8922A-03
SERVICE NOTE
SUPERSEDES: NONE
8922 GSM Test Sets.
Serial Numbers: 0000A00000 / 9999Z99999
Modulation Distribution board modified to account for Potential Audio
Diagnostic failures.
Duplicate Service Notes:
8922A-03
8922B-03
8922E-03
8922F-02
8922G-03
8922H-02
8922M-08
8922P-05
8922R-02
8922S-03
8922X-05
8922Y-02
To Be Performed By: Agilent-Qualified Personnel
Customer
Continued
DATE: September 1999
ADMINISTRATIVE INFORMATION
SERVICE NOTE CLASSIFICATION:
MODIFICATION RECOMMENDED
ACTION
CATEGORY:
LOCATION
CATEGORY:
AVAILABILITY:PRODUCT'S SUPPORT LIFE AGILENT RESPONSIBLE UNTIL: September 2001
AUTHOR: DO ENTITY: E600 ADDITIONAL INFORMATION:
IMMEDIATELY
ON SPECIFIED FAILURE
AGREEABLE TIME
CUSTOMER INSTALLABLE
ON-SITE
SERVICE CENTER
STANDARDS:
LABOR 0.5 Hours
SERVICE RETURN
INVENTORY: SCRAP
SEE TEXT
USED RETURN
PARTS: SCRAP
SEE TEXT
© 1999 AGILENT TECHNOLOGIES
PRINTED IN U.S.A.
1 of 2

Page 2 Service Note 8922A-03
Parts Required:
P/N Description Quantity
08922-69164 Refurb. Mod. Dist. Bd. 1
Situation:
Audio Diagnostic Test failures seen in the field
Audio Analyser 1 - Internal Paths Test
Lower Upper Actual
AC, Gain = 100, Range Det 1.553 1.898 2.271**
This test applies a 50mVpp sine wave to the Audio 1 board.Since it is a relatively low level signal
(compared with the Gain =10 (0.5Vpp), Gain =1 (5Vpp)) it is susceptible to noise. The fault was
diagnosed as due to a high noise contribution from U402 (TL072). Replacing with 1826-3000
(OP275GS) Low Noise Amp solved the problem. This fault does not affect the performance of
the unit.
Solution / Action:
Replace the Modulation Distribution Board (A4) ( 08922-60164) with refurbished board (08922-
69164) The replacement part is an exchange part. Old board must be returned to QMD to receive
the refurbished board.