8920A-03
SUPERSEDES: None
Continued
DATE: 12 May 1992
8920A RF Communications Test Set (Option 004)
Serial Numbers: 3010A00100 / 3207A01093
Firmware Bug in Option 004 Tone/Digital Signaling ROM
Parts Required: 08920-61058, Firmware Upgrade Kit
Situation:
Two firmware bugs have been discovered in the 8920A Option 004 Tone/Digital Signaling
ROM. These anomalies affect the tone sequential decoder portion of this option. They are
present in 8920As with Option 004 and firmware revision A.05.01 and below. Firmware
revision A.06.01 and above corrects this anomaly.
The first problem is that the decoder can be overly sensitive to noise which results in
repetitive measurements of a given tone. This can result in the decoded symbol being displayed twice. The second problem is that the decoder will not reliably decode multiple
sequences.
Solution/Action:
Display the current firmware revision by viewing the CONFIGURE screen. The CONFIGURE screen is accessed by pressing the SHIFT (blue) key and then the TESTS key. The
current firmware revision is displayed in the upper right hand corner of the CRT.
1 of 2
AUTHOR: LHL
ENTITY: 1000
© 1992 AGILENT TECHNOLOGIES
PRINTED IN U.S.A.
SERVICE NOTE
ADMINISTRATIVE INFORMATION
MODIFICATION RECOMMENDED
SERVICE NOTE CLASSIFICATION:
ACTION
CATEGORY:
LOCATION
CATEGORY:
CUSTOMER INSTALLABLE
ON-SITE
SERVICE CENTER
AGILENT RESPONSIBLE UNTIL: 01 July 1994
ADDITIONAL INFORMATION:
IMMEDIATELY
ON SPECIFIED FAILURE
AGREEABLE TIME
AVAILABILITY:
PRODUCT’S SUPPORT LIFE
RETURN
SCRAP
SEE TEXT
SERVICE
INVENTORY:
RETURN
SCRAP
SEE TEXT
USED
PARTS:
STANDARDS:
Labor 1.0 Hour

Page 2 Service Note 8920A-03
Check the current firmware revision installed in the instrument. If the firmware revision
is A.06.01 or above, do not implement this Service Note. If the firmware revision is
A.05.01 or below, replace the Tone Signaling ROM and Operating System ROMs by
ordering and installing P/N 08920-61058.