Hanna instruments HI 9026 User Manual

Australia:
Tel. (03) 9769.0666  Fax (03) 9769.0699
China:
Tel. (10) 88570068  Fax (10) 88570060
Egypt:
Tel. & Fax (02) 2758.683
Germany:
Tel. (07851) 9129-0  Fax (07851) 9129-99
Greece:
Tel. (210) 823.5192  Fax (210) 884.0210
Indonesia:
Tel. (21) 4584.2941  Fax (21) 4584.2942
Japan:
Tel. (03) 3258.9565  Fax (03) 3258.9567
Korea:
Tel. (02) 2278.5147  Fax (02) 2264.1729
Malaysia:
Tel. (603) 5638.9940  Fax (603) 5638.9829
Instruction Manual
HI 9026
Calibration Check
Waterproof
pH/mV/ºC Meter
Singapore:
Tel. 6296.7118  Fax 6291.6906
South Africa:
Tel. (011) 615.6076  Fax (011) 615.8582
Taiwan:
Tel. 886.2.2739.3014  Fax 886.2.2739.2983
Thailand:
Tel. (662) 619.0708.11  Fax (662) 619.0061
United Kingdom:
Tel. (01525) 850.855  Fax (01525) 853.668
USA:
Tel. (401) 765.7500  Fax (401) 765.7575
For e-mail contacts and a complete list of Sales and Technical offices, please see www.hannainst.com.
28
MAN9026
01/05
www.hannainst.com
1
Dear Customer, Thank you for choosing a Hanna Instruments product. Please read this instruction manual carefully before using the instrument. This manual will provide you with the necessary information for correct
use of the instrument, as well as a precise idea of its versatility. If you need additional technical information, do not hesitate to e-mail
us at tech@hannainst.com or turn to the back cover for our worldwide contact list. This instrument is in compliance with directives.
WARRANTY
HI 9026 is guaranteed for two years against defects in workmanship and materials when used for its intended purpose and maintained according to instructions. Electrodes and probes are guaranteed for six months. This warranty is limited to repair or replacement free of charge.
Damage due to accidents, misuse, tampering or lack of prescribed maintenance is not covered.
If service is required, contact the dealer from whom you purchased the instruments. If under warranty, report the model number, date of purchase, serial number and the nature of the problem. If the repair is not covered by the warranty, you will be notified of the charges incurred. If the instruments are to be returned to Hanna Instru­ments, first obtain a Returned Goods Authorization number from the Technical Service department and then send it with shipping costs prepaid. When shipping any instrument, make sure it is properly packed for complete protection.
RECOMMENDATIONS FOR USERS
Before using this product, make sure that it is entirely suitable for the environment in which it is used.
Operation of this instrument in residential areas could cause unacceptable interferences to radio and TV equipment, requiring the operator to follow all necessary steps to correct interferences.
The glass bulb at the end of the pH electrode is sensitive to electrostatic discharges. Avoid touching this glass bulb at all times.
During operation, ESD wrist straps should be worn to avoid possible damage to the electrode by electrostatic discharges.
Any variation introduced by the user to the supplied equipment may degrade the instruments EMC performance.
To avoid electrical shock, do not use this instrument when voltages at the measurement surface exceed 24 VAC or 60 VDC.
To avoid damage or burns, do not perform any measurement in microwave ovens.
TABLE OF CONTENTS
WARRANTY .................................................................................. 2
PRELIMINARY EXAMINATION ......................................................... 3
GENERAL DESCRIPTION ................................................................. 3
FUNCTIONAL DESCRIPTION ........................................................... 4
SPECIFICATIONS........................................................................... 5
OPERATIONAL GUIDE.................................................................... 6
pH CALIBRATION ......................................................................... 9
pH BUFFER TEMPERATURE DEPENDENCE ....................................... 13
SETUP MENU ............................................................................. 14
mV CALIBRATION ....................................................................... 15
TEMPERATURE CALIBRATION ....................................................... 15
BATTERY REPLACEMENT ............................................................... 15
LCD MESSAGE GUIDE .................................................................. 16
TEMPERATURE CORRELATION FOR pH SENSITIVE GLASS ................ 17
ELECTRODE CONDITIONING & MAINTENANCE ................................ 18
TROUBLESHOOTING GUIDE ......................................................... 20
ACCESSORIES ............................................................................. 21
2
Hanna Instruments reserves the right to modify the design,
construction and appearance of its products without advance notice.
27
OTHER ACCESSORIES
HI 721317 Rugged carrying case HI 740157 Plastic electrode refilling pipet (20 pcs) HI 76405 Electrode holder HI 7662 Stainless steel Temperature probe with 1 m (3.3')
screened cable
HI 8427 pH and ORP electrode simulator with 1 m (3.3')
coaxial cable ending in female BNC connectors
HI 931001 pH and ORP electrode simulator with LCD and 1 m
(3.3') coaxial cable ending in female BNC connectors
PRELIMINARY EXAMINATION
Remove the instrument from the packing material and examine it to make sure that no damage has occurred during shipping.
If there is any damage, notify your dealer or the nearest Hanna Customer Service Center.
The meter is supplied with:  HI 1230B combination double-junction, gel pH electrode  HI 7662 stainless steel temperature probe with 1 m (3.3') cable  pH 4.01 & pH 7.01 buffer solutions (20 mL sachet)  100 mL plastic beaker  1.5V AA alkaline batteries (4 pcs)  Instruction manual  Rugged carrying case
Note: Save all packing material until you are sure that the instrument
functions correctly. All defective items must be returned in the original packing with the supplied accessories.
GENERAL DESCRIPTION
HI 9026 is a state-of-the-art waterproof, heavy-duty pH meter designed to provide laboratory results and accuracy under harsh industrial conditions. This meter is provided with a series of new diagnostic features which add an entirely new dimension to the measuremeny of pH, by allowing the user to dramatically improve the reliability of the measurement:  7 memorized buffers (pH 1.68, 4.01, 6.86, 7.01, 9.18, 10.01
and 12.45) for calibration  Custom calibration by entering up to two custom buffers  Messages on the LCD to make the calibration easy and accurate  Diagnostic features to alert the user when the electrode needs
cleaning  Monitoring of the electrode aging  User-selectable calibration time-out to remind when a new
calibration is necessary HI 9026 can also measure with ORP electrodes, thanks to its capability to measure mV with a resolution up to 0.1 mV. Moreover, it offers an extended temperature range from 20 °C (4 °F) to 120 °C (248 °F), using HI 7662 interchangeable probes.
26
3
FUNCTIONAL DESCRIPTION
110 mm
4.3"
12 mm
0.5"
HI 1413
150 mm
5.9"
12 mm
0.5"
"S" VERSION
HI 3131
120 mm
4.7"
12 mm
0.5"
HI 3230
"S" VERSION
120 mm
4.7"
12 mm
0.5"
HI 4430
"S" VERSION
HI1413B
Glass-body, single junction, flat tip, Viscolene, non-refillable, combination pH electrode. Use: surface measurement.
ORP ELECTRODES
HI3131B
Glass-body, refillable, combination platinum ORP electrode. Use: titration.
HI3230B
Plastic-body (Ultem®), gel-filled, combination platinum ORP electrode. Use: general purpose.
HI4430B
Plastic-body (Ultem®), gel-filled, combination gold ORP electrode. Use: general purpose.
1) BNC electrode connector.
2) Temperature probe socket.
3) Liquid Crystal Display (LCD).
4) RANGE key, to select pH or mV.
5) ON/OFF key, to turn the meter ON and OFF.
6) CAL key, to enter/exit calibration mode.
7) SETUP/CFM key, to enter setup mode or to confirm calibration.
8) UP and DOWN arrow keys, for manual temperature setting,
entering menu parameters or changing buffer values.
9) MR/CUSTOM BUF key, to recall the stored value from memory or
to enter custom buffer values.
10) MEM/CLR key, to store the reading in memory or to clear calibration.
11) LIGHT key, to toggle display backlighting.
12) Secondary display.
13) Primary display.
4
Consult the Hanna General Catalog for a complete and wide selection of electrodes.
Ultem® is registered Trademark of General Electric Co. Kynar® is registered Trademark of Pennwalt Corp.
25
EGNAR
Hp00.61ot00.2
Vm9991±/Vm9.996±
)F°0.842ot0.4(C°0.021ot0.02
NOITULOSER
Hp10.0
Vm1/Vm1.0
)F°1.0(C°1.0
YCARUCCA
Fº86/Cº02@
Hp10.0±
Vm1±/Vm2.0±
)F°8.0±(C°4.0±
)rorreeborpgnidulcxe(
noitaiveDCMElacipyT
Hp20.0±
Vm1±/Vm2.0±
)F°8.0±(C°4.0±
noitarbilaCHp
,erudecorptniop-2ro1
2dnasreffubdeziromem7htiw
sreffubmotsuc
noitarbilaCtesffO Hp1±
noitarbilaCepolS %801ot08morF
noitasnepmoCerutarepmeT
C°0.021ot0.02morf,citamotuA
,launamro)F°0.842ot0.4(
eborperutarepmettuohtiw
edortcelEHp )dedulcni(B0321IH
eborPerutarepmeT )dedulcni(2667IH
ecnadepmItupnI 01
21
smho
efiL&epyTyrettaB
)seirettabenilakla(ezisAAV5.1x4
esusuounitnocfosruoh005.xorppa
ffo-otuA
rosetunim02:elbatcelesresU
delbasid
snoisnemiD )4.2x1.3x7.7(mm06x08x691
)ylnoretem(thgieW ).zo51(g524
tnemnorivnE
)F°22123(C°050
%001HRxam
ytnarraW sraey2
120 mm
4.7"
12 mm
0.5"
FC 100
75 mm
2.95"
6 mm
0.25"
FC 200
120 mm
4.7"
12 mm
0.5"
FC 210
120 mm
4.7"
12 mm
0.5"
9.5mm DIA
0.37"
FC 220
110 mm
4.3"
12 mm
0.5"
FC 911
FC100B
Plastic-body (Kynar®), double junction, refillable, combination pH electrode.
Use: general purpose for food industry.
FC200B
Plastic-body (Kynar®), open junction, conic, Viscolene, non-refillable, combination pH electrode. Use: meat & cheese.
FC210B
Glass-body, double junction, conic, Viscolene, non-refillable, combination pH electrode.
Use: milk, yogurt.
FC220B
Glass-body, triple-ceramic, single junction, refillable, combination pH electrode. Use: food processing.
FC911B
Plastic-body (Kynar®), double junction, refillable with built-in amplifier, combination pH electrode. Use: very high humidity.
24
SPECIFICATIONS
5
OPERATIONAL GUIDE
120 mm
4.7"
5mm
0.2"
5mm DIA
0.2"
"S" VERSION
HI 1330
120 mm
4.7"
12 mm
0.5"
HI 1230
75 mm
2.95"
6 mm
0.25"
HI 2031
120 mm
4.7"
12 mm
0.5"
"S" VERSION
HI 1332
210 mm
8.25"
8 mm
0.3"
7.5mm DIA
0.29"
"S" VERSION
HI 1331
INITIAL PREPARATION
The meter is supplied complete with batteries. Remove the back cover, unwrap the batteries and install them while paying attention to their polarity (see page 15).
To prepare the instrument for use, connect the pH electrode and the temperature probe to the BNC and temperature sockets on the top of the instrument. The temperature probe is used in conjunction with the pH electrode to utilize the meter's ATC capability, but it can also be used independently to take temperature measurements. If the probe is disconnected, temperature can be set manually with the UP and DOWN arrow keys (see page 7 for details).
Turn the instrument ON by pressing ON/OFF. At start-up the display will show all the used segments for a few seconds
(or while the button is held), followed by the percentage indication of the remaining battery charge, then enters measurement mode.
HI1330B
Glass-body, semimicro, single junction, refillable, combination pH electrode. Use: laboratory, vials.
HI1331B
Glass-body, semimicro, single junction, refillable, combination pH electrode. Use: flasks.
HI1230B
Plastic-body (Ultem®), double junction, gel-filled, combination pH electrode.
Use: general, field.
After measurement switch the meter off, clean the electrode and store it with a few drops of HI 70300 storage solution in the protection cap.
The auto-off feature turns the meter off after 20 minutes with no button pressed to save battery life. To disable this feature, see Setup Menu section on page 9.
pH MEASUREMENTS
To take a pH measurement remove the elec­trode protective cap and simply submerge the tip (4cm/1½") of the electrode and the temperature probe into the sample to be tested.
If necessary, press the RANGE key until the display changes to the pH mode.
Allow for the electrode to adjust and reading to stabilize (hourglass symbol turns off).
The LCD will show the pH measurement together with the temperature of the sample.
6
HI2031B
Glass-body, semimicro, conic, refillable, combination pH electrode. Use: semisolid products.
HI1332B
Plastic-body (Ultem®), double junction, refillable, combination pH electrode.
Use: general purpose.
23
120 mm
4.7"
12 mm
0.5"
9.5mm DIA
0.37"
"S" VERSION
HI 1043
120 mm
4.7"
12 mm
0.5"
"S" VERSION
HI 1053
120 mm
12 mm
0.5"
5 mm
0.2"
3 mm
0.12"
3.0 mm DIA
0.12"
HI 1083
120 mm
4.7"
12 mm
0.5"
9.5mm DIA
0.37"
"S" VERSION
HI 1131
ORP SOLUTIONS
HI7020L Test Solution 200-275 mV, 500 mL bottle HI7020M Test Solution 200-275 mV, 230 mL bottle HI7021L Test Solution 240 mV, 500 mL bottle HI7021M Test Solution 240 mV, 230 mL bottle
pH ELECTRODES
All electrodes part numbers ending in B are supplied with a BNC connector and 1 m (3.3') cable, as shown below:
HI1043B
Glass-body, double junction, refillable, combination pH electrode. Use: strong acid/alkali.
HI1053B
Glass-body, triple ceramic, conic shape, refillable, combination pH electrode. Use: emulsions.
In order to take more accurate pH measurements, make sure that the instrument is calibrated (see page 9 for details).
It is recommended that the electrode is always kept wet and rinsed thoroughly with the sample to be measured before use. The pH reading is directly affected by temperature. In order for the meter to measure the pH accurately, temperature must be taken into consideration. If the sample temperature is different from the tempera­ture at which the pH electrode was kept, allow a few minutes to reach thermal equilibrium. To use the meter's Automatic Temperature Compensation feature, submerge the temperature probe into the sample as close to the electrode as possible and wait for a few minutes.
If manual temperature compensation (MTC) is desired, the temperature probe must be disconnected from the instrument.
The display will show the default temperature of 25 °C or the last temperature set with the "°C" (or "°F") indicator blinking.
The MTC tag and up & down arrows symbols light up on the LCD to indicate that the meter is in MTC mode and the arrow keys can be used to enter the desired temperature value.
Note: When in MTC the user can press and hold the arrow keys, and
the meter will start incrementing/decrementing the temperature value. The meter keeps measuring and the display is updated every second.
HI1083B
Glass-body, micro, Viscolene, non-refillable, combination pH electrode. Use: biotechnology, micro titration.
HI1131B
Glass-body, single junction, refillable, combination pH electrode. Use: general purpose.
22
ORP MEASUREMENTS
To perform ORP measurements, connect an optional ORP electrode (see "Accessories" section) to the meter and turn it ON. If necessary, enter the "mV" mode by pressing RANGE until the display changes to mV. Submerge the ORP electrode tip (4cm/1½") into the sample to be tested and wait a few minutes for the reading to stabilize. Measurements within the ±699.9 mV range are displayed with 0.1 mV resolution, while outside this range the resolution automatically switches to 1 mV.
The ATC (or MTC) tag is turned off because mV readings are not temperature compensated.
7
For accurate ORP measurements, the surface of the electrode must be clean and smooth. Pretreatment solutions are available to condition the electrode and improve its response time (see "Accessories" section).
Notes:
 When the reading is out of range, the display will flash the closest
full-scale value.
 If using pH electrode while in mV mode, the meter will measure
the mV generated by the pH electrode.
MEM & MR FUNCTIONS
From normal measurement mode, pressing the MEM key will freeze the readings on the display and store into internal memory the current values (pH and temperature, or mV and temperature) together with condition and buffer segments.
The MEM tag blinks and the display remains frozen until the MEM key is pressed again.
Note: While in MEM mode, the user can switch between pH and mV
by pressing the RANGE key.
A stored value can be recalled by pressing MR: the display will show the value, together with the RCL & MEM tags, as long as the MR key is pressed.
Note: Pressing MR, only the range that was active at the time MEM
was pressed is shown.
BACKLIGHT FEATURE
The meter is provided with a Backlight feature, which can be easily toggled on and off through the keyboard by pressing the LIGHT key.
Note: The backlight automatically shuts off after approximately 1 minute
with no buttons pressed.
ACCESSORIES
pH CALIBRATION SOLUTIONS
HI70004P pH 4.01 Buffer Solution, 20 mL sachet, 25 pcs HI70006P pH 6.86 Buffer Solution, 20 mL sachet, 25 pcs HI70007P pH 7.01 Buffer Solution, 20 mL sachet, 25 pcs HI70009P pH 9.18 Buffer Solution, 20 mL sachet, 25 pcs HI70010P pH 10.01 Buffer Solution, 20 mL sachet, 25 pcs HI 7001L pH 1.68 Buffer Solution, 500 mL bottle HI 7001M pH 1.68 Buffer Solution, 230 mL bottle HI7004L pH 4.01 Buffer Solution, 500 mL bottle HI7004M pH 4.01 Buffer Solution, 230 mL bottle HI7006L pH 6.86 Buffer Solution, 500 mL bottle HI7006M pH 6.86 Buffer Solution, 230 mL bottle HI7007M pH 7.01 Buffer Solution, 500 mL bottle HI7007M pH 7.01 Buffer Solution, 230 mL bottle HI7009L pH 9.18 Buffer Solution, 500 mL bottle HI7009M pH 9.18 Buffer Solution, 230 mL bottle HI7010L pH 10.01 Buffer Solution, 500 mL bottle HI7010M pH 10.01 Buffer Solution, 230 mL bottle
ELECTRODE STORAGE SOLUTION
HI70300L Storage Solution, 460 mL bottle HI70300M Storage Solution, 230 mL bottle
ELECTRODE CLEANING SOLUTIONS
HI70000P Electrode Rinse Solution, 20 mL sachet, 25 pcs HI7061L General Cleaning Solution, 460 mL bottle HI7061M General Cleaning Solution, 230 mL bottle HI7073L Protein Cleaning Solution, 460 mL bottle HI7073M Protein Cleaning Solution, 230 mL bottle HI7074L Inorganic Cleaning Solution, 460 mL bottle HI7074M Inorganic Cleaning Solution, 230 mL bottle HI7077L Oil & Fat Cleaning Solution, 460 mL bottle HI7077M Oil & Fat Cleaning Solution, 230 mL bottle
REFILLING ELECTROLYTE SOLUTIONS (50 mL, 4 pcs)
HI7071 3.5M KCl+AgCl Electrolyte for single junction electrodes HI7072 1M KNO3 Electrolyte HI7082 3.5M KCl Electrolyte for double junction electrodes HI8093 1M KCl+AgCl Electrolyte
ORP PRETREATMENT SOLUTIONS
HI7091L Reducing Pretreatment Solution, 460 mL bottle HI7091M Reducing Pretreatment Solution, 230 mL bottle HI7092L Oxidizing Pretreatment Solution, 460 mL bottle HI7092M Oxidizing Pretreatment Solution, 230 mL bottle
8
21
or
TROUBLESHOOTING GUIDE
SMOTPMYS MELBORP NOITULOS
evissecxe/esnoperwolS
.tfird
.edortceleHpytriD IHnipitedortceleehtkaoS
03rofnoitulos1607
wollofnehtdnasetunim .erudecorPgninaelCeht
pusetautculfgnidaeR
.)esion(nwoddna
.noitcnujytrid/deggolC
leveletylortcelewoL
sedortceleelballifer(
.)ylno
.edortceleehtnaelC
etylortcelehserfhtiwllifeR
.)ylnosedortceleelballifer(
swohsyalpsiD
elacsllufgniknilb
.eulav
fotuognidaeR
.egnar
sielpmastahtkcehC
;egnarelbarusaemnihtiw
dnaleveletylortcelekcehC
.sutatsedortcelelareneg
fotuoelacsVm
.egnar
yrdroenarbmemyrD
.noitcnuj
00307IHniedortcelekaoS tsaeltarofnoitulosegarots
.setunim03
swohsyalpsiD
.F°roC°gniknilb
redrofotuO
.eborperutarepmet
erutarepmetecalpeR
.eborp
swohsyalpsiD
.gniknilbNAELC
neewtebecnereffiD
suoiverpdnawen
neebsahnoitarbilac
.detceted
dnaedortcelenaelC
melborpehtfI.etarbilacer reffubehtkcehc,sniamer
.snoitulos
swohsyalpsiD
yrettabgniknilb
.lobmys
.levelyrettabwoL .seirettabegnahC
krowtonseodreteM
erutarepmethtiw
.eborp
erutarepmetnekorB
.eborp
erutarepmetecalpeR
.eborp
etarbilacotsliafreteM
ytluafsevigro
.sgnidaer
.edortceleHpnekorB .edortceleecalpeR
siLACGNORW
Hpgniruddeyalpsid
.erudecorpnoitarbilac
rognorW
.reffubdetanimatnoc
noitulosreffubtahtkcehC
.hserfdnatcerrocsi
.ffostuhsreteM ;seirettabdaeD
sierutaefffo-otuA
,esacsihtni:delbane
retfaffostuhsretem
.esu-nonfonim02
;seirettabegnahC
.FFO/NOsserP
2rE,1rE,0rE
.putratstaegassem
.rorreMORPEE ynarorelaedruoytcatnoC
.retneCecivreSannaH
tratstaegassemrlC
.pu
HptluafeddedaoL
.seulavnoitarbilac
.noitarbilacHpmrofreP
pH CALIBRATION
It is recommended to calibrate the instrument frequently, especially if high accuracy is required.
The pH range should be recalibrated: a) Whenever the pH electrode is replaced. b) At least once a week. c) After testing aggressive chemicals. d) When extreme accuracy is required. e) When the calibration time-out is expired (if feature is enabled).
PROCEDURE
The meter offers a choice of 7 memorized buffers (1.68, 4.01, 6.86,
7.01, 9.18, 10.01 and 12.45 pH) and also allows the user to enter two more pH values for calibration, C1 and C2 (see Custom Buffer Selection section on page 12 for details).
 Pour small quantities of selected buffer solutions into clean beakers.
20
For accurate calibration use two beakers for each buffer solution, the first one for rinsing the electrode and the second one for calibration.
 Remove the protective cap and rinse the electrode with some of
the buffer solution to be used for the first calibration point.
TWO-POINT CALIBRATION
 Press the CAL key. The "CAL" and " " indicators will be
displayed. The secondary LCD will display buffer "7.01". If a different calibration buffer is desired (e.g. 6.86), use the UP and DOWN arrow keys to change the displayed value.
 Submerge the electrode approximately 4 cm (1½") into the
solution, place the temperature probe as close as possible to the electrode and stir gently.
 The LCD will flash "NOT READY" for 12
seconds, then: if the reading is not close to the selected buffer, "WRONG " and "WRONG " will blink alternatively; if it is close to the selected buffer the meter will advise the user with an acoustic signal (if enabled) when the reading becomes stable and the display will change to "READY" and blinking "CFM".
9
 Press the CFM key to confirm the calibra-
tion: the meter stores the first calibration point; the primary LCD will show the calibrated reading, while the secondary LCD will show the second buffer to be used for calibration (pH 4.01).
If youre going to calibrate with a different buffer, select the desired value by pressing the arrow keys.
 Submerge the electrode approximately 4 cm (1½") into the
second buffer solution, place the temperature probe as close as possible to the electrode and stir gently.
 The LCD will flash "NOT READY" for 12 seconds, then:
if the reading is not close to the selected buffer, "WRONG " and "WRONG " will blink alternatively; if it is close to the selected buffer the meter will advise the user with an acoustic signal (if enabled) when the reading becomes stable and the display will change to "READY" and blinking "CFM".
 Press the CFM key: the value is stored in memory and the meter
returns to normal mode. The tags corresponding to the buffers used for calibration will light up together with the condition bargraph (if enabled).
For refillable electrodes, if the refill solution (electrolyte) is more than 2½ cm (1") below the fill hole, add the appropriate Electrolyte Solution.
MEASUREMENT
Rinse the electrode tip with distilled water, immerse it (4 cm / 1½") in the sample and stir gently for a few seconds. For a faster response and to avoid cross contamination of the samples, rinse the electrode tip with the solution to be tested, before taking any measurements.
STORAGE PROCEDURE
To minimize clogging and ensure a quick response time, the glass bulb and the junction should always be kept moist. When not in use, store it with a few drops of HI 70300 storage solution in the protective cap.
NEVER STORE THE ELECTRODE IN DISTILLED OR DEIONIZED WATER.
PERIODIC MAINTENANCE
Inspect electrode and cable. The cable used for the connection to the meter must be intact and there must be no points of broken insulation on the cable or cracks on the electrode stem or bulb. If any scratches or cracks are present, replace the electrode. Rinse off any salt deposits with water. Connectors must be perfectly clean and dry.
For refillable electrodes:
Note: The meter automatically skips the buffer used for the first calibration
point to avoid erroneous procedure. A difference of at least 1.5 pH unit is required between the two buffers used for the offset and slope calibration: once calibrated at either pH 7.01 or 6.86, the instrument automatically ignores the other value for the second point (same for pH 10.01 and 9.18).
Note: During calibration, the secondary LCD shows the selected buffer
value; press RANGE to display the buffer temperature.
Note: To clear a previous calibration and return to the default
values, press CLR at any time after entering the calibration mode. The LCD will show CLr CAL for one second, and then will return to normal measurement mode. The LCD will show an empty bargraph to warn the user that the meter is not calibrated.
10
Refill the electrode with fresh electrolyte (see the electrodes specifications to select the correct refilling solution). Allow the electrode to stand upright for 1 hour. Follow the Storage Procedure above.
CLEANING PROCEDURE
General Soak in Hanna HI 7061 General Cleaning
Solution for approximately ½ hour.
Protein Soak in Hanna HI 7073 Protein Cleaning
Solution for 15 min.
Inorganic Soak in Hanna HI 7074 Inorganic
CleaningSolution for 15 min.
Oil/grease Rinse with Hanna HI 7077 Oil & Fat Cleaning
Solution for 1 min.
IMPORTANT: After performing any of the cleaning procedures, rinse the electrode thoroughly with distilled water and soak it in HI 70300 Storage Solution for at least 1 hour before taking measurements.
19
ELECTRODE CONDITIONING & MAINTENANCE
Not present in gel electrodes.
PREPARATION PROCEDURE
Remove the electrode protective cap. DO NOT BE ALARMED IF ANY SALT DEPOSITS ARE PRESENT. This is normal with electrodes and they will disappear when rinsed with water. During transport tiny bubbles of air may have formed inside the glass bulb. The electrode cannot function properly under these conditions. These bubbles can be removed by "shaking down" the electrode as you would do with a glass thermometer. If the bulb and/or junction are dry, soak the electrode in HI 70300 Storage Solution for at least one hour.
ONE-POINT CALIBRATION
For optimum accuracy it is always recommended to perform a two­point calibration, but for a faster operation it is also possible to carry out a single-point procedure. Buffers at pH 7.01 or pH 6.86 (NIST) are normally the most appropriate for this purpose, even though the meter can be calibrated with any of the memorized calibration values. After calibrating the first point (see above), press the CAL key to end the calibration procedure.
Note: With one-point calibration there is no Condition and only the
frame is shown. Calibration time-out is active.
CUSTOM BUFFER SELECTION
 The meter also allows the user to calibrate with custom buffers.  To select a custom buffer press the CUSTOM BUF key while in calibration
mode. Use the arrow keys to set the desired value for C1 (the default
value is 7.00).
Note: To increase the speed, keep pressed the UP or DOWN key.  Once the desired value is reached, the meter will continue
calibration as with memorized buffers.  Pressing again CUSTOM BUF, the meter
will select the second custom buffer. Pressing
CUSTOM BUF once again, the meter will
return to the standard buffers mode.
EXPIRED CALIBRATION
The instrument is provided with a real time clock (RTC), in order to monitor the time elapsed since the last pH calibration. The real time clock is reset every time the meter is calibrated and the expired calibration status is triggered when the meter detects a calibration time-out. The CAL tag will start blinking to warn the user that the meter should be recalibrated. The calibration time-out can be set (see Setup menu section on page 14) from 0 (function disabled) to 14 days. For example, if a 4 days time-out has been selected the meter will issue the alarm exactly 4 days after the last calibration. However, if at any moment the expiration value is changed (e.g. to 7 days), then the alarm will be immediately recalculated and appear 7 days after the last calibration.
18
11
Notes:
 When the meter is not calibrated or calibration is cleared (default
values loaded) there is no expired calibration, and the display always shows a blinking CAL tag.
 When an abnormal condition in the RTC is detected the meter
forces the expired calibration status.
CONDITION
The display is provided with a 5-dot bargraph (unless the feature is disabled) which gives an indication of the electrode status after calibration as follows:
Bargraph indication Condition value
All 5 dots steady 81 to 100% of life
4 dots steady 61 to 80%
TEMPERATURE CORRELATION
FOR pH SENSITIVE GLASS
The resistance of glass electrodes partially depends on the temperature. The lower the temperature, the higher the resistance. It takes more time for the reading to stabilize if the resistance is higher. In addition, the response time will suffer to a greater degree at temperatures below 25 °C.
3 dots steady 41 to 60%
2 dots steady 21 to 40%
1 dot steady 1 to 20%
1 dot blinking 0%
Only frame is ON No info available
The condition bargraph remains active for 12 hours after calibration, then only the frame is shown.
Note: When an abnormal condition in the RTC is detected, the condition
is cleared and only the bargraph frame is shown on the display.
CLEAN ELECTRODE
Each time pH calibration is performed, the meter internally compares the new calibration with the one previously stored.
When this comparison indicates a significant difference, the CLEAN message blinks on the LCD to advise the user that the pH electrode may need to be cleaned (see Electrode Conditioning & Maintenance section).
After cleaning, perform calibration. Note: If the calibration data are cleared, the comparison is done
with the default values.
12
Since the resistance of the pH electrode is in the range of 50  200 Mohms, the current across the membrane is in the pico Ampere range. Large currents can disturb the calibration of the electrode for many hours.
For these reasons high humidity environments, short circuits and static discharges can be detrimental to a stable pH reading.
The pH electrodes life also depends on the temperature. If constantly used at high temperatures, the electrode life is drastically reduced.
Typical Electrode Life
Ambient Temperature 1  3 years 90 °C Less than 4 months 120 °C Less than 1 month
Alkaline Error
High concentrations of sodium ions interfere with readings in alkaline solutions. The pH at which the interference starts to be significant depends upon the composition of the glass. This interference is called alkaline error and causes the pH to be underestimated. Hannas glass formulations have the indicated characteristics.
Sodium Ion Correction for Glass at 20-25 °C Concentration pH Error
0.1 Mol L-1 Na
1.0 Mol L-1 Na
+
+
13.00
13.50
14.00
12.50
13.00
13.50
14.00
0.10
0.14
0.20
0.10
0.18
0.29
0.40
17
PMET SREFFUBHp
86.1 10.4 68.6 10.7 81.9 10.01 54.21
0 23 76.1 10.4 89.6 31.7 64.9 23.01 83.31
5 14 76.1 00.4 59.6 01.7 93.9 42.01 81.31
01 05 76.1 00.4 29.6 70.7 33.9 81.01 99.21
51 95 76.1 00.4 09.6 50.7 72.9 21.01 08.21
02 86 86.1 00.4 88.6 30.7 22.9 60.01 26.21
52 77 86.1 10.4 68.6 10.7 81.9 10.01 54.21
03 68 86.1 20.4 58.6 00.7 41.9 69.9 92.21
53 59 96.1 30.4 48.6 99.6 11.9 29.9 31.21
04 401 96.1 40.4 48.6 89.6 70.9 88.9 89.11
54 311 07.1 50.4 38.6 89.6 40.9 58.9 38.11
05 221 17.1 60.4 38.6 89.6 10.9 28.9 07.11
55 131 27.1 80.4 48.6 89.6 99.8 97.9 75.11
06 041 27.1 90.4 48.6 89.6 79.8 77.9 44.11
56 941 37.1 11.4 48.6 99.6 59.8 67.9 23.11
07 851 47.1 21.4 58.6 99.6 39.8 57.9 12.11
57 761 67.1 41.4 68.6 00.7 19.8 47.9 01.11
08 671 77.1 61.4 78.6 10.7 98.8 47.9 00.11
58 581 87.1 71.4 78.6 20.7 78.8 47.9 19.01
09 491 97.1 91.4 88.6 30.7 58.8 57.9 28.01
59 302 18.1 02.4 98.6 40.7 38.8 67.9 37.01
LCD MESSAGE GUIDE
temperature compensation mode
pH or mV measurement
calibration buffers custom buffers active
temperature reading or pH buffer value
mode tags
instability indicator
battery symbol
calibration
messages
condition bargraph
arrow keys
active
first or second calibration point
calibration
messages
TAGS & SYMBOLS
Mode tags light up for indicating the corresponding active mode,
and blink for warning the user.
SETUP on: setup menu mode has been entered. CFM blinking: ask confirmation of calibration value. CAL on: calibration mode has been entered. CAL blinking: meter is not calibrated or calibration is expired. RCL, MEM on: when recalling stored values. MEM blinking: measurement stored in the internal memory and
frozen on the display.
 Indication of temperature compensation mode:
MTC for manual, ATC for automatic compensation.
Battery symbol blinking: low battery condition. Batteries should
be replaced soon.
Calibration messages.
NOT READY blinking: buffer has been recognized, but reading
is not stable.
READY on: buffer has been recognized and reading is stable. WRONG and WRONG blinking alternatively: wrong
buffer, value not recognized.
CLEAN blinking: an abnormal difference between new and
previous calibration has been detected. Electrode cleaning is suggested. Follow the cleaning procedure described in the Electrode conditioning & maintenance section. If the problem remains, check the buffer solutions.
pH BUFFER TEMPERATURE
DEPENDENCE
The temperature has an effect on pH. The calibration buffer solutions are affected by temperature changes to a lesser degree than normal solutions. During calibration the instrument will automatically calibrate to the pH value corresponding to the measured or set temperature.
16
During calibration the instrument will display the pH buffer value at 25 °C.
13
SETUP MENU
mV CALIBRATION
The instrument allows the user to configure several parameters through the Setup Menu.
To enter the Menu mode, while in normal measurement mode, press and hold the SETUP key for about 5 seconds.
Once the menu is entered, each parameter can be changed by using the arrow keys; then pressing the CFM key will confirm the value and scroll to the next parameter.
The sequence of programmable parameters is as follows:
1. Acoustic signal: On (default) or Off
2. Auto-off feature: 20 minutes (default) or disabled
3. Temperature measure unit: ºC (default) or ºF
4. Calibration segments: On (default) or Off (this enables or disables the bargraph, the CLEAN message and the buffer segments)
5. Calibration time-out: 1 (default) to 14 days or disabled (0 days)
After setting the last parameter, pressing the CFM key will confirm the value and return to normal measurement mode.
HI9026 has been accurately precalibrated for mV range at the factory.
For optimum accuracy, it is recommended to recalibrate the meter for mV readings at least once a year.
Contact your Dealer or the nearest Hanna Customer Service Center for more information.
TEMPERATURE CALIBRATION
HI9026 has been accurately precalibrated for temperature at the factory.
For optimum accuracy, it is recommended to recalibrate the meter for temperature at least once a year.
Contact your Dealer or the nearest Hanna Customer Service Center for more information.
BATTERY REPLACEMENT
If the batteries become weak, the display will flash the battery symbol to advise the user that approximately 25 hours of work­ing time is left. It is recommended to replace batteries soon.
The meter is also provided with the BEPS (Battery Error Prevention System) feature, which automatically turns the instrument off when the battery level is too low to ensure reliable readings. At start-up the display will show 0 bAtt for a few seconds, then the meter automatically turns off.
Battery replacement must only take place in a non hazardous area using the battery types specified in this instruction manual.
To replace rundown batteries, remove the rear cover of the instrument and substitute all four 1.5V AA alkaline batteries with new ones, while paying attention to the correct polarity. Reattach the cover and tighten the two screws.
14
15
Loading...