Fluke i30
AC/DC Current Clamp
The i30 current clamp is based
on Hall effect technology for use
in measurement of both dc and
ac current. The i30 may be used
in conjunction with multimeters,
recorders and other suitable
recording instruments for accurate
non-intrusive current measurement.
Technical Data
Electrical specifications
Specified current range: 30 mA to 30 A DC,
30 mA to 20 A AC rms
Usable current range: 5 mA to 30 A DC,
30 mA to 20 A AC rms
Crest factor: 1.4
Output sensitivity: 100 mV/A
Accuracy (at +25 °C): ± 1 % of reading ± 2 mA
Resolution: ± 1 mA
Load impedance: > 10 k Ohms and ≤ 100 pF
Conductor position sensitivity: ± 1 % relative to
center reading
Frequency range: DC to 20 kHz (- 0.5 dB)
Temperature coefficient: ± 0.01 % of reading/°C
Power supply: 9 V Alkaline, MN1604/PP3,
30 hours, low battery indicator
Working voltage (see Safety Standards section):
300 V ac rms or dc
General specifications
Maximum conductor size: 19 mm (.748 in)
diameter
Output connection: 4 mm (.157 in) safety
connector
Output zero: Manual adjust via thumbwheel
Cable length: 1.5 m (4.91 ft)
Operating temperature range: 0 °C to +50 °C
(-32 °F to 122 °F)
Storage temperature range (with battery
removed): -20 °C to +85 °C (-4 °F to 185 °F)
Operating humidity: 15 % to 85 %
(non-condensing)
Weight: 250 g (.55 lb)
Safety standards
BS EN 61010-1: 2001
BS EN 61010-2-032: 2002
BS EN 61010-031: 2002
300 Vrms, Category III, Pollution Degree 2
Use of the probe on uninsulated conductors is limited to 300 V acrms
or dc and frequencies below 1 kHz.
EMC Standards
EN 61326: 1998 +A1, A2, & A3
Ordering information
i30 AC/DC Current Clamp
Dimensions (HxWxD)
183 mm x 71 mm x 25 mm (7.2 in x 2.8 in x 1 in)
i30 connected to a Fluke 87V Digital Multimeter.
Fluke. Keeping your world
up and running.
Fluke Corporation
PO Box 9090, Everett, WA USA 98206
Fluke Europe B.V.
PO Box 1186, 5602 BD
Eindhoven, The Netherlands
For more information call:
In the U.S.A. (800) 443-5853 or
Fax (425) 446-5116
In Europe/M-East/Africa +31 (0) 40 2 675 200 or
Fax +31 (0) 40 2 675 222
In Canada (800) 36-FLUKE or
Fax (905) 890-6866
From other countries +1 (425) 446-5500 or
Fax +1 (425) 446-5116
Web access: http://www.fluke.com
©2006 Fluke Corporation. All rights reserved.
Specifications subject to change without notice.
Printed in U.S.A. 12/2006 2747126 D-EN-N Rev B