Fluke 8920A Service Note

8920A-04
SUPERSEDES: None
Continued
DATE: 26 May 1992
8920A RF Communications Test Set
Serial Numbers: 3010A00226 / 3123A00305
Modification to Resolve Low Sensitivity Problem of Antenna Input
Parts Required:
Situation:
A problem has been noted in the A23 Input Section assembly which causes the overpower protection circuitry on the Antenna Input to randomly trigger. This opens a relay in series with the Antenna Input which causes a drop in the measured signal level of 20 dB. Instruments with firmware revision A.05.01 or later will detect this condition and issue a two-tone beep and display the following message on the CRT:
“ANT IN OVERPOWER. REDUCE INPUT POWER AND PUSH MEAS RESET.”
Pressing the MEAS RESET key will reset this condition, but the next random trigger will initiate another overpower detection sequence. NOTE: Earlier versions of firmware will not indicate to the user that an overpower condition has been detected and triggered.
1 of 2
AUTHOR: L.H.L
ENTITY: 1000
© 1992 AGILENT TECHNOLOGIES PRINTED IN U.S.A.
SERVICE NOTE
ADMINISTRATIVE INFORMATION
MODIFICATION RECOMMENDED
SERVICE NOTE CLASSIFICATION:
ACTION CATEGORY:
LOCATION CATEGORY:
CUSTOMER INSTALLABLE ON-SITE SERVICE CENTER
AGILENT RESPONSIBLE UNTIL: 1 July 1997
ADDITIONAL INFORMATION: Utilize standard SMO Exchange Assembly process.
IMMEDIATELY ON SPECIFIED FAILURE AGREEABLE TIME
AVAILABILITY:
PRODUCT’S SUPPORT LIFE
RETURN SCRAP SEE TEXT
SERVICE INVENTORY:
RETURN SCRAP SEE TEXT
USED PARTS:
STANDARDS:
Labor 2.0 Hour
Page 2 Service Note 8920A-04
The following instruments are affected by this problem:
8920A Serial Numbers:
3010A00226 3010A00233-235 3010A00237-246 3119A00247-268 3119A00270-279 3123A00280-282 3123A00285 3123A00287-295 3123A00297 3123A00300 3123A00303-305
Solution/Action:
The A23 Input Assembly has been redesigned to solve this problem. Replace the A23 Input Assembly with an exchange assembly (see Parts Required). Chapter 10 of the 8920A Assembly Level Repair Manual contains removal and installation instructions for the A23 assembly. Re-calibration of the instrument is not required.
Check the date code on the A23 assembly before implementing this Service Note. Only A23 assemblies with date code 3049A, 3118A, 3120A, 3123A and 3124A have this problem. The date code is located on a label attached to the assembly’s aluminum hous­ing. Refer to the figure below to identify the date Code: