PortaFlow-X
Portable, Accurate Transit-Time Flow Measurement
The Portaow-X is a portable ultrasonic ow meter utilizing the transit-time difference
principle for measuring ow rates through pipes from the outside. It’s compact and
lightweight, and incorporates the latest breakthroughs in digital signal processing and
electronics realizing high performance and easy operation. The most attractive advantage
of the Portaow-X is the non-invasive setup, enabling multiple ow measurements of
different running systems, and offering a signicant cost advantage over time. Some
of the most common applications include: backup for an already installed ow meter
system, water supply and sewage treatment, ow measurement systems, boilerwater/
feedwater supplies, cooling water, cooling oil, and deionized water for semiconductor
manufacturing, among others. The meter may be set up with a wide variety of sensors
from small diameters to large diameters, over a wide temperature range.
Features
• Large LCD backlit display – both flow
rates and totalizing functions may be
viewed together
• Simple setup and easy operation –
typical setup takes approximately 10
minutes
• Sound velocity measuring function –
eliminates temperature and pressure
errors, and enables easy setup for
unknown materials
• Enhanced anti-bubble measurement
technology (ABM) – removes the effect of
entrapped and suspended flow bubbles
• Compact and light-weight – weighs only
3.3 lbs
• Built-in data logger – direct download to
PC spreadsheet applications
• High accuracy – ±0.5% to ±1.0% of
velocity typical on calibrated system
• Variety of detectors – for small and large
diameter pipes ranging from 0.50 to 235 in.
Performance Specications
for the Converter
Fluid Conditions
Measured fluid
(water, sea water, hydro-carbons or uid
of unknown sound velocity) capable of
ultrasonic wave propagation
Fluid turbidity 10000 deg. (mg/l) or less
State of flow Axis-symmetric ow in pipe
lled with uid
Velocity range 0.06 to 105 ft/sec (.018 to 32
m/sec) bi-directional ow
Fluid temperature
Standard temperature detector
212°F (-40 to 100°C)
High temperature detector -40 to 390°F
(-40 to 200°C)
Piping Conditions
Pipe material
Homogeneous liquids
-40 to
Carbon steel, stainless steel,
cast iron, copper, pvc, aluminum, ductile
iron, asbestos, frp, peek, pvdf, acrylic and
other. If other is selected, pipe materials
with a sonic velocity range of 3280 to
12136 ft/sec (1000 to 3700 m/sec) can
be selected via the keypad (sonic velocity
information for several pipe materials is
included in the Operator Manual)
Pipe size Small pipe range detector:
0.50 to 4.0 in. (13 to 100mm); Standard
pipe range detector: 2.0 to 16 in. (50 to
400mm); Universal pipe range detector:
8.0 to 48 in. (200 to 1200mm); Large
pipe range detector: 8.0 to 235 in. (200 to
6000mm); High-Temperature detector: 2.0
to 16 in. (50 to 400mm)
Lining material Tar, epoxy, mortar, rubber,
Teon, glass, pvc, other, or none. If other
is selected, liner materials with a sonic
velocity range of 3280 to 12136 ft/sec
(1000 to 3700 m/sec) can be selected via
the keypad (sonic velocity information for
several liner materials is included in the
Operator Manual)
Fluid type Water, sea water or other. If other
is selected, a sonic velocity range of 1640
to 8200 ft/sec (500 to 2500 m/sec) can
be selected via the keypad (sonic velocity
information for several uids is included in
the Operator Manual)
Measurement Accuracy
Accuracy
Linearity 0.1% of scale
Repeatability 0.5% or better
±0.5% to ±1.0% of velocity for
velocities >1.0 ft/sec. typical on calibrated
system; ±1.5% to ±2.0% of velocity for
velocities <1.0 ft/sec. typical on calibrated
system
(Calibrated system conditions
include a minimum of 10 inner pipe
diameters of upstream straight pipe run
and a minimum of 5 inner pipe diameters
of downstream pipe run. Longer runs may
be necessary due to pipe configurations.)
Ultrasonic Flow Meters Fuji PortaFlow-X
Key Features
• Large LCD backlit display
• Simple setup and easy operation
• Sound velocity measuring function
• Enhanced anti-bubble measurement
technology (ABM)
• Compact and light-weight
• Wide pipe range: 0.5 to 235 in.
Information subject to change without notice. Prices in USD.
082106
PortaFlow-X
Specications (continued)
Physical Specications
for the Converter
Ambient temperature Without printer: 14
to 131°F (-10 to 55°C) ; With printer: 14 to
113°F (-10 to 45°C)
Ambient Humidity less than 90% RH
Enclosure High impact thermoplastic, gray
in color, dust-proof (NEMA 3, IP52)
Dimensions Without printer: 9.45H x 5.00W
x 2.75D in. (240 x 127 x 70mm); With
printer: 14.13H x 5.00W x 2.75D in. (359 x
127 x 70mm)
Weight Without printer: 3.3 lbs. (1.5kg); With
printer: 4.4 lbs. (2.0kg)
Functional Specications
for the Converter
Power supply Built-in battery or external
power adapter
Built-in battery Special NiCd, Continuous
operation: 5 hrs. (without optional printer
and back light). Recharging time: 3 hrs.
(with supplied external power adapter)
Power adapter External power adapter,
input: 90 to 264V AC, 50/60 Hz
Output 17.5V DC, 1.2 A
Power consumption 12 W or less (without
optional printer)
LCD display 4.0H x 3.0W in. (102 x 76mm),
high resolution, 240 x 320 dot graphic with
back light (back light invoked via “LIGHT”
button on front panel)
Keypad 10 keys (ON, OFF, ESC, ENT,
LIGHT, PRINT
LED indicators DC IN (green), FAST
CHARGE (red)
Power down backup Ram memory backed
up by lithium battery (effective term, 5 yrs)
Response time 1 sec. or less
Output signal 0.8-20 mA DC, 4-20mA DC,
20-4-20mA DC, max. load resistance 1K
ohms
Information subject to change without notice. Prices in USD.
g,f,h,i)
Input signal 4-20mA DC
Serial communication RS-232C, even,
odd or no parity, 1 or 2 stop bits, keypad
selectable; Transmission speed: 300, 600,
1200, 2400, 4800 or 9600BPS, keypad
selectable; Transmission distance: max.
50 ft. (15m)
Printer (option) Thermal serial dot printing
(8 x 256 dots), mounts to top of converter
Display language English, German,
French, Japanese (Katakana) and keypad
selectable
Measurement display screen Flow velocity
and rate (with ow direction) displayed
simultaneously. Flow rate units are
selectable via keypad; Units: English or
Metric system keypad selectable; English
system units: ow velocity in ft/sec, ow
rate units: gal/sec, gal/min, gal/hr, Mgal/
day, ft3/sec, ft3/min, ft3/hr, Mft3/day, bbl/
sec, bbl/min, bbl/hr, Mbbl/day; (gal = U.S.
gallons); Metric system units: ow velocity
in m/sec, ow rate units: l/s, l/min, l/h, Ml/d,
m3/sec, m3/min, m3/h, m3/day, Mm3/day,
bbl/sec, bbl/min, bbl/hr, Mbbl/day
Totalizer display screen Forward and
reverse totalizer values displayed
simultaneously. Totalizer units are
selectable via keypad; English system
units: gal, Kgal, ft3, kft3, Mft3, bbl, Kbbl,
Mbbl (gal = U.S. gallons); Metric system
units: ml, l, m3, km3, Mm3, bbl, Kbbl, Mbbl
Clock display function Time (year, month,
day, hour, minute) display and setting,
keypad selectable
Damping 0 to 99 sec. (time constant),
keypad selectable
Low flow cutoff 0 to 3.30 ft/sec (0 to 1.0
m/sec), keypad selectable
Output setting function Current output
scaling, output type: (0.8-20 mA, 4-20 mA,
20-4-20 mA), output calibration and burn
out setting (output selectable to hold last
value, force zero, high or low), keypad
selectable
Communication function Data logger le
transfer, ow velocity, forward and reverse
totalizer values, date/time and error status
transmission on request from host
Data logging function Site data which
includes site name, piping, uid type,
sensor type and mounting method. Up to
20 sites and a maximum of 40,000 data
points (ow velocity, ow rate, forward and
reverse totalizer values) can be stored in
battery backed-up memory
Waveform display function Oscilloscope
display of upstream and downstream
received signals
Error message library Denes all error
messages and recommends solutions for
resolving error
Graph display function Display of logged
data and ow rate trend graphs
Printing function Printout of screen
(invoked from “PRINT” button on front
panel), xed cycle printout including ow
velocity, ow rate, forward and reverse
totalizer values, error status, logged data
trend graph, date/time, site setup and
diagnostic waveforms. Printing functions
are active only when optional printer is
installed.
Transmit output voltage X1, X2, X4 and
X8 transmit output voltages, keypad
selectable
Detector Models FLD
Mounting method V or Z method mounted
to outside of pipe by means of steel bands,
nylon belts or steel wire
Straight pipe length Upstream side, 10d or
more, downstream side, 5d or more (d =
inside pipe diameter).
Signal cable RG-58 or RG-58A/U, 50 ohm
coaxial cable
Cable length Standard length 16 ft.,
maximum length 1000 ft. (300m)
Connection Converter: BNC connectors
(female). Detector models FLD22,
FLD32, FLD11: BNC connectors (female).
Detector Models FLD12, FLD41, FLD50:
internal terminal screw, strain-relieved,
water-resistant, water-proof optional
Pipe diameter range
FLD22 (Small diam. pipe range): 0.50
to 4.0 in. (13 to 100mm). FLD12 (Small
diam. pipe range): 2.0 to 16 in. (50 to
400mm). FLD11 (Standard pipe range):
2.0 to 16 in. (50 to 400mm). FLD41
(Universal pipe range): 8.0 to 48 in. (200
to 1200mm). FLD50 (Large diameter pipe
range): 8.0 to 235 in. (200 to 6000mm).
082106