CE CERTIFICATE OF CONFORMITY ........................................................... 63
ISO 9000 Certified
Company since 1992
2
PRELIMINARY EXAMINATIONPRELIMINARY EXAMINATION
PRELIMINARY EXAMINATION
PRELIMINARY EXAMINATIONPRELIMINARY EXAMINATION
Remove the instrument from the packing material and examine it to
make sure that no damage has occurred during shipping. If there is
any damage, notify your Dealer.
Note: Save all packing material until you are sure that the instrument
functions correctly. All defective items must be returned in the
original packing with the supplied accessories.
GENERAL DESCRIPTIONGENERAL DESCRIPTION
GENERAL DESCRIPTION
GENERAL DESCRIPTIONGENERAL DESCRIPTION
HI9017, HI9218, HI9219 and HI9318 are microprocessor-based benchtop meters designed to meet the most stringent demands of the laboratory.
They feature:
•user friendly rubber keyboard
•automatic calibration with memorized buffers values
•automatic or manual temperature compensation
•storage of calibration data even after the meters are shut off
•error codes to guide the user in calibration and troubleshooting
All these models incorporate a standard RS232C output for communication with a computer.
The HI76405 electrode holder provided will hold your pH electrode
and temperature probe directly over the sample vessel which can be
placed on the enamel coated base of the holder. A magnetic stirrer
can also be placed on the enamel coated base if sample agitation is
required. The enamel coating is impervious to most chemicals and the
heavy steel base will not tip over easily.
These meters are perfect for use in Q.C. labs, R&D labs, environmental testing, industrial processing, chemical treatment, waste water
treatment, drinking water analysis, titration, hydroponics, general agriculture, food processing and anywhere accurate pH measurements
are required.
• HI9017 pH/mV/°C meter automatically recognizes three stan-
dard calibration buffer values (pH 4.01, 7.01, 10.01). The pH
and temperature values are shown separately on the LCDs. A wide
mV range and high resolution allow for ORP or ion-selective
measurements (the range automatically switches from ISE to ORP
when readings exceed ±399.9 mV).
HI9017 comes supplied with HI1131B glass-body combination
pH electrode (with 1m/3.3' cable), HI7669/2W temperature probe
3
(with 1m/3.3' cable), 12VDC adapter (HI710005 or HI710006),
pH 4.01 & pH 7.01 buffer solutions (20 mL each), HI7071
electrolyte solution (30 mL), HI76405 electrode holder and a
dust cover.
• HI9218 and HI9219 are programmable pH/°C meter with built-
in printer. They print pH values together with temperature, time
and date. These meter provide you with pH and temperature
readings simultaneously and separately on the LCDs and feature a
unique stability indicator. When the pH electrode stabilizes in the
sample, the "NO" on the LCD changes to "YES". This indicates that
the reading is correct and can be recorded. They automatically
recognize three standard buffers (4.01, 7.01 and 10.01 pH) for
calibration purposes. These meters are extremely user-friendly,
giving step-by-step instructions on the printer at every stage to
help the user.
HI9218 has been specifically designed for the food industry.
HI9218 comes supplied with FC100B plastic-body combination
pH electrode (with 1m/3.3' cable) while HI9219 comes with
HI1131B glass-body combination pH electrode (with 1m/3.3'
cable). HI9218 and HI9219 are also supplied with: HI7669/2W
temperature probe (with 1m/3.3' cable), 12VDC adapter (HI710005
or HI710006), pH 4.01 & pH 7.01 buffer solutions (20 mL each),
HI7071 electrolyte solution (30 mL), 5 replacement rolls of printer
paper, HI76405 electrode holder and a dust cover.
• HI9318 is a pH/mV/°C meter designed for logging data in various
applications. This programmable printing meter prints pH or mV
values as well as sample number, temperature, time and date.
The pH/mV and temperature values are displayed on separate
LCDs for ease of reading. It recognizes and memorizes three
different standard buffers (pH 4.01, 7.01 and 10.01) for calibration purposes. The printer can be programmed to print in one of 6
different languages (English, French, German, Italian, Spanish or
Swedish). Once your language is selected you can simply prompt
the meter to calibrate and it will print step-by-step instructions
through the entire process making this meter perfect for use by
even non-technical staff members.
HI9318 comes supplied with HI1131B glass-body combination
pH electrode (with 1m/3.3' cable), HI7669/2W temperature probe
(with 1m/3.3' cable), 12VDC adapter (HI710005 or HI710006),
pH 4.01 & pH 7.01 buffer solutions (20 mL each), HI7071
electrolyte solution (30 mL), 5 replacement rolls of printer paper,
HI76405 electrode holder and a dust cover.
4
FRONT PANEL ANDFRONT PANEL AND
FRONT PANEL AND
FRONT PANEL ANDFRONT PANEL AND
SPECIFICATIONS OF HSPECIFICATIONS OF H
SPECIFICATIONS OF H
SPECIFICATIONS OF HSPECIFICATIONS OF H
II
90179017
I
9017
II
90179017
1
to manually enter calibration buffervalue
PGM
1
to enter pH alarm
PGM
2
to enter mV alarm
PGM
3
to enter temperature alarm
PGM
4
to enter pH end point alarm
PGM
5
to enter RS 232 ESC code
PGM
6
to exit from any program
PGM
0
to clear entry or silence alarm
CLEAR
Standard pH buffer - 4.01, 7.01, 10.01
1. Program descriptions
Briefly describes the 6 programs
2. Temperature LCD
This LCD displays the temperature measured by the HI7669/2W
probe or the value that was manually set through the keyboard.
If the temperature has been set manually, the "°C" symbol will
flash.
When the temperature probe is connected the measured tempera-
ture is displayed and the "°C" symbol stops flashing.
When the temperature measured by the probe is out of range (-10
to 125°C) the "Err3" signal appears.
3. pH and mV LCD
This LCD displays the pH or mV reading (pH and mV is toggled
with the RANGE key). The corresponding pH or mV symbol will be
displayed along with the measured value.
An error signal ("Err1", "Err2", "Err4" or "Err5") will appear on this
display to indicate that the range is exceeded or that the buffer
used for calibration is not recognized by the meter's microprocessor.
4. Keyboard
ENTER A key used in programming.
ENTER B key used in programming.
CLEAR keyto clear entry or silence alarm.
PGM keyto select a program.
PROGRAMS
ENTER ENTER
1st buffer2nd buffer
ENTER ENTER
upper limit lower limit
ENTER ENTER
upper limit lower limit
ENTER ENTER
upper limit lower limit
ENTER ENTER
end point value alert value
ENTER
ESC code
A
BA
BA
BA
BA
BA
ENTERAENTERBCLEAR
7
4
1
0
pH°C
HI 9017 MICROPROCESSOR pH METER
ins rum n
ON
89
56
23
.
CAL
PGM
+/-
2
3
s
OFF
RANGE
CON
TEMP
4
5
ON keyto switch the instrument on.
OFF keyto switch the instrument off.
RANGE keyto switch range from pH to mV or vice versa.
CAL keyused in calibration with standard pH buffers (pH
4.01, 7.01, 10.01)
CON keyconfirmation of calibration data
+/- keyentering negative values
TEMP keyused for manual temperature adjustment
0 to 9 keysnumeric and Decimal keys for numeric data input
SPECIFICATIONSHI 9017
RangepH0.00 to 14.00
ISE± 399.9 mV
ORP±1999 mV
°C-10.0 to 125.0
ResolutionpH0.01
ISE0.1 mV
ORP1 mV
°C0.1
AccuracypH±0.01
(@ 20°C/68°F) ISE±0.2 mV
ORP±1 mV
°C±0.5
Typical EMC pH±0.02
DeviationISE±0.2 mV
ORP±1 mV
°C±0.5
Calibration2 point with automatic recognition of 3
buffers (pH 4.01, 7.01, 10.01)
Temperature-10° to +125°C (14 to 257 °F) completely
automatic withHI7669/2W temperature
probe; manual by means of keyboard
Comp. InterfaceRS232C
ElectrodeHI1131B Glass body combination
pH electrode (included)
Temperature ProbeHI7669/2W Temperature probe (included)
Input Impedance1012 Ohm
PowerPower socket for 12VDC
Environment0 to 50°C (32 to 122°F); RH 95%
Dimensions230 x 170 x 70mm (9.1 x 6.7 x 2.7")
Weight1.3 Kg (2.9 lb.); kit w/holder 3.2 Kg (7 lb.)
6
FRONT PANEL ANDFRONT PANEL AND
FRONT PANEL AND
FRONT PANEL ANDFRONT PANEL AND
SPECIFICATIONS OF HI9218 & HI9219SPECIFICATIONS OF HI9218 & HI9219
SPECIFICATIONS OF HI9218 & HI9219
SPECIFICATIONS OF HI9218 & HI9219SPECIFICATIONS OF HI9218 & HI9219
Foodcare
To change paper roll or ink ribbon
Slide cover to the left to open.
Push ribbon cassette as marked - to release
Insert new paper roll. Pass leading end under guide plate and
press PAPER to feed paper through the printer.
Pass leading end through ribbon cassette, with ribbon at the
front and press cassette into position
1
ins
rum n
e
PAPER
STABILITY INDICATOR
READY
NO
YES
PRINT
HI 9218
PRITING pH METER
WITH STABILITY INDICATOR
s
°C
PRINT
Press to check
reading stability.
Wait
Reading is stable.
Press to obtain
a copy.
pH
ON
DATE
CAL
UP
1. Printer
Plain paper printer
2. Temperature LCD
This LCD displays the temperature measured by the HI7669/2W
probe or the value that was manually set through the keyboard. If
the temperature has been set manually, the "°C" symbol will
flash. When the temperature probe is connected the measured
temperature is displayed and the "°C" symbol stops flashing.
When the temperature measured by the probe is out of range (-
10.0 to 125.0°C) the "Err3" signal appears.
3. pH LCD
This LCD displays the pH reading. The corresponding pH symbol
will be displayed along with the measured value.An error
signal ("Err1", "Err4" or "Err5") will appear on this display to
indicate that the range is exceeded or that the buffer used for
calibration is not recognized.
4. Small LCD
This LCD normally displays the time. When the DATE button is
pressed, the date will be displayed. When the READY key is
pressed, the display shows "NO" while the electrode is stabilizing
and "YES" once the reading is stable.
5. Keyboard
ON/OFF keyto switch the instrument "ON" or "OFF".
Note: Press OFF before unplugging instrument, otherwise, calibration
data and time setting could be lost.
2
3
4
OFF
5
TIME
CFM
DOWN
7
DATE/TIME key to display date and time.
CAL keyto enter calibration mode.
CFM keyto confirm memorized buffers 7.01 and 4.01/10.01.
UP/DOWN keyto set manual temperature compensation and date/time.
PAPER keymanual paper feed control.
PRINT keyprints pH value together with temperature, date
and time.
READY keystability indicator to ensure correct pH reading or
to enable calibration.
SPECIFICATIONSHI9218 HI9219
RangespH0.00 to 14.00
°C-10.0 to 125.0
Resolution pH0.01
°C0.1
AccuracypH±0.01
(@20°C/68°F) °C±0.5
Typical EMC pH±0.04
Deviation°C±1.5
Calibration2 point with automatic recognition of
3 buffers (pH 4.01, 7.01, 10.01)
Temperature-10° to 125°C (14 to 257°F)Completely
Compensationautomatic with HI 7669/2W temperature
probe; manual by means of keyboard
Comp. InterfaceRS232C
ElectrodeHI9218: FC100B plastic body combination
pH electrode (included)
HI9219:HI1131B Glass body combination
pH electrode (included)
TemperatureHI7669/2W Temperature probe (included)
Probe
Input Impedance1012 Ohm
Display2 large 4-digit LCD for pH and temperature
with symbols 1 small LCD for date/time and
stability indicator
Printer1 x 16 line dot matrix printer,
paper width: 44 mm,
roll diameter: 35 mm
Power SupplyPower socket for 12VDC
Environment0 to 50°C (32 to 122°F); RH 95%
Dimensions230x170x70mm (9.1x6.7x2.7")
Weight1.3 Kg (2.9 lb.);
kit w/holder 3.2 Kg (7 lb.)
8
FRONT PANEL ANDFRONT PANEL AND
FRONT PANEL AND
FRONT PANEL ANDFRONT PANEL AND
PROGRAMS
0 PGM to exit from any program
1
1 PGM to set sampling number, time and date
2 PGM to manually enter calibration buffervalue
3 PGM to print calibration data
4 PGM to print sample data until specified time
5 PGM to print sample data until end point
6 PGM to set pH/mV alarm
7 PGM to select different working language
8 PGM to print ion concentration
9 PGM to enter RS232C ESC code
CLEAR to clear entry
2
SPECIFICATIONS OF HSPECIFICATIONS OF H
SPECIFICATIONS OF H
SPECIFICATIONS OF HSPECIFICATIONS OF H
°C
PAPER
PRINTCLEAR
ABC
789
456
423
+/-.
rum n
e
s
ins
II
I
II
PGM
93189318
9318
93189318
pH
SAMPLE DATE
ON
DATERANGE
CAL
+/-
1. Program descriptions Briefly describes the 9 programs
2. PrinterPlain paper printer
3. Temperature LCDThis LCD displays the temperature. If
the temperature has been set manually, the "°C" symbol will
flash. When the temperature probe is connected the measured
temperature is displayed and the "°C" symbol stops flashing.
When the temperature measured by the probe is out of range (-10
to 125°C) the "Err3" signal appears.
4. pH and mV LCDThis LCD displays the pH or mV reading
(pH and mV is toggled with the RANGE key). The corresponding
pH or mV symbol will be displayed along with the measured
value. An error signal ("Err1", "Err2", "Err4" or "Err5") will appear
on this display to indicate that the range is exceeded or that the
buffer used for calibration is not recognized.
5. Sample LCD Displays the sample number of the current lot.
6. Time and Date LCDDisplays the Time as HH.MM. and the
date as MM.DD. This LCD also displays any numeric data entered
from the keyboard to show the user the entered values.
7. KeyboardPAPER key manual paper feed control. Press and hold this button
and the printer will advance the paper roll
PRINT key prints the time, date, sample #, and pH/mV values
CLEAR key cancels wrong data entered during programming
ON keyto switch the instrument on
OFF keyto switch the instrument off.
DATE keyallows to display the month and day or time in hours
and minutes
3
4
5
6
OFF
7
CON
TEMP.
9
RANGE key selects pH or mV readings to be displayed on the right LCD
CAL keyto enter into the calibration mode
CON keyfor confirmation of buffer value during calibration
+ / - keyselect + or - value for manual temperature compensation
TEMP keyallows to select temperature manually when the
temperature probe is not connected
PGM keyselects any of the 9 available programs
0 to 9 keys numeric and Decimal keys for numeric data input
SPECIFICATIONSHI9318
RangespH 0.00 to 14.00
ISE 0 to ±399.9 mV
ORP 0 to ±1999 mV
°C -10.0 to 125.0
Resolution pH 0.01
ISE 0.1mV
ORP 1 mV
°C 0.1
AccuracypH ±0.01
(@20°C/68°F) ISE±0.2 mV
ORP ±1 mV
°C ±0.5
Typical EMCpH ±0.04
DeviationISE ±2.5 mV
ORP ±3 mV
°C ±1.5
CalibrationAutomatic at pH 4, 7, 10 or at any two
buffer values with program 2
Offset±1pH
SlopeFrom 85 to 105%
Input Impedance1012 Ohm
Temperature-10° to 125°C (14 to 257°F) completely
Compensationautomatic with HI 7669/2W temp. probe;
manual by means of keyboard
Display2 large 4 digit LCDs for pH/mV and
Temperature with symbols and one
LCD for date/time and sample #
PrinterDot matrix, 44mm wide paper
Power SupplyPower socket for 12VDC
LanguagesEnglish, German, French,
Spanish, Italian & Swedish
ElectrodeHI1131B Glass body combination
pH electrode (included)
Temperature ProbeHI7669/2W Temperature probe (included)
Environment0 to 50°C (32 to 122°F); RH 95%
Dimensions230 x 170 x 70mm (9.1 x 6.7 x 2.7")
Weight1.3kg (2.9 lb.); Kit: 3.4 Kg (7.4lb.)
10
REAR PANEL CONNECTIONSREAR PANEL CONNECTIONS
REAR PANEL CONNECTIONS
REAR PANEL CONNECTIONSREAR PANEL CONNECTIONS
DTR
TXD
GND
COMBINATION
pH
REFERENCE
ELECTRODE
ELECTRODE
TEMPERATURE
PROBE
RESET
123 4 567
RXD
51
ooooo
oooo
96
DSR
RS-232
ooooo
oooo
1) BNC male connector for single or combination electrode. Will
accept any pH or ORP electrode with a standard BNC connector.
2) Jack bush connector for reference electrode
3) Banana bush connector for reference electrode
When the electrode attached to the BNC does not have a refer-
ence, connect the reference to one of these sockets (2 choices for
different diameters).
4) Temperature probe socket for use with the HI7669/2W.
5) Reset button.
6) RS 232C socket for serial communication.
7) DC Power Socket (for HI710005 or HI710006).
Electrode and probe connections
For combination pH (or ORP for HI9017 and HI9318) electrodes
(with internal reference) plug the electrode's BNC to the socket provided (#1).
For electrode with separate reference (see page 58), connect the
measuring electrode's BNC to the BNC socket (#1) and the reference
electrode's jack to the socket provided (#2 or #3).
For temperature measurements and automatic temperature compensation connect the temperature probe to the °C socket (#4).
RESET button
The RESET button (#5) is used when the instrument displays erroneous messages due to strong electrical interference or when the
instrument's power supply was disconnected before the meter was
switched off. Press the RESET button and restart the entire operation.
11
-
Power connection
Plug the 12VDC adapter (HI710005 or HI710006) into the DC socket
(#7).
Plug the adapter into the mains.
Note: The instrument uses the following configuration.
+
It is recommendable to use the Hanna HI7710005 or HI710006
voltage adapters (supplied witht the meters) that use the proper
polarity configuration.
The meters can also be used with other adapters. In this case, remember to check the correct polarity of your adapter before connecting it to
the meter.
Note: Make sure the main line is protected by a fuse.
Note: HI9017, HI9218, HI9219, HI9318 use a rechargeable battery
to retain the pH calibration data only. It lasts approximately one
month when fully charged. The battery is automatically charged
any time the meter is plugged-in (either turned on or off).
Notes:
1)If you need to unplug the instrument, press the OFF key before
disconnecting the meter from the mains. Only in this case will
the instrument retain the memorized data. Otherwise, the instrument needs to be re-calibrated. In the case of a power
failure while the instrument is operational , re-calibration is
required.
2)When the instrument is initially switched on, the microprocessor
checks if any calibration data is in memory (i.e. if the offset is
within ±1 pH and the slope between 70 - 105%). If the
calibration data are not memorized, the offset and slope of the
instrument will be set to 0 mV and 100% respectively.
12
OPERATIONAL GUIDEOPERATIONAL GUIDE
OPERATIONAL GUIDE
OPERATIONAL GUIDEOPERATIONAL GUIDE
TAKING PH MEASUREMENTS
Make sure that the instrument has been calibrated for pH before
taking pH measurements (see page 18).
Connect the pH electrode (combination-type) or the electrode half cells
to the appropriate connectors on the rear panel (see page 11).
If automatic temperature compensation is required, connect the
HI7669/2W temperature probe to the proper socket on the rear panel
(see page 11).
Press the ON key to turn the meter on. The meter will
automatically default to the pH measurement mode.
Immerse the electrode and the temperature probe
in the sample and shake briefly. Wait for 30
seconds before taking readings.
ON
ON
IN HI 9218 AND HI 9219
Press READY: "YES" confirms the reading is stable.
The LCD on the right shows the pH reading compensated for temperature. The temperature reading is displayed on the LCD on the left.
Note: the electrode should be submerged approxi-
mately 4cm (1½") into the solution.
If the temperature probe is used, it should be
located as close to the pH electrode as possible.
If measurements are taken in different samples successively, it is
recommended that the electrode is rinsed thoroughly for better conditioning and to eliminate cross-contamination of the sample.
13
READY
pH°C
(1½
")
4 cm
For the rinsing process, it is recommended to use a liberal amount of
the next solution to be measured.
Using refillable electrodes
Unscrew the small refill cap to open the refill hole during measurements. After measurements replace the refill hole cap.
This will increase the pressure and give a faster response.
FOR HI 9218, HI 9219, HI 9318 ONLY
PRINT
Press the PRINT key to get a printout at any time.
All the printouts will be complete with date, time,
sample No. (for HI9318 only), temperature and pH
(or mV with HI9318 only) value.
DATE: 01-02-96
TIME: 12:21 Hr
°C: 17.9
pH: 7.18
DATE 1/02
TIME 12:21
SAMPLE 1
°C 17.9
pH 7.18
HI9218/HI9219HI9318
Sample Printouts
MANUAL TEMPERATURE COMPENSATION
When the temperature probe is not connected, take the temperature of
the solution with a ChecktempC or a glass thermometer.
FOR HI 9017 ONLY
To manually set a temperature value (e.g. 37.2°C)
proceed as follows:
• press 3
• press 7
• press the key displaying decimal point "."
• press 2
• press TEMP
°C
32.7
TEMP
°C
14
With negative temperature values (e.g. -7.3°C) use the +/- key
• press 7
• press decimal point
• press 3
• press +/-
• press TEMP
7+/-3.
TEMP
ºC
Notes:
1)For negative value inputs, always key in the value followed by
the +/- key.
2)If the +/- key is pressed twice it will toggle back to the
°C
positive value.
The value will be displayed on the left LCD and "°C" symbol will be
flashing.
FOR HI 9218 AND HI 9219 ONLY
Use the UP or DOWN keys to manually adjust the display reading to
the value of the ChecktempC or the glass thermometer (e.g. 20°C).
The value will be displayed on the left LCD and "°C" symbol will be
flashing.
UP
DOWN
ºC
FOR HI 9318 ONLY
Press the +/- key (for negative figures) followed by the noted
temperature on the numeric keypad (e.g. 20°C). The value will be
displayed on the small LCD. Then press the TEMP. key and the
manually set temperature will be displayed on the left LCD with the
"°C" symbol blinking.
20
SAMPLE
Note: with the temperature probe connected, the "°C" symbol is
displayed continuously and it is not possible to manually set
the temperature.
°C
DATE
TEMP.
ºC
°C
15
TAKING TEMPERATURE MEASUREMENTS
Taking a temperature measurement is very easy.
Turn the instrument ON.
Make sure the temperature probe is connected and
allow the reading to stabilize (1 or 2 minutes).
The temperature value will be displayed on the left LCD.
pH°C
TAKING ORP MEASUREMENTS
(FOR HI 9017 AND HI 9318 ONLY)
HI9017 and HI9318 have the capability to take ORP measurements,
using an ORP electrode. Hanna Instruments offers a variety of ORP
electrodes for this purpose (see page 59). Contact your Dealer for more
information.
Just as pH measurements allow the acidity or the alkalinity of a
substance to be expressed in numbers, oxidationreduction potential (redox) measurements provide
the quantification of the oxidizing or reducing power
of any liquid.
Connect the ORP electrode to the meter and submerge the tip (4 cm / 1½") into the sample to be tested.
Note: ORP measurements are taken without temperature compensa-
tion.
Press the RANGE key once to enter the mV mode.
Allow a few minutes for the reading to stabilize.
The display will indicate the absolute mV value
(positive or negative).
When values are less than ±400 mV, tenths of
mV
mV are displayed; for values greater than ±400
mV, the decimal digit disappears.
Note that the change in scale (from 0.1 mV to 1 mV as resolution) is
automatic.
"Err2" appears if the value exceeds ±1999 mV
(out of range).
RANGE
(1½
")
4 cm
16
FOR HI 9318 ONLY
Press PRINT key to get a printout at any time.
All the printouts will be complete with date, time, sample No., tem-
perature and mV value.
DATE1/02
TIME 12:40
SAMPLE 5
°C 20.5
mV -350.0
AFTER MEASUREMENTS
Press the OFF key to switch the instrument OFF.
See page 50 for storage, cleaning and maintenance
procedures for electrodes.
OFF
OFF
17
PH CALIBRATIONPH CALIBRATION
PH CALIBRATION
PH CALIBRATIONPH CALIBRATION
IMPORTANT
The instrument's pH range should be re-calibrated:
••
• When the meter is new.
••
••
• Whenever the pH electrode is replaced.
••
••
• At least once a month.
••
••
• After testing aggressive chemicals.
••
••
• After cleaning procedure and changing the reference electrolyte.
••
••
• For greatest accuracy.
••
INITIAL PREPARATION
Pour small quantities (about 4cm / 1½" as level) of pH7.01 (HI7007
or HI8007) and pH4.01 (HI7004 or HI8004) or pH 10.01 (HI7010
or HI8010) solution into clean beakers. If possible use plastic beakers
to minimize any EMC interferences.
4
0
0
I 7
H
HI 7004HI 7007
If you are measuring in the acid range use pH4.01 as second buffer or
pH 10.01 if you are measuring in the alkaline range.
Note: with HI9017 and HI9318, different pH values can be selected
for calibration purposes. See page 29 and 39 respectively.
For accurate calibration, use two beakers for each buffer solution; the
first for rinsing the electrode, the second for calibration. In this way,
contamination of the buffer is minimized.
RINSE
HI 7007
CALIBRATION
HI 7007
PH CALIBRATION FOR HI 9017
Procedure
••
• Switch the instrument on after connecting the
••
pH electrode and the temperature probe.
18
ON
• Immerse the pH electrode and the ATC probe
into pH7.01 buffer solution. Shake briefly and
wait 1-2 minutes for thermal equilibrium.
HI 7007
(1½
4 cm
")
Note: the electrode should be submerged approximately 4 cm (1½")
into the solution. The temperature probe should be positioned
as close to the pH electrode as possible.
• Press the CAL key.
CAL
The temperature compensated value of the buffer
solution will appear on the left LCD and the pH
symbol will be blinking.
pHpH
E.g. if the buffer temperature is at 25°C, the display on the right will
show "pH7.01". If the buffer temperature is at 20°C, the display will
show "pH 7.03" (see page26).
pH
If a wrong buffer is used or the electrode is faulty, "Err4" will appear
on the left display.
pHpH
• Allow 1-2 minutes for the pH reading to stabilize.
Note: If you are not using the ATC probe, refer to page 25 to manually
correct for temperature.
• Press the CON key to accept the buffer value.
CON
"Err5" will appear on the left display and the
buffer value will be transferred to the right LCD.
The first point calibration is now complete.
pH
pH
Note: If a single point calibration is required, press the CAL key to
leave the calibration mode. However, it is always better to
proceed as follows for a 2-point calibration.
19
• Rinse the electrode and the temperature probe
before immersing them into pH 4.01 or pH10.01
calibration buffer, shake briefly and wait for
thermal equilibrium.
At this moment, "Err5" will disappear. The display will show the
temperature compensated value of the second buffer solution and
the pH symbol will be blinking.
pHpH
If a wrong buffer or a faulty electrode
has been used, "Err5" remains on the
left display.
pH
• Press the CON key to confirm the pH value.
The left display value is then transferred to
the right display. The left LCD will show the
temperature value to indicate that the meter
is now calibrated.
HI 7004
CON
(1½
4 cm
")
ºC
°C
pH
Note: The display on the left will indicate the temperature compen-
sated buffer value e.g. if the temperature is at 25°C, the
display will show "pH 4.01" or "pH 10.01", depending on the
solution used. If the temperature is at 20°C, the display will
show respectively "pH 4.00" or "pH 10.06" (see page 26).
Note: During calibration the pH symbol blinks on the left display, and
the right display always shows the measured value.
20
Loading...
+ 44 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.