Fluke 80T-E Service Manual

80T-IR/E
Extended Range Infrared
Temperature Probe
Instruction Sheet
Introduction
The Fluke 80T-IR/E Extended Range Infrared Temperature Probe (the probe) is a noncontact temperature measurement acc es sory for use with a test instrument capable of measuring DC volts in the millivolt range such as a digital multimeter (DMM). The probe has a temperature range of 32°F to 1000°F, with a basic accuracy of 3% of reading, and an output of 1 mV dc per °F.
Temperature is measured by pointi ng the probe at the surface to be measured, and reading the tem perature on the test instrument di s pl ay.
®
Box Contents
Temperature Probe, Battery (installed), Instruction Sheet, and Warranty Card.
Safety Information
The 80T-IR/E complies with IEC Publication 1010-1­1990 including Amendment 1, CSA C22.2 No. 231, ANSI/ISA-S82.01 and .03 Safety Standards.
W WARNING
IF TARGET EMMISSIVITY IS LESS THAN 0.95, THE PROBE CAN INDICATE A TEMPERATURE LOWER THAN THE ACTUAL TARGET TEMPERATURE. AVOID TOUCHING THE TARGET; THERMAL BURNS COULD RESULT.
· Do not place the probe on or around hot objects (70° C
/ 158°F). It will damage the probe case.
· If the probe is exposed to significant changes in
ambient temperature (hot to cold or cold to hot), allow 20 minutes for temperature stabilization, before taking measurements.
· Do not operate the probe near large electrical or
magnetic fields such as arc welders and induction heaters. These fields can cause measurement errors.
· Condensation may form on the lens when going
from a cold to hot environment - wait 10 minutes for condensation to dissipate before taking measurements.
· Connectors must only be plugged into voltage
measurement input jacks of the test instruments.
· Do not touch or hold by the front cone.
Temperature readings can be affected by heat from the hand.
· Equipment use not specified by manufacturer
may impair safety.
CAUTION
Compatibilit y
The probe is compatible with all DC milliv olt measuring instruments that have a minimum of 1 MW input impedance and accept safety shrouded, standard diameter 0.16 in. (4 mm) banana plugs.
Operati on
To take a measurement, perform t he f ol l owi ng s teps:
1. Plug the red connector into the VW dc input jack and the black connector into the common or ground input jack on the test i ns trument.
2. Select mV dc on the t es t instrument.
3. Slide the probe switch forward to the "ON" position.
4. Point the tip of the probe as close as pos sible to the object being measured without touching the object.
5. Read the test instrument di s pl ay.
Additional considerations are:
· After 10 minutes of use the probe will automatically shi ft to Sleep mode (the display will show 0°F). It can be restarted by sliding the switch to "OFF" and then to " O N" (s ee Tabl e 1).
· Sleep mode extends battery life. However, for maximum battery life, switch the probe to the "OFF" position.
· If the test instrument displays an overload condition, switch t he DMM range from mV dc t o V dc. Increasing the range to V dc moves the decimal position three places to the left (1000°F displays as 1.000V dc ).