6-2.PC and Printer Connections................................................................................... 6-7
7-1.Replacing the 11A (High Energy) Fuse.................................................................. 7-4
7-2.Replacing the Batteries .......................................................................................... 7-6
7-3.Replacing the 440 mA Fuse................................................................................... 7-7
viii
Page 10
Safety
mm
In this manual, the word “WARNING” identifies
conditions and actions that pose hazard(s) to the
user. The word “CAUTION” refers to conditions
and/or actions that can damage the instrument. Use
of the instrument in a manner not specified may impair safety. Read the following safety information
carefully before attempting to operate or service the
instrument.
• Use caution when working above 60V dc or 30V
ac rms. Such voltages pose a shock hazard.
• When using the probes, keep your fingers away
from probe contacts. Keep your fingers behind
the finger guards on the probes.
• Place test leads in proper input terminals.
• Disconnect the live test lead before disconnect-
ing the common test lead.
• Avoid working alone.
• Disconnect the power and discharge high-
voltage capacitors before testing in Ohms
Continuity, Diode, Capacitance, and COMPONENT Test.
• Inspect the test leads for damaged insulation or
exposed metal. Check test lead continuity.
Damaged leads should be replaced.
• Do not use the GMM if it looks damaged.
• Select the proper function and range for your
measurement.
• When measuring current, turn the power off to
the circuit being tested before connecting the
GMM test tool in that circuit.
• Check GMM test tool fuses before measuring
current transformer secondary or motor winding
current. (See Testing the Fuses in the User
Maintenance chapter.) An open fuse may allow
high voltage buildup, which is potentially hazardous.
• Use clamp-on probes (current clamps) when
measuring currents exceeding 10A.
ix
Page 11
867B/863
Users Manual
WARNING
m
Hazardous voltages are present on
the External Trigger input terminal
and any attached probe when the
GMM is measuring voltages greater
than 30V rms or 60V dc.
Replacing Fuses
The following fuses should be used:
• F 440 mA, 1000V, Minimum Interrupt Rating
10,000A.
• F 11A, 1000V, Minimum Interrupt Rating
17,000A.
Warning
Use of fuses with voltage ratings
below 1000V reduces protection.
Do not replace the 440 mA 1000V
fuse with a 4/10A 600V fuse.
m
$
WARNING
m
CAUTION
s
Equipment protected throughout by DOUBLE
INSULATION or REINFORCED INSULATION.
mm
Overload Protection
Voltage or Current Input: 1000V
Surge Protection: 8 kV
Maximum Voltage Isolation to Ground: 1000V
Protection Levels:
(fixed distribution-level installations which can
experience peak voltage transients up to
8000V.)
Protection Class: II
risk of electric shock.
see explanation in manual.
per IEC 1010-1; 1000V CAT III
s
x
Page 12
The NiCd battery pack must be fully charged before using the 867B. See Chapter 3.
A
COM
V
mA/
A
EXT TRIG
A
COMPONENT
TEST
SET UP
mV
V
V
OFF
LOGIC
HiZ
FREEZE
DISPLAY
MODE
SAVE
PRINT
MIN
MAX
RANGE
WAKE UP
HZ
TOUCH
HOLD
12345
mA
A
B
A
C
E
D
F
G
Your Graphical Multimeter
Figure 1-1. Graphical Multimeter Features
rita0091.eps
Chapter 1
A Quick Tour
Note
rita0092.eps
Figure 1-2. RS232 and AC Power Connections
1-1
Page 13
867B/863
Users Manual
Table 1-1. Introducing Your Graphical Multimeter
ITEMSYMBOLDESCRIPTION
1-2
A
j
W
B
C
Combo
Meter
View
Trend
1
2
3
TURN THE GMM ON BY ROTATING THE SELECTOR TO ANY FUNCTION
selector to the
then the screen sets up for this function, and you are ready to begin. You can toggle the backlight
power level (Model 867B only) by pressing W. Also note that the screen goes dark if you do not
change any controls for about 20 minutes and the GMM is on battery power; pressing W wakes
up the screen (all models.)
THE SCREEN
secondary display of smaller digits, and a graphical window showing either a waveform or an
analog NeedleGraph representation of the primary display. In addition, the upper screen line
displays status information. The bottom screen line (immediately above the softkey labels) shows
the display mode, range, and other information. For example, “Combo” appears on this line if a
waveform appears in the graphical window: the GMM is in Combo Mode.
THE SOFTKEYS.
along the bottom of the screen. These labels define the present uses for the five blue keys that
are positioned immediately below the screen. Together, the labels and the keys form a set of
Softkeys. These changing softkey definitions allow direct access to a wealth of GMM capabilities.
Refer to Figures 1-6, 1-7, and 1-8 at the end of this chapter for a roadmap to softkey use.
AC Volts (v)
. In AC Volts, the initial screen presents a primary display of large digits, a
Most control settings (selector change or key press) change the softkey labels
position. The input connections screen appears for a few seconds,
. Turn the rotary
4
5
Page 14
Table 1-1. Introducing Your Graphical Multimeter (cont)
ITEMSYMBOLDESCRIPTION
A Quick Tour
Your Graphical Multimeter
1
D
E
F
G
d
F
P
M
r
H
V
C
X
A
a
B
RS232
THE HARDKEYS
operations in all functions. Pressing W wakes up the screen from Sleep Mode (all models) and toggles
the backlight power level on Model 867B. Although hardkeys can generally be pressed at any time,
certain feature combinations are not allowed. The GMM test tool rejects any hardkey press that is not
allowed; a long beep sounds. Chapter 4 explains each hardkey in detail.
MEASUREMENT CONNECTIONS.
familiarization: no actual connections are necessary during the Quick Tour. Volts, Ohms,
Continuity, Conductance, Diode Test, Capacitance, and Component Test measurements use
for the Red lead. Amps measurements use A for the Red lead. mAµA measurements
V
use
a
functions use
views.
LINE POWER OR BATTERY POWER.
power through the Battery Eliminator. New alkaline (AA) cells provide a minimum of 4 hours
operation. A fully charged BP7217 NiCd battery pack typically provides a minimum of 8 hours
operation. The battery eliminator recharges the NiCd battery pack on Model 867B. Refer to
Figure 1-4.
SERIAL PC OR PRINTER CONNECTION.
interface cable to communicate with a PC or printer. Refer to Chapter 6.
. Hardkeys appear immediately below the softkeys. Hardkeys perform the same
Review the input connection scheme now only for
for the Red lead. LOGIC Test measurements use
for the Black lead. Refer to Figure 1-3 for TL70A test lead connection
C
You can power the GMM with batteries or from ac line
Your GMM can use an optional optical serial
X
for the Red lead. All
1-3
Page 15
867B/863
TL70A
Test Leads
Users Manual
1-4
Figure 1-3. Test Leads and Measurement Connections
rita0740.eps
Page 16
A Quick Tour
Europe PN 942602US PN 942599
JP PN 942610
UK PN 942607AU NZ
PN 944595
Your Graphical Multimeter
1
Figure 1-4. Using the Battery Eliminator
rita0770.eps
1-5
Page 17
867B/863
Users Manual
Table 1-2. A Practice Session
STEPACTIONDESCRIPTION
1-6
1
2
d
J
3
1
CHANGE THE DISPLAY MODE
With the GMM still in AC Volts, press first d, then
screen changes to the View Mode, where the entire screen is used for displaying a waveform.
You can repeat this procedure to select any of the display modes available in AC Volts (Combo,
Meter, View, and Trend.) Refer to Table 1-3 for an overview of display modes available in each
function. Examples of all display modes (Combo, Meter, View, Trend, Logic Test, Auto Diode
Test, Component Test, and Sleep) follow this table. Along with the flexibility provided by multiple
softkey definitions, display modes increase the versatility of your GMM considerably.
Many operations activated in one display mode are retained when you select a new mode. For
example, “Rel”, “dB”, Min Max, range, and frequency measurement type are retained.
CHANGE THE FUNCTION
Rotate the selector to DC Volts (E). Observe that the display shifts back to Combo Display Mode.
(This is the default mode, as established on the screen you see when you turn the selector to
SET UP.)
USE THE SOFTKEYS
With DC Volts and Combo Mode still active, you can use the DC Volts Softkeys. For example,
press
deactivate “Rel”. Explore all five softkeys in this fashion. Press a key first to activate a selection
(label highlighted) and second to disable the selection.
to activate “Rel”: difference readings only are now displayed. Press
1
to change the display mode. The
3
1
again to
Page 18
Table 1-2. A Practice Session (cont)
STEPACTIONDESCRIPTION
A Quick Tour
Your Graphical Multimeter
1
4
5
6
7
r
M
H
F
CHANGE THE MEASUREMENT RANGE
Press r to access the Range Softkeys, then press
“Manual” ranging.) Press
DC Volts Softkeys.
ENABLE MIN MAX READINGS
Press
readings. Press M again to deactivate Min Max.
ENABLE TOUCH HOLD
Press H to begin holding the last valid reading in the display, even if the test leads are
withdrawn from the measurement point. h appears in the upper status line. Press H again to
stop Touch Hold
CHANGE THE TYPE OF FREQUENCY DISPLAY
Press F to access the Frequency Softkeys.
The frequency reading (which was in the secondary display) and DC Volts reading (which was in
the primary display) are swapped.
The softkeys allow you to select different types of frequency measurements (Hz, Duty Cycle,
Pulse Width, or Period). Press
Volts Softkeys. Primary and secondary displays swap again, and the change to “Period” frequency
measurement continues in the secondary display.
to start displaying the maximum (“Max”), average (“Avg”), and minimum (“Min”)
M
to switch back to “Auto” ranging, then press r to return to the
1
to select “Period”, then press F again to return to the DC
4
to select a higher range (and activate
2
1-7
Page 19
867B/863
Users Manual
Table 1-2. A Practice Session (cont)
STEPACTIONDESCRIPTION
8
9
P
LOOK AT THE SAVE/PRINT SOFTKEYS
Press P to access the Save/Print Softkeys. You can now use the softkeys to exercise various
save, recall, and print features. Refer to Chapter 6 for more information. (Waveform save/recall is
not available on Model 863.)
TURN THE GMM OFF
o
About This Manual
This Quick Tour has introduced you to all the
features necessary to use your GMM. If you are
comfortable with this level of information, continue on
to Chapter 2 for specific information keyed to each
type of measurement. If you want more information
about any of the features, refer to Chapters 3
through 8, as follows:
Chapter 3
•
controls, screen, and input connections.
Chapter 4
•
hardkeys (
1-8
: General descriptions of the GMM
: Complete instructions for using the
d F P M r H
).
Chapter 5
•
detail.
Chapter 6
•
and Configuration Changes using SET UP.
Chapter 7
•
replacement, etc.)
Chapter 8
•
Index
•
: View Mode and Trend Mode, in
: Save, Recall, and Print using
. User Maintenance (battery and fuse
: Specifications
P
Page 20
A Quick Tour
Combo Mode
1
Combo Mode
Combo
MeterViewTrendExit
Combo Mode presents both
digital and graphical views of the
measurement and can be used
in VAC, VDC, mV DC, Amps,
and mAµA functions. Combo is
the default mode in these
functions.
Meter Mode
Combo
Meter is the only mode available in Diode and
Capacitance Test.
Meter
ViewTrendExit
Meter Mode presents both a
digital reading and an analog
position indication of the
measurement. You can select
Meter in VAC, VDC, mV DC,
Ohms, Amps, and mAµA
functions; you can also set Meter
on the SET UP screen as the
default mode in these functions.
View Mode
ComboMeter
Trend Mode
ComboMeterView
View
View Mode displays a full-screen
waveform. It is available in VAC,
VDC, mV DC, Amps, and mAµA
functions.
Trend Mode presents a digital
reading of the measurement and
plots that reading over time. It is
available in VAC, VDC, mV DC,
Ohms, Amps, and mAµA
functions.
TrendExit
Trend
Exit,
1-9
Page 21
867B/863
Users Manual
Auto Diode Test Mode
Meter
Auto Diode Test presents a DC
voltage reading and a symbol
representing the state of the
diode (forward, reverse, open,
short, unknown.)
Logic Test Mode
Logic
Logic Test shows icons
representing high and/or low logic
state along with frequency and
voltage readings of logic activity.
Exit
Exit
Component Test Mode
Comp.
Test
Component Test displays a
unique pattern representing the
characteristics of the in-circuit
component being tested. This
signature can resemble a
straight line for purely resistive
components or an ellipsoid for
frequency-sensitive components.
Sleep Mode
restores the screen.
W
Sleep Mode comes on
automatically during battery
operation if you do not press a
key or turn the selector for
approximately 20 minutes.
Recharge the NiCd battery pack before using the 867B. See Battery Considerations in Chapter 3.
Introduction
This chapter shows overall procedures for making
measurements in any of the rotary selector positions.
Necessary input connections are shown. Softkeys
used with each of the display modes available for the
selected function are also shown.
Using the Selector
To begin using the GMM, turn the selector to any
of the indicated functions. Softkey labels
momentarily identify the display modes available for
this function, then proceed to list the softkeys
available with this function.
Softkeys
Any selector setting or hardkey press provides you
with a unique set of softkey labels along the bottom
of the display. These labels define the present uses
for the five blue keys that are positioned immediately
below the display.
In Combo and Meter display modes, some labels are
used with several different selector positions; see
Table 2-1 for descriptions of these common labels.
Other labels are unique for the selector position;
these labels are defined in this chapter as needed.
View and TrendGraph softkey labels are common
for all functions that can use these modes; refer to
Chapter 5 for a full discussion.
Hardkeys
Although hardkeys can generally be pressed at any
time, certain feature combinations are not allowed.
The GMM rejects any hardkey press that is not
allowed and tells you so with a long beep.
2-1
Page 27
867B/863
Users Manual
Table 2-1. Common Combo/Meter Mode Softkeys
KEYFUNCTIONDESCRIPTION
1
2
3
3
4
2-2
Rel
Peak
Hold
rms
Average
DC &
AC rms
dB
Start relative readings in the primary display (REL is displayed). The relative reference (stored
when “Rel” is selected) appears with the softkey.
Peak Hold displays the highest and lowest peak values encountered in the present measurement
function. These values replace either the secondary reading in Meter Mode or the graphical
display in Combo Mode.
Toggle between true “rms” (“rms” annunciator on in display) and “Average” (average responding,
rms indicating - no annunciator) readings in any ac function. The rms/Average selection remains
in effect until you change the selector position. The rms/Average settings made in one display
mode are automatically made for other display modes in the same selector position. “Average” or
“rms” can be set as the standard selection on the SET UP display.
In VDC or mV DC, use the full display to show the numerical “DC” and “AC rms” values. Press
again to return to the original display.
3
Show dBm values in the primary display. The “dB” value shown above this softkey identifies the
resistance used in computing dBm; 600Ω is the standard value. You can change this value from
the SET UP display. (Rotate the selector to SET UP, change the reference selection, then rotate
the selector back to the original ac function.)
Page 28
Table 2-1. Common Combo/Meter Mode Softkeys (cont)
KEYFUNCTIONDESCRIPTION
Making Measurements
Introduction
2
5
5
5
4
Full
Auto
Hi ZEnter the high input impedance mode in mV DC. “Hi-Z” appears at the top of the display. Press
AC
DC
/
The GMM automatically selects range, trigger level, and time base (in VAC and VDC for an
optimum display. If “Manual” is selected for Range in the SET UP screen, “Full Auto” does not
come on. Selecting “Manual” from the Range Softkeys deselects “Full Auto”.
again to return to “mV”. Activating “DC & AC rms” automatically cancels “Hi-Z”. You cannot use
Peak Hold with HiZ.
Toggle between “DC” and “AC” amps or mAµA measurements. “DC” is the standard
configuration.
Activate the beeper for continuity or Manual Diode testing; a beep then sounds each time the
voltage drop passes below 0.7V (Continuity) or the logic activity indicator flashes (LOGIC
Testing). A short circuit results in a continuous tone. If the reading goes above 30 volts, a
possibly dangerous voltage exists on the inputs but is not displayed. Therefore, the safety
indicator (l) lights.
2-3
Page 29
867B/863
ACOM
V
mA
/A
EXT TRIG
1
4
3
FREEZE
DISPLAY
MODE
SAVE
PRINT
MIN
MAX
RANGE
WAKE UP
HZ
TOUCH
HOLD
125
34
V
Red
Circuit
2
Users Manual
Measuring AC Volts
Figure 2-1. AC Volts Measurements
6
Turn the Selector to Volts AC. Display mode
softkeys appear momentarily; press a
softkey to select a new mode, or wait briefly
to accept the presently highlighted selection.
7
8
Connect the test leads
Use the AC Volts Softkeys as discussed
as shown.
below.
9
At any time, press
(Combo, Meter, View, or Trend) to show the
measurement in a different display mode.
rita0550.eps
All hardkeys are available in Volts AC.
d
and one of the softkeys
2-4
Page 30
Using the AC Volts Softkeys
Combo or Meter Mode: refer to Table 2-1.
RelPeak
Hold
View Mode: refer to Chapter 5.
Time
>
>
Base
Trigger
Trend Mode: refer to Chapter 5.
Enable
RS232
Time
Longer
rms
Average
Single
Shot
Time
Shorter
>
dB
600
Glitch
Capture
Trend
Type
Ω
>
>
Restart
Trend
Full
Auto
Full
Auto
Making Measurements
Measuring AC Volts
2
2-5
Page 31
867B/863
FREEZE
DISPLAY
MODE
SAVE
PRINT
MIN
MAX
RANGE
WAKE UP
HZ
TOUCH
HOLD
125
34
V
ACOM
V
EXT TRIG
Red
Circuit
1
4
3
2
mA/
A
Users Manual
Measuring DC Volts
Figure 2-2. DC Volts Measurements
6
Turn the Selector to Volts DC. Display mode
softkeys appear momentarily; press a
softkey to select a new mode, or wait briefly
to accept the presently highlighted selection.
7
8
Connect the test leads
Use the DC Volts Softkeys as discussed
as shown.
below.
9
rita0560.eps
At any time, press
(Combo, Meter, View, or Trend) to show the
measurement in a different display mode.
All hardkeys are available in Volts DC.
d
and one of the softkeys
2-6
Page 32
Using the DC Volts or mV DC Softkeys
Combo or Meter Mode: refer to Table 2-1.
RelPeak
Hold
View Mode: refer to Chapter 5.
Time
>
>
Base
Trigger
Trend Mode: refer to Chapter 5.
Enable
RS232
Time
Longer
DC &
AC rms
Single
Shot
Time
Shorter
600
>
Trend
Type
dB
Ω
Glitch
Capture
>
Full
Auto
>
Restart
Trend
Full
Auto
Making Measurements
Measuring DC Volts
2
2-7
Page 33
867B/863
FREEZE
DISPLAY
MODE
SAVE
PRINT
MIN
MAX
RANGE
WAKE UP
HZ
TOUCH
HOLD
125
34
mV
HiZ
ACOM
V
EXT TRIG
1
4
3
Circuit
2
mA/
A
Red
Users Manual
Measuring DC Millivolts
Figure 2-3. DC Millivolt Measurements
6
Turn the Selector to mV DC. Display mode
softkeys appear momentarily; press a
softkey to select a new mode, or wait briefly
to accept the presently highlighted selection.
7
8
Connect the test leads
Use the mV DC Softkeys as discussed
as shown.
above.
rita0570.eps
9
At any time, press
All hardkeys are available in millivolts DC.
d
and one of the softkeys
(Combo, Meter, View, or Trend) to show the
measurement in a different display mode.
2-8
Page 34
Making Measurements
ACOM
V
EXT TRIG
1
4
3
FREEZE
DISPLAY
MODE
SAVE
PRINT
MIN
MAX
RANGE
WAKE UP
HZ
TOUCH
HOLD
125
34
Red
2
mA/
A
Measuring Ohms, Continuity, and Conductance
Measuring Ohms, Continuity, and
Conductance
2
Figure 2-4. Ohms, Continuity, and Conductance
6
Turn the Selector to Ohms. Softkeys for
Meter and Trend Modes appear
momentarily. Wait briefly to accept Meter
4
to select Trend Mode.
d
and one of the softkeys
Mode, or press
7
8
9
Connect the test leads as shown.
Use the Ohms (Meter or Trend Mode)
Softkeys as discussed below.
F
is not available in Ohms; all other
hardkeys can be used.
At any time, press
(Combo, Meter, View, or Trend) to show the
measurement in a different display mode.
rita0580.eps
Measurements
2-9
Page 35
867B/863
Users Manual
Using the Ohms Meter Mode Softkeys
Rel
2
(
beeps each time a low resistance continuity
measurement is encountered. For example,
beginning in the 300Ω range, the beeper goes on
when the resistance passes below 32Ω and off when
the resistance rises above 132Ω. You can select a
different range after activating the Continuity beeper;
each range uses different beeper on and off points.
For a list of these on and off points, see Continuity
Specifications in Chapter 8 of this manual.
3
(nS) begins conductance (nanosiemens “nS”)
readings in the primary display. Measurement
ranges are 300 nS and 3000 nS. (Megohm-only
readings continue in the secondary display.) Press
3
again to return to ohms (Ω) primary readings
(all ranges). Selecting conductance (nS)
automatically cancels active Min Max, “Peak Hold”,
Touch Hold, Continuity, and “Rel, and resets the
Trend plot. Frequency readings are not available in
conductance or ohms.
/
nS
/
) enables the Continuity beeper. The GMM
2-10
Page 36
Making Measurements
ACOM
V
EXT TRIG
1
4
3
FREEZE
DISPLAY
MODE
SAVE
PRINT
MIN
MAX
RANGE
WAKE UP
HZ
TOUCH
HOLD
125
34
Red
2
mA/
A
Testing Diodes and Measuring Capacitance
Testing Diodes and Measuring
Capacitance
2
Figure 2-5. Diode Test and Capacitance Measurements
6
Turn the Selector to Diode/Capacitance. The
GMM assumes Meter Mode.
7
8
Connect the test leads
Use the Diode/Capacitance Softkeys as
as shown.
discussed below. “Rel” and “/” (Continuity)
are not available in Auto Diode.
rita0590.eps
9
Pressing
beep sounds.
H
r
F
and
are not allowed in Diode Test.
is not allowed in Diode Test; a long
r is
allowed in Capacitance.
2-11
Page 37
867B/863
Users Manual
Using the Auto Diode Softkeys
Auto
Diode
4
toggles between AutoDiode and Manual Diode.
When selected, Auto Diode is highlighted and one of
the following icons is displayed:
N
k
g
u
? (
5 ( &
) is highlighted for Diode Testing.
(short)
(open)
(unknown)
Pressing this key toggles to Capacitance (|) and
disables Diode Testing.
“Rel” and “/” (Continuity) are not available in Auto
Diode.
In Manual Diode test, voltage is developed across
the components(s) by a test current (approximately 1
mA with the test leads shorted) from the meter.
Voltage is read on a 0 to +3.000V range that can
measure up to five silicon diode or transistor
junctions in series. For a silicon diode, the typical
forward voltage is about 0.6V. Voltages greater than
3.00V or open test leads produce an overload (OL)
reading. If the digital reading is the same in both
directions, the diode junction is probably shorted. If
the display reads OL in both directions, the diode
junction is probably open. To protect sensitive
devices the open test lead voltage will not exceed
3.1V. Negative inputs (from an external power
source) are not suppressed.
Auto
Diode
|
&
2-12
Page 38
Using the Capacitance Softkeys
ACOM
V
EXT TRIG
Red
1
4
3
FREEZE
DISPLAY
MODE
SAVE
PRINT
MIN
MAX
RANGE
WAKE UP
HZ
TOUCH
HOLD
125
34
A
Circuit
2
mA/
A
Making Measurements
Measuring AC and DC Amps
2
Rel
5 (|&
) toggles between Capacitance
Measurement and Diode Testing.
Measuring AC and DC Amps
6
7
8
9
At any time, press
(Combo, Meter, View, or Trend) to show the
measurement in a different display mode.
Turn the Selector to Amps. Display mode
softkeys appear momentarily; press a
softkey to select a new mode, or wait briefly
to accept the presently highlighted selection.
Connect the test leads
Use the Amps Softkeys as discussed below.
as shown.
All hardkeys are available in Amps.
d
and one of the softkeys
|
&
rita0600.eps
Figure 2-6. Amps Measurements
2-13
Page 39
867B/863
Users Manual
Using the Amps Softkeys
AC Amps (Combo/Meter Mode): refer to Table 2-1.
RelPeak
Hold
DC Amps (Combo/Meter Mode): refer to Table 2-1.
RelPeak
Hold
View Mode (AC or DC Amps): refer to Chapter 5.
Time
>
>
Base
Trigger
Trend Mode (AC or DC Amps): refer to Chapter 5.
Enable
RS232
Time
Longer
rms
AveragedB600
DC &
AC rms
Single
Shot
Time
Shorter
>
Trend
Ω
dB
600
Ω
Glitch
Capture
Type
>
>
AC
DC
AC
DC
Full
Auto
Restart
Trend
2-14
Page 40
Making Measurements
ACOM
V
EXT TRIG
Red
1
4
3
FREEZE
DISPLAY
MODE
SAVE
PRINT
MIN
MAX
RANGE
WAKE UP
HZ
TOUCH
HOLD
125
34
Circuit
2
mA
A
mA/
A
Measuring AC and DC Milliamps and Microamps
Measuring AC and DC Milliamps
and Microamps
2
Figure 2-7. Milliamp and Microamp Measurements
6
Turn the Selector to mAµA. Display mode
softkeys appear momentarily; press a
softkey to select a new mode, or wait briefly
to accept the presently highlighted selection.
7
8
Connect the test leads
Use the mAµA Softkeys as discussed below.
9
All hardkeys are available in mAµA.
At any time, press
(Combo, Meter, View, or Trend) to show the
measurement in a different display mode.
d
and one of the softkeys
rita0610.eps
as shown.
2-15
Page 41
867B/863
Users Manual
Using the mAµA Softkeys
AC mAµA (Combo/Meter Mode): refer to Table 2-1.
RelPeak
Hold
DC mAµA (Combo/Meter Mode): refer to Table 2-1.
RelPeak
Hold
rms
Average
DC &
AC rms
mA 1
µA 100
mA 1
µA 100
Ω
Ω
Ω
Ω
AC
DC
AC
DC
4
(mA 1Ω; µA 100Ω) In Combo or Meter Mode,
this softkey toggles between “mA” and “µA”
measurements. (“mA” is the standard configuration.)
Each toggle cancels Min Max, Touch Hold, “Peak
Hold”, and Rel and resets the Trend plot. “1Ω” and
“100Ω” represent the shunt resistances used for mA
and µA, respectively.
NOTE
If measuring <3 mA, using 3000.0 µA range
will give you better resolution.
View Mode (AC or DC mAµA): refer to Chapter 5.
Time
Base
>
>
Trigger
Single
Shot
>
Glitch
Capture
>
Full
Auto
Trend Mode (AC or DC mAµA): refer to Chapter 5.
Enable
RS232
Time
Longer
Time
Shorter
Trend
Type
>
Restart
Trend
2-16
The GMM autoranges in “mA” or “µA” (two ranges
are available for each selection) but does not
autorange between “mA” and “µA”. If manual range
is selected, “mA” assumes the 30 mA range, and
“µA” assumes the 300 µA range. The µA ranges are
not available on Model 863.
F
swaps the mA or µA in the primary display with
the frequency reading in the secondary display.
Subsequent changing between mA and µA or DC
and AC swaps the primary and secondary displays
back.
Page 42
Making Measurements
FREEZE
DISPLAY
MODE
SAVE
PRINT
MIN
MAX
RANGE
WAKE UP
HZ
TOUCH
HOLD
125
34
ACOM
V
EXT TRIG
Red
mA/
A
1
4
3
2
COMPONENT
TEST
Using Component Test
Using Component Test
2
Figure 2-8. Component Test
rita0630.eps
6
7
8
Turn the Selector to COMPONENT TEST.
Connect the test leads
as shown.
Use the COMPONENT TEST Softkeys as
discussed below.
9
Since COMPONENT TEST uses its own
display mode, only W,
P
, and
operate normally.
Caution
To avoid damage to the circuit,
Component Test should be used only on
unpowered circuits.
H
2-17
Page 43
867B/863
Users Manual
Using COMPONENT TEST Softkeys
2
Hz
Use COMPONENT TEST to measure the
characteristics of a passive component (in or out of
circuit, no power applied). When you use the
General Measurement Connection, the GMM
provides an ac stimulus signal to the component and
plots the voltage drop against the current
measurement. The resulting pattern provides
information about passive circuit conditions.
Generally, resistors exhibit some form of straight or
angled line. Diodes or transistors appear as straight
lines with right-angle ends. Frequency-sensitive
components (inductors, capacitors) portray some
form of an ellipsoid pattern based on the currentvoltage phase shift dictated by the stimulus
frequency and the inductance or capacitance.
Combinations of these components exhibit hybrid
patterns. Although the number of such
COMPONENT TEST signatures can be limitless,
you will become familiar with a few characteristic
signature patterns that will greatly aid in component
20
Hz
200
Hz
2
kHz
18.75
kHz
and circuit diagnosis. Some sample signature
patterns are shown on the following page.
Press one of the softkeys to select the frequency of
the stimulus signal. For components that do not
change impedance with frequency (resistors, diodes,
and transistors), the frequency selection is not
critical for a meaningful display. For components that
do change impedance with frequency (inductors and
capacitors), an appropriate COMPONENT TEST
frequency selection is necessary. Inductance greater
than 8.5 mH can be displayed. Capacitance can be
displayed according to the guidelines in Table 2-2.
Table 2-2. Component Test (Capacitance)
FrequencyCapacitance
2 Hz0.72 µF to 72 µF
20 Hz0.072 µF to 7.2 µF
200 Hz7200 pF to 0.72 µF
2 kHz720 pF to 0.072 µF
18.75 kHz77 pF to 7700 pF
2-18
Page 44
Making Measurements
Using Component Test
2
Figure 2-9. Component Test Patterns
rita0800.eps
2-19
Page 45
867B/863
ACOM
V
EXT TRIG
WAKE UP
LOGIC
1
FREEZE
DISPLAY
MODE
SAVE
PRINT
MIN
MAX
RANGE
HZ
TOUCH
HOLD
125
34
4
3
2
mA/
A
Red
Circuit
Users Manual
Testing for Logic Activity
(not available on Model 863)
2-20
Figure 2-10. LOGIC Testing
rita0620.eps
6
7
8
9
Turn the Selector to LOGIC.
Connect the test leads
as shown.
Use the LOGIC Testing Softkeys as shown.
The following hardkeys can be used
:
d P M
For the logic activity indicator, [ represents a logic
high, and ] represents a logic low. A horizontal bar
represents an indeterminate condition (no high or
low detected since the previous reading.) The logic
indicator flashes at a 10 Hz rate for activity faster
than 10 Hz or at the actual rate if activity is slower
than 10 Hz.
The actual logic frequency is shown in the primary
display. Frequencies to 10 MHz and higher can be
measured. DC Volts levels are shown in the
secondary display.
Page 46
Using the Logic Test Softkeys
TTL
3 V
CMOS
5 V
CMOS
/
Making Measurements
Testing for Logic Activity
2
1, 2
logic activity indicator and frequency reading. “5V
CMOS” is the standard configuration. A new
selection cancels the old selection. See Chapter 8
for threshold values.
4 (/
to indicate logic activity. The beeper is off in the
standard configuration. If the reading goes above 30
volts, the safety indicator (l) lights.
) activates the beeper; a beep then sounds
, and
3
select the trigger point for the
2-21
Page 47
867B/863
ACOM
V
EXT TRIG
1
4
3
2
FREEZE
DISPLAY
MODE
SAVE
PRINT
MIN
MAX
RANGE
WAKE UP
HZ
TOUCH
HOLD
125
34
A
mV
V
V
HiZ
mA/
A
mA
A
Users Manual
Measuring Frequency
You can measure low frequency signals (2 Hz to 2
MHz) with the procedure shown below. For higher
frequency signals (to 10 MHz and above), use the
Logic Test function.
6
Turn the Selector to any of the indicated
positions. (Frequency can be measured in
VAC, VDC, mV DC, A, and mAµA.)
7
Connect the test leads as required for the
selected measurement position.
8
9
The frequency counter trigger levels are set to
predetermined levels while the GMM is in the single
trigger mode. In dual trigger mode, the frequency
counter trigger levels are set initially (full auto) based
Press
F
.
Use the Frequency Softkeys as discussed
below.
on a percentage of the input signal. You can change
both levels with the Trigger Softkeys in the view
Figure 2-11. Frequency Measurements
rita0640.eps
mode. Manually setting these levels may cause the
frequency counter to be unstable or not count at all.
2-22
Page 48
Making Measurements
Measuring Frequency
2
Either press the Full Auto softkey or manually adjust
the trigger levels in view mode to obtain a stable
frequency reading.
Using the Frequency Softkeys
x
y
1
selects low side
for frequency, period, duty cycle, and pulse-width
measurements. Positive is the standard selection.
2
(Duty Cycle) selects duty cycle measurements
for the pulse (negative going or positive going)
selected with
3
(Pulse Width) selects pulse-width
measurements for the pulse (negative going or
positive going) selected with
4
(Period) selects period measurements of the
full signal (negative going or positive going).
5
(Exit) returns to the Display Mode Softkeys.
Duty
Cycle
1
Pulse
WidthPeriodExit
x
or high side
.
1
.
y
pulse
Frequency Operation
The frequency measurement type (Hz, duty cycle,
pulse width, or period) normally appears in the
secondary display.
secondary display readings and accesses the
Frequency Softkeys. You can then select a new
frequency type.
F
Press
function to the primary display and move the
frequency reading back to the secondary display.
F
Press
secondary readings.
at any time to swap primary and
F
swaps primary and
a second time to restore the selector
2-23
Page 49
867B/863
Users Manual
Frequency Considerations
F interacts with other features in several ways.
Generally, features that are active in the selector
function are retained when you press F. For
example, active features in VAC, such as “Peak
Hold”, “rms” or “Average”, or “dB” would stay
activated.
If you press any other key while the Frequency
Softkeys are active, the Frequency Softkeys are
deactivated and the frequency type is retained.
F cancels Min Max, Touch Hold, “Rel”, and the
Continuity functions and resets the plot in Trend
Display Mode.
Turning the selector to another position resets the
frequency measurement display to “Hz” in the
secondary display. Duty cycle, pulse width, and
period selections are not retained.
If you press F while the selector is set to Ohms,
Conductance, Diode Test, COMPONENT TEST,
LOGIC, or SET UP, the GMM emits a long beep and
ignores the key press.
The frequency measurement type (Hz, duty cycle,
pulse width, period) does not change when you
change display mode; the type remains in the
primary or secondary display as last selected.
In Amps functions, moving between “AC” and “DC”
or “mA” and “µA” swaps the frequency reading back
to the secondary display.
Pressing F cancels Touch Hold. While F is
active, you can press H to activate Touch Hold,
which then works normally with the frequency type
shown in the primary display.
Pressing F cancels Min Max. Pressing M when
F is active accesses the Min Max Softkeys but
retains frequency as the primary function.
If you press F when P is active, the GMM
beeps and ignores the key press. While F is
active, pressing P accesses the Print/Save
Softkeys, retaining the frequency primary display
selection.
2-24
Page 50
Chapter 3
Some General Descriptions
Introduction
This chapter adds some details about GMM operation and
measurement connections.
Battery Considerations
New alkaline (AA) cells provide a minimum of 4 hours
operation. A fully charged NiCd battery pack (BP7217)
typically provides 8 hours operation. Connect the battery
eliminator to recharge the NiCd battery pack for the 867B.
Fully charge the battery pack before using the 867B.
External charging (use Battery Charger BC-7210 - not
provided) must be used to charge a NiCd battery pack for
Model 863. Refer to Chapter 7 (User Maintenance). Model
867B uses an internal two-state charger. The initial
charge state is at approximately 170 mA (±30 mA) to
allow for full overnight charging (16 hours minimum). The
second charge state is at approximately 40 mA (±15 mA)
to allow for battery charge maintenance without full
charging from a discharged condition.
Display Blanking (Sleep Mode)
During battery operation the display will blank after about
20 minutes if no controls have been changed. Press the
wake up key once to restore the display.
If the battery eliminator is connected, these battery saving
features are not used. If a time-dependent measurement
(Min Max, Trend, etc.) is in progress during battery
operation, the backlight will go off after 20 minutes, but
the Model 863 display will not blank until a low battery
condition is reached. When the display blanks, no
measurements are taken.
3-1
Page 51
867B/863
Users Manual
Using the Backlight (Model 867B)
If the display is already on, pressing W toggles the
backlight power level.
NOTE
If the display is blank and cannot be recovered,
contrast may be off the scale. Restore contrast to
mid-scale with the following shortcut: turn the
selector to SET UP and avoid touching any key
until the display reappears (about 15 seconds).
1
Press
selector to the desired function.
to store this setting, then turn the
Adjusting Contrast
Adjust contrast as follows:
1. Rotate the selector to SET UP.
2. With “LCD Contrast” highlighted, press 2 or 3
to change the contrast. This setting becomes
effective immediately.
3. If you want to make this setting part of the power-up
defaults, press 5. The GMM then assumes this
contrast setting each time you turn it on.
4. Rotate the selector to the desired measurement or
test function.
When the GMM is left idle in Setup mode for 12 seconds,
the contrast will return to the LCD’s default setting.
General Features
Refer to Figure 3-1 and Table 3-1 for an overall
description of the controls, displays, and connections.
3-2
Page 52
Some General Descriptions
A
COM
V
mA /
A
EXT TRIG
COMPONENT
TEST
CAT
LOGIC
SEE WARNINGS ON BACK
A
COMPONENT
TEST
SET UP
mV
V
V
OFF
LOGIC
HiZ
10A MAX FUSED
1000V MAX
1000V MAX
ALL INPUTS
1000V MAX
FREEZE
DISPLAY
MODE
SAVE
PRINT
MIN
MAX
RANGE
WAKE UP
HZ
TOUCH
HOLD
12345
320mA MAX FUSED
12
11
10
9
2
3
4
5
7
1
6
13
14
15
8
mA
A
General Features
3
Figure 3-1. General Features
3-3
Page 53
867B/863
Users Manual
Table 3-1. General Features
NumberExampleNameDescription
1Combo
Meter
View
Trend
2RelSoftkey
3
4
5
1
r
M
3-4
Display
Labels
Softkeys
Range
Key
Min Max
Key
Combo, Meter, View, and Trend are the standard measurement display modes
for most functions. Auto Diode Test, LOGIC Test, and COMPONENT Test use
unique display modes. Refer to Figure 3-3 for descriptions of display features.
Labels define present uses for associated softkeys. A highlighted label means
that item is presently selected. Labels change with rotary selector positions and
key presses. If a small
an additional set of softkey labels (a submenu) by pressing this softkey.
The five blue softkeys serve different functions, depending on the rotary
selector position and other keys in effect.
Press at any time to change the range of a meter reading or the amplitude of a
waveform. Select “Auto” (the standard selection) for an optimal display or
“Manual” for selection of a specific range. Present range information is shown in
the lower right corner of the display. Refer to Range Softkeys in Chapter 4.
Press once to enable a display showing minimum, average, and maximum
values. Refer to Min Max Softkeys in Chapter 4.
symbol appears with a softkey label, you can access
>
Page 54
Table 3-1. General Features (cont)
T
NumberExampleNameDescription
Some General Descriptions
General Features
3
6
7
8
9
10
H
V
C
A
a
P
W
Touch Hold
Key
Rotary
Selector
Input
Connections
Save/Print
Key
Wake Up
Key
Press once to enable Touch Hold. The GMM test tool beeps and
automatically updates the display each time a change in the reading is
detected. If the input connections are withdrawn, the last reading (or
waveform) is held in the display. Toggles Touch Hold on and off.
Rotate the selector to any of the measurement or test positions. The input
connections screen appears briefly, when first turned on or when selecting
a function requiring an input connection change, then softkey labels
specific to the selector position appear.
Use
conductance, diode test, capacitance, COMPONENT TEST). Use A (or
mAµA) and COM for amps measurements. Use EXT TRIG and COM for
LOGIC Test or as an external trigger for waveform display.
Freeze the present reading (reverse contrast). You can then save or recall
measurement values and configurations or print measurement information.
In Model 867B, you can also save or recall waveforms and configurations.
Refer to Chapter 6.
Press to restore a blanked display; the display blanks after 20 minutes if no
controls are changed. The display will not blank if the battery eliminator is
connected. For Model 867B, this key toggles the backlight power level
when the GMM test tool is already awake.
and COM for most measurements and tests (volts, ohms,
V
3-5
Page 55
867B/863
Users Manual
Table 3-1. General Features (cont)
NumberExampleNameDescription
3-6
11
12
13
14
15
F
d
B
Frequency
Key
Display
Mode Key
RS-232
Optical Port
Battery
Eliminator
Connection
Calibration
Seal
Swaps primary and secondary displays. The frequency measurement type
(usually Hz) now occupies the primary display and the Frequency Softkeys
are active. You can now select a different frequency type with the softkeys.
Press F again to swap the primary and secondary displays back,
retaining the most recent frequency type.
Press to access the Display Mode Softkeys at any time. For volts and
amps functions, you can now select Combo, Meter, View, or Trend. Ohms
offers Meter or Trend. Other selector functions offer only Meter, while
LOGIC, COMPONENT Test, and SET UP use their own unique display
modes.
Connect the RS-232 serial interface cable here for communications with a
PC or printer.
Connect the battery eliminator for ac operation. A NiCd battery pack
installed in Model 867B charges whenever the battery eliminator is
connected.
Do not open this protective seal. Refer to the Service Manual.
Page 56
Understanding Display Features
125
34
ViewMeterComboTrendExit
Figure 3-3 illustrates typical display features. You will
encounter Primary and Secondary Displays and the
Graphical Window in Combo, Meter, and Trend Display
Modes. Other display features will appear in all display
modes.
Input Connections Screen
When you select a new function, the Input Connections
Screen (Figure 3-2) appears momentarily to identify
measurement connections to use, and displays modes
that are available. (You can press
3-second pause to proceed immediately to the function
display.) See Table 3-1 under Input Connections.
“Exit” during this
5
Some General Descriptions
Understanding Display Features
Figure 3-2. Input Connections Screen
3
rita0601.eps
Primary Display
Large digits represent the measurement or test based on
the selector position. This display always appears in
Meter, Combo, and Trend modes. The frequency reading
(usually Hz) can be swapped into the primary display.
When selected, “Rel” or “dB” automatically occupy the
primary display.
Secondary Display
Smaller digits show the type of frequency reading (usually
Hz) selected with the Frequency Softkeys. The primary
reading can also be swapped into this display.
Graphical Window
This window shows a waveform in Combo Mode, an
analog display in Meter Mode, or a plot in TrendGraph
Mode.
3-7
Page 57
867B/863
Window
Display
Secondary Display
Primary Display
Status Line (Lower)
Softkey Labels
Status Line (Upper)
Users Manual
Softkey Labels
The present uses for the five softkeys appear immediately
below the display.
Upper Status Line
The top display line identifies active operations as shown
in Table 3-2.
Lower Status Line
The bottom display line (immediately above the softkey
labels) identifies the display mode, function, and range.
safety indicator also appears here if dangerous
The
l
voltage is detected at the input terminals when the actual
reading is not displayed.
Message Boxes
Message boxes appear in the lower display when the
GMM test tool is performing an operation that you may not
want to interrupt (“Saving” or “Printing”).
Overload Indication (OL)
“OL” or “-OL” appears in the primary and secondary
displays when an input has exceeded the present range
(in Manual) or has exceeded the highest range available
in this function (in Auto). A clipped waveform can appear
in Combo or View mode (DC Volts, mV DC, DC Amps, DC
mAµA functions only) if a large ac signal is riding on an
overloaded dc signal.
Figure 3-3. Display Features
rita0390.eps
3-8
Page 58
Table 3-2. Display Status Information
SYMBOLDESCRIPTION
Some General Descriptions
Understanding Display Features
3
h
REL“Rel” (relative reading) is active
Min MaxMin Max is active. This is the only
RS232The serial interface port is enabled.
Hi-ZHi input impedance selected.
b
B
UncalGMM not calibrated. Contact a Fluke
InternalInternal trigger (or External.)
l
Touch Hold is active.
(reference minus actual reading.)
indication if Min Max is running in the
background.
Battery is low.
Line power operation.
Service Center.
Dangerous voltage is present at the
input terminals when the actual reading
is not displayed.
3-9
Page 59
867B/863
ACOM
V
EXT TRIG
Red
Circuit
mA/
A
ACOM
V
EXT TRIG
Red
Circuit
mA/
A
Users Manual
Measurement Connections
Warning
m
Never attempt a voltage measurement if a test
lead is in the amp (A) or milliamp microamp
(mAµA) input terminal. You might be injured
or the GMM TEST TOOL might be damaged.
General Measurement Connections
Volts, Ohms, Continuity, Conductance, Diode Test and
Capacitance connections are illustrated in Figure 3-4.
rita0384.eps
Figure 3-4. General Measurement Connections
Amps and mAµA Measurement Connections
Amps DC and AC and mAµA connections are illustrated in
Figures 3-5 and 3-6.
rita0382.eps
Figure 3-5. Amps Measurement Connections
3-10
Page 60
Some General Descriptions
ACOM
V
EXT TRIG
Red
Circuit
mA/
A
ACOM
V
EXT TRIG
Red
Circuit
mA/
A
Measurement Connections
3
Figure 3-6. Milliamp and Microamp Connections.
Logic Test Connections
LOGIC Test connections are illustrated in Figure 3-7.
rita0381.eps
rita0383.eps
Figure 3-7. LOGIC Testing Connections
Frequency Measurement Connections
You can measure frequency from 2 Hz to 2 MHz in AC
Volts, DC Volts, mV DC, Amps, and mAµA functions. For
higher frequency readings (to > 10 MHz), use the Logic
Test function.
3-11
Page 61
867B/863
Users Manual
3-12
Page 62
Chapter 4
Using the Hardkeys
Introduction
Hardkeys can be used in many function and display
mode combinations. Although hardkeys are
mentioned throughout this manual, this chapter
provides a single source of detailed descriptions for
using hardkeys and their accompanying softkeys.
Using Display Mode Softkeys
Press d at any time. The Display Mode Softkeys
appear, showing the modes available with the
present measurement function. Modes available with
VAC, VDC, mV DC, A, and mAµA are shown.
ComboMeterViewTrendExit
The GMM test tool now highlights the present
selection and allows you to do the following:
• Change display mode. Softkeys for the new
display mode now appear.
• Return without changing the display mode.
Press d again, or press the softkey for the
selected display mode, or press 5 to exit.
Combo (the default mode set at the factory) appears
automatically when the selector is set to VAC, VDC,
mV DC, Amps, or mAµA. You can also select Meter
as the default for these functions on the SET UP
screen. Meter is the only mode available for Diode
and Capacitance Test. Meter and Trend are the
4-1
Page 63
867B/863
Users Manual
choices for Ohms and Conductance. LOGIC TEST,
COMPONENT TEST, and SET UP each use a
single, unique mode.
The Display Mode Softkeys also appear briefly when
you rotate the selector to a new function. If you
make no mode selection within 2 seconds, the
softkeys shift automatically to the appropriate
function labels.
Using Frequency Softkeys
Softkeys accessed by pressing F are:
x
y
1 selects the low side x or high side y
pulse for frequency, period, duty cycle, and pulse
width measurements. Positive is the standard
selection.
2 (Duty Cycle) selects duty cycle measurements
for the pulse (negative going or positive going)
selected with 1.
3 (Pulse Width) selects pulse-width
measurements for the pulse (negative going or
positive going) selected with 1.
Duty
Cycle
Pulse
WidthPeriodExit
4 (Period) selects period measurements of the
full signal (negative going or positive going).
5 (Exit) returns to the Display Mode Softkeys.
Frequency Operation
The frequency measurement type (Hz, duty cycle,
pulse width, or period) normally appears in the
secondary display. F swaps primary and
secondary displays and accesses the Frequency
Softkeys, which you can use to select a new
frequency type.
Press F a second time to swap primary and
secondary displays back. The secondary reading
then continues to display the frequency
measurement type selected with the Frequency
Softkeys. The Frequency Softkeys remain active,
and you can change the frequency type shown in the
secondary display while these keys remain
displayed.
Press F at any time to swap primary and
secondary readings.
4-2
Page 64
Using the Hardkeys
Save/Print Softkeys
4
Save/Print Softkeys
Softkeys accessed by pressing
Save
Screen
1
(Save Screen) accesses the Save Screen
softkeys, which allow you to save the present display
to a memory location. (Model 867B only.)
2
(Recall) accesses the Recall softkeys, which
allow you to display the contents of a memory
location.
3
(Print) prints the display. “Freeze” changes to
“Printing”, then reverts to “Freeze” when printing is
complete. You can stop printing by pressing
(Stop Print) or
ignored.
4
(Save Config.) accesses the Save Config
softkeys.
5
(Exit) returns to the original display. You can
also press
display. You must press one of these two keys to
>
RecallPrint
5
P
>
(Exit); all other key presses are
again to return to the original
P
are:
Save
>
Config.Exit
3
return; other keys do not operate during the Freeze
display.
P
Press
updates. Display contrast is reversed, the “Freeze”
message appears, and the Save/Print Softkeys are
accessed. You can now Print display information or
Save and Recall display and configuration
information.
“Enable RS232” in TrendGraph Mode is canceled.
Touch Hold, Min Max, Trend, and “Peak Hold”
continue to update in the background (beeps
continue).
Refer to Chapter 6 for a full description of saving and
recalling operations available with Freeze.
to stop waveform and numerical value
4-3
Page 65
867B/863
Users Manual
Min Max Softkeys
Softkeys accessed by pressing M are:
Min Max
OffRestart
Minimum, maximum, and average measurements
are displayed (with timer information) in the graphical
display (Combo and Trend) or meter display (Meter).
In the View mode, Min Max is enabled and the MIN MAX icon appears in the upper status line. However,
Min Max measurement readings and timer
information cannot be viewed on the waveform
screen.
1 (Min Max Off) turns the Min Max feature off.
The GMM test tool returns to its normal display
mode. Measurement values and timer information
cannot be recovered at this point.
2 (Restart) resets all values/timers and begins
with new values and timers. Previous measurement
values and timer information cannot be recovered at
this point.
Run
PauseExit
3 (Run Pause) toggles between the normal “Run”
state and a temporary “Pause” condition. Changing
display modes when Min Max is active does not
change the “Run” or “Pause” state.
In Combo, Meter, and Trend modes, “Pause” turns
the “Min Max” icon at the top of the display dark, and
displayed values stop updating (although the
elapsed time continues counting.) In View Mode,
“Run” and “Pause” operate normally, but
measurement readings and timer information cannot
be viewed.
5 (Exit) returns the softkeys to the functions for
the selected measurement mode while the
remainder of the display remains unchanged. You
must use the Display mode key to return the window
and secondary displays to the Combo, Meter, or
Trend display. Min Max will remain enabled in the
background mode.
4-4
Page 66
Using the Hardkeys
Range Softkeys
4
You can use Min Max with LOGIC to keep track of
the range of logic levels being encountered. Use the
following procedure:
1. With the selector turned to LOGIC, select logic
type by pressing
2. Press
3. Place the Min Max display in the background by
4. Check the Min Max values at any time by
Cancel Min Max by reselecting or changing the logic
type. (Press
Softkeys.)
M
and times now appears.
pressing
reappears, and the Min Max icon indicates that
Min Max recording is continuing in the
background.
repeating steps 2 and 3.
d 1
1, 2
1, 2
. The full Min Max display of levels
. The LOGIC activity display
3
, or
3
, or
from the LOGIC
.
Range Softkeys
Softkeys accessed by pressing
Auto
Manual
1
(Auto Manual) toggles between Auto and
Manual ranging. In Auto, the GMM test tool
automatically selects the range presenting the
highest resolution. Auto is the standard
configuration; Manual can be selected as the powerup configuration when the selector is in the SET UP
position. The GMM test tool resets to Autoranging
when you select “Full Auto” (by pressing
the VAC, VDC, or mV DC softkeys.)
2
and
(if available). If Auto ranging is active, [ or ] selects
Manual ranging and the next higher (lower) range.
[]
3
select the next higher (lower) range
r
are:
Wave
ScaleExit
5
from
4-5
Page 67
867B/863
Users Manual
4 (Wave Scale - active in the View Display
Mode) enters manual ranging and cycles through
“units/Div” scaling factors of 1, 2, and 4 for the
vertical axis, effectively zooming in and out on the
waveform display. The scaling factors divide the
units/Div for the present range. “Wave Scale” is not
available in Meter Mode.
5 (Exit) returns to the softkeys previously in
effect. (You can also press r to return to the
previous softkeys.)
Entering Range cancels “Peak Hold”.
Touch Hold
This key allows you to hold a known good reading on
the display. (No special softkeys accompany Touch
Hold.) Use the following procedure:
1. Press H (
appears at top of the display.)
h
2. Apply measurement connections until you hear a
beep. This sound signifies that a stable reading
has been made. The display continues to update
with valid readings (and beeps) as long as you
maintain the measurement connections.
3. In Meter, Combo, and Trend display modes, the
last valid primary numeric reading is held; the
waveform in Combo and Trend is not held. In
View Mode and COMPONENT Test, the last
valid waveform is held.
Press H again to exit Touch Hold (
disappears.)
Touch Hold and “Peak Hold” are mutually exclusive:
activating one cancels the other. Touch Hold cannot
be activated when Frequency, SET UP, or LOGIC is
selected. Pressing F or d cancels Touch Hold.
Selecting “rms”, “Average”, or “Full Auto” cancels
Touch Hold. P stops Touch Hold display updates;
Touch Hold then continues to run in the background.
In view mode, selecting Touch Hold cancels “Glitch
Capture” and “Single Shot”.
h
4-6
Page 68
Chapter 5
Using View and Trend Display Modes
Introduction
This chapter describes how to display a waveform in
View Display Mode and how to plot and interpret
measurement data over time in TrendGraph
Display Mode.
View Display Mode Basics
The GMM must know when to begin displaying
(trigger) a waveform. Triggering requires that a valid
signal level is specified, a direction for that signal
(positive going or negative going) is selected, and
the trigger signal source (internal or external) is set.
Setting the range for the vertical divisions and the
timebase for the horizontal divisions directly
influences the proportions and legibility of the
triggered waveform.
The type of waveform acquisition may also be
important. A continuous, internal trigger enables an
ongoing display of the input signal whenever you
turn the selector to a new function.
Using Full Auto
“Full Auto” allows the GMM to automatically select
the Trigger Level, Trigger Slope, Trigger Source,
Range, and Time Base necessary for an optimum
display of the input signal. If a good input is present,
you will see a waveform that nearly fills the screen.
5-1
Page 69
867B/863
Users Manual
“Full Auto” (5) is enabled (highlighted) initially
when View Display Mode is activated if “Auto” is
selected in the SET UP screen and no manual
settings have been introduced.
“Full Auto” may not be automatically selected when
the GMM enters View Mode. If “Manual” is set as the
power-up default for Range on the SET UP screen,
“Full Auto” is not selected. Further, if you then
change to “Full Auto” by pressing 5 from the
View Mode Softkeys, “Manual” remains selected on
the SET UP screen and will be reapplied when you
change functions.
NOTE
Disabling any of the auto features turns the
“Full Auto” softkey label off (not highlighted.)
Other auto features then remain active.
Pressing
5
from the View Mode Softkeys
turns all auto features off at once.
Changing the Amplitude Range
You can use the softkeys accessed with r to
select a preset (Manual) amplitude range.
Using View Mode Softkeys
Time
>
>
Base
Trigger
1 (Time Base) accesses the Time Base
Softkeys, which allow you to zoom in or out on the
waveform by changing the horizontal (time) axis of
the View display.
2 (Trigger) accesses the Trigger Softkeys,
allowing you to control the trigger slope, level, and
source.
3 (Single Shot) accesses the Single Shot
Softkeys, which allow you to capture and examine a
single waveform that has met your trigger criteria.
The GMM test tool displays continuously updating
waveforms if this key is not pressed.
4 (Glitch Capture) accesses the Glitch Capture
Softkeys, allowing you to capture and examine a
waveform that has met a predetermined trigger
criteria with a briefly occurring spike. See Choosing
the Acquisition Type later in this manual for further
information on Trigger criteria for Glitch Capture.
Single
Shot
>
Glitch
Capture
>
Full
Auto
5-2
Page 70
5
5 s
4.2 µs
8.3 µs
20 µs
50 µs
100 µs
200 µs
500 µs
1 ms
2 ms5 ms
10 ms
20 ms
50 ms
100 ms
200 ms
500 ms
1 s
2 s
23
Slower
Faster
(Full Auto) selects the range, trigger level, and
time base automatically to produce an optimum
display of the waveform.
Setting Up the Time Base
Time
Base
>
>
Trigger
Single
Shot
>
Glitch
Capture
>
Full
Auto
Using View and Trend Display Modes
Setting Up the Time Base
5
Select “Time Base” (
1
) from the View Mode
Softkeys, then use the Time Base Softkeys as
follows:
SlowerFasterExit
2
(Slower) selects the next longer time per
division. More waveform activity then appears in one
time division, effectively zooming out on the
waveform.
3
division. Less waveform activity then appears in one
time division, effectively zooming in on the
waveform.
5
(Faster) selects the next shorter time per
(Exit) returns to the View display.
Figure 5-1. Time Base Selection
rita0710.eps
5-3
Page 71
867B/863
Users Manual
Setting Up the Trigger
Time
>
>
Base
Trigger
Select “Trigger” (
Softkeys, then use the Trigger Softkeys as shown in
the following paragraphs.
Selecting the Trigger Slope
Slope
K
Slope
t
1
(Slope) sets the waveform trigger to trip either
on the rising or falling edge of the signal for single
triggering. The counter triggering still use two (dual)
triggers for its operation and are set automatically
based on the input signal amplitude. In dual
triggering,
falling slope trigger points to allow their levels to be
set individually. The waveform trigger icon will switch
between two trigger points in the display as
selects each trigger. The rising slope setting will
always have a more positive setting than the falling
slope setting. However, whichever slope is left
selected will be the last criteria to be met before a
[]
1
Single
Shot
2
) from the View Mode
Glitch
>
Capture
Internal
ExternalExit
>
selects between the rising and
1
Full
Auto
trigger occurs, e.g., if the rising slope selection is left
highlighted, the input signal must pass through the
falling slope first and then through the rising slope to
trigger the GMM. In single trigger, Full Auto will
always set the trigger slope to rising.
The GMM trigger circuitry controls both the
waveform and counter displays. The counter always
uses a dual trigger setting while the waveform
display can be controlled by single or dual triggering.
The waveform trigger setting(s) are indicated in the
display with an icon in the left margin along the
vertical axis of the display grid (Figure 5-2). The
counter trigger settings are indicated by the two
icons along the right vertical axis and are marked
with “Hz”. The counter and waveform trigger settings
are the same when in Dual trigger operation. This is
the only condition where the counter trigger levels
are manually set.
5-4
Page 72
Using View and Trend Display Modes
Setting Up the Trigger
Changing the Trigger Level
Slope
K
Slope
t
[]
Internal
ExternalExit
5
Figure 5-2. Trigger Icons
86xsc011.tif
2
the waveform must pass through to trigger the GMM.
For Internal triggering, the waveform trigger icon
identifies the level and moves as you make the level
adjustment. The actual trigger level is displayed in
the upper right corner of the display. For Dual
triggering, you must first select which trigger point
you want to adjust with the
The waveform icon will move from one point to the
other in the display as the slope selection is
changed. The counter trigger levels are linked with
the waveform trigger levels in dual triggering and are
the only time they can be manually set. For external
triggering, a single trigger level is set with the (
and ]) softkeys and is indicated in the special
trigger level indicator in the upper right corner of the
display. There are 21 different preset trigger levels
available for external triggering.
If “Full Auto” is selected, the trigger levels are set on
percentages of the input signal.
and
3 ([
and ]) change the trigger level
1
Slope selection.
[
5-5
Page 73
867B/863
Users Manual
Selecting the Trigger Source
Slope
K
Slope
t
4
(Internal External) selects Internal or External
[]
trigger source. When “Internal” is selected, the
trigger circuit uses the measured input signal for
triggering. When “External” is selected, the trigger
circuit uses the signal placed on the EXT TRIG jack
for triggering. The waveform and counter trigger
icons are removed from the display and a special
trigger level indicator replaces the trigger level
readings in the upper right corner of the display
(Figure 5-3). There are 21 external trigger level
steps. You can provide a trigger from an external
source to synchronize the GMM display to some
other event. Refer to Specifications for External
trigger input levels.
5
(Exit) returns to the View Softkeys.
Internal
ExternalExit
Selecting Single or Dual Trigger
To accommodate various trigger situations, the
waveform can be triggered by one (single) or two
(dual) trigger settings. “Single” or “Dual” are trigger
selections you can be make from the SET UP
screen. When using external triggering, only one
trigger level may be set regardless of the set up
selection.
rita0392.eps
Figure 5-3. Setting Up the Trigger
5-6
Page 74
Using View and Trend Display Modes
Choosing the Acquisition Type
5
Single triggering establishes a fixed trigger point for
the waveform display. The input signal only needs to
cross the trigger amplitude set by the trigger level
and in the direction set by the slope setting. While in
single trigger mode, the frequency counter trigger
circuit sets two predetermined levels that are not
adjustable. This ensures a stable frequency reading
while triggering the waveform with a single trigger.
Dual triggering (the default setting) requires that the
signal pass through both upper and lower trigger
points to trigger the waveform and frequency
counter. These points are initially set based on
percentages of the input signal amplitude. You can
change both levels with the Trigger Softkeys. Dual
triggering provides a more stable waveform display
in noisy environments.
Choosing the Acquisition Type
Time
>
>
Base
Instead of using the continuous internal trigger, you
can instruct the GMM to update the screen only
once, yielding a retained, snapshot view (“Single
Shot”) of an input signal meeting the trigger
requirements. Press
Softkeys.
Yet another selection allows you to display and
monitor only randomly-occurring or fast transient
events or glitches (“Glitch Capture”). Trigger levels
are preset and are not ajustable. The “Glitch” must
have a change rate higher that 1 kHz and an
amplitude > 20% of the selected GMM range.
“Single Shot” and “Glitch Capture” use the same
submenu of softkeys as described in the following
paragraphs.
Trigger
Single
Shot
3
from the View Mode
>
Glitch
Capture
>
Full
Auto
5-7
Page 75
867B/863
Users Manual
Using Single Shot and Glitch Capture
Softkeys
Arm
1
(Arm) resets the trigger circuit and readies the
GMM test tool to capture another single shot
waveform or glitch capture transient. “Arm” is
selected (highlighted) initially. When the entire
waveform is captured, “Arm” reverts to normal
contrast.
The display shows elapsed time (hh:mm) since
arming or, if a trigger subsequently occurred, time
between arm and trigger. The time cycles from 00:00
through 99:59 continuously until a trigger is received.
<>
Exit
2
or right for viewing (available once the waveform is
completely captured). If you press
arm cycle starts over, < and > disappear, and the
arm-to-trigger time begins at 00:00. Waveform
position accomplished with < and > is retained
across arm cycles. Six time base divisions prior to
the trigger and 19 time base divisions after the
trigger can be viewed using < and >, respectively.
Refer to Figure 5-4 for an overview of time divisions
(on and off screen) that can be examined.
5
Free Run operation and autoranging are not
available during “Single Shot” and “Glitch Capture”.
Selecting Touch Hold cancels “Single Shot” and
“Glicth Capture”.
3
and
(Exit) returns to the View Softkeys.
(< and >) move the waveform left
1
again, the
5-8
Page 76
Using View and Trend Display Modes
619
Choosing the Acquisition Type
5
Figure 5-4. Viewable Time Divisions (Single Shot and Glitch Capture)
rita0780.eps
5-9
Page 77
867B/863
Users Manual
Trend Display Mode Basics
Trend can be used in VAC, VDC, mV DC, Ohms,
Amps, and mAµA functions. This mode uses the
graphical window to plot 120 data points over a
selectable time ranging from 1 second to 15 minutes.
Each data point can represent an actual reading, an
average of readings, or a pair of high and low
readings. You can also enable serial output of the
data points as they occur.
Using Trend Mode Softkeys
Enable
RS232
1
(Enable RS232) starts transmitting readings
(with time stamps) with the first press and stops
transmitting with the second press. (You can also
F
press
(hours:min:sec) resets each time RS232 is enabled.
The time rolls over at 24 hours. Refer to Chapter 6
for serial output connection descriptions.
2
or
the interval to use on the horizontal scale as data
Time
Longer
Time
Shorter
Trend
Type
>
to stop transmission.) The time stamp
3
(Time Longer, Time Shorter) selects
Restart
Trend
points are plotted on the trend graph. (Seconds: 1, 2,
5, 10, 15, 30, 45; Minutes: 1, 2, 5, 10, 15)
For “Sampled” and “Average” Trend Type selections,
the GMM plots the most recent 120 data points. For
“High Low” Trend Type, 120 data point pairs are
plotted. Earlier data points are not stored.
4
(Trend Type) accesses the Trend Type
softkeys, allowing you to plot average, sampled, or
high/low excursions of the measurement value.
5
(Restart Trend) starts plotting a new trend
graph as new samples are acquired. Primary
readings are displayed and plotted; there are no
secondary readings in Trend Mode.
Trend Mode Considerations
Only the value being trended is shown in the primary
display. For example, a “dB” or “Rel” value that
normally occupied the primary display in DC Volts
function will be replaced with a VDC reading.
Frequency values do remain in the primary display
during Trend.
The vertical scale in the Trend display depends on
the range setting and the type of measurement. A
5-10
Page 78
Using View and Trend Display Modes
Using Trend Mode Softkeys
5
unipolar arrangement (0 to full scale) is used for
VAC, AAC, and frequency. Bipolar readings
(negative to zero to positive) are used for VDC and
Ohms.
RS232 Trend Output Format
Trend data for each data point is sent out as it is
measured. ASCII format is used, as follows:
•
For “Sampled” or “Average” Trend Type:
[Value] [Units] [Time Stamp] <cr> <lf>
For “Glitch” Trend Type (Ohms only), the
following line is added at the end of each
“Sampled” or “Average” line or pair of lines:
[Glitch State] [Time Stamp] <cr> <lf>
[Time Stamp] is in hh:mm:ss format, representing
time since the Trend start. Time rolls over at 24
hours.
Selecting the Trend Type
High
AverageSampled
1
(Average) plots the average of all samples
taken by the GMM during the plot interval.
2
(Sampled) plots the last sample taken by the
GMM at the time the data point is plotted.
3
(High Low) plots the maximum and minimum
sample value taken during the time interval. A
vertical bar where the bottom end represents the
lowest sample and the top end represents the
highest sample is placed on the plot.
4
(Normal Glitch) selects “Normal” or “Glitch” in
Ohms function only. Glitch setting display and
highlight fast-occurring discontinuities on the trend
line. Markers along the bottom of the display identify
discontinuities as they appear on the Trend Plot. A
discontinuity is recorded when the measured
resistance rapidly passes (decreasing or increasing)
through a low resistance continuity point. Each range
uses a different continuity point (See Continuity
Beeper Values in Chapter 8 of this manual). Note:
When Glitch is first enabled, the GMM sets manual
Low
Normal
GlitchExit
5-11
Page 79
867B/863
Users Manual
ranging and sets the range to 300Ω. It may by
necessary to manually change the range for an onscale reading.
5 (Exit) returns to the Trend Softkeys.
5-12
Page 80
Chapter 6
Using Save, Recall, Print, and Set Up
Introduction
This chapter contains instructions for using the
Save/Print key to save, recall, and print GMM test
tool display information and the SET UP selector
position to change the GMM test tool power-up
configuration.
Refer to Figure 6-1 for an overview of operations
accessible through the Save/Print key.
For Model 867B, refer to Figure 6-2 for instructions
on connecting a printer or PC to the GMM test tool.
Using the Save/Print Softkeys
Pressing P stops waveform and numerical value
updates, freezing the screen. Display contrast is
reversed, the “Freeze” message appears, and the
Save/Print Softkeys are accessed. You can now
Print display information or save and recall display
and configuration information.
If on, the “Enable RS232” feature (used in Trend
Mode) is canceled. Otherwise, active Touch Hold,
Min Max, Trend, and “Peak Hold” features continue
to update in the background (beeps continue).
6-1
Page 81
867B/863
Save
PrintExit
ClearExitSave
Load
ScreenExit
ClearExit
Load
Config.ExitPrint
Stop
Print
Stop
Print
Users Manual
6-2
Figure 6-1. Freeze (Print/Save) Softkey Map
rita0370.eps
Page 82
Using Save, Recall, Print, and Set Up
Saving Screen and Configuration
6
P
Save
Screen
1
(Save Screen) accesses the Save Screen
softkeys, which allow you to save the present display
to a memory location. (Model 867B only.)
2
(Recall) accesses the Recall softkeys, which
allow display of a memory location.
3
(Print) prints the display. “Freeze” changes to
“Printing”, then reverts to “Freeze” when printing is
complete. You can stop printing by pressing
(Stop Print) or
4
(Save Config.) accesses the Save Config.
softkeys.
5
(Exit) (or pressing
original display. You must press one of these two
keys to return; other keys do not operate during the
Freeze display.
>
RecallPrint
5
>
(Exit); other keys are ignored.
P
Save
>
Config.Exit
again) returns to the
3
Saving Screen and Configuration
(Model 867B only)
Save
>
Screen
][
1
location. Basic information (a reading, etc.) is
displayed with the title for each memory location.
3
supporting configuration (message appears), then
returns to the Save/Print display.
4
appears only if something is saved in this memory
location.)
5
Save/Print display.
2 ([
and
(Save) saves the displayed information and
(Clear)) erases this memory location. (“Clear”
(Exit) (or pressing
SaveClearExit
and ]) select the Memory
P
again) returns to the
6-3
Page 83
867B/863
Users Manual
Recalling Screen or Configuration
>
Recall
View
>
][
1
and
2
cycle through memory locations
Memory
(seven configuration memories for all models and
three display memories on Model 867B only).
3
(View Memory) previews the selected memory
and accesses the View Memory softkeys. “View
Memory” does not appear if memory is empty.
Previewed memory always appears in reverse video.
4
(Load Screen or Load Config) directly replaces
the active display with the selected display from
memory. The selector function must match that of
the presently selected memory list item, and the
selected memory location must be occupied.
5
(Exit) returns to the Print/Save softkeys
(display frozen).
Load
ScreenExit
Previewing Screen or Configuration
>
Recall
][
1
and
2
previewing (reverse video).
3
(Print) begins printing the display. “Printing ...”
appears on the display and “Stop Print” defines the
softkey. Press
until “Print” becomes the softkey label again.
4
(Load Config. or Load Screen) can be used if
the selected memory matches the present selector
function configuration and the selected memory
location is occupied.
“Load Config” appears when you select one of the
seven configurations. The GMM loads the selected
configuration and returns to the active display.
View
>
Memory
Load
Print
Config.Exit
scroll through memory locations for
3
again to stop printing, or wait
6-4
Page 84
Using Save, Recall, Print, and Set Up
Saving the Present Configuration
6
“Load Screen” appears when you select one of the
three screens. Press
waveform along with the active waveform. The GMM
assumes the time base and amplitude settings of the
stored waveform. With the recalled and active
waveforms both displayed, you can press
access the following three softkeys:
the two-waveform display;
the recalled waveform and return to the active
waveform display;
waveform display.
If you press any other key, the recalled waveform
clears and the display returns to the active
waveform.
In View Mode, the stored waveform must have been
saved while the GMM was in View Mode. The loaded
screen appears as solid lines. In COMPONENT
TEST, the stored memory must have been saved
while the GMM was in COMPONENT TEST.
5
(Exit) returns to the Print/Save softkeys
(display frozen).
4
to display the stored
3
4
“Clear Exit” to clear
5
“Exit” to return to the two-
P
to
“Print” to
Saving the Present Configuration
Save
>
Config.
][
1
3
numerical data sufficient to recreate this display. A
message appears while the configuration is being
saved. The GMM test tool then returns to the
Save/Print display.
4
appears only if something has been saved in this
memory location.)
5
2
and
(Save) stores present configuration and
(Clear) clears this memory location. (“Clear”
(Exit) returns to the Save/Print display.
select the Memory location.
SaveClearExit
6-5
Page 85
867B/863
Users Manual
Printing
Stop
Print
Do not print if the GMM is presently
controlled by FlukeView 860 application
software.
Use the following procedure for printing:
1. Verify appropriate “Printer Type”, “Baud Rate”,
and “Parity”. These selections appear as the last
Print
NOTE
three items on the SET UP screen (selector
turned to SET UP.)
2. Connect the GMM to a printer as shown in
Figure 6-2. Or, download the screen to
FlukeView 860 Application Software running on
a PC. Then print from FlukeView 860.
3. Set the GMM to display the screen you want to
print. The screen can be the active display or
one of the recalled screens.
3
4. Press
5. You can now press
time to cancel the screen printing.
(Print) to begin printing.
3
(Stop Print) at any
6-6
Page 86
Correct
Incorrect
PAC91
PM9080
Using Save, Recall, Print, and Set Up
Printing
6
Figure 6-2. PC and Printer Connections
rita0270.eps
6-7
Page 87
867B/863
Users Manual
Changing the Configuration
U
Set upSelf
Test
(wait 2 seconds)
Next
Item
If no keys are pressed for 15 seconds after
the selector is turned to SET UP display
contrast returns to the default mid-scale
setting.
1. Beginning at the first line (LCD Contrast), press
1
<>
NOTE
to step through the other lines in this
Restore
All
Exit
Save
Set up
display. Continue pressing to wrap back to the
top line.
At each configuration line, press
make your selection. Then press
to the next line.
2. Press
3. Press
If you make SET UP changes and do not press
(Save Set up), the changes remain in effect only until
you turn the GMM off.
4
to restore all lines on this
configuration display to their standard settings.
You have two choices at this point. First, you can
5
press
information as the power-up configuration.
Second, you can use the retrieved standard
configuration as a starting point, change some of
the settings as described above, and then store
your customized settings as the new power-up
configuration.
presently shown on the display into nonvolatile
memory. These settings become effective
immediately and are reestablished each time the
GMM test tool is powered on.
now to save the displayed
5
to save all configuration settings
2
1
3
or
to move
5
to
6-8
Page 88
Using Save, Recall, Print, and Set Up
Changing the Configuration
Table 6-1. Configuration Selections (SET UP)
SET UP ITEMSELECTIONS AVAILABLEDEFAULT
LCD ContrastPress < or > to adjust display contrast.(mid-range)
6
AC Converter“rms” or “Average”. This setting controls “rms” or “Average” softkey selection
in ac volts or amps functions.
Display Mode“Combo” or “Meter”.Combo
dB ReferencedB ohms reference values: 2 4 8 16 50 75 93 110 125 135 150 300 600
900 1000 1200. Continue pressing > to select “Auto”, “Manual”, or “Disabled”.
Range“Auto” or “Manual”.Auto
Beeper“Enabled” or “Disabled”.Enabled
Sleep Mode“Enabled” or “Disabled”.Enabled
Trigger Mode“Dual” or “Single”.Dual
Printer Type“Epson”, “HP Graphics”, or “Text.”Epson
Baud Rate1200, 2400, 9600, or 19200.1200
Parity“no”, “even”, or “odd”. (Number of Bits is always 8; number of Stop Bits is
always 1.)
True rms
600
Ω
none
6-9
Page 89
867B/863
Users Manual
6-10
Page 90
Chapter 7
User Maintenance
Introduction
This chapter covers basic maintenance procedures
that can be performed by the user. For complete
service, disassembly, repair, and calibration
information, refer to the 867B/863 Service Manual
(PN 689312).
Cleaning
Clean your GMM test tool with a damp cloth and a
mild detergent. Do not use abrasives, solvents, or
alcohol.
Warning
m
To avoid ELECTRICAL shock, remove the
test leads and any input signals before
opening the case.
Testing the Fuses
Test the fuses with the following procedure:
1. Turn the selector to Ohms (
2. Connect a test lead between the
mAµA input terminals. Check that the display
reads less than 5Ω. This tests the 440 mA fuse.
If this reading is “OL”, refer to the following fuse
replacement information.
3. Move the test lead from the mAµA terminal to
the A terminal. Check that the display reads less
than 5Ω. This tests the high energy (11A) fuse. If
this reading is OL, refer to the following fuse
replacement information.
4. If either reading is outside the range shown,
have the GMM test tool serviced.
R
).
V
and
7-1
Page 91
867B/863
Users Manual
Replacing the Batteries
The GMM test tool uses 6 AA alkaline batteries
(included with Model 863) or a nickel-cadmium
battery pack (BP 7217 - included with Model 867B).
Internal charging for the battery pack is provided with
Model 867B. The NiCd battery pack must be fully
charged before using the GMM. External charging
(using BC7210) is necessary if the battery pack is
used with Model 863 (instructions are included with
BC7210.) Refer to Figure 7-2 for battery replacement
instructions. Replace the NiCd battery pack when it
no longer holds a charge for the rated interval. Order
replacement Model BP7217 Nickel-Cadmium Battery
Pack. In the USA and Canada, call Fluke Service
Parts at (800) 526-4731. Elsewhere, call +1 425356-5500 for instructions.
n
Note
If the Tester is dropped, it may not turn on.
Check the batteries to see if they are
damaged and need to be replaced.
Do not mix spent Nickel-Cadmium batteries with the
solid waste stream. Spent batteries should be
disposed of by a qualified recycler or hazardous
materials handler. Contact your authorized Fluke
Service Center for recycling information.
Replacing the 440 mA Fuses
m
Warning
Refer to SAFETY in the front of this
manual prior to replacing any fuse.
Refer to Figure 7-3 to locate and replace the 440
mA fuse(PN 934121). Note that the battery
compartment must be accessed first; refer to Figure
7-2 for these initial access procedures.
Replacing the 11A (High Energy)
Fuse
You must disassemble the GMM case to replace a
blown high energy (1000V, 11A) fuse (PN 943118).
The GMM’s safety design places this fuse inside the
case to protect the user and to encourage inspection
of the circuit board for damage if very high energies
have been encountered. Proceed as follows:
1. Set the GMM selector to OFF. Be sure to leave
the selector in this position throughout the
disassembly and reassembly procedure.
7-2
Page 92
User Maintenance
Reassembly
7
2. Remove the six screws attaching the case
halves. Detach the case halves.
3. Remove the two screws securing the shield as
shown in Figure 7-1. Then pry the shield loose
from two stanchions at the top of the circuit
board.
4. Gently pull up on the shield tab and rotate the
entire shield away from the circuit board.
5. Remove the three screws securing the circuit
board to the front half of the GMM.
6. Dislodge the circuit board by pushing in on the
Volts-Ohms input connector from the lower right
front of the GMM. (See Figure 7-1, Detail A).
7. Lift the entire circuit board up approximately one
inch, dislodging the slotted Battery Eliminator
connection.( See Figure 7-1, Detail B).
8. Rotate the circuit board 180 degrees.
9. Remove the fuse.
Note
For a more detailed drawing, see the
Service Manual (PN 689312).
Reassembly
Generally, reassembly reverses the disassembly
procedure. Specifically, do the following:
1. Verify that the selector switch still points to OFF.
2. Place the case top face down on a non-marring
surface.
Note
At all times, avoid stressing the ribbon cable.
3. Grasp the board at the top and bottom so that it
can be smoothly lowered into position in the
case top.
4. As you begin lowering the board into position,
make sure the battery eliminator connection
(Detail B in Figure 7-1) slips into its slot snug
against the top case wall.
5. Gently press the board into position at the top
and bottom
6. Replace the three screws securing the board.
7. Rotate the shield back into place on the circuit
board.
8. Replace the case bottom. Tighten the six screws
(middle screws first) securing the case bottom.
7-3
Page 93
867B/863
5
5
3
3
4
8
8
7
6
6
Users Manual
7-4
f7-1.eps
Figure 7-1. Replacing the 11A (High Energy) Fuse
Page 94
User Maintenance
Operational Test
7
Operational Test
Use the following procedure to test the basic
operability of your GMM test tool:
1. With no inputs connected, turn the GMM test tool
v
on by rotating the selector to VAC (
Meter display mode should appear, with a
random display in the mV range.
E
2. Rotate the selector to VDC (
primary display in the mV range.
3. Rotate the selector to Ohms (
display of “OL”.
4. Connect the test leads to
5. Short the test lead tips and observe a display of
approximately 0.250 ohms.
6. Press
For a complete performance test of GMM test tool
functions and specifications, refer to the 863/867B
Service Manual (PN 689312).
2 (/
listen for a beep.
), short the test leads again, and
) Observe a similar
R
V
). Combo or
). Observe a
C
and
.
Self Test
Begin an internal diagnostic routine by selecting SET
UP, wait two seconds, press
now presents the following information:
1. A test pattern.
2. An error message (if a problem is encountered
during the test). The GMM test tool Model and
serial numbers.
3. The calibration status.
Generally, you cannot use the GMM test tool keys
during Self Test. (You can press
diagnostic routine and return to the SET UP
Softkeys.)
d2
1
. The display
to exit the
7-5
Page 95
867B/863
Closed
AA
Batteries
Battery
Pack
Open
–
–
+
OR
+
Users Manual
7-6
Figure 7-2. Replacing the Batteries
rita0535.eps
Page 96
User Maintenance
Supplied
Spare Fuse
Fuse
Supplied
Spare Fuse
Self Test
7
rita0674.eps
Figure 7-3. Replacing the 440 mA Fuse
7-7
Page 97
867B/863
Users Manual
If Your GMM Does Not Work
NOTE
If the display remains blank after pressing
W
, contrast may be off the scale. Restore
contrast to mid-scale with the following
shortcut: set the selector to SET UP and
avoid touching any key until the display
reappears (about 15 seconds). Press
to store this setting, then turn the selector to
the desired function.
Examine the case for physical damage. If damage is
detected, contact a Fluke Service Center. Check and
replace (as needed) the battery, fuses, and test
leads. Review this manual to verify correct operation.
5
If the GMM still does not work, pack it securely and
forward it, postage paid, to the nearest service
center. Include a description of the problem. Fluke
assumes no responsibility for damage in transit.
A GMM under warranty will be repaired or replaced
(at Fluke’s option) and returned at no charge. See
the registration card for warranty terms. If the
warranty has lapsed, the GMM test tool will be
repaired and returned for a fixed fee. Contact the
nearest service center for information and prices.
If “Uncal” appears in the Upper Status Line, the
GMM is no longer calibrated; contact a Fluke Service
Center.
To locate an authorized service center, visit us on
the World Wide Web: www.fluke.com or call using
any of the phone numbers listed below:
USA and Canada:
1-800-44-FLUKE (1-800-443-5853)
Europe: +31 402-678-200
Japan: +81-3-3434-0181
Singapore: +65Anywhere in the world: +1-425-356-5500