The information in this manual applies only to MACTM 3500 Resting ECG Analysis Systems with product code SCA. It does
not apply to earlier software versions. Due to continuing product innovation, specifications in this manual are subject to
change without notice.
MUSE ,CASE, MAC, MARS, MULTI-LINK, and 12SL are trademarks owned by GE Medical Systems Information Technologies, Inc., a General Electric Company going to market as GE Healthcare. All other trademarks are owned by their
respective owners.
Each page of the document has the document part number and revision letter at the
bottom of the page. The revision letter identifies the document’s update level. The
revision history of this document is summarized in the table below.
RevisionDateComment
Introduction: Manual Information
Table 1. Revision History, PN 2021337-036
A1 August 2006Initial release of document.
B14 December 2006Added FRUs for MobileLink Silex serial server.
C1 February 2007Edited EMC section for IEC 60601-2-51 compliance.
D10 October 2008Added -007 board, 009C SD Card and -002 Comm
PC board. Included Functional Checkout Procedure.
E08 January 2010Added PN 2022328-002 PCB MAC 3500 CAMV2.
F10 May 2010Revised security on electronic file.
Manual Purpose
G12 December 2010Revised to add service disclaimer addendum.
H14 August 2012Added 2022332-003 to Parts List.
J19 June 2013Updated for the new CPU (801212-008),
communcations board (2022332-004) , and display
assembly (2026799-002).
K21 February 2014Updated several part numbers with new ROHS-
compliant versions.
L15 February 2019
This manual supplies technical information for service representative and technical
personnel so they can maintain the equipment to the assembly level. Use it as a
guide for maintenance and electrical repairs considered field repairable. Where
necessary the manual identifies additional sources of relevant information and or
technical assistance.
• Updated to remove CE markup.
• Updated to remove Authorized European
representative.
• Added the Instructions for Use.
• Updated UL symbol.
See the operator’s manual for the instructions
safely in accordance with its function and intended use.
necessary to operate the equipment
Intended Audience
This manual is intended for the person who uses, maintains, or troubleshoots this
equipment.
The following documents are referenced in this manual and provide additional
information that may be helpful in the installation, configuration, maintenance, and
use of this product.
Part NumberTitle
2021337-035MAC™ 3500 Resting ECG Analysis System Operator Manual
2036070-006Marquette™ 12SL™ ECG Analysis Program Physician's Guide
2020299-025LAN Option for MAC™ Resting ECG Systems Installation and
roubleshooting Guide
T
2044854–112Modular MAC™ ECG Trolley Service Manual
2056914–001Modular MAC™ ECG Trolley Assembly I
2056914–002Modular MAC™ ECG Trolley Assembly I
Warnings, Cautions, and Notes
The terms danger, warning, and caution are used throughout this manual to point out
hazards and to designate a degree or level or seriousness. Familiarize yourself with
their definitions and significance.
Hazard is defined as a source of potential injury to a
TermDefinition
DANGERIndicates an imminent hazard which, if not avoid
serious injury.
WARNINGIndicates a potential hazard or unsafe practice which, if not avoided, could
result in death o
CAUTIONIndicates a potential hazard or unsafe practice which, if not avoided, could
result in minor personal injury or product/property damage.
NOTEProvides application tips or other useful information to a
the most from your equipment.
nstructions (US)
nstructions (Non-US)
person.
ed, will result in death or
r serious injury.
ssure that you get
1-4MAC™ 3500 Resting ECG Analysis SystemRevision L
2021337-036
Page 13
Safety Messages
Introduction: Safety Messages
Additional safety messages may be found throughout this manual that provide
appropriate safe operation information.
DANGER
Do not use in the presence of flammable anesthetics.
WARNINGS
This is Class 1 equipment. The mains plug must be connected to
an appropriate power supply.
Operate the unit from its battery if the integrity of the protective
earth conductor is in doubt.
CAUTIONS
This equipment contains no serviceable parts. Refer servicing to
qualified service personnel.
U.S. Federal law restricts this device to the sale by or on the order
of a physician.
Responsibility of the Manufacturer
GE Medical Systems Information Technologies is responsible for the effects of
safety, reliability, and performance only if:
Assembly operations, extensions, readjustments, modifications, or repairs
are carried out by pers
The electrical installation of the relevant room complies with the
requirements
The equipment is used in accordance with the instructions for use.
The intended use of this device is to record ECG signals from surface ECG
electrodes. This device can analyze, record, and store electrocardiographic
information from adult and pediatric populations. This data can then be computer
analyzed with various algorithms such as interpretive ECG and signal averaging for
presentation to the user.
This device is intended for use under the direct supervision of
practitioner.
Failure on the part of the responsible individual, hospital, or institution using this
equipment to implement
equipment failure and possible health hazards.
To ensure patient safety, use only parts and accessories manufactured
recommended by GE Healthcare.
Contact GE Healthcare for i
equipment that are not recommended in this manual.
If the installation of this equipment, in t
the source must be a center-tapped, 240 V, single-phase circuit.
Parts and accessories used must meet the requireme
series safety standards, and/or the system configuration must meet the requirements
of the IEC 60601-1-1 medical electrical systems standard.
The use of ACCESSORY equipment not complying with the equivalent safety
requirements of this equipment may
system. Consideration relating to the choice shall include:
use of the accessory in the PATIENT VICINITY; and
evidence that the safety certification of the ACCESSORY has been
performed
60
601-1-1 harmonized national standard.
a satisfactory maintenance schedule may cause undue
nformation before connecting any devices to this
he USA, will use 240 V rather than 120 V,
nts of the applicable IEC 60601
lead to a reduced level of safety of the resulting
in accordance to the appropriate IEC 60601-1
a licensed health care
or
and/or IEC
1-6MAC™ 3500 Resting ECG Analysis SystemRevision L
2021337-036
Page 15
Equipment Symbols
The following symbols may appear on the product or its packaging.
Introduction: Equipment Symbols
Type BF equipment. The acquisition module is protected from defibrillation
shoc
ks.
Alternating current.
Equipotential.
Charge the battery. The flashing amber LED next to this symbol indicate
you must connect the system to AC power to re-charge the battery
LAN port for connecting an Ethernet cable with a standard RJ-45 jack.
Internal modem port for connecting a phone line with a standard RJ-11 jack.
Do NOT throw the battery into the garbage.
Recycle the battery.
Consult accompanying documents.
This position of the switch removes battery power from the equipment.
Classified with respect to electric
specified hazards only in accordance with UL 60601-1, CAN/CSA C22.2
No. 601-1, CAN/CSA C22.2 No. 601-2-25, EN 60601-2-25, EN 60601-1-1,
IEC 60601-1-2: 2001.
shock, fire, mechanica
l, and other
s
.
To reduce the risk of electric shock, do
servicing to qualified personnel.
This symbol indicates that the waste of electrical and electronic eq
must not be dispo
separately. Please contact an authorized representative of the manufacturer
for information concerning the decommissioning of your equipment.
sed as unsorted municipal waste and must be collected
NOT remove co
ver (or back). Refer
uipment
Page 16
Introduction: Equipment Symbols
This product consists of devices that may contain mercury, which must be
recycled or disposed of in accordance with local, state, or country laws.
(Within this system, the backlight lamps in the monitor display contain
mercury.)
Manufacturer name and address.
Consult instructions for use.
PCT. GOST marking symbolizing conformity with applicable Russian
Gosstandart tech
nical and safety standards.
1-8MAC™ 3500 Resting ECG Analysis SystemRevision L
2021337-036
Page 17
Service Information
9A
SERIAL NUMBER
LABEL
SERIAL NUMBER
LABEL
Service Requirements
Refer equipment servicing to GE authorized service personnel only. Any
unauthorized attempt to repair equipment under warranty voids that warranty.
It is the user’s responsibility to report the need for service to GE or to one of their
authorized agents
Equipment Identification
The serial number label is located inside the device as shown in the following
illustration.
2-2MAC™ 3500 Resting ECG Analysis SystemRevision L
2021337-036
Page 21
Equipment Overview: General Description
104A
CD
A
E
B
F
General Description
The MAC™ 3500 Resting ECG Analysis System is a 12-lead, 12-channel system
with a 6.5 inch (165 mm) diagonal display, active patient cable, and battery
operation. There are also options for communication capabilities.
Side View
NameDescription
AkeyboardPress the keyboard keys to control the system or to enter
data.
Bdisplay screenView the waveform and text data.
Cmodem portConnect the telephone cable here (optional feature)
DLAN portConnect to the LAN here (optional feature)
The green LED on the right side of this port indicates
a good ethernet link.
The amber LED on the left side of this port flashes to
3-2MAC™ 3500 Resting ECG Analysis SystemRevision L
2021337-036
Page 27
Introduction
Installation: Introduction
This chapter describes how to assemble the MAC 5500 system and optional
accessories on the optional MAC Series Trolley, and it identifies the requirements
and configurations for using select devices with the ST option.
Attaching the MAC device to the MAC Series trolley
Attaching the optional external modem
Attaching the magnetic card reader
Attaching the bar code reader
NOTE
These instructions describe the process only
for the MAC Series trolley (1). For
general assembly instructions for the modular MAC trolley (2), refer to the
Modular MAC Trolley Assembly Instructions identified in “Related
Documentation” on page 1-4.
3-4MAC™ 3500 Resting ECG Analysis SystemRevision L
2021337-036
Page 29
Installation: General Assembly
MAC 3500
Patient Cable
Arm and Holder
Swivel
Casters
Trolley
Serial Number
105A
Front Cover
Locking
Casters
Use the following photograph of a complete assembly as a reference when attaching
the optional accessories.
NOTE
Because the optional trolley is made by another vendor for GE,
the serial
number format is different from that shown in “Serial Number Format” on
page 1-10.
The optional MAC 3500 Trolley can be assembled for one of two heights, 92.07 cm
(36.25 inches) or 84.45 cm (33.25 inches). The trolley is normally shipped at the
92.07 cm (36.25 inches) height but can be changed to fit your needs. To change to
the lower height, use the following steps:
1.Tip the trolley on its side and using a 1/2-inch socket, remove the 4 outer 1/
2-inch
Installation: General Assembly
bolts and slide the base assemble up on the column.
2.Remove the remaining bolts and mounting plate.
3.Flip the mounting plate and
reverse the procedure.
3-6MAC™ 3500 Resting ECG Analysis SystemRevision L
2021337-036
Page 31
Installation: General Assembly
110A
109A
CAUTION
Do not over tighten. Over tightening the bolts may cause them to
strip.
The magnetic card reader and its mounting bracket are assembled and ready to
install on the trolley. Parts are included for two different trolley styles. Disregard
and do not use the parts indicated in the following illustration.
To install the magnetic card reader and its m
ounting bracket on the trolley, complete
the following steps:
1.Remove both end panels by pulling out and up at the bottom.
2.Using a Phillips screw driver, fasten the card
hand
le. Align with holes provided under front handle.
reader assembly under the front
3-12MAC™ 3500 Resting ECG Analysis SystemRevision L
2021337-036
Page 37
Installation: General Assembly
Cable Routing
Rear View
34A
77A
70A
3.Route the cable around the trolley column towards the rear as shown below.
4.At the front, hold the cable to the side
th
e panel.
so it clears the front panel as you replace
5.Plug the cable connector into port A then
replace the back panel.
6.Refer to the MAC 3500 Operator’s Manual for information on using the
Magnetic Card
The barcode reader and its mounting bracket are ready to install on the trolley. To
install the Bar Code Reader and its cable mounting bracket on the trolley, complete
the following steps:
1.Fasten the cable clamp bracket to the underside
Phill
ips screw driver and the self-tapping screws provided.
NOTE
DO NOT overtighten. Overtightening the s
strip and clamp to fail.
of the rear handle using a
crew may cause the screw to
3-14MAC™ 3500 Resting ECG Analysis SystemRevision L
2021337-036
Page 39
Installation: General Assembly
Internal Access
Button
9A
Port A
Not enough cable slack.
Correct amount of cable slack.
2.Press the internal access button to open the MAC 3500, then plug the cable
conn
ector into port A. Opening the MAC 3500 before
allo
ws you to place the correct amount of slack to free the cable from stress
the MAC 3500 needs to be re-opened.
when
attaching the cable clamp
3.Next fasten the cable and clamp to the clamp
Observe that there is e
nough slack to allow free movement of the cable when re-
bracket, then close the MAC 3500.
opening the MAC 3500.
4.Refer to the MAC 3500 Operator’s Manual for information on how to use the
barcode reader.
4-2MAC™ 3500 Resting ECG Analysis SystemRevision L
2021337-036
Page 43
Troubleshooting: Assembly Descriptions
Assembly Descriptions
The troubleshooting information in this chapter helps you narrow service problems
to one of the replaceable assemblies. These assemblies, illustrated in the following
diagrams, are discussed in more detail in the remainder of the chapter along with
replacement procedures.
PCB Block Diagram
The following diagram illustrates the logical relationship of the system components.
The following illustration diagrams the connections for the LVDS/LED display
assembly (2026799-002), and the table that follows it identifies those connections.
ItemGE Part NumberDesciption
A12024701-001ASSY DISPLAY CABLE MAC3500 (CMOS)
A22062075-001LVDS/LED LCD PANEL
A42024701-001MAC3500 PWR CABLE MAIN BOARD TO LVDS
A52061540-001PWA MAC5500 LVDS DRV BRD ROHS
A62059277-001ASSY MAC3500 BACKLIT CABLE-AUO
A72059322-001MAC3500 LCD CABLE
4-6MAC™ 3500 Resting ECG Analysis SystemRevision L
2021337-036
Page 47
Troubleshooting: General Fault Isolation
General Fault Isolation
Power-up Self-test
See the MAC 3500 Operator’s Manual, Chapter 2, “Equipment Overview: Getting
Started” to verify operation.
On power-up, the system automatical
good, the start up screen displays. If the equipment is not working properly, ask
yourself the following questions.
Is the unit turned on?
Have there been any changes in the use, location, or environment of the
equipment that
Has the equipment hardware or software been modified since last use?
Is operator error the cause of the problem? Try to repeat the scenario exactly
and com
ma
nual.
Is the battery installed?
When connected to the AC wall outlet, does the green AC power light glow?
Is the writer door closed?
could cause the failure?
pare that to the proper operation of the equipment described in th
ly runs an internal self-test. If all circuits test
4-8MAC™ 3500 Resting ECG Analysis SystemRevision L
2021337-036
Page 49
Poor Quality ECGs
Troubleshooting: General Fault Isolation
Poor quality ECGs can be caused by factors in the environment, inadequate patient
preparation, hardware failures related to the acquisition module, leadwires, cables,
or problems in the unit.
Use a simulator to obtain an ECG report.
external to the unit.
A thorough visual inspection of the equipment can save time. Small things—
disconnected cables, foreign debris on circuit boards, missing hardware, loose
components—can frequently cause symptoms and equipment failures that may
appear to be unrelated and difficult to track.
NOTE
Take the time to make all the recommended vi
detailed troubleshooting procedures.
Table 1. Visual Inspection List
AreaLook for the following problems
I/O Connectors and CablesFraying or other damage
Bent prongs or pins
Cracked housing
Loose screws in plugs
FusesType and rating. Replace as necessary.
sual checks before starting any
Interface CablesExcessive tension or wear
Loose connection
Strain reliefs out of place
Circuit BoardsMoisture, dust, or debris (top and bottom)
Loose or missing components
Burn damage or smell of over-heated components
Socketed components not firmly seated
PCB not seated properly in edge connectors
Solder problems: cracks, splashes on board, incomplete feedthrough, prior modifications or repairs
Ground Wires/WiringLoose wires or ground strap connections
Faulty wiring
Wires pinched or in vulnerable position
Mounting HardwareLoose or missing screws or other hardware, especially fasten
planes on PCBs
Power SourceFaulty wiring, especially AC outlet
Circuit not dedicated to system
(Power source problems can cause static discharge, resetting problems, and
ers used as connections to ground
noise.)
4-10MAC™ 3500 Resting ECG Analysis SystemRevision L
2021337-036
Page 51
Troubleshooting: Diagnostic Tests
13B
Diagnostic Tests
Verify that the MAC 3500 resting ECG analysis system operates properly by
running the diagnostic tests. These tests check the operation of the display screen,
speaker, keyboard, thermal writer, battery, and communication. Detailed
information displays on screen.
Loading the System Diagnostics Menu
1.Select Main Menu on the Resting screen.
2.Select SystemSetu
3.At the prompt type the word sy
th
e Enter key. If the password was not changed, the System Setup me
appears. If the men
NOTE
If the system’s unique password is ina
instructions in “Substitute Master Password” on page 4-26.
4.When the System SetuF5).
5.Type pr
6.The System Dia
od at the service password prompt.
p.
stem, the password set at the factory, then press
u does not appear, use the master
ccessible, create one following the
p menu displays, hold down Shift and press F5 (Shift +
gnostics menu appears.
password.
nu
NOTE
The Fl
oppy Drive Tests option does not apply to the MAC 3500 system.
For information on Display Test, go to “Display Test” on page 4-12.
Battery tests check the current battery status, battery discharge rate and battery
charge rate. Test results are stored in memory and can be printed out. The graphic
displayed shows the Battery Test menu. Each test is covered in detail below.
Battery Status. This test displays and constantly updates current information on
the battery voltage, battery current, percent of charge remaining and
battery
temperature.
Battery Voltage. With a reading of 80% or more for the percent of charge
rem
aining, the battery voltage should be between 15 and 24 vo
voltage is below 15 volts,
Battery Current. Disconnect AC power. If the battery current is less than -
0.7 amps, the main CPU may need to be replaced. For
-0.8 amps, consider
the battery may need to be replaced.
example if current is
replacing CPU. See “Main CPU Board Replac
lts. If battery
ement”
on page 5-22.
Battery Temperature. A temperature reading over 45° C indicates a failure.
If battery tem
consider replacing
Ambient Temperature. Indicates the temperature inside the unit. The
tempera
perature is more than
battery.
ture displayed is accurate to within
10° C over the ambient temperature
+5°C in the range of 0°C to
50°C. Ambient temperatures not within this range cannot be displayed.
4-16MAC™ 3500 Resting ECG Analysis SystemRevision L
2021337-036
Page 57
Troubleshooting: Diagnostic Tests
21A
Battery Discharge Test. This test monitors a full discharge cycle. This test will
take
several hours to run.
To run test:
Select Battery Dischar
to fully charge. Remove AC power when prompted. Select OK
Batter
y Discharge Test. Allow battery to fully discharge. Unit will
automatically shut
To view test results:
Connect MAC 3500 to AC power; return to the Battery T
ge Test; plug unit into AC power and allow battery
and reselect
off.
est menu and
select Print Battery Discharge Test. Test results remain in memory until th
Batter
y Discharge Test is run again.
e
Battery Discharge Test wind
ow
NOTE
Consider replacing the battery if discharge capacity is less than 2000 (2k)
mAH.
Battery Charge Test. This test completely discharges the battery and monitors a
char
ge cycle. This test can take up to six hours to run. Select Print Batter
Ch
arge Test to view test results.
y
NOTE
The Battery Dischar
ge Test takes less time to run and is a better indicator of
battery condition.
To run test:
Select Ba
d
ischarge; when prompted, plug unit into AC power and select OK
batt
To view test results:
Return to the Battery T
results remain in memory until the Battery Charge Test is
ttery Charge Test; unplug unit and allow battery to completely
. Allow
ery to fully charge until test is complete.
est menu and select Print Battery Charge Tes t. Test
Internal Modem Test. Select the Internal Modem Test. The test returns the
m
odem ID number, firmware rev, and current parameter setti
communication with the modem
splay N/A.
di
is unsuccessful, the ID
ngs. If
and firmware rev
NOTE
If the test fails, consider rep
Ethernet Module Test. Select the Ethernet Module Test. The test returns IP
lacing the communication board.
address, subnet mask, and MAC Address settings for the Ethernet modul
commu
ma
nication with the Ethernet module is unsuccessful, the IP addr
sk, and MAC address will display N/A.
ess, subnet
e. If
NOTE
If the test fails, consider replacing the communication board. See “COMM
Board Replacement” on
page 5-30 for more information.
If the LAN connection on the communication board returns information
bu
t network communication problems still exist, use the LED status
dicators on the LAN connection to see if the connection to the network is
in
active. You can also use a ping command from the MUSE server to see if
the MUSE system can find the unit on the network.
4-20MAC™ 3500 Resting ECG Analysis SystemRevision L
2021337-036
Page 61
Acq. Module Tests
Tests if the acquisition
board is communicating.
Displays the acquisition board noise floor
Displays the software
version of the acquisition
board
27A
Troubleshooting: Diagnostic Tests
To check the internal acquisition board, connect all leads to the RL lead on the
patient cable and keep them separated and away from any external power source.
NOTE
One way of doing this is to connect all leads to a patient simulator and leave
simulator turned off.
the
NOTE
The Bu
tton Pressed test does not apply to the MAC 3500 system.
To run test, connect ECGOut1 (pin 8) to Analog Input (pin 7) and QRSTrigger
(pin 3) to Analog Input (pin 9).
NOTE
If test fails you may need to replace
Board Replacement” on
page 5-22.)
the CPU board. (See “Main CPU
Floppy Drive Tests
This option does not apply to the MAC 3500 system.
4-24MAC™ 3500 Resting ECG Analysis SystemRevision L
2021337-036
Page 65
Internal Memory Tests
35A
36A
Troubleshooting: Diagnostic Tests
The Internal Memory Test checks for bad blocks and amount of free memory. After
the test is completed you are given the option of reformatting the internal memory.
CAUTION
Reformatting will erase all data in memory, including patient data.
Reformatting will not affect the system software or the System Setups.
SD Card Tests
Follow on-screen prompts to perform the Internal Memory Tests.
The SD Card Tests option performs a read/write test on the SD card currently
installed. Responds with a PASS or FAIL.
Follow on-screen prompts to perform the SD Card Tests.
If card is not formatted correctly, an error mes
sage will appear at the bottom of the
screen.
NOTE
To format a card, insert the SD card and copy data to the card using the Co
command from File Manager. The MAC 3500 software will then prompt you to
rmat the card. Follow on-screen prompts. Refer to the MAC 3500 Operator’s
Dry the surfaces with a clean cloth or paper towel.
Q: What is the capacity of the battery?
System Setup
Location Number
A: We recommend that the MAC 3
500 be plugged into a wall outlet whenever it is
not in use. However, the life of the battery is approximately 100 ECGs and onepage reports or six hours of continuous operation (without printing).
Q: When entering in the patient data, how do I get the Location field to
automatically populate with the same number?
A: The Location nu
mber can be set in System Setup to save you from entering it for
each test.
Go to System Setup.
Select Basic System.
Select Miscellaneous Setup.
Arrow down to Location and type in the number you want set as your
default
. Then press Enter.
Press the esc key until you return to System Setup.
Select Save Setup.
Select To S y s t e m .
You can select Main Menu to exit System Setup.
Patient Questions
Q: How do I change what questions I see when I am entering the patient data?
A: The patient questions you see on the Pati
were set up in System Setup.
Go to System Setup.
Select Basic System.
Select Patient Questions.
Select the patient questions you want to include when entering the patient
data for a test.
Press the esc key until you return to System Setup.
Select Save Setup.
Select To S y s t e m .
You can select Main Menu to exit System Setup.
4-30MAC™ 3500 Resting ECG Analysis SystemRevision L
2021337-036
ent Data window when starting a test
Page 71
Passwords
Troubleshooting: Frequently Asked Questions
Q: Can you set up a password for the Delete function that is different than the
System Setup password?
Clinical
Report Format
A: No. The password for the S
ystem Setup and the Delete function are the same.
Q: How do I change the way an ECG looks when it prints out?
A: Do the following steps.
Go to System Setup.
Select ECG.
Select which type of ECG report you want to change:
Select unconfirmed reports from the menu.
Find the report type you want the MAC 3500 to print.
Place the number of copies you want in the appropriate column.
If you want the MAC 3500 or 12SL interpretation included on the ECG, put the
num
ber of copies you want in the “with” column.
If you do not want the MAC 3500 interpretation to print on the ECG, put the
num
ber of copies you want in the “without” column.
Click view report type, to see the examples of the report formats.
Press the esc key until you return to System Setup.
Select Save Setup.
Select To Sy s te m .
You can select Main Menu to exit System Setup.
Editing
Q: Can you edit the interpretation at the MAC 3500, and then transmit the edited
record to the MUSE system as an unconfirmed record?
A: If you edit demographic information only the record is still transmitted to the
MUSE
system as an unconfirmed record. However, if you edit the
interpretation, the data will not be saved unless the record is confirmed at the
MAC 3500. The record is transmitted to the MUSE system as a confirmed
record as well.
Q: Do I have to enter all the Information I see on the Patient Data screen?
A: In S
patient identification number, or medical record number) be entered. It is not a
requirement to enter any other data. However, we recommend that you enter the
patient name and identification number, at the least. If you are transmitting to
the MUSE system you will want to enter the Location number as well. If an
emergency situation dictates that you must complete the test. without entering
the patient data, make sure you edit the record to add the missing information
before you transmit it to the MUSE system.
Transmission
Losing Fields When Transmitting
Q: Why do I lose the Referring MD and Technician names off of my reports when
I transmit records to the MUSE system?
A: Your MAC 3500 may be transmitting to the SDLC modem on the MUSE
system instead of the CSI
transmitting to the MUSE system CSI phone number.
ystem SetupBasic System Patient Questions you can require that the
modem. Check in System Setup to make sure you are
4-32MAC™ 3500 Resting ECG Analysis SystemRevision L
2021337-036
Page 73
Troubleshooting: Input and Output Connectors
2
1
6
5
3
4
66A
6
3
1
8
5
4
2
7
67A
Input and Output Connectors
The following pages detail the input/output signals for those connectors. The pin-bypin descriptions identify the signal names and pin outs for each connector on the
unit.
5-2MAC™ 3500 Resting ECG Analysis SystemRevision L
2021337-036
Page 85
Maintenance: Introduction
Introduction
Recommended Maintenance
Regular maintenance, irrespective of usage, is essential to ensure that the equipment
will always be functional when required.
WARNING
Failure on the part of all responsible individuals, hospitals or
institutions, employing the use of this device, to implement the
recommended maintenance schedule may cause equipment failure
and possible health hazards. The manufacturer does not in any
manner, assume the responsibility for performing the
recommended maintenance schedule, unless an Equipment
Maintenance Agreement exists. The sole responsibility rests with
the individuals, hospitals, or institutions utilizing the device.
Required Tools and Supplies
In addition to a standard set of hand tools, you will need the items listed in the table
below.
Perform a visual inspection of all equipment and peripheral devices daily. Turn off
the unit and remove power before making an inspection or cleaning the unit.
Check the case and display screen for cracks or other damage.
Regularly inspect all cords and cables for fraying or other damage.
Verify that all cords and connectors are securely seated.
Inspect keys and controls for proper operation and check that toggle keys do not
stick in one
Chap
ter 5 of this manual.
Exterior Cleaning
Clean the exterior surfaces monthly, or more frequently if needed.
position. If necessary, perform the
keyboard test described in the
Interior Cleaning
General
Thermal Printhead Cleaning
1.Use a clean, soft cloth and a
2.Wring the excess water from the cloth. Do not drip water or any liquid on the
equi
pment, and avoid contact with open vents, plugs, or connectors.
3.Dry the surfaces with a clean cloth or paper towel.
Check for dust buildup on the surfaces of the interior circuit boards, components,
and power supply. Use commercially available compressed air to blow away the
accumulated dust. Follow the manufacturers directions.
Clean the thermal printhead every three months or more often with heavy use. A
build-up of thermal paper coating on the printhead can cause light or uneven
printing.
Use a solution containing alcohol on a nonwoven, nonabras
Cloth to wipe off the printhead. Do not use paper toweling, as it can scratch the
printhead.
mild dish washing detergent diluted in water.
ive cloth such as Techni-
5-4MAC™ 3500 Resting ECG Analysis SystemRevision L
2021337-036
Page 87
Maintenance: Checking Electrical Safety
Thermal
Printhead
1A
Checking Electrical Safety
The device should be checked annually for current leakage and ground continuity.
For details, see “Electrical Safety Checks” on pag
Prior to performing any disassembly procedures, perform the following:
If possible, process any ECGs remaining in storage.
If possible, print out set-up for future reference.
Disconnect the unit from the AC wall outlet and remove the power cord from
th
e unit.
Remove the battery as described in “Battery Replacement”.
If the MAC 3500 system is mounted to the trolley, remove it from the trolley as
descri
bed in “Remove MAC 3500 System From Trolley”.
Remove the writer paper.
Take strict precautions against electrostatic discharge damage.
Battery Replacement
1.Press the writer button to open the unit.
2.Slide the battery release button in the dire
out.
ction of the arrow and lift the battery
5-6MAC™ 3500 Resting ECG Analysis SystemRevision L
2021337-036
Page 89
Maintenance: FRU Replacement Procedures
2A
5A
NOTE
Once the maintenance procedures are complete,
replace the battery.
Remove MAC 3500 System From Trolley
reverse these steps to
MAC Series Trolley
If your MAC system is mounted on a trolley, you must remove it from the trolley
before servicing it. The method for removing the device depends on the type of
trolley.
1.Lock the wheels, remove the rear trolley panel then loosen the three captive
screws locat
1.Remove the three fasteners from under the top cover at the front of the unit.
2.Remove the two fasteners that secu
uni
t and then remove the display cover.
re the display panel cover at the back of the
3.Remove the two fasteners holding the top of the keypad.
4.Pull up on the keypad assembly to release it from the top cover. You will
snapping noise
5-12MAC™ 3500 Resting ECG Analysis SystemRevision L
as each of the eight plastic standoffs is released.
2021337-036
hear a
Page 95
Maintenance: Keypad Replacement
NOTE
The eight plastic standoffs should remai
n with the keyboard. However, if
any do remain in the top cover, use a small pliers to extract from the top
cover. Do not damage the rubber cover which is permanently affixed to the
top cover.
5.Using a small pliers, remove the eight plastic standoffs that remained in the top
cover.
6.Place new keyboard in place on top cover. B
standoffs with
the appropriate holes in the top cover.
e sure you align the eight plastic
7.Once the keyboard has been properly aligned with the top cover, push down on
keyboard at each of the eight s
to
p cover.
tandoff locations until it snaps into place on
Reverse the procedures for removal of the keyb
whi
3.Remove the screws fastening the LVDS converter board to the display
ass
NOTE
.
embly
Dispose of the old LVDS converter board
local and federal laws.
in compliance with all applicable
4.Attach the new LVDS converter board to the display assembly using the tw
screws rem
5.Reconnect all the cables that were dis
6.Reattach the display assembly to the device.
See “Keyboard/Top Cover Assembly Reassembly” on page 5-17 for
instructions.
oved in step 3.
Keyboard/Top Cover Assembly Reassembly
Reverse the removal procedures to reassemble the keyboard/top cover assembly.
NOTE
When reconnecting the keyboard ribbon cable to its
PCB, lift both ends of the plastic locking bar as you insert the ribbon cable.
Once the ribbon cable has been inserted, push down on the both sides of the
locking bar to secure the cable.
o
connected in step 2.
connector on the main
Also, be sure to pull the two ground cables through so they can be reconnected
the display panel. If the unit has a KISS pump, you will also need to ensure
to
that the KISS pump suction tube and the KISS pump power supply cable are
pulled through as well.