GW-instek GPE-3060, GPE-6030 User Manual

Page 1
Multi-output DC Power Supply
GPE-3060/GPE-6030
USER MANUAL
GW INSTEK PART NO. 82GP360300301
ISO-9001 CERTIFIED MANUFACTURER
Page 2
This manual contains proprietary information, which is protected by copyrights. All rights are reserved. No part of this manual may be photocopied, reproduced or translated to another language without prior written consent of Good Will company.
The information in this manual was correct at the time of printing. However, Good Will continues to improve products and reserves the rights to change specification, equipment, and maintenance procedures at any time without notice.
Good Will Instrument Co., Ltd. No. 7-1, Jhongsing Rd., Tucheng Dist., New Taipei City 236, Taiwan
Page 3
TABLE OF CONTENTS
Table of Contents
SAFETY INSTRUCTIONS .................................................. 5
OVERVIEW ...................................................................... 8
Introduction ...................................................... 8
Series Lineup / Main Features ......................... 10
Principle of Operation ..................................... 11
Front Panel Overview ...................................... 12
Rear Panel Overview ........................................ 17
CV/CC Crossover Characteristics ..................... 18
SETUP ........................................................................... 19
Power Up ........................................................ 19
Load Cable Connection .................................... 20
Output On/Off ................................................ 21
Select CH1/CH2 series or parallel mode ......... 21
Setting Lock from Front Panel ......................... 23
Set the output state at startup ........................ 23
Selection voltage/current display digits........... 24
Remote Control Setting ................................... 25
OPERATION .................................................................. 26
CH1/CH2 Independent Mode .......................... 26
CH3 Independent Mode .................................. 28
CH1/CH2 Series Tracking Mode ...................... 30
CH1/CH2 Parallel Tracking Mode .................... 34
FAQ ............................................................................... 36
APPENDIX ..................................................................... 37
Fuse Replacement ........................................... 37
Specifications .................................................. 38
3
Page 4
GPE-3060/GPE-6030 User Manual
Declaration of Conformity ............................... 40
INDEX ............................................................................ 42
4
Page 5
SAFETY INSTRUCTIONS
WARNING
Warning: Identifies conditions or practices that could result in injury or loss of life.
CAUTION
Caution: Identifies conditions or practices that could result in damage to the GPE series or to other properties.
DANGER High Voltage
Attention Refer to the Manual
Protective Conductor Terminal
Earth (ground) Terminal
Do not dispose electronic equipment as unsorted municipal waste. Please use a separate collection facility or contact the supplier from which this instrument was purchased.
SAFETY INSTRUCTIONS
This chapter contains important safety instructions that you must follow when operating the GPE series and when keeping it in storage. Read the following before any operation to insure your safety and to keep the best condition.
Safety Symbols
These safety symbols may appear in this manual.
5
Page 6
GPE-3060/GPE-6030 User Manual
General Guidelines
CAUTION
Do not place any heavy object on the device. Avoid severe impacts or rough handling that
leads to damaging the device.
Do not discharge static electricity to the device. Do not block or obstruct the cooling fan vent
opening.
Do not perform measurement at circuits directly
connected to Mains (see note below).
Do not disassemble the device unless you are
qualified as service personnel.
Power Supply
WARNING
AC Input voltage:
100V/120V/220V/230VAC ±1 0%, 50/60Hz
Connect the protective grounding conductor of
the AC power cord to an earth ground, to avoid electrical shock.
Fuse
WARNING
Fuse type:
100V/120V: T12A/250V 220V/230V: T6.3A/250V
Make sure the correct type of fuse is installed
before power up.
To ensure fire protection, replace the fuse only
with the specified type and rating.
Disconnect the power cord before fuse
replacement.
Make sure the cause of fuse blowout is fixed
before fuse replacement.
Cleaning the device
Disconnect the power cord before cleaning. Use a soft cloth dampened in a solution of mild
detergent and water. Do not spray any liquid.
Do not use chemicals or cleaners containing
harsh products such as benzene, toluene, xylene, and acetone.
Safety Guidelines
6
Page 7
SAFETY INSTRUCTIONS
Operation Environment
Location: Indoor, no direct sunlight, dust free,
almost non-conductive pollution (note below)
Relative Humidity: < 80% Altitude: < 2000m Temperature: 0°C to 40°C
(Pollution Degree) EN 61010-1:2010 specifies the pollution degrees and their requirements as follows. The GPE series falls under degree
2.
Pollution refers to “addition of foreign matter, solid, liquid, or
gaseous (ionized gases), that may produce a reduction of dielectric strength or surface resistivity”.
Pollution degree 1: No pollution or only dry, non-conductive
pollution occurs. The pollution has no influence.
Pollution degree 2: Normally only non-conductive pollution
occurs. Occasionally, however, a temporary conductivity caused by condensation must be expected.
Pollution degree 3: Conductive pollution occurs, or dry, non-
conductive pollution occurs which becomes conductive due to condensation which is expected. In such conditions, equipment is normally protected against exposure to direct sunlight, precipitation, and full wind pressure, but neither temperature nor humidity is controlled.
Storage environment
Location: Indoor Relative Humidity: < 70% Temperature: −10°C to 70°C
Disposal
Do not dispose this instrument as unsorted municipal waste. Please use a separate collection facility or contact the supplier from which this instrument was purchased. Please make sure discarded electrical waste is properly recycled to reduce environmental impact.
7
Page 8
GPE-3060/GPE-6030 User Manual
Overview
The GPE series regulated DC power supply series are adjustable, multifunctional work stations. It has three independent outputs: two with adjustable voltage/current levels and one with fixed level 5V. The series can be used for logic circuits where various output voltage or current are needed, and for tracking mode definition systems where plus and minus voltages with insignificant error are required.
Independent / Series Tracking / Parallel Tracking
The three output modes, independent, series tracking and parallel tracking can be selected through pressing the TRACKING key on the front panel. In the independent mode, the output voltage and current of each channel are controlled separately. In the tracking modes, both the CH1 and CH2 outputs are automatically connected in series or parallel. CH1 is master and CH2 is slave; no need to connect output leads. In the series mode, the output voltage is doubled; in the parallel mode, the output current is doubled. The isolation degree, from output terminal to chassis or from output terminal to output terminal, is 500V.
OVERVIEW
This chapter describes the GPE series in a nutshell, including its main features and front/rear panel introduction. After going through the overview, follow the Setup chapter (page 19) to properly power up and set operation environment.
Introduction
8
Page 9
OVERVIEW
Constant Voltage/ Constant Current
Each output channel works in constant voltage (CV) or constant current (CC) mode. Even at the maximum output current, a fully rated, continuously adjustable output voltage is provided. For a big load, the power supply can be used as a CV source; while for a small load, a CC source. When in the CV mode (independent or tracking mode), output current (overload or short circuit) can be controlled via the front panel. When in the CC mode (independent mode only), the maximum (ceiling) output voltage can be controlled via the front panel. The power supply will automatically cross over from CV to CC operation when the output current reaches the target value. The power supply will automatically cross over from CC to CV when the output voltage reaches the target value. For more details about CV/CC mode operation, see page 18.
Automatic tracking mode
The front panel display (CH1, CH2) shows the output voltage or current. When operating in the tracking mode, the power supply will automatically connect to the auto- tracking mode. For more details about CH1/CH2 Series Tracking Mode, see page 30.
9
Page 10
GPE-3060/GPE-6030 User Manual
Performance
Low noise: Temperature controlled cooling fan
Operation
Constant Voltage / Constant Current operation Series Tracking / Parallel Tracking operation Output On/Off control Multi-output: Two channels V/I are adjustable,
and one is fixed;
Function for locking the setting Output voltage/ current setting view Set the displayed digit resolution for the voltage
& current output.
Protection
Overload protection Key misoperation protection (Lock) Reverse polarity protection
Interface
Remote control (Output ON/OFF)
Series Lineup / Main Features
Main Features
10
Page 11
OVERVIEW
Overview
The power supply consists of the following.
AC input circuit Transformer Bias power supply including rectifier, filter, pre-
regulator and reference voltage source.
Main regulator circuit including the main
rectifier and filter, series regulator, current comparator, voltage comparator, reference voltage amplifier, remote device and relay control circuit.
The block diagram below shows the CH1 circuit arrangement. The single phase input power is connected to the transformer through the input circuit. Details of each part are described in the next page.
Block diagram
Voltage
Amplifier
Voltage
Comparator
Series
Regulator
Instant Over
Load
Protection
Current
Comparator
Reference
Voltage
Source
Auxiliary
Rectifier &
Filter
“OR” Gate Amplifier
Relay
Control
Main
Rectifier &
Filter
AC Input
Transformer
O/P
Principle of Operation
11
Page 12
GPE-3060/GPE-6030 User Manual
Display
CH1 parameter display area
CH2 parameter display area
CH3 parameter display area
Status display area
LCD Voltage/OVP Knob CH1/CH2 Mode Setting Current Knob
Power Switch CH1 Output GND Terminal CH2 Output CH3 Output
Output key
OVP /Lock key
CH2 toggle key
CH1 toggle key
Front Panel Overview
The figure above is the front view of the GPE-3060.
12
Page 13
OVERVIEW
Output status display
Voltmeter
Displays output voltage of CH1/CH2 channel.
3 digits:
4 digits: CH3 display:
Ammeter
Displays output current of CH1/CH2 channel.
3 digits:
4 digits:
CV/CC/OVP indicators for CH1/CH2
You can view the constant current, constant voltage or OVP status for
Channel ( icon appears on the leaf-hand side of the LCD display). You can view the CV/CC status of the corresponding channel, which is only valid when outputting ON and not displayed when outputting OFF. When OVP is enabled, it is displayed in white, and when it is not enabled, there is no display. After startup, the OVP font flashes.
View setting value
When the channel output is OFF, it is displayed as the set value; When ON, it is displayed as the output value, when there is a set operation, it is displayed as the set value, and the status bar has a display..
GPE-4323


 
 




 
 


GPE-4323


 
 


GPE-4323


 
 


13
Page 14
GPE-3060/GPE-6030 User Manual
Channel indicator
Indicates the currently selected channel.
Output status of CH3
When the output current is over range, the overloaded indicator
will be lit on the LCD
display.
14
Page 15
OVERVIEW
Control Panel
CH1/CH2 parameter setting
Voltage/CurrentWhen rotate clockwise, the value increases, and rotate counterclockwise, the value decreases. Click on the Encode to enter the setting. Automatically confirm and exit the set state after stopping the operation for a few seconds.
OVP: Press and hold the Encode to enter the OVP setting. When rotating clockwise, the value increases, and rotating counterclockwise, the value decreases. Click the Encode to enter the setting. Automatically confirm and exit the set state after stopping the operation for a few seconds.
CH1/CH2 toggle
Click on the CH1 or CH2 button (there will be a flashing display of the corresponding channel number on the LCD) to set the voltage/current of this channel or view its readback value.
Parallel/Series Keys
Activates parallel/series tracking operation. For details, see page 30. The corresponding channel will be displayed on the LCD display.
15
Page 16
GPE-3060/GPE-6030 User Manual
Enable/Disable OVP, Lock key
Click this button to enable and disable the OVP function, which can refer to the icon display in the status bar; Long time is the Lock function, which can lock or unlock the panel buttons (except for serial and parallel operation and output key functions). The status bar
displays ,For more information, please refer to page 23.
Output Key
Turns the output on or off. For more details, see page 21.
Power Switch
Turns On or Off the main power. For the power up sequence, see page 19.
Terminals
GND Terminal
Accepts a grounding wire.
CH1 Output
Outputs CH1 voltage and current.
CH2 Output
Outputs CH2 voltage and current.
CH3 Output
Outputs CH3 voltage and current.
16
Page 17
OVERVIEW
Remote Control Terminal
For more information about the remote control terminal, please see page 25.
Power Cord / Fuse Socket
The power cord socket accepts the AC mains. For power up details, see page 19.
The fuse holder contains the AC mains fuse. For fuse replacement details, see page
37.
AC Selector
Selects AC input voltage: 100V/ 120V/ 220V/ 230V; 50/60Hz.









  



   







   
Remote control port Cooling fan
AC voltage selector Power cord/fuse socket
Rear Panel Overview
17
Page 18
GPE-3060/GPE-6030 User Manual
Background
According to load condition, The GPE can automatically switch between constant voltage mode (CV) and constant current mode (CC).
CV mode
When the current level is smaller than the output setting, the GPE series operates in Constant Voltage mode. The indicator for the corresponding channel appears on the LCD. The Voltage level is kept at the setting and the Current level fluctuates according to the load condition until it reaches the output current setting.
CC mode
When the current level reaches the output setting, the GPE series starts operating in Constant Current mode. The indicator for the corresponding channel appears on the LCD. The Current level is kept at the setting but the Voltage level becomes lower than the setting, in order to suppress the output power level from overload. When the current level becomes lower than the setting, the GPE series goes back to the Constant Voltage mode.
Diagram
Vmax
Imax
Constant
Voltage
Constant
Current
Vout
Iout
CV/CC Crossover Characteristics
18
Page 19
SETUP
Select AC voltage
Before powering up the power supply, select the AC input voltage from the rear panel.
Connect AC power cord
Connect the AC power cord to the rear panel socket.
Power On
Press the power switch to turn on the power. The display will first display all the LCD segments before showing settings for each channel.
Power Off
Press the power switch again to turn off the power.

   
SETUP
This chapter describes how to properly power up and configure the GPE series before operation.
Power Up
19
Page 20
GPE-3060/GPE-6030 User Manual
Standard accessories
1. Turn the terminal
counterclockwise and loosen the screw.
2. Insert the cable terminal.
3. Turn the terminal clockwise
and tighten the screw.
Banana plug
Insert the plug into the socket.
Wire type
When using load cables other than the attached, make sure they have enough current capacity for minimizing cable loss and load line impedance. Voltage drop across a wire should not excess 0.5V. The following list is the wire current rating at 450A/cm2.
Wire size (AWG)
Maximum current (A)
20
2.5 18
4
16
6
14
10 12
16
Load Cable Connection
20
Page 21
SETUP
Panel operation
Press the Output key to turn on all outputs in each channel. The
icon will become lit on the
LCD display.
Push the Output key again to turn off all outputs. The icon will become lit on the LCD display.
Automatic output off
Any of the following actions during output on automatically turns it off.
Change the operation mode between
independent/series tracking/parallel tracking, three channels will be closed together.
When a single channel OVP is started, the
corresponding output will be closed. When CH1/CH2 is simultaneously OVP, three channels will be closed together.
Background / Connection
When you need to output a higher voltage or current through the GPE-2323/3323/4323 series can be connected in series or parallel to achieve it. When connecting in series, the output voltage is twice than that of a single channel. When connecting in parallel, the output current is twice than that of a single channel. For details, please see page 30 through to 34.
Panel operation
You can toggle the connection mode of CH1/ CH2 by using different combinations of the mode selection key.
Output On/Off
Select CH1/CH2 series or parallel mode
21
Page 22
GPE-3060/GPE-6030 User Manual
For the independent mode,
the right key is not pressed
Toggle to parallel mode
when both keys are pressed.
Right key is pressed and the
left key is not pressed in series mode.
When CH1 / CH2 is in the
series or parallel mode, the corresponding series or parallel icon appears on the LCD display.






22
Page 23
SETUP
Background / Connection
The lock function can be used when you need to keep the output constant to avoid the load from being damaged due to inadvertent operation.
Panel operation
Press the LOCK key (for more than 2 seconds) to lock the operation in the front panel. The icon will become lit.
To unlock, press the LOCK key for more than 2 seconds. The icon will then turn off.
Note
The OUTPUT key is not affected by the lock
operation.
Series and parallel operation is not controlled
by the locking key, It will unlock the lock function.
Background / Connection
Through the following steps, you can set the output state at its next startup. There are two choices, ON and OFF available for selection.
Panel operation
1. Press and hold the Output key
and turn on the power until the
or icon flashes on the
LCD display.
2. Press the OVP key to select.
3. Press the ON/OFF key to
confirm.
Note
By default the output is set to OFF at startup.
Setting Lock from Front Panel
Set the output state at startup
23
Page 24
GPE-3060/GPE-6030 User Manual
Background / Connection
It can set the displayed digit resolution for the voltage and current settings/readings to 3 or 4 digits at startup.
Panel operation
1. Press and hold the "OVP" key and
turn the power until on the decimal point for the CH1 voltage flashes on the LCD display.
2. Press the OVP key to select the
number of displayed digits.
3. Press the ON/OFF key to
confirm the selection.
Note
By default the number of displayed digits is set to
3.
Selection voltage/current display digits
24
Page 25
SETUP
Background / Connection
Through the "Remote Control" terminal, the GPE can turn the power on or off.
Panel operation
1. Short pins 7 and 8 (remote
control setting). This will put the power state (ON/OFF) under remote control. At this moment, the or icon flashes on the LCD display.
2. Output control :
Pin 9 & 10 Open: ON state.
Pin 9 & 10 Short: OFF state.
Warring
The remote control terminal can only be controlled by shorting (external relay or jumper shunt) /opening the pins. Voltage cannot be applied to the pins. It is strictly prohibited to short pins 5 & 7 or 6 & 8. Pin 1~6 must be set to open.




   

   

   
ON/OFF setting
Remote control setting
1 3 5 7 9
2 4 6 8 10
Remote Control Setting
25
Page 26
GPE-3060/GPE-6030 User Manual
Background / Connection
CH1 and CH2 outputs work independent of each other.
Output rating
Please refer to the specification sheet.
Panel operation
1. Make sure the
Series/Parallel key is not activated (both the and
icons are off).



+ -
CH1
Load
+ -
CH2
Load
OPERATION
CH1/CH2 Independent Mode
26
Page 27
OPERATION
2. Connect the load to the front panel terminals,
CH1 +/−, CH2 +/−.
3. Use the voltage and current
knob to set the CH1 output voltage and current.
4. Use the voltage and current
knob to set the CH2 output voltage and current.
5. Press the Output key to turn
on the output. The Output key will be lit and the icon will appear on the LCD display. The or icon appears on the LCD to indicate the output status for each channel.
27
Page 28
GPE-3060/GPE-6030 User Manual
Output rating
Please refer to the specification sheet.
No Series/Parallel Tracking
CH3 doesnt have series/parallel tracking mode. Also, the CH3 output is not affected by the CH1 and CH2 modes.
Panel operation
1. Connect the load to the front
panel CH3 +/− terminal.
2. Press the Output key to turn
on the output. The Output key will be lit.
+ -
CH3
Load
CH3 Independent Mode
28
Page 29
OVERLOAD
When the output current value is too high, the icon appears on the LCD display and CH3 operation mode switches from constant voltage to constant current.
OPERATION
29
Page 30
GPE-3060/GPE-6030 User Manual
Background
Series tracking operation allows the GPE to combine the output by internally connecting CH1 (Master) and CH2 (Slave) in series. CH1 (Master) controls the combined output voltage/current level which is set independently.
The following describes two types of configurations, depending on how common ground is used.
Connection
Output rating
Please refer to the specification sheet
1. Press the Series/Parallel key
to activate the series tracking mode. The icon will be lit on the LCD display.
Load
+ -
CH1/CH2 Series Tracking Mode
Series Tracking without Common Terminal
30
Page 31
OPERATION
2. Connect the load to the front panel terminals,
CH1+ & CH2− (Single supply).
3. Use the current knob to set the
CH2 output current to the maximum level.
4. Use the voltage and current
knob to set the CH1 output voltage and current level.
5. Press the Output key to turn
on the output. The Output key will be lit.
6. Refer to the CH1 (Master) meter and indicators
for the output level and CV/CC status.
Output voltage level
Double the reading on the CH1 voltage meter.
Output current level
CH1 meter reading shows the output current.
31
Page 32
GPE-3060/GPE-6030 User Manual
Connection
Output rating
CH1+ ~ COM ~ CH2- ,
Please refer to the specification sheet.
1. Press the Series/Parallel key
to activate the series tracking mode. The icon will be lit on the LCD display.
2. Connect the load to the front panel terminals,
CH1+ & CH2−. Use the CH1 (−) terminal as the common line connection.
G N D
Common
Load
+ COM -
Series Tracking with Common Terminal
32
Page 33
OPERATION
3. Use the CH1 voltage knob to
set the master & slave output voltage (the same level for both channels).
4. Use the CH1 current knob to
set the master output current.
5. Use the CH2 current knob to
set the slave output current.
6. Press the Output key to turn
on the output. The Output key will be lit.
7. Refer to the CH1 (Master) meter and indicators
for the output level and CV/CC status.
CH1 (Master) voltage level
CH1 meter reading shows the output voltage.
CH1 (Master) current level
CH1 meter reading shows the output current.
8. Refer to the CH1/CH2 meter and CH2
indicators for the output level and CV/CC status.
CH2 (Slave) voltage level
The CH2 meter reading shows the output voltage.
CH2 (Slave) current level
The CH2 meter reading shows the output current.
33
Page 34
GPE-3060/GPE-6030 User Manual
Background / Connection
Parallel tracking operation allows the GPE to combine the output by internally connecting CH1 (Master) and CH2 (Slave) in parallel. CH1 (Master) controls the combined output voltage/current level.
Output rating
Please refer to the specification sheet
1. Press the Series/Parallel key
to activate the parallel tracking mode. The icon will be lit on the LCD display.
2. Connect the load to the CH1 +/− terminals.
+ -
Load
CH1/CH2 Parallel Tracking Mode
34
Page 35
OPERATION
3. Use the CH1 voltage and
current knobs to set the output voltage and current. CH2 control function is disabled.
4. Press the Output key to turn
on the output. The Output key will be lit.
5. The operating mode of CH2
will appear as the icon on the LCD display.
6. Refer to the CH1 meter and indicator for the
output level and CV/CC status.
Output voltage level
The CH1 meter reading shows the output voltage.
Output current level
Double the amount of CH1 current meter reading.
35
Page 36
FAQ
GPE-3060/GPE-6030 User Manual
Q1. I pressed the panel lock key but the output still turns on/off.
A1. For safety reasons the output key is not affected by the panel key lock feature.
Q2. The CH3 overload indicator turned on – is this an error?
A2. No, it simply means that the CH3 output current reached the maximum 5.2A and the operation mode turned from CV (constant voltage) to CC (constant current). You can continue using the power supply, although reducing the output load is recommended.
Q3. The specifications do not match the real accuracies.
A3. Make sure that the power supply is powered on for at least 30 minutes, within +20°C ~ +30°C.
For more information, contact your local dealer or GWInstek at www.gwinstek.com.tw / marketing@goodwill.com.tw
36
Page 37
APPENDIX
Steps
1. Take off the power cord and remove the fuse
socket using a minus driver.
2. Replace the fuse in the holder.
Rating
100V/120V:T12A/250V 220V/230V:T6.3A/250V
APPENDIX
Fuse Replacement
37
Page 38
GPE-3060/GPE-6030 User Manual
CH1/ CH2
Output
Ratings
Independent
0 ~ 30V , 0 ~ 6A (GPE-3060) 0 ~ 60V , 0 ~ 3A (GPE-6030)
Series
0 ~ 60V , 0 ~ 6A (GPE-3060) 0 ~ 120V , 0 ~ 3A (GPE-6030)
Parallel
0 ~ 30V , 0 ~ 12A (GPE-3060) 0 ~ 60V , 0 ~ 6A (GPE-6030)
Voltage
Regulation
Line Regulation
0.01% + 3mV
Load Regulation
0.01% + 5mV 0.02% + 5mV ( 10A)
Ripple & Noise
1mVrms (5Hz ~ 1MHz)
Recovery Time
100µs
(50% load change, minimum load 0.5A)
Temperature Coefficient
300ppm/°C
Current
Regulation
Line Regulation
0.1% + 3mA
Load Regulation
0.1% + 3mA
Ripple & Noise
2mArms
OVP
Range
OFF,ON(1V35V) (GPE-3060) OFF,ON(1V65V) (GPE-6030)
Resolution
1V
Accuracy
±1V
Tracking
Operation
Output tracking Error
0.1%+10mV of Master(GPE-3060) 0.2%+20mV of Master(GPE-6030)
(No Load, with load add load regulation 200mV)
Parallel Regulation
Line: 0.01% + 3mV
Load: 0.01% + 5mV
0.02% + 5mV ( 10A)
Ripple & Noise:
2mVrms (5Hz~1MHz)
Series Regulation
Line: 0.01% + 5mV
Load: 200mV
Ripple & Noise:
2mVrms (5Hz~1MHz)
Specifications
The specifications apply when the GPE are powered on for at least 30 minutes under +20°C – +30°C.
38
Page 39
APPENDIX
Meter
Display
LCD
4.3" single color LCD
Setting/Read back Resolution
Voltage(4digits): 10mV (GPE-3060) / 20mV (GPE-6030) Current(4digits): 2mA (GPE-3060) / 1mA (GPE-6030)
Setting/ Read back Accuracy
Voltage: ±(0.1% of reading+30mV) (4digits) ±(0.1% of reading+200mV) (3digits) Current: ±(0.3% of reading+10mA) (4digits) ±(0.3% of reading+20mA) (3digits)
CH3
Output
Voltage
5V ±5%
Current
5A
Line Regulation
3mV
Load Regulation
5mV
Ripple & Noise
1mVrms (5Hz ~ 1MHz)
Recovery Time
100µs
OVP
5.5V
Insulation Chassis and Terminal
20MΩ or above (DC 500V)
Chassis and AC cord
30MΩ or above (DC 500V)
Operation Environment
Indoor use, Altitude: 2000m Ambient temperature: 0 ~ 40°C Relative humidity: 80% Installation category: II Pollution degree: 2
Storage Environment
Ambient temperature: -10 ~ 70°C Relative humidity: 70%
Power Source
AC 100V/120V/220V/230V±10%, 50/60Hz
consumption
900VA, 680W
Accessories
Quick Start manual x1 Test lead: Non-European: GTL-104A x3 Test lead: European: GTL-204A x 3 , GTL-201A x1
Dimensions
210 (W) x 155 (H) x 360 (D) mm
Weight
Approx. 10kg
39
Page 40
GPE-3060/GPE-6030 User Manual
We
GOOD WILL INSTRUMENT CO., LTD.
declare that the CE marking mentioned product
satisfies all the technical relations application to the product within the scope of council:
Directive: EMC; LVD; WEEE; RoHS
The product is in conformity with the following standards or other normative documents:
EN 61326-1 :
Electrical equipment for measurement, control and laboratory use –– EMC requirements
Conducted & Radiated Emission EN 55011 / EN 55032
Electrical Fast Transients EN 61000-4-4
Current Harmonics EN 61000-3-2 / EN 61000-3-12
Surge Immunity EN 61000-4-5
Voltage Fluctuations EN 61000-3-3 / EN 61000-3-11
Conducted Susceptibility EN 61000-4-6
Electrostatic Discharge EN 61000-4-2
Power Frequency Magnetic Field EN 61000-4-8
Radiated Immunity EN 61000-4-3
Voltage Dip/ Interruption EN 61000-4-11 / EN 61000-4-34
EN 61010-1 :
Safety requirements for electrical equipment for measurement, control, and laboratory use - Part 1: General requirements
Declaration of Conformity
EMC
Safety
GOOD WILL INSTRUMENT CO., LTD. No. 7-1, Jhongsing Road, Tucheng Dist., New Taipei City 236, Taiwan Tel: +886-2-2268-0389 Fax: +866-2-2268-0639 Web: www.gwinstek.com Email: marketing@goodwill.com.tw
GOOD WILL INSTRUMENT (SUZHOU) CO., LTD. No. 521, Zhujiang Road, Snd, Suzhou Jiangsu 215011, China Tel: +86-512-6661-7177 Fax: +86-512-6661-7277 Web: www.instek.com.cn Email: marketing@instek.com.cn
40
Page 41
APPENDIX
GOOD WILL INSTRUMENT EURO B.V. De Run 5427A, 5504DG Veldhoven, The Netherlands Tel: +31(0)40-2557790 Fax: +31(0)40-2541194 Email: sales@gw-instek.eu
41
Page 42
GPE-3060/GPE-6030 User Manual
INDEX
Automatic out off ................. 21
Banana plug .......................... 20
Caution symbol ...................... 5
CC/CV ................................... 29
Cleaning the instrument ........ 6
Common terminal, series
tracking .............................. 30
Cooling fan ............................ 17
safety instruction ................ 6
CV/CC
CH1/CH2 indicator ......... 27
CH3 indicator.................... 29
Operation theory .......... 9, 18
Disposal instructions ............. 7
EN61010
Pollution degree ................. 7
Environment
Operation............................. 7
Specification ..................... 39
Storage ................................. 7
Front panel
Lock (manual) ................... 23
Overview ........................... 13
Fuse
Rating ................................. 37
Replacement ...................... 37
Safety instruction ................ 6
GPE series
Block diagram ................... 11
List of features .................. 10
Operation theory .............. 11
Ground symbol ...................... 5
Load connection ................... 20
Operation mode
Independent ..................... 26
Parallel Tracking .............. 34
Series Tracking ................. 30
Output current setting
Manual .............................. 27
Output on/off
FAQ ................................... 36
manual............................... 21
Output voltage setting
Manual .............................. 27
Over load indicator ............. 29
Power supply
Safety instruction ............... 6
Setup .................................. 19
Socket overview ............... 17
Specification .................... 39
Protective ground symbol .... 5
remote output control
terminal overview............ 17
Service operation
About disassembly ............ 6
Contact .............................. 36
tracking mode
parallel ................... 23, 24, 25
Tracking mode
Operation theory................ 8
Warning symbol .................... 5
Wire, load.............................. 20