GE MAC 400 User manual

5 (1)

MAC™ 400 Resting ECG Analysis System Safety and Warnings Guide

© 2007, 2008 General Electric Company. All Rights Reserved

The MAC™ 400 Resting ECG Analysis System Safety and Warnings Guide provides guide details, safety information and symbols definitions.

Guides information

The MAC 400 Resting ECG Analysis System Guides contain the instructions necessary for operating the MAC 400 system in accordance with its function and intended use.

The information in these guides only applies to MAC 400 system software version 1. It does not apply to earlier software versions. Due to continuing product innovation, specifications in these guides are subject to change without notice.

MAC™ and Mactrode™ are trademarks owned by GE Medical Systems Information Technologies, a General Electric Company going to market as GE Healthcare. All other marks are owned by their respective owners.

INTENDED AUDIENCE

The MAC 400 Resting ECG Analysis System Guides are intended for persons who use, maintain, or troubleshoot this equipment.

REVISION HISTORY

The document part number and revision letter are on the bottom of each page. The revision letter identifies the document’s update level.

Revision History 2032589-001

Revision

Date

Comment

A

15

May 2007

Initial release of document

B

15

June 2007

Revised per MVP feedback

C

16

July 2007

Revised IPD

D

11 September 2007

Revised heading.

E

17

March 2008

Updated “Equipment Symbols”.

PRODUCT REFERENCE

The product described in these guides is the MAC 400 resting ECG analysis system. It may be referred to as “the system” throughout these documents.

CONVENTIONS

These conventions are used in the MAC 400 system guides:

Bold text

Indicates keys on the keyboard, text to be entered, or

 

hardware items such as keys or switches on the

 

equipment.

 

 

Italicized text

Indicates terms that identify menu items or options in the

 

system display window.

 

 

CE marking information

The MAC 400 system bears the CE mark “CE-0459”, notified body GMED, indicating its conformity with the provisions of the Council Directive 93/42/EEC, concerning medical devices and fulfills the essential requirements of Annex I of this directive. The system delivers 3-channel ECG recordings in automatic and arrhythmia modes and 1 or 3-channel ECG recordings in manual mode. The country of manufacture can be found on the equipment labeling.

The safety and effectiveness of this device have been verified against previously distributed devices. All standards applicable to presently marketed devices may not be appropriate for prior devices (for example, electromagnetic compatibility standards). This device will not impair the safe and effective use of previously distributed devices.

NOTE: Electromagnetic compatibility information can be found in the “MAC™ 400 Resting ECG Analysis System Service Manual”.

Product use and classification

RECOMMENDATIONS

Operating the system near radio frequency (RF) electromagnetic interference (EMI) above the conditions defined in the electromagnetic compatibility (EMC) Standard EN60601-1-2 for Radiated Immunity (field strengths above 3 volts per meter) may cause waveform distortions. Portable and mobile RF communications equipment can affect medical electrical equipment. Users should consider RF sources, such as radio or TV stations and hand-held or mobile two-way radios, when installing a medical device or system. Adding accessories or components, or modifying the medical device or system may degrade the EMI performance. Consult with qualified personnel regarding changes to the system configuration.

Medical electrical equipment requires precautions regarding EMC and needs to be installed and used according to the EMC information provided in the product service manual. The use of accessories, transducers and cables other than those specified or sold by the manufacturer may result in increased emissions or decreased immunity of the system. The system should not be used nearby or stacked with other equipment. If stacking or using near other equipment cannot be avoided, observe to verify normal operation. Review the AAMI Committee Technical Information Report (TIR) 18, “Guidance on Electromagnetic Compatibility of Medical Devices for Clinical/Biomedical Engineers” for details on evaluating and managing an EMI environment in the hospital. Take the following actions to reduce the risk of medical device EMI and achieve EMC:

Assess the EMC environment of the facility and identify radio transmitters and/or areas where critical medical devices are used such as the emergency room and intensive care units.

Increase the distance between sources of EMI and susceptible devices.

Remove the devices that are highly susceptible to EMI.

Lower the power transmitted from electrical and electronic equipment (EMI sources) under hospital control (i.e. paging systems).

Label devices susceptible to EMI.

Educate facility staff (nurses and doctors) to identify, and recognize, potential EMI-related problems.

CLASSIFICATION

The device is classified, according to IEC 60601-1, as:

Type of protection against

Class I or internally powered equipment.

electrical shock

 

 

 

Degree of protection against

Type CF defibrillation-proof applied part.

electrical shock

 

 

 

Degree of protection against

Ordinary equipment (enclosed equipment

harmful ingress of water

without protection against ingress of water,

 

IPX0).

 

 

Degree of safety of

Equipment not suitable for use in the presence

application in the presence

of a flammable anesthetic mixture with air or

of a flammable anesthetic

with oxygen or nitrous oxide.

mixture with air, oxygen or

 

nitrous oxide

 

 

 

Methods of sterilization or

Not applicable.

disinfection recommended

 

by the manufacturer

 

 

 

Mode of operation

Continuous operation.

 

 

RESPONSIBILITY OF THE MANUFACTURER

GE is responsible for the effects of safety, reliability, and performance only if:

Assembly operations, extensions, readjustments, modifications, or repairs are carried out by persons authorized by GE.

The electrical installation of the relevant room complies with the requirements of the appropriate regulations.

The equipment is used in accordance with the instructions for use.

INTENDED USE

The MAC 400 device is for use under the direct supervision of a licensed healthcare practitioner. The system is intended to acquire, measure and record information from adult and pediatric populations. The basic system delivers 3-channel ECG recordings in automatic and arrhythmia modes and 1 or 3-channel ECG recordings in manual mode. The arrhythmia detection provides the convenience of automatic documentation. It is not designed to provide alarms for arrhythmia detection. This system is not intended for use as a vital signs physiological monitor, or for use during patient transport.

This device is not intended for use with high frequency surgical units. Disconnect the patient from the device before using the high frequency surgical units. This device is not intended for use with direct cardiac applications.

BIOCOMPATIBILITY

The parts of the product described in these guides, including all accessories, that come in contact with the patient during the intended use, fulfill the biocompatibility requirements of the applicable standards. Please contact GE or its representatives with any questions.

Safety information

DEFINITIONS

The terms danger, warning, and caution are used in these guides to point out hazards and to designate a degree or level of seriousness. Familiarize yourself with their definitions and significance. A hazard is defined as a source of potential injury to a person.

DANGER indicates an imminent hazard which, if not avoided, will result in death or serious injury.

WARNING indicates a potential hazard or unsafe practice which, if not avoided, could result in death or serious injury.

CAUTION indicates a potential hazard or unsafe practice which, if not avoided, could result in personal injury or product/property damage. NOTE provides application tips or other useful information to ensure that you get the most from your equipment.

The safety information given in this manual is classified as follows.

Warning

Description

 

 

Accidental spills

To avoid electric shock or device malfunction, liquids

 

must not enter the device.

 

If liquids enter the device, stop using it and have it

 

checked by a service technician before further use.

 

 

Battery

If the integrity of the protective earth conductor is in

operation

doubt, operate the unit from its battery.

 

 

Cables

To avoid possible strangulation, route all cables away

 

from patient's throat.

 

 

Connection to

This is Class I equipment. The mains plug must be

mains

connected to an appropriate power supply.

 

 

Warning

Description

 

 

Defibrillator

Do not come into contact with patients during

precautions

defibrillation. Serious injury or death could result. Patient

 

signal inputs labeled with the CF and BF symbols with

 

paddles are protected against damage resulting from

 

defibrillation voltages. To ensure proper defibrillator

 

protection, use only the recommended cables and lead

 

wires. Proper placement of defibrillator paddles in

 

relation to the electrodes is required to ensure successful

 

defibrillation.

 

 

Electrodes

Polarizing electrodes (stainless steel or silver

 

constructed) may cause the electrodes to retain a residual

 

charge after defibrillation. A residual charge will block

 

acquisition of the ECG signal. Whenever patient

 

defibrillation is a possibility, use nonpolarizing (silver/

 

silver chloride construction) electrodes for ECG

 

monitoring.

 

 

Magnetic and

Magnetic and electrical fields can interfere with proper

electrical

performance of the device. Make sure that all external

interference

devices operated near the device comply with the relevant

 

EMC requirements.

 

X-ray equipment or MRI devices may interfere with

 

system performance because they may emit higher levels

 

of electromagnetic radiation.

 

 

Explosion

Do NOT use in the presence of flammable anesthetics

hazard

vapors or liquids.

 

 

Interpretation

Computerized interpretation is only significant when

hazard

used in conjunction with clinical findings. A qualified

 

physician must over read all computer generated tracings.

 

 

Operator

Medical technical equipment such as this

 

electrocardiograph system must only be used by persons

 

who have received adequate training in the use of such

 

equipment and are capable of applying it properly.

 

 

Shock hazard

Improper use of this device presents a shock hazard.

 

Failure to observe the following warnings may endanger

 

the lives of the patient, the user, and bystanders.

 

Disconnect from the power source before disconnecting

 

the cable from the device to reduce the risk of

 

inadvertently introducing metal parts in the sockets of the

 

power cord and coming in contact with line voltage.

 

 

Site

Do not route cables in a way that they may present a

requirements

stumbling hazard. For safety reasons, connectors for

 

patient cables and lead wires are designed to prevent

 

disconnection if pulled on. For devices installed above

 

the patient, adequate precautions must be taken to prevent

 

them from dropping on the patient.

 

 

MAC™ 400 resting ECG analysis system guides

1 of 7

2032589-001

Revision E

GE MAC 400 User manual

MAC™ 400 Resting ECG Analysis System Safety and Warnings Guide

© 2007, 2008 General Electric Company. All Rights Reserved

Caution

Description

 

 

Accessories

To ensure patient safety, use only parts and accessories

(supplies)

manufactured or recommended by GE. Parts and

 

accessories must meet the requirements of the

 

applicable IEC 60601 series safety standards and

 

essential performance standards.

 

 

Proper lead wire

Improper connection will cause inaccuracies in the

connection

ECG. Trace each individual lead wire from its

 

acquisition module label to the colored connector and

 

then to the proper electrode to ensure that it is matched

 

to the correct label location.

 

 

Before installation

Compatibility is critical to safe and effective use of

 

this device. Please contact your local sales or service

 

representative prior to installation to verify equipment

 

compatibility.

 

 

Disposables

Disposable devices are intended for single use only.

 

Do NOT reuse, as performance may degrade or

 

contamination could occur.

 

 

Equipment damage

Devices intended for emergency application must not

 

be exposed to low temperatures during storage and

 

transport to avoid moisture condensation at the

 

application site. Wait until all moisture has vaporized

 

before using the device.

 

 

Electric shock

To reduce the risk of electric shock, do NOT remove

 

cover (or back). Refer servicing to qualified personnel.

 

 

Power

Before connecting the device to the power line, check

requirements

that the voltage and frequency ratings of the power

 

line are the same as those indicated on the unit's label.

 

If not, do not connect the system to the power line.

 

This equipment is suitable for connection to public

 

mains as defined in CISPR 11.

 

 

Low battery shut

If the battery is not charged for a long enough period

down

of time or after multiple attempts to power on

 

following a low-battery shut down, the system shifts

 

to a second level of deep discharge protection. If

 

device is turned on, it might call for a 1 to 2 minute

 

continuous self test. Charge the battery for a minimum

 

of half an hour before using device. We recommend

 

that you keep the unit charged to avoid a low-battery

 

shut down.

 

 

Serviceable parts

This equipment contains no user serviceable parts.

 

Refer servicing to qualified service personnel.

 

 

Supervised use

This equipment is intended for use under the direct

 

supervision of a licensed healthcare practitioner.

 

 

General information

RECORDING ECGS DURING DEFIBRILLATION

This equipment is protected against the effects of cardiac defibrillator discharge to ensure recovery, as required by test standards. The patient signal input of the acquisition module is defibrillation-proof. It is not necessary to remove the ECG electrodes prior to defibrillation. When using stainless steel or silver electrodes, a defibrillator discharge current may cause the electrodes to retain a residual charge causing a polarization or DC offset voltage. Electrode polarization blocks acquisition of the ECG signal. If a defibrillation procedure is necessary, use non-polarizing electrodes, such as silver/silver-chloride types, to avoid a DC offset voltage when subjected to a DC current. If using polarizing electrodes, disconnect the lead wires from the patient before delivering the shock. Electrode defibrillation recovery allows the ECG trace to return after defibrillation. We recommend using nonpolarizing disposable electrodes with defibrillation recovery ratings as specified in AAMI EC12 3.2.2.4. (MMS P/N 9623-103P Silver Mactrodes™, MMS spec. TP9623-003). AAMI EC12 requires that the polarization potential of an electrode pair does not exceed 100 mV, five seconds after a defibrillation discharge.

RECORDING ECGS OF PACEMAKER PATIENTS

The system does not support pacer pulse detection.

WARNING: PATIENT HAZARD If several adverse conditions exist at once, pacer pulses might be interpreted and counted as QRS complexes. Pacemaker patients should always be watched closely.

ACCURACY OF THE INPUT SIGNAL REPRODUCTION

Overall system error is tested using the method described in AAMI EC11 3.2.7.1. Overall system error is ±5 %.

Frequency response is tested using the method described in AAMI EC11 3.2.7.2 methods A and D.

MODULATING EFFECTS IN DIGITAL SYSTEMS

This device uses digital sampling techniques that may produce some variation in amplitudes of Q, R, and/or S waves from one heart beat to the next. These variations may occur more in pediatric recordings. If this variation is observed, be aware that the origin of amplitude variations is not entirely physiologic. For measuring voltages of Q, R, and S waves, use the QRS complexes with the largest deflection of the particular waves.

PARTS AND ACCESSORIES

To ensure patient safety, use only parts and accessories manufactured or recommended by GE. Parts and accessories must meet the requirements of the applicable IEC 60601 series safety standards and essential performance standards.

EQUIPMENT SYMBOLS

Symbol

Description/Function

Type CF equipment. The acquisition module is protected from defibrillation shocks.

The flashing yellow LED indicates you must connect to AC power to re-charge the battery.

Consult instructions for use.

Consult accompanying documents for cautions.

Classified with respect to electric shock, fire, mechanical, and other specified hazards only in accordance with UL 60601-1, CAN/CSA C22.2

No. 601.1, EN 60601-2-25, EN 60601-1, IEC 60601-1-2: 2001, IEC 60601-2-51.

Indicates that the device is classified as Ordinary Equipment (enclosed equipment without protection against ingress of water).

Indicates that the waste of electrical and electronic equipment must not be disposed as unsorted municipal waste and must be collected separately. Please contact an authorized representative of the manufacturer for information concerning the decommissioning of your equipment.

Fuse.

Do not throw or dispose of in fire!

Contains "Lithium Ion". This symbol indicates "General recovery/recyclable" and must not be disposed of as unsorted municipal waste and must be collected separately.

Manufacturer name and address.

Date of Manufacture (Year-Month).

Batch code of paper or battery.

Catalogue number (Part number).

Symbol

Description/Function

Serial number.

European authorized representative.

The packaging of this product can be recycled.

Humidity limitation.

Atmospheric pressure limitation.

Environment-friendly Use Period per Chinese standard SJ/

T11363-2006 (China specific).

PCT. GOST marking symbolizing conformity with applicable Russian Gosstandart technical and safety standards.

SERVICE REQUIREMENTS

Refer equipment servicing to GE authorized service personnel only. Any unauthorized attempt to repair equipment under warranty voids that warranty. It is the user’s responsibility to report the need for service to GE or an authorized agent.

EQUIPMENT IDENTIFICATION

Every GE device has a unique serial number for identification on the device label.

A

Product code is SCT

B

Year manufactured (00-99)

00 = 2000, 01 = 2001, and so on

 

 

 

C

Fiscal week manufactured

D

Production sequence number

E

Manufacturing site

F

Miscellaneous characteristic

g

GE Medical Systems

 

Asian Headquarters

Information

 

GE Medical Systems

Technologies, Inc.

 

Information Technologies Asia;

8200 West Tower

GE Medical Systems

GE (China) Co., Ltd.

Avenue

Information Technologies

11th Floor Shanghai MAXDO

Milwaukee, WI 53223 USA

GmbH

Centre

Tel: + 1414 355 5000

Munzinger Strabe 3-5

8 Xing Yi Road, Hong Qiao

1 800 558 5120 (US only)

D-79111 Freiburg

Development Zone

Fax: + 1414 355 3790

Germany

Shanghai 200336, People’s

 

Tel: + 49 761 4543 - 0

Republic of China

www.gehealthcare.com

Fax: + 49 761 4543 - 233

Tel: + 86-21-5257-4650

MAC™ 400 resting ECG analysis system guides

2 of 7

2032589-001

Revision E

MAC™ 400 Resting ECG Analysis System Installation and Setup Guide

© 2007, 2008 General Electric Company. All rights reserved.

The MAC™ 400 Resting ECG Analysis System Installation and Setup Guide provides instructions for initial installation and setup of your system.

Before you begin

Remove the device and accessories from the box, and keep on a flat dry surface, away from direct sunlight, heat sources & dust.

A.Confirming the box contents

MAC 400 resting ECG analysis device

Four limb clamp electrodes

Six bulb electrodes

Patient ECG cables/lead wires

Power cable

Pack of z-fold paper (280 sheets)

Electrode cream

CD containing a service manual and 12SL™ Physicians Guide

User guide

B.Charging the battery

The MAC 400 system needs a charged battery to print an ECG. Charge the battery for three hours for full power. A fully charged battery allows for 100 automatic mode ECGs or 100 minutes of arrhythmia/ manual mode recording.

1.Connect the device to a power source.

2.Turn on the power (a green light indicates that AC mains is connected).

3.After charging for three hours, unplug the power cord from both the device and the power source.

Recharge when you see a yellow light above the power key.

C. Loading paper

The MAC 400 system supports the use of standard thermal recording paper in either z-fold pads or roll paper.

1.Locate the printer assembly.

2.Gently pull back the printer assembly door to open the door latch.

3.Open the z-fold paper pack.

4.Ensure that the black paper guide is on the top and place the paper in the tray.

5.Advance the first sheet and close the door. Make sure that the paper is positioned on the pressure roller and that the door locks into place on both sides.

A red line at the top of the last ten sheets indicates that the paper supply is low. Change paper as needed.

NOTE: For instructions on loading roll paper, see the “MAC™ 400 Resting ECG Analysis System Maintenance Guide”.

Initial system setup

THE SETUP MENU

Press on/off to turn on the device, and then, press

configuration to access the setup menu. Use the setup menu to navigate between system settings and select the options you want. For more information, see System Symbols and System Menu Descriptions on page 4.

THE SYSTEM PARAMETERS TOOLS

Use these tools and guidelines to help you define the operating parameters.

1.Use up/down cursor to select a menu item.

2.Use right/left cursor to select the desired setting.

3.Always press enter to confirm your selection.

4.Always press configuration to save and exit the

setup menu.

SETTING OPTIONS FOR INITIAL USE

A. Selecting a language

Different languages are available for the display text and printed ECG reports.

1.Press configuration to display the language selection menu.

2.Use the right/left cursor to select the language.

3.Press enter to confirm your selection.

B. Selecting the lead notations

There are two different lead notation options: AHA and IEC.

1.Use the up/down cursor to select Notation.

2.Use the right/left cursor to select AHA or IEC.

3.Press enter to confirm your selection.

C. Setting the date and time

Set the date and time for printed ECG reports. To set the date:

1. Use the up/down cursor to select Date.

2.Use the right/left cursor to set the date.

3.Pressenter to confirm your selection.

To set the time:

1.Use the up/down cursor to select Time.

2.Use the right/left cursor to set the time.

3.Press enter to confirm your selection.

D. Selecting heart rate limit values

1.Use the up/down cursor to select HR control. The cursor flashes on the low limit value.

2.Use the right/left cursor to change the low limit value, in increments of 5 BPM, between 30 and 120 BPM.

3.Press enter to confirm your selection.

4.Use right/left cursor, change the high-limit value (between 80 and 240 BPM).

5.Press enter to confirm.

NOTE: After setting these options, you must press configuration to save the options and exit the setup menu.

Heart rate indication

THE BASICS

WARNING PATIENT HAZARD — The MAC 400 system is not intended for use as a vital signs physiological monitor. When needed, use a device intended for vital signs monitoring.

When the heart rate indication function is enabled, the automatic switchoff is disabled. Conditions of high and low heart rate are indicated in all operating modes, even when not recording. This function can be disabled in the setup menu. The default heart rate limits are 45 and 130 BPM, and can be modified in the setup menu. If the heart rate exceeds one of the set values, the system emits an audio signal. This audio signal ceases automatically when the heart rate returns to the permitted range or when

you press QRS beep. The audio signal will not recur if it was

silenced with QRS beep. The audio signal recurs only when the heart rate exceeds one of the limit values again.

E. Hooking up the starter kit

Once you have customized your settings, connect the patient cable to the device (A) on the right side panel.

Default settings for auto mode

When you turn on the device, the system default is automatic mode. Factory defaults have the following functions and settings (the most important settings are indicated on the display):

A B C D E

 

J

I H

G

F

 

 

A

Lead sequence: Standard = Einthoven (I, II, III),

Goldberger(aVR, aVL, aVF), Wilson 1 (V1, V2,

 

V3), Wilson 2 (V4, V5, V6).

 

 

 

B

Lead fail indicator: Indicates a disconnected

electrode. For example, LA indicates that the left

 

arm electrode is disconnected.

 

 

 

C

QRS indicator: The heart symbol blinks with every

detected systole.

 

 

 

 

 

 

 

 

 

D

Heart rate: Detects the patient’s heart rate. In this

example, it is 60 BPM.

 

 

 

 

 

 

 

 

E

AC filter: On (enabled) (50 Hz).

 

F

Muscle filter: Off (disabled).

 

G

Sensitivity (gain): 10 mm/mV.

 

H

Paper speed: 25 mm/s.

 

 

I

Rotating symbol: Displays when the ECG data

acquisition or recording is active.

 

 

 

 

 

 

J

Operating mode: Automatic.

 

Additional default settings (not indicated on the display):

Notation: AHA.

ADS: Enabled.

Report format: Simultaneous, short.

Override: Disabled.

NOTE: For further descriptions of the settings, see System Menu Descriptions on page 4.

MAC™ 400 resting ECG analysis system guides

3 of 7

2032589-001

Revision E

Loading...
+ 4 hidden pages