High pressure liquids and gases are potentially hazardous. Energy stored in these liquids and gases
can be released unexpectedly and with extreme force. High pressure systems should be assembled
and operated only by personnel who have been instructed in proper safety practices.
Information in this document is subject to change without notice. No part of this document may be reproduced or transmitted in any
form or by any means, electronic or mechanical, for any purpose, without the express written permission of DH Instruments, a
Fluke Company 4765 East Beautiful Lane Phoenix AZ 85044-5318 USA.
DH Instruments makes sincere efforts to ensure the accuracy and quality of its’ published materials; however, no warranty,
expressed or implied, is provided. DH Instruments disclaims any responsibility or liability for any direct or indirect damages
resulting from the use of the information in this manual or products described in it. Mention of any product does not constitute an
endorsement by DH Instruments of that product. This manual was originally composed in English and was subsequently translated
into other languages. The fidelity of the translation cannot be guaranteed. In case of conflict between the English version and other
language versions, the English version predominates.
Products described in this manual are manufactured under international patents and one or more of the following U.S. patents:
5,142,483, 5,257,640, 5,331,838, 5,445,035. Other U.S. and international patents pending.
DH Instruments, DH, DHI, molbox, molbox RFM, molbloc, molbloc-L, molbloc-S and CalTool are trademarks, registered and
otherwise, of DH Instruments, a Fluke Company.
LabVIEW is registered trademark of National Instruments Corporation.
Swagelok is a registered trademark of the Swagelok Company.
1.2.1 GENERAL SPECIFICATIONS ............................................................................................................................................................. 2
1.2.3 TEMPERATURE MEASUREMENT SPECIFICATIONS......................................................................................................................3
1.2.5 FRONT AND REAR PANELS............................................................................................................................................................ 13
1.2.5.1 FRONT PANEL..............................................................................................................................................................................13
2.1 UNPACKING AND INSPECTION ...............................................................................................................................................15
2.1.1 REMOVING FROM PACKAGING...................................................................................................................................................... 15
2.2 SITE REQUIREMENTS............................................................................................................................................................... 15
2.3.1 PREPARING FOR OPERATION ....................................................................................................................................................... 16
2.3.2 POWER CONNECTION ..................................................................................................................................................................... 16
2.3.3 MOLBOX RFM TO MOLBLOC CONNECTION.................................................................................................................................16
2.3.4 GAS SUPPLY AND FLOWPATH CONNECTIONS........................................................................................................................... 17
2.4 POWER UP AND VERIFICATION.............................................................................................................................................. 18
2.4.1 POWER UP.........................................................................................................................................................................................18
2.4.2 CH E CK P ROPER PRESS URE MEASU REMEN T OP ERATI O N .......................................................................................... 19
2.4.3 CHECK PROPER TEMPERATU R E M EASUR E M ENT OPERATIO N..........................................................................................19
2.5 ADDITIONAL PRECAUTIONS TO TAKE BEFORE MAKING FLOW MEASUREMENTS .......................................................20
2.6 SHORT TERM STORAGE...........................................................................................................................................................20
3.1 GENERAL OPERATING PRINCIPLES ......................................................................................................................................21
3.1.1 MOLBLOC-L AND MOLBLOC-S OPERATION................................................................................................................................ 21
3.2 MAIN RUN SCREEN................................................................................................................................................................... 25
3.3.4 DIRECT FUNCTION KEYS SUMMARY ............................................................................................................................................ 28
3.4 DIRECT FUNCTION KEYS.........................................................................................................................................................29
3.4.3.1 MASS FLOW VS. VOLUME FLOW...............................................................................................................................................35
3.4.3.2 VOLUMETRICALLY BASED MASS FLOW UNITS....................................................................................................................... 36
3.4.3.3 VOLUMETRICALLY BASED MASS FLOW UNITS AT VARIOUS REFERENCE TEMPERATURES (UXXX) ........................... 37
3.4.3.4 VOLUME FLOW UNITS (VLM)......................................................................................................................................................37
3.4.3.5 CUSTOMIZING FLOW UNITS AVAILABLE UNDER THE UNIT FUNCTION.............................................................................. 38
3.4.4.4.2 RUN AUTOZ............................................................................................................................................................................. 56
3.4.5 [P&T] (PRESSURE AND TEMPERATURE)......................................................................................................................................59
3.6.6.1 COM1 AND COM2......................................................................................................................................................................... 88
4.3.4.1 IEEE STD. 488.2 COMMON AND STATUS COMMANDS.........................................................................................................101
4.4 STATUS SYSTEM.....................................................................................................................................................................129
4.4.1 STATUS REPORTING SYSTEM ..................................................................................................................................................... 129
4.4.1.1 STATUS BYTE REGISTER.........................................................................................................................................................129
4.4.1.2 STANDARD EVENT REGISTER................................................................................................................................................. 131
5. MAINTENANCE, ADJUSTMENTS AND CALIBRATION ...........133
5.2.1.1 PA AND PM COEFFICIENTS...................................................................................................................................................... 135
5.2.3 SET-UP AND PREPARATION.........................................................................................................................................................136
5.2.3.1 UPSTREAM AND DOWNSTREAM ABSOLUTE RPTS.............................................................................................................. 136
5.2.4.2 VIEWING AND EDITING RPT PA, PM AND CALIBRATION DATE...........................................................................................140
5.2.5 RPT CALIBRATION/ADJUSTMENT PROCEDURE WITHOUT USING CALTOOL FOR RPTS SOFTWARE ............................ 141
5.3 OHMIC MEASUREMENT SYSTEM VERIFICATION................................................................................................................142
5.4 RELOADING EMBEDDED SOFTWARE INTO MOLBOX RFM FLASH MEMORY.................................................................143
5.6.1.5 POWER SUPPLY........................................................................................................................................................................ 146
5.6.1.8 MAIN BOARD ..............................................................................................................................................................................147
5.6.1.9 COOLING FAN ............................................................................................................................................................................147
7.1 CONVERSION OF NUMERICAL VALUES...............................................................................................................................155
This manual provides the user with the information necessary to operate a molbox RFM, Reference Flow
Monitor. It also includes a great deal of additional information provided to help you optimize molbox RFM
use and take full advantage of its many features and functions.
Before using the manual, take a moment to familiarize yourself with the Table of Contents structure: All first
time molbox RFM users should read Section
molbox RFM operating principles. Section 4 is for
provides maintenance and calibration information. Section 6 is a quick troubleshooting guide. Use
troubleshoot unexpected molbox RFM behavior based on the symptoms of that behavior.
Certain words and expressions have specific meaning as they pertain to molbox RFM. Section 8 is useful as
a quick reference for exact definition of specific words and expressions as they are used in this manual.
For those of you who “don’t read manuals”, go directly to section 2.3, initial setup, to set up your
molbox RFM. Then go to section 2.4, power up and verification. This will get you running quickly with
minimal risk of causing damage to yourself or your molbox RFM. THEN… when you have questions or
start to wonder about all the great features you might be missing, get into the manual!
T
T
HHIISS
M
AANNUUAAL
M
L
2. Section 3 provides a comprehensive description of general
remote operation from an external computer. Section 5
Manual Conventions
it to
(CAUTION) is used in throughout the manual to identify user warnings and cautions.
(NOTE) is used throughout the manual to identify operating and applications advice and
additional explanations.
[ ] indicates direct function keys (e.g., [RANGE]).
molbox RFM is a support unit for making low mass flow measurements using molbloc mass
flow elements. molbox RFM reads calibration data off the molbloc EEPROM and measures molbloc
upstream and downstream pressure using built-in high accuracy Reference Pressure
Transducers (RPTs). An ohmic measurement system reads the resistance of the molbloc platinum
resistance thermometers from which molbloc temperature is calculated. Using the molbloc calibration
data, pressures, temperature and gas properties stored in memory, the flow rate of the gas flowing
through the molbloc is calculated. A microrange option is available to increase resolution and accuracy
below 10 % FS of the molbloc-L flow range.
Internal molbox RFM valving supports on-board PRESSURE TRANSDUCER TARING, LEAK TESTING
and SELF PROTECTION functions as well as a gas purge routine.
molbox RFM provides a local user interface via a front panel key pad and display and includes advanced
on-board functions. Remote communication capability is supported with RS232 and IEEE-488 interfaces.
molbox RFM is intended for applications in which a highly compact presentation, high range ability and
lower cost are the primary considerations. A second model, molbox1, is available for applications where
higher accuracy is the most important requirement.
1.1.1 MOLBLOC FLOW ELEMENTS
Two different types of molblocs may be used with molbox RFM; molbloc-L (laminar) and
molbloc-S (sonic).
1.1.1.1 MOLBLOC-L FLOW ELEMENT
molbloc-L is the original molbloc laminar flow element. molbloc-L covers the
lower portion of the molbloc/molbox system flow range. The key molbloc-L
measurement is the differential pressure across the element, which is roughly
proportional to the mass flow rate through it. molbloc-L elements are calibrated
to be used at an absolute pressure that remains nearly constant, while the
differential pressure varies with flow rate. Different operating pressure options
and their effect on molbloc flow range are described in Section 1.2.4.1.2.
1.1.1.2 MOLBLOC-S FLOW ELEMENT
molbloc-S elements use critical (sonic) flow venturi nozzle technology to
measure flows, which overlap with the higher molbloc-L ranges and extend the
high end of the molbloc/molbox RFM system flow range. The mass flow rate
through a molbloc-S element is roughly proportional to the upstream absolute
pressure when the flow is “choked”, so the molbloc-S operating pressure can
vary widely as the mass flow rate is changed throughout the flow range. The
limits of molbloc-S operating pressure and flow ranges are defined by the
molbloc-S calibration type, described in Section 1.2.4.2.2.
Accuracy: ± 0.02 % of reading
Accuracy of 100 and 110 Ω reference resistors: ± 0.01 %
Stability of 100 and 110 Ω reference resistors: ± 0.005 % per three years
With molbloc-L Nitrogen (N2), Air, Argon (Ar), Butane (Butn), Carbon
With molbloc-S Nitrogen (N2), Air (dry and humid)
Quick connectors equivalent to Swagelok QM Series (-QM2-B200)
Maximum working pressure 600 kPa absolute (87 psia)
Maximum pressure without damage 800 kPa absolute (115 psia)
NOTE: The microrange option includes a 12.5 kPa (1.8 psi) differential
RPT which may be damaged by differential pressure greater than 100
kPa (15 psi).
See Sections 1.2.4.1.3, 1.2.4.1.4and 1.2.4.2.1
Available. Must be specified.
Piezoresistive silicon
0 to 600 kPa absolute (0 to 87 psia)
6.0 Pa (0.0009 psi)
± 0.01 % FS
Absolute measurements: ± 0.05 % FS for one year
Differential measurements: ± (20 Pa + 0.05 % ΔΡ). Taring sequence eliminates zero
error on measurement of difference between the two RPTs.
1. INTRODUCTION
1.2.2.2DIFFERENTIAL RPT (MICRORANGE OPTION)
Type
Range
Resolution
Repeatability
Accuracy
Piezoresistive silicon
0 to 12.5 kPa differential (0 to 1.8 psid)
0.14 Pa (0.00002 psi)
± 0.01 % FS
± 0.05 % FS for one year. Taring sequence eliminates zero error.
1.2.3 TEMPERATURE MEASUREMENT SPECIFICATIONS
Specifications are for molbloc mounted Platinum Resistance Thermometers (PRT) combined
with molbox RFM resistance measurement system and temperature calculation.
Range
Accuracy
Resolution
0 to 40 °C
± 0.05 °C
0.01 °C
The molbox RFM internal resistance measurement system is automatically calibrated using
reference 100 and 110 Ω (± 0.01 %) resistors (see Section
5.3).
1.2.4 FLOW MEASUREMENT SPECIFICATIONS
molbox RFM measures the flow through molbloc flow elements. There are two different
types of molblocs, molbloc-L (laminar) (see Section 1.1.1.1) and molbloc-S (sonic) (see
Section 1.1.1.2). Flow measurement specifications, calibration types, ranges and dimensions
are detailed separately for each molbloc type in Section 1.2.4.1 and 1.2.4.2.
1.2.4.1 molbloc-L
The flow range, useable operating pressure and absolute and differential
pressure associated with molbloc-L operation depend on the molbloc used and
its pressure-dependent calibration options (see Section
Measurement Update Rate
Range
Resolution
Linearity
Repeatability
Precision
Stability
(1 year)
Measurement Uncertainty
(1 year, N2 and any molbox RFM
supported gas for which the molbloc
1 Precision: Combined linearity, hysteresis, repeatability.
2 Stability: aximum change in zero and span over specified time period for typical molbox RFM and
molbloc used under typical conditions. As stability can only be predicted, stability for a specific
molbox RFM should be established from experience.
3 Measurement Uncertainty (Accuracy): Maximum deviation of the molbox RFM flow indication from
the true value of the flow through the molbloc including precision, stability and DHI calibration
standard measurement uncertainty.
in use is calibrated)
1 second
0 to molbloc full scale depending on gas and molbloc
pressure dependent calibration type
(see Section 1.2.4.1.2).
0.01 % FS
± 0.23 % of reading from 10 to 100 % FS,
± 0.023 % FS under 10 % FS
± 0.1 of reading from 10 to 100 % FS,
± 0.01 % FS under 10 % FS
1
± 0.25 % of reading from 10 to 100 % FS,
± 0.025 % FS under 10 % FS
2
± 0.15 % of reading from 10 to 100 % FS,
± 0.015 % FS under 10 % FS
3
± 0.5 % of reading from 10 to 100 % FS,
± 0.05 % FS under 10 % FS
The microrange option (see Section 1.2.2.2) improves molbloc-L flow
measurement specifications below 10 % FS of the molbloc range. With the
microrange option, the affected measurement specifications below 10 % FS
become:
Resolution
Linearity
Repeatability
Precision1
Stability
(1 year)
Measurement Uncertainty
(1 year, N2 and any molbox RFM
supported gas for which the molbloc in
use is calibrated)
1 Precision: Combined linearity, hysteresis, repeatability.
2 Stability: Maximum change in zero and span over specified time period for typical molbox RFM and
molbloc used under typical conditions. As stability can only be predicted, stability for a specific
molbox RFM should be established from experience.
3 Measurement Uncertainty (Accuracy): Maximum deviation of the molbox RFM flow indication from
the true value of the flow through the molbloc including precision, stability and DHI calibration
standard measurement uncertainty.
0.01 % of 10 % FS (0.001 % FS)
± 0.23 % of reading from 1 to 10 % FS
± 0.1 % of reading from 1 to 10 % FS
± 0.25 % of reading from 1 to 10 % FS
See your molbloc’s Calibration Report to determine the calibration type of
the molbloc you are using.
Different pressure dependent calibration options for molbloc-L elements determine
the range of operating pressures over which a molbloc can be used within its
mass flow measurement specifications. The calibration option also affects the
molbloc flow range and the differential pressure associated with the flow range.
Measurement uncertainty (accuracy) specifications for molbloc-L are valid only
for gases with which the molbloc has been calibrated. All molbloc-L elements
are calibrated for N2. Calibrations with other gases are optional. DHI calibration
capability is not maintained at all times for all gases on all molbloc designations.
Check for availability before ordering calibrations.
1.2.4.1.3molbloc-L Ranges with Low Pressure Calibrations
Table 2. molbloc-L Ranges with Low Pressure Calibrations
- full mod, low pressure
- full mod, downstream
- single P, low pressure
SIZE
Nitrogen N2
Argon Ar
Helium He
Sulfur Hexafluoride SF6
INERT
Xenon XE
Butane C4H10
Ethane C2H6
Ethylene C2H4
Hydrogen H2
FLAMMABLE
Methane CH4
Propane C3H8
Carbon
Tetrafluoride
Hexafluorethene C2F6
FLUORO-
CARBONS
Trifluoromethane CHF3
Air Air
Carbon Dioxide CO2
Carbon Monoxide CO
Nitrous Oxide N2O
OTHER
Octafluorocyclobutane
Oxygen O2
A bold value indicates that the maximum flow is limited by the maximum Reynolds number value of 1 200 which is reached
before the normal differential pressure range is reached. In that case, the second value gives the minimum flow for which
relative accuracy is ± 0.5 % of the measured value. With the microrange option, this value is divided by 10 (see Section
1.2.2.2).
Where there is no value in the field (--), this indicates that the maximum Reynolds number is reached before the differential
pressure reaches 5 kPa (1 kPa in the case of the 1E5 molbloc), therefore calibration with that gas is not useful.
[1] Due to low vapor pressure, only downstream calibration type is available.
A bold value indicates that the maximum flow is limited by the maximum Reynolds number value of 1 200 which is reached before
the normal differential pressure range is reached. In that case, the second value gives the minimum flow for which relative accuracy
is ± 0.5 % of the measured value. With the microrange option, this value is divided by 10 (see Section 1.2.2.2).
Where there is no value in the field (--), this indicates that the maximum Reynolds number is reached before the differential pressure
reaches 5 kPa (1 kPa in the case of the 1E5 molbloc), therefore calibration with that gas is not useful.
* * * * * * * * * N/A
25 125 250 500 1 250 2 500
11 000
1 500
20 000
3 000
20 000
9 000
N/A
[2] The operating pressure range is greater than the vapor pressure value for the gas.
The flow range and operating pressure associated with molbloc-S operation
depend on the molbloc used and its calibration options (see Section 1.2.4.2.2)
2. Stability: Maximum change in zero and span over specified time period for typical molbox RFM
and molbloc used under typical conditions. As stability can only be predicted, stability for a
specific molbloc and molbox RFM should be established from experience.
3. Measurement uncertainty (accuracy): Maximum deviation of the molbox RFM flow indication
from the true value of the flow through the molbloc including precision, stability and DHI
calibration standard measurement uncertainty.
1 second
Rate
Depends on molbloc-S pressure dependent calibration type
(see Section 1.2.4.2.2)
0.01 % of FS
± 0.25 % of reading
± 0.10 % of reading
1
± 0.30 % of reading
2
± 0.2 % of reading
± 0.5 % of reading from 50 to 500 kPa
3
± 0.5 % of reading from 50 to 200 kPa
3
± 0.5 % of 50 kPa flow from 20 to 50 kPa
1.2.4.2.1 molbloc-S RANGES
molbloc-S flow ranges are defined by the molbloc’s Pressure to Flow Conversion
Ratio, K
downstream pressure and the acceptable back pressure ratio (see Section
3.1.2). K
between mass flow and the absolute upstream pressure delivered to the
molbloc-S. molbloc-S sizes are defined by the nominal K
nozzle, using scientific notation, for example a 1E3 molbloc-S has a K
sccm/kPa. To differentiate from molbloc-L size designations, this molbloc size is
designated 1E3-S.
The molbox RFM pressure range, the molbloc-S calibration type (see Section
1.2.4.2.2) and the back pressure ratio (BPR) requirements (see Section 3.1.2)
limit the pressures, and flows, over which a molbloc-S can be used within known
measurement uncertainty limits. In practice, the usable range of a molbloc-S in a
given application also may depend on the available gas supply pressure, the presence
and flow capacity of a vacuum pump downstream or the allowable back pressure
on an upstream DUT.
The flow ranges for each molbloc-S size at various typical operating pressures
are summarized in Table 4 and the BPR limits are in Table 5.
, the absolute pressure that can be delivered upstream of molbloc-S, the
F
is expressed in units of sccm/kPa and defines the relationship
[1] When volumetrically based mass flow units with reference temperatures other than 0°C are used, flow values
will generally be higher; for example, the flow values for a given molbloc and upstream pressure are
approximately 7% higher when expressed in slm @ 20°C. Flow values at a given pressure may vary by up to
2% due to flowpath machining tolerances.
[2] Flow values in table are valid only when critical flow is established.
[3] Minimum upstream pressure to achieve critical flow with atmospheric pressure (approximately 100 kPa)
downstream of molbloc-S (no vacuum).
Table 5. Minimum molbloc-S Critical Flow (slm) at Various molbloc-S Downstream Pressures
[1] When volumetrically based mass flow units with reference temperatures other than 0 °C are used, flow values
will generally be higher; for example, the flow values for a given molbloc and upstream pressure are
approximately 7 % higher when expressed in slm @ 20 °C. Flow values at a given pressure may vary by up to
2 % due to flowpath machining tolerances.
[2] Limited by 20 kPa minimum calibration pressure rather than back pressure ratio.
See your molbloc’s Calibration Report to determine the calibration type of
the molbloc you are using.
Measurement uncertainty (accuracy) specifications for molblocs are valid only for
gases with which the molbloc has been calibrated. All molbloc-S elements are
calibrated in one standard gas, either air or N2, and may be calibrated in other
gases. Calibrations with other gases are optional. The set of gases which can
be measured by molbloc-S is separate from the list of molbloc-L gases, and may
be more limited. DHI calibration capability is not maintained at all times for all
gases on all molbloc sizes. Check for availability before ordering calibrations.
molbloc-S calibrations are performed over flow ranges corresponding to one of
two pressure ranges, summarized in Table 6.
Table 6. molbloc-S Calibration Types
CALIBRATION TYPE
LP
low pressure
SP
standard pressure
OPERATING PRESSURE
20 to 200 kPa absolute (3 to
30 psia
upstream of molbloc
50 to 500 kPa absolute (7 to
70 psia)
upstream of molbloc
molbloc-S flow measurements are valid only when the ratio of pressure
downstream to pressure upstream of the nozzle is high enough to assure a
critical (choked) flow (see Section 3.1.2).
the venturi nozzle extends
beyond the molbloc
downstream flange, making
the overall length dimension,
F, longer than the fitting to
fitting length dimension, K.The
nozzle overhang may interfere
with some molbloc-S
downstream connections or
the connection of a blank off
cap for leak testing, so a 40
mm diameter ISO-KF nipple is
supplied with 5E3-S and 1E4S molblocs.
[2] Default connector type is
listed. Additional upstream
connector options may be
available. Contact your DHI
Sales Representative for
details.
The rear panel assembly provides pressure and electrical connections to a
molbloc mass flow element, communications interfaces and the power
connection module.
1. Quick connectors to molbloc pressure connections (color coded)
2. Electrical power connector (IEC320-313)
3. IEEE-488 (GPIB) connector for host communications
4. COM1 (RS232) connector for host communications
5. Electrical connection to molbloc
6. COM2 (RS232) connector for communications with an external device
molbox RFM is delivered, along with its standard accessories, in a corrugated container with
polyurethane inserts to hold it in place.
Remove the molbox RFM and its accessories from the shipping container and remove each
element from its protective plastic bag.
2.1.2 INSPECTING CONTENTS
Check that all items are present and have NO visible damage.
A molbox RFM includes:
2. INSTALLATION
Table 7. molbox RFM Parts List
DESCRIPTION PART #
molbox RFM Reference Flow Monitor FAM0005
Report of Calibration 550100
ACCESSORIES, INCLUDING:
1 Operation and Maintenance Manual 550107
1 Power Cord (7.5 ft.) 100770
1 Set of (2) molbox to molbloc pressure connecting
tubes
1 molbox to molbloc electrical/data connection
cable
2 straight through quick connector stem 101889*
1 General Accessories CD (white)
(Important: Includes system support software
and documentation.)
* Equivalent to Swagelok P/N SS-QM2-S-200
2.2 SITE REQUIREMENTS
Install molbox RFM on any stable surface at a convenient height. The front feet are
extendible so that the unit can be inclined for easier viewing.
The molbox RFM can also be mounted in a standard 19-in. rack mount using the optional
rack mount kit (P/N 401465). For additional information, contact your DHI Sales
Representative.
401125
102096
102987
When installing molbox RFM, consideration should be given to where the molbloc flow
measuring element and associated hardware will be located. molbox RFM may be placed on
a shelf or cart at a different height than the molbloc, but the distance between the molbloc
and molbox is limited by the length of the cable and pneumatic lines connecting them. If you
will locate the molbox at a different height than the molbloc, the small errors that would be
associated with the difference in pressure can be removed using the molbox RFM head
correction (see Section 3.6.8).
If the molbloc/molbox system is being used to calibrate other devices (DUTs), the molbloc
may need to be connected upstream or downstream of the DUT to operate within the
pressure limits of the molbloc’s calibration type (see Sections 1.2.4.1.2, 2.2.4.2.2) and to
accommodate the pressure requirements of the DUT. Se e th e mo lb lo c’ s Ca l ib ra ti on R eport to
determine the calibration type of the molbloc you are using.
If the molbloc is connected upstream of the DUT, it is important to supply the molbloc with a
stable regulated gas source. The volume present between the molbloc and the device to be
calibrated should be minimized for low flows.
In some cases, molbloc-S is used with a vacuum source downstream to reduce the pressure
at w h i c h critical flow is reached. Consider the placement of the vacuum pump and connections.
Generally, a large vacu u m p u m p i s needed that should be isolated from the work area due to
noise and oil vapor considerations. If the vacuum pump and/or vacuum kit was purchased
from DHI, see the instruction sheets and/or manuals that are included with the hardware.
Optional molstics are offered for mounting molblocs. They provide a convenient means of
addressing supply regulation, filtering and interconnection issues with high quality, configured
hardware. For additional information, contact your DHI Representative.
If a DUT is located upstream of the molbloc and is contaminated, contaminates may flow
from the DUT to the molbloc and alter the molbloc calibration. If the DUT must be connected
upstream of the molbloc, be sure it is clean before flowing and consider installing a filter
between the DUT and the molbloc.
2.3 INITIAL SETUP
2.3.1 PREPARING FOR OPERATION
To prepare molbox RFM for check out and operation:
n Remove the plastic caps from the molbox RFM rear panel pressure connections.
o Remove the protective plastic sheet from the front panel display.
p Familiarize yourself briefly with the front and rear panels (see Section 1.2.5).
Follow the steps described in Sections 2.3.1 to 2.3.6
2.3.2POWER CONNECTION
Connect the power cable supplied to molbox RFM and to a power source. Power
requirements are 85 to 264 VAC, 50 to 60 Hz, 22 VA max. consumption.
molbox RFM is always powered and active when power is supplied through the rear panel
power connector. The front panel ON/OFF key controls a SOFT ON/OFF (see Section 3.3.3).
2.3.3 MOLBOX RFM TO MOLBLOC CONNECTION
There are two pressure connections (upstream and downstream) and one electrical/data
connection between molbox RFM and a molbloc.
For the pressure connections, use the molbox RFM to molbloc pressure tubes (P/N 401125)
supplied with the molbox RFM. Following the color coding on the pressure lines, connect the
upstream (HI) molbox RFM rear panel quick connector to the upstream port of the molbloc
and the downstream (LO) quick connector to the downstream port. Push firmly on the quick
connectors until they click into place to assure that the connection is properly completed.
For the electrical/data connection, use the molbox to molbloc electrical/data connection cable
(P/N 102096). Connect the cable to the molbloc and then to molbox RFM rear panel
connector labeled “molbloc”.
Avoid making molbloc electrical connections to molblocs while molbox RFM is plugged in.
Damage to the molbloc EEPROM may result (see Section 5.5.5)
2.3.4 GAS SUPPLY AND FLOWPATH CONNECTIONS
Connect a gas supply to the molbloc. Gas supply requirements are:
•The gas supply must be clean and dry (especially free from oil and particulates) to avoid
contaminating the molbloc.
•For correct measurements, the gas must be of the same species as that selected by the
molbox RFM GAS function (see Section 3.4.2). Gas purity affects the measurement
uncertainty of flow measurements as molbox RFM uses the thermodynamic properties
of the flowing gas in its flow calculations. Generally, gases with purity of 99.9 % or
better are used for molbloc measurements. Except when using ambient air with
molbloc-S, the test gas should be free of any humidity (dew point less than – 40 °C).
•If the molbloc is connected upstream of the DUT, the supply pressure must be regulated
and stable within the limits of the molbloc-L pressure dependent calibration type (see
Section
connected downstream of the DUT, use regulators and
1.2.4.1.2) or molbloc-S calibration type (see section 1.2.4.2.2). If the molbloc is
valves to make sure that the
pressure that is delivered to the molbloc will be within the limits of the molbloc calibration type.
•Care should also be taken to make sure that the pressure and flow supplied to the
molbloc are always low enough to avoid over pressuring the molbox RFM RPTs (see
Sections 1.2.2, 3.1.6). If a DUT upstream of the molbloc is operated at high pressure, a
pressure reducing regulator should be connected between the DUT and the molbloc to
ensure that even momentary high pressure spikes do not reach the molbox RPTs.
The gas supplied to the molbloc should be clean and dry. Contamination of the molbloc
flow passage with liquids, particulates or any other matter will alter the molbloc
calibration and can lead to out of tolerance flow measurements.
NEVER connect a pressure source to the molbloc that is greater than the overpressure
limit of your molbox RFM. molbox RFM overpressure limit is 660 kPa absolute/560 kPa
gauge (95 psia/80 psig). Overpressure can damage the molbox RFM internal RPTs (see
Section 1.2.2, 3.1.6).
If you are using a DHI molstic: Install the molbloc into the molstic and connect a gas
supply following the molstic instruction sheet or manual. The flow through the molbloc must
be in the direction of the arrow engraved on the molbloc.
If you are NOT using a DHI molstic: Connect a gas supply to the molbloc according to the
molbloc instruction sheet and the pressure limits of the molbloc calibration type. A valve
should be installed between the pressure supply and the molbloc to allow flow to the molbloc
to be interrupted. The flow through the molbloc must be in the direction of the arrow
engraved on the molbloc.
Adaptor kits are available from DHI to make connections from the molbloc or molstic fittings
to other common connector types. Ask you DHI Sales Representative about your specific
adaptor requirements.
Operating at pressures other than those of the molbloc-L pressure dependent calibration type
(see Section 1.2.4.1.2) or molbloc-S calibration type (see Section 1.2.4.2.2) may result in
out of tolerance flow measurements. Refer to the molbloc Calibration Report to
determine its calibration type.
2.3.5 VACUUM SUPPLY (molbloc-S ONLY)
In some cases molbloc-S is operated with a vacuum downstream of the molbloc to reduce the back
pressure so that critical flow can be reached at a lower upstream pressure (see Section 3.1.2).
There is no lower limit to the pressure that may safely be applied to molbox RFM RPTs.
If you are using DHI supplied downstream vacuum connect kit and vacuum pump:
Install the kit and pump following the supplied instruction sheet or manual. Carefully follow
the pump manufacturer’s recommendations for pump operation.
If you are NOT using DHI supplied downstream vacuum connect kit and vacuum
pump: Carefully evaluate vacuum pump specifications to be sure that the vacuum source
available has the pumping speed necessary to safely handle the planned flows and to
maintain low enough pressure at planned flow rates.
Be sure to provide facilities to avoid flowing into the pump when the pump is not ON as this
will cause pressure to build up on the pump and may damage it. Normally, a shut-off valve
should be included between the pump and the molbloc-S.
It is preferable to install a check valve with very low cracking pressure between the molbloc
and the vacuum shut-off valve.
Adaptor kits are available from DHI to make connections from the molbloc or molstic fittings
to other common connector types. Ask you DHI Sales Representative about your specific
adaptor requirements.
Operating at pressures other than those of the molbloc-S calibration type may result in
out of tolerance flow measurements (see Section 1.2.4.2.2). Refer to the molbloc
Calibration Report to determine its calibration type.
2.3.6 COMMUNICATIONS CONNECTIONS
If molbox RFM is being interfaced to a computer, connect an RS232 cable to molbox RFM
COM1 or an IEEE-488 cable (cables not supplied) to the molbox RFM IEEE-488 interface.
Configure the interface (see Section
3.6.6).
2.4 POWER UP AND VERIFICATION
2.4.1 POWER UP
Connect the molbox RFM power cable to an electric supply of 85 to 264 VAC (47 to 440 Hz).
Observe the front panel display as molbox RFM initializes, error checks, calibrates its internal
ohmic measurement system and goes to the main run screen (see Section 3.2). The top left
side of the main run screen should display a flow value near zero or <BPR HI>. If <NO BLOC> is displayed, molbox RFM has not been able to identify a molbloc connection and
load molbloc information. Verify that a valid molbloc is properly connected (see Section
2.3.3) and press [SETUP], <4molbloc> to load the molbloc data. If molbox RFM is still
unable to identify a molbloc, the molbloc may require reloading of EEPROM information or
molbox RFM may require repair.
If the molbox RFM fails to reach the main run screen: Service may be required. Record
the sequence of operations and displays observed and contact a DHI Authorized Service
Provider (see Section 7.2).
molbox RFM is always powered and active when power is supplied through the rear panel
power connector. The front panel ON/OFF key controls a SOFT ON/OFF only (see Section
3.1.4).
2.4.2 CHECK PROPER PRESSURE MEASUREMENT OPERATION
Check that the molbox RFM pressure measurements are operating properly. Proceed as
follows:
n Connect the molbloc to the molbox RFM (see Section
o Shut off the gas supply to the molbloc and open one or
2.3.3).
both molbloc ends to atmospheric
(ambient) pressure.
p Press [P&T] and observe the display of the pressure measured by the upstream and
downstream absolute RPTs. Observe the upstream and downstream pressures
(see Section 3.4.5). These should indicate current atmospheric pressure and be in
agreement within ± 0.5 kPa (0.1 psi). If the two readings disagree by more than ± 0.5
kPa (0.1 psi), one or both RPTs may be out of calibration and service may be required.
A difference in the RPT readings could also indicate that there is some flow through the
molbloc. Check the flowpath valve or disconnect the molbloc to ensure that there is
actually no flow.
q If the molbox RFM is equipped with the microrange option, and the option is currently
active, the bottom line left side of the display is <mDP>. Observe the differential
pressure value following this indication (see Section 3.4.5). It should be 0, ± 50 Pa
(0.004 psi). If the indication is different from zero by more than ± 50 Pa (0.004 psi), the
microrange differential RPT may be out of calibration and service may be required.
Again, a non-zero differential pressure could indicate an unexpected flow through the
molbloc.
r Press [ESCAPE] to return to the main run screen.
2.4.3 CHECK PROPER TEMPERATURE MEASUREMENT OPERATION
Check that the molbox RFM temperature measurement is operating properly.
Proceed as follows:
n Connect a molbloc to the molbox RFM (see Section
o From the molbox RFM main run
screen, press [P&T] twice to arrive at the temperature
2.3.3).
display screen. Observe the temperature readings of the two molbloc PRTs
(see Section 3.4.5). If the molbloc has been in a stable temperature environment for 30
to 60 minutes, the temperature indications should be roughly ambient temperature and
the two indications should agree within ± 0.2 °C. If the two readings disagree more than
± 0.2 °C, there may be a problem with the molbloc or the molbox RFM TEMPERATURE
MEASUREMENT function and service may be required.
2.4.4 LEAK CHECK
It is recommended that a new molbox RFM be leak checked at start-up to assure that no
internal leaks developed during shipping and handling. Run the molbox RFM on-board
INTERNAL LEAK CHECKING function (see Section 3.4.5).
molbox RFM has a security system based on user levels. By default, the security system is
set to “low”, which includes certain access restrictions, and there is no password required to
change the security level (see Sections 3.6.2). As part of the molbox RFM startup, set your
desired security level and a password.
molbox RFM is delivered with the security level set at low to avoid inadvertent altering of
critical internal settings but with access to changing security levels unrestricted. It is
recommended that the low security level be maintained at all times and password
protection be implemented if control over setting of security levels is desired.
2.5 ADDITIONAL PRECAUTIONS TO TAKE BEFORE MAKING
FLOW MEASUREMENTS
Before using molbox RFM to make meaningful flow measurements, consider the following:
• Be sure that the gas pressure connected to the molbloc is not great enough to overpressure the
molbox RFM internal RPTs.
•For molbloc-L operation, the pressure measuring RPTs should be tared at the operating line pressure
(see Section 3.4.4.1).
•Operating pressure should be within the limits of the molbloc calibration type (see Sections 1.2.4.1.2.
and 1.2.4.2.2).
• The gas type selected should be the gas flowing through the molbloc (see Section 3.4.2).
• For best accuracy, the gas type should be a gas with which molbloc has been calibrated. See the
molbloc Calibration Report or press [SETUP], <4molbloc>, [ENTER] to see if the gas is included in
the molbloc calibration gas list (see Section 3.5.4).
• Do not supply a gas or connect a device under test upstream of the molbloc that may contaminate
the molbloc.
• Be sure the flow unit of measure you are using is correct. Many different types of flow units of
measure are commonly used. Before selecting a unit of measure, familiarize yourself with Section
3.4.3 thoroughly.
• Troubleshooting: For information on typically encountered start-up and operational issues, see
Section 6.
2.6 SHORT TERM STORAGE
The following is recommended for short term storage of molbox RFM:
• Vent the molbox RFM pressure ports.
• Disconnect the power supply.
When molbox RFM will NOT be used for some time, it may be left powered. Use the SOFT ON/OFF key
to turn OFF the display.