more pH values for calibration, C1 and C2 (see Custom Buffer
Selection section on page 12 for details).
Pour small quantities of selected buffer solutions into clean beakers.
20
For accurate calibration use two beakers for each buffer solution, the
first one for rinsing the electrode and the second one for calibration.
Remove the protective cap and rinse the electrode with some of
the buffer solution to be used for the first calibration point.
TWO-POINT CALIBRATION
Press the CAL key. The "CAL" and "" indicators will be
displayed. The secondary LCD will display buffer "7.01". If a
different calibration buffer is desired (e.g. 6.86), use the UP
and DOWN arrow keys to change the displayed value.
Submerge the electrode approximately 4 cm (1½") into the
solution, place the temperature probe as close as possible to the
electrode and stir gently.
The LCD will flash "NOT READY" for 12
seconds, then:
if the reading is not close to the selected
buffer, "WRONG " and "WRONG "
will blink alternatively;
if it is close to the selected buffer the
meter will advise the user with an
acoustic signal (if enabled) when the
reading becomes stable and the display
will change to "READY" and blinking "CFM".
9
Press the CFM key to confirm the calibra-
tion: the meter stores the first calibration
point; the primary LCD will show the
calibrated reading, while the secondary
LCD will show the second buffer to be
used for calibration (pH 4.01).
If youre going to calibrate with a
different buffer, select the desired value
by pressing the arrow keys.
Submerge the electrode approximately 4 cm (1½") into the
second buffer solution, place the temperature probe as close as
possible to the electrode and stir gently.
The LCD will flash "NOT READY" for 12 seconds, then:
if the reading is not close to the selected buffer, "WRONG "
and "WRONG " will blink alternatively;
if it is close to the selected buffer the meter
will advise the user with an acoustic
signal (if enabled) when the reading
becomes stable and the display will change
to "READY" and blinking "CFM".
Press the CFM key: the value is stored in memory and the meter
returns to normal mode. The tags corresponding to the buffers
used for calibration will light up together with the condition
bargraph (if enabled).
For refillable electrodes, if the refill solution (electrolyte) is more than 2½
cm (1") below the fill hole, add the appropriate Electrolyte Solution.
MEASUREMENT
Rinse the electrode tip with distilled water, immerse it (4 cm / 1½")
in the sample and stir gently for a few seconds.
For a faster response and to avoid cross contamination of the samples,
rinse the electrode tip with the solution to be tested, before taking
any measurements.
STORAGE PROCEDURE
To minimize clogging and ensure a quick response time, the glass
bulb and the junction should always be kept moist.
When not in use, store it with a few drops of HI 70300 storage
solution in the protective cap.
NEVER STORE THE ELECTRODE IN DISTILLED OR DEIONIZED
WATER.
PERIODIC MAINTENANCE
Inspect electrode and cable. The cable used for the connection to the
meter must be intact and there must be no points of broken
insulation on the cable or cracks on the electrode stem or bulb. If any
scratches or cracks are present, replace the electrode. Rinse off any salt
deposits with water.
Connectors must be perfectly clean and dry.
For refillable electrodes:
Note: The meter automatically skips the buffer used for the first calibration
point to avoid erroneous procedure. A difference of at least 1.5 pH
unit is required between the two buffers used for the offset and slope
calibration: once calibrated at either pH 7.01 or 6.86, the
instrument automatically ignores the other value for the second point
(same for pH 10.01 and 9.18).
Note: During calibration, the secondary LCD shows the selected buffer
value; press RANGE to display the buffer temperature.
Note: To clear a previous calibration and return to the default
values, press CLR at any time after entering the calibration
mode. The LCD will show CLr CAL for one second, and then
will return to normal measurement mode.
The LCD will show an empty bargraph to warn the user that
the meter is not calibrated.
10
Refill the electrode with fresh electrolyte (see the electrodes specifications
to select the correct refilling solution). Allow the electrode to stand
upright for 1 hour. Follow the Storage Procedure above.
CLEANING PROCEDURE
GeneralSoak in Hanna HI 7061 General Cleaning
Solution for approximately ½ hour.
ProteinSoak in Hanna HI 7073 Protein Cleaning
Solution for 15 min.
InorganicSoak in Hanna HI 7074 Inorganic
CleaningSolution for 15 min.
Oil/greaseRinse with Hanna HI 7077 Oil & Fat Cleaning
Solution for 1 min.
IMPORTANT: After performing any of the cleaning procedures, rinse
the electrode thoroughly with distilled water and soak it in HI 70300
Storage Solution for at least 1 hour before taking measurements.
19
ELECTRODE CONDITIONING & MAINTENANCE
Not present in gel electrodes.
PREPARATION PROCEDURE
Remove the electrode protective cap.
DO NOT BE ALARMED IF ANY SALT DEPOSITS ARE PRESENT. This is
normal with electrodes and they will disappear when rinsed with water.
During transport tiny bubbles of air may have formed inside the glass
bulb. The electrode cannot function properly under these conditions.
These bubbles can be removed by "shaking down" the electrode as you
would do with a glass thermometer.
If the bulb and/or junction are dry, soak the electrode in HI 70300
Storage Solution for at least one hour.
ONE-POINT CALIBRATION
For optimum accuracy it is always recommended to perform a twopoint calibration, but for a faster operation it is also possible to carry
out a single-point procedure.
Buffers at pH 7.01 or pH 6.86 (NIST) are normally the most
appropriate for this purpose, even though the meter can be calibrated
with any of the memorized calibration values.
After calibrating the first point (see above), press the CAL key to end
the calibration procedure.
Note: With one-point calibration there is no Condition and only the
frame is shown. Calibration time-out is active.
CUSTOM BUFFER SELECTION
The meter also allows the user to calibrate with custom buffers.
To select a custom buffer press the CUSTOM BUF key while in calibration
mode. Use the arrow keys to set the desired value for C1 (the default
value is 7.00).
Note: To increase the speed, keep pressed the UP or DOWN key.
Once the desired value is reached, the meter will continue
calibration as with memorized buffers.
Pressing again CUSTOM BUF, the meter
will select the second custom buffer. Pressing
CUSTOM BUF once again, the meter will
return to the standard buffers mode.
EXPIRED CALIBRATION
The instrument is provided with a real time clock (RTC), in order to
monitor the time elapsed since the last pH calibration.
The real time clock is reset every time the meter is calibrated and the
expired calibration status is triggered when the meter detects a
calibration time-out. The CAL tag will start blinking to warn the user
that the meter should be recalibrated.
The calibration time-out can be set (see Setup menu section on
page 14) from 0 (function disabled) to 14 days.
For example, if a 4 days time-out has been selected the meter will
issue the alarm exactly 4 days after the last calibration.
However, if at any moment the expiration value is changed (e.g. to
7 days), then the alarm will be immediately recalculated and appear
7 days after the last calibration.
18
11
Notes:
When the meter is not calibrated or calibration is cleared (default
values loaded) there is no expired calibration, and the display
always shows a blinking CAL tag.
When an abnormal condition in the RTC is detected the meter
forces the expired calibration status.
CONDITION
The display is provided with a 5-dot bargraph (unless the feature is
disabled) which gives an indication of the electrode status after
calibration as follows:
Bargraph indicationCondition value
All 5 dots steady81 to 100% of life
4 dots steady61 to 80%
TEMPERATURE CORRELATION
FOR pH SENSITIVE GLASS
The resistance of glass electrodes partially depends on the temperature.
The lower the temperature, the higher the resistance. It takes more time
for the reading to stabilize if the resistance is higher. In addition, the
response time will suffer to a greater degree at temperatures below 25 °C.
3 dots steady41 to 60%
2 dots steady21 to 40%
1 dot steady1 to 20%
1 dot blinking0%
Only frame is ONNo info available
The condition bargraph remains active for 12 hours after calibration,
then only the frame is shown.
Note: When an abnormal condition in the RTC is detected, the condition
is cleared and only the bargraph frame is shown on the display.
CLEAN ELECTRODE
Each time pH calibration is performed, the meter internally compares
the new calibration with the one previously stored.
When this comparison indicates a significant
difference, the CLEAN message blinks on the
LCD to advise the user that the pH electrode
may need to be cleaned (see Electrode
Conditioning & Maintenance section).
After cleaning, perform calibration.
Note: If the calibration data are cleared, the comparison is done
with the default values.
12
Since the resistance of the pH electrode is in the range of 50 200
Mohms, the current across the membrane is in the pico Ampere range.
Large currents can disturb the calibration of the electrode for many
hours.
For these reasons high humidity environments, short circuits and static
discharges can be detrimental to a stable pH reading.
The pH electrodes life also depends on the temperature. If constantly
used at high temperatures, the electrode life is drastically reduced.
Typical Electrode Life
Ambient Temperature1 3 years
90 °CLess than 4 months
120 °CLess than 1 month
Alkaline Error
High concentrations of sodium ions interfere with readings in alkaline
solutions. The pH at which the interference starts to be significant
depends upon the composition of the glass. This interference is called
alkaline error and causes the pH to be underestimated. Hannas glass
formulations have the indicated characteristics.
Sodium Ion Correction for Glass at 20-25 °C
ConcentrationpH Error
0.1 Mol L-1 Na
1.0 Mol L-1 Na
+
+
13.00
13.50
14.00
12.50
13.00
13.50
14.00
0.10
0.14
0.20
0.10
0.18
0.29
0.40
17
PMETSREFFUBHp
CºFº86.110.468.610.781.910.0154.21
02376.110.489.631.764.923.0183.31
51476.100.459.601.793.942.0181.31
010576.100.429.670.733.981.0199.21
519576.100.409.650.772.921.0108.21
028686.100.488.630.722.960.0126.21
527786.110.468.610.781.910.0154.21
036886.120.458.600.741.969.992.21
535996.130.448.699.611.929.931.21
0440196.140.448.689.670.988.989.11
5431107.150.438.689.640.958.938.11
0522117.160.438.689.610.928.907.11
5513127.180.448.689.699.897.975.11
0604127.190.448.689.679.877.944.11
5694137.111.448.699.659.867.923.11
0785147.121.458.699.639.857.912.11
5776167.141.468.600.719.847.901.11
0867177.161.478.610.798.847.900.11
5858187.171.478.620.778.847.919.01
0949197.191.488.630.758.857.928.01
5930218.102.498.640.738.867.937.01
LCD MESSAGE GUIDE
temperature
compensation
mode
pH or mV
measurement
calibration buffers
custom buffers active
temperature reading
or pH buffer value
mode tags
instability
indicator
battery
symbol
calibration
messages
condition
bargraph
arrow keys
active
first or second
calibration point
calibration
messages
TAGS & SYMBOLS
Mode tags light up for indicating the corresponding active mode,
and blink for warning the user.
SETUP on: setup menu mode has been entered.
CFM blinking: ask confirmation of calibration value.
CAL on: calibration mode has been entered.
CAL blinking: meter is not calibrated or calibration is expired.
RCL, MEM on: when recalling stored values.
MEM blinking: measurement stored in the internal memory and
frozen on the display.
Indication of temperature compensation mode:
MTC for manual, ATC for automatic compensation.
Battery symbol blinking: low battery condition. Batteries should
be replaced soon.
Calibration messages.
NOT READY blinking: buffer has been recognized, but reading
is not stable.
READY on: buffer has been recognized and reading is stable.
WRONG and WRONG blinking alternatively: wrong
buffer, value not recognized.
CLEAN blinking: an abnormal difference between new and
previous calibration has been detected. Electrode cleaning is
suggested. Follow the cleaning procedure described in the
Electrode conditioning & maintenance section. If the problem
remains, check the buffer solutions.
pH BUFFER TEMPERATURE
DEPENDENCE
The temperature has an effect on pH. The calibration buffer solutions
are affected by temperature changes to a lesser degree than normal
solutions. During calibration the instrument will automatically calibrate
to the pH value corresponding to the measured or set temperature.
16
During calibration the instrument will display the pH buffer value at
25 °C.
13
SETUP MENU
mV CALIBRATION
The instrument allows the user to configure several parameters
through the Setup Menu.
To enter the Menu mode, while in normal measurement mode, press
and hold the SETUP key for about 5 seconds.
Once the menu is entered, each parameter can be changed by using
the arrow keys; then pressing the CFM key will confirm the value and
scroll to the next parameter.
The sequence of programmable parameters is as follows:
1. Acoustic signal: On (default) or Off
2. Auto-off feature: 20 minutes (default) or disabled
3. Temperature measure unit: ºC (default) or ºF
4. Calibration segments: On (default) or Off (this enables or disables
the bargraph, the CLEAN message and the buffer segments)
5. Calibration time-out: 1 (default) to 14 days or disabled (0 days)
After setting the last parameter, pressing the CFM key will confirm the
value and return to normal measurement mode.
HI9026 has been accurately precalibrated for mV range at the
factory.
For optimum accuracy, it is recommended to recalibrate the meter for
mV readings at least once a year.
Contact your Dealer or the nearest Hanna Customer Service Center for
more information.
TEMPERATURE CALIBRATION
HI9026 has been accurately precalibrated for temperature at the
factory.
For optimum accuracy, it is recommended to recalibrate the meter for
temperature at least once a year.
Contact your Dealer or the nearest Hanna Customer Service Center for
more information.
BATTERY REPLACEMENT
If the batteries become weak, the display
will flash the battery symbol to advise the
user that approximately 25 hours of working time is left. It is recommended to
replace batteries soon.
The meter is also provided with the BEPS
(Battery Error Prevention System) feature, which
automatically turns the instrument off when
the battery level is too low to ensure reliable
readings. At start-up the display will show
0 bAtt for a few seconds, then the meter
automatically turns off.
Battery replacement must only take place in a non hazardous area
using the battery types specified in this instruction manual.
To replace rundown batteries, remove the
rear cover of the instrument and substitute
all four 1.5V AA alkaline batteries with new
ones, while paying attention to the correct
polarity. Reattach the cover and tighten
the two screws.
14
15
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.