GE HEALTHCARE
PART NUMBER FN091019 REVISION 2VS5 N & VS6 N SERVICE MANUAL
Important Precautions
THIS SERVICE MANUAL IS AVAILABLE IN ENGLISH ONLY.
• IF A CUSTOMER’S SERVICE PROVIDER REQUIRES A LANGUAGE OTHER
THAN ENGLISH, IT IS THE CUSTOMER’S RESPONSIBILITY TO PROVIDE
TRANSLATION SERVICES.
WARNING
(EN)
AVERTISSEMENT
(FR)
• DO NOT ATTEMPT TO SERVICE THE EQUIPMENT UNLESS THIS SERVICE
MANUAL HAS BEEN CONSULTED AND IS UNDERSTOOD.
• FAILURE TO HEED THIS WARNING MAY RESULT IN INJURY TO THE SERVICE
PROVIDER, OPERATOR OR PATIENT FROM ELECTRIC SHOCK, MECHANICAL
OR OTHER HAZARDS.
CE MANUEL DE MAINTENANCE N’EST DISPONIBLE QU’EN ANGLAIS.
• SI LE TECHNICIEN DU CLIENT A BESOIN DE CE MANUEL DANS UNE AUTRE
LANGUE QUE L’ANGLAIS, C’EST AU CLIENT QU’IL INCOMBE DE LE FAIRE
TRADUIRE.
• NE PAS TENTER D’INTERVENTION SUR LES ÉQUIPEMENTS TANT QUE LE
MANUEL SERVICE N’A PAS ÉTÉ CONSULTÉ ET COMPRIS.
• LE NON-RESPECT DE CET AVERTISSEMENT PEUT ENTRAÎNER CHEZ LE
TECHNICIEN, L’OPÉRATEUR OU LE PATIENT DES BLESSURES DUES À DES
DANGERS ÉLECTRIQUES, MÉCANIQUES OU AUTRES.
WARNUNG
(DE)
DIESES KUNDENDIENST-HANDBUCH EXISTIERT NUR IN ENGLISCHER
SPRACHE.
• FALLS EIN FREMDER KUNDENDIENST EINE ANDERE SPRACHE BENÖTIGT,
IST ES AUFGABE DES KUNDEN FÜR EINE ENTSPRECHENDE ÜBERSETZUNG
ZU SORGEN.
• VERSUCHEN SIE NICHT, DAS GERÄT ZU REPARIEREN, BEVOR DIESES
KUNDENDIENST-HANDBUCH NICHT ZU RATE GEZOGEN UND VERSTANDEN
WURDE.
• WIRD DIESE WARNUNG NICHT BEACHTET, SO KANN ES ZU VERLETZUNGEN
DES KUNDENDIENSTTECHNIKERS, DES BEDIENERS ODER DES PATIENTEN
DURCH ELEKTRISCHE SCHLÄGE, MECHANISCHE ODER SONSTIGE
GEFAHREN KOMMEN.
- -i
Page 4
PRELIMINARY
GE HEALTHCARE
PART NUMBER FN091019 REVISION 2VS5 N & VS6 N SERVICE MANUAL
ESTE MANUAL DE SERVICIO SÓLO EXISTE EN INGLÉS.
• SI ALGÚN PROVEEDOR DE SERVICIOS AJENO A GEHC SOLICITA UN IDIOMA
QUE NO SEA EL INGLÉS, ES RESPONSABILIDAD DEL CLIENTE OFRECER UN
SERVICIO DE TRADUCCIÓN.
AVISO
(ES)
ATENÇÃO
(PT)
• NO SE DEBERÁ DAR SERVICIO TÉCNICO AL EQUIPO, SIN HABER
CONSULTADO Y COMPRENDIDO ESTE MANUAL DE SERVICIO.
• LA NO OBSERVANCIA DEL PRESENTE AVISO PUEDE DAR LUGAR A QUE EL
PROVEEDOR DE SERVICIOS, EL OPERADOR O EL PACIENTE SUFRAN
LESIONES PROVOCADAS POR CAUSAS ELÉCTRICAS, MECÁNICAS O DE
OTRA NATURALEZA.
ESTE MANUAL DE ASSISTÊNCIA TÉCNICA SÓ SE ENCONTRA DISPONÍVEL EM
INGLÊS.
• SE QUALQUER OUTRO SERVIÇO DE ASSISTÊNCIA TÉCNICA, QUE NÃO A
GEHC, SOLICITAR ESTES MANUAIS NOUTRO IDIOMA, É DA
RESPONSABILIDADE DO CLIENTE FORNECER OS SERVIÇOS DE TRADUÇÃO.
• NÃO TENTE REPARAR O EQUIPAMENTO SEM TER CONSULTADO E
COMPREENDIDO ESTE MANUAL DE ASSISTÊNCIA TÉCNICA.
• O NÃO CUMPRIMENTO DESTE AVISO PODE POR EM PERIGO A SEGURANÇA
DO TÉCNICO, OPERADOR OU PACIENTE DEVIDO A‘ CHOQUES ELÉTRICOS,
MECÂNICOS OU OUTROS.
AVVERTENZA
(IT)
HOIATUS
(ET)
IL PRESENTE MANUALE DI MANUTENZIONE È DISPONIBILE SOLTANTO IN
INGLESE.
• SE UN ADDETTO ALLA MANUTENZIONE ESTERNO ALLA GEHC RICHIEDE IL
MANUALE IN UNA LINGUA DIVERSA, IL CLIENTE È TENUTO A PROVVEDERE
DIRETTAMENTE ALLA TRADUZIONE.
• SI PROCEDA ALLA MANUTENZIONE DELL’APPARECCHIATURA SOLO DOPO
AVER CONSULTATO IL PRESENTE MANUALE ED AVERNE COMPRESO IL
CONTENUTO.
• NON TENERE CONTO DELLA PRESENTE AVVERTENZA POTREBBE FAR
COMPIERE OPERAZIONI DA CUI DERIVINO LESIONI ALL’ADDETTO ALLA
MANUTENZIONE, ALL’UTILIZZATORE ED AL PAZIENTE PER FOLGORAZIONE
ELETTRICA, PER URTI MECCANICI OD ALTRI RISCHI.
KÄESOLEV TEENINDUSJUHEND ON SAADAVAL AINULT INGLISE KEELES.
• KUI KLIENDITEENINDUSE OSUTAJA NÕUAB JUHENDIT INGLISE KEELEST
ERINEVAS KEELES, VASTUTAB KLIENT TÕLKETEENUSE OSUTAMISE EEST.
• ÄRGE ÜRITAGE SEADMEID TEENINDADA ENNE EELNEVALT KÄESOLEVA
TEENINDUSJUHENDIGA TUTVUMIST JA SELLEST ARU SAAMIST.
• KÄESOLEVA HOIATUSE EIRAMINE VÕIB PÕHJUSTADA TEENUSEOSUTAJA,
OPERAATORI VÕI PATSIENDI VIGASTAMIST ELEKTRILÖÖGI, MEHAANILISE
VÕI MUU OHU TAGAJÄRJEL.
-ii -
Page 5
PRELIMINARY
GE HEALTHCARE
PART NUMBER FN091019 REVISION 2VS5 N & VS6 N SERVICE MANUAL
TÄMÄ HUOLTO-OHJE ON SAATAVILLA VAIN ENGLANNIKSI.
• JOS ASIAKKAAN PALVELUNTARJOAJA VAATII MUUTA KUIN
ENGLANNINKIELISTÄ MATERIAALIA, TARVITTAVAN KÄÄNNÖKSEN
HANKKIMINEN ON ASIAKKAAN VASTUULLA.
VAROITUS
(FI)
ΠΡΟΕΙΔΟΠΟΙΗΣΗ
(EL)
• ÄLÄ YRITÄ KORJATA LAITTEISTOA ENNEN KUIN OLET VARMASTI LUKENUT
JA YMMÄRTÄNYT TÄMÄN HUOLTO-OHJEEN.
• MIKÄLI TÄTÄ VAROITUSTA EI NOUDATETA, SEURAUKSENA VOI OLLA
PALVELUNTARJOAJAN, LAITTEISTON KÄYTTÄJÄN TAI POTILAAN
VAHINGOITTUMINEN SÄHKÖISKUN, MEKAANISEN VIAN TAI MUUN
VAARATILANTEEN VUOKSI.
ΤΟΠΑΡΟΝΕΓΧΕΙΡΙΔΙΟΣΕΡΒΙΣΔΙΑΤΙΘΕΤΑΙΣΤΑΑΓΓΛΙΚΑΜΟΝΟ.
• ΕΑΝΤΟΑΤΟΜΟΠΑΡΟΧΗΣΣΕΡΒΙΣΕΝΟΣΠΕΛΑΤΗΑΠΑΙΤΕΙΤΟΠΑΡΟΝ
ΕΓΧΕΙΡΙΔΙΟ ΣΕ ΓΛΩΣΣΑ ΕΚΤΟΣ ΤΩΝ ΑΓΓΛΙΚΩΝ, ΑΠΟΤΕΛΕΙ ΕΥΘΥΝΗ ΤΟΥ
ΠΕΛΑΤΗ ΝΑ ΠΑΡΕΧΕΙ ΥΠΗΡΕΣΙΕΣ ΜΕΤΑΦΡΑΣΗΣ.
• ΜΗΝ ΕΠΙΧΕΙΡΗΣΕΤΕ ΤΗΝ ΕΚΤΕΛΕΣΗ ΕΡΓΑΣΙΩΝ ΣΕΡΒΙΣ ΣΤΟΝ ΕΞΟΠΛΙΣΜΟ
ΕΚΤΟΣ ΕΑΝ ΕΧΕΤΕ ΣΥΜΒΟΥΛΕΥΤΕΙ ΚΑΙ
ΕΧΕΤΕ ΚΑΤΑΝΟΗΣΕΙ ΤΟ ΠΑΡΟΝ
ΕΓΧΕΙΡΙΔΙΟ ΣΕΡΒΙΣ.
• ΕΑΝ ΔΕ ΛΑΒΕΤΕ ΥΠΟΨΗ ΤΗΝ ΠΡΟΕΙΔΟΠΟΙΗΣΗ ΑΥΤΗ, ΕΝΔΕΧΕΤΑΙ ΝΑ
ΠΡΟΚΛΗΘΕΙ ΤΡΑΥΜΑΤΙΣΜΟΣ ΣΤΟ ΑΤΟΜΟ ΠΑΡΟΧΗΣ ΣΕΡΒΙΣ, ΣΤΟ ΧΕΙΡΙΣΤΗ Ή
ΣΤΟΝ ΑΣΘΕΝΗ ΑΠΟ ΗΛΕΚΤΡΟΠΛΗΞΙΑ, ΜΗΧΑΝΙΚΟΥΣ Ή ΑΛΛΟΥΣ ΚΙΝΔΥΝΟΥΣ.
FIGYELMEZTETÉS
(HU)
VIÐVÖRUN
(IS)
EZEN KARBANTARTÁSI KÉZIKÖNYV KIZÁRÓLAG ANGOL NYELVEN ÉRHETŐ EL.
• HA A VEVŐ SZOLGÁLTATÓJA ANGOLTÓL ELTÉRŐ NYELVRE TART IGÉNYT,
AKKOR A VEVŐ FELELŐSSÉGE A FORDÍTÁS ELKÉSZÍTTETÉSE.
• NE PRÓBÁLJA ELKEZDENI HASZNÁLNI A BERENDEZÉST, AMÍG A
KARBANTARTÁSI KÉZIKÖNYVBEN LEÍRTAKAT NEM ÉRTELMEZTÉK.
• EZEN FIGYELMEZTETÉS FIGYELMEN KÍVÜL HAGYÁSA A SZOLGÁLTATÓ,
MŰKÖDTETŐ VAGY A BETEG ÁRAMÜTÉS, MECHANIKAI VAGY EGYÉB
VESZÉLYHELYZET MIATTI SÉRÜLÉSÉT EREDMÉNYEZHETI.
ÞESSI ÞJÓNUSTUHANDBÓK ER EINGÖNGU FÁANLEG Á ENSKU.
• EF ÞJÓNUSTUAÐILI VIÐSKIPTAMANNS ÞARFNAST ANNARS TUNGUMÁLS EN
ENSKU, ER ÞAÐ Á ÁBYRGÐ VIÐSKIPTAMANNS AÐ ÚTVEGA ÞÝÐINGU.
• REYNIÐ EKKI AÐ ÞJÓNUSTA TÆKIÐ NEMA EFTIR AÐ HAFA SKOÐAÐ OG
SKILIÐ ÞESSA ÞJÓNUSTUHANDBÓK.
• EF EKKI ER FARIÐ AÐ ÞESSARI VIÐVÖRUN GETUR ÞAÐ VALDIÐ MEIÐSLUM
ÞJÓNUSTUVEITANDA, STJÓRNANDA EÐA SJÚKLINGS VEGNA RAFLOSTS,
VÉLRÆNNAR EÐA ANNARRAR HÆTTU.
- -iii
Page 6
PRELIMINARY
GE HEALTHCARE
PART NUMBER FN091019 REVISION 2VS5 N & VS6 N SERVICE MANUAL
TENTO SERVISNÍ NÁVOD EXISTUJE POUZE V ANGLICKÉM JAZYCE.
•VPŘÍPADĚ, ŽE POSKYTOVATEL SLUŽEB ZÁKAZNÍKŮM POTŘEBUJE NÁVOD
V JINÉM JAZYCE, JE ZAJIŠTĚNÍ PŘEKLADU DO ODPOVÍDAJÍCÍHO JAZYKA
ÚKOLEM ZÁKAZNÍKA.
VÝSTRAHA
(CS)
ADVARSEL
(DA)
• NEPROVÁDĚJTE ÚDRŽBU TOHOTO ZAŘÍZENÍ, ANIŽ BYSTE SI PŘEČETLI
TENTO SERVISNÍ NÁVOD A POCHOPILI JEHO OBSAH.
•VPŘÍPADĚ NEDODRŽOVÁNÍ TÉTO VÝSTRAHY MŮŽE DOJÍT ÚRAZU
ELEKTRICKÁM PROUDEM PRACOVNÍKA POSKYTOVATELE SLUŽEB,
OBSLUŽNÉHO PERSONÁLU NEBO PACIENTŮ VLIVEM ELEKTRICKÉHOP
PROUDU, RESPEKTIVE VLIVEM K RIZIKU MECHANICKÉHO POŠKOZENÍ NEBO
JINÉMU RIZIKU.
DENNE SERVICEMANUAL FINDES KUN PÅ ENGELSK.
• HVIS EN KUNDES TEKNIKER HAR BRUG FOR ET ANDET SPROG END
ENGELSK, ER DET KUNDENS ANSVAR AT SØRGE FOR OVERSÆTTELSE.
• FORSØG IKKE AT SERVICERE UDSTYRET MEDMINDRE
DENNE SERVICEMANUAL ER BLEVET LÆST OG FORSTÅET.
• MANGLENDE OVERHOLDELSE AF DENNE ADVARSEL KAN MEDFØRE SKADE
PÅ GRUND AF ELEKTRISK, MEKANISK ELLER ANDEN FARE FOR
TEKNIKEREN, OPERATØREN ELLER PATIENTEN.
WAARSCHUWING
(NL)
BRĪDINĀJUMS
(LV)
DEZE ONDERHOUDSHANDLEIDING IS ENKEL IN HET ENGELS VERKRIJGBAAR.
• ALS HET ONDERHOUDSPERSONEEL EEN ANDERE TAAL VEREIST, DAN IS DE
KLANT VERANTWOORDELIJK VOOR DE VERTALING ERVAN.
• PROBEER DE APPARATUUR NIET TE ONDERHOUDEN VOORDAT DEZE
ONDERHOUDSHANDLEIDING WERD GERAADPLEEGD EN BEGREPEN IS.
• INDIEN DEZE WAARSCHUWING NIET WORDT OPGEVOLGD, ZOU HET
ONDERHOUDSPERSONEEL, DE OPERATOR OF EEN PATIËNT GEWOND
KUNNEN RAKEN ALS GEVOLG VAN EEN ELEKTRISCHE SCHOK,
MECHANISCHE OF ANDERE GEVAREN.
ŠĪ APKALPES ROKASGRĀMATA IR PIEEJAMA TIKAI ANGĻU VALODĀ.
• JA KLIENTA APKALPES SNIEDZĒJAM NEPIECIEŠAMA INFORMĀCIJA CITĀ
VALODĀ, NEVIS ANGĻU, KLIENTA PIENĀKUMS IR NODROŠINĀT TULKOŠANU.
• NEVEICIET APRĪKOJUMA APKALPI BEZ APKALPES ROKASGRĀMATAS
IZLASĪŠANAS UN SAPRAŠANAS.
•ŠĪ BRĪDINĀJUMA NEIEVĒROŠANA VAR RADĪT ELEKTRISK
ĀS STRĀVAS
TRIECIENA, MEHĀNISKU VAI CITU RISKU IZRAISĪTU TRAUMU APKALPES
SNIEDZĒJAM, OPERATORAM VAI PACIENTAM.
-iv -
Page 7
PRELIMINARY
GE HEALTHCARE
PART NUMBER FN091019 REVISION 2VS5 N & VS6 N SERVICE MANUAL
ŠIS EKSPLOATAVIMO VADOVAS YRA IŠLEISTAS TIK ANGLŲ KALBA.
• JEI KLIENTO PASLAUGŲ TEIKĖJUI REIKIA VADOVO KITA KALBA – NE ANGLŲ,
VERTIMU PASIRŪPINTI TURI KLIENTAS.
ĮSPĖJIMAS
(LT)
ADVARSEL
(NO)
•NEMĖGINKITE ATLIKTI ĮRANGOS TECHNINĖS PRIEŽIŪROS DARBŲ, NEBENT
VADOVAUTUMĖTĖS ŠIUO EKSPLOATAVIMO VADOVU IR JĮ SUPRASTUMĖTE
• NEPAISANT ŠIO PERSPĖJIMO, PASLAUGŲ TEIKĖJAS, OPERATORIUS AR
PACIENTAS GALI BŪTI SUŽEISTAS DĖL ELEKTROS SMŪGIO, MECHANINIŲ AR
KITŲ PAVOJŲ.
DENNE SERVICEHÅNDBOKEN FINNES BARE PÅ ENGELSK.
• HVIS KUNDENS SERVICELEVERANDØR TRENGER ET ANNET SPRÅK, ER DET
KUNDENS ANSVAR Å SØRGE FOR OVERSETTELSE.
• IKKE FORSØK Å REPARERE UTSTYRET UTEN AT DENNE
SERVICEHÅNDBOKEN ER LEST OG FORSTÅTT.
• MANGLENDE HENSYN TIL DENNE ADVARSELEN KAN FØRE TIL AT
SERVICELEVERANDØREN, OPERATØREN ELLER PASIENTEN SKADES PÅ
GRUNN AV ELEKTRISK STØT, MEKANISKE ELLER ANDRE FARER.
OSTRZEŻENIE
(PL)
ATE NŢIE
(RO)
NINIEJSZY PODRĘCZNIK SERWISOWY DOSTĘPNY JEST JEDYNIE W JĘZYKU
ANGIELSKIM.
•JEŚLI FIRMA ŚWIADCZĄCA KLIENTOWI USłUGI SERWISOWE WYMAGA
UDOSTĘPNIENIA PODRĘCZNIKA W JĘZYKU INNYM NIŻ ANGIELSKI,
OBOWIĄZEK ZAPEWNIENIA STOSOWNEGO TłUMACZENIA SPOCZYWA NA
KLIENCIE.
• NIE PRÓBOWAĆ SERWISOWAĆ NINIEJSZEGO SPRZĘTU BEZ UPRZEDNIEGO
ZAPOZNANIA SIĘ Z PODRĘCZNIKIEM SERWISOWYM.
• NIEZASTOSOWANIE SIĘ DO TEGO OSTRZEŻENIA MOżE GROZIĆ
OBRAŻENIAMI CIAłA SERWISANTA, OPERATORA LUB PACJENTA W WYNIKU
PORAŻENIA PRĄDEM, URAZU MECHANICZNEGO LUB INNEGO RODZAJU
ZAGROŻEŃ.
ACEST MANUAL DE SERVICE ESTE DISPONIBIL NUMAI ÎN LIMBA ENGLEZĂ.
• DACĂ UN FURNIZOR DE SERVICII PENTRU CLIENŢI NECESITĂ O ALTĂ LIMBĂ
DECÂT CEA ENGLEZĂ, ESTE DE DATORIA CLIENTULUI SĂ FURNIZEZE O
TRADUCERE.
• NU ÎNCERCAŢI SĂ REPARAŢI ECHIPAMENTUL DECÂT ULTERIOR
CONSULTĂRII ŞI ÎNŢELEGERII ACESTUI MANUAL DE SERVICE.
• IGNORAREA ACESTUI AVERTISMENT AR PUTEA DUCE LA RĂNIREA
DEPANATORULUI, OPERATORULUI SAU PACIENTULUI ÎN URMA
PERICOLELOR DE ELECTROCUTARE, MECANICE SAU DE ALTĂ NATURĂ.
- -v
Page 8
PRELIMINARY
GE HEALTHCARE
PART NUMBER FN091019 REVISION 2VS5 N & VS6 N SERVICE MANUAL
ДАННОЕ РУКОВОДСТВО ПО ОБСЛУЖИВАНИЮ ПРЕДОСТАВЛЯЕТСЯ ТОЛЬКО
НА АНГЛИЙСКОМ ЯЗЫКЕ.
• ЕСЛИ СЕРВИСНОМУ ПЕРСОНАЛУ КЛИЕНТА НЕОБХОДИМО РУКОВОДСТВО
НЕ НА АНГЛИЙСКОМ ЯЗЫКЕ, КЛИЕНТУ СЛЕДУЕТ САМОСТОЯТЕЛЬНО
ОБЕСПЕЧИТЬ ПЕРЕВОД.
ОПЕРАТОР ИЛИ ПАЦИЕНТ
ПОЛУЧАТ УДАР ЗЛЕКТРИЧЕСКИМ ТОКОМ, МЕХАНИЧЕСКУЮ ТРАВМУ ИЛИ
ДРУГОЕ ПОВРЕЖДЕНИЕ.
TÁTO SERVISNÁ PRÍRUČKA JE K DISPOZÍCII LEN V ANGLIČTINE.
• AK ZÁKAZNÍKOV POSKYTOVATEĽ SLUŽIEB VYŽADUJE INÝ JAZYK AKO
ANGLIČTINU, POSKYTNUTIE PREKLADATEĽSKÝCH SLUŽIEB JE
ZODPOVEDNOSŤOU ZÁKAZNÍKA.
• NEPOKÚŠAJTE SA VYKONÁVAŤ SERVIS ZARIADENIA SKÔR, AKO SI
NEPREČÍTATE SERVISNÚ PRÍRUČKU A NEPOROZUMIETE JEJ.
• ZANEDBANIE TOHTO UPOZORNENIA MÔŽE VYÚSTIŤ DO ZRANENIA
POSKYTOVATEĽA SLUŽIEB, OBSLUHUJÚCEJ OSOBY ALEBO PACIENTA
ELEKTRICKÝM PRÚDOM, PRÍPADNE DO MECHANICKÉHO ALEBO INÉHO
NEBEZPEČENSTVA.
VARNING
(SV)
DİKKAT
(TR)
DEN HÄR SERVICEHANDBOKEN FINNS BARA TILLGÄNGLIG PÅ ENGELSKA.
• OM EN KUNDS SERVICETEKNIKER HAR BEHOV AV ETT ANNAT SPRÅK ÄN
ENGELSKA ANSVARAR KUNDEN FÖR ATT TILLHANDAHÅLLA
ÖVERSÄTTNINGSTJÄNSTER.
• FÖRSÖK INTE UTFÖRA SERVICE PÅ UTRUSTNINGEN OM DU INTE HAR LÄST
OCH FÖRSTÅR DEN HÄR SERVICEHANDBOKEN.
• OM DU INTE TAR HÄNSYN TILL DEN HÄR VARNINGEN KAN DET RESULTERA I
SKADOR PÅ SERVICETEKNIKERN, OPERATÖREN ELLER PATIENTEN TILL
FÖLJD AV ELEKTRISKA STÖTAR, MEKANISKA FAROR ELLER ANDRA FAROR.
BU SERVİS KILAVUZU YALNIZCA İNGİLİZCE OLARAK SAĞLANMIŞTIR.
•EĞER MÜŞTERİ TEKNİSYENİ KILAVUZUN İNGİLİZCE DIŞINDAKİ B
• BU UYARININ GÖZ ARDI EDİLMESİ, ELEKTRİK ÇARPMASI YA DA MEKANİK
VEYA DİĞER TÜRDEN KAZALAR SONUCUNDA TEKNİSYENİN, OPERATÖRÜN
YA DA HASTANIN YARALANMASINA YOL AÇABİLİR.
-vi -
Page 9
PRELIMINARY
GE HEALTHCARE
PART NUMBER FN091019 REVISION 2VS5 N & VS6 N SERVICE MANUAL
(JA)
(ZH-CN)
(KO)
- -vii
Page 10
PRELIMINARY
GE HEALTHCARE
PART NUMBER FN091019 REVISION 2VS5 N & VS6 N SERVICE MANUAL
DAMAGE IN TRANSPORTATION
All packages should be closely examined at time of delivery. If damage is apparent write “Damage In
Shipment” on ALL copies of the freight or express bill BEFORE delivery is accepted or “signed for” by
a GE representative or hospital receiving agent. Whether noted or concealed, damage MUST be
reported to the carrier immediately upon discovery, or in any event, within 14 days after receipt, and the
contents and containers held for inspection by the carrier. A transportation company will not pay a claim
for damage if an inspection is not requested within this 14 day period.
CERTIFIED ELECTRICAL CONTRACTOR STATEMENT - FOR USA ONLY
All electrical Installations that are preliminary to position ing of the equipment at the site prepared for the
equipment shall be performed by licensed electrical contractors. Othe r conn ections between pieces o f
electrical equipment, calibrations and testing sha ll be performed by q ualified GE Healthcare person nel.
In performing all electrical work on these products, GE will use its own specially trained field engineers.
All of GE’s electrical work on these products will comply with the requirements of the applicable
electrical codes.
The purchaser of GE equipment shall only utilize qualified personnel (i.e., GE’s field engineers,
personnel of third-party service companies with equivalent training, or licensed electricians) to perform
electrical servicing on the equipment.
OMISSIONS & ERRORS
If there are any omissions, errors or suggestions for improving this documentation, please contact the
GE Healthcare Global Documentation Group with specific information listing the system type, manual
title, part number, revision number, page number and suggestion details.
Mail the information to:
Service Documentation,
GE Vingmed Ultrasound AS
P.O.Box: 141
NO 3191 HORTEN
NORWAY
SERVICE SAFETY CONSIDERATIONS
DANGER
DANGEROUS VOLTAGES, CAPABLE OF CAUSING DEATH, ARE PRESENT IN
THIS EQUIPMENT. USE EXTREME CAUTION WHEN HANDLING, TESTING AND
ADJUSTING.
WARNINGWARNING
Use all Personal Protection Equipment (PPE) such as gloves, safety shoes, safety
glasses, and kneeling pad, to reduce the risk of injury.
For a complete review of all safety requirements , see Section 1- 3 "Safety Considerations" on pa ge 1-7.
-viii -
Page 11
PRELIMINARY
GE HEALTHCARE
PART NUMBER FN091019 REVISION 2VS5 N & VS6 N SERVICE MANUAL
LEGAL NOTES
The contents of this publication may not be copied or duplicated in any form, in whole or in part, without
prior written permission of GE Healthcare.
GE Healthcare may revise this publication from time to time without written notice.
TRADEMARKS
All products and their name brands are trademarks of their respective holders.
GE HEALTHCARE
PART NUMBER FN091019 REVISION 2VS5 N & VS6 N SERVICE MANUAL
-xxvi -
Page 29
PRELIMINARY
GE HEALTHCARE
PART NUMBER FN091019 REVISION 2VS5 N & VS6 N SERVICE MANUAL
Chapter 1
Introduction
Section 1-1
Overview
1-1-1Purpose of Chapter 1
This chapter describes important issues related to safely servicing the Vivid S5 N and Vivid S6 N
ultrasound scanner. The service provider mu st read and under stand all the information presented here
before installing or servicing a unit.
Table 1-1 Contents in Chapter 1
SectionDescriptionPage Number
1-1
1-2
1-3
1-4
1-5
1-6
Overview
Important Conventions
Safety Considerations
Product Labels and Icons
EMC, EMI, and ESD
Customer Assistance
1-1
1-5
1-7
1-10
1-17
1-19
Chapter 1 - Introduction1-1
Page 30
PRELIMINARY
GE HEALTHCARE
PART NUMBER FN091019 REVISION 2VS5 N & VS6 N SERVICE MANUAL
1-1-2Purpose of Service Manual
This manual provides installation and service information for the Vivid S5 N and Vivid S6 N ul trasound
scanners, and contains the following chapters:
•Chapter 1 - Introduction
Contains a content summary and warnings.
•Chapter 2 - Site Preparations
Contains pre-installation requirements for the Vivid S5 N and Vivid S6 N ultrasound scanner.
•Chapter 3 - System Setup
Contains installation procedures and an installation checklist.
•Chapter 4 - Functional Checks
Contains functional checks that are reco m men de d as par t of the inst alla tio n procedure, or as
required during servicing and periodic maintenance.
•Chapter 5 - Components and Function (Theory)
Contains block diagrams and functional explanations of the electronic circuits.
•Chapter 6 - Service Adjustments
Contains instructions for performing service adjustments to the Vivid S5 N and Vivid S6 N
ultrasound scanner.
•Chapter 7 - Diagnostics/Troubleshooting
Provides instructions for setting up and running diagnostic, troubleshooting and other related
routines for the Vivid S5 N and Vivid S6 N ultrasound scanner.
•Chapter 8 - Replacement Procedures
Provides disassembly and reassembly procedures for all Field Replaceable Units (FRUs).
•Chapter 9 - Renewal Parts
Contains a complete list of field replaceable parts for the Vivid S5 N and Vivid S6 N ultrasound
scanner.
•Chapter 10 - Care and Maintenance
Provides periodic maintenance procedures for the Vivid S5 N and Vivid S6 N ultrasound scanner.
1-1-3Typical Users of the Service Manual
This manual is intended for the following categories of users:
•GE service personnel (installation, maintenance, etc.).
•Hospital service personnel.
•Contractors (some parts of Chapter 2 - Site Preparations).
1-2 -
Page 31
PRELIMINARY
GE HEALTHCARE
PART NUMBER FN091019 REVISION 2VS5 N & VS6 N SERVICE MANUAL
1-1-4Vivid S5 N and Vivid S6 N Models Covered in this Manual
The Vivid S5 N and Vivid S6 Nmodels documented in this manual are shown in Table 1-2 below.
Table 1-2 Vivid S5 N and Vivid S6 N Models and Hardware/Software Compatibility
MODEL
CAT NO.
Vivid S5 N
H45561HEFN000060
H45561FEFN000040
H45551WCFN000010
Vivid S6 N
H45561HFFN000070
H45561FFFN000050
H45551WBFN000020
NUMBERDESCRIPTION
Introduced 2010
Introduced 2009
Introduced 2008
Introduced 2010
Introduced 2009
Introduced 2008
SYSTEM
SOFTWARE
VERSION(S)
v3.0.10
v2.0.8v7.2.0.Build 100
v3.0.10
v2.0.8v7.2.0.Build 100
APPLICATION
SOFTWARE
VERSION(S)
v10.1.1 build 101
v10.1.0 build 65
v10.1.1 build 101
v10.1.0 build 65
Chapter 1 - Introduction1-3
Page 32
PRELIMINARY
GE HEALTHCARE
PART NUMBER FN091019 REVISION 2VS5 N & VS6 N SERVICE MANUAL
1-1-5System History - Hardware and Software Versions
The Vivid S5 N and Vivid S6 N is a compact, phased, linear array ultrasound imaging scanner.
Weighing only 65 Kgs (144 lbs), each system is extremely versatile and, depending upon the installed
software, can be used for a variety of applications.
The Vivid S5 N and Vivid S6 N ultrasound scanner is based on parallel products in the Vivid family (Vivid
3 n, Vivid 7, Vivid i n). Using the same advanced technology, the innovative features and versatility of
the system represent the future of ultrasound scanning.
NOTE:Upgrade options may be available with future versions.
NOTE:No upgrade from Vivid S5 N to Vivid S6 N is possible.
1-1-6Purpose of Operator Manual(s)
The Operator Manual(s) should be fully read and understood before operating the Vivid S5 N or Vivid
S6 N system, and also kept near the unit for quick reference.
1-4 -
Page 33
PRELIMINARY
GE HEALTHCARE
PART NUMBER FN091019 REVISION 2VS5 N & VS6 N SERVICE MANUAL
Section 1-2
Important Conventions
1-2-1Conventions Used in this Manual
1-2-1-1Model Designations
This manual covers the Vivid S5 N and Vivid S6 N ultrasound units listed in Table 1-2 "Vivid S5 N and
Vivid S6 N Models and Hardware/Software Compatibility" on page 1-3.
1-2-1-2Icons
Pictures, or icons, are used wherever they will reinforce the printed message. The icons, labels and
conventions used on the product and in the service information are described in this chapter.
1-2-1-3Safety Precaution Messages
Various levels of safety precautions are found on the equip ment and throughout this service manual.
Different levels of severity are identified by one of the following icons which precede precautionary
statements in the text.
DANGER
Indicates a hazard with a high level of risk which, if not avoided, will result in
death or serious injury.
DANGER
Indicates a hazard with a medium level of risk which, if not avoided, could result
in death or serious injury.
CAUTION
NOTE:Notes are used to provide important information about an item or a procedure. Be sure to read
Indicates a hazard with a low level of risk which, if not avoided, could result in minor or moderate
injury.
the notes as the information they contain can often save you time or effort.
Chapter 1 - Introduction1-5
Page 34
PRELIMINARY
N
GE HEALTHCARE
PART NUMBER FN091019 REVISION 2VS5 N & VS6 N SERVICE MANUAL
1-2-1-4Standard Hazard Icons
Important information will always be preceded by the exclamation point contained within
a triangle, as seen throughout this chapter. In addition to text, several different graphical
icons (symbols) may be used to make you aware of specific types of hazards that could
cause harm.
Table 1-3 Standard Hazard Icons
ELECTRICALMECHANICALRADIATION
LASERHEATPINCH
Other hazard icons make you aware of specific procedures that should be followed.
Table 1-4 Standard Icons Indicating a Special Procedure Be Used
AVOID STATIC ELECTRICITYTAG AND LOCK OUTWEAR EYE PROTECTION
TAG
&
EYE
EYE
PROTECTIO
PROTECTION
1-6Section 1-2 - Important Conventions
Page 35
PRELIMINARY
GE HEALTHCARE
PART NUMBER FN091019 REVISION 2VS5 N & VS6 N SERVICE MANUAL
Section 1-3
Safety Considerations
1-3-1Introduction
The following safety precautions must be observed duri ng all phases of operation, service and repair of
this equipment. Failure to comply with these precautions or with specific warnings elsewhere in this
manual, violates safety standards of design, manufacture and intended use of the equipment.
1-3-2Human Safety
Operating personnel must not remove the system covers.
Servicing should be performed by authorized personnel only.
Only personnel who have participated in Vivid S5 N or Vivid S6 N Training ar e authorized to service the
equipment.
1-3-3Mechanical Safety
DANGER
DANGER
DANGER
CAUTION
WHENENVER THE UNIT IS TO BE MOVED ALONG ANY INCLINE, USE EXTREME
CAUTION. MAKE SURE THAT THE VIVID S5 N OR VIVID S6 N SCANNER AND ALL
PERIPHERALS ARE SECURELY MOUNTED IN PLACE BEFORE ATTEMPTING TO MOVE
IT.
ULTRASOUND PROBES ARE HIGHLY SENSITIVE MEDICAL INSTRUMENTS THAT CAN
EASILY BE DAMAGED BY IMPROPER HANDLING. USE CARE WHEN HANDLING AND
PROTECT FROM DAMAGE WHEN NOT IN USE. DO NOT USE A DAMAGED OR
DEFECTIVE PROBE. FAILURE TO FOLLOW THESE PRECAUTIONS CAN RESULT IN
SERIOUS INJURY AND EQUIPMENT DAMAGE.
NEVER USE A PROBE THAT HAS BEEN SUBJECTED TO MECHANICAL SHOCK OR
IMPACT. EVEN IF THE PROBE APPEARS TO BE UNBROKEN, IT MAY IN FACT BE
DAMAGED.
The Vivid S5 N and Vivid S6 N ultrasound scanner weighs 65kg (144 lbs.) or more, depending on
carry-on peripherals when ready for use. Care must be used when moving the unit or replacing
its parts. Failure to follow the precautions listed could result in injury, uncontrolled motion and
costly damage.
ALWAYS:
•Use the handle to move the system
•Be sure the pathway is clear
•Use slow, careful motions
•Do not let the system strike walls or door frames
•When moving the system on inclines, make sure that the Vivid S5 N and Vivid S6 N scanner
and all peripherals are securely mounted in place before attempting to move the scanner.
•The rear handle should only be used for pushing the system. Do NOT use it for pulling the
ultrasound scanner (in any direction) as this may cause the system to become unstable in
the event of colliding with obstacles.
Chapter 1 - Introduction1-7
Page 36
PRELIMINARY
GE HEALTHCARE
PART NUMBER FN091019 REVISION 2VS5 N & VS6 N SERVICE MANUAL
1-3-3 Mechanical Safety (cont’d)
WARNINGWARNING
Always lock the system in its parked (locked) position after moving. Failure to do so
could result in personal injury or equipment damage.
WARNINGWARNING
Equipment damage could result if special care is not taken when transporting the Vivid
S5 N and Vivid S6 N system in a vehicle.
ALWAYS:
•Eject any media from the media storage devices DVD, MOD (if installed).
•Ensure that monitor is in folded and locked position.
•Ensure that the Vivid S5 N and Vivid S6 N system is well prepared and packed in its original
packaging before transporting. Special care must be taken to correctly position the packing
material, using all screws and brackets.
For further information, refer to Chapter 3 - System Setup.
•Place the probes in their carrying case.
•Secure the system in an full down position and lock the wheels (brake).
•Ensure that the Vivid S5 N and Vivid S6 N system is firmly secured while inside the vehicle.
•Secure the system with straps or as directed otherwise to prevent motion during transport.
•Prevent vibration damage by driving cautiously. Avoid unpaved roads, excessive speeds,
and erratic stops or starts.
1-3-4Electrical Safety
To minimize shock hazard, the equipment must be connected to a well grounded power source. The
system is equipped with a three-conductor AC power cable. This must be plugged into an approved
electrical outlet with safety grounding.
The power outlet used for this equipment should not be shared with other types of equipment. Both the
system power cable and the power connector must meet international electrical standards.
1-3-4-1Probes
All the probes for the Vivid S5 N and Vivid S6 N ultrasound unit are designed and manufactured to
provide trouble-free, reliable service. To ensure this, correct handling of probes is important and the
following points should be noted:
•Do not drop a probe or strike it against a hard surface, as this may damage the transducer elements,
acoustic lens, or housing.
•Do not use a cracked or damaged probe. In this event, call your field service representative
immediately to obtain a replacement.
•Avoid pulling, pinching or kinking the probe cable, since a damaged cable may compromise the
electrical safety of the probe.
•To avoid the risk of a probe accidentally falling, do not allow the probe cables to become entangled,
or to be caught in the system’s wheels.
NOTE:For detailed information on handling endocavity probes, refer to the appropriate supplementary
instructions for each probe. In addition, refer to the Vivid S5 N or Vivid S6 N User Manual for detailed
probe handling instructions.
1-8Section 1-3 - Safety Considerations
Page 37
PRELIMINARY
GE HEALTHCARE
PART NUMBER FN091019 REVISION 2VS5 N & VS6 N SERVICE MANUAL
1-3-5Dangerous Procedure Warnings
Warnings, such as the examples below, precede potentially dangerous procedures throughout this
manual. Instructions contained in the warnings must be followed.
DANGEROUS VOLTAGES, CAPABLE OF CAUSING DEATH, ARE PRESENT IN THIS
EQUIPMENT. USE EXTREME CAUTION WHEN HANDLING, TESTING AND ADJUSTING.
EXPLOSION WARNING
DO NOT OPERATE THE EQUIPMENT IN AN EXPLOSIVE ATMOSPHERE.
OPERATION OF ANY ELECTRICAL EQUIPMENT IN SUCH AN ENVIRONMENT CONSTITUTES A
DEFINITE SAFETY HAZARD.
WARNINGWARNING
DO NOT SUBSTITUTE PARTS OR MODIFY EQUIPMENT
BECAUSE OF THE DANGER OF INTRODUCING ADDITIONAL HAZARDS, DO NOT INSTALL
SUBSTITUTE PARTS OR PERFORM ANY UNAUTHORIZED MODIFICATION OF THE
EQUIPMENT.
Chapter 1 - Introduction1-9
Page 38
PRELIMINARY
GE HEALTHCARE
PART NUMBER FN091019 REVISION 2VS5 N & VS6 N SERVICE MANUAL
Section 1-4
Product Labels and Icons
The Vivid S5 N and Vivid S6 N ultrasound scanner comes equipped with product labels and icons.
These labels and icons represent pertinent information regarding the operation of the unit.
1-4-1Product Rating Plate Locations
The following diagrams show the label and icons found on the Vivid S5 N or Vivid S6 N ultrasound u nit.
All the labels and icons are described in Table 1-5 "Product Icons" on page 1-12.
Figure 1-1 Main Label Vivid S5 N and Vivid S6 N
(220-240V) International
Figure 1-2 Product label - Vivid S5 N
1-10Section 1-4 - Product Labels and Icons
Page 39
PRELIMINARY
GE HEALTHCARE
PART NUMBER FN091019 REVISION 2VS5 N & VS6 N SERVICE MANUAL
1-4-1 Product Rating Plate Locations (cont’d)
Figure 1-3 Product label - Vivid S6 N
Chapter 1 - Introduction1-11
Page 40
PRELIMINARY
GE HEALTHCARE
PART NUMBER FN091019 REVISION 2VS5 N & VS6 N SERVICE MANUAL
1-4-2Label Descriptions
The following table shows the labels and symbols that may be found on th e Vivid S5 N or Vivid S6 N
ultrasound unit, and provides a description of each label’s purpose and location.
Table 1-5 Product Icons
Label NameDescriptionLocation
Manufacturer’s name and address.
Identification and Rating
Plate
Date of Manufacture.
Model and Serial numbers.
Electrical ratings (Volts, Amps, phase, and
frequency.
On the base of the unit
Type/Class Label
Used to indicate the degree of safety or
protection.
Equipment Type BF (man in the box symbol)
IEC 878-02-03 indicates B Type equipment
having a floating applied part.
"CAUTION" The equilateral triangle is
usually used in combination with other
symbols to advise or warn the user.
“ATTENTION - Consult accompanying
documents” is intended to alert the user to
refer to the User Manual or other instructions
when complete information cannot be
provided on the label.
"CAUTION - Dangerous voltage" (the
lightning flash with arrowhead in equilateral
triangle) is used to indicate electric shock
hazards.
Bottom panel of the adapter
Probe connectors.
Various.
Various.
Various.
Device Listing/Certification
Labels
Laboratory logos or labels that denote
conformity with industry safety standards,
such as UL or IEC.
CE certification mark.On the base of the unit
Equipment Type CF IEC 878-02-05
indicates equipment having a floating
applied part that provides a degree of
protection suitable for direct cardiac contact.
1-12Section 1-4 - Product Labels and Icons
On the base of the unit
Above the ECG inlet, ECG connector
and surgical probes.
Page 41
PRELIMINARY
GE HEALTHCARE
PART NUMBER FN091019 REVISION 2VS5 N & VS6 N SERVICE MANUAL
Table 1-5 Product Icons (cont’d)
Label NameDescriptionLocation
CAUTION - This machine
weighs...Special care must
be used to avoid..."
This precaution is intended to prevent injury
that may be caused by the weight of the
machine if one person attempts to move it
considerable distances or on an incline.
Used in the Service and User Manual
which should be adjacent to
equipment at all times for quick
reference.
"DANGER - Risk of explosion
used in..."
The system is not designed for use with
flammable anesthetic gases.
"Protective Earth" Indicates the protective
earth (grounding) terminal.
"Equipotentiality" Indicates the terminal to be
used for connecting equipotential
conductors when interconnecting
(grounding) with other equipment.
“ON” indicates the power ON position of the
power switch. “Standby” indicates the power
stand by position of the power switch.
CAUTION This Power Switch DOES NOT
ISOLATE Mains Supply.
Waste Electrical and Electronic Equipment
(WEEE) Disposal
This symbol indicates that waste electrical
and electronic equipment must not be
disposed of as unsorted municipal waste
and must be collected separately.
Please contact an authorized representative
of the manufacturer for information
concerning the decommissioning of your
equipment.
Indicated in the Service Manual.
Rear of the system.
Peripherals
Adjacent to ON/Standby Switch
On the base of the unit.
On the rear of the system.
Chapter 1 - Introduction1-13
Page 42
PRELIMINARY
GE HEALTHCARE
PART NUMBER FN091019 REVISION 2VS5 N & VS6 N SERVICE MANUAL
1-4-3Vivid S6 N Battery Safety
NOTE:The Vivid S6 N ultrasound scanner is supplied with a lithium ion battery in the battery bay, as standard.
For the Vivid S5 N, it is supplied as an option.
The lithium ion battery provides power to safe ly shutdown the system or place it in Standby mode, when
an AC power source is not available. Lithium ion batteries last longer than conventional batteries and
do not require replacement as often. You ca n ex pect 4 h ours of batte ry life with a sing le, fully-charg ed
battery supporting Standby mode.
The lithium ion technology used in the system’s battery is significantly less hazardous to the
environment than the lithium metal technology used in some other batteries (such a watch batteries).
Used batteries should not be placed with common household waste products. Contact local authorities
for the location of a chemical waste collection program nearest you.
NOTE:Regulations vary for different countries. Dispose of a used battery in accordance with local regulat ions.
CAUTION
WARNINGWARNING
WARNINGWARNING
USE ONLY BATTERIES APPROVED BY GE HEALTHCARE AS SUITABLE FOR USE WITH
THE VIVID S6 N ULTRASOUND SCANNER
The Vivid S6 N battery has a safety device.
DO NOT ATTEMPT TO DIS-ASSEMBLE OR ALTER THE BATTERY!
Always observe the following precautions:
•Charge batteries only when the ambient temperature is between 0
and discharge the batteries between -10
•Do not short-circuit the battery by directly connecting the negative terminals with metal objects.
•Do not heat the battery or discard it in a fire.
•Do not expose the battery to temperatures over 60
other heat sources.
•Do not charge the battery near a heat source, such as, a fire or heater.
•Do not leave the battery in direct sunlight.
•Do not pierce the battery with a sharp object, hit it, or step on it.
•Do not use a damaged battery.
•Do not apply solder to a battery.
•Do not connect the battery to an electrical power outlet.
o
and 55o C (14o and 131o F)
o
C (140o F). Keep the battery away from fire and
o
and 65o C (32o and 149o F).
In the event that the Vivid S5 N or Vivid S6 N ultrasound scanner will not be used for a
long period, it is recommended to remove the battery from the system while not in use.
CAUTION
1-14Section 1-4 - Product Labels and Icons
To prevent the battery bursting, igniting, or fumes from the battery causing equipment damage,
always observe the following precautions:
•Do not immerse the battery in water or allow it to get wet.
•Do not place the battery into a microwave oven or pressurized container.
•If the battery leaks or emits an odor, remove it from all possible flammable sources.
•If the battery emits an odor or heat, is deformed or discolored, or in a way ap pears abnormal during
use, recharging or storage, immediately remove it and stop using it.
•If you have any questions about the battery, consult your local GE representative.
Page 43
PRELIMINARY
GE HEALTHCARE
PART NUMBER FN091019 REVISION 2VS5 N & VS6 N SERVICE MANUAL
1-4-3 Vivid S6 N Battery Safety (cont’d)
•Storage of the battery pack:
o
Short-term (less than 1 month):between 0
Long-term (more than 3 months): between 10
NOTE:When storing a battery pack for more than 6 months, it is necessary to charge the battery pack
at least once every 6 months in order to prevent leakage and deterioration in performance (as
a result of self-discharging).
NOTE:When charging the battery for the first time after long-term storage, recover the battery pack to
original performance through repeating several cycles of full charging and discharging.
C (32o F) and 50o C (122o F).
o
C (50o F) and 35o C (95o F)
NOTE:
When shipped, the Vivid S5 N or Vivid S6 N with the battery it is in a state of being approximately 30% charged.
Before use, it is necessary to fully charge and discharge the battery pack up to 3 times, in order to utilize Li-lon
smart packs.
Chapter 1 - Introduction1-15
Page 44
PRELIMINARY
GE HEALTHCARE
PART NUMBER FN091019 REVISION 2VS5 N & VS6 N SERVICE MANUAL
1-4-4Vivid S5 N and Vivid S6 N External Labels
In addition to the labels described in the previous section, additional labels may be found on the
Vivid S5 N or Vivid S6 N ultrasound unit, as described in the following sections:
•Product Labels section, see below.
•GND Label section, see below.
1-4-4-1Product Labels
Indicates the ultrasound unit’s basic power compliance and regulatory compliance information
The Product label shown below is located on the base assembly.
1-4-4-2GND Label
Indicates the protective earth (grounding) terminal. The GND label (sho wn Figure 1-5 below) is located
at the rear of the system.
Figure 1-4 Product Label example
Figure 1-5 GND Label
1-16Section 1-4 - Product Labels and Icons
Page 45
PRELIMINARY
GE HEALTHCARE
PART NUMBER FN091019 REVISION 2VS5 N & VS6 N SERVICE MANUAL
Section 1-5
EMC, EMI, and ESD
1-5-1Electromagnetic Compatibility (EMC)
Electromagnetic compatibility describes a level of performance of a device within its electromagnetic
environment. This environment consists of the device itself and its surroundings, including other
equipment, power sources and persons with which the device must interface. Inadequate compatibility
results when a susceptible device fails to perform as intended due to interference from its environment,
or when the device produces unacceptable levels of emission. This interference is often refe rred to as
radio–frequency or electromagnetic interference (RFI/EMI) and can be radiated through space or
conducted over interconnecting power or signal cables. In addition to electromagnetic energy, EMC
also includes possible effects from electrical fields, magnetic fields, electrostatic discharge and
disturbances in the electrical power supply.
1-5-2CE Compliance
The Vivid S5 N and Vivid S6 N ultrasound scanners conform to all applicable conducted and radiated
emission limits and to immunity from electrostatic discharge, radiated and conducted RF fields,
magnetic fields and power line transient requirements.
For applicable standards refer to the Safety Chapter in the Vivid S5 N or Vivid S6 N User Manual.
NOTE:For CE Compliance, it is critical that all covers, screws, shielding, gaskets, mesh and clamps are in good
condition and installed tightly without skew or stress. Proper installation following all comments noted
in this service manual is required in order to achieve full EMC performance.
1-5-3Electrostatic Discharge (ESD) Prevention
CAUTION: DO NOT TOUCH ANY BOARDS WITH INTEGRATED CIRCUITS PRIOR TO
TAKING THE NECESSARY ESD PRECAUTIONS:
1.ALWAYS CONNECT YOURSELF, VIA AN ARM-WRIST STRAP CONNECTED TO THE
CAGE ASSEMBLY OR ANY GROUND SCREW WHENEVER YOU OPEN THE SYSTEM
FOR MAINTENANCE.
2.FOLLOW GENERAL GUIDELINES FOR HANDLING OF ELECTROSTATIC SENSITIVE
EQUIPMENT.
Chapter 1 - Introduction1-17
Page 46
PRELIMINARY
GE HEALTHCARE
PART NUMBER FN091019 REVISION 2VS5 N & VS6 N SERVICE MANUAL
1-5-4Standards Used
To fulfill the requirements of relevant EC directives and/or European Harmonized/International
standards, the following documents/standards have been used:
Table 1-6 Standards Used
ScopeStandard/Directive
IEC 60601-1:1988+A1:1991+A2:1995/
Product safety requirements for Electrical Medical devices
Class I, Type B with BF and CF applied parts. (ICE catheter and
ECG leads)
EN 60601-1:1990+A1:1993+A2:1995+A13 :1996
UL 60601-1 (2003)
C 22.2 601-1 M90 (R2005)
Product safety requirements for Electrical Medical devices-safety
requirements for medical electrical systems
Product safety requirements for Electrical Medical devicesProgrammable electrical medical systems
Product safety requirements for Ultrasound system
Product EMC requirements
Group 1. Class B
Biological evaluation of medical devicesISO 10993-1:2003
Particular requirements for the safety of electrocardiograph
monitoring equipment (not for monitoring–only for
synchronization with ultrasound)
General requirements for basic safety and essential
performance–Usability
Patient privacyHIPPA
IEC 60601-1-1:2000
EN 60601-1-1:2001
IEC 60601-1-4:1996+A1:1999
EN 60601-1-4:1996+A1:1999
IEC 60601-2-37:2001+A1:2004+A2:2005
EN 60601-2-37: 2001+A1:2005 +A2:2005
IEC 60601-1-2:2001+A1:2004
EN 60601-1-2:200+A1:2006
IEC 60601-2-27:1994
EN 60601-2-27:1994
IEC 60601-1-6:2004
EN 60601-1-6:2004
NOTE:For CE Compliance, it is critical that all covers, screws, shielding, gaskets, mesh and clamps are in good
condition and installed tightly without skew or stress. Proper installation following all comments noted
in this service manual is required in order to achieve full EMC performance.
1-5-5Lockout/Tagout Requirements (For USA Only)
Follow OSHA Lockout/Tagout requirements to protect service personnel from injuries caused by
unexpected energizing or start-up of equipment during service, repair, or maintenance.
1-18Section 1-5 - EMC, EMI, and ESD
Page 47
PRELIMINARY
GE HEALTHCARE
PART NUMBER FN091019 REVISION 2VS5 N & VS6 N SERVICE MANUAL
Section 1-6
Customer Assistance
1-6-1Contact Information
If this equipment does not work as indicated in this service manual or in the Vivid S5 N or Vivid S6 N
User Manual, or if you require additional assistance, please contact the local distributor or appropriate
support resource, as listed below.
Prepare the following information before you call:
•System ID and/or serial number.
•Software version.
•Detailed description of any problem encountered.
•Where applicable, save the appropriate log files (by pressing <Alt +D>). It is preferable to save logs
directly to the GE ftp site, otherwise onto removable media.
Table 1-7 Phone numbers for Customer Assistance
LOCATIONPHONE NUMBER
USA
GE Healthcare
Ultrasound Service Engineering
9900 Innovation Drive
Wauwatosa, WI 53226
Canada1-800-668-0732
Latin America
Europe (OLC- EMEA)
GE Ultraschall Deutschland GmbH
Beethovenstraße 239
Postfach 11 05 60, D-42655 Solingen
Germany
GE HEALTHCARE
PART NUMBER FN091019 REVISION 2VS5 N & VS6 N SERVICE MANUAL
Chapter 2
Site Preparations
Section 2-1
Overview
2-1-1Purpose of Chapter 2
This chapter provides the information required to plan and p repare for the installation of a Vivid S5 N or
Vivid S6 N ultrasound unit. Included are descriptions of the electrical and facility requirements that must
be met by the purchaser. A worksheet is provided at th e en d of this cha p te r (refe r to Figure 2-2 on
page 2-10) to help ensure that all the required network information is available, prior to installation.
Table 2-1 Contents in Chapter 2
SectionDescriptionPage Number
2-1
2-2
2-3
Overview
Console Requirements
Facility Needs
2-1
2-2
2-6
Chapter 2 - Site Preparations2-1
Page 50
PRELIMINARY
GE HEALTHCARE
PART NUMBER FN091019 REVISION 2VS5 N & VS6 N SERVICE MANUAL
Section 2-2
Console Requirements
2-2-1Unit Environmental Requirements
Table 2-2 Environmental Requirements
Relative Humidity
RequirementTemperature
(non-condensing)
Air Pressure
o
C(50 — 104oF)
o
C(14 — 140oF)
o
C(14 — 140oF))
ON and/or using the system. 3-2-2 "Installation Warnings" on
30 — 75%700 — 1060 hPa
30 — 70%700 — 1060 hPa
30 — 70%700— 1060 hPa
CAUTION
Operational
Storage
Transport
10 — 40
-10 — 60
-10 — 60
If the system has been in storage or has been transported, please see the acclimation
requirements before powering
page 3-2.
NOTE:After a long period of storage, or after transportation of the system with the monitor in the folded-down
position, it is highly recommended to place the monitor in the upright position - and to leave it in this
position for a period of longer than 1 hour before use. This will enable it to properly adjust to the
environmental conditions.
2-2-2Cooling Requirements
The cooling requirement for the Vivid S5 N or Vivid S6 N ultr asound unit environment is 3500 BTU/hr.
This figure does not include the cooling required for lights, people, or other equipment in the room.
NOTE:Each person in the room places an additional 300 BTU/hr demand on the environmental
cooling.
2-2-3Lighting Requirements
For system installation, updates and repairs, bright lighting is required. However, operator and patient
comfort may be optimized if the room lighting is subdued and indirect when a scan is being per formed.
Therefore, a combination lighting system (dim/bright) is recommended. Keep in mind that monitor
lighting can be adjusted automatically on the Vivid S6 N if this option is used.
2-2-4Time and Manpower Requirements
Site preparation takes time. Begin pre-installation checks as soon as possible to allow sufficient time to
make any required changes. If possible, begin these checks as many as six weeks before system
delivery.
CAUTION
2-2Section 2-2 - Console Requirements
Only one person is required to unpack the Vivid S5 N or Vivid S6 N ultrasound unit; at least two
people must be available to roll the system down the wheeling ramp. Attempts to move the system
considerable distances (or on an incline) by one person alone, could result in personal injury, and/
or damage to the system.
Page 51
PRELIMINARY
GE HEALTHCARE
PART NUMBER FN091019 REVISION 2VS5 N & VS6 N SERVICE MANUAL
2-2-5Electrical Requirements
NOTE:GE HEALTHCARE requires a dedicated mains power line and Groun d for the proper operation
of its Ultrasound equipment. This dedicated power line shall originate at the last distribution
panel before the system.
Sites with a mains power system with defined Neutral and Live:
The dedicated line shall consist of one phase, a neutral (not shared with any other circuit), and a full
size Ground wire from the distribution panel to the Ultrasound outlet.
Sites with a mains power system without a defined Neutral:
The dedicated line shall consist of one phase (two lines), not shared with any other circuit, and a full
size Ground wire from the distribution panel to the Ultrasound outlet.
NOTE:Please note that image artifacts can occur, if at any time within the facility, the Ground from the
main facility's incoming power source to the Ultrasound unit is only a conduit.
2-2-5-1Vivid S5 N and Vivid S6 N Power Requirements
Electrical specifications for the Vivid S5 N and Vivid S6 N system are as follows:
Table 2-3 Electrical Requirements
Input VoltageTolerancesOp. CurrentFrequency
100V AC to 240V AC±10%0.5 to 1A50-60 Hz
2-2-5-2Inrush Current
Inrush current is not a factor for consideration, due to the inrush current limiting prop erties of the power
supplies.
Table 2-4 Inrush Current
Voltage
100 V2.3 A4.5 A
240 V4.5 A9 A
2-2-5-3Site Power Outlets
A dedicated AC power outlet must be within reach of the unit without requiring the use of extension
cords. Other outlets adequate for the external peripherals, medical and test equipment required to
support this unit must also be present and located within 1 m (3.2 ft) of the unit. Two dedicated outlets
for peripherals - located at the rear of the system - should be used for all peripherals attached to the
system, unless otherwise indicated. Electrical installation must meet all current local, state, and national
electrical codes.
Inrush Current
Console OnlyConsole with all Peripherals
2-2-5-4Mains Power Plug
The Vivid S5 N and Vivid S6 N portable ultrasound scanner is supplied with a mains power plug, as
standard. In the event that the unit arrives withou t a power plug, or with the wrong plug, contact your
GE dealer. When necessary, the installation engineer will supply the appropriate power plug to meet
the applicable local regulations.
Chapter 2 - Site Preparations2-3
Page 52
PRELIMINARY
GE HEALTHCARE
PART NUMBER FN091019 REVISION 2VS5 N & VS6 N SERVICE MANUAL
2-2-5 Electrical Requirements (cont’d)
2-2-5-5Power Stability Requirements
•Voltage drop-out
Max 10 ms.
•Power Transients
(All applications)
Less than 25% of nominal peak voltage for less than 1 millisecond for any type of transient, including
line frequency, synchronous, asynchronous, or aperiodic transients.
2-2-6EMI Limitations
Ultrasound machines are susceptible to Electromagnetic Interference (EMI) from radio frequencies,
magnetic fields, and transients in the air or wiring. They also generate EMI. The Vivid S5 N and Vivid
S6 N ultrasound unit complies with limits as stated on the EMC label. However, there is no guarantee
that interference will not occur in a particular installation.
NOTE:Possible EMI sources should be identified before the unit is installed, and should not be on the
same line as the ultrasound system. A dedicated line sh ould be used for the ultrasound system.
Electrical and electronic equipment may produce EMI unintentionally as the result of a defect. Sources
of EMI include the following:
•Medical lasers.
•Scanners.
•Cauterizing guns.
•Computers.
•Monitors.
•Fans.
•Gel warmers.
•Microwave ovens.
•Portable phones.
•Broadcast stations and mobile broadcasting machines.
2-4Section 2-2 - Console Requirements
Page 53
PRELIMINARY
GE HEALTHCARE
PART NUMBER FN091019 REVISION 2VS5 N & VS6 N SERVICE MANUAL
2-2-6 EMI Limitations (cont’d)
The following table lists recommendations for preventing EMI:
Table 2-5 EMI Prevention/ Abatement
EMI RuleDetails
Ground the unit.
Be aware of RF sources.
Replace and/or reassemble
all screws, RF gaskets,
covers and cores.
Replace broken RF gaskets.
Do not place labels where
RF gaskets touch metal.
Use GE-specified harnesses
and peripherals.
Take care with cellular
phones.
Properly address peripheral
cables.
Poor grounding is the most likely reason an ultrasound unit will have noisy images.
Check the grounding of the power cord and power outlet.
Keep the unit at least 5m (16.4 ft) away from other EMI sources. Special shielding may
be required to eliminate interference problems caused by high frequency, high powered
radio or video broadcast signals.
After you finish repairing or updating the system, replace all covers and tighten all
screws. Any cable with an external connection requires a magnet wrap at each end.
Install the shield over the front of the card cage. Loose or missing covers or RF gaskets
allow radio frequencies to interfere with the ultrasound signals.
If more than 20% or a pair of the fingers on an RF gasket are broken, replace the gasket.
Do not turn ON the unit until any loose metallic part is removed and replaced, if required.
Never place a label where RF gaskets meet the unit. Otherwise, the gap created will
permit RF leakage. In case a label has been found in such a location, move the label to
a different, appropriate location.
The interconnect cables are grounded and require ferrite beads and other shielding.
Cable length, material, and routing are all important; do not make any changes that do
not meet all specifications.
Cellular phones may transmit a 5 V/m signal that causes image artifacts.
Do not allow cables to lie across the top of the system. Loop any peripheral cable excess
length into one bundle.
2-2-7Probe Environmental Requirements
Table 2-6 Probe Operation and Storage Temperatures
Electronics
Operation
Storage
NOTE:System and electronic probes are designed for storage temperatures o f -20o to +50o C. When
exposed to large temperature variations, the probes should be kept at room temperature for a
minimum of 10 hours before use.
10 — 40oC (50 — 104oF)
-20 — 50oC (-4 — 122oF)
Chapter 2 - Site Preparations2-5
Page 54
PRELIMINARY
GE HEALTHCARE
PART NUMBER FN091019 REVISION 2VS5 N & VS6 N SERVICE MANUAL
Section 2-3
Facility Needs
2-3-1Purchaser Responsibilities
The work and materials required to prepare the site are the responsibility of the purchaser. To avoid
delay, complete all pre-installation work before delivery. Use the Pre-installation Check List (provided
in Table 2-7 "Vivid S5 N and Vivid S6 N Pre-Installation Check List" on page 2-12)to verify that all the
required steps have been completed.
Purchaser responsibilities include:
•Procuring the required materials.
•Completing the preparations prior to de livery of the ult ra so un d sys te m.
•Paying the costs of any alterations and modifications not specifically provided for in the sales
contract.
NOTE:All relevant preliminary electrical installations at the prepared site must be performed by
licensed electrical contractors. Other connections between electrical equipment, and
calibration and testing, must also be performed by qualified personnel. The products involved
(and the accompanying electrical installations) are h ighly sophisticated and special engineering
competence is required. All electrical work on these products must comply with the
requirements of applicable electrical codes. The purchaser of GE equipment must utilize only
qualified personnel to perform electrical servicing of the equipment.
To avoid delays during installation, the individual or team who will perform the installation should be
notified at the earliest possible date (preferably prior to installation), of the existence of any of the
following variances:
•Use of any non-listed product(s).
•Use of any customer provided product(s).
•Placement of an approved product further from the system than the interface kit allows.
The prepared site must be clean prior to delivery of the system. Carpeting is not recomme nded because
it collects dust and creates static. Potential sources of EMI should also be investigated before delivery.
Dirt, static, and EMI can negatively impact system reliability.
2-6Section 2-3 - Facility Needs
Page 55
PRELIMINARY
GE HEALTHCARE
PART NUMBER FN091019 REVISION 2VS5 N & VS6 N SERVICE MANUAL
2-3-2Mandatory Site Requirements
The following are mandatory site requirements. Additional (optional) recommendations, as well as a
recommended ultrasound room layout, are provided in section 2-3-3 - Site Recomme ndations (see
below).
•A dedicated single branch power outlet of adequate amperage (see Table 2-3 "Electrical
Requirements" on p age 2-3 ) that meets all local and national code s and is located less than 2.5 m
(8.2 ft) from the unit’s proposed location. 2-2-5 "Electrical Requirements" on page 2-3
•A door opening of at least 0.54 m (1.77 ft) in width.
•The proposed location for the unit is at least 0.2 m (0.67 ft) from the walls, to enable cooling.
•Power outlet and place for any external peripheral are within 2 m (6.5 ft.) of each other with
peripheral within 1 m of the unit to connect cables.
•Power outlets for other medical equipment and gel warmer.
•Power outlets for test equipment within 1 m (3.3 ft) of the ultrasound unit.
•Clean and protected space for storage of probes (either in their case or on a rack).
•Material to safely clean probes (perform ed usin g a pla stic container, never metal).
•In the case of a network option:
•An active network outlet in the vicinity of the ultrasound unit.
•A network cable of appropriate length (regular Pin-to-Pin network cable).
•An IT administrator who will assist in configuring the unit to work with your local network. A fixed
IP address may be required when using DICOM. Refer to the form provided in Figure 2-2
"Connectivity Installation Worksheet" on page 2-10 for network details that are required.
NOTE:All relevant preliminary network outlets installations at the prepared site must be performed by
authorized contractors. The purchaser of GE equipment must utilize only qualified personnel to
perform servicing of the equipment.
2-3-3Site Recommendations
The following are (optional) site recommendations. Mandatory site requirements are provided in the
Mandatory Site Requirements section, above.
•Door opening of 92 cm (3 ft) in width.
•Accessible circuit breaker for a dedicated power outlet.
•Sink with hot and cold running water.
•Receptacle for bio–hazardous waste, for example, used probe sheaths.
•Emergency oxygen supply.
•Storage area for linens and equipment.
•Nearby waiting room, lavatory, and dressing room.
•Dual level lighting (bright and dim).
•Lockable cabinet for software and manuals.
Chapter 2 - Site Preparations2-7
Page 56
PRELIMINARY
GE HEALTHCARE
PART NUMBER FN091019 REVISION 2VS5 N & VS6 N SERVICE MANUAL
2-3-3-1Recommended Ultrasound Room Layout
Figure 2-1 below shows a floor plan illustrating the recommended layout of the Ultrasound Room and
depicting the minimal room layout requirements.
VIVID 3
36 IN.
(92 CM)
Figure 2-1 Recommended Floor Plan 4.3m x 5.2m (14 ft x 17 ft)
Dedicated Power Outlets
Hospital Network
GE Cabinet for
Software and Manuals
(optional)
2-8Section 2-3 - Facility Needs
Page 57
PRELIMINARY
GE HEALTHCARE
PART NUMBER FN091019 REVISION 2VS5 N & VS6 N SERVICE MANUAL
2-3-4Networking Pre-Installation Requirements
2-3-4-1Stand-alone Unit (without Network Connection)
None.
2-3-4-2Unit Connected to Hospital’s Network
Supported networks:
•Wireless LAN
2-3-4-3Purpose of the DICOM Network Function
DICOM services provide the operator with clinically useful features for moving images and patient
information over a hospital network. Examples of DICOM services include the transfer of images to
workstations for viewing or transferring images to remote printers. As an added benefit, transferring
images in this manner frees up the on-board moni tor and peripherals, enabling viewing to be don e while
scanning continues. With DICOM, images can be archived, stored, and retrieved faste r, easier, a nd at
a lower cost.
2-3-4-4DICOM Option Pre-Installation Requirements
To configure the Vivid S5 N or Vivid S6 N ultrasound unit to work with other network connections, the
network administrator must provide the required information, which should include the following:
•Details:DICOM network details for the Vivid S5 N or Vivid S6 N unit,
including the host name, local port, IP address, AE title and net
mask.
•Routing Information:IP addresses for the default gateway and other routers in use at
the site.
•DICOM Application Information:Details of the DICOM devices in use at the site, including the
DICOM host name, AE title and IP addresses.
Chapter 2 - Site Preparations2-9
Page 58
PRELIMINARY
GE HEALTHCARE
PART NUMBER FN091019 REVISION 2VS5 N & VS6 N SERVICE MANUAL
GE HEALTHCARE
PART NUMBER FN091019 REVISION 2VS5 N & VS6 N SERVICE MANUAL
Chapter 3
System Setup
Section 3-1
Overview
3-1-1Purpose of Chapter 3
This chapter provides instructions for installing the Vivid S5 N and Vivid S6 N ultrasound unit. Before
beginning the installation process, an appropriate site must be prepared, as described in Chapter 2 -
Site Preparations. Once the site has been prepared, installation can proceed as described in this
chapter.
Included in this chapter are guidelines for transporting the unit to a new site, as well as procedures tha t
describe how to receive and unpack the equipment, and (if necessary) how to file a damage or loss
claim. Instructions for checking and testing the unit, probes, and external peripherals for electrical safety
are also provided.
NOTE:Depending on the customer’s specific requirements, the Vivid S5 N or Vivid S6 N portable
ultrasound scanner will have been supplied. Where applicable, the installation procedures for
each scenario are clearly identified in this chapter.
Table 3-1 Contents in Chapter 3
SectionDescriptionPage Number
3-1
3-2
3-3
3-4
3-5
3-6
3-7
3-8
3-9
Overview
Installation Reminders
Receiving and Unpacking the Equipment
Preparing for Installation
Completing the Hardware Installation
Configuration
Connectivity Setup
Storing and Transporting the Unit
Completing the Installation Paperwork
3-14
3-23
3-61
3-78
3-167
3-168
3-1
3-2
3-4
Chapter 3 - System Setup 3-1
Page 62
PRELIMINARY
GE HEALTHCARE
PART NUMBER FN091019 REVISION 2VS5 N & VS6 N SERVICE MANUAL
Section 3-2
Installation Reminders
3-2-1Average Installation Time
The Vivid S5 N or Vivid S6 N installation and functional checkout will take approximately one hour; Vivid
S5 N or Vivid S6 N consoles with optional equipment may take slightly longer.
Once the site has been prepared, the average installation time required is shown in Table 3-2 below.
Table 3-2 Average Installation Time
Average
Description
Unpacking the scanner30 minutes
Installing the scanner30 minutesTime may vary, according to the required configuration
DICOM Option
(connectivity)
Installation TimeComments
30 minutesTime may vary, according to the required configuration
3-2-2Installation Warnings
1.) The Vivid S5 N or Vivid S6 N ultrasound scanner weighs 65 Kgs (143 lbs), without
add-ons/peripherals.
Two persons are always required to unpack the system.
2.) There are no operator-serviceable components. To prevent shock, do not remove any covers or
panels. If problems or malfunctions occur, unplug the power cord. Only qualified service personnel
should carry out servicing and troubleshooting.
NOTE:For information on shipping carton labels, refer to Figure 3-1 "Drop and T ilt Indicators" on p age
3-5 and Figure 3-4 "Additional Labels on Shipping Carton" on page 3-7.
3-2-2-1System Acclimation Time
Following transport, the Vivid S5 N or Vivid S6 N system may be very cold, or hot. Allow time for the
system to acclimate before being switched ON. Acclimation requires 1 hour for each 2.5
o
when the temperature of the system is below 10
.
CAUTION
605550454035302520151050-5-10-15-20-25-30-35-40
°C
Turning the system ON after arrival at the site - without allowing time for acclimation - may
cause system damage!
Table 3-3 Vivid S5 N and Vivid S6 N System Acclimation Time
NOTE:After a long period of storage, or after transportation of the system with the monitor in the folded-down
position, it is highly recommended to place the monitor in the upright position - and to leave it in this
position for a period of longer than 1 hour before use. This will enable it to properly adjust to the
environmental conditions.
3-2Section 3-2 - Installation Reminders
Page 63
PRELIMINARY
GE HEALTHCARE
PART NUMBER FN091019 REVISION 2VS5 N & VS6 N SERVICE MANUAL
3-2-3Safety Reminders
DANGER
WARNINGWARNING
CAUTION
CAUTION
CAUTION
CAUTION
WHEN USING ANY TEST INSTRUMENT THAT IS CAPABLE OF OPENING THE
AC GROUND LINE (I.E., METER’S GROUND SWITCH IS OPEN), DO NOT TOUCH
THE UNIT!
Two people are required to unpack the system as it is heavy. Two people are always required
whenever a part weighing 19kg (35 lb.) or more must be lifted
If the unit is very cold or hot, do NOT turn ON power to the unit until it has had sufficient time to
acclimate to its operating environment.
To prevent electrical shock, connect the unit to a properly grounded power outlet.
Do NOT use a three-prong to two-prong adapter, as this defeats safety grounding.
Do NOT wear the ESD wrist strap when you work on live circuits where more than
30 V peak is present.
Do NOT operate the unit unless all board covers and frame panels are securely in place, to
ensure optimal system performance and cooling. (When covers are removed, EMI may be
present).
WARNINGWARNING
NOTE:The Vivid S5 N or Vivid S6 N User Manual should be fully read and understood before operating
ACOUSTIC OUTPUT HAZARD
Although the ultrasound energy transmitted from the Vivid S5 N or Vivid S6 N portable
ultrasound scanner is within AIUM/NEMA standards and FDA limitations, avoid unnecessary
exposure. Ultrasound energy can produce heat and mechanical damage.
the unit. Keep the manual near the unit for reference.
Chapter 3 - System Setup 3-3
Page 64
PRELIMINARY
GE HEALTHCARE
PART NUMBER FN091019 REVISION 2VS5 N & VS6 N SERVICE MANUAL
Section 3-3
Receiving and Unpacking the Equipment
CAUTION
NOTE:For details of weights and dimensions of all three shipping cartons, refer to Table 3-4 on page 7.
Please read this section fully before unpacking the Vivid S5 N or Vivid S6 N ultrasound unit.
The Vivid S5 N or Vivid S6 N ultrasound unit, together with the peripherals, probes, cables, and
accessories, are shipped from the factory in a single cardboard shipping carton mounted on a raised
wooden base - see Figure 3-2 "Shipping Carton - V ivid S5 N and V ivid S6 N with Per ipherals" on p age
3-6.
Where the specific customer requirements are for a Vivid S5 N or Vivid S6 N ultrasound unit with
peripherals, this is shipped in a smaller-sized cardboard shipping ca rton moun ted on a ra ised wo oden
base, or placed on top of the main cardboard carton th at contains the scanner- see Figure 3-3 "Shipping
Carton - Vivid S5 N and Vivid S6 N Only" on page 3-7.
Unpacking instructions for each type of shipping carton are provided separately in the following
sections:
•3-3-1-1 "Unpacking the Vivid S5 N or Vivid S6 N with Peripherals" on page 3-8
•3-3-1-2 "Unpacking the Vivid S5 N or Vivid S6 N" on page 3-9
Each cardboard shipping carton is sealed with plastic tape and secured on the outside with plastic
straps. The interior of the carton is fully lined with re-sealable water-proof wrapping to protect the
contents from external moisture; silica gel bags are also placed among the contents to absorb any
condensation or moisture within the carton.
NOTICE
Before unpacking the unit
•Ins pe ct the ca rto n for visibl e da m age .
•Inspect the Drop and Tilt indicators (on the Shock-watch and Tilt-watch labels, respectively) for
evidence of accidental shock or tilting during transit (damage incident) - refer to Figure 3-1
"Drop and Tilt Indicators" on page 3-5.
After unpacking the unit:
It is recommended to keep and store the shipping carton and all other packing materials (includin g the
support foams, anti-static plas ti c cove r, etc.) , in ca se the unit has to be mo ved to a d ifferen t location in
the future.
For warranty purposes, storage of the above is required for one year from date of purchase.
3-4Section 3-3 - Receiving and Unpacking the Equipment
Page 65
PRELIMINARY
GE HEALTHCARE
PART NUMBER FN091019 REVISION 2VS5 N & VS6 N SERVICE MANUAL
Section 3-3 Receiving and Unpacking the Equipment (cont’d)
Drop indicator turns Red
Tilt indicator turns Red
Figure 3-1 Drop and Tilt Indicators
If the carton is damaged, or if either the Drop (Shock) or Tilt indicators have turned red (indicating
damage incident), please inform the GE HEALTHCARE sales representative immediately. In addition,
mark on the shipping consignment note or packing slip/post-delivery checklist (in the “P ackage” column)
that the Tilt and/or Drop indicators show damage incident.
Chapter 3 - System Setup 3-5
Page 66
PRELIMINARY
GE HEALTHCARE
PART NUMBER FN091019 REVISION 2VS5 N & VS6 N SERVICE MANUAL
Section 3-3 Receiving and Unpacking the Equipment (cont’d)
Peripherals
Height
157 cm (62 in)
Plastic Straps
Ultrasound Scanner
Wooden Base
FRONT
Width
73 cm (29 in)
Note: Weight = 105 Kgs (231 lbs)
Depth
87 cm (34 in)
Figure 3-2 Shipping Carton - Vivid S5 N and Vivid S6 N with Peripherals
3-6Section 3-3 - Receiving and Unpacking the Equipment
Page 67
PRELIMINARY
GE HEALTHCARE
PART NUMBER FN091019 REVISION 2VS5 N & VS6 N SERVICE MANUAL
Section 3-3 Receiving and Unpacking the Equipment (cont’d)
Plastic Straps
Adhesive Tape
Height
119 cm (47 in)
Wooden Base
Note: Weight = 95 Kgs (209 lbs)
Figure 3-3 Shipping Carton - Vivid S5 N and Vivid S6 N Only
Table 3-4 Shipping Cartons - Dimensions and Weights
DescriptionHeightWidthDepth
Vivid S5 N or Vivid S6 N scanner
with peripherals and accessories
Vivid S5 N or Vivid S6 N scanner only
a. Weight is approximate and will vary depending upon the supplied peripherals
FRONT
Width
73 cm (29 in)
Depth
87 cm (34 in)
157 cm
62 ins
119 cm
47 ins
73 cm
29 ins
73 cm
29 ins
87 cm
34 ins
87cm
34 ins
Weight
105 kgs
231 lbs
95 kgs
209 lbs
a
Figure 3-4 Additional Labels on Shipping Carton
Chapter 3 - System Setup 3-7
Page 68
PRELIMINARY
GE HEALTHCARE
PART NUMBER FN091019 REVISION 2VS5 N & VS6 N SERVICE MANUAL
3-3-1Unpacking the Cardboard Shipping Carton
As previously described, the Vivid S5 N or Vivid S6 N ultrasound unit is packed in a single cardboard
shipping carton comprising a durable outer cardboard carton cover and a wooden platform base ; these
are firmly joined together with plastic straps. After completing a visual inspection of the shipping carton
and Drop/Tilt indicators, proceed to the unpacking instructions below, as appropriate
3-3-1-1Unpacking the Vivid S5 N or Vivid S6 N with Peripherals
1.) Cut the plastic straps securing the shipping carton to the wooden base - see Figure 3-2 "Shipping
Carton - Vivid S5 N and Vivid S6 N with Peripherals" on page 3-6.
CAUTION
When using sharp tools to open packing materials, take care to avoid cutting or damaging any
of the contents.
2.) Remove the options peripherals box (smaller-sized cardboard shipp ing ca rton ) from th e top of the
main cardboard shipping carton that contains the scanner.
3.) Carefully cut the adhesive tape used to close the shipping carton lids, and open the cartons from
the top.
4.) Undo the re-sealable water-proof carton liners.
5.) Remove the packing slip (shipping consignment note) detailing the co ntents of the shipping ca rton
and keep close to hand ready to mark the check list - refer to 3-4-2 "Verifying the Shipping Carton
Contents" on page 3-14.
6.) Continue with unpacking the main cardboard shipping carton that contains the scanner, as
indicated in 3-3-1-2 "Unpacking the Vivid S5 N or Vivid S6 N" on page 3-9.
3-8Section 3-3 - Receiving and Unpacking the Equipment
Page 69
PRELIMINARY
GE HEALTHCARE
PART NUMBER FN091019 REVISION 2VS5 N & VS6 N SERVICE MANUAL
3-3-1 Unpacking the Cardboard Shipping Carton (cont’d)
3-3-1-2Unpacking the Vivid S5 N or Vivid S6 N
1.) Cut the two plastic straps securing the shipping carton to the wooden base - see Figure 3-3
"Shipping Carton - Vivid S5 N and Vivid S6 N Only" on page 3-7.
CAUTION
When using sharp tools to open packing materials, take care to avoid cutting or damaging any
of the contents.
2.) Carefully cut the adhesive tape used to close the shipping carton upper lid and open the four folded
flaps - see Figure 3-5 "Opening the Shipping Carton" on page 3-9.
Remove Screws
Figure 3-5 Opening the Shipping Carton
3.) From each side, remove the screws securing the outer cardboard to the wooden base.
4.) Remove the outer cardboard walls pulling them gently upwards and away from the system
(Figure 3-6 "Wheeling Ramp in Upright Position" on page 3-10).
Chapter 3 - System Setup 3-9
Page 70
PRELIMINARY
GE HEALTHCARE
PART NUMBER FN091019 REVISION 2VS5 N & VS6 N SERVICE MANUAL
3-3-1 Unpacking the Cardboard Shipping Carton (cont’d)
Wheeling Ramp
(in upright position)
Guide Rail
Wooden Base
Figure 3-6 Wheeling Ramp in Upright Position
NOTE:The front side of the carton is re-inforced with wooden guide rails and, when dropped down in
the open position, these form a strong ramp used for wheeling the system out of the shipping
carton.
5.) Remove the packing slip (shipping consignment note) detailing the co ntents of the shipping ca rton
and keep close to hand ready to mark the check list - refer to 3-4-2 "Verifying the Shipping Carton
Contents" on page 3-14.
3-10Section 3-3 - Receiving and Unpacking the Equipment
Page 71
PRELIMINARY
GE HEALTHCARE
PART NUMBER FN091019 REVISION 2VS5 N & VS6 N SERVICE MANUAL
3-3-1 Unpacking the Cardboard Shipping Carton (cont’d)
6.) Lower the wheeling ramp into the unloading position, as shown in Figure 3-7 below.
Foam Supports
Adhesive tape
Console Holder
Unscrew to remove
Console Holder
Wheeling Ramp
7.) Carefully remove the foam supports - see Figure 3-8 below.
Unscrew to remove
Console Holder
Figure 3-7 Wheeling Ramp in Unloading Position for Scanner Removal
Figure 3-8 Foam Supports Protecting the LCD Screen
Chapter 3 - System Setup 3-11
Page 72
PRELIMINARY
GE HEALTHCARE
PART NUMBER FN091019 REVISION 2VS5 N & VS6 N SERVICE MANUAL
3-3-1 Unpacking the Cardboard Shipping Carton (cont’d)
8.) Remove the two screws that secure the console holder, then remove the holder, refer Figure 3-7
"Wheeling Ramp in Unloading Position for Scanner Removal" on page 3-11.
9.) Carefully remove the adhesive tape from the top of th e syste m and spr ead the a nti-static nylo n on
the floor.
10.)Remove any additional packing material from in front of and surrounding the system.
11.)Remove the two silica gel bags (Figure 3-9 below)
Locking brake
Anti-static nylon on floor
Silica gel bags
CAUTION
Figure 3-9 Locking Wheel Brake
12.)Press down on the locking brake to unlock the wheel. Repeat for the three other wheels.
13.)Using two people, carefully roll the scanner forwards down the unloading ramp to remove it
completely from the shipping carton.
Two people are required when moving the Vivid S5 N or Vivid S6 N scanner.
This precaution is intended to prevent injury that may be caused by the weight of the
system if one person attempts to move the scanner considerable distances or on an
incline.
14.)Make sure the system is standing firmly on the floor before applying the locking brake.
15.)Remove the foam sections from the top of the system and from between the main shelf and
probe shelf.
16.)Using the up/down handle, raise the main shelf up to appropriate height.
17.)Pull the monitor locking pin towards you and lift the monitor into the upright position.
18.)Before commencing the installation procedure, follow the instructions for 3-3-2 "Verifying the
Shipping Carton Contents" on page 3-13.
3-12Section 3-3 - Receiving and Unpacking the Equipment
Page 73
PRELIMINARY
GE HEALTHCARE
PART NUMBER FN091019 REVISION 2VS5 N & VS6 N SERVICE MANUAL
3-3-2Verifying the Shipping Carton Contents
After unpacking, it is important to verify that all items ordered by the customer have b een received (refer
to “Verifying the Shipping Carton Contents” on page 14). Compare all items listed on the packing slip
(shipping consignment note) with those received and report any items that are missing, b ack-ordere d,
or damaged, to your GE HEALTHCARE sales representative. The contact address is shown in 1-6-1
"Contact Information" on page 1-19.
NOTICE
It is recommended to keep and store the shipping carton and all other packing materials (includin g the
support foams, anti-static plas ti c cove r, etc.) , in case the unit has to be moved to a differ ent location in
the future.
For warranty purposes, storage of the above is required for one year from date of purchase.
3-3-3Physical Inspection
3-3-3-1System Voltage Settings
Verify that the Vivid S5 N or Vivid S6 N ultrasound AC is set to the correct voltage.
The Voltage settings are:
•220-240V AC - Europe, all Asia excluding Japan and Taiwan.
•100-120V AC - Americas, Japan and Taiwan.
WARNINGWARNING
Setting the Vivid S5 N or Vivid S6 N ultrasound components to the wrong voltage setting will
most likely destroy the equipment.
3-3-4EMI Protection
The Vivid S5 N or Vivid S6 N Ultrasound Unit has been designed to minimize the effects of Electro
Magnetic Interference (EMI). Many of the covers, shields, and screws are provided primarily to protect
the system from image artifacts caused by this interference. For this reason, it is imperative that all
covers and hardware are installed and secured before the unit is put into operation.
Chapter 3 - System Setup 3-13
Page 74
PRELIMINARY
GE HEALTHCARE
PART NUMBER FN091019 REVISION 2VS5 N & VS6 N SERVICE MANUAL
Section 3-4
Preparing for Installation
3-4-1Confirming Customer Order
When preparing for installation of a Vivid S5 N or Vivid S6 N system, it is important to verify that all items
ordered by the customer have been received. Compare all items listed on the packing slip (shipping
consignment note) with those received and report any items that are missing, back-ordered, or
damaged, to your GE HEALTHCARE sales representative.
3-4-2Verifying the Shipping Carton Contents
In addition to the Vivid S5N or Vivid S6N portable ultrasound Scanner there is a shipping Carton. The
shipping carton includes external accessories, probes, software CD's, user manual and optional
peripherals (as ordered). Ensure that all relevant components are presents before completing the
installation.
3-4-2-1Optional Peripherals
The shipping carton will only contain any of the optional peripherals if they have been ordered with the
Vivid S5 N or Vivid S6 N system. Table 3-5 "Optional Peripherals" on page 3-14 lists all possibilities.
Table 3-5 Optional Peripherals
Part DescriptionPart No
HP Network PrinterH45541MJ
Digital B&W Video Printer (for Vivid S6N)H45551YW
Digital B&W Video Printer (for Vivid S5N)H45551YX
Digital Color Video PrinterH45531XB
B&W Paper (for H45551YW)EN0473
Color Paper (for H45531XB)E70151SA
USB Flash Card (1.0GB)H45551XT
MODH45551XW
MO Media H45001JK
Ext ECG CableH45551WD
Foot SwitchH45551WE
Rear TrayH45551WF
3-14Section 3-4 - Preparing for Installation
Page 75
PRELIMINARY
GE HEALTHCARE
PART NUMBER FN091019 REVISION 2VS5 N & VS6 N SERVICE MANUAL
3-4-3Component Inspection
After verifying that all the required parts are included in the shipping crate, inspect the system
components using the checklist supplied below. In addition, ensure that all the labels described in
Chapter 1 - are present, accurate and in good condition, and enter the serial number printed on the
main label into the system installation details card, as described in 3-9-1 "System Installation Details"
on page 3-168.
3-4-3-1Damage Inspection Checklist
Visually inspect the contents of the shipping carton for damage. If any parts are damaged or missing,
contact an authorized GE Service Representative.
A Damage Inspection Checklist for the Vivid S5 N or Vivid S6 N portable ultrasound scanner is provided
in Table 3-6 "Damage Inspection Checklist - Vivid S5 N and Vivid S6 N Systems" on page 3-16 .
Chapter 3 - System Setup 3-15
Page 76
PRELIMINARY
GE HEALTHCARE
PART NUMBER FN091019 REVISION 2VS5 N & VS6 N SERVICE MANUAL
Table 3-6 Damage Inspection Checklist - Vivid S5 N and Vivid S6 N Systems
b
StepItemRecommended Procedure
1
Console
2
Control Console
3
Probes
4
LCD Display
5
Fans
6
Rear Panel
7
Covers
8
Peripherals
10
Probe HoldersClean the gel wells with warm water and a damp cloth to remove all traces of gel.
Verify that the system is switched OFF and unplugged. Clean the console and control
panel.
Physically inspect the control console for missing or damaged items. Verify the proper
illumination of all the control panel buttons.
Check all probes for wear and tear on the lens, cable, and connector. Look for bent
or damaged pins on the connector and in the connector socket on the unit. Verify that
the EMI fingers around the probe connector socket housing are intact. Check the
probe locking mechanism and probe switch.
Clean the LCD display by gently wiping with a dry, soft, lint-free non-abrasive folded
cloth. Inspect the monitor for scratches and raster burn.
Turn on the system and verify that the system’s cooling fans and peripheral fans are
operating.
Check the rear panel connectors for bent pins, loose connections and loose or
missing hardware. Screw all the cable connectors tightly to the connector sockets on
the panel. Verify that the labeling is in good condition.
Check that all screws are tightly secured in place, that there are no dents or scratches
and that no internal parts are exposed.
Check and clean the peripherals in accordance with the manufacturer’s directions. To
prevent EMI or system overheating, dress the peripheral cables inside the peripheral
cover.
13
CoversCheck that all screws are in place, all chassis and internal covers are installed.
14
15
16
17
18
Peripherals
AC System
Power Cord
Front Castors
Rear Castors
Check and clean the peripherals in accordance with the manufacturer’s directions. To
prevent EMI or system overheating, dress the peripheral cables inside the peripheral
cover.
Check the AC board connectors and the associated cabling for good connection and
proper insulation. Verify that the connections are secured.
Check the power cord for cuts, loose hardware, tire marks, exposed insulation, or any
deterioration. Verify continuity.
Tighten the clamps that secure the power cord to the unit and the outlet plug to the
cord. Replace the power cord and/or clamp, as required.
Clamp securing power cord
Check that the front castors can roll and swivel, and can be placed in the locked position by pressing the foot brake (lower lever) down on each.
Ensure that the wheels are locked and there is no movement in any direction.
Check that the front castors can roll and swivel, and can be placed in the locked position by pressing the foot brake (lower lever) down on each.
Ensure that the wheels are locked, unable to swivel left or right, and that there is no
movement forwards or backwards.
3-16Section 3-4 - Preparing for Installation
Page 77
PRELIMINARY
GE HEALTHCARE
PART NUMBER FN091019 REVISION 2VS5 N & VS6 N SERVICE MANUAL
3-4-3-2Front View of the Vivid S5 N or Vivid S6 N Ultrasound Unit
Figure 3-10 "Front and Side View of the Vivid S5 N or Vivid S6 N Ultrasound Scanner" on page 3-17
shows the Vivid S5 N or Vivid S6 N ultrasound unit components that are visible from the front and side
of the ultrasound unit.
Figure 3-10 Front and Side View of the Vivid S5 N or Vivid S6 N Ultrasound Scanner
1
Display Monitor:Tilts up and down and swivels left and right.
2Monitor Release pin:Enables locking of the monitor in the 90 degree position
3Rear Basket:Holds paper and other utensils that may be required
4Control PanelContains the alphanumeric keyboard and the buttons used to oper ate
5Probe holders:
the ultrasound unit
Provides a safe, rubber-cushioned resting place for probes.
.
6Control Panel swivel release handle:
Enables locking of the control panel in position; swivelling it left or right.
7Rear shelf (tray):Provides a surface that may be used for external peripherals.
Note: the rear shelf should be used only for small-footprint items of a
total weight not exceeding 1 Kg (2.2 lbs).
Chapter 3 - System Setup 3-17
Page 78
PRELIMINARY
GE HEALTHCARE
PART NUMBER FN091019 REVISION 2VS5 N & VS6 N SERVICE MANUAL
3-4-3-2 Front View of the Vivid S5 N or Vivid S6 N Ultrasound Unit (cont’d)
8Control Panel up/down release handle:
Enables locking of the control panel in position, or changing the
location of the arm by moving it upwards and away from the user, or
downwards and away from the system. This provides leg ro om for the
user who may then be seated beside the system.
9Rear Caster:Enables locking of swivel motion, or free rolling and swivelling.
10 Front Caster:Enables full-locking of swivel motion, or free rolling and swivelling.
11 ECG Connector:Provides External ECG input connection to an ECG monitor device
outlet, or direct patient three-contact ECG lead connection.
Important: Do not use the Vivid 7 ECG cable with the Vivid S5 N
or Vivid S6 N system.
12 Speakers:Two loudspeakers provided for Doppler sound.
13 Air inlet and Subwoofer:
Air flow is via a built-in filter on the side cover, for system air cooling.
A subwoofer speaker provides superb sound.
14 ON/OFF:Power ON/OFF knob enables the user turn ON the system (when
power is connected and the system is either in Shutdown or Standby
mode).
15 Probe ports:Vivid S6 N system:
5 active probe connectors (one for a pencil probe, and one Logiq-type
for TEE probes).
Vivid S5 N system:
4 active probe connectors (one for a pencil probe)
3-18Section 3-4 - Preparing for Installation
Page 79
PRELIMINARY
GE HEALTHCARE
PART NUMBER FN091019 REVISION 2VS5 N & VS6 N SERVICE MANUAL
3-4-3-3Rear View of the Vivid S5 N or Vivid S6 N Ultrasound Unit
Figure 3-11 Vivid S5 N or Vivid S6 N Rear View
Rear Cable Hooks:Provides a cable management solution for probes and ECG leads.
1
2AC Outlet:Two isolated AC outlets for peripherals attached to the scanner.
3I/O panel:Provides LAN, DVI, Audio and 2 USB 2.0 ports.
4Ground:
5Main AC IN:
6Main Circuit Breaker:Provides power to the scanner.
7Rear Handle:Used to pull/push the scanner and place it in the desired position.
8Gel Holders:Provides convenient storage for US Gel bottles. Holders are situated
Detachable Ground plug.
Main AC input plug, secured by a safety spring.
on either side of the control console, and on the top surface on e i ther
side of the monitor.
Chapter 3 - System Setup 3-19
Page 80
PRELIMINARY
GE HEALTHCARE
PART NUMBER FN091019 REVISION 2VS5 N & VS6 N SERVICE MANUAL
3-4-3-4Rear panel of the Vivid S5 N or Vivid S6 N Ultrasound Unit
Figure 3-12 "View of the Vivid S5 N or Vivid S6 N Rear Panel" on page 3-20 shows a view of the Vivid
S5 N or Vivid S6 N ultrasound unit rear panel and external peripheral/accessory connectors:
Figure 3-12 View of the Vivid S5 N or Vivid S6 N Rear Panel
1Peripherals 230VA AC outlet, isolated - provides power to external peripherals connected to
scanner.
2LAN 10/100 Base-TX Ethernet network connector.
3Two interchangeable USB ports (for digital printer, MOD and other peripherals).
4Audio outlet.
5DVI Output (VCR option or CRT monitor option).
3-20Section 3-4 - Preparing for Installation
Page 81
PRELIMINARY
GE HEALTHCARE
PART NUMBER FN091019 REVISION 2VS5 N & VS6 N SERVICE MANUAL
3-4-4System Voltage Confirmation
3-4-4-1System Voltage Settings
Verify that the Vivid S5 N or Vivid S6 N ultrasound scanner is set to the correct voltage.
The voltage settings are found on the system rear cover.
WARNINGWARNING
Figure 3-13 Safety Rating Plate - Example
Connecting the unit to the wrong voltage level will most likely destroy the scanner.
Chapter 3 - System Setup 3-21
Page 82
PRELIMINARY
GE HEALTHCARE
PART NUMBER FN091019 REVISION 2VS5 N & VS6 N SERVICE MANUAL
3-4-5Ensuring Protection from EMI
The Vivid S5 N or Vivid S6 N unit has been designed to minimize the effects of Electro-Magnetic
Interference (EMI). Many of the covers, shields, and screws are provided primarily to protect the system
from image artifacts caused by this interference. For this reason, it is imperative that all covers and
hardware are installed and secured before the unit is put into operation.
Ensure that the system is protected from electromagnetic interference (EMI), as follows:
•Operate the system at least 15 feet away from equipment that emits strong electromagnetic
radiation.
•Operate the system in an area enclosed by walls, floors and ceilings comprised of wood, plaster or
concrete, which help prevent EMI.
•Shield the system when operating it in the vicinity of radio broadcast equipment, if necessary.
•Do not operate mobile phones or other EMI emitting devices in the ultrasound room.
•Verify that all EMI rules listed in the following table are followe d :
NOTE:The Vivid S5 N and Vivid S6 N
environmentally qualified facilities, in terms of the prevention of radio wave interference.
Operation of the ultrasound unit
interference to radios and television sets situated near the medical equipment.
Table 3-7 EMI Prevention/ Abatement
EMI RuleDetails
Ground the unit.
Be aware of RF sources.
Replace and/or reassemble
all screws, RF gaskets,
covers and cores.
Replace broken RF gaskets.
Do not place labels where
RF gaskets touch metal.
Poor grounding is the most likely reason an ultrasound unit will have noisy images. Check
the grounding of the power cord and power outlet.
Keep the unit at least 5m (16.4 ft) away from other EMI sources. Special shielding may be
required to eliminate interference problems caused by high frequency, high powered radio
or video broadcast signals.
After you finish repairing or updating the system, replace all covers and tighten all screws.
Any cable with an external connection requires a magnet wrap at each end. Install the shield
over the front of the card cage. Loose or missing covers or RF gaskets allow radio
frequencies to interfere with the ultrasound signals.
If more than 20% or a pair of the fingers on an RF gasket are broken, replace the gasket.
Do not turn on the unit until any loose metallic part is removed and replaced if needed.
Never place a label where RF gaskets meet the unit. Otherwise, the gap created will permit
RF leakage. In case a label has been found in such a location, move the label to a different
appropriate location.
ultrasound unit is approved for use in hospitals, clinics and other
in an inappropriate environment can cause electronic
Use GE specified harnesses
and peripherals.
Take care with cellular
phones.
Properly address peripheral
cables.
The interconnect cables are grounded and require ferrite beads and other shielding. Cable
length, material, and routing are all important; do not make any changes that do not meet
all specifications.
Cellular phones may transmit a 5 V/m signal that causes image artifacts.
Do not allow cables to lie across the top of the card cage or hang out of the peripheral bays.
Loop any peripheral cable excess length inside the peripheral bays or hang on the hooks
provided below the console. Attach the monitor cables to the frame.
3-22Section 3-4 - Preparing for Installation
Page 83
PRELIMINARY
GE HEALTHCARE
PART NUMBER FN091019 REVISION 2VS5 N & VS6 N SERVICE MANUAL
Section 3-5
Completing the Hardware Installation
3-5-1Connecting Peripherals
NOTE:All peripherals mounted on the Vivid S5 N or Vivid S6 N system must be firmly secured in position using
the securing straps supplied with the scanner. These straps are fastened through the dedicated holes
located on each side of the rear tray and serve to hold the peripheral firmly in place.
3-5-1-1Wireless Network Adaptor
NOTICE
WARNINGWARNING
3-5-1-2MOD Drive
NOTICE
IMPORTANT During Stand-by mode, it is NOT recommended to introduce or remove USB
devices; this may cause the system to lock-up during the boot-up procedure.
The Wireless Network adaptor should be connected to the lower USB socket on the Vivid S5 N or
Vivid S6 N rear connectors panel.
One can connect the Wireless Network adaptor either once the system is powered ON, or after
shutdown. All software drivers for the Wireless Network are pre-installed for the designated
Wireless Network adaptor only. Follow the procedure described in th e 3-7-2-4 "Wireless Network
Configuration" on page 3-83.
Do not attempt to use a dif ferent type of Wireless Network adaptor . The ultrasound s ystem is an
extremely sensitive and complex medical system. Any unauthorized peripherals may cause
system failure or damage!
IMPORTANT During Stand-by mode, it is NOT recommended to introduce or remove USB
devices; this may cause the system to lock-up during the boot-up procedure.
Figure 3-14 MOD Drive
The MOD Drive can be connected to either of the USB ports (upper or lower) on the rear connectors
panel. This connection must be via a USB-to-SCSI adaptor; one side of the adaptor being
connected to the USB port, and the other side to the rear of the MOD drive.
Chapter 3 - System Setup 3-23
Page 84
PRELIMINARY
GE HEALTHCARE
PART NUMBER FN091019 REVISION 2VS5 N & VS6 N SERVICE MANUAL
3-5-1-2 MOD Drive (cont’d)
One must use the additional power source to activate the MOD Drive. However, this must be
powered from the scanner itself, via one of the isolated peripheral power outlets provided on the
rear panel of the scanner.
The MOD Drive may be connected once the system is powered ON, or after shutdown. All software
drivers for the MOD Drive are pre-configured and installed desig nated to be used with the sp ecific
MOD drive supplied by GE HEALTHCARE.
WARNINGWARNING
3-5-1-3USB Memory Stick
NOTICE
3-5-1-4Connecting the Black & White Printer
NOTICE
Do not attempt to use a different type of mass storage device other than the
MOD Drive supplied with the system. The ultrasound system is an extremely sensitive and
complex medical system. Any unauthorized peripherals may cause system failure or damag e!
IMPORTANT During Stand-by mode, it is NOT recommended to introduce or remove USB
devices; this may cause the system to lock-up during the boot-up procedure.
The USB Memory Stick may be connected to either of the USB ports on the keyboard, or on the
rear connectors panel.
The USB Memory Stick can be connected once the system is powered ON, or after shutdown. All
software drivers for the USB Memory Stick are pre-installed on the system. The system is capable
of detecting and working with other types of memory sticks. However, it is highly recommended to
use the memory stick that is provided and approved by GE HEALTHCARE. Other memory sticks
may be sensitive to EMC interference (this may affect system performance and/or image quality) or
may not be recognized by the system.
IMPORTANT During Stand-by mode, it is NOT recommended to introduce or remove USB
devices; this may cause the system to lock-up during the boot-up procedure.
Figure 3-15 Black and White Printer
3-24Section 3-5 - Completing the Hardware Installation
Page 85
PRELIMINARY
GE HEALTHCARE
PART NUMBER FN091019 REVISION 2VS5 N & VS6 N SERVICE MANUAL
3-5-1-4 Connecting the Black & White Printer (cont’d)
The B/W Printer connection in the different scenarios is explained below.
NOTE:After physical connection of the printer to the system, to configure the appro priate hot keys to activate
printing direct from the control panel, follow the instructions in 3-5-1-6 "Configuring Printin g Buttons P1
and P2" on page 3-28.
The ultrasound scanner has a designated B&W printer compartment located on the right side of the
system. When using a B&W printer with the scanner, af ter installation make sure the correct side cover
kits have been order for the Vivid S5 N or Vivid S6 N system, as appropriate:
•Vivid S6 N = P/N S2421612
•Vivid S5 N = P/N S2421611
The B&W Printer should connected to a designated USB port on the DIB board, as indicated in
Figure 3-16 "Cables Connected to DIB Box" on page 3-25.
KB Data
(J1)
USB
(J14A)
Display PWR
(J11)
AC Box
(J7)
DVI DVR
(J20)
USB Printer
(J15A)
Figure 3-16 Cables Connected to DIB Box
NOTE:Refer also to the schematic diagram in Figure 3-17 "Schematic Diagram of Internal Cables and
Connectors" on page 3-26.
Chapter 3 - System Setup 3-25
Page 86
PRELIMINARY
GE HEALTHCARE
PART NUMBER FN091019 REVISION 2VS5 N & VS6 N SERVICE MANUAL
3-5-1-4 Connecting the Black & White Printer (cont’d)
J19 -DVI DisplayR2419175-2
J19
J20-DVI DVRR2419151
J20
J23
J15 A
J15 B
J1
J14 A
J14 B
J11
Figure 3-17 Schematic Diagram of Internal Cables and Connectors
In addition, a dedicated power source and cable is provided within the dedicated B&W printer
compartment.
The B/W Printer should be connected once the system is powered OFF.
All software drivers for the B/W Printer are pre-configured and installed designated to be used with the
specific B/W Printer supplied by GE HEALTHCARE.
3-26Section 3-5 - Completing the Hardware Installation
Do not attempt to use a different type of B/W Printer (brand or model) other than the
B/W Printer provided by
and complex medical system. Any unauthorized peripherals may cause system failure or damage!
GE HEALTHCARE. The ultrasound system is an extremely sensitive
Page 87
PRELIMINARY
GE HEALTHCARE
PART NUMBER FN091019 REVISION 2VS5 N & VS6 N SERVICE MANUAL
3-5-1-5Connecting the Digital Color Printer
NOTICE
NOTE:After physical connection of the printer to the system, to configure the appropriate hot keys to activate
IMPORTANT During Stand-by mode, it is NOT recommended to introduce or remove USB
devices; this may cause the system to lock-up during the boot-up procedure.
printing direct from the control panel, follow the instructions in 3-5-1-6 "Configuring Printin g Buttons P1
and P2" on page 3-28.
Figure 3-18 Digital Color Printer
The Digital Color Printer can be connected to either of the USB ports (upper or lower) on the rear
connectors panel.
WARNINGWARNING
One must use the additional power source to activate the Digita l Color Printer. However, this must
be powered from the scanner itself, via one of the isolated peripheral power outlets provided on the
rear connectors panel.
The Digital Color Printer can be connected once the system is powered ON, or after shutdown. All
software drivers for the Digital Color Printer are pre-configured and installed designated to be used
with the specific Digital Color Printer supplied by GE HEALTHCARE.
Do not attempt to use a different type of Digital Color Printer (brand or model) other than the Digital
Color Printer provided by
and complex medical system. Any unauthorized peripherals may cause system failure or damage!
GE HEALTHCARE. The ultrasound system is an extremely sensitive
Chapter 3 - System Setup 3-27
Page 88
PRELIMINARY
GE HEALTHCARE
PART NUMBER FN091019 REVISION 2VS5 N & VS6 N SERVICE MANUAL
3-5-1-6Configuring Printing Buttons P1 and P2
Buttons P1 and P2 on the Vivid S5 N and Vivid S6 N operator consoles are used to activate the printer
directly from the console, as follows:
•Button P1 should be specifically configured independently to activate the Windows default printer.
By default, button P1 is designated to activate the Windows default printer. The printer that is
associated with the P1 button should be defined as the default printer via the Windows screens.
•Button P2 (which is fully configurable) should be separately configured to activate a second printer.
3-5-1-6-1P1 and P2 Hot Key Configuration via Vivid Vivid S5 N and Vivid S6 N Connectivity Settings
1.) Press Config (or F2).
2.) Select Connectivity (in the lower part of window).
3.) Select the Additional Outputs tab.
2
3
5
4
1
Figure 3-19 Configuring Control Panel Hot Keys for Printing
4.) From the Button drop-down menu (see Figure 3-19 "Configuring Control Panel Hot Keys for
Printing" on page 3-28), select the button to be configured - either the P1 (Print) button or
P2 (Alt+Print) to activate a variety of outputs.
5.) Use the left or right arrows below to add any of the available options (listed on the left) as Outputs
(listed on right).
6.) From the listed Outputs, select (highlight) the required output, and select Advanced to configure
the output device accordingly.
6
7
3-28Section 3-5 - Completing the Hardware Installation
Page 89
PRELIMINARY
GE HEALTHCARE
PART NUMBER FN091019 REVISION 2VS5 N & VS6 N SERVICE MANUAL
3-5-1-6 Configuring Printing Buttons P1 and P2 (cont’d)
.
Figure 3-20 Printer Properties Screen
7.) Select Configure.
Figure 3-21 Printer Setup Screen
8.) From the drop-down menu, select the desired printer and press OK.
9.) In the Printer Properties window, select OK.
Printing may now be activated direct from the Control Panel, using the hot key you have config ured
for this purpose.
Chapter 3 - System Setup 3-29
Page 90
PRELIMINARY
GE HEALTHCARE
PART NUMBER FN091019 REVISION 2VS5 N & VS6 N SERVICE MANUAL
3-5-1-6-2P1 Button Configuration via the Windows Control Panel
NOTE:If a customer has more than one printer and prefers to configure button P1 to activate a non-
configured printer, the non-configured printer must be configured as the default printer via the
Windows Control Panel, as described below.
The Sony UP-897 printer is used as an example in the following procedure.
1.) Press Config (or F2).
2.) Select Connectivity (in the lower part of window).
3.) From the Connectivity screen click the Tcpip tab.
Figure 3-22 Additional Outputs Screen
3-30Section 3-5 - Completing the Hardware Installation
Page 91
PRELIMINARY
GE HEALTHCARE
PART NUMBER FN091019 REVISION 2VS5 N & VS6 N SERVICE MANUAL
3-5-1-6-2 P1 Button Configuration via the Windows Control Panel (cont’d)
4.) In the Tcpip window, click Advanced Settings.
Click Advanced Settings
Click the "Up"
button to
navigate to the
Control Panel
Figure 3-23 Tcpip Tab Activated
The Network Connection screen opens as seen in Figure 3-24 below.
Figure 3-24 Network Connections showing Highlighted "Up One Level" Button
5.) Click the "Up" button to open the Windows Control Panel.
6.) Navigate to Printers and Faxes and double-click the printer that you want to choo se
(Sony UP-D897 in this example).
This opens the Printer Configuration dialog (Sony UP-D897 in this example - see F i gu r e 3- 25
"Printer Configuration Dialog Box Showing Set Default Printer" on page 3-32).
Chapter 3 - System Setup 3-31
Page 92
PRELIMINARY
GE HEALTHCARE
PART NUMBER FN091019 REVISION 2VS5 N & VS6 N SERVICE MANUAL
3-5-1-6-2 P1 Button Configuration via the Windows Control Panel (cont’d)
7.) From the menu, select Printer and select the Set As Default Printer option to select the printer as
the default printer or verify that the printer is the set as the default.
Figure 3-25 Printer Configuration Dialog Box Showing Set Default Printer
3-32Section 3-5 - Completing the Hardware Installation
Page 93
PRELIMINARY
GE HEALTHCARE
PART NUMBER FN091019 REVISION 2VS5 N & VS6 N SERVICE MANUAL
3-5-1-7Connecting the HP network printer - CP2025n
PRINTER CONTROL PANEL
Figure 3-26 HP CP2025n printer
WARNINGWARNING
WARNINGWARNING
The HP CP2020n printer must be connected directly to a mains AC power outlet and not
via the rear AC port on the VS5 N & VS6 N system.
When connecting the printer’s network cable, ensure that it is connected to the network
port at the rear of the Vivid S5/S6 system as shown in the picture below:
Do not connect the CP2025n
printer via the USB port
Chapter 3 - System Setup 3-33
Page 94
PRELIMINARY
GE HEALTHCARE
PART NUMBER FN091019 REVISION 2VS5 N & VS6 N SERVICE MANUAL
3-5-1-7 Connecting the HP network printer - CP2025n (cont’d)
Configuring of the HP network printer, model CP2025n, requires configuring the IP addresses for
the Vivid S5/S6 system and for the printer. There are two types of installations:
•"Peer-to-Peer" using a cross communication cable between the system and the printer
•Printer installation on a network (e.g. a hospita l netw or k)
Peer-to-Peer Configuration
A) Defining the system’s IP address:
1.) Using a service dongle, boot up the system in maintenance mode and exit to Windows Desktop.
2.) Click Start > Control Panel and navigate to Network Connections.
3.) In the Network Connections screen, select Local Area Connection 2.
Figure 3-27 Select Local Area Connection 2.
4.) On the console, press Update/Menu to access the context menu and then click Properties.
Figure 3-28 Local Are Connection 2 Properties
3-34Section 3-5 - Completing the Hardware Installation
Page 95
PRELIMINARY
GE HEALTHCARE
PART NUMBER FN091019 REVISION 2VS5 N & VS6 N SERVICE MANUAL
3-5-1-7 Connecting the HP network printer - CP2025n (cont’d)
5.) From "This connection uses the following items" scroll box, select Internet Protocol (TCP/IP) and
click Properties. The Internet Protocol (TCP/IP) Properties screen opens.
Figure 3-29 Use the following IP address
6.) Select Use the following IP address and in the IP address field type: 10.0.0.1 and in the Subnet
mask field type: 255.0.0.0.
7.) Click OK and click Close. The system’s IP address is defined.
B) Defining the printer’s IP address:
The printer’s settings are controlled via the printer’s control panel as identified in Figure 1-1 on page 1-1.
The control panel is illustrated Figure 3-30 below:
Figure 3-30 CP2025n printer control panel
The printer’s control panel is explained in the chart on the following page.
Chapter 3 - System Setup 3-35
Page 96
PRELIMINARY
GE HEALTHCARE
PART NUMBER FN091019 REVISION 2VS5 N & VS6 N SERVICE MANUAL
3-5-1-7 Connecting the HP network printer - CP2025n (cont’d)
Figure 3-31 Printer’s Control Panel Explaination
To configure the printer using the control panel, follow these guidelines:
1.) On the printer’s control panel, press OK (button 4) in Figure 3-30 "CP2025n printer control panel"
on page 3-35.
2.) Press the right arrow on the control panel four times to display "Network config" in the control
panel display and press OK. "TCP/IP config" appears in the control panel display.
3.) Press OK and on the control panel press the right arrow. "Manual" displays in the control panel
window.
4.) Press OK. An IP address displays.
NOTE:For the following step, use the above guidelines to enter the printer’s IP address.
5.) Using the control panel’s left and right arrow buttons, and the OK button, type in the following IP
address: 10.0.0.2.
6.) Press OK. "Yes" is displayed in the control panel. Press OK again.
The Subnet mask displays: 255.0.0.0
3-36Section 3-5 - Completing the Hardware Installation
Page 97
PRELIMINARY
GE HEALTHCARE
PART NUMBER FN091019 REVISION 2VS5 N & VS6 N SERVICE MANUAL
3-5-1-7 Connecting the HP network printer - CP2025n (cont’d)
7.) Press OK. Default Gateway displays with an IP address.
8.) Set the IP address to: 0.0.0.0.
9.) Press OK twice. The printer’s IP address is now configured.
C) Driver Installation:
1.) From Windows desktop, click Start > Settings and navigate to Printers.
2.) Click Add Printer. The Add Printer wizard opens.
3.) Click Next twice and select "Local Printer Attached to this Computer".
Figure 3-32 Select "Local Printer Attached to this Computer"
4.) Click Next and select "Create a new port".
In the "Type of Port" scrolldown list, select Standard TCP/IP port.
5.) Click Next and follow the instructions on the screen. Click Next.
NOTE:In the following step, you are required to type the printer’s IP address. In Peer-to-Peer mode,
the address will be: 10.0.0.2.
When installing the printer on a hospital/client network, type the IP address supplied by the
institution’s IT manager.
6.) In the "Printer Name or IP Address" field, type the printer’s IP address: 10.0.0.2. Click Next and in
the following screen click Next again.
Figure 3-33 Add standard TCP/IP Printer Port
Chapter 3 - System Setup 3-37
Page 98
PRELIMINARY
GE HEALTHCARE
PART NUMBER FN091019 REVISION 2VS5 N & VS6 N SERVICE MANUAL
3-5-1-7 Connecting the HP network printer - CP2025n (cont’d)
Click Finish. The Install Printer Software screen opens.
7.) Insert the printer’s driver CD into the CD and select "Have Disk".
The "Install From Disk" screen opens. Follow the instructions on the screen.
8.) In the "Copy manufacturer’s files from" field scrolldown list, select drive G:\ and click OK.
NOTE:When installing the printer on BT10 systems, select the corresponding DVD/HD drive letter and
click OK.
9.) Click Next and follow the instructions on the screen. The "Finish" screen displays.
10.)Print a test page.
NOTE:Wait a few minutes while the system copies the printer drivers.
If the printer does not respond timeously, press the cancel button (button no. 6 in Figure 1-3 above).
The printer will print a test page. If the printer does not respond, this could be due to the need to
configure the paper size.
Defining the paper size:
1.) From Windows Desktop, navigate to Printers and Faxes and select the HP-CP2025 printer.
2.) From the console, press Update/Menu and from the context menu select Properties.
3.) Select the required settings and click OK.
4.) Print a test page to confirm that the printer is properly installed.
Installing the CP2020n Printer on a Network
To install the CP2020n printer on a network, follow the procedure for "C) Driver Installation:" on page
3-37.
3-38Section 3-5 - Completing the Hardware Installation
Page 99
PRELIMINARY
GE HEALTHCARE
PART NUMBER FN091019 REVISION 2VS5 N & VS6 N SERVICE MANUAL
3-5-1-8Connecting the DeskJet Color Printer
NOTE:This printer is end of life, but may still be used, when available. Network Printer CP2025n has
been introduced as a replacement, see: 3-5-1-7 "Connecting the HP network printer -
CP2025n" on page 3-33.
NOTICE
IMPORTANT During Stand-by mode, it is NOT recommended to introduce or remove USB
devices; this may cause the system to lock-up during the boot-up procedure.
Figure 3-34 DeskJet Color Printer
This only applies to HP K5400 models which are supplied pre-installed with the appropriate
software drivers. The DeskJet Color Printer can be connected to either of the USB ports (upper or
lower) on the rear connectors panel.
WARNINGWARNING
One must use the additional power source to activate the DeskJet Color Printer. However, this must
be powered from the scanner itself, via one of the isolated peripheral power outlets provided on the
system. A special power cable and AC adaptor are provided with the printer, enabling connection
of the printer to the peripherals outlet.
The DeskJet Color Printer can be connected once the system is powered ON, or after shutdown.
Do not attempt to use a different type of DeskJet Color Printer (brand or model) other
than the DeskJet Color Printer provided by GE HEALTHCARE. The ultrasound system
is an extremely sensitive and complex medical system. Any unauthorized peripherals
may cause system failure or damage!
For instructions on:
-configuring the printer, see 3-5-1-8-1 "Configuring the HP 5400K Printer for Connection Via
USB Port" on page 3-40.
-setting the HP Office Pro K5400 as the default printer, see 3-5-1- 8- 2 "Se ttin g th e HP Office
Pro K5400 as the Default Report Printer" on page 3-44.
Chapter 3 - System Setup 3-39
Page 100
PRELIMINARY
GE HEALTHCARE
PART NUMBER FN091019 REVISION 2VS5 N & VS6 N SERVICE MANUAL
3-5-1-8-1Configuring the HP 5400K Printer for Connection Via USB Port
1.) Log in to Windows® with Administrator permission.
2.) On the console, press the Config button (or F2) and click the Service tab at the bottom of the
screen.
3.) From the Network Printer dialog box select the HP K5400 printer.
4.) Click Start Setup Program and follow the on-screen installation wizard instructions.
5.) Proceed as follows:
-If the software has not been installed, the HP installation wizard requests you to accept the
installation. Follow the instructions on the screen. Proceed to step 6 below to connect via a
USB cable.
-If the screen displays The Software has already been installed on your computer as
shown in Figure 3-35, then select the option to Add a device and click Next.
Figure 3-35 Adding a Device
3-40Section 3-5 - Completing the Hardware Installation
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.