GW Instek GPI-700A User Manual

Electrical Safety Tester
GPT/GPI-700A Series
USER MANUAL
GW INSTEK PART NO. 82PI-74500MG1
This manual contains proprietary information, which is protected by copyrights. All rights are reserved. No part of this manual may be photocopied, reproduced or translated to another language without prior written consent of Good Will company.
The information in this manual was correct at the time of printing. However, Good Will continues to improve products and reserves the rights to change specification, equipment, and maintenance procedures at any time without notice.
Good Will Instrument Co., Ltd.
No. 7-1, Jhongsing Rd., Tucheng City, Taipei County 236, Taiwan.
ISO-9001 CERTIFIED MANUFACTURER
ELECTRICAL SAFETY TESTER
USER MANUAL
MANUFACTURER ASSUMES NO LIABILITY
WARNING
IF UNIT OPERATED IN AN UNSAFE MANNER
WARNING THIS INSTRUMENT GENERATES AND DELIVERS A HAZARDOUSLY HIGH VOLTAGE (5kV). BE EXTREMELY CAREFUL WHEN USING THIS INSTRUMENT. BE SURE TO READ MANUAL SECTION 3 PRECAUTIONS BEFORE OPERATING.
ELECTRICAL SAFETY TESTER
USER MANUAL
EC Declaration of Conformity
We
GOOD WILL INSTRUMENT CO., LTD.
No. 7-1, Jhongsing Rd., Tucheng City, Taipei County 236, Taiwan.
GOOD WILL INSTRUMENT (SUZHOU) CO., LTD.
No. 69 Lushan Road, Suzhou New District Jiangsu, China.
declares that the below mentioned products
are herewith confirmed to comply with the requirements set out in the Council Directive on the Approximation of the Law of Member States relating to Electromagnetic Compatibility (89/336/EEC,92/31/EEC) and Low Voltage Equipment Directive (73/23/EEC, 93/68/EEC). For the evaluation regarding the Electromagnetic Compatibility and Low Voltage Equipment Directive, the following standards were applied:
EN 61326-1: Electrical equipment for measurement, control and laboratory use ––
Conducted and Radiated Emissions
EN 55011 Group I class A: 1998
Current Harmonic EN 61000-3-2: 2000 Voltage Fluctuation EN 61000-3-3: 1995 Surge Immunity
Conducted Susceptibility
Voltage Dips/ Interrupts
Low Voltage Equipment Directive 73/23/EEC & amended by 93/68/EEC
Safety Requirements
IEC/EN 61010-1: 2001
Remark: Also complied with Continuity of Protective Bonding Tester, Insulation
Resistance Test, Voltage Test, and Residual Voltage Test in accordance with
the Sub-Clauses 19.2, 19.3, 19.4 and 19.5 of EN 60204-1: 1997
GPT-705A/715A, GPI-725A/735A/745A
EMC requirements (1997+A1: 1998+A2:2001)
Electrostatic Discharge
EN 61000-4-2: 1995+A1 :1998
Radiated Immunity
EN 61000-4-3: 1996+A1 :1998
Electrical Fast Transients
EN 61000-4-4: 1995
EN 61000-4-5: 1995
EN 61000-4-6: 1996
EN 61000-4-11: 1994
i
ii
ELECTRICAL SAFETY TESTER
ELECTRICAL SAFETY TESTER
USER MANUAL
CONTENTS PAGE
1. PRODUCT INTRODUCTION............................................
1-1.Description…………………………..………………….
1-2.Features…………………………..……………………
1 1 2
SAFETY TERMS AND SYMBOLS
These terms may appear in this manual or on the product:
2. SPECIFICATION………………………………………….. 4
3. PRECAUTIONS BEFORE OPERATION……...………...
3-1.Unpacking the instrument……..…………….………... 3-2.Safety notice………………..…………………..………. 3-3.Environment……………………………………..……..
4. PANEL INTRODUCTION……………………..………….
4-1.Front Panel……………………………………………... 4-2.Rear Panel……………………………………….……...
11
12 12 13
9 9 9
5. OPERATION METHOD……..……………………………
5-1.Main display LCD..……………………………………. 5-2.Prepare the EST for Use….…………………………… 5-3.Structure of storage steps…..…………………………. 5-4.Menu parameter setup………………………………… 5-5.Functionality of remote interface……………………
6. MAINTENANCE………..………………………………….
6-1.Fuse rating and type.…………………………………... 6-2.Cleaning………………..………………………………..
16 16 19 29 31 39
40 40 40
The following symbols may appear in this manual or on the product:
DANGER ATTENTION Protective Earth (ground) Frame or High Voltage r efer to Manual Conductor Terminal Chassis Terminal Terminal
USER MANUAL
WARNING. Warning statements identify condition or practices that could result in injury or loss of life.
CAUTION. Caution statements identify conditions or practices that could result in damage to this product or other property.
The equipment shall not be used for measurements within category II, III and IV.
iii
iv
ELECTRICAL SAFETY TESTER
USER MANUAL
ELECTRICAL SAFETY TESTER
USER MANUAL
FOR UNITED KINGDOM ONLY
NOTE: This lead/appliance must only be wired by competent
persons.
WARNING: THIS APPLIANCE MUST BE EARTHED
IMPORTANT: The wires in this lead are colo ured in a cco r da n c e
with the following code:
Green/ Yellow: Earth Blue: Neutral Brown: Live (Phase)
As the colours of the wires in main leads may not correspond with the colours marking identified in your plug/appliance, proceed as follows:
The wire which is coloured Green & Yellow must be connected to the Earth terminal marked with the letter E or by the earth
symbol
or coloured Green or Green & Yellow.
The wire which is coloured Brown must be connected to the terminal marked with the letter L or P or coloured Brown or Red. If in doubt, consult the instructions provided with the equipment or contact the supplier.
This cable/appliance should be protected by a suitably rated and approved HBC mains fuse: refer to the rating information on the equipment and/or user instructions for details. As a guide, cable of 0.75mm
2
should be protected by a 3A or 5A fuse. Larger conductors would normally require 13A types, depending on the connection method used. Any moulded mains connector that requires removal /replacement must be destroyed by removal of any fuse & fuse carrier and disposed of immediately, as a plug with bared wires is hazardous if a engaged in live socket. Any re-wiring mu st be carried out in accordance with the information detailed on this label.
The wire which is coloured Blue must be connected to the terminal which is marked with the letter N or coloured Blue or Black.
v
vi
ELECTRICAL SAFETY TESTER
USER MANUAL
1. PRODUCT INTRODUCTION
1-1. Description
The GPT/GPI-700A series Electrical Safety Testers (EST) are
designed for AD/DC Withstanding Voltage test, Insulation Resistance test
(IR) and Ground Bond test (GB) in order to provide a safe and accurate
test environment for the operator. With thoughtful design described in
1-2. Feature insures a safe operation of high voltage test and protect user
from hazardous impact. Besides, the series can be used together with
scanner box of HSB-001-1 & HSB-001-2.
The Electrical Safety Testers comply with the requirement of the
electrical equipment & appliance control ordinances and JIS, CSA, UL,
BS and other overseas standards as well. The testers can be used for
withstanding voltage test of the various types of electrical and equipment
and components.
The GPT/GPI-700A series are based on the family of GW
withstanding voltage tester including AC Withstanding Voltage test, DC
Withstanding Voltage test, Insulation Resistance test and Ground Bond
test (GB).
Function
Model
GPI-745A GPI-735A
GPI-725A GPT-715A GPT-705A
AC DC IR GB
ˇ ˇ ˇ ˇ ˇ ˇ ˇ ˇ ˇ ˇ ˇ ˇ
ELECTRICAL SAFETY TESTER
USER MANUAL
1-2. Features
The GPT/GPI-700A series offer several other features:
1) No load setup of trip current and output voltage
A safe way to setup trip current and output voltage without high voltage activated.
2) A large 24×2 character LCD with adjustable LED backlight
Provide a clear display about test parameters, which including group, step, mode, status, output voltage, trip current and test time.
3) Easily and quickly setup by front panel
A user-friendly interface provides user an easy and quick way to
set all parameters.
4) Electronic ramping and testing
Digital controlled ramping time and testing time.
5) Line and load regulation
Linear amplifier and feedback-control maintain output voltage disregard of the variation of load.
6) Selectable output frequency
50/60 Hz is selectable by utility setup.
7) Adjustable ARC detect level
ARC detect level could be setup by utility setup.
8) 10 groups of storage and each group has 16 steps
Total 10 storage groups provided for testing of different products, and each group has 16 steps.
9) Adjustable AC/DC test voltage during CAC/CDC test
CAC and CDC are special test modes in the “COM” step, in which the output test voltage remains adjustable during test.
1
2
ELECTRICAL SAFETY TESTER
USER MANUAL
10) Flashing high voltage indicator
A flashing red LED indicates dangerous situation during high voltage output is activated.
11) Data lock function Front panel could be locked or unlocked by utility setup.
12) PLC remote control
The 9-pin interface provides inputs (START, RESET) and outputs (TEST, PASS, FAIL).
13) Compact size with multi-capability
The compact size is geared with the Safety Test capabilities including AC Withstanding test, DC Withstanding test, Insulation Resistance test and Ground Bond test.
14) R/P Outp ut (GPI-745A only)
The output terminal located on the rear panel.
15) Scanner Interface
16) RS-232 and GPIB Interface
The RS-232 and GPIB interface enables remote control operation
and signal processing via a PC.
ELECTRICAL SAFETY TESTER
USER MANUAL
2. SPECIFICATION (15℃~35℃ RH≦75%)
1) AC Hi-Pot Specifications
Voltage Range 0.100~5.000kV
Voltage Step 5V/step
Voltage Regulation (line & load) Voltage Accuracy ±1% of reading ±10V (above 500V)
Current Sourcing 30~40mA(above 300V, maximu test time:
Current Limit 0.10~40mA, 0.02mA/step Current Accuracy ±1% of reading ± 80μA
2) DC Hi-Pot Specifications (for GPT-715A/GPI-735A/745A only)
Voltage Range 0.100~6.000kV
Voltage Step 5V/step
Voltage Regulation (line & load) Voltage Accuracy ±1% of reading ±10V(above 500V)
Current Sourcing 0.10~10.00mA(above 300V, continuous
Current Limit 0.10~10mA, 0.01mA/step Current Accuracy ±1% of reading ± 50μA
1% + 5V
360 sec)
0.10~29.99mA(above 300V, continuous test)
0.10~10mA(below 300V, continuous test)
1% + 5V
test)
0.10~4mA(below 300V, continuous test)
3
4
ELECTRICAL SAFETY TESTER
USER MANUAL
3) Insulation Resistance Specifications (for GPI-725A/ 735A/ 745A only)
DC Voltage
50V/100V/500V/1000V
50V/100V: 1~50MΩ: ±5% of reading ±1 count 51~200MΩ: ±10% of reading ±1 count
Resistance Accuracy
201~2000MΩ: ±20% of reading±1 count 500V/1000V: 1~500MΩ: ±5% of reading ±1 count 501~2000MΩ: ±10% of reading±1 count 2001~9900MΩ: ±20% of reading±1 count
4) Ground Bond Specifications (for GPI-745A only)
Test Voltage Max. 8V(DC)
Current Range 3~42A
Current Resolution 0.02A
Current Accuracy
±1% of reading ±0.2A (3~8A) ±1% of reading ±50mA (9~42A)
Resistance Range
Resistance Resolution
Resistance Accuracy
0~600mΩ
0.1mΩ ±1% of reading ±3mΩ
Test Method F o u r Te r minals Test Mode
5) Continuity Check Specifications (for G P T-705A/715A/ GPI-725A/ 735A only)
Output Current Range 0.1A(DC)
Compliance Voltage Range 10%
Resistance Resolution
0.001Ω
ELECTRICAL SAFETY TESTER
USER MANUAL
6) Ramp Time and T est T i me
AC Hi-Pot Ramp/Test time 000.0~999.9s Ramp/000.5~999.9 test
DC Hi-Pot Ramp/Test time 000.0~999.9s Ramp/000.5~999.9 test
Insulation Resistance Test time
001.0~999.9s
Ground Bond Test Time 000.5~999.9s
Continuity Check Test Time 000.5~999.9s
7) ARC Detect
AC Detect Current 40 level (1~40mA)
DC Detect Current 10 level (1~10mA)
8) Storage
Groups
Steps
10
16
9) Interface
RS-232 Standard
GPIB Option
10) PLC Control
D-sub 9 pins female Standard
11) Scanner Interface
D-sub 9 pins female Standard
12) Real Plate Output
Hi-pot terminals and GB
Standard interface for GPI-745A only
terminals
5
6
ELECTRICAL SAFETY TESTER
USER MANUAL
13) General
Power Source
Operation Environment
Storage temperature & Humidity
Accessories
Dimension 446(D) × 330(W) × 149(H) (m/m)
Weigh
AC100V, 120V, 220V, 230V±10% 50/60Hz
Indoor use, altitude up to 2000m. Ambient Temperature 0 to 40℃.
Relative Humidity 80% (Maximum).
Installation category II Pollution Degree 2
-10 to 70℃. 70% (Maximum). GHT-205 × 1, GTL-115 × 1(only for GPI-745A),
Instruction manual × 1, Interface Manual × 1
Approx. 14.9 kgs
The main purpose provided by the series of instruments is for
Puncture Testing. The specification is not for continuous test. The temperature of heat-sink is monitored. The test procedure will stop if the heat-sink is too hot. Please refer to Figure 2-1.
Stop the instrument for 10 minutes when continuously proceeding
30~40mA withstanding test for one hour.
ELECTRICAL SAFETY TESTER
mA
40
35
30
25
20 15
10
6
4 2
0
100 200 300 400 500 1000 2000 3000 4000 5000 6000 V
USER MANUAL
AC Hi-Pot
Spec 180sec
AC Hi-Pot Spec
Continuous
DC Hi-Pot Spec
Continuous
Figure 2-1
7
8
ELECTRICAL SAFETY TESTER
USER MANUAL
3. PRECAUTIONS BEFORE OPERATION
3-1. Unpacking the instrument
The product has been fully inspected and tested before shipping from the factory. Upon receiving the instrument, please unpack and inspect it to check if there is any damage caused during transportation. If any sign of damage is found, notify the bearer and/or the dealer immediately.
3-2. Safety Notice
Working place
The working place must be isolated and when the high voltage is proceeding, it should be alerted with a warning sign of dangerous for special caution.
Checking the Line Voltage
The instruments can be applied with any kind of line voltage shown in the table below. Before connecting the power plug to an AC line outlet, make sure the voltage selector of the rear panel is set to the correct position corresponding to the line voltage. It might be damaged the instrument if connected to the wrong AC line voltage.
WARNING. To avoid electrical shock the power cord protective grounding conductor must be connected to ground.
The equipment shall not be used for measurements within category II, III and IV.
ELECTRICAL SAFETY TESTER
USER MANUAL
When line voltages are changed, replace the required fuses shown as below:
Line voltage Range
100V 120V
Operator’s Precaution
(1) With immense high output voltage and current of the puncture
tester, only qualified person can operate the tester in order to avoid fatal electric shock.
(2) On-job training is required for operator to better use the tester
smoothly and safely.
(3) The operator is prohibited to dress with metal ornaments or wear
metal decoration in order to avoid electric shock.
(4) The person with cardiac or wear a pacemaker must not to operate
the tester.
90-110V
108-132V
WARNING. To avoid personal injury, disconnect the power cord before removing the fuse holder.
WARNING: This is a Class A product. In a domestic environment this product may cause radio interference in which case the user may be required to take adequate measures.
Fuse Line voltage Range Fuse
T 7.0A
250V
220V 230V
198-242V 207-250V
T 4A
250V
9
10
ELECTRICAL SAFETY TESTER
USER MANUAL
Secure Testing
Never operate the tester in the place with electric circuit device around. The earth lead should be well connected in accordance with instruction. The Return Lead has to be connected to the tested object first before linking up test probe. Connect the test probe to the high voltage output terminal ONLY when conducting tests. Otherwise, keep the probe disconnected from the tester. Also, do not touch the electric conductor of test probe and the operator has to fully control the power on/off by using switch or remote control, which should not be lay aside carelessly.
WARNING: During the testing, do not touch the tested object or any other connec t ed objects.
ELECTRICAL SAFETY TESTER
USER MANUAL
4. PANEL INTRODUCTION
4-1. Front Panel
3-3.Environment
The normal ambient temperature range of this instrument is from 0° to 40°C (32° to 104°F). To operate the instrument over this specific temperature range may cause damage to the circuits. Do not use the instrument in a place where strong magnetic or electric field exists as it may disturb the measurement.
11
12
ELECTRICAL SAFETY TESTER
USER MANUAL
4-2. Rear Panel
ELECTRICAL SAFETY TESTER
USER MANUAL
1 Model Number Model number and description 2 FAIL Indicator LED The red LED indicates failure of test procedure 3 PASS Indicator
The green LED indicates pass of test procedure
LED
4 CAUTION
Indicator LED
During test the red LED will flash to indicate
dangerous. 5 Main Display LCD The LCD displays all message about test procedure. 6 START Button Press the green button to start a test procedure. 7 RESET Button Press the red button to reset/stop a test procedure.
8 MENU Key
When you press the MENU key, the status becomes
MENU and you can browse all groups.
When you press the EDIT/SAVE key, the status EDIT 9 EDIT/SAVE Key
is active and you can edit this step or setup. Press the
EDIT/SAVE key again will save this step or setup.
When you press the UTILITY key, the status 10 UTILITY Key
UTILITY is active and you can view all the utility
setups.
11 FIELD Key
When you edit the test step, press the FIELD key to
change the active parameter of stop. 12 Left Arrow Key Press the arrow key to adjust knob’s resolution. 13 Right Arrow Key Press the arrow key to adjust knob’s resolution.
If status EDIT is active, turn the knob to increase or
14 Knob
decrease the value of active parameter.
If status MENU is active, turn the knob to increase or
decrease active Step. 15 LCD Backlight
Turn the VR to adjust the LED backlight of LCD.
Adjustment
16 Buzzer Volume
Turn the VR to adjust the buzzer volume.
Adjustment
17 High Voltage Output
High voltage output terminal.
Seat
18 SOURCE-Terminal
(only for Ground
High current terminal for Ground Bond test.
Bond Test)
13
14
ELECTRICAL SAFETY TESTER
USER MANUAL
19 SOURCE+ Terminal
(only for Ground
Bond Test) 20 Power Switch Press the power switch to turn on the tester. 21 SENSE+ Terminal Voltage Terminal for Ground Bond test. 22 SENSE- & Return
Termi nal
23 Ground Terminal Connect Ground terminal to the earth ground. 24 Fuse Holder with
Voltage Selector 25 AC Inlet Connect the AC power line to the inlet. 26 Remote Interface The remote interface performs all the functions of
27 RS232 Terminal D-SUB 9 pin connector, Input/Output connector
28 GPIB Terminal Blue 24 connector, Input/Output connector for
29 Scanner Interface D-sub 9 pins female connector for scanner box.
High Voltage Output
30
on rear panel
Sense + Terminal on
31
rear panel
Source + Terminal
32
on rear panel
Source - Terminal
33
on rear panel
Sense - Terminal &
34
Return Terminal on
rear panel
*The instrument can be used together with scanner box of HSB-001-1 &
HSB-001-2.
*The Instrument can control one scanner box only.
High current terminal for Ground Bond test.
SENSE- Terminal is a voltage terminal for Ground Bond test, and Return Terminal is for all test.
To change AC source voltage, pull the fuse holder and rotate it to the proper value.
PLC control.
for RS232.
IEE-488.
High voltage output terminal.
Voltage terminal for Ground Bond test.
High current terminal for Ground Bond test.
High current terminal for Ground Bond test.
Sense - terminal is a voltage terminal for Ground Bond test, Return terminal is for all tests.
ELECTRICAL SAFETY TESTER
USER MANUAL
5. OPERATION METHOD
5-1. Main Display LCD
Storage Mode Output Voltage/Current ARC Status
1 ~ 0 1 A C I max=0
Measurement Limit Ramp/Test Time
WV=5
1.00m
ATEST:000 . 0 s
.000kV*R E A D Y
Table of parameters
Group/ Step: There are total 10 groups, and each group has 16
steps.
Storage
Mode
The first number represents group while the second number represents step. Ex. 3:01
3 stands for group number, 01 stands for step number.
The test mode of tester includes:
ACW: AC Withstanding voltage test.
DCW: DC Withstanding voltage test (only for GPT-715A/
GPI-735A/745A).
I R : Insulation Resistance test (only for GPI-725A/GPI-735A
/745A).
G B : Ground Bond test (only for GPI-745A).
C N T : Continuity Check test (only for GPT-705A/715A/
GPI-725A/735A)
The total types of mode will change for different model.
15
16
ELECTRICAL SAFETY TESTER
USER MANUAL
Output voltage or current for each step
A C : Output voltage (0.100~ 5.000 kV)
Output Voltage/Current
D C : Output voltage (0.100~ 6.000 kV) I R : Output voltage (50V/100V/500V/1000V) G B : Output current (3.00~42.00A) C N T : Output current (100mA) The status of tester includes: M E N U : Browse and check steps of test. E D I T : Edit parameters S A V E : Save parameters ERROR : The setting parameters are incorrect whilst saving. U T I L : Browse and check system utility.
Status
R E AD Y: Ready for test TEST: Testing P A S S : The result of test is pass F A I L : The result of test is fail S T O P : Stop the test R M T : Tester is under remote operation with a PC E r r : Tester receives an incorrect command from a PC
ARC
Measurement Limit
If the ARC function is enable, the sign “” means that there is ARC during test. Lower and upper limit of measurement I m a x / I m i n : Current measurement limit (ACW & DCW)
R m a x / Rm i n : Resistance measurement limit(IR&GB&CNT) Ramp/Test Time
Ramp time and test time
A C : Ramp (000.0~999.9s)
Test (000.5~999.9s)
D C : Ramp (000.0~999.9s)
Test (000.5~999.9s) I R : Test (001.0~999.9s) G B : Test (000.5~999.9s) C N T : Test (000.5~999.9s)
ELECTRICAL SAFETY TESTER
USER MANUAL
CHANNEL Hi: 00 Lo: 00
Set the Scanner Box to output and return function through the setting of Hi and Lo. If the setting is Hi=01 and Lo=02, that means the high voltage is output from CH1 and return from CH2. If Hi and Lo are set to 00, that means no signal is sent to Scanner Box.
When the measured load becomes capacitive, the test time must be
extended.
Ramp Time
Ramp Time setting provides a fast and safe way to detect short circuit or faulty wiring of your Device Under Test.
GPT/GPI-700A monitors the voltage level during the first stage of Ramp Time to find out short condition. If the voltage still remains less than 50V after 0.1s, GPT/GPI-700A cuts off the output and displays the message “Short”. Therefore the short current will be shut off before it grows beyond Imax, the current limit setting.
When checking GPT/GPI-700A current accuracy, if the DUT has less than 2Mohm input impedance, turn off the Ramp Time setting. GPT/GPI-700A may shut down the voltage taking the low impedance as a sign of short circuit.
voltage
time
ramp
time
test
17
18
0 - 0 0
A C
ELECTRICAL SAFETY TESTER
USER MANUAL
5-2. Prepare the EST for Use
z To view the Storage Steps
1. Press the MENU key to enter status MENU.
ELECTRICAL SAFETY TESTER
USER MANUAL
z To Edit/Save the Storage St eps
1. Follow the above procedure “To View the Storage Steps “ to select
a step.
2. Press the EDIT/SAVE key to enter status EDIT.
M E N U
E D I T
MENU
2. Use the left and right arrow keys to change knob’s resolution (group or step). Use the knob to change the active step.
EDIT/SAVE
3. Use knob to adjust parameter. Use arrow keys to change knob’s
resolution.
A C W
19
20
ELECTRICAL SAFETY TESTER
USER MANUAL
4. Use FIELD key to change active parameter.
V = 2 . 0 0 0 k V
ELECTRICAL SAFETY TESTER
USER MANUAL
z To Begin a Group Test
1. Repeat the procedure “To View the Storage Steps” to select a step.
2. Press RESET button to enter status READY.
R E A D Y
5. Repeat step 3 and 4 to adjust parameter.
6. After setting all parameters, press EDIT/SAVE key to save the step.
The status will become SAVE. After the step is saved, the status will return to EDIT.
FIELD
S A V E
EDIT/SAVE
7. Repeat the procedure “To View the Storage Steps” to select another step.
RESET
3. Make sure the test environment is safe.
4. Press START button to start the test while the status TEST is active
and the CAUTION LED flashes.
T E S T
START
21
22
P A S S
ELECTRICAL SAFETY TESTER
USER MANUAL
5. If you press RESET button the test will stop immediately.
ELECTRICAL SAFETY TESTER
USER MANUAL
7. If the result fails, the FAIL LED will be activated and the buzzer
will alarm operator. To stop the alarm, press RESET button again.
RESET
6. If the result is pass, the PASS LED will be active.
S T O P
RESET
8. Use knob to view the result of group step by step.
1 - 0 0
F A I L
23
24
T E S T F R O M
S T E P
1
ELECTRICAL SAFETY TESTER
USER MANUAL
z To View the System Utility
1. Press the UTILITY key to enter status UTIL.
U T I L
UTILITY
2. Use the knob or arrow keys to change the active parameter.
G R O U P
ELECTRICAL SAFETY TESTER
USER MANUAL
Table of system utility:
Parameter Option Description
From STEP 1 The group test procedure always begins from
TEST MODE
ARC MODE
ARC CURRENT AC FREQUENCY
TEST CONTROL MODE
DATA LOCK
CONTI. CALIBRATION
IR TEST MODE
ZERO CHECK (GB only)
Fail Set
From the present step DISABLE Disable the function arc detection. ENABLE & STOP ENABLE & CONTINUE Set the current level of arc detection. 50 Hz Set the AC hi-pot output frequency to 50 Hz. 60 Hz Set the AC hi-pot output frequency to 60 Hz. MODE 1: Double Click
MODE 2: Start Directly MODE 3: PLC enable MODE 4 : Reserved UNLOCKED Accept and save all parameters of test step
LOCKED Refuse to change any parameters of test step
Timer Reach the time of determination for pass or
Stop on Pass Stop on pass at the DUT detection. Stop on Fail Stop on fail at the DUT detection.
Zero check by shorting the resistance of test leads only for ground bond test.
Continue The operation will not stop until all 16
step 1 to end of group. e.g. 01~01 The group test procedure always begins from the step selected to end of group. e.g. 01~03
Enable the arc detection and stop the test when arc is active. Enable the arc detection and continue the test when arc is active.
Control mode of front panel. Mode 1: Reset first (press reset button twice before test) Mode 2: Press start button directly.
Mode 3: REMOTE I/O enable (the start button is disable) Mode 4: Reserved.
and utility.
and utility. Calibrate the short resistance of test leads for test of Continuity check.
fail.
group steps have been tested.
25
26
F R O M
ELECTRICAL SAFETY TESTER
USER MANUAL
Stop The operation will stop upon the failure
occurred at any step of the group tests.
Max off Holding maximum (resistance or current)
Measure Max
All group clear
Ramp Short Detect
Interface
z To Edit/Save the System Utility
1. Follow the above procedure “To View the System Utility” to select
a parameter.
2. Press the EDIT/SAVE key to enter status EDIT.
Max on Holding maximum (resistance or current)
Clear Group data Sure? Prompt to make sure of the decision to clear
Wait……. Erasing data. Short on
Short off RS-232 Baud
Rate GPIB Address 00~31
measurement off
measurement on.
Clears all the group and step data
all data.
Ramp short detection on. Ramp short detection off.
1200, 2400, 4800, 9600.
E D I T
EDIT/SAVE
3. Use knob to adjust parameter. Use arrow keys to change knob’s
resolution.
ELECTRICAL SAFETY TESTER
USER MANUAL
T E S T M O D E
4. After setting this parameter, press EDIT/SAVE key to save the
parameter. The status will become SAVE. After the parameter is saved, the status will return to EDIT.
S A V E
EDIT/SAVE
5. Repeat the procedure “To View the System Utility” to select another
parameter.
S T E P
27
28
ELECTRICAL SAFETY TESTER
USER MANUAL
5-3. Structure of storage steps
The storage steps of EST are total 10 groups (group 0 ~ group 9), 16
steps (step 01 ~ step 16) for each group. Except these steps, there is another step “COM” for special test. The presentation of storage steps is Group: Step. The first number represents group while the second represents step.
Group0 Group1 Group 2 Group3 Group4 Group
Step 1 0:01 1:01 2:01 3:01 4:01 5:01 9:01
Step 2 0:02 1:02 2:02 3:02 4:02 5:02 9:02
Step 3 0:03 1:03 2:03 3:03 4:03 5:03 9:03
Step 4 0:04 1:04 2:04 3:04 4:04 5:04 9:04
Step 5
Step 16
0:05
0:16
1:05
1:16
2:05
2:16
3:05
3:16
4:05
4:16
5:05
5:16
The special test step “COM” contains another two types of test: CAC and CDC. The function of CAC (CDC) is as same as ACW (DCW), except that users still can adjust the output voltage during test.
Group
~
5
9
9:05
9:16
ELECTRICAL SAFETY TESTER
USER MANUAL
After each step has been tested, the tested status will be shown on the screen:
Step123456789012345 6
TestPPPPPPPPPPPPPPP P
N: Empty P: Pass F: Fail Step 1: CH1 Step 2: CH2 Step 3: CH3 Step 4: CH4 Step 5: CH5 Step 6: CH6 Step 7: CH7 Step 8: CH8 Step 9: CH9 Step 0: CH10 Step 1: CH11 Step 2: CH12 Step 3: CH13 Step 4: CH15 Step 5: CH15 Step 6: CH16
29
30
ELECTRICAL SAFETY TESTER
USER MANUAL
5-4. Menu parameter setup
z AC/DC withstanding voltage test (ACW, DCW— only for
GPT-715A /GPI-735A/745A)
Press MENU key to enter status MENU then use knob and arrow keys to select a step. Press EDIT/SAVE key to enter status EDIT. Now the cursor stays at the “test mode” field. Use the knob to select mode ACW (DCW).
Functionality of FIELD key:
Test mode
ACW(DCW)
Output voltage
V=X.XXX kV
Measurement upper limit
Imax=XX.XX mA
Measurement lower limit
Imin=XX.XX mA
Ramping time
RAMP=XXX.XS
Testing time
TEST=XXX.XS
Channel
Hi=00 Lo=00
ELECTRICAL SAFETY TESTER
USER MANUAL
Press FIELD key to edit next field “output voltage”. Use the knob to adjust the desired output voltage while use the arrow keys to adjust the knob’s resolution. (0.100~5.000kV for ACW, 0.100~
6.000kV for DCW) Press FIELD key again to enter next field “measurement upper limit”. Use the knob to adjust the desired upper limit of leakage current while use the arrow keys to adjust the knob’s resolution. (0.10~40mA for ACW, 0.10~10mA for DCW) Press FIELD key again to enter next field “measurement lower limit”. Use the knob to adjust the desired lower limit of leakage current while use the arrow keys to adjust the knob’s resolution (0.10~40mA for ACW, 0.10~10mA for DCW). Press FIELD key again to enter next field “ramping time”. Use the knob to adjust the desired ramping time while use the arrow keys to adjust the knob’s resolution (0~999.9s). Press FIELD key again to enter next field “testing time”. Use the knob to adjust the desired testing time while use the arrow keys to adjust the knob’s resolution (0.5~999.9s). Press FIELD key again to enter next field “channel”. Use the knob to adjust the desired channel while use the arrow keys to adjust the knob’s resolution (00~16). [Note: When the Scanner Box is disconnected, the channel selected menu would not exist.] Press FIELD key again to return the first field “test mode” again. Press EDIT/SAVE key to save all the parameters.
AC/DC withstanding voltage Test Result Message.
I= “ShortmA” means the DUT current is over the metering range of ACW/DCW. The output could be short or over 40mA(ACW) or 10mA(DCW).
31
32
ELECTRICAL SAFETY TESTER
USER MANUAL
z Continuous AC/DC withstanding voltage test (CAC for all models,
CDC-only for GPT-715A/GPI-735A/745A)
The CAC/CDC test is available only on step “COM”. Like the traditional hi-pot tester, you can use the knob and arrow keys to adjust output voltage during test. All the parameters of CAC/CDC are the same as ACW/DCW, except the testing time. The testing of CAC/CDC is not limited.
z Insulation Resistance test (IR- only for GPI-725A/735A/745A)
Press MENU key to enter status MENU then use knob and arrow keys to select a step. Press EDIT/SAVE key to enter status EDIT. Now the cursor stays at the “test mode” field. Use the knob to select mode IR.
Functionality of FIELD key:
Test mode
IR
Output voltage
V=XXXX V
Measurement upper limit
Rmax=XXXX MΩ
Measurement lower limit
Rmin=XXXX MΩ
Testing time
TEST=XXX.XS
Channel
Hi=00
Lo=00
ELECTRICAL SAFETY TESTER
USER MANUAL
Press FIELD key to edit next field “output voltage”. Use the knob to adjust the desired output voltage (50V/100V/500V/1000V). Press FIELD key again to enter next field “measurement upper limit”. Use the knob to adjust the desired upper limit of resistance while use the arrow keys to adjust the knob’s resolution. (0~9900M@500V/1000V, 0~2000MΩ@50V/100V). Press FIELD key again to enter next field “measurement lower limit”. Use the knob to adjust the desired lower limit of resistance while use the arrow keys to adjust the knob’s resolution. (0~9900M@500V/1000V, 0~2000MΩ@50V/100V) Press FIELD key again to enter next field “testing time”. Use the knob to adjust the desired testing time while use the arrow keys to
adjust the knob’s resolution (1.0~999.9s). Press FIELD key again to enter next field “channel”. Use the knob to adjust the desired channel while use the arrow keys to adjust the knob’s resolution (00~16). [Note: When the Scanner Box is disconnected, the channel selected menu would not exist.]
Press FIELD key again to return the first field “test mode” again.
Press EDIT/SAVE key to save all the parameters.
33
34
ELECTRICAL SAFETY TESTER
USER MANUAL
z Ground Bonding Test (GB- only for GPI-745A)
Press MENU key to enter status MENU then use knob and arrow keys to select a step. Press EDIT/SAVE key to enter status EDIT. Now the cursor stays at the “test mode” field. Use the knob to select mode GB.
Functionality of FIELD key:
Test mode
GB
Output voltage
I=XX.XX A
Measurement upper limit
Rmax= XXX.X mΩ
Measurement lower limit
Rmin=XXX.X mΩ
Testing time
TEST=XXX.XS
Channel
Hi=00 Lo=00
ELECTRICAL SAFETY TESTER
USER MANUAL
Press FIELD key to edit next field “output current”. Use the knob to adjust the desired output current (3.00~42.00A). Press FIELD key again to enter next field “measurement upper limit”. Use the knob to adjust the desired upper limit of resistance while use the arrow keys to adjust the knob’s resolution (0~620.0mΩ).
Press FIELD key again to enter next field “measurement lower limit”. Use the knob to adjust the desired lower limit of resistance while use the arrow keys to adjust the knob’s resolution (0~620.0mΩ).
Press FIELD key again to enter next field “testing time”. Use the knob to adjust the desired testing time while use the arrow keys to
adjust the knob’s resolution (0.5~999.9s). Press FIELD key again to enter next field “channel”. Use the knob to adjust the desired channel while use the arrow keys to adjust the knob’s resolution (00~16). [Note: When the Scanner Box is disconnected, the channel selected menu would not exist.]
Press FIELD key again to return the first field “test mode” again.
Press EDIT/SAVE key to save all the parameters.
When the Scanner Box is disconnected, the channel selected menu
would not exist. When set channel to Hi:00 and Lo:10, the test
signal is sent to the instrument only, will not be sent to scanner.
35
36
ELECTRICAL SAFETY TESTER
USER MANUAL
z Continuity Check (Cnt)
Press MENU key to enter status MENU then use knob and arrow keys to select a step. Press EDIT/SAVE key to enter status EDIT. Now the cursor stays at the “test mode” field. Use the knob to select mode Cnt. The output current is fixed to 100mA.
Functionality of FIELD key:
Press FIELD key to enter next field “measurement upper limit”. Use the knob to adjust the desired upper limit of resistance while use the arrow keys to adjust the knob’s resolution (0~1.500Ω).
Press FIELD key to enter next field “measurement lower limit”. Use the knob to adjust the desired lower limit of resistance while use the arrow keys to adjust the knob’s resolution (0~1.500Ω).
Press FIELD key again to enter next field “testing time”. Use the knob to adjust the desired testing time while use the arrow keys to adjust the knob’s resolution (0.5~999.9s). Press FIELD key again to return the first field “test mode” again. Press EDIT/SAVE key to save all the parameters.
Test mode
Cnt
Measurement upper limit
Rmax=X.XXX Ω
Measurement lower limit
Rmin=X.XXX Ω
Testing time
TEST=XXX.XS
ELECTRICAL SAFETY TESTER
USER MANUAL
5-5. Functionality of remote interface
The remote interface provides two inputs (START and RESET) and three outputs (TESTING, PASS and FAIL).
TESTING 1 TESTING 2
RESET: Connect “RESET”(pin 1) to “COMMON”(pin 3)
START: In status READY, connect “START”(pin 2) to
TESTING 1, During test, the “TESTING 1”(pin 4) and TESTING 2 TESTING 2”(pin 5) will be shorted.
PASS 1, PASS 2 If the result of test is “pass”, the “PASS 1”(pin 6)
FAIL 1, FAIL 2 If the result of test is “fail”, the “FAIL 1”(pin 8) and
*The max. power rating of PASS/FAIL/TESTING outputs:
FAIL 2 FAIL 1
will reset or interrupt this machine (as same as the functionality of RESET button at front panel).
“COMMON”(pin 3) will start the test procedure (as same as the functionality of START button at front panel).
and “PASS 2”(pin 7) will be shorted.
“FAIL 2”(pin 9) will be shorted.
Apply voltage: +/-300V Carry current:: +/-100mA Power dissipation: 400mW
COMMON START RESET PASS 1 PASS 2
37
38
ELECTRICAL SAFETY TESTER
USER MANUAL
REMOTE CONTROLLER
RESET (pin 1)
START (pin 2)
COMMON (pin 3)
TESTING1 (pin 4)
TESTING2 (pin 5)
PASS1 (pin 6)
PASS2 (pin 7)
FAIL1 (pin 8)
FAIL2 (pin 9)
The function key of front panel can not be used when any two pins of the ResetStart or Common function of remote control appear
short circuit. Please check the remote control device.
EST REMOTE INTERFACE
ELECTRICAL SAFETY TESTER
USER MANUAL
6. MAINTENANCE
The following instructions are used by qualified person only to avoid electrical shock, do not perform any service other than contained in the operation instructions unless you are qualified to do so.
6-1. Fuse Rating and type
If the fuse blows, the product will not operate. Try to determine and correct the cause of the blown fuse, then replace the fuse with correct rating and type shown as below:
Line
voltage
100V 120V
90-110V
108-132V
Range
Fuse Line voltage Range
T 7.0A
250V
220V 230V
198-242V 207-250V
Fuse
T4A
250V
WARNING: For continued fire protection, replace only with 250V fuse of the specified type and rating, and disconnect the power cord before proceeding fuse replacement.
6-2. Cleaning
To keep the instrument clean, wipe the case with a damp cloth and detergent. Do not use abrasives or solvents.