®
i2000flex
2000A Flexible AC Current Probe
Instruction Sheet
Introducing the i2000flex
The i2000flex is a flexible AC current probe for use with
oscilloscopes and multimeters. The supplied Dual Banana to BNC
Adapter can be used for connection to instruments with banana
inputs. The i2000flex is optimized for current measurement on thick
and hard to reach conductors.
Unpacking
The following items should be included in the package:
• i2000flex Current Probe
• Dual Banana to BNC Adapter (model PM9081/001)
• Instruction Sheet (this paper)
Check the contents of the shipping box for completeness. If
something in the box has been damaged or missing, contact your
distributor or the nearest FLUKE sales or service office immediately.
Safety Information
Read First: Safety Information
To ensure safe operation and service of the current
clamp, follow these instructions:
• Read the operating instructions before use and follow all safety
instructions.
• Use the Current Probe only as specified in the operating
instructions; otherwise the probe’s safety features may not
protect you.
• Adhere to local and national safety codes. Individual protective
equipment must be used to prevent shock and arc blast injury
where hazardous live conductors are exposed.
• Before each use, inspect the Current Probe and its latching
system for any damage. Pay particular attention to the
insulation surrounding the flexible measuring head. Look for
cracks or missing portions of the probe housing or output cable
insulation. Also look for loose or weakened components.
• Do not use a probe that is cracked, damaged, or has a
defective cable.
• Never use the probe on a circuit with voltages higher than 600
V CAT III.
− CAT III equipment is designed to protect against transients
in equipment in fixed equipment installations, such as
distribution panels, feeders and short branch circuits, and
lighting systems in large buildings.
• De-energize the installation on which current will be measured
or adopt safe operating procedures during application and
removal of the current probe.
• Use extreme caution when working around bare
bus bars.
conductors or
• Do not use the Current Probe to measure bare conductors
carrying a voltage from 30V up to 600V unless you are wearing
protective clothing suitable for high-voltage work. Contact with
the conductor could result in electric shock. Always use
appropriate
• Use caution when working with voltages above 60 V dc, 30 V ac
rms or 42 V ac peak. Such voltages pose a shock hazard.
• Do not expose the i2000flex unit to water.
equipment for personal protection.
Symbols
Do not apply around or remove from HAZARDOUS
conductors.
LIVE
Product is protected by double insulation.
Risk of Danger. Important information. See
Instruction Sheet.
Risk of Electric Shock.
Conforms to relevant European standards.
Earth ground.
Specifications
SAFETY
i2000flex
(INPUT and OUTPUT)
EMC
ELECTRICAL SPECIFICATIONS
All Electrical Specifications are valid at the following reference
conditions:
• Ambient temperature
• Relative Humidity
• Frequency
• Continuous external field
• Load impedance on BNC
output
• The current may not contain any DC component
• No influence from adjacent currents
• Conductor centered within the probe aperture
• “POWER ON” indicator flashing (green LED)
“BATTERY LOW” indicator not flashing (red LED)
(see Figure 1)
Complies with American
industry standards UL61010B-1
& UL61010B-2-032 &
European standards EN/IEC
61010-1 2nd Edition & EN/IEC
61010-02-032.
Complies with:
EN/IEC 50081-1
EN/IEC 50082-2
23±3°C (73±5.4°F)
0 to 85%
48 to 65 Hz
< 40 A/m
Ω
// < 100 pF
> 1 k
Measuring range 2 to 200A ac 20 to 2000 A ac
Crest factor * up to 1.6 up to 1.6
Maximum nondestructive current:
Output signal 10 mV/A 1 mV/A
Basic accuracy ±1% of full scale ±1% of full scale
Noise < 2 mV rms
Additional errors:
• with temperature
(0 to +70 °C,
32 to 158 °F)
• with position of
conductor in the
probe window
• with external fields
(<40 A/m)
Phase shift
50 to 60 Hz
* This is the maximum permissible ratio between the peak value of
a superimposed transient and the ac rms value.
Bandwidth (-3dB) 1 Hz to 20 kHz
Battery 9V alkaline IEC 6LR61,
Battery life > 100 hr
Indicators (see Figure 1)
Power On
Battery Low
GENERAL
Coupling diameter
Weight 420 g (14.8 oz)
Transducer length 60 cm (31.5 in)
Transducer diameter 14.3 mm (0.563 in)
Minimum bending radius 38 mm (1.5 in)
Cable length from transducer to
box
Cable length from box to BNC
output
Maximum extension cable length 15 m (50 ft)
Weight (without battery) 420 gr (14.8 oz)
Temperature operating:
Transducer
i2000flex unit
Temperature non-operating:
Transducer and i2000flex unit -40 to + 105°C (-40 to +221°F)
Relative Humidity
Operating
Non-operating
Altitude
Operating
Non-operating
100 kA 100 kA
(200 mA rms)
0.02% of reading /°C,
0.036% of reading /°F
±2% of full scale
(bus ≥ 2.5 cm or
1 inch from coupling)
±1% of full scale
(adjacent conductor
≥ 20 cm or 8 inch from
head.)
See Figure 2
< ±0.5°
Current Ranges
200A 2000A
< 2 mV rms
(2A rms)
0.02% of reading /°C,
0.036% of reading /°F
±2% of full scale
(bus ≥ 2,5 cm or
1 inch from coupling)
±1% of full scale
(adjacent conductor
≥ 20 cm or 8 inch
from head.)
See Figure 3
< ±0.5°
550 mAhr
Green LED flashing
Red LED flashing
∅ 22.2 mm (0.874 in)
1.8 m (71 in)
0.5 m (19.7 in)
-20 to +90°C (-4 to +194°F)
0 to + 70°C (32 to +158°F)
0 to 85%, < +35°C (+95°F)
0 to 75%, < +90°C (+194°F)
0 to 85%, < -10°C (+14°F)
0 to 75%, < +105°C (+221°F)
to 3000 m (10000 ft)
to 12000 m (40000 ft)
Current Ranges
Power off
Flashing when a
current range is
selected. *
Flashing when the
battery is too weak
measurement
* Note: When the battery is almost empty, none of the LEDs will
PHASE SHIFT
for correct
results. *
flash.
{
Figure 1. Text on the i2000flex unit
200A Range
2.0
0.0
-2.0
-4.0
-6.0
-8.0
-10.0
-12.0
-14.0
1 10 100 1000 10000 100000
Frequenc y (H z)
Figure 2. Phase Shift versus Frequency (200A range)
2000A Range
2.0
0.0
-2.0
-4.0
-6.0
-8.0
-10.0
-12.0
-14.0
1 10 100 1000 10000 100000
Frequency (Hz)
Figure 3. Phase Shift versus Frequency (2000A range)
February 1998 rev 1, 08/05
© 1998-2005 Fluke Corporation.
All rights reserved.
Printed in the UK.
All product names are trademarks
of their respective companies.
Instrument Compatibility
The i2000flex can be used with any Fluke ScopeMeter test tool,
Power Quality Analyzer, Oscilloscope, or Multimeter that has the
following features:
• BNC input connector. The Dual Banana to BNC Adapter
included in the package, can be used to connect to standard
inputs on multimeters. For the 120 series ScopeMeters and
the Fluke 43 Power Quality Analyzer, use the BB120 Shielded
Banana to BNC Adapter (optional accessory).
• Input accuracy of 1% or better to take full advantage of the
accuracy of the Current Probe.
• Input impedance of greater than or equal to 1 kΩ, and for full
bandwidth and accuracy, a maximum input capacity of 100 pF.
• A passband of more than four times the frequency of the
waveform to be measured.
Using the i2000flex
To use the i2000flex, follow these instructions:
1. Connect the BNC connector of the i2000flex to the desired
input on the measuring instrument. When you are using a
multimeter, use the Dual Banana to BNC Adapter
(PM9081/001) to connect the i2000flex to the input. (See
Figure 5.)
2. On the i2000flex unit, select the least sensitive range (set the
slide switch to position ‘1 mV/A’).
3. Select the corresponding sensitivity (.. mV/A) on your
ScopeMeter test tool or oscilloscope. If you are using a
multimeter, select an appropriate AC voltage range.
Figure 4. Connecting the flexible probe around the conductor.
4. Connect the flexible probe around the conductor (See Figure 4.)
5. Make sure that the arrow marked on the probe coupling points
toward the correct orientation for correct phase display on the
oscilloscope. (See Figure 6.)
6. Keep the probe coupling more than 2.5 cm (1 inch) away from
the conductor.
7. Observe the current value and waveform on the instrument’s
display.
8. If desired, select the lower range on the i2000flex unit and set
the corresponding sensitivity on the measurement instrument.
If the i2000flex is used with a multimeter, the actual current value
can be calculated from the displayed AC voltage value.
Example 1 : Current Probe set to 1 mV/A (2000A range).
Multimeter displays 1.85Vac rms.
Actualcurrent
display value
sensitivityCurrent
Example 2: Current Probe set to 10 mV/A (200A range).
Multimeter displays 750 mVac rms.
Actualcurrent
display value
sensitivityCurrent
=
Pr obe
=
Pr obe 0
1.85V
== =
1 mV / A
750 mV
==
1 mV/A
1850 mV
1 mV / A
75A ac rms
1850A ac rms
Maintenance
• Before each use, assure continued safety by inspecting the
flexible probe and its latching system for any damage. Pay
I
particular attention to the insulation surrounding the flexible
probe. An i2000flex under warranty will be promptly repaired or
replaced (at Fluke's discretion) and returned at no charge.
Cleaning and Storage
• The flexible probe and its latching system require no special
care. Ensure that no foreign body obstructs the latching
mechanism.
• Clean the i2000flex unit with a damp cloth and a mild
detergent. Do not use abrasives, solvents, or alcohol.
• Remove the battery when storing the i2000flex for a long
period of time.
Replacing the Battery
Replace the 9V battery when the ‘BATTERY LOW’ LED on the
i2000flex unit flashes when the power is switched on (a range is
selected.)
Follow the steps 1 to 5 depicted in the following figure to replace the
battery.
• To avoid false readings which could cause injury,
replace battery as soon as the red Low Battery
Indicator flashes.
• Remove the flexible current probe, from any circuit to
be measured.
• Set the unit switch to OFF.
• Disconnect the unit from the measurement
instrument.
The i2000flex unit contains an alkaline battery. Do not
dispose of this battery with other solid waste. Used
batteries should be disposed of by a qualified recycler
or hazardous materials handler. Contact your authorized
FLUKE Service Center for recycling information.
Note:
Figure 5. Measurement Setup
Warning
If the sensitivity setting (mV/A) of the ScopeMeter test
tool or oscilloscope does not correspond with the
setting of the Current Probe, the ScopeMeter test tool
or oscilloscope may display a much lower current
than the actual value. This may result in a false and
misleading reading.
Measurement
at a phase
conductor
Measurement
at a neutral
conductor
Figure 6. Orientation of the flexible probe
ST8550_190
Measurement Considerations
Though the i2000flex is intended for measurements on hard to
reach conductors, observe the following guidelines to approach the
maximum measurement accuracy of ±1% of full scale as far as
possible:
• Center the conductor perpendicularly inside the flexible probe
area. If this is not possible, an additional measurement error of
±2% of full scale can occur.
• Avoid taking measurements close to other current-carrying
conductors if possible. An external field of maximum 40 A/m
can cause an additional measurement error of 1% of full scale.
The probe coupling must be at least 2.5 cm (1 inch) away from the
conductor to which the probe is attached, to meet specified
measurement accuracies.
Also see under “Additional errors” in the section “Specifications”.
Figure 7. Replacing the battery
If your i2000flex does not work
If the i2000flex does not perform properly, use the following steps to
help isolate the problem:
• Inspect the coupling system for any damage. If any foreign
material is present, the coupling system will not close properly
and errors will result.
• Inspect the cables between the flexible probe and the unit and
between the unit and the measurement instrument for any
damage.
• Check if the slide switch on the unit is set to the appropriate
current range, matching with the range (or sensitivity) of the
measurement instrument.
• Check if the “POWER ON” LED on the unit is flashing, and the
“BATTERY LOW” LED is not flashing. Otherwise replace the
battery. Also replace the battery if both LEDs stay off.
• Verify that the function selection on the measurement
instrument is correct.
• If your measurement instrument has a sensitivity setting for a
current probe, select the correct setting (1 mV/A or 10 mV/A).
• If you use a multimeter, preferably select the 2V ac range for
full range measurements with best accuracy results.
LIMITED WARRANTY & LIMITATION OF LIABILITY
This Fluke product will be free from defects in material and
workmanship for one year from the date of purchase. This warranty
does not cover fuses, disposable batteries or damage from
accident, neglect, misuse or abnormal conditions of operation or
handling. Resellers are not authorized to extend any other warranty
on Fluke’s behalf. To obtain service during the warranty period,
send your defective product to the nearest Fluke Authorized Service
Center with a description of the problem.
THIS WARRANTY IS YOUR ONLY REMEDY. NO OTHER
WARRANTIES, SUCH AS FITNESS FOR A PARTICULAR
PURPOSE, ARE EXPRESSED OR IMPLIED. FLUKE IS NOT
LIABLE FOR ANY SPECIAL, INDIRECT, INCIDENTAL OR
CONSEQUENTIAL DAMAGES OR LOSSES, ARISING FROM ANY
CAUSE OR THEORY.
Since some states or countries do not allow the exclusion or
limitation of an implied warranty or of incidental or consequential
damages, this limitation of liability may not apply to you.
Fluke Corporation Fluke Industrial B.V.
P.O. Box 9090 P.O. Box 90
Everett WA 7600 AB Almelo
98206-9090, USA The Netherlands
SERVICE CENTERS
To locate an authorized service center,
visit us on the World Wide Web:
http://www.fluke.com
or call Fluke using any of the phone numbers listed below:
+1-888-993-5853 in U.S.A. and Canada
+31-40-267-5200 in Europe
+1-425-446-5500 from other countries