This product complies with regulatory requirements of the following European
Directive 93/42/EEC concerning medical devices.
This manual is a reference for the Vivid S70 and Vivid S60 ultrasound systems and
covers the following models: Vivid S70 v203 and Vivid S60 v203. It applies to all
revisions of the 203 software for the Vivid S70 and Vivid S60 ultrasound systems,
which will hereafter be listed as Vivid S70 / S60. All information in this manual is
relevant for the two systems unless otherwise specified.
Nativ Ha’or Street no.1
3508510, Haifa
ISRAEL
Tel: (+972) 4 8519 555 Fax: (+972) 4 8519 500
GE Medical Systems SCS
283 rue de la Minière, 78530 BUC, France
Page 3
Revision History
Reason for change
DATE
REV
Rev. 012019-01-30Initial release
(YYYY-MM-DD)
List of Effective Pages
PAGE NUMBERREV
All pagesRev. 01
Please verify that you are using the latest revision of this document. Information
pertaining to this document is maintained on ePDM (GE electronic Product Data
Management). If you need to know the latest revision, contact your distributor, local GE
Sales Representative or in the USA call the GE Ultrasound Clinical Answer Center at
1 800 682 5327 or 1 262 524 5698.
REASON FOR CHANGE
Vivid S70 / S60 – User Manuali-1
BC092760-1EN01
Page 4
Regulatory Requirements
Conformance Standards
The GE Healthcare product families are tested to meet all
applicable requirements and relevant standards per the
countries in which the product will be sold. Any changes to
accessories, peripheral units, or any other part of the system
must be approved by the manufacturer: GE Vingmed Ultrasound
AS. Ignoring this advice may compromise the regulatory
approvals obtained for the product.
This product complies with the regulatory requirements of the
following:
Standard/DirectiveScope
93/42/EECMedical Devices Directive (MDD)
2012/19/EUWaste Electrical and Electronic Equipment (WEEE)
2011/65/EUDirective on the restriction of the use of certain hazardous substances in
IEC/EN 62366Medical Devices - Application of usability engineering to medical devices
ISO 10993-1Biological evaluation of medical devices
electrical and electronic equipment (ROHS)
Medical Electrical Equipment - Part 1. General requirements for basic
Safety and essential performance
basic safety and essential performance of ultrasonic medical diagnostic
and monitoring equipment
safety and essential performance - Collateral standard: Electromagnetic
disturbances - Requirements and tests
safety and essential performance - Collateral standard: Usability
indices on diagnostic ultrasound equipment.
i-2Vivid S70 / S60 – User Manual
BC092760-1EN
01
Page 5
Certifications
GE Medical Systems Türkiye Ltd. Şti.
Esentepe Mah. Harman Sok. No: 8
34394 Şişli İstanbul Türkiye
Türkiye
İthalatçısı /
Turkish
Importer
• Quality management standards for medical devices: GE
Importer Information
• TURKEY
• BRAZIL
Medical Systems, Israel Ltd. and GE Medical Systems,
China Co. Ltd. are ISO13485 certified.
Brazilian
Importer
GE Healthcare do Brasil Comércio e Serviços
para Equipamentos Médico- Hospitalares Ltda
Av. Magalhães de Castro, 4800
Andar 11 Conj. 111 e 112, Andar 12 Conj. 121 e
122, Torre 3 - Cidade Jardim
São Paulo SP – CEP: 05676-120
C.N.P.J.: 00.029.372/0001-40
Número de Registro ANVISA: 80071260360
Directives
The GE ultrasound product families are tested to meet all
applicable requirements in relevant EU Directives and
European/International standards.
• Council Directive 93/42/EEC concerning MDD (Medical
Devices Directive): the CE label affixed to the product
testifies compliance to this Directive.
The location of the CE marking is specified in ‘Device labels’
on page 2-26.
• Year of first CE mark: 2015
Classifications
According to 93/42/EEC Medical Device Directive, this is a
Class IIa Medical Device.
The following classifications are in accordance with the IEC/
EN 60601-1:
• According to IEC/EN 60601-1, Equipment is Class I, with BF
or CF Applied Parts.
Vivid S70 / S60 – User Manuali-3
BC092760-1EN01
Page 6
Class I Equipment
Type of protection against electric shock:
• Class I Equipment
Degree of protection against electric shock:
• Type BF Applied part (for Probes marked with BF symbol)
• Type CF Applied part (for ECG, and probes marked with CF
symbol)
• The ECG applied part is protected against defibrillation.
Continuous Operation
System is IPX0.
Footswitch is IPX8.
Transducers are IPX7.
ICE catheters are IPX8.
EQUIPMENT in which protection against electric shock does not
rely on BASIC INSULATION only, but also includes an earth
ground. This additional safety precaution prevents exposed
metal parts from becoming LIVE in the event of an insulation
failure.
Type BF Applied part
TYPE BF APPLIED PART providing a specified degree of
protection against electric shock with particular regard to
allowable LEAKAGE CURRENT.
Patient leakage current
Normal conditionSingle fault condition
d.c.<10 microA<50 microA
a.c.<100 microA<500 microA
i-4Vivid S70 / S60 – User Manual
BC092760-1EN
01
Page 7
Type CF Applied part
TYPE CF APPLIED PART providing a degree of protection
higher than that for Type BF Applied Part against electric shock
particularly regarding allowable LEAKAGE CURRENT.
The Vivid S70 / S60 is a high perfo rmance digital
ultrasound imaging system with tot al d at a manage ment.
The system provides image generation in 4D (option on
Vivid S70), 2D (B) Mode, Color Doppler, Power Doppler
(Angio), M-Mode, Color M-Mode, PW and CW Doppler
spectral, Tissue Velocity imaging, advanced Strain and
Contrast applications.
The fully digital architecture of the Vivid S70 / S60
system allows optimal usage of all scanning mo des and
probe types, throughout the full spectrum of operating
frequencies.
This chapter provides information concerning
indications for use/contraindications, contact
information, and how this documentation is organized.
Vivid S70 / S60 – User Manual1-1
BC092760-1EN01
Page 20
Introduction
Attention
Overview
This manual contains necessary and sufficient information to
operate the ultrasound system safely.
Read and understand all instructions in the User Manual before
attempting to use the ultrasound system.
Keep this manual with the equipment at all time. Periodically
review the procedures for operation and safety precautions.
Disregarding information on safety is considered abnormal use.
Not all features or products described in this document may be
available or cleared for sale in all markets. Please contact your
local GE Healthcare representative to get the latest information.
NOTE: Please note that orders are based on the individually agreed
upon specifications and may not contain all featur es listed in this
manual.
NOTE: All references to standards / regulations and their revisions are
valid at the time of publication of the user manual.
1-2Vivid S70 / S60 – User Manual
BC092760-1EN
01
Page 21
Documentation
CAUTION
Overview
Safety instructions must be reviewed before operating the
system.
Vivid S70 / S60 documentation consists of various manuals:
• The User Manual and Onboard Help (TRANSLATED)
provides information needed by the user to operate the
system safely. It describes the basic functions of the system,
safety features, operating modes, measurements/
calculations, probes, and user care and maintenance.
• The Advanced Reference Manual (ENGLISH ONLY)
contains data tables, such as measurements and
calculations, OB, and Acoustic Output.
• The Service Manual (ENGLISH ONLY) supplies block
diagrams, lists of spare parts, descriptions, adjustment
instructions, or similar information which helps qualified
technical personnel in repairing those parts of the system
which have been defined as repairable.
• Medical Ultrasound Safety publication from American
Institute of Ultrasound in Medicine (AIUM) (ENGLISH
ONLY). Provided as ALARA Educational Program, to
comply with US FDA Track 3 - Not available in all countries.
NOTE: The eDocumentation kit provides instructions on how to read the
user documenation via electronic media. All user manuals are
provided in electronic format. The eDocumenation media
includes English and all other translations.
The Vivid S70 / S60 manuals are written for users who are
familiar with basic ultrasound principles and techniques. They
do not include sonographic training or detailed clinical
procedures.
NOTE: The screen graphics in this manual are only for illustrational
purposes. Actual screen output may differ.
Conventions used in this manual
Bold type describes button names on the screen.
Italic type describes program windows, screens and dialogue
boxes.
Icons highlight safety issues as described in ‘Safety’ on
page 2-1.
Vivid S70 / S60 – User Manual1-3
BC092760-1EN01
Page 22
Introduction
Indications for use
The Vivid S70 / S60 is a general-purpose ultrasound system,
specialized for use in cardiac imaging. It is intended for use by,
or under the direction of a qualified physician for ultrasound
imaging and analysis of Fetal/Obstetrics; Abdominal (Including
Renal and GYN); Pediatric; Small Organ (breast, testes,
thyroid); Neonatal Cephalic; Adult Cephalic; Cardiac (adult and
pediatric); Peripheral Vascular; Musculo-skeletal Conventional;
Musculo-skeletal Superficial; Urology (including prostate);
Transesophageal; Transvaginal; Transrectal; Intra-cardiac;
Intra-luminal; and Intraoperative (abdominal, thoracic, &
vascular).
Clinical QuestionsFor information in the United States, Canada, Mexico, and parts
of the Caribbean, call the Customer Answer Center.
TEL: (1) 800-682-5327 or (1) 262-524-5698
In other locations, contact your local Applications, Sales, or
Service Representative.
Service QuestionsFor service in the United States, call GE CARES.
TEL: (1) 800-437-1171
In other locations, contact your local Service Representative.
Information
Requests
Placing an OrderTo order accessories, supplies, or service parts in the United
To request technical product information in the United States,
call GE.
TEL: (1) 800-643-6439
In other locations, contact your local Applications, Sales, or
Service Representative.
States, call the GE Technologies Contact Center.
TEL: (1) 800-558-5102
In other locations, contact your local Applications, Sales, or
Service Representative.
1-8Vivid S70 / S60 – User Manual
BC092760-1EN
01
Page 27
AMERICAS
A
A
GE Healthcare
Ultrasound Service Engineering
9900 Innovation Drive
Wauwatosa, WI 53226
TEL: (1) 800-668-0732
Customer Answer Center TEL: (1) 262-524-5698
GE Healthcare
Ultrasound Service Engineering
9900 Innovation Drive
Wauwatosa, WI 53226
TEL: (1) 262-524-5300
Customer Answer Center TEL: (1) 262-524-5698
GE Healthcare
Ultrasound Service Engineering
9900 Innovation Drive
Wauwatosa, WI 53226
TEL: (1) 800-437-1171 FAX: (1) 414-721-3865
ARGENTINA
CANADA
Contact Information
GE Healthcare TEL: 11-5298-2400
Nicolas Vedia 3616, piso 5
Buenos Aires
BRAZIL
GE Healthcare do Brasil Comércio e Serviços para
Equipamentos Médico- Hospitalares Ltda
v. Magalhães de Castro, 4800, Andar 11 Conj. 111 e 112,
ndar 12 Conj. 121 e 122, Torre 3 - Cidade Jardim
CEP: 05676-120 - São Paulo/SP - Brasil
C.N.P.J.: 00.029.372/0001-40
TEL: 3004 2525 (Capitais e regiões metropolitanas) /
0800 165 799 (Demais regiões)
LATIN & SOUTH
AMERICA
MEXICO
USA
GE Sistemas Medicos de Mexico S.A. de C.V.
Rio Lerma #302, 1º y 2º Pisos
Colonia Cuauhtemoc FAX: (5) 211-4631
06500-Mexico, D.F.
TEL: (5) 228-9600
Vivid S70 / S60 – User Manual1-9
BC092760-1EN01
Page 28
Introduction
A
ASIA
ASIA PACIFIC
AUSTRALIA
KOREA
NEW ZEALAND
SINGAPORE
JAPAN
CHINA
GE Healthcare Asia Pacific
4-7-127, Asahigaoka
Hinoshi, Tokyo
191-8503, Japan
TEL: +81 42 585 5111
GE Healthcare
32 Philip Street
Parramatta, NSW 2150
TEL: 1300 722 229 or +612 9846 4000
GE Healthcare - China
No. 1, Yongchang North Road
Beijing Economic & Technology Development Area
Beijing 100176, China
TEL: (8610) 5806 8888 FAX: (8610) 6787 1162
GE Healthcare Korea
15F, 416 Hangang Dae ro, Chung-gu
Seoul, 04637, Korea
TEL: +82 1544 6119
8 Tangihua Street
uckland 1010
New Zealand
TEL: 0800 434 325
GE HEALTHCARE PTE LTD
1 Maritime Square #13-01
HarbourFront Center
Singapore 099253
TEL: +65 6291 8528
BC092760-1EN
01
1-10Vivid S70 / S60 – User Manual
Page 29
EUROPE and MIDDLE EAST
GE Healthcare BVBA
Eagle Building
Kouterveldstraat 20
1831 DIEGEM
TEL: (+32) 2 719 7204 FAX: (+32) 2 719 7205
GE Healthcare
8-10 Sorou Str. Marousi
Athens 15125 Hellas
TEL: (+30) 210 8930600 FAX: (+30) 210 9625931
For all other European countries not listed, please contact your
local GE Healthcare distributor or the appropriate support
resource listed on www.gehealthcare.com.
AUSTRIA
General Healthcare Austria GmbH & Co OG
Technologiestr. 10, Euro Plaza Gebäude E
1120 Wien
TEL: (+43) 1 97272 0 FAX: (+43) 1 97272 2222
BELGIUM &
LUXEMBURG
CZECH REPUBLIC
GE Medical Systems
Bucharova 14/2641, Explora Business
158 00 Praha
DENMARK
GE Medical Systems Ultrasound
Park Alle 295, 2605 Brøndby
TEL: (+45) 43 295 400 FAX: (+45) 43 295 399
ESTONIA &
FINLAND
FRANCE
GERMANY
GREECE
GE Healthcare Finland Oy
Kuortaneenkatu 2, 000510 Helsinki
P.O.Box 330, 00031 GE Finland
TEL: (+358) 10 39 48 220 FAX: (+358) 10 39 48 221
GE Medical Systems SCS
24 Avenue de l'Europe - CS20529
78457 Velizy Villacoublay Cedex
TEL: (+33) 13 449 50 00 FAX: (+33) 13 44 95 202
GE Healthcare España
C/ Gobelas 35-37
28023 Madrid
TEL: (+34) 91 663 2500 FAX: (+34) 91 663 2501
BC092760-1EN
01
1-12Vivid S70 / S60 – User Manual
Page 31
SWEDEN
A
SWITZERLAND
TURKEY
UNITED ARAB
EMIRATES (UAE)
UNITED KINGDOM
Contact Information
GE Healthcare Sverige AB
FE 314, SE-182 82 Stockholm, Sweden
TEL: (+46) 8 559 50010
GE Medical Systems (Schweiz) AG
Europastrasse 31,
8152 Glattbrugg
TEL: (+41) 1 809 92 92 FAX: (+41) 1 809 92 22
GE Healthcare Türkiye
Istanbul Office
Levent Ofis
Esentepe Mah. Harman Sok.
No:8 Sisli-Istanbul
TEL: +90 212 398 07 00 FAKS: +90 212 284 67 00
GE Healthcare Holding ME SA
Dubai Internet City, Building No. 18
P. O. Box # 11549, Dubai
U.A.E
TEL: (+971) 4 429 6161
FAX (+971) 4 429 6200/01/02
GE Healthcare
mersham Place
Little Chalfont, Bucks, HP7 9NA
TEL: (+44) (0) 1494 544000
Vivid S70 / S60 – User Manual1-13
BC092760-1EN01
Page 32
Introduction
Manufacturer
GE Medical Systems, Israel Ltd.
Nativ Ha’or Street no.1
3508510, Haifa, ISRAEL
Tel.: (+972) 4 8519 555 Fax: (+972) 4 8519 500
1-14Vivid S70 / S60 – User Manual
BC092760-1EN
01
Page 33
Chapter 2
Safety
Describes the safety and regulatory information
pertinent to operating this ultrasound system as well as
procedures for simple care and maintenance of the
system.
‘Safety Precautions’ on page 2-2
‘Owner responsibility’ on page 2-3
‘Acoustic output’ on page 2-4
‘Important safety considerations’ on page 2-11
‘Device labels’ on page 2-26.
Vivid S70 / S60 – User Manual2-1
BC092760-1EN01
Page 34
Safety
DANGER
WARNING
CAUTION
Precaution Levels
Icon description
Safety Precautions
Various levels of safety precautions may be found on the
equipment, and different levels of severity are identified by one
of the following icons that precede precautionary statements in
the text.
Indicates that a specific hazard is known to exist which through
inappropriate conditions or actions will cause:
• Severe or fatal personal injury
• Substantial property damage
Indicates that a specific hazard is known to exist which through
inappropriate conditions or actions may cause:
• Severe personal injury
• Substantial property damage
Indicates that a potential hazard may exist which through
inappropriate conditions or actions will or can cause:
• Minor injury
• Property damage
NOTE: Indicates precautions or recommendations that should be used
in the operation of the ultrasound system, specifically:
• Maintaining an optimum system environment
• Using this Manual
• Notes to emphasize or clarify a point
2-2Vivid S70 / S60 – User Manual
BC092760-1EN
01
Page 35
Owner responsibility
CAUTION
Owner responsibility
For USA only:
CAUTION: Federal law restricts this device to sale by or on the
order of a physician.
It is the responsibility of the owner to ensure that anyone
operating the system reads and understands this section of the
manual. However, there is no representation that the act of
reading this manual renders the reader qualified to operate,
inspect, test, align, calibrate, troubleshoot, repair, or modify the
system. The owner should make certain that only properly
trained, fully-qualified service personnel undertake the
installation, maintenance, troubleshooting, calibration, and
repair of the equipment.
The owner of the ultrasound system should ensure that only
properly trained, fully qualified personnel are authorized to
operate the system. Before authorizing anyone to operate the
system, it should be verified that the person has read, and fully
understands, the operating instructions contained in this
manual. It is advisable to maintain a list of authorized operators.
Should the system fail to operate correctly, or not respond to the
commands described in this manual, the operator should
contact GE Service.
For information about specific requirements and regulations
applicable to the use of electronic medical equipment, consult
local, state, and federal agencies.
Notice against user modification
Never modify this product, including system components,
software, cables, etc. User modification may cause safety
hazards and degradation in system performance. All
modification must be done by a GE qualified person.
Vivid S70 / S60 – User Manual2-3
BC092760-1EN01
Page 36
Safety
Acoustic output
Definition of the acoustic output parameters
Thermal Index
TI is an estimate of the temperature increase of soft tissue or
bone. There are three thermal index categories:
• TIS: Soft tissue thermal index. The main TI category. Used
for applications that do not image bone.
• TIB: Bone thermal index (bone located in a focal region).
Used for fetal application.
• TIC: Cranial bone thermal index (bone located close to the
surface). Used for transcranial application.
Mechanical Index
Ispta
The Vivid S70 / S60 has an internal limit of 3.0 on TI for the
chosen index category.
MI is the estimated likelihood of tissue damage due to cavitation.
The absolute maximum regulatory limit of the MI is 1.9 as set by
the FDA 510(k) guidance of September 9, 2008.
The maximum MI values obtainable with Vivid S70 / S60 in
relevant operating modes, and with available ultrasound probes,
are found in the Reference manual.
The Ispta is the Spatial Peak Temporal Average Intensity. The
absolute maximum regulatory limit of Ispta is 720 mW/cm
set by the FDA 510(k) guidance of September 9, 2008.
2
as
2-4Vivid S70 / S60 – User Manual
BC092760-1EN
01
Page 37
Acoustic output and display on the Vivid S70 / S60
In the title bar, two fields are allocated for the display of power
values as shown in Figure 2-1.
1. Title bar
2. MI
3. TI
Figure 2-1. The display of MI and TI on the screen
The Vivid S70 / S60 chooses the correct category based on
mode of operation and chosen application, and presents the
relevant TI category to the operator. It is therefore important that
the operator chooses the right application. The system also
provides the ability to select the display of any of the TI
categories regardless of the current application.
The British Medical Ultrasound Society has suggested some
maximum scanning times relative to displayed TI as follows:
Acoustic output
Obstetric scanning
TItimeNote
0.0–0.7UnlimitedMonitor TIS up to 10 weeks post LMP, TIB thereafter
0.7–1.0< 60 min
1.0–1.5< 30 min
1.5–2.0< 15 min
2.0–2.5< 4 min
2.5–3.0< 1 min
Vivid S70 / S60 – User Manual2-5
BC092760-1EN01
Page 38
Safety
Neonatal trans-cranial & spinal scanning
TItimeNote
0.0–0.7UnlimitedMonitor TIC. MI>0.7 should be used with caution in the presence of
0.7–1.0< 60 min
1.0–1.5< 30 min
1.5–2.0< 15 min
2.0–2.5< 4 min
2.5–3.0< 1 min
contrast agents.
Neonatal general and cardiac scanning
TItimeNote
0.0–1.0UnlimitedMonitor TIB. MI>0.7 should be used with caution in the presence of
1.0–1.5< 120 min
1.5–2.0< 60 min
2.0–2.5< 15 min
2.5–3.0< 4 min
contrast agents.
Adult trans-cranial scanning
TItimeNote
0.0–1.0UnlimitedMonitor TIC. MI>0.7 should be used with caution in the presence of
1.0–1.5< 30 min
1.5–2.0< 15 min
2.0–2.5< 4 min
2.5–3.0< 1 min
contrast agents.
General Abdominal, Peripheral Vascular and other scanning (excluding the eye)
TItimeNote
0.0–1.0UnlimitedMonitor TIB or TIC if bone closer than 1 cm, TIS if no bone is in the
1.0–1.5< 120 min
1.5–2.0< 60 min
2.0–2.5< 15 min
2.5–3.0< 4 min
image. MI>0.7 should be used with caution in the presence of
contrast agents.
2-6Vivid S70 / S60 – User Manual
BC092760-1EN
01
Page 39
ALARA
Training
Acoustic output
NOTE: The Vivid S70 / S60 does not monitor the duration of thermal
exposure.
References
• The British Medical Ultrasound Society. Guidelines for the
safe use of diagnostic ultrasound equipment.
• American Institute of Ultrasound in Medicine Consensus
Report on Potential Bioeffects of Diagnostic Ultrasound.
The maximum possible MI and Ispta on the Vivid S70 / S60 is
within the limits set in Track 3 in the FDA 510(k) guide of
September 9, 2008, MI < 1.9 and Ispta < 720 mW/cm
Ultrasound procedures should be performed using output levels
and exposure times As Low As Reasonably Achievable
(ALARA) while acquiring clinical information.
During each ultrasound examination the user is expected to
weigh the medical benefit of the diagnostic information that
would be obtained against the risk of potentially harmful effects.
Once an optimal image is achieved, the need for increasing
acoustic output or prolonging the exposure cannot be justified. It
is recommended that all users receive proper training in
applications before performing them in a clinical setting. Contact
the GE sales representative for training assistance.
2
.
Safety statement
GE safety statement
Although no harmful biological effects have been demonstrated
for ultrasound frequencies, intensities, or exposure times used
in examination with the GE system, GE recommends using the
lowest acoustic output settings which will produce diagnostically
acceptable information.
Vivid S70 / S60 – User Manual2-7
BC092760-1EN01
Page 40
Safety
System controls affecting acoustic output
The operator controls that directly affect the acoustic output are
discussed in the Acoustic Output Data Tables in the Reference
Manual. These tables show the highest possible acoustic
intensity for a given mode, obtainable only when the maximum
combination of control settings is selected. Most settings result
in a much lower output. It is important to note the following:
• The duration of an ultrasound examination is as important
as the acoustic output, since patient exposure to output is
directly related to the exposure time.
• Better image quality yields faster clinical results, making it
possible to complete the relevant ultrasound examination
more rapidly. Therefore, any control that improves the
quality of the examination can help to reduce patient
exposure, even though it may not directly affect acoustic
output.
Probe selection
As long as the appropriate application is available, any probe
can be used with the knowledge that the intensities fall at, or
below, those stated in the Acoustic Output Data Tables. The
duration of patient exposure is most likely minimized with the
use of a probe that is optimized to provide resolution and focal
depth, appropriate to the examination.
Application selection
Selecting the probe and application preset appropriate to a
particular ultrasound examination automatically provides
acoustic output limits within FDA guidelines for that application.
Other parameters which optimize performance for the selected
application are also set automatically, and should assist in
reducing the patient exposure time. See ‘Connect and
disconnect probes’ on page 3-26, for information on selecting
probes and application presets.
2-8Vivid S70 / S60 – User Manual
BC092760-1EN
01
Page 41
Changing imaging modes
Acoustic output depends on the imaging mode selected. The
choice of mode (2D, M-Mode, Doppler or Color Flow)
determines whether the ultrasound beam is stationary or in
motion. This greatly affects the energy absorbed by the tissue.
See ‘Image Optimization’ on page 5-1, for complete information
on changing imaging modes.
When operating in a combined mode, such as 2D and M-Mode,
the total acoustic output comprises contributions from each
individual mode. Depending on the modes in use, either or both
output indices may be affected.
The user can override the default settings, but care should be
taken to observe the displayed MI and TI values.
Power
It is possible to change the power in all operating modes so that
the operator can use the ALARA principle.
Acoustic output
Vivid S70 / S60 – User Manual2-9
BC092760-1EN01
Page 42
Safety
OB Exam
Exam Preparation
Prior to an ultrasound examination, the patient should be
informed of the clinical indication, specific benefits, potential
risks, and alternatives, if any. In addition, if the patient requests
information about the exposure time and intensity, it should be
provided. Patient access to educational materials regarding
ultrasound is strongly encouraged to supplement the information
communicated directly to the patient. Furthermore, these
examinations should be conducted in a manner and take place
in a setting which ensures patient dignity and privacy:
• Prior material knowledge and approval of the presence of
nonessential personnel with the number of such personnel
kept to a minimum.
• An intent to share with the parents per the physician's
judgment, either during the examination or shortly hereafter,
the information derived.
• An offer of choice about viewing the fetus.
• An offer of choice about learning the sex of the fetus, if such
information becomes available.
Ultrasound examinations performed solely to satisfy the family's
desire to know the fetal sex, to view the fetus, or to obtain a
picture of the fetus should be discouraged.
2-10Vivid S70 / S60 – User Manual
BC092760-1EN
01
Page 43
Important safety considerations
WARNING
Important safety considerations
This section includes considerations for the following:
• ‘Patient safety’ on page 2-11
• ‘Personnel and equipment safety’ on page 2-15
• ‘Electromagnetic Compatibility (EMC)’ on page 2-19
• ‘Environmental protection’ on page 2-25
The information contained in this section is intended to
familiarize the user with the hazards associated with the use of
the system, and to alert them to the extent to which injury and
damage may occur if the precautions are not observed.
Users are obligated to familiarize themselves with these safety
considerations and to avoid conditions that could result in injury
or damage.
Patient safety
Patient identification
The concerns listed in this section can seriously affect the
safety of the patient undergoing a diagnostic ultrasound
examination.
Always include proper identification with all patient data and
verify the accuracy of the patient's name and/or identity number
when entering such data. Ensure that the correct patient ID is
provided on all recorded data and hard copy prints. Identification
errors could result in an incorrect diagnosis.
Vivid S70 / S60 – User Manual2-11
BC092760-1EN01
Page 44
Safety
CAUTION
CAUTION
Diagnostic information
The images and calculations provided by the system are
intended as a diagnostic tool for competent users. They are
explicitly not to be regarded as the sole, irrefutable basis for
clinical diagnosis. Users are encouraged to study the literature
and reach their own professional conclusions regarding the
clinical utility of the system.
The system provides calculations (e.g. estimated fetal weight)
and charts based on published scientific literature. The
selection of the appropriate chart and clinical interpretation of
calculations and charts is the sole responsibility of the user.
The user must consider contraindications for the use of a
calculation or chart as described in the scientific literature. The
diagnosis, decision for further examinations, and medical
treatment must be performed by qualified personnel following
good clinical practice.
The user should be aware of the product specifications and
system accuracy and stability limitations. These limitations must
be considered before making any decision based on quantitative
values. If in doubt, consult the nearest GE Ultrasound Service
Office.
Equipment malfunction or incorrect settings can result in
measurement errors or failure to detect details in the image. The
user must become thoroughly familiar with the operation of the
system in order to optimize its performance and to recognize
possible malfunctions. Application training is available through
the sales representative.
Be certain to ensure the privacy of patient information.
2-12Vivid S70 / S60 – User Manual
BC092760-1EN
01
Page 45
Mechanical hazards
WARNING
WARNING
CAUTION
WARNING
WARNING
Important safety considerations
Damaged probes or improper use and manipulation of the
transesophageal probe may result in injury or increased risk of
infection. Inspect probes frequently for sharp, pointed, or rough
surface damage that could cause injury or tear protective
barriers (gloves and sheaths). Never use excessive force when
manipulating intracavity probes. Become familiar with all
instructions and precautions provided with special purpose
probes.
Observe probe immersion levels (see Figure 13-5 on page 13-21).
Inspect probes for sharp edges or rough surfaces that could
injure sensitive tissue.
DO NOT bend or pull the cable forcefully, to avoid mechanical
shock or impact to the probe.
Ultrasound transducers are sensitive instruments which can
easily be damaged by rough handling. Take extra care not to
drop transducers and avoid contact with sharp or abrasive
surfaces. A damaged housing, lens, or cable can result in
patient injury or serious impairment of operation.
Transesophageal probe safety
Never use excessive force when manipulating intracavity
probes. Become familiar with all instructions and precautions
provided with special purpose probes.
Electrical Hazard
A damaged probe may increase the risk of electric shock if
conductive solutions come in contact with internal live pads.
Inspect probes often for cracks or openings in the housing and
holes in and around the acoustic lens, or other damage that
could allow moisture to enter. Become familiar with the probe's
use and care precautions outlined in ‘Probes’ on page 13-1.
Vivid S70 / S60 – User Manual2-13
BC092760-1EN01
Page 46
Safety
CAUTION
WARNING
CAUTION
Ultrasound can produce harmful effects in tissue and
potentially result in patient injury. Always minimize exposure
time and keep ultrasound levels low when there is no medical
benefit. Use the principle of ALARA (As Low As Reasonably
Achievable), increasing output only when needed to obtain
diagnostic image quality. Observe the acoustic output display
and be familiar with all controls affecting the output level. See
the Bioeffects section of the Acoustic Output chapter in the
Advanced Reference Manual for more information.
The system and Electrosurgical units
This equipment provides no special means of protection from
high frequency (HF) burns that may result from using an
electrosurgical unit (ESU). To reduce the risk of HF burns,
avoid contact between the patient and ultrasound transducer or
ECG electrodes while operating the ESU. Where contact
cannot be avoided, as in the case of TEE monitoring during
surgery, make sure the transducer or ECG electrodes are not
located between the ESU active and dispersive electrodes and
keep the ESU cables away from the transducer or ECG cables
Defibrillation
Remove any sensors on the patient other than the ECG before
defibrillation.
2-14Vivid S70 / S60 – User Manual
BC092760-1EN
01
Page 47
Personnel and equipment safety
CAUTION
CAUTION
WARNING
WARNING
The hazards listed below can seriously affect the safety of
personnel and equipment during a diagnostic ultrasound
examination.
General hazard
Do not use this equipment if a safety problem is known to exist.
Have the system repaired and performance verified by
qualified service personnel before returning to use.
Only approved and recommended peripherals and accessories
should be used.
All peripherals and accessories must be securely mounted to
the Vivid S70 / S60.
Explosion hazard
Important safety considerations
Electrical hazard
Never operate the equipment in the presence of flammable or
explosive liquids, vapors, or gases. Malfunctions in the system
or sparks generated by fan motors can electrically ignite these
substances. Operators should be aware of the following points
to prevent such explosion hazards.
• If flammable substances are detected in the environment,
do not plug in or turn on the system.
• If flammable substances are detected after the system has
been turned on, do not attempt to turn off the system, or to
unplug it.
• If flammable substances are detected, evacuate and
ventilate the area before turning off the system.
The internal circuits of the system use high voltages, capable of
causing serious injury or death by electrical shock.
Vivid S70 / S60 – User Manual2-15
BC092760-1EN01
Page 48
Safety
CAUTION
NOTE: Any rest energy within our scanners or their components will be
To avoid injury
Moving hazard
below 60 V DC or 2 mJ.
• Do not remove the system's protective covers. No
user-serviceable parts are inside. If servicing is required,
contact qualified technical personnel.
• Connect the attachment plug to a hospital-grade grounding
outlet to ensure adequate grounding.
• Do not place liquids on or above the system. Conductive
fluids seeping into the active circuit components may cause
short circuiting, which could result in an electrical fire.
• An electrical hazard may exist if any light, monitor, or visual
indicator remains on after the system is turned off.
Fuses blown within 36 hours of being replaced may indicate a
malfunctioning electrical circuit within the system. In this event,
the system must be checked by GE Ultrasound service
personnel. No attempt should be made to replace the fuses with
others of a higher rating.
The ultrasound system weighs approximately 75 Kg (165 lb.).
Special care must be used to avoid injury when moving or
transporting the system.
• Always be sure the pathway is clear.
• Limit the speed of movement to a careful walk.
• Use at least two people when moving the system on
inclines.
Ensure that the system is well prepared before transporting.
Refer to ‘Moving and transporting the system’ on page 3-48 for
more information.
2-16Vivid S70 / S60 – User Manual
BC092760-1EN
01
Page 49
Biological hazard
CAUTION
Pacemaker hazard
Important safety considerations
For patient and personnel safety, beware of biological hazards
while performing transesophageal procedures. To avoid the
risk of disease transmission:
• Use protective barriers (gloves and probe sheaths)
whenever necessary. Follow sterile procedures as
required.
• Thoroughly clean and disinfect probes after each patient
examination. Refer to Chapter ‘Probes’ on page 13-1, for
probe use and care instructions.
• Reusable accessories should be cleaned and disinfected
or sterilized as stated by the manufacturer, after each
patient.
• Follow all in-house infection control policies as they apply
to personnel and equipment.
Electrical safety
Device classifications
The possibility of the system interfering with pacemakers is
minimal. However, as this system generates high frequency
electrical signals, the operator should be aware of the potential
hazard this could cause.
See ‘Classifications’ on page i-3 for more information.
Vivid S70 / S60 – User Manual2-17
BC092760-1EN01
Page 50
Safety
CAUTION
CAUTION
Peripheral devices
Use only GE Healthcare approved internal equipment when
replacing an internal
peripheral.
External
related IEC/EN standards for safety. The electrical medical
systems conformance to IEC/EN60601-1 Clause 16 must be
verified.
All non-medical electrical equipment must be kept outside of
the patient environment, as defined in IEC/EN60601-1, unless
it is equipped with additional protective earth or extra
separating transformer. Commercial devices such as laser
cameras, printers, VCRs and external monitors, usually exceed
allowable leakage current limits per IEC/EN60601-1 and, when
plugged into separate AC outlets, are in violation of patient
safety standards. Suitable electrical isolation of such external
AC outlets, or providing the device with extra protective earth,
will be required in order to meet IEC/EN60601-1 standards for
electrical leakage currents.
Internally connected peripheral devices
The system, together with peripheral devices, such as video
printer, meets ANSI AAMI ES60601-1 and IEC/EN 60601-1
standards for electrical isolation and safety. These standards are
applicable only when the specified peripheral devices are
plugged into the AC outlets provided on the system.
External Connection of other peripheral devices
peripheral equipment must be in compliance with
External devices can be used only if CE marked and in
compliance with related IEC or ISO Safety standards.
Conformance to IEC/EN60601-1 Clause 16 requirements for
Medical Electrical systems must be verified.
Accessory equipment connected to the analog and digital
interfaces must be certified according to the respective IEC or
ISO Safety standards (e.g. IEC/EN60950-1 for data processing
equipment or IEC/EN60601-1 for medical equipment).
Furthermore all complete configurations shall comply with the
requirements for Medical Electrical systems of IEC/EN60601-1.
Anybody connecting additional equipment to the signal input
part or signal output part of the ultrasound system configures a
medical electrical system, and is therefore responsible that the
system complies with the requirements of the IEC/EN60601-1
Clause 16. If in doubt, consult the technical service department
or your local GE representative.
2-18Vivid S70 / S60 – User Manual
BC092760-1EN
01
Page 51
Important safety considerations
WARNING
CAUTION
CAUTION
Any devices or cables, other than those sold with the
ultrasound system, connected to the Peripheral /accessory
connector panel or to an USB port on the system may result in
an increase of the electromagnetic emission from the system,
or a decrease of the electromagnetic immunity of the system.
When using peripheral device, observe all warnings and
cautions given in peripheral operator manuals.
Allergic reactions to latex-containing medical devices
Contact with natural rubber latex may cause a severe
anaphylactic reaction in persons sensitive to the natural latex
protein. Sensitive users and patients must avoid contact with
these items.
Due to reports of severe allergic reactions to medical devices
containing latex (natural rubber), the FDA advises healthcare
professionals to identify latex-sensitive patients, and be
prepared to treat allergic reactions promptly. Latex is a
component of many medical devices, including surgical and
examination gloves, catheters, incubation tubes, anesthesia
masks, and dental dams. Patient reaction to latex has ranged
from contact urticaria to systemic anaphylaxis.
For more details regarding allergic reaction to latex, refer to FDA Medical Alert MDA91-1, March 29.
Electromagnetic Compatibility (EMC)
NOTE: This system carries the CE mark. It complies with regulatory
requirements of the European Directive 93/42/EEC concerning
medical devices. It also complies with emission limits for a
Group 1, Class A Medical Device as stated in EN 60601-1-2
(IEC 60601-1-2).
NOTE: The emissions characteristics of this equipment make it suitable
for use in industrial areas and hospitals, e.g. professional
healthcare environment (CISPR 11 class A). If it is used in a
residential environment (for which CISPR 11 class B is normally
required) this equipment might not offer adequate protection to
radio-frequency communication services. The user might need
Vivid S70 / S60 – User Manual2-19
BC092760-1EN01
Page 52
Safety
to take mitigation measures, such as relocating or re-orienting
the equipment.
Electrical medical equipment requires special precautions
regarding EMC and must be installed and implemented
according to the EMC information provided in this manual.
AII types of electronic equipment may characteristically cause
electromagnetic interference with other equipment, transmitted
either through air or connecting cables. The term
Electromagnetic Compatibility (EMC) indicates the capability of
the equipment to curb electromagnetic influence from other
equipment, while at the same time not affecting other equipment
with similar electromagnetic radiation.
Radiated or conducted electromagnetic signals can cause
distortion, degradation, or artifacts in the ultrasound image
which may impair the ultrasound system’s essential
performance (see page 2-25).
There is no guarantee that interference will not occur in a
particular installation. If this equipment is found to cause or
respond to interference, attempt to correct the problem by one
or more of the following measures:
• Re-orient or re-locate the affected device.
• Increase the separation between the system and the
affected device.
• Power the equipment from a source other than that of the
affected device.
• Consult your service representative for further suggestions.
The manufacturer is not responsible for any interference or
response caused by the use of interconnecting cables other
than those recommended or by unauthorized changes or
modifications to this system. Unauthorized changes or
modifications could void the user's authority to operate the
equipment.
To comply with the regulations on electromagnetic interference,
all interconnecting cables to peripheral devices must be
shielded and properly grounded. Use of cables not properly
shielded and grounded may result in the equipment causing or
responding to radio frequency interference, in violation of the
European Union Medical Device Directive and FCC regulations.
2-20Vivid S70 / S60 – User Manual
BC092760-1EN
01
Page 53
Interference caution
CAUTION
WARNING
WARNING
WARNING
Important safety considerations
Use of devices that transmit radio waves near the system could
cause it to malfunction.
Devices which intrinsically transmit radio waves, such as cellular
phones, radio transceivers, mobile radio transmitters,
radio-controlled toys, etc., should preferably not be operated
near the system.
Medical staff in charge of the system are required to instruct
technicians, patients, and other people who may be around the
system to fully comply with the above recommendations.
Any electrical device can unintentionally emit electromagnetic
waves. However, minimum device separation distances cannot
be calculated for such unspecified radiation. When the
ultrasound system is used adjacent to or in close proximity to
other equipment the user should be attentive to unexpected
device behavior which may be caused by such radiation.
The ultrasound system is intended for use in the
electromagnetic environment specified in the tables below.
The user of the ultrasound system should ensure that the device
is used in such an environment.
The use of accessories, transducers and cables other than
those specified, with the exception of transducers and cables
sold by the manufacturer of the Vivid S70 / S60 as replacement
parts for internal components, may result in increased
electromagnetic emissions or decreased electromagnetic
immunity of the Vivid S70 / S60.
The use of the accessory, transducer or cable with Vivid S70 /
S60 other than those specified may result in increased
emissions or decreased immunity of the Vivid S70 / S60.
The Vivid S70 / S60 should not be used adjacent to or stacked
with other equipment and that if adjacent or stacked use is
necessary, the Vivid S70 / S60 should be observed to verify
normal operation in the configuration in which it will be used.
Vivid S70 / S60 – User Manual2-21
BC092760-1EN01
Page 54
Safety
WARNING
Portable RF communications equipment (including peripherals
such as antenna cables and external antennas) should be
used no closer than 30 cm (12 inches) to any part of the Vivid
S70 / S60 series system, including cables specified by the
manufacturer. Otherwise, degradation of the performance of
this equipment could result.
Electromagnetic emissions
The Vivid S70 / S60 is intended for use in the electromagnetic
environment specified below. The customer or the user of the
Vivid S70 / S60 should ensure that it is used in such an
environment.
Guidance and manufacturer’s declaration – electromagnetic emissions.
Voltage fluctuations/flicker
emissions
IEC 61000-3-3
Group 1The ultrasound system uses RF energy only for its
Class AThis system is suitable for use in all establishments,
Class A
Complies
Electromagnetic immunity
The Vivid S70 / S60 is intended for use in the electromagnetic
environment specified below. The customer or the user of the
Vivid S70 / S60 should ensure that it is used in such an
environment,
internal function. Therefore, its RF emissions are very
low and are not likely to cause any interference in nearby
electronic equipment.
other than domestic establishments and those directly
connected to the public low-voltage power supply
network that supplies buildings used for domestic
purposes, provided the following warning is heeded:
Warning: This system is intended for use by healthcare
professionals only. This system may cause radio
interference or disrupt the operation of nearby
equipment. It may be necessary to take mitigating
measures, such as re-orienting or relocating the system
or shielding the location.
2-22Vivid S70 / S60 – User Manual
BC092760-1EN
01
Page 55
Important safety considerations
Guidance and manufacturer’s decleration - electromagnetic immunity.
IEC 60601-1-2 test
Immunity test
level
Compliance level
Electromagnetic
environment -
guidance
Electrostatic discharge
(ESD)
IEC 61000-4-2
Electrical transients/
bursts
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short
interruptions, and
voltage variations on
power supply input lines
IEC 61000-4-11
±8 kV contact
±15 kV air (max)
±2 kV for power-supply
lines
±1 kV for input/output
lines
100 kHz Repetition
Frequency
Controlled shutdown
with return to
pre-disturbance
condition after
operator‘s intervention.
(Power-on switch)
Floors should be wood,
concrete or ceramic tile.
If floors are covered
with synthetic material,
the relative humidity
should be at least 30%.
Mains power quality
should be that of a
typical commercial or
hospital environment
Mains power quality
should be that of a
typical commercial or
hospital environment.
Mains power quality
should be that of a
typical commercial or
hospital environment. If
the user of the
ultrasound system
requires continued
operation during power
mains interruptions, it is
recommended that the
ultrasound system is
powered from an
uninterruptible power
supply or a battery.
Power frequency (50/60
Hz) magnetic field
IEC 61000-4-830 A/m
NOTE: U
is the a. c. mains voltage prior to application of the test level.
T
50 Hz or 60 Hz
30 A/m
50 and 60Hz
Power frequency
magnetic fields should
be at levels
characteristic of a
typical location in a
typical commercial or
hospital environment.
Vivid S70 / S60 – User Manual2-23
BC092760-1EN01
Page 56
Safety
Guidance and manufacturer’s decleration - electromagnetic immunity.
Immunity testIEC 60601-1-2 test levelCompliance level
Conducted RF
IEC 61000-4-6
Radiated RF and Proximity fields
from RF wireless
communications equipment
IEC 61000-4-3
3 Vrms
150 kHz to 80 MHz
3 V/m; 80 MHz to 2.7 GHz
80% AM at 1 kHz
385 MHz (18 Hz Pulse
Modulation)
450 MHz (FM +/ -5 kHz deviation
1 kHz sine or 18 Hz Pulse
Modulation)
710 MHz (217 Hz PM)9 V/m
745 MHz (217 Hz PM)9 V/m
780 MHz (217 Hz PM)9 V/m
810 MHz (18 Hz PM)28 V/m
870 MHz (18 Hz PM)28 V/m
930 MHz (18 Hz PM)28 V/m
1720 MHz (217 Hz PM)28 V/m
1845 MHz (217 Hz PM)28 V/m
1970 MHz (217 Hz PM)28 V/m
2450 MHz (217 Hz PM)28 V/m
3 Vrms
150 kHz to 80 MHz
3 V/m; 80 MHz to 2.7 GHz
80% AM at 1 kHz
27 V/m
28 V/m
5240 MHz (217 Hz PM)9 V/m
5500 MHz (217 Hz PM)9 V/m
5785 MHz (217 Hz PM)9 V/m
NOTE: Electromagnetic propagation is affected by absorption and reflection from structures, objects and
people.
2-24Vivid S70 / S60 – User Manual
BC092760-1EN
01
Page 57
Essential performance
The essential performance of the ultrasound system is:
• the ability to display physiological images as input for
diagnosis by trained physician.
• the ability to display physiological traces as aid for diagnosis
by trained physician.
• the ability to display quantified data as input for diagnosis by
trained physician.
• the display of ultrasound indexes as aid for safe use of the
system.
• the display of probe surface temperature as aid for safe use
of the system (Probe dependent).
Environmental protection
System disposal
Please follow the disassembly and part disposal procedure
attached inside the system. To access to the procedure, remove
the right side panel by unscrewing the two screws on the lower
part.
Important safety considerations
Vivid S70 / S60 – User Manual2-25
BC092760-1EN01
Page 58
Safety
0459
Symbols
Device labels
The following table describes the purpose and location of safety
labels and other important information provided on the
equipment.
LabelPurposeLocation
Identification PlateManufacturer's name and address
Model
Device Listing/Certification Labels
On/off button
Warning: System shutdown using the
On/Off button does not disconnect the
ultrasound system from mains
voltage. For disconnecting the
ultrasound system from mains voltage
after system shutdown, please set the
circuit breaker close to the mains inlet to
OFF as described on page 3-23.
Equipment Type BF, in which protection
against electric shock does not rely on
basic insulation only. Provides additional
safety precautions such as double
insulation or reinforced insulation,
because there is no provision for
protective earthing or reliance upon
installation conditions.
Defibrillator-proof Type CF equipment.ECG
Standard
RearN/A- by GE
Healthcare
Control panelIEC 60417-5010
Probes / Rear
of system
connector
IEC 60417-5333
IEC 60417-5336
Indicates that the product is in
compliance with all relevant European
Directives and under surveillance by
Notified Body 0459.
Rear of
system
N/A- by
certification
body
2-26Vivid S70 / S60 – User Manual
BC092760-1EN
01
Page 59
Device labels
LabelPurposeLocation
For USA only:
Caution: Federal law restricts this device
to sale by or on the order of a physician.
TUV SÜD
Classification Label
Mark name: "Eurasian Conformity" mark;
the single conformity mark for circulation
of products on the markets of memberstates of Customs Union.
Mark meaning: This product passed all
conformity assessment (approval)
procedures that correspond to the
requirements of applicable technical
reguations of the Customs Union.
GOST-R Mark: per Law of the Russian
Federation No. 184-FZ.
Non-ionizing electromagnetic radiation.Rear of
Rear of
system
Rear of
system
Rear of
system
Rear of
system
system
Standard
N/A- by GE
Healthcare
N/A- by
certification
body
N/A- by
certification
body
N/A- by
certification
body
IEC 60417-5140
Alternating currentVariousIEC 60417-5032
Protective earth (ground)InternalIEC 60417-5019
Equipotentiality: Indicates terminal to be
used for connecting equipotential
conductors when interconnecting
(grounding) with other equipment as
described in IEC60601-1.
Follow instructions for use. Read and
understand all instructions in the User's
Manual before attempting to use the
ultrasound system.
Symbol indicating that the Instructions for
Use are supplied in electronic form.
CAUTION - Dangerous voltage: Used to
indicate electric shock hazards.
Rear of
system
Rear of
system
Rear of
system
Rear of
system
IEC 60417-5021
ISO 7010-M002
ISO 7000-1641
ISO 7010-W012
Vivid S70 / S60 – User Manual2-27
BC092760-1EN01
Page 60
Safety
LabelPurposeLocation
Attention - Consult accompanying
documents: Alerts the user to refer to the
user documentation when complete
information cannot be provided on the
label.
The system is not designed for use with
flammable anesthetic gases.
CAUTION - Do not push the system
sideways when casters are in break
position. Instability may occur.
DO NOT place objects on the surface of
the rear of the main display while folded.
This symbol indicates that the waste of
electrical and electronic equipment must
not be disposed as unsorted municipal
waste and must be collected separately.
Please contact the manufacturer or other
authorized disposal company to
decommission your equipment.
The disassembly and parts disposition
procedure is located on the card cage
front cover. To access to the procedure,
remove the right side panel.
VariousISO 7010-W001
Rear of
system
Top console
(both sides)
Display rear
panel
Rear of
system
Standard
N/A- by GE
Healthcare
ISO 7010-P017
ISO 7010-P012
EN 50419
System weightRear of
Date of manufacture: The date could be a
year, year and month, or year, month and
day, as appropriate. See ISO 8601 for
date formats.
Manufacturer name and addressRear of
Serial numberRear of
Brand and model identifierRear of
Unique Device Identification (UDI). Every
system has a unique marking for
identification. Scan or enter the UDI
information into the patient health record
as required by governing laws.
Device part number identifierRear of
system
Rear of
system
system
system
system
Rear of
system
system
N/A- by GE
Healthcare
ISO 7000-2497
ISO 7000-3082
ISO 7000-2498
ISO 7000-2493
N/A- by GE
Healthcare
N/A- by GE
Healthcare
2-28Vivid S70 / S60 – User Manual
BC092760-1EN
01
Page 61
Device labels
LabelPurposeLocation
Assembled in XXXXX
(XXXXX represents the
country name)
Identify the Customs Country of Origin of
the materials.
To indicate the acceptable upper and
lower limits of atmospheric pressure for
transport and storage.
To identify the temperature limits, for
example on transport packaging to
indicate limits within which the package
has to be kept and handled. The
temperature values may be shown
adjacent to the symbol.
To indicate the acceptable upper and
lower limits of relative humidity for
transport and storage.
On transport packaging. To indicate that
the items are not to be vertically stacked.
On transport packaging. To indicate the
correct upright position.
Rear of
system
Package
labeling
Package
labeling
Package
labeling
Package
labeling
Package
labeling
Standard
N/A- by GE
Healthcare
ISO 7000-2621
ISO 7000-0632
ISO 7000-2620
ISO 7000-2402
ISO 7000-0623
On transport packaging. To indicate that
the content of the package is fragile and
that the package must be handled with
care.
On transport packaging. To indicate that
the the package must be handled with
care.
On transport packaging. To indicate that
the package must be kept in dry
conditions.
Package
labeling
Package
labeling
Package
labeling
ISO 7000-0621
N/A
ISO 7000-0626
Vivid S70 / S60 – User Manual2-29
BC092760-1EN01
Page 62
Safety
For China only
2-30Vivid S70 / S60 – User Manual
BC092760-1EN
01
Page 63
Device labels
Vivid S70 / S60 – User Manual2-31
BC092760-1EN01
Page 64
Safety
2-32Vivid S70 / S60 – User Manual
BC092760-1EN
01
Page 65
Chapter 3
Prepare the System for Use
Describes the site requirements, system overview,
powering on the system, system positioning/
transporting, operator controls, probes, and adjusting
the main display.
‘Site requirements’ on page 3-2
‘System Overview’ on page 3-9
‘Switching On/Off’ on page 3-23
‘Connect and disconnect probes’ on page 3-26
‘The Scanning screen’ on page 3-32.
‘Control panel’ on page 3-35.
‘Main display adjustment’ on page 3-44.
‘Moving and transporting the system’ on page 3-48.
Vivid S70 / S60 – User Manual3-1
BC092760-1EN01
Page 66
Prepare the System for Use
WARNING
WARNING
Only qualified physicians or ultrasound sonographers should
perform scans of patients for medical diagnostic reasons.
Request training, if needed. Ensure that unauthorized personnel
do not tamper with the system.
Service representatives authorized by GE Ultrasound will
unpack and install the system. Do not attempt to install the
system alone.
All the warnings in Chapter ‘Safety’ on page 2-1, should be
read and understood before operating the system.
Site requirements
Power requirements
To carry out regular preventative maintenance refer to Chapter
‘System Care and Maintenance’ on page 15-2.
Maintain a clean environment. Turn off the circuit breaker before
cleaning the system. Refer to ‘System Care and Maintenance’
on page 15-2 for cleaning instructions.
The ultrasound system must operate within the proper
environment and in accordance with the requirements described
in this section. Before using the system, ensure that the
requirements are met.
Optimal operation of the system can be obtained by
implementing the following requirements:
The ultrasound system uses a separate power outlet for 100–
240 VAC, 50/60 Hz.
Operating the system with the wrong voltage range causes
damages, voiding the factory warranty.
3-2Vivid S70 / S60 – User Manual
BC092760-1EN
01
Page 67
Site requirements
Operating Environment
Ensure that there is sufficient air flow around the ultrasound
system when installed or operated.
Environmental requirements
The ultrasound system requires constant maintenance of its
operational environment. Different temperature and humidity
requirements are specified for operation, storage and
transportation.
RequirementTempe ratureHumidityAir Pressure
Operational10–35 °C30–85%700–1060 hPa
Storage-20–60 °C10–95%700–1060 hPa
Transport-20–60 °C10–95%700–1060 hPa
NOTE: System may be operated at an altitude of up to 3000 meters.
Electromagnetic interferences
The ultrasound system is approved for use in hospitals, clinics,
and other environmentally qualified facilities, in terms of the
prevention of radio wave interference. Operation of the system
in an inappropriate environment can cause electronic
interference to radios and television sets situated near the
medical equipment.
Ensure that the system is protected from electromagnetic
interferences as follows:
• Operate the system at least 4.5 meters (15 feet) away from
equipment that emits strong electromagnetic radiation.
• Shield the system when operating it in the vicinity of radio
broadcasting equipment, if necessary.
Vivid S70 / S60 – User Manual3-3
BC092760-1EN01
Page 68
Prepare the System for Use
WARNING
WARNING
WARNING
Connecting the system
A GE-qualified person should perform the initial system
installation.
Connecting the ultrasound system involves preliminary checks
of the power cord, voltage level, and compliance with electrical
safety requirements.
Use only power supply cords, cables, and plugs provided by or
designated by GE.
Ensure that the power cord and plug are intact and that the
power plug is the proper hospital-grade type (where required).
The system should be connected to a fixed power socket which
has the protective grounding connector. Never use an
extension cord or adapter plug.
Failure to provide an adequate earth circuit can cause electrical
shock, resulting in serious injury.
Connection of additional protective earth conductors or
potential equalization conductors is not necessary in most
cases and is only recommended for situations involving
multiple equipment in a high-risk patient environment to
provide assurance that all equipment is at the same potential
and operates within acceptable leakage current limits. An
example of a high-risk patient would be a special procedure
during which the patient has an accessible conductive path to
the heart such as exposed cardiac pacing leads.
3-4Vivid S70 / S60 – User Manual
BC092760-1EN
01
Page 69
Voltage level check
WARNING
Site requirements
Check the rating label on the rear side of the system (see
Figure 3-1).
Check the voltage range indicated on the label:
• 100–240 VAC, 50/60 Hz, 500 VA
At 110 V: 6.4 A
At 115 V: 6.0 A
At 230 V: 3.0 A
Figure 3-1. The rating label location
If the mains supply is not within the specified range, do not
connect the system to the power source. Contact your local
distributor or the appropriate support resource.
Vivid S70 / S60 – User Manual3-5
BC092760-1EN01
Page 70
Prepare the System for Use
WARNING
Connecting to the electrical outlet
POWER OUTAGE MAY OCCUR. The ultrasound system
requires a dedicated single branch circuit. To avoid circuit
overload and possible loss of critical care equipment, make
sure you DO NOT have other equipment operating on the
same circuit.
The system’s power must be supplied from a separate,
properly rated outlet to avoid risk of fire. Refer to ‘Power
requirements’ on page 3-2 for rating information.
The power cord should not, under any circumstances, be
altered to a configuration rated less than that specified for the
current.
Do not use an extension cord or adapter plug.
1. Ensure that the wall outlet is of appropriate type, and that
the power switch is turned off.
2. Uncoil the power cable, allowing sufficient slack so that the
system can be moved slightly.
3. Attach the power plug to the system and secure it in place
by using the retaining clamp.
4. Secure the power plug to the wall outlet.
1. Ground socket
2. Retaining clamp for power plug
3. Circuit breaker
Figure 3-2. Power plug
3-6Vivid S70 / S60 – User Manual
BC092760-1EN
01
Page 71
Site requirements
CAUTION
CAUTION
WARNING
Ensure that the retaining clamp for the power plug is fixed
firmly.
Use caution to ensure that the power cable does not
disconnect during system use. If the system is accidentally
unplugged and the optional battery is not installed, data may be
lost.
To avoid risk of fire, the system power must be supplied from a
separate, properly rated outlet.
Under no circumstances should the AC power plug be altered,
changed, or adapted to a configuration rated less than
specified. Never use an extension cord or adapter plug.
To help ensure grounding reliability, connect to a "hospitalgrade"
or "hospital only" grounded power outlet.
1. 100-120 VAC, 1200 VA Plug and Outlet Configuration
2. 220-240 VAC, 1200 VA Plug and Outlet Configuration
Figure 3-3. Example Plug and Outlet Configurations
Battery
The System contains an internal battery (optional), which is not
user-accessible. The internal battery acts as a limited alternative
power source, which becomes active when the AC power fails or
AC cable is abruptly pulled out.
Display turns off and no scanning is possible while in battery
mode.
Battery technology
The lithium ion technology used in the system’s battery is
significantly less hazardous to the environment than the lithium
metal technology used in some other batteries (such as watch
Vivid S70 / S60 – User Manual3-7
BC092760-1EN01
Page 72
Prepare the System for Use
batteries). Used batteries should not be placed with common
household waste products. Contact local authorities for the
location of a chemical waste collection program nearest you.
NOTE: The battery may only be disassembled and re-assembled by an
authorized field-service engineer.
Viewing Current Battery Status
When the system is running, a status icon is displayed in the
system Status bar to indicate the current battery status.
IconStatus Description
Table 3-1: Battery status icons
Battery is fully charged (80%-100%)
Battery is partially charged (40%-80%)
Battery is almost empty (10%-40%)
NOTE: The % values mentioned above may fluctuate by up to +/- 3%
points.
View Detailed Battery Status
The status of the battery charge (% charge) appears on the
touch panel during Tran sportation mode.
3-8Vivid S70 / S60 – User Manual
BC092760-1EN
01
Page 73
Getting to know the system
System Overview
System Overview
1. Main display monitor
2. Touch panel
3. Rear handle
4. Probe and gel holder on both sides of control panel
5. Pull-out alphanumeric Keyboard (option)
6. Front handle
7. Up/down "Flex-Fit" arm
8. Probe connectors (DLP type)
9. Probe-cable management tray (removable)
10. Front-wheel with foot-break
11. Probe connector (RS type for TEE probe)
12. ECG cable connector
13. Doppler Pencil Probe Connector
14. Probe-cable hooks
15. Up/Down & Left/right swivel release lever
16. Front USB port
17. Control panel
18. Articulated arm for display monitor
Figure 3-4. Vivid S70 / S60 (front)
Vivid S70 / S60 – User Manual3-9
BC092760-1EN01
Page 74
Prepare the System for Use
1. Main display monitor
2. Touch panel
3. Probe and gel holder on both sides of control panel
4. Rear handle
5. Rear I/O connectors panel
6. System labels
7. Swivel rear wheels
8. Circuit-breaker, power and ground
connectors
9. Ventilation port
10. Black and white video printer (optional)
11. CD/DVD drive (option)
12. Doppler speakers
13. Articulated arm for display monitor
Figure 3-5. Vivid S70 / S60 (rear)
3-10Vivid S70 / S60 – User Manual
BC092760-1EN
01
Page 75
Removable media
CAUTION
Intended use
System Overview
Removable media can be used for the following purposes:
• Long-term image storage: The final destination of the
images, after they are moved out of the system harddisk by
using the Disk Management feature (see page 10-48).
• Backup of patient database and system configuration
presets (see page 10-56)
• Patient archive sneakernet: Copy a set of patient records
between a system and EchoPAC Software Only using the
Transfer feature (see page 10-40) with removable media.
• DICOM transfer to copy a set of patient records to/from a
third party DICOM review station.
• XML export: Exports demographics, measurements, and
reporting data from the system to a third party reporting
application using removable media (see page 10-40).
• Copy of system configuration presets between two units
using the Backup/Restore feature (see page 10-56).
• Save images as JPEG, MPEG, AVI, DICOM, or RawDICOM
for review on a regular computer.
Supported removable media
The following removable media are supported:
• CD-R (option)
• DVD-R (option)
• USB Flash card
Archive removable media written from a Vivid S70 / S60 using
the 203 software cannot be read by a Vivid S70 / S60 using
older software.
NOTE: Depending on the system configuration, USB mass storage may
be disabled on the system. Enabling/disabling of USB mass
storage can only be done by users with administrator right s: Sys
Admin or GE Admin users (see ‘Local System Users’ on
page 12-8). When USB mass storage is disabled, the following
icon is shown on the status bar:
Vivid S70 / S60 – User Manual3-11
BC092760-1EN01
Page 76
Prepare the System for Use
CAUTION
USB Flash card:
• Use only shielded USB Flash cards that are verified for
EMC performance according to EN55011/EN55022. The
use of other USB Flash cards may cause interference on
the system itself or on other electronic devices.
Recommendation concerning CD and DVD handling
To avoid data loss, never touch the recordable surface of a disk.
Handle the disk only by the outer edge. Do not place it face
down on a hard surface. Fingerprints or scratches will make the
disk unusable. Before usage, verify that the disk surface has no
visible scratches. If there are any scratches, do NOT use the
disk.
Formatting removable media
To format removable media:
1. Insert the media in the drive.
2. Press Utility/Config on the Touch panel.
3. If required, log on to the system.
4. Select the category Connectivity and select the sheet
Tools (Figure 3-6).
Figure 3-6. The Tools sheet
5. Select the removable media from the Media drop-down
menu (CD-R, DVD-R or USB device).
3-12Vivid S70 / S60 – User Manual
BC092760-1EN
01
Page 77
System Overview
CAUTION
NOTE: Select Refresh if the media does not appear on the list.
6. Enter a name for the removable media in the Label field.
NOTE: Only the following characters and signs can be used when
labeling a media: A - Z, a - z, 0 - 9, “_” and “-”. Do not use
more than 11 characters or signs. Do not use space.
7. Select Format.
A confirmation window is displayed.
The formatting process will erase any data present on the
media.
8. Select OK to continue.
9. Wait for the display of the Information window indicating that
the formatting process is completed.
10. Select OK.
11. Eject the media as described below.
NOTE: Removable media used during Disk space management,
Backup, Export, or Save as do not need to be formatted in
advance as the formatting process is part of these procedures if
required.
Ejecting removable media
1. Press Utility/Eject on the Touch panel.
NOTE: Do not eject CD/DVD using the button on the CD/DVD drive.
The Eject device menu is displayed (Figure 3-7).
2. Select the relevant media.
Wait for the display of the Information window indicating that
the device can be safely removed before removing the
media.
Figure 3-7. The Eject device menu
Vivid S70 / S60 – User Manual3-13
BC092760-1EN01
Page 78
Prepare the System for Use
CAUTION
CAUTION
Physiological traces
The physiological module consists of two channels: ECG and
Respiration. The scanned image that is displayed is
synchronized with the ECG and respiration traces. In M-Mode or
Doppler, the traces are synchronized to that particular mode's
sweep.
Use only GE Healthcare accessories.
Conductive parts of electrodes and associated connectors for
applied parts, including neutral electrodes should not contact
other conductive parts, including earth.
Simultaneous use of two or more applied parts will cause
summation of patient leakage currents.
Heart rate may be adversely affected by cardiac arrhythmias or
the operation of cardiac pacemaker pulses.
Figure 3-8. The patient (I/O) connection panel
3-14Vivid S70 / S60 – User Manual
BC092760-1EN
01
Page 79
ECG/Respiration
System Overview
The ECG cable is a modular cable consisting of two distinct
parts:
• The Trunk: A single cable connecting to the system at one
end, providing a cable splitter device at the other end (see
Figure 3-9).
• The triple color-coded electrode cable: To be inserted in
the splitter device. Each electrode cable hooks up to the
appropriate stick-on electrode by a color-coded clip type
connector.
The color-coding of the electrodes follows one of two standards
that are common in different parts of the world. The cable splitter
device illustrates the color codes, names, and body locations for
the two standard color codes (see Figure 3-9).
AHA (USA)IEC (Europe, Asia, ROW)
1. RA: White
2. LA: Black
3. LL: Red
1. R: Red
2. L: Yellow
3. F: Green
Figure 3-9. The cable splitter device with electrode placement conventions
Vivid S70 / S60 – User Manual3-15
BC092760-1EN01
Page 80
Prepare the System for Use
CAUTION
To connect the internal ECG
1. Connect the ECG trunk cable to the rectangular socket
marked ECG on the patient trace (I/O) panel. The patient
trace (I/O) panel is located at the front left of the ultrasound
system (see Figure 3-8).
2. Hook up the electrode cables to the electrodes, following the
appropriate convention (see Figure 3-9).
Using Physio
The use of respiratory monitoring might interfere with the
proper operation of impedance-based rate-responsive
pacemakers. When examining a patient with a pacemaker, it is
required to turn off the respiratory trace on the system. This will
block off the respiratory monitoring signals which might
interfere with the pacemaker.
Exception: When using an external respiratory monitor, turning
off the respiratory trace on the system WILL NOT block the
operation of the external monitor.
Display ECG trace
Figure 3-10. ECG controls
3-16Vivid S70 / S60 – User Manual
BC092760-1EN
01
Page 81
1. The ECG is displayed by default in all cardiac applications.
NOTE: The ECG button shows/hides the ECG trace. It does not
2. The following controls can be adjusted:
Removing/adding Triggers
In case of a noisy ECG the system could generate an erroneous
QRS marker which produces a heart cycle loop in a wrong
location. The user can correct this manually by adjusting the
actual location of the relevant QRS marker.
System Overview
For all other applications, press the tab Physio on the Touch
panel. and press ECG (Figure 3-10).
turn off the ECG.
• Horiz. Sweep: Adjusts the sweep speed stepwise (1, 2,
3, 4, 6, 8, 12, and 16 sec. The default setting is 4 sec.)
• Gain: Adjusts the trace amplitude.
• Position: Moves the trace vertically.
• ECG Lead: Selects the desired lead:
- Lead 1: RA (-) to LA (+) (right, left, or lateral)
- Lead 2: RA (-) to LL (+) (superior inferior)
- Lead 3: LA (-) to LL (+) (superior inferior)
• QRS visible: Show/Hide the QRS marker on the ECG.
To remove a QRS marker on a replayed loop:
1. Press QRS visible to display the QRS markers and identify
2. Press 2D Freeze to stop the cineloop and use the trackball
3. Press Remove Trig.
To add a trigger:
1. Place the small red bar on the place where a QRS marker is
2. Press Add Trig.
3. Press 2D Freeze and store a copy of the cineloop with the
Display respiratory trace
1. Press the Physio tab on the Touch panel.
2. Press Respiratory.
3. The following controls can be adjusted:
any extra QRS marker to be removed.
to place the small red bar in the vicinity of the QRS marker
to be removed.
missing.
The new QRS marker is added.
new QRS markers.
The respiratory signal appears in addition to the ECG signal.
Vivid S70 / S60 – User Manual3-17
BC092760-1EN01
Page 82
Prepare the System for Use
• Lead selection: Select the one that provides a stronger
• Horiz. Sweep: Adjusts the sweep speed.
• Gain: Adjusts the trace amplitude.
• Position: Moves the trace vertically.
signal
3-18Vivid S70 / S60 – User Manual
BC092760-1EN
01
Page 83
Peripheral/accessory connector panel
WARNING
The Peripheral/accessory connector panel is located at the rear
of the system (see Figure 3-11).
System Overview
1. Ethernet LAN connector
2. Insulated USB connector (supporting USB 1.0)
3. Dual USB connector (supporting USB 2.0)
4. DVI display-out connector
Figure 3-11. Peripheral/accessory connector panel
Accessory equipment connected to the analog and digital
interfaces must be certified according to the respective IEC or
ISO Safety standards (e.g. IEC/EN 60950 for data processing
equipment and IEC/EN 60601-1 for medical equipment). Any
person connecting additional equipment to the signal input part
or output part of the ultrasound system configures the medical
system and is therefore responsible that the system complies
with the requirements of the medical electrical systems of IEC/
EN 60601-1 Clause 16. If in doubt, consult the technical
service department or your local representative.
Do not touch the conducting parts of the USB or Ethernet
cables when connecting equipment to the system.
Vivid S70 / S60 – User Manual3-19
BC092760-1EN01
Page 84
Prepare the System for Use
WARNING
Any devices or cables, other than those sold with the
ultrasound system, connected to the Peripheral /accessory
connector panel, or to an USB port on the system may result in
an increase in the electromagnetic emission from the system,
or a decrease in the electromagnetic immunity of the system.
SocketSignal typeDevice typeNote
Insulated USBUniversal serial bus
supporting USB 1.0
USB - Dual socketUniversal serial bus
supporting USB 2.0
DVIDigital signals onlyDigital monitorThis is a DVI-I connector, but
Ethernet1000 Base-TX
Ethernet IEEE 802.3
PrinterPrinters with external power
supply will connect here
Memory devicesDevices without
supply will connect here
there is no analog signal
transmission. Only digital
monitors can be connected.
Network device
external power
3-20Vivid S70 / S60 – User Manual
BC092760-1EN
01
Page 85
Wired Footswitch (Option)
CAUTION
You can attach this Footswitch to the system by connecting it to
one of the USB port on the rear of the system.
To avoid damage of the cable, keep the cable away from the
wheels. Disconnect the Footswitch before moving the system.
System Overview
Figure 3-12. Footswitch and USB Cable
This is a 3-pedal Footswitch. You can configure its functionality
from the Config/Imaging/Application (see page 12-22).
The graphic on the bottom of the main screen can be configured
to show the functional assignment of the different pedals. See
the example in Figure 3-13.
Figure 3-13. Footswitch functional assignment
Vivid S70 / S60 – User Manual3-21
BC092760-1EN01
Page 86
Prepare the System for Use
CAUTION
When using the Footswitch, DO NOT hold down the footswitch
pedal. Press and release the Footswitch pedal. Pushing and
holding down the pedal behaves the same way as pushing and
holding down a key on the keyboard.
3-22Vivid S70 / S60 – User Manual
BC092760-1EN
01
Page 87
Switch on the system
1. Switch on the circuit breaker at the rear of the system (see
2. Press the on/off button on the top left of the control panel
Switching On/Off
Switching On/Off
Figure 3-14).
(see Figure 3-14).
After initialization the default scanning screen is displayed.
1. On/Off switch
2. Control Panel
3. Circuit breaker
Figure 3-14. The Circuit breaker and On/Off button
Log In
During the boot-up process the system may require the user to
enter a password to operate it. Personal IDs and associated
passwords can be configured in the Vivid S70 / S60 (‘Users and
Security’ on page 12-8) .
If IDs and passwords have been entered and "Use Auto Logon"
is Off (‘Auto logon’ on page 12-11), the Operator Login window
appears, requesting for an ID and password when Power up
sequence is completed, or when it is required.
NOTE: By factory Login default, the Operator ID is USR and there is no
need to enter a password.
Vivid S70 / S60 – User Manual3-23
BC092760-1EN01
Page 88
Prepare the System for Use
Switch off the system
When the ultrasound system is switched off, the system
performs an automatic shutdown sequence. It is recommended
to perform a full shutdown at least once per week. This will
prevent extreme memory fragmentation, which might cause
some computer slowdown.
Shutdown
1. Press the on/off button on the top left of the control panel.
The Exit dialogue window is displayed.
Figure 3-15. The Exit dialogue window
The system can be switched off into one of the three states:
• Shutdown: the entire system is shut down. Full
shutdown is recommended if the system is not intended
to be used for a whole day or longer.
• Standby mode: In Standby mode most of the system is
powered down, but a certain portion of the system
remains energized. Standby mode allows a shorter
reboot time when the system is used on a daily basis or
moved from one place to another.
• Transportation mode: When AC power is interrupted,
the main screen and keyboard are powered down. The
touch panel and the rest of the system is still powered
by an internal battery. When AC power is resumed, the
system will turn ON in less than 5 seconds (‘Using
Transportation mode’ on page 3-51).
3-24Vivid S70 / S60 – User Manual
BC092760-1EN
01
Page 89
Switching On/Off
WARNING
Battery performance may degrade over time and limit the use
of transportation mode. In case the battery can not be fully
charged, or can not support transportation mode for more than
20 minutes please arrange for service of the battery.
NOTE: In case of total lockup of the system, hold the on/off button
down a few seconds to turn the system off.
2. Select Shutdown.
The shutdown process takes a few seconds and is
completed when the color of the on/off button changes from
green to amber.
3. Select Standby.
The systems enters Standby mode.
Booting up from Standby mode takes less than a minute.
NOTE: After switching off the system, wait at least ten seconds before
turning it on again.
NOTE: Whenever the system is fully shut down or put in Stand by mode,
the system automatically performs "End-Exam" to save all data
and images of the current patient into the archiving system.
When switching the system off to move it, follow the additional
steps below:
1. Wait until the on/off button has turned amber and set the
circuit breaker to OFF.
2. Remove the plug from the mains power socket.
3. Secure the system power cable around the cable storage
hooks at the rear of the system.
Vivid S70 / S60 – User Manual3-25
BC092760-1EN01
Page 90
Prepare the System for Use
Connect and disconnect probes
Connect the probe
Probes can be connected at any time, whether the system is on
or off.
The system has two types of probe ports: one RS probe port
and three DLP probe ports (Figure 3-16).
1. DLP Probe port
2. RS Probe port
Figure 3-16. Probe ports
1. Before connecting the probe:
• Inspect the probe and the probe cable for any damage.
• Do a visual check of the probe pins and system sockets.
Remove any dust or foam remains from the probe pins.
2. Hold the probe connector vertically with the cable pointing
upward.
3. Turn the connector locking handle counterclockwise.
4. Align the connector with the probe port and carefully push
into place.
5. Turn the locking handle clockwise to the full vertical position
to lock in place.
6. Position the probe cable so that it is not resting on the floor.
3-26Vivid S70 / S60 – User Manual
BC092760-1EN
01
Page 91
Connect and disconnect probes
CAUTION
CAUTION
WARNING
CAUTION
Do not allow the probe head to hang freely. Impact to the probe
head may result in irreparable damage.
Take the following precautions with the probe cables:
• Keep free from the wheels.
• Do not bend.
• Do not cross cables between probes.
Do NOT touch the patient and any of the connectors on the
ultrasound system simultaneously, including ultrasound probe
connectors.
NOTE: Further information on handling of ICE and TEE probes is
located in the particular user manuals supplied with these
probes.
Connect the 6Tc-RS or the 9T-RS TEE probe
Transesophageal probes require a special handling. Refer to
the user documentation enclosed with these probes.
The 6Tc-RS and the 9T-RS probes are equipped with a Vivid i
connector. To connect these probes to the ultrasound system,
use the RS probe port(see Figure 3-16).
To connect a probe (Small connector - type RS)
1. Hold the probe connector vertically with the cable pointing
upward.
2. Push the connector locking handle to the right-most
position.
3. Align the connector with the probe port and carefully push
into place.
4. Push the connector locking handle to the left-most position.
5. Position the probe cable so that it is not resting on the floor.
Vivid S70 / S60 – User Manual3-27
BC092760-1EN01
Page 92
Prepare the System for Use
CAUTION
WARNING
Figure 3-17. RS Probe Connection Locking Lever
Handle the probes gently while connecting and disconnecting
Do NOT touch the patient and any of the connectors on the
ultrasound unit simultaneously, including ultrasound probe
connectors.
Activate the probe
When a probe is connected to the system it is automatically
detected.
Selecting a probe and an application
1. Press Probe on the control panel. A list of the connected
probes and application presets is displayed.
3-28Vivid S70 / S60 – User Manual
BC092760-1EN
01
Page 93
Connect and disconnect probes
CAUTION
Figure 3-18. Probe selection
2. Select the desired probe application preset to start
scanning.
Make sure that the probe and application names displayed on
the screen correspond to the actual probe and application
selection.
Check that the correct TI category is displayed. TIB must be
displayed when a fetal application is selected.
Switching Application Preset using QuickApps
QuickApps provides an easy access to specific imaging
settings, such as coronary or contrast imaging, without leaving
the selected imaging preset.
QuickApps keeps the current scanning mode and image
geometry parameters (such as 2D Depth, 2D Width, Color ROI
size and position) as adjusted by the user, while optimizing
imaging parameters for that selected scanning situation.
The Vivid S70 / S60 comes with a set of factory QuickApps
presets for most supported probes and Application Presets.
Factory QuickApps cannot be modified, but user-defined
QuickApps presets can be created as needed.
QuickApps can be managed on the fly as described below:
Vivid S70 / S60 – User Manual3-29
BC092760-1EN01
Page 94
Prepare the System for Use
1. New: To create a new QuickApps preset based on the currently active QuickApps preset or Imaging
preset. Press New and name the new QuickApps preset as desired.
2. Reload: Has two different functions:
• If a factory QuickApp is active, Reload will restore the default factory settings.
• If a user-defined QuickApp is active, Reload will restore the most recently saved settings on that
QuickApp.
3. Save: To overwrite the currently active QuickApps preset. It is only possible to use Save when
working with a user defined QuickApps preset (Save is not available for factory QuickApps presets).
4. Delete: To delete a user defined QuickApps preset. Press Delete and select the QuickApp that
should be eliminated.
Figure 3-19. Manage QuickApps
Deactivate the probe
Press the Freeze key to deactivate the probe. When
deactivating the probe, the probe is automatically placed in
Standby mode.
Disconnect the probe
Probes can be disconnected at any time. It is recommended that
the probe should not be active when being disconnected.
3-30Vivid S70 / S60 – User Manual
BC092760-1EN
01
Page 95
Connect and disconnect probes
To disconnect a probe (Small connector - type RS):
1. Freeze the image by pressing Freeze.
2. Press the connector locking lever towards the left to unlock
the connector.
3. Pull the probe and connector straight out of the probe port.
4. Carefully slide the probe and connector away from the
probe port and around the right side of the keyboard.
5. Ensure the cable is free.
6. Be sure that the probe head is clean before placing the
probe in its storage case.
To disconnect a probe (Large connector - type DLP):
1. Rotate the lock handle counter-clockwise to the horizontal
position to unlock the connector.
2. Pull the connector straight out of the probe port.
Transporting probes
When transporting a probe a long distance, store it in its carrying
case.
Storing probes
It is recommended that all probes be stored in the carrying case
provided.
• First place the probe connector into the carrying case.
• Carefully wind the cable into the carrying case.
• Carefully place the probe head into the carrying case. DO
NOT use excessive force or impact the probe face.
Intra Cardiac Ultrasound Catheters
See ‘Intracardiac Ultrasound Catheters’ on page 13-11 for
information.
Vivid S70 / S60 – User Manual3-31
BC092760-1EN01
Page 96
Prepare the System for Use
The Scanning screen
1. Title bar
- Current patient data
- Institution
- Operator ID
-Probe
- Application
- Mechanical & Thermal Index
2. Measurement result table
3. Depth scale
4. Physiological traces, ECG, Phono, Resp
5. Probe orientation marker
6. Measurement
7. Scanplane indicator (TEE probe/4D probe)
8. Greyscale/Color bar
9. Parameter window
10. Clipboard
11. Prompt/Status information
12. - Trackball assignment
- Footswitch pedal assignments
- TEE probe button assignments
13. - Caps lock on/off
- Access to service platform
- Network status
14. Frame counter and timer
15. Probe surface temperature indicator (only
appears when a TEE probe is selected)
16. Current, real-time display of Heart-rate, Date
and time
Figure 3-20. The scanning screen (composite)
The scanning screen is divided in several areas as follows:
3-32Vivid S70 / S60 – User Manual
BC092760-1EN
01
Page 97
The Title bar
NOTE: The patient information displayed on the Title bar is configurable
Patient Information
Institution name
Operator ID
The Scanning screen
(‘Global imaging settings’ on page 12-17).
From the left:
Displays the information that uniquely identifies the patient, such
as patient name, identification number and birth date. This
information is entered in the New patient window (‘Creating a
new patient record’ on page 4-2).
The institution name is entered from the configuration package
(‘General system settings’ on page 12-4).
Identification code of the operator (‘Users and Security’ on
page 12-8).
Date and time
Displays the current date and time or for a retrieved image, the
date and time at which it was stored.
Probe and Application
Displays the currently selected probe and application or for
retrieved image the probe and application that were used
(‘Selecting a probe and an application’ on page 3-28).
Live scanning related information
Displays, if available, the current values for
• Mechanical Index (MI), for the current active image
• Thermal Index (TI), for the current active image
• Probe temperature (for TEE probe)
• Heart rate (HR)
Vivid S70 / S60 – User Manual3-33
BC092760-1EN01
Page 98
Prepare the System for Use
Archive Information
Displays the currently selected patient and image archives.
Parameters window
Displays scan mode or application specific parameters. In
scanning mode the parameters for the active mode are
highlighted. This window also displays zoom information, stress
template, and image groups in image browser.
Clipboard
Displays the thumbnail images representing the acquired data
during the current examination.
3-34Vivid S70 / S60 – User Manual
BC092760-1EN
01
Page 99
Control Panel overview
Control panel
Control panel
1. Touch panel with adjustment Softkey rotaries
(see page 3-42)
2. On/Off button (see page 3-23)
3. Active mode gain
4. TGC Screen activation
5. 2D Gain
6. Auto (Auto-tissue, ASO)
7. Cursor
8. Scan mode selection
9. Trackball and related buttons (see page 3-37)
10. 4D & Multi D controls (Vivid S70 only)
11. Loudspeaker volume control
12. Flex - programmable button
13. Zoom
14. Clear
15. Depth
16. Print, capture
17. Freeze, 2D Freeze
18. Measurement, Caliper
19. Store
Figure 3-21. The control panel
Vivid S70 / S60 – User Manual3-35
BC092760-1EN01
Page 100
Prepare the System for Use
Key illumination
The keys on the control panel are illuminated according to their
availability:
• Illumination in green: The key function is currently active.
• Illumination in yellow: The key function is available (but
not active) in the current state of the system.
• No illumination: The key is not available in the current state
of the system.
Adjust the Control Panel
The system Control panel can be freely adjusted to swivel or
move up/down. There are two brake handles located under the
control panel.
To swivel the control panel left or right
1. Pull and hold the left brake handle, located under the control
panel.
The control panel can now be freely rotated left or right.
2. When reaching the desired swivel angle, release the
swivel-brake handle.
The control panel will remain at the set angle.
NOTE: The main display will swivel together with the control-panel. In
addition the main display may be swiveled independently of the
control-panel.
To move the control panel up or down
1. Pull and hold the right-hand handle, located under the
control panel.
The control panel can now be freely adjusted up or down.
2. When reaching the desired height, release the height-adjust
handle.
The control panel will remain at the set height.
NOTE: When the control panel is lowered it moves towards the
operator. When panel is raised it also moves away from the
operator.
NOTE: When preparing the system to be moved, pull the left handle and
bring the control panel to a center position. Swivel it slightly till a
locking click is heard.
3-36Vivid S70 / S60 – User Manual
BC092760-1EN
01
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.