Hanna Instruments HI 8614 User Manual

Instruction Manual
HI 8614 • HI 8614L
HI 8615 • HI 8615L
pH and ORP T ransmitters
HI8614L
pH INDICATOR
& TRANSMITTER
www.hannainst.com
1
Dear Customer, Thank you for choosing a Hanna Instruments
Product. Please read this instruction manual carefully
before using the instrument. This manual will provide you with all the neces-
sary information for the correct use of the instrument.
If you need additional technical information, do not hesitate to e-mail us at tech@hannainst.com.
These instruments are in compliance with the
directives.
TABLE OF CONTENTS
Preliminary Examination............................. 3
General Description .................................... 3
Specifications of HI8614 & HI8614L .......... 5
Specifications of HI8615 & HI8615L........... 6
Terminal Board Connections....................... 7
pH Calibration with Automatic Temperature
Compensation (HI8614 & HI8614L) .......... 9
pH Calibration with Manual Temperature
Compensation (HI8614 & HI 8614L) ....... 13
ORP Calibration (HI8615 & HI 8615L).... 14
Electrode Conditioning and Maintenance.. 19 Temperature-Resistance Correlation for
HANNA pH Sensitive Glass .................... 23
Installation Procedure and Examples ...... 25
Accessories .............................................. 28
Warranty .................................................. 34
CE Declaration of Conformity ................. 35
PRELIMINARY EXAMINATION
Remove the instrument from the packing ma­terial and examine it carefully to make sure that no damage has occurred during ship­ping. If there is any noticeable damage, notify your Dealer.
Note: Save all packing material until you are
sure that the instrument functions cor­rectly. All defective items must be re­turned in the original packing material together with the supplied accessories.
GENERAL DESCRIPTION
HI 8614 / HI 8614L (pH) and HI 8615 / HI 8615L
(ORP) are 2-wire water-resistant transmitters specially designed for long distance mea­surement of pH or ORP in industrial applica­tions.
Two versions are available: the standard
HI8614 and HI8615, and the HI8614L and HI8615L models with LCD.
The LCD allows easy verification and moni­toring of measured values, and provides easier calibration and maintainance.
The pH or ORP signal is transmitted in a 2­wire current loop in the range of 4 to 20 mA.
HI8614 and HI8614L can be connected to the Hanna process instruments HI8510T, HI8710T or HI 8711T, or to any recorder, computer or data monitoring device that ac­cepts 4 to 20 mA input.
HI8615 and HI8615L can be connected to the Hanna meters HI8512T, HI8720T or to any recorder, computer or data monitor that accepts 4 to 20 mA input.
32
The transmitters use a universal BNC socket for quick and secure connection to any elec­trode with a BNC connector.
For HI8614 and HI8614L the temperature compensation is performed by the transmitter's ATC circuitry if the temperature probe (HI76608, optional) is attached. If ATC is not required, it is possible to substitute the tem­perature probe with a fixed resistor.
The input is isolated from the current loop to eliminate problems related to ground loop, low insulation cables, multiple electrode con­nections, and a common mode voltage of up to 100 V ensures true differential readings.
The terminal board of the transmitter provides for connection of power supply, pH or ORP electrodes and temperature probe (for HI8614 and HI8614L models only).
The unit is enclosed in a protective casing conforming to IP 65 standards.
SPECIFICATIONS OF HI8614 & HI8614L
Range 4 to 20 mA / 0.00 to 14.00 pH Resolution 0.01 mA / 0.01 pH Accuracy (@20°C/68°F)
±0.02 mA / ±0.02 pH
Typical EMC Deviation
±0.25 mA / ±0.2 pH
Calibration Offset: ±2.2 mA / ±2 pH
Slope: ±0.5 mA / 86 to 116%
Temperature Compensation
Fixed or automatic from 0 to 100°C
(32 to 212°F) with HI76608 probe
Input Impedance 1012 Ohm Output 4-20 mA (isolated) Installation Category I I Protection IP 65 Environment 0 to 50°C (32 to 122°F);
RH max 95% non-condensing
Power Supply
HI8614: 18 to 30 Vdc
HI8614L: 20 to 36 Vdc Load Max 500 Ohm Dimensions 165 x 110 x 90 mm
(6.5 x 4.3 x 3.5'')
Weight 1 kg (2.2 lb.)
HI8614
pH TRANSMITTER
pH
HI8614L
pH INDICATOR
& TRANSMITTER
54
SPECIFICATIONS OF HI8615 & HI8615L
TERMINAL BOARD CONNECTIONS
Range 4 to 20 mA / ±1000 mV Resolution 0.01 mA / 1 mV Accuracy (@20°C/68°F)
±0.02 mA / ±5 mV
Typical EMC Deviation
±0.25 mA / ±15 mV
Calibration Offset: ±0.8 mA / ±100 mV
Slope: ±0.8 mA / 90 to 110%
Input Impedance 1012 Ohm Output 4-20 mA (isolated) Installation Category I I Protection IP 65 Environment 0 to 50°C (32 to 122°F);
RH max 95% non-condensing
Power Supply
HI8615: 18 to 30 Vdc
HI8615L: 20 to 36 Vdc Load Max 500 Ohm Dimensions 165 x 110 x 90 mm
(6.5 x 4.3 x 3.5'')
Weight 1 kg (2.2 lb.)
Unscrew the 4 screws and remove the top cover. There are three cable glands on the transmitter
cover. The large cable gland with the split in the rubber is for the electrode.
ELECTRODE
CABLE GLAND
WIRE CABLE GLANDS
Connect the positive supply to the strip terminal "+VE LOOP" and the negative supply to the terminal "-VE LOOP" of the transmitter terminal block.
HI 8615 ORP TRANSMITTER
mV
HI 8615 ORP TRANSMITTER
76
The wire between the transmitter and the recorder/indicator/controller should be a PVC insulated two wire with a wire diameter of at least 0.7 mm. This wire is fed through one of the smaller cable glands. The maximum dis­tance between the power supply and the amplifier is 300 m (1000'). It is not necessary to use shielded cable. The transmitter is protected against inversion of supply voltage.
Electrode connection: connect the BNC of the cable to the BNC socket on the transmitter.
pH CALIBRATION WITH AUTOMATIC
TEMPERATURE COMPENSATION
(HI 8614 & HI 8614L)
Initial preparation
Connect the pH electrode to the BNC socket.
Connect the temperature probe to the transmitter.
Pour small quantities of pH 7.01 and pH 4.01 solution into two clean beakers.
For HI8614 & HI8614L only: for automatic temperature com­pensation, connect the 2 termi-
nals of the temperature probe (HI76608, optional) to "TEMP. PROBE" terminals. If automatic temperature compensation is not required, short the "TEMP. PROBE" terminals with a resis­tance according to the external temperature:
Temperature (°C) Resistance (Ohms)
0 1634 10 1774 20 1922 30 2078 40 2242 50 2412 60 2590
A 2 kohm resistor is factory mounted for 25°C temperature compensation.
HI 7004
HI 7004 HI 7007
For accurate calibration use two beakers for each buffer solution, the first one for rinsing the electrode, the second one for calibration.
RINSE
HI 7007
CALIBRATION
HI 7007
To get accurate readings, use pH7.01 and pH 4.01 if you are going to measure acid samples or pH 7.01 and pH 10.01 for alkaline measurements.
Note: with HI8614L the instruments display
can be used during calibration without the need to connect the ammeter and the reading is directly expressed in pH units.
98
Procedure
Disconnect the +ve supply cable from "+VE LOOP" terminal and connect a 20mA f.s. ammeter between the +ve cable and "+VE LOOP" terminal.
Remove the protective cap from the electrode, rinse it with some pH 7.01 solution or immerse it in the pH 7 rinse solution, then immerse the pH electrode and tem-
HI 7007
perature probe into pH 7.01 calibration buffer solution; shake briefly and wait for the reading to stabilize.
For other buffer temperatures, refer to page 13 for the appropriate mA / pH reading.
Rinse the electrode with tap water or distilled water and a small amount of pH 4.01 or 10.01 buffer solution or immerse it in the pH4 rinse solution (2nd calibration point). Dip the electrode and
HI 7004
the temperature probe into pH 4.01 (or 10.01) calibra­tion buffer solution, shake briefly and wait for a few minutes for reading to stabi­lize.
Adjust the slope trimmer until the amme­ter reads 8.58 mA or the display shows "4.01" at 25°C. For other buffer tempera­tures, refer to page 13 for the appropriate mA / pH reading (HI 8614L).
Note: the tip of the elec-
trode should be sub­merged approximately 4 cm (1½") into the
4 cm
1½"
solution. The tempera­ture probe should be located as close to the pH electrode as possible.
Adjust the offset trimmer until the amme­ter reads 12mA or the display shows "7.01" (HI 8614L only) if the temperature of the buffer is at 25°C.
SLOPE ADJUSTMENT OFFSET ADJUSTMENT
mA
SLOPE ADJUSTMENT OFFSET ADJUSTMENT
mA
If you are using pH 10.01 buffer solution adjust the slope trimmer until the am-
mA
meter read 15.43mA at 25°C (77°F) or the value in­dicated at page 13.
FOR HI8614L ONLY
The Display Module is factory calibrated, so that the LCD display results are referred to the 4-20 mA loop current (e.g. LCD displays
0.00 pH when loop current is 4.00 mA and displays 14.00 pH when current is 20.00mA).
1110
Under normal application, adjustment on this module may not be necessary.
If routine check is required, the following pro­cedures shall be performed.
Calibrate the transmitter module using a ammeter, as explained in the previous pages.
Simulate a 4.00 mA loop current for the transmitter (i.e. pH 0.00 as Electrode in­put) and check for display reading.
Simulate a 20.00 mA loop current for the transmitter (i.e. pH 14.00 as Electrode input) and check for display reading.
Note: when the instrument is used in con-
junction with the Hanna indicator
HI 8510T or controllers HI 8710T, HI 8711T, the calibration can also be
performed on the indicator/controller. In this case slight calibration adjust­ment can be made on the indicator/ controller even if the whole system cali­bration is advised, always starting from the transmitter.
pH CALIBRATION WITH FIXED
TEMPERATURE COMPENSATION
(HI8614 & HI 8614L only)
Take the temperature of the
°C
buffer solutions using a Checktemp or a thermom­eter with a resolution of at least 1°.
Connect the appropriate re­sistor to the "TEMP. PROBE" terminals (see page 6) depending on the temperature of the calibra­tion solution.
Follow the procedure as outlined in the calibration with automatic temperature com­pensation and use the thermometer read­ing to adjust the trimmer until the amme­ter reads the value according to the follow­ing table.
Temperature pH 4.01 Buffer pH 7.01 Buffer pH 10.01 Buffer
°C °F pH m A pH m A pH mA
0
32
4.01
8.576
7.13
12.137
10.32
10 15 20 25 30 35 40 45 50 55 60
5
41
4.00
8.565
7.10 50 59 68 77 86 95
104 113 122 131 140
4.00
4.00
4.00
4.01
4.02
4.03
4.04
4.05
4.06
4.07
4.09
8.565
8.565
8.565
8.576
8.587
8.599
8.610
8.622
8.633
8.645
8.667
7.07
7.04
7.03
7.01
7.00
6.99
6.98
6.98
6.98
6.98
6.98
12.103
12.069
12.034
12.023
12.000
11.989
11.977
11.966
11.966
11.966
11.966
11.966
10.24
10.18
10.12
10.06
10.01
9.96
9.92
9.88
9.85
9.82
9.79
9.77
15.789
15.697
15.629
15.560
15.491
15.434
15.377
15.331
15.285
15.251
15.217
15.182
15.159
1312
ORP CALIBRATION (HI8615 & HI8615L)
Initial preparation
Disconnect the +ve supply cable from the "+VE LOOP" terminal and connect a 20 mA f.s. ammeter between the +ve cable and the "+VE LOOP" terminal. With HI 8615L the instrument display can be used during cali­bration without the need to connect the am­meter. In this case the values are directly expressed in mV units.
Pour a small quantity of HI7020
HI 7020
ORP solution into a beaker.
Procedure
Connect the shorting BNC connector to the ORP trans­mitter.
Adjust the OFFSET AD­JUSTMENT trimmer on the module for a display of 12 mA on the ammeter or 0 mV on the instrument display (HI 8615L only). This sets the zero point for the transmit­ter.
If the reading lies outside this range, adjust the slope adjustment trimmer on the
SLOPE ADJUSTMENT OFFSET ADJUSTMENT
transmitter for a reading just within this range.
The unit is now calibrated. A complete calibration of the transmitter mod-
ule is advised periodically. This calibration procedure requires the HI 8427
or the HI 931001 pH and ORP simulator to simulate the ORP electrode.
HI 8427 or HI 931001 produce a known signal into the system so that the faults of the system can be isolated.
Connect the ORP transmitter to the simu­lator.
SLOPE ADJUSTMENT OFFSET ADJUSTMENT
mA
Connect the ORP electrode to the transmit­ter and immerse the tip of the electrode into a beaker of HI 7020 ORP calibration solution, and check that the ammeter read­ing lies between 13.6 and 14.2 mA or the instrument reading is between 200 and 275 mV at 25°C (HI8615L).
HI 7020
mA
Set the simulator to 0 mV and adjust the offset trimmer to read 12 mA on the am­meter or 0 mV on the HI 8615L display (HI 8615L only).
SLOPE ADJUSTMENT OFFSET ADJUSTMENT
mA
1514
Set the simulator to 350 mV and adjust the slope trimmer to read 14.8 mA on the ammeter or 350 mV on the HI 8615L display (HI 8615L only).
SLOPE ADJUSTMENT OFFSET ADJUSTMENT
mA
Connect the ORP electrode to the module and immerse the tip of the electrode into the beaker of HI 7020 ORP solution and check that the ammeter reading lies be­tween 13.6 and 14.2 mA or the instrument reading is between 200 and 275mV at 25°C (HI 8615L only).
HI 7020
mA
Only if the reading lies out­side this range, adjust the slope adjustment trimmer on the transmitter to reflect a
SLOPE ADJUSTMENT OFFSET ADJUSTMENT
reading within this range.
FOR HI8615L ONLY
The HI 8615L is factory calibrated, and the displayed values are referenced to the 4-20 mA loop current (e.g. LCD displays -1000 mV when loop current is 4.00 mA and dis­plays +1000 mV when current is 20.00mA). Under normal application, adjustment on this module may not be necessary. If routine check is required, the following procedures shall be performed.
Follow the above procedure for HI 8615 (see page 12).
Simulate a 12.00 mA loop current for the transmitter (i.e. 0 mV at Electrode input) and check display reading.
Simulate a 20.00 mA loop current for the transmitter (e.g. +1000 mV at Electrode input) and check display reading.
1716
Please note:
-1000 mV = 4 mA
1000 mV = 20 mA
350 mV = 14.8 mA 200 mV = 13.6 mA 275 mV = 14.2 mA
ELECTRODE CONDITIONING
AND MAINTENANCE
0 mV = 12 mA
Reference Filling Hole
Reference Filling Hole
1 mV = 12.008 mA
Note: when the meter is used in conjunction
with the Hanna indicator HI 8512T, or the controller HI 8720T, the calibration can also be performed on the indicator/ controller. In this case slight adjust­ment can be made on the indicator/ controller even if the whole system cali­bration is advised, always starting from the transmitter.
Sensitive Wire
Reference Wire
Reference Junction
Glass Bulb
Reference Wire
Reference Junction
Platinum or Gold tip
Plastic Body
pH Electrode
Reference
Plastic Body
ORP Electrode
Glass Body
pH Electrode
Wire
Glass Body
ORP Electrode
Reference
Wire
Sensitive
Wire
Reference
Junction
Glass
Bulb
Reference
Junction
Platinum or
Gold tip
PREPARATION
Remove the protective cap. DO NOT BE ALARMED IF ANY SALT DE-
POSITS ARE PRESENT. This is normal with electrodes and they will
disappear when rinsed with water. During transport tiny bubbles of air may have
formed inside the glass bulb. The electrode cannot function properly under these condi­tions. These bubbles can be removed by
1918
"shaking down" the electrode as you would do with a glass thermometer.
If the bulb and/or junction are dry, soak the electrode in HI70300 storage solution for at least one hour.
For refillable electrodes
If the fill solution (electrolyte) is more than 1 cm (½") below the fill hole, add HI7082
3.5M KCl electrolyte solution for double junc­tion or HI7071 3.5M KCl+AgCl electrolyte solution for single junction electrodes. For a faster response unscrew the fill hole screw during measurements.
For AmpHel® electrodes
If the electrode does not respond to pH changes, the battery is run down and the electrode should be replaced.
MEASUREMENT
Rinse the electrode tip with distilled water. Immerse the tip (4 cm / 1½") in the sample
and stir gently for approx. 30 seconds. For a faster response and to avoid cross
contamination of the samples, rinse the elec­trode tip with a few drops of the solution to be tested, before taking measurements.
STORAGE
To minimize clogging and assuring a quick response time, the glass bulb and the junc­tion should be kept moist and not allowed to dry out. Replace the solution in the protective cap with a few drops of HI70300 storage solution or, in its absence, filling solution (HI7071 for single junction or HI7082 for double junction electrodes). Follow the Prepa­ration Procedure above before taking mea­surements.
Note: NEVER STORE THE ELECTRODE IN
DISTILLED WATER OR DRY.
PERIODIC MAINTENANCE
Inspect the electrode and the cable. The cable used for connection to the meter must be intact and there must be no points of broken insulation on the cable or cracks on the elec­trode stem or bulb. Connectors must be per­fectly clean and dry. If any scratches or cracks are present on the electrode body, replace the electrode. Rinse off any salt deposits with water.
For refillable electrodes
Refill it with fresh electrolyte (HI7071 for single junction or HI7082 for double junction elec­trodes). Allow the electrode to stand upright for 1 hour.
Follow the Storage Procedure above.
CLEANING PROCEDURE
General Soak in Hanna HI7061 general
cleaning solution for approxi­mately 1 hour.
Removal of films, dirt or deposits on the mem­brane/junction:
Protein Soak in Hanna HI7073 protein
cleaning solution for 15 minutes.
Inorganic Soak in Hanna HI7074 inorganic
cleaning solution for 15 minutes.
Oil/grease Rinse with Hanna HI7077 Oil &
Fat cleaning solution.
IMPORTANT: After performing any of the cleaning procedures rinse the electrode thor­oughly with distilled water, drain and refill the reference chamber with fresh electrolyte, (not necessary for GEL filled electrodes) and soak the electrode in HI70300 storage solution for at least 1 hour before taking measurements.
2120
TROUBLESHOOTING
-20 -20
-
0 +10+20 +30+40+50+60+70+80+90
°C
-
-
-
-
-
-
Evaluate your electrode performance based on the following.
Noise (Readings fluctuate up and down) could be due to:
- Clogged/Dirty Junction: refer to the
Cleaning Procedure above.
- Loss of shielding due to low electrolyte
level (in refillable electrodes only): HI7071 for single junction or HI7082 for double junction electrodes.
• Dry Membrane/Junction: soak in HI70300 storage solution for at least one
hour.
Drifting: soak the electrode tip in warm Hanna solution HI7082 for one hour and rinse the tip with distilled water. Refill with fresh HI7071 for single junction electrodes and HI7082 for double junction electrodes.
Low Slope: refer to the Cleaning Proce­dure above.
No Slope: check the electrode for cracks in glass stem or bulb and replace the electrode.
Slow Response/Excessive Drift: soak the tip in Hanna HI7061 cleaning solution for 30 minutes, rinse thoroughly in distilled water and then follow the Cleaning Proce­dure above.
TEMPERATURE-RESISTANCE
CORRELATION FOR HANNA
pH SENSITIVE GLASS
The resistance of glass electrodes partially depends on the temperature. The lower the temperature, the higher the resistance. It takes longer time for the reading to stabilize if the resistance is higher. In addition, the response time will suffer to a greater degree at tem­peratures below 10°C.
Ω
9
2x10
10
9
1x10
10
8
10
2x10
8
1x10
10
7
2x10
10
10
7
1x10
10
Since the resistance of the pH electrode is in the range of 200 Mohm, the current across the membrane is in the pico Ampere range. Large currents can disturb the calibration of the electrode for many hours.
For these reasons high humidity environ-
ments, short circuits and static discharges
are detrimental for a stable pH reading.
The pH electrode's life also depends on the temperature. If constantly used at high tem­peratures, the electrode life is drastically re­duced.
2322
Typical Electrode Life Ambient Temperature 1- 3 years 90 °C Less than 4 months 120°C Less than 1 month
High concentrations of sodium ions interfere with readings in alkaline solutions; the pH at which the interference starts to be significant depends upon the composition of the glass. This interference is the alkaline error and causes the pH to be underestimated. Hanna's glass formulations have the indicated charac­teristics.
Alkaline Error
Sodium Ion Correction for the Glass
at 20-25°C
Concentration pH Error
0.1 Mol L-1 Na
1.0 Mol L-1 Na
+
13.00
13.50
14.00
+
12.50
13.00
13.50
14.00
0.10
0.14
0.20
0.10
0.18
0.29
0.40
INSTALLATION PROCEDURE
AND EXAMPLES
The HI 8614, HI8614L, HI8615 and HI8615L transmitters may be wall mounted in any con­venient location near the measurement point. To minimize thermal drift due to extreme tem­perature fluctuations during the measurement process, particularly if the measurement is conducted outdoors, it is best to protect the transmitter in an enclosed casing.
General Installation Procedure
For most industrial application involving long term monitoring and control, it is also recom­mended to use tank electrode holders (HI6050 or HI6051) to protect the pH electrode and the temperature probe from contamination by the test solution.
Controlling the pH / ORP with a recorder
2524
Monitoring the pH/ORP with Panel Mounting
pH (HI8510)/ORP (HI8512) Indicator
Controlling the pH/ORP with
an Industrial Regulator
Monitoring and Controlling the pH/ORP
with Panel Mounting Indicator/Regulator
and Dosage Control of either Acid or Base
Monitoring and Controlling the pH with
(HI 8711) Panel Mounting Indicator/
Regulator with Independent Dosage
Control for Acid and Base
2726
PG13.5 THREAD
m
110mm30mm
DIA 9.5mm
3
110mm
D
M13 x 1.5
DIA
16
mm
7
mm
25
mm
PG13.5 THREAD
m
110mm30mm
D
3/4 x 16 UNF
M13 x 1.5
DIA
16
mm
7
mm
25
mm
150mm
DIA 16.5mm
M13 x 1.5
A 6 m
mm
mm
ACCESSORIES
pH CALIBRATION SOLUTIONS
HI7004M pH 4.01 buffer solution, 230 mL HI7004L pH 4.01 buffer solution, 500 mL HI7006M pH 6.86 buffer solution, 230 mL HI7006L pH 6.86 buffer solution, 500 mL HI7007M pH 7.01 buffer solution, 230 mL HI7007L pH 7.01 buffer solution, 500 mL HI7009M pH 9.18 buffer solution, 230 mL HI7009L pH 9.18 buffer solution, 500 mL HI7010M pH 10.01 buffer solution, 230mL HI7010L pH 10.01 buffer solution, 500 mL
ORP SOLUTIONS
HI 7020M 200-275 mV test solution, 230 mL HI 7020L 200-275 mV test solution, 500 mL HI 7091M Reducing pretreatment solution, 230 mL HI 7091L Reducing pretreatment solution, 500 mL HI 7092M Oxidizing pretreatment solution, 230 mL HI 7092L Oxidizing pretreatment solution, 500 mL
ELECTRODE MAINTENANCE SOLUTIONS
HI70300M Storage solution, 230 mL HI70300L Storage solution, 500 mL HI7061M General cleaning solution, 230 mL HI7061L General cleaning solution, 500 mL HI7073M Protein cleaning solution, 230 mL HI7073L Protein cleaning solution, 500 mL HI7074M Inorganic cleaning solution, 230 mL HI7074L Inorganic cleaning solution,500 mL HI7077M Oil & Fat cleaning solution, 230 mL HI7077L Oil & Fat cleaning solution,500 mL HI7071 3.5M KCl + AgCl electrolyte solu-
tion, 4 x 5 0 mL bottle, for single junc­tion electrodes
HI7072 1M KNO3 electrolyte solution, 4 x
50 mL bottle
HI7082 3.5M KCl electrolyte solution, 4 x
50 mL bottle, for double junction electrodes
pH ELECTRODES
HI 1090T Screwcap PG13.5 connector,
double junction, glass body
φ 12mm φ 9.5m
HI 1110S Screw connector, single junction,
glass body
HI 1130B/3 BNC connector, 3 m (9.9') cable,
single junction, glass body
/4 x 16 UNF
IA 20.5mm
38.5mm
HI 1110S HI 1130B/3
DIA 12mm
HI 1110T Screwcap PG13.5 connector,
double junction, glass body
φ 12mm φ 9.5m
HI 1114S Screw connector, double junction
plastic body
HI 1134B/3 BNC connector, 3 m (9.9') cable,
double junction plastic body
DIA 12mm
IA 20.5mm
38.5mm
HI 1114S HI 1134B/3
110mm
HI 1115S Screw connector, single junction,
glass body
HI 1135B/3 BNC connector, 3 m (9.9') cable,
single junction, glass body
5mm DIA 7.6mm
DI
1
m
25
7
HI 1115S HI 1135B/3
25mm
2928
DIA 12mm
D
3/4 x 16 UNF
D
3/4 x 16 UNF
D
3/4 x 16 UNF
110mm
M13 x 1.5
DIA
16
mm
7
mm
25
mm
PG13.5 THREAD
m
110mm30mm
DIA 16.5mm
M13 x 1.5
A 6 m
mm
mm
HI 1210T Screwcap PG13.5 connector,
PG13.5 THREAD
110mm30mm
D
3/4 x 16 UNF
D
3/4 x 16 UNF
D
DIA 9.5mm
110mm
D
3/4 x 16 UNF
double junction, plastic body
φ 12mm
HI 2910B/5 BNC connector, 5 m (16.5') cable,
double junction, plastic body, built­in amplifier
DIA 12mm
IA 20.5mm
HI 1910B BNC connector, 1 m (3.3') cable,
double junction, plastic body, built­in amplifier
DIA 12mm
IA 20.5mm
38.5mm
110mm
HI 1911B BNC connector, 1 m (3.3') cable,
double junction, plastic body, built­in amplifier
DIA 12mm
IA 20.5mm
38.5mm
110mm
HI 1912B BNC connector, 1 m (3.3') cable,
double junction, plastic body, built­in amplifier
HI 1912B/5 BNC connector, 5 m (16.5') cable,
double junction, plastic body, built­in amplifier
3/4 x 16 UNF
IA 20.5mm
DIA 12mm
38.5mm
110mm
ORP ELECTRODES
HI 2930B/5 BNC connector, 5 m (16.5') cable,
Pt, plastic body, built-in amplifier
DIA 12mm
IA 20.5mm
38.5mm
110mm
HI 3110S Screw-type connector, Pt, glass
body
HI 3130B/3 BNC connector, 3 m (9.9') cable,
Pt, glass body
DIA 12mm
IA 20.5mm
38.5mm
HI 3110S HI 3130B/3
HI 3110T Screwcap PG13.5 connector, Pt,
glass body
φ 12mm φ 1m
38.5mm
HI 2114B/5 BNC connector, 5 m (16.5') cable,
double junction, plastic body
DIA 12mm
IA 20.5mm
38.5mm
110mm
HI 3115S Screw-type connector, side-arm,
Pt, glass body
HI 3135B/3 BNC connector, 3 m (9.9') cable,
side-arm, Pt, glass body
5mm DIA 7.6mm
DI
1
m
25
7
HI 3115S HI 3135B/3
25mm
3130
DIA 12mm
150mm
HI 3210T Screwcap PG13.5 connector, Pt,
PG13.5 THREAD
110mm30mm
D
3/4 x 16 UNF
M13 x 1.5
DIA
16
mm
7
mm
25
mm
3/4 x 16 UNF
110mm
D
3/4 x 16 UNF
110mm
M13 x 1.5
A 6 m
mm
mm
3/4 x 16 UNF
110mm
plastic body
φ 12mm
EXTENSION CABLES FOR SCREW-TYPE ELECTRODES ONLY (SCREW TO BNC CON­NECTOR)
HI 7855 SERIES CABLE CONNECTORS
CONNECTOR AND 3.0 mm (0.12") CABLE WITH BNC
CONNECT TO SCREW TYPE
ELECTRODES
CONNECT TO THE
BNC SOCKET
OF THE METER
HI 3410S Screw connector, Pt, plastic body HI 3430B/3 BNC connector, 3 m (9.9') cable,
Pt, plastic body
DIA 12mm
IA 20.5mm
38.5mm
HI 3410S HI 3430B/3
110mm
HI 3932B/5 BNC connector, 5 m (16.5') cable,
Pt, plastic body, built-in amplifier
DIA 12mm
DIA 20.5mm
38.5mm
HI 4110S Screw-type connector, Au, glass
body
HI 4130B/3 BNC connector, 3 m (9.9') cable,
Au, glass body
DIA 12mm
DI
IA 20.5mm
1
m
25
7
HI 4110S HI 4130B/3
38.5mm
HI 4932B/5 BNC connector, 5 m (16.5') cable,
Au, plastic body, built-in amplifier
HI7855/1 Extension cable 1 m (3.3') long HI7855/3 Extension cable 3 m (9.9') long HI7855/5 Extension cable 5 m (16.5') long HI7855/10 Extension cable 10 m (33') long HI7855/15 Extension cable 15 m (49.5') long
OTHER ACCESSORIES
BL PUMPS Dosing pumps with flow rate from
1.5 to 20 lph
HI 98501 Pocket-size, thermometer with
penetration probe and 0.1°C resolution (range -50.0 to
150.0°C)
HI6050 & HI 6051 Submersible electrode
holders
HI 6054 & HI 6057 Electrode holders for in-line
applications
HI 7871 & HI 7873 Level controllers HI8427 pH / ORP electrode simulator HI931001 pH / ORP electrode simulator
with LCD
DIA 12mm
DIA 20.5mm
38.5mm
3332
WARRANTY
All Hanna Instruments meters are warranted for two years against defects in workman-
ship and materials when used for their in­tended purpose and maintained according to instructions.
The probes and the electrodes are war­ranted for a period of six months.
Damages due to accident, misuse, tamper­ing or lack of prescribed maintenance are not covered. This warranty is limited to repair or replacement free of charge.
If service is required, contact the dealer from whom you purchased the instrument. If under warranty, report the model number, date of purchase, serial number and the nature of the failure. Obtain a Returned Goods Authoriza­tion from the Customer Service department first and then return the instrument with the Authorization # included along with shipment costs prepaid. If the repair is not covered by the warranty, you will be notified of the charge for repair or replacement. When shipping any instrument, make sure it is properly pack­aged for complete protection.
CE DECLARATION OF CONFORMITY
All rights are reserved. Reproduction in whole or in part is prohibited without the written consent of the copyright owner.
Hanna Instruments reserves the right to
modify the design, construction and appear-
ance of its products without advance notice.
Recommendations for Users
Before using these products, make sure that they are entirely suitable for the environment in which they are used.
Operation of these instruments in residential area could cause unacceptable interferences to radio and TV equipments, requiring the operator to take all necessary steps to correct interferences.
Any variation introduced by the user to the supplied equipment may degrade the instruments' EMC performance.
To avoid damages or burns, do not perform any measurement in microwave ovens.
3534
SALES AND TECHNICAL SERVICE CONTACTS
Australia:
Tel. (03) 9769.0666 • Fax (03) 9769.0699
China:
Tel. (10) 88570068 • Fax (10) 88570060
Egypt:
Tel. & Fax (02) 2758.683
Germany:
Tel. (07851) 9129-0 • Fax (07851) 9129-99
Greece:
Tel. (210) 823.5192 • Fax (210) 884.0210
Indonesia:
Tel. (21) 4584.2941 • Fax (21) 4584.2942
Japan:
Tel. (03) 3258.9565 • Fax (03) 3258.9567
Korea:
Tel. (02) 2278.5147 • Fax (02) 2264.1729
Malaysia:
Tel. (603) 5638.9940 • Fax (603) 5638.9829
Singapore:
Tel. 6296.7118 • Fax 6291.6906
South Africa:
Tel. (011) 615.6076 • Fax (011) 615.8582
Taiwan:
Tel. 886.2.2739.3014 • Fax 886.2.2739.2983
Thailand:
Tel. 66.2619.0708 • Fax 66.2619.0061
United Kingdom:
Tel. (01525) 850.855 • Fax (01525) 853.668
USA:
Tel. (401) 765.7500 • Fax (401) 765.7575
For e-mail contacts and complete list of Sales and Technical offices, please see www.hannainst.com
36
MANPHTRR3 11/05
Loading...