Defibrillator Analyzer Specifications
Energy Output Measurement
Compatible Defibrillator Waveshapes ................ Lown, Edmark, Trapezoidal, DC Bi-phasic, and AC Pulsed Bi-phasic
Note
AC Pulsed Biphasic waveform has not been approved in the United States.
Autoranged Measurement.................................... 0.1 to 600 J
Accuracy
0.1 to 360 J ..................................................... ±(1 % of reading + 0.1 J)
360 to 600 J .................................................... ±(1 % of reading + 0.1 J), typical
Note
For Pulsed Bi-Phasic defibrillator, specified accuracy is ±(1.5 % of reading + 0.3 J) on both ranges.
Load resistance
Resistance .......................................................... 50 Ω
Accuracy ......................................................... ±1 %, non-inductive (<2 μH)
Pulse trigger level................................................. 20 V
Pulse width
Range ................................................................. 1.0 to 50.0 ms
4/08 1
Manual Supplement Impulse 6000D/7000DP Users
Accuracy............................................................. ±0.1 ms
Voltage
Range ................................................................. 20 to 5000 V
Accuracy............................................................. ±(1 % of reading + 2 V)
Current
Range ................................................................. 0.4 to 100.0 A
Accuracy............................................................. ±(1 % of reading + 0.1 A)
Tilt (biphasic and pulsed biphasic)
Range ................................................................. 1 % to 99 %
Accuracy............................................................. ±1 digit
Interphase delay (biphasic and pulsed biphasic)
Range ................................................................. 0.1 ms to 9.9 ms
Accuracy............................................................. ±0.1 ms
Frequency (pulsed biphasic only)
Range ................................................................. 2000 Hz to 8000 Hz
Accuracy............................................................. ±1 % of reading
Duty cycle (pulse biphasic only)
Range ................................................................. 1 % to 99 %
Accuracy............................................................. ±1 digit
Sample rate ........................................................... 250 kHz (4 μs sample)
Maximum Average Power.................................... 12 W, equivalent to 10 defib pulses of 360 J every 5 minutes
2 8/08
Impulse 6000D/7000DP Users Manual Supplement
Scope Output
Autorange ........................................................... 2000:1, 400:1 and 80:1: dependant on the range
Waveform Playback
Output ............................................................. BNC
Output impedance........................................... 50 Ω
Amplitude Accuracy ........................................ ±5 %
Charge Time Measurement
Range ................................................................. 0.1 to 100.0 s
Accuracy ............................................................. ±0.05 s, typical
Synchronization Test (Elective Cardioversion)
Delay Time Measurement
Timing window ................................................ ECG R-wave peak to the defib pulse peak
Range .............................................................-120 to +380 ms; measures timing from 120 ms prior to the R-wave peak to up to 380
ms following the R-wave peak.
Resolution ....................................................... 1 ms
Accuracy ......................................................... ±1 ms
ECG waves
Normal Sinus Rhythm (NSR) .......................... 10 to 180 (by 1) BPM
Atrial fibrillation................................................ Coarse and fine
Monomorphic Ventricular Tachycardia ...........120 to 240 (by 5) BPM
Asystole ..........................................................Flat line
Automated Defibrillator Test ECG Waves
Normal Sinus ......................................................10 to 300 (by 1) BPM
4/08 3