SPECIFICATIONSSPECIFICATIONS
SPECIFICATIONS
SPECIFICATIONSSPECIFICATIONS
Range 0 to 3.5 mg/L (ppm) as Total Chlorine Smallest Increment 0.1 mg/L Analysis Method Colorimetric Sample Size 5 mL Number of Tests 100 Case Dimensions 235x175x115 mm (9.2x6.9x4.5") Shipping Weight 977 g (34.5 oz.)
INSTRUCTIONSINSTRUCTIONS
INSTRUCTIONS
INSTRUCTIONSINSTRUCTIONS
READ THE ENTIRE INSTRUCTIONS BEFORE USING THE KIT
• Use the pipette to fill each glass vial with 5 mL of sample (up to the mark).
5 mL
• Insert one of the vials into the left hand opening of the checker disc. This is the blank.
• Keep the checker disc at a distance of 30-40 cm (12­16") from the eyes to match the color. Rotate the disc while looking at the color test windows and stop when you find the color match. Read the value in the result window directly in mg/L (or ppm) of Chlorine.
Checker Disc
www.hannainst.com
Dear Customer, Thank you for choosing a Hanna Product. Please read the instruction sheet carefully before using the
test kit. It will provide you with the necessary information for correct use of the kit. If you need additional information, do not hesitate to e-mail us at tech@hannainst.com.
Each kit is supplied with:
HI 93711-0 Total Chlorine Reagent, packets (100 pcs);
• Deionized Water, 1 bottle (500 mL);
• 1 checker disc (containing the disc);
• 2 glass vials with caps;
• 1 plastic pipette (3 mL).
Note: Any damaged or defective item must be returned in
its original packing materials.
SIGNIFICANCE AND USESIGNIFICANCE AND USE
SIGNIFICANCE AND USE
SIGNIFICANCE AND USESIGNIFICANCE AND USE
Chlorine is the most commonly used water disinfectant in applications that vary from treatment of drinking water and wastewater, pool and spa sanitization, to food process­ing and sterilization. Chlorine present in water binds with bacteria, leaving only a part of the original quantity (free chlorine) to continue its disinfecting action. If the free chlorine level is improper with respect to pH, water will have an unpleasant taste and odor and the disinfecting poten­tial of the chlorine will be diminished.
Free chlorine reacts with ammonium ions and organic compounds to form chlorine compounds resulting in dimin­ished disinfecting capabilities compared with free chlorine. Chlorine compounds together with chloramines form com­bined chlorine. Combined chlorine and free chlorine together result in total chlorine.
While free chlorine has a much higher disinfectant potential, combined chlorine has a much higher stability and has a lesser volatility.
Note: mg/L is equivalent to ppm (parts per million).
CHEMICAL REACTIONCHEMICAL REACTION
CHEMICAL REACTION
CHEMICAL REACTIONCHEMICAL REACTION
The reaction between chlorine and the DPD reagent causes a pink tint in the sample which is proportional to the chlorine concentration.
ISTR38016 02/00 PRINTED IN ITALY
• Add to the other vial deionized water up to the 10 mL mark. Replace the cap and shake to mix.
• Remove the cap and add 1 packet of HI 93711-0 reagent. Replace the cap and mix.
• Wait for 2 minutes to allow reaction to occur. This is the reacted sample.
• Remove the cap and insert the reacted sample into the right hand opening of the checker disc.
• Hold the checker disc so that a light source illuminates the samples from the back of the windows.
For best results: Perform the reading three times and
take the average value (divide by 3 the sum of the three numbers). Intensely colored samples will make the color matching difficult and they should be adequately treated before performing the test. Sus­pended matter in large amounts should be removed by prior filtration.
Caution: Ultraviolet radiation may cause fading of colors.
When not in use, keep the disc protected from light, in a cool and dry place.
Interferences: bromine, iodine, fluorine, ozone, oxidized
manganese and chromium.
REFERENCESREFERENCES
REFERENCES
REFERENCESREFERENCES
Adaptation of the EPA recommended DPD method 330.5.
HEALTH AND SAFETYHEALTH AND SAFETY
HEALTH AND SAFETY
HEALTH AND SAFETYHEALTH AND SAFETY
The chemicals contained in this kit may be hazardous if improperly handled. Read the relevant Health and Safety Data Sheet before performing this test.