This product complies with regulatory requirements of the
following European Directive 93/42/EEC concerning medical
devices.
This manual is a reference for the Vivid S5 and Vivid S6. It
applies to all versions of the 10.2.x software for the Vivid S5
and Vivid S6 ultrasound systems.
g
GE Medical Systems
MANUAL STATUS
R2424458-100
07 February 2010
Doc ID: DOC0600341
COMPANY DATAGE Medical Systems, Israel Ltd.
PRELIMINARY
GE Medical Systems. All rights reserved. No part of this
manual may be reproduced, stored in a retrieval system, or
transmitted, in any form or by any means, electronic,
mechanical, photocopying, recording, or otherwise,
without the prior written permission of GE Medical
Systems.
4 Etgar Street
39120 Tirat Carmel
Israel
Tel: (+972) 4851 9555 Fax: (+972) 4851 9500
GE Medical Systems Information Technologies GmbH,
Munzinger Strasse 5 D-79111 Freiburg, Germany
Tel: (+49) 761 45 43 0 Fax: (+49) 76145 43 233
Page 3
Table of Contents
Revision History
List of effective pages................................................................. xv
Please verify that you are using the latest revision of this
document. If you need to know the latest revision, contact your
distributor, local GE Sales Representative or in the USA call the
GE Medical Systems Clinical Answer Center at:
1-800-682-5327 or 1-262-524-5698.
Revision History
Vivid S5/Vivid S6 User Manualxv
R2424458-100 Rev. 2
Page 18
Revision History
xviVivid S5/Vivid S6 User Manual
R2424458-100 Rev. 2
Page 19
Introduction
The Vivid S5/Vivid S6 ultrasound systems are light-weight, high
performance innovative ergonomic mobile digital ultrasound
imaging systems.
Both systems provide image generation in 2D (B) Mode, Color
Doppler, Power Doppler (Angio), M-Mode, Color M-Mode, PW
and CW Doppler spectra, LVO Contrast, B-Flow, BFI option
applications.
Vivid S6 offers additional modes, like TVI or Tissue Tracking,
and options like TSI (Tissue Synchronization Imaging) and
SI/SRI (Strain/Strain-rate imaging.
The fully digital architecture of the Vivid S5/Vivid S6 unit allows
optimal usage of all scanning modes and probe types,
throughout the full spectrum of operating frequencies.
Introduction
Attention
This manual contains necessary and sufficient information to
operate the system safely. Advanced equipment training may
be provided by a factory trained Applications Specialist for the
agreed-upon time period.
Read and understand all instructions in the User's Manual
before attempting to use the Vivid S5/Vivid S6 ultrasound unit.
Keep the manual with the equipment at all time. Periodically
review the procedures for operation and safety precautions.
Prescription Device
For USA only:
CAUTION
United States law restricts this device to sale or use by, or on the
order of a physician.
Safety
All information in Chapter 1, "Safety" on page 13, should be
read and understood before operating the Vivid S5/Vivid S6
ultrasound unit.
Vivid S5/Vivid S6 User Manual1
R2424458-100 Rev. 2
Page 20
Introduction
Principles of Operation
Medical ultrasound images are created by computer and digital
memory from the transmission and reception of mechanical
high-frequency waves applied through a transducer. The
mechanical ultrasound waves spread through the body,
producing an echo where density changes occur. For example,
in the case of human tissue, an echo is created where a signal
passes from an adipose tissue (fat) region to a muscular tissue
region. The echoes return to the transducer where they are
converted back into electrical signals.
These echo signals are highly amplified and processed by
several analog and digital circuits having filters with many
frequency and time response options, transforming the
high-frequency electrical signals into a series of digital image
signals which are stored in memory. Once in memory, the
image can be displayed in real-time on the image monitor. All
signal transmission, reception and processing characteristics
are controlled by the main computer. By selection from the
system control panel, the user can alter the characteristics and
features of the system, allowing a wide range of uses, from
obstetrics to peripheral vascular examinations.
Transducers are accurate, solid-state devices, providing
multiple image formats. The digital design and use of
solid-state components provides highly stable and consistent
imaging performance with minimal required maintenance. A
sophisticated system design with computer controlled
extensive features and functions make the Vivid S5 and
Vivid S6 easy systems to use and very user friendly.
Interference caution
Use of devices that transmit radio waves near the unit could
cause it to malfunction.
CAUTION
Devices not to be used near this equipment:
Devices which intrinsically transmit radio waves such as
cellular phones, radio transceivers, mobile radio transmitters,
radio-controlled toys, and so on, should not be operated near
the unit.
2Vivid S5/Vivid S6 User Manual
R2424458-100 Rev. 2
Page 21
Medical staff in charge of the unit are required to instruct
technicians, patients, and other people who may be around the
unit, to fully comply with the above recommendations.
Indications for use
The Vivid S5/Vivid S6 ultrasound unit is intended for the
following applications:
• Abdominal
• Cardiac
• Musculoskeletal including Superficial
• Small Organ
• Pediatric
• OB/Gyn
• Fetal Heart
• Transesophageal
• Peripheral Vascular
• Neonatal Cephalic
• Adult Cephalic
• Intraoperative
Introduction
This machine should be used in compliance with law. Some
jurisdictions restrict certain uses, such as gender determination.
CAUTION
Contraindications
The Vivid S5/Vivid S6 ultrasound unit is not intended for
ophthalmic use or any use causing the acoustic beam to pass
DANGER
through the eye.
Documentation
Vivid S5/Vivid S6 documentation consists of two manuals:
• The Basic User Manual (TRANSLATED) provides
information needed by the user to operate the system
safely. It describes the basic functions of the system, safety
features, operating modes, measurements/calculations,
Vivid S5/Vivid S6 User Manual3
R2424458-100 Rev. 2
Page 22
Introduction
probes, and user care and maintenance.
Note: Probe information displayed on screen examples
does not necessarily reflect the probes available on your
ultrasound system. Please refer to the Probes chapter for
a listing of available probes and features.
• The Advanced Reference Manual (ENGLISH ONLY)
contains M&A formulas and parameter-names, data tables,
such as OB and Acoustic Output tables.
Note: The documentation kit provides the Basic User
Manual and Advanced Reference Manual in electronic
format only. The CD-ROM includes English and all
translations. Paper documentation may be ordered.
The Vivid S5/Vivid S6 documentation is written for users who
are familiar with basic ultrasound principles and techniques.
They do not include sonographic training or detailed clinical
procedures.
Note: The original documentation was written in English.
CAUTION
Manual contents
The Vivid S5/Vivid S6 User's Manual is organized to provide
the information needed to start scanning immediately.
Some of the functions or features described in this manual are
optional and may not be available in the configuration of your
specific system.
The safety instruction must be reviewed before operation of the
unit.
Finding information
Table of Contents, lists the main topics and their location.
Headers and Footers, give the chapter name and page
number.
Index, provides an alphabetical and contextual list of topics.
4Vivid S5/Vivid S6 User Manual
R2424458-100 Rev. 2
Page 23
Conventions used in this manual
2-column layout, the right column contains the main text. The
left column contains notes, hints, and warnings.
Keys and buttons, on the control panel are indicated by over
and underlined text (ex.
Bold type, describes button names on the screen.
Italic type: describes program windows, screens and dialogue
boxes.
Safety icons, highlight safety issues as described in
"Introduction" on page 15.
Product icons, indicate product variant features as follows:
Indicates that the relevant feature exists in the standard configuration
of Vivid S6 and is not available on Vivid S5.
2D refers to the 2D mode key)
Introduction
Indicates that the relevant feature exists as an option of the Vivid S6
and is not available on Vivid S5.
Indicates that the relevant feature exists as an option on both Vivid S6
and Vivid S5 systems.
Indicates that the relevant feature exists in the standard configuration
of Vivid S6 and is available as an option on Vivid S5.
Indicates that the relevant feature exists in the standard configuration
of Vivid S5 and is not available on Vivid S6.
Vivid S5/Vivid S6 User Manual5
R2424458-100 Rev. 2
Page 24
Introduction
Indicates that a specific hazard is known to exist which through
inappropriate conditions or actions, will cause:
DANGER
WARNING
CAUTION
• Severe or fatal personal injury
• Substantial property damage
Indicates that a specific hazard is known to exist which through
inappropriate conditions or actions, will cause:
• Severe personal injury
• Substantial property damage
Indicates that a potential hazard may exist which through
inappropriate conditions or actions, will or can cause:
• Minor injury
• Property damage
Regulatory requirements
The Vivid S5/Vivid S6 ultrasound unit confirms to directives,
classifications, and standards, as described in "Regulatory
information" on page 19.
6Vivid S5/Vivid S6 User Manual
R2424458-100 Rev. 2
Page 25
Contact information
If additional information or assistance is needed, please
contact the local distributor or the appropriate support resource
listed bellow:
This section describes the important safety measures which
should be taken before operating the Vivid S5/Vivid S6
ultrasound unit. Procedures for simple care and maintenance
of the unit are also described.
Various levels of safety precautions may be found on the
equipment, and different levels of severity are identified by one
of the following icons that precede precautionary statements in
the text.
The following icons and conventions are used to indicate
precautions:
Indicates that a specific hazard is known to exist which through
inappropriate conditions or actions, will cause:
DANGER
WARNING
• Severe or fatal personal injury
• Substantial property damage
Indicates that a specific hazard is known to exist which through
inappropriate conditions or actions, will cause:
• Severe personal injury
• Substantial property damage
Safety
Indicates that a potential hazard may exist which through
inappropriate conditions or actions, will or can cause:
CAUTION
• Minor injury
• Property damage
Note:
used in the operation of the ultrasound system, specifically:
Indicates precautions or recommendations that should be
• Maintaining an optimum system environment
• Using this Manual
• Notes to emphasize or clarify a point
Other precautions or prudent-use recommendations are
indicated in the note sections in the left column. These are:
• Use of the Vivid S5/Vivid S6 ultrasound unit as a
prescription device, under the order of a physician.
• Maintaining an optimum unit environment.
• Reference to the User's Manual.
Vivid S5/Vivid S6 User Manual15
R2424458-100 Rev. 2
Page 34
Safety
Hazard symbols
Icon description
Potential hazards are indicated by the following icons:
IconPotential hazardUsageSource
• Patient/user infection due to
contaminated equipment
• Electrical micro-shock to
patient, e.g., ventricular
• Console, accessories or
optional storage devices
that can fall on patient, user,
or others.
• Collision with persons or
objects may result in injury
while maneuvering or during
system transport.
• Injury to user from moving
the console
• Patient injury or tissue
damage from ultrasound
radiation
• Cleaning and care
instructions
• Sheath and glove
guidelines
• Probes
• ECG (if applicable)
• Connections to back panel
• Moving
• Using brakes
• Transporting
• ALARA, the use of Power
Output following the 'as
low as reasonably
achievable' principle
ISO 7000
No. 0659
• Risk of explosion if used in
the presence of flammable
anesthetics
• Patient/user injury or
adverse reaction from fire or
smoke
• Patient/user injury from
explosion and fire
• Flammable anesthetic
• Replacing fuses
• Outlet guidelines
16Vivid S5/Vivid S6 User Manual
R2424458-100 Rev. 2
Page 35
Owner responsibility
For USA only:
Safety
CAUTION
Federal law restricts this device to use by, or on the orders of, a
physician.
It is the responsibility of the owner to ensure that anyone
operating the system reads and understands this section of the
manual. However, there is no representation that the act of
reading this manual renders the reader qualified to operate,
inspect, test, align, calibrate, troubleshoot, repair or modify the
system. The owner should make certain that only properly
trained, fully-qualified service personnel undertake the
installation, maintenance, troubleshooting, calibration and
repair of the equipment.
The owner of the Vivid S5/Vivid S6 ultrasound unit should
ensure that only properly trained, fully qualified personnel are
authorized to operate the system. Before authorizing anyone to
operate the system, it should be verified that the person has
read, and fully understands, the operating instructions
contained in this manual. It is advisable to maintain a list of
authorized operators.
Should the system fail to operate correctly, or if the unit does
not respond to the commands described in this manual, the
operator should contact the nearest field GE Medical Systems
Service Office.
For information about specific requirements and regulations
applicable to the use of electronic medical equipment, consult
the local, state and federal agencies.
Vivid S5/Vivid S6 User Manual17
R2424458-100 Rev. 2
Page 36
Safety
Important safety considerations
Notice against user modification
Never modify this product, including system components,
software, cables, and so on. User modification may cause
safety hazards and degradation in system performance. All
modification must be done by a GE qualified person.
The equipment is not suitable for use in the presence of
flammable anesthetic mixture with air or with Oxygen or Nitrous
Oxide.
This section includes considerations for the
following
• Patient safety
• Personnel and equipment safety
The information contained in this section is intended to
familiarize the user with the hazards associated with the use of
the unit, and to alert them to the extent to which injury and
damage may occur if the precautions are not observed.
CAUTION
Improper use can result in serious injury. The user must be
thoroughly familiar with the instructions and potential hazards
involving ultrasound examination before attempting to use the
device. Training assistance is available from GE Medical
Systems if needed.
Users are obligated to familiarize themselves with these safety
considerations and to avoid conditions that could result in injury
or damage.
18Vivid S5/Vivid S6 User Manual
R2424458-100 Rev. 2
Page 37
Regulatory information
Any changes to accessories, peripheral units or any other part of
the system must be approved by the manufacturer. Ignoring this
CAUTION
advice may compromise the regulatory approvals obtained for
the product.
Directives
The GE Healthcare Ultrasound product families are tested to
meet all applicable requirements in relevant EU Directives and
European/International standards.
• Council Directive 93/42/EEC concerning MDD (Medical
Devices Directive): the CE label affixed to the product
testifies compliance to this Directive.
The location of the CE marking is specified in "Device
labels" on page 22.
Product Classifications
Safety
The Vivid S5/Vivid S6 ultrasound unit confirms to the following
classifications, in accordance with the IEC/EN 60601-1:6.8.1:
• According to 93/42/EEC Medical Device Directive, this is
Class IIa Medical Device.
• According to IEC/EN 60601-1, Equipment is Class I, Type
B with BF or CF Applied Parts.
• According to CISPR 11, this is Group 1, Class B ISM
Equipment.
• According to IEC 60529, the footswitch rate IPx8 is suitable
for use in surgical rooms.
• Classification according to the degree of protection against
ingress of water as detailed in the current edition of IEC 529
(section 6.1.1): Ordinary for Main Unit (PC), IPX1, IPX7,
IPX8 for ultrasound probes (transducers).
Conformity to Standards
To fulfill the requirements of relevant EC directives and/or
European Harmonized/International standards, the following
documents/standards have been used:
• International Electrotechnical Commission (IEC).
Vivid S5/Vivid S6 User Manual19
R2424458-100 Rev. 2
Page 38
Safety
• IEC/EN 60601-1: Medical Electrical Eqiupment, Part 1
General Requirements for Safety
• IEC/EN 60601-1-2: Medical electrical equipment Part 1-2: General requirements for safety - Collateral
standard: Electromagnetic compatibility - Requirements
and tests
• EN 55011/ CISPR11+A2 ED3.2: Emitted noise
according to Class B requirements + Electromagnetic
Susceptibility
• IEC/EN 60601-1-4: Medical electrical equipment - Part
1-4: General requirements for safety - Collateral
standard: Programmable electrical medical systems
• IEC/EN 60601-1-6: Medical electrical equipment - Part
1-6: General requirements for basic safety and essential
performance - Collateral Standard: Usability
• IEC/EN 60601-2-37: Medical Electrical Equipment Part
2-37: Particular Requirements for the Safety of
Ultrasonic Medical Diagnostic and Monitoring
Equipment
• EN60601-1-1: 2001 - Medical Electrical Equipment Part
1-1: General Requirements for Safety Collateral
Standard: Safety Requirements for Medical Electrical
Systems
• International Organization of Standards (ISO)
• ISO 10993-1: Biological Evaluation of Medical Devices Part 1: Evaluation & Testing
• ISO 10993-5: Biological Evaluation of Medical Devices Part 5: Tests for In Vitro Cytotoxicity
• ISO 10993-10:Biological evaluation of medical devices Part 10: Tests for irritation and delayed-type
hypersensitivity
• Underwriters' Laboratories, Inc. (UL), an independent
testing laboratory.
• UL 60601-1 Medical Electrical Equipment, Part 1
General Requirements for Safety.
• Canadian Standards Association (CSA).
• CSA 22.2, 601.1 Medical Electrical Equipment, Part 1
General Requirements for Safety.
20Vivid S5/Vivid S6 User Manual
R2424458-100 Rev. 2
Page 39
Safety
• NEMA/AIUM Acoustic Output Display Standard (NEMA
US-3, 1998).
• Medical Device Good Manufacturing Practice Manual
issued by the FDA (Food and Drug Administration,
Department of Health, USA).
Certifications
• Quality management standards for medical devices:
General Electric Medical Systems is ISO 9001 and
ISO13485 certified.
Label 5 - Universal Label, for all systemsVivid S5/Vivid S6 label locations
Figure 1-1: System Label and Location
22Vivid S5/Vivid S6 User Manual
R2424458-100 Rev. 2
Page 41
Label Icon Description
The following table describes the purpose and location of
safety labels and other important information provided on the
equipment.
LabelPurposeLocation
Safety
Identification and
Rating Plate
• Manufacturer's name and address
• Date of Manufacture
• Model and serial numbers
• Electrical ratings (Volts, Amps,
phase, and frequency)
• Device Listing/Certification Labels
Equipment Type BF, in which
protection against electric shock does
not rely on basic insulation only.
Provides additional safety
precautions such as double insulation
or reinforced insulation, because
there is no provision for protective
earthing or reliance upon installation
conditions.
Equipment Type CF, indicates
equipment having a floating applied
part having a degree of protection
suitable for direct cardiac contact.
Bottom of unit
Probe connectors.
ECG connector
Bottom of unit.
Alternating currentVarious
This symbol indicates that the waste
of electrical and electronic equipment
must not be disposed as unsorted
municipal waste and must be
collected separately. Please contact
an authorized representative of the
manufacturer for information
concerning the decommissioning of
your equipment.
Vivid S5/Vivid S6 User Manual23
R2424458-100 Rev. 2
Bottom Cover
Page 42
Safety
LabelPurposeLocation
Protective earth (ground)Internal
Earth (ground)Internal
Equipotentiality: indicates terminal to
be used for connecting equipotential
conductors when interconnecting
(grounding) with other equipment.
Connection of additional protective
earth conductors or potential
equalization conductors is not
necessary in most cases and is only
recommended for situations involving
multiple equipment in a high-risk
patient environment to provide
assurance that all equipment is at the
same potential and operates within
acceptable leakage current limits. An
example of a high-risk patient would
be a special procedure where the
patient has an accessible conductive
path to the heart such as exposed
cardiac pacing leads.
Attention - Consult accompanying
documents: alerts the user to refer to
the user documentation when
complete information cannot be
provided on the label.
This marking on the control panel is
especially intended to alert the user
to consult the user manual for use
BEFORE operation of the system.
Bottom of unit
Various
CAUTION - Dangerous voltage: used
to indicate electric shock hazards.
Various
24Vivid S5/Vivid S6 User Manual
R2424458-100 Rev. 2
Page 43
LabelPurposeLocation
Safety
Apply a short push on the ON/OFF
button to shut down the system.
CAUTION: This Power Switch DOES
NOT ISOLATE Mains Supply.
This product consists of devices that
may contain mercury, which must be
recycled or disposed of in
accordance with local, state, or
country laws. (Within this system, the
backlight lamps in the monitor display
contain mercury.)
Date of manufacture: The date
could be a year, year and month, or
year, month and day, as appropriate.
See ISO 8601 for date formats.
Catalog or model numberRear of unit
Serial numberRear of unit
Direct Current: For products to be
powered from a DC supply.
Keyboard
Rear of unit
Rear of unit
GOST-R Mark: per Law of the
Russian Federation No. 184-FZ. The
field 0000 will contain the number of
the institute that issued the GOST
label.
Prescription Device Label for United
States per 21 CFR 801.109(b)(1)
Vivid S5/Vivid S6 User Manual25
R2424458-100 Rev. 2
Bottom of unit
Bottom of unit
Page 44
Safety
Acoustic output
Definition of the acoustic output
parameters
Thermal Index
TI is an estimate of the temperature increase of soft tissue or
bone. There are three thermal index categories:
• TIS: Soft tissue thermal index. The main TI category. Used
for applications that do not image bone.
• TIB: Bone thermal index (bone located in a focal region).
Used for fetal application.
• TIC: Cranial bone thermal index (bone located close to the
surface). Used for transcranial application.
Mechanical Index
MI is the estimated likelihood of tissue damage due to
cavitation. The absolute maximum limits of the MI is 1.9 as set
by the FDA 510 (k) guidance of 1997.
Note: Further explanation on "cavitation" appears in the
Reference Manual Chapter 3 - Nonthermal Bioeffects.
Ispta
The Ispta is the Spatial Peak Temporal Average Intensity. The
absolute maximum limit of Ispta is 720 MW/cm
FDA 510(k) guidance of 1997.
2
as set by the
ALARA
Ultrasound procedures should be performed using output
levels and exposure times As Low As Reasonably Achievable
(ALARA) while acquiring clinical information.
Training
During each ultrasound examination the user is expected to
weigh the medical benefit of the diagnostic information that
would be obtained against the risk of potential harmful effects.
Once an optimal image is achieved, the need for increasing
acoustic output or prolonging the exposure cannot be justified.
26Vivid S5/Vivid S6 User Manual
R2424458-100 Rev. 2
Page 45
Safety
It is recommended that all users receive proper training in
applications before performing them in a clinical setting. Please
contact the local GE sales representative for training
assistance.
Safety statement
GE Medical Systems safety statement
Although no harmful biological effects have been demonstrated
for ultrasound frequencies, intensities and exposure times used
in examination with the GE Vivid S5/Vivid S6 system, GE
Medical Systems recommends using the lowest acoustic
output settings which will produce diagnostically acceptable
information.
System controls affecting acoustic
output
The operator controls that directly affect the acoustic output are
discussed in the Acoustic Output Data Tables in the Reference
Manual. These tables show the highest possible acoustic
intensity for a given mode, obtainable only when the maximum
combination of control settings is selected. Most settings result
in a much lower output. It is important to note the following:
• The duration of an ultrasound examination is as important
as the acoustic output, since patient exposure to output is
directly related to the exposure time.
• Better image quality yields faster clinical results, making it
possible to complete the relevant ultrasound examination
more rapidly. Therefore, any control that improves the
quality of the examination can help to reduce patient
exposure, even though it may not directly affect acoustic
output.
Probe selection
As long as the appropriate application is available, any probe
can be used with the knowledge that the intensities fall at, or
below, those stated in the Acoustic Output Data Tables. The
duration of patient exposure is most likely minimized with the
use of a probe that is optimized to provide resolution and focal
depth, appropriate to the examination.
Vivid S5/Vivid S6 User Manual27
R2424458-100 Rev. 2
Page 46
Safety
Application selection
Selecting the probe and application preset appropriate to a
particular ultrasound examination automatically provides
acoustic output limits within FDA guidelines for that application.
Other parameters which optimize performance for the selected
application are also set automatically, and should assist in
reducing the patient exposure time. See page 100, for
information on selecting probes and application presets.
Changing imaging modes
Acoustic output depends on the imaging mode selected. The
choice of mode (2D, M-Mode, Doppler or Color Flow)
determines whether the ultrasound beam is stationary or in
motion. This greatly affects the energy absorbed by the tissue.
See Chapter 4, "Scanning Modes" on page 137, for complete
information on changing imaging modes.
When operating in a combined mode, such as 2D and M-Mode,
the total acoustic output comprises contributions from each
individual mode. Depending on the modes in use, either or both
output indices may be affected.
The user can override the default settings, but care should be
taken to observe the displayed MI and TI values.
Power
It is possible to change the power in all operating modes so that
the operator can use the ALARA principle.
28Vivid S5/Vivid S6 User Manual
R2424458-100 Rev. 2
Page 47
OB Exam
Safety
Exam Preparation
Prior to an ultrasound examination, the patient should be
informed of the clinical indication, specific benefits, potential
risks, and alternatives, if any. In addition, if the patient requests
information about the exposure time and intensity, it should be
provided. Patient access to educational materials regarding
ultrasound is strongly encouraged to supplement the
information communicated directly to the patient. Furthermore,
these examinations should be conducted in a manner and take
place in a setting which assures patient dignity and privacy.
• Prior material knowledge and approval of the presence of
nonessential personnel with the number of such personnel
kept to a minimum.
• An intent to share with the parents per the physician's
judgment, either during the examination or shortly
hereafter, the information derived.
• An offer of choice about viewing the fetus.
• An offer of choice about learning the sex of the fetus, if such
information becomes available.
Ultrasound examinations performed solely to satisfy the
family's desire to know the fetal sex, to view the fetus, or to
obtain a picture of the fetus should be discouraged.
Vivid S5/Vivid S6 User Manual29
R2424458-100 Rev. 2
Page 48
Safety
Acoustic Output Considerations
The Vivid S5/Vivid S6 system is a multi-use device which is
capable of exceeding FDA Pre-enactment acoustic output
WARNING
CAUTION
(spatial peak-temporal average) intensity limits for fetal
applications.
It is prudent to conduct an exam with the minimum amount and
duration of acoustic output necessary to optimize the image's
diagnostic value.
Concerns surrounding fetal exposure
Always be aware of the acoustic output level by observing the
Acoustic Output Display. In addition, become thoroughly
familiar with the Acoustic Output Display and equipment
controls affecting output.
30Vivid S5/Vivid S6 User Manual
R2424458-100 Rev. 2
Page 49
Patient safety
Patient identification
The concerns listed in this section can seriously affect the safety
of the patient undergoing a diagnostic ultrasound examination.
WARNING
Always include proper identification with all patient data and
verify the accuracy of the patient's name and/or identity number
when entering such data. Ensure that the correct patient ID is
provided on all recorded data and hard copy prints.
Identification errors could result in an incorrect diagnosis.
Note: For further information on patient identification, see
"Starting an examination" on page 95.
Diagnostic information
The images and calculations provided by the system are
intended for use by competent users, as a diagnostic tool. They
are explicitly not to be regarded as the sole, irrefutable basis for
clinical diagnosis. Users are encouraged to study the literature
and reach their own professional conclusions regarding the
clinical utility of the system.
Safety
The user should be aware of the product specifications and of
the system accuracy and stability limitations. These limitations
must be considered before making any decision based on
quantitative values. If in doubt, the nearest GE Medical
Systems Service Office should be consulted.
Equipment malfunction or incorrect settings can result in
measurement errors or failure to detect details in the image.
The user must become thoroughly familiar with the operation of
the unit in order to optimize its performance and to recognize
possible malfunctions. Application training is available through
the sales representative.
Be certain to ensure privacy data of patient information.
CAUTION
Vivid S5/Vivid S6 User Manual31
R2424458-100 Rev. 2
Page 50
Safety
Patient guidance
Remember to assist the patient when moving up to the
examination bed, moving down from the bed, or when moving in
CAUTION
the vicinity of the system.
Make sure to keep the patient in full view at all times and never
leave the patient unattended while on the examination bed.
Probe Safety
This section includes information on hazards to both the user
and the equipment, as follow:
• Mechanical hazards
• Electrical hazards
• Biological hazards
Mechanical hazards
Damaged probes or improper use and manipulation of the
transesophageal probe may result in injury or increased risk of
infection. Inspect probes frequently for sharp, pointed or rough
surface damage that could cause injury or tear protective
barriers (gloves and sheaths).
Never use excessive force when manipulating intracavity
probes. Become familiar with all instructions and precautions
provided with special purpose probes.
The use of damaged probes can result in injury or increased
risk of infection. Inspect probes often for sharp, pointed, or
rough surface damage that could cause injury or tear protective
barriers. Become familiar with all instructions and precautions
provided with special purpose probes.
Observe immersion levels as displayed in Figure 11-5, page 559.
WARNING
Inspect probes for sharp edges or rough surfaces that could
injure sensitive tissue.
DO NOT bend or pull the cable forcefully, to avoid mechanical
shock or impact to the probe.
32Vivid S5/Vivid S6 User Manual
R2424458-100 Rev. 2
Page 51
CAUTION
CAUTION
Safety
Allowing the machine to transmit acoustic output with the probe
not in use (or in its holder) can cause the transducer to build up
heat. Always turn off acoustic output or freeze the image when
not in use. The system's acoustic output remains transmitting
when the user controls are being used.
The system is equipped with an Auto Freeze feature which
disables acoustic output and freezes the image when the
machine is not in use.
Ultrasound transducers are sensitive instruments which can
easily be damaged by rough handling. Take extra care not to
drop transducers and avoid contact with sharp or abrasive
surfaces. A damaged housing, lens or cable can result in patient
injury or serious impairment or operation.
Transesophageal probe safety
Never use excessive force when manipulating the
transesophageal probe. The detailed operator manual
enclosed with the transesophageal probe must be read
carefully.
Electrical Hazard
Probes are driven by electricity, which can injure the patient or
user when exposed to contact with conductive solution.
A damaged probe may increase the risk of electric shock if
conductive solutions come in contact with internal live pads.
Inspect probes often for cracks or openings in the housing and
holes in and around the acoustic lens, or other damage that
could allow moisture to enter. Become familiar with the use and
care precautions described in Chapter 11, "Probes" on
page 535.
Vivid S5/Vivid S6 User Manual33
R2424458-100 Rev. 2
Page 52
Safety
WARNING
CAUTION
Do not immerse the probe into any liquid beyond the level shown
in Figure 11-5, page 559. Never immerse the probe connector or
adaptors into any liquid.
Do not subject the probe to mechanical shock or impact, which
may result in cracks or chips in the housing and degrade
performance.
Inspect the probe before and after each use, as described on
page 553, for damage or degradation to the housing, strain relief,
lens and seal.
DO NOT apply excessive force to the probe cable, to prevent
insulation failure.
Electrical leakage checks should be performed regularly by a GE
service representative or qualified hospital personnel, according
to the procedures described in EN 60601-1/IEC 60601-1 §19.
Ultrasound can produce harmful effects in tissue and potentially
result in patient injury. Always minimize exposure time and keep
ultrasound levels low when there is no medical benefit. Use the
principle of ALARA (As Low As Reasonably Achievable),
increasing output only when needed to obtain diagnostic image
quality. Observe the acoustic output display and be familiar with
all controls affecting the output level. See the Bioeffects section
of the Acoustic Output chapter in the Advanced Reference
Manual for more information.
Biological hazards
Transesophageal probes require a special handling. Refer to the
user documentation enclosed with these probes.
CAUTION
To minimize disease transmission, legally marketed and sterile
pyrogen-free sheaths should be used for each probe
recommended for intra-cavity procedures.
Adequate cleaning and disinfection are essential to prevent
disease transmission. It is the responsibility of the user to verify
and maintain the effectiveness of the infection control
procedures in use.
34Vivid S5/Vivid S6 User Manual
R2424458-100 Rev. 2
Page 53
Personnel and equipment safety
The hazards listed below can seriously affect the safety of
personnel and equipment during a diagnostic ultrasound
DANGER
examination.
Explosion hazard
Never operate the equipment in the presence of flammable or
explosive liquids, vapors or gases. Malfunctions in the unit, or
sparks generated by fan motors, can electrically ignite these
substances. Operators should be aware of the following points
to prevent such explosion hazards.
• If flammable substances are detected in the environment,
do not plug in or turn on the system.
• If flammable substances are detected after the system has
been turned on, do not attempt to turn off the unit, or to
unplug it.
• If flammable substances are detected, evacuate and
ventilate the area before turning off the unit.
Safety
WARNING
Electrical hazard
The internal circuits of the unit use high voltages, capable of
causing serious injury or death by electrical shock.
If any defects are observed or malfunctions occur, stop
operating the equipment and perform the proper action for the
patient. Inform a qualified service person and contact a Service
Representative for information.
To avoid injury
• Do not remove the unit's protective covers. No
user-serviceable parts are inside. If servicing is required,
contact qualified technical personnel.
• Connect the attachment plug to a hospital-grade grounding
outlet to ensure adequate grounding.
• Never use any adaptor or converter of a three-prong-to
two-prong type to connect with a mains power plug. The
protective earth connection will loosen.
Vivid S5/Vivid S6 User Manual35
R2424458-100 Rev. 2
Page 54
Safety
CAUTION
• Do not place liquids on or above the unit. Conductive fluids
seeping into the active circuit components may cause a
short-circuit, which could result in an electrical fire.
• An electrical hazard may exist if any light, monitor or visual
indicator remains on after the unit is turned off.
Fuses blown within 36 hours of being replaced may indicate a
malfunctioning electrical circuit within the system. In this event,
the unit must be checked by GE Medical Systems service
personnel. No attempt should be made to replace the fuses
with others of a higher rating.
Do not use this equipment if a safety problem is known to exist.
Have the unit repaired and performance verified by qualified
service personnel before returning to use.
Smoke and fire hazard
The system must be supplied from an adequately rated
electrical circuit. The capacity of the supply circuit must be as
specified.
Biological hazard
For patient and personnel safety, beware of biological hazards
while performing transesophageal procedures. To avoid the
risk of disease transmission:
• Use protective barriers (gloves and probe sheaths)
whenever necessary. Follow sterile procedures as
required.
• Thoroughly clean probes and reusable accessories after
each patient examination and disinfect or sterilize as
needed. Refer to Chapter 11, "Probes" on page 535, for
probe use and care instructions.
• Follow all in-house infection control policies as they apply
to personnel and equipment.
Pacemaker hazard
The possibility of the system interfering with pacemakers is
minimal. However, as this system generates high frequency
electrical signals, the operator should be aware of the potential
hazard this could cause.
36Vivid S5/Vivid S6 User Manual
R2424458-100 Rev. 2
Page 55
CAUTION
CAUTION
CAUTION
CAUTION
Safety
To avoid injury or system damage, NEVER place any object or
liquid on the monitor.
Archived data is managed at the individual sites. Performing data
backup (to any device) is recommended.
Do not unpack the Vivid S5/Vivid S6. This must be performed by
qualified service personnel only.
Do not use the Vivid S5/Vivid S6 Ultrasound system ECG wave
for diagnosis and monitoring.
DO NOT load non-system software on the system computer.
CAUTION
LCD Monitor
• DO NOT place a finger, hand or any object on the joint of
the monitor or monitor arm to avoid injury when moving the
monitor and monitor arm.
• To avoid result of injury or system damage, NEVER place
any object or liquid on the monitor, whether in the home or
flip down/transport position.
• DO NOT place any object on the ventilation slots on the
upper rear side of the monitor cabinet. Blocking the
ventilation slots prevents proper airflow and may result in
fire, electric shock, or equipment damage.
• Keep small objects or liquids away from the LCD monitor.
Vivid S5/Vivid S6 User Manual37
R2424458-100 Rev. 2
Page 56
Safety
Small objects or liquid may accidentally fall through the
ventilation slots into the cabinet or spill into the cabinet
which may result in fire, electric shock, or equipment
damage.
If an object or liquid falls/spills into the cabinet, unplug the
system immediately. Call a Service Representative for
information.
• DO NOT scratch or press on the panel with any sharp
objects, such as a pencil or pen, as this may result in
damage to the panel.
38Vivid S5/Vivid S6 User Manual
R2424458-100 Rev. 2
Page 57
Electrical safety
Internally connected peripheral devices
The system, together with peripheral devices, such as video
tape recorders and printers, meets UL 60601-1 and
IEC 60601-1 (1988) standards for electrical isolation and
safety. These standards are applicable only when the specified
peripheral devices are plugged into the AC outlets provided in
the unit.
External Connection of other peripheral
devices
External devices can be used only if CE marked and in
compliance with related standards (EN 60601-1 or EN 60950).
CAUTION
Conformance to EN 60601-1-1 (2000) must be verified.
External devices meeting EN 60950 should be kept outside of the
patient environment, as defined in IEC 60601-1-1 (2000).
Safety
Other external devices, such as laser cameras, printers, VCRs
and external monitors, usually exceed allowable leakage limits
and, when plugged into separate AC outlets that are then
connected to the unit, are in violation of patient safety
standards. Suitable electrical isolation of such external AC
outlets may be required in order to meet UL2601-1 and
IEC 60601-1 (1988) standards for electrical leakage.
Vivid S5/Vivid S6 User Manual39
R2424458-100 Rev. 2
Page 58
Safety
Allergic reactions to latex-containing
medical devices
Due to reports of severe allergic reactions to medical devices
containing latex (natural rubber), the FDA advises health-care
professionals to identify latex-sensitive patients, and be
prepared to treat allergic reactions promptly. Latex is a
component of many medical devices, including surgical and
examination gloves, catheters, incubation tubes, anesthesia
masks and dental dams. Patient reaction to latex has ranged
from contact urticaria, to systemic anaphylaxis.
For more details regarding allergic reaction to latex, refer to
FDA Medical Alert MDA91-1, March 29, 1991 Medical Alert on
latex products.
40Vivid S5/Vivid S6 User Manual
R2424458-100 Rev. 2
Page 59
Use of ECG
Do not use the Vivid S5/Vivid S6 Ultrasound system ECG wave
for diagnosis and monitoring.
CAUTION
Use of Defibrillator
Do not use the Vivid S5/Vivid S6 Ultrasound system with
Defibrillator. This equipment does not have defibrillator-
CAUTION
CAUTION
approved applied parts.
Remove the TEE probe from the patient when defibrillators are
used.
Use of Electrosurgical Unit
To avoid skin burns in surgical use, do not place ECG electrodes
in current path between Electrosurgical Unit (ESU) active and
CAUTION
dispersive electrodes. Keep ESU cables away from ECG leads.
Safety
This equipment provides no special means of protection from
high frequency (HF) burns that may result from using an
electrosurgical unit (ESU). To reduce the risk of HF burns,
avoid contact between the patient and ultrasound transducer
while operating the ESU. Where contact cannot be avoided, as
in the case of TEE monitoring during surgery, make sure the
transducer is not located between the ESU active and
dispersive electrodes and keep the ESU cables away from the
transducer cable.
Vivid S5/Vivid S6 User Manual41
R2424458-100 Rev. 2
Page 60
Safety
Electromagnetic Compatibility (EMC)
NOTE: This unit carries the CE mark. It complies with
regulatory requirements of the European Directive 93/42/EEC
concerning medical devices. It also complies with emission
limits for a Group 1, Class B Medical Device as stated in
IEC/EN 60601-1-2 (2001).
All types of electronic equipment may characteristically cause
electromagnetic interference with other equipment, transmitted
either through air or connecting cables. The term
Electromagnetic Compatibility (EMC), indicates the capability
of the equipment to curb electromagnetic influence from other
equipment, while at the same time not affecting other
equipment with similar electromagnetic radiation.
Radiated or conducted EMC can cause distortion, degradation,
or artifacts in the ultrasound image which could potentially
obscure diagnostic information.
There is no guarantee that interference will not occur in a
particular installation. If this equipment is found to cause or
respond to interference, which may be determined by turning
equipment on and off, qualified service personnel should
attempt to correct the problem by one or more of the following
measures:
• Re-orient or re-locate the affected device.
• Increase the separation between the unit and the affected
device.
• Power the equipment from a source other than that of the
affected device.
• Consult the service representative for further suggestions.
The manufacturer is not responsible for any interference or
responses caused by the use of interconnecting cables other
than those recommended, or by unauthorized changes or
modifications to this unit. Unauthorized changes or
modifications could void the user's authority to operate the
equipment.
To comply with the regulations on electromagnetic interference,
all interconnecting cables to peripheral devices must be
shielded and properly grounded. Use of cables not properly
shielded and grounded may result in the equipment causing or
42Vivid S5/Vivid S6 User Manual
R2424458-100 Rev. 2
Page 61
Safety
responding to radio frequency interference, in violation of the
European Union Medical Device Directive and FCC
regulations.
Do not use devices which intentionally transmit RF signals, for
example, cellular phones, transceivers, or radio controlled
products, in the vicinity of this equipment as it may cause
performance outside the published specifications. Keep the
power to these types of devices turned off when near this
equipment.
EMC performance
Portable and mobile radio communications equipment (e.g.
two-way radio, cellular/cordless telephones, wireless computer
networks), other than those supplied by GE, should be used no
closer to any part of this system, including cables, than
determined according to the following method:
Table 1-1: Portable and mobile radio communications equipment distance requirements
Where: d= separation distance in meters, P = rated power of the transmitter,
V
=compliance value for conducted RF, E1 = compliance value for radiated RF
1
If the maximum
transmitter power
in watts is ratedThe separation distance in meters should be:
52.62.65.2
205.25.210.5
10012.012.024.0
d=[3.5/V
root of P
] square
1
d = [3.5/E1] square
root of P
d = [7/E1] square
root of P
Vivid S5/Vivid S6 User Manual43
R2424458-100 Rev. 2
Page 62
Safety
Declaration of Emissions
This system is suitable for use in the following environment.
The user must assure that it is used only in the electromagnetic
environment as specified.
This system uses RF energy only for its
internal function. Therefore, RF
emissions are very low and are not likely
to cause any interference in nearby
electronic equipment. It is suitable for use
in all establishments, including domestic
establishments and those directly
connected to the public low-voltage
power supply network that supplies
buildings used for domestic purposes.
Declaration of Immunity
This system is suitable for use in the following environment.
The user must assure that the system is used according to the
specified guidance and only in the electromagnetic
environment listed.
Notice upon Installation of Product
Separation distance and effect from fixed radio
communications equipment: field strengths from fixed
transmitters, such as base stations for radio (cellular/cordless)
telephones and land mobile radios, amateur radio, AM and FM
radio broadcast, and TV broadcast transmitter cannot be
predicted theoretically with accuracy. To assess the
44Vivid S5/Vivid S6 User Manual
R2424458-100 Rev. 2
Page 63
Safety
electromagnetic environment due to fixed RF transmitters, an
electromagnetic site survey should be considered. If the
measured field strength in the location in which the ultrasound
system is used exceeds the applicable RF compliance level as
stated in the immunity declaration, the ultrasound system
should be observed to verify normal operation. If abnormal
operation is observed, additional measures may be necessary,
such as re-orienting or relocating the ultrasound system or
using an RF shielded examination room may be necessary.
1.Use either power supply cords provided by GE Medical
Systems or ones designated by GE Medical Systems.
Products equipped with a power source plug should be
plugged into the fixed power socket which has the
protective grounding conductor. Never use any adaptor or
converter to connect with a power source plug (e.g.
three-prong-to-two-prong converter).
2.Locate the equipment as far away as possible from other
electronic equipment.
3.Be sure to use only the cables provided by or designated by
GE Medical Systems. Connect these cables following the
installation procedures (e.g. wire power cables separately
from signal cables).
4.Lay out the main equipment and other peripherals following
the installation procedures described in the Option
Installation manuals.
General notice
1.Designation of Peripheral Equipment Connectable to This
Product.
The equipment indicated in Chapter 12, "Peripherals" on
page 575 can be hooked up to the product without
compromising its EMC performance. Avoid using
equipment not designated in the list. Failure to comply with
this instruction may result in poor EMC performance of the
product.
2.Notice against User Modification The user should never
modify this product. User modifications may cause
degradation in EMC performance. Modification of the
product includes changes in:
• Cables (length, material, wiring, etc.)
• System installation/layout
Vivid S5/Vivid S6 User Manual45
R2424458-100 Rev. 2
Page 64
Safety
• System configuration/components
• Securing system parts (cover open/close, cover
screwing)
3.Operate the system with all covers closed. If a cover is
opened for some reason, be sure to shut it before starting/
resuming operation.
4.Operating the system with any cover open may affect EMC
performance.
Peripheral Update for EC countries
The following is intended to provide the users in EC countries
with updated information concerning the connection of the
Vivid S5/Vivid S6 to image recording and other devices or
communication networks.
The Vivid S5/Vivid S6 has been verified for overall safety,
compatibility and compliance with the following on-board image
recording devices:
• Sony UP-D895 MD/D897MD Digital Printer
• Sony UP-D23 MD Digital Color Printer
• HP DeskJet 460 Series Inkjet printer
• HP OfficeJet Pro K5400 Series Inkjet printer
The Vivid S5/Vivid S6 has also been verified for compatibility,
and compliance for connection to a local area network (LAN)
via the rear panel Ethernet connection, provided the LAN
components are IEC/EN 60950 compliant.
Connection may also be made to a CE Marked and IEC/ EN
60950 compliant modem using one of the serial or USB ports
on the system.
The Vivid S5/Vivid S6 may also be used safely while connected
to devices other than those recommended above if the devices
and their specifications, installation, and interconnection with
the system conform to the requirements of IEC/EN 60601-1-1.
General precautions for installing an alternate on-board device
would include:
• The added device must have appropriate safety standard
conformance and CE Marking.
• The total power consumption of the added devices, which
46Vivid S5/Vivid S6 User Manual
R2424458-100 Rev. 2
Page 65
Safety
connect to Vivid S5/Vivid S6 and are used simultaneously,
must be less than or equal to the rated supply of the
VividS5/VividS6.
• There must be adequate heat dissipation and ventilation to
prevent overheating of the device.
• There must be adequate mechanical mounting of the
device and stability of the combination.
• Risk and leakage current of the combination must comply
with IEC/EN 60601-1.
• Electromagnetic emissions and immunity of the
combination must conform to IEC/EN 60601-1-2.
General precautions for installing an alternate off-board,
remote device or a network would include:
• The added device(s) must have appropriate safety
standard conformance and CE Marking.
• The added device(s) must be used for their intended
purpose having a compatible interface.
• Signal or mains isolation devices and additional protective
earth may be needed to assure compliance with IEC/EN
60601-1-1.
The connection of equipment or transmission networks other
than as specified in the user instructions can result in an electric
CAUTION
Vivid S5/Vivid S6 User Manual47
R2424458-100 Rev. 2
shock hazard or equipment malfunction. Substitute or alternate
equipment and connections requires verification of compatibility
and conformity to IEC/EN 60601-1-1 by the installer. Equipment
modifications and possible resulting malfunctions and
electromagnetic interference are the responsibility of the owner.
Page 66
Safety
Patient Environmental Devices
1. Rear Panel
2. Peripheral Devices Power
3. Ethernet network
4. Rear USB Ports
5. Audio Output
6. DVI Video output
7. Power Line Circuit Breaker
8. Ground Line
9. Power Line (AC~)
10. DVD Drive
11. Hard-Copy Printer
(numbers 12 through19 not used in figure)
20. LCD
21. LCD Folding Lever
22. Up/Down & Left/right swivel release levers
23. Front USB Ports
24. RS Probe Connectors
25. LOGIQ Probe Connector (Vivid S6 only)
26. Non-imaging Pencil Probe Connector
27. ECG connector
Figure 1-2: Patient Environmental Devices
48Vivid S5/Vivid S6 User Manual
R2424458-100 Rev. 2
Page 67
CAUTION
Safety
Acceptable devices
The Patient Environmental devices shown on the previous
page are specified to be suitable for use within the PATIENT
ENVIRONMENT.
DO NOT connect any probes or accessories without approval by
GE within the PATIENT ENVIRONMENT.
See "Peripheral Update for EC countries" on page 46 for more
information.
Unapproved devices
DO NOT use unapproved devices.
CAUTION
CAUTION
If devices are connected without the approval of GE, the warranty
will be INVALID.
Any device connected to the Vivid S5/Vivid S6 must conform to
one or more of the requirements listed below:
• IEC standard or equivalent standards appropriate to devices.
• The devices shall be connected to PROTECTIVE EARTH
(GROUND).
Accessories, options, and supplies
Unsafe operation or malfunction may result. Use only the
accessories, options and supplies approved or recommended in
these instructions for use.
Vivid S5/Vivid S6 User Manual49
R2424458-100 Rev. 2
Page 68
Safety
Environmental protection
System and battery disposal
the waste of electrical and electronic equipment must not be
disposed as unsorted municipal waste and must be collected
separately. Please contact an authorized representative of the
manufacturer for information concerning the decommissioning
of your equipment.
This product consists of devices that may contain mercury,
which must be recycled or disposed of in accordance with local,
state, or country laws. (Within this system, the backlight lamps
in the monitor display contain mercury.)
• Connecting and disconnecting probes ....................................... 87
• Adjusting the Display Monitor ..................................................... 90
• Starting an examination ............................................................... .... 95
• Creating a new Patient record or starting an examination from an
existing patient record ................................................................ 95
• Selecting a Probe and an Application ...................................... 100
Vivid S5/Vivid S6 User Manual51
R2424458-100 Rev. 2
Page 70
Getting started
Introduction
Only qualified physicians or ultrasound sonographers should
perform scans of patients for medical diagnostic reasons.
Request training, if needed.
Ensure that unauthorized personnel do not tamper with the
unit.
Service representatives authorized by GE Medical Systems will
unpack and install the unit. Do not attempt to install the unit
alone.
Never set liquids on the unit in order to avoid spillage into the
unit or the control panel. Maintain a clean environment. Turn off
the circuit breaker before cleaning the unit. Refer to "System
Care and Maintenance" on page 642 for cleaning instructions.
To carry out regular preventative maintenance refer to
Chapter 14, "User maintenance" on page 641.
All the warnings in "Important safety considerations" on
page 18, should be read and understood before operating the
WARNING
unit.
52Vivid S5/Vivid S6 User Manual
R2424458-100 Rev. 2
Page 71
Preparing the unit for use
The Vivid S5/Vivid S6 ultrasound unit must operate within the
proper environment and in accordance with the requirements
described in this section. Before using the system, ensure that
the requirements are met.
Site requirements
Optimal operation of the unit can be obtained by implementing
the following requirements:
Power requirements
The Vivid S5/Vivid S6 ultrasound unit uses a separate power
outlet for 100 – 120 VAC or 240 VAC, 50–60 Hz.
Operating the unit with the wrong voltage range causes
damages, voiding the factory warranty.
WARNING
Getting started
Operating Environment
Ensure that there is sufficient air flow around the
Vivid S5/Vivid S6 ultrasound unit when installed or operated.
Environmental requirements
The Vivid S5/Vivid S6 ultrasound unit requires constant
maintenance of its operational environment. Different
temperature and humidity requirements are specified for
operation, storage and transportation.
Table 2-1: Vivid S5/Vivid S6 temperature and humidity requirements
RequirementTemperatureHumidityAir Pressure
Operational 10–40 °C30–85%700–1060 hPa
Storage-10–60 °C30–70%700–1060 hPa
Transport-10–60 °C30–70%700–1060 hPa
Vivid S5/Vivid S6 User Manual53
R2424458-100 Rev. 2
Page 72
Getting started
Electromagnetic interferences
The
Vivid S5/Vivid S6
ultrasound unit is
approved for use in
hospitals, clinics
and other environmentally qualified
facilities, in terms of
the prevention of radio wave interference. Operation of
the unit in an inappropriate environment can cause
electronic interference to radios and
television sets situated near the medical equipment.
Ensure that the unit is protected from electromagnetic
interferences as follows:
• Operate the unit at least 4.5 meters (fifteen feet) away from
equipment that emits strong electromagnetic radiation.
• Shield the unit when operating it in the vicinity of radio
broadcasting equipment, if necessary.
Connecting the unit
A GE-qualified person should perform the initial system
installation.
Connecting the Vivid S5/Vivid S6 ultrasound unit involves
preliminary checks of the power adaptor unit and cord, voltage
level and compliance with electrical safety requirements.
Use only power supply cords, cables and plugs provided by or
designated by GE Medical Systems.
Ensure that the power cord and plug are intact and that the
power plug is the proper hospital-grade type (where required).
The unit should be connected to a fixed power socket which
has the protective grounding connector. Never use an
extension cord or adapter plug.
Failure to provide an adequate earth circuit can cause electrical
shock, resulting in serious injury.
WARNING
Connection of additional protective earth conductors or
potential equalization conductors is not necessary in most cases
WARNING
and is only recommended for situations involving multiple
equipment in a high-risk patient environment to provide
assurance that all equipment is at the same potential and
operates within acceptable leakage current limits. An example of
a high-risk patient would be a special procedure where the
patient has an accessible conductive path to the heart such as
exposed cardiac pacing leads
54Vivid S5/Vivid S6 User Manual
R2424458-100 Rev. 2
Page 73
Getting started
Voltage level check
Check the label near the circuit breaker on the rear side of the
system (Figure 2-1 or Figure 2-2).
Figure 2-1: Vivid S5 rating labels
WARNING
Figure 2-2: Vivid S6 rating labels
Check the voltage range indicated on the label:
• 100–120 VAC, 50/60 Hz, 5 A
Or
• 230 VAC, 50/60 Hz, 2.5 A
If the mains supply is not within the specified range, do not
connect the unit to the power source. Contact the dealer to have
the unit adjusted to the specific mains supply.
Figure 2-3: Universal label
Vivid S5/Vivid S6 User Manual55
R2424458-100 Rev. 2
Page 74
Getting started
WARNING
Connecting to the electrical outlet
POWER OUTAGE MAY OCCUR. The Vivid S5/Vivid S6 requires a
dedicated single branch circuit. To avoid circuit overload and
possible loss of critical care equipment, make sure you DO NOT
have other equipment operating on the same circuit.
The unit’s power must be supplied from a separate, properly
rated outlet to avoid risk of fire. Refer to "Power requirements"
on page 53 for rating information.
The power cord should not, under any circumstances, be altered
to a configuration rated less than that specified for the current.
Do not use an extension cord or adapter plug.
1.Ensure that the wall outlet is of appropriate type, and that
the power switch is turned off.
2.Uncoil the power cable, allowing sufficient slack so that the
unit can be moved slightly.
3.Attach the power plug to the system and secure it in place
by using the retaining clamp.
4.Secure the power plug in the wall outlet.
a. Retaining clamp for power plug
Figure 2-4: Power plug
Ensure that the retaining clamp for the power plug is fixed firmly.
CAUTION
56Vivid S5/Vivid S6 User Manual
R2424458-100 Rev. 2
Page 75
CAUTION
WARNING
Getting started
Use caution to ensure that the power cable does not disconnect
during system use. If the system is accidentally unplugged, data
may be lost.
To avoid risk of fire, the system power must be supplied from a
separate, properly rated outlet.
Under no circumstances should the AC power plug be altered,
changed, or adapted to a configuration rated less than specified.
Never use an extension cord or adapter plug.
To help assure grounding reliability, connect to a "hospital
grade" or "hospital only" grounded power outlet.
The battery may
only be disassembled and re-assembled by an
authorized
field-service engineer.
1. 100-120 VAC, 1200 VA Plug and Outlet Configuration
2. 220-240 VAC, 1200 VA Plug and Outlet Configuration
Figure 2-5: Example Plug and Outlet Configurations
Battery
The System contains an internal battery (optional), which is not
user-accessible. The internal battery acts as a limited
alternative power source, which becomes active when the AC
power fails or AC cable is abruptly pulled out.
Display turns off and no scanning is possible while in battery
mode. For further information about use of the battery see
section "Battery Power Mode" on page 63.
Battery technology
The lithium ion technology used in the system’s battery is
significantly less hazardous to the environment than the lithium
metal technology used in some other batteries (such as watch
batteries). Used batteries should not be placed with common
household waste products. Contact local authorities for the
location of a chemical waste collection program nearest you.
Vivid S5/Vivid S6 User Manual57
R2424458-100 Rev. 2
Page 76
Getting started
Viewing Current Battery Status
When the system is running, a status icon is displayed in the
system Status bar to indicate the current battery status.
Table 2-2: Battery status icons
IconStatus Description
AC Power is ON; no battery present
AC Power is ON; battery is fully charged (80%-100%)
AC Power is ON; battery is partially charged (40%-80%)
AC Power is ON; battery is almost empty (10%-40%)
Note: The % values mentioned above may fluctuate by up to +/- 3 %
points.
View Detailed Battery Status
Refer to page 575
for further information on peripherals.
CAUTION
In order to view further details about the battery status, click on
the battery-status icon. A more detailed status description
appears.
A special message may appear, suggestion to replace the
battery soon. This message may appear when the battery has
aged to such an extent that even after prolonged charging it will
not hold enough charge.
Peripherals/Accessory connection
The external Peripherals / accessories connectors are situated
on the rear side of the unit See Figure 2-6.
Peripherals/Accessory Connector Panel
Vivid S5/Vivid S6 peripherals and accessories can be properly
connected using the rear connector panel.
Use only approved peripherals, accessories or probes.
DO NOT connect any accessories or probes without approval by GE
58Vivid S5/Vivid S6 User Manual
R2424458-100 Rev. 2
Page 77
Getting started
1. Two isolated AC sockets for peripherals (Max. 250VA total)
2. LAN 10/100 Base-TX Ethernet network connector
3. Two interchangeable USB ports (digital printers or other peripherals)
4. Audio output
5. DVI video output connection to external display
Figure 2-6: Peripherals/Accessory Connector Panel
Accessory equipment connected to the analogue and digital
interfaces must be certified according to the respective IEC
WARNING
standards (e.g. IEC 60950 for data processing equipment and
IEC 60601-1 (1988) for medical equipment). Any person
connecting additional equipment to the signal input part or
output part is configuring the medical system, and is therefore
responsible that the system complies with the requirements of
the system standard IEC 60601-1-1 (2000). If in doubt, consult the
technical service department or your local representative.
Do not touch the conducting parts of the USB or Ethernet cables
when connecting equipment to the unit.
Vivid S5/Vivid S6 User Manual59
R2424458-100 Rev. 2
Page 78
Getting started
The connection of equipment or transmission networks other
than as specified in these instructions can result in electric
CAUTION
Table 2-3: Contents of the Rear Panel
SocketSignal typeDevice typeNote
shock hazard. Alternate connections will require verification of
compatibility and conformity to IEC/EN 60601-1-1 by the installer.
DVI-I OutDVI-I output high
resolution video
USBUniversal serial
bus x2
Ethernet10/100 Base-TX
Ethernet IEEE
8023
Network device
External monitor
USB Cable /
Device
Network device
60Vivid S5/Vivid S6 User Manual
R2424458-100 Rev. 2
Page 79
Getting started
Switching On/Off
To switch on the unit:
1.Switch on the circuit breaker on the rear of the unit (see
Figure 2-6).
2.Press (on/off button) on the top left of the control panel
(see Figure 2-11).
After initialization the default scanning screen (2D mode) is
displayed, the active probe being the one connected to the
left most connector socket.
When turning ON
a system while system is in standby, it
takes a few seconds
before it responds.
Do not push the
on/off button again
during this period
(A second push will
initiate a full shut-
down).
By factory Login
default, the Operator ID is USR and
there is no need to
enter a password
LEDs
There are two LEDs above the On/Off switch that indicate the
status of the system.
Indicates power status.
After pressing the On/Standby switch, the system
power is ON and this LED is lit.
Color: Green
Indicates battery status.
When the battery is charged, the LED is green.
When the battery power is low, the LED is orange.
Color: Green and Orange
Password Protection
Log In
During the turn-on process the system may require the user to
enter a password in order to operate it. Personal IDs and
associated passwords can be configured in the
Vivid S5/Vivid S6. See "Presets and System setup" on
page 583 for more information.
If IDs and passwords have been entered and "Use Auto Logon"
is Off (see "Users" on page 637 for details), the Operator Login
window appears, requesting for an ID and password when
Power up sequence is completed, or when it is required.
Vivid S5/Vivid S6 User Manual61
R2424458-100 Rev. 2
Page 80
Getting started
1. Operator: Select the relevant Operator name
2. Password: Enter the Operator’s password
3. Emergency: Data stored for the duration of the current examination only
4. Log on: Select type of Log on (for example, Standard logon)
5. Cancel: Cancel Log on
Figure 2-7: Operator Login Window
Switching off the unit
When the Vivid S5/Vivid S6 is switched off, the system
performs an automatic shutdown sequence. The unit can be
switched off into one of three states.
• Full shutdown: the entire system is shut down. Full
shutdown is recommended if the system is not intended to
be used for a whole day or longer. It is recommended to
perform a full shutdown at least once a week. This will
prevent extreme memory fragmentation, which might cause
some computer slowdown.
• Standby mode: most of the system is powered down, but
a certain portion of the unit remains energized. The standby
mode allows a shorter reboot time when the system is used
on a daily basis or moved from one place to another.
• Battery mode: When AC power is interrupted, the screen
and keyboard are powered down. The rest of the system is
still powered ON. The fans are heard. When AC power is
resumed, the system will turn ON in less than 10 seconds
(see "Battery Power Mode" on page 63).
62Vivid S5/Vivid S6 User Manual
R2424458-100 Rev. 2
Page 81
After switching off
the system, wait at
least ten seconds before turning it on
again.
Getting started
Full shutdown
1.Press (on/off button) on the top left of the control panel.
The Exit dialogue window is displayed.
Figure 2-8: The Exit dialogue window
In case of total lockup of the system,
hold the on/off button down a few seconds to turn the
system off.
2.Select Shutdown.
The shutdown process takes a few seconds and is
completed when the control panel illumination is turned off.
Booting up from Full Shut-down mode may take about 60
seconds.
Standby mode
1.Press (on/off button) on the top left of the control panel.
The Exit dialogue window is displayed (Figure 2-8).
2.Select Standby.
The system enters Standby mode.
Booting up from Standby mode may take about 20
seconds.
Note: Whenever the system is fully shut down or put in
standby mode, the system will automatically perform "End
Exam" to save all data and images of the current patient
into the archiving system.
Turning on the system at the new location
1.Plug the AC cable into a proper power outlet.
2.Press (on/off button) (Figure 2-11).
Battery Power Mode
When the system is operating normally, if the AC power is
interrupted or the power cable is removed from the wall outlet
the screen and keyboard will turn off, but the rest of the system
Vivid S5/Vivid S6 User Manual63
R2424458-100 Rev. 2
Page 82
Getting started
will continue to operate in "Battery Mode", powered by the
internal battery.
While in battery mode, if the AC power is restored within 10
minutes, the system will turn back on within less than 10
seconds.
While in Battery-power mode the system will maintain the
current patient active and not perform "End-Exam". As a result,
when power is restored, the current patient's details and stored
images are readily available on display.
If the AC power is not restored within 10 minutes, the system
will shut-down into Standby mode.
If AC is not restored, the system can remain in Standby mode
for approximately 1 hour using the internal battery power
(assuming battery is fully charged and relatively new). If left for
more than about 1 hour while in standby-mode without AC
power, the system will switch from Standby in to full Shut-down
mode.
Moving and transporting the unit
Wheels
The front wheels of the unit are controlled by the break-pedals
situated above each wheel.
The rear wheels of the unit are controlled by the swivel lock /
release pedals situated above each wheel (see item 10 in
Figure 2-11).
Examine the wheels frequently for defects to avoid breaking or
jamming.
Moving the unit
To prepare the unit to be moved
1.Turn system Off to Full shut-down or Standby mode.
2.Remove the AC plug from the power outlet.
3.Wrap the AC power cord to ensure cord is not hanging in
the wheel area or beyond the sides of the system.
64Vivid S5/Vivid S6 User Manual
R2424458-100 Rev. 2
Page 83
CAUTION
Getting started
4.Pull the left handle and swivel the keyboard so it will lock in
its central position (see page 93).
5.Disconnect all external cables linking the unit to any
off-board peripheral devices and network. (Note the marks
on each cable to reconnect them later).
6.Place all probes securely in proper probe holders. Ensure
that the probe cables do not protrude from the side of the
unit or interfere with the wheels, or foot area. Hang the
cords on respective hooks provided, ensuring cords are
secure and out of lower portion of the system and do not
extend beyond the sides of system.
7.Ensure that no loose items are left on the unit.
8.If intending to travel over bumpy surface or if more visibility
is required during transportation, fold the LCD screen
forward to a horizontal position (see "Position Adjustment"
on page 92).
Before folding the LCD monitor, please make sure that the gel
bottle is placed with the dispenser tip down into gel holder or
moved sideways in order not touch the monitor's faceplate.
WARNING
9.Unlock the front-wheel’s brakes.
10. Lock the rear wheel brakes (Swivel Brake) for long travel or
free it for short distance traveling where maneuvering is
needed.
Once all cables and cords are wrapped and secure out of
the lower portion of the system, not extending beyond
sides of system, you are ready for transport.
To ensure safety while moving the unit
1.Ensure that the keyboard console is in central-locked
position.
Ensure that the hands of the patient are away from the console
when moving the system or keyboard console.
2.Proceed cautiously when crossing door or elevator
thresholds. Grasp the front handle grips and push or pull or
use the rear handle bar for pushing the system. Do not
attempt to move the unit using cables or probe connectors.
Vivid S5/Vivid S6 User Manual65
R2424458-100 Rev. 2
Page 84
Getting started
CAUTION
CAUTION
WARNING
Take extra care while moving the unit on inclines. Do not
hang or lean body weight on the handles.
The rear handle should only be used for pushing the system, not
for pulling, The system might become unstable when hitting an
obstacle while pulled with the rear handle.
3.Ensure that the unit does not strike the walls or door
frames.
4.Ensure the pathway is clear, and probe cables are secure
not to catch onto anything in the path (including wheels and
feet).
5.Move the unit slowly and carefully.
Avoid ramps that are steeper than 10 degrees.
Probe cables may get caught on external devices, such as doors,
medical devices in the transporting path, wheels, or feet. Please
ensure cords are wrapped properly, not extended beyond sides
of system and out of the way for portables.
Transporting the unit
Take extra care when transporting the unit by vehicle. In
addition to the moving precautions listed on page 64, follow the
procedure described below.
1.Disconnect all probes and secure them in their boxes.
2.Lower the system's keyboard to its minimum height.
3.Park the vehicle on a level surface for loading and
unloading.
4.Secure the unit while it is on the lift, to prevent rolling. Do
not attempt to hold it in place by hand. Cushion the unit and
strap the lower part so that it does not break loose.
5.Ensure that the unit is secured inside the vehicle. Secure it
with straps to prevent movement while in transit.
6.Drive cautiously to prevent vibration damage.
66Vivid S5/Vivid S6 User Manual
R2424458-100 Rev. 2
Page 85
Getting started
Reinstalling at a new location
1.When the unit is in place at a new location, lock the wheel
brakes (see "Moving and transporting the unit" on page 64).
2.Follow the installation procedure described on "Connecting
the unit" on page 54.
Preparing Vivid S5/Vivid S6 for scanning
1.Lock front Wheel brakes
2.Un-Wrap the AC power cord and plug into power source.
3.Lift the LCD screen to the fully open position while pulling
out the lock. Secure the Monitor at full upright position.
4.Turn the system on.
5.Connect external cables if needed (LAN, USB, etc.).
6.Probe being used to scan should be ready, with cables
clear of foot area and wheels by wrapping un-used portion
of the cord on respective hooks provided.
7.Adjust keyboard's height and swivel (see page 93).
Note: Ensure that the probe cables do not protrude from the
unit or interfere with the wheels, or foot area. The Hanging
cords may cause damage to system if lodged or hooked onto
nearby items. By ensuring cords are secure and out of lower
portion of the system, this may provide a more secure
atmosphere and avoid unnecessary damage.
Avoiding possible hazards
• Avoid dragging cables on the floor. In cases where the
wheel rolls over cable or person pushing the system steps
on cables, it may generate severe force to break the cable
or the connector.
• Avoid hanging cables beyond sides of the system. When
system is moved, these cables may be caught in
door-knobs or other medical equipment and apply severe
forces that may break probe cables or connectors.
• Check that the Gel bottle is not interfering with the folding
screen. The gel bottle needs to be placed with dispenser tip
down into gel holder or moved sideways in order not to
push on the LCD frame (see Figure 2-9).
Vivid S5/Vivid S6 User Manual67
R2424458-100 Rev. 2
Page 86
Getting started
Proper Gel bottle storageImproper Gel bottle storage
Figure 2-9: Proper Gel bottle storage
Unit acclimation time
Following transport the unit may be very cold or hot. Allow the
unit to acclimate before being switched on. Acclimation will
take one hour for each 2.5
temperature is below 10
o
C
o
F
Hours43210023
o
C
o
F
Hours45678910
02.557.510354042.5
3236.54145.55095104108.5
4547.55052.55557.560
113117.5122126.5131135.5140
o
C increment when the unit’s
o
C or above 40oC.
68Vivid S5/Vivid S6 User Manual
R2424458-100 Rev. 2
Page 87
System description
System overview
Getting started
Figure 2-10: Vivid S5/Vivid S6 System Overview
Note: Above figure is Vivid S6.
Vivid S5/Vivid S6 User Manual69
R2424458-100 Rev. 2
Page 88
Getting started
1.LCD Display Monitor: Swivels to the left and right, tilts
backwards and forwards.
2.Display Monitor Spring-lock Release: Pull knob and fold
screen forward to a horizontal position. To unfold the
screen pull the same knob and lift the screen up.
3.Accessory tray (removable)
4.Rear Handle
5.Probe and gel holders: Removable for cleaning. Situated
on either side of the Control panel.
6.Up/down "Flex-Fit" arm
7.DVD Drive
8.Black and white video printer
9.Fans Service Cover
10. Rear Wheels: Swivel wheels. May be swivel-locked
11. Foot Rest (removable)
12. Front Wheels: Each wheel may be braked.
13. Control panel: Contains all the buttons, softkeys and the
alphanumeric keyboard used to operate the system.
14. USB Port
15. Probe cable hooks: Located on each side in the front and
rear of system
16. Front Handle
17. Console swivel operating handle: Situated on the right
side under the control panel
Console lifting operating handle: Situated on the left
side under the control panel
18. Probe ports: Three RS-type Probe sockets
19. Pencil probe socket
20. ECG connector socket: Also supporting the external ECG
input
21. Large probe socket: (Available on Vivid S6 only)
Supporting the TEE probes only
22. RS-Socket locking latch
70Vivid S5/Vivid S6 User Manual
R2424458-100 Rev. 2
Page 89
Getting started
Control panel
The following pictures illustrate the layout of the
Vivid S5/Vivid S6 control panel. The buttons and controls are
grouped together for ease of use. A detailed description of the
buttons is provided on the following pages.
1. Power status LEDs and light-sensor
2. On/off button
3. Softmenu keys and rotaries
4. Alphanumeric Keyboard
5. 2D-gain and active-gain rotaries
6. TGC sliders
7. Speakers
8. Softmenu 4-way Rocker
9. Alphanumeric Function keys
10. Speaker volume rotary
11. Scan-mode keys
12. Trackball and related keys
13. Freeze keys
Figure 2-11: Vivid S5/Vivid S6 Control Panel
Note: Above figure is Vivid S6.
Vivid S5/Vivid S6 User Manual71
R2424458-100 Rev. 2
Page 90
Getting started
Key illumination
The keys on the control panel are illuminated according to their
availability:
• Illumination in green: the key function is currently active.
• Illumination in white: the key function is available (but not
active) in the current state of the scanner.
• No illumination: The key is not available in the current
state of the scanner.
Power On/Off key
KeyDescription
Turns the unit ON and OFF.
Sets the unit to Standby.
Navigation keys
The following buttons on the top left of the control panel are
used for navigating different screens and packages on the
scanner. They are related to either pre or post-examination
operations. Each of these operations are described in more
detail in the following chapters.
KeyDescription
Displays the Archiving opening page. Enables
the user to perform the following functions:
• Create a new patient record
• Edit the current patient’s information
• Browse the Patient List to search for patient
records
• End the current examination
For further information, refer to page 95.
Displays the Select Probe and Application dialog
box that enables the users to select the desired
probe and application preset for the current
examination. For information about selecting
probes, refer to page 100 and page 535.
72Vivid S5/Vivid S6 User Manual
R2424458-100 Rev. 2
Page 91
KeyDescription
Enters Stress Echo mode (optional). The
Protocol screen is displayed showing the default
stress protocol for the current probe.
Displays the Measurement worksheet where the
user may edit or delete measurements, change
averaging etc. Refer to page 353 for details on
how to operate the worksheet.
Brings the scanner into the Image review mode,
that enables the user to select images from the
clipboard for analysis, activate the image
browser or enter the Image Review screen
where bigger previews of the images are shown
for image selection. Refer to page 426 for details
on the review of images.
Displays the examination report.
Getting started
Vivid S5/Vivid S6 User Manual73
R2424458-100 Rev. 2
Page 92
Getting started
NoteKeyDescription
Scan Mode Selection keys
The following keys are used to select the required scan mode,
and to select additional tools that enhance the application’s
capabilities. Refer to page 137 for detailed information about
scanning.
Note: The TVI button is blank on Vivid S5.
Displays the 2D live acquisition mode that is the
default scanning screen for the unit. For further
information on 2D scanning, refer to page 140.
M-Mode can be
added from a
2D scan also in
replay.
CW mode is not
available on all
scanning
probes.
Displays the M-Mode examination screen and
enables M-Mode functions. Used for viewing motion
patterns. For further information, refer to page 150.
Displays the examination screen in Color Flow
Mapping mode. Used to display color-coded blood
flow information. For further information, refer to
page 157.
Displays the examination screen in Pulsed Wave
Doppler mode. Used for displaying the Doppler
spectrum of blood flow at a selected part of the
anatomy. For further information, refer to page 164.
Displays the examination screen in Continuous
Wave Doppler mode. Allows examination of blood
flow data all along the Doppler CW cursor. For
further information, refer to page 164.
Displays the tissue velocity overlay on 2D and
M-Mode scans. If TVI is on, the Doppler modes
(PW/CW) will also be optimized for tissue velocity.
For further information, refer to page 169.
Note: TVI is optional on Vivid S5. The TVI button is
always installed even if option is not installed.
74Vivid S5/Vivid S6 User Manual
R2424458-100 Rev. 2
Page 93
NoteKeyDescription
Depending on the options installed on the scanner,
this key will bring up the softkey menu for selection
of additional optional scanning modes, such as
AMM, TVI, DTI, TSI, SRI, SI, etc.
The Quantitative Analysis option will also appear (if
installed) as a softkey button named Q Analysis.
Note: Some functions are available only on
Vivid S6.
Basic Mode Parameter Adjustment Controls
The following controls are used to modify and adjust the unit’s
display to best suit the user’s requirements, such as color, gain,
zoom and image depth, according to the mode being operated
by the user.
ControlsDescription
Gain rotaryControls the total gain of the gray scale images
in 2D Mode.
ATO (Automatic tissue optimization) - while in 2D
press the rotary knob to optimize 2D image
automatically.
Getting started
Active mode
rotary
TGCSix sliding keys that compensate for
Vivid S5/Vivid S6 User Manual75
R2424458-100 Rev. 2
Controls the total gain of other activated modes,
such as, M-Mode, Color, PW, or CW Doppler
modes.
In combined mode, switches between the mode
specific assignable controls without having to
exit each mode in order to control it.
depth-related attenuation in an image. The upper
slider corresponds to the smallest depth.
Page 94
Getting started
ControlsDescription
DepthControls the displayed depth of tissue scan. Has
no effect in replay.
Toggles the cursor display on/off in 2D scanning
mode.
Changes the Doppler unltrasound beam angle
on linear probes. The steering angles are fixed
for each linear probe. This key has no effect with
sector imaging probes.
ZoomControls image magnification. Press to activate
"HR-Zoom" mode. Rotate clockwise to activate
"Display-zoom" mode and increase zoom factor.
Rotate counterclockwise to decrease and turn off
zoom (see "Zoom" on page 120).
AutoWhile in 2D it activates ATO - Automatic Tissue
Optimization to optimize 2D image automatically
(see page 144).
While in CW or PW Doppler it activates ASO Automatic Spectrum Optimization to optimize the
Doppler spectrum.
FlexThis button can be configured to function as any
button on the keyboard (see page 590).
Note: with the exception of alphanumeric and
soft-menu sections.
76Vivid S5/Vivid S6 User Manual
R2424458-100 Rev. 2
Page 95
Getting started
Freeze keys
The freeze keys are used to freeze images and cine loops in all
modes for on-line analysis.
KeyDescription
Stops or restarts all data acquisition. When scan
is frozen, the Trackball can be used to scroll
through the cine loop.
Activates or freezes 2D mode. In simultaneous
mode, pressing
deactivate the 2D image, leaving the other mode
display unchanged. In freeze mode, stops/starts
the cineloop.
Display Format Key
2D FREEZE will activate or
KeyDescription
Toggles the display priority of 2D-Mode or
Doppler/M-Mode and top/bottom or side by side
display when working in combined mode.
Enables multiple image display windows in
which two or four images can be viewed
simultaneously. When reducing the number of
images, the active window will always been kept.
Toggles the active window in multiple imaging
mode. The active window is indicated by a
highlighted frame. All imaging controls are
normally applied only to the active window.
Same arrows are used while reviewing images
from clipboard, the active frame may be stepped
forward or backwards to review the next or
previous image.
Vivid S5/Vivid S6 User Manual77
R2424458-100 Rev. 2
Page 96
Getting started
Store Button
KeyDescription
StoreStores the currently active imaging window to
disk. The stored information depends on the
configuration of the current application.
Stored images are shown on the clipboard.
Measurement controls
The following keys are used to take measurements and
perform calculations.
KeyDescription
Activates the Measurement & Analysis (M&A)
calculation program. This program is context
sensitive and will display relevant measurements
to the current mode and application. See
page 243 for further details on M&A.
Activates measurement tools (unassigned
measurement). See page 243 for further details
on M&A.
Print and Record Control
KeyDescription
RecordWhen the DVR option is installed and DVR
media has been mounted, the Record button will
activate or pause the real-time recording of the
displayed image (See page 578 for details)
Note: This button is available even if the DVR
option is not installed.
P1Prints the current imaging screen content to a
selected (configurable) printer. For more
information about printing (see page 577). The
PRINT key can also be configured for alternative
storing of images (see page 623).
78Vivid S5/Vivid S6 User Manual
R2424458-100 Rev. 2
Page 97
Getting started
KeyDescription
P2Prints the current imaging screen content to a
printer assigned to a second port (configurable).
The default factory configuration is set to
"Secondary Capture". For more information see
page 623.
Trackball operation
The Trackball area consists of the trackball and five
surrounding keys. Three of these have the very same function
SET function) for ergonomic reasons.
(the
KeyDescription
TrackballUsed for navigation and together with the
surrounding keys, to move, select or activate
objects on the screen.
Trackball
Button
SetDepending on the situation (see Figure 3-3,
Controls the trackball assignments between the
mode-specific options. By pressing
the trackball function will cycle through the
possible assignments, which are indicated in the
lower right corner of the screen (see page 108).
page 109):
• Performs the selected control or highlighted
menu item.
• Toggles between the Trackball functions within
the active group.
The key is duplicated for ergonomic reasons.
TRACKBALL,
Vivid S5/Vivid S6 User Manual79
R2424458-100 Rev. 2
Page 98
Getting started
KeyDescription
Update/MenuIn Freeze, activates menu with additional options
and controls not available from the assignable
keys.
In live mode, toggles between 2D imaging and
live time-motion imaging (Doppler/M-Mode).
Assignable keys and rotary knobs (soft keys)
The functions of the assignable keys and rotaries vary
according to the mode and/or module in which the user is
working.
KeyDescription
Soft RotariesFour assignable rotary knobs, whose
mode-specific functions vary according to the
scan mode and position that is currently active.
These assignable rotaries are used both as
rotary-control knobs and as on/off toggles for
different controls on the menu The assigned
functions are indicated above the rotary on the
LCD display. The mode-specific functions for
these rotary knobs are described in Chapter 4,
"Scanning Modes" on page 137.
Assignable
Buttons
Four assignable buttons, whose mode-specific
functions vary according to the scan mode and
position that is currently active. These
assignable buttons are used as on/off toggles for
different controls on the menu. The assigned
functions are indicated above the button on the
LCD display. The mode-specific functions for
these buttons are described in Chapter 4,
"Scanning Modes" on page 137.
80Vivid S5/Vivid S6 User Manual
R2424458-100 Rev. 2
Page 99
The soft menu rocker
KeyDescription
Getting started
Soft Menu
Rocker
A 4-way rocker used to access mode-specific
menus, select a menu option and adjust
option-related values.
• The vertical arrows are used to select the menu
options.
• The horizontal arrows are used to adjust the
values.
The mode-specific menus are described in
Chapter 4, "Scanning Modes" on page 137.
The Alphanumeric Function keys
KeyDescription
HelpDisplays the on-line version of the user manual.
Config/DiagDisplays the configuration dialog box, allowing
user configuration of various settings on the
scanner.
Diagnostics of the system is activated by
pressing Shift >
CONFIG.
LCDTo activate LCD adjustments controlled by
softkeys.
BiopsyDisplays the biopsy path marker.
BodymarkDisplays the available body marks for the current
application.
Page EraseErases all previously-typed annotations (and
body marks).
Vivid S5/Vivid S6 User Manual81
R2424458-100 Rev. 2
Page 100
Getting started
KeyDescription
PhysioProvides access to controls for ECG trace. The
ECG controls appear on the soft-menu.
ArrowDisplays an arrow that can be used to point at a
specific structure in the image.
TextEnables text annotation to be inserted on the
image. The annotations can be typed or selected
from a (configurable) menu.
PlaybackFor future use
DeleteCan be used to delete text during text
annotation.
List of shortcuts on alphanumeric keyboard
KeyDescription
Alt+EAllows to eject a device like MOD, CD or memory
card.
Alt+LAllows to adjust intensity of keyboard backlight.
Alt+PAllows to view and control printer spooler.
Alt+SAllows to view and control DICOM spooler. (see
"DICOM spooler" on page 479) The DICOM
spooler is used for checking the current job's
status when a job is saved or when the total
spooler status on the right of the Archive
windows displays an error.
Alt+DAllows to comment and save logfiles for
diagnostics (see Chapter 14, "Adding Problem
description" on page 646).
Alt+BAllows to insert a "bookmark" into the system
failure log ("Adding bookmarks" on page 646)
Shift+ConfigAllows to run various diagnostics.
82Vivid S5/Vivid S6 User Manual
R2424458-100 Rev. 2
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.