Fluke 8922A Service Note

8922A-03
SERVICE NOTE
SUPERSEDES: NONE
8922 GSM Test Sets.
Serial Numbers: 0000A00000 / 9999Z99999
Modulation Distribution board modified to account for Potential Audio Diagnostic failures.
Duplicate Service Notes:
8922A-03 8922B-03 8922E-03 8922F-02 8922G-03 8922H-02 8922M-08 8922P-05 8922R-02 8922S-03 8922X-05 8922Y-02
To Be Performed By: Agilent-Qualified Personnel
Customer
Continued
DATE: September 1999
ADMINISTRATIVE INFORMATION
SERVICE NOTE CLASSIFICATION:
MODIFICATION RECOMMENDED
ACTION CATEGORY:
LOCATION CATEGORY:
AVAILABILITY:PRODUCT'S SUPPORT LIFE AGILENT RESPONSIBLE UNTIL: September 2001
AUTHOR: DO ENTITY: E600 ADDITIONAL INFORMATION:
IMMEDIATELY ON SPECIFIED FAILURE AGREEABLE TIME
CUSTOMER INSTALLABLE ON-SITE SERVICE CENTER
STANDARDS: LABOR 0.5 Hours
SERVICE RETURN INVENTORY: SCRAP
SEE TEXT
USED RETURN PARTS: SCRAP
SEE TEXT
© 1999 AGILENT TECHNOLOGIES PRINTED IN U.S.A.
1 of 2
Page 2 Service Note 8922A-03
Parts Required: P/N Description Quantity
08922-69164 Refurb. Mod. Dist. Bd. 1
Situation:
Audio Diagnostic Test failures seen in the field Audio Analyser 1 - Internal Paths Test
Lower Upper Actual
AC, Gain = 100, Range Det 1.553 1.898 2.271** This test applies a 50mVpp sine wave to the Audio 1 board.Since it is a relatively low level signal
(compared with the Gain =10 (0.5Vpp), Gain =1 (5Vpp)) it is susceptible to noise. The fault was diagnosed as due to a high noise contribution from U402 (TL072). Replacing with 1826-3000 (OP275GS) Low Noise Amp solved the problem. This fault does not affect the performance of the unit.
Solution / Action:
Replace the Modulation Distribution Board (A4) ( 08922-60164) with refurbished board (08922-
69164) The replacement part is an exchange part. Old board must be returned to QMD to receive the refurbished board.