GE Vivid 7 User manual

g
Technical Publications
Vivid 7/EchoPAC PC
Dimension - version 7.x.x
0470
GE Medical Systems
4D and Multi-plane Imaging
GEVU #: FD092081
MHLW No: 21300BZY00416000
Operating Documentation
Copyright © 2007 By General Electric Co.
g
GE Medical Systems
MANUAL STATUS
FD092081-01
01/08/2007
COMPANY DATA GE VINGMED ULTRASOUND A/S
© GE Medical Systems. All rights reserved. No part of this manual may be reproduced, stored in a retrieval system, or transmitted, in any form or by any means, electronic, mechanical, photocopying, recording, or otherwise, without the prior written permission of GE Medical Systems.
Strandpromenaden 45, N-3191 Horten, Norway
Tel.: (+47) 3302 1100 Fax: (+47) 3302 1350

Table of Contents

Table of Contents Introduction
4D imaging ..................................................................................... 1
Multi-plane Imaging....................................................................... 2
Measurement and analysis........................................................... 2
Multi-plane stress echo................................................................. 2
Remark concerning the 3V probe ................................................ 2
Important........................................................................................ 3
Conventions used in this manual ................................................ 4
Chapter 1 4D Imaging
Table of Contents
Introduction.................................................................................... 6
4D mode overview - Vivid 7 .......................................................... 7
Volume rendering mode screen............................................ 7
Slice mode screen ................................................................ 8
4D mode controls.................................................................. 9
Assigned rotaries and keys................................................. 10
4D mode assigned controls ................................................ 11
Additional 4D mode assigned controls ............................... 14
Soft menu controls.............................................................. 16
Trackball controls................................................................ 17
Display controls .................................................................. 19
Using 4D mode - Vivid 7.............................................................. 21
Parasternal view acquisition ............................................... 21
Apical view acquisition........................................................ 22
Full volume acquisition ....................................................... 23
Rotating/Translating the 4D image ..................................... 25
Zooming.............................................................................. 26
Cropping ............................................................................. 26
9 Slice................................................................................. 29
4D mode overview - EchoPAC PC ............................................. 32
4D/Multiplane Imaging User's Manual 1 FD092081-01
Table of Contents
Volume rendering mode screen ..........................................32
Slice mode screen...............................................................33
4D mode control panel ........................................................34
Display controls................................................................... 36
Cropping..............................................................................37
Working with 4D acquisitions - EchoPAC PC ...........................39
9 Slice..................................................................................39
4D LV Volume application ...........................................................41
Starting the 4D LV Volume application — Vivid 7 ...............41
Starting the 4D LV Volume application — EchoPAC PC ....41
Chapter 2 4D Color Flow Imaging
Introduction ..................................................................................44
4D Color Flow mode overview - Vivid 7 .....................................45
Color Flow Volume rendering mode screen ........................45
Color Flow Slice mode screen.............................................46
4D Color Flow mode controls .............................................. 47
Assigned rotaries and keys .................................................48
4D Color Flow mode assigned controls...............................49
Additional 4D mode assigned controls ................................50
Soft menu controls ..............................................................51
Trackball controls ................................................................52
Display controls................................................................... 52
Using 4D Color Flow mode - Vivid 7...........................................54
Real time 4D Color flow acquisition.....................................54
Full volume Color Flow acquisition......................................55
6 Slice..................................................................................57
4D Color Flow mode overview - EchoPAC PC ..........................60
Volume rendering mode screen ..........................................60
Slice mode screen...............................................................61
4D mode control panel ........................................................62
Chapter 3 Multi-plane imaging
Introduction ..................................................................................66
2 4D/Multiplane Imaging User's Manual
FD092081-01
Multi-plane mode overview - Vivid 7.......................................... 67
Bi-plane mode screen......................................................... 67
Tri-plane mode screen........................................................ 68
Multi-plane mode controls................................................... 69
Using Multi-plane mode imaging - Vivid 7 ................................ 77
Scan plane rotation............................................................. 78
Tilting scan plane 2............................................................. 78
Zooming.............................................................................. 79
Multi-plane mode overview - EchoPAC PC ............................... 82
Bi-plane mode screen......................................................... 82
Tri-plane mode screen........................................................ 83
The multi-plane control panel ............................................. 84
Working with multi-plane acquisitions - EchoPAC PC ............ 85
Chapter 4 Measurements and Analysis
Introduction.................................................................................. 88
Left ventricular volume measurements..................................... 89
Tri-plane acquisition............................................................ 89
Full volume acquisition ....................................................... 92
Rotation of the Volume reconstruction................................ 93
Bi-plane acquisition............................................................. 94
TSI surface model........................................................................ 96
To edit the sampling path ................................................... 97
Quantitative analysis................................................................... 98
Starting Quantitative analysis from a multi-plane
acquisition........................................................................... 98
Table of Contents
Chapter 5 Multi-plane Stress Echo
Introduction................................................................................ 102
Creating a Multi-plane stress test template ............................ 103
Launching the Template editor ......................................... 104
Stress Template setup...................................................... 104
Stress test acquisition .............................................................. 106
Baseline acquisitions ........................................................ 106
4D/Multiplane Imaging User's Manual 3 FD092081-01
Table of Contents
Low dose and Peak dose level acquisitions......................109
Image analysis............................................................................112
Index
4 4D/Multiplane Imaging User's Manual
FD092081-01

Introduction

This user manual describes the 4D and Multi-plane Imaging applications for the Vivid 7 Dimension and EchoPACPCDimension.
The 2D matrix probe 3V enables real time volume rendering, simultaneous bi-plane or tri-plane data acquisition (multi-plane acquisition).
Introduction
1. Real time Volume acquisition (4D imaging)
2. Real time Bi-plane acquisition
3. Real time Tri-plane acquisition

4D imaging

4D imaging enables real time acquisition and rendering of volume ultrasound data. Free rotation of the 3-dimensional image combined with the zoom function and 4D image optimization controls enhance spatial understanding of the anatomical structure and function of the heart.
4D imaging is available in combination with B-Mode only.
4D/Multiplane Imaging User's Manual 1 FD092081-01
Introduction

Multi-plane Imaging

Multi-plane imaging displays two (Bi-plane) or three (Tri-plane) rotated scan planes acquired simultaneously. Free rotation, tilting (in Bi-plane) of the scan planes and zoom enable the investigation of anatomical structures from different angles.
Multi-plane imaging is available from B-Mode, Color flow mode and TVI related modes.
If not otherwise specified, the term multi-plane means either Bi-plane or Tri-plane.

Measurement and analysis

All cardiac measurements available in B mode are also available in 4D and Multi-plane.
The 4D and Multi-plane modes enable the creation of a left ventricular volume reconstruction based on contours drawn from three cross sections at both end-systole and end-diastole with calculation of end-systolic and end-diastolic volumes and ejection fraction.

Multi-plane stress echo

Combined with specially designed Stress echo protocols, Multi-plane mode enables faster stress tests as several views can be acquired simultaneously.

Remark concerning the 3V probe

The 3V probe temperature will increase during extensive use, due to heating from the probe electronics. If the probe reaches its temperature limit, this will be detected by the system and scanning will stop. Currently, this limit is set to 42.3 degrees Celsius, which complies with regulatory instructions relating to patient comfort and safety. This is reported with a dialog box and the user has to wait until the temperature is below acceptable limits before scanning can continue. This does not in any way indicate probe malfunction. The user has to unfreeze to start scanning again.
In order to reduce the chance for full freeze due to heating, the system will reduce power slightly in a temperature interval just
2 4D/Multiplane Imaging User's Manual
FD092081-01

Important

Introduction
below the full freeze limit. Changes to the power level are reported in the status bar as Setting power level down and Setting power level up.
The recommended use of this probe for live 4D and full volume 4D scanning is to enter full freeze whenever the probe is not used for imaging. For other modalities no special actions are required.
Read and understand all instructions in the Vivid 7 and EchoPAC PC User manuals before attempting to use the devices.
4D/Multiplane Imaging User's Manual 3 FD092081-01
Introduction

Conventions used in this manual

Keys and button, on the control panel or alphanumeric keyboard are indicated by over and underlined text (ex. refers to the 2D mode key)
Bold type, describes button names on the screen.
Italic type: describes program windows, screens and dialogue
boxes.
Icons, highlight safety issues as follow:
Indicates that a specific hazard exists that, given inappropriate conditions or actions, will cause:
DANGER
WARNING
• Severe or fatal personal injury
• Substantial property damage
Indicates that a specific hazard exists that, given inappropriate conditions or actions, will cause:
• Severe personal injury
• Substantial property damage
2D
CAUTION
Indicates that a potential hazard may exist that, given inappropriate conditions or actions, can cause:
• Minor injury
• Property damage
4 4D/Multiplane Imaging User's Manual
FD092081-01

4D Imaging

Chapter 1 4D Imaging
• Introduction ................................................................................... ...... 6
• 4D mode overview - Vivid 7 .......................................................... ...... 7
• Volume rendering mode screen ................................................... 7
• Slice mode screen ........................................................................ 8
• 4D mode controls ......................................................................... 9
• Using 4D mode - Vivid 7 ............................................................... .... 21
• Parasternal view acquisition ....................................................... 21
• Apical view acquisition ............................................................... 22
• Full volume acquisition ............................................................... 23
• Rotating/Translating the 4D image ............................................. 25
• Zooming ..................................................................................... 26
• Cropping ..................................................................................... 26
• 9 Slice ........................................................................................ 29
• 4D mode overview - EchoPAC PC ............................................... .... 32
• Volume rendering mode screen ................................................. 32
• Slice mode screen ...................................................................... 33
• 4D mode control panel ............................................................... 34
• Display controls .......................................................................... 36
• Working with 4D acquisitions - EchoPAC PC ............................ .... 39
• 9 Slice ........................................................................................ 39
• 4D LV Volume application ............................................................ .... 41
• Starting the 4D LV Volume application — EchoPAC PC ........... 41
4D/Multiplane Imaging User's Manual 5 FD092081-01
4D Imaging

Introduction

The 3V probe enables real time acquisition of volume ultrasound data. Free rotation of the three-dimensional image combined with zooming and 4D image optimization controls enhance spatial understanding of the anatomical structure and function of the heart.
Two display modes are available, Volume rendering mode for three dimensional scanning and Slice mode for measurements and volume reconstruction purpose.
4D imaging is available from B mode only.
6 4D/Multiplane Imaging User's Manual
FD092081-01

4D mode overview - Vivid 7

Volume rendering mode screen

The Volume rendering mode displays a volume rendering and 2D images from two perpendicular cut-planes.
4D Imaging
1. Volume rendering display from cut-plane 1 (yellow). The volume rendering may be adjusted by rotating and translating the cut-plane 1.
2. Cut-plane 2 (white): 2D image in the azimuth plane.
3. Cut-plane 3 (green): 2D image in the elevation plane.
4. Orientation window: displays a three-dimensional model with cut-planes position and orientation.
5. Color coded cut-plane markers indicate the other cut-planes position relative to the displayed cut-plane.
6. View direction marker.
7. Soft menu controls (see page 16)
8. Trackball functions (see page 17)
Figure 1-1: The 4D screen (Volume rendering)
4D/Multiplane Imaging User's Manual 7 FD092081-01
4D Imaging

Slice mode screen

The Slice mode displays three cut-planes. The cut-planes can be rotated and translated independently of each other. This mode is use to perform measurements and volume reconstruction based on contour traces done on several cut-planes.
1. Cut-plane 1 (yellow)
2. Cut-plane 2 (white)
3. Cut-plane 3 (green)
4. Color coded cut-plane markers indicate the position of the other cut-planes relative to the displayed cut-plane.
5. Orientation window: displays a three-dimensional model with cut-planes position.
6. View direction marker.
7. Trackball functions (see page 17)
Figure 1-2: The 4D screen (Slice mode)
8 4D/Multiplane Imaging User's Manual
FD092081-01

4D mode controls

Alt. Alt. Alt. Alt.
4D Imaging
3
4
Width
Update /
ck
ll
a
a
Tr
B
Menu
1. 4D key
2. Multiplane key (see page 65)
3. Assigned 4D rotaries: see next page.
4. Assigned 4D keys: see next page.
5. Zoom/HR Zoom (see page 20)
6. Layout (see page 19): toggles the display between:
• Multi screen with volume/slice
• Single screen volume/slice
7. Soft menu
8. Trackball (see page 17)
• Rotate/translate volume rendering or selected cut-plane (Slice mode)
• Scroll through the cineloop
9. Clear: resets the orientation to default positions.
10. Angle: predefined orientations optimized for volume rendering.
11. 4D Gain
12. 2D Gain
Figure 1-3: The 4D controls on the control panel
4D/Multiplane Imaging User's Manual 9 FD092081-01
4D Imaging

Assigned rotaries and keys

Volume rendering assigned controls
1. Assigned rotaries
• Width
• Frequency
• Focus position
• Volume size
MORE menu
• Translate R
Slice mode assigned controls
1. Assigned rotaries
• Width
• Rotate R
• Translate (in Freeze only) R
• Focus position
• Volume size
MORE menu
• Frequency
Controls marked with R are also available in cine replay.
Figure 1-4: Volume rendering and Slice mode assigned controls
2. Assigned keys
• Slice R
• Front/Back
• Crop (in Freeze only) R
• Box (in Freeze only) R
• Cineloop (in Freeze only) R
• Full volume
• 4D Colorize R
• 9 Slice (in Full volume Freeze only) R
MORE menu
• Up/Down R
• Flip R
• Orientation window R
• Cine rotate (in Freeze only) R
2. Assigned keys
• Slice exit R
• Reference plane R
• Cineloop (in Freeze only) R
MORE menu
• Up/Down R
• Orientation window R
10 4D/Multiplane Imaging User's Manual
FD092081-01
4D Imaging

4D mode assigned controls

This section describes only the 4D mode controls. The scanning mode controls are described in the system User manual.
Width
(Volume rendering and Slice mode, Live)
Controls both elevation and azimuth widths, an increase of the elevation width results in a decrease of the azimuth width.
Volume size
(Volume rendering and Slice mode, Live)
Controls the size of the volume. Adjusting the volume size may affect the volume rate.
Rotate
(Slice mode, Live and Replay)
1. Rotate control
Rotate the selected cut-plane around the z-axis (see Figure 1-5).
Figure 1-5: Cut-plane rotation around the z-axis
Slice/Slice exit
(Volume rendering and Slice mode, Live and Replay)
Toggles the display between Volume rendering (Figure 1-1) and Slice mode (Figure 1-2).
4D/Multiplane Imaging User's Manual 11 FD092081-01
4D Imaging
Front/Back
(Volume rendering, Live and Replay)
Tilts the volume in the elevation direction and rotates the view position in one operation. rendering display from different angles (Figure 1-6).
FRONT/BACK enables volume
1. Front/Back: Volume tilt and View position rotation
A. Volume tilted to the left, view position looking towards the anterior wall.
B. Volume tilted to the right, view position looking towards the inferior wall.
Figure 1-6: Front/Back control on a PLAX volume
Reference plane
(Slice mode, Live and Replay)
Toggles the cut-plane selection between cut-plane 1, 2 or 3.
Crop (Free cropping)
(Volume rendering, Freeze and Replay)
12 4D/Multiplane Imaging User's Manual
FD092081-01
4D Imaging
Removes all data up front of the active cut-plane. Cropping can be applied on several parts of the volume by rotating/translating the active cut-plane (see page 26).
Box (Box cropping)
(Volume rendering, Freeze and Replay)
Removes all data around a user-adjustable box (see page 27).
Full volume
(Volume rendering, Live)
Activates the ECG triggered sub-volume acquisition. This technique enables the acquisition of a larger volume without compromising the resolution, by combining several sub-volumes acquired over two to six heart cycles (see Figure 1-7). When acquisition is done for the number of heart cycles set, the process is repeated replacing the oldest sub-volumes.
+=
Figure 1-7: ECG triggered volume acquisition (two heart cycles)
4D Colorize
(Volume rendering, Live and Replay)
Adjusts the volume rendering color from a color map menu.
Depth encoded color maps
From this menu the user can also select a depth encoded color map. These color maps use colors to improve the perception of depth. Selecting the bronze/blue color map will display structures that are close to the view plane with a bronze color. Structures that are farther behind will be colored with a gray
4D/Multiplane Imaging User's Manual 13 FD092081-01
4D Imaging
color, while the structures that are farthest behind will be colored in blue. Very bright colors are almost white, independent of the depth.
Stereo vision
4D Stereo vision is a display technique that enhances the perception of depth in 3D renderings. This is achieved by mixing two different volume renderings with slightly separated viewing angles and presenting them separately to the user’s left and right eyes. This feature requires the use of anaglyph stereo glasses (glasses with one red and one cyan lens).
Normally you should be able to see the stereoscopic effect after a few seconds. The effect may gradually improve after a while. If you are already wearing glasses or lenses you should not take them off, since the stereo effect then may be greatly reduced.
Note: Not all users may be able to perceive depth using stereoscopic display techniques.

Additional 4D mode assigned controls

The following controls are available after pressing MORE.
Up/Down
(Volume rendering and Slice mode, Live and Replay)
Flips the volume 180 degrees.
Flip
(Volume rendering, Live and Replay)
Flips the view position (Figure 1-8).
14 4D/Multiplane Imaging User's Manual
FD092081-01
1. Flip control
A. View position looking downward
B. View position looking upward
Figure 1-8: Flip control
Orientation window
4D Imaging
(Volume rendering and Slice mode, Live and Replay)
Shows/hides the Orientation window.
Cine rotate
(Volume rendering, Replay)
Rotates back and forth the volume rendering.
4D/Multiplane Imaging User's Manual 15 FD092081-01
4D Imaging

Soft menu controls

Soft menu controls are related to image quality adjustment. These controls are accessed using the 4-way rocker on the control panel. Only the 4D controls are described in this section, refer to the system user manual for general imaging controls.
Smoothness
(Volume rendering, Live and Replay)
Affects continuity of structures and image noise. Too much smoothness will blur the image, too little will leave too much noise.
Shading
(Volume rendering, Live and Replay)
Adjusts the shading effect on the image. Shading may improve three dimensional perception.
Tilt
(Volume rendering, Live)
Tilts the volume in the elevation direction.
1. Tilt control: volume tilting in the elevation direction
Figure 1-9: Tilt control
16 4D/Multiplane Imaging User's Manual
FD092081-01
4D Imaging
Number of Cycles
(Volume rendering, Live, Full volume acquisition)
Controls the number of cycles the ECG triggered full volume acquisition is based on. Select between two, three, four or six cycles. Four cycles is default setting.
DDP (Data Dependent Processing)
(Volume rendering, Live)
Performs temporal processing which reduces random noise without affecting the motion of significant tissue structures.
UD Clarity
(Volume rendering, Freeze and Replay)
Enables the user to create a personalized appearance of the tissue rendering by reducing noise and enhancing boundaries between different structures. Adjustment toward the left creates a smoother image. Adjustment toward the right creates a crisper image.
Volume optimize
(Volume rendering, Live and Replay)
Optimizes the volume rendering by adjusting several display controls simultaneously (e.g Shading, Smoothness... etc.).
Gamma
Adjusts the brightness of midtone values. A higher gamma value produces an overall darker image, a lower gamma value a lighter image.

Trackball controls

The trackball has multiple functions. The trackball functions are organized in several functional groups as shown in the table below.
The function selected is displayed in the lower right corner of the screen (Figure 1-10).
• Press within the active functional group. Groups with several
SELECT to toggle between the trackball functions
4D/Multiplane Imaging User's Manual 17 FD092081-01
4D Imaging
functions are marked with a + symbol.
• Press
1. Trackball key: toggles between trackball functional groups.
2. Select key: toggles between the functions within the active group. Groups with several functions are marked with a + symbol.
TRACKBALL to toggle between the functional groups.
Figure 1-10: The Trackball area
Orientation controls group
Function Description
Press
SELECT to toggle between the controls.
Rotate Rotates the volume rendering (see page 25).
Translate Translates the volume rendering (see
page 25).
Cineloop control group
Function Description
Scroll Scrolls through a cineloop.
18 4D/Multiplane Imaging User's Manual
FD092081-01
4D Imaging

Display controls

Layout key on the front panel
Toggles the display between multi-screen and single-screen.
• Multi-screen:
• In Volume rendering: displays a volume rendering and 2D images from two perpendicular cut-planes.
• In Slice mode: displays 2D images from three cut-planes with the selected cut-plane in the main window.
• Single-screen:
• In Volume rendering: displays the volume rendering.
• In Slice mode: display the 2D image of the selected cut-plane.
1. Display options for Volume rendering
2. Display options for Slice mode
Figure 1-11: 4D mode display options
4D Gain (Active Gain rotary)
Affects the image “depth” or transparency. Too much 4D Gain applied will take away structures, too little will leave opaque “Gray clouds” in the ventricle.
4D/Multiplane Imaging User's Manual 19 FD092081-01
4D Imaging
Zoom
Display zoom
Activated and adjusted by rotating the rectangular shape of the zoomed area is displayed in the Orientation window.
High Resolution (HR) zoom
HR zoom concentrates the image processing to a magnified, user selectable area in the image, resulting in a higher volume rate in the selected image area.
To be able to use the HR zoom in 4D imaging the HR zoom function must be activated in B mode before imaging mode.
ZOOM rotary ( ). The
entering the 4D
20 4D/Multiplane Imaging User's Manual
FD092081-01

Using 4D mode - Vivid 7

1. Select the 3V probe and cardiac application.
2. Create an examination.

Parasternal view acquisition

1. In 2D mode, acquire a PLAX view and optimize the image quality, using
2. Press
4D.
3. Move the probe up or down so that the anterior wall is just inside the image in the lower left window in the 4D screen (Figure 1-12).
DEPTH, GAIN, TGC...etc.
4D Imaging
Note: In this position the volume is tilted to the left and the view position is looking towards the anterior wall.
Figure 1-12: PLAX volume rendering
4. Adjust 4D Gain to optimize the depth impression of the back wall.
5. To display a volume looking towards the inferior wall, press
FRONT/BACK.
The volume is tilted to the right and the view position is rotated 180 degrees.
4D/Multiplane Imaging User's Manual 21 FD092081-01
4D Imaging
6. Press FREEZE, 2D FREEZE and rotate the volume to check the result.
7. Press
IMG. STORE.

Apical view acquisition

1. In 2D mode, acquire a Apical view and optimize the image quality, using
2. Press
4D and follow the same procedure as for the PLAX
view (see page 21). Make sure to assess the image quality in the reference views and the volume rendering. Use display the volume correctly.
3. Press
FREEZE, 2D FREEZE and rotate the volume to check
the result.
4. Press
IMG. STORE.
DEPTH, GAIN, TGC...etc.
FRONT/BACK to
22 4D/Multiplane Imaging User's Manual
FD092081-01
4D Imaging

Full volume acquisition

Full volume acquisition is based on ECG triggering acquisition of sub-volumes. This technique enables the acquisition of a larger volume without compromising the resolution, by combining several sub-volumes acquired over two, three, four or six heart cycles (see Figure 1-7, page 13). When acquisition is done for the number of heart cycles set, the process is repeated replacing the oldest sub-volumes.
ECG triggering acquisition may by nature contain artifacts.
CAUTION
Artifacts may be caused by:
Movements of the probe caused by the operator during
acquisition.
Movements of the patient during acquisition.
Irregular heart rate during acquisition.
To validate the acquisition, perform a visual inspection in both the volume rendering and the elevation plane. Stitching artifacts are shown as visible transitions between the sub-volumes (Figure 1-13)
To avoid spatial artifacts, make sure that the probe and the patient are not moving during the acquisition. The patient should, if possible hold his/her breath. The ECG trace should be stable.
1. Connect the ECG device and make sure to obtain a stable ECG trace.
2. In 2D mode, acquire an Apical view and optimize the image quality, using
3. Press
4. Press
4D.
FULL VOLUME.
DEPTH, GAIN, TGC...etc.
The Full volume acquisition is started.
5. You may adjust
ANGLE to get different view. The default
top/down view is best for stitching quality assessment.
6. Ask the patient to hold her/his breath at end expiration. Keep the probe steady and look for stitching artifacts in both the volume rendering and the elevation plane in the lower left window of the screen (Figure 1-13). Attention should be brought on stitching quality during acquisition rather than volume rendering quality.
4D/Multiplane Imaging User's Manual 23 FD092081-01
4D Imaging
A: Acquisition with stitching artifacts
1. Elevation plane
2. Volume rendering
B: Acquisition without stitching artifacts
1. Elevation plane
2. Volume rendering
Figure 1-13: Stitching quality assessment
When no artifacts are seen over a few heart cycles, press
FREEZE.
24 4D/Multiplane Imaging User's Manual
FD092081-01
4D Imaging
Note: it is recommended to acquire several heart cycles and use
7. Once in Freeze, adjust the result and tress
SELECT CYCLE to select the best one.
4D GAIN. Rotate the volume to check
IMG. STORE.

Rotating/Translating the 4D image

Translation and rotation can be performed in either Volume rendering or Slice mode. Press modes.
1. Move the trackball to rotate or translate the cut-plane according to the trackball function selected.
2. Press
SELECT to toggle between Rotate and Translate.
Move the trackball to apply the selected control (see Figure 1-14). The default position can be displayed again by pressing
CLEAR.
SLICE to toggle between the two
1. Select the trackball control “Translate” and use the trackball to translate the
cut-plane.
2. Select the trackball control “Rotate” and use the trackball to rotate the
cut-plane (all directions).
Figure 1-14: Rotating/translating the 4D image
4D/Multiplane Imaging User's Manual 25 FD092081-01
4D Imaging
1. Zoomed area

Zooming

1. Rotate the ZOOM knob clockwise on the control panel. The volume rendering is magnified. The frame of the Volume rendering cut-plane in the Orientation window is updated showing the magnified portion of the volume (Figure 1-15).
Figure 1-15: The Orientation window in zoom mode

Cropping

There are two cropping tools available: Free cropping tool and Box cropping tool.
The Free cropping tool removes all data up front of the active cut-plane. Cropping can be applied on several parts of the volume by rotating/translating the active cut-plane.
The Box cropping tool removes all data around a user-adjustable box.
It is not possible to use both cropping tools on the same acquisition.
Free cropping
The Free cropping tool is available in Volume rendering mode, when in Freeze or in Replay.
1. While in Volume rendering live mode, press
2. Press Note: Pressing cropping operations.
3. Rotate and translate the active cut-plane. All data above the cut-plane will be removed when applying crop.
CROP.
CROP will remove any previous Box
FREEZE.
26 4D/Multiplane Imaging User's Manual
FD092081-01
4D Imaging
4. Press SET to apply crop. All data up front of the active cut-plane is removed.
5. Repeat steps 3 and 4 to remove other parts of the volume. Note: press stepwise. Press operation.
Box cropping
The Box cropping tool is available in Volume rendering mode, when in Freeze or in Replay.
1. While in Volume rendering live mode, press
2. Press A box frame is displayed with two adjustable sides (highlighted as red and blue), see Figure 1-16. Note: Pressing cropping operations.
3. Rotate the corresponding sides around the structure of interest. All data outside the box is removed.
4. Rotate the sides of the box.
5. Repeat steps 3 and 4 to adjust the selected sides of the box until the structure of interest is within the box. Note: press
6. Press
UNDO CROP to undo the cropping actions
REMOVE CROP to undo all cropping actions in one
FREEZE.
BOX.
BOX will remove any previous Free
RED and BLUE assigned rotaries to adjust the
BOX SIDES assigned rotary to select two other
RESET BOX to undo the box adjustments.
BOX EXIT to apply and exit box cropping.
4D/Multiplane Imaging User's Manual 27 FD092081-01
4D Imaging
1. Default cropping box with two active sides (red and blue). The other sides can be selected for adjustment
2. Cropped volume (from upper and lower sides)
Figure 1-16: Box cropping
28 4D/Multiplane Imaging User's Manual
FD092081-01
4D Imaging

9 Slice

9 Slice enables simultaneous display of nine short axis views generated from an Apical full volume acquisition.
9 Slice is available only from a full volume acquisition when in Freeze.
1. Acquire an apical Full volume view (see page 23).
2. Press
3. Press The 9Slice screen is displayed showing nine short axis views (Figure 1-17). The slices are evenly distributed and maximized in size for best assessment.
FREEZE.
9SLICE.
1. Upper slice
2. Lower slice
Figure 1-17: 9 Slice screen
If required, zoom in. All slices are zoomed in simultaneously.
4. The following adjustments can be done:
• Adjust
TOP and BOTTOM controls to change the slicing
area (Figure 1-18).
4D/Multiplane Imaging User's Manual 29 FD092081-01
4D Imaging
1. Top: upper slice adjustment
2. Bottom: lower slice adjustment
Figure 1-18: Slice distribution adjustment
• Adjust the AXIS 1 and AXIS 2 controls to rotate the slices backward/forward and sideways (Figure 1-19), to align the slices with the anatomical structure.
1. Axis 1: slice rotation backward/forward
2. Axis 2: slice rotation sideways
Figure 1-19: Slice rotation adjustment
30 4D/Multiplane Imaging User's Manual
FD092081-01
5. Press IMG. STORE to save.
4D Imaging
4D/Multiplane Imaging User's Manual 31 FD092081-01
4D Imaging

4D mode overview - EchoPAC PC

4D acquisitions from Vivid 7 can be opened and post processed on the EchoPAC PC workstation. This section describes the controls and procedures related to 4D mode on the workstation. Refer to the EchoPAC PC User manual about general use of the workstation.

Volume rendering mode screen

1. Volume rendering display from cut-plane 1 (yellow). The volume rendering may be adjusted by rotating and translating the cut-plane 1.
2. Cut-plane 2 (white): 2D image in the azimuth plane.
3. Cut-plane 3 (green): 2D image in the elevation plane.
4. Orientation window: displays a three-dimensional model with cut-planes position and orientation.
5. Color coded cut-plane markers indicate the other cut-planes position relative to the displayed cut-plane.
6. 4D control panel (Volume rendering mode)
Figure 1-20: The 4D screen (Volume rendering)
32 4D/Multiplane Imaging User's Manual
FD092081-01

Slice mode screen

4D Imaging
1. Cut-plane 1 (yellow)
2. Cut-plane 2 (white)
3. Cut-plane 3 (green)
4. Color coded cut-plane markers indicate the position of the other cut-planes relative to the displayed cut-plane.
5. View direction marker.
6. Orientation window: displays a three-dimensional model with cut-planes position.
7. 4D control panel (Slice mode)
Figure 1-21: The 4D screen (Slice mode)
4D/Multiplane Imaging User's Manual 33 FD092081-01
4D Imaging

4D mode control panel

Refer to ’4D mode controls’ on page 9 for information on the 4D related controls. All other general imaging controls are described in the workstation User manual.
Figure 1-22: The 4D Control panel
34 4D/Multiplane Imaging User's Manual
FD092081-01
4D Imaging
Rendering controls
In volume rendering, select Rendering to display the rendering controls.
Figure 1-23: Rendering and orientation panel
4D/Multiplane Imaging User's Manual 35 FD092081-01
4D Imaging

Display controls

Rotation
Rotation using the mouse
1. Place the mouse cursor in the image area.
2. Press and hold down the Left mouse button and drag the mouse in any direction. The volume rendering is rotated following the mouse cursor movement. To rotate around the Z axis, press and hold down drag the mouse to the left or right.
Rotation from the control panel
1. In Volume rendering, Select Rendering. The Rendering control panel is displayed (see Figure 1-23).
2. Adjust Rotate X, Rotate Y or Rotate Z to rotate the volume rendering around the corresponding axis.
ALT and
Figure 1-24: Rotation around the X, Y and Z axis
Translation
Translation using the mouse
1. Place the mouse cursor in the image area.
2. Press and hold down down. The volume rendering cut-plane is translated.
Translation from the control panel
1. In Volume rendering, Select Rendering. The Rendering control panel is displayed (see Figure 1-23).
2. Adjust the control Translate.
SHIFT and drag the mouse up or
36 4D/Multiplane Imaging User's Manual
FD092081-01
4D Imaging
Zoom
Zooming
1. Rotate the Mouse wheel to zoom in and out. OR Press OR Adjust Zoom on the control panel.
Zoom panning
1. While in zoom mode, press and hold down the mouse wheel and drag the mouse. OR Press and hold down
ARROW UP to zoom in, ARROW DOWN to zoom out.
SHIFT and press the ARROW keys.

Cropping

Free cropping
1. Select Rendering. The Rendering control panel is displayed (see Figure 1-23).
2. Select Crop. Note: Pressing Crop will remove any previous Box cropping operations.
3. Rotate and translate the active cut-plane. All data above the cut-plane will be removed when applying crop.
4. Press Set to apply crop. All data up front of the active cut-plane is removed.
5. Repeat steps 3 and 4 to remove other parts of the volume. Note: press Undo to undo the cropping actions stepwise. Press Remove crop to undo all cropping actions in one operation.
6. Press Crop to exit the Free cropping tool.
Box cropping
1. Select Rendering. The Rendering control panel is displayed (see Figure 1-23).
2. Select Box. A box frame is displayed with two adjustable sides (highlighted as red and blue), see Figure 1-25.
4D/Multiplane Imaging User's Manual 37 FD092081-01
4D Imaging
Note: Pressing Box will remove any previous Free cropping operations.
3. Use the Red and Blue controls to adjust the corresponding sides around the structure of interest. All data outside the box is removed.
4. Use the Box Sides control to select two other sides of the box.
5. Repeat steps 3 and 4 to adjust the selected sides of the box until the structure of interest is within the box. Note: Press Reset Box to undo all cropping actions in one operation.
6. Press Box to apply and exit Box cropping tool.
1. Default cropping box with two active sides (red and blue). The other sides can
be selected for adjustment
2. Cropped volume (from upper and lower sides)
Figure 1-25: Box cropping
38 4D/Multiplane Imaging User's Manual
FD092081-01
4D Imaging

Working with 4D acquisitions - EchoPAC PC

1. In the Search/Create patient window (Archive menu), select a patient record with 4D acquisitions. The Examination List window is displayed.
2. Select the desired examination and open (double-click) a 4D acquisition from the clipboard. OR Press Image browser and open (double-click) the 4D acquisition in the Image browser screen. The Volume rendering mode screen is displayed (see Figure 1-20, page 32).
3. Rotate and translate the volume rendering to display the structure of interest (see page 36).
4. Optimize the volume rendering using 4D Gain, Smoothness, Shading controls.
5. If desired, double-click on the image area to display the volume rendering in a single screen.
6. If desired, zoom in the structure of interest (see page 37).
7. If necessary, use the cropping tools to remove unwanted structures (see page 37).
8. If desired, press Cine rotate to rotate the volume rendering back and forth.
9. Press Store to save the changes.

9 Slice

9 Slice enables simultaneous display of nine short axis views generated from an Apical full volume acquisition.
1. Select an Apical full volume acquisition.
2. Press 9 Slice. The 9Slice screen is displayed showing nine short axis views. The slices are evenly distributed and maximized in size for best assessment. If required, zoom in. All slices are zoomed in simultaneously.
3. The following adjustments can be done:
• Adjust Top and Bottom controls to change the slicing
area.
• Adjust the Axis 1 and Axis 2 controls to rotate the slices
4D/Multiplane Imaging User's Manual 39 FD092081-01
4D Imaging
backward/forward and sideways, to align the slices with the anatomical structure.
4. Press Store to save.
40 4D/Multiplane Imaging User's Manual
FD092081-01

4D LV Volume application

The 4D LV Volume function may be analyzed on the Vivid 7 system or EchoPAC PC workstation using the TomTec 4D LV-Volume application. The 4D LV-Volume application enables analysis of global and regional volume measurement from a Full volume tissue acquisition.
The 4D LV-Volume application is an option.
Measurements generated in the 4D LV-Volume application are saved to the worksheet. The measurements have a specific TomTec label.
In addition 4D tissue acquisitions can be stored on a removable media to a format compatible with the TomTec application using the “Save as” function on the Vivid 7 system or the EchoPAC PC workstation (File format: VolDicom (*.dcm)). Refer to the Vivid 7 or EchoPAC PC user manual for additional information on the “Save as” function.
Starting the 4D LV Volume application — Vivid 7
4D Imaging
1. In a 4D Full volume tissue acquisition, press MEASURE.
2. Select Volume and 4D LV Volume in the Measurement menu. The 4D LV-Volume application is started. Refer to the TomTec 4D LV-Volume User manual for proper use of the application.
Starting the 4D LV Volume application — EchoPAC PC
1. In a 4D Full volume tissue acquisition, select 4D LV Volume on the control panel.
The 4D LV-Volume application is started. Refer to the TomTec 4D LV-Volume User manual for proper use of the application. Note: the 4D LV Volume application can also be started from the Measurement menu as described for the Vivid 7 above, and from the Rendering and orientation panel (Figure 1-23).
4D/Multiplane Imaging User's Manual 41 FD092081-01
4D Imaging
42 4D/Multiplane Imaging User's Manual
FD092081-01

4D Color Flow Imaging

Chapter 2 4D Color Flow Imaging
• Introduction ................................................................................... .... 44
• 4D Color Flow mode overview - Vivid 7 ...................................... .... 45
• Color Flow Volume rendering mode screen ............................... 45
• Color Flow Slice mode screen ................................................... 46
• 4D Color Flow mode controls ..................................................... 47
• Using 4D Color Flow mode - Vivid 7 ........................................... .... 54
• Full volume Color Flow acquisition ............................................. 55
• 6 Slice ........................................................................................ 57
• 4D Color Flow mode overview - EchoPAC PC ........................... .... 60
• Real time 4D Color flow acquisition ........................................... 54
• Volume rendering mode screen ................................................. 60
• Slice mode screen ...................................................................... 61
• 4D mode control panel ............................................................... 62
4D/Multiplane Imaging User's Manual 43 FD092081-01
4D Color Flow Imaging

Introduction

The 4D Color Flow Imaging mode enables real time volume rendering display with color flow data. 4D Color Flow imaging based on Full volume acquisition is available.
44 4D/Multiplane Imaging User's Manual
FD092081-01
4D Color Flow Imaging

4D Color Flow mode overview - Vivid 7

Color Flow Volume rendering mode screen

1. Volume rendering display with Color Flow
2. 2D Color Flow image in the azimuth plane
3. 2D Color Flow image in the elevation plane
4. Orientation window: displays a three-dimensional model with cut-planes position and orientation.
5. Soft menu controls (see page 51)
6. Trackball functions (see page 52)
Figure 2-1: 4D Color Flow screen (Volume rendering)
4D/Multiplane Imaging User's Manual 45 FD092081-01
4D Color Flow Imaging

Color Flow Slice mode screen

1. Cut-plane 1 (yellow)
2. Cut-plane 2 (white)
3. Cut-plane 3 (green)
4. Trackball functions (see page 52)
Figure 2-2: 4D Color Flow screen (Slice mode)
46 4D/Multiplane Imaging User's Manual
FD092081-01

4D Color Flow mode controls

Both Tissue and Color flow mode controls are available. Press
ACTIVE MODE to toggle between the two modes.
Alt. Alt. Alt. Alt.
4D Color Flow Imaging
3
4
Width
13
Update /
ck
ll
a
a
Tr
B
Menu
1. 4D key
2. Multiplane key
3. Assigned 4D rotaries: see next page.
4. Assigned 4D keys: see next page.
5. Zoom (see page 20)
6. Layout (see page 19): toggles the display between:
• Multi screen with volume/slice
• Single screen volume/slice
7. Soft menu (see page 51)
8. Trackball (see page 52)
• Rotate/translate volume rendering or selected cut-plane (Slice mode)
• Scroll through the cineloop
9. Clear: resets the orientation to default positions.
10. Angle: predefined orientations optimized for volume rendering.
11. Color Gain
12. 2D Gain
13. Color key
Figure 2-3: 4D Color Flow mode controls on the control panel
4D/Multiplane Imaging User's Manual 47 FD092081-01
4D Color Flow Imaging

Assigned rotaries and keys

Color Flow Volume rendering assigned controls
1. Assigned rotaries
• Scale
• Baseline R
• Volume size
MORE menu
• Translate R
Color Flow Slice mode assigned controls
1. Assigned rotaries
• Rotate R
• Translate (in Freeze only) R
Controls marked with R are also available in cine replay.
2. Assigned keys
• Slice R
• Invert
• Crop (in Freeze only) R
• Box (in Freeze only) R
• Cineloop (in Freeze only) R
• 4D CF Prepare
• Full Volume
• Color map R
• 6 Slice (in Full volume Freeze only) R (page 57)
MORE menu
• Up/Down R
• Flip R
• Orientation window R
• Cine rotate (in Freeze only) R
• Variance
• 4D Colorize R
2. Assigned keys
• Slice exit R
• Reference plane R
• Cineloop (in Freeze only) R
MORE menu
• Up/Down R
• Orientation window R
Figure 2-4: Color Flow Volume rendering and Slice mode assigned controls
48 4D/Multiplane Imaging User's Manual
FD092081-01
4D Color Flow Imaging

4D Color Flow mode assigned controls

Scale
Adjusts the repetition rate of the Doppler pulses transmitted to acquire the data for color flow mapping. The Scale (Nyquist limit) should be adjusted so that no aliasing occurs, while still having good resolution of velocities. The Nyquist limit should be somewhat above the maximum velocity found in the data.
Baseline
Adjusts the color map to emphasize flow either toward or away from the probe.
Volume size
The volume size control enables the adjustment of the color ROI. ROI adjustment may affect volume rate and/or resolution.
Slice/Slice exit
Toggles the display between Volume rendering and Slice mode.
Invert
Enables the color scheme assigned to positive and negative velocities to be inverted.
Variance
Controls the amount of variance data added to a color display. Variance enables computer-aided detection of turbulent flow (e.g. jets or regurgitation).
4D Colorize
(Volume rendering, Live and Replay)
Adjusts the volume rendering color from a color map menu.
Reference plane
(Slice mode, Live and Replay)
Toggles the cut-plane selection between cut-plane 1, 2 or 3.
4D/Multiplane Imaging User's Manual 49 FD092081-01
4D Color Flow Imaging
Crop (Free cropping)
See page 12.
Box (Box cropping)
See page 13.
4D CF Prepare and Full volume
Enables preparation and acquisition of a color flow full volume based on ECG triggered sub-volume acquisitions of color flow data. Depending on the adjustment, four or seven heart cycles are acquired. The 4D CF Prepare control enables the user to adjust the probe placement for optimized color flow acquisition from a bi-plane display before acquiring the Full volume with color flow data (see page 55).
Color maps
Displays a menu of color map options. Each color map is assigning different color hues to different velocities.
6Slice
6 Slice enables simultaneous display of six evenly distributed short axis views and two long axis views (in the azimuth and elevation planes) generated from an Apical full volume color flow acquisition (see page 57).

Additional 4D mode assigned controls

The following controls are available after pressing MORE.
Up/Down
Flips the volume 180 degrees.
Flip
Flips the view position.
Orientation window
Shows/hides the Orientation window.
50 4D/Multiplane Imaging User's Manual
FD092081-01
4D Color Flow Imaging
Cine rotate
Rotates back and forth the volume rendering.

Soft menu controls

Soft menu controls are related to image quality adjustment. These controls are accessed using the 4-way rocker on the control panel. Only the 4D controls are described in this section, refer to the system user manual for general imaging controls.
Number of cycles
(Volume rendering, Live)
Controls the number of heart cycles the ECG triggered color flow full volume acquisition is based on. Select between four or seven heart cycles.
Low velocity reject
Enables the extent of low velocity removal to be adjusted.
Flow transparency
(Volume rendering, Replay only)
Adjusts the color flow data transparency level. Higher setting may provide a better visualization of flow turbulences.
Tissue transparency
(Volume rendering, Replay only)
Adjusts the transparency level of the tissue structure. Maximum setting makes the tissue information invisible, leaving only color flow data displayed.
Flow direction
(Volume rendering, Replay only)
Enables to show/hide positive and/or negative velocities.
Smoothness
(Volume rendering, Replay only)
4D/Multiplane Imaging User's Manual 51 FD092081-01
4D Color Flow Imaging
Smooths color rendering. Smoothness affects continuity of color display and image noise.

Trackball controls

The trackball has multiple functions. The trackball functions are organized in several functional groups as shown in the table below.
The function selected is displayed in the lower right corner of the screen.
• Press within the active functional group. Groups with several functions are marked with a + symbol.
• Press
Function Description
SELECT to toggle between the trackball functions
TRACKBALL to toggle between the functional groups.
Orientation controls group
Press
SELECT to toggle between the controls.
Rotate Rotates the volume rendering/slice.
Translate Translates the volume rendering/slice.
Cineloop control group
Function Description
Scroll Scrolls through a cineloop.
Speed Adjust cine replay speed

Display controls

Color key
Starts/stops the color mode. In 4D color freeze, pressing COLOR show/hide the color data.
Layout key on the front panel
This control is working the same way as in 4D Mode (see page 19).
52 4D/Multiplane Imaging User's Manual
FD092081-01
4D Color Flow Imaging
Active mode Gain
With Color flow mode activated: gain amplifies the overall strength of echoes processed in the Color Flow window
Zoom
See page 20.
4D/Multiplane Imaging User's Manual 53 FD092081-01
4D Color Flow Imaging

Using 4D Color Flow mode - Vivid 7

Color Flow can be enabled in 2D mode before entering 4D mode or when in 4D mode.

Real time 4D Color flow acquisition

1. Select the 3V probe and cardiac application.
2. Create an examination.
CAUTION
3. Press The 4D Color Flow mode is started.
4. Acquire a volume.
5. Press the result.
6. Press
Tissue structures may obscure relevant flow information. If required, increase
Color flow data may obscure other relevant color flow information (e.g jet). If required adjust setting.
4D and COLOR.
FREEZE, 2D FREEZE and rotate the volume to check
IMG. STORE.
TISSUE TRANSPARENCY setting.
FLOW TRANSPARENCY
In some settings the volume rate can be less than 10 volumes per second. This may lead to small display mismatch between the color and tissue data. This is because of the rapid movement of structures (e.g. valves) compared to the time lag between the tissue and the color volume acquisitions.
54 4D/Multiplane Imaging User's Manual
FD092081-01
CAUTION
4D Color Flow Imaging

Full volume Color Flow acquisition

Full volume Color Flow acquisition is based on ECG triggered sub-volume acquisitions of color flow data. Depending on the adjustment, four or seven heart cycles are acquired.
Full volume Color Flow acquisition is done in two steps:
Step 1: positioning of the probe for optimal full volume acquisition with color flow data. This is done from a Bi-plane screen displaying azimuth and elevation planes.
Step 2: acquisition over seven heart beats.
ECG triggering acquisition may by nature contain artifacts. The triggering works by acquiring the whole tissue volume during the first heart beat, followed by a series of color sub-volumes that are stitched together (three or six). Consequently, the tissue volume is updated only each fourth or seventh heart beat, and only the color data may have stitching artifacts.
Artifacts may be caused by:
Movements of the probe caused by the operator during
acquisition.
Movements of the patient during acquisition.
Irregular heart rate during acquisition.
To validate the acquisition, perform a visual inspection in both the volume rendering and the elevation plane. Stitching artifacts are shown as visible transitions between the sub-volumes in the color flow data.
1. Connect the ECG device and make sure to obtain a stable ECG trace.
2. Adjust Num. Cycles to either four or seven cycles.
3. In 4D Color Flow mode, press
4D CF PREPARE.
A Bi-plane screen is displayed. Acquire an apical view.
4. Adjust the probe position to display optimal color flow image in both planes. You may:
• adjust
VOLUME SIZE to change the color ROI size.
• change the position of the ROI using the trackball (POS).
5. Ask the patient to hold her/his breath at end expiration. Keep the probe steady and press
FULL VOLUME. The Full
volume acquisition is started. the tissue color volume are build up over four/seven heart cycles. Make sure to keep the probe steady during the entire acquisition.
4D/Multiplane Imaging User's Manual 55 FD092081-01
4D Color Flow Imaging
Press FREEZE.
6. In Freeze, look for stitching artifacts in the color flow data in both the volume rendering and the elevation plane in the lower left window of the screen.
7. Press Depending on the result, you may press and make a new acquisition.
8. Press
2D FREEZE, rotate the volume to check the result.
4D CF PREPARE
IMG. STORE.
56 4D/Multiplane Imaging User's Manual
FD092081-01
4D Color Flow Imaging

6 Slice

The purpose of the 6 slice mode is to be able to measure the size of the minimal area of a regurgitant flow jet, typically through the mitral valves.
6 Slice enables simultaneous display of six short axis views generated from an Apical full volume Color Flow acquisition and two apical long axis views.
6 Slice is available only from a full volume color flow acquisition when in Freeze.
1. Acquire an apical Full volume view (see page 55).
2. Press
3. Press The 6 Slice screen is displayed (Figure 2-5). The six slices are evenly distributed and maximized in size for best assessment.
FREEZE.
6SLICE.
1. Upper slice
2. Lower slice
Figure 2-5: 6 Slice screen
If required, zoom in. All slices are zoomed in simultaneously.
4D/Multiplane Imaging User's Manual 57 FD092081-01
4D Color Flow Imaging
4. The following adjustments can be done:
• Adjust area (Figure 2-6).
1. Top: upper slice adjustment
2. Bottom: lower slice adjustment
Figure 2-6: Slice distribution adjustment
TOP and BOTTOM controls to change the slicing
• Adjust the AXIS 1 and AXIS 2 controls to rotate the slices backward/forward and sideways (Figure 2-7), to align the slices with the anatomical structure.
58 4D/Multiplane Imaging User's Manual
FD092081-01
1. Axis 1: slice rotation backward/forward
2. Axis 2: slice rotation sideways
4D Color Flow Imaging
Figure 2-7: Slice rotation adjustment
• Adjust ROTATE to rotate all six slices around the center axis.
• Adjust
5. Press
6. Press
TRANSLATE to move all six slices up or down.
IMG. STORE to save. 6 SLICE EXIT.
4D/Multiplane Imaging User's Manual 59 FD092081-01
4D Color Flow Imaging

4D Color Flow mode overview - EchoPAC PC

4D Color Flow mode acquisitions from Vivid 7 can be opened and post processed on the EchoPAC PC workstation. The controls are the same as in the Vivid 7 (see page 49). The procedures related to the 4D Color Flow mode are similar to the 4D mode (see page 32).

Volume rendering mode screen

1. Volume rendering display with Color Flow
2. Cut-plane 2 (white): 2D image in the azimuth plane
3. Cut-plane 3 (green): 2D image in the elevation plane
4. Orientation window: displays a three-dimensional model with cut-planes position and orientation.
5. 4D control panel (Color Flow Volume rendering mode)
Figure 2-8: 4D Color Flow screen (Volume rendering)
60 4D/Multiplane Imaging User's Manual
FD092081-01

Slice mode screen

4D Color Flow Imaging
1. Cut-plane 1 (yellow)
2. Cut-plane 2 (white)
3. Cut-plane 3 (green)
4. Orientation window: displays a three-dimensional model with cut-planes position.
5. 4D control panel (Color Flow Slice mode)
Figure 2-9: 4D Color Flow screen (Slice mode)
4D/Multiplane Imaging User's Manual 61 FD092081-01
4D Color Flow Imaging

4D mode control panel

Refer to ’4D Color Flow mode assigned controls’ on page 49 for information on the 4D Color Flow related controls. All other general imaging controls are described in the workstation User manual.
Figure 2-10: 4D Color Flow Control panel
62 4D/Multiplane Imaging User's Manual
FD092081-01
4D Color Flow Imaging
Rendering controls
In volume rendering, select Rendering to display the rendering controls.
Figure 2-11: Rendering and orientation panels
4D/Multiplane Imaging User's Manual 63 FD092081-01
4D Color Flow Imaging
64 4D/Multiplane Imaging User's Manual
FD092081-01

Multi-plane imaging

Chapter 3 Multi-plane imaging
• Introduction ................................................................................... .... 66
• Multi-plane mode overview - Vivid 7 ........................................... .... 67
• Bi-plane mode screen ................................................................ 67
• Tri-plane mode screen ............................................................... 68
• Multi-plane mode controls .......................................................... 69
• Using Multi-plane mode imaging - Vivid 7 .................................. .... 77
• Scan plane rotation .................................................................... 78
• Tilting scan plane 2 .................................................................... 78
• Zooming ..................................................................................... 79
• Multi-plane mode overview - EchoPAC PC ................................ .... 82
• Bi-plane mode screen ................................................................ 82
• Tri-plane mode screen ............................................................... 83
• The multi-plane control panel ..................................................... 84
• Working with multi-plane acquisitions - EchoPAC PC .............. .... 85
4D/Multiplane Imaging User's Manual 65 FD092081-01
Multi-plane imaging

Introduction

Multi-plane mode displays two (Bi-plane) or three (Tri-plane) rotated cross sections acquired simultaneously. This mode enables the investigation of anatomical structures from different angles.
Multi-plane mode enables the creation of a left ventricular volume reconstruction based on contours drawn from three cross sections at both end-systole and end-diastole with calculation of end-systolic and end-diastolic volumes and ejection fraction (see Chapter 4, ’Measurements and Analysis’ on page 87).
Combined to specially designed Stress echo protocols, the Multi-plane mode enables faster stress examination as two projections can be acquired simultaneously (see Chapter 5, ’Multi-plane Stress Echo’ on page 101).
Multi-plane mode is available from B-Mode, Color flow mode and TVI related modes.
66 4D/Multiplane Imaging User's Manual
FD092081-01
Multi-plane imaging

Multi-plane mode overview - Vivid 7

Bi-plane mode screen

1. Scan plane 1 (yellow): default reference scan plane. This scan plane is fixed, cannot be tilted or rotated.
2. Scan plane 2 (white): this scan plane is by default perpendicular to scan plane 1 along the scanning axis. This scan plane can be either tilted or rotated.
3. Geometric model: displays both scan planes in a projection and the rotation angle (A1) and tilt angle (T) values for the scan plane 2 relative to the scan plane 1.
4. Scan plane marker for scan plane 2: crossing line of scan plane 2 in scan plane 1.
5. Trackball functions (see page 73)
Figure 3-1: The Bi-plane imaging mode screen
4D/Multiplane Imaging User's Manual 67 FD092081-01
Multi-plane imaging

Tri-plane mode screen

1. Scan plane 1 (yellow): default reference scan plane. This scan plane is fixed, cannot be rotated.
2. Scan plane 2 (white): this scan plane can be rotated using the trackball.
3. Scan plane 3 (green): this scan plane can be rotated using the trackball.
4. Geometric model: displays all the scan planes in a projection.
5. Navigator: displays rotation angle values for the scan planes 2 (A1) and 3 (A2) relative to the scan plane 1.
6. Trackball functions (see page 73)
Figure 3-2: The Tri-plane imaging mode screen
68 4D/Multiplane Imaging User's Manual
FD092081-01

Multi-plane mode controls

Alt. Alt. Alt. Alt.
Multi-plane imaging
2
3
4
Width
1. Multi-plane key
2. Mode specific assigned rotaries: see next pages.
3. Mode specific assigned keys: see next pages.
4. Zoom
5. Trackball
• Tilt scan plane 2 (Bi-plane mode only).
• Rotate scan plane 2 and 3
• Zoom adjustment.
• Scroll through the cineloop.
6. Clear: resets the rotation and/or tilt angle of scan plane 2 and 3.
7. Angle: in Bi-plane mode, rotates scan plane 2 to a pre-defined angle relatively to scan plane 1.
Figure 3-3: The Multi-plane controls on the control panel
4D/Multiplane Imaging User's Manual 69 FD092081-01
Multi-plane imaging
Assigned rotaries and keys
¨Multi-plane B mode assigned controls
1. Assigned rotaries
• Angle 1 (Tri-plane)
• Angle 2 (Tri-plane)
• Width (Bi-plane)
• Frequency (Bi-plane)
• Focus position
• Frame rate
ALT menu
• Angle 1 (Bi-plane)
• Width (Tri-plane)
• Frequency (Tri-plane)
Multi-plane Color flow mode assigned controls
1. Assigned rotaries
• 2D Width
• Scale
• Baseline R
• Frame rate
MORE menu
• Angle 1
• Angle 2 (Tri-plane)
2. Assigned keys
• Up/Down R
• Reference Plane R
• Tri-plane
• Cineloop (in Freeze only) R
• B Color maps R
ALT menu
• Curved Anatomical M Mode R
• Q Analysis (in Freeze only)
• Anatomical M Mode
• Tissue Tracking
• Strain rate
• Strain
ALT + MORE menu
• TSI
2. Assigned keys
• Invert R
• Variance R
• Reference Plane R
• Tri-plane
• Cineloop (in Freeze only) R
• Color map R
MORE menu
• Up/Down R
ALT menu
• Same as Multi-plane B mode
ALT + MORE menu
• Same as Multi-plane B mode
Controls marked with R are also available in cine replay.
Figure 3-4: Multi-plane B mode and CF modes assigned controls
70 4D/Multiplane Imaging User's Manual
FD092081-01
Multi-plane TVI mode assigned controls
Multi-plane imaging
1. Assigned rotaries
• 2D Width
• Scale
• Baseline R
• Frame rate
MORE menu
• Angle 1
• Angle 2 (Tri-plane)
Multi-plane Tissue Tracking mode assigned controls
1. Assigned rotaries
• 2D Width
• Track scale R
• Track end R
• Frame rate
MORE menu
• Angle 1
• Angle 2 (Tri-plane)
2. Assigned keys
• Invert R
• TSI R
• Reference Plane R
• Tri-plane
• Cineloop (in Freeze only) R
• TVI visible R
ALT menu
• Curved Anatomical M Mode R
• Q Analysis (in Freeze only)
• Anatomical M Mode
• Tissue Tracking
• Strain rate
• Strain
ALT + MORE menu
• TSI
2. Assigned keys
• Invert R
• TSI R
• Reference Plane R
• Tri-plane
• Cineloop (in Freeze only) R
• Color maps R
ALT menu
• Same as Multi-plane TVI
Controls marked with R are also available in cine replay.
Figure 3-5: Multi-plane TVI and TT modes assigned controls
4D/Multiplane Imaging User's Manual 71 FD092081-01
Multi-plane imaging
Multi-plane Strain imaging mode assigned controls
1. Assigned rotaries
• 2D Width
• Strain scale R
• Strain end R
• Frame rate
MORE menu
• Angle 1
• Angle 2 (Tri-plane)
Multi-plane Strain rate imaging mode assigned controls
1. Assigned rotaries
• 2D Width
• SRI scale R
• Frame rate
MORE menu
• Same as Multi-plane Strain imaging
Multi-plane Tissue Synchronization Imaging mode assigned controls
1. Assigned rotaries
• 2D Width
• Frame rate
MORE menu
• TSI start
• TSI end
2. Assigned keys
• Invert R
• TSI R
• Reference Plane R
• Tri-plane
• Cineloop (in Freeze only) R
• Color maps R
ALT menu
• Curved Anatomical M Mode
• Q Analysis (in Freeze only)
• Anatomical M Mode
• Tissue Tracking
• Strain rate
• Strain
ALT + MORE menu
• TSI
2. Assigned keys
• Same as Multi-plane Strain imaging
2. Assigned keys
• TSI R
• Reference Plane R
• Tri-plane
• Cineloop (in Freeze only)
• Color maps R
ALT menu
• Same as Multi-plane Strain imaging
Controls marked with R are also available in cine replay.
Figure 3-6: Multi-plane SI, SRI and TSI modes assigned controls
72 4D/Multiplane Imaging User's Manual
FD092081-01
Multi-plane imaging
Multi-plane mode assigned controls
This section describes only the Multi-plane mode controls. The scanning mode controls are described in the system User manual.
Tri-plane
(All modes, Live)
Toggles between Bi-plane and Tri-plane mode.
Reference Plane
(All modes, Live and Replay)
In Tri-plane mode, toggles the reference plane between scan plane 1, 2 or 3.
The reference plane may also be selected using the trackball when the Pointer trackball tool selected (see below).
Trackball controls
The trackball has multiple functions depending on the scanner’s state (Live/Freeze, zoom/unzoom, Bi-plane/Tri-plane mode and selected scanning mode). The trackball functions are organized in several functional groups as shown in the table below. The function selected is highlighted in the lower right corner of the screen.
• Press within the active functional group.
• Press
SELECT to toggle between the trackball functions
TRACKBALL to toggle between the functional groups.
Scan planes rotation controls group
Function State Description
Press
SELECT to toggle between the controls.
Pos Bi-plane, Live, Unzoom Tilts the scan plane 2 (white) around
the top of the sector.
Angle 1 Bi-plane/Tri-plane, Live,
Unzoom
4D/Multiplane Imaging User's Manual 73 FD092081-01
Adjusts the angle between the scan planes 2 (white) and scan plane 1 (yellow). Scan plane 2 is rotated around the crossing line between the scan planes. Scan plane 1 is fixed.
Multi-plane imaging
Scan planes rotation controls group
Function State Description
Angle 2 Tri-plane, Live, Unzoom Adjusts the angle between the scan
planes 3 (green) and scan plane 1 (yellow). Scan plane 1 is fixed.
Scan plane selection
Function State Description
Pointer All states Selects the reference scan plane.
Zoom mode controls group
Function State Description
Press
SELECT to toggle between the controls.
Pos Zoom, Live/Freeze Moves the zoom area.
Size Zoom, Live/Freeze, 2D Resizes the zoom area.
Color/TVI modes ROI controls group
Function State Description
Press
SELECT to toggle between the controls.
Pos CFM, TVI modes,
Live/Freeze, Zoom/Unzoom
Size CFM, TVI modes,
Live/Freeze, Zoom/Unzoom
Moves the color sector.
In Bi-plane CFM mode, if scan plane 1 (yellow) is selected as reference scan plane, both color sectors are moved. If scan plane 2 (white) is selected as reference scan plane its color sector is moved independently. In Tri-plane CFM mode, all sectors are moved at the same time.
Resizes the color sector.
All sectors are resized at the same time
74 4D/Multiplane Imaging User's Manual
FD092081-01
Multi-plane imaging
Cineloop control group
Function State Description
Scroll Freeze, Zoom/Unzoom Scrolls through a cineloop.
Display controls
Layout key
In Bi-plane mode, toggles the display between the default Bi-plane dual screen and a single screen showing the selected scan plane.
In Tri-plane mode, toggles the display between the default Tri-plane quad screen, a quad screen with enlarged Geometric model and a single screen showing the selected scan plane.
1. Display options for Bi-plane mode
2. Display options for Tri-plane mode
Figure 3-7: Multi-plane mode display options
4D/Multiplane Imaging User's Manual 75 FD092081-01
Multi-plane imaging
Zoom
Display zoom
Activated and adjusted by rotating the orientation preview showing the outlined magnified area is displayed in the upper right corner of the screen. The position and size of the magnified area are adjusted with the trackball when in B mode (see page 79).
High Resolution (HR) zoom
HR zoom concentrates the image processing to a magnified user selectable portion of the image, resulting in an improved image quality in the selected image portion.
HR zoom is activated and adjusted by pressing and rotating the zoom rotary ( ). An orientation preview showing the outlined
magnified area is displayed in the upper right corner of the screen. The position and size of the magnified area are adjusted with the trackball.
ZOOM rotary ( ). An
Other controls
Clear key
Resets all scan planes to the default position.
Angle key
In Bi-plane mode, toggles the position of scan plane 2 between the default position and a pre-defined angle relatively to scan plane 1.
76 4D/Multiplane Imaging User's Manual
FD092081-01
Multi-plane imaging

Using Multi-plane mode imaging - Vivid 7

1. Select the 3V probe.
2. Create an examination.
3. Press The Bi-plane B mode screen is displayed.
4. Press acquisition.
5. Adjust the angle increment between scan planes as described below.
6. If in Bi-plane mode, adjust the scan plane 2 tilt angle as described below.
7. To activate another scanning mode:
• Press
• Press
• In TVI mode, press
• Press
8. Zoom in the structure of interest (see page 79). Note: scan plane rotation or tilting cannot be done when in zoom mode.
9. Press
MP on the control panel.
TRI-PLANE to select between bi-plane or tri-plane
COLOR to activate CF mode.
TVI to activate TVI mode.
TSI to activate TSI mode.
ALT and either TISSUE TRACKING, STRAIN or STRAIN
RATE to activate the corresponding mode in Multi-plane
imaging.
IMG. STORE to save the acquisition.
4D/Multiplane Imaging User's Manual 77 FD092081-01
Multi-plane imaging

Scan plane rotation

The angle increment of scan plane 2 or 3 relative to the scan plane 1 (fixed) can be adjusted with the trackball. The rotation is done around the crossing line between the scan planes. Scan plane rotation is available only in Unzoom live mode.
1. Press selected:
Angle 1: rotation of scan plane 2 (white)
Angle 2: rotation of scan plane 3 (green, Tri-plane)
2. Use the trackball to rotate the corresponding scan plane around the probe center axis.
SELECT until the desired trackball function is
Figure 3-8: Rotation of Scan plane 2 (Bi-plane mode)
3. To reset the scan planes to the original default position,
CLEAR.
press

Tilting scan plane 2

In Bi-plane mode, the scan plane 2 can be tilted around the top of the scanning sector using the trackball. Tilting is available only in Unzoom live mode.
1. In Bi-plane mode, press Pos is selected.
2. Use the trackball to tilt the scan plane 2.
SELECT until the trackball function
78 4D/Multiplane Imaging User's Manual
FD092081-01
Multi-plane imaging
Figure 3-9: Tilting of scan plane 2 (Bi-plane mode)
3. To reset the scan planes to the original default position,
CLEAR.
press

Zooming

Activating the zoom
1. Rotate the ZOOM knob clockwise on the control panel. All scan plane images are zoomed in. A Navigation window is displayed with a frame highlighting the zoomed area (see Figure 3-10).
4D/Multiplane Imaging User's Manual 79 FD092081-01
Multi-plane imaging
1. Magnified scan planes
2. Navigation window in Zoom mode
3. Zoomed area
Figure 3-10: Zoom mode screen (Tri-plane mode)
Adjusting the zoomed area
The zoomed are can be moved within the sector or resized.
1. In zoom mode, press
SELECT to toggle the trackball function
to Pos and use the trackball to freely move the zoomed area within the scan plane.
2. Press
SELECT to toggle the trackball function to Size and
use the trackball to adjust the size of the zoomed area (B mode only).
80 4D/Multiplane Imaging User's Manual
FD092081-01
Multi-plane imaging
Note: If the trackball function Ptr (mouse pointer) or Scroll (when in Freeze) is selected, press trackball functions Pos or Size is selected.
TRACKBALL until the
4D/Multiplane Imaging User's Manual 81 FD092081-01
Multi-plane imaging

Multi-plane mode overview - EchoPAC PC

Multi-plane acquisitions from Vivid 7 can be opened and post processed on the EchoPAC PC workstation. This section describes the controls and procedures related to Multi-plane mode on the workstation. Refer to the EchoPAC PC User manual about general use of the workstation.

Bi-plane mode screen

1. Scan plane 1 (yellow): default reference scan plane.
2. Scan plane 2 (white): the tilt or rotation angle is set during acquisition.
3. Geometric model: displays both scan planes in a projection and the rotation angle (A1) and tilt angle (T) values for the scan plane 2 relative to the scan plane 1.
4. Scan plane marker for scan plane 2: crossing line of scan plane 2 in scan plane 1.
5. Multi-plane control panel
Figure 3-11: The Bi-plane mode screen
82 4D/Multiplane Imaging User's Manual
FD092081-01

Tri-plane mode screen

Multi-plane imaging
1. Scan plane 1 (yellow): default reference scan plane.
2. Scan plane 2 (white): the rotation angle is set during acquisition.
3. Scan plane 3 (green): the rotation angle is set during acquisition.
4. Geometric model: displays all three scan planes in a projection and/or reconstructed volume.
5. Navigator: displays rotation angle values for the scan planes 2 (A1) and 3 (A2) relative to the scan plane 1.
6. Multi-plane control panel
Figure 3-12: The Tri-plane mode screen
4D/Multiplane Imaging User's Manual 83 FD092081-01
Multi-plane imaging

The multi-plane control panel

The controls in multi-plane mode are similar to the controls for the associated scanning mode (B mode, Color or TVI related modes). Refer to the workstation User manual for additional informations on these controls.
Figure 3-13: The Multi-plane Control panel
84 4D/Multiplane Imaging User's Manual
FD092081-01
Multi-plane imaging
Working with multi-plane acquisitions ­EchoPAC PC
1. In the Search/Create patient window (Archive menu), select a patient record with multi-plane acquisitions. The Examination List window is displayed.
2. Select the desired examination and open (double-click) a multi-plane acquisition from the clipboard. OR Press Image browser and open (double-click) the multi-plane acquisition in the Image browser screen. The Multi-plane mode screen (bi-plane or tri-plane) is displayed.
3. Perform image optimization and measurements.
4. Press Store to save the changes.
4D/Multiplane Imaging User's Manual 85 FD092081-01
Multi-plane imaging
86 4D/Multiplane Imaging User's Manual
FD092081-01

Measurements and Analysis

Chapter 4 Measurements and Analysis
• Introduction ................................................................................... .... 88
• Left ventricular volume measurements ...................................... .... 89
• Tri-plane acquisition ................................................................... 89
• Full volume acquisition ............................................................... 92
• Rotation of the Volume reconstruction ....................................... 93
• Bi-plane acquisition .................................................................... 94
• TSI surface model ......................................................................... .... 96
• Quantitative analysis .................................................................... .... 98
• Starting Quantitative analysis from a multi-plane acquisition ..... 98
4D/Multiplane Imaging User's Manual 87 FD092081-01
Measurements and Analysis

Introduction

All cardiac measurements available in 2D are also available in 4D (when in Slice mode) and Multi-plane modes. In addition, the 4D and Multi-plane modes enable the creation of a left ventricular volume reconstruction based on contours drawn from three cross-sections at both end-systole and end-diastole with calculation of the end-systolic and end-diastolic volumes and the ejection fraction.
Left ventricular volume measurement may also be performed in Bi-plane mode based on contours drawn at both end-systole and end-diastole.
Refer to the scanner’s or the workstation’s User Manual for more information about 2D measurements.
From the TSI Tri-plane mode the user can create a TSI surface model of the left ventricle.
All these features are available on both Vivid 7 and EchoPAC PC.
88 4D/Multiplane Imaging User's Manual
FD092081-01
Measurements and Analysis

Left ventricular volume measurements

Tri-plane acquisition

This procedure describes the calculation and reconstruction of the left ventricular volume from a Tri-plane acquisition.
1. In Tri-plane mode, acquire an Apical 4 chamber view in scan plane 1 (yellow).
2. Rotate scan plane 2 and 3 to display an Apical 2 chamber view in scan plane 2 (white) and an Apical long axis view in scan plane 3 (green).
3. Press
4. Press The Measurement menu is displayed.
5. In the Measurement menu select Volume and Tri-plane. The Measurement screen is displayed with the Ejection fraction tool selected (see Figure 4-1). The end diastolic frame of the current cardiac cycle is displayed and the cursor is moved to the reference scan plane.
FREEZE.
MEASURE.
4D/Multiplane Imaging User's Manual 89 FD092081-01
Measurements and Analysis
1. Ejection fraction tool for Tri-plane
2. Scan plane 1 (yellow): Apical 4 chamber view
3. Scan plane 2 (white): Apical 2 chamber view
4. Scan plane 3 (green): Apical long axis view
5. Volume reconstruction
Figure 4-1: The Tri-plane measurement screen
6. If desired, press LAYOUT twice to display the reference scan plane in a single screen.
7. Place the cursor to the start point for the trace.
8. Press
SELECT and draw a contour of the left ventricle.
To edit the contour while drawing:
• Follow the contour backward to erase it and redraw.
• Press
UNDO or BACKSPACE to erase the contour
stepwise and redraw.
• Press
9. Press
DELETE to remove the entire contour and redraw.
SELECT to complete the contour.
The cursor is automatically moved to the next scan plane. The crossing point of the trace done in the first scan plane is marked in the second scan plane.
90 4D/Multiplane Imaging User's Manual
FD092081-01
Measurements and Analysis
10. Draw a contour of the left ventricle in scan plane 2 and 3 following the same procedure. When the last end-diastolic contour is drawn, an end-systolic frame is automatically displayed. The system automatically enters scroll mode. Using the trackball, ensure that the correct end-systolic frame is displayed. Press
11. Repeat steps 7 to 10 to draw a contour of the left ventricle at end-systole in all the scan planes. The measurement results, including end-diastolic and end-systolic volumes and the left ventricular ejection fraction, are displayed in the Measurement result table. Note: Other measurements may be displayed by configuring the Measurement menu, refer to the scanner’s or workstation’s User manual for more information about Measurement menu configuration.
12. If in single screen mode, press Volume reconstruction of the left ventricle in the Geometric model.
13. Press model. The Volume reconstruction can be rotated in all directions (see page 93).
SELECT to leave scroll mode.
LAYOUT to display the
LAYOUT again to display an enlarged Geometric
4D/Multiplane Imaging User's Manual 91 FD092081-01
Measurements and Analysis

Full volume acquisition

This procedure describes the calculation and reconstruction of the left ventricular volume from a Full volume acquisition.
ECG triggering acquisition may by nature contain artifacts, that may have impact on the measurements.
CAUTION
Artifacts may be caused by:
Movements of the probe caused by the operator during
acquisition.
Movements of the patient during acquisition.
Irregular heart rate during acquisition.
See recommendations on page 23 to avoid stitching artifacts during full volume acquisition.
It is recommended to specify in the Comments for the examination that the measurements are performed in a Full volume acquisition.
1. Using the Full volume acquisition mode (see page 23), acquire a 4D Apical 4 chamber image.
2. Press
FREEZE.
3. Orientate the reference cut-plane to display an Apical 4 chamber (see ’Rotating/Translating the 4D image’ on page 25).
4. Press
MEASURE.
The Measurement menu is displayed.
5. In the Measurement menu select Vol u m e and Tri-plane. The Measurement screen is displayed with the Ejection fraction tool selected.
6. Follow the procedure described in ’Tri-plane acquisition’ on page 89from step 6.
92 4D/Multiplane Imaging User's Manual
FD092081-01
Measurements and Analysis

Rotation of the Volume reconstruction

The volume reconstruction displayed in the Geometric model can be rotated in any directions.
1. Place the pointer in the Geometric model.
2. Press and hold down the volume reconstruction.
SELECT and use the trackball to rotate
4D/Multiplane Imaging User's Manual 93 FD092081-01
Measurements and Analysis

Bi-plane acquisition

This procedure describes the calculations of the left ventricular volume from a Bi-plane acquisition. The volume calculation is based on the Method of disk.
1. In Bi-plane mode, acquire an Apical 4 chamber view in scan plane 1(yellow).
2. If required, rotate scan plane 2 to display an Apical 2 chamber view.
3. Press
4. Using the trackball, scroll through the cineloop to display the end-diastolic frame.
5. Press The Measurement menu is displayed.
6. In the Measurement menu select Vol u m e and Bi-plane. The Trace tool for the Left ventricular end-diastolic volume for the Apical 4 chamber view is selected (see Figure 4-2).
FREEZE.
MEASURE.
1. Trace tools
Figure 4-2: The Volume measurement screen (Bi-plane)
7. In scan plane 1 (yellow), place the cursor to the start point for the trace.
94 4D/Multiplane Imaging User's Manual
FD092081-01
Loading...