Fluke 8922A Service Note

8922A-01
SUPERSEDES: None
DATE: 02 June 1994
8922A GSM RF Test Set
Serial Numbers: 3050A00101 / 3240A00137
Recommended Removal of Hardware
Duplicate Service Notes:
8922A-01 8922B-01 8922E-01 8922G-01
Parts Required: None
Situation:
Instruments with serial numbers noted above may have internal hardware (screw and nut) which is improperly attached to the A34 RTI Bypass Assembly. Refer to Figure 1 for loca­tion of this assembly. The hardware can become separated from the assembly, contact other circuitry and may cause electrical damage.
Solution/Action:
Remove and discard the hardware. Specific locations are indicated in Figure 2. Check for and remove any loose hardware within the instrument. Re-calibration is not required.
1 of 3
AUTHOR: LHL
ENTITY: 1000
© 1994 AGILENT TECHNOLOGIES PRINTED IN U.S.A.
SERVICE NOTE
ADMINISTRATIVE INFORMATION
MODIFICATION RECOMMENDED
SERVICE NOTE CLASSIFICATION:
ACTION CATEGORY:
LOCATION CATEGORY:
CUSTOMER INSTALLABLE ON-SITE SERVICE CENTER
AGILENT RESPONSIBLE UNTIL: 01 June 1999
ADDITIONAL INFORMATION:
IMMEDIATELY ON SPECIFIED FAILURE AGREEABLE TIME
AVAILABILITY:
PRODUCT’S SUPPORT LIFE
RETURN SCRAP SEE TEXT
SERVICE INVENTORY:
RETURN SCRAP SEE TEXT
USED PARTS:
STANDARDS:
Labor 1.0 Hour
Continued
Page 2 Service Note 8922A-01