Fluke 8508A User Manual

8508A Reference Multimeter
Extended Specifications
Interpreting and applying
the specifications
Introduction
The Fluke 8508A has been designed specifically for metrologists. Not only does it provide the performance metrologists need, but it is specified in a way to allow users to really understand the uncertainties of the measurements, and easily make allowance for those uncertainty contribu­tions when performing measurement uncertainty analyses and compiling uncertainty budgets. Contemporary metrology practices, including ISO 17025 based laboratory accreditation schemes, require uncertainty analysis to be performed in accordance with the statistically based techniques described in the ISO Guide to the Expression of Uncertainty in Measurement (often referred to as the 'GUM'). For convenience, the 8508A specifications are quoted at a coverage factor of k=2, equivalent to a confidence level of approximately 95 %, as required by these methods. Specifications are also provided at a confidence level of 99 %.
Performance specifications for the 8508A consist of two elements; the first is a contribution expressed as parts-per-million of the Reading, and the second contribution is expressed as parts-per-million of the Range. These must be evaluated and combined for the relevant reading and range values applicable to the measurement being made, ensuring that both elements are evaluated on the same basis, such as parts per million of the measured value or in absolute terms (volts, amps, ohms, etc). The two elements are combined by adding algebraically. For example, measuring 10 V on the 20 VDC range and applying the 365 day ±1 °C specifications:
First, expressing the contributions in terms of parts-per-million of the measured value.
Second, expressing the contributions in volts:
To realize the full potential of the 8508A per­formance, accepted metrology practices should be employed, such as performing a zeroing or null operation to remove any offsets present in the measurement setup when making DC meas­urements. The 8508A specifications assume that these methods are employed.
Absolute and Relative specifications
The Relative to Calibration Standards specifica­tions describe the performance of the 8508A itself for the time periods and temperature range listed, excluding the uncertainty of the standards used to perform calibration of the 8508A during manufacture. The Absolute specifications include the uncertainty of the standards used to perform calibration of the 8508A at manufacture and may be used to determine the uncertainty of measurements made with the 8508A for periods up to 1 year and over a temperature range of ±5 °C from calibration. If the user has their 8508A calibrated with different uncertainties, the Relative specifications can be combined with the uncertainties applicable to that calibration to determine the effective absolute uncertainty following that calibration.
Applying user’s calibration uncertainties
When the 8508A is calibrated by another laboratory, the uncertainties of the calibration standards used may be applied by combining those uncertainties with the 8508A's Relative to Standards specifications. The applicable calibration uncertainties and the 8508A relative specifications must both be expressed at the same confidence level, and be combined in an RSS (Root Sum Square) summation. Accepted metrology practice mandates that calibration uncertainties are stated at 95 %. Check the applicable calibration uncertainties are stated at 95 % and then combine them with the 8508A 95 % Relative specifications. For example, if the 8508A is calibrated at 10 VDC with an uncertainty of 1.5 ppm at 95 % the absolute uncertainty at 10 V for a period of 90 days and ±1 °C from calibration is:
The 8508A is designed to provide accuracy and stability without the need for internal auto or self calibration routines which may otherwise compromise the continuity and traceability of measurement performance history.
2 Fluke Corporation 8508A Extended Specifications
Applying the specifications
Operating and calibration temperature ranges
As a metrology tool, the 8508A will commonly be used in a calibration laboratory where the temperature would be controlled to ±1 °C, and the 8508A ±1 °C specifications are applicable to those situations. The majority of electrical calibration laboratories operate at a nominal temperature of 23 °C, the temperature at which the 8508A is calibrated by Fluke during manu­facture and service. The 8508A is also capable of being calibrated at any temperature between 20 °C and 25 °C and the ±1 °C specifications will apply to operation within ±1 °C of that calibra­tion temperature. In the 8508A specification tables, the temperature of calibration is referred to as TCal. Specifications for ±5 °C are provided for situations where the 8508A is operated in environments with wider temperature variations up to ±5 °C. For applications where the knowledge of the effect of temperature on 8508A perform­ance is important, temperature coefficients are listed in the 8508A specifications. If the operating temperature is within the range 15 °C to 30 °C, the 15 °C to 30 °C temperature coefficient specifications are applicable; otherwise use the 5 °C to 15 °C/30 °C to 40 °C figures, provided the temperature lies within that range. The 8508A may be operated at temperatures between 0 °C and 50 °C, but performance is not specified outside the range 5 °C to 40 °C.
Applying temperature coefficient specifications
The 8508A specification tables include information for the typical operating conditions of ±1 °C for calibration laboratories with tight temperature control, and ±5 °C for calibration laboratories with looser temperature control or uncontrolled environments within that tempera­ture range. For the majority of applications, choosing the Absolute specifications for the most appropriate operating temperature range will be adequate. However performance at other temperatures may be determined by including an allowance for temperature coefficient over the additional temperature range. Care should be taken when making this calculation, as an amount of temperature coefficient is already included in the 8508A specifications, and those specifications are themselves based on combining contributions using techniques similar to those employed in uncertainty analysis. For example, consider operating at 33 °C, 10 °C from the 23 °C calibration temperature. The ±5 °C specifications already include a contribution for 5 °C of tem­perature difference, so this amount of temperature effect must be removed before the effect of the 10 °C difference is added. Consider 10 V on the 20 VDC range: 365 day absolute specification (95 %) at 33 °C expressed in parts-per-million of 10 V is:
The Fluke 8508A and 8508A-SPRT Standard Platinum Resistance Thermometer
Fluke Corporation 8508A Extended Specifications 3
Applying the specifications
Ratio measurements
The 8508A Ratio mode will automatically take measurements of inputs applied to the front and rear terminals and display the result as a ratio in the voltage and resistance functions. The measurements can be made on the same range or different ranges. When making measurements on different ranges, the error in each measure­ment is evaluated by applying the relevant specification for each range and combining the two specifications in an RSS summation, expressing the contributions in parts-per-million of the measured values. For example, making measurements of the ratio of 100 mV on the 200 mVDC range and 100 V on the 200 VDC range, applying the 365 day ±1 °C Absolute specifications:
Making measurements on the same range will eliminate range-to-range errors, such as drift since the time of calibration, and improve the result. When making measurements on the same range, these errors will affect both meas­urements and effectively cancel, leaving short term noise and linearity as the dominant errors. The 20 minute Transfer Uncertainty specifica­tions are provided to describe the performance obtained when making ratio measurements on the same range. The error in each measurement is evaluated by applying the relevant 20 minute Transfer Uncertainty Specification for each value and combining the two specifications in an RSS summation, expressing the contributions in parts-per-million of the measured values. If the measurements are made within the same range, but independently (not using the ratio mode) with an elapsed time greater than 20 minutes but less than 24 hours between the measurements, then the 24 hour specifications should be applied instead.
For example, making measurements of the ratio of 5 V and 10 V on the 20 VDC range, applying the 20 minute Transfer Uncertainty Specifications:
Additional errors
The 8508A specifications are listed for the maximum resolution in each function, using the Normal reading mode. For measurements taken in other resolutions or the Fast read mode additional error contributions listed in the Read Rate and Additional Uncertainty table must be included. These additional contributions must be added algebraically to the relevant specifica­tions. For example, measuring 10 V on the 20 VDC range at 5 digit resolution in Fast mode and applying the 365 day ±1 °C Absolute specifications:
Other additional contributions apply in certain situations and are also to be added algebraically to the relevant specifications. These additional contributions include the DC Accuracy specifica­tion to be applied when making DC measure­ments on the AC Voltage function when DC coupled, and the High Voltage Adder when making measurements above 300 V on the AC function.
4 Fluke Corporation 8508A Extended Specifications
DC Voltage specifications
DC Voltage
Range
[1] [2] [3]
Full Scale Uncertainty Relative to Cal Stds
± (ppm Reading + ppm Range)
24 hour
TCal ±1 °C
90 day
TCal ±1 °C
365 day
TCal ±1 °C
Absolute Uncertainties
[4]
365 day
TCal ±1 °C
365 day
TCal ±5 °C
95 % Confidence Level
200 mV 199.999 999 0.7 + 0.5 1.4 + 0.5 2.7 + 0.5 4.5 + 0.5 5.0 + 0.5 2 V 1.999 999 99 0.5 + 0.2 1.4 + 0.2 2.7 + 0.2 3.0 + 0.2 3.5 + 0.2 20 V 19.999 999 9 0.5 + 0.2 1.4 + 0.2 2.7 + 0.2 3.0 + 0.2 3.5 + 0.2 200 V 199.999 999 1.0 + 0.2 2.6 + 0.2 4.0 + 0.2 4.5 + 0.2 5.5 + 0.2 1000 V 1050.000 00 1.0 + 0.5 2.6 + 0.5 4.0 + 0.5 4.5 + 0.5 5.5 + 0.5
99 % Confidence Level
200 mV 199.999 999 0.8 + 0.6 2.0 + 0.6 3.5 + 0.6 6.0 + 0.6 6.5 + 0.6 2 V 1.999 999 99 0.6+ 0.25 1.8 + 0.25 3.5 + 0.25 4.0 + 0.25 4.5 + 0.25 20 V 19.999 999 9 0.6 + 0.25 1.8 + 0.25 3.5 + 0.25 4.0 + 0.25 4.5 + 0.25 200 V 199.999 999 1.2 + 0.25 3.5 + 0.25 5.2 + 0.25 6.0 + 0.25 7.0 + 0.25 1000 V 1050.000 00 1.2 + 0.6 3.5 + 0.6 5.2 + 0.6 6.0 + 0.6 7.0 + 0.6
DC Voltage
Range
Transfer
Uncertainty
20 mins ±1 °C
± (ppm Reading
+ ppm Range)
(Secondary Specifications)
Temperature Coefficient
15 °C - 30 °C
± ppm Reading/°C
[1] [2] [3]
5 °C - 15 °C
30 °C - 40 °C
200 mV 0.4 + 0.3 0.4 0.6 2 V 0.12 + 0.1 0.3 0.5 20 V 0.12 + 0.1 0.3 0.5 200 V 0.4 + 0.1 0.7 1.0 1000 V 0.4 + 0.3 0.7 1.0
Type Multi-slope, multi-cycle A-D Converter CMRR (1 kunbalance)
[5]
NMRR
[5]
140 dB at DC and 1 - 60 Hz
Filter Out 60 dB at 50/60 Hz ±0.09 % Filter In 110 dB at 50/60 Hz ±0.09 %
Protection (All ranges) 1 kV rms Input Impedance
200 mV to 20 V Ranges >100 G 200 V & 1000 V Ranges 10.1 M± 1 %
Max Input Current 50 pA Ratio Accuracy
Range to Range ±(Net Front Input Accuracy + Net Rear Input Accuracy) Within Range Apply 24 hour or 20 minute Transfer
Uncertainty specifications
Settling Time (to 10 ppm step size) Filter Out <50 ms Filter In <1 s
Fluke Corporation 8508A Extended Specifications 5
Loading...
+ 11 hidden pages