SAFETY INSTRUCTIONS
WARNING
Warning: Identifies conditions or practices that
could result in injury or loss of life.
CAUTION
Caution: Identifies conditions or practices that
could result in damage to the GPE series or to
other properties.
DANGER High Voltage
Attention Refer to the Manual
Protective Conductor Terminal
Earth (ground) Terminal
Do not dispose electronic equipment as unsorted
municipal waste. Please use a separate collection
facility or contact the supplier from which this
instrument was purchased.
SAFETY INSTRUCTIONS
This chapter contains important safety instructions
that you must follow when operating the GPE
series and when keeping it in storage. Read the
following before any operation to insure your
safety and to keep the best condition.
Safety Symbols
These safety symbols may appear in this manual.
5
Page 6
GPE-3060/GPE-6030 User Manual
General
Guidelines
CAUTION
Do not place any heavy object on the device.
Avoid severe impacts or rough handling that
leads to damaging the device.
Do not discharge static electricity to the device.
Do not block or obstruct the cooling fan vent
opening.
Do not perform measurement at circuits directly
connected to Mains (see note below).
Do not disassemble the device unless you are
qualified as service personnel.
Power Supply
WARNING
AC Input voltage:
100V/120V/220V/230VAC ±1 0%, 50/60Hz
Connect the protective grounding conductor of
the AC power cord to an earth ground, to avoid
electrical shock.
Fuse
WARNING
Fuse type:
100V/120V: T12A/250V
220V/230V: T6.3A/250V
Make sure the correct type of fuse is installed
before power up.
To ensure fire protection, replace the fuse only
with the specified type and rating.
Disconnect the power cord before fuse
replacement.
Make sure the cause of fuse blowout is fixed
before fuse replacement.
Cleaning the
device
Disconnect the power cord before cleaning.
Use a soft cloth dampened in a solution of mild
detergent and water. Do not spray any liquid.
Do not use chemicals or cleaners containing
harsh products such as benzene, toluene, xylene,
and acetone.
Safety Guidelines
6
Page 7
SAFETY INSTRUCTIONS
Operation
Environment
Location: Indoor, no direct sunlight, dust free,
(Pollution Degree) EN 61010-1:2010 specifies the pollution degrees
and their requirements as follows. The GPE series falls under degree
2.
Pollution refers to “addition of foreign matter, solid, liquid, or
gaseous (ionized gases), that may produce a reduction of dielectric
strength or surface resistivity”.
Pollution degree 1: No pollution or only dry, non-conductive
pollution occurs. The pollution has no influence.
Pollution degree 2: Normally only non-conductive pollution
occurs. Occasionally, however, a temporary conductivity caused
by condensation must be expected.
Pollution degree 3: Conductive pollution occurs, or dry, non-
conductive pollution occurs which becomes conductive due to
condensation which is expected. In such conditions, equipment
is normally protected against exposure to direct sunlight,
precipitation, and full wind pressure, but neither temperature
nor humidity is controlled.
Do not dispose this instrument as unsorted
municipal waste. Please use a separate collection
facility or contact the supplier from which this
instrument was purchased. Please make sure
discarded electrical waste is properly recycled to
reduce environmental impact.
7
Page 8
GPE-3060/GPE-6030 User Manual
Overview
The GPE series regulated DC power supply series
are adjustable, multifunctional work stations. It
has three independent outputs: two with
adjustable voltage/current levels and one with
fixed level 5V. The series can be used for logic
circuits where various output voltage or current
are needed, and for tracking mode definition
systems where plus and minus voltages with
insignificant error are required.
Independent /
Series Tracking /
Parallel Tracking
The three output modes, independent, series
tracking and parallel tracking can be selected
through pressing the TRACKING key on the front
panel. In the independent mode, the output
voltage and current of each channel are controlled
separately. In the tracking modes, both the CH1
and CH2 outputs are automatically connected in
series or parallel. CH1 is master and CH2 is slave;
no need to connect output leads. In the series
mode, the output voltage is doubled; in the
parallel mode, the output current is doubled. The
isolation degree, from output terminal to chassis or
from output terminal to output terminal, is 500V.
OVERVIEW
This chapter describes the GPE series in a nutshell,
including its main features and front/rear panel
introduction. After going through the overview,
follow the Setup chapter (page 19) to properly
power up and set operation environment.
Introduction
8
Page 9
OVERVIEW
Constant Voltage/
Constant Current
Each output channel works in constant voltage
(CV) or constant current (CC) mode. Even at the
maximum output current, a fully rated,
continuously adjustable output voltage is
provided. For a big load, the power supply can be
used as a CV source; while for a small load, a CC
source. When in the CV mode (independent or
tracking mode), output current (overload or short
circuit) can be controlled via the front panel. When
in the CC mode (independent mode only), the
maximum (ceiling) output voltage can be
controlled via the front panel. The power supply
will automatically cross over from CV to CC
operation when the output current reaches the
target value. The power supply will automatically
cross over from CC to CV when the output voltage
reaches the target value. For more details about
CV/CC mode operation, see page 18.
Automatic
tracking mode
The front panel display (CH1, CH2) shows the
output voltage or current. When operating in the
tracking mode, the power supply will
automatically connect to the auto- tracking mode.
For more details about CH1/CH2 Series Tracking
Mode, see page 30.
9
Page 10
GPE-3060/GPE-6030 User Manual
Performance
Low noise: Temperature controlled cooling fan
Operation
Constant Voltage / Constant Current operation
Series Tracking / Parallel Tracking operation
Output On/Off control
Multi-output: Two channels V/I are adjustable,
and one is fixed;
Function for locking the setting
Output voltage/ current setting view
Set the displayed digit resolution for the voltage
AC input circuit
Transformer
Bias power supply including rectifier, filter, pre-
regulator and reference voltage source.
Main regulator circuit including the main
rectifier and filter, series regulator, current
comparator, voltage comparator, reference
voltage amplifier, remote device and relay
control circuit.
The block diagram below shows the CH1 circuit
arrangement. The single phase input power is
connected to the transformer through the input
circuit. Details of each part are described in the
next page.
Block diagram
Voltage
Amplifier
Voltage
Comparator
Series
Regulator
Instant Over
Load
Protection
Current
Comparator
Reference
Voltage
Source
Auxiliary
Rectifier &
Filter
“OR” GateAmplifier
Relay
Control
Main
Rectifier &
Filter
AC Input
Transformer
O/P
Principle of Operation
11
Page 12
GPE-3060/GPE-6030 User Manual
Display
CH1 parameter
display area
CH2 parameter
display area
CH3 parameter
display area
Status display
area
LCD Voltage/OVP Knob CH1/CH2 Mode Setting Current Knob
Power Switch CH1 Output GND Terminal CH2 Output CH3 Output
Output key
OVP /Lock key
CH2 toggle key
CH1 toggle key
Front Panel Overview
The figure above is the front view of the GPE-3060.
12
Page 13
OVERVIEW
Output status
display
Voltmeter
Displays output voltage of CH1/CH2 channel.
3 digits:
4 digits:
CH3 display:
Ammeter
Displays output current of CH1/CH2 channel.
3 digits:
4 digits:
CV/CC/OVP
indicators for
CH1/CH2
You can view the constant current,
constant voltage or OVP status for
Channel (icon appears on the
leaf-hand side of the LCD display).
You can view the CV/CC status of
the corresponding channel, which is
only valid when outputting ON and
not displayed when outputting OFF.
When OVP is enabled, it is
displayed in white, and when it is
not enabled, there is no display.
After startup, the OVP font flashes.
View setting value
When the channel output is OFF, it
is displayed as the set value; When
ON, it is displayed as the output
value, when there is a set operation,
it is displayed as the set value, and
the status bar has a display..
GPE-4323
GPE-4323
GPE-4323
13
Page 14
GPE-3060/GPE-6030 User Manual
Channel indicator
Indicates the currently selected
channel.
Output status of
CH3
When the output current is over
range, the overloaded indicator
will be lit on the LCD
display.
14
Page 15
OVERVIEW
Control Panel
CH1/CH2
parameter setting
Voltage/Current:When rotate
clockwise, the value increases, and
rotate counterclockwise, the value
decreases. Click on the Encode to
enter the setting. Automatically
confirm and exit the set state after
stopping the operation for a few
seconds.
OVP: Press and hold the Encode to
enter the OVP setting. When
rotating clockwise, the value
increases, and rotating
counterclockwise, the value
decreases. Click the Encode to enter
the setting. Automatically confirm
and exit the set state after stopping
the operation for a few seconds.
CH1/CH2 toggle
Click on the CH1 or CH2 button
(there will be a flashing display of
the corresponding channel number
on the LCD) to set the
voltage/current of this channel or
view its readback value.
Parallel/Series
Keys
Activates parallel/series tracking
operation. For details, see page 30.
The corresponding channel will be
displayed on the LCD display.
15
Page 16
GPE-3060/GPE-6030 User Manual
Enable/Disable
OVP, Lock key
Click this button to enable and
disable the OVP function, which can
refer to the icon display in the
status bar; Long time is the Lock
function, which can lock or unlock
the panel buttons (except for serial
and parallel operation and output
key functions). The status bar
displays ,For more information,
please refer to page 23.
Output Key
Turns the output on or off. For more
details, see page 21.
Power Switch
Turns On or Off the main
power. For the power up sequence,
see page 19.
Terminals
GND Terminal
Accepts a grounding wire.
CH1 Output
Outputs CH1 voltage and current.
CH2 Output
Outputs CH2 voltage and current.
CH3 Output
Outputs CH3 voltage and current.
16
Page 17
OVERVIEW
Remote
Control
Terminal
For more information about the
remote control terminal, please
see page 25.
Power Cord /
Fuse Socket
The power cord socket accepts
the AC mains. For power up
details, see page 19.
The fuse holder contains the
AC mains fuse. For fuse
replacement details, see page
37.
AC Selector
Selects AC input voltage:
100V/ 120V/ 220V/ 230V;
50/60Hz.
Remote control port Cooling fan
AC voltage selector Power cord/fuse socket
Rear Panel Overview
17
Page 18
GPE-3060/GPE-6030 User Manual
Background
According to load condition, The GPE can
automatically switch between constant voltage mode
(CV) and constant current mode (CC).
CV mode
When the current level is smaller than the output
setting, the GPE series operates in Constant Voltage
mode. The indicator for the corresponding
channel appears on the LCD. The Voltage level is kept
at the setting and the Current level fluctuates
according to the load condition until it reaches the
output current setting.
CC mode
When the current level reaches the output setting, the
GPE series starts operating in Constant Current mode.
The indicator for the corresponding channel
appears on the LCD. The Current level is kept at the
setting but the Voltage level becomes lower than the
setting, in order to suppress the output power level
from overload. When the current level becomes lower
than the setting, the GPE series goes back to the
Constant Voltage mode.
Diagram
Vmax
Imax
Constant
Voltage
Constant
Current
Vout
Iout
CV/CC Crossover Characteristics
18
Page 19
SETUP
Select AC voltage
Before powering up the power
supply, select the AC input
voltage from the rear panel.
Connect AC
power cord
Connect the AC power cord to
the rear panel socket.
Power On
Press the power switch to turn
on the power. The display will
first display all the LCD
segments before showing
settings for each channel.
Power Off
Press the power switch again to
turn off the power.
SETUP
This chapter describes how to properly power up
and configure the GPE series before operation.
Power Up
19
Page 20
GPE-3060/GPE-6030 User Manual
Standard
accessories
1. Turn the terminal
counterclockwise and loosen
the screw.
2. Insert the cable terminal.
3. Turn the terminal clockwise
and tighten the screw.
Banana plug
Insert the plug into the socket.
Wire type
When using load cables other than the attached,
make sure they have enough current capacity for
minimizing cable loss and load line impedance.
Voltage drop across a wire should not excess 0.5V.
The following list is the wire current rating at
450A/cm2.
Wire size (AWG)
Maximum current (A)
20
2.5 18
4
16
6
14
10 12
16
Load Cable Connection
20
Page 21
SETUP
Panel operation
Press the Output key to turn on
all outputs in each channel. The
icon will become lit on the
LCD display.
Push the Output key again to
turn off all outputs. The icon
will become lit on the LCD
display.
Automatic output
off
Any of the following actions during output on
automatically turns it off.
Change the operation mode between
independent/series tracking/parallel tracking,
three channels will be closed together.
When a single channel OVP is started, the
corresponding output will be closed.
When CH1/CH2 is simultaneously OVP, three
channels will be closed together.
Background /
Connection
When you need to output a higher voltage or
current through the GPE-2323/3323/4323 series
can be connected in series or parallel to achieve it.
When connecting in series, the output voltage is
twice than that of a single channel. When
connecting in parallel, the output current is twice
than that of a single channel. For details, please see
page 30 through to 34.
Panel operation
You can toggle the connection
mode of CH1/ CH2 by using
different combinations of the
mode selection key.
Output On/Off
Select CH1/CH2 series or parallel mode
21
Page 22
GPE-3060/GPE-6030 User Manual
For the independent mode,
the right key is not pressed
Toggle to parallel mode
when both keys are pressed.
Right key is pressed and the
left key is not pressed in
series mode.
When CH1 / CH2 is in the
series or parallel mode, the
corresponding series or
parallel icon appears on the
LCD display.
22
Page 23
SETUP
Background /
Connection
The lock function can be used when you need to
keep the output constant to avoid the load from
being damaged due to inadvertent operation.
Panel operation
Press the LOCK key (for more than 2
seconds) to lock the operation in the
front panel. The icon will
become lit.
To unlock, press the LOCK key for more than 2
seconds. The icon will then turn off.
Note
The OUTPUT key is not affected by the lock
operation.
Series and parallel operation is not controlled
by the locking key, It will unlock the lock
function.
Background /
Connection
Through the following steps, you can set the
output state at its next startup. There are two
choices, ON and OFF available for selection.
Panel operation
1. Press and hold the Output key
and turn on the power until the
or icon flashes on the
LCD display.
2. Press the “OVP” key to select.
3. Press the “ON/OFF” key to
confirm.
Note
By default the output is set to OFF at startup.
Setting Lock from Front Panel
Set the output state at startup
23
Page 24
GPE-3060/GPE-6030 User Manual
Background /
Connection
It can set the displayed digit resolution for the
voltage and current settings/readings to 3 or 4
digits at startup.
Panel operation
1. Press and hold the "OVP" key and
turn the power until on the decimal
point for the CH1 voltage flashes on
the LCD display.
2. Press the “OVP” key to select the
number of displayed digits.
3. Press the “ON/OFF” key to
confirm the selection.
Note
By default the number of displayed digits is set to
3.
Selection voltage/current display digits
24
Page 25
SETUP
Background /
Connection
Through the "Remote Control" terminal, the GPE
can turn the power on or off.
Panel operation
1. Short pins 7 and 8 (remote
control setting). This will put
the power state (ON/OFF)
under remote control. At this
moment, the or icon
flashes on the LCD display.
2. Output control :
Pin 9 & 10 Open: ON state.
Pin 9 & 10 Short: OFF state.
Warring
The remote control terminal can only be controlled
by shorting (external relay or jumper shunt)
/opening the pins. Voltage cannot be applied to
the pins. It is strictly prohibited to short pins 5 & 7
or 6 & 8. Pin 1~6 must be set to open.
ON/OFF setting
Remote control setting
1 3 5 7 9
2 4 6 8 10
Remote Control Setting
25
Page 26
GPE-3060/GPE-6030 User Manual
Background /
Connection
CH1 and CH2 outputs work independent of each
other.
Output rating
Please refer to the specification sheet.
Panel operation
1. Make sure the
Series/Parallel key is not
activated (both the and
icons are off).
+ -
CH1
Load
+ -
CH2
Load
OPERATION
CH1/CH2 Independent Mode
26
Page 27
OPERATION
2. Connect the load to the front panel terminals,
CH1 +/−, CH2 +/−.
3. Use the voltage and current
knob to set the CH1 output
voltage and current.
4. Use the voltage and current
knob to set the CH2 output
voltage and current.
5. Press the Output key to turn
on the output. The Output
key will be lit and the
icon will appear on the LCD
display. The or icon
appears on the LCD to
indicate the output status
for each channel.
27
Page 28
GPE-3060/GPE-6030 User Manual
Output rating
Please refer to the specification sheet.
No
Series/Parallel
Tracking
CH3 doesn’t have series/parallel tracking mode.
Also, the CH3 output is not affected by the CH1 and
CH2 modes.
Panel operation
1. Connect the load to the front
panel CH3 +/− terminal.
2. Press the Output key to turn
on the output. The Output
key will be lit.
+ -
CH3
Load
CH3 Independent Mode
28
Page 29
OVERLOAD
When the output current value
is too high, the icon
appears on the LCD display and
CH3 operation mode switches
from constant voltage to
constant current.
OPERATION
29
Page 30
GPE-3060/GPE-6030 User Manual
Background
Series tracking operation allows the GPE to
combine the output by internally connecting CH1
(Master) and CH2 (Slave) in series. CH1 (Master)
controls the combined output voltage/current
level which is set independently.
The following describes two types of
configurations, depending on how common
ground is used.
Connection
Output rating
Please refer to the specification sheet
1. Press the Series/Parallel key
to activate the series tracking
mode. The icon will be lit
on the LCD display.
Load
+ -
CH1/CH2 Series Tracking Mode
Series Tracking without Common Terminal
30
Page 31
OPERATION
2. Connect the load to the front panel terminals,
CH1+ & CH2− (Single supply).
3. Use the current knob to set the
CH2 output current to the
maximum level.
4. Use the voltage and current
knob to set the CH1 output
voltage and current level.
5. Press the Output key to turn
on the output. The Output key
will be lit.
6. Refer to the CH1 (Master) meter and indicators
for the output level and CV/CC status.
Output voltage
level
Double the reading on the CH1
voltage meter.
Output current
level
CH1 meter reading shows the
output current.
31
Page 32
GPE-3060/GPE-6030 User Manual
Connection
Output rating
CH1+ ~ COM ~ CH2- ,
Please refer to the specification sheet.
1. Press the Series/Parallel key
to activate the series tracking
mode. The icon will be
lit on the LCD display.
2. Connect the load to the front panel terminals,
CH1+ & CH2−. Use the CH1 (−) terminal as the
common line connection.
G N D
Common
Load
+ COM -
Series Tracking with Common Terminal
32
Page 33
OPERATION
3. Use the CH1 voltage knob to
set the master & slave output
voltage (the same level for
both channels).
4. Use the CH1 current knob to
set the master output current.
5. Use the CH2 current knob to
set the slave output current.
6. Press the Output key to turn
on the output. The Output
key will be lit.
7. Refer to the CH1 (Master) meter and indicators
for the output level and CV/CC status.
CH1 (Master)
voltage level
CH1 meter reading shows the
output voltage.
CH1 (Master)
current level
CH1 meter reading shows the
output current.
8. Refer to the CH1/CH2 meter and CH2
indicators for the output level and CV/CC
status.
CH2 (Slave)
voltage level
The CH2 meter reading shows
the output voltage.
CH2 (Slave)
current level
The CH2 meter reading shows
the output current.
33
Page 34
GPE-3060/GPE-6030 User Manual
Background /
Connection
Parallel tracking operation allows the GPE to
combine the output by internally connecting CH1
(Master) and CH2 (Slave) in parallel. CH1 (Master)
controls the combined output voltage/current
level.
Output rating
Please refer to the specification sheet
1. Press the Series/Parallel key
to activate the parallel
tracking mode. The
icon will be lit on the LCD
display.
2. Connect the load to the CH1 +/− terminals.
+ -
Load
CH1/CH2 Parallel Tracking Mode
34
Page 35
OPERATION
3. Use the CH1 voltage and
current knobs to set the
output voltage and current.
CH2 control function is
disabled.
4. Press the Output key to turn
on the output. The Output
key will be lit.
5. The operating mode of CH2
will appear as the icon
on the LCD display.
6. Refer to the CH1 meter and indicator for the
output level and CV/CC status.
Output voltage
level
The CH1 meter reading shows
the output voltage.
Output current
level
Double the amount of CH1
current meter reading.
35
Page 36
FAQ
GPE-3060/GPE-6030 User Manual
Q1. I pressed the panel lock key but the output still turns on/off.
A1. For safety reasons the output key is not affected by the panel key
lock feature.
Q2. The CH3 overload indicator turned on – is this an error?
A2. No, it simply means that the CH3 output current reached the
maximum 5.2A and the operation mode turned from CV (constant
voltage) to CC (constant current). You can continue using the power
supply, although reducing the output load is recommended.
Q3. The specifications do not match the real accuracies.
A3. Make sure that the power supply is powered on for at least 30
minutes, within +20°C ~ +30°C.
For more information, contact your local dealer or GWInstek at
www.gwinstek.com.tw / marketing@goodwill.com.tw
Voltage:
±(0.1% of reading+30mV) (4digits)
±(0.1% of reading+200mV) (3digits)
Current:
±(0.3% of reading+10mA) (4digits)
±(0.3% of reading+20mA) (3digits)