GE HEALTHCARE
DIRECTION FL091021, REVISION 5 VIVID I N SERVICE MANUAL
Important Precautions
• THIS SERVICE MANUAL IS AVAILABLE IN ENGLISH ONLY.
• IF A CUSTOMER’S SERVICE PROVIDER REQUIRES A LANGUAGE OTHER
THAN ENGLISH, IT IS THE CUSTOMER’S RESPONSIBILITY TO PROVIDE
TRANSLATION SERVICES.
WARNING
(EN)
AVERTISSEMENT
(FR)
• DO NOT ATTEMPT TO SERVICE THE EQUIPMENT UNLESS THIS SERVICE
MANUAL HAS BEEN CONSULTED AND IS UNDERSTOOD.
• FAILURE TO HEED THIS WARNING MAY RESULT IN INJURY TO THE SERVICE
PROVIDER, OPERATOR OR PATIENT FROM ELECTRIC SHOCK, MECHANICAL
OR OTHER HAZARDS.
• CE MANUEL DE MAINTENANCE N’EST DISPONIBLE QU’EN ANGLAIS.
• SI LE TECHNICIEN DU CLIENT A BESOIN DE CE MANUEL DANS UNE AUTRE
LANGUE QUE L’ANGLAIS, C’EST AU CLIENT QU’IL INCOMBE DE LE FAIRE
TRADUIRE.
• NE PAS TENTER D’INTERVENTION SUR LES ÉQUIPEMENTS TANT QUE LE
MANUEL SERVICE N’A PAS ÉTÉ CONSULTÉ ET COMPRIS.
• LE NON-RESPECT DE CET AVERTISSEMENT PEUT ENTRAÎNER CHEZ LE
TECHNICIEN, L’OPÉRATEUR OU LE PATIENT DES BLESSURES DUES À DES
DANGERS ÉLECTRIQUES, MÉCANIQUES OU AUTRES.
WARNUNG
(DE)
• DIESES KUNDENDIENST-HAND BUCH EXISTIERT NUR IN ENGLISCHER
SPRACHE.
• FALLS EIN FREMDER KUNDENDIENST EINE ANDERE SPRACHE BENÖTIGT,
IST ES AUFGABE DES KUNDEN FÜR EINE ENTSPRECHENDE ÜBERSETZUNG
ZU SORGEN.
• VERSUCHEN SIE NICHT, DAS GERÄT ZU REPARIEREN, BEVOR DIESES
KUNDENDIENST-HANDBUCH NICHT ZU RATE GEZOGEN UND VERSTANDEN
WURDE.
• WIRD DIESE WARNUNG NICHT BEACHTET, SO KANN ES ZU VERLETZUNGEN
DES KUNDENDIENSTTECHNIKERS, DES BEDIENERS ODER DES PATIENTEN
DURCH ELEKTRISCHE SCHLÄGE, MECHANISCHE ODER SONSTIGE
GEFAHREN KOMMEN.
- -i
Page 4
GE HEALTHCARE
DIRECTION FL091021, REVISION 5 VIVID I N SERVICE MANUAL
• ESTE MANUAL DE SERVICIO SÓLO EXISTE EN INGLÉS.
• SI ALGÚN PROVEEDOR DE SERVICIOS AJENO A GEHC SOLICITA UN IDIOMA
QUE NO SEA EL INGLÉS, ES RESPONSABILIDAD DEL CLIENTE OFRECER UN
SERVICIO DE TRADUCCIÓN.
AVISO
(ES)
• NO SE DEBERÁ DAR SERVICIO TÉCNICO AL EQUIP O, SIN HABER
CONSULTADO Y COMPRENDIDO ESTE MANUAL DE SERVICIO.
• LA NO OBSERVANCIA DEL PRESENTE AVISO PUEDE DAR LUGAR A QUE EL
PROVEEDOR DE SERVICIOS, EL OPERADOR O EL PACIENTE SUFRAN
LESIONES PROVOCADAS POR CAUSAS ELÉCTRICAS, MECÁNICAS O DE
OTRA NATURALEZA.
• ESTE MANUAL DE ASSISTÊNCIA TÉCNICA SÓ SE ENCONTRA DISPONÍVEL EM
INGLÊS.
• SE QUALQUER OUTRO SERVIÇO DE ASSISTÊNCIA TÉCNICA, QUE NÃO A
GEHC, SOLICITAR ESTES MANUAIS NOUTRO IDIOMA, É DA
ATENÇÃO
(PT)
RESPONSABILIDADE DO CLIENTE FORNECER OS SERVIÇOS DE TRADUÇÃO.
• NÃO TENTE REPARAR O EQUIPAMENTO SEM TER CONSULTADO E
COMPREENDIDO ESTE MANUAL DE ASSISTÊNCIA TÉCNICA.
• O NÃO CUMPRIMENTO DESTE AVISO PODE POR EM PERIGO A SEGURANÇA
DO TÉCNICO, OPERADOR OU PACIENTE DEVIDO A‘ CHOQUES ELÉTRICOS,
MECÂNICOS OU OUTROS.
AVVERTENZA
(IT)
HOIATUS
(ET)
• IL PRESENTE MANUALE DI MANUTENZIONE È DISPONIBILE SOLTANTO IN
INGLESE.
• SE UN ADDETTO ALLA MANUTENZIONE ESTERNO ALLA GEHC RICHIEDE IL
MANUALE IN UNA LINGUA DIVERSA, IL CLIENTE È TENUTO A PROVVEDERE
DIRETTAMENTE ALLA TRADUZIONE.
• SI PROCEDA ALLA MANUTENZIONE DELL’APPARECCHIATURA SOLO DOPO
AVER CONSULTATO IL PRESENTE MANUALE ED AVERNE COMPRESO IL
CONTENUTO.
• NON TENERE CONTO DELLA PRESENTE AVVERTENZA POTREBBE FAR
COMPIERE OPERAZIONI DA CUI DERIVINO LESIONI ALL’ADDETTO ALLA
MANUTENZIONE, ALL’UTILIZZATORE ED AL PAZIENTE PER FOLGORAZIONE
ELETTRICA, PER URTI MECCANICI OD ALTRI RISCHI.
• KÄESOLEV TEENINDUSJUHEND ON SAADAVAL AINULT INGLISE KEELES.
• KUI KLIENDITEENINDUSE OSUTAJA NÕUAB JUHENDIT INGLISE KEELEST
ERINEVAS KEELES, VASTUTAB KLIENT TÕLKETEENUSE OSUTAMISE EEST.
• ÄRGE ÜRITAGE SEADMEID TEENINDADA ENNE EELNEVALT KÄESOLEVA
TEENINDUSJUHENDIGA TUTVUMIST JA SELLEST ARU SAAMIST.
• KÄESOLEVA HOIATUSE EIRAMINE VÕIB PÕHJUSTADA TEENUSEOSUTAJA,
OPERAATORI VÕI PATSIENDI VIGASTAMIST ELEKTRILÖÖGI, MEHAANILISE
VÕI MUU OHU TAGAJÄRJEL.
-ii -
Page 5
GE HEALTHCARE
DIRECTION FL091021, REVISION 5 VIVID I N SERVICE MANUAL
• TÄMÄ HUOLTO-OHJE ON SAATAVILLA VAIN ENGLANNIKSI.
• JOS ASIAKKAAN PALVELUNTARJOAJA VAATII MU UTA KUIN
ENGLANNINKIELISTÄ MATERIAALIA, TARVITTAVAN KÄÄNNÖKSEN
HANKKIMINEN ON ASIAKKAAN VASTUULLA.
VAROITUS
(FI)
• ÄLÄ YRITÄ KORJATA LAITTEISTOA ENNEN KUIN OLET VARMASTI LUKENUT
JA YMMÄRTÄNYT TÄMÄN HUOLTO-OHJEEN.
• MIKÄLI TÄTÄ VAROITUSTA EI NOUDATETA , SEURAUKSENA VOI OLLA
PALVELUNTARJOAJAN, LAITTEISTON KÄYTTÄJÄN TAI POTILAAN
VAHINGOITTUMINEN SÄHKÖISKUN, MEKAANISEN VIAN TAI MUUN
VAARATILANTEEN VUOKSI.
ΠΡΟΚΛΗΘΕΙ ΤΡΑΥΜΑΤΙΣΜΟΣ ΣΤΟ ΑΤΟΜΟ ΠΑΡΟΧΗΣ ΣΕΡΒΙΣ, ΣΤΟ ΧΕΙΡΙΣΤΗ Ή
ΣΤΟΝ ΑΣΘΕΝΗ ΑΠΟ ΗΛΕΚΤΡΟΠΛΗΞΙΑ, ΜΗΧΑΝΙΚΟΥΣ Ή ΑΛΛΟΥΣ ΚΙΝΔΥΝΟΥΣ.
FIGYELMEZTETÉS
(HU)
VIÐVÖRUN
(IS)
• EZEN KARBANTARTÁSI K ÉZIKÖNYV KIZÁRÓLAG ANGOL NYELVEN ÉRHETŐ
EL.
• HA A VEVŐ SZOLGÁLTATÓJA ANGOLTÓL ELTÉRŐ NYELVRE TART IGÉNYT,
AKKOR A VEVŐ FELELŐSSÉGE A FORDÍTÁS ELKÉSZÍTTETÉSE.
• NE PRÓBÁLJA ELKEZDENI HASZNÁLNI A BERENDEZÉST, AMÍG A
KARBANTARTÁSI KÉZIKÖNYVBEN LEÍRTAKAT NEM ÉRTELMEZTÉK.
• EZEN FIGYELMEZTETÉS FIGYELMEN KÍVÜL HAGYÁSA A SZOLGÁLTATÓ,
MŰKÖDTETŐ VAGY A BETEG ÁRAMÜTÉS, MECHANIKAI VAGY EGYÉB
VESZÉLYHELYZET MIATTI SÉRÜLÉSÉT EREDMÉNYEZHETI.
• ÞESSI ÞJÓNUSTUHANDBÓK ER EINGÖNGU FÁANLEG Á ENSKU.
• EF ÞJÓNUSTUAÐILI VIÐSKIPTAMANNS ÞARFNAST ANNARS TUNGUMÁLS EN
ENSKU, ER ÞAÐ Á ÁBYRGÐ VIÐSKIPTAMANNS AÐ ÚTVEGA ÞÝÐINGU.
• REYNIÐ EKKI AÐ ÞJÓNUSTA TÆKIÐ NEMA EFTIR AÐ HAFA SKOÐAÐ OG
SKILIÐ ÞESSA ÞJÓNUSTUHANDBÓK.
• EF EKKI ER FARIÐ AÐ ÞESSARI VIÐVÖRUN GETUR ÞAÐ VALDIÐ MEIÐSLUM
ÞJÓNUSTUVEITANDA, STJÓRNANDA EÐA SJÚKLINGS VEGNA RAFLOSTS,
VÉLRÆNNAR EÐA ANNARRAR HÆTTU.
- -iii
Page 6
GE HEALTHCARE
DIRECTION FL091021, REVISION 5 VIVID I N SERVICE MANUAL
• TENTO SERVISNÍ NÁVOD EXISTUJE POUZE V ANGLICKÉM JAZYCE.
•VPŘÍPADĚ, ŽE POSKYTOVATEL SLUŽEB ZÁKAZNÍKŮM POTŘEBUJE NÁVOD
V JINÉM JAZYCE, JE ZAJIŠTĚNÍ PŘEKLADU DO ODPOVÍDAJÍCÍHO JAZYKA
ÚKOLEM ZÁKAZNÍKA.
• NEPROVÁDĚJTE ÚDRŽBU TOHOTO ZAŘÍZENÍ, ANIŽ BYSTE SI PŘEČETLI
VÝSTRAHA
(CS)
TENTO SERVISNÍ NÁVOD A POCHOPILI JEHO OBSAH.
•VPŘÍPADĚ NEDODRŽOVÁNÍ TÉTO VÝSTRAHY MŮŽE DOJÍT ÚRAZU
ELEKTRICKÁM PROUDEM PRACOVNÍKA POSKYTOVATELE SLUŽEB,
OBSLUŽNÉHO PERSONÁLU NEBO PACIENTŮ VLIVEM ELEKTRICKÉHOP
PROUDU, RESPEKTIVE VLIVEM K RIZIKU MECHANICKÉHO POŠKOZENÍ NEBO
JINÉMU RIZIKU.
• DENNE SERVICEMANUAL FIND ES KUN PÅ ENGELSK.
• HVIS EN KUNDES TEKNIKER HAR BRUG FOR ET ANDET SPROG END
ENGELSK, ER DET KUNDENS ANSVAR AT SØRGE FOR OVERSÆTTELSE.
ADVARSEL
(DA)
• FORSØG IKKE AT SERVICERE UDSTYRET MEDMINDRE
DENNE SERVICEMANUAL ER BLEVET LÆST OG FORSTÅET.
• MANGLENDE OVERHOLDELSE AF DENNE ADVARSEL KAN MEDFØRE SKADE
PÅ GRUND AF ELEKTRISK, MEKANISK ELLER ANDEN FARE FOR
TEKNIKEREN, OPERATØREN ELLER PATIENTEN.
WAARSCHUWING
(NL)
BRĪDINĀJUMS
(LV)
• DEZE ONDERHOUDSHANDLEIDING IS ENKEL IN HET ENGELS
VERKRIJGBAAR.
• ALS HET ONDERHOUDSPERSONEEL EEN ANDERE TAAL VEREIST, DAN IS DE
KLANT VERANTWOORDELIJK VOOR DE VERTALING ERVAN.
• PROBEER DE APPARATUUR NIET TE ONDERHOUDEN VOORDAT DEZE
ONDERHOUDSHANDLEIDING WERD GERAADPLEEGD EN BEGREPEN IS.
• INDIEN DEZE WAARSCHUWING NIET WORDT OPGEVOLGD, ZOU HET
ONDERHOUDSPERSONEEL, DE OPERATOR OF EEN PATIËNT GEWOND
KUNNEN RAKEN ALS GEVOLG VAN EEN ELEKTRISCHE SCHOK,
MECHANISCHE OF ANDERE GEVAREN.
•ŠĪ APKALPES ROKASGRĀMATA IR PIEEJAMA TIKAI ANGĻU VALODĀ.
• JA KLIENTA APKALPES SNIEDZĒJAM NEPIECIEŠAMA INFORMĀCIJA CITĀ
VALODĀ, NEVIS ANGĻU, KLIENTA PIENĀKUMS IR NODROŠINĀT TULKOŠANU.
• NEVEICIET APRĪKOJUMA APKALPI BEZ APKALPES ROKASGRĀMATAS
IZLASĪŠANAS UN SAPRAŠANAS.
•ŠĪ BRĪDINĀJUMA NEIEVĒROŠANA VAR RADĪT ELEKTRISK
ĀS STRĀVAS
TRIECIENA, MEHĀNISKU VAI CITU RISKU IZRAISĪTU TRAUMU APKALPES
SNIEDZĒJAM, OPERATORAM VAI PACIENTAM.
-iv -
Page 7
GE HEALTHCARE
DIRECTION FL091021, REVISION 5 VIVID I N SERVICE MANUAL
• ŠIS EKSPLOATAVIMO VADOVAS YRA IŠLEISTAS TIK ANGLŲ KALBA.
• JEI KLIENTO PASLAUGŲ TEIKĖJUI REIKIA VADOVO KITA KALBA – NE ANGLŲ,
VERTIMU PASIRŪPINTI TURI KLIENTAS.
ĮSPĖJIMAS
(LT)
•NEMĖGINKITE ATLIKTI ĮRANGOS TECHNINĖS PRIEŽIŪROS DARBŲ, NEBENT
VADOVAUTUMĖTĖS ŠIUO EKSPLOATAVIMO VADOVU IR JĮ SUPRASTUMĖTE
• NEPAISANT ŠIO PERSPĖJIMO, PASLAUGŲ TEIKĖJAS, OPERATOR IUS AR
PACIENTAS GALI BŪTI SUŽEISTAS DĖL ELEKTROS SMŪGIO, MECHANINIŲ AR
KITŲ PAVOJŲ.
• DENNE SERVICEHÅNDBOKEN FINNES BARE PÅ ENGELSK.
• HVIS KUNDENS SERVICELEVERANDØR TRENGER ET ANNET SPRÅK, ER DET
KUNDENS ANSVAR Å SØRGE FOR OVERSETTELSE.
ADVARSEL
(NO)
• IKKE FORSØK Å REPARERE UTSTYRET UTEN AT DENNE
SERVICEHÅNDBOKEN ER LEST OG FORSTÅTT.
• MANGLENDE HENSYN TIL DENNE ADVARSELEN KAN FØRE TIL AT
SERVICELEVERANDØREN, OPERATØREN ELLER PASIENTEN SKADES PÅ
GRUNN AV ELEKTRISK STØT, MEKANISKE ELLER ANDRE FARER.
OSTRZEŻENIE
(PL)
ATENŢIE
(RO)
• NINIEJSZY PODRĘCZNIK SERWISOWY DOSTĘPNY JEST JEDYNIE W JĘZYKU
ANGIELSKIM.
•JEŚLI FIRMA ŚWIADCZĄCA KLIENTOWI USłUGI SERWISOWE WYMAGA
UDOSTĘPNIENIA PODRĘCZNIKA W JĘZYKU INNYM NIŻ ANGIELSKI,
OBOWIĄZEK ZAPEWNIENIA STOSOWNEGO TłUMACZENIA SPOCZYWA NA
KLIENCIE.
• NIE PRÓBOWAĆ SERWISOWAĆ NINIEJSZEGO SPRZĘTU BEZ UPRZEDNIEGO
ZAPOZNANIA SI
Ę Z PODRĘCZNIKIEM SERWISOWYM.
• NIEZASTOSOWANIE SIĘ DO TEGO OSTRZEŻENIA MOżE GROZIĆ
OBRAŻENIAMI CIAłA SERWISANTA, OPERATORA LUB PACJENTA W WYNIKU
PORAŻENIA PRĄDEM, URAZU MECHANICZNEGO LUB INNEGO RODZAJU
ZAGROŻEŃ.
• ACEST MANUAL DE SERVICE ESTE DISPONIBIL NUMAI ÎN LIMBA ENGLEZĂ.
• DACĂ UN FURNIZOR DE SERVICII PENTRU CLIENŢI NECESITĂ O ALTĂ LIMBĂ
DECÂT CEA ENGLEZĂ, ESTE DE DATORIA CLIENTULUI SĂ FURNIZEZE O
TRADUCERE.
• NU ÎNCERCAŢI SĂ REPARAŢI ECHIPAMENTUL DECÂT ULTERIOR
CONSULTĂRII ŞI ÎNŢELEGERII ACESTUI MANUAL DE SERVICE.
• IGNORAREA ACESTUI AVERTISMENT A R PUTEA DUCE LA RĂNIREA
DEPANATORULUI, OPERATORULUI SAU PACIENTULUI ÎN URMA
PERICOLELOR DE ELECTROCUTARE, MECANICE SAU DE ALTĂ NATURĂ.
- -v
Page 8
GE HEALTHCARE
DIRECTION FL091021, REVISION 5 VIVID I N SERVICE MANUAL
ПОЛУЧАТ УДАР ЗЛЕКТРИЧЕСКИМ ТОКОМ, МЕХАНИЧЕСКУЮ ТРАВМУ ИЛИ
ДРУГОЕ ПОВРЕЖДЕНИЕ.
• TÁTO SERVISNÁ PRÍRUČKA JE K DISPOZÍCII LEN V ANGLIČTINE.
• AK ZÁKAZNÍKOV POSKYTOVATEĽ SLUŽIEB VYŽADUJE INÝ JAZYK AKO
ANGLIČTINU, POSKYTNUTIE PREKLADATEĽSKÝCH SLUŽIEB JE
ZODPOVEDNOSŤOU ZÁKAZNÍKA.
UPOZORNENIE
(SK)
• NEPOKÚŠAJTE SA VYKONÁVAŤ SERVIS ZARIADENIA SKÔR, AKO SI
NEPREČÍTATE SERVISNÚ PRÍRUČKU A NEPOROZUMIETE JEJ.
• ZANEDBANIE TOHTO UPOZORNENIA MÔŽE VYÚSTIŤ DO ZRANENIA
POSKYTOVATEĽA SLUŽIEB, OBSLUHUJÚCEJ OSOBY ALEBO PACIENTA
ELEKTRICKÝM PRÚDOM, PRÍPADNE DO MECHANICKÉHO ALEBO INÉHO
NEBEZPEČENSTVA.
VARNING
(SV)
DİKKAT
(TR)
• DEN HÄR SERVICEHANDBOKEN FINNS BARA TILLGÄNGLIG PÅ ENGELSKA.
• OM EN KUNDS SERVICETEKNIKER HAR BEHOV AV ETT ANNAT SPRÅK ÄN
ENGELSKA ANSVARAR KUNDEN FÖR ATT TILLHANDAHÅLLA
ÖVERSÄTTNINGSTJÄNSTER.
• FÖRSÖK INTE UTFÖRA SERVICE PÅ UTRUSTNINGEN OM DU INTE HAR LÄST
OCH FÖRSTÅR DEN HÄR SERVICEHANDBOKEN.
• OM DU INTE TAR HÄNSYN TILL DEN HÄR VARNINGEN KAN DET RESULTERA I
SKADOR PÅ SERVICETEKNIKERN, OPERATÖREN ELLER PATIENTEN TILL
FÖLJD AV ELEKTRISKA STÖTAR, MEKANISKA FAROR ELLER ANDRA FAROR.
• BU SERVİS KILAVUZU YALNIZCA İNGİLİZCE OLARAK SAĞLANMIŞTIR.
• BU UYARININ GÖZ ARDI EDİLMESİ, ELEKTRİK ÇARPMASI YA DA MEKANİK
VEYA DİĞER TÜRDEN KAZALAR SONUCUNDA TEKNİSYENİN, OPERATÖRÜN
YA DA HASTANIN YARALANMASINA YOL AÇABİLİR.
-vi -
Page 9
GE HEALTHCARE
DIRECTION FL091021, REVISION 5 VIVID I N SERVICE MANUAL
(JA)
(ZH-CN)
(KO)
- -vii
Page 10
GE HEALTHCARE
DIRECTION FL091021, REVISION 5 VIVID I N SERVICE MANUAL
DAMAGE IN TRANSPORTATION
All packages should be closely examined at time of delivery. If damage is apparent write “Damage In
Shipment” on ALL copies of the freight or express bill BEFORE delivery is accepted or “signed for” by
a GE representative or hospital receiving agent. Whether noted or concealed, damage MUST be
reported to the carrier immediately upon discovery, or in any event, within 14 days after receipt, and the
contents and containers held for inspection by the carrier. A transportation company will not pay a claim
for damage if an inspection is not requested within this 14 day period.
CERTIFIED ELECTRICAL CONTRACTOR STATEMENT - FOR USA ONLY
All electrical Installations that are preliminary to position ing of the equipment at the site prepar ed for the
equipment shall be performed by licensed electrical contractors. Othe r co nnections b etween pieces o f
electrical equipment, calibrations and testing shall be performed by q ualified GE Healthcare personne l.
In performing all electrical work on these products, GE will use its own specially trained field engineers.
All of GE’s electrical work on these products will comply with the requirements of the applicable
electrical codes.
The purchaser of GE equipment shall only utilize qualified personnel (i.e., GE’s field engineers,
personnel of third-party service companies with equivalent training, or licensed electricians) to perform
electrical servicing on the equipment.
OMISSIONS & ERRORS
If there are any omissions, errors or suggestions for improving this documentation, please contact the
GE Healthcare Global Documentation Group with specific information listing the system type, manual
title, part number, revision number, page number and suggestion details. Mail the information to:
Service Documentation, 9900 Innovation Drive (RP-2123), Wauwatosa, WI 53226, USA.
GE Healthcare employees should use the iTrak System to report all documentation errors or omissions.
SERVICE SAFETY CONSIDERATIONS
DANGER
DANGEROUS VOLTAGES, CAPABLE OF CAUSING DEATH, ARE PRESENT IN
THIS EQUIPMENT. USE EXTREME CAUTION WHEN HANDLING, TESTING AND
ADJUSTING.
WARNINGWARNING
Use all Personal Protection Equipment (PPE) such as gloves, safety shoes, safety
glasses, and kneeling pad, to reduce the risk of injury.
For a complete review of all safety requirements, see the Safety Considerations section in Chapter 1.
-viii -
Page 11
GE HEALTHCARE
DIRECTION FL091021, REVISION 5 VIVID I N SERVICE MANUAL
LEGAL NOTES
The contents of this publication may not be copied or duplicated in any form, in whole or in part, without
prior written permission of GE Healthcare.
GE Healthcare may revise this publication from time to time without written notice.
TRADEMARKS
All products and their name brands are trademarks of their respective holders.
GE HEALTHCARE
DIRECTION FL091021, REVISION 5 VIVID I N SERVICE MANUAL
Revision History
RevisionDateReason for change
1MAY 2005Initial Release.
2NOV 2005Revised Spare Parts List update
3FEB. 2006Added new cart and transportation Box
4JAN. 2007Removed all traces of Cart and Transportation Box
5JUL 2009Updated per BT’09 release.
List of Effected Pages (LOEP)
PagesRevisionPagesRevisionPagesRevision
Title Page5 3-1 to 3-1585 8-1 to 8-1085
Warnings i to x5 4-1 to 4-305 9-1 to 9-165
TOC (generated file)5 5-1 to 5-425 10-1 to 10-365
1-1 to 1-205 6-1 to 6-125 Back CoverN/A
2-1 to 2-125 7-1 to 7-205
-x -
Page 13
GE HEALTHCARE DIRECTION FL091021, REVISION 5 VIVID I NSERVICE MANUAL
GE HEALTHCARE
DIRECTION FL091021, REVISION 5 VIVID I N SERVICE MANUAL
Chapter 1
Introduction
Section 1-1
Overview
1-1-1Purpose of Chapter 1
This chapter describes important issues related to safely servicing the Vivid i n portable ultrasound
scanner. The service provider must read and understand all the information presented here before
installing or servicing a unit.
Table 1-1 Contents in Chapter 1
SectionDescriptionPage Number
1-1
1-2
1-3
1-4
1-5
1-6
1-7
1-8
Overview
Important Conventions
Safety Considerations
Product Labels and Icons
EMC, EMI, and ESD
Lockout/Tagout (LOTO) Requirements
Returning/Shipping Probes and Repair Parts
Customer Assistance
1-1
1-4
1-6
1-10
1-16
1-18
1-18
1-19
Chapter 1 - Introduction 1-1
Page 28
GE HEALTHCARE
DIRECTION FL091021, REVISION 5 VIVID I N SERVICE MANUAL
1-1-2Purpose of Service Manual
This manual provides installation and service information for the Vivid i n portab le ultraso und scann er,
and contains the following chapters:
•Chapter 1 - Introduction
Contains a content summary and warnings.
•Chapter 2 - Site Preparations
Contains site requirements for the Vivid i n portable ultrasound scanner.
•Chapter 3 - System Setup
Contains installation procedures and an installation checklist.
•Chapter 4 - General Procedures and Functional Checks
Contains functional checks that are reco m men de d as par t of the installa tio n pr oc ed ur e , or as
required during servicing and periodic maintenance.
•Chapter 5 - Components and Function (Theory)
Contains block diagrams and functional explanations of the electronic circuits.
•Chapter 6 - Service Adjustments
Contains instructions for performing service adjustments to the Vivid i n portable ultrasound
scanner.
•Chapter 7 - Diagnostics/Troubleshooting
Provides instructions for setting up and running diagnostic, troubleshooting and other related
routines for the Vivid i n portable ultrasound scanner.
•Chapter 8 - Replacement Procedures
Provides disassembly and reassembly procedures for all Field Replaceable Units (FRUs).
•Chapter 9 - Renewal Parts
Contains a complete list of field replaceable parts for the Vivid i n portable ultrasound scanner.
•Chapter 10 - Care and Maintenance
Provides periodic maintenance procedures for the Vivid i n portable ultrasound scanner.
1-1-3Typical Users of the Service Manual
This manual is intended for the following categories of users:
•Hospital service personnel.
•Contractors (some parts of Chapter 2).
1-1-4Vivid i n Models Covered in this Manual
The Vivid i n models documented in this manual is shown in Table 1-2 below.
GE HEALTHCARE
DIRECTION FL091021, REVISION 5 VIVID I N SERVICE MANUAL
1-1-5System History - Hardware and Software Versions
The Vivid i n is a compact, fully portable, phased, linear array ultrasound imaging scanner.
Weighing only 5 Kgs (11 lbs), this laptop-style system is extremely versatile and, depending upon the
installed software, can be used for a variety of applications.
The Vivid i n portable ultrasound scanner is based on parallel products in the Vivid™ family. Using the
same advanced technology, the innovative features and versatility of the system represent the future of
ultrasound scanning.
NOTE:Upgrade options may be available with future versions.
1-1-6Purpose of Operator Manual(s)
The Operator Manual(s) should be fully read and understood before operating the Vivid i n system, and
also kept near the unit for quick reference.
Chapter 1 - Introduction 1-3
Page 30
GE HEALTHCARE
DIRECTION FL091021, REVISION 5 VIVID I N SERVICE MANUAL
Section 1-2
Important Conventions
1-2-1Conventions Used in this Manual
1-2-1-1Model Designations
This manual covers the Vivid i n ultrasound units listed in Table 1-2 on page 1-2.
1-2-1-2Icons
Pictures, or icons, are used wherever they will reinforce the printed message. The icons, labels and
conventions used on the product and in the service information are described in this chapter.
1-2-1-3Safety Precaution Messages
Various levels of safety precautions are found on the equipment and throughout this service manual.
Different levels of severity are identified by one of the following icons which precede precautionary
statements in the text.
DANGER: Indicates the presence of a hazard that will cause severe personal injury or death if the
instructions are ignored.
WARNING: Indicates the presence of a hazard that can cause severe personal injury and property
damage if the instructions are ignored.
CAUTION: Indicates the presence of a hazard that can cause property damag e but ha s a bsolu te ly n o
personal injury risk.
Note:Notes are used to provide important information about an item or a procedure. Be sure to read
the notes as the information they contain can often save you time or effort.
1-4Section 1-2 - Important Conventions
Page 31
GE HEALTHCARE
N
DIRECTION FL091021, REVISION 5 VIVID I N SERVICE MANUAL
1-2-1-4Standard Hazard Icons
Important information will always be preceded by the exclamation point contained within
a triangle, as seen throughout this chapter. In addition to text, several different graphical
icons (symbols) may be used to make you aware of specific types of hazards that could
cause harm.
Table 1-3 Standard Hazard Icons
ELECTRICALMECHANICALRADIATION
LASERHEATPINCH
Other hazard icons make you aware of specific procedures that should be followed.
Table 1-4 Standard Icons Indicating a Special Procedure Be Used
AVOID STATIC ELECTRICITYTAG AND LOCK OUTWEAR EYE PROTECTION
TAG
&
EYE
EYE
PROTECTIO
PROTECTION
Chapter 1 - Introduction 1-5
Page 32
GE HEALTHCARE
DIRECTION FL091021, REVISION 5 VIVID I N SERVICE MANUAL
Section 1-3
Safety Considerations
1-3-1Introduction
The following safety precautions must be observed duri ng all phases of operation, service and repair of
this equipment. Failure to comply with these precautions or with specific warnings elsewhere in this
manual, violates safety standards of design, manufacture and intended use of the equipment.
1-3-2Human Safety
Operating personnel must not remove the system covers.
Servicing should be performed by authorized personnel only.
Only personnel who have participated in Vivid i n training are authorized to service the equipment.
Local laws may restrict this device for sale or use by or on the order of a physician.
1-3-3Mechanical Safety
DANGER: WHEN THE VIVID I N SCANNER IS USED WITH A CART, MAKE SURE
THE UNIT IS FIRMLY SECURED IN THE CORRECT POSITION ON THE CART
SINCE IT MAY BECOME UNSTABLE AND TIP OVER.
DANGER: WHENENVER A CART IS TO BE MOVED ALONG ANY INCLINE, USE
EXTREME CAUTION. MAKE SURE THAT THE VIVID I N SCANNER AND ALL
PERIPHERALS ARE SECURELY MOUNTED ON THE CART BEFORE
ATTEMPTING TO MOVE IT.
DANGER: ULTRASOUND PROBES ARE HIGHLY SENSITIVE MEDICAL
INSTRUMENTS THAT CAN EASILY BE DAMAGED BY IMPROPER HANDLING.
USE CARE WHEN HANDLING AND PROTECT FROM DAMAGE WHEN NOT IN
USE. DO NOT USE A DAMAGED OR DEFECTIVE PROBE. FAILURE TO FOLLOW
THESE PRECAUTIONS CAN RESULT IN SERIOUS INJURY AND EQUIPMENT
DAMAGE.
DANGER: NEVER USE A PROBE THAT HAS BEEN SUBJECTED TO
MECHANICAL SHOCK OR IMPACT. EVEN IF THE PROBE APPEARS TO BE
UNBROKEN, IT MAY IN FACT BE DAMAGED.
CAUTION: The Vivid i n portable ultrasound scanner weighs 5kg (11 lbs.) or more, depending on
carry-on peripherals, when ready for use.
In addition, the Cart weighs approximately 30kg (66 lbs.), including peripherals. Care must be
used when moving the unit or replacing its parts. Failure to follow the precautions listed could
result in injury, uncontrolled motion and costly damage.
ALWAYS:
• Be sure the pathway is clear
• Use slow and careful motions
• Use two people when moving or lifting more then 16 kg (35 lbs)
1-6Section 1-3 - Safety Considerations
Page 33
GE HEALTHCARE
DIRECTION FL091021, REVISION 5 VIVID I N SERVICE MANUAL
1-3-3Mechanical Safety (cont’d)
ALWAYS:
•Use the handle to move the system
•Be sure the pathway is clear.
•Use slow, careful motions.
•Do not let the system strike walls or door frames.
•Whenever the Vivid i n scanner is mounted on the Cart and being moved on in clines, make
sure that the scanner and all peripherals are securely mounted on the Cart before
attempting to move it.
WARNING: ALWAYS LOCK THE CART IN ITS PARKED (LOCKED) POSITION AFTER
MOVING THE SYSTEM. FAILURE TO DO SO COULD RESULT IN PERSONAL INJURY OR
EQUIPMENT DAMAGE.
WARNING: EQUIPMENT DAMAGE COULD RESULT IF SPECIAL CARE IS NOT TAKEN
WHEN TRANSPORTING THE VIVID I N SYSTEM (AND CART, IF APPLICABLE) IN A
VEHICLE.
ALWAYS:
•Eject any disks from the MOD (if installed).
•Ensure that the Vivid i n system is well prepared and packed in its original packaging
before transporting. Special care must be taken to correctly position the packing material.
For further information, refer to Chapter 3- System Setup.
•Place the probes in their carrying case.
•Secure the Cart in an upright position and lock the wheels (brake).
•Ensure that the Vivid i n system (and Cart if applicable) is firmly secured while inside the
vehicle.
•Secure the system with straps or as directed otherwise to prevent motion during transport.
•Prevent vibration damage by driving cautiously. Avoid unpaved roads, excessive speeds,
and erratic stops or starts.
Chapter 1 - Introduction 1-7
Page 34
GE HEALTHCARE
DIRECTION FL091021, REVISION 5 VIVID I N SERVICE MANUAL
1-3-4Electrical Safety
To minimize shock hazard, the equipment must be connected to a well grounded power source. The
system is equipped with a three-conductor AC power cable. This must be plugged into an approved
electrical outlet with safety grounding.
To ensure proper grounding, connect this eq uip m en t to a rece pt acle marked "HOSPITAL ONLY" OR
"HOSPITAL GRADE".
The power outlet used for this equipment should not be shared with other type s of equipment. Bo th the
system power cable and the power connector must meet international electrical standards.
1-3-4-1Probes
All the probes for the Vivid i n ultrasound unit are designed and manufactured to provide trouble-free,
reliable service. To ensure this, correct handling of probes is important and the following points should
be noted:
•Do not drop a probe or strike it against a ha rd surface, as this may damage the transducer elements,
acoustic lens, or housing.
•Do not use a cracked or damaged probe. In this event, call your field service representative
immediately to obtain a replacement.
•Avoid pulling, pinching or kinking the probe cable, since a damaged cable may compromise the
electrical safety of the probe.
•To avoid the risk of a probe accidentally falling, do not allow the probe cables to become entangled,
or to be caught in the machine’s wheels.
NOTE:For detailed information on handling endocavity probes and invasive intracardiac probes, refer to the
appropriate supplementary instructions for each pr obe. In addition, refer to the Vivid i n User Manua l for
detailed probe handling instructions.
1-8Section 1-3 - Safety Considerations
Page 35
GE HEALTHCARE
DIRECTION FL091021, REVISION 5 VIVID I N SERVICE MANUAL
1-3-5Dangerous Procedure Warnings
Warnings, such as the warnings below, precede potentially dangerous proce dures throughout this
manual. Instructions contained in the warnings must be followed.
DANGEROUS VOLTAGES, CAPABLE OF CAUSING DEATH, ARE PRESENT IN THIS
EQUIPMENT. USE EXTREME CAUTION WHEN HANDLING, TESTING AND ADJUSTING.
EXPLOSION WARNING
DO NOT OPERATE THE EQUIPMENT IN AN EXPLOSIVE ATMOSPHERE.
OPERATION OF ANY ELECTRICAL EQUIPMENT IN SUCH AN ENVIRONMENT CONSTITUTES A
DEFINITE SAFETY HAZARD.
EQUIPMENT IS NOT SUITABLE FOR USE IN THE PRESENCE OF A FLAMMABLE ANAESTHETIC
MIXTURE WITH AIR OR WITH OXYGEN OR NITROUS OXIDE.
DO NOT SUBSTITUTE PARTS OR MODIFY EQUIPMENT
BECAUSE OF THE DANGER OF INTRODUCING ADDITIONAL HAZARDS, DO NOT INSTALL
SUBSTITUTE PARTS OR PERFORM ANY UNAUTHORIZED MODIFICATION OF THE
EQUIPMENT.
Chapter 1 - Introduction 1-9
Page 36
GE HEALTHCARE
DIRECTION FL091021, REVISION 5 VIVID I N SERVICE MANUAL
Section 1-4
Product Labels and Icons
The Vivid i n portable ultrasound scanner comes equipped wi th product l abels and icon s. These labels
and icons represent pertinent information regarding the operation of the unit.
1-4-1Product Label Locations
The following diagrams show the label and icons found on the Vivid i nultrasound unit. All the labels an d
icons are described in Table 1-5 "Product Icons Vivid i n" on page 1-11
Figure 1-1 Main Label for Vivid i n
1-10Section 1-4 - Product Labels and Icons
Page 37
GE HEALTHCARE
DIRECTION FL091021, REVISION 5 VIVID I N SERVICE MANUAL
1-4-2Label Descriptions
The following table shows the labels and symbols that may be found on the Vivid i n ultrasound unit,
and provides a description of each label’s purpose and location.
Table 1-5 Product Icons Vivid i n
Label NameDescriptionLocation
Manufacturer’s name and location.
Identification and Rating Plate
Type/Class Label Used to indicate the degree of safety or protection. Bottom panel of the adapter
IP Code (IPX1)
Date of Manufacture.
Model and Serial numbers.
Electrical ratings (Volts and Amps)
Indicates the degree of protection provided by the
enclosure per IEC60 529. Cannot be used in
operating room environment.
"CAUTION" The equilateral triangle is usually used
in combination with other symbols to advise or warn
the user.
On the base of the unit
Foot Switch
Various.
Device Listing/Certification
Labels
“ATTENTION - Consult accompanying documents”
is intended to alert the user to refer to the User
Manual or other instructions when complete
information cannot be provided on the label.
"CAUTION - Dangerous voltage" (the lightning flash
with arrowhead in equilateral triangle) is used to
indicate electric shock hazards.
Laboratory logos or labels that denote conformity
with industry safety standards, such as UL or IEC.
Indicates Equipment Type BF applied part for
medical equipment. Identifies a BF type applied part
complying with IEC 60601-1.
CE certification mark.On the base of the unit
Various.
Various.
On the base of the unit
Probe connectors.
On the base of the unit
Chapter 1 - Introduction 1-11
Page 38
GE HEALTHCARE
DIRECTION FL091021, REVISION 5 VIVID I N SERVICE MANUAL
Table 1-5 Product Icons Vivid i n
Label NameDescriptionLocation
Equipment Type CF IEC 60601-1 indicates
equipment having a floating applied part that
provides a degree of protection suitable for direct
cardiac contact.
Consult instructions for use.On the base of the unit
Waste Electrical and Electronic Equipment (WEEE)
Disposal
This symbol indicates that waste electrical and
electronic equipment must not be disposed of as
unsorted municipal waste and must be collected
separately.
Please contact an authorized representative of the
manufacturer for information concerning the
decommissioning of your equipment.
Above the ECG inlet, ECG connector
and surgical probes.
On the base of the unit.
CAUTION - This machine
weighs...Special care must be
used to avoid..."
"DANGER - Risk of explosion
used in..."
This precaution is intended to prevent injury that may
be caused by the weight of the machine if one
person attempts to move it considerable distances or
on an incline.
The system is not designed for use with flammable
anesthetic gases.
"Protective Earth" Indicates the protective earth
(grounding) terminal.
"Equipotentiality" Indicates the terminal to be used
for connecting equipotential conductors when
interconnecting (grounding) with other equipment.
“ON” indicates the power ON position of the power
switch. “Standby” indicates the power stand by
position of the power switch.
CAUTION This Power Switch DOES NOT ISOLATE
Mains Supply.
Used in the Service and User Manual
which should be adjacent to equipment
at all times for quick reference.
See "EXPLOSION WARNING" on page
1-9.
Rear of the Cart.
Peripherals
Adjacent to ON/Standby Switch
1-12Section 1-4 - Product Labels and Icons
Page 39
GE HEALTHCARE
DIRECTION FL091021, REVISION 5 VIVID I N SERVICE MANUAL
1-4-3Vivid i n Battery Safety
NOTE:The Vivid i n portable ultrasound scanner is supplied with a dummy battery case, and a lithium ion
battery in the battery bay, as standard.
The lithium ion battery provides power when an AC power so urce is not available. L ithium ion batteries
last longer than conventional batteries and do not require replacement as often. You can expect one
hour of battery life with a single, fully-charged battery.
The lithium ion technology used in the system’s battery is significantly less hazardous to the
environment than the lithium metal technology used in some other batteries (such a watch batteries).
Used batteries should not be placed with common household waste products. Contact local authorities
for the location of a chemical waste collection program nearest you.
NOTE:Regulations vary for different countries. Dispose of a used battery in accordance with local regula tions.
CAUTION: use only batteries approved by GE Healthcare as suitab le for use with the Vivid i n
portable ultrasound scanner.
WARNING: THE VIVID I N BATTERY HAS A SAFETY DEVICE.
DO NOT ATTEMPT TO DIS-ASSEMBLE OR ALTER THE BATTERY!
ALWAYS OBSERVE THE FOLLOWING PRECAUTIONS:
•Charge batteries only when the ambient temperature is between 0o and 65o C (32o and 149o F).
o
and discharge the batteries between -10
•Do not short-circuit the battery by directly connecting the negative terminals with metal objects.
•Do not heat the battery or discard it in a fire.
•Do not expose the battery to temperatures over 60
other heat sources.
•Do not charge the battery near a heat source, such as, a fire or heater.
•Do not leave the battery in direct sunlight.
•Do not pierce the battery with a sharp object, hit it, or step on it.
•Do not use a damaged battery.
•Do not apply solder to a battery.
•Do not connect the battery to an electrical power outlet.
and 55o C (14o and 131o F)
o
C (140o F). Keep the battery away from fire and
WARNING: IN THE EVENT THAT THE VIVID I N PORTABLE ULTRASOUND SCANNER
WILL NOT BE USED FOR A LONG PERIOD, IT IS NECESSARY TO REMOVE THE
BATTERY FROM THE SYSTEM WHILE NOT IN USE.
CAUTION: To prevent the battery bursting, igniting, or fumes from the battery causing
equipment damage, always observe the following precautions:
•Do not immerse the battery in water or allow it to get wet.
•Do not place the battery into a microwave oven or pressurized container.
•If the battery leaks or emits an odor, remove it from all possible flammable sources.
•If the battery emits an odor or heat, is deformed or discolored, or in a way appea rs abnormal during
use, recharging or storage, immediately remove it and stop using it.
•If you have any questions about the battery, consult your local GE representative.
Chapter 1 - Introduction 1-13
Page 40
GE HEALTHCARE
DIRECTION FL091021, REVISION 5 VIVID I N SERVICE MANUAL
1-4-3Vivid i n Battery Safety (cont’d)
•Storage of the battery pack:
-Short term (less than 1 month): between 0
-Long-term (more than 3 months): between 10
Note:When storing a battery pack for more than 6 months, it is necessary to charge the battery pack
at least once every 6 months in order to prevent leakage and deterioration in performance (as
a result of self-discharging).
Note:When charging the battery for the first time after long-term storage, recover the battery pack to
original performance through repeating several cycles of full charging and discharging.
o
C (32 oF) and 50 oC (122 oF).
o
C (50 oF) and 35 oC (95 oF)
Note:
When shipped, the Vivid i n battery is in a state of being approximately 30% charged. Before use, it is
necessary to fully charge and discharge the battery pack up to 3 times, in order to utilize Li-lon smart
packs.
Important:
•The Vivid i n portable ultrasound scanner is supplied from the facto ry with a dummy battery. Wh en
unpacking the unit, do not discard the dummy battery.
•When charging the Vivid i n lithium ion battery on an external charger, since the battery in fact
forms one of the four legs of the scanner, removing it will leave the unit standing unbalanced. For
this reason, the dummy battery should be kept, so it may be inserted in position to provide stability
to the scanner while the lithium ion battery is being charged.
•In addition, use of the dummy battery is recommended during transportation or long-term storage
of the Vivid i n portable ultrasound scanner.
1-14Section 1-4 - Product Labels and Icons
Page 41
GE HEALTHCARE
DIRECTION FL091021, REVISION 5 VIVID I N SERVICE MANUAL
1-4-4Vivid i n External Labels
In addition to the labels described in the previous section, additional labels may be found on the
Vivid i n ultrasound unit, as described in the following sections:
•Rating Labels section, see below.
1-4-4-1Rating Labels
Indicates the ultrasound unit’s basic power compliance and regulatory compliance information
The rating label shown below is located on the base assembly (underneath the ultrasound unit).
Figure 1-2 Vivid i n Rating Label
Chapter 1 - Introduction 1-15
Page 42
GE HEALTHCARE
DIRECTION FL091021, REVISION 5 VIVID I N SERVICE MANUAL
Section 1-5
EMC, EMI, and ESD
1-5-1Electromagnetic Compatibility (EMC)
Electromagnetic compatibility describes a level of performance of a device within its electromagnetic
environment. This environment consists of the device itself and its surroundings, including other
equipment, power sources and persons with which the device must interface. Inadequate compatibility
results when a susceptible device fails to perform as intended due to interference from its environment,
or when the device produces unacceptable levels of emission. This interference is often referred to as
radio–frequency or electromagnetic interference (RFI/EMI) and can be radiated thr ough space or
conducted over interconnecting power or signal cables. In addition to electromagnetic energy, EMC
also includes possible effects from electrical fields, magnetic fields, electrostatic discharge and
disturbances in the electrical power supply.
NOTE:The Vivid i n system should be operated at a distance of more than 15 feet from any magnetic
field.
1-5-2CE Compliance
The Vivid i n portable ultrasound scanner conforms to all applicable conducted and radiated emission
limits and to immunity from electrostatic discharge, radiated and conducted RF fields, magnetic fields
and power line transient requirements.
For applicable standards refer to the Safety Chapter in the Vivid i n User Manual.
NOTE:For CE Compliance, it is critical that all covers, screws, shielding, gaskets, mesh and clamps are in good
condition and installed tightly without skew or stress. Proper installation following all comments noted
in this service manual is required in order to achieve full EMC performance.
1-5-3Electrostatic Discharge (ESD) Prevention
CAUTION: Do not touch any boards with integrated circuits prior to taking the necessary ESD
precautions:
Always connect yourself, via an arm-wrist strap connected to the bottom cover whenever you
open the system for maintenance.
Follow general guidelines for handling of electrostatic sensitive equipment.
1-16Section 1-5 - EMC, EMI, and ESD
Page 43
GE HEALTHCARE
DIRECTION FL091021, REVISION 5 VIVID I N SERVICE MANUAL
1-5-4Standards Used
To fulfill the requirements of relevant EC directives and/or European Harmonized/International
standards, the following documents/standards have been used :
Table 1-6 Standards Used
ScopeStandard/Directive
IEC 60601-1:1988+A1:1991+A2:1995
Product safety requirements for Electrical Medical devices
Class I, Type BF with CF applied parts (ICE catheter and
ECG leads).
EN 60601-1:1990+A1:1993+A2:1995+A13:1996
UL 60601-1 (2003)
C 22.2 601-1 M90 (R2005)
Product safety requirements for Electrical Medical devicessafety requirements for medical electrical systems
Product safety requirements for Electrical Medical devicesProgrammable electrical medical systems
Product safety requirements for Ultrasound system
Product EMC requirements
Group 1. Class A
Biological evaluation of medical devicesISO 10993-1:2003
Particular requirements for the safety of electrocardiograph
monitoring equipment (Not for monitoring-only for
synchronization with ultrasound)
General requirements for basic safety and essential
performance-Usability
Patient privacyHIPPA Safety.
IEC 60601-1-1:2000
EN 60601-1-1:2001
IEC 60601-1-4:1996+A1:1999
EN 60601-1-4:1996+A1:1999
IEC 60601-2-37:2001+A1:2004+A2:2005
EN 60601-2-37: 2001+A1:2005 +A2:2005
IEC 60601-1-2:2001+A1:2004
EN 60601-1-2:200+A1:2006
IEC60601-2-27:1994
EN60601-2-27:1994
IEC60601-1-6:2004
EN60601-1-6:2004
NOTE:For CE Compliance, it is critical that all covers, screws, shielding, gaskets, mesh and clamps are in good
condition and installed tightly without skew or stress. Proper installation following all comments noted
in this service manual is required in order to achieve full EMC performance.
Chapter 1 - Introduction 1-17
Page 44
GE HEALTHCARE
DIRECTION FL091021, REVISION 5 VIVID I N SERVICE MANUAL
Section 1-6
Lockout/Tagout (LOTO) Requirements
Follow local Lockout/Tagout (LOTO) requirements by ensuring you a re in total control of the AC power
plug at all times during the service process.
To apply Lockout/Tagout:
1.) Plan and prepare for shutdown.
2.) Shutdown the equipment.
3.) Isolate the equipment.
4.) Apply Lockout/Tagout Devices.
5.) Control all stored and residual energy.
6.) Verify isolation.
All potentially hazardous stored or residual energy is relieved.
NOTICE
Energy Control and Power Lockout for Vivid i n.
When servicing parts of the system where there is exposure to voltage greater than 30 Volts:
1.) Turn off the breaker.
2.) Unplug the system.
3.) Maintain control of the system power plug.
4.) Wait for at least 20 seconds for capacitors to discharge as there are no test points to verify isolation.
Beware that parts inside the unit may be energized even if the power is turned off when the cord is still
plugged into the AC Outlet.
Section 1-7
Returning/Shipping Probes and Repair Parts
Equipment being returned must be clean and free of blood and other in fectious substances.
GE Healthcare policy states that body fluids must be properly removed from any part or equipment prior
to shipment. GE Healthcare employees, as well as customers, are responsible for ensuring that parts/
equipment have been properly decontaminated prior to shipment. Under no circumstance should a part
or equipment with visible body fluids be taken or shipped from a clinic or site (for example, body coils
or an ultrasound probe). The purpose of the regulation is to protect employees in the transportation
industry, as well as the people who will receive or open this package.
GE HEALTHCARE
DIRECTION FL091021, REVISION 5 VIVID I N SERVICE MANUAL
Section 1-8
Customer Assistance
1-8-1Contact Information
If this equipment does not work as indicated in this service manual or in the Vivid i n User Manual, or if
you require additional assistance, please contact the local distributor or appropriate support resource,
as listed below.
Prepare the following information before you call:
•System ID and/or serial number.
•Software version.
•Date and time of occurrence.
•Sequence of events leading to issue.
•Is the issue reproduceable?
•Imaging mode, probe, preset/application.
•Media brand, speed, capacity, type.
•Detailed description of any problem encountered.
•Save secondary image capture, cine loop.
•Where applicable, save the appropriate log files (by pressing <Ctrl F> [or <Alt D]).
Remember to save the log files for each day on a separate floppy disk, labelled accordingly.
Note:Restart the application before resuming clinical scanning.
Table 1-7 Phone numbers for Customer Assistance
LOCATIONPHONE NUMBER
USA
GE Healthcare
Ultrasound Service Engineering
9900 Innovation Drive
Wauwatosa, WI 53226
Canada1-800-668-0732
Latin America
Europe (OLC- EMEA)
GE Ultraschall Deutschland GmbH
Beethovenstraße 239
Postfach 11 05 60, D-42655 Solingen
Germany
GE HEALTHCARE
DIRECTION FL091021, REVISION 5 VIVID I N SERVICE MANUAL
This page was intentionally left blank.
1-20Section 1-8 - Customer Assistance
Page 47
GE HEALTHCARE
DIRECTION FL091021, REVISION 5 VIVID I N SERVICE MANUAL
Chapter 2
Site Preparations
Section 2-1
Overview
2-1-1Purpose of Chapter 2
This chapter provides the information required to plan and prepare for the installation of a Vivid i n
ultrasound unit. Included are descriptions of the electrical and facility requirements that must be met by
the purchaser. A worksheet is provided at the end of this chapter (see Figure 2-2 on page 2-10 ) to help
ensure that all the required network information is available, prior to installation.
GE HEALTHCARE
DIRECTION FL091021, REVISION 5 VIVID I N SERVICE MANUAL
Section 2-2
Console Requirements
2-2-1Unit Environmental Requirements
Relative Humidity
RequirementTemperature
(non-condensing)Air Pressure
Operational
Storage
Transport
10 — 40
- 10 — 60
- 10 — 60
o
C(50 — 104 oF)
o
C (14 — 140 oF)
o
C (14 — 140 oF)
30 — 85%700 — 1060 hPa
30 — 95%700 — 1060 hPa
30 — 95%700— 1060 hPa
Table 2-2 Environmental Requirements
CAUTION: If the system has been in storage or has been transported, please see the acclimation
requirements before powering
ON and/or using the system. See “Installation Warnings” on page 3-2.
2-2-2Cooling Requirements
The cooling requirement for the Vivid i n ultrasound unit environment is 3500 BTU/hr. This figure does
not include the cooling required for lights, people, or other equipment in the room.
Note:Each person in the room places an additional 300 BTU/hr demand on the environmental cooling.
2-2-3Lighting Requirements
For system installation, updates and repairs, bright lighting is required. However, operator and patient
comfort may be optimized if the room lighting is subdued and indirect when a scan is being perfo rmed.
Therefore, a combination lighting system (dim/bright) is recommended. Keep in mind that lighting
controls and dimmers can be a source of EMI which could degrade image quality. These controls shou ld
be selected to minimize possible interference.
2-2-4Time and Manpower Requirements
Site preparation takes time. Begin pre-installation checks as soon as possible to allow sufficient time to
make any required changes. If possible, begin these checks as many as six weeks before system
delivery.
CAUTION: Only one person is required to unpack the Vivid i n ultrasound unit; at least two people
must be available to roll the system down the wheeling ramp. Attempts to move the system
considerable distances (or on an incline) by one person alone, could result in personal injury,
and/or damage to the system.
2-2Section 2-2 - Console Requirements
Page 49
GE HEALTHCARE
DIRECTION FL091021, REVISION 5 VIVID I N SERVICE MANUAL
2-2-5Electrical Requirements
NOTE:GE Healthcare requires a dedicated mains power line and Ground for the proper operation of
its Ultrasound equipment. This dedicated power line shall originate at the last distribution panel
before the system.
Sites with a mains power system with defined Neutral and Live:
The dedicated line shall consist of one phase, a neutral (not shared with any other circuit), and a full
size Ground wire from the distribution panel to the Ultrasound outlet.
Sites with a mains power system without a defined Neutral:
The dedicated line shall consist of one phase (two lines), not shared with any other circuit, and a full
size Ground wire from the distribution panel to the Ultrasound outlet.
NOTE:Please note that image artifacts can occur, if at any time within the facility, the Ground from the
main facility's incoming power source to the Ultrasound unit is only a conduit.
2-2-5-1Vivid i n Power Requirements
Electrical specifications for the Vivid i n system are as follows:
Input
Adaptor
AC DC converter
Voltage
100V AC to
240V AC
Table 2-3 Electrical Requirements
2-2-5-2Inrush Current
Inrush current is not a factor for consideration, due to the inrush current limiting prop erties of the power
supplies.
Voltage
100 V4.5 A9 A
240 V2.3 A4.5 A
2-2-5-3Site Power Outlets
A dedicated AC power outlet must be within reach of the unit without requiring the use of extension
cords. Other outlets adequate for the external peripherals, medical and test equipment required to
support this unit must also be present and located within 1 m (3.2 ft) of the unit. Electrical installation
must meet all current local, state, and national electrical codes.
Output
VoltageTolerancesOp. CurrentFrequency
20V±10%0.5 to 1A50-60 Hz
Inrush Current
Console OnlyConsole with all Peripherals
Table 2-4 Inrush Current
Chapter 2 - Site Preparations 2-3
Page 50
GE HEALTHCARE
DIRECTION FL091021, REVISION 5 VIVID I N SERVICE MANUAL
2-2-5-4Mains Power Plug
The Vivid i n portable ultrasound scanner are supplied with a mains power plug, as standard . In the
event that the unit arrives without a power plug, or with the wrong plug, contact your GE dealer. When
necessary, the installation engineer will supply the appropriate power plug to meet the applicable local
regulations.
2-2-5-5Power Stability Requirements
•Voltage drop-out
Max 10 ms.
Note:Only relevant if system is operating without a battery/
•Power Transients
(All applications)
Less than 25% of nominal peak voltage for less than 1 millisecond for any type of transient,
including line frequency, synchronous, asynchronous, or aperiodic transients.
2-2-6EMI Limitations
Ultrasound machines are susceptible to Electromagnetic Interference (EMI) from radio frequencies,
magnetic fields, and transients in the air or wiring. They also gener ate EMI. The Vivid i n ultrasound unit
comply with limits as stated on the EMC label. However, there is no guarantee that interference will not
occur in a particular installation.
NOTE:Possible EMI sources should be identified before the unit is installed, and should not be on the
same line as the ultrasound system. A dedicated line sh ould be used for the ultrasound system.
Electrical and electronic equipment may produce EMI unintentionally as the result of a defect. Sources
of EMI include the following:
•Medical lasers.
•Scanners.
•Cauterizing guns.
•Computers.
•Monitors.
•Fans.
•Gel warmers.
•Microwave ovens.
•Portable phones.
•Broadcast stations and mobile broadcasting machines.
2-4Section 2-2 - Console Requirements
Page 51
GE HEALTHCARE
DIRECTION FL091021, REVISION 5 VIVID I N SERVICE MANUAL
2-2-6EMI Limitations (cont’d)
The following table lists recommendations for preventing EMI:
EMI RuleDetails
Ground the unit.
Be aware of RF sources.
Replace and/or reassemble
all screws, RF gaskets,
covers and cores.
Replace broken RF gaskets.
Do not place labels where
RF gaskets touch metal.
Use GE-specified harnesses
and peripherals.
Take care with cellular
phones.
Properly address peripheral
cables.
Table 2-5 EMI Prevention/ Abatemen t
Poor grounding is the most likely reason an ultrasound unit will have noisy images.
Check the grounding of the power cord and power outlet.
Keep the unit at least 5m (16.4 ft) away from other EMI sources. Special shielding may
be required to eliminate interference problems caused by high frequency, high powered
radio or video broadcast signals.
After you finish repairing or updating the system, replace all covers and tighten all
screws. Any cable with an external connection requires a magnet wrap at each end.
Install the shield over the front of the card cage. Loose or missing covers or RF gaskets
allow radio frequencies to interfere with the ultrasound signals.
If more than 20% or a pair of the fingers on an RF gasket are broken, replace the gasket.
Do not turn ON the unit until any loose metallic part is removed and replaced, if required.
Never place a label where RF gaskets meet the unit. Otherwise, the gap created will
permit RF leakage. In case a label has been found in such a location, move the label to
a different, appropriate location.
The interconnect cables are grounded and require ferrite beads and other shielding.
Cable length, material, and routing are all important; do not make any changes that do
not meet all specifications.
Cellular phones may transmit a 5 V/m signal that causes image artifacts.
Do not allow cables to lie across the top of the system. Loop any peripheral cable excess
length into one bundle.
2-2-7Probe Environmental Requirements
Electronics
Operation
Storage
Note:System and electronic probes are designed for storage temper atures of -10
exposed to large temperature variations, the probes should be kept at room temperature for a
minimum of 10 hours before use.
Chapter 2 - Site Preparations 2-5
10 — 40oC (50 — 104oF)
-10 — 60oC (14 — 140oF)
Table 2-6 Probe Operation and Storage Temperatures
oC
to +60 oC. When
Page 52
GE HEALTHCARE
DIRECTION FL091021, REVISION 5 VIVID I N SERVICE MANUAL
Section 2-3
Facility Needs
2-3-1Purchaser Responsibilities
The work and materials required to prepare the site are the responsibility of the purchaser. To avoid
delay, complete all pre-installation work before delivery. Use the Pre-installation Check List (provided
in Table 2-7 on page 2-12)to verify that all the required steps have been completed.
Purchaser responsibilities include:
•Procuring the required materials.
•Completing the preparations prior to delive ry of the ult ra so un d sys te m.
•Paying the costs of any alterations and modifications not specifically provided for in the sales
contract.
Note:All relevant preliminary electrical installations at the prepared site must be performed by
licensed electrical contractors. Other connections between electrical equipment, and
calibration and testing, must also be performed by qualified personnel. The products involved
(and the accompanying electrical installations) are h ighly sophisticated and special engineering
competence is required. All electrical work on these products must comply with the
requirements of applicable electrical codes. The purchaser of GE equipment must utilize only
qualified personnel to perform electrical servicing of the equipment.
To avoid delays during installation, the individual or team who will perform the installation should be
notified at the earliest possible date (preferably prior to installation), of the existence of any of the
following variances:
•Use of any non-listed product(s).
•Use of any customer provided product(s).
•Placement of an approved product further from the system than the interface kit allows.
The prepared site must be clean prior to delivery of the system. Carpeting is not recomme nded because
it collects dust and creates static. Potential sources of EMI should also be investigated before delivery.
Dirt, static, and EMI can negatively impact system reliability.
2-6Section 2-3 - Facility Needs
Page 53
GE HEALTHCARE
DIRECTION FL091021, REVISION 5 VIVID I N SERVICE MANUAL
2-3-2Mandatory Site Requirements
The following are mandatory site requirements. Additional (optional) recommendations, as well as a
recommended ultrasound room layout, are provided in section 2-3-3 - Site Recommendations (see
below).
•A dedicated single branch power outlet of adequate amperage (see Table 2- 3, “Electrical
Requirements,” on page 3) that meets all local and national codes and is located less than 2.5 m
(8.2 ft) from the unit’s proposed location. See “Electrical Requirements” on page 2-3.
•A door opening of at least 76 cm (2.5 ft) in width (if using the Cart).
•The proposed location for the unit is at least 0.2 m (0.67 ft) from the walls, to enable cooling.
•Power outlet and place for any external peripheral are within 2 m (6.5 ft.) of each other with
peripheral within 1 m of the unit to connect cables.
•Power outlets for other medical equipment and gel warmer.
•Power outlets for test equipment within 1 m (3.3 ft) of the ultrasound unit.
•Clean and protected space for storage of probes (either in their case or on a rack).
•Material to safely clean probes (perform ed usin g a pla stic container, never metal).
•In the case of a network option:
-An active network outlet in the vicinity of the ultrasound unit.
-A network cable of appropriate length (regular Pin-to-Pin network cable).
-An IT administrator who will assist in configuring the unit to work with your local network. A
fixed IP address is required. Refer to the form provided in Figure 2-2 on page 2-10 for network
details that are required.
Note:All relevant preliminary network outlets installations at the prepared site must be per formed by
authorized contractors. The purchaser of GE equipment must utilize only qualified personnel to
perform servicing of the equipment.
2-3-3Site Recommendations
The following are (optional) site recommendations. Mandatory site requirements are provided in the
Mandatory Site Requirements section, above.
•Door opening of 92 cm (3 ft) in width (if using a cart).
•Accessible circuit breaker for a dedicated power outlet.
•Sink with hot and cold running water.
•Receptacle for bio–hazardous waste, for example, used probe sheaths.
•Emergency oxygen supply.
•Storage area for linens and equipment.
•Nearby waiting room, lavatory, and dressing room.
•Dual level lighting (bright and dim).
•Lockable cabinet for software and manuals.
Chapter 2 - Site Preparations 2-7
Page 54
GE HEALTHCARE
DIRECTION FL091021, REVISION 5 VIVID I N SERVICE MANUAL
2-3-3-1Recommended Ultrasound Room Layout
Figure 2-1 below shows a floor plan illustrating the recommended layout of the Ultrasound Room and
depicting the minimal room layout requirements.
Vivid i n
36 IN.
Dedicated Power Outlets
Dedicated Analog Telephone Line
for Connection to InSite
GE Cabinet for
Software and Manuals
(optional)
(92 CM)
Figure 2-1 Recommended Floor Plan 4.3m x 5.2m (14 ft x 17 ft)
Hospital Network
2-8Section 2-3 - Facility Needs
Page 55
GE HEALTHCARE
DIRECTION FL091021, REVISION 5 VIVID I N SERVICE MANUAL
2-3-4Networking Pre-Installation Requirements
2-3-4-1Stand-alone Unit (without Network Connection)
None.
2-3-4-2Unit Connected to Hospital’s Network
Supported networks:
•Wired LAN
•Wireless LAN
2-3-4-3Purpose of the DICOM Network Function
DICOM services provide the operator with clinically useful features for moving images and patient
information over a hospital network. Examples of DICOM services include the transfer of images to
workstations for viewing or transferring images to remote prin ters. As an added benefit, transferring
images in this manner frees up the on-board moni tor and peripherals, enabling viewing to be don e while
scanning continues. With DICOM, images can be archived, stored, and retrieved faste r, easier , and at
a lower cost.
2-3-4-4DICOM Option Pre-Installation Requirements
To configure the Vivid i nultrasound unit to work with other network connections, the network
administrator must provide the required information, which should include the following:
•Vivid i n details:DICOM network details for the Vivid i nunit, including the host
name, local port, IP address, AE title and net mask.
•Routing Information:IP addresses for the default gateway and other routers in use at
the site.
•DICOM Application Information:Details of the DICOM devices in use at the site, including the
DICOM host name, AE title and IP addresses.
Chapter 2 - Site Preparations 2-9
Page 56
GE HEALTHCARE
DIRECTION FL091021, REVISION 5 VIVID I N SERVICE MANUAL
GE HEALTHCARE
DIRECTION FL091021, REVISION 5 VIVID I N SERVICE MANUAL
Chapter 3
System Setup
Section 3-1
Overview
3-1-1Purpose of Chapter 3
This chapter provides instructions for installing the Vivid i n ultrasound unit. Before beginning the
installation process, an appropriate site must be prepared, as described in Chapter 2. Once the site has
been prepared, installation can proceed as described in this chapter.
Included in this chapter are guidelines for transporting the unit to a new site, as well as procedures that
describe how to receive and unpack the equipment, and (if necessary) how to file a damag e or loss
claim. Instructions for checking and testing the unit, probes, and external peripherals for electrical safety
are also provided.
Table 3-1 Contents in Chapter 3
SectionDescriptionPage Number
3-1
3-2
3-3
3-4
3-5
3-6
3-7
3-8
3-9
3-10
Overview
Installation Reminders
Receiving the Vivid i n
Unpacking the Equipment
Preparing for Installation
Completing the Hardware Installation
Configuration
Connectivity Setup
Storing and Transporting the Unit
Completing the Installation Paperwork
3-1
3-2
3-5
3-8
3-12
3-19
3-47
3-62
3-154
3-155
Chapter 3 - System Setup 3-1
Page 60
GE HEALTHCARE
DIRECTION FL091021, REVISION 5 VIVID I N SERVICE MANUAL
Section 3-2
Installation Reminders
3-2-1Average Installation Time
The Vivid i n installation and functional checkout will take approximately one hour; Vivid i nconsoles
with optional equipment may take slightly longer.
Once the site has been prepared, the average installation time required is shown in Table 3-2 below.
Table 3-2 Average Installation Time
Average
Description
Unpacking the scanner20 minutes
Installing the scanner30 minutesTime may vary, according to the required configuration
DICOM Option
(connectivity)
3-2-2Installation Warnings
Installation TimeComments
30 minutesTime may vary, according to the required configuration
1.) The Vivid i n portable ultrasound scanner weighs only 5 Kgs (11 lbs), without options.
2.) There are no operator-serviceable components. To prevent shock, do not remove any covers or
panels. Should problems or malfunctions occur, unplug the power cord. Only qualified service
personnel should carry out servicing and troubleshooting.
NOTE:For information on transportation box labels, refer to Figure 3-1 on page 3- 4 and Figure 3-3 on
page 3-7 .
3-2-2-1System Acclimation Time
Following transport, the Vivid i n system may be very cold, or hot. Allow time for the system to acclimate
o
C increment, when the
CAUTION
before being switched ON. Acclimation requires 1 hour for each 2.5
o
temperature of the system is below 10
C or above 35 oC.
Turning the system ON after arrival at the site - without allowing time for acclimation - may
cause system damage!
Table 3-3 Vivid i n System Acclimation Time
ALLOWED TRANSPORTATION AND STORAGE TEMPERATURESOUT OF SPEC!
GE HEALTHCARE
DIRECTION FL091021, REVISION 5 VIVID I N SERVICE MANUAL
3-2-3Safety Reminders
DANGER: WHEN USING ANY TEST INSTRUMENT THAT IS CAPABLE OF
OPENING THE AC GROUND LINE (I.E., METER’S GROUND SWITCH IS OPEN),
DO NOT TOUCH THE UNIT!
CAUTION: If the unit is very cold or hot, do NOT turn ON power to the unit until it has had
sufficient time to acclimate to its operating environment.
CAUTION: To prevent electrical shock, connect the unit to a properly grounded power outlet.
Do NOT use a three-prong to two-prong adapter, as this defeats safety grounding.
CAUTION: To ensure proper grounding, connect this equipment to a receptacle marked
"HOSPITAL ONLY" OR "HOSPITAL GRADE".
CAUTION: Do NOT wear the ESD wrist strap when you work on live circuits where more than
30 V peak is present.
CAUTION: Do NOT operate the unit unless all board covers and frame panels are securely in
place, to ensure optimal system performance and cooling. (When cov ers are removed, EMI may
be present).
WARNING: ACOUSTIC OUTPUT HAZARD
ALTHOUGH THE ULTRASOUND ENERGY TRANSMITTED FROM THE VIVID I N IS
WITHIN AIUM/NEMA STANDARDS AND FDA LIMITATIONS, AVOID UNNECESSARY
EXPOSURE. ULTRASOUND ENERGY CAN PRODUCE HEAT AND MECHANICAL
DAMAGE.
Note:The Vivid i n User Manual should be fully read and under stood before operating the unit. Keep
the manual near the unit for reference.
Chapter 3 - System Setup 3-3
Page 62
GE HEALTHCARE
DIRECTION FL091021, REVISION 5 VIVID I N SERVICE MANUAL
3-2-4The Tilt & Shock Indicators
3-2-4-1Overview
Unproper handeling during transportation may harm the equepment inside the package even if the
package itself is undamaged.
To make it easier to detection if the handeling during transportation has been unproper, a set of Tilt &
Shock indicators have been attached to the transportation box.
3-2-4-2Position of the Tilt & Shock Indicators
The Tilt & Shock indicators have been attached to the right side of the transportation box as illustrated
in the figure below.
Figure 3-1 Tiltwatch and Shockwatch positions on Transportation Box
3-4Section 3-2 - Installation Reminders
Page 63
GE HEALTHCARE
DIRECTION FL091021, REVISION 5 VIVID I N SERVICE MANUAL
Section 3-3
Receiving the Vivid i n
3-3-1Examin All Packages
Examin all packages closely at time of delivery, as described in the procedure below.
Table 3-4 Examin All Packages
STEPTASKILLUSTRATIONS
1
2Is the Shock Indicator red colored inside the middle of the
3Is the Tilt Indicator red colored inside the middle of the
Is dammage apparent?
•If yes; continue with the instructions in subsection
3-3-2 - Damage in Transportation.
•If no; continue with step 2.
indicator?
•If yes: The Shock Indicator has been trigged.
Make a remark on the Post Delivery Checklist about the
trigged indicator before you continue with step 2.
•If no: Continue with step 2.
indicator?
•If yes: The Tilt Indicator has been trigged.
Make a remark on the Post Delivery Checklist about the
trigged indicator and then follow the rest of the
instructions in subsection 3-3-2 - Damage in
Transportation.
•If no: Continue with the instructions in Section 3-3 -
Receiving the Vivid i n.
RED COLOR
RED COLOR
Chapter 3 - System Setup 3-5
Page 64
GE HEALTHCARE
DIRECTION FL091021, REVISION 5 VIVID I N SERVICE MANUAL
3-3-2Damage in Transportation
Follow this procedure if damage is apparent, or if any of the Tilt & Drop Indicators show failure:
Table 3-5 Damage in Transportation
STEPTASK
1
2
3
Write “Damage In Shipment” on ALL copies of the freight or express bill BEFORE delivery is accepted or “signed
for” by a GE representative or hospital receiving agent.
Report the damage to the carrier.
•Whether noted or concealed, damage MUST be reported to the carrier immediately upon discovery, or in
any event, within 14 days after receipt, and the contents and containers held for inspection by the carrier.
•A transportation company will not pay a claim for damage if an inspection is not requested within this 14 day
period.
Report the damage on the Post Delivery Checklist.
Specify if the tilt & drop indicators show failure in the “Packing” field on the Post Delivery Checklist.
3-6Section 3-3 - Receiving the Vivid i n
Page 65
GE HEALTHCARE
DIRECTION FL091021, REVISION 5 VIVID I N SERVICE MANUAL
3-3-2-1Vivid i n Transportation Box Label
The Vivid i n Transportation Box Label is located at the front of the transportation box.
RELATIVE HUMIDITY
BETWEEN 30 and 95%
TOP.
UPRIGHT
TRANSPORTATION
& STORAGE
FRAGILE,
HANDLE WITH CARE
KEEP TRANSPORTATION
TEMPERATURE BETWEEN
-20°C and +60°C
RECYCLING.
RECYCLABLE WOOD
HANDLE WITH CARE
Figure 3-2 Vivid i n Transportation Box Label
KEEP DRY,
PROTECT FROM MOISTURE
Figure 3-3 Label mounted
Chapter 3 - System Setup 3-7
Page 66
GE HEALTHCARE
DIRECTION FL091021, REVISION 5 VIVID I N SERVICE MANUAL
Section 3-4
Unpacking the Equipment
CAUTION: Please read this section fully before unpacking the Vivid i n ultrasound unit.
Figure 3-4 Shipping Box for the Vivid i n only
Figure 3-5 To remove the Lid of the Box
Open the Lock
Break the seal
3-8Section 3-4 - Unpacking the Equipment
Page 67
GE HEALTHCARE
DIRECTION FL091021, REVISION 5 VIVID I N SERVICE MANUAL
Section 3-4 Unpacking the Equipment (cont’d)
Figure 3-6 Contents in Transportation box when Lid is removed
Table 3-6 Shipping Carton Dimensions and Weights
DescriptionHeightWidthDepth
Vivid i n scanner with
Probes, Peripherals and accessories
a.Weight is approximate and will vary depending on supplied peripherals
3-4-1Examin All Packages
Examin the Transportation box closely at time of delivery, as described in the procedure that follow.
a
Weight
85 cm87 cm66cm35 kgs. Empty
Chapter 3 - System Setup 3-9
Page 68
GE HEALTHCARE
DIRECTION FL091021, REVISION 5 VIVID I N SERVICE MANUAL
3-4-2Unpacking the Wooden Transportation box
After completing a visual inspection of the Transportation box and Drop/Tilt indicators, proceed
unpacking as follows:
1.) Remove sealers and unlock two locking levers at two sides of the Box as shown in Figure 3-5 on
page 3-8 .
2.) Remove the packing slip (shipping consignment note) detailing the contents of the Transportation
Box, keep close to hand and be ready to mark the check list - refer to Table 3-7 "Damage Inspection
Checklist - Vivid i n System" on page 3-13.
3.) Remove the surface packing material (paper pad, silica gel, etc.). Continue to remove additional
packing material as applicable during the following procedure steps.
4.) Carefully remove the box containing the probes.
5.) Take out the package containing the following accessories: cables (AD/DC; ECG; Network),
Isolation Box, and software CD.
6.) Verify the content against the Delivery Note.
7.) Remove each of the boxes (one, two, or more, depending on options ordered) containing the
peripherals.
8.) Carefully remove the Vivid i n ultrasound unit from the Cardb oard box it is delivered in as shown in
Figure 3-7 on page 3-10 .
Vivid i n Scanner
Figure 3-7 Unpacked the Vivid i n.
3-10Section 3-4 - Unpacking the Equipment
Page 69
GE HEALTHCARE
DIRECTION FL091021, REVISION 5 VIVID I N SERVICE MANUAL
3-4-3Verifying the Transportation box Contents
After unpacking, it is important to verify that all items ordered by the customer have b een received (refer
to "Verifying the Transportation Box Contents" o n p age 3-12 ). Comp are a ll items listed on the packing
slip (shipping consignment note) with those received and re port any items that are missing, backordered, or damaged, to your GE Healthcare sales representative. The contact address is shown in
1-8-1 "Contact Information" on page 1-19 .
NOTICE
It is recommended to keep and store the Transportation Box and all other p acking materials in case the
unit has to be moved to a different location in the future.
For warranty purposes, storage of the above is required for one year from date of purchase.
3-4-4Physical Inspection
3-4-4-1System Voltage Settings
Verify that the Vivid i n ultrasound AC adapter and Cart (if applicable) are set to the correct voltage.
The Voltage settings are 220-240V AC - (Europe, Latin America, and China).
WARNING Setting the Vivid i n ultrasound components to the wrong voltage setting will most
likely destroy the equipment.
3-4-5EMI Protection
The Vivid i n Ultrasound Unit has been designed to minimize the effects of Electro Magnetic
Interference (EMI). Many of the covers, shields, and screws are provided primarily to protect the system
from image artifacts caused by this interference. For this reason, it is imperative that all covers and
hardware are installed and secured before the unit is put into operation.
Chapter 3 - System Setup 3-11
Page 70
GE HEALTHCARE
DIRECTION FL091021, REVISION 5 VIVID I N SERVICE MANUAL
Section 3-5
Preparing for Installation
3-5-1Confirming Customer Order
When preparing for installation of a Vivid i n system, it is important to verify that all items ordered by the
customer have been received. Compare all items listed on the packing slip (shipping consignment note)
with those received and report any items that are missing, back-ordered, or damaged, to your GE
Healthcare sales representative.
3-5-2Verifying the Transportation Box Contents
The following sections list the contents of the shipping carton that are additi onal to the Vivid i n. These
include external accessories and probes, as well as a regional language support kit, and optional
peripherals (as ordered). Ensure that all relevant components are present before comp leting the
installation.
NOTE:The transportation box contains an External Accessory Kit and Optional Peripherals Check
List. When checking the contents, make sure the Check List is completed. In the event that any
items are missing, contact your local GE Healthcare representative.
The Transportation box will contain the probes that have been ordered with the system.
For a list of probes available for use with the Vivid i n portable ultrasound scanner, refer to Section 9-7
"Probes" on page 9-8.
3-5-2-1Probes
The shipping carton will contain the probes that have been ordered with the system.
For a list of probes available for use with the Vivid i n, refer to "Available Probes" on page 3-42.
3-12Section 3-5 - Preparing for Installation
Page 71
GE HEALTHCARE
DIRECTION FL091021, REVISION 5 VIVID I N SERVICE MANUAL
3-5-3Component Inspection
After verifying that all the required parts are included in the shipping crate, inspect the system
components using the checklist supplied below. In addition, ensure that all the labels described in
Chapter 1 - Introduction are present, accurate and in good condition, and enter the serial nu mber printed
on the main label into the system installation details card, as described in System Installation Details on
page 3-155.
3-5-3-1Damage Inspection Checklist
Visually inspect the contents of the shipping carton for damage. If any parts are damaged or missing,
contact an authorized GE Service Representative.
A Damage Inspection Checklist for the Vivid i n is provided in Table 3-7 below.
Table 3-7 Damage Inspection Checklist - Vivid i n System
b
StepItemRecommended Procedure
1
Console
2
Control Console
3
Probes
4
LCD Display
5
FansVerify that the system’s cooling fans and peripheral fans are operating.
6
Rear Panel
7
Covers
8
Peripherals
9
AC DC System
Verify that the system is switched OFF and unplugged. Clean the console and control
panel.
Physically inspect the control console for missing or damaged items. Verify the proper
illumination of all the control panel buttons.
Check all probes for wear and tear on the lens, cable, and connector. Look for bent
or damaged pins on the connector and in the connector socket on the unit. Verify that
the EMI fingers around the probe connector socket housing are intact. Check the
probe locking mechanism and probe switch.
Clean the LCD display by gently wiping with a dry, soft, lint-free non-abrasive folded
cloth. Inspect the monitor for scratches and raster burn.
Check the rear panel connectors for bent pins, loose connections and loose or
missing hardware. Screw all the cable connectors tightly to the connector sockets on
the panel. Verify that the labeling is in good condition.
Check that all screws are tightly secured in place, that there are no dents or scratches
and that no internal parts are exposed.
Check and clean the peripherals in accordance with the manufacturer’s directions. To
prevent EMI or system overheating, dress the peripheral cables inside the peripheral
cover.
Check the AC DC unit is not damaged or cracked board and verify that the output
cable is properly secured.
10
Power Cord
Check the power cord for cuts, loose hardware, tire marks, exposed insulation, or any
deterioration. Verify continuity. Replace the power cord, as required.
Chapter 3 - System Setup 3-13
Page 72
GE HEALTHCARE
DIRECTION FL091021, REVISION 5 VIVID I N SERVICE MANUAL
3-5-3-2Front and Side View of the Vivid i n Ultrasound Unit
Figure 3-8 below shows the Vivid i n ultrasound unit components that are visible from the front and side
of the ultrasound unit.
Figure 3-8 Front and Side View of the Vivid i n
Display Monitor: tilts up and down.
1
2Speakers: Two loudspeakers for Doppler sound
Control Panel: Contains the alphanumeric keyboard and the buttons used to operate the
3
ultrasound unit.
4Rear Handle
5Air Vents
6Probe Ports: Two active probe connectors (one for a pencil probe).
GE HEALTHCARE
DIRECTION FL091021, REVISION 5 VIVID I N SERVICE MANUAL
3-5-3-3Rear View of the Vivid i n Ultrasound Unit
Figure 3-9 shows a view of the Vivid i n ultrasound unit rear panel and external peripheral/accessory
connectors
Figure 3-9 View of the Vivid i n Rear Panel
1Two interchangeable USB ports (digital printer, CD-RW and other peripherals)
2Docking connector (currently not in use)
3Port for DC IN (AC Adapter)
4SVGA Output (VCR option or CRT monitor option)
5LAN 10/100 Base-TX Ethernet network connector
6PCMCIA port for PC card.
7Ejection lever for PCMCIA device
Chapter 3 - System Setup 3-15
Page 74
GE HEALTHCARE
DIRECTION FL091021, REVISION 5 VIVID I N SERVICE MANUAL
3-5-4Connection and Usage of the optional Modo Cart
Please refer to the “Vivid i n Cart, User Manual – Supplement to Vivid i n User manual”,
Direction Number FL092096.
3-16Section 3-5 - Preparing for Installation
Page 75
GE HEALTHCARE
DIRECTION FL091021, REVISION 5 VIVID I N SERVICE MANUAL
3-5-5System Voltage Confirmation
3-5-5-1System Voltage Settings
Verify that the Vivid i n is set to the correct voltage.
The Vivid i n voltage settings are found on the base (underside) of the system.
Figure 3-10 Safety Rating Plate - Example
WARNING: CONNECTING THE VIVID I N TO THE WRONG VOLTAGE LEVEL WILL MOST
LIKELY DESTROY THE SCANNER.
Chapter 3 - System Setup 3-17
Page 76
GE HEALTHCARE
DIRECTION FL091021, REVISION 5 VIVID I N SERVICE MANUAL
3-5-6Ensuring Protection from EMI
The Vivid i n unit has been designed to minimize the effects of Electro-Magnetic Interference (EMI).
Many of the covers, shields, and screws are prov ide d prim ar ily to pr ot ect the syste m fr om ima ge
artifacts caused by this interference. For this reason, it is imperative that all covers and hardware are
installed and secured before the unit is put into operation.
Ensure that the system is protected from electromagnetic interference (EMI), as follows:
•Operate the system at least 15 feet away from equipment that emits strong electromagnetic
radiation.
•Operate the system in an area enclosed by walls, floors and ceilings comprised of wood, plaster or
concrete, which help prevent EMI.
•Shield the system when operating it in the vicinity of radio broadcast equipment, if necessary.
•Do not operate mobile phones or other EMI emitting devices in the ultrasound room.
•Verify that all EMI rules listed in the following table are followe d:
Note:The Vivid i n
qualified facilities, in terms of the prevention of radio wave interference. Operation of the
ultrasound unit
television sets situated near the medical equipment.
EMI RuleDetails
Ground the unit.
Be aware of RF sources.
Replace and/or reassemble
all screws, RF gaskets,
covers and cores.
Replace broken RF gaskets.
Do not place labels where
RF gaskets touch metal.
Use GE specified harnesses
and peripherals.
ultrasound unit is approved for use in hosp itals, clinics and other environmentally
in an inappropriate environment can cause electronic interference to ra dios and
Poor grounding is the most likely reason an ultrasound unit will have noisy images. Check
the grounding of the power cord and power outlet.
Keep the unit at least 5m (16.4 ft) away from other EMI sources. Special shielding may be
required to eliminate interference problems caused by high frequency, high powered radio
or video broadcast signals.
After you finish repairing or updating the system, replace all covers and tighten all screws.
Any cable with an external connection requires a magnet wrap at each end. Install the shield
over the front of the card cage. Loose or missing covers or RF gaskets allow radio
frequencies to interfere with the ultrasound signals.
If more than 20% or a pair of the fingers on an RF gasket are broken, replace the gasket.
Do not turn on the unit until any loose metallic part is removed and replaced if needed.
Never place a label where RF gaskets meet the unit. Otherwise, the gap created will permit
RF leakage. In case a label has been found in such a location, move the label to a different
appropriate location.
The interconnect cables are grounded and require ferrite beads and other shielding. Cable
length, material, and routing are all important; do not make any changes that do not meet
all specifications.
Take care with cellular
phones.
Properly address peripheral
cables.
Cellular phones may transmit a 5 V/m signal that causes image artifacts.
Do not allow cables to lie across the top of the card cage or hang out of the peripheral bays.
Loop any peripheral cable excess length inside the peripheral bays or hang on the hooks
provided below the console. Attach the monitor cables to the frame.
Table 3-8 EMI Prevention/ Abatement
3-18Section 3-5 - Preparing for Installation
Page 77
GE HEALTHCARE
DIRECTION FL091021, REVISION 5 VIVID I N SERVICE MANUAL
Section 3-6
Completing the Hardware Installation
The hardware installation procedures described in this section provide instructions and information for
the possible scenarios, as follows:
•Vivid i n scanner - connections direct to the system
•Vivid i n scanner - connections via USB hub
3-6-1Connecting Peripherals
3-6-1-1USB Global Modem
NOTICE
IMPORTANT During Stand-by mode, it is NOT recommended to introduce or remove USB
devices; this may cause the system to lock-up during the boot-up procedure.
The USB Global Modem connection in the different scenarios is explained below.
•Connection directly to the Vivid i n system
The USB Global modem can be powered via the USB port on the Vivid i n system with no need for
additional power. The modem should be connected to the lower USB port on the rear connectors panel.
When connecting the modem, there is no preferen ce as to which cable to connect first. One end of the
telephone cable can be connected to the wall outlet and the other side to the modem, followed by
connecting one end of the USB cable to the modem and the other side to the lower USB port on the
rear connectors panel.
All software drivers for the modem are pre-installed for the specific modem type. One must follow the
i-Linq configuration procedure to complete the check-out process.
WARNING Do not attempt to use a different type of modem either via the PC MCIA or USB ports.
The ultrasound system is an extremely sensitive and complex medical system. Any
unauthorized peripherals may cause system failure or damage!
•Connection via USB hub
This is not an approved configuration and, in most cases, will not work and will cause software
complications.
Chapter 3 - System Setup 3-19
Page 78
GE HEALTHCARE
DIRECTION FL091021, REVISION 5 VIVID I N SERVICE MANUAL
3-6-1-2Wireless Network Adaptor
The Wireless Network adaptor connection is explained below.
•Connection directly to the Vivid i n system
The Wireless Network adaptor should be connected to the PC MCIA socket on the
Vivid i n rear connectors panel.
You can connect the Wireless Network adaptor either once the system is powered ON, or after
shutdown. All software drivers for the Wireless Network are pre-installed for the designated
Wireless Network adaptor only. Follow the procedure described in the "Wireless Network
Configuration" on page 3-67.
WARNING Do not attempt to use a different type of Wireless Network adaptor either via the
PC MCIA or USB ports. The ultrasound system is an extremely sensitive and complex medical
system. Any unauthorized peripherals may cause system failure or damage!
3-20Section 3-6 - Completing the Hardware Installation
Page 79
GE HEALTHCARE
DIRECTION FL091021, REVISION 5 VIVID I N SERVICE MANUAL
3-6-1-3DVD/CD-RW Drive
NOTE:This procedure describes replacement of the DVD/CD-RW drive.
NOTE:This procedure describes replacement of the DVD/CD-RW drive. The older-type DVD/CD-RW drive
(Addonics) is no longer available.
NOTE:For instructions regarding older-type DVD/CD-RW drives (Addonics), refer to Section 3-6-1-4 on page 3-24.
NOTICE
NOTICE
IMPORTANT By default, the DVD's AC adaptor provided is not medical grade therefore it is
required to disconnect the DVD from the Vivid i n system while scanning a patient.
A medical grade AC adaptor may be ordered (P/N 2416318- 2) in c ase ther e is a ne ed to con nect
the DVD to the scanner while scanning.
IMPORTANT: Use only CD-R, DVD-R or DVD+R media from a reputable vendor and only
high-grade media types.
IMPORTANT: For BT06 systems, it is not possible to use DVD+R media (only CD-R or DVD-R).
USB Cable
DVD/CD-RW Drive
Emergency Release Pin
Protective Pouch
Figure 3-11 Samsung DVD/CD-RW Drive
NOTE:The Samsung DVD-R/RW drive requires BT06 software installation (6.1.0 build 109 or h igher)
in order to support the media finalization process with the Vivid i n application. However, this
drive may be used with a lower software version for the purpose of upgrading the software to
the required version. The BT06 software 6.1.0 build 109 Software CD can be obtained by
ordering the Vivid i n BT06 SW (M4) CD kit - P/N: 2421595 . For instructions, refer to "Software
Loading" on page 8-56.
Chapter 3 - System Setup 3-21
Page 80
GE HEALTHCARE
DIRECTION FL091021, REVISION 5 VIVID I N SERVICE MANUAL
3-6-1-3 DVD/CD-RW Drive (cont’d)
.
NOTICE
IMPORTANT - Always follow the safety instructions provided with the Samsung DVD-R/RW drive and
supplied by the manufacturer.
The Samsung DVD-R/RW drive, which is designed to operate with the Vivid i n system using the USB
port, combines the functions of a DVD or CD recorde r, and a DVD or CD play er .
This DVD device may be used on the Vivid i n system for the following purposes:
•To format and label each media before use
•To export Exams from the Vivid i n system to media
•To import Exams from media to the Vivid i n system
•To perform the Database Backup pr ocedure - in order to store back-up copies of patient Exams
(data Archive) and also to create free space on the local hard disk.
•To export Exams in MPEGVue mode, whereby the me dia can be viewed on any external PC
•To load or upgrade system software, and for the use of GE Healthcare Field Engineers when
performing various other funct ion s.
Note:For more details, refer to the latest version of the Vivid i n (P/N 2378958-100). Refer in
particular to the Archive chapter.
3-22Section 3-6 - Completing the Hardware Installation
Page 81
GE HEALTHCARE
DIRECTION FL091021, REVISION 5 VIVID I N SERVICE MANUAL
3-6-1-3-1DVD/CD-RW Drive Replacement on Vivid i n Systems
NOTE:If you are replacing a DVD/CD-RW Drive and the current system software version is below 6.1.0
build 109, refer to the Note on pa ge 3-21.
NOTE:When replacing a DVD/CD-RW Drive on a stand-alone Vivid i n syste m, it is necessary to order
the following parts:
Table 3-9 Parts to be Ordered
Part NoPart Description
2421141-2Samsung DVD/CD-RW Drive
2421169
a
a. the 5V PS is supplied without a power cable; the appropriate cable must
also be ordered.
5V Power Supply (medical grade) with universal connector
(suitable for the Samsung DVD/CD-RW Drive)
NOTICE
IMPORTANT - the following steps must be performed in the order stated.
(If the USB cable is connected first while power to the Vivid i n is ON, the drive will not be recognized.
In this event, it will be necessary to re-boot the system).
1.) Connect the 5V power supply output cable to the power port on the Samsung DVD-R/RW drive rear
panel.
2.) Connect the USB cable (supplied) between the Vivid i n rear panel and the Samsung DVD-R/RW
drive rear panel.
CAUTION: Do not disrupt power to the Samsung DVD-R/RW drive from its AC/DC adapte r while
the USB cable is connected to the Vivid i n system.
In this event, the system will stop recognizing the CD/DVD driv e or any oth er device con nected
to this USB port. A complete system re-boot will then be required to make the USB port
operational again.
Chapter 3 - System Setup 3-23
Page 82
GE HEALTHCARE
DIRECTION FL091021, REVISION 5 VIVID I N SERVICE MANUAL
3-6-1-4DVD/CD-RW Drive (Older Type)
NOTE:This procedure describes replacement of the older-type DVD/CD-RW drive (Addonics) only.
Figure 3-12 DVD/CD-RW Drive
The DVD/CD-RW drive connection in the different scenarios is explained below.
•No Cart - connection directly to the Vivid i n system or via USB Hub
The DVD/CD-RW Drive can be connected to either of the USB ports (upper or lower) on the rear
connectors panel.
Note:The DVD/CD-RW power switch on the rear of the DVD must be set in the External position,
as shown below:
ExternalPower OFFInternal
When connecting directly to the system, it is necessary to use the additional power supply supp lied
with the DVD/CD-RW Drive.
3-24Section 3-6 - Completing the Hardware Installation
Page 83
GE HEALTHCARE
DIRECTION FL091021, REVISION 5 VIVID I N SERVICE MANUAL
3-6-1-4 DVD/CD-RW Drive (Older Type) (cont’d)
Note:When plugging the DVD's AC adaptor directly into the mains supply, there is no isolation
from leakage currents since the USB connector is not isolated. It is therefore necessary to
use an isolation transformer to isolate the mains power supplied to the DVD’s AC adaptor.
If an isolation transformer is not available, the DVD must be disconnected from the
Vivid i n system while scanning a patient.
The DVD/CD-RW Drive can be connected once the system is powered ON, or after shutdown. All
software drivers for the DVD/CD-RW Drive are pre-configured and installed designated to be used
with the specific DVD/CD-RW supplied with the system.
WARNING Do not attempt to use a different type of mass storage device other than the
DVD/CD-RW Drive supplied with the system. The ultrasound system is an extremely sensitive and
complex medical system. Any unauthorized peripherals may cause system failure or damage!
Chapter 3 - System Setup 3-25
Page 84
GE HEALTHCARE
DIRECTION FL091021, REVISION 5 VIVID I N SERVICE MANUAL
3-6-1-5MOD Drive
NOTICE
NOTICE
IMPORTANT Do not connect the MOD Drive to the Vivid i n system while scanning is in progress!
IMPORTANT During Stand-by mode, it is NOT recommended to introduce or remove USB
devices; this may cause the system to lock-up during the boot-up procedure.
Figure 3-13 MOD Drive
The MOD drive connection in the different scenarios is explained below.
•Connection directly to the Vivid i n system
The MOD Drive can be connected to either of the USB ports (upper or lower) on the rear connectors
panel. This connection must be via a USB-to-SCSI adaptor; one side of the adaptor be ing
connected to the Vivid i n USB port, and the other side to the rear of the MOD drive.
When connecting the MOD drive directly to the system, it is necessary to use an additional power
source. As a safety precaution, this must be done via an isolation transformer.
The MOD Drive may be connected once the system is powered ON, or after shutdown. All software
drivers for the MOD Drive are pre-configured and installed desig nated to be used with the specific
MOD drive supplied by GE Healthcare.
WARNING Do not attempt to use a different type of mass storage device other than the
MOD Drive supplied with the system. The ultrasound system is an extremely sensitive and
complex medical system. Any unauthorized peripherals may cause system failure or damage!
•Connection via USB hub
As stated above. Safety considerations must be taken into account. The MOD Drive must be
powered via an isolation transformer. Either USB port may be used to communicate with the
MOD device.
3-26Section 3-6 - Completing the Hardware Installation
Page 85
GE HEALTHCARE
DIRECTION FL091021, REVISION 5 VIVID I N SERVICE MANUAL
3-6-1-6USB Memory Stick
NOTICE
IMPORTANT During Stand-by mode, it is NOT recommended to introduce or remove USB
devices; this may cause the system to lock-up during the boot-up procedure.
The USB Memory Stick (disk-on-key) connection in the different scenarios is explained below.
•Connection directly to the Vivid i n system
The USB Memory Stick may be connected to either of the USB ports (upper or lower) on the rear
connectors panel.
The USB Memory Stick can be connected once the system is powered ON, or after shutdown. All
software drivers for the USB Memory Stick are pre-installed on the system. The system is capable
of detecting and working with other types of memory sticks. However, it is highly recommended to
use the memory stick that is provided and approved by GE Healthcare. Other memory sticks may
be sensitive to EMC interference - this may affect system performance and/or image quality.
•Connection via USB hub
As above. If applicable, the USB hub must be powered via the isolation transformer.
Chapter 3 - System Setup 3-27
Page 86
GE HEALTHCARE
DIRECTION FL091021, REVISION 5 VIVID I N SERVICE MANUAL
3-6-1-7Connecting the Black & White Printer
NOTICE
IMPORTANT During Stand-by mode, it is NOT recommended to introduce or remove USB
devices; this may cause the system to lock-up during the boot-up procedure.
Figure 3-14 Black and White Printer
The B/W Printer connection in the different scenarios is explained below.
NOTE:After physical connection of the printer to the Vivid i n system, to configure the appropriate hot keys to
activate printing direct from the control panel, follow the instr uctions in "Configuring Hot Keys to Activate
Printing Direct from the Control Panel" on page 3-30.
•Connection directly to the Vivid i n system
The B/W Printer can be connected to either of the USB ports (upper or lower) on the rear connectors panel.
When connecting the B/W Printer directly to the system, it is necessary to use an additional power
source. As a safety precaution, this must be done via an isolation transformer.
The B/W Printer can be connected once the system is powered ON, or after shutdown. All software
drivers for the B/W Printer are pre-configured and installed designated to be used with the specific
B/W Printer supplied by GE Healthcare.
WARNING Do not attempt t o use a different type of B/W Printer (brand or model) other than the
B/W Printer provided by GE Healthcare. The ultrasound system is an extremely sensitive and
complex medical system. Any unauthorized peripherals may cause system failure or damage!
•Connection via USB hub
As stated above. Safety considerations must be taken into account. The B/W Printer must be
powered via an isolation transformer. Either USB port may be used to communicate with the device.
NOTE:Once the printer is connected, perform the following functionality check:
•"B/W and Color Printer Test" on page 4-23 should be performed
3-28Section 3-6 - Completing the Hardware Installation
Page 87
GE HEALTHCARE
DIRECTION FL091021, REVISION 5 VIVID I N SERVICE MANUAL
3-6-1-8Connecting the Digital Color Printer
NOTICE
IMPORTANT During Stand-by mode, it is NOT recommended to introduce or remove USB
devices; this may cause the system to lock-up during the boot-up procedure.
NOTE:After physical connection of the printer to the Vivid i n system, to configure the appropriate hot keys to
activate printing direct from the control panel, follow the instructions in "Config uring Hot Keys to Activate
Printing Direct from the Control Panel" on page 3-30.
Figure 3-15 Digital Color Printer
The Digital Color Printer connection in the different scenarios is explained below.
•Connection directly to the Vivid i n system
The Digital Color Printer can be connected to either of the USB ports (upper or lower) on the rear
connectors panel.
When connecting the Digital Color Printer directly to the system, it is necessary to use an additional
power source. As a safety precaution, this must be done via an isolation transformer.
The Digital Color Printer can be connected once the system is powered ON, or after shutdown. All
software drivers for the Digital Color Printer are pre-configured and installed designated to be used
with the specific Digital Color Printer supplied by GE Healthcare.
WARNING Do not attempt to use a different type of Digital Color Printer (brand or model) other
than the Digital Color Printer provided by GE Healthcare. The ultrasound system is an extremely
sensitive and complex medical system. Any unauthorized peripherals may cause system f ailure
or damage!
•Connection via USB hub
Asstated above. Safety conside rations must be t aken into account. T he Digital Color Printer
must be powered via an isolation transformer. Either USB port may be used to communicate
with the device.
NOTE:Once the printer is connected, perform the following functionality check:
•"Color Printer Test" on page 4-23
Chapter 3 - System Setup 3-29
Page 88
GE HEALTHCARE
DIRECTION FL091021, REVISION 5 VIVID I N SERVICE MANUAL
3-6-1-9Configuring Hot Keys to Activate Printing Direct from the Control Panel
1.) Press Config (or F2).
2.) Select Connectivity (in the lower part of window).
3.) Select the Additional Outputs tab.
Figure 3-16 Configuring Control Panel Hot Keys for Printing
4.) From the Button drop-down menu (see Figure 3-16), select to configure either the Print button or
Alt+Print to activate a variety of outputs.
5.) Use the left or right arrows below to add any of the available options (listed on the left) as Outputs
(listed on right).
6.) From the listed Outputs, select (highlight) the required output, right-click and select Properties to
configure the output device accordingly.
Printing may now be activated direct from the Vivid i n Control Panel, using the hot key you have
configured for this purpose.
7.) Proceed to perform the "B/W and Color Printer Test" on page 4-23.
3-30Section 3-6 - Completing the Hardware Installation
Page 89
GE HEALTHCARE
DIRECTION FL091021, REVISION 5 VIVID I N SERVICE MANUAL
3-6-1-10Connecting the DeskJet Color Printer
NOTICE
NOTICE
IMPORTANT Do not connect the DeskJet Color Printer to the Vivid i n system while scanning is
in progress!
IMPORTANT During Stand-by mode, it is NOT recommended to introduce or remove USB
devices; this may cause the system to lock-up during the boot-up procedure.
Figure 3-17 DeskJet Color Printer
The DeskJet Color Printer connection in the different scenarios is explained below.
•Connection via LAN
There are a few DeskJet Color Printers that can be conne cted to the system via LAN - refer to "Add
Printer" on page 3-57. The DeskJet Color Printer can be connected to an external, non-isolated
power source.
WARNING Do not attempt to use a different type of DeskJet Color Printer (brand or model) other
than the DeskJet Color Printer provided by GE Healthcare. The ultrasound system is an
extremely sensitive and complex medical system. Any unauthorized peripherals may cause
system failure or damage!
Chapter 3 - System Setup 3-31
Page 90
GE HEALTHCARE
DIRECTION FL091021, REVISION 5 VIVID I N SERVICE MANUAL
3-6-1-10 Connecting the DeskJet Color Printer (co nt’d)
•Connection directly to the Vivid i n system
This only applies to HP 460 models which are supplied pre-installed with the appropriate software
drivers. The DeskJet Color Printer can be con nected to either of the USB ports ( upper or lower) on
the rear connectors panel.
When connecting directly to the system it is necessary to use an additional power source. As a safety precaution, this must be done via an isolation transformer.
The DeskJet Color Printer can be connected once the system is powered ON, or after shutdown.
WARNING Do not attempt to use a different type of DeskJet Color Printer (brand or model) other
than the DeskJet Color Printer provided by GE Healthcare. The ultrasound system is an extremely
sensitive and complex medical system. Any unauthorized peripherals may cause system failure
or damage!
•Connection via USB hub
As stated above. Safety considerations must be taken into account . The DeskJet Color
Printer must be powered via an isolation transformer. Either USB port may be used to
communicate with the device.
3-32Section 3-6 - Completing the Hardware Installation
Page 91
GE HEALTHCARE
DIRECTION FL091021, REVISION 5 VIVID I N SERVICE MANUAL
3-6-1-11Connecting the VCR
Figure 3-18 Video Casette Recorder
There are two types of VCR - PAL or NTSC approved by GE Healthcare. The VCR connection in the
different scenarios is explained below.
•Direct Connection to the Vivid i n system
A Mitsubishi VCR is used in this configuration - use the type that is suited to your region. The video
signal is received through a VGA-to-video converter.
1) Connect one end of the VGA cable to the VCR OUTPUT located on the Vivid i n rear connector
panel (see note below).
Note:When performing step 1, it may be necessary to use the VGA adaptor (P/N 2420728 - see
Figure 3-19) to connect the end of the VGA cable to the Vivid i n rear connector panel.
Since some VGA plugs have thicker housing, they do not fit properly into the
Vivid i n VGA connector socket on the rear panel (because the plug is not centrally a ligned
within the frame). In this event, use of the VGA adaptor will eliminate the possibility of
causing damage to the BEP by the use of unnecessary force.
Figure 3-19 VGA Adaptor
2) Connect the other end of the VGA cable to the VGA INPUT on the video co nverter - see Figure 3-20.
Figure 3-20 Video Converter
Chapter 3 - System Setup 3-33
Page 92
GE HEALTHCARE
DIRECTION FL091021, REVISION 5 VIVID I N SERVICE MANUAL
3-6-1-11 Connecting the VCR (cont’d)
3.) Connect one end of the S-Video cable to the S-Video OUTPUT on the video converter.
4.) Connect the other end of the S-Video cable to the VCR S-Video INPUT at the rear of the VCR.
Note:Dip switches are located beneath the video converter, as shown in Figure 3-21 below.
Dip Switches
Figure 3-21 Video Converter Dip Switches
5.) Follow the manufacturer’s guidelines to set the switches to the correct positions, as shown in Figure
3-22 below.
Figure 3-22 Video Converter Dip Switch Positions
3-34Section 3-6 - Completing the Hardware Installation
Page 93
GE HEALTHCARE
DIRECTION FL091021, REVISION 5 VIVID I N SERVICE MANUAL
3-6-1-11 Connecting the VCR (cont’d)
6.) When all power and signal cables are conne cted to the system, VCR, and video converter, proceed
as follows (in the order stated):
a.) Press the power ON switch on the VCR.
b.) Press the power ON switch on the video converter.
c.) Turn ON power to the Vivid i n system.
7.) Press the Input Select button (see Figure 3-20 on page 3-33 ) on the video converter three times
in order to select VGA Input as your signal source.
The LED located on the video converter front panel will be illuminated, accordingly.
8.) Press on the three buttons <CTRL> <ALT> <Page Erase> simultaneously.
The following screen will appear.
Figure 3-23 Intel® Dual Display Clone
9.) Select the Intel® Dual Display Clone icon and press OK - see Figure 3-23.
Chapter 3 - System Setup 3-35
Page 94
GE HEALTHCARE
DIRECTION FL091021, REVISION 5 VIVID I N SERVICE MANUAL
3-6-1-11 Connecting the VCR (cont’d)
NOTE:If the VGA cable is not properly connected to the rear panel, or in the event that the
scan-converter device is not switched ON, you may see a single icon displayed, as shown in
Figure 3-24.
Figure 3-24 Single Icon
10.)Confirm Desktop Change by clicking OK.
The dialog box will be closed and the VGA output will become activated, see Figure 3-25
Figure 3-25 Confirm Desktop Change
NOTE:When the VGA cable is disconnected and reconnected while the system is ON, the VGA output
is maintained.
NOTE:If system is shut-down to standby or Full-shut-down while the VGA device is connected
continuously, the VGA output will appear after the system is turned ON
NOTE:If the system is shut-down to Standby or Full-shut-down and the VGA device is disconnected,
the VGA output will not appear after power is turned ON, and the above procedure will have to
be repeated.
11.)Proceed to perform the VCR Test on 4-3-2-1 "VCR Test" on page 4-23.
3-36Section 3-6 - Completing the Hardware Installation
Page 95
GE HEALTHCARE
DIRECTION FL091021, REVISION 5 VIVID I N SERVICE MANUAL
3-6-2Charging the Battery
NOTE:The CH5000GE is a stand-alone desktop smart-battery charger with the added ability to recalibrate th e
fuel gauge when detected to be out of calibration. The charger is specifically designed to operate with
the Vivid i n smart-battery (P/N 2378963-2).
The charger is supplied in a carton containing the following items:
•CH5000 desktop charger/calibrator
•24V 2.5A AC-adapter power supply, universal mains input
•Instruction Manual
CAUTION: Always observe the following safety precautions:
•Do not expose the charger or power supply to water or conductive liquids (this is not a sealed case).
•Do not open the charger or power supply case. There are no user-serviceable parts inside.
•Do not cover the fan exhaust or obstruct the airflow, as this will cause overheating.
•Use only the manufacturer’s 24V 2.5A power supply.
•Place the charger in a cool location, away from exte rnal he at sources.
•During recalibration, the battery connector and base of the charger may become warm.
3-6-2-1Setting up the Charger
1) Place the charger unit on a flat, level surface away from sources of heat and moisture.
2) Plug the DC connector from the power supply into the back of the charger.
Figure 3-26 Battery Charger Unit
3.) Connect the power supply to the mains AC supply, using the cable supplied.
Chapter 3 - System Setup 3-37
Page 96
GE HEALTHCARE
DIRECTION FL091021, REVISION 5 VIVID I N SERVICE MANUAL
3-6-2-2Charging the Battery
1) Place the battery into the battery bay making sure that the 5-way connector is fully seated.
The LEDs in the status window (see Figure 3-26) will provide status information as detailed in
Table 3-10 below, and the charger will automatically begin to charge the battery.
Table 3-10 Battery Charger LED Indicators and Status
LED IndicatorStatus
Green FlashingBattery charging
Green Solid Battery fully charged
Blue flashingBattery in Calibration mode
Blue SolidBattery fuel gauge calibrated
Red FlashingBattery fuel gauge in need of recalibration
Red SolidError
•Recharge Time:
The time for a full charge from 0% to 100% state of charge is typically 4.5 hours.
•Recalibration Time
The recalibration cycle begins by discharging away any residual capacity. Then a calibration
charge is delivered to the battery. This is followed by a calibration discharge. Finally the battery
is given a regular charge. A calibration cycle will be faster if the battery is fully discharged to
begin with.
The recalibration time is approximately 9 hours for an empty battery, and approximately
14 hours for a fully-charged battery.
Note:Calibration is initiated each time the button is pressed, so it is not recommended to press
the recalibration button part way through th e re ca libr ati on cycle .
During the charging procedure, the green LED will be seen flashing.
At the end of this procedure, when the battery is fully charged, the solid green LED will be
illuminated in the status window, indicating a fully charged battery.
2.) Remove the battery from the charger unit battery bay and install it in the
Vivid i n.
3-38Section 3-6 - Completing the Hardware Installation
Page 97
GE HEALTHCARE
DIRECTION FL091021, REVISION 5 VIVID I N SERVICE MANUAL
3-6-2-3Battery Recalibration
NOTE:For more details on recalibration and an explanation of why it is necessary, refer to Recalibration on
page 5 - 82.
If the battery is in need of “fuel gauge” recalibration, the red LED will flash upon insertion of the battery.
This indicator provides feedback to the user on the accuracy of the fuel gauge and avoids unnecessary
battery calibration cycles.
The user has the option to calibrate the fuel gauge and charge the b attery, or to only charge the battery.
This option is given because a recalibration cycle is longer than a charge cycle.
1) To recalibrate the battery, press the calibrate button on the front of the battery charger unit.
No action is required if only a recharge is required, as the charger will automatically begin to charge
the battery.
The blue calibration LED will flash to indicate that the battery is undergoing the recalibration cycle.
There may be a short delay before the calibration begins. During calibratio n the discharge resistors
will heat up and the fan will operate to maintain temperature within acceptable limits.
At the end of this procedure, when the battery is fully charged and the battery fuel gauge is
calibrated, the solid green and blue LEDs will be illuminated in the status window, indicating a fully
charged, fully calibrated battery.
2.) Remove the battery from the charger unit battery bay and insta ll it in the Vivid i n.
Note:The most common cause of calibration failure is overheating of the pack during discharge.
For this reason, keep the charger away from direct sunlight or heat sources.
Chapter 3 - System Setup 3-39
Page 98
GE HEALTHCARE
DIRECTION FL091021, REVISION 5 VIVID I N SERVICE MANUAL
3-6-3Connecting Probes
The Vivid i n ultrasound unit operates with various types of probes that are used for scanning patients,
including flat phased, convex and linear electronic array probes. Once connected, the probes can be
selected for different applications.
Probe connectors on the Vivid i n unit are as follows:
•Two active probe connectors (one for a pencil probe) on the right side of the unit
Probes can be connected or changed any time, as described belo w, regardless of whether the system
is powered ON or OFF.
CAUTION: Handle the probe gently while connecting and disconnecting.
Do NOT touch the patient and any of the connectors on the ultrasound unit simultaneously, including
ultrasound probe connectors.
1) Place the probe’s carrying case on a stable surface and open the case.
2) Inspect the probe socket to verify that it is free of debris.
3) Carefully remove the probe and unwrap the probe cord.
NOTICE
DO NOT allow the probe head to hang freely. Impact to the probe head could result in irreparable
damage.
4) Press the probe connector locking lever downwards (refer to Figure 3-27).
Figure 3-27 Probe Connection Locking Lever
3-40Section 3-6 - Completing the Hardware Installation
Page 99
GE HEALTHCARE
DIRECTION FL091021, REVISION 5 VIVID I N SERVICE MANUAL
3-6-3Connecting Probes (cont’d)
5) Align the connector with the probe port and carefu lly push into place, as shown in Figure 3-28.
Figure 3-28 Connecting Probe to the Vivid i n
6) Press the connector locking lever upwards to the full vertical position to lock in place, as shown in
Figure 3-29.
Figure 3-29 Securing Probe with Locking Lever
7) Carefully position the probe cord so that it is free to move and is not resting on the floor.
When the probe is connected, it is automatically activated.
CAUTION: Do not bend the probe cable acutely.
Fault conditions can result in electric shock hazard. Do not touch the surface of probe connectors which
are exposed when the probe is remove d. Do not touc h the patient when connecting or disconnecting a
probe.
NOTE:It is not necessary to turn OFF power to connect or disconn e ct a pr ob e.
Chapter 3 - System Setup 3-41
Page 100
GE HEALTHCARE
DIRECTION FL091021, REVISION 5 VIVID I N SERVICE MANUAL
3-6-3-1Available Probes
For a list of the available probes, please see: Section 9-7 "Probes" on page 9-8.
3-42Section 3-6 - Completing the Hardware Installation
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.