written consent of the copyright owner, Hanna Instruments Inc., 584 Park East Drive,
Woonsocket, Rhode Island, 02895 , USA.
32
PRELIMINARY EXAMINATION
Remove the instrument from the packing material and examine it carefully to make sure that no damage has occurred
during shipping. If there is any noticeable damage, notify
your Dealer or the nearest Hanna Customer Service Center
immediately.
NoteSave all packing materials until you are sure that the instru-
ment functions correctly. Any damaged or defective items must
be returned in their original packing materials together with
the supplied accessories.
GENERAL DESCRIPTION
The product is a real time microprocessor-based pH/ORP
controller. It provides accurate measurements, flexible
ON/OFF, proportional or PID control capabilities and dual
alarm signals.
The system is composed of a case inside which the signal
conversion circuitry, the microprocessor circuitry and the
output power drivers are contained.
MAIN FEATURES OF DIFFERENT MODELS
• Display: large LCD with 4 ½ 13 mm digits and 3 ½ 7.7
mm digits.
• LEDs: three (HI 22) or four (HI 21) LEDs are provided for
signaling the energizing of relay 1 (a yellow LED), relay 2
(a yellow LED in HI 21 Series only) and alarm relays (a
green and a red LED).
• Relays: Electromechanical type, working through NO,
COM and NC contacts; 1 or 2 output relays for acid or
base dosage; 1 output relay for alarm condition.
• RS485 isolated communication link.
• Calibration and Setup procedures allowed only through
an unlock password.
• Calibration: for HI 21 Series in 1, 2 or 3 points with buffers 4.01, 7.01 and 10.01 pH (25 °C); for HI 22 Series in
1 or 2 points at 0, 350 and 1900 mV.
• Temperature compensation of the HANNA standard buffers (for HI 21 Series only).
• Temperature compensation of the pH reading (for HI 21
Series only).
• Manual temperature setting when the temperature probe is
not inserted or temperature exceeds the upper range.
• Internally recorded last calibration data (non-volatile EEPROM memory): calibration date and time, pH offset, pH
slopes, number of calibration points and correspondent
pH values (for HI 21 Series only) or calibration date and
time and the mV calibration points used (for HI 22 Series
only).
• Input: pH electrode with BNC connector.
• ±5 VDC output for amplified electrodes (max 10 mA).
• Output:
- isolated 0-1 mA, 10 KW maximum load (optional);
- isolated 0-20 mA, 750 W maximum load (optional);
- isolated 4-20 mA, 750 W maximum load (optional);
2. LCD keyexits from setup and calibration modes and reverts back to
normal mode (in idle or control phases with the measurement on the display). In HI 21 series, during pH calibration,
alternately displays pH buffer value or current temperature
3. CAL DATA keylast calibration data viewing (enters and exits)
4. SETUP keyenters setup mode
5. CAL keyinitiates and exits calibration mode
6. ñ keyincreases the blinking digit/letter by one when selecting a
parameter. Advances forward while in last calibration data
viewing mode. Increases the temperature setting when
temperature probe is not inserted
7. ð keymoves to the next digit/letter (circular buffer) when selecting a parameter. Same as ñ key during last calibration
data viewing mode
8. ò keydecreases the blinking digit/letter by one when selecting a
parameter. Reverts backward while in last calibration data
viewing mode. Decreases the temperature setting when
temperature probe is not inserted
9. CFM keyconfirms current choice (and skips to the next item)
10. BNC Socket
11. LEDs
CONNECTIONS PANEL
1. 4-pin RS485 terminal (not for HI 21XY1 and HI 22XY1)
Unplug the meter before starting any electrical connections.
MECHANICAL DIMENSIONS
FRONT VIEWBOTTOM VIEW
76
SPECIFICATIONS HI 21 & HI 22
INSTALLATION
Range0.00 to 14.00 pH (HI 21 Series only)
±2000 mV (HI 22 Series only)
-9.9 to 120.0 °C
Resolution0.01 pH (HI 21 Series only)
1 mV (HI 22 Series only)
0.1 °C
Accuracy±0.02 pH (HI 21 Series only)
(@20°C/68°F)±2 mV (HI 22 Series only)
±0.5 °C
Typical EMC Deviation±0.05 pH (HI 21 Series only)
±4 mV (HI 22 Series only)
±1.0 °C
Analog Output0 to 1 mA, 0 to 20 mA, 4 to 20 mA
0 to 5 VDC, 1 to 5 VDC, 0 to 10 VDC
Digital OutputRS485
Power Supply Input±5V (for amplified electrodes)
Power Supply230 ±10% VAC or 115 ±10% VAC, 50/60 Hz
Power Consumption15 VA
Over Current Protection400 mA, 250V, FAST FUSE
Relays 1 and 2Electromechanical relay SPDT contact outputs,
5A-250 VAC, 5A - 30 VDC (resistive load)
(HI 211YZ, HI 212YZ and HI 221YZ)
HI 21 and HI 22 offer
a multitude of possibilities, from single and
dual setpoints to ON/
OFF, proportional or
PID dosage, isolated
outputs with user-selectable zoom,
bidirectional RS485, recorder outputs in
mAmps and Volts.
In addition, HI 21 andHI 22 are both
equipped with the exclusive differential
input.
In a system with poor
grounding, it is possible
to have a ground current flowing through
the reference junction.
This can cause a rapid
degradation of the electrode. The Hanna
differential input reduces the likelihood of
ground loops.
See the diagram for a
recommended installation.
Environment0-50 °C; max 85% RH not condensing
ProtectionIP 54
Dimensions221 x 181 x 86 mm (8.7 x 7.1 x 3.4’’)
Weightapproximately 1.4 Kg (3.1 lb.)
98
••
• Power Supply: Connect a 3-wire power cable
••
to the terminal strip, while paying attention to
the correct line (L), earth (PE) and neutral (N)
terminal connections.
Power: 115VAC - 100 mA / 230VAC - 50 mA.
Line Contact: fused inside 400 mA.
PE must be connected to ground; leakage current 1
mA.
••
• Electrode: Connect the pH or ORP electrode to the BNC
••
socket (#10 at page 6).
To benefit from the differential input, connect the proper
electrode wire (if available)
or a cable with a potential
matching pin (grounding bar) to the relevant terminal (#5
at page 7).
NoteWhen it is not possible to immerse the Potential Matching Pin
together with the pH electrode in the solution, disable the
differential input by connecting the Connection for Potential
Matching Pin (#5 page 7)
with the Connection for Electrode Reference (#6 on
page 7) with a jumper wire.
••
• Pt 100 Terminals: these contacts (#3 at page 7) connect
••
the Pt 100 temperature sensor for automatic temperature
compensation of pH measurement. In the case of shielded wire,
connect the shield to pin 4.
In the case of a 2-wire sensor connect the Pt 100 to pins 1 and 3,
and short pins 2 and 3 with a
jumper wire.
If the Pt 100 has more than 2
wires, connect the two wires of one
end to pins 2 and 3 (pin 2 is an
auxiliary input to compensate for the cable resistance) and
one wire from the other end to pin 1.
Leave the fourth wire unconnected, if
present.
••
• Power Supply Output: these terminals provide +5V and -
••
5V DC (max current: 10 mA) signals to supply power to
amplified electrodes.
NoteAll cables connected to the panel should end with cable lugs.
SETUP MODE
HI 21 and HI 22 offer a multitude of possibilities from ON/
OFF, proportional or PID dosage to analog recorder output
and from alarm to selftest features.
The Setup Mode allows the user to set all needed characteristics of the meter.
The setup mode is entered by pressing SETUP
and entering the password when the device
is in idle or control mode.
Generally speaking, if the password is not inserted the user
can only view the setup parameters (except for password)
without modifying them (and the device remains in control
mode). An exception is certain setup items, or flags, which
can activate special tasks when set and confirmed.
Each setup parameter (or setup item) is assigned a twodigit setup code which is entered and displayed on the
secondary LCD.
The setup codes can be selected after password and CFM
are pressed. When CFM is pressed, the current setup item is
saved on EEPROM and the following item is displayed.
Whenever LCD is pressed, the device reverts back to control mode.
The same is true when CFM is
pressed on the last setup item.
The possible transitions in setup mode are the following:
ENTERING THE PASSWORD
• Press SETUP to enter the setup mode. The LCD will display
“0000” on the upper part and “PAS” on the lower. The first
digit of the upper part of the LCD will blink.
• Enter the first digit of the password by the ñ or ò keys.
1110
• Then confirm the displayed digit with
ð and move to the next one.
• When the whole password has been
inserted, press CFM to confirm it.
NoteThe default password is set at “0000”.
• The LCD will display “SET” on the
upper part and “c.00” on the lower,
allowing the user to edit setup parameters (see table below).
• Enter the code of the parameter you want to set, using the
arrow keys as per the password procedure above (e.g. 41).
• Confirm the code by pressing CFM and the default or the
previously memorized value will be displayed with the first
digit blinking.
NoteWhen the password is not inserted or a wrong password is
confirmed, the display will only show the previously memorized value, without blinking (read only mode). In this case,
the value cannot be set. Press LCD and start again.
• Enter the desired value using the arrow keys and then press
CFM.
• After confirmation, the selected parameter is displayed. The user can scroll
through the parameters by pressing CFM.
In order to directly set another parameter, press SETUP again and enter the
code or scroll to it by pressing CFM.
The following table lists the setup codes along with the description of the specific setup items, their valid values and whether
password is required to view that item (“PW” column):
CodeValid ValuesDefaultPW
00 Factory ID0 to 99990000no
01 Process ID0 to 9999 (models w/o RS485) 0000no
15 Relay 1 reset time0.1 to 999.9 minutes999.9 minsno
(HI 21523 only)
16 Relay 1 rate time0.0 to 999.9 minutes0.0 minsno
(HI 21523 only)
21* Relay 2 mode (M2)same as relay 10no
22* Relay 2 setpoint (S2)0.00 to 14.00 pH6.00 pHno
-2000 to 2000 mV-500 mV
* Available only in models with two relays
1312
CodeValid ValuesDefaultPW
CodeValid ValuesDefaultPW
23* Relay 2 hysteresis (H2)0.00 to 14.00 pH1 pHno
0 to 4000 mV50 mV
24* Relay 2 deviation (D2)0.50 to 14.00 pH1 pHno
25 to 4000 mV50 mV
25* Relay 2 reset time0.1 to 999.9 minutes999.9 minsno
(HI 21523 only)
26* Relay 2 rate time0.0 to 999.9 minutes0.0 minsno
(HI 21523 only)
30 Relay 3 high alarm (HA) 0.00 to 14.00 pH9.00 pHno
-2000 to 2000 mV600 mV
HA>LA, HA>S1 or HA>S2
31 Relay 3 low alarm (LA)0.00 to 14.00 pH5.00 pHno
-2000 to 2000 mV-600 mV
LA<HA, LA<S1 or LA<S2
32 Proportional control1 to 30 min (
not for HI 21523) 5 minno
mode period5 s to 30 min (HI 21523 only) 30 sno
33 Maximum relay ON time 1 to 60 min (
10 to 999 min (
HI 21523 only) 60no
not for HI 21523) 60no
(after which an alarm mode is entered)
34 Alarm mask time00:00 to 30:00 00:00no
(HI 21523 only)
40 Analog output selection0: 0-1mA 2no
1: 0-20 mA
2: 4-20 mA
3: 0-5 VDC
4: 1-5 VDC
5: 0-10 VDC
Note: this parameter refer to internal hardware configuration of the instrument. To
change it from the factory setting, see page 26.
41 Analog output0.00 to 13.00 pH0.00 pHno
lower limit-2000 to 2000 mV-2000 mV
(O_VARMIN)(O_VARMIN < O_VARMAX-(1.00pH or 50mV))
42 Analog output1.00 to 14.00 pH14.00 pHno
upper limit-2000 to 2000 mV2000 mV
(O_VARMAX)(O_VARMIN < O_VARMAX- (1.00pH or 50mV))
60 Current day01 to 31from RTCno
61 Current month01 to 12from RTCno
62 Current year1997 to 9999from RTCno
63 Current time00:00 to 23:59from RTCno
71 Baud rate1200, 2400, 4800, 96009600 (RS485) no
90 Display selftest0: off0yes
1: on
91 Keyboard selftest0: off0yes
1: on
92 EEPROM selftest0: off0yes
1: on
93 Relays and LEDs selftest0: off0yes
1: on
94 Watchdog selftest0: off0yes
1: on
99 Unlock password0000 to 99990000yes
NoteThe process controller automatically checks to ensure that
the entered data matches other related variables. If a wrong
configuration is entered, “WRONG” blinks on the LCD to
prompt the user. The correct configurations are the following:
If M1=/ 0 then S1
If M2=/ 0 then S2
For models other than HI 21523:
If M1= 1 then S1-H1
If M1= 2 then S1+H1
If M1= 3 then S1+D1
If M1= 4 then S1-D1
If M2= 1 then S2-H2
If M2= 2 then S2+H2
If M2= 3 then S2+D2
If M2= 4 then S2-D2
If M1= 1 and M2 = 2
then S1-H1
If M1 = 2 and M2 = 1
then S2-H2
If M1 = 3 and M2 = 2
>S2+H2, S2>LA, HA>S1+D1;
then S1
<HA, S1>LA;
<HA, S2>LA;
>LA;
<HA;
<HA;
>LA;
>LA;
<HA;
<HA;
>LA;
>S2+H2, S2>LA, HA>S1;
>S1+H1, S1>LA, HA>S2;
1514
If M1 = 2 and M2 = 3
then S1+H1
If M1 = 4 and M2 = 1
then S1
If M1 = 1 and M2 = 4
then S1–H1
If M1 = 3 and M2 = 4
then S1
If M1 = 4 and M2 = 3
then S2
For HI 21523 only:
If M1 = 1 then S1+D1
If M1 = 2 then S1–D1
where the minimum deviation (D1 or D2) is 0.5 pH (for HI
21) or 25 mV (for HI 22).
<S2, S1>LA, HA>S2+D2;
<S2–H2, S1–D1>LA, HA>S2;
>S2, S2–D2>LA, HA>S1;
>S2, S2–D2>LA, HA>S1+D1;
>S1, S1–D1>LA, HA>S2+D2;
<HA;
>LA;
NoteThe password cannot be viewed when SETUP is pressed with-
out entering the original password first. The default password
is set at “0000”. In the event that the user forgets the password, this can be reset to “0000” by pressing and holding
CFM and then pressing LCD and CAL DATA at the same time
when the pH controller is in normal operating mode (idle or
control with measurement displaying).
CONTROL MODE
The control mode is the normal operational mode for these
meters. During control mode HI 21 and HI 22 fulfill the
following main tasks:
• convert information from pH/ORP and temperature inputs
• control relays and generate the analog outputs as deter-
• RS485 management (if the feature is included).
In addition, HI 21 and HI 22 can log working data through
RS485 connection. This data includes:
• pH, mV and
• last calibration data;
• setup configuration (also from PC).
The status of the meter is shown by the LED’s on the right
to digital values;
mined by the setup configuration, display alarm condition;
o
C measured values;
STATUSLEDs
ControlAlarmAlarm LED (green) Relay LED (yellow)Red LED
OF F----O NO FFO N
ONOFFONON or OFFOFF
ONONOFFON or OFFBlinking
NoteWhen a wrong setup value is con-
firmed, the pH controller does not skip
to the next setup item but remains in
the current item displaying a flashing
“WRONG” indicator until the parameter value is changed by the user (the same is also true for
the setup code selection). In some circumstances, the user
cannot succeed in setting a parameter to a desired value if
the related parameters are not changed beforehand; e.g. to
set a pH high setpoint to 10.00 the high alarm must be set to
a value greater than pH 10.00 first.
NoteFor code numbers 40, 41, 42, the output is related to pH or mV units
depending on the model (pH process meters or mV process meters).
Items 41 and 42 are not available in HI 215YZ .
Meter exits control mode by pressing SETUP or CAL and confirming the password. Note that this command generates a
temporary exit. To deactivate the
control mode definitively, set
CONTROL ENABLE to “0” (item
# 02).
RELAY MODES
Once enabled, the relays 1 and 2 can be used in four different modes:
1) ON/OFF, high setpoint (acid dosage) (not for HI 21523);
2) ON/OFF, low setpoint (base dosage) (not for HI 21523);
3) Proportional or PID (HI 21523 only), low setpoint (base
dosage, if available);
4) Proportional or PID (HI 21523 only), high setpoint (acid
1716
An upper boundary is imposed for acid/base dosage time
when relays are energized continuously, i.e. when relay works
in ON/OFF mode or in PID mode but in the latter case only
if the relay is always ON. This parameter can be set through
setup procedure. When the maximum boundary is reached,
an alarm is generated; device stays in alarm condition until
relay is de-energized.
ON/OFF CONTROL MODE
Either for mode 1 or 2 (base or acid dosage) the user has to
define the following values through setup:
• relay setpoint (pH/mV value);
• relay hysteresis (pH/mV value).
Connect your device to the COM and
NO (Normally Open) or NC (Normally
Closed) terminals.
The ON relay state occurs when relay is energized (NO and
COM connected, NC and COM disconnected).
The OFF relay state occurs when relay is de-energized (NO
and COM disconnected, NC and COM connected).
The following graphs show relay states along with pH measured value (similar graph can be derived for mV control).
As shown below, a high setpoint relay is activated when the
measured pH exceeds the setpoint and is deactivated when it
is below the setpoint value minus hysteresis.
Such a behavior is suitable to control an acid dosing pump.
A low setpoint relay as can be seen from the following graphs
is energized when the pH value is below the setpoint and is
de-energized when the pH value is above the sum of setpoint
and the hysteresis. The low setpoint relay may be used to
control an alkaline dosing pump.
dosage, if available).
ON
OFF
Setpoint –
Hysteresis
Setpoint
14
ON
OFF
Setpoint
P.I.D. CONTROL MODE (HI 21523 ONLY)
PID control is designed to eliminate the cycling associated
with ON/OFF control in a rapid and steady way by means of
the combination of the proportional, integral and derivative
control methods.
With the proportional function, the duration of the activated
control is proportional to the error value (Duty Cycle Control
Mode): as the measurement approaches setpoint, the ON
period diminishes.
The following graph describes the pH process controller behavior. Similar graph may apply to the mV controller.
t
0
During proportional control the process controller calculates
the relay activation time at certain moments t
etc. The ON interval (the shaded areas) is then dependent to
the error amplitude.
With the integral function (reset), the controller will reach a
more stable output around the setpoint providing a more
accurate control than with the ON/OFF or proportional action only.
The derivative function (rate action) compensates for rapid
changes in the system reducing undershoot and overshoot of
the pH value.
During PID control, the ON interval is dependent not only to
the error amplitude but even to the previous measurements.
Definitely PID control provides more accurate and stable control than ON/OFF controllers and it is best suitable in system
with a fast response, quickly reacting to changes due to addition of acid or base solution.
t0+T
Setpoint +
Hysteresis
c
t0+2T
14
t0+3T
c
c
, t0+Tc, t0+2T
0
c
1918
An example of how the response overshoot can be improved with
a proper rate action setting is depicted in the following graphic.
pH
RATE ACTION COMPENSATES FOR RAPID CHANGESt
PID TRANSFER FUNCTION
The transfer function of a PID control is as follows:
with Ti = Kp/Ki, Td = Kd/Kp,
where the first term represents the proportional action, the
second is the integrative action and the third is the derivative
action.
Proportional action can be set by means of the Proportional
Band (PB). Proportional Band is expressed in percentage of the
input range and is related to Kp according to the following:
Kp = 100/PB.
Kp + Ki/s + s Kd = Kp (1 + 1/(s Ti) +s Td)
100%
Controller
output
In HI 21 and HI 22 the proportional action is set directly as
“Deviation” in pH and mV units respectively. Relation between Deviation (D) and PB is:
D = Range * PB/100
Each setpoint has a selectable proportional band: PB1 for
setpoint1 and PB2 for setpoint2. Two further parameters must
be provided for both setpoints:
Ti = Kp/Ki, reset time, measured in minutes
Td = Kd/Kp, rate time, measured in minutes.
Ti1 and Td1 will be the reset time and rate time for setpoint1,
while Ti2 and Td2 will be the reset time and the rate time for
setpoint2.
TUNING A PID CONTROLLER
The proportional, integrative, derivative terms must be tuned,
i.e. adjusted to a particular process. Since usually the process variables are not completely known, a “trial and error”
tuning procedure must be applied to get the best possible
control for the particular process. The target is to achieve a
fast response time and a small overshoot.
Many tuning procedures are available and can be applied
to HI 21. A simple and profitable procedure is reported in
this manual and can be used in almost all applications.
The user can vary five different parameters, i.e. the setpoint
(S1 or S2), the deviation (D1 or D2), the reset time, the rate
time and the proportional control mode period T
30 minutes).
NoteUser can disable the derivative and/or integrative action (for
P or PI controllers) by setting Td = 0 and/or Ti = MAX (Ti)
respectively through the setup procedure.
SIMPLE TUNING PROCEDURE
The following procedure uses a graphical technique of analyzing a process response curve to a step input.
NoteConnecting an external device (chart recorder or PC) to the
controller, the procedure is easier and doesn’t need the use
of hand plotting the process variable (pH or mV).
(from 1 to
c
0
Proportional Band
Error
2120
1. Starting from a solution with a pH or mV value different
from the dosed liquid (at least a 3 pH or 150 mV difference) turn on the dosing device at its maximum capacity
without the controller in the loop (open loop process). Note
the starting time.
2. After some delay the pH or mV starts to vary. After more
delay, the pH or mV will reach a maximum rate of change
(slope). Note the time that this maximum slope occurs and
the pH or mV value at which it occurs. Note the maximum
slope in pH or mV per minute. Turn the system power off.
3. On the chart draw a tangent to the maximum slope point until
intersection with the horizontal line corresponding to the initial
pH or mV value. Read the system time delay Tx on the time axis.
4. The deviation, Ti and Td can be calculated from the following:
• Deviation = Tx * max. slope (pH or mV)
• Ti = Tx / 0.4 (minutes)
• Td = Tx * 0.4 (minutes).
5. Set the above parameters and restart the system with the
controller in the loop. If the response has too much overshoot or is oscillating, then the system can be fine-tuned
slightly increasing or decreasing the PID parameters one
at a time.
Example:
the chart recording in the figure
aside was obtained continuously
dosing an alkaline solution to a
weak acid solution in a tank. The
initial settings will be:
Max. slope = 3 pH/5 mins = 0.6
pH/min
Time delay = Tx = approx. 7 mins
Deviation = Tx * 0.6 = 4.2 pH
Ti = Tx / 0.4 = 17.5 mins
Td = Tx * 0.4 = 2.8 mins
ALARM RELAYThe alarm relay functions in the following manner:
During alarm condition, the relay is de-energized. When not
in alarm condition, the relay is energized.
FS•C = NO (Normally Open)
Energized Relay
COM
FS•O = NC (Normally Closed)
De-energized Relay
Example:High alarm set at 10 pH
Low alarm set at 4 pH
An hysteresis will eliminate the possibility of continuous sequences ‘energizing/de-energizing’ of the alarm relay when
the measured value is close to the alarm setpoint. The hysteresis amplitude is 0.2 pH in HI 21 and 30 mV in HI 22.
Moreover the alarm signal is generated only after a user selectable time period (alarm mask) has elapsed since the
controlled value has overtaken one alarm threshold. This
additional feature will avoid fake or temporary alarm conditions.
NoteIf the power supply is interrupted, the relay is de-energized as
if in alarm condition to alert the operator.
In addition to the user-selectable alarm relays, all HI 21 and
HI 22 models are equipped with the Fail Safe alarm fea-
ture.
The Fail Safe feature protects the process against critical
errors arising from power interruptions, surges and human
errors. This sophisticated yet easy-to-use system resolves these
predicaments on two fronts: hardware and software. To eliminate problems of blackout and line failure, the alarm function
operates in a “Normally Closed” state and hence alarm is
triggered if the wires are tripped, or when the power is down.
2322
Loading...
+ 25 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.