ELECTRICAL SAFETY TESTER
USER MANUAL
MANUFACTURER ASSUMES NO LIABILITY
WARNING
IF UNIT OPERATED IN AN UNSAFE MANNER
WARNING
THIS INSTRUMENT GENERATES AND
DELIVERS A HAZARDOUSLY HIGH
VOLTAGE (5kV). BE EXTREMELY CAREFUL
WHEN USING THIS INSTRUMENT. BE SURE
TO READ MANUAL SECTION 3
PRECAUTIONS BEFORE OPERATING.
ELECTRICAL SAFETY TESTER
USER MANUAL
EC Declaration of Conformity
We
GOOD WILL INSTRUMENT CO., LTD.
No. 7-1, Jhongsing Rd., Tucheng City, Taipei County 236, Taiwan.
GOOD WILL INSTRUMENT (SUZHOU) CO., LTD.
No. 69 Lushan Road, Suzhou New District Jiangsu, China.
declares that the below mentioned products
are herewith confirmed to comply with the requirements set out in the Council
Directive on the Approximation of the Law of Member States relating to
Electromagnetic Compatibility (89/336/EEC,92/31/EEC) and Low Voltage
Equipment Directive (73/23/EEC, 93/68/EEC).
For the evaluation regarding the Electromagnetic Compatibility and Low Voltage
Equipment Directive, the following standards were applied:
EN 61326-1:Electrical equipment for measurement, control and laboratory use ––
Conducted and Radiated Emissions
EN 55011 Group I class A: 1998
Current Harmonic
EN 61000-3-2: 2000
Voltage Fluctuation
EN 61000-3-3: 1995
Surge Immunity
Conducted Susceptibility
Voltage Dips/ Interrupts
Low Voltage Equipment Directive 73/23/EEC & amended by 93/68/EEC
Safety Requirements
IEC/EN 61010-1: 2001
Remark: Also complied with Continuity of Protective Bonding Tester, Insulation
Resistance Test, Voltage Test, and Residual Voltage Test in accordance with
the Sub-Clauses 19.2, 19.3, 19.4 and 19.5 of EN 60204-1: 1997
5-1.Main display LCD..…………………………………….
5-2.Prepare the EST for Use….……………………………
5-3.Structure of storage steps…..………………………….
5-4.Menu parameter setup…………………………………
5-5.Functionality of remote interface……………………
6. MAINTENANCE………..………………………………….
6-1.Fuse rating and type.…………………………………...
6-2.Cleaning………………..………………………………..
16
16
19
29
31
39
40
40
40
The following symbols may appear in this manual or on
the product:
DANGER ATTENTION Protective Earth (ground) Frame or
High Voltage r efer to Manual Conductor Terminal Chassis
Terminal Terminal
USER MANUAL
WARNING. Warning statements identify condition or
practices that could result in injury or loss of life.
CAUTION. Caution statements identify conditions
or practices that could result in damage to this
product or other property.
The equipment shall not be used for measurements
within category II, III and IV.
iii
iv
ELECTRICAL SAFETY TESTER
USER MANUAL
ELECTRICAL SAFETY TESTER
USER MANUAL
FOR UNITED KINGDOM ONLY
NOTE: This lead/appliance must only be wired by competent
persons.
WARNING: THIS APPLIANCE MUST BE EARTHED
IMPORTANT: The wires in this lead are colo ured in a cco r da n c e
with the following code:
Green/ Yellow: Earth
Blue: Neutral
Brown: Live (Phase)
As the colours of the wires in main leads may not correspond
with the colours marking identified in your plug/appliance,
proceed as follows:
The wire which is coloured Green & Yellow must be connected
to the Earth terminal marked with the letter E or by the earth
symbol
or coloured Green or Green & Yellow.
The wire which is coloured Brown must be connected to the
terminal marked with the letter L or P or coloured Brown or
Red.
If in doubt, consult the instructions provided with the equipment
or contact the supplier.
This cable/appliance should be protected by a suitably rated and
approved HBC mains fuse: refer to the rating information on
the equipment and/or user instructions for details. As a guide,
cable of 0.75mm
2
should be protected by a 3A or 5A fuse. Larger
conductors would normally require 13A types, depending on the
connection method used.
Any moulded mains connector that requires removal
/replacement must be destroyed by removal of any fuse & fuse
carrier and disposed of immediately, as a plug with bared wires
is hazardous if a engaged in live socket. Any re-wiring mu st be
carried out in accordance with the information detailed on this
label.
The wire which is coloured Blue must be connected to the
terminal which is marked with the letter N or coloured Blue or
Black.
v
vi
ELECTRICAL SAFETY TESTER
USER MANUAL
1. PRODUCT INTRODUCTION
1-1. Description
The GPT/GPI-700A series Electrical Safety Testers (EST) are
designed for AD/DC Withstanding Voltage test, Insulation Resistance test
(IR) and Ground Bond test (GB) in order to provide a safe and accurate
test environment for the operator. With thoughtful design described in
1-2. Feature insures a safe operation of high voltage test and protect user
from hazardous impact. Besides, the series can be used together with
scanner box of HSB-001-1 & HSB-001-2.
The Electrical Safety Testers comply with the requirement of the
electrical equipment & appliance control ordinances and JIS, CSA, UL,
BS and other overseas standards as well. The testers can be used for
withstanding voltage test of the various types of electrical and equipment
and components.
The GPT/GPI-700A series are based on the family of GW
withstanding voltage tester including AC Withstanding Voltage test, DC
Withstanding Voltage test, Insulation Resistance test and Ground Bond
test (GB).
Function
Model
GPI-745A
GPI-735A
GPI-725A
GPT-715A
GPT-705A
AC DC IR GB
ˇ ˇ ˇ ˇ
ˇ ˇ ˇ
ˇ ˇ
ˇ ˇ
ˇ
ELECTRICAL SAFETY TESTER
USER MANUAL
1-2. Features
The GPT/GPI-700A series offer several other features:
1) No load setup of trip current and output voltage
A safe way to setup trip current and output voltage without high
voltage activated.
2) A large 24×2 character LCD with adjustable LED backlight
Provide a clear display about test parameters, which including
group, step, mode, status, output voltage, trip current and test
time.
3) Easily and quickly setup by front panel
A user-friendly interface provides user an easy and quick way to
set all parameters.
4) Electronic ramping and testing
Digital controlled ramping time and testing time.
5) Line and load regulation
Linear amplifier and feedback-control maintain output voltage
disregard of the variation of load.
6) Selectable output frequency
50/60 Hz is selectable by utility setup.
7) Adjustable ARC detect level
ARC detect level could be setup by utility setup.
8) 10 groups of storage and each group has 16 steps
Total 10 storage groups provided for testing of different products,
and each group has 16 steps.
9) Adjustable AC/DC test voltage during CAC/CDC test
CAC and CDC are special test modes in the “COM” step, in
which the output test voltage remains adjustable during test.
1
2
ELECTRICAL SAFETY TESTER
USER MANUAL
10) Flashing high voltage indicator
A flashing red LED indicates dangerous situation during high
voltage output is activated.
11) Data lock function
Front panel could be locked or unlocked by utility setup.
12) PLC remote control
The 9-pin interface provides inputs (START, RESET) and
outputs (TEST, PASS, FAIL).
13) Compact size with multi-capability
The compact size is geared with the Safety Test capabilities
including AC Withstanding test, DC Withstanding test,
Insulation Resistance test and Ground Bond test.
14) R/P Outp ut (GPI-745A only)
The output terminal located on the rear panel.
15) Scanner Interface
16) RS-232 and GPIB Interface
The RS-232 and GPIB interface enables remote control operation
and signal processing via a PC.
ELECTRICAL SAFETY TESTER
USER MANUAL
2. SPECIFICATION (15℃~35℃ RH≦75%)
1) AC Hi-Pot Specifications
Voltage Range 0.100~5.000kV
Voltage Step 5V/step
Voltage Regulation
(line & load)
Voltage Accuracy ±1% of reading ±10V (above 500V)
Current Sourcing*30~40mA(above 300V, maximu test time:
Current Limit 0.10~40mA, 0.02mA/step
Current Accuracy ±1% of reading ± 80μA
2) DC Hi-Pot Specifications (for GPT-715A/GPI-735A/745A only)
Voltage Range 0.100~6.000kV
Voltage Step 5V/step
Voltage Regulation
(line & load)
Voltage Accuracy ±1% of reading ±10V(above 500V)
Current Sourcing*0.10~10.00mA(above 300V, continuous
Current Limit 0.10~10mA, 0.01mA/step
Current Accuracy ±1% of reading ± 50μA
50V/100V:
1~50MΩ: ±5% of reading ±1 count
51~200MΩ: ±10% of reading ±1 count
Resistance Accuracy
201~2000MΩ: ±20% of reading±1 count
500V/1000V:
1~500MΩ: ±5% of reading ±1 count
501~2000MΩ: ±10% of reading±1 count
2001~9900MΩ: ±20% of reading±1 count
4) Ground Bond Specifications (for GPI-745A only)
Test Voltage Max. 8V(DC)
Current Range 3~42A
Current Resolution 0.02A
Current Accuracy
±1% of reading ±0.2A (3~8A)
±1% of reading ±50mA (9~42A)
Resistance Range
Resistance Resolution
Resistance Accuracy
0~600mΩ
0.1mΩ ±1% of reading ±3mΩ
Test Method F o u r Te r minals Test Mode
5) Continuity Check Specifications (for G P T-705A/715A/ GPI-725A/
735A only)
Output Current Range 0.1A(DC)
Compliance Voltage Range 10%
Resistance Resolution
0.001Ω
ELECTRICAL SAFETY TESTER
USER MANUAL
6) Ramp Time and T est T i me
AC Hi-Pot Ramp/Test time 000.0~999.9s Ramp/000.5~999.9 test
DC Hi-Pot Ramp/Test time 000.0~999.9s Ramp/000.5~999.9 test
Insulation Resistance Test
time
001.0~999.9s
Ground Bond Test Time 000.5~999.9s
Continuity Check Test Time 000.5~999.9s
7) ARC Detect
AC Detect Current 40 level (1~40mA)
DC Detect Current 10 level (1~10mA)
8) Storage
Groups
Steps
10
16
9) Interface
RS-232 Standard
GPIB Option
10) PLC Control
D-sub 9 pins female Standard
11) Scanner Interface
D-sub 9 pins female Standard
12) Real Plate Output
Hi-pot terminals and GB
Standard interface for GPI-745A only
terminals
5
6
ELECTRICAL SAFETY TESTER
USER MANUAL
13) General
Power Source
Operation
Environment
Storage temperature
& Humidity
Accessories
Dimension 446(D) × 330(W) × 149(H) (m/m)
Weigh
AC100V, 120V, 220V, 230V±10% 50/60Hz
Indoor use, altitude up to 2000m.
Ambient Temperature 0℃ to 40℃.
Relative Humidity 80% (Maximum).
Installation category II
Pollution Degree 2
-10℃ to 70℃.
70% (Maximum).
GHT-205 × 1,
GTL-115 × 1(only for GPI-745A),
Instruction manual × 1,
Interface Manual × 1
Approx. 14.9 kgs
* The main purpose provided by the series of instruments is for
Puncture Testing. The specification is not for continuous test. The
temperature of heat-sink is monitored. The test procedure will stop if
the heat-sink is too hot. Please refer to Figure 2-1.
* Stop the instrument for 10 minutes when continuously proceeding