GE Vivid 7 User manual

g
Technical Publications
Vivid 7/EchoPAC PC
Dimension - version 7.x.x
0470
GE Medical Systems
4D and Multi-plane Imaging
GEVU #: FD092081
MHLW No: 21300BZY00416000
Operating Documentation
Copyright © 2007 By General Electric Co.
g
GE Medical Systems
MANUAL STATUS
FD092081-01
01/08/2007
COMPANY DATA GE VINGMED ULTRASOUND A/S
© GE Medical Systems. All rights reserved. No part of this manual may be reproduced, stored in a retrieval system, or transmitted, in any form or by any means, electronic, mechanical, photocopying, recording, or otherwise, without the prior written permission of GE Medical Systems.
Strandpromenaden 45, N-3191 Horten, Norway
Tel.: (+47) 3302 1100 Fax: (+47) 3302 1350

Table of Contents

Table of Contents Introduction
4D imaging ..................................................................................... 1
Multi-plane Imaging....................................................................... 2
Measurement and analysis........................................................... 2
Multi-plane stress echo................................................................. 2
Remark concerning the 3V probe ................................................ 2
Important........................................................................................ 3
Conventions used in this manual ................................................ 4
Chapter 1 4D Imaging
Table of Contents
Introduction.................................................................................... 6
4D mode overview - Vivid 7 .......................................................... 7
Volume rendering mode screen............................................ 7
Slice mode screen ................................................................ 8
4D mode controls.................................................................. 9
Assigned rotaries and keys................................................. 10
4D mode assigned controls ................................................ 11
Additional 4D mode assigned controls ............................... 14
Soft menu controls.............................................................. 16
Trackball controls................................................................ 17
Display controls .................................................................. 19
Using 4D mode - Vivid 7.............................................................. 21
Parasternal view acquisition ............................................... 21
Apical view acquisition........................................................ 22
Full volume acquisition ....................................................... 23
Rotating/Translating the 4D image ..................................... 25
Zooming.............................................................................. 26
Cropping ............................................................................. 26
9 Slice................................................................................. 29
4D mode overview - EchoPAC PC ............................................. 32
4D/Multiplane Imaging User's Manual 1 FD092081-01
Table of Contents
Volume rendering mode screen ..........................................32
Slice mode screen...............................................................33
4D mode control panel ........................................................34
Display controls................................................................... 36
Cropping..............................................................................37
Working with 4D acquisitions - EchoPAC PC ...........................39
9 Slice..................................................................................39
4D LV Volume application ...........................................................41
Starting the 4D LV Volume application — Vivid 7 ...............41
Starting the 4D LV Volume application — EchoPAC PC ....41
Chapter 2 4D Color Flow Imaging
Introduction ..................................................................................44
4D Color Flow mode overview - Vivid 7 .....................................45
Color Flow Volume rendering mode screen ........................45
Color Flow Slice mode screen.............................................46
4D Color Flow mode controls .............................................. 47
Assigned rotaries and keys .................................................48
4D Color Flow mode assigned controls...............................49
Additional 4D mode assigned controls ................................50
Soft menu controls ..............................................................51
Trackball controls ................................................................52
Display controls................................................................... 52
Using 4D Color Flow mode - Vivid 7...........................................54
Real time 4D Color flow acquisition.....................................54
Full volume Color Flow acquisition......................................55
6 Slice..................................................................................57
4D Color Flow mode overview - EchoPAC PC ..........................60
Volume rendering mode screen ..........................................60
Slice mode screen...............................................................61
4D mode control panel ........................................................62
Chapter 3 Multi-plane imaging
Introduction ..................................................................................66
2 4D/Multiplane Imaging User's Manual
FD092081-01
Multi-plane mode overview - Vivid 7.......................................... 67
Bi-plane mode screen......................................................... 67
Tri-plane mode screen........................................................ 68
Multi-plane mode controls................................................... 69
Using Multi-plane mode imaging - Vivid 7 ................................ 77
Scan plane rotation............................................................. 78
Tilting scan plane 2............................................................. 78
Zooming.............................................................................. 79
Multi-plane mode overview - EchoPAC PC ............................... 82
Bi-plane mode screen......................................................... 82
Tri-plane mode screen........................................................ 83
The multi-plane control panel ............................................. 84
Working with multi-plane acquisitions - EchoPAC PC ............ 85
Chapter 4 Measurements and Analysis
Introduction.................................................................................. 88
Left ventricular volume measurements..................................... 89
Tri-plane acquisition............................................................ 89
Full volume acquisition ....................................................... 92
Rotation of the Volume reconstruction................................ 93
Bi-plane acquisition............................................................. 94
TSI surface model........................................................................ 96
To edit the sampling path ................................................... 97
Quantitative analysis................................................................... 98
Starting Quantitative analysis from a multi-plane
acquisition........................................................................... 98
Table of Contents
Chapter 5 Multi-plane Stress Echo
Introduction................................................................................ 102
Creating a Multi-plane stress test template ............................ 103
Launching the Template editor ......................................... 104
Stress Template setup...................................................... 104
Stress test acquisition .............................................................. 106
Baseline acquisitions ........................................................ 106
4D/Multiplane Imaging User's Manual 3 FD092081-01
Table of Contents
Low dose and Peak dose level acquisitions......................109
Image analysis............................................................................112
Index
4 4D/Multiplane Imaging User's Manual
FD092081-01

Introduction

This user manual describes the 4D and Multi-plane Imaging applications for the Vivid 7 Dimension and EchoPACPCDimension.
The 2D matrix probe 3V enables real time volume rendering, simultaneous bi-plane or tri-plane data acquisition (multi-plane acquisition).
Introduction
1. Real time Volume acquisition (4D imaging)
2. Real time Bi-plane acquisition
3. Real time Tri-plane acquisition

4D imaging

4D imaging enables real time acquisition and rendering of volume ultrasound data. Free rotation of the 3-dimensional image combined with the zoom function and 4D image optimization controls enhance spatial understanding of the anatomical structure and function of the heart.
4D imaging is available in combination with B-Mode only.
4D/Multiplane Imaging User's Manual 1 FD092081-01
Introduction

Multi-plane Imaging

Multi-plane imaging displays two (Bi-plane) or three (Tri-plane) rotated scan planes acquired simultaneously. Free rotation, tilting (in Bi-plane) of the scan planes and zoom enable the investigation of anatomical structures from different angles.
Multi-plane imaging is available from B-Mode, Color flow mode and TVI related modes.
If not otherwise specified, the term multi-plane means either Bi-plane or Tri-plane.

Measurement and analysis

All cardiac measurements available in B mode are also available in 4D and Multi-plane.
The 4D and Multi-plane modes enable the creation of a left ventricular volume reconstruction based on contours drawn from three cross sections at both end-systole and end-diastole with calculation of end-systolic and end-diastolic volumes and ejection fraction.

Multi-plane stress echo

Combined with specially designed Stress echo protocols, Multi-plane mode enables faster stress tests as several views can be acquired simultaneously.

Remark concerning the 3V probe

The 3V probe temperature will increase during extensive use, due to heating from the probe electronics. If the probe reaches its temperature limit, this will be detected by the system and scanning will stop. Currently, this limit is set to 42.3 degrees Celsius, which complies with regulatory instructions relating to patient comfort and safety. This is reported with a dialog box and the user has to wait until the temperature is below acceptable limits before scanning can continue. This does not in any way indicate probe malfunction. The user has to unfreeze to start scanning again.
In order to reduce the chance for full freeze due to heating, the system will reduce power slightly in a temperature interval just
2 4D/Multiplane Imaging User's Manual
FD092081-01

Important

Introduction
below the full freeze limit. Changes to the power level are reported in the status bar as Setting power level down and Setting power level up.
The recommended use of this probe for live 4D and full volume 4D scanning is to enter full freeze whenever the probe is not used for imaging. For other modalities no special actions are required.
Read and understand all instructions in the Vivid 7 and EchoPAC PC User manuals before attempting to use the devices.
4D/Multiplane Imaging User's Manual 3 FD092081-01
Introduction

Conventions used in this manual

Keys and button, on the control panel or alphanumeric keyboard are indicated by over and underlined text (ex. refers to the 2D mode key)
Bold type, describes button names on the screen.
Italic type: describes program windows, screens and dialogue
boxes.
Icons, highlight safety issues as follow:
Indicates that a specific hazard exists that, given inappropriate conditions or actions, will cause:
DANGER
WARNING
• Severe or fatal personal injury
• Substantial property damage
Indicates that a specific hazard exists that, given inappropriate conditions or actions, will cause:
• Severe personal injury
• Substantial property damage
2D
CAUTION
Indicates that a potential hazard may exist that, given inappropriate conditions or actions, can cause:
• Minor injury
• Property damage
4 4D/Multiplane Imaging User's Manual
FD092081-01

4D Imaging

Chapter 1 4D Imaging
• Introduction ................................................................................... ...... 6
• 4D mode overview - Vivid 7 .......................................................... ...... 7
• Volume rendering mode screen ................................................... 7
• Slice mode screen ........................................................................ 8
• 4D mode controls ......................................................................... 9
• Using 4D mode - Vivid 7 ............................................................... .... 21
• Parasternal view acquisition ....................................................... 21
• Apical view acquisition ............................................................... 22
• Full volume acquisition ............................................................... 23
• Rotating/Translating the 4D image ............................................. 25
• Zooming ..................................................................................... 26
• Cropping ..................................................................................... 26
• 9 Slice ........................................................................................ 29
• 4D mode overview - EchoPAC PC ............................................... .... 32
• Volume rendering mode screen ................................................. 32
• Slice mode screen ...................................................................... 33
• 4D mode control panel ............................................................... 34
• Display controls .......................................................................... 36
• Working with 4D acquisitions - EchoPAC PC ............................ .... 39
• 9 Slice ........................................................................................ 39
• 4D LV Volume application ............................................................ .... 41
• Starting the 4D LV Volume application — EchoPAC PC ........... 41
4D/Multiplane Imaging User's Manual 5 FD092081-01
4D Imaging

Introduction

The 3V probe enables real time acquisition of volume ultrasound data. Free rotation of the three-dimensional image combined with zooming and 4D image optimization controls enhance spatial understanding of the anatomical structure and function of the heart.
Two display modes are available, Volume rendering mode for three dimensional scanning and Slice mode for measurements and volume reconstruction purpose.
4D imaging is available from B mode only.
6 4D/Multiplane Imaging User's Manual
FD092081-01

4D mode overview - Vivid 7

Volume rendering mode screen

The Volume rendering mode displays a volume rendering and 2D images from two perpendicular cut-planes.
4D Imaging
1. Volume rendering display from cut-plane 1 (yellow). The volume rendering may be adjusted by rotating and translating the cut-plane 1.
2. Cut-plane 2 (white): 2D image in the azimuth plane.
3. Cut-plane 3 (green): 2D image in the elevation plane.
4. Orientation window: displays a three-dimensional model with cut-planes position and orientation.
5. Color coded cut-plane markers indicate the other cut-planes position relative to the displayed cut-plane.
6. View direction marker.
7. Soft menu controls (see page 16)
8. Trackball functions (see page 17)
Figure 1-1: The 4D screen (Volume rendering)
4D/Multiplane Imaging User's Manual 7 FD092081-01
4D Imaging

Slice mode screen

The Slice mode displays three cut-planes. The cut-planes can be rotated and translated independently of each other. This mode is use to perform measurements and volume reconstruction based on contour traces done on several cut-planes.
1. Cut-plane 1 (yellow)
2. Cut-plane 2 (white)
3. Cut-plane 3 (green)
4. Color coded cut-plane markers indicate the position of the other cut-planes relative to the displayed cut-plane.
5. Orientation window: displays a three-dimensional model with cut-planes position.
6. View direction marker.
7. Trackball functions (see page 17)
Figure 1-2: The 4D screen (Slice mode)
8 4D/Multiplane Imaging User's Manual
FD092081-01

4D mode controls

Alt. Alt. Alt. Alt.
4D Imaging
3
4
Width
Update /
ck
ll
a
a
Tr
B
Menu
1. 4D key
2. Multiplane key (see page 65)
3. Assigned 4D rotaries: see next page.
4. Assigned 4D keys: see next page.
5. Zoom/HR Zoom (see page 20)
6. Layout (see page 19): toggles the display between:
• Multi screen with volume/slice
• Single screen volume/slice
7. Soft menu
8. Trackball (see page 17)
• Rotate/translate volume rendering or selected cut-plane (Slice mode)
• Scroll through the cineloop
9. Clear: resets the orientation to default positions.
10. Angle: predefined orientations optimized for volume rendering.
11. 4D Gain
12. 2D Gain
Figure 1-3: The 4D controls on the control panel
4D/Multiplane Imaging User's Manual 9 FD092081-01
4D Imaging

Assigned rotaries and keys

Volume rendering assigned controls
1. Assigned rotaries
• Width
• Frequency
• Focus position
• Volume size
MORE menu
• Translate R
Slice mode assigned controls
1. Assigned rotaries
• Width
• Rotate R
• Translate (in Freeze only) R
• Focus position
• Volume size
MORE menu
• Frequency
Controls marked with R are also available in cine replay.
Figure 1-4: Volume rendering and Slice mode assigned controls
2. Assigned keys
• Slice R
• Front/Back
• Crop (in Freeze only) R
• Box (in Freeze only) R
• Cineloop (in Freeze only) R
• Full volume
• 4D Colorize R
• 9 Slice (in Full volume Freeze only) R
MORE menu
• Up/Down R
• Flip R
• Orientation window R
• Cine rotate (in Freeze only) R
2. Assigned keys
• Slice exit R
• Reference plane R
• Cineloop (in Freeze only) R
MORE menu
• Up/Down R
• Orientation window R
10 4D/Multiplane Imaging User's Manual
FD092081-01
4D Imaging

4D mode assigned controls

This section describes only the 4D mode controls. The scanning mode controls are described in the system User manual.
Width
(Volume rendering and Slice mode, Live)
Controls both elevation and azimuth widths, an increase of the elevation width results in a decrease of the azimuth width.
Volume size
(Volume rendering and Slice mode, Live)
Controls the size of the volume. Adjusting the volume size may affect the volume rate.
Rotate
(Slice mode, Live and Replay)
1. Rotate control
Rotate the selected cut-plane around the z-axis (see Figure 1-5).
Figure 1-5: Cut-plane rotation around the z-axis
Slice/Slice exit
(Volume rendering and Slice mode, Live and Replay)
Toggles the display between Volume rendering (Figure 1-1) and Slice mode (Figure 1-2).
4D/Multiplane Imaging User's Manual 11 FD092081-01
4D Imaging
Front/Back
(Volume rendering, Live and Replay)
Tilts the volume in the elevation direction and rotates the view position in one operation. rendering display from different angles (Figure 1-6).
FRONT/BACK enables volume
1. Front/Back: Volume tilt and View position rotation
A. Volume tilted to the left, view position looking towards the anterior wall.
B. Volume tilted to the right, view position looking towards the inferior wall.
Figure 1-6: Front/Back control on a PLAX volume
Reference plane
(Slice mode, Live and Replay)
Toggles the cut-plane selection between cut-plane 1, 2 or 3.
Crop (Free cropping)
(Volume rendering, Freeze and Replay)
12 4D/Multiplane Imaging User's Manual
FD092081-01
4D Imaging
Removes all data up front of the active cut-plane. Cropping can be applied on several parts of the volume by rotating/translating the active cut-plane (see page 26).
Box (Box cropping)
(Volume rendering, Freeze and Replay)
Removes all data around a user-adjustable box (see page 27).
Full volume
(Volume rendering, Live)
Activates the ECG triggered sub-volume acquisition. This technique enables the acquisition of a larger volume without compromising the resolution, by combining several sub-volumes acquired over two to six heart cycles (see Figure 1-7). When acquisition is done for the number of heart cycles set, the process is repeated replacing the oldest sub-volumes.
+=
Figure 1-7: ECG triggered volume acquisition (two heart cycles)
4D Colorize
(Volume rendering, Live and Replay)
Adjusts the volume rendering color from a color map menu.
Depth encoded color maps
From this menu the user can also select a depth encoded color map. These color maps use colors to improve the perception of depth. Selecting the bronze/blue color map will display structures that are close to the view plane with a bronze color. Structures that are farther behind will be colored with a gray
4D/Multiplane Imaging User's Manual 13 FD092081-01
4D Imaging
color, while the structures that are farthest behind will be colored in blue. Very bright colors are almost white, independent of the depth.
Stereo vision
4D Stereo vision is a display technique that enhances the perception of depth in 3D renderings. This is achieved by mixing two different volume renderings with slightly separated viewing angles and presenting them separately to the user’s left and right eyes. This feature requires the use of anaglyph stereo glasses (glasses with one red and one cyan lens).
Normally you should be able to see the stereoscopic effect after a few seconds. The effect may gradually improve after a while. If you are already wearing glasses or lenses you should not take them off, since the stereo effect then may be greatly reduced.
Note: Not all users may be able to perceive depth using stereoscopic display techniques.

Additional 4D mode assigned controls

The following controls are available after pressing MORE.
Up/Down
(Volume rendering and Slice mode, Live and Replay)
Flips the volume 180 degrees.
Flip
(Volume rendering, Live and Replay)
Flips the view position (Figure 1-8).
14 4D/Multiplane Imaging User's Manual
FD092081-01
1. Flip control
A. View position looking downward
B. View position looking upward
Figure 1-8: Flip control
Orientation window
4D Imaging
(Volume rendering and Slice mode, Live and Replay)
Shows/hides the Orientation window.
Cine rotate
(Volume rendering, Replay)
Rotates back and forth the volume rendering.
4D/Multiplane Imaging User's Manual 15 FD092081-01
4D Imaging

Soft menu controls

Soft menu controls are related to image quality adjustment. These controls are accessed using the 4-way rocker on the control panel. Only the 4D controls are described in this section, refer to the system user manual for general imaging controls.
Smoothness
(Volume rendering, Live and Replay)
Affects continuity of structures and image noise. Too much smoothness will blur the image, too little will leave too much noise.
Shading
(Volume rendering, Live and Replay)
Adjusts the shading effect on the image. Shading may improve three dimensional perception.
Tilt
(Volume rendering, Live)
Tilts the volume in the elevation direction.
1. Tilt control: volume tilting in the elevation direction
Figure 1-9: Tilt control
16 4D/Multiplane Imaging User's Manual
FD092081-01
4D Imaging
Number of Cycles
(Volume rendering, Live, Full volume acquisition)
Controls the number of cycles the ECG triggered full volume acquisition is based on. Select between two, three, four or six cycles. Four cycles is default setting.
DDP (Data Dependent Processing)
(Volume rendering, Live)
Performs temporal processing which reduces random noise without affecting the motion of significant tissue structures.
UD Clarity
(Volume rendering, Freeze and Replay)
Enables the user to create a personalized appearance of the tissue rendering by reducing noise and enhancing boundaries between different structures. Adjustment toward the left creates a smoother image. Adjustment toward the right creates a crisper image.
Volume optimize
(Volume rendering, Live and Replay)
Optimizes the volume rendering by adjusting several display controls simultaneously (e.g Shading, Smoothness... etc.).
Gamma
Adjusts the brightness of midtone values. A higher gamma value produces an overall darker image, a lower gamma value a lighter image.

Trackball controls

The trackball has multiple functions. The trackball functions are organized in several functional groups as shown in the table below.
The function selected is displayed in the lower right corner of the screen (Figure 1-10).
• Press within the active functional group. Groups with several
SELECT to toggle between the trackball functions
4D/Multiplane Imaging User's Manual 17 FD092081-01
4D Imaging
functions are marked with a + symbol.
• Press
1. Trackball key: toggles between trackball functional groups.
2. Select key: toggles between the functions within the active group. Groups with several functions are marked with a + symbol.
TRACKBALL to toggle between the functional groups.
Figure 1-10: The Trackball area
Orientation controls group
Function Description
Press
SELECT to toggle between the controls.
Rotate Rotates the volume rendering (see page 25).
Translate Translates the volume rendering (see
page 25).
Cineloop control group
Function Description
Scroll Scrolls through a cineloop.
18 4D/Multiplane Imaging User's Manual
FD092081-01
4D Imaging

Display controls

Layout key on the front panel
Toggles the display between multi-screen and single-screen.
• Multi-screen:
• In Volume rendering: displays a volume rendering and 2D images from two perpendicular cut-planes.
• In Slice mode: displays 2D images from three cut-planes with the selected cut-plane in the main window.
• Single-screen:
• In Volume rendering: displays the volume rendering.
• In Slice mode: display the 2D image of the selected cut-plane.
1. Display options for Volume rendering
2. Display options for Slice mode
Figure 1-11: 4D mode display options
4D Gain (Active Gain rotary)
Affects the image “depth” or transparency. Too much 4D Gain applied will take away structures, too little will leave opaque “Gray clouds” in the ventricle.
4D/Multiplane Imaging User's Manual 19 FD092081-01
4D Imaging
Zoom
Display zoom
Activated and adjusted by rotating the rectangular shape of the zoomed area is displayed in the Orientation window.
High Resolution (HR) zoom
HR zoom concentrates the image processing to a magnified, user selectable area in the image, resulting in a higher volume rate in the selected image area.
To be able to use the HR zoom in 4D imaging the HR zoom function must be activated in B mode before imaging mode.
ZOOM rotary ( ). The
entering the 4D
20 4D/Multiplane Imaging User's Manual
FD092081-01

Using 4D mode - Vivid 7

1. Select the 3V probe and cardiac application.
2. Create an examination.

Parasternal view acquisition

1. In 2D mode, acquire a PLAX view and optimize the image quality, using
2. Press
4D.
3. Move the probe up or down so that the anterior wall is just inside the image in the lower left window in the 4D screen (Figure 1-12).
DEPTH, GAIN, TGC...etc.
4D Imaging
Note: In this position the volume is tilted to the left and the view position is looking towards the anterior wall.
Figure 1-12: PLAX volume rendering
4. Adjust 4D Gain to optimize the depth impression of the back wall.
5. To display a volume looking towards the inferior wall, press
FRONT/BACK.
The volume is tilted to the right and the view position is rotated 180 degrees.
4D/Multiplane Imaging User's Manual 21 FD092081-01
4D Imaging
6. Press FREEZE, 2D FREEZE and rotate the volume to check the result.
7. Press
IMG. STORE.

Apical view acquisition

1. In 2D mode, acquire a Apical view and optimize the image quality, using
2. Press
4D and follow the same procedure as for the PLAX
view (see page 21). Make sure to assess the image quality in the reference views and the volume rendering. Use display the volume correctly.
3. Press
FREEZE, 2D FREEZE and rotate the volume to check
the result.
4. Press
IMG. STORE.
DEPTH, GAIN, TGC...etc.
FRONT/BACK to
22 4D/Multiplane Imaging User's Manual
FD092081-01
4D Imaging

Full volume acquisition

Full volume acquisition is based on ECG triggering acquisition of sub-volumes. This technique enables the acquisition of a larger volume without compromising the resolution, by combining several sub-volumes acquired over two, three, four or six heart cycles (see Figure 1-7, page 13). When acquisition is done for the number of heart cycles set, the process is repeated replacing the oldest sub-volumes.
ECG triggering acquisition may by nature contain artifacts.
CAUTION
Artifacts may be caused by:
Movements of the probe caused by the operator during
acquisition.
Movements of the patient during acquisition.
Irregular heart rate during acquisition.
To validate the acquisition, perform a visual inspection in both the volume rendering and the elevation plane. Stitching artifacts are shown as visible transitions between the sub-volumes (Figure 1-13)
To avoid spatial artifacts, make sure that the probe and the patient are not moving during the acquisition. The patient should, if possible hold his/her breath. The ECG trace should be stable.
1. Connect the ECG device and make sure to obtain a stable ECG trace.
2. In 2D mode, acquire an Apical view and optimize the image quality, using
3. Press
4. Press
4D.
FULL VOLUME.
DEPTH, GAIN, TGC...etc.
The Full volume acquisition is started.
5. You may adjust
ANGLE to get different view. The default
top/down view is best for stitching quality assessment.
6. Ask the patient to hold her/his breath at end expiration. Keep the probe steady and look for stitching artifacts in both the volume rendering and the elevation plane in the lower left window of the screen (Figure 1-13). Attention should be brought on stitching quality during acquisition rather than volume rendering quality.
4D/Multiplane Imaging User's Manual 23 FD092081-01
4D Imaging
A: Acquisition with stitching artifacts
1. Elevation plane
2. Volume rendering
B: Acquisition without stitching artifacts
1. Elevation plane
2. Volume rendering
Figure 1-13: Stitching quality assessment
When no artifacts are seen over a few heart cycles, press
FREEZE.
24 4D/Multiplane Imaging User's Manual
FD092081-01
Loading...
+ 94 hidden pages