Fluke i410, i1010 Service Manual

i410/i1010
®
AC/DC Current Clamp
Service Information
WWarning
To avoid electric shock:
Do not service this Current Clamp unless you are qualified to do so.
The service information provided in this document is f or use by qual i ­fied personnel only.
Always remove input signals from the Current Cl amp before opening
the case.
Be careful when working with vol t ages t hat exceed 30V ac rms, 42.4V
pk, or 60V dc.
The Current Clamp is protected throughout by double insul ation. When
servicing the Current Clamp, use only specified replacement parts.
If the Current Clamp is to be Calibrat ed, you must have a replacement
decal. See the Replacement Parts section for further det ai l s.
This Current Clamp contains parts that can be damaged by stat i c dis-
charge. Follow the standard practices for handling static sensitive de­vices.

Introduction

Service Information covers the following service-related areas for the i410/i1010 AC/DC Current Clamps:
Product specifications
Required equipment
Performance test procedure and test values
Preparing for calibration
Calibration procedure
Reassembly procedure
Replacement parts
Usually, service information in this document applies to both models (i410 and i1010). Information ap­plicable to only one model is so noted. For operating information, refer to the Instruction Sheet that came with the Current Clamp.
Caution
PN 642733 December, 1996 1996 Fluke Corporation. All rights reserved. Printed in U.S.A. Product names are trademarks of their respective companies.
i410/i1010
AC/DC Current Clamp

Service Under Warranty

The Current Clamp is warranted to be free from defects in material and workmanship for one year. The warranty does not cover abuse, batteries, or damage caused by using the Current Clamp outside its
rated specifications or operating environment. The complete warranty statement appears in the Current Clamp Instruction Sheet.
For repairs:
In the USA or Canada, call 1-800-323-5700
Outside the USA, contact your nearest Fluke Distributor or Service Center.
Service, repairs, and replacement parts are warranted for 90 days.

Specifications

General specifications are shown in Table 1. Range and accuracy specifications are shown in Table 2.
The accuracy of the Current Clamp is specified for one year after calibration, at 18°C to 28°C (64°F to 82°F), with relative humidity to 90%.
Table 1. General Specifications
Operating Temperature -10 to 50°C (14 to 122°F) Current x Hz Product 240,000 Storage Temperature -20 to 60°C (-4 to 140°F) Output Signal 1 mV per amp dc or ac Relative Humidity 0 to 95% (0 to 30°C)
0 to 75% (30 to 40°C) 0 to 45% (40 to 50°C)
Altitude Operating Non-operating
Battery Type 9V (alkaline) Load Impedance 1 M, 100 pF Battery Life 60 hrs typical (continuous
Safety
Specified Current Range 1 - 400A ac rms *
Usable Current Range 0.5 - 400A 0.5 - 1000A DC Accuracy (zero adjusted, conductor centered) AC Accuracy 3.5% + 0.5A, 45 - 400 Hz, Crest
Bandwidth 3 kHz 10 kHz
*
With a true-rms DMM, the minimum ac current is limited to the low end of the specified mV ac range.
0 - 2000m 0 - 12000m
with alkaline)
s Double Insulation, 600V,
CAT. III
Table 2. Range and Accuracy Specifications
i410 i1010
1 - 400A dc
3.5% + 0.5A (0 - 400A) 2.0% + 0.5A (0-1000A)
Factor 3. (1 - 400A)
Working Voltage 600V CAT. III
maximum at input
Maximum Conductor Size
Temperature Coefficient Size 209 x 78 x 48 mm (3.09 x
Weight 0.5 kg (17.6 oz.)
1 - 600A ac rms * 1 - 1000A dc
2.0% + 0.5A, 45 - 400 Hz, Crest Factor 3.
3.0% + 0.5A, 400 Hz - 2 kHz sine wave. (1 - 600A)
1 ea. 30 mm (1.18 in.) diameter 2 ea. 25 mm (0.98 in.) diameter
+/-(0.05 x accuracy per °C (0 -18°C, 28-50°C)
8.21 x 1.87 in.)
2

Required Equipment

Equipment required for the performance test and calibration procedure is shown in Table 3.
Equipment Recommended Model
Calibrator Fluke 5700A, 5720A, or 5500A Transconductance Amplifier Fluke 5220A Digital Multimeter (DMM) Fluke 45 or equivalent Current Coil Fluke 5500A/Coil (50-Turn) Calibration tool (slotted, as required) Decal (replace to calibrate) PN 642584 (i410) or 642592 (i1010)

Performance Test

To avoid electric shock, do not execute the performance t est s unless the Current Clamp is fully assembled.
Service Information
Required Equipment
Table 3. Required Equipment
WWarning
The Battery LED and Measurement Accuracy tests verify that the Current Clamp is working properly and performs to specification.

Battery LED Test

Complete the following procedure to verify that the Battery LED threshold is set correctly:
1. Turn on the Calibrator and allow it to warm up and stabilize for 30 minutes.
2. Place the Current Clamp face down.
3. Remove the screw securing the battery cover to the back of the Current Clamp.
4. Remove the battery cover.
5. Connect the Calibrator to the Current Clamp battery snap connectors.
6. Set the Calibrator to 6.70V dc. Check that the Current Clamp LED is off.
7. Set the Calibrator to 7.30V dc. Check that the LED is on.
8. Replace the battery cover and securing screw.

Measurement Accuracy Tests

Connect the Current Clamp and test equipment as shown in Figure 1. Then perform the appropriate steps shown in Table 4. Most tests apply to both the i410 and the i1010, with resulting DMM readings listed separately for each model. Some tests apply to only one model; the model not being tested is then
marked with “NA”.
3
i410/i1010
AC/DC Current Clamp
5700A/5720A 5500A
HI
LO
1V PK
MAX
A, -SENSE,
AUX V
20V PK
SCOPE
200V PK
MAX
20V RMS MAX
TRIG OUT
MAX
Calibrator
NORMAL AUX
V, ,
RTD
1000V
RMS MAX
20V PK
MAX
TC
V
1000V
MAX
750V
COM
ALL INPUTS
10A
FUSED
!
100 mA
500mA F 250V
1kV MAX
Conductor
45
DUAL DISPLAY MULTIMETER
45
AUTO
REL
SMF MAX
REMOTE
MINdBHOLD
EXT TRG
V
V
mA
DC AC
mV
mA
mVDCAC
UNCAL
CAL
Hz
Mk
Hz
Mk
A
AUTO
A
FREQ
COMP
HI
REL
REF# LOCAL
HOLD
LO
THRESH ADDR BAUD
POWER
ENABLE
2ND
dB
REF
RATE
MN MX
50-Turn Current Coil
5220A

Preparing for Calibration

Before calibrating the Current Clamp, do the following:
1. Calibration necessitates removal of the decal on the case top. Plan to install a new decal when cali-
bration is complete. See “Replacement Parts” for decal ordering information.
2. Turn on the Calibrator and allow it to warm up and stabilize for 30 minutes.
3. Place the Current Clamp face down.
4. Remove the screw securing the battery cover to the back of the Current Clamp.
5. Remove the battery cover.
6. Connect the DMM to the Current Clamp battery terminals.
Figure 1. Equipment Connections
el001f.eps
7. Press the green power button on and verify that the battery voltage is at least 7V. If the battery volt-
age is too low, replace the battery before proceeding.
8. Carefully remove the Current Clamp decal to uncover a set of five holes in the case top. These holes
allow direct access to calibration potentiometers.
9. Continue with the calibration procedure that follows.
4
Table 4. Performance Test Values
Service Information
Preparing for Calibration
Step Calibrator
Setting
DC Amps Zero Test
10.0 (ZERO knob CW) NA Pos. span 3.0 15.0 3.0 15.0
20.0 (ZERO knob CCW) NA Neg. span 15.0 3.0 15.0 3.0
DC Amps Test
3 0.04A dc 2.0A dc 2.00 mV 1.43 2.57 1.46 2.54 4 0.4A dc 20.0A dc 20.0 mV 18.8 21.2 19.1 20.9 5 -0.4A dc -20.0A dc -20.0 mV -21.2 -18.8 -20.9 -19.1 6 1.2A dc 60.0A dc 60.0 mV 57.4 62.6 58.3 61.7
AC Amps Test
7 0.40A, 45 Hz, 400 Hz 20.0A ac 20.0 mV 18.8 21.2 19.1 20.9 8 0.40A, 2 kHz 20.0A ac 20.0 mV NA NA 18.9 21.1 9 1.0A 45 Hz, 400 Hz 50.0A ac 50.0 mV 47.7 52.3 48.5 51.5 10 1.0A 2 kHz 50.0A ac 50.0 mV NA NA 48.0 52.0 11 2.0A, 45 Hz, 400 Hz 100.0A ac 100.0 mV 96.0 104.0 97.5 102.5
1
50T Coil Current
Measure i410
Lo mV
i410 Hi mV
i1010 Lo mV
i1010 Hi mV
12 2.0A, 1 kHz 100.0A ac 100.0 mV NA NA 96.5 103.5
DC Amps Balance Test
13 4.0A dc 200.0A dc 200.0 mV
DC Amps Test
14 4.0A dc 200.0A dc 200.0 mV 192.5 207.5 195.5 204.5 15 8.0A dc 400.0A dc 0.4000V 0.3855V 0.4145V 0.3915V 0.4085V 16 -8.0A dc -400.0A dc -0.4000V -0.4145V -0.3855V 17 -12.0A dc -600.0A dc -0.6000V NA NA 18 20.0A dc 1000.0A dc 1.0000V NA NA 19 -20.0A dc -1000.0A dc -1.0000V NA NA
AC Amps Test
20 2.4A, 45 Hz, 400 Hz 120.0A ac 120.0 mV 115.3 124.7 117.1 122.9 21 2.4A, 800 Hz 120.0A ac 120.0 mV NA NA 115.9 124.1 22 5.0A, 45 Hz, 400 Hz 250.0A ac 250.0 mV 240.7 259.3 244.5 255.5 23 8.0A, 45 Hz, 400 Hz 400.0A ac 0.4000V 0.3855V 0.4145V NA NA 24 12.0A, 45 Hz, 65 Hz 600.0A ac 0.6000V NA NA 0.5875V 0.6125V
1.0A maximum change from Center of Jaw Reading to Top or Bottom of Jaw Readings.
NA NA
-0.6125V -0.5875V
0.9795V 1.0205V
-1.0205V -0.9795V
Notes:
1. Prior to performing DC Amp tests, adjust ZERO so that the DMM reads 00.0 +/- 00.1 mV without current applied.
5
i410/i1010
AC/DC Current Clamp

Calibration and Adjustment

Always complete the Performance Test prior to Calibration. If Performance Test results are satisfactory, calibration is not necessary. Otherwise, to ensure that the Current Clamp performs to specification, cali­brate the Battery Indicator, DC Zero, and DC Amps and check the AC Amps function on a yearly basis.
When it is needed, be sure to complete the full calibration procedure (Battery Indicator, DC Zero, DC Amps Calibration and AC Amps Check.) Partial completion of this procedure will not ensure that the Current Clamp performs to specification.
Adjustment points are shown in Figure 2.
Decal
AC/DC CURRENT CLAMP
i1010
OUTPUT 1mV/A
600V
Figure 2. Calibration Adjustment Points

Calibrating the Battery Indicator

Use the following procedure to calibrate the low battery indicator:
1. Set the DC voltage generator to 9.0V dc.
CAT III
CAT III
600V 600A
1000A
~
Power On Indicator Power On/Off Button
ZERO Knob
VR4
A
ZERO
VR5 VR1
VR6
VR3
el004f.eps
2. On the Current Clamp, disconnect the battery from the battery clip.
3. Connect the DC voltage generator to the battery snap.
4. Connect the Current Clamp output cable to a DMM. Check that the DMM reads 9.0V dc.
5. On the Current Clamp, press the green power button on and check that the LED is on.
6. Set the DC voltage generator to 7.0V dc. (The LED may be on or off at this point.)
7. Adjust VR4 on the Current Clamp to the point where the LED cycles from on to off.
6

Calibrating DC Zero

Use the following procedure to calibrate the ZERO control:
1. Set the DMM to measure 3V dc to four decimal places.
2. Rotate the ZERO knob to mid-range (ZERO knob bar vertical.)
3. Adjust VR3 so that the DMM reads 0.0000.
4. Turn ZERO (VR2) fully CCW. Check that the DMM reads 0.0050 or higher.
5. Turn ZERO (VR2) full CW. Check that the DMM reads -0.0050 or lower.

Calibrating the DC Amps Function

Use the following procedure to calibrate the DC Amps function:
1. Clamp and center the Current Clamp onto the current coil.
2. Set the DMM to 300 mV dc range at medium speed.
3. Rotate the ZERO knob so that the DMM reads 000.0.
4. Apply 4A dc (200 ampere-turns) to the current coil. Check that the DMM reads 200 mV.
Service Information
Calibration and Adjustment
5. Move the coil conductor within the Current Clamp jaws as you observe reading differences on the
DMM.
6. Adjust VR5 so that multimeter readings vary by no more than 0.2% (199.6 to 200.4 mV).
7. Remove the Current Clamp jaws from the current coil and check that the DMM reads 000.0. If nec-
essary, adjust the ZERO knob to obtain 000.0.
8. Apply 4A dc (200 ampere-turns) to the current coil.
9. Clamp the Current Clamp jaws onto the current coil.
10. With the coil conductor centered in the jaws, adjust VR6 to its mid-range. Then adjust VR1 so that
the DMM reads 199.9 to 200.1. If necessary, adjust VR6 to fine tune this reading.

Checking the AC Amps Function

To check the AC Amps function, do the following:
1. Make sure that the Calibrator, Transconductance Amplifier, DMM, and Current Coil are connected
as shown in Figure 1.
2. Set the DMM to the 300 mV ac range.
3. Clamp the Current Clamp around the Current Coil, taking care to align the conductor with the align-
ment marks on the jaws of the Current Clamp.
4. Apply 5A ac (250 ampere-turns) at 50 or 60 Hz to the Current Coil.
For the i410, the DMM should read between 240.7 and 259.3 mV ac. For the i1010, the DMM should
read between 244.50 and 255.50 mV ac.
5. If you do not obtain readings within the limits of step 4, repeat "Calibrating the DC Amps Function."
7
i410/i1010
AC/DC Current Clamp

Reassembling the Current Clamp

After you have completed the last calibration step and checked the AC Amps function, do the following:
1. Replace the decal over the potentiometer access holes.
2. Replace the battery cover and its securing screw. The Current Clamp is now ready for normal operation.

Replacement Parts

Replacement parts are listed in Table 5 and shown in Figure 3. To obtain replacement parts, call Fluke Corporation at 1-800-526-4731.
Table 5. Replacement Parts
Part Description Part No.
i410 Case Top Decal 642584 i1010 Case Top Decal 642592 Battery Door 642600 Top Case and Zero Knob 642618 PCB Assembly 642626 Case Screw 642634 Jaw Spring 642642 Bottom Case 642645 Brass Cable Clamp 642659 Power On/Off Button (green) 642675 Output Cable Assembly 642683 Jaw Assembly 603602 Poron Ring 642535
8
Zero Knob
Service Information
Replacement Parts
Jaw Assembly
Battery Door
CAT
Decal
i1010
CAT III
OUTPUT 1mV/A
ZERO
AC/DC CURRENT CLAMP
600V
Top Case
600V
600A
1000A
A
Power ON/OFF Button
Poron Ring
Jaw Spring
Case Bottom Screw
(1 of 2)
Figure 3. Replacement Parts
Case Bottom
Brass Cable Clamp
Cable Output Assembly
el002f.eps
9