Fluke 9009 Specifications

Industrial Dual-Block Calibrator
Features
in one unit
Two wells in each block for
simultaneous comparison calibrations
Rugged, lightweight, water-resistant
enclosure
The Hart Scientific Model 9009 Industrial Dual-Block Calibrator lets you calibrate at hot and cold temperatures at the same time. Double your productivity or cut your calibration time in half— either way you look at it, your in-field temperature calibrations just got easier.
The Model 9009 includes two independently controlled temperature blocks. The hot block provides temperatures from 50°C to 350°C, while the cold block covers the range –15°C to 110°C. Each block is controlled by a precision Hart Scientific temperature controller.
Each temperature block includes two wells with removable inserts. You can calibrate four probes at once, or you can calibrate two probes at the same time with an external reference, or you can use the two temperature wells to get quick “zero” and “span” references for transmitter calibrations.
Need portability and durability? The Hart Scientific 9009 is housed in a tough Pelican™ case that is both airtight and watertight. It’s a small package weighing only 10 pounds, yet it fits everything you need, including a power cord and four extra inserts. Inserts are available to accommodate sensors of any size from 1/16" (1.6 mm) to 7/16” (11.1 mm). This rugged system can go anywhere.
Of course the Hart Scientific 9009 also delivers the performance you expect from a Hart Scientific temperature source. The cold block is calibrated to within ±0.2°C with stability of ±0.05°C. The hot block’s display is accurate to ±0.6°C with stability of ±0.05°C. A NIST-traceable calibration is included for each of the two test blocks.
PO Box 1073, 8 Leroy Road, Williston, VT 05495 USA • fax 802-863-1193 • ph 802-863-0085 • toll free 800-235-8367
For use with automated systems, the 9009 comes with an RS-232 connection and our Model 9930 Interface-it software, which allows you to control and monitor temperatures from your PC.
Two blocks in one unit, a total range of –15°C to 350°C, portability, durability, versatility, performance, automation. The Hart Scientific 9009 Industrial Dual-Block Calibrator delivers it all.
Ordering Information
9009-X 3102-0
3102-1 3102-2 3102-3 3102-4 3102-5 3102-6 3102-7 3102-8
Industrial Dual-Block Dry-Well (X = case color. Specify “B” for black or “Y” for yellow.) Includes two 1/4 in (6.4 mm) and two 3/16 in (4.8 mm) inserts.
Insert, Blank Insert, 1/16 in (1.6 mm) Insert, 1/8 in (3.2 mm) Insert, 3/16 in (4.8 mm) Insert, 1/4 in (6.4 mm) Insert, 5/16 in (7.9 mm) Insert, 3/8 in (9.5 mm) Insert, 7/16 in (11.1 mm) Insert, 5/32 in (4 mm)
Specifications
Range 50 °C to 350 °C (122 °F to 662 °F) –15 °C to 110 °C (5 °F to 230 °F)
Accuracy ±0.6 °C ±0.2 °C Stability ±0.05 °C Well-to-Well Uniformity
Display Resolution 0.1 ° Heating Times 10 minutes from 25 °C to 350 °C 15 minutes from 25 °C to 110 °C Cooling Times 30 minutes from 350 °C to 100 °C 16 minutes from 25 °C to –15 °C Stabilization Times 8 minutes Well Depth 4" (102 mm) Removable Inserts Two 6.4 mm (1/4 in) and two 4.8 mm (3/16 in) inserts included; see Ordering Information for
Computer Interface RS-232 included with free Interface-it software Power 115 VAC (±10 %), 3 A, or 230 VAC (±10 %), 2 A, specify, 50/60 Hz, 280 W Size (HxWxD) 178 x 267 x 248 mm (7 x 10.5 x 9.75 in) Weight 4.5 kg (10 lb.) NIST-Traceable
Calibration
Data at 50 °C, 100 °C, 150 °C, 200 °C, 250 °C,
Hot Block Cold Block
(–8 °C [18 °F] with hot block at 350 °C
±0.1 °C
other available inserts
Data at –8 °C, 0 °C, 25 °C, 50 °C, 75 °C,
300 °C, and 350 °C
[662 °F])
100 °C, and 110 °C
PO Box 1073, 8 Leroy Road, Williston, VT 05495 USA • fax 802-863-1193 • ph 802-863-0085 • toll free 800-235-8367
Loading...