Fluke i1000s Quick Start Guide

Page 1
i1000s
AC CURRENT PROBE FOR OSCILLOSCOPES
Quick Reference Card
Read Safety Information found in the i1000s Users Manual before use.
OPERATION
To make a measurement, follow these steps and refer to the illustration on the back. (Steps are identified in the illustration.)
Connect the i1000s Current Probe to the desired input
1
channel on the oscilloscope.
2
On the Current Probe, select the least sensitive range (1 mV/A).
3
On your oscilloscope, select an appropriate range.
Clamp the probe around the conductor to be measured,
4
and observe the current waveform on your oscilloscope display. For standard oscilloscopes, use a 1:1 probe setting. Not needed for Fluke ScopeMeter 123.
MEASUREMENT CONSIDERATIONS
Carefully move the probe to center the conductor inside
•
the jaws.
Make sure the probe is perpendicular to the conductor.
•
If possible, avoid measurements close to other current­carrying conductors.
•
On the Current Probe, the 1 mV/A range offers the best accuracy and least phase shift. When using the 1 mV/A probe range, use the “volts per division” adjustment on your oscilloscope for best display results.
PN 1575200 June 2000, Rev. 1, 7/05 ©
2000-2005 Fluke Corporation. All rights reserved. Litho in U.S.A.
®
Page 2
4
2
1000
100
10
3
1
CHANNEL A