GE Sensing Aurora TransPort Specifications

TransPort Aurora
Reliable, Portable Moisture Measurement
BHGE’s Aurora series of moisture analyzers uses tunable diode laser absorption spectroscopy (TDLAS) to rapidly and accurately measure moisture in a variety of background gases. The Aurora TransPort, the latest addition to the series, is a battery operated, moveable
analyzer that can be taken into the eld to directly
BHGE’s patented temperature and pressure compensated TDLAS provides repeatable, accurate and drift-free moisture measurement with fast response.
The TransPort is ideal for spot checking the performance of natural gas processing and drying systems, gas storage facilities, compression stations,
renery processes, heat treating furnaces, instrument air and more. It is ideally suited for eld verication
of permanently installed moisture analyzers and transmitters.
The unit is equipped with a rechargeable battery that provides 8 to 10 hours of operation as well as integrated sample conditioning components. An easy-to-use display/keypad enables direct display of moisture, temperature and pressure. The unit also provides both analog (4-20 mA) and digital interfaces (RS232/485 and Ethernet) for data recording.
bhge.com
Specications
Range
Calibrated Range 0 to 5000 ppm
For CO2 applications: 0 to 1000 ppm
Lower Detection Level 2 ppmv (-71.7°C)
For CO2 applications: 20 ppmv (-55.3°C)
Dew/Frost Point -97.1° to 27.3°F (-71.7°C to –2.6°C) frost point
@ STP of 25°C, 14.696 psia
Process Dew/Frost Point Process/Equivalent dew Point/Frost point calculation
at user-dened constant pressure or external pressure specied by Modbus
Absolute Humidity 0.095 to 237 lbs/MMSCF (1.52 to 3,803 mg/m3)
v
v
Accuracy
Parts Per Million by Volume ±1% of reading or ±2 ppmv , whichever is greater; for
Repeatability ±0.2 ppmv or ±0.1%, whichever is greater
Calibration Certication NIST or equivalent NMI traceable certication
Calibration Options Nitrogen, standard natural gas and
>1000 ppmv ±5% of reading
For CO2 applications: ±3% of reading or ±5 ppmv
For H2 recycle applications: ±1% of reading or ±2 ppmv (for up to ±5% H2 and ±1% C2H6 variation from nominal calibration composition)
(Individual instrument calibrated accuracy conditions
provided in Certicate of Conformance. Accuracy of
other parameters derived from ppmv.)
For CO2 applications: ±1.0 ppmv or ±0.5%, whichever is greater
3 customizable calibration curves
Response Time
Response Time Optical system <2 seconds
System Response The system response is dependent on the length of
sample tubing, sample system components, ow
rate and pressure, as well as the change in moisture concentration.
Pressure
Operating Sample Cell Pressure
Maximum Pressure 30 psig (308.2 kPa)
10 to 25 psia (69 to 172 kPa)
Flow Rate
Sample Cell Flow Rate 0.1 to 1.0 SLPM (0.2 to 2.1 SCFH )
Bypass Fast Loop 5 to 10X of ow rate through sample cell
0.5 SLPM (1.1 SCFH) nominal
I/O
Display Backlit transective display. Three programable
simultaneous parameters. Alphanumeric status and diagnostic display. LEDs for power, laser temperature stability, keypad lockout
Power Standard rechargeable lithium-ion battery pack
(14.4V/6600 mAh); Universal 100 - 240 VAC, 50/60 Hz, 120W adapter with 24 VDC output
Analog Outputs Three 0/4-20 mA DC (source) with maximum 500
ohm impedance. User programmable for any parameter and scalable. Complies with NAMUR protocol for analog signals.
Digital Interfaces Two programmable digital communications ports:
RS232, RS485 with multidrop capability and assignable address, MODBUS RTU protocol. One Ethernet port: Modbus TCP/IP protocol
User Interface Tactile keypad with status/fault indicator LEDs
Laser Class 1 product. Conforms to IEC 60825-1. Edition 2.0
Safety of Laser Products
Enclosure
Net Weight 49 lbs. (22.3 kg)
Dimensions (H x L x W) 24.6 in X 19.7 in X 11.7 in
(624.84 mm X 500.38 mm X 297.18 mm)
Temperature
Operating -20°C to 50°C (-4°F to 149°F) with battery discharging;
0°C to 45°C (32°F to 113°F) with battery charging
Storage -20°C to 50°C (-4°F to 149°F)
Certication
USA/Canada General Purpose (No hazardous area certication)
EU and Elsewhere General Purpose; Complies with 2014/30/EU EMC
and 2014/35/EU LVD harmonizing directives
bhge.com
© 2017 Baker Hughes, a GE company – All rights reserved.
Baker Hughes reserves the right to make changes in specications and features shown herein, or discontinue the product described at any time without notice or obligation. Contact your BHGE representative for the most current information. The Baker Hughes logo is a trade mark of Baker Hughes, a GE company. The GE Monogram is a trademark of the General Electric Company.
920-685A