Remove the instrument from the packing material and examine it carefully to make sure
that no damage has occurred during shipping. If there is any noticeable damage, notify
your Dealer.
Note: Save all packing material until you are
sure that the instrument functions correctly. All defective items must be returned in the original packing material
together with the supplied accessories.
GENERAL DESCRIPTION
HI 8614/HI 8614L (pH) and HI 8615/ HI 8615L
(ORP) are 2-wire water-resistant transmitters
specially designed for long distance measurement of pH or ORP in industrial applications.
Two versions are available: the standard
HI8614 and HI8615, and the HI8614L and
HI8615L models with LCD.
The LCD allows easy verification and monitoring of measured values, and provides easier
calibration and maintainance.
The pH or ORP signal is transmitted in a 2wire current loop in the range of 4 to 20 mA.
HI8614 and HI8614L can be connected to
the Hanna process instruments HI8510T,HI8710T or HI 8711T, or to any recorder,
computer or data monitoring device that accepts 4 to 20 mA input.
HI8615 and HI8615L can be connected to
the Hanna meters HI8512T, HI8720T or to
any recorder, computer or data monitor that
accepts 4 to 20 mA input.
32
The transmitters use a universal BNC socket
for quick and secure connection to any electrode with a BNC connector.
For HI8614 and HI8614L the temperature
compensation is performed by the transmitter's
ATC circuitry if the temperature probe
(HI76608, optional) is attached. If ATC is not
required, it is possible to substitute the temperature probe with a fixed resistor.
The input is isolated from the current loop to
eliminate problems related to ground loop,
low insulation cables, multiple electrode connections, and a common mode voltage of up
to 100 V ensures true differential readings.
The terminal board of the transmitter provides
for connection of power supply, pH or ORP
electrodes and temperature probe (for HI8614
and HI8614L models only).
The unit is enclosed in a protective casing
conforming to IP 65 standards.
SPECIFICATIONS OF HI8614 & HI8614L
Range4 to 20 mA / 0.00 to 14.00 pH
Resolution0.01 mA / 0.01 pH
Accuracy (@20°C/68°F)
±0.02 mA / ±0.02 pH
Typical EMC Deviation
±0.25 mA / ±0.2 pH
CalibrationOffset: ±2.2 mA / ±2 pH
Slope: ±0.5 mA / 86 to 116%
Temperature Compensation
Fixed or automatic from 0 to 100°C
(32 to 212°F) with HI76608 probe
Input Impedance1012 Ohm
Output4-20 mA (isolated)
Installation CategoryI I
ProtectionIP 65
Environment0 to 50°C (32 to 122°F);
RH max 95% non-condensing
Power Supply
HI8614: 18 to 30 Vdc
HI8614L: 20 to 36 Vdc
LoadMax 500 Ohm
Dimensions165 x 110 x 90 mm
(6.5 x 4.3 x 3.5'')
Weight1 kg (2.2 lb.)
HI8614
pH TRANSMITTER
pH
HI8614L
pH INDICATOR
& TRANSMITTER
54
SPECIFICATIONS OF HI8615 & HI8615L
TERMINAL BOARD CONNECTIONS
Range4 to 20 mA / ±1000 mV
Resolution0.01 mA / 1 mV
Accuracy (@20°C/68°F)
±0.02 mA / ±5 mV
Typical EMC Deviation
±0.25 mA / ±15 mV
CalibrationOffset: ±0.8 mA / ±100 mV
Slope: ±0.8 mA / 90 to 110%
Input Impedance1012 Ohm
Output4-20 mA (isolated)
Installation CategoryI I
ProtectionIP 65
Environment0 to 50°C (32 to 122°F);
RH max 95% non-condensing
Power Supply
HI8615: 18 to 30 Vdc
HI8615L: 20 to 36 Vdc
LoadMax 500 Ohm
Dimensions165 x 110 x 90 mm
(6.5 x 4.3 x 3.5'')
Weight1 kg (2.2 lb.)
Unscrew the 4 screws and remove the top cover.
There are three cable glands on the transmitter
cover. The large cable gland with the split in the
rubber is for the electrode.
ELECTRODE
CABLE GLAND
WIRE CABLE GLANDS
Connect the positive supply to the strip terminal
"+VE LOOP" and the negative supply to the
terminal "-VE LOOP" of the transmitter terminal
block.
HI 8615
ORP TRANSMITTER
mV
HI 8615
ORP TRANSMITTER
76
The wire between the transmitter and the
recorder/indicator/controller should be a PVC
insulated two wire with a wire diameter of at
least 0.7 mm. This wire is fed through one of
the smaller cable glands. The maximum distance between the power supply and the
amplifier is 300 m (1000'). It is not necessary
to use shielded cable. The transmitter is
protected against inversion of supply voltage.
Electrode connection: connect
the BNC of the cable to the BNC
socket on the transmitter.
pH CALIBRATION WITH AUTOMATIC
TEMPERATURE COMPENSATION
(HI 8614 & HI 8614L)
Initial preparation
•Connect the pH electrode
to the BNC socket.
•Connect the temperature
probe to the transmitter.
Pour small quantities of pH 7.01 and pH 4.01
solution into two clean beakers.
For HI8614 & HI8614L only: for
automatic temperature compensation, connect the 2 termi-
nals of the temperature probe
(HI76608, optional) to "TEMP.
PROBE" terminals.
If automatic temperature compensation is not
required, short the "TEMP.
PROBE" terminals with a resistance according to the external
temperature:
Temperature (°C) Resistance (Ohms)
01634
101774
201922
302078
402242
502412
602590
A 2 kohm resistor is factory mounted for 25°C
temperature compensation.
HI 7004
HI 7004HI 7007
For accurate calibration use two beakers for
each buffer solution, the first one for rinsing
the electrode, the second one for calibration.
RINSE
HI 7007
CALIBRATION
HI 7007
To get accurate readings, use pH7.01 and
pH 4.01 if you are going to measure acid
samples or pH 7.01 and pH 10.01 for alkaline
measurements.
Note: with HI8614L the instruments display
can be used during calibration without
the need to connect the ammeter and
the reading is directly expressed in pH
units.
98
Procedure
•Disconnect the +ve supply cable from
"+VE LOOP" terminal and connect a 20mA
f.s. ammeter between the +ve cable and
"+VE LOOP" terminal.
•Remove the protective cap
from the electrode, rinse it
with some pH 7.01 solution
or immerse it in the pH 7
rinse solution, then immerse
the pH electrode and tem-
HI 7007
perature probe into pH 7.01
calibration buffer solution;
shake briefly and wait for
the reading to stabilize.
For other buffer temperatures, refer to page
13 for the appropriate mA / pH reading.
•Rinse the electrode with tap
water or distilled water and
a small amount of pH 4.01
or 10.01 buffer solution or
immerse it in the pH4 rinse
solution (2nd calibration
point). Dip the electrode and
HI 7004
the temperature probe into
pH 4.01 (or 10.01) calibration buffer solution, shake
briefly and wait for a few
minutes for reading to stabilize.
•Adjust the slope trimmer until the ammeter reads 8.58 mA or the display shows
"4.01" at 25°C. For other buffer temperatures, refer to page 13 for the appropriate
mA / pH reading (HI 8614L).
Note: the tip of the elec-
trode should be submerged approximately
4 cm (1½") into the
4 cm
1½"
solution. The temperature probe should be
located as close to
the pH electrode as
possible.
•Adjust the offset trimmer until the ammeter reads 12mA or the display shows "7.01"
(HI 8614L only) if the temperature of the
buffer is at 25°C.
SLOPE ADJUSTMENT
OFFSET ADJUSTMENT
mA
SLOPE ADJUSTMENT
OFFSET ADJUSTMENT
mA
If you are using pH 10.01
buffer solution adjust the
slope trimmer until the am-
mA
meter read 15.43mA at
25°C (77°F) or the value indicated at page 13.
FOR HI8614L ONLY
The Display Module is factory calibrated, so
that the LCD display results are referred to
the 4-20 mA loop current (e.g. LCD displays
0.00 pH when loop current is 4.00 mA and
displays 14.00 pH when current is 20.00mA).
1110
Under normal application, adjustment on this
module may not be necessary.
If routine check is required, the following procedures shall be performed.
•Calibrate the transmitter module using a
ammeter, as explained in the previous
pages.
•Simulate a 4.00 mA loop current for the
transmitter (i.e. pH 0.00 as Electrode input) and check for display reading.
•Simulate a 20.00 mA loop current for the
transmitter (i.e. pH 14.00 as Electrode
input) and check for display reading.
Note: when the instrument is used in con-
junction with the Hanna indicator
HI 8510T or controllers HI 8710T,
HI 8711T, the calibration can also be
performed on the indicator/controller.
In this case slight calibration adjustment can be made on the indicator/
controller even if the whole system calibration is advised, always starting from
the transmitter.
pH CALIBRATION WITH FIXED
TEMPERATURE COMPENSATION
(HI8614 & HI 8614L only)
•Take the temperature of the
°C
buffer solutions using a
Checktemp or a thermometer with a resolution of at
least 1°.
•Connect the appropriate resistor to the "TEMP.
PROBE" terminals (see
page 6) depending on the
temperature of the calibration solution.
•Follow the procedure as outlined in the
calibration with automatic temperature compensation and use the thermometer reading to adjust the trimmer until the ammeter reads the value according to the following table.
Temperature pH 4.01 BufferpH 7.01 BufferpH 10.01 Buffer
°C°FpHm ApHm ApHmA
0
32
4.01
8.576
7.13
12.137
10.32
10
15
20
25
30
35
40
45
50
55
60
5
41
4.00
8.565
7.10
50
59
68
77
86
95
104
113
122
131
140
4.00
4.00
4.00
4.01
4.02
4.03
4.04
4.05
4.06
4.07
4.09
8.565
8.565
8.565
8.576
8.587
8.599
8.610
8.622
8.633
8.645
8.667
7.07
7.04
7.03
7.01
7.00
6.99
6.98
6.98
6.98
6.98
6.98
12.103
12.069
12.034
12.023
12.000
11.989
11.977
11.966
11.966
11.966
11.966
11.966
10.24
10.18
10.12
10.06
10.01
9.96
9.92
9.88
9.85
9.82
9.79
9.77
15.789
15.697
15.629
15.560
15.491
15.434
15.377
15.331
15.285
15.251
15.217
15.182
15.159
1312
ORP CALIBRATION (HI8615 & HI8615L)
Initial preparation
Disconnect the +ve supply cable from the
"+VE LOOP" terminal and connect a 20 mA
f.s. ammeter between the +ve cable and the
"+VE LOOP" terminal. With HI 8615L the
instrument display can be used during calibration without the need to connect the ammeter. In this case the values are directly
expressed in mV units.
Pour a small quantity of HI7020
HI 7020
ORP solution into a beaker.
Procedure
•Connect the shorting BNC
connector to the ORP transmitter.
•Adjust the OFFSET ADJUSTMENT trimmer on the module for a
display of 12 mA on the ammeter or 0 mV
on the instrument display (HI 8615L only).
This sets the zero point for the transmitter.
•If the reading lies outside
this range, adjust the slope
adjustment trimmer on the
SLOPE ADJUSTMENT
OFFSET ADJUSTMENT
transmitter for a reading just
within this range.
The unit is now calibrated.
A complete calibration of the transmitter mod-
ule is advised periodically.
This calibration procedure requires the HI 8427
or the HI 931001 pH and ORP simulator to
simulate the ORP electrode.
HI 8427 or HI 931001 produce a known signal
into the system so that the faults of the
system can be isolated.
•Connect the ORP transmitter to the simulator.
SLOPE ADJUSTMENT
OFFSET ADJUSTMENT
mA
•Connect the ORP electrode to the transmitter and immerse the tip of the electrode
into a beaker of HI 7020 ORP calibration
solution, and check that the ammeter reading lies between 13.6 and 14.2 mA or the
instrument reading is between 200 and 275
mV at 25°C (HI8615L).
HI 7020
mA
•Set the simulator to 0 mV and adjust the
offset trimmer to read 12 mA on the ammeter or 0 mV on the HI 8615L display (HI
8615L only).
SLOPE ADJUSTMENT
OFFSET ADJUSTMENT
mA
1514
•Set the simulator to 350 mV and adjust
the slope trimmer to read 14.8 mA on the
ammeter or 350 mV on the HI 8615L
display (HI 8615L only).
SLOPE ADJUSTMENT
OFFSET ADJUSTMENT
mA
•Connect the ORP electrode to the module
and immerse the tip of the electrode into
the beaker of HI 7020 ORP solution and
check that the ammeter reading lies between 13.6 and 14.2 mA or the instrument
reading is between 200 and 275mV at
25°C (HI 8615L only).
HI 7020
mA
•Only if the reading lies outside this range, adjust the
slope adjustment trimmer on
the transmitter to reflect a
SLOPE ADJUSTMENT
OFFSET ADJUSTMENT
reading within this range.
FOR HI8615L ONLY
The HI 8615L is factory calibrated, and the
displayed values are referenced to the 4-20
mA loop current (e.g. LCD displays -1000
mV when loop current is 4.00 mA and displays +1000 mV when current is 20.00mA).
Under normal application, adjustment on this
module may not be necessary. If routine
check is required, the following procedures
shall be performed.
•Follow the above procedure for HI 8615
(see page 12).
•Simulate a 12.00 mA loop current for the
transmitter (i.e. 0 mV at Electrode input)
and check display reading.
•Simulate a 20.00 mA loop current for the
transmitter (e.g. +1000 mV at Electrode
input) and check display reading.
1716
Please note:
-1000 mV=4 mA
1000 mV=20 mA
350 mV=14.8 mA
200 mV=13.6 mA
275 mV=14.2 mA
ELECTRODE CONDITIONING
AND MAINTENANCE
0 mV=12 mA
Reference
Filling Hole
Reference
Filling Hole
1 mV=12.008 mA
Note: when the meter is used in conjunction
with the Hanna indicator HI 8512T, or
the controller HI 8720T, the calibration
can also be performed on the indicator/
controller. In this case slight adjustment can be made on the indicator/
controller even if the whole system calibration is advised, always starting from
the transmitter.
Sensitive
Wire
Reference
Wire
Reference
Junction
Glass
Bulb
Reference
Wire
Reference
Junction
Platinum or
Gold tip
Plastic Body
pH Electrode
Reference
Plastic Body
ORP Electrode
Glass Body
pH Electrode
Wire
Glass Body
ORP Electrode
Reference
Wire
Sensitive
Wire
Reference
Junction
Glass
Bulb
Reference
Junction
Platinum or
Gold tip
PREPARATION
Remove the protective cap.
DO NOT BE ALARMED IF ANY SALT DE-
POSITS ARE PRESENT.
This is normal with electrodes and they will
disappear when rinsed with water.
During transport tiny bubbles of air may have
formed inside the glass bulb. The electrode
cannot function properly under these conditions. These bubbles can be removed by
1918
"shaking down" the electrode as you would
do with a glass thermometer.
If the bulb and/or junction are dry, soak the
electrode in HI70300 storage solution for at
least one hour.
For refillable electrodes
If the fill solution (electrolyte) is more than
1 cm (½") below the fill hole, add HI7082
3.5M KCl electrolyte solution for double junction or HI7071 3.5M KCl+AgCl electrolyte
solution for single junction electrodes. For a
faster response unscrew the fill hole screw
during measurements.
For AmpHel® electrodes
If the electrode does not respond to pH
changes, the battery is run down and the
electrode should be replaced.
MEASUREMENT
Rinse the electrode tip with distilled water.
Immerse the tip (4 cm / 1½") in the sample
and stir gently for approx. 30 seconds.
For a faster response and to avoid cross
contamination of the samples, rinse the electrode tip with a few drops of the solution to be
tested, before taking measurements.
STORAGE
To minimize clogging and assuring a quick
response time, the glass bulb and the junction should be kept moist and not allowed to
dry out. Replace the solution in the protective
cap with a few drops of HI70300 storage
solution or, in its absence, filling solution
(HI7071 for single junction or HI7082 for
double junction electrodes). Follow the Preparation Procedure above before taking measurements.
Note: NEVER STORE THE ELECTRODE IN
DISTILLED WATER OR DRY.
PERIODIC MAINTENANCE
Inspect the electrode and the cable. The cable
used for connection to the meter must be
intact and there must be no points of broken
insulation on the cable or cracks on the electrode stem or bulb. Connectors must be perfectly clean and dry. If any scratches or cracks
are present on the electrode body, replace
the electrode. Rinse off any salt deposits with
water.
For refillable electrodes
Refill it with fresh electrolyte (HI7071 for single
junction or HI7082 for double junction electrodes). Allow the electrode to stand upright
for 1 hour.
Follow the Storage Procedure above.
CLEANING PROCEDURE
GeneralSoak in Hanna HI7061 general
cleaning solution for approximately 1 hour.
Removal of films, dirt or deposits on the membrane/junction:
ProteinSoak in Hanna HI7073 protein
cleaning solution for 15 minutes.
InorganicSoak in Hanna HI7074 inorganic
cleaning solution for 15 minutes.
Oil/greaseRinse with Hanna HI7077 Oil &
Fat cleaning solution.
IMPORTANT: After performing any of the
cleaning procedures rinse the electrode thoroughly with distilled water, drain and refill the
reference chamber with fresh electrolyte, (not
necessary for GEL filled electrodes) and soak
the electrode in HI70300 storage solution for
at least 1 hour before taking measurements.
2120
TROUBLESHOOTING
-20 -20
-
0 +10+20 +30+40+50+60+70+80+90
°C
-
-
-
-
-
-
Evaluate your electrode performance based
on the following.
•Noise (Readings fluctuate up and down)
could be due to:
- Clogged/Dirty Junction: refer to the
Cleaning Procedure above.
- Loss of shielding due to low electrolyte
level (in refillable electrodes only): HI7071
for single junction or HI7082 for double
junction electrodes.
• Dry Membrane/Junction: soak in
HI70300 storage solution for at least one
hour.
•Drifting: soak the electrode tip in warm
Hanna solution HI7082 for one hour and
rinse the tip with distilled water. Refill with
fresh HI7071 for single junction electrodes
and HI7082 for double junction electrodes.
•Low Slope: refer to the Cleaning Procedure above.
•No Slope: check the electrode for cracks
in glass stem or bulb and replace the
electrode.
•Slow Response/Excessive Drift: soak the
tip in Hanna HI7061 cleaning solution for
30 minutes, rinse thoroughly in distilled
water and then follow the Cleaning Procedure above.
TEMPERATURE-RESISTANCE
CORRELATION FOR HANNA
pH SENSITIVE GLASS
The resistance of glass electrodes partially
depends on the temperature. The lower the
temperature, the higher the resistance. It takes
longer time for the reading to stabilize if the
resistance is higher. In addition, the response
time will suffer to a greater degree at temperatures below 10°C.
Ω
9
2x10
10
9
1x10
10
8
10
2x10
8
1x10
10
7
2x10
10
10
7
1x10
10
Since the resistance of the pH electrode is in
the range of 200 Mohm, the current across
the membrane is in the pico Ampere range.
Large currents can disturb the calibration of
the electrode for many hours.
For these reasons high humidity environ-
ments, short circuits and static discharges
are detrimental for a stable pH reading.
The pH electrode's life also depends on the
temperature. If constantly used at high temperatures, the electrode life is drastically reduced.
2322
Typical Electrode Life
Ambient Temperature 1- 3 years
90 °C Less than 4 months
120°C Less than 1 month
High concentrations of sodium ions interfere
with readings in alkaline solutions; the pH at
which the interference starts to be significant
depends upon the composition of the glass.
This interference is the alkaline error and
causes the pH to be underestimated. Hanna's
glass formulations have the indicated characteristics.
Alkaline Error
Sodium Ion Correction for the Glass
at 20-25°C
ConcentrationpH Error
0.1 Mol L-1 Na
1.0 Mol L-1 Na
+
13.00
13.50
14.00
+
12.50
13.00
13.50
14.00
0.10
0.14
0.20
0.10
0.18
0.29
0.40
INSTALLATION PROCEDURE
AND EXAMPLES
The HI 8614, HI8614L, HI8615 and HI8615L
transmitters may be wall mounted in any convenient location near the measurement point.
To minimize thermal drift due to extreme temperature fluctuations during the measurement
process, particularly if the measurement is
conducted outdoors, it is best to protect the
transmitter in an enclosed casing.
General Installation Procedure
For most industrial application involving long
term monitoring and control, it is also recommended to use tank electrode holders (HI6050
or HI6051) to protect the pH electrode and the
temperature probe from contamination by the
test solution.
Controlling the pH / ORP with a recorder
2524
Monitoring the pH/ORP with Panel Mounting
pH (HI8510)/ORP (HI8512) Indicator
Controlling the pH/ORP with
an Industrial Regulator
Monitoring and Controlling the pH/ORP
with Panel Mounting Indicator/Regulator
and Dosage Control of either Acid or Base
Monitoring and Controlling the pH with
(HI 8711) Panel Mounting Indicator/
Regulator with Independent Dosage
Control for Acid and Base
2726
PG13.5 THREAD
m
110mm30mm
DIA 9.5mm
3
110mm
D
M13 x 1.5
DIA
16
mm
7
mm
25
mm
PG13.5 THREAD
m
110mm30mm
D
3/4 x 16 UNF
M13 x 1.5
DIA
16
mm
7
mm
25
mm
150mm
DIA 16.5mm
M13 x 1.5
A
6
m
mm
mm
ACCESSORIES
pH CALIBRATION SOLUTIONS
HI7004M pH 4.01 buffer solution, 230 mL
HI7004L pH 4.01 buffer solution, 500 mL
HI7006M pH 6.86 buffer solution, 230 mL
HI7006L pH 6.86 buffer solution, 500 mL
HI7007M pH 7.01 buffer solution, 230 mL
HI7007L pH 7.01 buffer solution, 500 mL
HI7009M pH 9.18 buffer solution, 230 mL
HI7009L pH 9.18 buffer solution, 500 mL
HI7010M pH 10.01 buffer solution, 230mL
HI7010L pH 10.01 buffer solution, 500 mL
ORP SOLUTIONS
HI 7020M 200-275 mV test solution, 230 mL
HI 7020L 200-275 mV test solution, 500 mL
HI 7091M Reducing pretreatment solution, 230 mL
HI 7091L Reducing pretreatment solution, 500 mL
HI 7092M Oxidizing pretreatment solution, 230 mL
HI 7092L Oxidizing pretreatment solution, 500 mL
ELECTRODE MAINTENANCE SOLUTIONS
HI70300M Storage solution, 230 mL
HI70300L Storage solution, 500 mL
HI7061M General cleaning solution, 230 mL
HI7061L General cleaning solution, 500 mL
HI7073M Protein cleaning solution, 230 mL
HI7073L Protein cleaning solution, 500 mL
HI7074M Inorganic cleaning solution, 230 mL
HI7074L Inorganic cleaning solution,500 mL
HI7077M Oil & Fat cleaning solution, 230 mL
HI7077L Oil & Fat cleaning solution,500 mL
HI70713.5M KCl + AgCl electrolyte solu-
tion, 4 x 5 0 mL bottle, for single junction electrodes
HI70721M KNO3 electrolyte solution, 4 x
50 mL bottle
HI70823.5M KCl electrolyte solution, 4 x
50 mL bottle, for double junction
electrodes
pH ELECTRODES
HI 1090TScrewcap PG13.5 connector,
double junction, glass body
φ 12mmφ 9.5m
HI 1110SScrew connector, single junction,
glass body
HI 1130B/3 BNC connector, 3 m (9.9') cable,
single junction, glass body
/4 x 16 UNF
IA 20.5mm
38.5mm
HI 1110SHI 1130B/3
DIA 12mm
HI 1110TScrewcap PG13.5 connector,
double junction, glass body
φ 12mmφ 9.5m
HI 1114SScrew connector, double junction
plastic body
HI 1134B/3 BNC connector, 3 m (9.9') cable,
double junction plastic body
DIA 12mm
IA 20.5mm
38.5mm
HI 1114SHI 1134B/3
110mm
HI 1115SScrew connector, single junction,
glass body
HI 1135B/3 BNC connector, 3 m (9.9') cable,
single junction, glass body
5mm
DIA 7.6mm
DI
1
m
25
7
HI 1115SHI 1135B/3
25mm
2928
DIA 12mm
D
3/4 x 16 UNF
D
3/4 x 16 UNF
D
3/4 x 16 UNF
110mm
M13 x 1.5
DIA
16
mm
7
mm
25
mm
PG13.5 THREAD
m
110mm30mm
DIA 16.5mm
M13 x 1.5
A
6
m
mm
mm
HI 1210TScrewcap PG13.5 connector,
PG13.5 THREAD
110mm30mm
D
3/4 x 16 UNF
D
3/4 x 16 UNF
D
DIA 9.5mm
110mm
D
3/4 x 16 UNF
double junction, plastic body
φ 12mm
HI 2910B/5 BNC connector, 5 m (16.5') cable,
double junction, plastic body, builtin amplifier
DIA 12mm
IA 20.5mm
HI 1910BBNC connector, 1 m (3.3') cable,
double junction, plastic body, builtin amplifier
DIA 12mm
IA 20.5mm
38.5mm
110mm
HI 1911BBNC connector, 1 m (3.3') cable,
double junction, plastic body, builtin amplifier
DIA 12mm
IA 20.5mm
38.5mm
110mm
HI 1912BBNC connector, 1 m (3.3') cable,
double junction, plastic body, builtin amplifier
HI 1912B/5 BNC connector, 5 m (16.5') cable,
double junction, plastic body, builtin amplifier
3/4 x 16 UNF
IA 20.5mm
DIA 12mm
38.5mm
110mm
ORP ELECTRODES
HI 2930B/5 BNC connector, 5 m (16.5') cable,
Pt, plastic body, built-in amplifier
DIA 12mm
IA 20.5mm
38.5mm
110mm
HI 3110SScrew-type connector, Pt, glass
body
HI 3130B/3 BNC connector, 3 m (9.9') cable,
Pt, glass body
DIA 12mm
IA 20.5mm
38.5mm
HI 3110SHI 3130B/3
HI 3110TScrewcap PG13.5 connector, Pt,
glass body
φ 12mmφ 1m
38.5mm
HI 2114B/5 BNC connector, 5 m (16.5') cable,
double junction, plastic body
DIA 12mm
IA 20.5mm
38.5mm
110mm
HI 3115SScrew-type connector, side-arm,
Pt, glass body
HI 3135B/3 BNC connector, 3 m (9.9') cable,
side-arm, Pt, glass body
5mm
DIA 7.6mm
DI
1
m
25
7
HI 3115SHI 3135B/3
25mm
3130
DIA 12mm
150mm
HI 3210TScrewcap PG13.5 connector, Pt,
PG13.5 THREAD
110mm30mm
D
3/4 x 16 UNF
M13 x 1.5
DIA
16
mm
7
mm
25
mm
3/4 x 16 UNF
110mm
D
3/4 x 16 UNF
110mm
M13 x 1.5
A
6
m
mm
mm
3/4 x 16 UNF
110mm
plastic body
φ 12mm
EXTENSION CABLES FOR SCREW-TYPE
ELECTRODES ONLY (SCREW TO BNC CONNECTOR)
HI 7855 SERIES CABLE CONNECTORS
CONNECTOR AND 3.0 mm (0.12") CABLE WITH BNC
CONNECT TO
SCREW TYPE
ELECTRODES
CONNECT TO THE
BNC SOCKET
OF THE METER
HI 3410SScrew connector, Pt, plastic body
HI 3430B/3 BNC connector, 3 m (9.9') cable,
Pt, plastic body
DIA 12mm
IA 20.5mm
38.5mm
HI 3410SHI 3430B/3
110mm
HI 3932B/5 BNC connector, 5 m (16.5') cable,
Pt, plastic body, built-in amplifier
DIA 12mm
DIA 20.5mm
38.5mm
HI 4110SScrew-type connector, Au, glass
body
HI 4130B/3 BNC connector, 3 m (9.9') cable,
Au, glass body
DIA 12mm
DI
IA 20.5mm
1
m
25
7
HI 4110SHI 4130B/3
38.5mm
HI 4932B/5 BNC connector, 5 m (16.5') cable,
Au, plastic body, built-in amplifier
HI7855/1Extension cable 1 m (3.3') long
HI7855/3Extension cable 3 m (9.9') long
HI7855/5Extension cable 5 m (16.5') long
HI7855/10Extension cable 10 m (33') long
HI7855/15Extension cable 15 m (49.5') long
OTHER ACCESSORIES
BL PUMPSDosing pumps with flow rate from
1.5 to 20 lph
HI 98501Pocket-size, thermometer with
penetration probe and 0.1°C
resolution (range -50.0 to
150.0°C)
HI6050 & HI 6051 Submersible electrode
holders
HI 6054 & HI 6057 Electrode holders for in-line
applications
HI 7871 & HI 7873 Level controllers
HI8427pH / ORP electrode simulator
HI931001pH / ORP electrode simulator
with LCD
DIA 12mm
DIA 20.5mm
38.5mm
3332
WARRANTY
All Hanna Instruments meters are warranted
for two years against defects in workman-
ship and materials when used for their intended purpose and maintained according to
instructions.
The probes and the electrodes are warranted for a period of six months.
Damages due to accident, misuse, tampering or lack of prescribed maintenance are not
covered. This warranty is limited to repair or
replacement free of charge.
If service is required, contact the dealer from
whom you purchased the instrument. If under
warranty, report the model number, date of
purchase, serial number and the nature of the
failure. Obtain a Returned Goods Authorization from the Customer Service department
first and then return the instrument with the
Authorization # included along with shipment
costs prepaid. If the repair is not covered by
the warranty, you will be notified of the charge
for repair or replacement. When shipping any
instrument, make sure it is properly packaged for complete protection.
CE DECLARATION OF CONFORMITY
All rights are reserved. Reproduction in whole
or in part is prohibited without the written
consent of the copyright owner.
Hanna Instruments reserves the right to
modify the design, construction and appear-
ance of its products without advance notice.
Recommendations for Users
Before using these products, make sure that they are entirely suitable for the
environment in which they are used.
Operation of these instruments in residential area could cause unacceptable
interferences to radio and TV equipments, requiring the operator to take all necessary
steps to correct interferences.
Any variation introduced by the user to the supplied equipment may degrade the
instruments' EMC performance.
To avoid damages or burns, do not perform any measurement in microwave ovens.
3534
SALES AND TECHNICAL SERVICE CONTACTS
Australia:
Tel. (03) 9769.0666 • Fax (03) 9769.0699
China:
Tel. (10) 88570068 • Fax (10) 88570060
Egypt:
Tel. & Fax (02) 2758.683
Germany:
Tel. (07851) 9129-0 • Fax (07851) 9129-99
Greece:
Tel. (210) 823.5192 • Fax (210) 884.0210
Indonesia:
Tel. (21) 4584.2941 • Fax (21) 4584.2942
Japan:
Tel. (03) 3258.9565 • Fax (03) 3258.9567
Korea:
Tel. (02) 2278.5147 • Fax (02) 2264.1729
Malaysia:
Tel. (603) 5638.9940 • Fax (603) 5638.9829
Singapore:
Tel. 6296.7118 • Fax 6291.6906
South Africa:
Tel. (011) 615.6076 • Fax (011) 615.8582
Taiwan:
Tel. 886.2.2739.3014 • Fax 886.2.2739.2983
Thailand:
Tel. 66.2619.0708 • Fax 66.2619.0061
United Kingdom:
Tel. (01525) 850.855 • Fax (01525) 853.668
USA:
Tel. (401) 765.7500 • Fax (401) 765.7575
For e-mail contacts and complete list of Sales and
Technical offices, please see www.hannainst.com
36
MANPHTRR3 11/05
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.