Fluke 863, 867B User Manual

Page 1
867B/863
Graphical Multimeter
Users Manual
®
4822 872 00894 November 1997 , Rev. 3, 10/98
© 1997, 1998 Fluke Corporation. All rights reserved. Printed in the Netherlands. All product names are trademarks of their respective companies.
Page 2
Table of Contents
Chapter Title Page
1 A Quick Tour.................................................................................................... 1-1
Your Graphical Multimeter................................................................................. 1-1
About This Manual............................................................................................. 1-8
Combo Mode..................................................................................................... 1-9
Meter Mode ...................................................................................................... 1-9
View Mode......................................................................................................... 1-9
Trend Mode....................................................................................................... 1-9
Auto Diode Test Mode....................................................................................... 1-10
Logic Test Mode................................................................................................ 1-10
Component Test Mode...................................................................................... 1-10
Sleep Mode........................................................................................................ 1-10
2 Making Measurements.................................................................................... 2-1
Introduction........................................................................................................ 2-1
Measuring AC Volts........................................................................................... 2-4
Page 3
867B/863
Users Manual
Measuring DC Volts........................................................................................... 2-6
Measuring DC Millivolts...................................................................................... 2-8
Measuring Ohms, Continuity, and Conductance................................................ 2-9
Testing Diodes and Measuring Capacitance ..................................................... 2-11
Measuring AC and DC Amps............................................................................. 2-13
Measuring AC and DC Milliamps and Microamps.............................................. 2-15
Using Component Test ...................................................................................... 2-17
Testing for Logic Activity.................................................................................... 2-20
Measuring Frequency ........................................................................................ 2-22
3 Some General Descriptions............................................................................ 3-1
Introduction ........................................................................................................ 3-1
Battery Considerations....................................................................................... 3-1
Display Blanking (Sleep Mode).......................................................................... 3-1
Using the............................................................................................................ 3-2
Backlight (Model 867B)...................................................................................... 3-2
Adjusting Contrast.............................................................................................. 3-2
General Features............................................................................................... 3-2
Understanding Display Features........................................................................ 3-7
Measurement Connections................................................................................ 3-10
4 Using the Hardkeys.......................................................................................... 4-1
Introduction ........................................................................................................ 4-1
Using Display Mode Softkeys ............................................................................ 4-1
Using Frequency Softkeys................................................................................. 4-2
Save/Print Softkeys............................................................................................ 4-3
ii
Page 4
Contents
(continued)
Min Max Softkeys.............................................................................................. 4-4
Range Softkeys................................................................................................. 4-5
Touch Hold........................................................................................................ 4-6
5 Using View and Trend Display Modes........................................................... 5-1
Introduction........................................................................................................ 5-1
View Display Mode Basics ................................................................................ 5-1
Using View Mode Softkeys................................................................................ 5-2
Setting Up the Time Base.................................................................................. 5-3
Setting Up the Trigger ....................................................................................... 5-4
Choosing the Acquisition Type.......................................................................... 5-7
Trend Display Mode Basics............................................................................... 5-10
Using Trend Mode Softkeys.............................................................................. 5-10
6 Using Save, Recall, Print, and Set Up............................................................ 6-1
Introduction........................................................................................................ 6-1
Using the Save/Print Softkeys........................................................................... 6-1
Saving Screen and Configuration...................................................................... 6-3
Recalling Screen or Configuration..................................................................... 6-4
Previewing Screen or Configuration.................................................................................. 6-4
Saving the Present Configuration...................................................................... 6-5
Printing............................................................................................................... 6-6
Changing the Configuration............................................................................... 6-8
7 User Maintenance............................................................................................ 7-1
Introduction........................................................................................................ 7-1
Page 5
867B/863
Users Manual
Cleaning............................................................................................................. 7-1
Testing the Fuses............................................................................................... 7-1
Replacing the Batteries...................................................................................... 7-2
Replacing the 440 mA Fuses............................................................................. 7-2
Replacing the 11A (High Energy) Fuse.............................................................. 7-2
Reassembly ....................................................................................................... 7-3
Operational Test................................................................................................. 7-5
Self Test............................................................................................................. 7-5
If Your GMM Does Not Work ............................................................................. 7-8
8 Specifications................................................................................................... 8-1
General Specifications....................................................................................... 8-1
iv
Page 6
List of Tables
Table Title Page
1-1. Introducing Your Graphical Multimeter .................................................................. 1-2
1-2. A Practice Session................................................................................................. 1-6
1-3. Functions and Display Modes................................................................................ 1-11
2-1. Common Combo/Meter Mode Softkeys................................................................. 2-2
2-2. Component Testing (Capacitance)........................................................................ 2-18
3-1. General Features................................................................................................... 3-4
3-2. Upper Status Line.................................................................................................. 3-9
6-1. Set Up Selections .................................................................................................. 6-9
v
Page 7
867B/863
Users Manual
vi
Page 8
List of Figures
Figure Title Page
1-1. Graphical Multimeter Features............................................................................... 1-1
1-2. RS232 and AC Power Connections....................................................................... 1-1
1-3. Test Leads and Measurement Connections.......................................................... 1-4
1-4. Using the Battery Eliminator .................................................................................. 1-5
1-5. Volts and Amps Measurement Map....................................................................... 1-12
1-6. Ohms, Conductance, Diode Test, Capacitance Map............................................. 1-13
1-7. Logic Test, Component Test, and Set Up.............................................................. 1-14
1-8. Hz Min Max, and Range Map ................................................................................ 1-14
2-1. AC Volts Measurements........................................................................................ 2-4
2-2. DC Volts Measurements........................................................................................ 2-6
2-3. Millivolt Measurements .......................................................................................... 2-8
2-4. Ohms, Continuity, and Conductance Measurements............................................ 2-9
2-5. Diode Test and Capacitance Measurements......................................................... 2-11
2-6. Amps Measurements............................................................................................. 2-13
vii
Page 9
867B/863
Users Manual
2-7. Milliamp and Microamp Measurements.................................................................. 2-15
2-8. Component Testing................................................................................................ 2-17
2-9. Component Test Patterns....................................................................................... 2-19
2-10. LOGIC Testing........................................................................................................ 2-20
2-11. Frequency Measurements...................................................................................... 2-22
3-1. General Features.................................................................................................... 3-3
3-2. Input Connections Screen...................................................................................... 3-7
3-3. Display Features..................................................................................................... 3-8
3-4. General Measurement Connections....................................................................... 3-10
3-5. Amps Measurement Connections .......................................................................... 3-10
3-6. Milliamp and Microamp Connections...................................................................... 3-11
3-7. LOGIC Testing Connections................................................................................... 3-11
5-1. Time Base Selection............................................................................................... 5-3
5-2. Trigger Icons........................................................................................................... 5-5
5-3. Setting Up the Trigger ............................................................................................ 5-6
5-4. Viewable Time Divisions (Single Shot and Glitch Capture).................................... 5-9
6-1. Freeze (Print/Save) Softkey Map ........................................................................... 6-2
6-2. PC and Printer Connections................................................................................... 6-7
7-1. Replacing the 11A (High Energy) Fuse.................................................................. 7-4
7-2. Replacing the Batteries .......................................................................................... 7-6
7-3. Replacing the 440 mA Fuse................................................................................... 7-7
viii
Page 10
Safety
mm
In this manual, the word “WARNING” identifies conditions and actions that pose hazard(s) to the user. The word “CAUTION” refers to conditions and/or actions that can damage the instrument. Use of the instrument in a manner not specified may im­pair safety. Read the following safety information carefully before attempting to operate or service the instrument.
Use caution when working above 60V dc or 30V ac rms. Such voltages pose a shock hazard.
When using the probes, keep your fingers away from probe contacts. Keep your fingers behind the finger guards on the probes.
Place test leads in proper input terminals.
Disconnect the live test lead before disconnect-
ing the common test lead.
Avoid working alone.
Disconnect the power and discharge high-
voltage capacitors before testing in Ohms Continuity, Diode, Capacitance, and COMPO­NENT Test.
Inspect the test leads for damaged insulation or exposed metal. Check test lead continuity. Damaged leads should be replaced.
Do not use the GMM if it looks damaged.
Select the proper function and range for your
measurement.
When measuring current, turn the power off to the circuit being tested before connecting the GMM test tool in that circuit.
Check GMM test tool fuses before measuring current transformer secondary or motor winding current. (See Testing the Fuses in the User Maintenance chapter.) An open fuse may allow high voltage buildup, which is potentially haz­ardous.
Use clamp-on probes (current clamps) when measuring currents exceeding 10A.
ix
Page 11
867B/863
Users Manual
WARNING
m
Hazardous voltages are present on the External Trigger input terminal and any attached probe when the GMM is measuring voltages greater than 30V rms or 60V dc.
Replacing Fuses
The following fuses should be used:
F 440 mA, 1000V, Minimum Interrupt Rating 10,000A.
F 11A, 1000V, Minimum Interrupt Rating 17,000A.
Warning
Use of fuses with voltage ratings below 1000V reduces protection. Do not replace the 440 mA 1000V fuse with a 4/10A 600V fuse.
m
$
WARNING
m
CAUTION
s
Equipment protected throughout by DOUBLE INSULATION or REINFORCED INSULA­TION.
mm
Overload Protection
Voltage or Current Input: 1000V Surge Protection: 8 kV Maximum Voltage Isolation to Ground: 1000V
Protection Levels:
(fixed distribution-level installations which can experience peak voltage transients up to 8000V.)
Protection Class: II
risk of electric shock.
see explanation in manual.
per IEC 1010-1; 1000V CAT III
s
x
Page 12
The NiCd battery pack must be fully charged before using the 867B. See Chapter 3.
A
COM
V
mA/
A
EXT TRIG
A
COMPONENT
TEST
SET UP
mV
V
V
OFF
LOGIC
HiZ
FREEZE
DISPLAY
MODE
SAVE
PRINT
MIN
MAX
RANGE
WAKE UP
HZ
TOUCH
HOLD
12345
mA
A
B
A
C
E
D
F
G
Your Graphical Multimeter
Figure 1-1. Graphical Multimeter Features
rita0091.eps
Chapter 1
A Quick Tour
Note
rita0092.eps
Figure 1-2. RS232 and AC Power Connections
1-1
Page 13
867B/863
Users Manual
Table 1-1. Introducing Your Graphical Multimeter
ITEM SYMBOL DESCRIPTION
1-2
A
j
W
B
C
Combo
Meter
View
Trend
1 2 3
TURN THE GMM ON BY ROTATING THE SELECTOR TO ANY FUNCTION
selector to the then the screen sets up for this function, and you are ready to begin. You can toggle the backlight power level (Model 867B only) by pressing W. Also note that the screen goes dark if you do not change any controls for about 20 minutes and the GMM is on battery power; pressing W wakes up the screen (all models.)
THE SCREEN
secondary display of smaller digits, and a graphical window showing either a waveform or an analog NeedleGraph representation of the primary display. In addition, the upper screen line displays status information. The bottom screen line (immediately above the softkey labels) shows the display mode, range, and other information. For example, “Combo” appears on this line if a waveform appears in the graphical window: the GMM is in Combo Mode.
THE SOFTKEYS.
along the bottom of the screen. These labels define the present uses for the five blue keys that are positioned immediately below the screen. Together, the labels and the keys form a set of Softkeys. These changing softkey definitions allow direct access to a wealth of GMM capabilities. Refer to Figures 1-6, 1-7, and 1-8 at the end of this chapter for a roadmap to softkey use.
AC Volts (v)
. In AC Volts, the initial screen presents a primary display of large digits, a
Most control settings (selector change or key press) change the softkey labels
position. The input connections screen appears for a few seconds,
. Turn the rotary
4 5
Page 14
Table 1-1. Introducing Your Graphical Multimeter (cont)
ITEM SYMBOL DESCRIPTION
A Quick Tour
Your Graphical Multimeter
1
D
E
F
G
d F P M r H
V
C
X
A
a
B
RS232
THE HARDKEYS
operations in all functions. Pressing W wakes up the screen from Sleep Mode (all models) and toggles the backlight power level on Model 867B. Although hardkeys can generally be pressed at any time, certain feature combinations are not allowed. The GMM test tool rejects any hardkey press that is not allowed; a long beep sounds. Chapter 4 explains each hardkey in detail.
MEASUREMENT CONNECTIONS.
familiarization: no actual connections are necessary during the Quick Tour. Volts, Ohms, Continuity, Conductance, Diode Test, Capacitance, and Component Test measurements use
for the Red lead. Amps measurements use A for the Red lead. mAµA measurements
V
use
a
functions use views.
LINE POWER OR BATTERY POWER.
power through the Battery Eliminator. New alkaline (AA) cells provide a minimum of 4 hours operation. A fully charged BP7217 NiCd battery pack typically provides a minimum of 8 hours operation. The battery eliminator recharges the NiCd battery pack on Model 867B. Refer to Figure 1-4.
SERIAL PC OR PRINTER CONNECTION.
interface cable to communicate with a PC or printer. Refer to Chapter 6.
. Hardkeys appear immediately below the softkeys. Hardkeys perform the same
Review the input connection scheme now only for
for the Red lead. LOGIC Test measurements use
for the Black lead. Refer to Figure 1-3 for TL70A test lead connection
C
You can power the GMM with batteries or from ac line
Your GMM can use an optional optical serial
X
for the Red lead. All
1-3
Page 15
867B/863
TL70A
Test Leads
Users Manual
1-4
Figure 1-3. Test Leads and Measurement Connections
rita0740.eps
Page 16
A Quick Tour
Europe PN 942602 US PN 942599
JP PN 942610
UK PN 942607AU NZ
PN 944595
Your Graphical Multimeter
1
Figure 1-4. Using the Battery Eliminator
rita0770.eps
1-5
Page 17
867B/863
Users Manual
Table 1-2. A Practice Session
STEP ACTION DESCRIPTION
1-6
1
2
d
J
3
1
CHANGE THE DISPLAY MODE
With the GMM still in AC Volts, press first d, then screen changes to the View Mode, where the entire screen is used for displaying a waveform. You can repeat this procedure to select any of the display modes available in AC Volts (Combo, Meter, View, and Trend.) Refer to Table 1-3 for an overview of display modes available in each function. Examples of all display modes (Combo, Meter, View, Trend, Logic Test, Auto Diode Test, Component Test, and Sleep) follow this table. Along with the flexibility provided by multiple softkey definitions, display modes increase the versatility of your GMM considerably.
Many operations activated in one display mode are retained when you select a new mode. For example, “Rel”, “dB”, Min Max, range, and frequency measurement type are retained.
CHANGE THE FUNCTION
Rotate the selector to DC Volts (E). Observe that the display shifts back to Combo Display Mode. (This is the default mode, as established on the screen you see when you turn the selector to SET UP.)
USE THE SOFTKEYS
With DC Volts and Combo Mode still active, you can use the DC Volts Softkeys. For example, press deactivate “Rel”. Explore all five softkeys in this fashion. Press a key first to activate a selection (label highlighted) and second to disable the selection.
to activate “Rel”: difference readings only are now displayed. Press
1
to change the display mode. The
3
1
again to
Page 18
Table 1-2. A Practice Session (cont)
STEP ACTION DESCRIPTION
A Quick Tour
Your Graphical Multimeter
1
4
5
6
7
r
M
H
F
CHANGE THE MEASUREMENT RANGE
Press r to access the Range Softkeys, then press “Manual” ranging.) Press DC Volts Softkeys.
ENABLE MIN MAX READINGS
Press readings. Press M again to deactivate Min Max.
ENABLE TOUCH HOLD
Press H to begin holding the last valid reading in the display, even if the test leads are withdrawn from the measurement point. h appears in the upper status line. Press H again to stop Touch Hold
CHANGE THE TYPE OF FREQUENCY DISPLAY
Press F to access the Frequency Softkeys. The frequency reading (which was in the secondary display) and DC Volts reading (which was in the primary display) are swapped. The softkeys allow you to select different types of frequency measurements (Hz, Duty Cycle, Pulse Width, or Period). Press Volts Softkeys. Primary and secondary displays swap again, and the change to “Period” frequency measurement continues in the secondary display.
to start displaying the maximum (“Max”), average (“Avg”), and minimum (“Min”)
M
to switch back to “Auto” ranging, then press r to return to the
1
to select “Period”, then press F again to return to the DC
4
to select a higher range (and activate
2
1-7
Page 19
867B/863
Users Manual
Table 1-2. A Practice Session (cont)
STEP ACTION DESCRIPTION
8
9
P
LOOK AT THE SAVE/PRINT SOFTKEYS
Press P to access the Save/Print Softkeys. You can now use the softkeys to exercise various save, recall, and print features. Refer to Chapter 6 for more information. (Waveform save/recall is not available on Model 863.)
TURN THE GMM OFF
o
About This Manual
This Quick Tour has introduced you to all the features necessary to use your GMM. If you are comfortable with this level of information, continue on to Chapter 2 for specific information keyed to each type of measurement. If you want more information about any of the features, refer to Chapters 3 through 8, as follows:
Chapter 3
controls, screen, and input connections.
Chapter 4
hardkeys (
1-8
: General descriptions of the GMM
: Complete instructions for using the
d F P M r H
).
Chapter 5
detail.
Chapter 6
and Configuration Changes using SET UP.
Chapter 7
replacement, etc.)
Chapter 8
Index
: View Mode and Trend Mode, in
: Save, Recall, and Print using
. User Maintenance (battery and fuse
: Specifications
P
Page 20
A Quick Tour
Combo Mode
1
Combo Mode
Combo
Meter View Trend Exit
Combo Mode presents both digital and graphical views of the measurement and can be used in VAC, VDC, mV DC, Amps, and mAµA functions. Combo is the default mode in these functions.
Meter Mode
Combo
Meter is the only mode available in Diode and Capacitance Test.
Meter
View Trend Exit
Meter Mode presents both a digital reading and an analog position indication of the measurement. You can select Meter in VAC, VDC, mV DC, Ohms, Amps, and mAµA functions; you can also set Meter on the SET UP screen as the default mode in these functions.
View Mode
Combo Meter
Trend Mode
Combo Meter View
View
View Mode displays a full-screen waveform. It is available in VAC, VDC, mV DC, Amps, and mAµA functions.
Trend Mode presents a digital reading of the measurement and plots that reading over time. It is available in VAC, VDC, mV DC, Ohms, Amps, and mAµA functions.
Trend Exit
Trend
Exit,
1-9
Page 21
867B/863
Users Manual
Auto Diode Test Mode
Meter
Auto Diode Test presents a DC voltage reading and a symbol representing the state of the diode (forward, reverse, open, short, unknown.)
Logic Test Mode
Logic
Logic Test shows icons representing high and/or low logic state along with frequency and voltage readings of logic activity.
Exit
Exit
Component Test Mode
Comp.
Test
Component Test displays a unique pattern representing the characteristics of the in-circuit component being tested. This signature can resemble a straight line for purely resistive components or an ellipsoid for
frequency-sensitive components.
Sleep Mode
restores the screen.
W
Sleep Mode comes on automatically during battery operation if you do not press a key or turn the selector for approximately 20 minutes.
Exit
1-10
Page 22
Table 1-3. Functions and Display Modes
FUNCTION
A Quick Tour
Sleep Mode
1
Combo
Meter
View
Trend
Auto Diode
Logic
Comp.Test
vEe RDIi
•• • • •
•• • • • • •
•• • • •
•• • • •
Lc
1-11
Page 23
867B/863
ViewCombo Meter Trend Exit
Rel Peak
Hold
dB
150
Rel Peak
Hold
dB
150
Rel Peak
Hold
dB
150 Hi Z
Rel Peak
Hold
dB
150
Rel Peak
Hold
dB
150
DC &
AC rms
AC
DC
rms
Average
Rel Peak
Hold
DC &
AC rms
DC
AC
Rel Peak
Hold
rms
Average
Enable RS232
Time
Longer
Restart
Trend
Time
Shorter
Average Sampled Exit
High
Low
Arm
Exit
Full
Auto
DC &
AC rms
Internal
External Exit
Arm
Exit
Slower ExitFaster
rms
Average
Full
Auto
Full
Auto
DC &
AC rms
AC
DC
DISPLAY
MODE
mA 1
uA 100
mA 1
uA 100
DC
AC
Users Manual
1-12
Figure 1-5. Volts and Amps Measurement Map
rita0330.eps
Page 24
Meter Trend Exit
Rel
nS
Rel
Rel Auto
Diode
Rel
Enable RS232
Time
Longer
Restart
Trend
Time
Shorter
DISPLAY
MODE
Average Sample
Normal
Glitch Exit
High
Low
nS
Auto
Diode
A Quick Tour
Sleep Mode
1
Figure 1-6. Ohms, Conductance, Diode Test, Capacitance Map
rita0340.eps
1-13
Page 25
867B/863
TTL3VCMOS
5V
CMOS
2
Hz
20
Hz
2
kHz
18.75 kHz
200
Hz
Set Up
Self
Test Exit
Next Item
Restore
All
Save
Set Up
Duty
Cycle
Period Exit
Pulse Width
Min Max
Off
Restart Exit
Wave Scale
Exit
Run
Pause
Auto
Manual
Users Manual
Figure 1-7. LOGIC Test, Component Test, Set Up
1-14
Figure 1-8. Hz Min Max, and Range Map
rita0360.eps
rita0350.eps
Page 26
Chapter 2
Making Measurements
Note
Recharge the NiCd battery pack before using the 867B. See Battery Considerations in Chapter 3.
Introduction
This chapter shows overall procedures for making measurements in any of the rotary selector positions. Necessary input connections are shown. Softkeys used with each of the display modes available for the selected function are also shown.
Using the Selector
To begin using the GMM, turn the selector to any of the indicated functions. Softkey labels momentarily identify the display modes available for this function, then proceed to list the softkeys available with this function.
Softkeys
Any selector setting or hardkey press provides you with a unique set of softkey labels along the bottom
of the display. These labels define the present uses for the five blue keys that are positioned immediately below the display.
In Combo and Meter display modes, some labels are used with several different selector positions; see Table 2-1 for descriptions of these common labels. Other labels are unique for the selector position; these labels are defined in this chapter as needed.
View and TrendGraph softkey labels are common for all functions that can use these modes; refer to Chapter 5 for a full discussion.
Hardkeys
Although hardkeys can generally be pressed at any time, certain feature combinations are not allowed. The GMM rejects any hardkey press that is not allowed and tells you so with a long beep.
2-1
Page 27
867B/863
Users Manual
Table 2-1. Common Combo/Meter Mode Softkeys
KEY FUNCTION DESCRIPTION
1 2
3
3
4
2-2
Rel
Peak
Hold
rms
Average
DC &
AC rms
dB
Start relative readings in the primary display (REL is displayed). The relative reference (stored when “Rel” is selected) appears with the softkey.
Peak Hold displays the highest and lowest peak values encountered in the present measurement function. These values replace either the secondary reading in Meter Mode or the graphical display in Combo Mode.
Toggle between true “rms” (“rms” annunciator on in display) and “Average” (average responding, rms indicating - no annunciator) readings in any ac function. The rms/Average selection remains in effect until you change the selector position. The rms/Average settings made in one display mode are automatically made for other display modes in the same selector position. “Average” or “rms” can be set as the standard selection on the SET UP display.
In VDC or mV DC, use the full display to show the numerical “DC” and “AC rms” values. Press
again to return to the original display.
3
Show dBm values in the primary display. The “dB” value shown above this softkey identifies the resistance used in computing dBm; 600Ω is the standard value. You can change this value from the SET UP display. (Rotate the selector to SET UP, change the reference selection, then rotate the selector back to the original ac function.)
Page 28
Table 2-1. Common Combo/Meter Mode Softkeys (cont)
KEY FUNCTION DESCRIPTION
Making Measurements
Introduction
2
5
5
5 4
Full
Auto
Hi Z Enter the high input impedance mode in mV DC. “Hi-Z” appears at the top of the display. Press
AC DC
/
The GMM automatically selects range, trigger level, and time base (in VAC and VDC for an optimum display. If “Manual” is selected for Range in the SET UP screen, “Full Auto” does not come on. Selecting “Manual” from the Range Softkeys deselects “Full Auto”.
again to return to “mV”. Activating “DC & AC rms” automatically cancels “Hi-Z”. You cannot use Peak Hold with HiZ.
Toggle between “DC” and “AC” amps or mAµA measurements. “DC” is the standard configuration.
Activate the beeper for continuity or Manual Diode testing; a beep then sounds each time the voltage drop passes below 0.7V (Continuity) or the logic activity indicator flashes (LOGIC Testing). A short circuit results in a continuous tone. If the reading goes above 30 volts, a possibly dangerous voltage exists on the inputs but is not displayed. Therefore, the safety indicator (l) lights.
2-3
Page 29
867B/863
A COM
V
mA
/A
EXT TRIG
1
4
3
FREEZE
DISPLAY
MODE
SAVE
PRINT
MIN
MAX
RANGE
WAKE UP
HZ
TOUCH
HOLD
12 5
34
V
Red
Circuit
2
Users Manual
Measuring AC Volts
Figure 2-1. AC Volts Measurements
6
Turn the Selector to Volts AC. Display mode softkeys appear momentarily; press a softkey to select a new mode, or wait briefly to accept the presently highlighted selection.
7 8
Connect the test leads Use the AC Volts Softkeys as discussed
as shown.
below.
9
At any time, press (Combo, Meter, View, or Trend) to show the measurement in a different display mode.
rita0550.eps
All hardkeys are available in Volts AC.
d
and one of the softkeys
2-4
Page 30
Using the AC Volts Softkeys
Combo or Meter Mode: refer to Table 2-1.
Rel Peak
Hold
View Mode: refer to Chapter 5.
Time
>
>
Base
Trigger
Trend Mode: refer to Chapter 5.
Enable RS232
Time Longer
rms
Average
Single
Shot
Time Shorter
>
dB
600
Glitch
Capture
Trend Type
>
>
Restart Trend
Full
Auto
Full
Auto
Making Measurements
Measuring AC Volts
2
2-5
Page 31
867B/863
FREEZE
DISPLAY
MODE
SAVE PRINT
MIN
MAX
RANGE
WAKE UP
HZ
TOUCH
HOLD
12 5
34
V
A COM
V
EXT TRIG
Red
Circuit
1
4
3
2
mA/
A
Users Manual
Measuring DC Volts
Figure 2-2. DC Volts Measurements
6
Turn the Selector to Volts DC. Display mode softkeys appear momentarily; press a softkey to select a new mode, or wait briefly to accept the presently highlighted selection.
7 8
Connect the test leads Use the DC Volts Softkeys as discussed
as shown.
below.
9
rita0560.eps
At any time, press (Combo, Meter, View, or Trend) to show the measurement in a different display mode.
All hardkeys are available in Volts DC.
d
and one of the softkeys
2-6
Page 32
Using the DC Volts or mV DC Softkeys
Combo or Meter Mode: refer to Table 2-1.
Rel Peak
Hold
View Mode: refer to Chapter 5.
Time
>
>
Base
Trigger
Trend Mode: refer to Chapter 5.
Enable RS232
Time
Longer
DC &
AC rms
Single
Shot
Time
Shorter
600
>
Trend
Type
dB
Glitch
Capture
>
Full
Auto
>
Restart
Trend
Full
Auto
Making Measurements
Measuring DC Volts
2
2-7
Page 33
867B/863
FREEZE
DISPLAY
MODE
SAVE
PRINT
MIN
MAX
RANGE
WAKE UP
HZ
TOUCH
HOLD
12 5
34
mV
HiZ
A COM
V
EXT TRIG
1
4
3
Circuit
2
mA/
A
Red
Users Manual
Measuring DC Millivolts
Figure 2-3. DC Millivolt Measurements
6
Turn the Selector to mV DC. Display mode softkeys appear momentarily; press a softkey to select a new mode, or wait briefly to accept the presently highlighted selection.
7 8
Connect the test leads Use the mV DC Softkeys as discussed
as shown.
above.
rita0570.eps
9
At any time, press
All hardkeys are available in millivolts DC.
d
and one of the softkeys (Combo, Meter, View, or Trend) to show the measurement in a different display mode.
2-8
Page 34
Making Measurements
A COM
V
EXT TRIG
1
4
3
FREEZE
DISPLAY
MODE
SAVE
PRINT
MIN
MAX
RANGE
WAKE UP
HZ
TOUCH
HOLD
12 5
34
Red
2
mA/
A
Measuring Ohms, Continuity, and Conductance
Measuring Ohms, Continuity, and Conductance
2
Figure 2-4. Ohms, Continuity, and Conductance
6
Turn the Selector to Ohms. Softkeys for Meter and Trend Modes appear momentarily. Wait briefly to accept Meter
4
to select Trend Mode.
d
and one of the softkeys
Mode, or press
7 8
9
Connect the test leads as shown. Use the Ohms (Meter or Trend Mode)
Softkeys as discussed below.
F
is not available in Ohms; all other
hardkeys can be used.
At any time, press (Combo, Meter, View, or Trend) to show the measurement in a different display mode.
rita0580.eps
Measurements
2-9
Page 35
867B/863
Users Manual
Using the Ohms Meter Mode Softkeys
Rel
2
( beeps each time a low resistance continuity measurement is encountered. For example, beginning in the 300 range, the beeper goes on when the resistance passes below 32 and off when the resistance rises above 132. You can select a different range after activating the Continuity beeper; each range uses different beeper on and off points. For a list of these on and off points, see Continuity Specifications in Chapter 8 of this manual.
3
(nS) begins conductance (nanosiemens “nS”) readings in the primary display. Measurement ranges are 300 nS and 3000 nS. (Megohm-only readings continue in the secondary display.) Press
3
again to return to ohms () primary readings (all ranges). Selecting conductance (nS) automatically cancels active Min Max, “Peak Hold”, Touch Hold, Continuity, and “Rel, and resets the Trend plot. Frequency readings are not available in conductance or ohms.
/
nS
/
) enables the Continuity beeper. The GMM
2-10
Page 36
Making Measurements
A COM
V
EXT TRIG
1
4
3
FREEZE
DISPLAY
MODE
SAVE
PRINT
MIN MAX
RANGE
WAKE UP
HZ
TOUCH
HOLD
12 5
34
Red
2
mA/
A
Testing Diodes and Measuring Capacitance
Testing Diodes and Measuring Capacitance
2
Figure 2-5. Diode Test and Capacitance Measurements
6
Turn the Selector to Diode/Capacitance. The GMM assumes Meter Mode.
7 8
Connect the test leads Use the Diode/Capacitance Softkeys as
as shown.
discussed below. “Rel” and “/” (Continuity) are not available in Auto Diode.
rita0590.eps
9
Pressing beep sounds.
H
r
F
and
are not allowed in Diode Test.
is not allowed in Diode Test; a long
r is
allowed in Capacitance.
2-11
Page 37
867B/863
Users Manual
Using the Auto Diode Softkeys
Auto
Diode
4
toggles between AutoDiode and Manual Diode. When selected, Auto Diode is highlighted and one of the following icons is displayed:
N k g u ? (
5 ( &
) is highlighted for Diode Testing.
(short)
(open) (unknown)
Pressing this key toggles to Capacitance (|) and disables Diode Testing.
“Rel” and “/” (Continuity) are not available in Auto Diode.
|
&
Using the Manual Diode Softkeys
Rel
4
selects Manual Diode testing (Auto Diode not
/
highlighted). Selecting Manual Diode initially sets “Rel” and “/” (Continuity) off.
In Manual Diode test, voltage is developed across the components(s) by a test current (approximately 1 mA with the test leads shorted) from the meter. Voltage is read on a 0 to +3.000V range that can measure up to five silicon diode or transistor junctions in series. For a silicon diode, the typical forward voltage is about 0.6V. Voltages greater than
3.00V or open test leads produce an overload (OL) reading. If the digital reading is the same in both directions, the diode junction is probably shorted. If the display reads OL in both directions, the diode junction is probably open. To protect sensitive devices the open test lead voltage will not exceed
3.1V. Negative inputs (from an external power source) are not suppressed.
Auto
Diode
|
&
2-12
Page 38
Using the Capacitance Softkeys
A COM
V
EXT TRIG
Red
1
4
3
FREEZE
DISPLAY
MODE
SAVE
PRINT
MIN
MAX
RANGE
WAKE UP
HZ
TOUCH
HOLD
12 5
34
A
Circuit
2
mA/
A
Making Measurements
Measuring AC and DC Amps
2
Rel
5 (|&
) toggles between Capacitance
Measurement and Diode Testing.
Measuring AC and DC Amps
6
7 8 9
At any time, press (Combo, Meter, View, or Trend) to show the measurement in a different display mode.
Turn the Selector to Amps. Display mode softkeys appear momentarily; press a softkey to select a new mode, or wait briefly to accept the presently highlighted selection.
Connect the test leads Use the Amps Softkeys as discussed below.
as shown.
All hardkeys are available in Amps.
d
and one of the softkeys
|
&
rita0600.eps
Figure 2-6. Amps Measurements
2-13
Page 39
867B/863
Users Manual
Using the Amps Softkeys
AC Amps (Combo/Meter Mode): refer to Table 2-1.
Rel Peak
Hold
DC Amps (Combo/Meter Mode): refer to Table 2-1.
Rel Peak
Hold
View Mode (AC or DC Amps): refer to Chapter 5.
Time
>
>
Base
Trigger
Trend Mode (AC or DC Amps): refer to Chapter 5.
Enable RS232
Time
Longer
rms
AveragedB600
DC &
AC rms
Single
Shot
Time
Shorter
>
Trend
dB
600
Glitch Capture
Type
>
>
AC DC
AC DC
Full
Auto
Restart
Trend
2-14
Page 40
Making Measurements
A COM
V
EXT TRIG
Red
1
4
3
FREEZE
DISPLAY
MODE
SAVE
PRINT
MIN MAX
RANGE
WAKE UP
HZ
TOUCH
HOLD
12 5
34
Circuit
2
mA
A
mA/
A
Measuring AC and DC Milliamps and Microamps
Measuring AC and DC Milliamps and Microamps
2
Figure 2-7. Milliamp and Microamp Measurements
6
Turn the Selector to mAµA. Display mode softkeys appear momentarily; press a softkey to select a new mode, or wait briefly to accept the presently highlighted selection.
7 8
Connect the test leads Use the mAµA Softkeys as discussed below.
9
All hardkeys are available in mAµA.
At any time, press (Combo, Meter, View, or Trend) to show the measurement in a different display mode.
d
and one of the softkeys
rita0610.eps
as shown.
2-15
Page 41
867B/863
Users Manual
Using the mAµA Softkeys
AC mAµA (Combo/Meter Mode): refer to Table 2-1.
Rel Peak
Hold
DC mAµA (Combo/Meter Mode): refer to Table 2-1.
Rel Peak
Hold
rms
Average
DC &
AC rms
mA 1
µA 100
mA 1
µA 100
AC DC
AC DC
4
(mA 1; µA 100) In Combo or Meter Mode, this softkey toggles between “mA” and “µA” measurements. (“mA” is the standard configuration.) Each toggle cancels Min Max, Touch Hold, “Peak Hold”, and Rel and resets the Trend plot. “1” and “100” represent the shunt resistances used for mA and µA, respectively.
NOTE
If measuring <3 mA, using 3000.0 µA range will give you better resolution.
View Mode (AC or DC mAµA): refer to Chapter 5.
Time
Base
>
>
Trigger
Single
Shot
>
Glitch Capture
>
Full
Auto
Trend Mode (AC or DC mAµA): refer to Chapter 5.
Enable RS232
Time
Longer
Time
Shorter
Trend
Type
>
Restart
Trend
2-16
The GMM autoranges in “mA” or “µA” (two ranges are available for each selection) but does not autorange between “mA” and “µA”. If manual range is selected, “mA” assumes the 30 mA range, and “µA” assumes the 300 µA range. The µA ranges are not available on Model 863.
F
swaps the mA or µA in the primary display with the frequency reading in the secondary display. Subsequent changing between mA and µA or DC and AC swaps the primary and secondary displays back.
Page 42
Making Measurements
FREEZE
DISPLAY
MODE
SAVE
PRINT
MIN MAX
RANGE
WAKE UP
HZ
TOUCH
HOLD
12 5
34
A COM
V
EXT TRIG
Red
mA/
A
1
4
3
2
COMPONENT
TEST
Using Component Test
Using Component Test
2
Figure 2-8. Component Test
rita0630.eps
6 7 8
Turn the Selector to COMPONENT TEST. Connect the test leads
as shown.
Use the COMPONENT TEST Softkeys as discussed below.
9
Since COMPONENT TEST uses its own display mode, only W,
P
, and
operate normally.
Caution
To avoid damage to the circuit, Component Test should be used only on unpowered circuits.
H
2-17
Page 43
867B/863
Users Manual
Using COMPONENT TEST Softkeys
2
Hz
Use COMPONENT TEST to measure the characteristics of a passive component (in or out of circuit, no power applied). When you use the General Measurement Connection, the GMM provides an ac stimulus signal to the component and plots the voltage drop against the current measurement. The resulting pattern provides information about passive circuit conditions.
Generally, resistors exhibit some form of straight or angled line. Diodes or transistors appear as straight lines with right-angle ends. Frequency-sensitive components (inductors, capacitors) portray some form of an ellipsoid pattern based on the current­voltage phase shift dictated by the stimulus frequency and the inductance or capacitance. Combinations of these components exhibit hybrid patterns. Although the number of such COMPONENT TEST signatures can be limitless, you will become familiar with a few characteristic signature patterns that will greatly aid in component
20
Hz
200
Hz
2
kHz
18.75 kHz
and circuit diagnosis. Some sample signature patterns are shown on the following page.
Press one of the softkeys to select the frequency of the stimulus signal. For components that do not change impedance with frequency (resistors, diodes, and transistors), the frequency selection is not critical for a meaningful display. For components that do change impedance with frequency (inductors and capacitors), an appropriate COMPONENT TEST frequency selection is necessary. Inductance greater than 8.5 mH can be displayed. Capacitance can be displayed according to the guidelines in Table 2-2.
Table 2-2. Component Test (Capacitance)
Frequency Capacitance
2 Hz 0.72 µF to 72 µF 20 Hz 0.072 µF to 7.2 µF 200 Hz 7200 pF to 0.72 µF 2 kHz 720 pF to 0.072 µF
18.75 kHz 77 pF to 7700 pF
2-18
Page 44
Making Measurements
Using Component Test
2
Figure 2-9. Component Test Patterns
rita0800.eps
2-19
Page 45
867B/863
A COM
V
EXT TRIG
WAKE UP
LOGIC
1
FREEZE
DISPLAY
MODE
SAVE
PRINT
MIN MAX
RANGE
HZ
TOUCH
HOLD
12 5
34
4
3
2
mA/
A
Red
Circuit
Users Manual
Testing for Logic Activity
(not available on Model 863)
2-20
Figure 2-10. LOGIC Testing
rita0620.eps
6 7 8 9
Turn the Selector to LOGIC. Connect the test leads
as shown.
Use the LOGIC Testing Softkeys as shown. The following hardkeys can be used
:
d P M
For the logic activity indicator, [ represents a logic high, and ] represents a logic low. A horizontal bar represents an indeterminate condition (no high or low detected since the previous reading.) The logic indicator flashes at a 10 Hz rate for activity faster than 10 Hz or at the actual rate if activity is slower than 10 Hz.
The actual logic frequency is shown in the primary display. Frequencies to 10 MHz and higher can be measured. DC Volts levels are shown in the secondary display.
Page 46
Using the Logic Test Softkeys
TTL
3 V
CMOS
5 V
CMOS
/
Making Measurements
Testing for Logic Activity
2
1, 2
logic activity indicator and frequency reading. “5V CMOS” is the standard configuration. A new selection cancels the old selection. See Chapter 8 for threshold values.
4 (/
to indicate logic activity. The beeper is off in the standard configuration. If the reading goes above 30 volts, the safety indicator (l) lights.
) activates the beeper; a beep then sounds
, and
3
select the trigger point for the
2-21
Page 47
867B/863
A COM
V
EXT TRIG
1
4
3
2
FREEZE
DISPLAY
MODE
SAVE PRINT
MIN
MAX
RANGE
WAKE UP
HZ
TOUCH
HOLD
12 5
34
A
mV
V
V
HiZ
mA/
A
mA
A
Users Manual
Measuring Frequency
You can measure low frequency signals (2 Hz to 2 MHz) with the procedure shown below. For higher frequency signals (to 10 MHz and above), use the Logic Test function.
6
Turn the Selector to any of the indicated positions. (Frequency can be measured in VAC, VDC, mV DC, A, and mAµA.)
7
Connect the test leads as required for the selected measurement position.
8 9
The frequency counter trigger levels are set to predetermined levels while the GMM is in the single trigger mode. In dual trigger mode, the frequency counter trigger levels are set initially (full auto) based
Press
F
.
Use the Frequency Softkeys as discussed below.
on a percentage of the input signal. You can change both levels with the Trigger Softkeys in the view
Figure 2-11. Frequency Measurements
rita0640.eps
mode. Manually setting these levels may cause the frequency counter to be unstable or not count at all.
2-22
Page 48
Making Measurements
Measuring Frequency
2
Either press the Full Auto softkey or manually adjust the trigger levels in view mode to obtain a stable frequency reading.
Using the Frequency Softkeys
x y
1
selects low side for frequency, period, duty cycle, and pulse-width measurements. Positive is the standard selection.
2
(Duty Cycle) selects duty cycle measurements for the pulse (negative going or positive going) selected with
3
(Pulse Width) selects pulse-width measurements for the pulse (negative going or positive going) selected with
4
(Period) selects period measurements of the full signal (negative going or positive going).
5
(Exit) returns to the Display Mode Softkeys.
Duty
Cycle
1
Pulse
Width Period Exit
x
or high side
.
1
.
y
pulse
Frequency Operation
The frequency measurement type (Hz, duty cycle, pulse width, or period) normally appears in the secondary display. secondary display readings and accesses the Frequency Softkeys. You can then select a new frequency type.
F
Press function to the primary display and move the frequency reading back to the secondary display.
F
Press secondary readings.
at any time to swap primary and
F
swaps primary and
a second time to restore the selector
2-23
Page 49
867B/863
Users Manual
Frequency Considerations
F interacts with other features in several ways. Generally, features that are active in the selector function are retained when you press F. For example, active features in VAC, such as “Peak Hold”, “rms” or “Average”, or “dB” would stay activated.
If you press any other key while the Frequency Softkeys are active, the Frequency Softkeys are deactivated and the frequency type is retained.
F cancels Min Max, Touch Hold, “Rel”, and the Continuity functions and resets the plot in Trend Display Mode.
Turning the selector to another position resets the frequency measurement display to “Hz” in the secondary display. Duty cycle, pulse width, and period selections are not retained.
If you press F while the selector is set to Ohms, Conductance, Diode Test, COMPONENT TEST, LOGIC, or SET UP, the GMM emits a long beep and ignores the key press.
The frequency measurement type (Hz, duty cycle, pulse width, period) does not change when you change display mode; the type remains in the primary or secondary display as last selected.
In Amps functions, moving between “AC” and “DC” or “mA” and “µA” swaps the frequency reading back to the secondary display.
Pressing F cancels Touch Hold. While F is active, you can press H to activate Touch Hold, which then works normally with the frequency type shown in the primary display.
Pressing F cancels Min Max. Pressing M when F is active accesses the Min Max Softkeys but retains frequency as the primary function.
If you press F when P is active, the GMM beeps and ignores the key press. While F is active, pressing P accesses the Print/Save Softkeys, retaining the frequency primary display selection.
2-24
Page 50
Chapter 3
Some General Descriptions
Introduction
This chapter adds some details about GMM operation and measurement connections.
Battery Considerations
New alkaline (AA) cells provide a minimum of 4 hours operation. A fully charged NiCd battery pack (BP7217) typically provides 8 hours operation. Connect the battery eliminator to recharge the NiCd battery pack for the 867B. Fully charge the battery pack before using the 867B. External charging (use Battery Charger BC-7210 - not provided) must be used to charge a NiCd battery pack for Model 863. Refer to Chapter 7 (User Maintenance). Model 867B uses an internal two-state charger. The initial charge state is at approximately 170 mA (±30 mA) to allow for full overnight charging (16 hours minimum). The
second charge state is at approximately 40 mA (±15 mA) to allow for battery charge maintenance without full charging from a discharged condition.
Display Blanking (Sleep Mode)
During battery operation the display will blank after about 20 minutes if no controls have been changed. Press the wake up key once to restore the display.
If the battery eliminator is connected, these battery saving features are not used. If a time-dependent measurement (Min Max, Trend, etc.) is in progress during battery operation, the backlight will go off after 20 minutes, but the Model 863 display will not blank until a low battery condition is reached. When the display blanks, no measurements are taken.
3-1
Page 51
867B/863
Users Manual
Using the Backlight (Model 867B)
If the display is already on, pressing W toggles the backlight power level.
NOTE
If the display is blank and cannot be recovered, contrast may be off the scale. Restore contrast to mid-scale with the following shortcut: turn the selector to SET UP and avoid touching any key until the display reappears (about 15 seconds).
1
Press selector to the desired function.
to store this setting, then turn the
Adjusting Contrast
Adjust contrast as follows:
1. Rotate the selector to SET UP.
2. With “LCD Contrast” highlighted, press 2 or 3
to change the contrast. This setting becomes effective immediately.
3. If you want to make this setting part of the power-up defaults, press 5. The GMM then assumes this contrast setting each time you turn it on.
4. Rotate the selector to the desired measurement or test function.
When the GMM is left idle in Setup mode for 12 seconds, the contrast will return to the LCD’s default setting.
General Features
Refer to Figure 3-1 and Table 3-1 for an overall description of the controls, displays, and connections.
3-2
Page 52
Some General Descriptions
A
COM
V
mA /
A
EXT TRIG
COMPONENT
TEST
CAT
LOGIC
SEE WARNINGS ON BACK
A
COMPONENT
TEST
SET UP
mV
V
V
OFF
LOGIC
HiZ
10A MAX FUSED
1000V MAX
1000V MAX
ALL INPUTS
1000V MAX
FREEZE
DISPLAY
MODE
SAVE
PRINT
MIN MAX
RANGE
WAKE UP
HZ
TOUCH HOLD
12345
320mA MAX FUSED
12
11
10
9
2
3
4
5
7
1
6
13
14
15
8
mA
A
General Features
3
Figure 3-1. General Features
3-3
Page 53
867B/863
Users Manual
Table 3-1. General Features
Number Example Name Description
1 Combo
Meter
View
Trend
2 Rel Softkey
3
4
5
1
r
M
3-4
Display
Labels
Softkeys
Range Key
Min Max Key
Combo, Meter, View, and Trend are the standard measurement display modes for most functions. Auto Diode Test, LOGIC Test, and COMPONENT Test use unique display modes. Refer to Figure 3-3 for descriptions of display features.
Labels define present uses for associated softkeys. A highlighted label means that item is presently selected. Labels change with rotary selector positions and key presses. If a small an additional set of softkey labels (a submenu) by pressing this softkey.
The five blue softkeys serve different functions, depending on the rotary selector position and other keys in effect.
Press at any time to change the range of a meter reading or the amplitude of a waveform. Select “Auto” (the standard selection) for an optimal display or “Manual” for selection of a specific range. Present range information is shown in the lower right corner of the display. Refer to Range Softkeys in Chapter 4.
Press once to enable a display showing minimum, average, and maximum values. Refer to Min Max Softkeys in Chapter 4.
symbol appears with a softkey label, you can access
>
Page 54
Table 3-1. General Features (cont)
T
Number Example Name Description
Some General Descriptions
General Features
3
6
7
8
9
10
H
V
C
A
a
P
W
Touch Hold Key
Rotary Selector
Input Connections
Save/Print Key
Wake Up Key
Press once to enable Touch Hold. The GMM test tool beeps and automatically updates the display each time a change in the reading is detected. If the input connections are withdrawn, the last reading (or waveform) is held in the display. Toggles Touch Hold on and off.
Rotate the selector to any of the measurement or test positions. The input connections screen appears briefly, when first turned on or when selecting a function requiring an input connection change, then softkey labels specific to the selector position appear.
Use conductance, diode test, capacitance, COMPONENT TEST). Use A (or mAµA) and COM for amps measurements. Use EXT TRIG and COM for LOGIC Test or as an external trigger for waveform display.
Freeze the present reading (reverse contrast). You can then save or recall measurement values and configurations or print measurement information. In Model 867B, you can also save or recall waveforms and configurations. Refer to Chapter 6.
Press to restore a blanked display; the display blanks after 20 minutes if no controls are changed. The display will not blank if the battery eliminator is connected. For Model 867B, this key toggles the backlight power level when the GMM test tool is already awake.
and COM for most measurements and tests (volts, ohms,
V
3-5
Page 55
867B/863
Users Manual
Table 3-1. General Features (cont)
Number Example Name Description
3-6
11
12
13
14
15
F
d
B
Frequency Key
Display Mode Key
RS-232 Optical Port
Battery Eliminator Connection
Calibration Seal
Swaps primary and secondary displays. The frequency measurement type (usually Hz) now occupies the primary display and the Frequency Softkeys are active. You can now select a different frequency type with the softkeys. Press F again to swap the primary and secondary displays back, retaining the most recent frequency type.
Press to access the Display Mode Softkeys at any time. For volts and amps functions, you can now select Combo, Meter, View, or Trend. Ohms offers Meter or Trend. Other selector functions offer only Meter, while LOGIC, COMPONENT Test, and SET UP use their own unique display modes.
Connect the RS-232 serial interface cable here for communications with a PC or printer.
Connect the battery eliminator for ac operation. A NiCd battery pack installed in Model 867B charges whenever the battery eliminator is connected.
Do not open this protective seal. Refer to the Service Manual.
Page 56
Understanding Display Features
12 5
34
ViewMeterCombo Trend Exit
Figure 3-3 illustrates typical display features. You will encounter Primary and Secondary Displays and the Graphical Window in Combo, Meter, and Trend Display Modes. Other display features will appear in all display modes.
Input Connections Screen
When you select a new function, the Input Connections Screen (Figure 3-2) appears momentarily to identify measurement connections to use, and displays modes that are available. (You can press 3-second pause to proceed immediately to the function display.) See Table 3-1 under Input Connections.
“Exit” during this
5
Some General Descriptions
Understanding Display Features
Figure 3-2. Input Connections Screen
3
rita0601.eps
Primary Display
Large digits represent the measurement or test based on the selector position. This display always appears in Meter, Combo, and Trend modes. The frequency reading (usually Hz) can be swapped into the primary display. When selected, “Rel” or “dB” automatically occupy the primary display.
Secondary Display
Smaller digits show the type of frequency reading (usually Hz) selected with the Frequency Softkeys. The primary reading can also be swapped into this display.
Graphical Window
This window shows a waveform in Combo Mode, an analog display in Meter Mode, or a plot in TrendGraph Mode.
3-7
Page 57
867B/863
Window
Display
Secondary Display
Primary Display
Status Line (Lower)
Softkey Labels
Status Line (Upper)
Users Manual
Softkey Labels
The present uses for the five softkeys appear immediately below the display.
Upper Status Line
The top display line identifies active operations as shown in Table 3-2.
Lower Status Line
The bottom display line (immediately above the softkey labels) identifies the display mode, function, and range.
safety indicator also appears here if dangerous
The
l
voltage is detected at the input terminals when the actual reading is not displayed.
Message Boxes
Message boxes appear in the lower display when the GMM test tool is performing an operation that you may not want to interrupt (“Saving” or “Printing”).
Overload Indication (OL)
“OL” or “-OL” appears in the primary and secondary displays when an input has exceeded the present range (in Manual) or has exceeded the highest range available in this function (in Auto). A clipped waveform can appear in Combo or View mode (DC Volts, mV DC, DC Amps, DC mAµA functions only) if a large ac signal is riding on an overloaded dc signal.
Figure 3-3. Display Features
rita0390.eps
3-8
Page 58
Table 3-2. Display Status Information
SYMBOL DESCRIPTION
Some General Descriptions
Understanding Display Features
3
h
REL “Rel” (relative reading) is active
Min Max Min Max is active. This is the only
RS232 The serial interface port is enabled.
Hi-Z Hi input impedance selected.
b B
Uncal GMM not calibrated. Contact a Fluke
Internal Internal trigger (or External.)
l
Touch Hold is active.
(reference minus actual reading.)
indication if Min Max is running in the background.
Battery is low. Line power operation.
Service Center.
Dangerous voltage is present at the input terminals when the actual reading is not displayed.
3-9
Page 59
867B/863
A COM
V
EXT TRIG
Red
Circuit
mA/
A
A COM
V
EXT TRIG
Red
Circuit
mA/
A
Users Manual
Measurement Connections
Warning
m
Never attempt a voltage measurement if a test lead is in the amp (A) or milliamp microamp (mAµA) input terminal. You might be injured or the GMM TEST TOOL might be damaged.
General Measurement Connections
Volts, Ohms, Continuity, Conductance, Diode Test and Capacitance connections are illustrated in Figure 3-4.
rita0384.eps
Figure 3-4. General Measurement Connections
Amps and mAµA Measurement Connections
Amps DC and AC and mAµA connections are illustrated in Figures 3-5 and 3-6.
rita0382.eps
Figure 3-5. Amps Measurement Connections
3-10
Page 60
Some General Descriptions
A COM
V
EXT TRIG
Red
Circuit
mA/
A
A COM
V
EXT TRIG
Red
Circuit
mA/
A
Measurement Connections
3
Figure 3-6. Milliamp and Microamp Connections.
Logic Test Connections
LOGIC Test connections are illustrated in Figure 3-7.
rita0381.eps
rita0383.eps
Figure 3-7. LOGIC Testing Connections
Frequency Measurement Connections
You can measure frequency from 2 Hz to 2 MHz in AC Volts, DC Volts, mV DC, Amps, and mAµA functions. For higher frequency readings (to > 10 MHz), use the Logic Test function.
3-11
Page 61
867B/863
Users Manual
3-12
Page 62
Chapter 4
Using the Hardkeys
Introduction
Hardkeys can be used in many function and display mode combinations. Although hardkeys are mentioned throughout this manual, this chapter provides a single source of detailed descriptions for using hardkeys and their accompanying softkeys.
Using Display Mode Softkeys
Press d at any time. The Display Mode Softkeys appear, showing the modes available with the present measurement function. Modes available with VAC, VDC, mV DC, A, and mAµA are shown.
Combo Meter View Trend Exit
The GMM test tool now highlights the present selection and allows you to do the following:
Change display mode. Softkeys for the new display mode now appear.
Return without changing the display mode. Press d again, or press the softkey for the selected display mode, or press 5 to exit.
Combo (the default mode set at the factory) appears automatically when the selector is set to VAC, VDC, mV DC, Amps, or mAµA. You can also select Meter as the default for these functions on the SET UP screen. Meter is the only mode available for Diode and Capacitance Test. Meter and Trend are the
4-1
Page 63
867B/863
Users Manual
choices for Ohms and Conductance. LOGIC TEST, COMPONENT TEST, and SET UP each use a single, unique mode.
The Display Mode Softkeys also appear briefly when you rotate the selector to a new function. If you make no mode selection within 2 seconds, the softkeys shift automatically to the appropriate function labels.
Using Frequency Softkeys
Softkeys accessed by pressing F are:
x y
1 selects the low side x or high side y
pulse for frequency, period, duty cycle, and pulse width measurements. Positive is the standard selection.
2 (Duty Cycle) selects duty cycle measurements for the pulse (negative going or positive going) selected with 1.
3 (Pulse Width) selects pulse-width measurements for the pulse (negative going or positive going) selected with 1.
Duty
Cycle
Pulse Width Period Exit
4 (Period) selects period measurements of the full signal (negative going or positive going).
5 (Exit) returns to the Display Mode Softkeys.
Frequency Operation
The frequency measurement type (Hz, duty cycle, pulse width, or period) normally appears in the secondary display. F swaps primary and secondary displays and accesses the Frequency Softkeys, which you can use to select a new frequency type.
Press F a second time to swap primary and secondary displays back. The secondary reading then continues to display the frequency measurement type selected with the Frequency Softkeys. The Frequency Softkeys remain active, and you can change the frequency type shown in the secondary display while these keys remain displayed.
Press F at any time to swap primary and secondary readings.
4-2
Page 64
Using the Hardkeys
Save/Print Softkeys
4
Save/Print Softkeys
Softkeys accessed by pressing
Save Screen
1
(Save Screen) accesses the Save Screen softkeys, which allow you to save the present display to a memory location. (Model 867B only.)
2
(Recall) accesses the Recall softkeys, which allow you to display the contents of a memory location.
3
(Print) prints the display. “Freeze” changes to “Printing”, then reverts to “Freeze” when printing is complete. You can stop printing by pressing (Stop Print) or ignored.
4
(Save Config.) accesses the Save Config softkeys.
5
(Exit) returns to the original display. You can also press display. You must press one of these two keys to
>
Recall Print
5
P
>
(Exit); all other key presses are
again to return to the original
P
are:
Save
>
Config. Exit
3
return; other keys do not operate during the Freeze display.
P
Press updates. Display contrast is reversed, the “Freeze” message appears, and the Save/Print Softkeys are accessed. You can now Print display information or Save and Recall display and configuration information.
“Enable RS232” in TrendGraph Mode is canceled. Touch Hold, Min Max, Trend, and “Peak Hold” continue to update in the background (beeps continue).
Refer to Chapter 6 for a full description of saving and recalling operations available with Freeze.
to stop waveform and numerical value
4-3
Page 65
867B/863
Users Manual
Min Max Softkeys
Softkeys accessed by pressing M are:
Min Max Off Restart
Minimum, maximum, and average measurements are displayed (with timer information) in the graphical display (Combo and Trend) or meter display (Meter).
In the View mode, Min Max is enabled and the MIN ­MAX icon appears in the upper status line. However, Min Max measurement readings and timer information cannot be viewed on the waveform screen.
1 (Min Max Off) turns the Min Max feature off. The GMM test tool returns to its normal display mode. Measurement values and timer information cannot be recovered at this point.
2 (Restart) resets all values/timers and begins with new values and timers. Previous measurement values and timer information cannot be recovered at this point.
Run Pause Exit
3 (Run Pause) toggles between the normal “Run” state and a temporary “Pause” condition. Changing display modes when Min Max is active does not change the “Run” or “Pause” state.
In Combo, Meter, and Trend modes, “Pause” turns the “Min Max” icon at the top of the display dark, and displayed values stop updating (although the elapsed time continues counting.) In View Mode, “Run” and “Pause” operate normally, but measurement readings and timer information cannot be viewed.
5 (Exit) returns the softkeys to the functions for the selected measurement mode while the remainder of the display remains unchanged. You must use the Display mode key to return the window and secondary displays to the Combo, Meter, or Trend display. Min Max will remain enabled in the background mode.
4-4
Page 66
Using the Hardkeys
Range Softkeys
4
You can use Min Max with LOGIC to keep track of the range of logic levels being encountered. Use the following procedure:
1. With the selector turned to LOGIC, select logic
type by pressing
2. Press
3. Place the Min Max display in the background by
4. Check the Min Max values at any time by
Cancel Min Max by reselecting or changing the logic type. (Press Softkeys.)
M
and times now appears.
pressing reappears, and the Min Max icon indicates that Min Max recording is continuing in the background.
repeating steps 2 and 3.
d 1
1, 2
1, 2
. The full Min Max display of levels
. The LOGIC activity display
3
, or
3
, or
from the LOGIC
Range Softkeys
Softkeys accessed by pressing
Auto
Manual
1
(Auto Manual) toggles between Auto and Manual ranging. In Auto, the GMM test tool automatically selects the range presenting the highest resolution. Auto is the standard configuration; Manual can be selected as the power­up configuration when the selector is in the SET UP position. The GMM test tool resets to Autoranging when you select “Full Auto” (by pressing the VAC, VDC, or mV DC softkeys.)
2
and (if available). If Auto ranging is active, [ or ] selects Manual ranging and the next higher (lower) range.
[]
3
select the next higher (lower) range
r
are:
Wave Scale Exit
5
from
4-5
Page 67
867B/863
Users Manual
4 (Wave Scale - active in the View Display Mode) enters manual ranging and cycles through “units/Div” scaling factors of 1, 2, and 4 for the vertical axis, effectively zooming in and out on the waveform display. The scaling factors divide the units/Div for the present range. “Wave Scale” is not available in Meter Mode.
5 (Exit) returns to the softkeys previously in effect. (You can also press r to return to the previous softkeys.)
Entering Range cancels “Peak Hold”.
Touch Hold
This key allows you to hold a known good reading on the display. (No special softkeys accompany Touch Hold.) Use the following procedure:
1. Press H (
appears at top of the display.)
h
2. Apply measurement connections until you hear a beep. This sound signifies that a stable reading has been made. The display continues to update with valid readings (and beeps) as long as you maintain the measurement connections.
3. In Meter, Combo, and Trend display modes, the last valid primary numeric reading is held; the waveform in Combo and Trend is not held. In View Mode and COMPONENT Test, the last valid waveform is held.
Press H again to exit Touch Hold ( disappears.)
Touch Hold and “Peak Hold” are mutually exclusive: activating one cancels the other. Touch Hold cannot be activated when Frequency, SET UP, or LOGIC is selected. Pressing F or d cancels Touch Hold. Selecting “rms”, “Average”, or “Full Auto” cancels Touch Hold. P stops Touch Hold display updates; Touch Hold then continues to run in the background. In view mode, selecting Touch Hold cancels “Glitch Capture” and “Single Shot”.
h
4-6
Page 68
Chapter 5
Using View and Trend Display Modes
Introduction
This chapter describes how to display a waveform in View Display Mode and how to plot and interpret measurement data over time in TrendGraph Display Mode.
View Display Mode Basics
The GMM must know when to begin displaying (trigger) a waveform. Triggering requires that a valid signal level is specified, a direction for that signal (positive going or negative going) is selected, and the trigger signal source (internal or external) is set.
Setting the range for the vertical divisions and the timebase for the horizontal divisions directly
influences the proportions and legibility of the triggered waveform.
The type of waveform acquisition may also be important. A continuous, internal trigger enables an ongoing display of the input signal whenever you turn the selector to a new function.
Using Full Auto
“Full Auto” allows the GMM to automatically select the Trigger Level, Trigger Slope, Trigger Source, Range, and Time Base necessary for an optimum display of the input signal. If a good input is present, you will see a waveform that nearly fills the screen.
5-1
Page 69
867B/863
Users Manual
“Full Auto” (5) is enabled (highlighted) initially when View Display Mode is activated if “Auto” is selected in the SET UP screen and no manual settings have been introduced.
“Full Auto” may not be automatically selected when the GMM enters View Mode. If “Manual” is set as the power-up default for Range on the SET UP screen, “Full Auto” is not selected. Further, if you then change to “Full Auto” by pressing 5 from the View Mode Softkeys, “Manual” remains selected on the SET UP screen and will be reapplied when you change functions.
NOTE
Disabling any of the auto features turns the “Full Auto” softkey label off (not highlighted.) Other auto features then remain active. Pressing
5
from the View Mode Softkeys
turns all auto features off at once.
Changing the Amplitude Range
You can use the softkeys accessed with r to select a preset (Manual) amplitude range.
Using View Mode Softkeys
Time
>
>
Base
Trigger
1 (Time Base) accesses the Time Base Softkeys, which allow you to zoom in or out on the waveform by changing the horizontal (time) axis of the View display.
2 (Trigger) accesses the Trigger Softkeys, allowing you to control the trigger slope, level, and source.
3 (Single Shot) accesses the Single Shot Softkeys, which allow you to capture and examine a single waveform that has met your trigger criteria. The GMM test tool displays continuously updating waveforms if this key is not pressed.
4 (Glitch Capture) accesses the Glitch Capture Softkeys, allowing you to capture and examine a waveform that has met a predetermined trigger criteria with a briefly occurring spike. See Choosing the Acquisition Type later in this manual for further information on Trigger criteria for Glitch Capture.
Single
Shot
>
Glitch
Capture
>
Full
Auto
5-2
Page 70
5
5 s
4.2 µs
8.3 µs 20 µs
50 µs
100 µs
200 µs
500 µs
1 ms
2 ms5 ms
10 ms
20 ms
50 ms
100 ms
200 ms
500 ms
1 s
2 s
2 3
Slower
Faster
(Full Auto) selects the range, trigger level, and time base automatically to produce an optimum display of the waveform.
Setting Up the Time Base
Time
Base
>
>
Trigger
Single
Shot
>
Glitch
Capture
>
Full
Auto
Using View and Trend Display Modes
Setting Up the Time Base
5
Select “Time Base” (
1
) from the View Mode Softkeys, then use the Time Base Softkeys as follows:
Slower Faster Exit
2
(Slower) selects the next longer time per division. More waveform activity then appears in one time division, effectively zooming out on the waveform.
3
division. Less waveform activity then appears in one time division, effectively zooming in on the waveform.
5
(Faster) selects the next shorter time per
(Exit) returns to the View display.
Figure 5-1. Time Base Selection
rita0710.eps
5-3
Page 71
867B/863
Users Manual
Setting Up the Trigger
Time
>
>
Base
Trigger
Select “Trigger” ( Softkeys, then use the Trigger Softkeys as shown in the following paragraphs.
Selecting the Trigger Slope
Slope
K
Slope
t
1
(Slope) sets the waveform trigger to trip either on the rising or falling edge of the signal for single triggering. The counter triggering still use two (dual) triggers for its operation and are set automatically based on the input signal amplitude. In dual triggering, falling slope trigger points to allow their levels to be set individually. The waveform trigger icon will switch between two trigger points in the display as selects each trigger. The rising slope setting will always have a more positive setting than the falling slope setting. However, whichever slope is left selected will be the last criteria to be met before a
[]
1
Single
Shot
2
) from the View Mode
Glitch
>
Capture
Internal
External Exit
>
selects between the rising and
1
Full
Auto
trigger occurs, e.g., if the rising slope selection is left highlighted, the input signal must pass through the falling slope first and then through the rising slope to trigger the GMM. In single trigger, Full Auto will always set the trigger slope to rising.
The GMM trigger circuitry controls both the waveform and counter displays. The counter always uses a dual trigger setting while the waveform display can be controlled by single or dual triggering. The waveform trigger setting(s) are indicated in the display with an icon in the left margin along the vertical axis of the display grid (Figure 5-2). The counter trigger settings are indicated by the two icons along the right vertical axis and are marked with “Hz”. The counter and waveform trigger settings are the same when in Dual trigger operation. This is the only condition where the counter trigger levels are manually set.
5-4
Page 72
Using View and Trend Display Modes
Setting Up the Trigger
Changing the Trigger Level
Slope
K
Slope
t
[ ]
Internal
External Exit
5
Figure 5-2. Trigger Icons
86xsc011.tif
2
the waveform must pass through to trigger the GMM. For Internal triggering, the waveform trigger icon identifies the level and moves as you make the level adjustment. The actual trigger level is displayed in the upper right corner of the display. For Dual triggering, you must first select which trigger point you want to adjust with the The waveform icon will move from one point to the other in the display as the slope selection is changed. The counter trigger levels are linked with the waveform trigger levels in dual triggering and are the only time they can be manually set. For external triggering, a single trigger level is set with the ( and ]) softkeys and is indicated in the special trigger level indicator in the upper right corner of the display. There are 21 different preset trigger levels available for external triggering.
If “Full Auto” is selected, the trigger levels are set on percentages of the input signal.
and
3 ([
and ]) change the trigger level
1
Slope selection.
[
5-5
Page 73
867B/863
Users Manual
Selecting the Trigger Source
Slope
K
Slope
t
4
(Internal External) selects Internal or External
[]
trigger source. When “Internal” is selected, the trigger circuit uses the measured input signal for triggering. When “External” is selected, the trigger circuit uses the signal placed on the EXT TRIG jack for triggering. The waveform and counter trigger icons are removed from the display and a special trigger level indicator replaces the trigger level readings in the upper right corner of the display (Figure 5-3). There are 21 external trigger level steps. You can provide a trigger from an external source to synchronize the GMM display to some other event. Refer to Specifications for External trigger input levels.
5
(Exit) returns to the View Softkeys.
Internal
External Exit
Selecting Single or Dual Trigger
To accommodate various trigger situations, the waveform can be triggered by one (single) or two (dual) trigger settings. “Single” or “Dual” are trigger selections you can be make from the SET UP
screen. When using external triggering, only one trigger level may be set regardless of the set up selection.
rita0392.eps
Figure 5-3. Setting Up the Trigger
5-6
Page 74
Using View and Trend Display Modes
Choosing the Acquisition Type
5
Single triggering establishes a fixed trigger point for the waveform display. The input signal only needs to cross the trigger amplitude set by the trigger level and in the direction set by the slope setting. While in single trigger mode, the frequency counter trigger circuit sets two predetermined levels that are not adjustable. This ensures a stable frequency reading while triggering the waveform with a single trigger.
Dual triggering (the default setting) requires that the signal pass through both upper and lower trigger points to trigger the waveform and frequency counter. These points are initially set based on percentages of the input signal amplitude. You can change both levels with the Trigger Softkeys. Dual triggering provides a more stable waveform display in noisy environments.
Choosing the Acquisition Type
Time
>
>
Base
Instead of using the continuous internal trigger, you can instruct the GMM to update the screen only once, yielding a retained, snapshot view (“Single Shot”) of an input signal meeting the trigger requirements. Press Softkeys.
Yet another selection allows you to display and monitor only randomly-occurring or fast transient events or glitches (“Glitch Capture”). Trigger levels are preset and are not ajustable. The “Glitch” must have a change rate higher that 1 kHz and an amplitude > 20% of the selected GMM range.
“Single Shot” and “Glitch Capture” use the same submenu of softkeys as described in the following paragraphs.
Trigger
Single
Shot
3
from the View Mode
>
Glitch
Capture
>
Full
Auto
5-7
Page 75
867B/863
Users Manual
Using Single Shot and Glitch Capture Softkeys
Arm
1
(Arm) resets the trigger circuit and readies the GMM test tool to capture another single shot waveform or glitch capture transient. “Arm” is selected (highlighted) initially. When the entire waveform is captured, “Arm” reverts to normal contrast.
The display shows elapsed time (hh:mm) since arming or, if a trigger subsequently occurred, time between arm and trigger. The time cycles from 00:00 through 99:59 continuously until a trigger is received.
<>
Exit
2
or right for viewing (available once the waveform is completely captured). If you press arm cycle starts over, < and > disappear, and the arm-to-trigger time begins at 00:00. Waveform position accomplished with < and > is retained across arm cycles. Six time base divisions prior to the trigger and 19 time base divisions after the trigger can be viewed using < and >, respectively. Refer to Figure 5-4 for an overview of time divisions (on and off screen) that can be examined.
5
Free Run operation and autoranging are not available during “Single Shot” and “Glitch Capture”. Selecting Touch Hold cancels “Single Shot” and “Glicth Capture”.
3
and
(Exit) returns to the View Softkeys.
(< and >) move the waveform left
1
again, the
5-8
Page 76
Using View and Trend Display Modes
619
Choosing the Acquisition Type
5
Figure 5-4. Viewable Time Divisions (Single Shot and Glitch Capture)
rita0780.eps
5-9
Page 77
867B/863
Users Manual
Trend Display Mode Basics
Trend can be used in VAC, VDC, mV DC, Ohms, Amps, and mAµA functions. This mode uses the graphical window to plot 120 data points over a selectable time ranging from 1 second to 15 minutes. Each data point can represent an actual reading, an average of readings, or a pair of high and low readings. You can also enable serial output of the data points as they occur.
Using Trend Mode Softkeys
Enable RS232
1
(Enable RS232) starts transmitting readings (with time stamps) with the first press and stops transmitting with the second press. (You can also
F
press (hours:min:sec) resets each time RS232 is enabled. The time rolls over at 24 hours. Refer to Chapter 6 for serial output connection descriptions.
2
or the interval to use on the horizontal scale as data
Time
Longer
Time
Shorter
Trend
Type
>
to stop transmission.) The time stamp
3
(Time Longer, Time Shorter) selects
Restart
Trend
points are plotted on the trend graph. (Seconds: 1, 2, 5, 10, 15, 30, 45; Minutes: 1, 2, 5, 10, 15)
For “Sampled” and “Average” Trend Type selections, the GMM plots the most recent 120 data points. For “High Low” Trend Type, 120 data point pairs are plotted. Earlier data points are not stored.
4
(Trend Type) accesses the Trend Type softkeys, allowing you to plot average, sampled, or high/low excursions of the measurement value.
5
(Restart Trend) starts plotting a new trend graph as new samples are acquired. Primary readings are displayed and plotted; there are no secondary readings in Trend Mode.
Trend Mode Considerations
Only the value being trended is shown in the primary display. For example, a “dB” or “Rel” value that normally occupied the primary display in DC Volts function will be replaced with a VDC reading. Frequency values do remain in the primary display during Trend.
The vertical scale in the Trend display depends on the range setting and the type of measurement. A
5-10
Page 78
Using View and Trend Display Modes
Using Trend Mode Softkeys
5
unipolar arrangement (0 to full scale) is used for VAC, AAC, and frequency. Bipolar readings (negative to zero to positive) are used for VDC and Ohms.
RS232 Trend Output Format
Trend data for each data point is sent out as it is measured. ASCII format is used, as follows:
For “Sampled” or “Average” Trend Type: [Value] [Units] [Time Stamp] <cr> <lf>
For “High Low” Trend Type: [Value High] [Units] [Time Stamp] <cr> <lf> [Value Low] [Units] [Time Stamp] <cr> <lf>
For “Glitch” Trend Type (Ohms only), the following line is added at the end of each “Sampled” or “Average” line or pair of lines:
[Glitch State] [Time Stamp] <cr> <lf>
[Time Stamp] is in hh:mm:ss format, representing time since the Trend start. Time rolls over at 24 hours.
Selecting the Trend Type
High
Average Sampled
1
(Average) plots the average of all samples
taken by the GMM during the plot interval.
2
(Sampled) plots the last sample taken by the
GMM at the time the data point is plotted.
3
(High Low) plots the maximum and minimum sample value taken during the time interval. A vertical bar where the bottom end represents the lowest sample and the top end represents the highest sample is placed on the plot.
4
(Normal Glitch) selects “Normal” or “Glitch” in Ohms function only. Glitch setting display and highlight fast-occurring discontinuities on the trend line. Markers along the bottom of the display identify discontinuities as they appear on the Trend Plot. A discontinuity is recorded when the measured resistance rapidly passes (decreasing or increasing) through a low resistance continuity point. Each range uses a different continuity point (See Continuity Beeper Values in Chapter 8 of this manual). Note: When Glitch is first enabled, the GMM sets manual
Low
Normal
Glitch Exit
5-11
Page 79
867B/863
Users Manual
ranging and sets the range to 300. It may by necessary to manually change the range for an on­scale reading.
5 (Exit) returns to the Trend Softkeys.
5-12
Page 80
Chapter 6
Using Save, Recall, Print, and Set Up
Introduction
This chapter contains instructions for using the Save/Print key to save, recall, and print GMM test tool display information and the SET UP selector position to change the GMM test tool power-up configuration.
Refer to Figure 6-1 for an overview of operations accessible through the Save/Print key.
For Model 867B, refer to Figure 6-2 for instructions on connecting a printer or PC to the GMM test tool.
Using the Save/Print Softkeys
Pressing P stops waveform and numerical value updates, freezing the screen. Display contrast is reversed, the “Freeze” message appears, and the Save/Print Softkeys are accessed. You can now Print display information or save and recall display and configuration information.
If on, the “Enable RS232” feature (used in Trend Mode) is canceled. Otherwise, active Touch Hold, Min Max, Trend, and “Peak Hold” features continue to update in the background (beeps continue).
6-1
Page 81
867B/863
Save
Print Exit
Clear ExitSave
Load
Screen Exit
Clear Exit
Load
Config. ExitPrint
Stop Print
Stop Print
Users Manual
6-2
Figure 6-1. Freeze (Print/Save) Softkey Map
rita0370.eps
Page 82
Using Save, Recall, Print, and Set Up
Saving Screen and Configuration
6
P
Save
Screen
1
(Save Screen) accesses the Save Screen softkeys, which allow you to save the present display to a memory location. (Model 867B only.)
2
(Recall) accesses the Recall softkeys, which allow display of a memory location.
3
(Print) prints the display. “Freeze” changes to “Printing”, then reverts to “Freeze” when printing is complete. You can stop printing by pressing (Stop Print) or
4
(Save Config.) accesses the Save Config. softkeys.
5
(Exit) (or pressing original display. You must press one of these two keys to return; other keys do not operate during the Freeze display.
>
Recall Print
5
>
(Exit); other keys are ignored.
P
Save
>
Config. Exit
again) returns to the
3
Saving Screen and Configuration
(Model 867B only)
Save
>
Screen
][
1
location. Basic information (a reading, etc.) is displayed with the title for each memory location.
3
supporting configuration (message appears), then returns to the Save/Print display.
4
appears only if something is saved in this memory location.)
5
Save/Print display.
2 ([
and
(Save) saves the displayed information and
(Clear)) erases this memory location. (“Clear”
(Exit) (or pressing
Save Clear Exit
and ]) select the Memory
P
again) returns to the
6-3
Page 83
867B/863
Users Manual
Recalling Screen or Configuration
>
Recall
View
>
][
1
and
2
cycle through memory locations
Memory
(seven configuration memories for all models and three display memories on Model 867B only).
3
(View Memory) previews the selected memory and accesses the View Memory softkeys. “View Memory” does not appear if memory is empty. Previewed memory always appears in reverse video.
4
(Load Screen or Load Config) directly replaces the active display with the selected display from memory. The selector function must match that of the presently selected memory list item, and the selected memory location must be occupied.
5
(Exit) returns to the Print/Save softkeys (display frozen).
Load
Screen Exit
Previewing Screen or Configuration
>
Recall
][
1
and
2
previewing (reverse video).
3
(Print) begins printing the display. “Printing ...” appears on the display and “Stop Print” defines the softkey. Press until “Print” becomes the softkey label again.
4
(Load Config. or Load Screen) can be used if the selected memory matches the present selector function configuration and the selected memory location is occupied.
“Load Config” appears when you select one of the seven configurations. The GMM loads the selected configuration and returns to the active display.
View
>
Memory
Load
Print
Config. Exit
scroll through memory locations for
3
again to stop printing, or wait
6-4
Page 84
Using Save, Recall, Print, and Set Up
Saving the Present Configuration
6
“Load Screen” appears when you select one of the three screens. Press waveform along with the active waveform. The GMM assumes the time base and amplitude settings of the stored waveform. With the recalled and active waveforms both displayed, you can press access the following three softkeys: the two-waveform display; the recalled waveform and return to the active waveform display; waveform display.
If you press any other key, the recalled waveform clears and the display returns to the active waveform.
In View Mode, the stored waveform must have been saved while the GMM was in View Mode. The loaded screen appears as solid lines. In COMPONENT TEST, the stored memory must have been saved while the GMM was in COMPONENT TEST.
5
(Exit) returns to the Print/Save softkeys
(display frozen).
4
to display the stored
3
4
“Clear Exit” to clear
5
“Exit” to return to the two-
P
to
“Print” to
Saving the Present Configuration
Save
>
Config.
][
1
3
numerical data sufficient to recreate this display. A message appears while the configuration is being saved. The GMM test tool then returns to the Save/Print display.
4
appears only if something has been saved in this memory location.)
5
2
and (Save) stores present configuration and
(Clear) clears this memory location. (“Clear”
(Exit) returns to the Save/Print display.
select the Memory location.
Save Clear Exit
6-5
Page 85
867B/863
Users Manual
Printing
Stop
Print
Do not print if the GMM is presently controlled by FlukeView 860 application software.
Use the following procedure for printing:
1. Verify appropriate “Printer Type”, “Baud Rate”, and “Parity”. These selections appear as the last
Print
NOTE
three items on the SET UP screen (selector turned to SET UP.)
2. Connect the GMM to a printer as shown in Figure 6-2. Or, download the screen to FlukeView 860 Application Software running on a PC. Then print from FlukeView 860.
3. Set the GMM to display the screen you want to print. The screen can be the active display or one of the recalled screens.
3
4. Press
5. You can now press time to cancel the screen printing.
(Print) to begin printing.
3
(Stop Print) at any
6-6
Page 86
Correct
Incorrect
PAC91
PM9080
Using Save, Recall, Print, and Set Up
Printing
6
Figure 6-2. PC and Printer Connections
rita0270.eps
6-7
Page 87
867B/863
Users Manual
Changing the Configuration
U
Set up Self
Test
(wait 2 seconds)
Next Item
If no keys are pressed for 15 seconds after the selector is turned to SET UP display contrast returns to the default mid-scale setting.
1. Beginning at the first line (LCD Contrast), press
1
<>
NOTE
to step through the other lines in this
Restore
All
Exit
Save
Set up
display. Continue pressing to wrap back to the top line. At each configuration line, press make your selection. Then press to the next line.
2. Press
3. Press
If you make SET UP changes and do not press (Save Set up), the changes remain in effect only until you turn the GMM off.
4
to restore all lines on this configuration display to their standard settings. You have two choices at this point. First, you can
5
press information as the power-up configuration. Second, you can use the retrieved standard configuration as a starting point, change some of the settings as described above, and then store your customized settings as the new power-up configuration.
presently shown on the display into nonvolatile memory. These settings become effective immediately and are reestablished each time the GMM test tool is powered on.
now to save the displayed
5
to save all configuration settings
2
1
3
or
to move
5
to
6-8
Page 88
Using Save, Recall, Print, and Set Up
Changing the Configuration
Table 6-1. Configuration Selections (SET UP)
SET UP ITEM SELECTIONS AVAILABLE DEFAULT
LCD Contrast Press < or > to adjust display contrast. (mid-range)
6
AC Converter “rms” or “Average”. This setting controls “rms” or “Average” softkey selection
in ac volts or amps functions. Display Mode “Combo” or “Meter”. Combo dB Reference dB ohms reference values: 2 4 8 16 50 75 93 110 125 135 150 300 600
900 1000 1200. Continue pressing > to select “Auto”, “Manual”, or “Disabled”. Range “Auto” or “Manual”. Auto Beeper “Enabled” or “Disabled”. Enabled Sleep Mode “Enabled” or “Disabled”. Enabled Trigger Mode “Dual” or “Single”. Dual Printer Type “Epson”, “HP Graphics”, or “Text.” Epson Baud Rate 1200, 2400, 9600, or 19200. 1200 Parity “no”, “even”, or “odd”. (Number of Bits is always 8; number of Stop Bits is
always 1.)
True rms
600
none
6-9
Page 89
867B/863
Users Manual
6-10
Page 90
Chapter 7
User Maintenance
Introduction
This chapter covers basic maintenance procedures that can be performed by the user. For complete service, disassembly, repair, and calibration information, refer to the 867B/863 Service Manual (PN 689312).
Cleaning
Clean your GMM test tool with a damp cloth and a mild detergent. Do not use abrasives, solvents, or alcohol.
Warning
m
To avoid ELECTRICAL shock, remove the test leads and any input signals before opening the case.
Testing the Fuses
Test the fuses with the following procedure:
1. Turn the selector to Ohms (
2. Connect a test lead between the mAµA input terminals. Check that the display reads less than 5. This tests the 440 mA fuse. If this reading is “OL”, refer to the following fuse replacement information.
3. Move the test lead from the mAµA terminal to the A terminal. Check that the display reads less than 5Ω. This tests the high energy (11A) fuse. If this reading is OL, refer to the following fuse replacement information.
4. If either reading is outside the range shown, have the GMM test tool serviced.
R
).
V
and
7-1
Page 91
867B/863
Users Manual
Replacing the Batteries
The GMM test tool uses 6 AA alkaline batteries (included with Model 863) or a nickel-cadmium battery pack (BP 7217 - included with Model 867B). Internal charging for the battery pack is provided with Model 867B. The NiCd battery pack must be fully charged before using the GMM. External charging (using BC7210) is necessary if the battery pack is used with Model 863 (instructions are included with BC7210.) Refer to Figure 7-2 for battery replacement instructions. Replace the NiCd battery pack when it no longer holds a charge for the rated interval. Order replacement Model BP7217 Nickel-Cadmium Battery Pack. In the USA and Canada, call Fluke Service Parts at (800) 526-4731. Elsewhere, call +1 425­356-5500 for instructions.
n
Note
If the Tester is dropped, it may not turn on. Check the batteries to see if they are damaged and need to be replaced.
Do not mix spent Nickel-Cadmium batteries with the solid waste stream. Spent batteries should be
disposed of by a qualified recycler or hazardous materials handler. Contact your authorized Fluke Service Center for recycling information.
Replacing the 440 mA Fuses
m
Warning
Refer to SAFETY in the front of this manual prior to replacing any fuse.
Refer to Figure 7-3 to locate and replace the 440 mA fuse(PN 934121). Note that the battery compartment must be accessed first; refer to Figure 7-2 for these initial access procedures.
Replacing the 11A (High Energy) Fuse
You must disassemble the GMM case to replace a blown high energy (1000V, 11A) fuse (PN 943118). The GMM’s safety design places this fuse inside the case to protect the user and to encourage inspection of the circuit board for damage if very high energies have been encountered. Proceed as follows:
1. Set the GMM selector to OFF. Be sure to leave the selector in this position throughout the disassembly and reassembly procedure.
7-2
Page 92
User Maintenance
Reassembly
7
2. Remove the six screws attaching the case halves. Detach the case halves.
3. Remove the two screws securing the shield as shown in Figure 7-1. Then pry the shield loose from two stanchions at the top of the circuit board.
4. Gently pull up on the shield tab and rotate the entire shield away from the circuit board.
5. Remove the three screws securing the circuit board to the front half of the GMM.
6. Dislodge the circuit board by pushing in on the Volts-Ohms input connector from the lower right front of the GMM. (See Figure 7-1, Detail A).
7. Lift the entire circuit board up approximately one inch, dislodging the slotted Battery Eliminator connection.( See Figure 7-1, Detail B).
8. Rotate the circuit board 180 degrees.
9. Remove the fuse.
Note
For a more detailed drawing, see the Service Manual (PN 689312).
Reassembly
Generally, reassembly reverses the disassembly procedure. Specifically, do the following:
1. Verify that the selector switch still points to OFF.
2. Place the case top face down on a non-marring surface.
Note
At all times, avoid stressing the ribbon cable.
3. Grasp the board at the top and bottom so that it can be smoothly lowered into position in the case top.
4. As you begin lowering the board into position, make sure the battery eliminator connection (Detail B in Figure 7-1) slips into its slot snug against the top case wall.
5. Gently press the board into position at the top and bottom
6. Replace the three screws securing the board.
7. Rotate the shield back into place on the circuit board.
8. Replace the case bottom. Tighten the six screws (middle screws first) securing the case bottom.
7-3
Page 93
867B/863
5
5
3
3
4
8
8
7
6
6
Users Manual
7-4
f7-1.eps
Figure 7-1. Replacing the 11A (High Energy) Fuse
Page 94
User Maintenance
Operational Test
7
Operational Test
Use the following procedure to test the basic operability of your GMM test tool:
1. With no inputs connected, turn the GMM test tool
v
on by rotating the selector to VAC ( Meter display mode should appear, with a random display in the mV range.
E
2. Rotate the selector to VDC ( primary display in the mV range.
3. Rotate the selector to Ohms ( display of “OL”.
4. Connect the test leads to
5. Short the test lead tips and observe a display of approximately 0.250 ohms.
6. Press
For a complete performance test of GMM test tool functions and specifications, refer to the 863/867B Service Manual (PN 689312).
2 (/
listen for a beep.
), short the test leads again, and
) Observe a similar
R
V
). Combo or
). Observe a
C
and
.
Self Test
Begin an internal diagnostic routine by selecting SET UP, wait two seconds, press now presents the following information:
1. A test pattern.
2. An error message (if a problem is encountered during the test). The GMM test tool Model and serial numbers.
3. The calibration status.
Generally, you cannot use the GMM test tool keys during Self Test. (You can press diagnostic routine and return to the SET UP Softkeys.)
d2
1
. The display
to exit the
7-5
Page 95
867B/863
Closed
AA
Batteries
Battery
Pack
Open
+
OR
+
Users Manual
7-6
Figure 7-2. Replacing the Batteries
rita0535.eps
Page 96
User Maintenance
Supplied
Spare Fuse
Fuse
Supplied
Spare Fuse
Self Test
7
rita0674.eps
Figure 7-3. Replacing the 440 mA Fuse
7-7
Page 97
867B/863
Users Manual
If Your GMM Does Not Work
NOTE
If the display remains blank after pressing
W
, contrast may be off the scale. Restore contrast to mid-scale with the following shortcut: set the selector to SET UP and avoid touching any key until the display reappears (about 15 seconds). Press to store this setting, then turn the selector to the desired function.
Examine the case for physical damage. If damage is detected, contact a Fluke Service Center. Check and replace (as needed) the battery, fuses, and test leads. Review this manual to verify correct operation.
5
If the GMM still does not work, pack it securely and forward it, postage paid, to the nearest service center. Include a description of the problem. Fluke assumes no responsibility for damage in transit.
A GMM under warranty will be repaired or replaced (at Fluke’s option) and returned at no charge. See the registration card for warranty terms. If the warranty has lapsed, the GMM test tool will be repaired and returned for a fixed fee. Contact the nearest service center for information and prices.
If “Uncal” appears in the Upper Status Line, the GMM is no longer calibrated; contact a Fluke Service Center.
To locate an authorized service center, visit us on the World Wide Web: www.fluke.com or call using any of the phone numbers listed below:
USA and Canada:
1-800-44-FLUKE (1-800-443-5853) Europe: +31 402-678-200 Japan: +81-3-3434-0181 Singapore: +65­Anywhere in the world: +1-425-356-5500
-276-6196
*
7-8
Page 98
General Specifications
100
90 80 70 60 50 40 30 20 10
0
-20 0 20 40 60 80 100 120 140 Temperature (deg F)
Temperature (deg C)
%RH
-4
-20
32
0
30 40 50 60
= Normal Operation (0˚C — 50˚C)
= Storage (-20˚C — 60˚C)
Chapter 8
Specifications
Display:
Temperature Operating:
Temperature Coefficient:
(0°C to 18°C and 28°C to 59°C)
Relative Humidity: Altitude Operating/Non-operating:
meters)/ 40,000 ft. (12,200 meters)
Input Impedance:
LCD - Dot Matrix, 240 X 200 pixels
Fluke 867B: Fluke 863:
Transmissive, Backlit
Reflective
0°C to 50°C (See Chart)
Storage: Charging:
-20°C to 60°C (Batteries Removed) 0 to 45°C
(0.1 X % Accuracy) per °C
0% to 90% non-condensing
10 M
6,562 ft. (2,000
f34.eps
8-1
Page 99
867B/863
Users Manual
Shock and Vibration: per MIL-T-28800, class 3, sinusoidal, non-operating
Dimensions: 5.4 x 9.7 x 2.7 in. (137 x 246 x 68 mm) Weight: 3 lbs (1.35 kg) Battery Operating Time (backlight off)
Alkaline: 6 hours typical NiCd: 10 hours typical
Battery Recharge Time: 16 hours minimum from full
discharge
Drip Proof Case: per IEC 529; IP 52, Drip Proof
Safety: Designed to meet IEC 1010-1 Category III,
UL3111, CSA-C22.2. 1010-1-92, CE and TUV requirements
Certification:
Electromagnetic Interference
P
* +
%
RF Emissions EN-50081-1 Commercial Limits
VFG 243-1991 FCC Part 15 Class B,
RF Susceptibility: EN 50082-1 Industrial Limits
Power:
Fluke 867B Fluke 863
Battery Eliminator/Charger Yes Optional Eliminator only
NiCd Battery Pack Yes
Alkaline Batteries 6 AA Cells Optional Customer Supplied Yes Battery Life: NiCad
Alkaline
8 hrs (supplied)
4 hrs (optional)
Optional BP7217 Battery Pack Optional BC7210 Ext. Charger
10 hrs (optional)
6 hrs (supplied)
8-2
Page 100
Specifications
General Specifications
8
Specifications:
The following specifications apply within 1 year of calibration when operating
in a temperature range of 18°C (64°F) to 28°C (82°F).
AC Volts (True RMS, AC-Coupled) [±(percent of reading + counts)]
Range
20-50 Hz 50-100 Hz 1 kHz-30 kHz 30 kHz-100 kHz 100kHz-200kHz 200 kHz-300 kHz
300.00 mV
3.0000V 1.5% + 10
30.000V 1.5% + 10
300.00V 1.5% + 10
1000.0V 1.5% + 10
Crest Factor:
Measurement Range:
1.5% + 10
0.19 dB
0.19 dB
0.19 dB
0.19 dB
0.19 dB
300 mV - 300V ranges 3:1, 1000V range 3:1 decreasing to 1.41:1 as input voltage increases to
1000V (peak voltage not to exceed 1414V).
0.5% + 10
0.10 dB
0.5% + 10
0.10 dB
0.5% + 10
0.10 dB
0.5% + 10
0.10 dB
1.5% + 10
0.19 dB
300 mV range from 10% to 100% of range. 3V - 1000V ranges from 5% to 100% of range. For frequencies > 100 kHz 30% to 100% of range (all ranges).
0.5% + 10
0.10 dB
0.5% + 10
0.10 dB
0.5% + 10
0.10 dB
0.5% + 10
0.10 dB NA NA NA NA
Frequency
4% + 200
1.39 dB
4% + 200
1.39 dB
4% + 200
1.39 dB
4% + 200
1.39 dB
8% + 200
1.68 dB
8% + 200
1.68 dB
8% + 200
1.68 dB
8% + 200
1.68 dB
10%+ 200
1.82 dB
10%+ 200
1.82 dB
10%+ 200
1.82 dB
10%+ 200
1.82 dB
8-3
Loading...